Abstract
Well established long ago for bacterial and fungal detection, Matrix-Assisted Laser Desorption/Ionization–Time of Flight (MALDI-TOF) technique is not so well established in the virology field, and taking care of its advantages (speed, precision and low cost), this can be a powerful method for viral identification. To explore the feasibility and potential of MALDI-TOF for viral detection, this study shows the development of an in-house spectral library including several uninfected cell cultures and cultures infected with different clinically relevant viruses, such as SARS-CoV-2. This library was applied to the identification of viral infections directly on cell cultures, assessing the ability of the technique to discriminate between infected and non-infected profiles. Additionally, bioinformatic analyses were conducted to evaluate the structure, specificity, and reproducibility of the in-house library, and to understand its strengths and limitations. Sensitivity and specificity of the method were estimated by testing multiple culture batches from selected viruses included in the library. Together, these results provide a deeper understanding of the performance and applicability of MALDI-TOF in the virological context, highlighting its potential as a valuable research platform and a prospective tool for clinical viral detection.
1. Introduction
Since its development for bacterial identification, Matrix-Assisted Laser Desorption/Ionization–Time of Flight Mass Spectrometry (MALDI-TOF/MS) has been widely used in clinical laboratories, especially after the commercialization of standardized databases [1,2].
Based on the comparison of protein fingerprints obtained from microbial cells with reference spectra stored in databases, using algorithms integrated into the system, MALDI-TOF/MS is considered a “soft” ionization technique. It produces little or no fragmentation of the sample and allows accurate identification of analyte ions even in complex mixtures of biopolymers [3,4].
Originally designed for bacterial identification, MALDI-TOF has since been extended to other microorganisms such as mycobacteria and filamentous fungi [5]. However, the development of a viral database remains an unresolved challenge, despite the ongoing efforts of several research groups worldwide [6,7,8,9] and the new advances in the field [10,11,12,13,14].
On the one hand, regarding viruses, their culture in cell lines has traditionally been the gold standard for their purification and detection. Nevertheless, the presence of cellular debris in these cultures may hinder the direct identification of viral proteins [15], which could explain the limited progress toward comprehensive viral spectral databases.
And on the other hand, in the broader field of biology, as in other scientific disciplines, bioinformatic analysis of large datasets has become an essential tool for improving experimental understanding, including in mass spectrometry-based studies.
In this context, this study combines both the practical analysis of the ability of the Bruker Daltonics MALDI-TOF system to correctly discriminate between non-infected and virus-infected cell cultures based on an in-house library, and the bioinformatic analysis of the spectra that compose this library. The latter includes, by one side, evaluating how Bruker’s software itself groups the spectra into clusters using its default method, and on the other, employing the MassUp analysis tool developed by the Next Generation Computer Systems Group (SING) and the Bioscope Group at the University of Vigo [16] that enables the search for specific biomarkers within the spectra.
2. Materials and Methods
2.1. In-House Library Elaboration
Samples: The Cultures Used for the Composition of This Library Were (Table 1)
Table 1.
Distribution of cultures used for in-house library creation.
Non-infected cell cultures: Vero E6, MRC-5, LLC-MK2, HEP-2, A549 and MDCK (3 independent cultures of each one).
Infected Vero E6 cell cultures: 3 cultures with Herpes Simplex Virus 1 or HSV-1, 2 cultures with Acyclovir-resistant HSV-1, 3 cultures with Herpes Simplex Virus 2 orHSV-2, 1 culture with Varicela-Zoster Virus or VZV, 3 cultures with SARS-CoV-2 virus and 2 cultures with Mumps Virus or MuV.
Infected MRC-5 cell cultures: 3 cultures with HSV1, 3 cultures with HSV2, 1 culture with VZV, 3 cultures with Cytomegalovirus or CMV, 4 cultures with Adenovirus or ADV (one for each of the following genotypes: 2, 3, 7 and 8) and 3 cultures with Respiratory Syncytial Virus or RSV.
2.2. Culture Obtention
Non-infected cell cultures: all cell lines were cultivated on growth medium, based on: MEM medium (Dominique Dutscher; Brumath, France) supplemented with Bovine Fetal Serum (Dominique Dutscher; Brumath, France) at 10%, L-glutamine (Lonza; Verviers, Belgium) at 0.1% and 250 µL of gentamicin 40 mg/mL (Normon; Madrid, Spain). These cell lines were cultivated on flask tubes and aliquoted on sterile tubes for later work. All cell lines were incubated at 37 °C with CO2 reduction of 5%.
For the MALDI-TOF analysis for library acquisition all cell lines were collected at 100% of cell confluence.
Infected cell cultures: each selected virus was inoculated in a different cell culture on a sterile tube, according as it was explained previously.
For inoculation, the growth medium was discarded and replaced with maintenance medium (MEM medium supplemented with L-glutamine at 0.1% and 1250 µL of gentamicin 40 mg/mL), after that, 150 µL of the positive sample was added. These tubes were incubated at 37 °C until cytopathic effects (CPE) appear, and then 150 µL of the tube was passed to a new tube for pure viral culture isolation.
After this step, all viral cultures intended for MALDI-TOF library acquisition were harvested upon observation of cytopathic effect (CPE). Viral presence was confirmed by both RT-qPCR and immunofluorescence assay prior to spectral acquisition.
2.3. Viral Presence Confirmation on Cultures
- By RT-qPCR: a volume of 200 μL of each culture was extracted and purified using MagNa Pure 96 system (Roche; Geneva, Switzerland), following the manufacturer instructions.
An in-house designed RT-qPCR test was performed after that.
Briefly, 5 µL of each sample was added to 6 µL of TaqMan Fast 1-Step Master Mix (Life Technologies, Carlsbad, CA, USA) and supplemented with 4 µL of the suitable mixture of primers (Thermo Fisher Scientific; Waltham, MA, USA) and taqman MGB probes (Applied Biosystems; Foster City, CA, USA) (Table 2).
Table 2.
Primers and probes used for RT-qPCR assays.
The cycling protocol was as follows: (50 °C, 20 min; 95 °C, 5 min; 45 cycles of 95 °C, 10 s; 55 °C, 15 s and 60 °C, 30 s) [17].
- Immunofluorescence assay: the monolayer of each culture was peeled off the tube and resuspended on the maintenance medium. In total, 500 µL of this resuspension was washed adding phosphate buffer Difco FA Buffer (Becton Dickinson; Sparks, MD, USA) and centrifugated at 2500 rpm for 10 min to collect the pellet. The supernatant was discarded and, with 100 µL of phosphate buffer, the pellet was resuspended and passed to a slide, where it was dried and fixed by immersion on cold acetone for 10 min.
For SARS-CoV2 cultures, this slide was incubated for 30 min at 37 °C with 5 µL of a 1/128 dilution of IgG positive serum from a COVID-19 patient. This serum was measured on 232 AU/mL by the Liaison XL System (DiaSorin Molecular LLC; Cypress, CA, USA) and titrated on dilution 1/512. After this, the slide was washed on phosphate buffer and incubated again (in the same conditions) with an anti-human IgG conjugate, washed again, and prepared for his visualization on a fluorescence microscope.
For the rest of the viral cultures, the slide was incubated for 30 min at 37 °C with 5 µL of an anti-Ag conjugate specific for each virus (Table 3). After washing, the slide was also prepared for its visualization on a fluorescence microscope [18].
Table 3.
Anti-Ag used for immunofluorescence assays.
2.4. Sample Processing for MALDI-TOF Analysis
The monolayer of each culture was peeled off the tube and resuspended in the maintenance medium. Approximately 750 µL of this resuspension was transferred to an Eppendorf tube and centrifugated at 6600 rpm for 10 min for the pellet obtention. The supernatant was discarded.
To the pellet, 300 µL of HPLC water was added, vortexing for a few seconds, and 900 µL of pure ethanol was added. Then, the tube was centrifugated at 13,000 rpm for 2 min and the supernatant was discarded. This step was repeated one more time and the excess of ethanol was dried in the air.
Once the pellet was dry, acetonitrile and formic acid 70% were added on equal volumes (15 µL) for protein extraction; of this volume 1 μL was taken and co-crystallized with the HCCA matrix (Bruker Daltonics; Billerica, MA, USA) on the work plate of the MALDI-TOF system for the analysis.
2.5. Sample Integration on In-House Library
According to Bruker Diagnostics protocol for elaboration and library amplification.
Nine reading positions were used for each sample, including one position with 1 μL of Bacterial Test Standard (BTS) (Bruker Daltonics; MA, USA), and one more with 1 μL of BTS together with 1 μL of the target sample. All these volumes were co-crystallized with 1 μL of the HCCA matrix at an approximate 1:1 ratio, and were used for peak discrimination by the MALDI-TOF softwares following the manufacturer protocol.
From each of these positions, three independent readings were acquired using a MALDI Biotyper MBT Smart (Bruker Daltonics; MA, USA) operated with the flexControl 3.4v software (method MBT_FC.par). The molecular spectra generated in the range between 2000 m/z and 20,000 m/z were reviewed with flexAnalysis 3.4v software. After revision and optimization, the final PMF (Peptide Mass Fingerprint) of each sample was obtained from a total of at least 21 readings and subsequently added to the library. This PMF was generated using the MALDI Biotyper Compass Explorer 4.1.100v software. All programs used were provided by Bruker Daltonics.
Spectra were acquired using a nitrogen laser (337 nm) MALDI-TOF mass spectrometer (Bruker Daltonics). Laser energy was automatically adjusted by the Bruker FlexControl software to optimize ion yield, as routinely applied in MALDI Biotyper workflows. Ion extraction delay was controlled automatically by the instrument according to the manufacturer’s default acquisition parameters.
2.6. Sensitivity and Specificity Assay
- Samples: for this assay 90 HSV cultures (47 HSV-1 and 43 HSV-2), 100 ADV, 50 SARS-CoV-2 cultures and 100 non-infected cell cultures (50 Vero E6 and 50 MRC-5) were used. Of the 47 HSV-1 cultures, 25 were performed on Vero E6 cells and 22 on MRC-5 cells. Likewise, of the 43 HSV-2 cultures, 20 were performed on Vero E6 cells and 23 on MRC-5 cells.
All cultures were carried out and processed for analysis following the previously described protocols.
Viral presence on all samples was confirmed by RT-qPCR and immunofluorescence assays (IF); about RT-qPCR results, for HSV the mean, expressed in Ct, was 22.46 ± 2.11, with a range between 18 and 27 cycles and a 95% CI between 22.03 and 22.89. For ADV, the mean was 20.52 ± 1.98, with a range between 17 and 26 cycles and a 95% CI between 20.14 and 20.90 cycles. For SARS-CoV-2, the mean was 22.89 ± 2.23, with a range between 20 and 28 cycles and a 95% CI between 22.28 and 23.50 cycles.
- 2.
- Sample analysis: each sample was read with MBT Compass 4.1.100 program and compared with the following libraries: BDAL (contains 8468 PMFs), IVD (contains 8326 PMFs), Mycobacteria library (contains 952 PMFs), Filamentous Fungi (contains 577 PMFs) and the in-house virus library, which contains the following spectrums:
- Non-infected cell cultures: Vero E6, MRC-5, LLC-MK2, HEP-2, A549 and MDCK (3 independent PMF of each one).
- Infected Vero E6 cell cultures: 3 PMF of HSV1, 2 PMF of Acyclovir-resistant HSV1, 3 PMF of HSV2, 1 PMF of VZV, 3 PMF of SARS-CoV-2 and 2 PMF of MuV.
- Infected MRC-5 cell cultures: 3 PMF of HSV1, 3 PMF of HSV2, 3 PMF of CMV, 4 PMF of ADV (one for each of the following genotypes: 2, 3, 7 and 8) and 3 PMF of RSV.
To consider a sample as positive, the following criteria were followed:
- ▪
- Evaluate only the first result offered by the program.
- ▪
- If the first result obtained corresponds to the inoculated virus under analysis, but is different in another cell line from that of the sample analyzed, it will also be considered as a valid result.
- ▪
- All the results that agree with the analyzed sample will be valued as positive, regardless of their score.
- ▪
- For infected cultures, results identified correctly as non-infected or as infected with a different virus were considered negative. For non-infected cultures, the same criterion was applied.
- ▪
- No peak detection (No Peaks Found or NPF) was considered as a false negative.
2.7. Bioinformatic Assays
Samples: for the different assays performed, the mass spectrums contained in the in-house library were analyzed, being 21 for each one of the PMF of the cultures stored on it (all of them were obtained in a MALDI Biotyper (Bruker Daltonics; Billerica, MA, USA).
Due to the limitations of the computer capacity of our laboratory, three different analyses were made, one about clustering analysis and two about markers discovery.
- ○
- Clustering analysis: the total of 1008 mass spectrums that make up the library were analyzed.
- ○
- Markers discovery on HSV: a group of 504 mass spectrums was analyzed from uninfected Vero E6 and MRC-5 cells, HSV-1 and HSV-2 grown on Vero E6 and MRC-5, CMV and SARS-CoV-2. From each culture, 3 samples were used, from each of which 21 independent spectra were obtained (n = 63 for each culture).
- ○
- Markers discovery on ADV: a group of 210 mass spectrums from the following cultures in MRC-5 cells were analyzed: uninfected cells (n = 63), ADV of genotypes 2 (n = 21), 3 (related to respiratory symptoms) (n = 21), 7 (pneumonia) (n = 21) and 8 (keratoconjunctivitis) (n = 21) and RSV (n = 63).
2.8. Bioinformatic Analysis
- Clustering analysis: all PMF included in the library were analyzed using the default clustering (correlation-based) method implemented in the MALDI Biotyper Compass 4.1.100v software (Bruker Daltonics, USA). This approach was applied to explore how the MALDI-TOF system processes and groups spectral data, providing insight into its internal classification criteria during sample identification.
- Markers discovery: It was carried out using the MassUp program. For this, the raw data of the PMF obtained were used. In the pre-processing of these, the Savitzky–Golay smoothing method was used and the base correction using the TopHat method. For the pairing of the different readings of a mass spectrum, tolerances of 300 ppm were considered, based on the mean, both intra-spectrum (between the different readings) and inter-spectrum. The percentage of presence established to consider a peak as representative was 95%.
For each viral culture, a consensus PMF was constructed by combining all mass spectra obtained for the same virus in the same cell culture, generating an averaged representative spectrum. Cell line differentiation was maintained. These consensus PMFs were subsequently analyzed using the Biomarker Discovery function to identify significant peaks and potential discriminant markers between viral cultures. Raw data used for this analysis has been included as Supplementary Material.
3. Results
First, about sensitivity and specificity assays, considering the analyzed cultures, the MALDI-TOF system demonstrated a sensitivity for the detection of HSV infected cells of 94.32% (95.56% for HSV-1 and 93.02% for HSV-2), of the 81.05% for ADV and 82.93% for SARS-CoV-2. Likewise, specificity was higher than 90% in all cases, reaching 100% for ADV and SARS-CoV-2.
In non-infected cultures, it was able to detect all the MRC-5 cell samples and 49 of the Vero E6 samples.
For the 16 samples considered as false positives, the detection performed by MALDI-TOF system for them was as follows:
- HSV-1 infected Vero E6 cells: 2 cultures identified as non-infected Vero E6 cells
- ADV infected MRC-5 cells: 3 cultures identified as CMV infected MRC-5 cells (not considered for calculus), 1 culture identified as HSV-1 infected MRC-5 cells and 1 culture identified as non-infected MRC-5 cells.
- SARS-CoV-2 infected Vero E6 cells: 8 cultures identified as HSV-1 infected Vero E6 cells and 1 culture identified as HSV-2 infected Vero E6 cells.
All these false positives were reviewed again by RT-qPCR finding they were only positive for the virus inoculated at the beginning. Also, there were 30 samples, considered false negatives, for which the MALDI-TOF system was unable to generate a clear reading (NPF).
Table 4 shows the diagnostic performance of the MALDI-TOF system for the different samples tested. Figure 1 and Figure 2 show different ECP and immunofluorescence assay results.
Table 4.
Detection performance of the MALDI-TOF system with the in-house library of cultures.
Figure 1.
(1) Non-infected Vero E6 cells; (2) Immunofluorescence of non-infected Vero E6 cells; (3) SARS-CoV-2 cytopathic effect; (4) Immunofluorescence of SARS-CoV-2; (5) HSV1 cytopathic effect; (6) Immunofluorescence of HSV1; (7) HSV2 cytopathic effect; (8) Immunofluorescence of HSV2.
Figure 2.
(1) Non-infected MRC5 cells; (2) Immunofluorescence of non-infected MRC5 cells; (3) ADV cytopathic effect; (4) Immunofluorescence of ADV.
The results of the clustering analysis are shown in Figure 3. In this clustering, a total of five clades are distinguished. In the first, colored in dark blue, are included virus cultured in MRC-5 cells (HSV2, VZV and ADV genotype 7); in the second, in green, next to non-infected MRC-5 cells, RSV, CMV and the rest of ADV genotypes; the third, in red, includes VZV grown on Vero E6, MuV, and Acyclovir-resistant herpes; the fourth, in purple, groups only MDCK cells, and the fifth includes the rest of the viruses grown in Vero E6, together with HSV1 in MRC-5 and the rest of non-infected cells.
Figure 3.
Clustering analysis of the total spectrum of the in-house library developed in the laboratory.
On biomarker analysis, the markers detected by comparing the spectra of the HSV-1 and HSV-2 viruses, grown on both MRC-5 and Vero E6, when comparing with each other and with other cultures, are shown, expressed as a function of their mass/charge (m/z), in Table 5.
Table 5.
Detected markers for HSV-1 and HSV-2 viruses, ordered by their m/z value.
After the analysis, a total of 44 peaks (out of 1535 detected) were found with a p-value less than 0.05, present only in the Herpes Simplex Virus cultures. Of these 44 peaks, only one, at position 6859.5464 m/z, is present in all HSV cultures, regardless of the type or cell line in which it was grown (Table 5).
On the other hand, 6 typical HSV-1 peaks have been detected, present both in its culture in MRC-5 cells and in Vero E6 cells. Similarly, HSV-2 also has another 6 identical self-peaks between cultures in both cell lines. If discriminated by culture cell line, in MRC-5 cells, HSV-1 and HSV-2 share 4 peaks that are not found in the same viruses grown on Vero E6. In the opposite case, HSV-1 and HSV-2 show 2 identical peaks when grown in Vero E6 cells that are not found when grown in MRC-5 cells (Table 5).
Finally, the markers detected when comparing the ADV spectra with each other and with other cultures are shown, expressed as a function of their mass/charge (m/z), in Table 6.
Table 6.
Detected markers for ADV genotypes, ordered by their m/z value.
After the analysis of the spectra of the 4 Adenovirus genotypes, 39 different peaks were found (out of 1515 detected) with a p-value less than 0.05. Of these peaks, 14 are shared by the four tested genotypes. Regarding genotype 7, 4 peaks present exclusively in this genotype were detected, as well as a region marked by the absence of peaks between 7600 and 9500 m/z.
Although m/z values are reported with four decimal places as provided by the acquisition and analysis software, this numerical precision does not reflect absolute mass accuracy, which in MALDI-TOF MS is limited to several Daltons. These values are used solely for peak comparison and reproducible pattern identification.
4. Discussion
As MALDI-TOF is one of the most widely used techniques for microorganism identification and is known for its versatility and capacity for optimization [6,7,8,9], its adaptation to the virology field should be considered a promising future goal.
To establish the basis for its potential use in this area, in the first part, this study demonstrates the identification performance of an in-house library generated from different infected and non-infected reference cell lines. The system achieved acceptable sensitivity values for several clinically relevant viruses, including SARS-CoV-2, showing that MALDI-TOF can be adapted beyond bacterial identification to detect viral infection in cell cultures.
Regarding system performance, the MALDI-TOF platform correctly identified 94.32% of HSV-infected samples (95.56% for HSV-1 and 93.02% for HSV-2), 81.05% for Adenovirus, and 82.93% for SARS-CoV-2. These results are particularly noteworthy considering that the analyzed material was “dirty”, containing a large amount of host cell proteins dragged from the cultures.
This background could explain not only some of the false negatives (NPF) due to potential signal saturation that prevented accurate readings, but also part of the false positives detected, where misclassification may result from overlapping host cell spectral features.
This phenomenon becomes more evident when comparing false positive data with clustering results.
The clustering analysis revealed that the primary grouping factor was the host cell line rather than the specific virus. This suggests that the cellular proteomic background exerts a strong influence on the resulting spectra, potentially masking virus-specific signals. Although the analytical protocol was designed to minimize these effects—standardizing growth factors, incubation temperature, and culture time—the host contribution remains significant.
Also, from a clinical perspective, it is worth noting that several studies have evaluated the performance of MALDI-TOF directly on patient samples (nasopharyngeal swabs, saliva or cerebrospinal fluid) for viral identification [19,20,21], particularly for SARS-CoV-2. These works, with a similar approach to that presented in this study, have reported good sensitivity levels (between 62.50% and 96.67%), especially in the case of Chivte et al. [20] reinforcing the potential of MALDI-TOF system for rapid pathogen detection. However, it should be noted that these studies only analyzed results for a single virus, without comparison to others, something that could lead to an overestimation of the sensitivity values obtained.
It must also be remembered that, although the use of direct samples would be ideal for rapid pathogen identification and for making this method viable for clinical diagnosis, specimens such as nasopharyngeal swabs or saliva remain complex and heterogeneous samples, possibly even more difficult to standardize than cell cultures, as they depend on additional factors such as interindividual differences in microbial flora [22,23].
In addition to the above, one of the main limitations of the MALDI-TOF system cannot be ignored: it can only identify microorganisms that are present in the reference library. If a microorganism is not included in the library, the algorithm will nevertheless assign it to an existing class [1,2]. This inevitably leads to false positive identifications.
See the following example: if the library is constructed to recognize only microorganism (A) and microorganism (B), and a different microorganism (Δ) is analyzed, the system will still assign it to either A or B based on spectral similarity. In practice, Δ may appear more similar to A than to B, and the system will therefore report it as A, albeit with a lower score than that obtained when analyzing a true A isolate.
Of course, faced with this problem, if the intention is to use the MALDI-TOF system for viral identification in patient samples, there are two options: one is to create a library extensive enough to contain each virus in circulation and, at least, its most notable variants, which could never stop being updated. The second is to introduce a confidence or similarity threshold below which the system reports that no class can be assigned, similar to the one used for fungi and bacteria applied automatically in every Bruker Daltonics MALDI-TOF.
Both would be feasible options to achieve with work, resources and effort, although it is worth asking whether they are necessary, given existing rapid and efficient techniques such as PCR.
In the second part, this study explores the contribution of computational analysis to a deeper understanding of the obtained data. Bioinformatics has had a transformative impact on data processing and interpretation in recent years, and its integration with established analytical techniques such as MALDI-TOF provides new interpretative potential, enabling the identification of subtle proteomic patterns that would otherwise remain unnoticed [24,25,26].
As previously mentioned, the clustering analysis revealed several notable patterns. Specifically, two main groups were observed: one (including the first two clades) that contained almost all viral cultures grown on MRC-5 cells—except for HSV-1—and another comprising those grown on Vero E6 cells. Within each group, cultures of the same virus are clustered together in distinct subclades, clearly separated from others by significant spectral distances.
These distributions help explain the occurrence of false positives. Except for the case of an ADV culture in MRC-5 cells misidentified as HSV-1 in the same cell type, all false positives appeared within their corresponding cell line cluster. Even in that exception, the misclassification can likely be attributed to the influence of the cellular background.
Regarding the distribution of non-infected cells, all of them formed unified subclades within each major cluster. The MRC-5 cells (human fetal lung fibroblasts) grouped together with most of the infected cultures derived from the same lineage. Separately, an isolated cluster was formed by MDCK cells (canine kidney epithelial line).
The last main cluster included, at one end, Vero E6 cells (immortalized renal epithelium from Cercopithecus aethiops), and at the other, three subclusters corresponding to HEP-2 (human laryngeal carcinoma), A549 (human lung adenocarcinoma), and LLC-MK2 (immortalized renal epithelium from Macaca rhesus) cells.
These clustering patterns reinforce the idea that cellular processes and the proteomic expression profile of the host cells are key determinants in MALDI-TOF analysis. Cell lines of different origins but with similar immortalization processes tend to group together, and the partial clustering of specific viral groups—such as HSV-2, VZV, and ADV genotype 7 within MRC-5-derived clades—suggests that viral replication induces reproducible and detectable proteomic alterations in host cells.
Even so, certain inconsistencies remain, such as the separate grouping of VZV cultures or the disaggregation of acyclovir-resistant HSV-1 strains from the other herpes cultures. These findings suggest that some stages of the process, including culture quality control and sample processing, still require optimization.
It is also important to emphasize that the clustering observed here does not represent a phylogenetic relationship, as MALDI-TOF detects global proteomic profiles rather than genetic sequences. Therefore, the clustering reflects phenotypic similarities and can be influenced by the physiological state of the cells supporting viral replication, even when the protocol was specifically designed to minimize this.
For all of these, additional bioinformatic tools beyond those provided by the manufacturer are valuable. One such tool is the MassUp application [11], previously used in several microbiological studies [21,24,25,26], such as that by Torres et al., where it was applied to analyze spectra from cerebrospinal fluid samples of healthy and infected patients to identify enteroviral biomarkers [21].
This software allowed a deeper exploration of the obtained spectra, supporting the sensitivity findings of the MALDI-TOF system and reinforcing the idea that the technique can differentiate not only between infected and non-infected cultures but also among infections caused by different viruses.
The viruses chosen for biomarker analysis were selected trying to conjugate both their characteristics as clinical relevance, structural diversity (enveloped vs. non-enveloped), differences in replication strategies, etc., and the data processing capacity of the computer system available.
In the Herpes Simplex virus analysis, several characteristic peaks were detected. A unique peak at 6859.5464 m/z was present in all HSV cultures, regardless of the type or host cell line, suggesting a conserved viral marker. Additionally, several specific peaks were identified for HSV-1 and HSV-2 individually, consistent across both MRC-5 and Vero E6 cell cultures.
Similarly, for Adenovirus, all four genotypes shared a common set of m/z peaks, except genotype 7 (known for its association with severe pneumonia [27]) which was distinguished by four unique peaks and a characteristic absence of peaks in the 7600–9500 m/z region.
The presence of unique markers for each virus supports the hypothesis that MALDI-TOF can detect virus-derived proteins even in the presence of abundant cellular background. Moreover, the MassUp-based analysis opens new possibilities for investigating proteomic differences between viral genotypes and for identifying potential markers linked to infection prognosis (as in the case of ADV genotype 7) although further validation is needed.
However, one inherent limitation of this approach is the inability to determine the exact molecular identity of each detected peak, as they may correspond to either viral peptides or cellular proteins expressed in response to infection, a problem that others like Chivte et al. [20] have expressed before and that resides in the very way the system measures it.
Although MALDI-TOF provides a characteristic proteomic fingerprint of each sample, it does not directly measure the absolute molecular mass of the detected compounds. The system determines the mass-to-charge ratio (m/z) from the time of flight of the ions, which depends on their charge state (z). Since this charge is not always known with certainty and may vary due to the formation of adducts or ion clusters, the m/z value cannot be unambiguously converted into an exact molecular mass in Daltons (Da) [28,29,30].
Additionally, instrumental limitations such as mass resolution, calibration accuracy, and the influence of the matrix or ionization conditions introduce further variability. As a result, each detected peak represents a reproducible but relative signal rather than an absolute molecular identity. Therefore, while MALDI-TOF is a powerful tool for pattern recognition and comparative proteomic analysis, it cannot on its own determine or physically isolate the specific peptide or protein responsible for a given peak with complete certainty [28,29,30].
Studies such as that by Lohnes et al. [31] have attempted to address this issue using complementary techniques like gel electrophoresis followed by MALDI-MS imaging. Nevertheless, these approaches require specialized equipment and resources not available in many clinical laboratories as this case.
In conclusion, this work represents one of the first attempts [32,33,34] to construct a specific and standardized viral reference database for MALDI-TOF, using infected cell cultures as a foundation. While, in addition to the inherent limitations of the technique, this culture-based approach has its own constraints (dependence on the base cell culture, culture time, high resource costs, etc.) that limit its direct applicability to clinical diagnosis, it provides a robust and reproducible framework for the identification and characterization of viral biomarkers under controlled conditions. These findings lay the groundwork for refining a method for the reliable detection and analysis of viruses in complex samples, ultimately paving the way for their future implementation in clinical research.
Supplementary Materials
The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/applmicrobiol6030042/s1, (spectra raw data used for bioinformatic analysis).
Funding
This research received no external funding.
Institutional Review Board Statement
This study was approved by Comité de Ética de la Investigación del Principado de Asturias with code CEImPA 2021.188.
Informed Consent Statement
Not applicable.
Data Availability Statement
All research data are provided in the article.
Acknowledgments
There was an initial contribution from the team at the Virology Section of the Central University Hospital of Asturias (HUCA) during an early phase of the project, but as the work evolved and opinions on the conclusions diverge, they chose not to remain involved. The development, analysis, and final writing of the article have been carried out exclusively by the author.
Conflicts of Interest
The author declares no potential conflicts of interest or competing interests.
References
- Welker, M.; Moore, E.R. Applications of whole-cell matrix-assisted laser-desorption/ionization time-of-flight mass spectrometry in systematic microbiology. Syst. Appl. Microbiol. 2011, 34, 2–11. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lavigne, J.P.; Espinal, P.; Dunyach-Remy, C.; Messad, N.; Pantel, A.; Sotto, A. Mass spectrometry: A revolution in clinical microbiology? Clin. Chem. Lab. Med. 2013, 51, 257–270. [Google Scholar] [CrossRef] [Scilit]
- Wieser, A.; Schneider, L.; Jung, J.; Schubert, S. MALDI-TOF MS in microbiological diagnostics-identification of microorganisms and beyond (mini review). Appl. Microbiol. Biotechnol. 2012, 93, 965–974. [Google Scholar] [CrossRef] [Scilit]
- Bizzini, A.; Greub, G. Matrix-assisted laser desorption ionization time-of-flight mass spectrometry, a revolution in clinical microbial identification. Clin. Microbiol. Infect. 2010, 16, 1614–1619. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Dingle, T.C.; Butler-Wu, S.M. Maldi-tof mass spectrometry for microorganism identification. Clin. Lab. Med. 2013, 33, 589–609. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Cobo, F. Application of maldi-tof mass spectrometry in clinical virology: A review. Open Virol. J. 2013, 27, 84–90. [Google Scholar] [CrossRef] [Scilit]
- Calderaro, A.; Arcangeletti, M.C.; Rodighiero, I.; Buttrini, M.; Gorrini, C.; Motta, F.; Germini, D.; Medici, M.-C.; Chezzi, C.; De Conto, F. Matrix-assisted laser desorption/ionization time-of-flight (MALDI-TOF) mass spectrometry applied to virus identification. Sci. Rep. 2014, 30, 6803. [Google Scholar] [CrossRef] [Scilit]
- Erukhimovitch, V.; Karpasasa, M.; Huleihel, M. Spectroscopic detection and identification of infected cells with herpes viruses. Biopolymers 2009, 91, 61–67. [Google Scholar] [CrossRef] [Scilit]
- Schwahn, A.B.; Wong, J.W.H.; Downard, K.M. Subtyping of the Influenza Virus by High Resolution Mass Spectrometry. Anal. Chem. 2009, 81, 3500–3506. [Google Scholar] [CrossRef] [Scilit]
- Lei, R.; Zhao, Z.; Dong, Z.; Tian, Q.; Wang, S.; Sun, Y.; Wang, X.; Li, H.; Zhang, Y. Simultaneous surveillance of 30 phytopathogens in potato using a high-throughput diagnostic system combining multiplex PCR with MALDI-TOF MS. Pest Manag. Sci. 2025, 81, 7090–7101. [Google Scholar] [CrossRef] [Scilit]
- Dederko, P.; Monedeiro, F.; Walczak-Skierska, J.; Ludwiczak, A.; Zienkiewicz, A.; Zienkiewicz, K.; van Eldik, R.; Treder, K.; Pomastowski, P.P. MALDI-TOF MS technique as a new approach for simultaneous detection and differentiation of potato virus Y strains. Sci. Rep. 2025, 15, 29928. [Google Scholar] [CrossRef] [Scilit]
- Song, R.; Li, D.; Hao, X.; Lyu, Q.; Ma, Q.; Chen, X.; Qiao, L. MALDI-TOF MS analysis of nasal swabs for the characterization of patients infected with SARS-CoV-2 Omicron. View 2024, 5, 20240015. [Google Scholar] [CrossRef] [Scilit]
- Deulofeu, M.; García-Cuesta, E.; Peña-Méndez, E.M.; Conde-González, J.E.; Jiménez-Romero, O.; Verdú, E.; Serrando, M.T.; Salvadó, V.; Boadas-Vaello, P. MALDI-TOF MS and machine learning explanations for the detection of SARS-CoV-2 infection in human plasma: Fingerprints as a strategy for risk assessment. ACS Omega 2025, 10, 41413–41424. [Google Scholar] [CrossRef] [Scilit]
- Xu, Z.; Xiong, B.; Cheng, W.; Hu, B.; Tang, Y. Rapid Analysis of Nucleic Acids by Matrix-Assisted Laser Desorption Ionization Time-Of-Flight Mass Spectrometry With a Data Algorithm. Rapid Commun. Mass Spectrom. 2025, 39, e9981. [Google Scholar] [CrossRef] [Scilit]
- Storch, G.A. Diagnostic virology. Clin. Infect. Dis. 2000, 31, 739–751. [Google Scholar] [CrossRef] [Scilit]
- Web Page of the MassUp Application. Available online: https://www.sing-group.org/mass-up/ (accessed on 13 April 2023).
- Victoriano, C.M.; Pask, M.E.; Malofsky, N.A.; Seegmiller, A.; Simmons, S.; Schmitz, J.E.; Haselton, F.R.; Adams, N.M. Direct PCR with the CDC 2019 SARS-CoV-2 assay: Optimization for limited-resource settings. Sci. Rep. 2022, 12, 11756. [Google Scholar] [CrossRef] [Scilit]
- Martin, G.; Abreu-Salinas, F. Development of an Indirect Immunofluorescence Assay for SARS-CoV-2 Nasopharyngeal Samples, 16 April 2025, PREPRINT (Version 1) Available at Research Square. Available online: https://www.researchsquare.com/article/rs-6456821/v1 (accessed on 15 June 2025).
- Rocca, M.F.; Zintgraff, J.C.; Dattero, M.E.; Santos, L.S.; Ledesma, M.; Vay, C.; Prieto, M.; Benedetti, E.; Avaro, M.; Russo, M.; et al. A combined approach of MALDI-TOF mass spectrometry and multivariate analysis as a potential tool for the detection of SARS-CoV-2 virus in nasopharyngeal swabs. J. Virol. Methods 2020, 286, 113991. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Chivte, P.; LaCasse, Z.; Seethi, V.D.R.; Bharti, P.; Bland, J.; Kadkol, S.S.; Gaillard, E.R. MALDI-ToF protein profiling as a potential rapid diagnostic platform for COVID-19. J. Mass Spectrom. Adv. Clin. Lab. 2021, 21, 31–41. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Torres, I.; Giménez, E.; Vinuesa, V.; Pascual, T.; Moya, J.M.; Alberola, J.; Martínez-Sapiña, A.; Navarro, D. Matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF-MS) proteomic profiling of cerebrospinal fluid in the diagnosis of enteroviral meningitis: A proof-of-principle study. Eur. J. Clin. Microbiol. Infect. Dis. 2018, 37, 2331–2339. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Verhagen, L.M.; van den Bergh, M.R.; Bogaert, D. The upper respiratory tract microbiota of healthy adults is affected by Streptococcus pneumoniae carriage, smoking habits, and contact with children. Microbiome 2023, 11, 146. [Google Scholar] [CrossRef] [Scilit]
- Allen, E.K.; Koeppel, A.F.; Hendley, J.O.; Turner, S.D.; Winther, B.; Sale, M.M. Characterization of the nasopharyngeal microbiota in health and during rhinovirus challenge. Microbiome 2014, 2, 22. [Google Scholar] [CrossRef] [Scilit]
- Torres-Corral, Y.; Santos, Y. Predicting antimicrobial resistance of Lactococcus garvieae: PCR detection of resistance genes versus MALDI-TOF protein profiling. Aquaculture 2022, 553, 738098. [Google Scholar] [CrossRef] [Scilit]
- Chen, M.; Wei, X.; Zhang, J. Differentiation of Bacillus cereus and Bacillus thuringiensis Using Genome-Guided MALDI-TOF MS Based on Variations in Ribosomal Proteins. Microorganisms 2022, 10, 918. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Hamidi, H.; Bagheri Nejad, R.; Es-haghi, A.; Ghassempour, A. A Combination of MALDI-TOF MS Proteomics and Species-Unique Biomarkers’ Discovery for Rapid Screening of Brucellosis. J. Am. Soc. Mass Spectrom. 2022, 33, 1530–1540. [Google Scholar] [CrossRef] [Scilit]
- Chen, Q.; Liu, J.; Liang, W.; Chen, Y.; Dou, M.; Liu, Z.; Chen, Y.; Zheng, Z.; Zhu, B.; Lin, Y. Clinical Features, Replication Competence, and Innate Immune Responses of Human Adenovirus Type 7 Infection. J. Infect. Dis. 2021, 223, 1390–1399. [Google Scholar] [CrossRef] [Scilit]
- Jarrold, M.F. Applications of Charge Detection Mass Spectrometry in Molecular Biology and Biotechnology. Chem. Rev. 2022, 122, 7415–7441. [Google Scholar] [CrossRef] [Scilit]
- Yu, J.; Chudinov, A.; Li, L.; Liu, J.; Gao, W.; Huang, Z.; Zhou, Z.; Nikiforov, S.; Kozlovskiy, V. Improvement of m/z accuracy for linear matrix-assisted laser desorption/ionization time-of-flight mass spectrometer. Rapid Commun. Mass Spectrom. 2020, 34, e8701. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Calderaro, A.; Chezzi, C. MALDI-TOF MS: A Reliable Tool in the Real Life of the Clinical Microbiology Laboratory. Microorganisms 2024, 12, 322. [Google Scholar] [CrossRef] [Scilit]
- Lohnes, K.; Quebbemann, N.R.; Liu, K.; Kobzeff, F.; Loo, J.A.; Ogorzalek Loo, R.R. Combining high-throughput MALDI-TOF mass spectrometry and isoelectric focusing gel electrophoresis for virtual 2D gel-based proteomics. Methods 2016, 104, 163–169. [Google Scholar] [CrossRef] [Scilit]
- Cheng, C.; Zong, Y.; Duan, F.; Chen, Z.; Liu, X.; Wu, K. Identification of herpes simplex virus type 1 infection by matrix-assisted laser desorption/ionization time-of-flight mass spectrometry. Exp. Eye Res. 2025, 255, 110391. [Google Scholar] [CrossRef] [Scilit]
- Martín, G.; Rojo, S.; Campo, R.; Costales, I.; Perez, Z.; Garcia, X.; Alavarez-Arguelles, M.; Melon-Garcia, S. Identification of DNA virus in conventional culture by matrix assisted laser desorption-ionization time of flight/mass spectrometry (MALDI-TOF/MS). In Abstract Book of the European Congress of Clinical Microbiology and Infectious Diseases; ECCMID: Paris, France, 2020; p. 3733. [Google Scholar]
- Novo, M.G. Virus Respiratorio Sincitial. Marcadores Virales y Relación Con Otros Virus. Ph.D. Thesis, Universidad de Oviedo, Oviedo, Spain, 2019. [Google Scholar]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the author. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.