Abstract
In X-ray imaging, tissue scattering is an important factor that degrades image clarity, especially using a portable gridless X-ray imaging device. This study focuses on using Monte Carlo simulation to quantify the effect of scatter radiation on image resolution, by analyzing the point spread function (PSF) and the corresponding modulation transfer function (MTF). Lateral energy absorption profiles in tissue and a cesium iodide (CsI) scintillator were calculated at different X-ray tube voltages (70–90 kV) and filter configurations. Results showed that 85.7% of the total scattered radiation is concentrated at a distance of 4 cm from the central axis for the tissue and 67.37% for the CsI scintillator. The MTF remained high at low spatial frequencies (23% at 0.04 cycles/cm) but dropped at mid frequencies (0.015–0.025 at 0.3–0.6 cycles/cm) and was almost zero at high frequencies (0.004 at 0.8 cycles/cm), indicating loss of detail due to scattering. Increasing the thickness of the filter or adding a copper (Cu) filter reduced the contrast at low spatial frequencies (from 23% to 21%). The study quantitatively investigated the MTF degradation in portable X-ray imaging devices without grid, due to scatter. These results may aid in the development of scatter correction algorithms to improve image quality without the need for an anti-scatter grid.
1. Introduction
X-ray imaging is the most common diagnostic method worldwide. A prerequisite for the derived anatomical image is clarity, that is, the ability of accurately representing the anatomical information. However, the detector characteristics impose an upper limit on the desired spatial resolution and image noise [1]. Additional blur in the final image may be the result of the X-ray interactions within the tissues, especially when scattering events (elastic and inelastic) are dominant [2,3,4,5,6]. The scattered X-rays appear in the final image as events with dislocated interaction sites with respect to the primary interaction position within the tissue. This phenomenon degrades image clarity and may hinder diagnostic information. The occurrence is more frequent during imaging of thicker body regions, where blurriness is created, leading to reduced contrast and spatial resolution [3,5,7]. During chest or abdominal X-rays, the scattered radiation produced on the patient’s body can be partially reduced by anti-scatter grids. To compensate for the simultaneous attenuation of primary radiation by the grid, patient exposure is increased. However, this practice remains controversial, especially with digital systems [8].
Fixed radiographic units typically use focused high-ratio grids (12:1) to control scatter. In contrast, such high-ratio grids are rarely used in portable radiographic units due to their sensitivity to misalignment. In portable radiography units, grids typically have lower ratios ranging from 6:1 to 10:1. However, these grids are often misaligned, either due to off focus positioning, tilting or lateral displacement, which can increase image noise and reduce grid transmission and image contrast [9,10]. The effect of misalignment becomes more pronounced as the grid ratio increases. Improper alignment, especially during portable chest radiography or in emergency situations, can lead to asymmetric image attenuation or even mimic pathology, such us changes in lung lucency. For these reasons, grids are not commonly used in portable radiography [9,10].
In exposures where grids are not implemented, the degradation of image quality depends on the spatial distribution of the secondary radiation. To address the problem of grid removal, software methods have been proposed for gridless imaging [2,11]. The use and future development of such methods require the knowledge of scatter radiation spread properties or a quantification of the corresponding degradation of image quality.
The point spread function (PSF) is the response of an imaging system to a point object (or point source). It essentially expresses the blurring in the image of a point during image formation. The PSF is affected by factors such as radiation scattering in the human body and in the detector, as well as the spread of light within the phosphor material (scintillator). However, the visualization of the PSF alone does not provide complete information about the width or shape, which are necessary for a complete description of the blurring characteristics of the system. The effect of scatter on image quality may be estimated by the PSF and by the modulation transfer function (MTF) in the spatial frequency domain. The MTF provides additional information, describing how different spatial frequencies are transferred by the system, allowing a more compete assessment of blur and resolution [12]. Specifically, it evaluates how the system maintains and reproduces high-frequency elements of the image. Increased MTF values means that fine details are retained and reproduced well, while decreased MTF values shows that detail is lost or deteriorated. The effect of secondary radiation spread, which incorporates mainly scattered X-rays and in some extended K X-rays, can be considered as an additional blur that further degrades the spatial resolution [13]. In this study, the PSF and the corresponding MTF of the secondary radiation spread, which is obtained from Monte Carlo simulations, are calculated. The resulting MTF reflects the degradation caused mainly by scattering and does not including the optical response of the scintillator.
The distribution of scatter has been widely investigated in medical imaging, with particular focus on its impact on image quality. In 1977, Wagner’s work laid the foundation for future studies, emphasizing the need for a holistic approach to understanding how scattered radiation and noise affect object detection [14]. Shortly thereafter, in 1979, Muntz’s work focused on the relationship between the focus size, magnification, and scatter and developed a model that incorporates all of these factors in order to improve image quality assessment [15]. In 1980, Rudin focused on the effectiveness of scatter rejection using rotating beam technology, while Johns in 1983 attempted to generalize the scatter distribution, investigating the angular scatter distribution and the role of gratings in reducing its effect. Boone et al. in 1986 introduced the MTF to analyze the spatial effects of scatter and to remove blur from clinical images [16,17,18]. Sorenson in 1985 and Krol in 1996, respectively, examined how scattered radiation interacts with magnification, using contrast enhancement factors and detective quantum efficiency (DQE) to improve the performance of X-ray systems [19,20]. These studies demonstrate the progress in understanding the scattering distribution and the development of methods to improve image quality in medical imaging.
In modern research, the accurate modeling of scattering remains a critical element. For example, in 2010, Kuhls-Gilchrist et al. proposed an accurate method to measure the MTF using the noise response to understand and quantify scattering phenomena in digital radiography [21]. In the same year, Gao et al. introduced in computed tomography (CT) an innovative approach to correct scattering using primary modulation to mitigate its effects [22]. In 2019, Cheng et al. proposed a beam propagation technique to simulate how scattering degrades the PSF. This has enabled more accurate modeling of scattering phenomena in medical imaging systems [23]. In 2025, Tivnan et al. used Fourier diffusion models to examine MTF and the noise power spectrum (NPS), which provided valuable insights into controlling image characteristics in stochastic imaging systems [24]. These experimental and theoretical methods continue to improve the understanding of scattering phenomena, contributing to the development of more accurate and reliable imaging results in the medical field.
The present study complements earlier findings by the published literature, focusing on the effect of scattered radiation on image sharpness and resolution in portable X-ray imaging devices without a grid. In this work, the secondary radiation absorption distribution, resulting from Monte Carlo simulations is used to calculate the related PSF function. Hence, the corresponding MTF is derived, which can be used as a straightforward metric of how scatter radiation degrades image resolution.
Specifically, this study used the PENELOPE Monte Carlo software package (version 2018) [25,26,27] to calculate the secondary radiation lateral absorbed energy distribution of a human tissue closely followed by a cesium iodide (CsI) scintillator. In this way, the spatial distribution of the transmitted scattered photons was obtained. The examined X-ray tube voltage ranged from 70 to 90 kV with Al and Cu filtration combinations. These parameters correspond to abdomen and pelvis examinations that may be performed with a mobile X-ray system. The results of our study may contribute to further understanding secondary radiation spread in a detector under gridless examination conditions and may assist in exposure parameters’ optimization or the development of specific patient scatter removal software.
2. Materials and Methods
The simulation of ionizing radiation was performed using the Monte Carlo method by employing the PENELOPE based simulation package PenEasy (version 2020) [25,26,27]. X-ray spectra were used as input in the simulations in order to find the profile of the lateral energy absorption corresponding to the scattered radiation for each spectrum. The maximum number of histories considered in the simulation was 109; however, a simulation time limit of 1200 s was set to ensure computational efficiency. The cut-off energies for electrons and photons were chosen as 2 keV and 1 keV, respectively. The examined tungsten generated X-ray spectra [28] ranged from 70 kV to 90 kV, with standard aluminum filters of 2 mm and 2.5 mm. In addition, a combination of aluminum–copper filters with values of 2 mm Al + 0.1 mm Cu and 2.5 mm Al + 0.1 mm Cu was used. All these values are commonly used in portable radiography and adjusted according to the patient’s weight. The spektr 3.0 computational kit was used to calculate the X-ray spectra [28]. To find the scattered radiation profile in the simulations, a PENELOPE built-in tally for calculating cylindrical dose distribution was considered. The distribution of the absorbed energy resulting from the secondary radiation was laterally sampled with a step size of 0.1 cm. The sample bin was calculated assuming a maximum radius of a cylinder divided into , as depicted in Equation (1):
In the simulation procedure, , corresponding to the secondary radiation spread length, was set to 12 cm, while the number of bins () was adjusted to ensure a constant sampling distance of 0.1 cm.
The lateral distribution was determined through Monte Carlo simulations performed for a 10 cm tissue. The simulated tissue used was an international commission on radiation units and measurements (ICRU) four-component tissue, available from PENELOPE data. Its density was 1.0 g/cm3, and its elemental composition was as follows: hydrogen (H) (10.1%), carbon (C) (11.1%), nitrogen (N) (2.6%) and oxygen (O) (76.2%). Its composition reflects a typical approximation of average human soft tissue and provides more realistic interaction probabilities than water, particularly for photoelectric and Compton effects. The X-rays emerging from the tissue were assumed to impinge on a 0.1 cm thick CsI scintillator placed 5 cm away from the tissue. PenEasy (version 2020) was used to calculate the absorption of secondary radiation at different lateral positions along the central beam’s path [24,25]. For this purpose, the thickness of the scintillator was subdivided into sublayers of 0.02 cm, where the 0.02 cm CsI thickness is within the range of the CsI thickness studied in the literature [29]. The study focused on the substrate closer to the tissue exit and the adjacent CsI scintillator layer of 0.02 cm thickness. The source was 0.1 cm away from the tissue and had an angular aperture of 0.05 degrees. The average statistical error for these conditions was 1.19% ± 0.08%, with values ranging between 1.08% and 1.30%, as automatically determined by PenEasy.
A simple representation of the geometry is shown in Figure 1.
Figure 1.
The geometry of the experimental setup.
A common power law equation describing each scatter profile was assumed, and its parameters were determined by means of fitting for each profile. The mathematical equation had the following form:
This fitting was applied to both the absorbed dose to the tissue and the energy absorbed in the CsI scintillator (Table 1 and Table 2, respectively).
Table 1.
Secondary radiation fitted equations for 10 cm tissue thickness.
Table 2.
Secondary radiation fitted equations for 0.1 cm CsI scintillator thickness.
Although, the simulation output provided discrete data points, the use of analytical equations allowed the construction of a smooth and continuous scattering profile, facilitating further analysis. It allowed the reconstruction of radially symmetric PSF by axial rotation of fitted curve, the calculation of MTF and the quantitative comparison of different spectra and filter conditions using the fitted parameters.
For this reason, each equation was mathematically rotated around the central axis, and a two-dimensional distribution of scattered radiation was determined corresponding to the PSF. A Matlab Version 9.13.0.2193358 (R2022b) code was developed for the computation of the scattered radiation distribution [30]. From the PSF, the MTF was calculated as the Fourier transform of the PSF integral in one dimension, which corresponded to a line spread function (LSF), as shown in Equation (3):
where denotes the Fourier transform, and f is the spatial frequency.
A flowchart of the method is presented in Figure 2.
Figure 2.
Flowchart of the method.
3. Results
The spectra of 70, 80, and 90 kV tube voltages that were used for the simulations have a significant degree of overlap, especially in the lower energies as shown in Figure 3. The peaks at approximately 60 keV and 70 keV correspond to tungsten characteristic radiation. In addition, the average energy of the spectra is 41.6 keV, 45.2 keV and 48.5 keV for the 70 kV, 80 kV and 90 kV X-ray tube voltages, respectively. A similar deduction may be obtained by the different filtration used in clinical routine. For instance, as shown in Figure 3 for the 90 kV tube voltage generated spectra, the different filters 2 mm Al, 2.5 mm Al, 2 mm Al + 0.1 mm Cu, 2.5 mm Al + 0,1 mm Cu result in similar average energy values of 48.5 keV, 49.1 keV, 52.4 keV and 52.8 keV, respectively.
Figure 3.
X-ray spectra of 70 kV tube voltage filtered by 2 mm Al, 80 kV tube voltage filtered with 2 mm Al and 90 kV tube voltage filtered by 2 mm Al, 2.5 mm Al, 2 mm Al + 0.1 mm Cu and 2.5 mm Al + 0.1 mm Cu.
In addition, Table 1 and Table 2 list the mathematical equations that were fitted to the data and their corresponding coefficient of determination R2 for each spectrum.
It can be observed that the energy absorbed laterally from the secondary radiation in the tissue (Table 1) can be always fitted with a monotonically decreasing equation with R2 > 0.92. In practice, it was found that the radiation absorbed energy spreads laterally up to 12 cm distance for a tissue thickness of 10 cm. Similarly, in Table 2, the corresponding R2 values of the laterally absorbed energy distribution are all higher than 0.90. It was found that the lateral spread of the absorbed secondary radiation had no zero values up to 12 cm. All the lateral energy profiles were normalized to their maximum value for each simulation case. In Figure 4 and Figure 5, some of the normalized absorbed energy scatter lateral distributions are indicatively demonstrated.
Figure 4.
Normalized scatter lateral distribution of 70 kV, 80 kV and 90 kV tube voltage-generated X-rays filtered by 2 mm Al in the tissue and CsI scintillator.
Figure 5.
Normalized secondary scatter lateral distribution of 90 kV voltage-generated X-rays filtered by 2 mm Al, 2.5 mm Al, 2 mm Al + 0.1 mm Cu and 2.5 mm Al + 0.1 mm Cu in the tissue and CsI scintillator.
In Figure 6, the MTF curves corresponding to the spatial distribution of the energy deposited by secondary scattered radiation in the CsI scintillator are presented. These curves were calculated from the PSFs derived from the equations in Table 2, based on the X-ray spectra of the 70, 80, 90 kV tube voltages, with a 2 mm aluminum filter. Figure 6 shows that the system demonstrates excellent performance at low spatial frequencies (0 to 0.2 cycles/cm), maintaining high contrast for larger structures. A gradual decline is observed at mid spatial frequencies (0.3 to 0.6 cycles/cm), while at high spatial frequencies (0.7 to 1 cycles/cm), the MTF drops sharply. The variation in tube voltages (70, 80, 90 kV) does not significantly affect the system’s performance across this frequency range, as the MTF curves remain nearly identical. Specifically, at all spatial frequencies, the MTF curves have a percentage deviation of less than 2.7%, while the standard deviation of the MTF values remains below 0.0014. This indicates a negligible effect of the tube voltage on the spatial resolution of the system.
Figure 6.
MTF curves of secondary scattered radiation contribution for the X-ray spectra at 70, 80, 90 kV by a 2 mm aluminum filter for CsI scintillator.
At low spatial frequencies (0 to 0.2 cycles/cm), the system exhibits high MTF values, reaching approximately 23% at a frequency of 0.04 cycles/cm for the different kV values. Specifically, at 0.04 cycles/cm, for 70, 80 and 90 kV, the MTF values are 22.9%, 23.2% and 23.1%, respectively. This indicates that the system has excellent preservation of contrast for larger structures. Further, the MTF values at 70 kV are slightly lower than at 90 kV, and although one might theoretically expect higher contrast at lower kV, the resulting differences are very small. The effect of the tube voltage is practically negligible compared to other dominant factors, such as scattering and light propagation within the CsI scintillator. Therefore, these small deviations do not substantially affect the overall spatial performance of the system.
At low-to-mid spatial frequencies (0.3 to 0.6 cycles/cm), a significant drop in MTF is observed across all voltage settings (70, 80, 90 kV). The MTF values fall below 0.02 at approximately 0.5 cycles/cm. This indicates that the system gradually loses its ability to reproduce details of smaller dimensions; however, the performance remains similar between the different kV values (70, 80, 90 kV).
At high frequencies above 0.7 cycles/cm, the MTF is almost zero (0.01), regardless of the X-ray tube voltage considered. Specifically, the MTF drops from 0.23 at 0.04 cycles/mm to 0.004 at 0.8 cycles/cm. This corresponds to a reduction of approximately 98%. This drop indicates that the system has significant difficulty in reproducing fine details at higher spatial frequencies. In addition, the spatial resolution is clearly affected by the lateral spread caused by scattered radiation within the CsI scintillator, which limits the scintillator’s performance for fine details in gridless examinations.
In Figure 7, the MTF curves of the X-ray spectrum at 90 kV tube voltage with aluminum filters of 2 mm, 2.5 mm and with a combination of aluminum–copper filters with values 2 mm Al + 0.1 mm Cu, and 2.5 mm Al + 0.1 mm Cu of the energy absorbed in the detector are presented for the CsI scintillator.
Figure 7.
MTF curves of secondary scattered radiation contribution for the X-ray spectra at 90 kV by aluminum filters of 2 mm, 2.5 mm and by a combination of aluminum–copper filters with values 2 mm Al + 0.1 mm Cu, and 2.5 mm Al + 0.1 mm Cu for CsI scintillator.
At low spatial frequencies (0 to 0.2 cycles/cm), the MTF curve of 90 kV with a 2 mm aluminum filter exhibits a slightly higher MTF compared to the others, which suggests better contrast preservation in this region.
This may be due to the changes in the X-ray spectra caused by different filter combinations, which lead to beam hardening and spectral shaping. These spectral variations slightly change the spatial distribution of secondary scattered radiation within the tissue and its lateral spread when it reaches the scintillator. This affects the corresponding PSF caused by scattering, resulting in small differences in MTF values at low spatial frequencies.
In contrast, by increasing the thickness of the aluminum or adding the copper filter, the X-ray spectrum components are shifted into higher energies, yielding a higher average energy. Despite the fact that the maximum spectrum energy does not change, the energy shift increases the average photon energy. The increase in energy reduces the probability of a photoelectric effect and increases the probability of Compton scattering. Therefore, the lateral energy deposition events are increased, and a slightly broader lateral scatter distribution occurs in the tissue facing the scintillator, and consequently lower MTF values are generally anticipated. However, the lower MTF values (21%) for the tube voltage of 70 kV shown in Figure 6 may be attributed to the characteristic X-rays of CsI. More specifically, the average energy of the 70 kV X-ray tube voltage spectrum equals 41.6 keV, and it is above the CsI k-edges corresponding to 34 keV and 36 keV [31]. Therefore, the 70 kV tube voltage induces a higher number of characteristic X-rays, which further degrade the MTF.
At low-to-mid spatial frequencies (0.3 to 0.6 cycles/cm), a decrease in MTF is observed for all filter settings. The 2 mm Al filter also maintains slightly higher values in the frequency range compared to thicker or copper-added filters. In addition, Cu filters (2 mm or 2.5 mm Al + 1 mm Cu) have slightly reduced MTF values. This is due to the spectral hardening and the cutoff of low-energy photons, which contribute to contrast in the mid-frequency regions.
At high frequencies above 0.7 cycles/cm, all the curves approach near zero. This indicates the limitation of the system to reproduce fine structures regardless of filter thickness or combination. This was expected, since the range of light photons created by X-rays within the scintillator, acts as a blurring factor affecting spatial resolution. Moreover, the use of a copper filter shows slightly worse performance, even if the curves are very close together. The reason is the same as the case of low and medium spatial frequencies, and this is probably due to the slightly lower X-ray absorption within the human body, which in turn corresponds to lower values of X-ray absorption coefficients, resulting in lower contrast, which seems to appear even at high frequencies.
4. Discussion
In Figure 4, the scatter lateral distribution decreases as the distance increases, both in the tissue and in the CsI scintillator. The largest percentage of scattering is at a distance 0–4 cm, while beyond 4 cm, the scattered radiation has very low values. For example, the percentage of total scattered radiation within 4 cm in the tissue is 85.66%, and in the CsI scintillator, it is 67.37%, while beyond 4 cm only 14.34% is distributed in the tissue and 32.63% in the CsI scintillator. As the kV increases, Compton scattering is enhanced, but without a significant increase in the scatter lateral distribution, maintaining an almost constant distribution shape. For example, in the tissue, the value of the scatter at a 2 cm lateral distance is 0.073 at 70 kV and 0.079 at 90 kV. Respectively, in the CsI scintillator, the value at 70 kV is 0.105, and at 90 kV, it is 0.129.
This can be explained since the secondary radiation observed in both the tissue and the detector is mainly produced by Compton scattering. In human tissue, Compton scattering is the dominant mechanism within the examined energy range (70–90 kV), as it predominantly occurs in low atomic number elements such as water and biological tissues. In contrast, Rayleigh scattering does not contribute significantly to the production of the secondary radiation, but within the studied spectrum, it may slightly affect its lateral and angular spread.
Furthermore, regardless of the filter type, as seen in Figure 5, the scatter lateral distribution decreases as the distance increases, both in the tissue and in the CsI scintillator. Similar to Figure 4, the largest percentage of scattering is found at a distance from 0 up to 4 cm, while beyond 4 cm, the values of the scattered radiation have a decrease of approximately 3–6% in the tissue and 6–9% in the CsI scintillator. Further, the average value of the scattered radiation energy absorbed in the CsI scintillator is higher compared to the corresponding energy deposited within the tissue (e.g., for 2 mm Al, 0.072 in the detector and 0.050 in the tissue, an increase of 40% can be seen). The higher energy absorption is due to the higher density and effective atomic number of CsI. As a result, a large part of the secondary radiation exiting from the tissue is absorbed by the scintillator, leading to higher values in CsI. These differences show the contribution of scatter radiation to the detector signal, which negatively affects the image quality by reducing the MTF values, as seen in Figure 7.
In Figure 6, it can be demonstrated that at low spatial frequencies (0 to 0.2 cycles/cm), expressing the system’s ability to resolve larger structures, the system maintains high contrast in the examined energy range (70–90 kV), while at mid (0.3–0.6 cycles/cm) and high (0.7–1 cycles/cm) spatial frequencies, the MTF drops sharply. For example, from 0 to 0.2 cycles/cm, the MTF ranges from 0.18 to 0.23. However, from 0.3 to 0.6 cycles/cm, the MTF ranges from 0.015 to 0.025, and at frequencies from 0.7 to 1 cycles/cm, the MTF approaches values close to zero. This shows the strong influence of scattered radiation on the detector’s resolution and can be explained by the fact that scattered radiation affects lower frequencies less due to the nature of the lateral spread of the scattered radiation. With higher energy X-rays, penetration through tissue increases, resulting in a broader PSF compared to lower energy X-rays. This impacts the MTF curve, leading to the reduction in image contrast.
Additionally, in Figure 7, the MTF values are slightly lower with the addition of the copper filter (0.1 mm) or with increasing the thickness of the aluminum filter (2.5 mm). Increasing the thickness of the filter removes more low-energy photons, resulting in reduced absorption in the detector, increased scattering probability and finally wider lateral spread of the absorbed energy. Thus, the choice of the filter probably affects the image quality, and this should also be evaluated when choosing the irradiation conditions for the diagnostic requirements of each examination.
Moreover, the analysis of attenuation coefficients, using the XmuDat photon attenuation database software (version 1.0.1), for the ICRU four-component soft tissue used in the Monte Carlo simulations, provides additional insights into how these phenomena influence the radiation spread and the final image quality [31]. In the tissue, the total attenuation coefficient decreases from 0.19 cm2/g at 70 keV to 0.17 cm2/g at 90 keV. The Compton scattering is the dominant mechanism in this energy range. Specifically, it accounts for approximately 89.7% of the total attenuation at 70 keV, increasing to 94.31% at 90 keV. In contrast, the photoelectric absorption (p.e.) is low and decreases significantly from 4.33% at 70 keV to 2.06% at 90 keV. Similarly, Rayleigh scattering also shows a small decrease, from 5.28% to 3.59%. These ratios clearly indicate that as the energy increases, the primary radiation absorption decreases, while the scattering effects become more pronounced. This change significantly affects the image contrast and radiation penetration, especially in soft tissues. In addition, the photoelectric effect implies the presence of K characteristic radiation. These data can be observed in Table 3.
Table 3.
Attenuation coefficients for the ICRU four-component soft tissue at energies of 70, 80 and 90 keV.
Michail et al., in 2024, showed that the use of CsI in X-ray detectors provides images with high spatial resolution and corresponding MTF curves extending to high spatial frequencies, depending upon the X-ray energy and the scintillator thickness (e.g., at a spatial frequency of around 0.9 cycles/mm, the MTF is approximately 0.2) [12]. This is expected when CsI is irradiated with normally incident X-rays or with the utilization of an X-ray bucky. In the absence of it, like the cases examined in this work, a significant drop in the maximum resolution is expected. Inspecting Figure 6 and Figure 7, a significant drop is expected at approximately 0.2 cycles/cm, where our calculations resulted in a value for the secondary radiation MTF approximately equal to 0.03. The accurate interpretation of the MTF values should however consider the statistical uncertainty of the Monte Carlo simulation of this study, which is equal to 1.3% and is demonstrated in the error bars of Figure 4, Figure 5, Figure 6 and Figure 7, as well as the deviation between the Monte Carlo data and that calculated by the fitted equation. The former can be reduced by increasing the simulation histories and simulation time. Regarding the fitting error, this work, for consistency, has assumed a power law fitting yielding an R2 > 0.9 in each case. For the case of the 90 kV + 2.5 mm Al, where the minimum R2 value (0.90) was observed, the average percent deviation between the Monte Carlo derived and the fitted data points was 2.93%. For the 80 kV + 2.5 mm Al + 0.1 mm Cu, where the maximum R2 value (0.94) was observed, the average percent deviation was 1.87%. If each profile could be fitted with a combination of functions, then higher R2 values and smaller deviations between the data would be observed.
The secondary radiation escaping from the tissue is expected to have less energy than that of the incident X-ray photons. The density and the effective atomic number of CsI 4.51 g/cm3 and 54.04, respectively, are much higher than the corresponding ones of the tissue, density: 1.06 g/cm3 and Effective Z: 7.64. Thus, the probability of photoelectric interaction in the CsI is increased, suggesting that the absorbed energy spatial distribution calculated in the CsI is comparable with the exiting ionizing energy spatial distribution. This may be even valid for the lower energy X-rays subject to Rayleigh scattering in the tissue. Figure 8 and Table 4 demonstrate the related CsI coefficients.
Figure 8.
CsI attenuation coefficients from 10 keV to 90 keV.
Table 4.
Attenuation coefficients for CsI at energies of 70, 80 and 90 keV.
It should, however, be emphasized that the energy deposition calculated within the CsI could not take into account phenomena like Rayleigh scattering events that occur in the scintillator. Their effect may be of importance for higher scintillator thicknesses, due to the increased probability of absorption.
5. Conclusions
This study shows quantitatively how secondary scattered radiation affects the image quality in portable X-ray imaging devices without a grid. Firstly, it provides a combined analysis of the lateral spread of secondary radiation in tissue and in a CsI scintillator, due to its extensive use in indirect digital detectors. Then, through the adapted mathematical equations (Table 1 and Table 2) for the reconstruction of the PSF and the calculation of the MTF at various X-ray energies and filter configurations, the basis is formed for the design of an algorithm that will correct the image degradation due to scattering without the use of a grid.
The results show that the spatial distribution of absorbed energy in the CsI closely follows the lateral spread of secondary radiation emerging from the tissue, while the MTF minimizes rapidly to 0.4 cycles/cm and approximates zero at 0.6 cycles/cm. This implies that any scatter correction methods can increase the available image information in the range from 0.4 cycles/cm up to 0.6 cycles/cm, leading to better visualization of the tissue dimensions between 1.7 cm and 2.5 cm. Further image detail improvement may be achievable if current gridless scatter correction software based on standard [32,33,34] or artificial intelligence methods [35,36] utilizes the results of scatter distribution and the corresponding frequency response provided by personalized X-ray tube voltage, filtration and tissue thickness.
Author Contributions
Conceptualization, N.K. and I.A.; methodology, N.K., I.A. and P.L.; software, I.A.; validation, C.M., N.K. and P.L.; formal analysis, I.K.; investigation, I.A.; resources, G.F., I.V. and N.K.; data curation, P.L.; writing—original draft preparation, I.A.; writing—review and editing, I.A., I.V., C.M., P.L., N.K., G.F. and I.K.; visualization, N.K. and I.A.; supervision, N.K., P.L. and C.M.; project administration, N.K.; All authors have read and agreed to the published version of the manuscript.
Funding
The publication fees of this study (APC) was partially funded by the University of West Attica, Athens, Greece, through IOAP discount. This research received no other external funding.
Institutional Review Board Statement
Not applicable.
Informed Consent Statement
Not applicable.
Data Availability Statement
The original contributions presented in the study are included in the article; further inquiries can be directed to the corresponding author.
Conflicts of Interest
The authors declare no conflicts of interest.
References
- Michail, C.; Liaparinos, P.; Kalyvas, N.; Kandarakis, I.; Fountos, G.; Valais, I. Radiation Detectors and Sensors in Medical Imaging. Sensors 2024, 24, 6251. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Sayed, M.; Knapp, K.M.; Fulford, J.; Heales, C.; Alqahtani, S.J. The principles and effectiveness of X-ray scatter correction software for diagnostic X-ray imaging: A scoping review. Eur. J. Radiol. 2023, 158, 110600. [Google Scholar] [CrossRef] [Scilit]
- Aichinger, H.; Dierker, J.; Joite-Barfuß, S.; Säbel, M. Radiation Exposure and Image Quality in X-Ray Diagnostic Radiology: Physical Principles and Clinical Applications; Springer: Berlin/Heidelberg, Germany, 2004. [Google Scholar] [CrossRef] [Scilit]
- Aichinger, H.; Dierker, J.; Joite-Barfuß, S.; Säbel, M. Penetration of X-Rays. In Radiation Exposure and Image Quality in X-Ray Diagnostic Radiology; Springer: Berlin/Heidelberg, Germany, 2004. [Google Scholar] [CrossRef] [Scilit]
- Hendee, W.R.; Ritenour, E.R. Medical Imaging Physics, 4th ed.; John Wiley & Sons: Hoboken, NJ, USA, 2002. [Google Scholar] [CrossRef] [Scilit]
- Jain, A.; Bednarek, D.R.; Rudin, S. Scatter reduction for high resolution image detectors with a region of interest attenuator. Proc. SPIE—Int. Soc. Opt. Eng. 2014, 19, 9033–903364. [Google Scholar] [CrossRef] [Scilit]
- Sprawls, P. Physical Principles of Medical Imaging, 2nd ed.; Medical Physics Publishing: Madison, WI, USA, 1993. [Google Scholar]
- Lehnert, T.; Naguib, N.N.; Wutzler, S.; Bauer, R.W.; Kerl, J.M.; Burkhard, T.; Schulz, B.; Larson, M.C.; Ackermann, H.; Vogl, T.J.; et al. Comparative Study Between Mobile Computed Radiography and Mobile Flat-Panel Radiography for Bedside Chest Radiography. Investig. Radiol. 2014, 49, 1–6. [Google Scholar] [CrossRef] [Scilit]
- Gauntt, D.M.; Barnes, G.T. An automatic and accurate X-ray tube focal spot/grid alignment system for mobile radiography: System description and alignment accuracy. Med. Phys. 2010, 37, 6402–6410. [Google Scholar] [CrossRef] [Scilit]
- Scott, A.W.; Gauntt, D.M.; Yester, M.V.; Barnes, G.T. High-ratio grid considerations in mobile chest radiography. Med. Phys. 2012, 39, 3142–3153. [Google Scholar] [CrossRef] [Scilit]
- Neamin, H. X-ray scatter correcting methods for digital radiographic imaging. J. Assoc. Med. Sci. 2018, 51, 128–137. [Google Scholar]
- Michail, C.; Liaparinos, P.; Kalyvas, N.; Kandarakis, I.; Fountos, G.; Valais, I. Phosphors and Scintillators in Biomedical Imaging. Crystals 2024, 14, 169. [Google Scholar] [CrossRef] [Scilit]
- Monnin, P.; Verdun, F.R.; Bosmans, H.; Pérez, S.R.; Marshall, N.W. A comprehensive model for X-ray projection imaging system efficiency and image quality characterization in the presence of scattered radiation. Phys. Med. Biol. 2017, 62, 5691–5722. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Wagner, R.F. Toward a unified view of radiological imaging systems. Part II: Noisy images. Med. Phys. 1977, 4, 279–296. [Google Scholar] [CrossRef] [Scilit]
- Muntz, E.P. Analysis of the significance of scattered radiation in reduced dose mammography, including magnification effects, scatter sup- pression, and focal spot and detector blurring. Med. Phys. 1979, 6, 110–117. [Google Scholar] [CrossRef] [Scilit]
- Rudin, S.; Bednarek, D.R. Improved contrast in special procedures using a rotating aperture wheel sRAWd device. Radiology 1980, 137, 505–510. [Google Scholar] [CrossRef] [Scilit]
- Johns, P.C.; Yaffe, M.J. Coherent scatter in diagnostic radiology. Med. Phys. 1983, 10, 40–50. [Google Scholar] [CrossRef] [Scilit]
- Boone, J.M.; Arnold, B.A.; Seibert, J.A. Characterization of the point spread function and modulation transfer function of scattered radiation using a digital imaging system. Med. Phys. 1986, 13, 254–256. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Sorenson, J.A.; Floch, J. Scatter rejection by air gaps: An empirical model. Med. Phys. 1985, 12, 308–316. [Google Scholar] [CrossRef] [Scilit]
- Krol, A.; Bassano, D.A.; Chamberlain, C.C.; Prasad, S.C. Scatter reduction in mammography with air gap. Med. Phys. 1996, 23, 1263–1270. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Kuhls-Gilcrist, A.; Jain, A.; Bednarek, D.R.; Hoffmann, K.R.; Rudin, S. Accurate MTF measurement in digital radiography using noise response. Med. Phys. 2010, 37, 724–735. [Google Scholar] [CrossRef] [Scilit]
- Gao, H.; Fahrig, R.; Bennett, N.R.; Sun, M.; Star-Lack, J.; Zhu, L. Scatter correction method for X-ray CT using primary modulation: Phantom studies. Med. Phys. 2010, 37, 934–946. [Google Scholar] [CrossRef] [Scilit]
- Cheng, X.; Li, Y.; Mertz, J.; Sakadžić, S.; Devor, A.; Boas, D.A.; Tian, L. Development of a beam propagation method to simulate the point spread function degradation in scattering media. Opt. Lett. 2019, 44, 4989–4992. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Tivnan, M.; Teneggi, J.; Lee, T.-C.; Zhang, R.; Boedeker, K.; Cai, L.; Gang, G.J.; Sulam, J.; Stayman, J.W. Fourier Diffusion Models: A Method to Control MTF and NPS in Score-Based Stochastic Image Generation. IEEE Trans. Med. Imaging 2025, 44, 3694–3704. [Google Scholar] [CrossRef] [Scilit]
- Baró, J.; Sempau, J.; Fernández-Varea, J.M.; Salvat, F. PENELOPE: An algorithm for Monte Carlo simulation of the penetration and energy loss of electrons and positrons in matter. Nucl. Instrum. Methods Phys. Res. B 1995, 100, 31–46. [Google Scholar] [CrossRef] [Scilit]
- Sempau, J.; Acosta, E.; Baró, J.; Fernández-Varea, J.M.; Salvat, F. An algorithm for Monte Carlo simulation of coupled electron-photon transport. Nucl. Instrum. Methods Phys. Res. B 1997, 132, 377–390. [Google Scholar] [CrossRef] [Scilit]
- Sempau, J.; Badal, A.; Brualla, L. A PENELOPE-based system for the automated Monte Carlo simulation of clinacs and voxelized geometries-application to far-from-axis fields. Med. Phys. 2011, 38, 5887–5895. [Google Scholar] [CrossRef] [Scilit]
- Punnoose, J.; Xu, J.; Sisniega, A.; Zbijewski, W.; Siewerdsen, J.H. Technical Note: Spektr 3.0—A computational tool for X-ray spectrum modeling and analysis. Med. Phys. 2016, 43, 4711–4717. [Google Scholar] [CrossRef] [Scilit]
- Badano, A.; Gagne, R.M.; Gallas, B.D.; Jennings, R.J.; Boswell, J.S.; Myers, K.J. Lubberts Effect in Columnar Phosphors: Lubberts Effect in Columnar Phosphors. Med. Phys. 2004, 31, 3122–3131. [Google Scholar] [CrossRef] [Scilit]
- The MathWorks, Inc. Matlab R2022b; The MathWorks, Inc.: Natick, MA, USA, 2022. Available online: https://ch.mathworks.com/?s_tid=gn_logo (accessed on 15 November 2025).
- Nowotny, R. XMuDat: Photon Attenuation Data on PC Version 1.0.1 of August 1998—Summary; International Atomic Energy Agency: Vienna, Austria, 1998; Available online: https://nds.iaea.org/records/crrr1-rkn63 (accessed on 21 November 2025).
- Mentrup, D.; Jockel, S.; Menser, B.; Neitzel, U. Iterative scatter correction for grid-less bedside chest radiography: Performance for a chest phantom. Radiat. Prot. Dosim. 2016, 169, 308–312. [Google Scholar] [CrossRef] [Scilit]
- Lisson, C.G.; Lisson, C.S.; Kleiner, S.; Regier, M.; Beer, M.; Schmidt, S.A. Iterative scatter correction for grid-less skeletal radiography allows improved image quality equal to an antiscatter grid in adjunct with dose reduction: A visual grading study of 20 body donors. Acta Radiol. 2019, 60, 735–741. [Google Scholar] [CrossRef] [Scilit]
- Kim, K.; Kang, S.; Kim, W.; Park, C.; Lee, D.; Cho, H.; Kang, W.; Park, S.; Kim, G.; Lim, H.; et al. A new software scheme for scatter correction based on a simple radiographic scattering model. Med. Biol. Eng. Comput. 2019, 57, 489–503. [Google Scholar] [CrossRef] [Scilit]
- Lee, H.; Lee, J. A Deep Learning-Based Scatter Correction of Simulated X-ray Images. Electronics 2019, 8, 944. [Google Scholar] [CrossRef] [Scilit]
- Onodera, S.; Lee, Y.; Tanaka, Y. Evaluation of dose reduction potential in scatter-corrected bedside chest radiography using U-net. Radiol. Phys. Technol. 2020, 13, 336–347. [Google Scholar] [CrossRef] [Scilit] [PubMed]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |