Abstract
A scalable and facile fabrication strategy is presented for developing a flexible, nanostructured, non-enzymatic electrochemical sensor for lactate detection based on copper-modified laser-induced graphene (CuNPs/LIG). A one-step electrodeposition process was employed to uniformly decorate the porous LIG framework with copper nanostructures, offering a cost-effective and reproducible approach for constructing enzyme-free sensing platforms. Scanning electron microscopy and energy-dispersive X-ray spectroscopy confirmed dense Cu nanostructure loading and efficient interfacial integration across the conductive LIG surface. The resulting CuNPs/LIG electrode exhibited excellent electrocatalytic performance, achieving a sensitivity of 8.56 μA µM−1 cm−2 with a low detection limit of 42.65 μM and a linear response toward lactate concentrations ranging from 100 to 1100 μM in artificial saliva under physiological conditions. The sensor maintained high selectivity in the presence of physiologically relevant interferents. Practical applicability was demonstrated through recovery studies, where recovery rates exceeding 104% showcase the sensor’s analytical reliability in complex biological matrices. Overall, this work establishes a robust, sensitive, and cost-efficient Cu-nanostructured LIG sensing platform, offering strong potential for non-invasive lactate monitoring in real-world biomedical and wearable applications.
1. Introduction
Lactate is a key metabolic intermediate generated during glycolysis under both aerobic and anaerobic conditions, making it an essential biomarker for assessing cellular metabolism, tissue oxygenation, and overall physiological status [1]. Elevated lactate levels are closely associated with hypoxia, sepsis, cardiovascular complications, and exercise-induced fatigue, emphasizing the clinical value of accurate and continuous lactate monitoring [2,3]. Conventional analytical techniques provide reliable quantification but are hindered by several limitations, including reliance on centralized laboratories, complex sample preparation, long turnaround times, and the inability to support real-time or point-of-care analysis [4,5,6]. These constraints have accelerated the demand for flexible, miniaturized, and noninvasive electrochemical platforms capable of delivering rapid, sensitive, and field-deployable lactate detection.
Electrochemical biosensors traditionally rely on enzymatic pathways particularly lactate oxidase (LOx) to achieve high selectivity towards lactate [7]. While effective, enzyme-based sensors suffer from intrinsic drawbacks such as poor stability, susceptibility to denaturation, and limited operational lifetime under variable environmental and physiological conditions [8]. To mitigate these issues, LOx immobilization strategies on unconventional substrates, including cotton fabric [9] and laser-induced graphene (LIG) [10], have been explored. Although these sensors exhibit encouraging sensitivity, the instability and degradation of enzymes remain major obstacles to long-term reliability and practical deployment.
Therefore, there has been a pronounced shift toward non-enzymatic electrocatalysts, leveraging nanomaterials with high intrinsic activity, enhanced electron transport properties, and structural robustness. Materials such as metal nanostructures [11], metal oxides [12], noble metal nanoparticles [13], metal–organic frameworks (MOFs) [14], MXenes (Ti3C2Tx) [15], and carbon-based nanostructures [16] have demonstrated significant promise for enzyme-free lactate oxidation. Among noble metals, platinum-based composites, such as sepiolite-supported Pt nanoparticles synthesized through chemical reduction, have achieved reliable detection across 10–1100 µM [17]. However, the high cost, limited availability, and fabrication complexity associated with noble metals restrict their scalability. In contrast, transition-metal-based catalysts, particularly nickel derivatives, offer compelling advantages due to their rich redox activity, strong affinity toward lactate, cost-effectiveness, and tunable electrochemical behavior. Ni(OH)2 modified indium tin oxide electrodes have demonstrated high sensitivity (0.1325 mA mM−1 cm−2) and broad detection ranges [18], while NiO-based platforms [19,20], and nickel nanostructures [21] have further improved catalytic efficiency. Additionally, hybrid structures combining nickel with carbon nanomaterials have shown synergistic enhancements in electron transfer and catalytic performance.
Among transition metals, copper-based nanomaterials have emerged as one of the most powerful electrocatalysts for non-enzymatic lactate sensing due to their accessible redox states (Cu0, Cu+, Cu2+, Cu3+), low cost, structural diversity, and excellent electrocatalytic capability in both alkaline and neutral media [22]. Several Cu-based systems including mesoporous CuO nanostructures were synthesized via an inverse micelle sol–gel process [23], coprecipitated CuO screen-printed electrodes for sweat analysis [24], and flexible PET electrodes functionalized with MWCNTs and Cu-HHTP MOFs [25] have shown favorable detection ranges and low limits of detection towards lactate. More advanced designs, such as self-supported copper foam structures with 3D open-pore architectures, have delivered ultrahigh sensitivities (800 μA−1 cm−2 and broad detection windows (1–120 mM) [26], while Cu-Au alloy catalysts have achieved detection limits as low as 5 μM [27].
The existing approaches have been directed towards lactate sensing in serum or sweat primarily, exploring rigid glassy carbon electrodes and drop-cast catalyst inks or composite powders, which suffer from binder dependence, nonuniform porosity, and batch-to-batch variability. Moreover, many non-enzymatic lactate sensors require strong alkaline electrolytes to activate Cu redox chemistry, limiting their applicability in real-world biofluids. Here we explored the use of laser-induced graphene (LIG) as a promising, cost-efficient, and flexible sensing platform, owing to its unique structural and electronic characteristics [28,29,30]. The fabrication process is convenient and straightforward, involving the localized photothermal conversion of polyimide to form a three-dimensional porous graphene network with abundant edge-plane defects, which enhances the electrochemical sensing capabilities [31]. LIG sensors possess excellent electrical conductivity, a large electroactive surface area, and defect-rich active sites that facilitate redox reactions and catalytic processes. However, their electrochemical performance is often limited by structural heterogeneity, high capacitive background currents, and the difficulty in accurately defining the electroactive surface area. Additionally, the defect-rich and porous nature of LIG can lead to surface instability, biofouling, and limited intrinsic selectivity in complex biological media, necessitating surface modification [32].
Saliva is an attractive non-invasive biofluid due to its ease of collection and suitability for frequent monitoring [33]; however, enzyme-free lactate detection in saliva under physiological conditions remains scarcely reported. Animashaun and coworkers reported a flexible (LIG-based non-enzymatic lactate sensor incorporating electrodeposited Au nanoparticles and a PEDOT-based molecularly imprinted polymer (MIP), which achieved excellent sensitivity, a wide dynamic range, and high molecular selectivity through synergistic conductivity enhancement and template-guided recognition [29]. While this approach demonstrates strong analytical performance in artificial saliva, it relies on multi-step fabrication and polymer imprinting strategies that add complexity and may pose challenges for large-scale manufacturability and long-term stability.
Therefore, we present a flexible, LIG sensor modified via a simple, one-step electroreduction of copper nanostructures for non-enzymatic lactate detection in artificial saliva under neutral pH. A systematic investigation of deposition parameters including applied potential and precursor concentration was conducted to optimize Cu nanostructure formation on the conductive graphene architecture. The synergistic integration of electrocatalytically active Cu nanoparticles with the high surface area and excellent charge transport characteristics of LIG enabled sensitive lactate oxidation across 100–1100 µM, achieving a sensitivity of 8.56 μA µM−1 cm−2 and a low detection limit of 42.65 µM. Scanning electron microscopy imaging revealed dense, uniformly distributed Cu nanoparticles anchored throughout the LIG framework, while elemental mapping confirmed homogeneous deposition. Beyond remarkable sensitivity, the sensor demonstrated strong selectivity against physiologically relevant interferents, highlighting its suitability for complex biological matrices. The ability of CuNPs/LIG to sustain strong electrocatalytic oxidation at pH 7.5 reflects superior material design, as the redox behavior of copper is typically suppressed outside alkaline environments. Thereby demonstrating a proof of concept for flexible lactate monitoring system using an engineered Cu-nanostructured architecture that enables scalable manufacturability, cost-effective fabrication, and stable operation in saliva under near-neutral pH conditions, facilitating direct lactate detection without pH adjustment. This work marks a significant advancement toward realistic, biocompatible, and field-deployable non-enzymatic lactate sensing.
2. Experimental
2.1. Chemicals and Materials
Polyethylene terephthalate sheets (PET) and polyimide (PI) sheets (thickness = 50 μm) were used for the synthesis of LIG. Potassium ferricyanide [K3Fe(CN)6] was used as a redox mediator for electrochemical characterization. The synthesis of copper nanostructures on LIG sensor surface involves the use of copper (II) bromide (CuBr2), sodium sulphate (Na2SO4) and sulphuric acid (H2SO4), which were obtained from Sigma Aldrich, St. Louis, MO, USA. Detailed interference studies were carried out using glucose (Sigma Aldrich, St. Louis, MO, USA), fructose (Acros Scientific, Waltham, MA, USA), uric acid (Sigma Aldrich, St. Louis, MO, USA), dopamine hydrochloride (Alfa Aesar, Haverhill, MA, USA), and Ascorbic Acid (Alfa Aesar, Haverhill, MA, USA). Phosphate-buffered saline (PBS) was prepared using an established PBS recipe from Sigma Aldrich, which involves sodium chloride (NaCl) (Sigma Aldrich, St. Louis, MO, USA), potassium chloride (KCl) (Sigma Aldrich, St. Louis, MO, USA), sodium dihydrogen phosphate (NaH2PO4, Sigma Aldrich, St. Louis, MO, USA), and dipotassium hydrogen phosphate (K2HPO4) (ThermoScientific, Waltham, MA, USA), and was used as a supporting electrolyte during all electrochemical measurements. Deionized water (DI) 18.2 MΩ cm−1 was used to prepare all solutions. The artificial saliva was prepared using KH2PO4 (Sigma Aldrich, St. Louis, MO, USA), K2HPO4 (ThermoScientific, Waltham, MA, USA), CaCl2 (Sigma-Aldrich, St. Louis, MO, USA), NaHCO3, and Citric acid from Fisher Scientific (Waltham, MA, USA).
2.2. Apparatus
The electrochemical performance of the copper-nanostructured LIG sensor was assessed using a three-electrode configuration on a Metrohm Dropsens µStat-i-MultiX system. In this setup, the CuNPs/LIG served as the working electrode (WE), a platinum (Pt) electrode as the counter electrode (CE), and an Ag/AgCl/3 M NaCl electrode was used as the reference electrode (RE) for all measurements. Morphological and elemental characterization of the electrodeposited CuNPs, and LIG was performed using field-emission scanning electron microscopy (FE-SEM, JSM-IT700HR InTouchScope™, JEOL Ltd., Tokyo, Japan), providing insights into nanoparticle distribution and surface composition.
2.3. Fabrication of Laser Induced Graphene Electrodes
A rectangular LIG working electrode (WE) was designed in CorelDraw 2019 and patterned onto polyimide (PI) tape (width: 1.25 cm). The PI tape was carefully laminated onto a flexible PET substrate. The LIG was produced via laser-assisted photothermal carbonization of laminated PI tape using a BOSS LS1416 CO2 pulsed laser system operated through LightBurn software (v. 0.909) [34]. The laser parameters including laser speed, maximum power settings, and focal height were systematically optimized to ensure high conductivity, mechanical stability, and reproducibility. The final conditions: 250 mm s−1 scan speed, 18% maximum power, 20 mm focal height, 0.1 s laser interval, a single raster pass, and a total irradiation duration of 9 s. This yielded a reproducible 4 mm rectangular WE pattern with a well-defined graphene-based sensor. Following carbonization, a thin layer of nail enamel was applied to precisely define the electroactive area and to insulate the contact pads.
2.4. Electrochemical Modification of Cu-Nanostructured LIG Electrodes
For electrode modification, CuNPs were electrochemically deposited onto the LIG surface using a chronoamperometric technique under constant potential. The electrodeposition was performed in an optimized electrolyte bath composed of CuBr2, sodium sulphate, and sulfuric acid. To determine the conditions that produced uniform and catalytically active Cu nanostructures, the growth of the CuNPs was optimized under variable deposition potentials and Cu precursor concentrations using a constant potential chronoamperometric technique. CV measurements were recorded from −0.9 to 0.9 V at a scan rate of 50 mV s−1 using a standard three-electrode setup.
3. Results and Discussion
3.1. Electrochemical Synthesis of the Cu Nanostructures on LIG Electrodes
The electrocatalytic performance of the CuNPs/LIG electrode toward lactate oxidation was systematically evaluated following controlled electrochemical deposition of Cu nanostructures onto the LIG framework. The electrodeposition parameters, specifically the applied electroreduction potential and Cu precursor concentration were optimized to tailor the surface morphology and catalytic activity of the resulting CuNPs/LIG interface (Figure 1). These parameters are known to strongly influence nucleation kinetics, crystal growth, and the density of catalytically active sites on carbon-based supports. To determine the ideal deposition potential, Cu nanostructures were electrodeposited using CuBr2 (10 mM) at varied reduction potentials (−1.1, −1.3, and −1.5 V) for 3 min in an electrolyte containing 10 mM sodium sulfate (Figure S1a, Supporting Information). The corresponding CVs displayed two well-defined anodic peaks at approximately −0.09 V and 0.17 V in the presence of 700 µM lactate, indicating the formation of catalytically active Cu nanostructures at the optimized deposition potential of −1.3 V. In contrast, at a lower deposition potential (−1.1 V), insufficient nucleation was observed, as evidenced by a weaker Cu oxidation peak current (Figure S1a, Supporting Information), leading to reduced electrocatalytic lactate oxidation. Following identification of the optimal deposition potential, the effect of precursor concentration was assessed by varying CuBr2 levels while maintaining deposition at −1.3 V (Figure S1b, Supporting Information). Cyclic voltammetry revealed that uniform and highly active Cu nanostructures formed most effectively at 10 mM CuBr2, as evidenced by the most prominent Cu oxidation peak and the strongest lactate response. At higher CuBr2 concentrations, the CVs displayed broadened and shifted oxidation peaks, indicating mass-transfer–limited growth and reduced electroactive surface area. This behavior is consistent with concentration-dependent mass transport limitations, which at elevated precursor levels may favor aggregation rather than controlled nanoparticle formation [30]. These electrochemical analyses confirm that Cu nanoparticles growth at −1.3 V using 10 mM CuBr2 for 3 min yields the most favorable balance of nucleation density, particle size, and catalytic surface exposure and therefore has been considered as optimal. We observed identical chronoamperograms corresponding to the uniform electrochemical modification of the LIG sensors with Cu NPs at the optimized electrodeposition conditions (Figure S2, Supporting Information).
Figure 1.
Schematic illustration detailing the steps involved in the fabrication of copper-nanostructured laser-induced graphene (CuNPs/LIG) for non-enzymatic lactate sensing.
3.2. Morphological and Elemental Characterization of CuNPs/LIG Electrodes
The structural characteristics responsible for the enhanced electrocatalytic performance of the CuNPs/LIG electrodes were investigated via detailed SEM and energy-dispersive X-ray spectroscopy (EDXS) analyses. The SEM micrograph of the pristine laser-induced graphene (Figure 2a) reveals a highly porous, interconnected 3D network formed because of laser-induced photothermal decomposition of polyimide. This process is known to cleave C–O, C–N, and N–C bonds, resulting in turbostratic graphene layers with expanded micro- and mesopores [31,34]. The abundance of large pores and edge-plane defect sites contributes to the high specific surface area of LIG, which enhances electrolyte accessibility and promotes rapid electron-transfer kinetics. Such structural features are essential for improving the conductivity and catalytic performance of LIG-based electrochemical sensors.
Figure 2.
Scanning electron microscopy images of (a) LIG; (b) CuNPs/LIG; (c) Corresponding EDAX analysis for the as-fabricated CuNPs/LIG.
Following electrochemical deposition, the morphology of the Cu-modified LIG surface changes markedly, as shown in Figure 2b. The Cu nanostructures appear as densely distributed granular clusters and chain-like assemblies anchored firmly onto the LIG ridges. This configuration suggests an efficient and uniform nucleation-growth process during electroreduction at the optimized potential. The chain-like agglomerates interconnect across the LIG microchannels, forming continuous conductive pathways that facilitate rapid charge transport during lactate oxidation. The intimate contact between Cu nanostructures and the conductive graphene backbone is expected to reduce interfacial resistance, enhance redox cycling, and accelerate reaction kinetics. The granular and clustered morphology is also advantageous because it exposes abundant catalytically active surface sites. Such nanoscale roughness has been widely associated with improved adsorption of lactate molecules and enhanced catalytic turnover on Cu-based sensors. In this context, the observed Cu morphology is consistent with the improved electrocatalytic response recorded under the optimized deposition conditions described earlier.
Elemental composition further supports successful Cu integration onto the LIG framework. EDXS spectra (Figure 2c) confirmed the presence of carbon, oxygen, and copper, with mass percentages of 55.16%, 6.24%, and 27.94%, respectively. The substantial copper content verifies the dense surface loading of Cu nanostructures, while the presence of oxygen is attributed to both residual oxygen functionalities on the LIG surface and native oxide species formed on the Cu nanostructures. Such mixed-valence copper-oxygen species (Cu0/Cu+/Cu2+) are known to contribute synergistically to catalytic activity for lactate oxidation, which further reinforces the suitability of the CuNPs/LIG interface for non-enzymatic sensing applications.
3.3. Scan Rate-Dependent Electrochemical Kinetics of Lactate Oxidation at CuNPs/LIG
The effect of scan rate on the electrochemical behavior of the Cu-nanostructured LIG electrode was investigated using CV within the potential window of −0.9 to 0.9 V in the presence of 500 µM lactate. As the scan rate increased from 20 to 100 mV s−1, both the anodic and cathodic peak currents exhibited a proportional increase, reflecting the dynamic response of the CuNPs/LIG interface under kinetically accelerated conditions. The anodic peak associated with lactate oxidation, centered at −0.05 V, exhibited a clear linear increase in oxidative current with increasing scan rate (Figure 3a), indicating the rapid electron-transfer capability of the Cu-modified electrode surface. This enhancement in current response can be attributed to the high electrochemically active surface area provided by the uniformly distributed Cu nanostructures embedded within the porous LIG framework. The hierarchical porosity and conductive graphitic domains of the LIG substrate facilitate efficient charge transport, while the Cu nanoparticles serve as catalytic sites, promoting rapid redox transitions during lactate oxidation. The resulting linear dependence between the anodic peak current and scan rate follows the regression:
indicating strong correlation and excellent electron-transfer performance under varying scan conditions.
Figure 3.
CVs of (a) CuNPs/LIG at variable scan rates ranging from 20 to 100 mV s−1 and (b) Corresponding linear relation between anodic peak current and scan rate v1/2.
Furthermore, the linear relationship between the peak current and the square root of the scan rate (Figure 3b) confirms that the electrooxidation of lactate at the CuNPs/LIG interface is governed by a diffusion-controlled mechanism (Figure S3, Supporting Information) [35]. This behavior is characteristic of systems in which mass transport of the analyte to the electrode surface is the rate-limiting step, rather than surface-limited adsorption processes. The corresponding peak potential shifts positively for anodic peaks and negatively for cathodic peaks, further implying quasi-reversible redox kinetics with increasing scan rate. These shifts are consistent with multistep copper redox transitions (Cu0/Cu+/Cu2+/Cu3+) that mediate the electrocatalytic oxidation of lactate, as previously reported for Cu-based catalytic systems [24]. Such behavior highlights the complexity of the electron-transfer processes involved and the suitability of the CuNPs/LIG electrode for sensitive lactate monitoring under dynamic electrochemical conditions.
3.4. Electrochemical Characterization and Lactate Sensing Performance of CuNPs/LIG Electrodes
The optimized CuNPs/LIG electrodes were first evaluated using the standard redox mediator potassium ferricyanide ([K3Fe(CN)6]) and subsequently examined for lactate oxidation prior to detailed sensing studies across variable lactate concentrations. The oxidation and reduction peaks for CuNPs/LIG electrodes are characteristic of a quasi-reversible to irreversible redox system with anodic and cathodic peak potentials of approximately 0.392 V and 0.001 V, respectively, and accompanied by significantly elevated peak currents (Ipa = 215.42 μA and Ipc = −252.82 μA) (Figure 4a). The large peak-to-peak separation, asymmetric peak shapes, and unequal anodic and cathodic peak currents confirm non-ideal reversibility. This behavior is expected for Cu-based surfaces, where multistep Cu/Cu+/Cu2+ transitions, oxide formation, and surface restructuring limit full regeneration of the initial redox state during the reverse scan [23]. This marked enhancement in redox activity compared to unmodified LIG indicates a substantial improvement in electron-transfer kinetics, arising from the intimate integration of copper nanostructures with the defect-rich, conductive LIG scaffold. The porous graphene network provides a high density of nucleation sites for Cu deposition, while the resulting nanoscale metallic clusters amplify the number of electroactive sites available for redox-mediated processes.
Figure 4.
Cyclic voltammograms (CV)of (a) CuNPs/LIG and LIG electrodes in 5 mM K3Fe(CN)6 at a scan rate of 50 mV/s; (b) CuNPs/LIG towards electrocatalytic oxidation of lactate; (c) CuNPs/LIG under variable concentrations of lactate; and (d) Corresponding calibration response between anodic peak current and lactate concentrations (n = 3).
To quantitatively assess this enhancement, the electroactive surface area of the CuNPs/LIG electrode was calculated using the Randles-Sevcik equation from Figure 3a [32]:
where Ip is the peak current, n = 1 for the one-electron process, C is the concentration of the redox probe, D is the diffusion coefficient, and ν is the scan rate [30]. A comparative analysis demonstrated that the electroactive surface area of CuNPs/LIG is approximately 2.75-fold higher than that of the unmodified LIG electrode. This result confirms that the electrodeposition process substantially increases the available catalytic surface by forming a densely interconnected copper nanostructured layer. The porous 3D structure of LIG enhances double-layer capacitance due to its large electroactive surface area and abundant edge-plane defects, resulting in elevated capacitive background currents that can partially overlap with Faradaic signals. Additionally, hierarchical porosity modifies mass transport, enhancing local analyte accumulation while potentially introducing diffusion limitations in thicker Cu-deposited regions. These effects together account for the broadened peaks and quasi-reversible behavior observed in the CVs [36]. The optimized CuNPs/LIG electrodes were subsequently assessed for lactate oxidation (Figure 4b). Two prominent oxidative peaks were observed at approximately −0.05 V and 0.209 V, consistent with characteristic copper redox transitions involved in lactate electrooxidation. These features are attributed to the cycling of Cu+/Cu2+ species at the electrode surface and reflect the active participation of copper nanostructures in mediating the lactate oxidation pathway. When tested across lactate concentrations ranging from 100 to 1100 µM, the CuNPs/LIG electrodes exhibited a progressive, concentration-dependent increase in both oxidative peaks (Figure 4c). This monotonic response confirms electrocatalytic activity and charge-transfer processes facilitated by the Cu-LIG hybrid architecture. The corresponding calibration plot (Figure 4d) demonstrates a strong linear relationship between the anodic peak current and lactate concentration, expressed by:
The slope of the calibration curve and the active surface area were used to calculate the sensor sensitivity (8.56 μA µM−1 cm−2). The limit of detection (LOD) was determined using 3.3* ϭ/S [37], where σ is the standard deviation of the blank (n = 3) and S is the calibration slope. The proposed CuNPs/LIG sensor demonstrates a detection limit of 42.65 µM and a linear range of 100–1100 µM, which fully encompass reported physiological salivary lactate concentrations, supporting its suitability for salivary lactate monitoring.
A comparative analysis with previously reported lactate sensors (Table 1) reveals that the CuNPs/LIG electrode offers competitive sensitivity and a favorable detection range, while providing the advantages of a flexible substrate, low-cost fabrication, and non-enzymatic sensing capability at near-neutral pH. Table 1 presents a comparative benchmarking overview of various non-enzymatic lactate sensors reported in the literature, highlighting their differences in electrode materials, electrochemical techniques, and sample matrices. Therefore, direct comparisons of detection limits and sensitivities is interpreted in the context of the sensing mechanism and experimental conditions.
Table 1.
Comparative performance analysis of the nanostructured electrodes for non-enzymatic lactate detection.
3.5. Selectivity, Reusability, and Operational Stability of the CuNPs/LIG Lactate Sensor
Evaluation of sensor selectivity is a critical requirement for determining its suitability for real-sample analysis, particularly in complex biological matrices where multiple electroactive species coexist. To assess interference effects, the electrochemical response of the CuNPs/LIG electrode toward lactate oxidation was evaluated in the presence of common endogenous interferents, such as glucose, fructose, ascorbic acid, dopamine, and uric acid.
The normal physiological levels of common salivary interferents such as ascorbic acid (40–80 µM), dopamine (0.12 nM), uric acid (100–400 µM), and glucose (30–200 µM), and fructose (1.7–1.9 µM) were considered prior to designing the selectivity study. However, to assess potential cross-reactivity, each interferent was introduced at significantly higher than their physiological concentrations (600 µM each), thereby imposing a more stringent testing condition than real saliva samples. As shown in Figure 5a, all interferents produced an increase in the oxidative peak current; however, the magnitude of this enhancement varied substantially depending on the interferent. Fructose produced only a modest 7.98% increase in current, indicating minimal interference. Glucose and ascorbic acid generated moderate signal increases of 22.29% and 12.5%, respectively, consistent with the partial overlap of their oxidation pathways with copper-mediated redox processes. In contrast, dopamine and uric acid induced more pronounced current elevations of 57.29% and 42.50%, respectively. The significant influence of these species is attributable to their strong electroactivity and their known propensity to undergo facile anodic oxidation on transition-metal-modified electrodes. Moreover, uric acid produced a distinct anodic peak at 0.482 V with a peak current of 244.70 μA, confirming the occurrence of overlapping oxidation processes coexisting with the lactate oxidation response. Additionally, the interference studies were performed using three independently fabricated CuNPs/LIG sensors, and the variability across replicates is represented by error bars in Figure 5b. Although the concentrations of the interfering species are significantly higher than typical salivary levels, the observed current increases of approximately 40–60% for dopamine and uric acid indicate that, if unmitigated, fluctuations in these electroactive species could lead to overestimation of lactate concentration under in vivo conditions. To address this, further improvements in anti-interference performance are required, including optimization of the operating potential window to favor lactate oxidation and the incorporation of thin protective or size-exclusion coatings compatible with flexible substrates. These strategies are expected to suppress interfering signals, enhance selectivity, and improve long-term stability for reliable wearable lactate monitoring [44].
Figure 5.
CVs of (a) CuNPs/LIG towards lactate (300 µM) and 2-fold co-interferents (ascorbic acid, dopamine, uric acid, fructose, and glucose) n = 3; (b) Corresponding bar graph illustrating the influence of interferents toward the electrocatalytic oxidation of lactate; (c) Stability of CuNPs/LIG sensor (300 µM lactate); (d) Operational stability of CuNPs/LIG under repetitive cycles (300 µM lactate; 20 cycles).
Sensor stability was further evaluated by performing repeated measurements of 300 µM lactate using the same CuNPs/LIG electrode. As shown in Figure 5c and Figure S4, Supporting Information, the first five consecutive measurements produced nearly identical oxidative peak currents, demonstrating minimal signal deviation during early reuse cycles. The decline in peak current observed after the fifth cycle is attributed to progressive surface oxidation and partial passivation of catalytically active copper sites during repeated redox cycling. In addition, adsorption of lactate oxidation intermediates and surface restructuring of the nanostructured electrode can further reduce the number of electrochemically accessible active sites. The observed electrochemical response is consistent with the existing literature [12]. Overall, the sensor retained stable response characteristics for up to five repeated assays, confirming its short-term reusability. Further, the operational stability was assessed subjecting the CuNPs/LIG electrode to 20 consecutive cyclic voltammetric cycles in 300 µM lactate (Figure 5d). The oxidative peak current retained 83.35% of its initial value, indicating strong electrochemical durability and robust adhesion of the electrodeposited Cu nanostructures to the LIG scaffold. This retention reflects a stable catalytic interface capable of sustaining redox cycling without substantial loss of activity.
3.6. Practical Evaluation of a CuNPs/LIG Lactate Sensor in Artificial Saliva
The practical applicability of the fabricated CuNPs/LIG sensor was evaluated in artificial saliva prepared following the composition reported by Animashaun et al. [44], with the pH carefully adjusted to 7.5 to closely mimic physiological conditions. For analytical evaluation, the artificial saliva was diluted 1000-fold and spiked with varying concentrations of lactate. Quantification was performed using the standard addition method, and percentage recoveries were calculated to evaluate analytical accuracy. As summarized in Table 2, the sensor demonstrated excellent recovery values ranging from 104.5% to 108%, indicating robust performance in biofluid matrices. Relative standard deviation (R.S.D.) values ranged from 17.5% to 20.83% were observed, which are higher than expected and is attributed to the presence of ionic species and residual organic components that can influence electrode wetting, and nonspecific adsorption, introducing variability in the measured currents.
Table 2.
Real sample analysis using CuNPs/LIG for lactate detection in artificial saliva.
4. Conclusions
Here, we present a facile and cost-effective approach for fabricating Cu nanostructured laser-induced graphene (CuNPs/LIG) electrodes for the detection of lactate. The electrode modification strategy is straightforward, reproducible, and demonstrates a significant electrochemical response toward lactate over a broad detection range, highlighting its high sensitivity and analytical reliability. Importantly, the practical utility of the CuNPs/LIG sensor was validated through lactate detection in artificial saliva, with recovery values ranging from 104% to 108%, demonstrating reliable performance in physiologically relevant matrices. The CuNPs/LIG sensor demonstrates a proof-of-concept study for non-invasive, real-time monitoring of lactate, with potential applications in clinical diagnostics, sports medicine, and personalized health monitoring. The simplicity and versatility of the proposed fabrication approach provide a proof-of-concept flexible platform for lactate detection under the tested conditions. Future development will focus on evaluation in human saliva samples, employing advanced selectivity strategies, and assessment of mechanical robustness under repeated deformation aimed at validating its performance for flexible and wearable applications.
Supplementary Materials
The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/bios16010019/s1, Figure S1: Cyclic voltammograms (CV) in the presence of lactate (700 µM) using CuNPs/LIG fabricated under (a) variable electroreduction potentials and (b) variable CuBr2 concentrations; Figure S2: Chronoamperograms of two identically prepared CuNPs/LIG using 10 mM CuBr2 for 3 min at a deposition potential of −1.3 V; Figure S3: Schematic representation of the mechanism associated with the electrocatalytic oxidation of lactate at CuNPs/LIG sensors; Figure S4: Cyclic voltammograms of CuNPs/LIG in the presence of lactate (300 µM).
Author Contributions
Conceptualization, methodology, validation, formal analysis, A.J. and G.S.; investigation, A.J.; resources, G.S.; data curation, writing—original draft, A.J.; writing—review and editing, G.S.; visualization, A.J.; supervision, project administration, funding acquisition, G.S. All authors have read and agreed to the published version of the manuscript.
Funding
This research received no external funding.
Institutional Review Board Statement
Not applicable.
Informed Consent Statement
Not applicable.
Data Availability Statement
Data is contained within the article/Supplementary Material.
Conflicts of Interest
The authors declare no conflicts of interest.
References
- Li, X.; Yang, Y.; Zhang, B.; Lin, X.; Fu, X.; An, Y.; Zou, Y.; Wang, J.X.; Wang, Z.; Yu, T. Lactate metabolism in human health and disease. Signal Transduct. Target. Ther. 2022, 7, 305. [Google Scholar] [CrossRef] [Scilit]
- Deulkar, P.; Singam, A.; Mudiganti, V.; Jain, A. Lactate Monitoring in Intensive Care: A Comprehensive Review of Its Utility and Interpretation. Cureus 2024, 16, e66356. [Google Scholar] [CrossRef] [Scilit]
- Yang, H.; Yan, S. A comprehensive review of electrochemical lactate biosensors: Principles, innovations, and future perspectives. Int. J. Electrochem. Sci. 2025, 20, 101132. [Google Scholar] [CrossRef] [Scilit]
- Biagi, S.; Ghimenti, S.; Onor, M.; Bramanti, E. Simultaneous determination of lactate and pyruvate in human sweat using reversed-phase high-performance liquid chromatography: A noninvasive approach. Biomed. Chromatogr. 2012, 26, 1408–1415. [Google Scholar] [CrossRef] [Scilit]
- Tong, W.; Shi, J.; Yu, Z.; Ran, B.; Chen, H.; Zhu, Y. High sensitivity and automatic chemiluminescence detection of glucose and lactate using a spin-disc paper-based device. Lab A Chip 2024, 24, 810–818. [Google Scholar] [CrossRef] [Scilit]
- Schmiedeknecht, K.; Kaufmann, A.; Bauer, S.; Solis, F.V. L-lactate as an indicator for cellular metabolic status: An easy and cost-effective colorimetric L-lactate assay. PLoS ONE 2022, 17, e0271818. [Google Scholar] [CrossRef] [Scilit]
- Miesse, P.; Slaughter, G. Flexible Electrochemical Lactate Biosensor. In Proceedings of the 2020 IEEE 15th International Conference on Nano/Micro Engineered and Molecular System (NEMS), San Diego, CA, USA, 27–30 September 2020; IEEE: Piscataway, NJ, USA, 2020; pp. 580–583. [Google Scholar]
- He, Q.; Wang, C.; Jain, R.; Byrnes, J.; Farquhar, E.R.; Reed, E.; Berezovsky, E.; Chance, M.R.; Lodowski, D.; An, R. An engineered lactate oxidase based electrochemical sensor for continuous detection of biomarker lactic acid in human sweat and serum. Heliyon 2024, 10, e34301. [Google Scholar] [CrossRef] [Scilit]
- Malon, R.S.P.; Chua, K.Y.; Wicaksono, D.H.B.; Córcoles, E.P. Cotton fabric-based electrochemical device for lactate measurement in saliva. Analyst 2014, 139, 3009–3016. [Google Scholar] [CrossRef] [Scilit]
- Madden, J.; Vaughan, E.; Thompson, M.; Riordan, A.O.; Galvin, P.; Iacopino, D.; Teixeira, S.R. Electrochemical sensor for enzymatic lactate detection based on laser-scribed graphitic carbon modified with platinum, chitosan and lactate oxidase. Talanta 2022, 246, 123492. [Google Scholar] [CrossRef] [Scilit]
- Wang, C.; Li, G. Recent Advancement in Non-Enzymatic Electrochemical Detection of Lactate Based on Metal Nanomaterials: A Review. Sensors 2025, 25, 6194. [Google Scholar] [CrossRef] [Scilit]
- Kim, S.; Yang, W.S.; Kim, H.-J.; Lee, H.-N.; Park, T.J.; Seo, S.-J.; Park, Y.M. Highly sensitive non-enzymatic lactate biosensor driven by porous nanostructured nickel oxide. Ceram. Int. 2019, 45, 23370–23376. [Google Scholar] [CrossRef] [Scilit]
- Nien, Y.-H.; Kang, Z.-X.; Chou, J.-C.; Yang, P.-H.; Kuo, P.-Y.; Lai, C.-H.; Dong, Z.-X.; Chen, Y.-Y.; Wu, Y.-T.; Lee, K.-T. Application of non-enzymatic lactate sensor modified by graphitic carbon nitride/iron–platinum nanoparticles and combined with the low power consumption instrumentation amplifier and calibration readout circuit. IEEE Trans. Instrum. Meas. 2023, 72, 1–8. [Google Scholar] [CrossRef] [Scilit]
- Ji, G.; Zhu, W.; Jia, X.; Ji, S.; Han, D.; Gao, Z.; Liu, H.; Wang, Y.; Han, T. AuNP/Cu-TCPP (Fe) metal–organic framework nanofilm: A paper-based electrochemical sensor for non-invasive detection of lactate in sweat. Nanoscale 2023, 15, 5023–5035. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Xiao, D.; Zou, M.; Huang, M. Wearable self-powered electrochemical sensor with MXene/PU composite and advanced electrodes for non-invasive lactate detection during physical activity. Mater. Sci. Eng. B 2025, 322, 118647. [Google Scholar] [CrossRef] [Scilit]
- Hussain, M.M.; Asiri, A.M.; Rahman, M.M. A non-enzymatic electrochemical approach for l-lactic acid sensor development based on CuO·MWCNT nanocomposites modified with a Nafion matrix. New J. Chem. 2020, 44, 9775–9787. [Google Scholar] [CrossRef] [Scilit]
- Liu, H.; Zare, N. A Novel Clay Based Electrocatalyst for Lactate Detection in Saliva and Blood Samples during Exercise in Athletes. Top. Catal. 2025, 68, 698–704. [Google Scholar] [CrossRef] [Scilit]
- Li, Y.; Wu, Y.; Miao, C.; Ma, Y.; Rong, S.; Guo, J.; Dai, Z. A novel nonenzymatic sensor of lactic acid based on nickel hydroxide nanostructure. Ionics 2025, 31, 6425–6434. [Google Scholar] [CrossRef] [Scilit]
- Gao, W.; Guan, X.; Wang, J.; Heinig, N.F.; Thomas, J.P.; Zhang, L.; Ding, K.; Leung, K.T. Supported NiOx Nanocatalysts on Graphene for Nonenzymatic Lactate Sensing. ACS Appl. Nano Mater. 2024, 7, 7162–7171. [Google Scholar] [CrossRef] [Scilit]
- Kim, S.; Kim, K.; Kim, H.-J.; Lee, H.-N.; Park, T.J.; Park, Y.M. Non-enzymatic electrochemical lactate sensing by NiO and Ni(OH)2 electrodes: A mechanistic investigation. Electrochim. Acta 2018, 276, 240–246. [Google Scholar] [CrossRef] [Scilit]
- Wu, Y.-T.; Tsao, P.-K.; Chen, K.-J.; Lin, Y.-C.; Aulia, S.; Chang, L.-Y.; Ho, K.-C.; Chang, C.Y.; Mizuguchi, H.; Yeh, M.-H. Designing bimetallic Ni-based layered double hydroxides for enzyme-free electrochemical lactate biosensors. Sens. Actuators B Chem. 2021, 346, 130505. [Google Scholar] [CrossRef] [Scilit]
- Li, C.; He, J.; Xiao, Y.; Li, Y.; Delaunay, J.-J. Earth-abundant Cu-based metal oxide photocathodes for photoelectrochemical water splitting. Energy Environ. Sci. 2020, 13, 3269–3306. [Google Scholar] [CrossRef] [Scilit]
- Heo, S.G.; Yang, W.-S.; Kim, S.; Park, Y.M.; Park, K.-T.; Oh, S.J.; Seo, S.-J. Synthesis, characterization and non-enzymatic lactate sensing performance investigation of mesoporous copper oxide (CuO) using inverse micelle method. Appl. Surf. Sci. 2021, 555, 149638. [Google Scholar] [CrossRef] [Scilit]
- Sajna, M.S.; Cabibihan, J.-J.; Malik, R.A.; Sadasivuni, K.K.; Geetha, M.; Alahmad, J.K.; Hijazi, D.A.; Alsaedi, F. Nonenzymatic electrochemical lactic acid sensor using CuO nanocomposite. Mater. Sci. Eng. B 2023, 288, 116217. [Google Scholar] [CrossRef] [Scilit]
- Pei, S.; Ji, W.; Yang, Y.; Yang, S.; Liu, T.; Dai, J.; Fu, C.; Wu, J.; Yin, Y.; Wu, Q.; et al. An enzyme-free wearable sweat sensor based on a Cu metal-organic frame composite material for monitoring lactate metabolism. Sens. Actuators B Chem. 2025, 433, 137512. [Google Scholar] [CrossRef] [Scilit]
- Gao, W.; Thomas, J.P.; Leung, K.T. Three-Dimensional Copper Foam Nanocatalysts for Nonenzymatic Electrochemical Lactate Sensing. ACS Appl. Nano Mater. 2025, 8, 18479–18486. [Google Scholar] [CrossRef] [Scilit]
- Lu, C.; Lu, Y.; Xu, M.; Zhang, Z.; Han, W.; Ghanei, M. Fabrication of Nanostructured Cu-Au Materials as an Efficient Electrocatalyst for Lactate Determination in Athletes Biological Fluid During Exercise. Top. Catal. 2025, 68, 636–643. [Google Scholar] [CrossRef] [Scilit]
- Avinash, K.; Patolsky, F. Laser-induced graphene structures: From synthesis and applications to future prospects. Mater. Today 2023, 70, 104–136. [Google Scholar] [CrossRef] [Scilit]
- Animashaun, C.; Lahcen, A.A.; Slaughter, G. Gold nanoparticle-enhanced molecularly imprinted polymer electrode for non-enzymatic lactate sensing. Biosensors 2025, 15, 384. [Google Scholar] [CrossRef] [Scilit]
- Joshi, A.; Slaughter, G. Cost-Effective Hierarchical Cobalt Nanostructured Laser-Induced Graphene for Enhanced Uric Acid Detection. Adv. Sens. Res. 2025, 4, e70003. [Google Scholar] [CrossRef] [Scilit]
- Torati, S.R.; Hanson, B.; Shinde, M.; Slaughter, G. Gold-deposited laser-induced graphene electrode for detection of miRNA-141. IEEE Sens. J. 2023, 24, 2154–2161. [Google Scholar] [CrossRef] [Scilit]
- Aftab, S.; Koyyada, G.; Mukhtar, M.; Kabir, F.; Nazir, G.; Memon, S.A.; Aslam, M.; Assiri, M.A.; Kim, J.H. Laser-induced graphene for advanced sensing: Comprehensive review of applications. ACS Sens. 2024, 9, 4536–4554. [Google Scholar] [CrossRef] [Scilit]
- Ghimenti, S.; Lomonaco, T.; Bellagambi, F.G.; Biagini, D.; Salvo, P.; Trivella, M.G.; Scali, M.C.; Barletta, V.; Marzilli, M.; Di Francesco, F. Salivary lactate and 8-isoprostaglandin F2α as potential non-invasive biomarkers for monitoring heart failure: A pilot study. Sci. Rep. 2020, 10, 7441. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Liu, X.; Cheng, H.; Zhao, Y.; Wang, Y.; Li, F. Portable electrochemical biosensor based on laser-induced graphene and MnO2 switch-bridged DNA signal amplification for sensitive detection of pesticide. Biosens. Bioelectron. 2022, 199, 113906. [Google Scholar] [CrossRef] [Scilit]
- Ahmad, R.; Tripathy, N.; Ahn, M.S.; Bhat, K.S.; Mahmoudi, T.; Wang, Y.; Yoo, J.Y.; Kwon, D.W.; Yang, H.Y.; Hahn, Y.B. Highly efficient non-enzymatic glucose sensor based on CuO-modified vertically grown ZnO nanorods on electrodes. Sci. Rep. 2017, 7, 5715. [Google Scholar] [CrossRef] [Scilit]
- Lin, J.; Peng, Z.; Liu, Y.; Ruiz-Zepeda, F.; Ye, R.; Samuel, E.L.; Yacaman, M.J.; Yakobson, B.I.; Tour, J.M. Laser-induced porous graphene films from commercial polymers. Nat. Commun. 2014, 5, 5714. [Google Scholar] [CrossRef] [Scilit]
- Weber, S.G.; Long, J.T. Detection limits and selectivity in electrochemical detectors. Anal. Chem. 1988, 60, 903A–913A. [Google Scholar] [CrossRef]
- Kishnani, V.; Makadia, R.A.; Natarajan, S.; Joseph, J.; Gupta, A. Development of a silver–polyaniline functionalized biosensor for non-enzymatic lactic acid detection. Mater. Adv. 2025, 6, 766–776. [Google Scholar] [CrossRef] [Scilit]
- Xiao, H.; Cao, L.; Qin, H.; Wei, S.; Gu, M.; Zhao, F.; Chen, Z. Non-enzymatic lactic acid sensor based on AuPtNPs functionalized MoS2 nanosheet as electrode modified materials. J. Electroanal. Chem. 2021, 903, 115806. [Google Scholar] [CrossRef] [Scilit]
- Faisal, M.; Alam, M.; Ahmed, J.; Asiri, A.M.; Alsareii, S.; Alruwais, R.S.; Alqahtani, N.F.; Rahman, M.M.; Harraz, F.A. Efficient electrochemical detection of L-lactic acid using platinum nanoparticle decorated Chitosan/ZnTiO3 nanocomposites. J. Ind. Eng. Chem. 2023, 118, 362–371. [Google Scholar] [CrossRef] [Scilit]
- Wali, R.; Moulaee, K.; Qasymeh, M.; Maalej, R.; Neri, G. Atomic layer deposition of NiO-coated CNT nanocomposites: Tailoring electrochemical properties for salivary lactate detection. J. Electroanal. Chem. 2024, 971, 118594. [Google Scholar] [CrossRef] [Scilit]
- Shao, W.; Mai, J.; Wei, Z. Nonenzymatic lactic acid detection using cobalt polyphthalocyanine/carboxylated multiwalled carbon nanotube nanocomposites modified sensor. Chemosensors 2022, 10, 83. [Google Scholar] [CrossRef] [Scilit]
- Huang, Y.S.; Chen, K.Y.; Cheng, Y.T.; Lee, C.K.; Tsai, H.E. An inkjet-printed flexible non-enzymatic lactate sensor for clinical blood plasma test. IEEE Electron Device Lett. 2020, 41, 597–600. [Google Scholar] [CrossRef] [Scilit]
- Animashaun, C.; Lahcen, A.A.; Slaughter, G. Development of flexible regenerable lactate-specific molecularly imprinted polymers based on laser-induced graphene electrochemical biosensor. Bioelectrochemistry 2025, 166, 109017. [Google Scholar] [CrossRef] [Scilit] [PubMed]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2025 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.