Experimental Characterization and Numerical Optimization of 3D-Printed PA6-CF External Fixator Rings

,
,
and
1
Department of Robots and Production Systems, National University of Science and Technology POLITEHNICA Bucharest, 060042 Bucharest, Romania
2
Department of Materials Strength, National University of Science and Technology POLITEHNICA Bucharest, 060042 Bucharest, Romania
*
Authors to whom correspondence should be addressed.

Abstract

This research investigated the feasibility of 3D-printed external fixator (EF) rings made from carbon fiber reinforced polyamide 6 (PA6-CF) as an alternative to the conventional metallic counterpart. The study integrated tensile testing with digital image correlation (DIC) in as-printed and cold plasma-sterilized conditions, finite-element analysis (FEA) under wire loading, topology optimization for material and energy reduction, and evaluation of printability limits for large PA6-CF rings. The average Young’s modulus was 4.76 GPa and the maximum tensile strength was 60.5 MPa for as-printed samples, decreasing by 6.4% and 10.4% after sterilization, respectively. Using these properties as model inputs, FEA predicted safety factors larger than 1.42 for all configurations under 1000 N wire pretension, while topology optimization targeted up to 50% mass reduction without compromising ring stiffness. The study also revealed challenges in the printability of PA6-CF for large and thin components, including dimensional contraction, significant warping and moisture-induced defects, requiring an experienced 3D printer operator.

1. Introduction

External fixators (EFs) are commonly used in the management of severe tibial and fibular fractures [1], as they provide mechanical stability while allowing soft tissue preservation and access to the injury site [2]. Depending on the patient’s condition, the fracture type and its location, the orthopedic surgeon can choose between linear, hybrid or circular EFs [3]. Among these categories, several configurations are commercially available, differing in their fixation method, components’ type, spatial arrangement of connecting rods (e.g., unilateral, bilateral or multilevel), number and orientation of fixation pins or screws (e.g., parallel, convergent, or divergent) and presence and type of circular elements such as full rings, 5/8 rings or arches [1,4,5]. These design aspects have a direct influence not only on the overall EF–bone construct stiffness and the ability to control micromotions at the fracture site (a critical factor in bone healing) but also on the surgeon’s capability to ensure fracture reduction while minimizing soft tissue damage.
EF rings are usually manufactured from metallic materials, having a modular design, but they are costly [6]. The availability of these traditional EFs can be challenging in settings where resources or logistics are limited, such as remote or conflict war zones or low-income countries [7]. In this context, polymer-based MEX 3D printing technology is offering the opportunity to locally produce EF components such as rings [8,9] in settings where basic 3D-printing infrastructure and staff are available, and to customize them as lightweight and personalized for specific anatomical needs, thus reducing the dependence on centralized manufacturing and supply chains [10]. However, one of the main limitations of 3D-printed polymer components is their reduced stiffness compared to metallic structures, which represents a problem for load-bearing orthopedic elements. Moreover, manufacturing thin and large circular rings from filaments such as polyamide (PA)-based composites, by material extrusion process (MEX) is technically demanding due to thermal contraction and moisture sensitivity, which may negatively impact the build quality and dimensional stability.
In a prior study [10], it was shown that PLA-CF (polylactic acid reinforced with carbon fibers) has the potential to be used for producing rings for hybrid fixators via the MEX, but only if the ring’s design is optimized to increase its stiffness. By applying topological optimization and adjusting parameters such as ring shape, cross-sectional area and structural reinforcements, the overall stiffness of the EF was improved even for this relatively low-stiffness material. Based on these initial promising results, the present study proposed to extend the investigation over PA6-CF, which is a more mechanically robust material compared with PLA-CF [11]. PA6-CF has higher tensile strength, toughness, fatigue resistance and ductility [12], making it more suitable for load-bearing components subjected to cyclic loading and sustained wire pretension. Its higher ductility also reduces the risk of brittle fracture under over-tightening.
While literature showed that there are works investigating 3D-printed components for EF, only a limited number of studies address full-scale lower-limb external fixator rings produced by MEX, and these focused on a relatively narrow set of materials compared with the wide range of polymers and composite filaments that are currently available. Alqahtani et al. [13] developed a topology-optimized, patient-specific circular EF consisting of a 4 mm thick porous shell, which anatomically fit around the leg. Qiao et al. [14] also designed a patient-specific, 3D-printed EF, optimized based on simulation data from the anatomical contouring and positional registration of external fixation pins, obtaining a personalized, wrap-around shell-type fixator designed around the patient’s limb. This customized fixator enabled automated fracture reduction with successful healing and no reported complications, but no mechanical evaluation was presented using FEA or experiments, nor was the effect of the fixator’s sterilization considered. Other studies in the field were focused on clinical implementation of 3D-printed EF without topological optimization [15,16]. Although these studies demonstrated the feasibility of 3D printing for EF components and patient-specific customization, they remained limited in mechanical validation, material selection and sterilization effects. Most 3D-printed EFs have been produced from low-stiffness polymers such as PLA or ABS, which are easy to process but provide insufficient strength for lightweight functional structures. Short CF-reinforced nylon (Onyx) has been used successfully for clamps [6], but limited evidence exists on the development and validation of thin, topology-optimized rings manufactured from high-performance composites such as PA6-CF. Therefore, the main limitations identified in prior studies lie in the lack of use of higher-strength composite materials for EF rings, the lack of approaches that combine experimental material characterization, topology optimization and manufacturability evaluation into a unified workflow, and the absence of mechanical testing under sterilized conditions to provide realistic input data for simulation and performance verification.
To address these research gaps, the present study investigates PA6-CF topology-optimized EF rings manufactured via MEX through a combined experimental–numerical workflow. The tensile properties before and after cold plasma sterilization were quantified using DIC and used as direct input for finite element analysis under wire pretension. This represents the first evaluation of the sterilization effect on the mechanical performance of topology-optimized EF rings made from a high-performance composite. In addition, this study provides the first feasibility assessment of printing full-scale, thin PA6-CF rings, demonstrating the contrast between the material’s promising strength and its demanding printability for large, load-bearing components.

2. Materials and Methods

2.1. Research Methodology

Figure 1 presents the flowchart of the research methodology in which are emphasized the main processes, sub-processes, outputs and decision blocks. Its purpose is to provide the reader a clear overview of the main constitutive layers and the information transfers that take place.
Figure 1. The proposed research methodology.
The start of the workflow consists of tensile testing (process block) for evaluating the mechanical properties of the material on a batch of PA6-CF specimens, both as-printed and sterilized, to reproduce clinically relevant conditions. DIC is used for gathering full-field strain data (sub-process block). The resulting experimental material properties, represented as a green output block, form the input for the FE-based simulation layer, since both static analysis and topology optimization rely on accurate constitutive characterization. All these steps are included in the first layer in Figure 1, namely the experimental layer.
The FE-based simulation layer starts in parallel (Figure 1) from the initial 3D CAD models (process block) of two EF-ring configurations. The design spaces are defined according to the build plate dimensions of the 3D printer. Static analysis (process block) is completed for evaluating the stress and displacement distribution based on 2D shell elements. Topology optimization is further employed with the objective of minimizing the mass of the EF-rings by up to 50% by iterative material removal (process block). The decision block “Target reached?” rules the cycle. If the target is not achieved, the optimization continues, while a positive outcome leads to reverse engineering (sub-process block) to smooth irregularities and obtain an improved 3D CAD model (output block) suitable for manufacturing.
Finally, in the 3D printing layer (Figure 1), process parameters are set to accommodate the moisture-sensitive nature of the PA6-CF (sub-process block). These settings were used for 3D printing both the tensile specimens as well as the optimized EF rings (process block). The parts are printed, and the decision block “3DP Issues?” determines whether reprinting is required. Once acceptable dimensional accuracy and surface quality are reached, the workflow yields the physical prototype (output block) used to assess printability.

2.2. Theoretical Considerations

In ANSYS Workbench 19.0 (ANSYS Inc., Canonsburg, PA, USA), the solution procedure for the external fixator ring follows the standard finite element method (FEM). The global system of equations is assembled as in [17],
K E , ν u = F
where K is the global stiffness matrix, E is the modulus of elasticity of the material, υ is the Poisson’s ratio, u is the vector of nodal displacements—the primary unknowns, and F is the vector of the applied nodal loads.
The solver first calculates u. Once displacements are known, the strain field is obtained through the strain–displacement matrix B,
ε = B u
Stress results are than computed by employing Hooke’s law,
σ = D ε
In case of isotropic linear elastic shells,
D = E 1 + ν 1 − 2 ν 1 ν 0 ν 1 0 0 0 1 − ν 2
Compliance is the objective function in topology optimization [18],
C = F T u
Given the fact that u is the first result of static analysis, the same displacement solution becomes the foundation for topology optimization. The optimization process simply redistributes material via density variables, while ensuring the displacement-based compliance remains within acceptable bounds. It is therefore imperative to accurately capture material stiffness values (E, ν) as any error propagates from displacement to strain to stress.

2.3. FE-Based Simulation Layer

  • Initial 3D CAD model
The EF’s ring was modeled as a mid-surface representation to enable efficient FEA. Two design spaces were defined according to the build plate dimensions of the 3D printer, namely the K1C (220 mm × 220 mm) of Creality and an alternative X1C with a larger printing bed (256 mm × 256 mm). To ensure adequate adhesion, a 10 mm margin was excluded from the usable area of the printer.
The baseline ring geometry had a thickness of 10 mm and an internal diameter of 140 mm, with external diameters of 210 mm for the K1C case, and 246 mm for the X1C case. Thirty-six uniformly distributed holes of 6 mm diameter were positioned every 10° to accommodate the fixation bolts for the fixator Kirschner wires (K-wires).
  • Simplified FEM Simulation Model
The FE mesh was generated with shell elements, using a global element size of 1 mm and local refinement to 0.8 mm around the holes to capture stress concentrations. The rings were assigned isotropic elastic properties representative of PA6-CF, with Young’s modulus and Poisson’s ratio derived experimentally from tensile testing with DIC. The K-wires were modeled as stainless steel 316LVM.
Static structural analysis considered the dominant load cases produced by K-wire pretension. Three wires are considered in both cases, each tensioned at 1000 N as reported in the literature (Figure 2) [19].
Figure 2. Details regarding the applied boundary conditions for the studied load case: (a) K1C Load case; (b) X1C Load case.
  • Topology optimization
Topology optimization was performed in ANSYS Workbench 19.0 (ANSYS Inc., Canonsburg, PA, USA) using the density-based SIMP algorithm with compliance minimization as the objective function. The resulting coarse geometries were subsequently processed and refined in CATIA V5 (Dassault Systèmes, FR, Vélizy-Villacoublay, France) to generate the final designs.
The workflow involves coupling between the static structural analysis and the topology optimization module. This allows the algorithm to evaluate the structural response of the ring under the defined loads and boundary conditions. Thus, the material layout can be iteratively adjusted in the design space until the optimization target is met.
In the next stage, the design and exclusion regions are defined (Figure 3). In both cases, a region extending 1.5 times the hole diameter was defined in the proximity of the functional holes. This constraint is essential to ensure manufacturability while preserving adequate bearing capacity under the applied shear loading.
Figure 3. Details regarding the optimization separation regions: (a) K1C design and exclusion region; (b) X1C design and exclusion region.
In accordance with the theoretical considerations, compliance minimization was decided as the optimization objective to promote material retention along the dominant load paths induced by wire pretension, while enabling the removal of material from regions that contribute minimally to global stiffness.
The response constraint is set to retain 50% of the global mass, a critical indicator of the printing time and filament usage.
Following convergence of the optimization process, the resulting output consists of coarsely optimized geometries, which are exported as STL files for subsequent processing in CATIA V5. In this step, smoothing techniques were applied to generate more symmetric arch-bridge features. Lateral connecting holes were also introduced in the CAD reconstruction to allow the assembly of multiple ring stages if required. The refined geometries were then prepared for finite element modeling and subsequent additive manufacturing.
Static linear analyses were set up in ANSYS Mechanical for both K1C and X1C configurations. The rings were modeled in a ring-only configuration and subjected to three-wire pretension loads of 1000 N. Boundary conditions followed the procedure described in [10] with a reference node clamped at the location corresponding to minimal displacement in the EF–bone construct model. For consistency, both optimized and baseline geometries were analyzed under identical loading and boundary conditions.

2.4. Experimental Layer

Tensile tests on PA6-CF specimens were conducted in both as-printed and sterilized conditions. A Dantec Dynamics DIC system (Dantec Dynamics Ltd., Skovlunde, Denmark) was used to determine the mechanical parameters required for the FEM simulation model. For each condition (as-printed and sterilized), three specimens were tested, and their averaged mechanical properties with associated variability are reported in the Section 3. One face of each specimen was coated with a white base and overlaid with a black speckle pattern to enable DIC tracking.
The tensile testing was conducted on Instron 8872 (Instron Corp., Norwood, MA, USA) at a constant rate of 1 mm/min. Mechanical loading was applied, and strain measurements were captured with DIC (Figure 4).
Figure 4. Details regarding the experimental setup for DIC.
The strain field was analyzed over a selected calibrated region, allowing the extraction of key mechanical properties, including the stress–strain response, Young’s modulus, Poisson’s ratio, and ultimate tensile strength.

2.5. 3D Printing Layer

The following process parameters (Table 1) were used for manufacturing the PA6-CF Type I specimens for tensile testing (Figure 5), in accordance with ASTM D638 [20]. Even with an infill density of 100%, the infill pattern is relevant. For instance, the gyroid pattern was selected for all specimens and the ring, as it provides a more uniform, nearly isotropic internal structure and consistent thermal behavior during printing [21].
Table 1. Printing parameters for PA6-CF samples and rings.
Figure 5. Tensile specimen dimensions.
PA6-CF20 filament from Polymaker [22] was used to manufacture all the specimens and parts in this study. As mentioned, the printing was performed on a Creality K1C 3D printer, using a 0.6 mm hardened steel (SKD11) nozzle with a copper alloy coating, that ensured uniform heat transfer and wear resistance during extrusion.
As mentioned in the introduction, PA6-CF20 was selected because, according to the manufacturer’s technical data sheet, PA6-CF filament has a tensile strength of approximately 75 MPa, a tensile modulus of 6.5 GPa, and an elongation at break of 2.1% [12]. For comparison, the PLA-CF filament employed in our previous study [10] presents a tensile strength of around 48 MPa, a modulus of 4.95 GPa, and an elongation of 1.5%. Both materials contain short carbon fibers (20%) dispersed within the polymer matrix, but PA6-CF provides higher ductility and improved fatigue resistance, making it more suitable for load-bearing components.
To minimize the effects of moisture absorption during processing, the PA6-CF filament was dehumidified for 10 h, with the filament spool placed inside the dehumidifier, from where the filament was directly fed into the extruder.
PA6-CF specimens were sterilized with the Sterrad system using vaporized hydrogen peroxide plasma (0.23–0.8 kPa, 45–50 °C, 28 min). No preliminary cleaning with medical-grade cleaning agents was performed at this stage of the research. While this represents a limitation with respect to clinical translation, future studies will integrate standardized cleaning protocols to more closely replicate real hospital protocols.
For consistency, the same 3D printer, environment and process parameters were used for the 3D-printed optimized ring.

3. Results and Discussion

3.1. Mechanical Test Results on PA6-CF Specimens

Figure 6 shows the stress–strain curves obtained from the tensile tests, displaying the PA6-CF mechanical behavior under tensile load, for both as printed (Dia) and sterilized (DiaSt) specimens.
Figure 6. Stress–strain test results.
Figure 7 shows frames from the mechanical test of representative as-printed and sterilized specimens. Figure 7a,b represent the principal strains 1 and 2 for the specimen as-printed at the position on the curve corresponding to the maximum strength reached. Figure 7c,d show the principal strains 1 and 2 for the specimen sterilized at the position on the curve corresponding to the maximum strength reached.
Figure 7. Frames from the mechanical test of representative: (a) as-printed specimen—principal strain 1; (b) as-printed specimen—principal strain 2; (c) sterilized specimen—principal strain 1; (d) sterilized specimen—principal strain 2.
For both as-printed and sterilized specimens, one can observe that the strain repartitions are uniform along the specimen length without significant strain concentration areas; thus, test data can further be successfully used for FEA.
As summarized in Table 2, the average Young’s modulus decreased from 4.76 ± 0.19 GPa to 4.45 ± 0.17 GPa (−6.4%), while the tensile strength decreased from 60.5 ± 2.9 MPa to 54.3 ± 3.9 MPa (−10.2%) after sterilization. Poisson’s ratio remained nearly constant at around 0.24. Thus, cold plasma sterilization of the PA6-CF specimens has decreased the stiffness of the part by 6.4%, while the material strength was reduced by 10.2%. The low COV values (<5% for E and ν) confirm good reproducibility of the mechanical tests, while the slightly higher COV in tensile strength after sterilization (7.3%) reflects minor local inhomogeneities typical of MEX-printed composites.
Table 2. Tensile test results for as-printed and sterilized specimens.
The decrease in mechanical performance after sterilization was in line with literature evidence, as cold plasma sterilization is a non-thermal, low-pressure process; therefore, the mechanical decrease cannot be attributed to thermal or oxidative degradation, but rather to moisture adsorption, which is a problem for PA6 material [23]. For instance, Bonjo et al. [24] reported that 3D-printed PA and PA-CF composites exhibited up to a 60% decrease in flexural modulus after water immersion, with CF only partially mitigating the negative effect of absorption. Also, Gong et al. [25] noted that even low moisture contents (0.1–0.2 wt %) in PA filaments can determine microvoids and pseudo-crosslinking during MEX and degrade interlayer adhesion and mechanical uniformity. Popescu et al. [26] confirmed dimensional changes caused by the crystallinity and hygroscopic nature of the material for surgical guides subjected to disinfection and sterilization, suggesting the necessity to perform surface treatments as a mitigation means.
The experimental values were lower than the material characteristics provided by the filament manufacturer, Polymaker [22]. This is mainly due to the fact that no thermal treatment of the specimens was performed, as specified in the producer’s data sheets.
Here it should be noted that the problem of sterilization was not addressed in any of the other studies focused on 3D printing EF, regardless of material.

3.2. FEM Simulation Results

  • Simplified FEM simulation model results
The results of interest from the static FEA were the total deformation and the equivalent (von Mises) stress distributions. These two outputs were used to assess both the structural integrity and the stiffness of the baseline EF rings. Figure 8 presents the von Mises stress distribution.
Figure 8. The equivalent von Mises stress distribution for the studied cases.
The maximum stress was observed in the K1C configuration, reaching a magnitude of 35.5 MPa. Considering the allowable stress of 60.5 MPa for non-sterilized PA6-CF, the corresponding safety factor was calculated as 1.70.
Figure 9 shows the displacement vector sum. The largest deflection, with a magnitude of 1.6 mm, was recorded for the K1C configuration. Moreover, the resultant reaction force was extracted from the analysis, producing a magnitude of 1000 N. From this, the global stiffness of the rings was derived as the ratio between the applied load and the measured displacement, giving a stiffness of 625 N/mm.
Figure 9. Displacement vector sum of the baseline K1C and X1C configurations.
Compliance was further evaluated as the indicator of the global flexibility of the structure. Unlike the maximum displacement (which reflects local behavior), compliance accounts for the overall deformation response under loading and is therefore useful for comparing baseline and optimized geometries. In this study, compliance was calculated from the strain energy as C = 2U.
Table 3 is a summary of the results achieved for both studied cases.
Table 3. Baseline FEA results of the non-optimized PA6-CF EF ring.
  • Topology optimization results
Figure 10 presents the raw topology-optimization outputs for K1C and X1C and the corresponding manufacturable ring designs, with the associated mass values (before/after reconstruction) and the optimization metrics (compliance/strain energy criterion).
Figure 10. Raw topology-optimization results and corresponding manufacturable ring designs for K1C and X1C.
Figure 11 shows the dimensioned drawings of the baseline reference ring from previous work [10] and of the topology-optimized K1C ring selected for fabrication in PA6-CF.
Figure 11. Drawings of the baseline (non-optimized) reference ring (a) and the topology-optimized K1C ring (b), shown with identical overall dimensions for direct comparison of mechanical performance.
Following optimization, the coarse density-based geometries were processed in CATIA V5 (Dassault Systemes, Fr) to obtain manufacturable designs. In this step, smoothing techniques were applied to eliminate irregular edges and discontinuities generated by the optimization algorithm. Symmetry constraints were introduced, and arch-like features were refined into continuous curves, resulting in smoother and more uniform ring geometries. Additional lateral holes were integrated into the reconstructed models to allow the assembly of multiple ring stages if required (Figure 11).
The optimized geometries were evaluated numerically under the same FEM conditions and parameters as those used for the baseline configuration of [10]. The results of the von Mises stress are shown in Figure 12.
Figure 12. The equivalent von Mises stress distribution for the optimized cases.
In comparison to the baseline design, the optimized geometries showed slightly higher stress magnitudes, with maximum values of 42.6 MPa for the K1C case, and 36.8 MPa for the X1C case. Despite this increase, the stress levels remain well within the allowable limit for PA6-CF, indicating that no material failure is expected under the investigated loading conditions.
The stress fields are smoothly distributed across the ring cross-sections, and no sharp gradients or localized peaks are observed that would indicate critical stress concentration zones. On the other hand, the regions of maximum stress remain uniformly spread rather than concentrated in isolated elements.
Figure 13 presents the displacement vector sum of the optimized configurations. Displacements of the ring are quite significant, up to 3.88 mm and are slightly higher for the K1C design compared to X1C. This means that the correct alignment of the fractured bones needs to be checked also after the application of the wire prestress, as it may be impacted by the ring displacement.
Figure 13. Displacement vector sum of the optimized K1C and X1C configurations.
Table 4 shows the impact of material change on a non-optimized full ring. The PLA-CF material properties from the filament technical datasheet used in the previous study were replaced with experimentally determined PA6-CF properties. The safety factor increases by 24%, mainly due to the higher allowable stress/strength of PA6-CF, while stiffness and stress values remain relatively similar, with the PLA-CF ring being slightly stiffer.
Table 4. Summary of the results achieved for the material change on a non-optimized design.
A comparison between the initial non-optimized configuration of [10] and optimized configurations is depicted in Table 5. Ring-level experimental validation is planned using a dedicated pretension test fixture.
Table 5. Summary of the results of studied cases for K1C and X1C design compared to initial design.
As observed, the safety factors are significantly increased by the use of optimized designs. For K1C, as the outer ring size and shape are similar, the stiffness is very similar. However, the stress is reduced by 25% by the redistribution of material in the radial net section achieved by only 13% more filament material consumed in a similar printing time. For the X1C design, as the ring is bulkier, the stiffness is increased by about 20%, and the maximum stress is reduced by 35%. This is achieved with 66% more material consumption. Also, the printing time is increased from 12 h to 19 h.
The purpose of the topology optimization analysis conducted in this study was to define a feasible full-scale PA6-CF ring configuration, rather than to investigate multiple alternative design variants. Future optimizations may explore different mass reduction targets (30% or 40% mass reduction, for instance) when the clinical requirements are established.
Regarding the safety factors, in the absence of industry standards prescribing specific numerical values, references from the literature were considered. Tomanec et al. [27] reported a safety factor of 1.72 for supporting rings of EFs, while Le et al. [28] reported a safety factor of 1.5. Caburnay et al. [29] analyzed an EF using FEM with safety factors computed from composite failure criteria. Also, there are previous studies [30] focused on the topological optimization of medical devices that use 1.4 as the minimum acceptable safety factor. The critical values were around 1, and the final design was considered safe as long as no elements with a safety factor lower than 1 were present. In our designs, the minimum safety factor is higher than 1.4. If subsequent mechanical tests indicate that this value is not sufficient to prevent ring failure, the design can be further refined using the same topology-optimization framework, for example, by adjusting the ring thickness.

3.3. PA6-CF Ring Printability Assessment

Another objective of this subsection was to document the qualitative printability issues observed during the manufacturing of the full-scale PA6-CF rings, based on visible defects that directly affect structural feasibility, such as layer delamination, local hole undersizing, build plate warping and filament jamming.
The 3D printing of the full-scale K1C ring from PA6-CF (Figure 14) proved challenging despite extensive prior experience with the MEX and this material, and regardless of implementing corrective actions initiated during the specimen manufacturing stage.
Figure 14. Physical parts: (a) K1C ring 3D-printed from PA6-CF; (b) failures from PA6-CF ring, K1C configuration, and other manufacturing defects.
The difficulties were related to the dimensions and the geometry of the ring (large, thin and with large pockets and many small holes); higher extrusion temperature, which induced contractions (for instance, build plate buckling occurred frequently, necessitating mechanical reinforcement with Kapton insulation tape added on the build plate edges); and the high sensitivity to moisture of the material, this latter property requiring continuous storage of the filament in a dryer and feeding through a Bowden tube. Still, jamming and filament breakage occasionally occurred due to tube misalignment or excessive filament stiffness.
In addition to the known undersizing of printed holes [31], which was addressed by manufacturing calibration parts with different diameters, ultimately prescribing a nominal diameter of 6.25 mm that resulted in a 6.0 mm printed hole, other issues such as poor Z-seam alignment, layer delamination, and discontinuities from retraction waste frequently occurred.
Consequently, a full-scale PA6-CF ring was successfully manufactured, though only after several trials (Figure 14a) and by an experienced operator adjusting the printer to avoid defects (Figure 14b).
In a second step, a global dimensional accuracy assessment was conducted by comparing the initial CAD with the 3D scan of the K1C printed ring, with the aim to evaluate full tolerance compliance of the ring. The intent was to determine whether the material and MEX can reliably produce large and thin rings without unacceptable deviations in these critical features.
The scanning process was conducted using a Revopoint Metro X 3D scanner (Revopoint 3D, Shenzhen, China) with 0.01 mm precision (repeatability) and up to 0.03 mm single-frame accuracy [32], while CloudCompare software v2.14 [33] was used for evaluating the dimensional deviations between the nominal ring and the 3D-printed ring. Prior to scanning, a matting spray was applied to the ring surface to reduce specular reflections during optical acquisition.
As observed in Figure 15, PA6-CF ring deviations compared with initial CAD were between 0.51 mm and −0.08 mm for most regions of the ring. However, a significant warping is observed at the exterior edges of the ring with a maximum measured error of 1.29 mm. This occurred due to the significant contraction forces generated during the cool-down of the part, causing the buckling of the part following the removal from the printing bed.
Figure 15. K1C 3D-printed ring measured error compared to nominal model, and ring split proposal.
Considering the targeted application context, remote areas and low-income countries, where printing is likely to be performed by operators with limited experience and minimal parameter optimization, further work will focus on designing modular PA6-CF rings composed of four smaller segments fastened together at the ends, as shown in Figure 15, that are less prone to printing defects. An additional advantage of modular design is that the manufacturing time of the ring can be reduced by printing the components in parallel on multiple printers.
Another solution to limiting the occurrence of internal compressive forces due to layers cooling down is to use a controlled-temperature heat chamber. Jaber et al. [34] have shown that even using an enclosure temperature of 60 °C for printing of ABS filament can significantly reduce the warping and layer shifting at vertical levels, thus improving the overall quality and precision of the print.

4. Conclusions and Further Work

This research proposed and validated a methodology for developing external fixator rings manufactured in MEX from PA6-CF, a material not previously investigated for this medical device. Thus, the research combined experimental characterization, FEA, and topology optimization into an integrated workflow for assessing mechanical feasibility and printability.
Characterization of PA6-CF was conducted on both as-printed and cold plasma-sterilized specimens, using DIC investigations, which represented another novelty of the work. The results showed moderate reductions in stiffness (−6.4%) and strength (−10.2%) after sterilization. Although the sterilization was conducted without prior washing of specimens, this controlled simplification allowed the isolation of material–cold plasma sterilization effects, providing baseline data for future clinically standardized protocols. It is important to mention that the present study represents an engineering feasibility assessment and does not address clinical performance. Moisture absorption was discussed as a possible contributor, as moisture content was not measured and no post-sterilization drying/conditioning was applied before tensile testing. Future work will quantify mass change and test specimens conditioned to constant mass (desiccator/vacuum oven, controlled temperature) to separate moisture effects from thermal-conditioning effects on crystallinity. This study conveyed a safety-factor improvement of 24% due to the material change alone and 68% and 94% due to topology optimization for two rings’ configurations, respectively, compared with the PLA-CF non-optimized ring geometry of the previous work. The optimized PA6-CF rings were shown to have safety factors above 1.4 under 1000 N wire pretension while achieving up to 50% mass reduction without compromising stiffness. Thus, the study demonstrated that PA6-CF is a promising material for lightweight, load-bearing external fixator components and establishes a reproducible experimental–numerical framework applicable to other high-performance polymer composites and medical devices.
Nonetheless, the 3D printing of large, thin PA6-CF components was challenging because of thermal contraction and moisture sensitivity, which required careful process control and experienced operators. Future work will address the modular redesign of PA6-CF rings and include quantitative moisture monitoring and complete clinical sterilization protocols to enhance reliability and reproducibility.

Author Contributions

Conceptualization, I.B., T.-G.A. and D.P.; data curation, I.B. and F.B.; formal analysis, I.B., T.-G.A., D.P. and F.B.; investigation, I.B., T.-G.A., D.P. and F.B.; methodology, T.-G.A. and D.P.; software, I.B.; validation, I.B., T.-G.A., D.P. and F.B.; writing—original draft, I.B. and D.P.; writing—review and editing, T.-G.A., D.P. and F.B. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Data Availability Statement

Data available upon request due to research in progress.

Conflicts of Interest

The authors declare no conflicts of interest.

Abbreviations

The following abbreviations are used in this manuscript:
MEXMaterial Extrusion Process
DICDigital Image Correlation
EFExternal Fixator
FEAFinite Element Analysis

References

  1. Sen, D. Principles and Overview of External Fixators in Orthopaedic Traumatology. In Handbook of Orthopaedic Trauma Implantology; Banerjee, A., Biberthaler, P., Shanmugasundaram, S., Eds.; Springer: Singapore, 2023; pp. 1–20. [Google Scholar]
  2. Bibbo, C.; Bauder, A.R.; Nelson, J.; Ahn, J.; Levin, L.S.; Mehra, S.; Kovach, S.J., III. Reconstruction of traumatic defects of the tibia with free fibula flap and external fixation. Ann. Plast. Surg. 2020, 85, 516–521. [Google Scholar] [CrossRef] [Scilit]
  3. Bible, J.E.; Mir, H.R. External fixation: Principles and applications. JAAOS-J. Am. Acad. Orthop. Surg. 2020, 85, 683–690. [Google Scholar] [CrossRef] [Scilit]
  4. Bliven, E.K.; Greinwald, M.; Hackl, S.; Augat, P. External fixation of the lower extremities: Biomechanical perspective and recent innovations. Injury 2019, 50, S10–S17. [Google Scholar] [CrossRef] [Scilit]
  5. Watson, M.A.; Mathias, K.J.; Maffulli, N. External ring fixators: An overview. Proc. Inst. Mech. Eng. Part H J. Eng. Med. 2000, 214, 459–470. [Google Scholar] [CrossRef] [Scilit]
  6. Chaus, G.W.; Dukes, C.; Hak, D.J.; Mauffrey, C.; Hammerberg, E.M. Analysis of usage and associated cost of external fixators at an urban level 1 trauma centre. Injury 2014, 45, 1611–1613. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  7. Manon, J.; Pletser, V.; Saint-Guillain, M.; Vanderdonckt, J.; Wain, C.; Jacobs, J.; Cornu, O. An easy-to-use external fixator for all hostile environments, from space to war medicine: Is it meant for everyone’s hands? J. Clin. Med. 2023, 12, 4764. [Google Scholar] [CrossRef] [Scilit]
  8. Cinthuja, P.; Wijesinghe, P.C.I.; Silva, P. Use of external fixators in developing countries: A short socioeconomic analysis. Cost Eff. Resour. Alloc. 2022, 20, 14. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  9. Landaeta, F.J.; Shiozawa, J.N.; Erdman, A.; Piazza, C. Low cost 3D printed clamps for external fixator for developing countries: A biomechanical study. 3D Print. Med. 2020, 6, 31. [Google Scholar] [CrossRef] [Scilit]
  10. Badea, I.; Alexandru, T.G.; Popescu, D. Feasibility and mechanical performance of 3D-Printed polymer composite external fixators for tibial fractures. Appl. Sci. 2025, 15, 4007. [Google Scholar] [CrossRef] [Scilit]
  11. Dimitrellou, S.; Iakovidis, I.; Psarianos, D.R. Mechanical characterization of polylactic acid, polycarbonate, and carbon fiber-reinforced polyamide specimens fabricated by fused deposition modeling. J. Mater. Eng. Perform. 2024, 33, 3613–3626. [Google Scholar] [CrossRef] [Scilit]
  12. Sun, B.; Mubarak, S.; Zhang, G.; Peng, K.; Hu, X.; Zhang, Q.; Wang, J. Fused-deposition modeling 3D printing of short-cut carbon-fiber-reinforced PA6 composites for strengthening, toughening, and light weighting. Polymers 2023, 15, 3722. [Google Scholar] [CrossRef] [Scilit]
  13. Alqahtani, M.S.; Al-Tamimi, A.A.; Hassan, M.H.; Liu, F.; Bartolo, P. Optimization of a patient-specific external fixation device for lower limb injuries. Polymers 2021, 13, 2661. [Google Scholar] [CrossRef] [Scilit]
  14. Qiao, F.; Li, D.; Jin, Z.; Gao, Y.; Zhou, T.; He, J.; Cheng, L. Application of 3D printed customized external fixator in fracture reduction. Injury 2015, 46, 1150–1155. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  15. Wu, J.; Zhou, P.; Zhang, Y.; Xiao, L.; Li, L.; Xu, S. A new technology using a customized 3D-printed fixator to assist fracture reduction and fixation: Technical note. Int. J. Med. Robot. Comput. Assist. Surg. 2021, 17, e2270. [Google Scholar] [CrossRef] [Scilit]
  16. Li, H.; Li, D.; Qiao, F.; Tang, L.; Han, Q. 3D printing adjustable stiffness external fixator for mechanically stimulated healing of tibial fractures. BioMed Res. Int. 2021, 1, 8539416. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  17. Huang, M.; Xiao, M.; Gao, L.; Zhou, M.; Sha, W.; Zhang, J. Full-Scale Isogeometric Topology Optimization of Cellular Structures Based on Kirchhoff-Love Shells. Comput. Model. Eng. Sci. (CMES) 2024, 139, 2478–2505. [Google Scholar] [CrossRef] [Scilit]
  18. Feng, F.; Xiong, S.; Kobayashi, H.; Zhou, Y.; Tanaka, M.; Kawamoto, A.; Zhu, B. Nonlinear topology optimization on thin shells using a reduced-order elastic shell model. Thin-Walled Struct. 2024, 197, 111566. [Google Scholar] [CrossRef] [Scilit]
  19. Paley, D.; Fleming, B.; Catagni, M.; Kristiansen, T.; Pope, M. Mechanical evaluation of external fixators used in limb lengthening. Clin. Orthop. Relat. Res. 1990, 250, 50–57. [Google Scholar] [CrossRef] [Scilit]
  20. Standard Test Method for Tensile Properties of Plastics. Available online: https://store.astm.org/d0638-14.html (accessed on 17 October 2025).
  21. Bean, P.; Lopez-Anido, R.A.; Vel, S. Numerical modeling and experimental investigation of effective elastic properties of the 3D-printed gyroid infill. Appl. Sci. 2022, 12, 2180. [Google Scholar] [CrossRef] [Scilit]
  22. Fiberon™ PA6-CF20. Available online: https://polymaker.com/product/polymide-pa6-cf/ (accessed on 17 October 2025).
  23. Kubyshkina, G.; Zupančič, B.; Štukelj, M.; Grošelj, D.; Marion, L.; Emri, I. Sterilization effect on structure, thermal and time-dependent properties of polyamides. In Mechanics of Time-Dependent Materials and Processes in Conventional and Multifunctional Materials, Volume 3: Proceedings of the 2011 Annual Conference on Experimental and Applied Mechanics; Springer: New York, NY, USA, 2011; Volume 3. [Google Scholar]
  24. Banjo, A.D.; Agrawal, V.; Auad, M.L.; Celestine, A.D.N. Moisture-induced changes in the mechanical behavior of 3D printed polymers. Compos. Part C Open Access 2022, 7, 100243. [Google Scholar] [CrossRef] [Scilit]
  25. Gong, H.; Runzi, M.; Wang, Z.; Wu, L.; Zhang, Y. Influence of Filament Moisture on 3D Printing Nylon. Technologies 2025, 13, 376. [Google Scholar] [CrossRef] [Scilit]
  26. Popescu, D.; Iacob, M.C.; Marinescu, R. Dimensional accuracy of 3D-printed surgical cutting guides after hospital sterilization: A comparative evaluation of ten MEX materials. 3D Print. Med. 2025, 11, 44. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  27. Tomanec, F.; Rusnáková, S.; Žaludek, M. Optimization of the Material of External Fixator with FEM Simulation. Mater. Sci. Forum 2018, 919, 275–281. [Google Scholar] [CrossRef] [Scilit]
  28. Le, Q.-T.; Nguyen, T.-B.-A. Designing composite external fixator for intertrochanteric fracture treatment based on Vietnamese femur morphological parameters. Russ. Open Med. J. 2021, 10, e0214. [Google Scholar] [CrossRef] [Scilit]
  29. Caburnay, E.V.; Danao, L.M.; Magdaluyo, E.D., Jr. Finite Element Analysis of an External Fixator with Composite Connecting Rods. In Proceedings of the International MultiConference of Engineers and Computer Scientists 2023 IMECS 2023, Hong Kong, China, 5–7 July 2023. [Google Scholar]
  30. Profitiliotis, T.; Kladovasilakis, N.; Pechlivani, E.M.; Tzetzis, D. Finite element analysis and topology optimization of SLM additive manufactured orthodontic brackets. Int. J. Mod. Manuf. Technol. 2025, 17, 63–69. [Google Scholar] [CrossRef] [Scilit]
  31. Knoop, F.; Schoeppner, V. Geometrical Accuracy of Holes and Cylinders Manufactured with Fused Deposition Modeling. In Proceedings of the 28th Annual International Solid Freeform Fabrication Symposium, Austin, TX, USA, 2–5 August 2017; pp. 2757–2776. [Google Scholar]
  32. Available online: https://www.revopoint3d.com/products/3d-laser-scanner-metrox (accessed on 21 January 2026).
  33. CloudCompare—3D Point Cloud and Mesh Processing Software Open Source Project. Available online: https://www.cloudcompare.org/ (accessed on 22 December 2025).
  34. Jaber, A.S.; Saleh, A.M.; Ibraheem, M.Q. A Study on the Influence of Enclosure Temperature Control on the Printing of ABS Filament in a Three-Dimension Printer. Eng. Technol. Appl. Sci. Res. 2025, 15, 20681–20686. [Google Scholar] [CrossRef] [Scilit]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Article Metrics

Citations

Article Access Statistics

Multiple requests from the same IP address are counted as one view.