Abstract
Label-free and non-contact manipulation methods capable of actively regulating particle trajectories are important for particle handling, selective separation, and targeted collection. However, conventional magnetic-levitation approaches primarily rely on passive equilibrium positioning and provide limited capability for actively transporting particles to prescribed locations. Here, we propose a position-amplified active MagLev method that uses a moving magnetic field as a controllable non-contact actuator for particle manipulation, separation, and targeted placement. A magneto-manipulation model incorporating gravity, buoyancy, magnetic force, and hydrodynamic drag is established to predict particle trajectories and the final center distance, enabling parameter optimization. For particles of the same density with diameters of 3, 4, and 5 mm, the predicted horizontal center-to-center distances between the 3 and 4 mm particles and between the 4 and 5 mm particles are 13.16 and 3.40 mm, respectively, closely matching experimental values of 12.46 and 3.56 mm (deviations of 0.70 and 0.16 mm, respectively). The three particles form a size-ordered spatial distribution and are successfully directed into separate collection troughs. Vertically, increasing magnet spacing enhances vertical density sensitivity and enlarges inter-material distances; for TPEE, PA66, and PC/ABS, the vertical distance between PA66 and TPEE increases from 3.99 mm to 5.70 mm, and that between PA66 and PC/ABS increases from 3.08 mm to 3.99 mm, with all three guided into distinct troughs. These results demonstrate that the proposed method extends MagLev from passive equilibrium positioning to an active, model-guided manipulation platform in which particle-property differences are converted into predictable spatial responses and subsequently exploited for targeted collection. The method therefore integrates non-contact transport, positional separation, sensitivity tuning, and prescribed particle placement within a single magnetic-manipulation process.
1. Introduction
Particle sorting and collection are essential across analytical detection, materials processing, and resource recycling. However, existing techniques face inherent limitations. Cell or microbead sorting risks sample damage and demands high accuracy and recovery [1]; solid-waste recycling requires rapid identification from complex mixtures [2]; centrifugation, though widely applicable, may compromise fragile samples under high shear [3]; flotation and electrostatic methods rely on surface properties, charge states, or pretreatment [4,5]; and microfluidic systems, while precise, are often constrained in throughput and device complexity [6]. Consequently, there remains a pressing need for separation approaches that are gentle, label-free, contactless, observable, and readily compatible with collection systems—particularly for particles with similar densities, irregular morphologies, or labeling incompatibility.
Magnetic levitation (MagLev) exploits the difference in magnetic susceptibility between a sample and a paramagnetic medium, so that nonmagnetic or weakly magnetic particles experience a magneto-Archimedes force in a nonuniform magnetic field and levitate at a position related to their own density. Mirica et al. established a MagLev-based density measurement method, providing a basis for non-contact density analysis [7]. Since then, MagLev has been used for density characterization in analytical chemistry [8], analysis of foods and water [9], biological detection [10,11], and the separation of crystal polymorphs [12]. Recent reviews have summarized the expanding design space and applications of MagLev, including ring-magnet, rotational, and self-assembly platforms [13], while radial MagLev has been applied to density measurement, separation, and detection of microplastics [14]. By virtue of requiring no labels or mechanical clamping, MagLev can convert differences in particle properties into observable positional differences.
Improvements in magnetic-field configuration have further expanded the accessible working space and sensitivity of MagLev. The axial ring-magnet configuration improved sample accessibility and enabled density-based analysis, separation, and manipulation [15]. High-sensitivity MagLev has demonstrated that modification of the magnetic-field gradient and magnet spacing can substantially tune the positional response to small density differences [16]. Our previous axial–circular MagLev configuration further expanded the levitation space and demonstrated three-dimensional manipulation of a particle assembly [17]. Independently, a tunable MagLev system has demonstrated adjustable measurement range and sensitivity through magnetic-field-configuration control [18]. Dynamic MagLev strategies have also emerged; dynamically rotating MagLev couples centrifugal effects with nonlinear magnetic forces to regulate levitation equilibria [19]. These developments show that magnetic-field structural design and external actuation can provide richer and tunable spatial responses beyond conventional one-dimensional equilibrium-height readout. Independent dynamic and multidimensional studies have further demonstrated density- and size-dependent sequential separation under controlled flow [20] and three-dimensional spatial manipulation in nonlinear magnetic fields [21].
Among our previous publications, Refs. [17,22,23] are the most directly related to the present study, but their primary manipulation objectives differ from that pursued here. The axial-circular configuration [17] demonstrated collective manipulation and placement of a particle assembly, whereas magnetic projection [22] used particle trajectories generated in a prescribed stationary magnetic field for separation and transfer. In tunable-sensitivity MagLev [23], magnet-spacing adjustment was primarily used to enlarge density-dependent equilibrium gaps for sequential separation and extraction. In contrast, the present work uses controlled translation of the dual-ring magnets as the active manipulation input and couples the prescribed magnet speed vm and magnet spacing d with particle-dependent dynamics. A coupled model predicts how particles with different sizes or densities respond to the moving magnetic field, allowing suitable operating parameters to be selected in advance according to the desired manipulation response. The resulting spatial differences are not treated simply as final separation outputs, but are retained as intermediate control states for subsequent selective manipulation. The same moving dual-ring magnetic field is then programmed in speed, direction, and spacing to guide different particles along different trajectories and deliver them to prescribed collection troughs. Thus, the present method establishes a model-guided manipulation process that links particle-property-dependent response prediction, controllable spatial differentiation, active trajectory regulation, and different-destination targeted placement.
Building on these studies, the methodological novelty of the present work is the establishment of a model-guided active MagLev framework in which controlled translation of the dual-ring magnets is introduced directly as the manipulation input and quantitatively coupled with particle dynamics. The coupled magneto-manipulation model incorporates the magnetic field, magneto-Archimedes force, gravity, buoyancy, and hydrodynamic drag to predict particle trajectories, levitation-region transitions, and final spatial positions under prescribed magnet speed vm and magnet spacing d. For nonspherical particles, particle size and shape are represented in the drag model using the equivalent diameter and sphericity, providing an effective description within the present experimental scope. This framework allows suitable operating parameters to be selected before manipulation according to the desired particle response. In the horizontal route, the magnet speed is selected so that particle-dependent dynamic responses and transfer between the central and annular levitation regions convert small property differences into enlarged and predictable spatial differences. Magnet spacing provides an additional control dimension for active manipulation: a smaller d strengthens centered-levitation confinement and favors stable long-distance transport, whereas increasing d weakens the central confinement while enlarging density-dependent vertical distances and thereby creates additional spatial clearance near the collection region. These model-guided operating characteristics form the basis of the staged manipulation strategy used in this work, in which stable transport, controllable spatial differentiation, and subsequent targeted placement are integrated within the same moving dual-ring magnetic platform.
2. Experiments and Theoretical Model
2.1. Active MagLev Platform with Dual Ring Magnets and Experimental Methods
As shown in Figure 1a, the experimental platform mainly consists of a dual-ring MagLev assembly, a transparent glass container, and a motion-control platform. Two axially magnetized N35-grade NdFeB ring magnets (provided by Ningbo Ciabao Technology Co., Ltd., Ningbo, China) were arranged vertically and coaxially with their N poles facing each other, and the magnet spacing d could be adjusted using the vertical positioning mechanism. Each magnet had an outer diameter of 60 mm, an inner diameter of 30 mm, and a thickness of 10 mm, with a dimensional tolerance of ±0.1 mm. The nominal remanent flux density of the magnets was approximately 1.17 T, and this value was used in the magnetic-field calculations. The magnet assembly was mounted on the three-dimensional motion-control platform (provided by Shanghai Puhuasen Intelligent Technology Co., Ltd., Shanghai, China) to enable controlled horizontal and vertical translation. A rectangular transparent glass container (provided by Donghua Quartz Products Co., Ltd., Lianyungang, China) filled with the paramagnetic solution was positioned between the two ring magnets.
Figure 1.
Active manipulation, sorting, and targeted placement based on MagLev with dual ring magnets. (a) experimental platform; (b) cross-sectional radial magnetic-force distribution and schematic annular levitation position; (c) particle force analysis and experimental annular levitation state; (d) magnetic flux density distribution; (e) theoretical static levitation height–density curves at different magnet spacings d; (f) staged workflow of material sorting and targeted placement, including transport at a small magnet spacing, sensitivity enhancement and sequential collection at an enlarged spacing, and restoration of the initial spacing for the final placement of the remaining material.
The spacing d between the upper and lower magnets was adjusted using a screw–nut mechanism and measured with a vernier caliper with an uncertainty of ±0.1 mm, thereby changing the superimposed magnetic-field distribution and the axial gradient of the two magnets. The magnet assembly was mounted on the horizontal motion stage and driven by a motor at a prescribed speed vm along a prescribed direction. During the experiments, the container and the solution remain stationary relative to the laboratory frame, and only the magnetic field moves together with the magnets; the motion of a particle in three-dimensional space is jointly determined by magnetic force, gravity, buoyancy, and hydrodynamic drag, thereby achieving non-contact manipulation of object motion. Figure 1b shows the cross-sectional radial magnetic-force distribution and the annular levitation position; Figure 1c presents the particle force analysis and the corresponding annular levitation state; Figure 1d presents the magnetic flux density distribution; Figure 1e gives the static levitation-height-density relationship under different d; and Figure 1f schematically illustrates the staged sorting-and-placement workflow for engineering plastics, including transport at d = 20.5 mm, enlargement of the magnet spacing to 23.5 mm to increase the vertical spatial separation and enable sequential collection of PC/ABS and PA66, and restoration of d to 20.5 mm before the final placement of TPEE.
The positions of the particles were recorded using a camera. Scale bars in the experimental images were calibrated using known reference dimensions, including the diameters of the standard spherical particles, the width of the collection trough, and the height of the transparent container. The dimensional tolerances of these reference dimensions were ±0.1 mm, and the independently obtained calibration results were cross-checked for consistency. Particle center-to-center distances were measured from the calibrated images. Each experimental condition used for model calibration was repeated three times, and the horizontal center distances are reported as the mean ± standard deviation of three replicate experiments (n = 3). In the results that follow, the horizontal analysis mainly concerns the horizontal center distance between a particle and the magnet centerline and the final horizontal center distance between different particles, whereas the vertical analysis mainly concerns the vertical center distance before and after adjusting d and its correspondence with the collection troughs. The densities of the standard particles were taken from the supplier-certified values, the densities of the engineering plastics were pre-calibrated before the experiments by the standard MagLev density-measurement method, and the density of the MnCl2 solution was measured with a pycnometer; all served as known input parameters of the theoretical model.
2.2. Magnetic Field, Magneto-Archimedes Force, and Active-Manipulation Dynamics of the Dual Ring Magnets
The levitation height, vertical density sensitivity, and active-manipulation trajectory are all closely related to the magnetic-field distribution within the working region. Given that the device is axisymmetric about the magnet axis, we take the cross-section through the axis as the object of analysis and use cylindrical coordinates (r, z) to describe the magnetic field within this cross-section. The magnetic field of a single ring magnet can be expressed as the difference between the fields of a cylindrical magnet of its outer radius and a cylindrical magnet of its inner radius; the total field produced by two coaxial ring magnets with opposing N poles is given by vector superposition:
where BA and BB denote the magnetic flux density vectors produced by the lower and upper ring magnets within the working region, respectively, and d is the adjustable spacing between the two magnets. Because the two magnets have opposing N poles, the axial magnetization directions of the upper and lower magnets are opposite. In the numerical calculation, both BA and BB are obtained from the analytical field of a ring-shaped permanent magnet, and the spatial distributions of Br(r, z), Bz(r, z), and B2(r, z) within the cross-section are further computed.
Nonmagnetic or weakly magnetic particles levitating in a paramagnetic medium generally tend toward regions of weaker magnetic field. The key feature of the dual ring magnets is reflected in the spatial distribution of the radial magnetic force within the cross-section: near the central axis, the radial magnetic force points toward the axis, so that particles can form a stable centered levitation; when a particle deviates from the axis beyond a certain distance, the radial magnetic force turns to point toward the off-axis weak-field region, so that particles can form an annular levitation. As shown in Figure 1b, there exists a boundary between the two regions at which the radial magnetic force reverses direction, providing the magnetic-field basis for a particle to transfer from the centered levitation region to the annular levitation region. The magneto-Archimedes force experienced by a particle in a paramagnetic medium can be expressed as:
where χs and χm are the volume magnetic susceptibilities of the particle and the surrounding medium, respectively, V is the particle volume, and μ0 is the vacuum permeability. In the theoretical calculations, the particles were treated as weakly diamagnetic, with a representative volume magnetic susceptibility of χs = −5.0 × 10−6 (SI, dimensionless), following the approximation adopted in previous MagLev work [7]. Equation (2) represents the effective magneto-Archimedes force on a particle immersed in a magnetic medium, where the contribution of the displaced medium enters through the susceptibility difference χs − χm. Because ∣χs∣ ≪ 1 and ∣χm∣ ≪ 1 under the present conditions, the linear-susceptibility approximation is adopted, with χ < 0 for diamagnetic materials and χ > 0 for paramagnetic materials.
During static levitation, a particle is simultaneously subjected to gravity, buoyancy, and the magneto-Archimedes force, and the relationship between density and levitation height can be obtained from the force balance in the vertical direction:
Equation (3) relates the particle density ρs to the levitation position. Near the central axis, the Br term can be approximately neglected, and Equation (3) reduces to the commonly used one-dimensional density–height relationship. For the dual-ring-magnet structure in this work, this relationship determines not only the static levitation height of a particle but also the local radial magnetic-field region in which the particle resides under different d, thereby affecting whether the particle remains centered and follows the magnets or transfers to the annular levitation region during active manipulation.
When the magnets translate uniformly along a prescribed direction, the solution and the container remain stationary, and a particle moves relative to the solution and is subjected to hydrodynamic drag. For an equivalent spherical particle, the drag can be written as
where CD is the drag coefficient, A = πds2/4 is the frontal area, ds is the equivalent particle diameter, and k is a dimensionless system-specific calibration coefficient. CD is calculated using the Haider–Levenspiel correlation as a function of the Reynolds number Re = ρm|v|ds/η and the particle sphericity ψ [24]. Here, v denotes the magnitude of the particle velocity relative to the surrounding fluid. Since the MnCl2 solution remains stationary, where r and z are the radial and axial coordinates of the particle, respectively. This velocity is distinct from the prescribed magnet translation speed vm. For irregular particles, the equivalent diameter ds and sphericity ψ provide an effective representation of particle size and shape in the drag correlation. However, particle rotation and time-dependent changes in orientation are not explicitly resolved in the present model. Therefore, trajectory predictions for irregular fragments should be interpreted as effective estimates under the investigated low-particle-number conditions rather than as a complete description of orientation-dependent particle dynamics. In the dual-ring-magnet configuration, hydrodynamic drag affects both the radial displacement within the central levitation region and particle transfer across the radial magnetic-force reversal boundary. Therefore, k was determined separately for each particle diameter using the corresponding experimental calibration data. For particles with ds = 3, 4, and 5 mm, the calibrated k values were 0.95, 0.73, and 0.67, respectively. The correction factors for the 3 and 4 mm particles were obtained from their corresponding experimental datasets following the same calibration procedure. The 5 mm particle was selected as a representative case for explicit demonstration of the calibration–validation procedure. For this particle, only the experimental data at vm = 2.5 and 5 mm/s were used to calibrate k. After k was fixed, the measurements at vm = 7.5, 10, and 12.5 mm/s were not used for any further parameter adjustment and were retained exclusively for independent model validation. These coefficients were treated as size-specific empirical correction factors for the present experimental system and were used only for the corresponding investigated particle sizes.
To describe the complete motion of a particle from its initial levitation position to the manipulation end state, its equation of motion is decomposed into radial and axial components:
where r and z are the radial and axial coordinates of the particle, respectively, rm is the radial displacement of the magnet center, Fmag,r and Fmag,z are the radial and axial components of the magneto-Archimedes force, and Fd,r and Fd,z are the corresponding components of the hydrodynamic drag; the instantaneous coordinates of the particle relative to the magnets must be substituted when computing the magneto-Archimedes force. At the start of an experiment, a particle levitates statically near the magnet centerline; as the magnets translate uniformly at speed vm, the horizontal center distance between the particle and the magnet centerline varies with time, and the particle may either follow the magnets stably within the centered levitation region or transfer to the annular levitation region after exceeding a certain critical speed. Because the actual transfer behavior is jointly affected by particle size, drag, levitation height, local magnetic-field gradient, and magnet spacing, we numerically solve the field-dynamics model constituted by Equations (1)–(6) to obtain the variation in the horizontal center distance between the particle and the magnet centerline with time, and accordingly select the manipulation parameters that can form the target center distance and realize sorting and targeted placement.
The vertical density sensitivity is characterized by S, defined as the change in levitation height per unit change in particle density:
When d increases, the axial magnetic-field gradient between the two magnets becomes gentler, and the vertical center distance corresponding to the same density difference increases accordingly. Compared with changing the MnCl2 concentration, adjusting d does not alter the chemical environment of the solution and is therefore more suitable for tuning the vertical density sensitivity in stages within a single active-manipulation process.
All experiments were conducted at room temperature (21 ± 1 °C). The density and dynamic viscosity, η, of the MnCl2 aqueous solutions were measured using a pycnometer and a viscometer, respectively. The volume magnetic susceptibility, χm, was obtained by linear interpolation from previously reported concentration-dependent data for MnCl2 aqueous solutions [25]. The physical properties used in the experiments and theoretical calculations are summarized in Table 1.
Table 1.
Physical properties of the MnCl2 aqueous solutions used in the experiments and theoretical calculations.
3. Results and Discussion
3.1. Dual-Region Magnetic-Field Characteristics and Validation of the Dynamics Model
Figure 2 shows the cross-sectional magnetic-field distribution and the corresponding static levitation states at two magnet spacings, d = 27.5 mm and d = 30.5 mm. At d = 27.5 mm, the radial magnetic force near the central axis points toward the axis, forming a stable and continuous centered levitation region. In the experiment, particles with densities of approximately 1.1365, 1.150, and 1.170 g/cm3 levitate stably in the central region in descending order, with the levitation height decreasing monotonically with increasing density. For the 1.1365 g/cm3 and 1.150 g/cm3 particles, the theoretical vertical center distance is 5.48 mm, and the experimentally measured center distance is about 5.62 mm. In addition, another particle with a density also of 1.1365 g/cm3 does not levitate at the center but instead levitates in the off-axis annular region, at almost the same height as the centered 1.1365 g/cm3 particle. That two particles of the same density stabilize respectively in the central levitation region and the annular levitation region directly demonstrates that the dual-ring magnetic field can simultaneously maintain both centered and annular levitation states.
Figure 2.
Radial dual-region magnetic field and static levitation characteristics of the dual ring magnets. (a) theoretical levitation height–density relationship curves; (b) radial-magnetic-force direction distribution at d = 27.5 mm; (c) actual levitation positions of particles at d = 27.5 mm; (d) radial-magnetic-force direction distribution at d = 30.5 mm; (e) actual levitation states of the particles at d = 30.5 mm. The vertical center-to-center distance between the particles with densities of 1.1365 and 1.150 g/cm3 increased from approximately 5.62 mm to 7.83 mm.
When d increases to 30.5 mm, the magnetic-field distribution changes markedly: the radial confinement in the central region weakens, the previously continuous centered levitation channel becomes discontinuous, and the annular levitation region expands accordingly. At this point, the 1.1365 g/cm3 particle that was previously centered remains centered, whereas the 1.150 g/cm3 standard particle is no longer centered and is pushed to the annular levitation position. This indicates that when the spacing between the upper and lower magnets exceeds a certain value, the axial magnetic-field gradient distribution changes, the centered levitation region that was previously continuous along the central axis becomes divided, and the density range that can be continuously levitated is no longer continuous. Meanwhile, the theoretical vertical center distance between the 1.1365 g/cm3 and 1.150 g/cm3 particles increases to 7.96 mm, and the experimentally measured center distance is about 7.83 mm; estimated with a density difference of 0.0135 g/cm3, the theoretical sensitivity increases from about 406 to about 590 mm·cm3·g−1, corresponding to an increase of approximately 45.3%. This result shows that increasing the magnet spacing not only changes the centered-levitation condition of particles but also increases the vertical center distance corresponding to the same density difference, providing a basis for the subsequent vertical sorting and targeted placement.
Figure 2 illustrates that the dual ring magnets do not merely strengthen the magnetic field but simultaneously provide, on the same platform, two regulation approaches: radial levitation-region transfer and axial sensitivity tuning. The coexistence of centered levitation and annular levitation within a single device provides the condition for a particle to cross the horizontal magnetic-force boundary under active manipulation and transfer to the off-axis annular levitation region, while the spacing d between the upper and lower magnets can serve as an external control parameter for tuning the vertical density sensitivity.
After confirming the dual-region magnetic-field characteristics, 5 mm standard particles with a density of 1.14 g/cm3 were used to evaluate the predictive capability of the active-manipulation dynamics model established in Section 2.2. With a fixed magnet spacing of d = 22.5 mm and an MnCl2 concentration of 1.43 mol/L, experiments were conducted at vm = 2.5, 5, 7.5, 10, and 12.5 mm/s. The measurements at 2.5 and 5 mm/s were used exclusively for calibration of k. After k was fixed, no further parameter adjustment was performed, and the experimental results at 7.5, 10, and 12.5 mm/s were used only for independent validation. As shown in Figure 3a,b, the model predictions obtained using the fixed calibrated parameters show good agreement with the withheld validation data, indicating that the model can reasonably predict the horizontal response of the particle under manipulation conditions not used for parameter calibration. Specifically, in the low-speed regime, the particle remains within the centered levitation region, and the horizontal center distance increases continuously with vm; as the speed approaches the critical value, the particle approaches the limiting position that can be stably manipulated within the centered levitation region, and beyond this position it is subjected to an outward-pointing radial magnetic force and rapidly reaches the annular levitation region, with the horizontal center distance abruptly increasing from a few millimeters to more than ten millimeters. This transition-induced enlargement of the horizontal positional response is referred to here as the position-amplification effect. This abrupt change can be explained by the direction change in the radial magnetic force in the dual-ring magnetic field: near the central axis, the radial magnetic force points toward the axis and imposes a restoring constraint on the particle; when the particle is pushed away from the center by the hydrodynamic drag introduced by magnet translation and crosses the first horizontal magnetic-force boundary, it is subjected to an outward-pointing force, and the center distance rapidly increases and reaches the annular levitation region. There therefore exists a critical manipulation speed: below this speed, the particle follows the magnets with small continuous movements within the centered levitation region, and above this speed the particle crosses the boundary and rapidly transfers to the annular levitation region.
Figure 3.
Validation of the active-manipulation dynamics model and the effect of magnet spacing on manipulation stability. (a) Model calibration and independent validation of the horizontal center distance of a 5 mm standard particle with a density of 1.14 g/cm3 over vm = 2.5–12.5 mm/s in 1.43 mol/L MnCl2 solution at d = 22.5 mm. The experimental data at 2.5 and 5 mm/s were used for calibration of k, whereas those at 7.5, 10, and 12.5 mm/s were excluded from calibration and used only for independent validation. The solid curve represents the model prediction obtained using the fixed calibrated value of k. (b) Experimental images at the corresponding speeds; (c) variation in the horizontal center distance with time at d = 23, 25, and 27 mm under vm = 5 mm/s; (d) experimental images at the corresponding magnet spacings. In panel (a), the experimental data points represent the mean values of three replicate experiments, and the error bars indicate the standard deviation (n = 3).
The magnet spacing d itself also significantly affects the magneto-manipulation behavior. Figure 3c,d compare the magneto-manipulation responses of the same particle at the same speed vm = 5 mm/s for three spacings of d = 23, 25, and 27 mm. The results show that as d increases, the horizontal center distance between the particle and the magnet centerline gradually increases; at d = 27 mm and vm = 5 mm/s, the particle transfers from the centered levitation region to the annular levitation region during magnet translation. This indicates that an excessively large d weakens the radial restoring constraint of the central levitation region and reduces the manipulation stability. Therefore, a smaller d should be chosen when long-distance stable manipulation of particles is required, and a larger d should then be used to increase the vertical center distance between materials of different densities during the collection stage.
3.2. Magneto-Manipulation Targeted Placement and Horizontal Sorting and Collection
A central function of the proposed platform is the ability to use the moving magnetic field as a controllable non-contact actuator rather than solely as a means of generating particle separation. Once the particle trajectory and its relative position with respect to the magnet center are predicted, the magnet motion can be programmed to transport the particle along a prescribed path and deliver it to a designated location. To demonstrate this capability, the targeted-placement experiment shown in Figure 4 was first conducted using a single particle with a density of 1.16 g/cm3 at a magnet spacing of d = 23 mm. Under the selected operating parameters, the model predicts that the particle remains within the centered levitation region and follows the translating magnets stably. The magnets were first translated leftward at 5 mm/s over 50 mm to transport the particle to the front of the target collection trough. Because the levitation height of the particle was lower than the trough opening, the magnet assembly was then adjusted upward by 1.6 mm at 1 mm/s to align the particle with the trough opening. Finally, the magnets were translated by 10 mm toward the collection trough at 2.5 mm/s, guiding the particle into the designated collection location. This programmed sequence demonstrates that the moving magnetic field can be used to transport a levitated particle along a prescribed path without mechanical contact and subsequently place it at a predefined position.
Figure 4.
Validation of magneto-manipulation targeted placement. A particle with d = 23 mm and a density of 1.16 g/cm3: (a) translated leftward by 50 mm at 5 mm/s to the front of the target collection trough; (b) adjusted upward by 1.6 mm at 1 mm/s to align with the trough opening; (c) translated 10 mm toward the collection trough at 2.5 mm/s to guide the particle into the collection trough; (d) oblique view after completion of the targeted particle placement.
As shown in Figure 4, magneto-manipulation can stably deliver a single particle into a designated collection trough. In a multi-particle system, particles of different sizes or densities develop different horizontal center distances under active manipulation, which can be predicted by the theoretical model. When the resulting center distance is sufficiently large, the direction of magnet motion can be adjusted according to the particle positions, allowing different particles to be delivered into separate collection troughs.
The following horizontal sorting experiments were conducted at a fixed magnet spacing of d = 23 mm in a 1.4 mol/L MnCl2 solution. Figure 5a schematically illustrates the overall magnetic-manipulation separation and targeted-placement procedure. During the initial separation stage, the magnets translate leftward at a prescribed speed, causing particles with different sizes or densities to develop different horizontal center distances. Once a sufficiently large separation is established, the direction of magnet motion is adjusted according to the final particle positions, and the particles are subsequently guided into their corresponding collection troughs.
Figure 5.
Horizontal sorting and targeted placement with the dual ring magnets. (a) Schematic of the magnetic-manipulation separation and targeted-placement procedure; (b) theoretical calculation and sorting-and-placement experiment for 4 and 5 mm particles of the same density; (c) theoretical calculation and sorting-and-placement experiment for particles with densities of 1.140 and 1.150 g/cm3 and the same diameter; (d) theoretical calculation and ordered sorting-and-placement experiment for the three-particle system with diameters of 3, 4, and 5 mm. In panels (b–d), the left plots show the theoretically calculated horizontal center distance as a function of time during the initial separation stage. The photographs show the initial particle state, the separated state produced by leftward magnet translation, and the final targeted-placement state after the direction of magnet motion was adjusted toward the corresponding collection troughs.
The critical speed required for a particle to cross the horizontal magnetic-force boundary depends on its size and density. Therefore, under the same manipulation conditions, particles with different sizes or densities exhibit different critical speeds and consequently different horizontal responses. For each experiment, an appropriate manipulation speed was first determined using the theoretical model to generate a sufficiently large horizontal center distance for subsequent sorting and targeted placement. Based on this strategy, sorting experiments involving particles of different sizes, particles of different densities, and a three-particle system were performed, as shown in Figure 5b–d, respectively.
In the size-difference-based dual-particle experiment, the two particles both had a density of about 1.1365 g/cm3 and diameters of 4 mm and 5 mm. Because their densities were identical, their static levitation-height difference was extremely small, making it difficult to distinguish them by height stratification alone; the size difference mainly changed the critical speed required for levitation-region transfer through the frontal area and the drag response. The theoretical model shows that at vm = 6.8 mm/s, with the magnets moving leftward for 12 s, the 4 mm particle first crosses the horizontal magnetic-force boundary and enters the annular levitation region, while the 5 mm particle remains within the centered levitation region, at which point the horizontal center distance between the two particles is about 14.63 mm. In the experiment, after the magnets translated leftward by 80 mm at 6.8 mm/s, the two particles formed a horizontal center distance of about 14.72 mm, differing from the theoretical prediction by about 0.09 mm. Subsequently, the direction of magnet motion was changed according to this center distance, and the magnets were moved toward the collection troughs, successfully guiding the two particles of different sizes and identical density into two different collection troughs.
In the density-difference-based dual-particle experiment, the two particles both had a diameter of 5 mm and densities of 1.140 and 1.150 g/cm3, with a density difference of only 0.010 g/cm3. The density difference changed the levitation height of the particles and the local radial magnetic-field environment in which they resided, so that the two particles corresponded to different horizontal center distances under active manipulation. The theoretical model predicts that at vm = 7.1 mm/s the denser particle enters the annular levitation region while the less dense particle still follows the magnets within the centered levitation region, with a horizontal center distance of about 13.94 mm. In the experiment, after the magnets translated leftward by 80 mm at the same speed, the horizontal center distance between the two particles was about 13.16 mm; subsequently, the direction of magnet motion was changed according to this center distance and the magnets were moved toward the collection troughs, successfully guiding the two particles of different density and identical size into two different collection troughs.
In the multi-particle experiment, the three particles all had a density of about 1.1365 g/cm3 and diameters of 3, 4, and 5 mm, with size difference as the only variable. The theoretical model shows that at vm = 5.8 mm/s, with the magnets moving leftward for 12 s, the 3 mm particle with the most pronounced drag response first enters the outer position, the 4 mm particle is in the middle, and the 5 mm particle remains closer to the center; the calculated spacing between the 3 mm and 4 mm particles is about 13.16 mm, and that between the 4 mm and 5 mm particles is about 3.4 mm. In the experiment, after the magnets translated leftward by 80 mm at 5.8 mm/s, the horizontal center distances between the 3 mm and 4 mm particles and between the 4 mm and 5 mm particles were about 12.46 mm and 3.56 mm, respectively, and the three particles formed a spatial distribution ordered by size. The direction of magnet motion was then changed according to their respective horizontal center distances, sequentially guiding the three particles of different sizes and identical density into three different collection troughs.
The above experiments show that, in the dual ring magnets, the size difference changes the critical speed required for a particle to cross the horizontal magnetic-force boundary through the frontal area and the drag response, and the density difference does so through the levitation height and the local magnetic-field environment; as long as a suitable manipulation speed is selected for different particles, the small differences among particles can be converted into model-predictable horizontal center distances on the order of ten millimeters, according to which the direction of magnet motion is adjusted to guide different particles into their respective collection troughs.
3.3. Targeted Placement Assisted by Tunable Vertical Separation
The adjustable magnet spacing is used here primarily as a means of creating sufficient spatial clearance for subsequent selective manipulation and targeted collection, rather than as an independent separation endpoint. When particles with different densities remain too close in the vertical direction for reliable sequential collection, increasing the magnet spacing enlarges their density-dependent vertical center distance. The resulting positional separation can then be exploited by the active magnetic-manipulation system to guide the particles individually into prescribed collection troughs. Accordingly, the vertical strategy combines two functions within the same platform: first, adjustment of the magnetic-field configuration increases the spatial distinction between particles; second, programmed magnet motion converts the enlarged positional difference into selective transport and targeted placement. According to the static levitation results above and the manipulation experiments in Figure 3c,d, a larger d is not always better: a smaller magnet spacing provides a stronger centered-levitation constraint and allows a higher upper limit of manipulation speed, and is therefore suitable for long-distance, relatively high-speed stable manipulation; a larger magnet spacing can enlarge the vertical center distance between materials but reduces the manipulation stability, and is therefore more suitable for use with a slower speed during the collection stage. Based on these observations, we adopt a staged strategy consisting of long-distance and relatively high-speed transport at a small magnet spacing, enlargement of the magnet spacing near the collection region to enhance the vertical separation and enable low-speed sequential collection, and, when required, restoration of the smaller magnet spacing to recover stronger centered-levitation confinement for the final delivery of the remaining material.
As shown in Figure 6a,b, standard spheres with densities of 1.130 and 1.150 g/cm3 were used to validate the vertical sorting strategy. At the initial magnet spacing of d = 23.5 mm, the magnets translated at 3 mm/s, transporting both particles stably and synchronously to a position directly in front of the collection troughs. At this position, the vertical center-to-center distance between the two particles was approximately 4.82 mm. The magnet spacing was then increased to 27 mm, which reduced the axial magnetic-field gradient, increased the vertical density sensitivity, and enlarged the vertical center-to-center distance to approximately 6.47 mm. Using this enlarged vertical separation, the magnets were subsequently moved toward the collection troughs at a lower speed, sequentially guiding the 1.150 g/cm3 particle into the first trough and the 1.130 g/cm3 particle into the second trough, as shown in Figure 6c.
Figure 6.
Vertical sorting and targeted placement of standard particles based on tunable vertical density sensitivity. (a) Magnetic-field distribution and corresponding particle positions at d = 23.5 mm, where the vertical center-to-center distance is approximately 4.82 mm; (b) magnetic-field distribution and corresponding particle positions after d is increased to 27 mm, where the distance increases to approximately 6.47 mm; (c) final sorting and targeted-placement result for the 1.130 and 1.150 g/cm3 standard particles. The increased vertical separation enables the two particles to be sequentially guided into different collection troughs by magneto-manipulation.
As schematically illustrated in Figure 7a, the above strategy is further applied to the multicomponent sorting and targeted placement of engineering plastics. Three types of materials, TPEE, PA66, and PC/ABS, were selected for the experiment, with densities of about 1.108, 1.133, and 1.155 g/cm3 measured by the standard MagLev density-measurement method, among which there were three pieces each of TPEE and PC/ABS and one regular sphere of PA66. Compared with standard density spheres, engineering plastics are more complex in shape, surface roughness, and levitation attitude, and are closer to actual waste-plastic recycling scenarios. A 1.3 mol/L MnCl2 solution was used in the experiment, with a magnet manipulation speed of 2 mm/s.
Figure 7.
Vertical sorting and targeted placement of multicomponent engineering plastics. (a) Schematic of vertical sorting and targeted placement through staged adjustment of the magnet spacing; (b) magnetic-field distribution at d = 20.5 mm; (c) magnetic-field distribution at d = 23.5 mm; (d) final configuration of TPEE, PA66, and PC/ABS after sorting and targeted placement. Images (1)–(6) show the experimental sequence. In image (1), the three materials are transported to a position directly in front of the leftmost collection trough at d = 20.5 mm. In image (2), d is increased to 23.5 mm, increasing the vertical center-to-center distances of PA66–PC/ABS and PA66–TPEE from 3.08 and 3.99 mm to 3.99 and 5.70 mm, respectively. In images (3)–(5), PC/ABS and PA66 are sequentially guided into the first and second collection troughs while d remains at 23.5 mm. Before the final placement shown in image (6), d is reduced back to 20.5 mm, after which the remaining TPEE is guided into the third collection trough.
As shown in Figure 7b, at the initial magnet spacing of d = 20.5 mm, the three materials had already formed a vertical distribution according to their densities. The magnets first transported the three materials together at 2 mm/s to a position directly in front of the leftmost collection trough. The magnet spacing was then increased to d = 23.5 mm, as shown in Figure 7c, increasing the vertical center-to-center distances between PA66 and PC/ABS and between PA66 and TPEE from approximately 3.08 and 3.99 mm to approximately 3.99 and 5.70 mm, respectively. At d = 23.5 mm, PC/ABS was first guided into the leftmost collection trough. The magnets were then translated to align PA66 with the second collection trough, into which PA66 was subsequently guided. After PC/ABS and PA66 had been collected, the magnet spacing was reduced from 23.5 mm back to 20.5 mm, restoring the stronger centered-levitation confinement required for stable manipulation of the remaining TPEE. The TPEE particles were then guided into the third collection trough. The final sorted configuration is shown in Figure 7d, demonstrating the density-based sorting and targeted placement of the three-component engineering-plastic system.
4. Conclusions
In this work, we developed a position-amplified active magnetic levitation method for particle sorting and targeted placement using dual ring magnets. The magnetic configuration provides coexisting central and annular levitation regions, while magnet translation and spacing adjustment enable active trajectory regulation and tunable vertical density sensitivity. A coupled magneto-manipulation model incorporating magnetic force, gravity, buoyancy, and hydrodynamic drag was established to predict particle trajectories and final separation distances.
Two complementary sorting-and-targeted-placement routes were experimentally validated. In the horizontal direction, standard particles transfer from the central levitation region to the annular levitation region near the critical speed, causing the horizontal center distance to increase from a few millimeters to more than ten millimeters. Based on this position-amplification effect, the 4 mm and 5 mm particles of the same density, the 1.140 and 1.150 g/cm3 particles of the same size, and the multi-particle system composed of 3, 4, and 5 mm particles all formed clear horizontal center distances and were successfully guided into their respective collection troughs. In the vertical direction, adjusting the magnet spacing increased the vertical center distance of the 1.130 and 1.150 g/cm3 standard particles from about 4.82 mm to about 6.47 mm, and particles of different densities were guided into their respective collection troughs by magneto-manipulation; applying this strategy to TPEE, PA66, and PC/ABS engineering plastics achieved non-contact sorting and targeted placement of the three components.
In both manipulation routes, particle position is not treated merely as the final outcome of magnetic levitation. Instead, particle-dependent responses are first converted into predictable spatial differences, after which the moving magnetic field is actively programmed to transport the separated particles to prescribed collection locations. The horizontal route generates large positional differences through particle-dependent dynamic responses and levitation-region transfer, whereas the vertical route uses magnet-spacing adjustment to create sufficient spatial clearance for selective sequential collection. Overall, the principal contribution of this work is the transformation of MagLev from a passive position-readout method into a model-guided, non-contact particle-manipulation approach in which particle properties, dynamic trajectories, spatial separation, and final placement are linked within a single controllable process. This capability provides a new route for the selective handling and targeted collection of particles with similar properties and engineering plastics. The present study primarily focuses on a proof-of-concept demonstration involving millimeter-scale particles under low-particle-number conditions. Further extension toward smaller particle scales and higher-throughput continuous processing will require systematic investigation of size-dependent levitation dynamics, Brownian effects, and imaging and tracking capabilities. The MnCl2-based paramagnetic medium was selected for the present proof-of-concept experiments, and its material compatibility with polymer samples was not systematically evaluated. Future work may tune the working density range by adjusting the medium concentration and explore alternative paramagnetic media considering material compatibility, safety, and cost.
Author Contributions
H.Z.: Conceptualization, Methodology, Investigation, Writing—Original draft preparation, Writing—Reviewing and Editing; H.W.: Methodology, Investigation, Writing—Reviewing and Editing; C.Z.: Conceptualization, Methodology, Writing—Reviewing and Editing; C.L.: Methodology, Validation, Writing—Reviewing and Editing; B.Z.: Methodology, Validation; P.Z.: Conceptualization, Methodology, Supervision, Funding acquisition. All authors have read and agreed to the published version of the manuscript.
Funding
This research was funded by the National Natural Science Foundation of China (Grant Nos. 52575460 and 52535010), the Zhejiang Provincial Teams of Leading Talents in Innovation and Entrepreneurship (Grant No. 2024R01002), and the Key Project of Science and Technology of “Innovation Yongjiang 2035” of Ningbo (Grant No. 2025Z010).
Data Availability Statement
The original contributions presented in this study are included in the article. Further inquiries can be directed to the corresponding author.
Conflicts of Interest
The authors declare no conflicts of interest.
References
- Witek, M.A.; Freed, I.M.; Soper, S.A. Cell separations and sorting. Anal. Chem. 2020, 92, 105–131. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Gundupalli, S.P.; Hait, S.; Thakur, A. A review on automated sorting of source-separated municipal solid waste for recycling. Waste Manag. 2017, 60, 56–74. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Gh, M.S.; Norouzi, F. Guidelines for an optimized differential centrifugation of cells. Biochem. Biophys. Rep. 2023, 36, 101585. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Pawliszak, P.; Bradshaw-Hajek, B.H.; Skinner, W.; Beattie, D.A.; Krasowska, M. Frothers in flotation: A review of performance and function in the context of chemical classification. Miner. Eng. 2024, 207, 108567. [Google Scholar] [CrossRef] [Scilit]
- Dascalescu, L.; Zeghloul, T.; Medles, K.; Iuga, A. Recent advances in the electrostatic separation of particulate matter. J. Electrost. 2025, 134, 104036. [Google Scholar] [CrossRef] [Scilit]
- Sajeesh, P.; Sen, A.K. Particle separation and sorting in microfluidic devices: A review. Microfluid. Nanofluid. 2014, 17, 1–52. [Google Scholar] [CrossRef] [Scilit]
- Mirica, K.A.; Shevkoplyas, S.S.; Phillips, S.T.; Gupta, M.; Whitesides, G.M. Measuring densities of solids and liquids using magnetic levitation: Fundamentals. J. Am. Chem. Soc. 2009, 131, 10049–10058. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ge, S.; Nemiroski, A.; Mirica, K.A.; Mace, C.R.; Hennek, J.W.; Kumar, A.A.; Whitesides, G.M. Magnetic levitation in chemistry, materials science, and biochemistry. Angew. Chem. Int. Ed. 2020, 59, 17810–17855. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Mirica, K.A.; Phillips, S.T.; Mace, C.R.; Whitesides, G.M. Magnetic levitation in the analysis of foods and water. J. Agric. Food Chem. 2010, 58, 6565–6569. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ashkarran, A.A.; Mahmoudi, M. Magnetic levitation systems for disease diagnostics. Trends Biotechnol. 2021, 39, 311–321. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Baday, M.; Calamak, S.; Durmus, N.G.; Davis, R.W.; Steinmetz, L.M.; Demirci, U. Integrating cell phone imaging with magnetic levitation (i-LEV) for label-free blood analysis at the point-of-living. Small 2016, 12, 1222–1229. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Atkinson, M.B.J.; Bwambok, D.K.; Chen, J.; Chopade, P.D.; Thuo, M.M.; Mace, C.R.; Mirica, K.A.; Kumar, A.A.; Myerson, A.S.; Whitesides, G.M. Using magnetic levitation to separate mixtures of crystal polymorphs. Angew. Chem. Int. Ed. 2013, 52, 10208–10211. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Dabbagh, S.R.; Alseed, M.M.; Saadat, M.; Sitti, M.; Tasoglu, S. Biomedical applications of magnetic levitation. Adv. NanoBiomed. Res. 2022, 2, 2100103. [Google Scholar] [CrossRef] [Scilit]
- Xia, L.; Liu, R.; Liu, J.; Zhu, X.; Ding, A.; Cao, Q. Radial magnetic levitation and its application to density measurement, separation, and detection of microplastics. Anal. Chem. 2023, 95, 8660–8667. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ge, S.; Whitesides, G.M. Axial magnetic levitation using ring magnets enables simple density-based analysis, separation, and manipulation. Anal. Chem. 2018, 90, 12239–12245. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Nemiroski, A.; Kumar, A.A.; Soh, S.; Harburg, D.V.; Yu, H.-D.; Whitesides, G.M. High-sensitivity measurement of density by magnetic levitation. Anal. Chem. 2016, 88, 2666–2674. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Zhang, C.; Zhao, P.; Gu, F.; Zhang, X.; Xie, J.; He, Y.; Zhou, H.; Fu, J.; Turng, L.S. Axial-circular magnetic levitation: A three-dimensional density measurement and manipulation approach. Anal. Chem. 2020, 92, 6925–6931. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Yu, J.; Li, D.; Zhu, C.; Ouyang, Q.; Miao, C.; Yu, H. A magnetic levitation system for range/sensitivity-tunable measurement of density. Sensors 2023, 23, 3955. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Gao, Q.-H.; Song, P.-H.; Zou, H.-X.; Wu, Z.-Y.; Zhao, L.-C.; Zhang, W.-M. Dynamically rotating magnetic levitation to characterize the spatial density heterogeneity of materials. Adv. Sci. 2023, 10, 2300219. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ren, X.; Breadmore, M.C.; Maya, F. Bidimensional dynamic magnetic levitation: Sequential separation of microplastics by density and size. Anal. Chem. 2024, 96, 3259–3266. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Gao, Q.-H.; Song, P.-H.; Zou, H.-X.; Wu, Z.-Y.; Zhao, L.-C.; Zhang, W.-M. Three-dimensional manipulation via magnetic levitation. Int. J. Mech. Sci. 2025, 287, 109949. [Google Scholar] [CrossRef] [Scilit]
- Zhang, X.; Gu, F.; Xie, J.; Zhang, C.; Fu, J.; Zhao, P. Magnetic projection: A novel separation method and its first application on separating mixed plastics. Waste Manag. 2019, 87, 805–813. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Xie, J.; Lin, W.; Lyu, C.; Zhang, L.; Zhao, P.; Li, J. Total separation of multi-plastic wastes using magnetic levitation with adjustable sensitivity. Sep. Purif. Technol. 2025, 357, 130118. [Google Scholar] [CrossRef] [Scilit]
- Haider, A.; Levenspiel, O. Drag coefficient and terminal velocity of spherical and nonspherical particles. Powder Technol. 1989, 58, 63–70. [Google Scholar] [CrossRef] [Scilit]
- Zhang, B.; Zhang, C.; Lyu, C.; Wu, M.; Gao, Y.; Fu, J.; Zhao, P. Quantitative characterization of dynamic evolution in paramagnetic solutions using magnetic levitation. Sens. Actuators B Chem. 2026, 448, 139007. [Google Scholar] [CrossRef] [Scilit]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.