Highlights
What are the main findings?
- Errors in non-cardiac neonatal POCUS can be grouped into four domains (indication, acquisition, interpretation, and clinical integration), and the most harmful often follow a technically adequate scan.
- Look-alike findings, bedside cognitive bias, and gaps in probe hygiene and output control are recurring and largely preventable sources of harm.
What are the implications of the main findings?
- Units that use POCUS need a stepwise, competency-based pathway for training, supervision, credentialing, image archiving, and quality review.
- Structured reporting, simple debiasing habits, and routine collaboration with radiology and cardiology can make neonatal POCUS safer without slowing bedside care.
Abstract
Point-of-care ultrasound (POCUS) is now part of everyday work in neonatal intensive care, supporting diagnosis, procedures, and decisions in acutely unwell infants. Its spread has outpaced the development of training, credentialing, and quality assurance structures, and the neonatal literature rarely describes how and why the technique fails. In this perspective, we focus on non-cardiac neonatal POCUS (lung, cranial, and abdominal). We adapt existing frameworks into four error domains: indication, image acquisition, image interpretation, and clinical integration. We also separate errors that can occur in any ultrasound examination from those tied to the point-of-care setting, including confirmation, anchoring, and availability bias, time pressure, and fatigue. Using illustrative cases and comparative tables, we describe common look-alike findings, summarize thermal and mechanical bioeffects by organ, and discuss probe and gel hygiene in the neonatal unit. We then outline a stepwise pathway for training, supervision, credentialing, and quality assurance, and suggest practical safeguards: structured protocols and reports, simple debiasing habits, image archiving with periodic expert review, and close collaboration with radiology and cardiology. Seen in this way, investment in safety is part of the value of neonatal POCUS rather than an added cost.
1. Introduction
Point-of-care ultrasound (POCUS) is an examination performed and interpreted at the bedside by the clinician caring for the patient to answer a focused diagnostic, therapeutic, or procedural question [1]. After decades of use in adult emergency and critical care, it has become a routine tool in the neonatal intensive care unit (NICU). Current neonatal applications include lung, cranial, abdominal, urogenital, and vascular imaging, as well as guidance for vascular access, lumbar puncture, thoracentesis, and paracentesis and confirmation of endotracheal tube position [2]. Lung ultrasound remains the most thoroughly validated diagnostic application [2,3].
Guidance has followed this growth. The POCUS Working Group of the European Society of Paediatric and Neonatal Intensive Care (ESPNIC) issued international evidence-based guidelines in 2020 [1], and North American experts have recently published neonatal recommendations for training, credentialing, and quality assurance [4]. Concerns about safety persist nonetheless. In its list of health technology hazards for 2020, the Emergency Care Research Institute (ECRI) ranked POCUS second and warned that safeguards for user training, experience, and skill had not kept pace with the speed of adoption [5]. In response to these concerns, a risk assessment framework for pediatric and neonatal POCUS was later proposed [6]. It is a useful starting point, but it was built mainly around procedural and cardiac applications and around institutional and medicolegal risk. It does not give the bedside clinician a working classification of the errors themselves.
Ultrasound is also not free of biological risk. Wider use of lung ultrasound has reduced the number of chest radiographs, and therefore radiation exposure, in ventilated and CPAP-supported neonates [7]. The thermal and mechanical effects of ultrasound still deserve attention, especially in preterm infants.
This perspective proposes a structured way of thinking about the pitfalls of non-cardiac neonatal POCUS (lung, cranial, and abdominal), with the aim of supporting safer practice, training, and quality assurance. Cardiac POCUS is left aside because it has its own extensive consensus literature and because our own experience lies mainly with non-cardiac applications. Its particular position is discussed in Section 10.
2. A Gap in the Literature: Under-Reporting of POCUS Errors
Despite a large body of work on diagnostic accuracy, scoring systems, and procedural guidance, the neonatal POCUS literature says relatively little about when and why the technique fails [2]. Two reviews of the limitations of neonatal lung ultrasound are informative, but they deal mainly with situations in which air reflects the entire beam, such as deep congenital pulmonary airway malformations, bullae, subcutaneous or interstitial emphysema, and estimation of pneumothorax size [8]. Errors throughout the whole examination, from the decision to scan to the final clinical decision, have received much less attention.
Radiology offers a useful contrast. The American College of Radiology’s RADPEER program, introduced in 2002, asks radiologists to re-score a sample of their colleagues’ previous interpretations and has produced continuous, quantified data on discrepancy rates [9]. Neonatal POCUS has no comparable system, and to our knowledge no systematic estimate of its error rate has been published. We read this absence as a sign of under-reporting rather than as evidence that misdiagnosis is rare.
The supporting infrastructure is also incomplete. Among academic pediatric critical care programs performing diagnostic ultrasound, fewer than one in four had the core elements of a sustainable program (image storage, documentation, quality assurance, and credentialing), and no division had credentialed more than a quarter of its clinicians [6]. Neonatologists report fear of missing a diagnosis, legal concerns, difficult coordination with subspecialists, and a lack of training, experts, equipment, time, and funding as the main barriers to implementation [2].
Litigation data are sometimes reassuring. Reviews of lawsuits involving emergency physicians, and of those involving neonatologists and pediatric subspecialists, found that the usual allegation was a failure to perform an ultrasound, or to perform it in time, rather than misinterpretation [10,11]. This shows that misinterpretation rarely reaches the courts or the journals. It does not show that it rarely happens.
3. A Framework of Error Domains
Several authors have grouped POCUS errors along the examination process. Brockman and Wong recently classified pitfalls in adult POCUS into image acquisition, interpretation, and clinical integration [12]. The I-AIM model (indication, acquisition, interpretation, medical decision-making), developed for teaching and performing focused sonography, adds the decision to scan as a separate step [13]. We combine the two and adapt them to the newborn, using four error domains (Table 1). Our contribution lies in the neonatal content of each domain rather than in the structure itself.
Table 1.
Error domains in non-cardiac neonatal POCUS, adapted from the I-AIM model [13] and from Brockman and Wong [12].
Acquisition errors in the newborn often reflect patient size and equipment choice. The open anterior fontanelle is the natural window for neurosonography, not an obstacle. Relying on it alone, however, leaves the posterior fossa poorly seen, and supplementary mastoid and posterior fontanelle windows are needed to detect lesions such as cerebellar hemorrhage [14]. Dedicated neonatal cerebral ultrasound uses high-frequency transducers, generally in the 7.5 to 18 MHz range, usually combining a small-footprint probe for the standard coronal and sagittal planes with a linear probe for the cortex, the extra-axial spaces, and the superior sagittal sinus [14]. A general-purpose abdominal curvilinear probe gives poor near-field resolution in a small head. In older infants, a narrowing fontanelle progressively limits the window.
Interpretation errors are illustrated in Figure 1 and Figure 2. In both cases a normal structure or an artifact was taken for disease, and a more experienced examiner resolved the question with a different acoustic window or a simple anatomical landmark.
Figure 1.
An artifact misinterpreted as pathology in a neonate with a cephalhematoma. (a) A bedside scan with a linear probe placed over the cephalhematoma. A reflection of the hematoma through the skull was misread as a possible epidural hematoma, and a specialist was called. (b) Specialist examination using a transtemporal view from the side opposite the cephalhematoma, which avoids the reflection. The brain parenchyma is intact and no epidural collection is seen.
Figure 2.
A normal variant misread as pathology. Coronal cranial ultrasound in a preterm infant with bilateral connatal (frontal horn) cysts, initially taken for periventricular white matter lesions. The arrow marks the cyst on the right-hand side of the image. A horizontal line drawn through the lateral angles of the frontal horns (red line) helps the distinction: the connatal cyst lies below this line, next to the tip of the frontal horn, whereas cystic white matter lesions usually lie above it. Connatal cysts are a benign finding in preterm infants [15].
Clinical integration errors are, in our view, the most dangerous, because they follow an examination that was technically acceptable and correctly described. The case in Figure 3 shows how a near-normal lung scan can create false reassurance. Congenital diaphragmatic hernia offers a second example. Corsini et al. described the typical lung ultrasound findings as absent A-lines over the affected area, partial absence of the pleural and diaphragmatic lines, and moving bowel loops within the thorax [16]. An operator who scans only one or two anterior zones, or who is not familiar with these signs, may miss them or take the bowel into consolidation. Therefore, these findings should be read against the clinical presentation and, when available, prenatal imaging. Coexisting conditions such as pneumonia, early-onset sepsis, or air leak can also change both the appearance of the lung and its clinical meaning [3].
Figure 3.
A reassuring lung scan given too much weight: an error of clinical integration. (a) Right lung; (b) left lung [scanning zone—anterior upper zone]. A late preterm infant (34 weeks) with respiratory distress was examined at 2 h of life while on CPAP. Both sides show a regular pleural line with lung sliding and A-lines, with only occasional B-lines, an appearance close to normal aeration. The scan was read as compatible with transient tachypnea of the newborn (TTN), and the infant was kept on CPAP with parenteral nutrition only. Three facts were not taken into account: maternal vaginal colonization with group B Streptococcus, rising hypercarbia, and worsening respiratory distress. The infant developed respiratory acidosis and progressive distress, was intubated and ventilated, and antibiotics were started. The blood culture grew group B Streptococcus at 72 h. Ventilation lasted 5 days and the outcome was favorable. An A-predominant pattern obtained under CPAP does not by itself establish TTN, and a near-normal scan early in life does not exclude evolving sepsis or pneumonia. The positive culture confirms the infection but says nothing about the accuracy of the scan. The error lay in letting a reassuring image outweigh the risk factors and the clinical trajectory. (c) A typical image for transient tachypnea, left lung, lateral scan—the primary finding is the double lung point, which is a highly specific sonographic sign for Transient Tachypnea of the Newborn; Left Side (Superior): You can observe distinct horizontal A-lines beneath the pleural line, representing a normal, well-aerated lung. Right Side (Inferior): Prominent vertical B-lines extend downward, indicating interstitial fluid or alveolar edema.
4. General Ultrasound Errors Versus POCUS-Specific Errors
Some errors can occur in any ultrasound examination. Others are specific to POCUS, which non-radiologists usually perform at the bedside, under time pressure, and without a second reader. Table 2 summarizes this distinction.
Table 2.
General ultrasound pitfalls versus pitfalls specific to point-of-care use in the neonate.
4.1. Physics- and Technique-Related Errors
Incorrect gain, focus, depth, or frequency and inappropriate Doppler settings are errors common to all ultrasound users. Neonatal scanning usually calls for high-frequency transducers, because the structures of interest lie only a few centimeters deep [2]. Lung ultrasound relies on artifacts generated at the pleural interface. Post-processing designed for solid organs, such as speckle reduction (“smoothing”) filters and spatial compound imaging, alters these artifacts and can distort the appearance and apparent number of B-lines. A dedicated lung preset with these functions switched off is therefore advisable [17,18]. Doppler has its own trap: a velocity scale set for adult vessels can suggest that flow is absent when it is simply too slow to register.
4.2. Artifacts and Anatomical Variants
Mistaking an artifact for disease is a recurring problem in every organ system. In the lung, A-lines and isolated B-lines are normal or near-normal findings, and B-lines are expected during the first day or two of life while fetal lung fluid clears [19]. Even the lung point, the most specific sign of pneumothorax, has variants such as the double lung point, septated pneumothorax, and the “hydro-point” of hydropneumothorax, which can lead an inexperienced operator to over- or under-call the diagnosis [20].
Absent lung sliding needs particular care. Sliding, B-lines, or a lung pulse rule out pneumothorax at the site examined, but absent sliding alone is not specific [21]. It is also seen with apnea, intubation of one main bronchus, pleural adhesions, extensive consolidation, and high distending pressure. In ventilated adults, lung sliding was falsely absent in 4.5% of evaluations at a positive end-expiratory pressure (PEEP) of 15 cmH2O and in 26.9% at 20 cmH2O [22]. These pressures are well above those used in neonates, so the figures cannot be transferred directly. The direction of the effect remains relevant, however, since high mean airway pressure and gas trapping may reduce sliding in ventilated newborns as well. A pneumothorax diagnosis should rest on a lung point, a consistent pattern across zones, and the clinical picture, rather than a single sign [23]. Lung ultrasound also cannot measure hyperinflation, which may mimic pneumothorax.
Neonate-specific variants add a further layer of difficulty. Ventricular asymmetry, a wide anterior fontanelle, the shape of the thymus, and the normal appearance of the preterm bowel can all be misread by an inexperienced examiner. Cranial findings also need their clinical and developmental context. Studies correlating head ultrasound at term-equivalent age with general movements assessment, and with neuromotor outcome at 12 to 24 months corrected age, show that ventricular and other cerebral measurements become meaningful when combined with functional neurological assessment rather than read in isolation [24,25].
In the abdomen, gas within the bowel lumen produces bright reverberations that can resemble the echogenic foci of pneumatosis intestinalis. Gentle probe pressure helps, because intraluminal gas moves while gas trapped in the bowel wall does not [26]. Absent color Doppler signal in the bowel wall should also be read with caution: the velocity scale must be lowered as far as possible before absent perfusion is accepted [26,27]. Peristalsis is a further trap. Normal preterm bowel may show variable and sometimes sluggish motility, while reduced or absent peristalsis is a feature of established necrotizing enterocolitis (NEC) [26]. Motility should therefore be judged together with wall thickness, echogenicity, perfusion, free fluid, and portal venous gas, never on its own.
4.3. Cognitive Biases and Human Factors
This is where POCUS differs most from a formal radiological examination with a second reader. Confirmation bias arises when the operator starts with a hypothesis and notices supporting findings more readily than contradicting ones. Context or expectation bias is closely related: knowing the clinical question changes how an ambiguous image is read. Time pressure adds to both. Scans are often performed during an acute event, on an infant who is moving, crying, or surrounded by tubes and lines.
Three short scenarios show how these biases look in practice. A neonatologist who suspects sepsis finds a few B-lines in a 6-h-old infant and reads them as pneumonia, discounting the more likely explanation of normal lung fluid clearance. This is confirmation bias. A preterm infant with respiratory distress syndrome (RDS) who has received surfactant suddenly desaturates. The operator, anchored on the admission diagnosis, sees posterior consolidation, attributes the deterioration to worsening RDS or atelectasis, and does not look for a lung point in the anterior zones, so a pneumothorax is missed. This is anchoring. An operator who has just diagnosed a severe intraventricular hemorrhage over-reads a normal but prominent choroid plexus in the next infant. This is availability bias, in which a recent and memorable case shapes the next interpretation.
Fatigue and uneven experience compound these effects. The pediatric POCUS risk framework lists inadequate training, inadequate supervision, poor documentation, and the lack of any longitudinal assessment of individual skill as factors that raise the risk to patients, providers, and institutions even when the equipment performs as expected [6]. The same authors point out that the severity of risk varies widely between applications. A failed ultrasound-guided peripheral cannulation carries a very different risk from a misread assessment of cardiac function in a child later found to have myocarditis [6].
4.4. Normal Findings and Their Look-Alikes
Teaching normal and abnormal findings side by side is one of the most direct ways to prevent interpretation errors. Table 3 summarizes the discriminating features for three pairs that are frequently confused at the bedside.
Table 3.
Normal findings and look-alike pathology: discriminating features for teaching.
5. Patient Safety: Biological Effects and the ALARA Principle
Ultrasound does not use ionizing radiation, but it is not free of risk. The recognized concerns are tissue heating and mechanical effects, including inertial and non-inertial cavitation. No study has shown actual harm from diagnostic ultrasound in human neonates, but extremely preterm infants are, in theory, more vulnerable because of their immature tissues [6].
The European Federation of Societies for Ultrasound in Medicine and Biology (EFSUMB) Clinical Safety Statement, last revised in 2019, presents the mechanical index (MI) and thermal index (TI) as the safety indices displayed to the operator and confirms that the ALARA principle (as low as reasonably achievable) should guide every diagnostic examination [28]. The British Medical Ultrasound Society gives more concrete advice for neonatal scanning. The TI and MI should be displayed whenever either exceeds 0.4. Exposure time of the neonatal central nervous system should be restricted once TI exceeds 0.7, and that of other neonatal structures once TI exceeds 1.0. An MI above 0.3 carries a theoretical possibility of minor damage to the neonatal lung or intestine and should prompt a shorter exposure if that setting is clinically necessary [29]. Published neonatal lung ultrasound guidelines accordingly recommend a preset that keeps the MI at or below 0.3, since B-mode lung imaging gains nothing from higher output [17]. Table 4 summarizes the main concerns by tissue.
Table 4.
Thermal and mechanical bioeffects of diagnostic ultrasound by tissue, with practical guidance for neonatal POCUS. No harmful effect has been demonstrated in human neonates at diagnostic output; the table reflects theoretical and experimental concerns [28,29].
Doppler delivers considerably more acoustic energy than B-mode imaging. It should be kept short and, in the most immature infants, avoided when not strictly needed. We also propose, as our own suggestion rather than an established standard, that the peak MI and TI reached during a neonatal POCUS examination be recorded in the report. This would parallel the recording of dose in radiology, create an auditable record, and reinforce the ALARA habit at the bedside.
6. Infection Control and Equipment Safety
Probes and gel shared between patients can carry pathogens, and preterm infants are among the most vulnerable patients in any hospital. Probes can become contaminated with body fluids and act as vectors for cross-transmission [6]. The European Society of Radiology recommends cleaning and disinfection of the transducer before first use and after every examination. Low-level disinfection is considered sufficient for standard examinations on intact skin, whereas all interventions require high-level disinfection, a dedicated transducer cover, and sterile gel [31].
These rules need some adaptation in the NICU. The skin of extremely preterm infants is thin and easily damaged, and many scans are performed close to umbilical catheters, central lines, or surgical wounds. In these situations, we consider it prudent to treat the examination as contact with non-intact skin and to apply the higher standard, using disinfectants compatible with the manufacturer’s instructions. Gel deserves the same attention. Contaminated gel, including unopened non-sterile product, has been linked to outbreaks of Burkholderia cepacia complex bacteremia in neonatal and adult intensive care units [32]. Single-use sterile gel sachets are a simple safeguard. The keyboard, cable, and trolley should be cleaned between patients, and a machine should not move from an isolation room to other incubators without cleaning. Recording probe cleaning in the examination report closes the loop.
Equipment also ages. Transducer elements can fail and processing speed can fall over the working life of a machine, gradually degrading image quality [6]. Periodic technical checks of probes and presets belong in the quality assurance program described below.
7. Training, Credentialing, and Quality Assurance
7.1. Where the Field Stands
At present, no national or international body certifies neonatal POCUS operators or defines who may perform which examination. The authority to scan rests with individual institutions, through local credentialing and the granting of privileges [4]. Recent guidance from the National Neonatal POCUS Collaborative is the most detailed proposal so far [4]. It separates basic from advanced applications and suggests 25 to 50 studies per diagnostic application, with a mix of normal and abnormal examinations, excluding cardiac POCUS. The authors stress that these numbers are expert recommendations rather than strict requirements, and they favor competency-based over time- or number-based assessment [4]. The pathway we propose below is consistent with this guidance and is intended to be adapted locally.
7.2. A Stepwise Pathway
Figure 4 outlines the pathway. Training begins with theory and simulation: ultrasound physics, machine controls and presets, bioeffects and ALARA, probe and gel hygiene, and the error domains and cognitive biases described above. Basic probe handling and machine operation can be practiced on phantoms, simulators, and healthy adult volunteers, where scanning time carries no clinical cost. Neonatal patterns, however, can only be learned on neonates.
Figure 4.
The proposed pathway for training, credentialing, and quality assurance in non-cardiac neonatal POCUS, informed by recent expert guidance [4]. Scan numbers are indicative. Competency-based assessment is preferred to fixed thresholds. LP, lumbar puncture; MI, mechanical index; QA, quality assurance; TI, thermal index.
Supervised scanning then begins on stable infants, with the trainer at the bedside, a logbook, and archiving of every image for review. Competence is assessed by direct observation, at the bedside or in simulation, together with review of archived images, and it should cover all four domains: indication, acquisition, interpretation, and integration. A trainee who is not yet competent returns to supervised practice. Credentialing follows, with privileges defined application by application.
Critically ill or unstable neonates need a stricter rule. Until competence in the relevant application is documented, a trainee should scan such infants only with a credentialed operator present, and no management decision should rest on a trainee’s scan alone. Procedures such as central line positioning, lumbar puncture, or drainage of an effusion benefit from real-time senior review [2].
Supervision should be provided by trainers who are credentialed for the application they teach, have sustained independent experience with it, and take part in the unit’s quality review. In practice, this means an experienced neonatologist–sonographer or a pediatric radiologist, and a pediatric cardiologist or TnECHO-trained neonatologist for cardiac work. Experience as an instructor in a recognized course is an advantage.
7.3. Quality Assurance
Quality assurance should cover the indication, image quality and protocol completeness, the accuracy of interpretation, and documentation. Fraga et al. recommend archiving all POCUS images with the report, auditing interpretations against outcomes, other tests, and surgical or pathological findings, and giving timely, non-judgmental feedback. Until a unit has its own quality team, they advise auditing every image, with help from radiology, cardiology, and local experts [4]. Once a program matures, a RADPEER-like approach becomes feasible: a fixed random sample of studies is re-scored by a senior reader each quarter, giving the unit its own discrepancy rate over time and helping to detect technical drift [9]. Asynchronous review suits most trainee scans, but some situations call for real-time feedback because a decision is being made at that moment [2]. Quality assurance also covers equipment: probe integrity, preset configuration, and cleaning records. Discrepancies should lead to targeted retraining rather than blame.
8. Further Safeguards
8.1. Standardized Protocols and Checklists
Written protocols and checklists for each type of neonatal examination reduce errors of omission and make the examination less dependent on individual experience. The SAFE protocol (Sonographic Algorithm for liFe threatening Emergencies) and its successor SAFE-R apply this approach to the highest-acuity setting. They offer a simple, ordered rule-in and rule-out sequence covering cardiac tamponade, pneumothorax, pleural effusion, and myocardial function, designed for the average neonatologist with limited additional training [2,3,33].
8.2. Debiasing Strategies
Debiasing belongs at the interpretive stage of every examination. Two questions asked before an interpretation is finalized, “what else could this be?” and “what would I conclude if I knew nothing about this patient?”, directly counter confirmation and context bias. They also fit the general rule that clinical and laboratory data should be integrated with the image rather than replaced by it [3]. A related cognitive-forcing strategy, the “STOP” mnemonic (stop, think, observe, then proceed), builds a short pause into the workflow before a finding is acted upon. In our own practice, we also try to prove the first hypothesis wrong before accepting it.
8.3. Interdisciplinary Collaboration
Collaboration with radiology remains essential. POCUS does not replace a comprehensive ultrasound examination, and a specialist opinion should be sought when a finding is equivocal or when the clinical picture does not match the image. Reviews of the limitations of neonatal lung ultrasound, for example, recommend chest radiography or computed tomography to evaluate bullae, subcutaneous and interstitial emphysema, and the size of a pneumothorax, because air makes these entities partly or wholly invisible to ultrasound [8]. A regular joint case review between neonatal POCUS operators and radiology or cardiology colleagues is a simple, low-cost step. More broadly, ultrasound works best as an extension of the physical examination that answers a focused clinical question, not as a substitute for clinical skill. Used this way, adult studies show it improves diagnostic accuracy beyond physical examination alone and reduces unnecessary downstream testing [34].
8.4. Documentation and Structured Reporting
Documentation and image storage are central to safety and remain underrepresented in neonatal POCUS. A truly structured report uses a predefined, IT-supported format and standardized terminology rather than free-text dictation. This improves the consistency and reproducibility of any imaging report and makes later data analysis possible. The distinction between standardization (agreement on content and vocabulary) and structured reporting (the tool that produces the report) is directly relevant to the design of neonatal POCUS templates [35]. Neonatal lung ultrasound now has specific proposals: recent standard-setting work defines a common terminology and examination approach [18], and the Pleura-ABCDE protocol offers a region-by-region scheme for documenting the pleura, A-lines, B-lines, consolidation, dynamic signs, and effusion [36]. At a minimum, each neonatal POCUS report should state the indication, the operator, the probe and preset, the zones or windows examined, the findings and their interpretation, any limitations, and the archiving reference. Probe cleaning and, as proposed above, the peak MI and TI can be added. The key takeaways from this review are summarized in Box 1.
Box 1. Key messages.
- The absence of published data on neonatal POCUS errors is not evidence that errors are rare. Errors can be grouped into four domains, and failures of clinical integration are the most dangerous because they follow a technically adequate scan.
- Bedside cognitive bias, look-alike findings, and gaps in output control and probe hygiene are largely preventable with structured training, checklists, and simple debiasing habits.
- Competency-based credentialing, image archiving with regular expert review, and structured reporting should be seen as part of a POCUS program, not as optional additions.
9. A Note on Value: Safety Investment Is Not Simply a Cost
Cost is a frequent objection to formal POCUS training, credentialing, and quality infrastructure. The available economic evidence, drawn largely from adult internal medicine, points the other way. A prospective cost–benefit analysis of bedside ultrasound in an internal medicine department found a break-even point of 734 examinations, corresponding to about €82,000 and reached within 406 days, after which the program produced consistent savings through more appropriate diagnosis-related group coding, shorter hospital stays, and less downstream imaging [37]. This analysis was not performed in neonates. It does suggest, however, that the investment needed for training, equipment, and quality assurance, the very infrastructure this article argues for, can be recovered through better diagnostic efficiency.
In the neonatal setting, the economic argument is strengthened by the potential of lung ultrasound to reduce chest radiographs, with their radiation exposure and handling of fragile infants, as shown in a quality management project in ventilated and CPAP-supported neonates [7]. The same logic plausibly extends to fewer radiographs for confirming central line position when position is checked by ultrasound, and occasionally to avoiding head computed tomography. Whether structured POCUS also shortens length of stay through earlier diagnosis remains a hypothesis. Dedicated neonatal cost–benefit studies are needed before firmer conclusions can be drawn.
10. Limitations
This perspective is limited to lung, cranial, and abdominal POCUS. Cardiac ultrasound in the NICU nevertheless deserves more than a passing mention, and the distinction between cardiac POCUS and targeted neonatal echocardiography (TnECHO) matters for safety. TnECHO is a comprehensive, quantitative hemodynamic assessment performed by neonatologists with advanced training, ordinarily after an initial comprehensive echocardiogram has excluded congenital heart disease. It guides decisions about the ductus arteriosus, pulmonary hypertension, and cardiovascular therapy, and it has its own training standards and quality requirements [38]. The distinction does not mean that all cardiac ultrasound should be left to TnECHO specialists. With structured training, a neonatologist can reasonably acquire basic, routinely useful skills: qualitative assessment of systolic function and cardiac filling, detection of pericardial effusion, recognition of a patent ductus arteriosus with shunt direction, and signs suggesting pulmonary hypertension [38,39]. Recent recommendations list qualitative cardiac function, filling, and pericardial effusion among the basic neonatal POCUS applications [4]. Quantitative measures such as ejection fraction and cardiac output can be learned with further training, but full hemodynamic interpretation and, above all, the exclusion of structural heart disease require TnECHO or pediatric cardiology support. The errors specific to cardiac imaging deserve a separate analysis, which we have not attempted here.
The four-domain framework is a conceptual synthesis, not a validated instrument, and we have not tested it for inter-rater reliability or its ability to reduce errors in practice. As noted in Section 2, the evidence on POCUS errors is sparse, and several observations discussed here are extrapolated from adult critical care and general pediatric literature rather than from neonatal error-reporting studies. The pathway in Figure 4 reflects expert opinion and our own experience and should be adapted to local resources.
Two further limitations deserve mention. First, artificial intelligence is increasingly being built into image acquisition and interpretation. It may reduce some acquisition and interpretation errors, but it will also introduce new, automation-related errors that this framework does not yet address. Second, our proposals assume a level of institutional resources and subspecialty support that is not available everywhere. Units with limited access to radiology or cardiology will need to adapt them.
11. Conclusions
Neonatal POCUS is best regarded not as a broad screening tool but as a focused, question-driven examination performed by well-trained and appropriately supervised clinicians. Making sound, rapid decisions under clinical pressure is much harder without solid grounding in both the technique and its limits. POCUS is a skill. It requires structured training, disciplined technique, awareness of one’s own cognitive biases, and a willingness to ask for a second opinion when the image and the clinical picture disagree. A framework that separates errors of indication, acquisition, interpretation, and clinical integration, combined with attention to the human factors of bedside imaging, offers a practical starting point for the training curricula, quality assurance programs, and structured reporting systems that non-cardiac neonatal POCUS still largely lacks.
Author Contributions
Conceptualization, A.C. and A.M.D.; methodology, D.I.V., S.L.V. and A.R.G.; software, A.R.G.; validation, A.M.D. and S.L.V.; formal analysis, D.I.V. and A.C.; investigation, A.R.G. and A.C.; resources, S.L.V. and A.M.D.; data curation, A.C. and D.I.V.; writing—original draft preparation, A.M.D. and A.C.; writing—review and editing, D.I.V. and A.R.G.; supervision, A.C., A.M.D., D.I.V. and S.L.V.; project administration, A.C. All authors have read and agreed to the published version of the manuscript.
Funding
The publication of this paper was supported by the University of Medicine and Pharmacy Carol Davila, through the institutional program Publish not Perish.
Institutional Review Board Statement
Not applicable.
Informed Consent Statement
Not applicable.
Data Availability Statement
No new data were created or analyzed in this study. Data sharing is not applicable to this article.
Acknowledgments
During the preparation of this manuscript, the authors used Claude (Anthropic) Opus 5.5 for creating Figure 4, and Grammarly for language corrections. The authors have reviewed and edited the output and take full responsibility for the content of this publication.
Conflicts of Interest
The authors declare no conflicts of interest.
References
- Singh, Y.; Tissot, C.; Fraga, M.V.; Yousef, N.; Cortes, R.G.; Lopez, J.; Sanchez-De-Toledo, J.; Brierley, J.; Colunga, J.M.; Raffaj, D.; et al. International evidence-based guidelines on Point of Care Ultrasound (POCUS) for critically ill neonates and children issued by the POCUS Working Group of the European Society of Paediatric and Neonatal Intensive Care (ESPNIC). Crit. Care 2020, 24, 65. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Osborne, A.D.; Corder, W.; Stoller, J.Z.; Lawrence, K.; Carr, C.B.; Fraga, M.V.; Johng, S. Technical considerations and applications of point-of-care ultrasound in the NICU. Semin Perinatol. 2026, 50, 152190. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Raimondi, F.; Yousef, N.; Migliaro, F.; Capasso, L.; De Luca, D. Point-of-care lung ultrasound in neonatology: Classification into descriptive and functional applications. Pediatr. Res. 2021, 90, 524–531. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Fraga, M.V.; Bhombal, S.; Juliano, C.; Kim, M.; Groves, A.M.; Chan, B.; Singh, Y.; National Neonatal POCUS Collaborative. Neonatal point-of-care ultrasound-guidelines for training, credentialing and quality assurance. J. Perinatol. 2026, 46, 113–118. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- ECRI Institute. 2020 Top 10 Health Technology Hazards: Executive Brief; ECRI Institute: Fort Washington, PA, USA, 2019; Available online: https://home.ecri.org/blogs/guidance-insights-tools/2020-top-10-health-technology-hazards-executive-brief?srsltid=AU7gw4Wfux4bg2ErlJYcjTd6NRKvaubPNbV18gTZSbL2Zai-EbdvFNzE (accessed on 27 September 2026).
- Conlon, T.W.; Yousef, N.; Mayordomo-Colunga, J.; Tissot, C.; Fraga, M.V.; Bhombal, S.; Suryawanshi, P.; Medina Villanueva, A.; Siassi, B.; Singh, Y. Establishing a risk assessment framework for point-of-care ultrasound. Eur. J. Pediatr. 2022, 181, 1449–1457. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Nemes, A.F.; Toma, A.I.; Dima, V.; Serboiu, S.C.; Necula, A.I.; Stoiciu, R.; Ulmeanu, A.I.; Marinescu, A.; Ulmeanu, C. Use of Lung Ultrasound in Reducing Radiation Exposure in Neonates with Respiratory Distress: A Quality Management Project. Medicina 2024, 60, 308. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Liu, X.; Si, S.; Guo, Y.; Wu, H. Limitations of Bedside Lung Ultrasound in Neonatal Lung Diseases. Front Pediatr. 2022, 10, 855958. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Borgstede, J.P.; Lewis, R.S.; Bhargavan, M.; Sunshine, J.H. RADPEER quality assurance program: A multifacility study of interpretive disagreement rates. J. Am. Coll. Radiol. 2004, 1, 59–65. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Stolz, L.; O’Brien, K.M.; Miller, M.L.; Winters-Brown, N.D.; Blaivas, M.; Adhikari, S. A review of lawsuits related to point-of-care emergency ultrasound applications. West J. Emerg. Med. 2015, 16, 1–4. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Nguyen, J.; Cascione, M.; Noori, S. Analysis of lawsuits related to point-of-care ultrasonography in neonatology and pediatric subspecialties. J. Perinatol. 2016, 36, 784–786. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Brockman, N.; Wong, T. Pitfalls in Point-of-Care Ultrasound. Med. Clin. N. Am. 2025, 109, 253–283. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Bahner, D.P.; Hughes, D.; Royall, N.A. I-AIM: A novel model for teaching and performing focused sonography. J. Ultrasound Med. 2012, 31, 295–300. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Dudink, J.; Jeanne Steggerda, S.; Horsch, S.; eurUS. Brain Group. State-of-the-art neonatal cerebral ultrasound: Technique and reporting. Pediatr. Res. 2020, 87, 3–12. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Pal, B.R.; Preston, P.R.; Morgan, M.E.; Rushton, D.I.; Durbin, G.M. Frontal horn thin walled cysts in preterm neonates are benign. Arch. Dis. Child Fetal Neonatal Ed. 2001, 85, F187–F193. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Corsini, I.; Parri, N.; Coviello, C.; Leonardi, V.; Dani, C. Lung ultrasound findings in congenital diaphragmatic hernia. Eur. J. Pediatr. 2019, 178, 491–495. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Kurepa, D.; Zaghloul, N.; Watkins, L.; Liu, J. Neonatal lung ultrasound exam guidelines. J. Perinatol. 2018, 38, 11–22. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Singh, Y.; Dauengauer-Kirliene, S.; Yousef, N. Setting the Standards: Neonatal Lung Ultrasound in Clinical Practice. Diagnostics 2024, 14, 1413. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ruoss, J.L.; Bazacliu, C.; Cacho, N.; De Luca, D. Lung Ultrasound in the Neonatal Intensive Care Unit: Does It Impact Clinical Care? Children 2021, 8, 1098. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Volpicelli, G.; Boero, E.; Stefanone, V.; Storti, E. Unusual new signs of pneumothorax at lung ultrasound. Crit. Ultrasound J. 2013, 5, 10. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Volpicelli, G.; Elbarbary, M.; Blaivas, M.; Lichtenstein, D.A.; Mathis, G.; Kirkpatrick, A.W.; Melniker, L.; Gargani, L.; Noble, V.E.; Via, G.; et al. International evidence-based recommendations for point-of-care lung ultrasound. Intensive Care Med. 2012, 38, 577–591. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Markota, A.; Golub, J.; Stožer, A.; Fluher, J.; Prosen, G.; Bergauer, A.; Svenšek, F.; Sinkovič, A. Absence of lung sliding is not a reliable sign of pneumothorax in patients with high positive end-expiratory pressure. Am. J. Emerg. Med. 2016, 34, 2034–2036. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Raimondi, F.; Rodriguez Fanjul, J.; Aversa, S.; Chirico, G.; Yousef, N.; De Luca, D.; Corsini, I.; Dani, C.; Grappone, L.; Orfeo, L.; et al. Lung Ultrasound for Diagnosing Pneumothorax in the Critically Ill Neonate. J. Pediatr. 2016, 175, 74–78.e1. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Toma, A.I.; Dima, V.; Alexe, A.; Rusu, L.; Nemeș, A.F.; Gonț, B.F.; Arghirescu, A.; Necula, A.; Fieraru, A.; Stoiciu, R. Correlations between Head Ultrasounds Performed at Term-Equivalent Age in Premature Neonates and General Movements Neurologic Examination Patterns. Life 2024, 14, 46. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Toma, A.I.; Dima, V.; Rusu, L.; Nemeș, A.F.; Gonț, B.F.; Arghirescu, A.; Necula, A.; Fieraru, A.; Stoiciu, R.; Andrășoaie, L.; et al. Cerebral Ultrasound at Term-Equivalent Age: Correlations with Neuro-Motor Outcomes at 12–24 Months Corrected Age. Children 2025, 12, 30. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Priyadarshi, A.; Rogerson, S.; Hinder, M.; Tracy, M. Neonatologist performed point-of-care bowel ultrasound: Is the time right? Australas. J. Ultrasound Med. 2019, 22, 15–25. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Faingold, R.; Daneman, A.; Tomlinson, G.; Babyn, P.S.; Manson, D.E.; Mohanta, A.; Moore, A.M.; Hellmann, J.; Smith, C.; Gerstle, T.; et al. Necrotizing enterocolitis: Assessment of bowel viability with color Doppler US. Radiology 2005, 235, 587–594. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Kollmann, C.; Jenderka, K.V.; Moran, C.M.; Draghi, F.; Jimenez Diaz, J.F.; Sande, R.K. EFSUMB Clinical Safety Statement for Diagnostic Ultrasound (2019 revision). Ultraschall Med. 2020, 41, 387–389. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- British Medical Ultrasound Society (BMUS). Guidelines for the Safe Use of Diagnostic Ultrasound Equipment; BMUS: London, UK, 2009. [Google Scholar]
- U.S. Food and Drug Administration. Marketing Clearance of Diagnostic Ultrasound Systems and Transducers: Guidance for Industry and Food and Drug Administration Staff; Docket FDA-2017-D-5372; FDA: Silver Spring, MD, USA, 2023. [Google Scholar]
- Nyhsen, C.M.; Humphreys, H.; Koerner, R.J.; Grenier, N.; Brady, A.; Sidhu, P.; Nicolau, C.; Mostbeck, G.; D’onofrio, M.; Gangi, A.; et al. Infection prevention and control in ultrasound: Best practice recommendations from the European Society of Radiology Ultrasound Working Group. Insights Imaging 2017, 8, 523–535. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Nannini, E.C.; Ponessa, A.; Muratori, R.; Marchiaro, P.; Ballerini, V.; Flynn, L.; Limansky, A.S. Polyclonal outbreak of bacteremia caused by Burkholderia cepacia complex and the presumptive role of ultrasound gel. Braz. J. Infect. Dis. 2015, 19, 543–545. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Yousef, N.; Singh, Y.; De Luca, D. “Playing it SAFE in the NICU” SAFE-R: A targeted diagnostic ultrasound protocol for the suddenly decompensating infant in the NICU. Eur. J. Pediatr. 2022, 181, 393–398. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Narula, J.; Chandrashekhar, Y.; Braunwald, E. Time to Add a Fifth Pillar to Bedside Physical Examination: Inspection, Palpation, Percussion, Auscultation, and Insonation. JAMA Cardiol. 2018, 3, 346–350. [Google Scholar] [PubMed]
- Nobel, J.M.; Kok, E.M.; Robben, S.G.F. Redefining the structure of structured reporting in radiology. Insights Imaging 2020, 11, 10. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Sandig, J.; Küng, E.; Aichhorn, L.; Schwaberger, B.; Klemme, M.; Wagner, N.; Bührer, C.; Sternal, F. Pleura-ABCDE: A structured expert-based protocol for neonatal lung ultrasound documentation and interpretation. Ultrasound J. 2025, 17, 48. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Testa, A.; Francesconi, A.; Giannuzzi, R.; Berardi, S.; Sbraccia, P. Economic analysis of bedside ultrasonography (US) implementation in an Internal Medicine department. Intern Emerg. Med. 2015, 10, 1015–1024. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- McNamara, P.J.; Jain, A.; El-Khuffash, A.; Giesinger, R.; Weisz, D.; Freud, L.; Levy, P.T.; Bhombal, S.; de Boode, W.; Leone, T.; et al. Guidelines and recommendations for targeted neonatal echocardiography and cardiac point-of-care ultrasound in the neonatal intensive care unit: An update from the American Society of Echocardiography. J. Am. Soc. Echocardiogr. 2024, 37, 171–215. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Singh, Y.; Bhombal, S.; Katheria, A.; Tissot, C.; Fraga, M.V. The evolution of cardiac point of care ultrasound for the neonatologist. Eur. J. Pediatr. 2021, 180, 3565–3575. [Google Scholar] [CrossRef] [Scilit] [PubMed]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.