Abstract
Background: Luminex-based single antigen bead assays are widely used for HLA antibody assessment in transplantation, but their semi-quantitative fluorescence outputs may be affected by technical, signal-dependent, and bead-specific factors. This single-center study evaluated inter-operator variability under routine laboratory conditions. Methods: Twenty serum samples were tested for each assay format: Class I SAB, Class II SAB, and screening. For each assay type, the same serum set was independently tested by four trained operators using the same Luminex FlexMap 3D instrument and xPONENT software version 4.3. Sera were EDTA treated, and exported trimmed mean fluorescence intensity values were analyzed at the sample–bead level. Variability was assessed using log-transformed inter-operator spread, threshold-based discordance, variance decomposition, direct item-level operator deviations, and bead-level variability profiling. Results: Overall variability was limited, with median item-level log-range values of 0.185, 0.172, and 0.181 in Class I, Class II, and screening assays, respectively. The analytical item accounted for most total variance. However, global discordance of 1.84–5.57% increased markedly near fixed thresholds, reaching 58.7–88.9% in Class I, 26.6–96.2% in Class II, and 57.1–90.9% in screening assays. Variability was signal dependent and unevenly distributed across beads. Conclusions: Luminex-based HLA antibody assays were globally reproducible but showed localized, signal- and bead-dependent vulnerability, particularly in threshold-sensitive regions.
1. Introduction
Luminex-based solid-phase immunoassays, particularly single antigen bead (SAB) assays, are central to HLA antibody assessment in transplantation because they provide high analytical sensitivity and antigen-level resolution. SAB and related screening assays support immunological risk assessment, virtual crossmatching, donor selection, and post-transplant monitoring. However, their fluorescence outputs are semi-quantitative and should not be interpreted as simple linear surrogates of antibody strength. Signal intensity may be influenced by antigen density and presentation, bead-specific properties, reagent-related effects, and post-acquisition analytical decisions, particularly in weak and intermediate signal ranges [1,2,3,4,5,6,7,8,9,10].
Previous studies have demonstrated inter-laboratory, methodological, and lot-to-lot variation in SAB results, while bead-by-bead normalization approaches have highlighted the importance of target-specific analytical behavior [2,3,11,12,13,14,15].
Inter-operator variability represents an additional, potentially modifiable source of technical dispersion. Even when small at the global level, operator-associated effects may remain relevant if they are concentrated in selected beads or near interpretation thresholds [2,16,17,18,19].
Assay-control beads and antigen-coated beads also require distinct interpretation. The uncoated negative control reflects background behavior, whereas the positive control primarily evaluates secondary-antibody detection. Separating these components can help distinguish antigen-specific variability from background- or detection-related effects. It also provides a framework for complementing classical reproducibility metrics with residual-pattern analysis to determine whether operator-associated structure persists after removal of the dominant sample–bead signal [1,6,20,21,22].
The present study evaluated inter-operator variability in Class I SAB, Class II SAB, and screening assays performed under routine single-center conditions. We assessed item-level spread, threshold-based discordance, variance decomposition, bead-specific variability, and assay-control bead behavior. An exploratory supervised machine-learning analysis was also applied to residualized bead profiles. We hypothesized that inter-operator variability would be globally limited but non-random, signal dependent, and concentrated in selected beads and borderline reactivity ranges.
2. Materials and Methods
2.1. Study Design
This study was designed as a single-center technical evaluation of inter-operator variability in Luminex-based HLA antibody testing under routine laboratory conditions. The primary objective was to quantify the magnitude, structure, and signal dependence of operator-associated variability across different assay formats.
A total of 24 operator-level assay runs were analyzed. The study included three assay formats: Class I SAB, Class II SAB, and screening (One Lambda, West Hills, CA, USA). For each assay format, the same set of 20 serum samples was independently tested by four operators, allowing direct within-sample comparison between operators. Thus, the study included 20 sera for Class I SAB, 20 sera for Class II SAB, and 20 sera for screening. The analytical design is summarized in Table 1.
Table 1.
Study design and analytical structure.
Because the serum sets differed between acquisition periods, comparisons between years were not treated as paired temporal comparisons. Instead, the primary analysis was performed using a pooled design within each assay type.
2.2. Samples and Pre-Analytical Treatment
The study was based on serum samples selected from routine laboratory activity and internal analytical controls. For each assay format, 20 serum samples were included and the same serum set was tested independently by the four operators. Samples were not selected according to operator performance. Previous antibody results, transplant history, sensitizing events, donor information, and clinical interpretation were not used for the analytical comparison performed in this study. Before testing, samples were treated with EDTA in order to reduce complement-related interference and mitigate prozone-related phenomena, which are recognized sources of signal distortion in solid-phase HLA antibody assays. No additional heat treatment was applied.
This pre-analytical EDTA treatment was part of the routine laboratory workflow and was consistently applied before analysis. Negative and positive internal controls were also included in the analytical dataset.
2.3. Luminex Assays
All assays were performed according to the manufacturer’s instructions and local routine laboratory procedures. The protocol was fully manual. In general terms, the workflow consisted of incubation of EDTA-treated serum with the bead populations, washing steps to remove unbound material, incubation with fluorochrome-conjugated secondary antibody, additional washing, bead resuspension, and final acquisition on the instrument.
Because the aim of the study was to evaluate inter-operator variability under real-world laboratory conditions, the technical workflow was not modified for research purposes. Instead, the study captured the performance of the assay as routinely executed in practice.
2.4. Operator Training and Assay Execution
All four operators were trained laboratory personnel routinely involved in Luminex-based HLA antibody testing. Their practical experience with the assay workflow ranged from approximately 3 to 10 years. The individual experience levels were not linked to the anonymized operator labels used in the analyses because the study was not designed to evaluate seniority as a determinant of analytical performance. All operators had completed the local laboratory training process and were considered competent according to predefined internal competency criteria. They followed the same manufacturer-based and local standard operating procedure, including sample preparation, bead incubation, washing, secondary-antibody incubation, bead resuspension, and acquisition.
The laboratory participates in external proficiency testing for histocompatibility assays through GECLID. The study did not aim to rank operators according to seniority or individual proficiency performance, and operator identity was anonymized for analysis.
2.5. Data Acquisition Platform
Fluorescence data were acquired on the same Luminex FlexMap 3D instrument (Luminex Corporation, Austin, TX, USA) throughout the study. Raw data were exported from xPONENT software version 4.3 as comma-separated value files (Luminex Corporation, Austin, TX, USA).
The quantitative signal used for analysis was the Trimmed Mean fluorescence intensity, extracted directly from the “DataType: Trimmed Mean” section of each export file. This signal metric was selected because it is less sensitive to outlier events than raw mean fluorescence and is commonly used for robust assay-level signal extraction.
For the screening assay, the exported xPONENT files did not provide an explicit normalized background (NBG) ratio. Therefore, the screening analysis in this study was based on trimmed mean values rather than on the routine laboratory interpretive rule based on NBG ratio. This distinction should be considered when interpreting screening-specific results.
2.6. Assay-Control Beads
Two bead populations were treated as assay-control beads rather than antigen-coated analytical targets:
- Bead 1, corresponding to the internal negative control and consisting of an uncoated bead
- Bead 2, corresponding to the positive control used to assess secondary-antibody detection
These two beads were not interpreted as antigen-specific anti-HLA targets and were therefore analyzed separately from the main antigen-coated bead panel in the bead-ranking analyses. Their behavior was specifically evaluated in order to distinguish background-related effects from variability associated with the detection phase of the assay.
2.7. Data Structure and Analytical Unit
The exported data were restructured into long format with the following variables: assay type, acquisition period, anonymized operator, sample identifier, bead identifier, and fluorescence intensity value.
The primary analytical unit was defined as a unique sample–bead combination, hereafter referred to as an item. For each item, up to four measurements were available, corresponding to the independent readings produced by the four operators.
This item-level definition was chosen because it provides the appropriate framework for evaluating inter-operator variability for the same biological and analytical target.
2.8. Signal Transformation
Because fluorescence intensity values were strongly right-skewed, all trimmed mean values were transformed using the base-10 logarithm before variability analysis. Values equal to or below zero were clipped to a minimal positive constant prior to transformation.
The transformed signal was therefore defined as:
This transformation was applied consistently across all assays and all downstream analyses.
Log transformation was used for variability metrics and residual-based analyses because raw fluorescence intensity values were right-skewed and heteroscedastic. In contrast, threshold-based discordance analyses were performed using the original tMFI scale because the thresholds of 1000, 2000, and 3000 tMFI are defined on the untransformed fluorescence scale. Thus, log-transformed values were used to quantify relative inter-operator spread, whereas absolute tMFI values were retained for threshold-based interpretive analyses.
2.9. Quantification of Inter-Operator Variability
Inter-operator variability was quantified at the item level using the log-range, defined as the difference between the maximum and minimum log-transformed signal values observed across operators for the same item:
This measure was selected as the primary descriptor of operator-dependent spread because it directly captures the maximum divergence between operators for the same analytical target and is appropriate for skewed fluorescence data. Medians and interquartile ranges were used to summarize variability distributions.
2.10. Pooled Analytical Strategy
Because serum sets differed between acquisition periods, the principal analysis was performed using pooled data within each assay type. Pooling was performed at the sample–bead item level. For each assay type, all available antigen-coated bead measurements from complete four-operator comparisons were combined after exclusion of assay-control beads 1 and 2, unless otherwise specified. Therefore, pooled values represent distributions of item-level variability across all serum–bead combinations within an assay type, rather than averages of patient-level results or year-level summaries. This approach was chosen to characterize the global analytical structure of inter-operator variability without introducing bias from non-matched sample comparisons between acquisition periods.
2.11. Threshold-Based Discordance Analysis
To evaluate the practical effect of operator-dependent variability on signal interpretation, threshold-based discordance was assessed using fixed cutoffs of 1000, 2000, and 3000 trimmed mean fluorescence intensity units.
For each item, discordance was defined as disagreement in binary classification across operators, that is, when at least one operator produced a value above the threshold and another produced a value below it. Discordance was assessed both globally and in items with mean signal intensity close to each threshold. Near-threshold items were defined as those with a mean item-level tMFI within ±500 units of the corresponding threshold, namely 500–1500 for the 1000 cutoff, 1500–2500 for the 2000 cutoff, and 2500–3500 for the 3000 cutoff. Continuous discordance behavior across the signal range was visualized using locally weighted regression.
Threshold-based discordance was calculated analytically after data export and did not represent a clinical interpretation made independently by the operators. Operators performed the assay workflow and data acquisition under routine laboratory procedures, but the present analysis did not incorporate previous patient results, transplant history, sensitizing events, donor information, or clinical antibody interpretation rules. Therefore, discordance in this study refers to post hoc analytical disagreement around predefined fluorescence thresholds rather than to operator-specific clinical reporting decisions.
2.12. Signal-Dependent Variability
To determine whether inter-operator spread varied as a function of signal intensity, items were stratified according to their mean signal values into predefined intensity bins. The distribution of the item-level log-range was then evaluated within each bin for each assay type.
This analysis was intended to identify whether variability was uniformly distributed across the signal range or preferentially concentrated in low, intermediate, or high-intensity regions.
2.13. Variance Decomposition
Variance decomposition was performed on pooled log-transformed signal values to estimate the relative contribution of three components:
- Item-associated variance, reflecting the sample–bead signal
- Operator-associated variance
- Residual unexplained variance
The purpose of this analysis was to quantify the extent to which total signal variability was driven by the analytical item itself rather than by operator-associated technical effects.
2.14. Bead-Level Variability Analysis
Bead-level variability was assessed by summarizing, for each bead, the distribution of item-level inter-operator log-range across all available samples. The median was used as the primary ranking statistic because it is robust to outliers and better reflects the typical variability associated with a bead than a mean-based measure.
Bead-ranking analyses were performed separately for each assay type. Beads 1 and 2 were excluded from the main antigen-bead ranking and were analyzed separately as assay-control beads.
2.15. Residualization for Machine Learning
To isolate operator-specific technical structure from the dominant analytical signal, fluorescence values were residualized by subtracting the mean log-transformed signal of each item from each corresponding operator measurement:
This step removed the principal sample–bead effect and retained only residual deviations attributable to operator-associated and unexplained variability.
2.16. Supervised Machine-Learning Analysis
Residual bead profiles were used in an exploratory supervised classification analysis of operator identity. For each assay type, the input matrix consisted of residualized bead-level values, with operator identity as the class label. Residualization was performed before model training by subtracting the item-level mean, thereby reducing the dominant sample–bead signal. Logistic regression was used as the primary classifier. As sensitivity analyses, linear support vector machines and random forests were also evaluated.
To minimize sample-level information leakage, model performance was assessed using leave-one-sample-out cross-validation, ensuring that all measurements from a given sample were kept within the same validation fold. Performance was summarized using balanced accuracy because the number of observations per operator could differ across assay types. Chance performance was defined as 0.25 for the four-operator classification task. Models were evaluated with all beads included and after exclusion of assay-control beads 1 and 2.
2.17. Statistical Analysis
All statistical analyses, data preprocessing, and machine-learning workflows were performed using Python (version 3.13).
Data handling and transformation were conducted using standard scientific computing libraries, including NumPy and pandas, while statistical analyses and modeling were performed using SciPy and scikit-learn.
Continuous variables were summarized using medians and interquartile ranges because fluorescence-based assay data are not normally distributed. Threshold-based discordance was expressed as a percentage of analytical items classified differently across operators.
The analytical framework of the study was primarily descriptive and structural. Accordingly, the statistical strategy focused on robust variability quantification, threshold-based discordance profiling, variance partitioning, bead-level ranking, and residual pattern detection rather than on classical hypothesis testing alone.
All analyses were performed separately for each assay type.
3. Results
3.1. Pooled Inter-Operator Variability Remained Limited Across Assay Types
The primary pooled analysis was performed separately for Class I SAB, Class II SAB, and screening assays after exclusion of beads 1 and 2 from the main antigen-bead analysis. At the analytical-item level, defined as a unique sample-bead combination, inter-operator variability remained globally limited across all assay types when quantified as the log-range of log-transformed trimmed mean fluorescence intensity values.
The median item-level log-range was 0.185 in Class I assays, 0.172 in Class II assays, and 0.181 in screening assays (Table 2). Thus, although all three assay formats showed broadly comparable pooled reproducibility, Class I exhibited the widest overall spread, whereas Class II showed the most compact distribution. Screening assays displayed an intermediate pooled variability profile. The corresponding pooled variability distributions are shown in Figure 1.
Table 2.
Pooled inter-operator variability across assay types after exclusion of beads 1 and 2 from the main antigen-bead analysis. Values are expressed as the median and interquartile range of the item-level log-range of log-transformed trimmed mean fluorescence intensity values.
Figure 1.
Distribution of pooled inter-operator variability across assay types after exclusion of control beads 1 and 2. Histograms show the item-level log-range across operators in Class I, Class II, and screening assays.
These findings indicate that, at the global level, none of the assay types displayed generalized analytical instability. Instead, the pooled distributions were relatively compact, with only modest differences between assay formats. This establishes that the principal analytical issue is not broad operator-driven assay failure, but the possibility that relatively small technical effects become amplified in specific analytical contexts.
3.2. Low Global Discordance Concealed Substantial Local Instability near Interpretation Thresholds
Threshold-based discordance analysis demonstrated high overall agreement between operators across all assay types. In Class I assays, pooled discordance rates were 5.57%, 4.64%, and 3.61% at thresholds of 1000, 2000, and 3000 trimmed mean fluorescence intensity (tMFI), respectively (Table 3). In Class II assays, the corresponding values were 2.21%, 2.84%, and 2.63%, confirming consistently low global disagreement across thresholds. Screening assays showed discordance rates of 3.42%, 2.63%, and 1.84%.
Table 3.
Global and near-threshold discordance rates at 1000, 2000, and 3000 tMFI after exclusion of control beads 1 and 2 from the main antigen-bead analysis.
However, this global agreement masked a strongly localized pattern. When discordance was examined only for items with mean signal values near the relevant thresholds, disagreement increased sharply. In Class I assays, near-threshold discordance rose to 58.7%, 82.5%, and 88.9% at the 1000, 2000, and 3000 tMFI cutoffs, respectively. In Class II assays, the corresponding values were 26.6%, 96.2%, and 95.5%. Screening assays also showed marked increases in local discordance (Table 3).
This contrast between low global discordance and high near-threshold discordance is one of the central findings of the study. It demonstrates that inter-operator variability is not analytically important because it broadly destabilizes the assay, but because it becomes highly consequential in precisely those signal regions where threshold-based interpretation is most vulnerable. This behavior is illustrated by the continuous discordance curves in Figure 2, which show that discordance probability rises sharply in the borderline reactivity range rather than across the full signal spectrum.
Figure 2.
Continuous probability of inter-operator discordance across the pooled signal range. Smoothed discordance curves are shown for the 1000, 2000, and 3000 tMFI thresholds.
3.3. Inter-Operator Spread Showed a Clear Signal-Dependent Pattern
When item-level variability was stratified according to signal intensity, a consistent signal-dependent structure emerged across all assay types. Inter-operator spread was broader in low-to-intermediate signal ranges and progressively narrower at higher signal intensities. This effect was most pronounced in Class I assays, where the dispersion of item-level log-range values was visibly greater in weak and intermediate reactivities than in strongly reactive items.
As shown in Figure 3, the assays did not behave as systems with homogeneous technical precision across the full signal range. Instead, variability was preferentially concentrated in signal regions that are analytically difficult and interpretively unstable. High-intensity signals remained comparatively robust, whereas lower and intermediate values were more susceptible to operator-dependent divergence.
Figure 3.
Within-item inter-operator spread across signal bins for Class I, Class II, and screening assays. Violin plots show the distribution of item-level log-range according to mean signal intensity category.
These findings reinforce the threshold-based results. The main analytical vulnerability of the assay lies not in global reproducibility, but in the instability of borderline and intermediate signals. To examine this behavior more directly, operator-specific deviations from the item-level mean were visualized across all complete four-operator items (Figure 4). This comparison showed that most measurements remained closely centered around the item-specific mean, arguing against a strong systematic operator bias across the dataset. However, a subset of items exhibited substantially wider dispersion, indicating that inter-operator differences were localized rather than uniformly distributed. This direct item-level comparison is important because it shows that the analytical problem is not broad operator drift, but selective divergence affecting specific sample–bead measurements.
Figure 4.
Direct comparison of inter-operator deviations at the item level. Each thin gray line represents one complete sample–bead item measured by the four operators after exclusion of control beads 1 and 2. The y-axis shows the deviation of each operator from the item-specific mean on the log-transformed trimmed mean fluorescence scale. Symbols indicate the operator-specific median and vertical bars the interquartile range. Most observations remain centered close to zero, whereas a subset of items shows wider inter-operator dispersion, highlighting localized analytical variability in the absence of strong global operator bias.
Although the operator-specific medians showed small directional offsets, these patterns should not be interpreted as an experience-related performance gradient. Deviations were calculated relative to the item-specific mean of the four operators, and no independent reference measurement was available to establish whether a positive or negative displacement represented greater analytical accuracy. In addition, with only four operators, practical experience could not be separated from individual operator identity. Figure 4 therefore demonstrates modest operator-specific signal offsets, without establishing whether these offsets were related to practical experience, operator proficiency, or other individual technical characteristics.
3.4. Item-Level Signal Overwhelmingly Dominated Total Variance
Variance decomposition showed that the dominant source of signal variation in all pooled datasets was the analytical item itself. Item-associated variance accounted for 99.0% of total variance in Class I assays, 95.3% in Class II assays, and 99.4% in screening assays (Table 4). By contrast, operator-associated variance remained very small in absolute terms, ranging from 0.09% to 0.66%, with the remaining variance assigned to the residual component.
Table 4.
Variance decomposition of pooled log-transformed fluorescence values, showing the proportion of variance attributable to the analytical item, operator, and residual component.
These results are analytically reassuring because they indicate that the measured signal is overwhelmingly determined by the sample-bead combination rather than by the operator. At the same time, the operator contribution was not null, meaning that a measurable technical component persisted even in globally robust datasets.
This quantitative framework provides an important reference for interpreting subsequent analyses. Although the assays are globally stable, the presence of a small but measurable operator-associated component is sufficient to support the localized discordance and residual structure observed in later sections.
3.5. Variability Was Not Uniformly Distributed Across Antigen-Coated Beads
Bead-level ranking of the antigen-coated panel showed that inter-operator variability was unevenly distributed across targets. Instead of a homogeneous panel-wide behavior, a restricted subset of beads consistently exhibited higher pooled inter-operator spread.
To avoid bias introduced by selective reporting of highly variable targets, bead-level variability was evaluated across the complete antigen-coated bead panel after exclusion of assay-control beads 1 and 2 (Figure 5). Most beads clustered within a relatively narrow variability range, supporting the overall analytical robustness of the assays. However, a smaller subset consistently occupied the higher-variability tail of the distribution, indicating that inter-operator variability was not uniformly distributed across antigen-coated targets. The highest-ranking beads and their corresponding HLA antigen specificities are detailed in Supplementary Table S1.
Figure 5.
Distribution of bead-level inter-operator variability across the complete antigen-coated bead panel after exclusion of assay-control beads 1 and 2. Each point represents one antigen-coated bead ranked according to the median item-level log-range across all available samples. Most beads clustered within a relatively narrow variability range, whereas a smaller subset formed a higher-variability tail, demonstrating that inter-operator variability was unevenly distributed across the antigen-coated panel rather than restricted to isolated targets.
This result is analytically important because it shows that the panel does not behave as a uniform set of equivalent targets. Rather, a subset of beads contributes disproportionately to the overall variability landscape. In practical terms, these findings imply that summary reproducibility metrics based only on global agreement may underestimate localized analytical vulnerability if bead-level heterogeneity is ignored.
3.6. Assay-Control Beads Showed Distinct Technical Behavior and Did Not Account for the Full Residual Structure
Because bead 1 and bead 2 corresponded to assay-control targets, they were analyzed separately from the antigen-coated beads. Bead 1, the internal negative control consisting of an uncoated bead, showed higher pooled within-item variability than bead 2 in all assay types. In Class I assays, median log-range values were 0.165 for bead 1 and 0.070 for bead 2. In Class II assays, the corresponding values were 0.221 and 0.101. In screening assays, they were 0.295 and 0.090 (Table 5).
Table 5.
Pooled variability of assay-control beads 1 and 2, analyzed separately from antigen-coated beads.
The comparative distributions of control-bead variability relative to the antigen-coated panel are shown in Figure 6. These results refine the interpretation of assay-control behavior. Although bead 2 is analytically important because it reflects the secondary-detection phase, it did not emerge as the dominant source of inter-operator spread when variability was quantified correctly at the within-item level. This finding argues against an overly simplistic explanation in which the residual variability of the assay would be attributed exclusively to the secondary-antibody detection step.
Figure 6.
Variability of assay-control beads relative to the antigen-coated bead panel.
At the same time, the relatively higher variability of bead 1 suggests that background-related behavior also contributes to the technical profile of the assay. Taken together, the control-bead data indicate that both background and detection-phase effects are relevant, but neither alone explains the full structure observed across antigen-coated beads.
3.7. Exploratory Residual-Pattern Analysis
As an exploratory secondary analysis, residualized bead profiles were used to assess whether operator-associated residual patterns remained detectable after removal of the dominant sample–bead signal. Balanced accuracy remained above the four-class chance level of 0.25 in all assay types, both before and after exclusion of assay-control beads 1 and 2. After exclusion of beads 1 and 2, balanced accuracy was 0.700 for Class I, 0.788 for Class II, and 0.613 for screening assays. These findings suggest that residual operator-associated variation was not entirely random. However, this analysis was considered supportive and exploratory, and the main conclusions of the study are based on descriptive reproducibility, threshold-discordance, variance-decomposition, direct item-level, and bead-level analyses. Detailed model performance and confusion matrices are provided in Supplementary Figures S1 and S2.
3.8. Integrated Interpretation of Pooled Assay Behavior
Taken together, the pooled results define a coherent and internally consistent analytical profile across all assay types. First, inter-operator variability was globally limited, and the analytical item accounted for the vast majority of total variance. Second, this apparent robustness concealed a strong local effect, with discordance rising sharply near interpretation thresholds. Third, variability was clearly signal dependent, being concentrated in low-to-intermediate intensity ranges. Fourth, it was also bead dependent, with a subset of antigen-coated targets showing disproportionately high inter-operator spread. Fifth, assay-control beads showed distinct behavior but did not explain the full residual variability pattern. Finally, supervised machine learning confirmed that the remaining operator-associated component was structured rather than random.
In summary, the assays studied here were globally robust but analytically structured. Their residual inter-operator variability was small in absolute terms, yet structured, signal dependent, bead dependent, and particularly relevant in borderline reactivity zones. This distinction between global robustness and local vulnerability represents the central analytical message of the study.
4. Discussion
This single-center study shows that inter-operator variability in Luminex-based HLA antibody testing is limited at the global level but becomes analytically relevant in specific signal and bead contexts. Across Class I SAB, Class II SAB, and screening assays, overall spread remained limited, the sample–bead item accounted for most total variance, and direct item-level comparison did not indicate a strong global operator bias. However, this overall robustness coexisted with sharply increased discordance in borderline signal regions, clear signal dependence, heterogeneous bead-level behavior, and residual operator-associated structure. These findings support a distinction between global assay robustness and localized analytical vulnerability. SAB assays remain highly informative, but their fluorescence values should not be treated as absolute quantitative surrogates of antibody strength [9,13,23,24,25,26,27].
The contrast between low global discordance and high near-threshold discordance is central to the interpretation of the results. Operator-related variation was not distributed uniformly across the signal range; its practical consequences were concentrated in low-to-intermediate regions, where small numerical differences may alter classification around a fixed cutoff. This supports recommendations that SAB results should be interpreted using signal context, reactivity patterns, and threshold choice rather than isolated MFI values [8,9,23,28]. Importantly, the discordance evaluated here was analytical and post hoc. It did not represent independent clinical reporting by the operators and did not incorporate sensitization history, previous antibody results, donor information, or other elements used in routine interpretation.
The mechanisms underlying localized discordance are likely multifactorial. All operators followed the same manual procedure and used the same instrument and reagent lots within each assay type, making broad systematic differences unlikely. More plausible contributors include small differences in pipetting, washing efficiency, bead resuspension, incubation timing, background signal, and bead-specific antigen presentation. These effects may have limited influence on strongly positive or clearly negative results but become more consequential near fixed thresholds. All sera were EDTA-treated as part of the routine workflow, reducing complement-mediated prozone interference. The residual variability therefore cannot be attributed solely to inadequate control of this recognized confounder and likely reflects technical execution and signal behavior beyond the prozone phenomenon [29,30,31,32,33,34].
Although practical experience differed among operators, the present design cannot determine whether experience contributed to the observed directional offsets. Experience was available for only four individuals and was completely confounded with operator identity, while deviations were calculated relative to the four-operator item mean rather than to an independent reference. The findings therefore support the presence of small operator-specific technical patterns among trained personnel but do not establish a seniority-related performance gradient.
Variability was also bead dependent rather than uniformly distributed across the antigen-coated panel. This is consistent with previous reports of bead-level heterogeneity, altered antigen display, lot-specific effects, and the potential value of bead-specific normalization [21,29,35]. Most beads showed relatively similar reproducibility, whereas a restricted subset contributed disproportionately to the higher-variability tail. Global agreement measures may therefore underestimate localized assay vulnerability when bead identity is ignored, particularly for weak or borderline reactivities.
The separate evaluation of assay-control beads further refined the interpretation. Bead 1, the uncoated negative control, and bead 2, the positive control for secondary-antibody detection, do not have the same analytical meaning as antigen-coated targets. Bead 2 did not dominate inter-operator spread when variability was assessed at the within-item level, arguing against an explanation based exclusively on the secondary-detection phase. The greater variability observed for bead 1 indicates that background-related behavior may also contribute. Nevertheless, neither control bead explained the full structure observed across antigen-coated beads, suggesting that residual variability arises from several interacting technical and bead-specific processes [23,24,29].
Variance decomposition placed these findings in the correct quantitative context. Operator-associated variance was small, confirming that operator identity was not a dominant determinant of fluorescence signal. However, small structured effects may remain relevant when the decision space is narrow, the signal is borderline, or the bead is intrinsically more variable [9,20,35,36]. The residual-pattern analysis should therefore be regarded as exploratory and supportive rather than as an operator-classification tool. Its purpose was to determine whether small residual deviations retained non-random organization after removal of the dominant sample–bead effect. The principal conclusions remain based on descriptive variability, threshold discordance, variance decomposition, direct item-level comparison, and bead-level profiling [1,35].
The study has several strengths. It was conducted under routine laboratory conditions, included three assay formats, distinguished assay-control from antigen-coated beads, and used the sample–bead item as the analytical unit. It also combined complementary methods to characterize both the magnitude and structure of variability [30].
Several limitations should be acknowledged. Screening analyses used trimmed mean signal exported from xPONENT rather than the normalized background ratio used for routine interpretation; these results therefore describe signal behavior rather than direct reproduction of the laboratory positivity rule. In addition, the serum sets from the two acquisition periods were not identical, preventing paired assessment of temporal operator improvement. The inclusion of only four operators, with practical experience completely confounded with individual operator identity, also prevented a formal assessment of experience as an independent determinant of analytical variability. Finally, this was an assay-evaluation study without clinical outcome data, and the findings do not support direct conclusions regarding graft outcome or rejection risk [9,24].
5. Conclusions
This single-center study showed that inter-operator variability in Luminex-based HLA antibody testing was globally limited, with the sample–bead item accounting for most of the total signal variance. However, variability was not uniformly distributed. It was more pronounced in low-to-intermediate signal ranges, increased markedly near fixed interpretation thresholds, and differed across antigen-coated beads. These findings indicate that the main analytical vulnerability of SAB testing lies not in broad assay instability, but in localized signal- and bead-dependent variability. This supports cautious interpretation of threshold-proximal MFI values and reinforces the need to consider assay context, bead behavior, and technical variability when evaluating HLA antibody results.
Supplementary Materials
The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/joma3020020/s1, Supplementary Table S1. Supplementary Figure S1. Supervised machine-learning performance for operator prediction from residual bead signatures. Balanced accuracy is shown for Class I, Class II, and screening assays before and after exclusion of control beads 1 and 2. Supplementary Figure S2. Confusion matrices for operator classification from residual bead signatures after exclusion of control beads 1 and 2. Normalized confusion matrices are shown separately for Class I, Class II, and screening assays.
Author Contributions
Conceptualization, L.R. and C.T.; methodology, L.R., C.T., I.P. and I.S.; formal analysis, L.R. and I.P.; investigation, L.R.; writing—original draft preparation, L.R.; writing—review and editing, L.R., C.T., I.P. and I.S.; supervision, L.R. All authors have read and agreed to the published version of the manuscript.
Funding
This research received no external funding.
Institutional Review Board Statement
This study was a technical evaluation of routine laboratory assay performance using pre-existing serum samples that had been irreversibly anonymized before research analysis. The study involved no participant recruitment, intervention, additional research sampling, or analysis of identifiable or re-identifiable patient-level clinical information. No formal Ethics Committee application was undertaken for this technical laboratory evaluation.
Informed Consent Statement
Individual informed consent was not required for this secondary analytical study. The study used pre-existing serum samples obtained during routine laboratory activity that had been irreversibly anonymized before research analysis, in accordance with Article 18(4) of Portuguese Law No. 12/2005 of 26 January.
Data Availability Statement
The raw data supporting the conclusions of this article will be made available by the authors on request.
Conflicts of Interest
The authors declare no conflicts of interest.
Abbreviations
The following abbreviations are used in this manuscript:
| SAB | Single antigen bead |
| EDTA | Ethylenediamine tetraacetic acid |
| HLA | Human Leukocyte Antigen |
| NBG | Normalized background |
| LSM | Labscreen Mixed Class I & II |
| MFI | Mean Fluorescence Intensity |
References
- Ramalhete, L.; Araújo, R.; Vieira, M.B.; Vigia, E.; Calado, C.R.C.; Ferreira, A. Rejection-Focused Precision Medicine in Kidney Transplantation: Biology, Biomarkers, and Artificial Intelligence. Life 2026, 16, 674. [Google Scholar] [CrossRef] [Scilit]
- Ramalhete, L.; Araújo, R.; Teixeira, C.; Pedro, I.; Silva, I.; Ferreira, A. Mapping Anti-HLA Class I Cross-Reactivity for Transplantation Using Interpretable Embedding and Clustering of SAB MFI. AI Med. 2025, 1, 1. [Google Scholar] [CrossRef] [Scilit]
- Ramalhete, L.; Araújo, R.; Teixeira, C.; Teixeira, A.; Almeida, P.; Silva, I.; Lima, A. Evaluation of Rapid Optimized Flow Cytometry Crossmatch (Halifaster) in Living Donor Kidney Transplantation. HLA 2024, 103, e15391. [Google Scholar] [CrossRef] [Scilit]
- Phanish, M.K. Immunological Risk Assessment and Human Leukocyte Antigen Antibody Testing in Kidney Transplantation. Indian J. Nephrol. 2016, 26, 80–85. [Google Scholar] [CrossRef] [Scilit]
- Lachmann, N.; Todorova, K.; Schulze, H.; Schönemann, C. Luminex(®) and Its Applications for Solid Organ Transplantation, Hematopoietic Stem Cell Transplantation, and Transfusion. Transfus. Med. Hemother. 2013, 40, 182–189. [Google Scholar] [CrossRef] [Scilit]
- Lehmann, C.; Landgraf, R.; Doxiadis, I. The Single Antigen Luminex Bead Assay for the Definition of HLA-Specific Antibodies Revisited: Improved Reactivity by Incubation at 37 Degrees Celsius. Biologics 2025, 5, 31. [Google Scholar] [CrossRef] [Scilit]
- Karahan, G.E.; Haasnoot, G.W.; Voogt-Bakker, K.; Claas, F.H.J.; Roelen, D.; Heidt, S. A Modeling Approach for Mean Fluorescence Intensity Value Harmonization and Cutoff Prediction for Luminex Single Antigen Bead Assays of Two Different Vendors. HLA 2023, 102, 557–569. [Google Scholar] [CrossRef] [Scilit]
- Sullivan, H.C.; Liwski, R.S.; Bray, R.A.; Gebel, H.M. The Road to HLA Antibody Evaluation: Do Not Rely on MFI. Am. J. Transplant. 2017, 17, 1455–1461. [Google Scholar] [CrossRef] [Scilit]
- Sullivan, H.C.; Gebel, H.M.; Bray, R.A. Understanding Solid-Phase HLA Antibody Assays and the Value of MFI. Hum. Immunol. 2017, 78, 471–480. [Google Scholar] [CrossRef] [Scilit]
- Dean, C.L.; Krummey, S.M.; Gebel, H.M.; Bray, R.A.; Sullivan, H.C. Identification of a Recurrent Pattern of False-Positivity by Luminex HLA MHC Class I Single Antigen Bead Testing. Hum. Immunol. 2020, 81, 73–78. [Google Scholar] [CrossRef] [Scilit]
- Liwski, R.S.; Greenshields, A.L.; Conrad, D.M.; Murphey, C.; Bray, R.A.; Neumann, J.; Gebel, H.M. Rapid Optimized Flow Cytometric Crossmatch (FCXM) Assays: The Halifax and Halifaster Protocols. Hum. Immunol. 2018, 79, 28–38. [Google Scholar] [CrossRef] [Scilit]
- Carey, B.S.; Poulton, K.V.; Poles, A. Factors Affecting HLA Expression: A Review. Int. J. Immunogenet. 2019, 46, 307–320. [Google Scholar] [CrossRef] [Scilit]
- Gandhi, M.J.; Carrick, D.M.; Jenkins, S.; De Goey, S.; Ploeger, N.A.; Wilson, G.A.; Lee, J.H.; Winters, J.L.; Stubbs, J.R.; Toy, P.; et al. Lot-to-Lot Variability in HLA Antibody Screening Using a Multiplexed Bead-Based Assay. Transfusion 2013, 53, 1940–1947. [Google Scholar] [CrossRef] [Scilit]
- Mishra, V.C.; Raina, V. Enhancing Precision of the Single-Antigen Bead Assay: Considerations and Challenges. J. Clin. Transl. Pathol. 2024, 4, 12–17. [Google Scholar] [CrossRef] [Scilit]
- Bettinotti, M.P.; Zachary, A.A.; Leffell, M.S. Clinically Relevant Interpretation of Solid Phase Assays for HLA Antibody. Curr. Opin. Organ Transplant. 2016, 21, 453–458. [Google Scholar] [CrossRef] [Scilit]
- Ramalhete, L.; Almeida, P.; Araújo, R.; Espada, E. Integrative Population Analysis of MICA and MICB Using Unsupervised Machine Learning in a Large Histocompatibility Laboratory Cohort. J 2026, 9, 8. [Google Scholar] [CrossRef] [Scilit]
- Xu, X.; Giulini, M.; Bonvin, A.M.J.J. Improved Prediction of Antibody and Their Complexes with Clustered Generative Modelling Ensembles. Bioinform. Adv. 2025, 5, vbaf161. [Google Scholar] [CrossRef] [Scilit]
- Kim, D.N.; McNaughton, A.D.; Kumar, N. Leveraging Artificial Intelligence to Expedite Antibody Design and Enhance Antibody–Antigen Interactions. Bioengineering 2024, 11, 185. [Google Scholar] [CrossRef] [Scilit]
- Chomicz, D.; Kończak, J.; Wróbel, S.; Satława, T.; Dudzic, P.; Janusz, B.; Tarkowski, M.; Deszyński, P.; Gawłowski, T.; Kostyn, A.; et al. Benchmarking Antibody Clustering Methods Using Sequence, Structural, and Machine Learning Similarity Measures for Antibody Discovery Applications. Front. Mol. Biosci. 2024, 11, 1352508. [Google Scholar] [CrossRef] [Scilit]
- Tait, B.D. Detection of HLA Antibodies in Organ Transplant Recipients—Triumphs and Challenges of the Solid Phase Bead Assay. Front. Immunol. 2016, 7, 570. [Google Scholar] [CrossRef] [Scilit]
- Usureau, C.; Lhotte, R.; Devriese, M.; Siemowski, J.; Gabet, L.; Letort, V.; Taupin, J.-L. Bead-by-Bead Normalization of Single Antigen Assays: A Necessary Step for Accurate Detection of Weak Anti-HLA Antibodies. Eur. J. Immunol. 2024, 54, 2451181. [Google Scholar] [CrossRef] [Scilit]
- Ravindranath, M.H.; Ravindranath, N.M.; Amato-Menker, C.J. Luminex Multiplex Bead Assay Monitoring HLA IgG Antibodies in Sensitized Pre- and Post-Transplant Patients: Clonality of the Detection Antibody Impacts Specificity and Sensitivity. Appl. Sci. 2021, 11, 6430. [Google Scholar] [CrossRef] [Scilit]
- Ellis, T.M. Interpretation of HLA Single Antigen Bead Assays. Transplant. Rev. 2013, 27, 108–111. [Google Scholar] [CrossRef] [Scilit]
- Schinstock, C.A.; Gandhi, M.J.; Stegall, M.D. Interpreting Anti-HLA Antibody Testing Data. Transplantation 2016, 100, 1619–1628. [Google Scholar] [CrossRef] [Scilit]
- Rajalingam, R.; Cho, Y.; Da Gente, G.; Lai, J.; Buenaventura, O.; Kong, D.; Bratton, T. Assessing the Interoperability of Single Antigen Bead Assays across Vendors to Enhance the Accuracy and Efficiency of HLA Antibody Measurement in Transplant Patients. Hum. Immunol. 2025, 86, 111583. [Google Scholar] [CrossRef] [Scilit]
- Reed, E.F.; Rao, P.; Zhang, Z.; Gebel, H.; Bray, R.A.; Guleria, I.; Lunz, J.; Mohanakumar, T.; Nickerson, P.; Tambur, A.R.; et al. Comprehensive Assessment and Standardization of Solid Phase Multiplex-Bead Arrays for the Detection of Antibodies to HLA. Am. J. Transplant. 2013, 13, 1859–1870. [Google Scholar] [CrossRef] [Scilit]
- Schacker, T.; Landgraf, R.; Doxiadis, I.; Loeffler-Wirth, H.; Lehmann, C. Luminex MFI—Efforts from a Qualitative to a Quantitative Analysis. Biology 2025, 14, 686. [Google Scholar] [CrossRef] [Scilit]
- Middleton, D.; Jones, J.; Lowe, D. Nothing’s Perfect: The Art of Defining HLA-Specific Antibodies. Transpl. Immunol. 2014, 30, 115–121. [Google Scholar] [CrossRef] [Scilit]
- Friedlander, R.; Putheti, P.; Diaz, E.; Menon, A.; Ponce, B.; Muthukumar, T.; Sharma, V.K.; Dadhania, D.; Suthanthiran, M. On the Detection of Anti-HLA Antibodies Using Single Antigen Bead Luminex Assay. Transplantation 2013, 96, e24–e26. [Google Scholar] [CrossRef] [Scilit]
- Gandhi, M.J.; DeGoey, S.; Falbo, D.; Jenkins, S.; Stubbs, J.R.; Noreen, H.; Lorentzen, D.F.; Lee, J.; Stegall, M. Inter and Intra Laboratory Concordance of HLA Antibody Results Obtained by Single Antigen Bead Based Assay. Hum. Immunol. 2013, 74, 310–317. [Google Scholar] [CrossRef] [Scilit]
- Weinstock, C.; Schnaidt, M. The Complement-Mediated Prozone Effect in the Luminex Single-Antigen Bead Assay and Its Impact on HLA Antibody Determination in Patient Sera. Int. J. Immunogenet. 2013, 40, 171–177. [Google Scholar] [CrossRef] [Scilit]
- Anani, W.Q.; Zeevi, A.; Lunz, J.G. EDTA Treatment of Serum Unmasks Complement-Mediated Prozone Inhibition in Human Leukocyte Antigen Antibody Testing. Am. J. Clin. Pathol. 2016, 146, 346–352. [Google Scholar] [CrossRef] [Scilit]
- Kim, H.S.; Choi, A.R.; Yang, M.; Oh, E.J. EDTA Treatment for Overcoming the Prozone Effect and for Predicting C1q Binding in HLA Antibody Testing. Ann. Lab. Med. 2019, 39, 572–576. [Google Scholar] [CrossRef] [Scilit]
- Wang, J.; Meade, J.R.; Brown, N.K.; Weidner, J.G.; Marino, S.R. EDTA Is Superior to DTT Treatment for Overcoming the Prozone Effect in HLA Antibody Testing. HLA 2017, 89, 82–89. [Google Scholar] [CrossRef] [Scilit]
- Wehmeier, C.; Hönger, G.; Schaub, S. Caveats of HLA Antibody Detection by Solid-Phase Assays. Transpl. Int. 2020, 33, 18–29. [Google Scholar] [CrossRef] [Scilit]
- Carey, B.S.; Boswijk, K.; Mabrok, M.; Rowe, P.A.; Connor, A.; Saif, I.; Poles, A. A Reliable Method for Avoiding False Negative Results with Luminex Single Antigen Beads; Evidence of the Prozone Effect. Transpl. Immunol. 2016, 37, 23–27. [Google Scholar] [CrossRef] [Scilit]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Published by MDPI on behalf of the Oman Medical Association. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.