Abstract
Hydraulic resistance is needed for the design, characterisation and quality control of microfluidic and organ-on-chip (OoC) devices, and standardised procedures for its measurement are an objective of the European metrology project MFMET II (24NRM03). This study combines a theoretical description of hydraulic resistance with an experimental validation on a polymeric microfluidic chip. Starting from the Navier–Stokes equations, the relations for circular pipes (Hagen–Poiseuille equation) and rectangular channels are derived, and it is shown that the Darcy–Weisbach equation with shape-dependent friction factors reproduces the analytical solutions. Previously published pressure drop data of a 50 µm × 50 µm × 40 mm microchannel, measured at nominal flow rates between 0.025 and 0.400 mL/h, were re-evaluated for two methods: a pressure-driven flow controller from the Microfluidics Innovation Center (MIC) with an Elveflow MFS2 flow sensor, using measurements with and without the microfluidic chip, and a high-precision syringe pump from Cetoni with two external pressure sensors. The measured hydraulic resistances of 507.8 and 513.6 mbar h/mL agree with the theoretical value of 506.7 mbar h/mL within the expanded uncertainties. The uncertainty budget shows that the theoretical value is dominated by the channel dimensions and the measured values by the regression and the zero offset of the pressure data.
1. Introduction
The hydraulic resistance is needed for the design, characterisation and quality control of microfluidic devices (This article is a revised and expanded version of a paper entitled “Investigations on the Performance of a Newly Developed Pressure-Driven Flow Controller Used in Microfluidic Applications”, which was presented at the XXIV IMEKO World Congress, Hamburg, Germany, 26–29 August 2024 [1]. A preliminary version of this work was also presented at the 20th International Flow Measurement Conference (FLOMEKO 2026), Nara, Japan, 17–20 May 2026). In analogy to the electrical resistance, it relates the pressure drop along a channel to the volumetric flow rate, , and thus determines the flow distribution in channel networks and the driving pressure required for a given flow rate. Applications range from lab-on-a-chip and organ-on-chip (OoC) systems to microchannel cooling and microreactors. Since is obtained from two measured quantities, the pressure difference and the flow rate, its traceable determination requires traceable pressure and flow measurements at flow rates in the nanolitre to microlitre per minute range. Within the European Metrology Programme for Innovation and Research (EMPIR) project MFMET (20NRM02) [2] and its follow-up project MFMET II (24NRM03) [3], metrological procedures for the characterisation of microfluidic devices, including the measurement of their flow resistance, are being developed. First guidance on the evaluation of flow-related quantities in microfluidic devices has been published as EURAMET Technical Guide No. 4 [4], and MFMET II aims, among other things, to develop standard procedures for the characterisation of microfluidic devices and to establish guidelines for the quality control and validation of microfluidic systems. Recent work has addressed the measurement of the hydrodynamic resistance of microfluidic chips in a case study [5] and calibration techniques for micro- and nanoflow measurements [6], and the U.S. Food and Drug Administration has published a regulatory science tool for measuring the flow resistivity of microfluidics-based medical devices [7]. The present work contributes to these efforts.
For the fully developed laminar flow of Newtonian fluids, the velocity field in a straight channel is governed by the Poisson equation. Its classical series solution for rectangular ducts dates back to Boussinesq [8], and exact solutions of the Navier–Stokes equations for further cross-sections have been compiled by Berker [9]. Many textbooks on viscous flow state only the resulting series for the velocity profile and the flow rate; in [10], for example, the derivation is left as an exercise in which the series coefficients need not be determined. The influence of the cross-sectional shape on the hydraulic resistance of microchannels has been analysed by Mortensen et al. [11]. The continuum description based on the Navier–Stokes equations remains valid for liquid flows in microchannels of the size considered here [12], and the hydraulic resistances of common channel geometries are summarised by Bruus [13], who also gives a complete derivation of the series solution for rectangular channels based on a Fourier expansion of the pressure-gradient term. For non-circular cross-sections, the standard reference is the compilation by Shah and London [14], which lists friction factors and Nusselt numbers for fully developed laminar flow in rectangular, triangular, trapezoidal, elliptical, annular and many other ducts. Because fully developed laminar forced convection is governed by equations of the same elliptic (Poisson) type, the analytical and numerical methods used to determine the hydraulic resistance are largely the same as those used for heat transfer in ducts. Since the hydraulic resistance scales with the inverse fourth power of the characteristic channel dimension (and with the inverse third power of the smallest dimension in flat channels), small geometric deviations, for example due to manufacturing tolerances, have a large effect on the flow. The hydraulic resistance also depends on the temperature through the viscosity of the fluid; at 20 °C, the viscosity of water decreases by about 2.4 % per kelvin. Design calculations therefore have to be validated by accurate measurements.
The hydraulic resistance is used in the design of lab-on-a-chip (LoC) systems, droplet-based (multi-phase) microfluidics and OoC devices. In OoC devices, the hydraulic resistance of the channel network determines the flow rates and thus the wall shear stress and the nutrient supply, which have to match physiological conditions and keep the cells viable. Some of the liquids handled in these applications, such as blood or concentrated cell suspensions, are soft matter systems with non-Newtonian behaviour, for which the Newtonian description presented here serves as a reference case. In contrast, typical cell culture media are nearly Newtonian, so that the description applies to them directly.
This study combines a theoretical derivation of the hydraulic resistance of circular pipes and rectangular channels, starting from the Navier–Stokes equations, with an experimental validation on a polymer microfluidic chip. Two practical measurement methods, a pressure-driven flow controller with a flow sensor and a high-precision syringe pump with two external pressure sensors, are compared with each other and with the theoretical prediction. Preliminary measurement results obtained with both systems were presented at the XXIV IMEKO World Congress [1]. Compared with [1], the present work contributes (i) a self-contained step-by-step derivation of the hydraulic resistance of rectangular channels from the Navier–Stokes equations which, in contrast to [13], is based on the decomposition of the velocity field into the plane Poiseuille flow and a homogeneous correction term, including the determination of the series coefficients and an analysis of the convergence of the series; (ii) the demonstration that the Darcy–Weisbach equation with shape-dependent friction factors yields the same hydraulic resistance; (iii) a re-evaluation of the measurement data with a corrected theoretical value, a gain correction of the pressure sensors and a consideration of zero offsets; (iv) an uncertainty budget according to the Guide to the Expression of Uncertainty in Measurement (GUM) [15] and an assessment of the agreement by the normalised error; and (v) a discussion of the limitations and the range of application of the theoretical description and of the measurement methods, including manufacturing-related deviations from the rectangular cross-section, curved channels and deformable polydimethylsiloxane (PDMS) channels.
2. Theoretical Framework
The hydraulic resistance for the steady, incompressible, laminar flow of Newtonian fluids can be derived from the Navier–Stokes equation. For incompressible flows, the Navier–Stokes equations in cylindrical coordinates (r, , z) are solved together with the continuity equation (mass conservation).
The continuity equation for an incompressible fluid is expressed as:
The momentum equations for the radial direction (r-component), azimuthal direction (-component) and axial direction (z-component) are:
where:
- , , : velocity components in the radial, azimuthal, and axial directions;
- : fluid density (constant for incompressible flow);
- : dynamic viscosity;
- p: pressure;
- , , : body force components (e.g., gravity) in the respective directions.
Assuming that the flow is steady (all time derivatives are zero), there is no gravity, no radial flow (), and no swirl (), the equations are greatly simplified:
Equation (8) can be further simplified with Equation (5) and the assumption of an axially symmetric velocity profile ():
Since the left-hand side of Equation (9) depends only on z and the right-hand side only on r, both sides must be equal to the same constant, i.e., the pressure gradient is constant along the pipe. For a channel with a non-circular (e.g., rectangular) cross-section, the same assumptions applied to Equation (8), but without assuming axial symmetry, give the corresponding equation in Cartesian coordinates, with as the axial velocity:
Equations (9) and (10) are known as Poisson equations. For pipe flow, the Poisson equation reduces to an ordinary differential equation (ODE), whereas for rectangular channels, it reduces to a partial differential equation (PDE).
2.1. Mathematical Derivation for Hydraulic Resistance in (Cylindrical) Pipes
The starting point is Equation (9) and the assumption with :
Both sides are integrated twice. The first integration yields:
The second integration provides:
With the boundary conditions, symmetry at and is finite at , it follows that . Due to the no slip boundary condition at the wall, i.e., at , it follows that :
The volumetric flow rate can be calculated by integrating the velocity profile over the cross-section of the pipe, with :
This is the Hagen–Poiseuille equation, and the hydraulic resistance of a circular pipe is:
For circular pipes, the Hagen–Poiseuille equation provides an analytical solution. For rectangular channels, the velocity profile is determined using the Poisson equation. Analytical solutions for channels with a high aspect ratio approximate plane Poiseuille flow, whereas the exact solutions involve Fourier series. For more complex geometries, numerical methods such as the finite difference method (FDM), the finite element method (FEM) or the finite volume method (FVM) are used to solve the Poisson equation by discretizing the domain to calculate the velocity profile and integrating it to determine the volume flow and, consequently, the hydraulic resistance. Figure 1 shows the influence of the geometry on the velocity profile.
Figure 1.
Pipe and channel flows ( and ).
2.2. Mathematical Derivation for Hydraulic Resistance in Rectangular Channels
In contrast to the derivation in [13], in which the pressure-gradient term is expanded in a Fourier series, the velocity is decomposed here into the plane Poiseuille flow between two parallel plates and a homogeneous correction term, so that the physical meaning of the two parts of the flow rate in Equation (36) becomes directly visible. The starting point is Equation (10) and the assumption with C as the (constant) forcing function:
Solving a non-homogeneous linear PDE using the homogeneous (complementary) solution and a particular solution:
The cross-section of the channel is described by and , and the no-slip condition requires on all four walls. A simple particular solution which fulfils the Poisson equation and the boundary conditions on the vertical walls ( and ) is chosen:
For the homogeneous solution , a solution of the Laplace equation is sought that exactly eliminates the ‘errors’ of at the boundaries and :
The homogeneous solution is obtained by separation of variables, which decomposes the PDE into two ODEs:
Substituting this into the Laplace equation gives:
In order for at the side walls ( and ), must be an oscillation, therefore the negative sign is selected for the separation parameter.
From follows and from follows for
As the flow profile must be symmetrical, the cosine hyperbolic function is chosen around the axis of symmetry (), i.e., the sinh term vanishes (), and the product of the remaining constants is denoted by :
The solutions of the ODE result in:
With the boundary conditions from Equation (20) and since the homogeneous equation is a sum of sine terms, the parabola must also be represented as a Fourier sine series:
The coefficients follow from the Euler–Fourier integral:
Since for even n, only the odd terms contribute, and the sums are restricted to these terms from here on. For the determination of the coefficients the series of and the boundary condition are set equal. For , the following applies:
Because the cosh term is symmetric about , the boundary condition at is then fulfilled automatically. By comparing coefficients, one obtains:
This results in the following final homogeneous solution:
This results in the following complete solution:
To calculate the volume flow Q, the velocity profile must be integrated over the entire cross-sectional area A of the channel:
The result for the particular solution part, Equation (19), is:
This part corresponds to the flow between two infinitely extended plates (parabolic profile). The result for the homogeneous (complementary) fraction, Equation (31), is:
The volume flow can be calculated from this:
In the literature, w and h are often swapped so that h is the smaller dimension; this is common but not necessary:
With follows:
The hydraulic resistance of the rectangular channel (duct) follows directly:
The series in Equation (39) is exact for any aspect ratio. Choosing , i.e., with h as the smaller dimension, avoids the subtraction of two nearly equal large numbers in the bracket. Since , the terms decrease at least as , and the series converges very rapidly. For a square channel (), the bracket converges to (see Equation (40)). If only the first term () is taken into account, K is overestimated and is underestimated by 0.67 %; with two terms () and three terms (), the relative errors decrease to 0.061 % and 0.013 %, respectively. For flatter cross-sections (), the error of the first-term approximation is even smaller. All theoretical values in this work were calculated with 50 terms (relative truncation error below ).
If , the height and width can simply be swapped. A comparable form that already considers this and simplifies the sum over the odd numbers is as follows:
with:
The tubing in microfluidic setups usually has a circular cross-section, and the channels on microfluidic chips are mostly rectangular. The relations for circular pipes and rectangular channels therefore cover most microfluidic setups.
2.3. Alternative Method for the Determination of Hydraulic Resistance
An alternative theoretical approach is the Darcy–Weisbach equation, which is widely used in fluid dynamics and hydraulics. In turbulent flow, the friction factor has to be determined empirically, whereas in fully developed laminar flow it follows analytically from the Navier–Stokes equations:
where:
- : the dimensionless Darcy–Weisbach (friction) factor;
- : fluid density (constant for incompressible flow);
- : hydraulic diameter, ;
- u: mean flow velocity ();
- L: length of the channel.
2.3.1. Cylindrical Pipes
The Moody diagram plots for different values of relative pipe roughness against the Reynolds number . For fully developed laminar pipe flows, a fixed value of can be obtained. In this specific case, Equation (41) can be simplified with and to:
With and and , this gives:
For fully developed laminar pipe flow, the Darcy–Weisbach equation thus reduces to the Hagen–Poiseuille equation.
2.3.2. Rectangular Channels
The value of changes depending on the number and the cross-section shape of the conduit. Values of for rectangular, triangular, trapezoidal, elliptical and many other cross-sections have been compiled by Shah and London [14]; the values of for rectangular ducts are also summarised in standard textbooks, e.g., Table 6.4 in [16]. For very thin channels (low height h and large width w), a value of can be found. For square channels () the value is (Figure 2).
Figure 2.
Darcy friction factor–Reynolds number product () as a function of the aspect ratio (values according to [12,14,16]).
For intermediate values, there is a fifth-degree polynomial fit [14] to the analytical values, based on the aspect ratio between the height and width of the channel:
The equation is only valid for values between . If the height h is greater than the width w of the channel, the values for the calculation must be exchanged. This equation agrees with the analytically derived Equation (39) to within 0.1 % (maximum deviation of about 0.06 % for ).
Using , and , Equation (41) can be rearranged for an arbitrary cross-section into an expression for the hydraulic resistance:
with and for rectangular channels. For circular pipes (, , ), Equation (45) reduces to Equation (16). For channel no. 2 used in this study ( µm, mm, , mPa s), Equation (45) gives mbar h/mL, which is identical to the value obtained from the analytical Equation (40).
2.4. Validity and Limitations of the Theoretical Description
The relations derived above assume a steady, incompressible, fully developed laminar flow of a Newtonian fluid in a straight channel with rigid walls and the no-slip boundary condition. In a straight channel with fully developed flow, the convective terms of the Navier–Stokes equations vanish identically, so that the Poisson equation, Equation (17), is exact for laminar flow and not restricted to ; inertia only affects the entrance region, whose length is negligible for the channel considered here (Section 3.2).
In curved channels, the centrifugal force induces a secondary flow (Dean vortices), which is superimposed on the axial flow and increases the pressure drop. The relevant parameter is the Dean number , where is the radius of curvature of the channel centreline [17,18,19]. According to Dean’s perturbation solution for a curved pipe of circular cross-section [17], the reduction in the flow rate at a given pressure gradient compared with a straight pipe is proportional to for small Dean numbers; it amounts to about 0.4 % for and about 2 % for . The secondary flows exploited in inertial microfluidics occur at considerably higher Reynolds and Dean numbers [19]. As a rule of thumb, the theory for straight channels can therefore be applied to curved planar channels as long as , e.g., up to for , with L taken as the length of the channel centreline. In first approximation, the same criterion can also be applied to channels with other cross-sections and to three-dimensional (e.g., helical) channels. According to the manufacturer’s drawing, channel no. 2 used in this study is a meander with two 180° bends and two 90° bends (see Section 3.2), all with a radius of curvature of the centreline of mm; the nominal length of 40 mm corresponds to the length of the centreline. At the highest flow rate (), the Dean number is , so that the influence of the bends on is far below 0.1 % and can be neglected.
Further limitations are the assumption of a Newtonian fluid (for shear-thinning fluids such as blood, depends on the flow rate), the assumption of rigid walls (see Section 5.3 for deformable PDMS channels), deviations of the real cross-section from the ideal rectangle (Section 3.2) and the temperature dependence of the viscosity. Reported slip lengths for liquids range from tens of nanometres to about 1 µm, the larger values for coated, hydrophobic surfaces [12]; for the uncoated channel considered here, wall slip is neglected, and the agreement between measured and theoretical values does not indicate a significant slip effect.
3. Materials and Methods
This study compares two methods for measuring the hydraulic resistance of a microchannel on a microfluidic test chip (Section 3.2) with each other and with the theoretical value derived in Section 2.
Pressure-driven flow controllers apply a controlled gas pressure to a sealed liquid reservoir, which drives the liquid through the connected fluidic system. If a flow sensor is integrated, the inlet pressure is adjusted automatically until the target flow rate is reached, so that the pressure required for a given flow rate is directly available.
High-precision syringe pumps generate accurate flow rates, particularly at low flow rates and volumes, but do not measure the pressure. To determine the hydraulic resistance, pressure sensors must therefore be installed upstream and downstream of the microfluidic chip.
3.1. Experimental Setup
3.1.1. Pressure-Driven Flow Controller System
As shown in Figure 3, a pressure-driven flow controller system from the Microfluidics Innovation Center (MIC, France) was used for the investigation. It consisted of a prototype pressure-driven flow controller and a commercially available Elveflow Microfluidic Flow Sensor (MFS), model MFS2 (Elveflow (Elvesys), Paris, France). The controller is supplied with gas from an external pressurised gas cylinder and regulates the gas pressure above the sealed liquid reservoir in a closed loop with a proportional solenoid valve and an integrated pressure sensor. According to the manufacturer, the pressure range is 0 to 2000 mbar, the accuracy of the pressure control (deviation from the set point at steady state) is 2.5 mbar, i.e., 0.125 % of the full scale, and the response time of the regulator is about 50 ms, depending on the pneumatic system. The prototype used here has the same characteristics as the commercial version of the controller. In flow-control mode, the pressure is readjusted continuously on the basis of the MFS signal; the specified flow range of the system is 0.4 to 1000 µL/min with an accuracy of 5 % of the reading, depending on the flow sensor used.
Figure 3.
Pressure-driven flow controller of the Microfluidics Innovation Center (MIC) with an Elveflow MFS2 flow meter.
The MFS2 is a thermal mass flow meter (TMFM) time-of-flight (ToF) conductivity flow sensor designed for a flow rate range of up to 7 µL/min. The liquid is heated locally in a capillary inside the sensor, and the flow rate is determined from the temperature difference measured upstream and downstream of the heating element. Since this relationship depends on the properties of the liquid, the sensor must be calibrated for the liquid used. The MIC system was operated using the Elveflow Smart Interface (ESI) software, in which the target values were set.
3.1.2. High-Precision Syringe Pump
As shown in Figure 4, the volume flow rates were generated using a high-precision syringe pump (Cetoni neMESYS Base 120 controller and neMESYS Low Pressure Module 290N, CETONI GmbH, Korbussen, Germany) in conjunction with a calibrated 1 mL glass syringe; with its integrated encoder, the pump can be regarded as a type of piston prover. For the pressure drop measurements, two Cetoni pressure sensors with a flat ceramic membrane and a nominal pressure of 500 kPa (5.0 bar) were used to measure the pressure upstream and downstream of the microfluidic chip. The sensors were connected to the pump via a Cetoni input/output (I/O) splitter, and the pressure values were recorded together with the volume flow rate (dosed amount) using the Cetoni Qmix Elements software. The measurement setup consisted of the syringe pump, the MFS2 flow sensor of the MIC system (used to display the current volume flow rate), the upstream pressure sensor, the microfluidic chip and the downstream pressure sensor; the measuring line ended at an infusion needle that protruded into a beaker placed on a weighing scale (Mettler Toledo XPR10 ultra-micro comparator scale, Mettler Toledo, Greifensee, Switzerland).
Figure 4.
Cetoni neMESYS Base 120 and neMESYS Low Pressure module 290N with two external pressure sensors.
3.2. Microfluidic Chip
The measurements were performed on a microfluidic chip made of TOPAS (Thermoplastic Olefin Polymer with Amorphous Structure), as illustrated in Figure 5. TOPAS is a thermoplastic polymer widely used for microfluidic chips because of its high transparency, chemical resistance and biocompatibility. The microfluidic chip was manufactured by microfluidic ChipShop (Germany) for the EMPIR project MFMET (20NRM02) as a test chip and has eight channels with dimensions down to 5 µm × 5 µm; the theoretical and metrological work presented here was carried out within the follow-up project MFMET II (24NRM03). As indicated in Table 1, given the maximum pump pressure of 2000 mbar, only channels no. 1, 2, and 5 were suitable for the intended flow rate range. Channel no. 2 was selected for the measurements because of the resulting pressure drops at the given flow rates. At the highest flow rate, the Reynolds number in channel no. 2 is about , i.e., the flow is clearly laminar. At such low Reynolds numbers, the hydrodynamic entrance length (about 35 µm) is of the order of the hydraulic diameter (50 µm) and therefore negligible compared with the channel length of 40 mm, so that the assumption of fully developed flow is justified.
Figure 5.
Microfluidic polymer chip with 8 different channels, chip dimensions 75.5 mm × 25.5 mm and Mini-Luer access ports.
Table 1.
Dimensions of the 8 channels on the microfluidic chip.
In addition to the values listed in Table 1, each channel on the microfluidic chip had a direct connection between the inlet and outlet ports (port-to-port connection) with a height and width of approximately 500 µm. Due to its dimensions, the pressure drop across this direct connection is negligible for the given flow rates.
The real channel geometry can deviate from the ideal rectangular cross-section assumed in Section 2. Polymer chips are typically fabricated by moulding processes such as injection moulding or hot embossing, which require a draft angle of typically 3–5° for demoulding. As a result, the channels usually have a slightly trapezoidal cross-section, and the corners of the channels are usually rounded [20]. For such cross-sections, an analytical solution for is no longer possible. Numerically determined friction factors for trapezoidal ducts with sharp corners are tabulated in [14]. For the actual geometry, including rounded corners, can be determined numerically, e.g., with the finite difference method (FDM), the finite element method (FEM) or the finite volume method (FVM); approximate analytical models for laminar flow in microchannels of arbitrary cross-section are also available [21]. The actual cross-section of the channels should therefore be checked before measured and theoretical values are compared.
4. Results
The measurement data evaluated in this section were first published in [1]. Here, they are re-evaluated using the updated theoretical value (Section 2.2), a correction of the Cetoni pressure sensors, a consideration of the zero offsets of the pressure measurement and the uncertainty analysis in Section 5. The MFS used in both measurement setups came with two pre-calibrated factory settings, one for water and the other for isopropyl alcohol, each at 23 °C. All measurements in this study were performed at a laboratory temperature of (20.0 ± 0.5) °C. The temperature of (23.0 ± 0.5) °C stated in [1] for these measurements is incorrect. The theoretical value is therefore calculated with the viscosity of water at 20 °C. The supplied MFS2 has a flow specification of 0.42 µL/min to 7 µL/min (0.0252 mL/h to 0.42 mL/h). To cover a wide flow range, three flow rates (0.025 mL/h, 0.100 mL/h, and 0.400 mL/h) with a factor of 4 between each value were selected for the measurement campaign. Prior to the measurements, the MFS was calibrated at these three points, that is, at the same flow rates used for the hydraulic resistance measurements. The RISE Nanoflow calibration facility, which is based on the dynamic weighing method, was used for flow calibrations with ultrapure degassed water [22]. As shown in Table 2, there were large measurement deviations. Part of these deviations may be due to the difference between the measurement temperature (20 °C) and the reference temperature of the factory calibration (23 °C), since the response of a thermal flow sensor depends on the thermophysical properties of the liquid. Since the deviations change sign over the flow rate range, this temperature difference is unlikely to be the only cause. In addition, at the highest set point, the actual flow rate of about 0.45 mL/h exceeded the specified measuring range of the MFS2. The new calibration data were not implemented in the software; all measurements were performed with the preset factory calibration.
Table 2.
Calibration of the MFS using the MIC system as a flow generator.
The measurements were performed using channel no. 2 of the microfluidic chip. For measurements using the pressure-driven flow controller and high-precision syringe pump systems, the value displayed by the MFS was used to set the target flow rate, whereas the reference flow rates were determined gravimetrically (MIC system) and with the calibrated syringe (Cetoni system). The MIC system adjusts the flow rate on the basis of the MFS signal. For the Cetoni system, the flow rate in the Qmix Elements software was set such that the MFS display in the ESI software showed the target flow rate, so that both systems were operated at comparable flow rates. With both systems, each measurement series was performed twice, and each set point was approached twice. After a stabilisation time of 5 min, the pressure and flow values were averaged over 20 min. The values given in Table 3 and Table 4 are the mean values of these repetitions; since the individual values are no longer available, the repeatability is taken into account through the scatter of the mean values around the regression line (Section 5.1).
4.1. Results with Pressure-Driven Flow Controller System
The advantage of the MIC system is that the pressure required for a certain flow rate is directly available as a value provided by the system. However, this pressure includes the pressure drop over the entire measurement setup, so two measurements were required for each flow rate: one with channel no. 2 of the microfluidic chip included in the measurement setup and one with the tubing connected directly via the port-to-port connection, whose pressure drop is negligible (Section 3.2). Since the measurement setup and the back pressure were otherwise identical in both configurations, the difference between the two inlet pressures corresponds to the pressure drop over channel no. 2. Table 3 shows the pressure values for both configurations. Since the two measurements were not performed at exactly the same flow rates, the pressure difference was assigned to the average of the respective reference flow rates from the measurements against the weighing scale. As can be seen in Figure 6, the measured value ( mbar h/mL) agrees well with the theoretical value ( mbar h/mL; calculated with Equation (40) for µm, µm, mm and the dynamic viscosity of water at the measurement temperature of 20 °C, mPa s).
Figure 6.
Comparison of the measurement results obtained using the MIC system with the theoretical value. Error bars: expanded uncertainty of the measured hydraulic resistance (coverage probability 95 %, ); dashed lines: theoretical value (, ).
Table 3.
Measurement results obtained using the pressure-driven flow controller (MIC) system (data from [1]).
4.2. Results with High-Precision Syringe Pump
With the Cetoni system, the pressure drop was measured directly over channel no. 2 of the microfluidic chip, so that only one measurement per flow rate is required. As shown in Table 4, the Cetoni pump flow rate must be set to about 0.450 mL/h for the MFS to display a flow rate of 0.400 mL/h. According to the MFS log file, the mean value was about 0.400 mL/h in these cases. These values also corresponded well with the MFS measurements against the weighing scale, as shown in Table 2. In addition, the pressure drop over the port-to-port connection was measured. The pressure sensors have a linear voltage output from 0.5 V to 4.5 V, a common industrial standard. With a resolution of 1.25 mbar (12-bit, 4096 discrete values), the pressure sensors are not suitable for resolving such small pressure differences. The measured values of up to 2.2 mbar (Table 4) are nevertheless considerably larger than the theoretical pressure drop of the port-to-port connection itself (less than 0.01 mbar at 0.45 mL/h); they mainly reflect the pressure losses of the tubing, fittings and ports between the two pressure sensors and the limited resolution of the sensors. The measured pressure values were corrected for the gain error of the pressure sensors, which was determined in a calibration against a reference pressure standard and amounts to about +0.85 % in the relevant pressure range of 0 to 0.25 bar; the zero offsets of the sensors were set to zero by taring at zero flow before the measurements. The residual zero offsets are discussed in Section 4.3. Table 4 and Figure 7 summarize the pressure drop measurements obtained using the Cetoni setup and channel no. 2 of the microfluidic chip. The result ( mbar h/mL) agrees well with the theoretical value ( mbar h/mL). The measured value is slightly higher, which can largely be explained by the pressure losses between the two sensors mentioned above. These losses are also contained in the measurement over channel no. 2 and amount to about 1 % of the measured pressure drop; if the port-to-port values are subtracted, as in the evaluation of the MIC measurements, a hydraulic resistance of 507.7 mbar h/mL is obtained. A remaining difference could indicate that the actual channel cross-section (width and/or height) was slightly smaller or the channel slightly longer than specified, for example due to the manufacturing-related deviations from the rectangular cross-section discussed in Section 3.2.
Figure 7.
Comparison of the measurement results obtained using the Cetoni system with the theoretical value. Error bars: expanded uncertainty of the measured hydraulic resistance (coverage probability 95 %, ); dashed lines: theoretical value (, ).
Table 4.
Measurement results obtained using the high-precision syringe pump (Cetoni) system (data from [1]; Δp values corrected for the gain error of the pressure sensors).
4.3. Deviations at Low Flow Rates
At the lowest flow rate, the measured pressure drops of both systems are considerably higher than the theoretical values, by about 56 % (MIC) and 47 % (Cetoni), whereas the relative deviations decrease with increasing flow rate (MIC: +3.9 % and −0.1 %; Cetoni: +13.0 %, +8.9 %, +1.2 % and −0.8 %). In absolute terms, the excess pressure at the lowest flow rate is only about 7 mbar (MIC) and 6 mbar (Cetoni). An approximately constant excess pressure of a few mbar indicates a zero offset of the pressure measurement rather than a flow-dependent effect: a constant offset leads to a relative deviation of , which is large at small flow rates and decreases with increasing flow rate. Such offsets are not completely eliminated by taring. The zero readings of the Cetoni pressure sensors recorded at zero flow during the measurement campaign varied by about 1 mbar for the upstream sensor and by up to about 13 mbar for the downstream sensor. For the MIC system, the pressure drop is the difference in two inlet pressures measured at different times, so that a drift of the pressure readout between the two measurements has the same effect.
In a regression through the origin, the slope is dominated by the measurements at higher flow rates, where the relative effect of an offset is small. As a plausibility check, a linear regression with intercept, including the zero-flow points, was carried out. It yields slopes of 499.3 mbar h/mL (MIC; intercept 3.2 mbar) and 497.0 mbar h/mL (Cetoni; intercept 5.3 mbar), which deviate from the theoretical value by −1.4 % and −1.9 %, respectively, and agree with the results of the regression through the origin within the uncertainties. The regression through the origin is retained as the main result, since the pressure drop vanishes at zero flow by definition; the residual zero offset is taken into account as an uncertainty contribution (Section 5.1). For future measurements, zero-point measurements before and after each set point and a larger number of flow rates are recommended.
Measuring the hydraulic resistance of a microfluidic chip or system with a pressure-driven flow controller and a flow sensor requires less effort than the conventional method with a high-precision syringe pump (or prover) and pressure sensors, and its accuracy is at least comparable.
5. Discussion
5.1. Measurement Uncertainty
For a metrological assessment of the agreement between the measured and the theoretical hydraulic resistance, the uncertainties of both values have to be considered. The measured hydraulic resistance follows from the model equation . Assuming uncorrelated input quantities, its relative standard uncertainty is given by:
For the theoretical value of a square channel (), Equation (39) yields the relative sensitivity coefficients , and . Assuming uncorrelated input quantities, the relative standard uncertainty of the theoretical value is therefore:
The viscosity contribution is dominated by the uncertainty of the fluid temperature: at 20 °C, a temperature uncertainty of 1 K corresponds to a relative uncertainty of the viscosity of water of about 2.4 %. If width and height of the channel are determined by the same manufacturing process, their deviations may be correlated, which increases the geometric contribution. The individual contributions are summarised in Table 5.
Table 5.
Uncertaintycontributions to the measured and theoretical hydraulic resistance of channel no. 2 (relative standard uncertainties; theoretical value for a maximum deviation of ±1 µm of width and height).
The geometric uncertainty is estimated from an assumed maximum deviation of ±1 µm of the channel width and height (rectangular distribution, µm). The manufacturer did not disclose tolerances for the channel dimensions, and the cross-section of the channel was not measured. The assumption is based on the fact that channels as small as 5 µm × 5 µm were fabricated on the same chip, but it is not a verified tolerance. To show the influence of this assumption, the evaluation was repeated with a maximum deviation of ±2 µm: the expanded uncertainty of the theoretical value then increases from 6.7 % to 13.1 % (from 9.3 % to 18.5 % for fully correlated deviations of width and height), and the conclusions on the agreement with the measured values remain unchanged (). A dimensional measurement of the channel cross-section, e.g., by optical profilometry or confocal microscopy, is recommended to reduce this contribution. For the channel length, a maximum deviation of ±0.1 mm is assumed. For the fluid temperature, the temperature uncertainty of the laboratory of ±0.5 K is used, which is considered conservative, since each measurement series took only a short time. In addition, the draft angle required for moulding (Section 3.2) leads to a trapezoidal cross-section. Numerical calculations (FDM and FEM) for a draft angle of 5° show that the hydraulic resistance of channel no. 2 differs by only about 0.3 % from that of the ideal rectangular channel if the nominal width corresponds to the mean channel width, whereas deviations of about +21 % or −15 % would result if the nominal width corresponded to the top or bottom width of the channel, respectively. The measured values therefore indicate that the effective cross-section of channel no. 2 corresponds to the nominal cross-section within the stated uncertainty.
On the measurement side, the measured hydraulic resistance is the slope of the regression line through the origin, . The following contributions are considered (Table 5). The gain error of the Cetoni pressure sensors was corrected; the remaining uncertainty of this correction, including a possible long-term drift since the calibration in 2021, is estimated as ±0.5 % (rectangular distribution). For the pressure readout of the MIC system, no accuracy specification is available; the manufacturer only specifies the accuracy of the pressure control of 2.5 mbar (0.125 % of the full scale). This value is used as an estimate and taken into account as a rectangular distribution for each of the two inlet pressures ( mbar), which are assumed to be uncorrelated, i.e., 2.04 mbar for their difference; its effect on the slope of the regression line amounts to 0.87 %. The resolution of the Cetoni pressure sensors of 1.25 mbar is taken into account by a rectangular distribution with a standard uncertainty of mbar per sensor, i.e., 0.51 mbar for the difference in the two sensors; its effect on the slope, , amounts to 0.17 %. For the MIC system, the resolution of the pressure readout is not known, and its effect is contained in the scatter of the measured values. Zero offsets of the pressure measurement are not completely eliminated by taring (Cetoni) or by the difference in two measurements (MIC), as shown in Section 4.3. A residual offset changes the slope of the regression line through the origin by . The intercept of the regression with intercept (3.2 mbar for MIC and 5.3 mbar for Cetoni) is used as the half-width of a rectangular distribution of the residual offset, which results in relative standard uncertainties of 0.97 % (MIC) and 1.86 % (Cetoni). In addition, the pressure drops of the port-to-port and channel configurations of the MIC system were measured at slightly different flow rates, which leads to an overestimation of the resulting hydraulic resistance of up to about 1.6 %. This is taken into account as an additional uncertainty contribution (rectangular distribution). The flow rate reference contributes 0.25 % (, , valid for flow rates of 500 nL/min and above [22]) for both systems; at the lowest flow rate (about 400 nL/min), the uncertainty of the reference is slightly larger (up to 1 %), which has a negligible effect on the slope of the regression line.
The repeatability is evaluated from the scatter of the measured values around the regression line (Type A evaluation). Since each value in Table 3 and Table 4 is the mean of the repeated measurements, the standard uncertainty of the slope, with the residual standard deviation , contains both the repeatability and deviations from linearity. It is the largest contribution on the measurement side and amounts to 2.2 % for the MIC system ( flow rates, degrees of freedom) and 2.0 % for the Cetoni system (, ). Because of the small number of degrees of freedom, a coverage factor of is not justified for the measured values. The effective degrees of freedom are therefore calculated with the Welch–Satterthwaite formula [15], , treating the Type B contributions as having infinite degrees of freedom. This gives (MIC) and 14.5 (Cetoni) and, for a coverage probability of approximately 95 % (95.45 %), coverage factors of (MIC) and (Cetoni) from the t-distribution. For the theoretical value, all contributions are Type B estimates with rectangular distributions, and is used.
The agreement between measured and theoretical values is assessed by the normalised error
where U denotes the expanded uncertainties for a coverage probability of approximately 95 %; indicates agreement within the stated uncertainties. For the MIC system, , and for the Cetoni system, (0.12 if the deviations of width and height are assumed to be fully correlated). Both measurement methods therefore agree with the theoretical value within the stated uncertainties.
5.2. Comparability of the Two Measurement Methods
The two methods do not determine exactly the same pressure-drop boundaries. With the MIC system, the pressure drop over channel no. 2 is obtained as the difference between the inlet pressures with the channel and with the port-to-port connection. All other pressure losses of the setup (tubing, fittings, flow sensor and outlet) cancel, and the result refers to the channel between the chip ports, reduced by the negligible pressure drop of the port-to-port connection. With the Cetoni system, the pressure difference is measured directly with two sensors, but it also includes the pressure losses of the tubing, fittings and chip ports between the sensors. These additional losses were measured with the port-to-port connection (0.4 mbar to 2.2 mbar, Table 4) and amount to about 1 % of the pressure drop over channel no. 2. If they are subtracted, both methods refer to the same boundaries and yield practically identical results (507.7 mbar h/mL for Cetoni and 507.8 mbar h/mL for MIC; for both with respect to the theoretical value). The second difference between the methods, the slightly different flow rates of the two MIC configurations, is considered as an uncertainty contribution (Table 5). For comparisons between different methods or laboratories, the measurand, in particular the pressure-drop boundaries, should therefore be defined explicitly.
5.3. Range of Application
The measuring range of the methods is determined by the flow range of the flow sensor or the syringe pump and by the measurable pressure drop. With the MFS2 flow sensor (0.025 mL/h to 0.42 mL/h), a maximum pressure of 2000 mbar and a minimum pressure drop of about 60 mbar, for which a residual zero offset of a few mbar contributes only a few percent, hydraulic resistances between about 140 mbar h/mL and 80 000 mbar h/mL can be measured with the present setups. Of the channels on the chip (Table 1), channel no. 2 (507 mbar h/mL) and channel no. 5 (253 mbar h/mL) are within this range at the higher flow rates, and channels no. 3 ( mbar h/mL) and no. 6 ( mbar h/mL) can be measured at flow rates up to about 0.1 mL/h and 0.2 mL/h, respectively. For channel no. 1 (32 mbar h/mL), the pressure drop is only about 1 mbar to 13 mbar, which requires higher flow rates, i.e., a flow sensor with a higher measuring range, or a differential pressure sensor with a considerably better resolution and zero stability. Channels no. 4, 7 and 8 ( mbar h/mL to mbar h/mL) require flow rates below 0.025 mL/h (0.4 µL/min), i.e., a flow sensor with a lower measuring range or a syringe pump with a gravimetric reference [22], and higher pressures. For these small channels, a deviation of 1 µm in width and height changes by several tens of percent, so that the measured hydraulic resistance is primarily a sensitive measure of the actual cross-section. In this study, channel no. 2 was measured on one chip only. Chip-to-chip variations were therefore not assessed experimentally and are only indirectly covered by the geometric uncertainty of the theoretical value; measurements on several chips of the same batch are recommended.
Many microfluidic devices are made of polydimethylsiloxane (PDMS), whose Young’s modulus (about 1 MPa to 3 MPa) is about three orders of magnitude lower than that of TOPAS (about 2.6 GPa to 3 GPa). Under pressure, PDMS channels deform, and the hydraulic resistance decreases with increasing pressure, which results in a non-linear relationship between pressure drop and flow rate [23,24,25]. For shallow channels, the deformation of the channel wall scales with , where p is the local gauge pressure, w the channel width and E Young’s modulus [23,24]. For a channel with the dimensions of channel no. 2 in PDMS ( MPa) and a local pressure of 200 mbar, this gives a wall displacement of the order of 0.5 µm, i.e., about 1 % of the channel height at the inlet, decreasing towards the outlet. Since , this corresponds to a decrease in the local resistance of about 3 % at the inlet and of the order of 1 % to 2 % averaged over the channel length. For shallow channels with large aspect ratios (), the relative deformation and thus the effect on are correspondingly larger and can reach several tens of percent. For TOPAS, the corresponding displacement is below 1 nm and negligible. The measurement methods presented here can in principle be applied to PDMS channels, but the measured hydraulic resistance then depends on the pressure level. For such channels, we recommend measuring at several flow rates and checking the linearity of , reporting together with the pressure range or as the limit for , which can be compared with the rigid-wall theory, and keeping the back pressure low, because the deformation depends on the absolute pressure in the channel as well as on the pressure difference.
6. Conclusions and Outlook
This study investigates the hydraulic resistance, which determines the flow control in microfluidic and organ-on-chip systems, by combining theoretical derivations with experimental validation. The theoretical framework is based on the Navier–Stokes equations, which are reduced to a Poisson form under standard microfluidic assumptions. For circular pipes, this results in the Hagen–Poiseuille equation and a hydraulic resistance in closed form, whereas for rectangular channels, the Poisson equation is solved using a particular solution plus a homogeneous Laplace solution obtained by separating the variables, resulting in a Fourier series velocity profile and a corresponding series expression for hydraulic resistance. As an alternative approach, the Darcy–Weisbach equation is presented and shown to reduce to the Hagen–Poiseuille equation for pipe flow. For rectangular channels, the hydraulic resistance values obtained using shape-dependent friction factors agreed with the analytically derived values. Experimentally, two methods for measuring hydraulic resistance on a polymer microfluidic chip were compared, whereby a channel was selected based on theoretical considerations. A pressure-driven flow controller combined with a flow sensor was compared with a high-precision syringe pump using two external pressure sensors for pressure measurement. Both setups used the flow sensor display to set comparable flow rates. Using the pressure-driven flow controller system, the hydraulic resistance of the microchannel was determined by measuring the pressure in two configurations: with and without the chip. This resulted in a measured hydraulic resistance that was in very good agreement with the theoretical value. With the high-precision syringe pump system, a single direct chip pressure drop measurement per flow rate was sufficient. The hydraulic resistance determined was slightly higher but also agreed well with the theoretical value; after subtraction of the pressure losses between the sensors, both methods yielded practically identical results. Both measured values agree with the theoretical value within the expanded uncertainties (7.2 % for MIC, 6.1 % for Cetoni and 6.7 % for the theoretical value; ). On the measurement side, the largest contributions result from the scatter of the measured values and from the residual zero offset of the pressure measurement, which dominates the deviations at low flow rates; zero-point measurements before and after each set point and a larger number of flow rates are therefore recommended. Overall, the study shows that both methods can be used to measure the hydraulic resistance of OoC and microfluidic platforms with comparable accuracy, and that the pressure-driven flow controller requires less effort. Future work will address non-rectangular channel cross-sections resulting from the manufacturing process, chip-to-chip variations, deformable (PDMS) channels, non-Newtonian fluids and the transfer of the methods to thermal-fluid applications such as microchannel cooling.
Author Contributions
Conceptualization, O.B., K.S., E.B. and R.M.; methodology, O.B., K.S., E.B. and R.M.; software, O.B.; validation, O.B., K.S., E.B. and R.M.; formal analysis, O.B., E.B. and R.M.; investigation, O.B. and K.S.; resources, O.B., K.S., E.B. and R.M.; data curation, O.B., E.B. and R.M.; writing—original draft preparation, O.B. and E.B.; writing—review and editing, K.S. and R.M.; visualization, O.B.; project administration, E.B.; funding acquisition, O.B. and E.B. All authors have read and agreed to the published version of the manuscript.
Funding
The project 20NRM02 MFMET has received funding from the EMPIR programme co-financed by the Participating States and from the European Union’s Horizon 2020 research and innovation programme. The project 24NRM03 MFMET II has received funding from the European Partnership on Metrology, co-financed from the European Union’s Horizon Europe Research and Innovation Programme and by the Participating States. The pressure-driven flow controller was developed in the Horizon 2020 project “Tumor and Lymph Node on Chip for cancer studies” (TUMOR-LN-oC), which has received funding from the European Union’s Horizon 2020 research and innovation programme under grant agreement No. 953234.
Data Availability Statement
The data presented in this study are available on request from the corresponding author.
Acknowledgments
A preliminary version of this work was presented at the 20th International Flow Measurement Conference (FLOMEKO 2026), held in Nara, Japan, on 17–20 May 2026. Since the conference paper was not published in the official proceedings, the conference organisers agreed that the work could be submitted for publication in a scientific journal. The authors used Claude (Anthropic) for language editing and shortening of the text and the conversion of the manuscript into the MDPI template. All output was reviewed and verified by the authors, who take full responsibility for the content.
Conflicts of Interest
The authors declare no conflicts of interest.
References
- Büker, O.; Stolt, K.; Batista, E.; Sepulveda, J.; Betterelli Giuliano, C.; Morin, A.; Nobre, D.; Navas, H. Investigations on the Performance of a Newly Developed Pressure-Driven Flow Controller Used in Microfluidic Applications. Meas. Sens. 2025, 38, 101645. [Google Scholar] [CrossRef] [Scilit]
- 20NRM02 MFMET Project. Establishing Metrology Standards in Microfluidic Devices. Available online: https://mfmet.eu/ (accessed on 27 October 2026).
- EURAMET. Establishing Metrology Standards in Microfluidic Devices II (24NRM03 MFMET II). Available online: https://www.euramet.org/research-innovation/search-research-projects/details/project/establishing-metrology-standards-in-microfluidic-devices-ii (accessed on 27 October 2026).
- EURAMET e.V., Technical Committee for Flow. Evaluation of Flow-Related Quantities in Microfluidic Devices; EURAMET Technical Guide No. 4, Version 1.0; EURAMET e.V.: Braunschweig, Germany, 2024. [Google Scholar]
- Ogheard, F.; Daugbjerg, T.S.; Sluše, J.; Geršl, J.; Wippold, J.A.; Silverio, V. Assessing Hydrodynamic Resistance in Microfluidics: A Case Study. Preprints 2024, 2024060049. [Google Scholar] [CrossRef] [Scilit]
- Batista, E.; Martins, R.F.; Silverio, V.; Godinho, I. Advancing Calibration Techniques for Accurate Micro and Nanoflow Measurements. Front. Nanotechnol. 2025, 7, 1600426. [Google Scholar] [CrossRef] [Scilit]
- U.S. Food and Drug Administration. Measuring Flow Resistivity in Microfluidics-Based Medical Devices; Regulatory Science Tool RST24MF01.01; U.S. Food and Drug Administration: Silver Spring, MD, USA, 2025. Available online: https://cdrh-rst.fda.gov/measuring-flow-resistivity-microfluidic-based-medical-devices (accessed on 13 November 2026).
- Boussinesq, J. Mémoire sur l’influence des frottements dans les mouvements réguliers des fluides. J. Math. Pures Appl. 1868, 13, 377–424. [Google Scholar]
- Berker, R. Intégration des équations du mouvement d’un fluide visqueux incompressible. In Fluid Dynamics II, Encyclopedia of Physics; Flügge, S., Truesdell, C., Eds.; Springer: Berlin/Göttingen/Heidelberg, Germany, 1963; Volume VIII/2, pp. 1–384. [Google Scholar]
- White, F.M.; Majdalani, J. Viscous Fluid Flow, 4th ed.; McGraw Hill: New York, NY, USA, 2022; Section 3-3.3 and Problem 3-38. [Google Scholar]
- Mortensen, N.A.; Okkels, F.; Bruus, H. Reexamination of Hagen–Poiseuille Flow: Shape Dependence of the Hydraulic Resistance in Microchannels. Phys. Rev. E 2005, 71, 057301. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Sharp, K.V.; Adrian, R.J.; Santiago, J.G.; Molho, J.I. Liquid Flows in Microchannels. In MEMS: Introduction and Fundamentals, 2nd ed.; Gad-el-Hak, M., Ed.; CRC Press/Taylor & Francis: Boca Raton, FL, USA, 2005; pp. 10-1–10-46. [Google Scholar]
- Bruus, H. Theoretical Microfluidics; Oxford University Press: Oxford, UK, 2008. [Google Scholar]
- Shah, R.K.; London, A.L. Laminar Flow Forced Convection in Ducts: A Source Book for Compact Heat Exchanger Analytical Data; Academic Press: New York, NY, USA, 1978. [Google Scholar]
- JCGM. Evaluation of Measurement Data—Guide to the Expression of Uncertainty in Measurement; JCGM 100:2008; BIPM: Sèvres, France, 2008. [Google Scholar]
- White, F.M.; Xue, H. Fluid Mechanics, 9th ed.; McGraw Hill: New York, NY, USA, 2021; Section 6.8 and Table 6.4. [Google Scholar]
- Dean, W.R. The Stream-Line Motion of Fluid in a Curved Pipe. Lond. Edinb. Dubl. Phil. Mag. J. Sci. 1928, 5, 673–695. [Google Scholar] [CrossRef] [Scilit]
- Dean, W.R. Note on the Motion of Fluid in a Curved Pipe. Lond. Edinb. Dubl. Phil. Mag. J. Sci. 1927, 4, 208–223. [Google Scholar] [CrossRef] [Scilit]
- Di Carlo, D. Inertial Microfluidics. Lab Chip 2009, 9, 3038–3046. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- van Heeren, H.; Hewkin, P. (Eds.) Design for Microfluidic Device Manufacture Guidelines; Version 5; Microfluidics Consortium: Cambridge, UK, 2014; Available online: https://www.microfluidicsinfo.com/wp-content/uploads/2017/08/DesignforManufacture-1.pdf (accessed on 27 October 2026).
- Bahrami, M.; Yovanovich, M.M.; Culham, J.R. Pressure Drop of Fully-Developed, Laminar Flow in Microchannels of Arbitrary Cross-Section. In Proceedings of the 3rd International Conference on Microchannels and Minichannels (ICMM2005), Toronto, ON, Canada, 13–15 June 2005. Paper ICMM2005-75109. [Google Scholar]
- Büker, O.; Stolt, K. RISE Test Facilities for the Measurement of Ultra-Low Flow Rates and Volumes with a Focus on Medical Applications. Appl. Sci. 2022, 12, 8332. [Google Scholar] [CrossRef] [Scilit]
- Gervais, T.; El-Ali, J.; Günther, A.; Jensen, K.F. Flow-Induced Deformation of Shallow Microfluidic Channels. Lab Chip 2006, 6, 500–507. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Christov, I.C.; Cognet, V.; Shidhore, T.C.; Stone, H.A. Flow Rate–Pressure Drop Relation for Deformable Shallow Microfluidic Channels. J. Fluid Mech. 2018, 841, 267–286. [Google Scholar] [CrossRef] [Scilit]
- Turcitu, T.; Armstrong, C.J.K.; Lee-Yow, N.; Salame, M.; Le, A.V.; Fenech, M. Comparison of PDMS and NOA Microfluidic Chips: Deformation, Roughness, Hydrophilicity and Flow Performance. Micromachines 2023, 14, 2033. [Google Scholar] [CrossRef] [Scilit] [PubMed]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |