Development and High-Count-Rate Performance of Multi-Channel HPGe Spectrometers for X- and Gamma-Ray Spectroscopy

,
,
,
,
,
,
and
Baltic Scientific Instruments, LV-1005 Rīga, Latvia
*
Author to whom correspondence should be addressed.

Abstract

This paper presents the development and comparative investigation of multi-detector HPGe spectrometers for operation at high count rates, with emphasis on K-series X-ray lines of rare-earth and heavy elements in X-ray fluorescence analysis. Two detector architectures were developed and tested: a six-channel multi-crystal assembly based on individual P-type semi-planar HPGe detectors and a six-strip segmented planar HPGe detector with amorphous-germanium contacts. The segmented detector with cooled FETs demonstrated energy resolutions of 560 eV at 5.9 keV, 700 eV at 59.5 keV, and 1.29 keV at 662 keV at low count rates. At 661.7 keV, the FWHM increased to 3.2–3.5 keV at input count rates of 40–57 kcps per segment. The multi-crystal assembly achieved average FWHM values of 370 eV at 14 keV, 450 eV at 59.5 keV, 560 eV at 122 keV, and 1.16 keV at 662 keV at low count rates. At maximum throughput, FWHM values at 59.5 keV were 580 and 805 eV under 238Pu and 137Cs loading, respectively (ICR 100–120 kcps; OCR 57 and 46 kcps per channel). The 600 eV target was not consistently met. The multi-crystal assembly also recorded preliminary tube-excited K-line XRF spectra of Sm, Ho, and W samples.

1. Introduction

High-purity germanium (HPGe) detectors provide a useful combination of energy resolution and detection efficiency for X-ray and gamma-ray spectroscopy [1,2,3]. These properties support their use in scientific experiments and applied radiation analysis.
Experimental applications of HPGe detectors include passive gamma spectroscopy of spent nuclear fuel [4], in-beam nuclear spectroscopy with tracking arrays [5], and searches for rare decays [6]. For intense X-ray and gamma-ray fields, high count-rate capability must be balanced against energy resolution. Measurements with segmented HPGe detectors illustrate how the achievable throughput and peak width depend on the detector and readout configuration [2,3].
One possible way to solve this problem is to use multi-channel spectrometers in which each input channel is equipped with a planar or quasi-planar low-capacitance HPGe detector. Such systems distribute the detected radiation flux among several registration channels and thereby increase the total throughput of the multi-channel system. The spectra acquired by individual channels can subsequently be combined to obtain a spectrum of the complete detector assembly.
A traditional implementation uses multiple individual HPGe crystals in a common detector assembly, as demonstrated by the 13-element fluorescence EXAFS array of Cramer et al. [7]. This arrangement permits each detector channel to be characterized separately. The design of such assemblies must also account for cryostat complexity, detector packing, and the distribution of irradiation across the channels.
Another promising direction for creating multi-detector assemblies for multi-channel spectrometers is the formation of a multi-channel detector structure within a single HPGe monocrystal by using various segmentation technologies [2,3,8,9,10,11]. Detector segmentation not only increases the speed of the detection system but also provides additional position or angular sensitivity and reduces the background level by spatially separating events. This is especially important, for example, in the analysis of spectra in EDXRF spectrometry.
Experimental multi-element developments include the 19-channel hexagonal-pad demonstrator with CMOS preamplifiers [12], the LEAPS-INNOV/XAFS-DET prototypes [13,14], the 96-channel Gaia system with dedicated ASIC readout [15], and the compact seven-channel Gepta-EX detector [1]. These studies show that sensor segmentation, charge-sharing control, front-end electronics and digital processing must be considered together when assessing resolution and throughput.
High-energy K-line XRF has been demonstrated for rare-earth and heavy-element analysis using synchrotron excitation [16,17] and laboratory 241Am excitation with a germanium detector [18]. The principal Kα and Kβ1 lines of lanthanides lie approximately within 33–64 keV; the broader 10–100 keV band considered here also includes Y K-lines and selected emissions from other heavy elements [19].
A target FWHM of 600 eV at 59.5 keV was adopted as a preliminary resolution-throughput benchmark. It does not guarantee complete separation of every rare-earth interference: La Kβ1 at 37.801 keV and Nd Kα1 at 37.361 keV are separated by only about 0.440 keV [19]. Its adequacy for quantitative XRF requires validation with reference materials under the intended excitation conditions.
The aim of this work was to develop multi-detector structures of different types based on HPGe crystals and to study and compare their spectrometric characteristics both at low and high count rates. The main intended application of the developed spectrometer is the detection of characteristic K-shell X-ray emission lines of rare-earth elements in X-ray fluorescence analysis. The contribution is the implementation and comparative characterization of two six-channel detector blocks, including thin entrance contacts, cooled input stages and rate-dependent spectrometric measurements for laboratory XRF development.

2. Development of a Multi-Detector Assembly Based on Single HPGe Crystals

Historically, multi-detector assemblies were first manufactured on the basis of single crystals; therefore, the characteristics of single detectors can serve as a reference point when evaluating the characteristics of the developed multi-detector assemblies. Considering the energy range of interest, 10 to 100 keV, the simplest solution would have been to develop a multi-detector assembly based on a conventional planar HPGe crystal. However, because dense detector packing is required in the assembly, we chose a quasi-spherical (semi-planar) crystal design, which has a lower operating voltage, simpler manufacturing technology, and higher reliability.
Low detector capacitance can be obtained by combining a small readout electrode with suitable field shaping, as demonstrated by shaped-field and p-type point-contact HPGe detectors [20,21]. The recessed central contact in Figure 1 shares this design principle. Its measured capacitance of approximately 3 pF is specific to the complete electrode geometry; the preamplifier input load also includes the input transistor and connections.
Figure 1. Cross-section of the quasi-spherical HPGe detector. The Li-diffused region and the a-Ge/Al entrance contact form a common n+ electrode. The central p+ contact serves as the readout electrode, and the annular guard groove is passivated with SiO.
The multi-crystal detector block consists of six separate P-type HPGe quasi-spherical (semi-planar) crystals, each 19.2 mm in diameter and 12.2 mm thick. The thin a-Ge/Al entrance-contact area is 211.1 mm2 per crystal, giving a total area of 1266.6 mm2 for the six-crystal assembly. The crystal geometry and photograph are shown in Figure 1 and Figure 2, respectively. The impurity concentration was (1.0–1.2) × 1010 cm−3.
Figure 2. Photograph of a single quasi-spherical (semi-planar) HPGe detector.
The Li-diffused n+ contact has a diffusion depth of 0.7 mm. It surrounds the thin entrance window, covers the lateral surface, and extends over the peripheral rear surface up to the guard groove. The entrance contact consists of a-Ge and Al layers with thicknesses of 0.3 and 0.1 µm, respectively. During deposition, the a-Ge/Al contact partially overlaps the Li-diffused region, providing direct electrical continuity and forming a common n+ electrode. The nominal entrance-window dead-layer thickness is 0.5 µm.
The a-Ge layer was deposited by magnetron sputtering in an Ar + 7% H2 mixture at a pressure of 1.5 × 10−2 Torr.
The central p+ readout contact was formed by boron ion implantation in a well 4.0 mm in diameter and 3.3 mm deep. An annular guard groove surrounds this contact to limit leakage currents associated with edge effects. Its outer and inner diameters are 10.7 and 7.0 mm, respectively, and its depth is 2.0 mm. The groove surface is passivated with silicon monoxide (SiO).
The current–voltage and capacitance–voltage characteristics of the individual detectors were studied in an immersion cryostat cooled with liquid nitrogen. Each crystal of the multi-crystal assembly had a capacitance of 3.1 pF. The full-depletion voltage was 200 V. The leakage currents of each detector under cooling and at bias voltages in the range of 200–800 V were below 10 pA. Each detector was operated with a bias of +400 V applied to the wrap-around contact.
All six crystals, together with the input stages of the charge-sensitive preamplifiers (see below), were installed in an LN2-cooled vacuum cryostat. The preamplifier inputs were connected to the rear contacts of the detectors by mechanical clamps. Each detector was operated with a positive bias applied to the wrap-around contact. A photograph of the cryostat and a model of the vacuum chamber are shown in Figure 3. The cryostat had six beryllium entrance windows 18 mm in diameter and 0.2 mm thick, allowing X-ray radiation to be detected from approximately 3 keV.
Figure 3. Photograph of the cryostat and model of the vacuum chamber.
Signals from the individual detectors in the multi-crystal assembly were read out using serial charge-sensitive preamplifiers (CSPs) with resistive feedback, which are used by BSI in commercially produced X-ray spectrometers. The input-stage elements of the preamplifier placed in the cryostat near the detectors were optimized to ensure the best spectrometric characteristics of the system. Since the total capacitance at the preamplifier input must be minimized, an MX20 FET from MOXTEK (Orem, UT, USA), with an input capacitance of 0.5 pF and a transconductance of 12.4 mS [22], was selected for a single-detector capacitance of 3 pF. The feedback resistor value was 1 GΩ to ensure operation at high count rates. The MX20 JFET also incorporates a dedicated Substrate electrode, which provides an additional degree of freedom for optimizing its noise performance under cryogenic operating conditions.
Applying a negative bias to the Substrate electrode with respect to the JFET channel promotes the formation of a more uniform conduction channel along its entire length. This increases the transconductance of the JFET while maintaining its low input capacitance, thereby reducing the equivalent input series noise. The optimum substrate bias depends on the drain-to-source voltage, drain current, and operating temperature of the JFET. During detector assembly, this bias is adjusted experimentally to achieve the minimum electronic noise at the preamplifier output.
For a JFET operating at a temperature of 80–100 K, with a source voltage of 0 V, a drain voltage of +3 V, and a drain current of 12–15 mA, the optimum low-noise substrate bias is typically −6.5 to −7 V.

3. Development of the Segmented Detector

3.1. Crystal Segmentation Using Amorphous Germanium

The development of technologies for forming contacts on semiconductor detectors based on amorphous germanium (a-Ge) has significantly simplified the creation of multi-channel segmented planar HPGe detectors [8]. Such contacts provide good blocking properties and low leakage currents and make it possible to form multi-segment structures using relatively simple technological methods, for example, by applying metallization through shadow masks.
Amorphous-semiconductor contacts also support orthogonal-strip Ge detectors for gamma-ray spectroscopy and imaging. Amman and Luke demonstrated that separately biased field-shaping electrodes can reduce incomplete charge collection in the gaps between collecting electrodes [23]. Experimental comparisons of conventional and amorphous contacts, hybrid Li-diffused/a-Ge configurations, and spatially resolved charge-collection measurements provide further context for the present design [9,10,11].
Leakage-current behaviour must be assessed together with contact preparation, surface passivation, temperature and operating environment. Measurements of a-Ge-contact detectors in cryogenic liquids illustrate the separate roles of bulk charge injection and surface leakage [24]. Thus, low leakage current alone does not establish spatially uniform charge collection or the effective collection area of a segmented detector.
The segmented detector developed in this work is based on a planar HPGe crystal 83.4 mm in diameter and 16.8 mm thick, with a P-type impurity concentration of (5.6–6.4) × 109 cm−3. The electrode pattern is a single-sided strip-electrode configuration consisting of six charge-collection strips 41.4 mm long and 2.9 mm wide, each surrounded by a steering-grid strip 1.0 mm wide. The pitch of the collection strip is 7.7 mm. The steering grid is used to maintain good charge-collection efficiency in the regions between the collection strips. The steering grid is surrounded by a 3.0 mm-wide guard ring. The geometric area of one collecting electrode is 41.4 × 2.9 = 120.06 mm2, giving a total collecting-electrode area of 720.36 mm2 for six strips. These values should be distinguished from the effective charge-collection area because the steering grid directs charge generated in the interstrip regions towards the collecting electrodes. Accounting for the interstrip spacing, a nominal geometric area of pL = 7.7 × 41.4 = 318.78 mm2 can be assigned to each segment, giving 6pL = 1912.68 mm2 for the six-segment detector. This estimate assumes ideal charge steering and neglects boundary effects at the outer strips and strip ends, as well as possible local charge-collection losses. It has not been experimentally validated as an effective sensitive area. The nominal estimate is approximately 1.51 times the total thin-window area of 1266.6 mm2 reported for the multi-crystal assembly in Section 2. This geometric ratio does not establish the relative photon-detection efficiencies of the two assemblies.
The present device combines a Li-diffused bias contact with patterned a-Ge/Al collecting electrodes. Related hybrid contacts configurations have been demonstrated experimentally [10]. This choice retains a conventional full-area bias contact while using amorphous-contact processing for the segmented surface.
Figure 4 shows (a) the detector crystal structure with the dimensions of its elements and (b) a photograph of the fabricated crystal with deposited contacts. The contact on the back side over the entire area, that is, the contact on the side opposite the strips, was formed by lithium diffusion. The electrodes on the front side have an a-Ge/Al structure with layer thicknesses of 0.3 and 0.1 µm, respectively. Metallization was performed by deposition through a mask. The open surface of the crystal was passivated by a layer of amorphous germanium. The detector operates with a positive bias applied to the full-area contact; the collection strips are connected to virtual ground through charge-sensitive preamplifiers, and a positive bias is also applied to the steering grid. Each collection strip is read out by an individual charge-sensitive preamplifier. The detector is mounted in a holder that provides the necessary electrical and thermal connections.
Figure 4. (a) Structure of the detector crystal with element dimensions; (b) photograph of the segmented planar crystal with deposited contacts.

3.2. Crystal Modeling and Electrical Characterization

The crystal was modeled using the SolidStateDetectors.jl (version 0.11.9) package [25]. For this purpose, a 3D model of the crystal was created in FreeCAD (version 1.0.2) and converted into a format compatible with the simulation package using CSGStripper (version 1.0), an application developed at BSI.
Since the crystal has two planes of mirror symmetry, the simulations were performed for only one quarter of the crystal, taking advantage of the symmetry to reduce computational time.
The simulation package can accurately determine the depletion voltage only when the potential applied to the steering grid is 0 V. Therefore, for simulations with positive potentials applied to the steering grid, the depletion voltage values calculated for a zero grid potential were used.
Capacitance calculations for the detector electrode geometry yielded a self-capacitance of approximately 15.5 pF for a collecting strip. The calculated capacitance components are summarized in Table 1.
Table 1. Calculated self-capacitance and interelectrode capacitances of the collecting strips.
These calculated values describe the detector electrodes. The input-transistor and connection capacitances constitute additional contributions to the complete front-end input capacitance.
For the cylindrical reference model, the calculated full-depletion voltages at zero steering-grid potential were 316, 653, and 963 V for impurity concentrations of 2.5 × 109, 5.0 × 109, and 7.5 × 109 cm−3, respectively (Figure 5a).
Figure 5. Electrostatic simulations of HPGe detectors. (a) Calculated full-depletion voltage versus impurity concentration for the cylindrical reference model at zero steering-grid potential. (b–i) Electric-field distributions for the actual detector geometry shown in Figure 4, with an impurity concentration of 6 × 109 cm−3. (b–e) Detector bias of +700 V and steering-grid potentials of 0, +50, +100, and +200 V, respectively. (f–i) Detector biases of +700, +1200, +1700, and +2200 V, respectively, with the steering grid held at +50 V. Colours indicate the electric-field magnitude using a common scale; white curves represent electric-field lines.
Increasing the steering-grid potential expands the low-field region beneath the grid (Figure 5b–e). Conversely, increasing the detector bias at a fixed steering-grid potential of +50 V reduces this region in segments, as illustrated in Figure 5f–i.
The simulations provided qualitative guidance for selecting a detector bias above the calculated depletion voltages while limiting the steering-grid potential. The final operating settings of +1700 V and +50 V were selected using experimental spectrometric measurements. The electric-field calculations were performed for the actual detector geometry shown in Figure 4, with an impurity concentration of 6 × 109 cm−3. Weighting-potential maps and quantitative charge-collection boundaries are not presented here.
The detector was characterized in a liquid-nitrogen-cooled immersion cryostat. The leakage currents remained below 10 pA per strip at bias voltages between 200 and 2000 V.

3.3. Charge-Sensitive Preamplifier for the Segmented Detector

Segmented HPGe detectors require multi-channel readout while preserving spectrometric performance. Dedicated ASIC front ends have been demonstrated with multi-element Ge sensors in [1,15].
In this setup, signals from the preamplifier outputs of the Verdi ASIC [26] were processed and digitized using a BOSON multichannel analyzer, which was connected to each preamplifier output in turn.
The preamplifier input stage was implemented using a cooled J309 JFET with an input capacitance of 5 pF and a transconductance of 20 mS. The input stage was located inside the cryostat, while the main amplifying and output stages operated at room temperature outside the cryostat. The feedback components were Rfb = 1 GΩ and Cfb = 0.5 pF.
The connection diagram for the Verdi ASIC is shown in Figure 6.
Figure 6. Circuit diagram of the detector readout using the Verdi ASIC with cooled JFET input stages.
Figure 7a shows a photograph of the segmented detector and cooled preamplifier board assembly in the cryostat. The corresponding connection of the detector segments to the preamplifier inputs was made by wire bonding with 30 µm thick gold wire manufactured by Heraeus Electronics GmbH & Co. KG (Hanau, Germany). The outputs of the board were wired to the cryostat connectors with 0.3 mm diameter manganin wires manufactured by Isabellenhütte Heusler GmbH & Co. KG (Dillenburg, Germany). A photograph of the full segmented HPGe detector assembly, manufactured by Baltic Scientific Instruments (Riga, Latvia), with the installed thermal shield is shown in Figure 7b. The cryostat is closed from above by a vacuum cover with a 0.8 mm thick carbon entrance window manufactured by R&G Faserverbundwerkstoffe GmbH (Waldenbuch, Germany), which makes it possible to perform spectrometric measurements in the low-energy region. The distance between the entrance window and the surface of the detector crystal was 6 mm.
Figure 7. (a) Photograph of the segmented detector and cooled board assembly in the cryostat; (b) photograph of the complete segmented detector assembly in the cryostat.

4. Results

The spectrometric characteristics of the developed HPGe multi-detector assemblies with their signal preamplifiers were subsequently tested as part of a spectrometer with a BOSON multichannel analyzer, manufactured by Baltic Scientific Instruments (Riga, Latvia), equipped with GammaPro spectrometric software (version 2026.10.03) for spectrum acquisition and analysis. The segmented-detector readout used a shaping time constant of 1 µs. The processing parameters were adjusted experimentally, with the operating range limited by the knee of the output-count-rate (OCR) versus input-count-rate (ICR) characteristic, corresponding to approximately 63% dead time. The available pulse-processing settings used for each measurement series are specified in the relevant sections of the manuscript.

4.1. Segmented Detector at Low Count Rates

The spectra of 137Cs, 57Co, 241Am, and 55Fe with a cooled FET displayed on a logarithmic scale are shown in Figure 8a, Figure 8b, Figure 8c and Figure 8d, respectively. The 55Fe peak has a complex shape because it consists of two overlapping peaks at 5.9 and 6.5 keV; therefore, to determine the resolution at 5.9 keV, the standard double-Gaussian fitting function in GammaPro was used, taking into account the relative intensities of the two lines.
Figure 8. Spectra of radionuclide sources 137Cs (a), 57Co (b), 241Am (c), and 55Fe (d), obtained from one detector segment. HV = +1700 V.
The results of the energy-resolution measurements at a detector bias voltage of +1700 V and a steering-grid voltage of +50 V are presented in Table 2. The bias voltage and steering grid voltage were selected based on both simulations and experimental measurements. The simulations showed that increasing the steering-grid voltage expands the low-field regions beneath the grid and modifies the electric-field distribution near the collecting strips. Increasing the detector bias at a fixed grid potential reduces these low-field regions (Figure 5b–i). Experimental measurements performed with steering-grid voltages between 0 and +100 V showed no significant differences in detector performance. Therefore, a steering-grid voltage of +50 V was selected for all measurements.
Table 2. Energy resolution of the six-strip segmented HPGe detector at different photon energies.
Cooling can reduce electronic noise, but the resulting resolution also depends on total input capacitance, the input transistor and shaping conditions [21,27]. In this work we used the J309 JFET, which is employed in BSI’s standard production because of its optimal balance of characteristics and cost. The MX20 junction field-effect transistor (manufactured by Moxtek) [22] can be considered an alternative due to its lower input capacitance. This would allow for even better energy resolution at high count rates.

4.2. Segmented Detector at High Count Rates

Measurements with a 137Cs source showed that increasing the count rate leads to the expected degradation of energy resolution. In particular, at a count rate of 40 kcps per segment, a resolution of 3.2 keV (FWHM) was obtained, whereas increasing the rate to 57 kcps per segment worsened the resolution to 3.5 keV (FWHM). As the count rate increases, the probability of pulse pile-up increases and the influence of noise and baseline instability grows, resulting in peak broadening. These measurements were performed using a short shaping time of approximately 1 µs, which is a typical operating mode of spectrometric analyzers at high count rates.
The high-count-rate performance reported here characterizes the tested segmented detector together with its readout electronics and pulse processing. The available measurements do not separately quantify the contributions of detector geometry and electronics. Further optimization of the complete detection chain may improve its performance.

4.3. Multi-Crystal Assembly at Low Count Rates

For each HPGe detector included in the multi-crystal assembly, X-ray and gamma-ray spectra of 55Fe, 241Am, 57Co, 137Cs, and 152Eu radionuclides from point sources were recorded, as well as a spectrum of a mixed source. Measurements were performed under identical conditions for all system channels. The distance between the point source and the detector was selected so that the count rate of the spectrometric chain was in the range of 0.9–1 kcps. The low-count-rate spectra were acquired with BOSON settings (shaping time constant—6 µs, BLR = 32 and BLR = 64). For the 55Fe spectrum, the overlapping peaks were fitted with two Gaussian components. The 152Eu spectrum in Figure 9 shows the 121.8, 244.7, 344.0, and 778.9 keV lines and illustrates the investigated energy range. Table 3 summarizes the FWHM values obtained for selected spectral lines of 55Fe, 57Co, 241Am, 152Eu, and 137Cs.
Figure 9. Spectra of 55Fe, 241Am, 57Co, 137Cs, and 152Eu for a single detector in low-count-rate mode. In the 55Fe spectrum, the grey solid line represents the measured data, the black solid line represents the overall fitted profile, and the grey dashed lines represent the individual fitting components.
Table 3. Energy resolution of the six-channel multi-crystal HPGe detector at different photon energies.
Typical recorded spectra of 55Fe, 57Co, 241Am, 152Eu, and 137Cs for a single detector in the low-count-rate mode are shown in Figure 9. Processing of the acquired spectra yielded the energy resolution (FWHM) of the detectors for characteristic X-ray and gamma-ray lines of the indicated radionuclides. For each nuclide, the resolution values were averaged over all channels of the assembly. The obtained mean values are given in Table 3.
The results show that the detectors demonstrate high energy resolution over the entire investigated energy range. This indicates a low level of electronic noise, efficient charge collection, and optimized operation of the spectrometric chain. The measurements also confirm stable operation of all channels of the multi-crystal assembly and high uniformity of their spectrometric characteristics, which is an important condition for application in multi-crystal spectrometers at high count rates.

4.4. Multi-Crystal Assembly at High Count Rates

When studying the operation of the multi-crystal assembly at increased count rates, particular attention was paid to energy resolution in the low-energy region, especially at the 59.5 keV line. This line was chosen because, in tasks involving control of rare-earth and heavy elements, the 10–100 keV energy range is the most informative, since it contains the characteristic X-ray lines of these elements.
The studies were performed using a 241Am source placed at a distance of 5 cm, which provided a basic count rate of about 2 kcps. To investigate operation at elevated count rates, a high-activity 137Cs source was added, increasing the combined input count rate in the measured detector channel to 200 kcps.
The dependence of energy resolution on the input count rate is shown in Figure 10. With increasing input count rate, a monotonic degradation of the resolution for the 59.5 keV line was observed: the FWHM value increased from approximately 444 eV at 2 kcps to approximately 1080 eV at 200 kcps. The most pronounced broadening was observed above ICRs of 60–80 kcps, consistent with increasing count-rate-dependent processing distortions. The additional Compton continuum raises the background beneath the photopeak and can increase the uncertainty of peak fitting, but does not by itself broaden the intrinsic peak response.
Figure 10. FWHM versus per-channel ICR for the 661.7 keV 137Cs peak and the 59.5 keV 241Am peak under 137Cs and 238Pu loading in a single channel of the multi-crystal assembly. The shaping time constant was 1 µs. The baseline-restoration settings were BLR = 16 for the 238Pu loading series and BLR = 32 for the 137Cs loading series. Error bars represent the statistical uncertainties (1σ) of the FWHM values obtained from peak fitting.
Similar measurements were performed for a combination of 241Am and 238Pu sources (Figure 10). In this case, the increase in FWHM with increasing input count rate was much weaker: from approximately 459 eV to approximately 680 eV in the range up to 200 kcps. The weaker increase in FWHM under 238Pu loading was observed with BLR = 16, whereas the 137Cs loading series was acquired with BLR = 32. The effects of the loading conditions and BLR settings were not independently separated, so the observed difference cannot be attributed solely to the loading source.
For the 137Cs line at 661.7 keV, degradation of energy resolution with increasing input count rate was also observed, from approximately 1240 eV to approximately 1720 eV. However, the relative change in resolution for the high-energy line was less pronounced than for the low-energy line, which is attributed to the operating characteristics of the gated baseline restorer at high count rates.
The energy resolution of the full-energy peak is determined by the quadratic sum of the electronic-noise contribution and the statistical fluctuations in the number of charge carriers generated in the detector:
FWHM =   η 2 + 2.35 2 ∗ F ∗ ω ∗ E
Here, FWHM is the energy resolution of the spectral line; η is the noise-related contribution expressed as an energy FWHM; F is the Fano factor, with a value of 0.118 adopted for germanium; ω is the mean electron–hole pair creation energy, equal to 2.96 eV for germanium; and E is the energy of the detected photons. Consistent energy units are used throughout the equation.
With increasing ICR, the effective noise contribution can increase because shorter intervals between successive pulses reduce the time available for accurate baseline estimation. The statistical contribution depends on photon energy and is independent of the baseline-restoration process.
For illustration, using the approximate value F ≈ 0.118 and ω = 2.96 eV gives statistical FWHM contributions of approximately 0.339 keV at 59.5 keV and 1.13 keV at 662 keV.
For a comparable rate-dependent noise contribution, its relative effect on the total peak width is smaller at 661.7 keV than at 59.5 keV, resulting in less pronounced relative degradation of energy resolution. Overall, the measurements show that increasing the input count rate is accompanied by deterioration in energy resolution associated with pulse pile-up and limitations of the signal-processing chain.
At an input count rate of 200 kcps and a mean deposited energy of approximately 300 keV, the average detector current is approximately 3 nA. With a feedback resistance of 1 GΩ, this corresponds to a mean feedback-voltage shift of approximately 3 V. Baseline excursions of 3–5 V under 137Cs loading must therefore be considered when assessing the available output headroom. The different responses under 137Cs and 238Pu loading may partly reflect differences in the mean deposited energy and the resulting baseline disturbances.
The BOSON analyzer uses an analog spectroscopic amplifier with input CR differentiation and a short differentiation time constant, followed by an ADC with an input range of ±10 V. This arrangement reduces the sensitivity of the downstream electronics to slow baseline variations, provided that the preamplifier remains within its linear operating range.
Resistive feedback was selected for its reliability, implementation simplicity, and established use in our commercial systems. Despite its dynamic-range limitations at high deposited-energy rates, this configuration remains a practical choice for the intended low-energy XRF application. The potential improvements achievable with transistor-reset preamplifiers will be investigated in a subsequent study. Experimental work with adaptive digital processing and transistor-reset preamplifiers provides a relevant precedent for this direction [28].
In the single-channel rate scans shown in Figure 10, the FWHM of the 59.5 keV peak reached approximately 680 eV under 238Pu loading and 1080 eV under 137Cs loading at an ICR of 200 kcps. These results characterize the respective loading and processing configurations, with BLR = 16 and BLR = 32, respectively. The effects of the incident energy distribution and baseline-restoration settings were not independently separated.
The influence of one spectrometric channel on another was investigated using the above-mentioned 238Pu source. The cross-influence was evaluated by the change in energy resolution of the 59.5 keV line (241Am) in one detector while the remaining detectors of the multi-crystal assembly were loaded with a high-activity 137Cs point source. In this case, the count rate was in the range of 100–150 kcps.
Figure 11 provides additional characterization of the same detector channel under the two loading and baseline-restoration configurations. OCR was estimated as OCR = ICR × (1 − D), where D is the dead-time fraction.
Figure 11. High-count-rate characteristics of a single channel of the multi-crystal HPGe assembly under 238Pu (shaping time constant—1 µs, BLR—16) and 137Cs (shaping time constant—1 µs, BLR—32) loading: (a) estimated output count rate (OCR) versus ICR; (b) dead time versus ICR; and (c) shift in the 59.5 keV 241Am peak centroid, in eV, relative to the lowest-ICR spectrum for each loading condition.
The estimated OCR reached approximately 57 kcps at an ICR of 102–121 kcps under 238Pu loading and approximately 46 kcps at an ICR of 100 kcps under 137Cs loading. At an ICR of 200 kcps, the dead-time fractions were 87% and 98%, corresponding to estimated OCR values of 26 and 4 kcps, respectively. As shown in Figure 11c, the centroid of the 59.5 keV peak under 238Pu loading fluctuated around an approximately constant level at ICRs up to about 120 kcps per channel. Above the selected optimal operating range, it shifted towards lower energies. Under 137Cs loading, the centroid showed a consistent shift towards lower energies with increasing ICR. The maximum negative shifts reached approximately −130 eV under 238Pu loading and −340 eV under 137Cs loading at 200 kcps per channel, relative to the lowest-ICR spectrum for each loading condition.
Beyond the OCR maximum, reached at ICRs of approximately 100–120 kcps per channel, further increases in ICR lead to higher dead time and reduced throughput. These points characterize an overload regime because reduced event acceptance can limit photopeak statistics and increase peak-fitting uncertainty. The ICR range of 100–120 kcps per channel was therefore selected as the optimal operating range.
It is known that the BLR setting does not affect the throughput characteristic but can significantly influence the baseline offset and, consequently, the position of the full-energy peak. The baseline is monitored during the intervals between spectrometric pulses, and deviations from zero are integrated and used to compensate for baseline drift. Large-amplitude pulses can therefore result in larger residual baseline-restoration errors.
Figure 11c shows the peak-centroid shifts observed under the two loading conditions. The differences between the throughput characteristics under 238Pu and 137Cs loading may be associated with the longer effective duration of the spectrometric pulses produced by high-energy 137Cs photons.
Figure 12 shows a typical spectrum of the 238Pu/241Am source recorded by one detector of the multi-crystal assembly (red curve). The same figure also shows the spectrum obtained by the same detector under conditions in which the other detectors of the assembly were loaded with 137Cs sources (blue curve). The figure shows that no noticeable peak shift is observed, and the energy resolution of the 59.5 keV line (241Am) remains practically unchanged both when measuring with one detector and when the entire multi-crystal system operates simultaneously with additional loading of other detectors by the 137Cs source. The test characterizes the combined influence of neighbouring-detector loading and does not separate electronic cross-talk from additional irradiation of the detector under study. Future work will address these contributions through dedicated pulser and shielding measurements, together with investigations of interstrip cross-talk and charge sharing in the segmented detector. At the same time, the spectrum clearly shows an increase in the Compton-scattering level over the entire measured energy range, which is associated with additional gamma radiation from the 137Cs source placed near neighboring detectors.
Figure 12. Typical spectrum of the 238Pu/241Am source of a single detector in the absence (red spectrum) and presence (blue spectrum) of cross-channel loading.
Figure 13 compares the 59.5 keV peak resolution of the six detector channels and their summed spectra. At per-channel ICRs of approximately 100–160 kcps, the mean FWHM values of individual channels ranged from 596 to 704 eV without additional loading of neighbouring detectors and from 630 to 733 eV with additional 137Cs loading. The corresponding FWHM values of the summed spectra were 648 and 662 eV, respectively. The energy resolution was measured for all six individual detector channels of the assembly. The same 238Pu/241Am source was placed directly adjacent to each detector in turn, without changing the source-to-detector geometry. Owing to differences in the individual detector efficiencies and responses, the resulting count rates ranged from approximately 100 to 160 kcps. For each detector, measurements were performed under two conditions: without additional loading of the neighboring channels and with each of the remaining five channels loaded by a separate point 137Cs source placed directly adjacent to that channel. The spectra used for the energy-resolution evaluation were acquired only from the detector irradiated by the 238Pu/241Am source.
Figure 13. FWHM of the 59.5 keV 241Am line for the six detector channels, identified as GF15, GF16, GF17, GF19, GF20, and GF21, and their summed spectrum (SUM), without (circles) and with (triangles) additional loading of neighbouring detectors. Each group contains five measurements of 200 s each; labels show the mean FWHM. The broken horizontal axis accommodates the higher total ICR of the summed spectrum compared with individual-channel ICRs.
Each measurement presented in Figure 13 was repeated five times under identical experimental conditions with an acquisition time of 200 s per spectrum in order to verify the repeatability and stability of the energy-resolution values determined at the 59.5 keV gamma-ray line of Am-241. All five individual measurements are shown in Figure 13, while the labeled FWHM values represent the mean of the five measurements for each detector and for the summed spectrum. The spectra from the six detector channels were acquired independently. Before summation, each spectrum was energy-calibrated separately and converted to a common energy scale. The calibrated spectra were then summed bin by bin; no event-by-event coincidence summation was applied.
Table 4 presents the mean FWHM values and sample standard deviations calculated from five repeated measurements under both loading conditions for each detector channel and the summed spectrum.
Table 4. Mean FWHM values and sample standard deviations.
Additional loading of neighbouring detectors increased the mean FWHM by approximately 8–34 eV across the six channels, corresponding to 1.2–5.7%. For the summed spectrum, the mean FWHM increased from 648 to 662 eV, while the sample standard deviations of five repeated measurements were 1.9 and 2.7 eV, respectively. The change in the summed spectrum therefore exceeded the observed repeat-to-repeat scatter. Because electronic cross-talk and direct additional irradiation were not separated, the broadening cannot be assigned to a single mechanism.

4.5. Testing of the Detector Assembly for X-Ray Fluorescence Detection

The detector assembly was tested as part of an operational prototype X-ray fluorescence (XRF) analyzer by recording characteristic K-series X-rays from samarium (Sm), holmium (Ho), and tungsten (W).
The samples were placed in plastic tubes 20 mm in diameter and 180 mm in length. Characteristic radiation was excited using a tungsten-anode X-ray tube operated at 145 kV and 7.5 mA. The primary radiation passed through a 6.2 mm thick tin filter. Each measurement lasted 900 s (15 min).
The analyte concentrations were 500 ppm for Sm and W and 2000 ppm for Ho. Spectra recorded by the independent detector channels were converted to a common energy scale, summed, and processed as a single spectrum.
Figure 14 presents the summed spectra in the 30–70 keV energy range. The characteristic K-series lines of the investigated elements are clearly visible.
Figure 14. X-ray fluorescence spectra of samarium (Sm), holmium (Ho), and tungsten (W) samples acquired using the multi-channel HPGe detector assembly. The samples were irradiated using a tungsten-anode X-ray tube operated at 145 kV and 7.5 mA, with a 6.2 mm Sn filter and a measurement time of 900 s.
Detection limits (DLs) were estimated from the number of counts in the selected region of the characteristic peak and the corresponding background contribution. The results are presented in Table 5.
Table 5. Estimated detection limits.
These results demonstrate the capability of the multi-channel HPGe detector assembly to record characteristic K-series X-rays from the investigated elements. Under the measurement conditions considered, the estimated detection limits ranged from 5.57 to 12.03 ppm.

5. Discussion

Table 6 places the present implementations alongside representative Ge detector systems. The listed resolutions and rate measurements have different shaping and irradiation conditions and therefore do not define a matched performance ranking. In particular, Gepta-EX reports 218 eV at 5.9 keV and 373 eV at 59.5 keV under low-flux conditions [1], while Gaia reports 252 eV and 406 eV, respectively, for a representative channel [15].
Table 6. Representative Ge detector measurements. FWHM values refer to low-rate characterization and are not asserted to hold at the rate points listed in the same row.
Published rate results also require careful interpretation. The Diamond hexagonal-pad demonstrator was reported to operate near 1 Mcps per channel with less than 13% count loss in a preliminary beam test [29]. A later LEAPS-INNOV evaluation reports throughput of approximately 150–200 kcps per pixel, or 1.5–2 Mcps for ten elements [14]. These measurements should be distinguished from design targets and from calibration rates used to measure the best low-rate FWHM.
For the present strip prototype, an additional comparison can be made with the segmented planar HPGe detectors developed for high-count-rate gamma-ray spectroscopy in [2,3]. A comparison of detector topologies shows that the detector designs differ in several respects that directly affect their spectrometric parameters. The collecting electrodes in the present detector are 41.4 mm long and 2.9 mm wide, corresponding to an electrode area of 120.06 mm2. In Refs. [2,3], the collecting strips are 71 mm long and 2.85 mm wide, corresponding to an electrode area of 202.35 mm2. To include the interstrip regions consistently, the same nominal geometric estimate pL can be used for both layouts. This gives 318.78 mm2 per segment for the present detector and 539.6 mm2 per strip for the detector in Refs. [2,3], which has a strip pitch of 7.6 mm. These estimates neglect edge and end effects and should not be interpreted as measured effective sensitive areas. Differences in capacitive loading cannot be inferred from these area estimates alone.
The calculated self-capacitance of a collecting strip in the present detector is approximately 15.5 pF, as described in Section 3.2. In Ref. [2], the measured capacitance of a central strip was approximately 26 pF, compared with a calculated value of 20.6 pF, including an interstrip contribution of 18.8 pF. The authors attributed the higher measured value to stray capacitances associated with cables and feedthroughs and to measurement uncertainty. Ref. [3] does not report a new numerical measurement of strip capacitance. These detector-level capacitances should be distinguished from the total input capacitance of the assembled front-end electronics, which also includes contributions from the input transistor and connections. Differences in electronic noise therefore depend on both detector capacitances and front-end operating conditions.
Established semiconductor-electronics studies describe the coupled dependence of electronic noise on total input capacitance and shaping time [27]. Short shaping times increase the importance of series noise, while long shaping times increase parallel-noise contributions and pulse overlap. Digital trapezoidal shaping, deconvolution and adaptive processing provide additional ways to balance resolution, ballistic effects, pile-up and throughput [28,30].
Contact-related leakage and interstrip charge collection should be assessed under specified bias, temperature and irradiation conditions [11,23,24]. The leakage currents measured here were below 10 pA per strip, consistent with a small leakage-current contribution to parallel noise under those conditions. This result alone does not establish the total electronic noise or spatial charge-collection uniformity.
The present segmented detector was developed for high-resolution low-energy spectroscopy using cooled front-end electronics and low-leakage-current contacts. The detectors in Refs. [2,3] were developed for high-rate gamma-ray spectroscopy at total input count rates of approximately 1–2 Mcps. Their performance differences reflect both detector characteristics and readout configurations. At low count rates, Ref. [2] reported an energy resolution of 2.16 keV FWHM at 662 keV. When transitioning to high count rates, the resolution gradually deteriorated; for example, at count rates of about 100,000 cps per channel, the FWHM reached about 3.3 keV, and with further rate increases to hundreds of thousands of counts per second per strip and total rates up to mega counts per second, the resolution deteriorated to 4 keV and above.
At low count rates, the present segmented prototype with cooled FETs achieved: 560 eV at 5.9 keV, 649–655 eV at 13.9–14.4 keV, 669 eV at 26.34 keV, 700 eV at 59.5 keV, 783 eV at 122 keV, and 1290 eV at 662 keV. Thus, at the 662 keV line our detector demonstrates a resolution approximately 1.7–1.8 times better than that reported in [2,3]. However, it should be noted that the segmented detector developed in this work already provided an energy resolution of 3.2–3.5 keV at substantially lower count rates (40–57 thousand cps), which is comparable to the results of [2,3] at higher count rates.
The contribution of this work is the fabrication and comparative characterization of two complete six-channel detector implementations for laboratory XRF development. The results document the performance obtained with the chosen contacts, cooled input stages and processing configurations. The modern monolithic systems summarized above show that improved low-rate resolution and high throughput are achievable with other implementations; the present comparison does not establish an inherent architectural advantage. For the segmented prototype, the high-rate measurements remain limited to 661.7 keV and do not establish performance in the intended 10–100 keV range.
The two spectrometers were tested as complete systems with different input FETs, preamplifier implementations, and entrance windows. The multi-crystal assembly used MX20 input FETs and 0.2 mm thick beryllium windows, whereas the segmented prototype used J309 input FETs and a 0.8 mm thick carbon window. The higher total capacitance at the input of the segmented detector’s preamplifier is one of the factors contributing to its poorer energy resolution at low energies compared with the multi-crystal assembly. The present comparison therefore does not isolate the contribution of detector geometry from that of the readout electronics. An architectural comparison would require matched front-end characterization and consistent pulse-processing settings, including the same peaking time. Entrance-window transmission should also be considered when comparing the low-energy detection response and the incomplete record of the pulse-processing settings limits the reproducibility of the segmented-detector measurements.
The tested multi-crystal assembly achieved average FWHM values of 346 eV at 5.9 keV and 450 eV at 59.5 keV at low count rates, with the latter below the target value of 600 eV. In the separate six-channel comparison at high count rates, individual-channel mean FWHM values ranged from 596 to 704 eV without additional loading of neighbouring detectors and from 630 to 733 eV with additional loading. The summed-spectrum FWHM values were 648 and 662 eV, respectively. These results do not demonstrate consistent achievement of the 600 eV target across all channels under high-count-rate conditions.
Further work will address optimization of both implementations, including the input transistors, preamplifiers, and pulse-processing parameters, with the goal of consistently achieving 600 eV FWHM at 59.5 keV across all channels at high input count rates.
The spectra presented in Figure 14 provide a preliminary experimental demonstration of rare-earth and heavy-element XRF measurements using the multi-crystal assembly. A comprehensive presentation of the XRF results, including quantitative analysis of reference materials, will be given in a separate publication currently in preparation.

Author Contributions

Conceptualization, V.G., A.S. and S.P.; methodology, S.P., T.M. and M.K.; software, S.P.; formal analysis, V.K.; investigation, R.N., O.S. and T.M.; writing—original draft preparation, T.M. and V.G.; writing—review and editing, T.M. and V.G.; project administration, A.S. and V.G.; visualization, M.K.; supervision, V.K., A.S. and V.G. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by the European Regional Development Fund within the research project No.P1 “Research and development of an X-ray fluorescence analyzer for rare earths and heavy metals,” supported by European Regional Development Fund (project No. 1.2.1.2.i.2/1/24/A/CFLA/003 between the “Energy and Transport Competence Center” (http://etkc.lv/) and the “Central Finance and Contracting Agency of Latvia” (http://www.cfla.gov.lv/lv/).

Data Availability Statement

The raw data supporting the conclusions of this article will be made available by the authors on request.

Acknowledgments

The authors express their gratitude to the Energy and Transport Competence Center and the Central Finance and Contracting Agency of Latvia for their support of this research and development work funding.

Conflicts of Interest

All authors are employees of Baltic Scientific Instruments (BSI), the commercial manufacturer of the detectors and spectrometers described in this paper, and BSI provided the equipment and facilities used in this study. All authors had full access to all data in the study and take responsibility for the integrity of the data and the accuracy of the data analysis.

References

  1. Rumaiz, A.K.; Capocasa, F.; Kuczewski, A.J.; Pinaroli, G.; Krings, T.; Ravel, B.; Siddons, D.P. Gepta-EX: A multi-channel germanium detector for X-ray absorption fine structure. J. Synchrotron Radiat. 2026, 33, 1016–1021. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  2. Cooper, R.J.; Amman, M.; Luke, P.N.; Vetter, K. A prototype high purity germanium detector for high resolution gamma-ray spectroscopy at high count rates. Nucl. Instrum. Methods Phys. Res. Sect. A 2015, 795, 167–173. [Google Scholar] [CrossRef] [Scilit]
  3. Cooper, R.J.; Amman, M.; Vetter, K. High resolution gamma-ray spectroscopy at high count rates with a prototype high purity germanium detector. Nucl. Instrum. Methods Phys. Res. Sect. A 2018, 886, 1–6. [Google Scholar] [CrossRef] [Scilit]
  4. Jansson, P.; Bengtsson, M.; Bäckström, U.; Grape, S.; Branger, E.; Sjöland, A. Time stamped list mode data from gamma-ray spectroscopic measurements on 47 nuclear fuel assemblies performed at Clab, Sweden, September 2016 through March 2019. Data Brief 2020, 31, 106039. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  5. Akkoyun, S.; Algora, A.; Alikhani, B.; Ameil, F.; de Angelis, G.; Arnold, L.; Astier, A.; Ataç, A.; Aubert, Y.; Aufranc, C.; et al. AGATA—Advanced GAmma Tracking Array. Nucl. Instrum. Methods Phys. Res. Sect. A 2012, 668, 26–58. [Google Scholar] [CrossRef] [Scilit]
  6. Agostini, M.; Araujo, G.R.; Bakalyarov, A.M.; Balata, M.; Barabanov, I.; Baudis, L.; Bauer, C.; Bellotti, E.; Belogurov, S.; Bettini, A.; et al. Final Results of GERDA on the Search for Neutrinoless Double-β Decay. Phys. Rev. Lett. 2020, 125, 252502. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  7. Cramer, S.P.; Tench, O.; Yocum, M.; George, G.N. A 13-element Ge detector for fluorescence EXAFS. Nucl. Instrum. Methods Phys. Res. Sect. A 1988, 266, 586–591. [Google Scholar] [CrossRef] [Scilit]
  8. Luke, P.N.; Cork, C.; Madden, N.; Rossington, C.; Wesela, M. Amorphous Ge bipolar blocking contacts on Ge detectors. IEEE Trans. Nucl. Sci. 1992, 39, 590–594. [Google Scholar] [CrossRef] [Scilit]
  9. Gros, S.; Hammond, N.; Lister, C.; Chowdhury, P.; Fischer, S.; Freeman, S. Performance tests of large-area position-sensitive planar germanium detectors with conventional and amorphous contacts. Nucl. Instrum. Methods Phys. Res. Sect. A 2009, 602, 467–476. [Google Scholar] [CrossRef] [Scilit]
  10. Yang, M.; Li, Y.; Zeng, Z.; Tian, Y.; Xue, T.; Zeng, M.; Tang, W.; Chang, J. A prototype segmented planar high-purity germanium detector using wraparound lithium diffusion electrode and amorphous germanium blocking contact. Radiat. Detect. Technol. Methods 2022, 6, 433–438. [Google Scholar] [CrossRef] [Scilit]
  11. Rintoul, E.; Boston, A.J.; Boston, H.C.; Caffrey, A.; Cresswell, J.; Harkness-Brennan, L.; Judson, D.; Nolan, P.; Platt, J.; Unsworth, C.; et al. Characterisation of the charge collection properties in a segmented planar HPGe detector. Nucl. Instrum. Methods Phys. Res. Sect. A 2021, 987, 164804. [Google Scholar] [CrossRef] [Scilit]
  12. Tartoni, N.; Chatterji, S.; Crook, R.; Krings, T.; Bombelli, L.; Alborini, A. Hexagonal Pad Multichannel Ge X-Ray Spectroscopy Detector Demonstrator: Comprehensive Characterization. IEEE Trans. Nucl. Sci. 2020, 67, 1952–1961. [Google Scholar] [CrossRef] [Scilit]
  13. Goyal, N.; Aplin, S.; Balerna, A.; Bell, P.; Casas, J.; Cascella, M.; Chatterji, S.; Cohen, C.; Collet, E.; Fajardo, P.; et al. Next generation multi-element monolithic Germanium detectors for spectroscopy: First integration at ESRF facility. Nucl. Instrum. Methods Phys. Res. Sect. A 2026, 1081, 170835. [Google Scholar] [CrossRef] [Scilit]
  14. Goyal, N.; Iguaz, F.J.; Aplin, S.; Balerna, A.; Bell, P.; Casas, J.; Cascella, M.; Chatterji, S.; Cohen, C.; Collet, E.; et al. Development and first performance evaluation of multi-element monolithic HPGe detector for X-ray spectroscopy. Nucl. Instrum. Methods Phys. Res. Sect. A 2026, 1092, 171854. [Google Scholar] [CrossRef] [Scilit]
  15. Rumaiz, A.K.; Capocasa, F.; Kuczewski, A.J.; Pinaroli, G.; Li, J.; Kuczewski, J.; Finnelli, K.; Koh, K.; Shutlz-Johnson, L.; Fitzgerald, C.; et al. Gaia: Segmented germanium detector for high-energy X-ray fluorescence and spectroscopic imaging. J. Synchrotron Radiat. 2026, 33, 91–97. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  16. Nakai, I.; Terada, Y.; Itou, M.; Sakurai, Y. Use of highly energetic (116 keV) synchrotron radiation for X-ray fluorescence analysis of trace rare-earth and heavy elements. J. Synchrotron Radiat. 2001, 8, 1078–1081. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  17. van der Ent, A.; Brueckner, D.; Spiers, K.M.; Falch, K.V.; Falkenberg, G.; Layet, C.; Liu, W.-S.; Zheng, H.-X.; Le Jean, M.; Blaudez, D. High-energy interference-free K-lines synchrotron X-ray fluorescence microscopy of rare earth elements in hyperaccumulator plants. Metallomics 2023, 15, mfad050. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  18. Vizcaya, S.; Gutierrez, M.; Greaves, E.D. 241Am Radioisotope-Excited Spectrometer for K-Line X-ray Fluorescence (XRF) Analysis of Rare Earth Elements. Appl. Spectrosc. 2025, 79, 1680–1687. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  19. Deslattes, R.D.; Kessler, E.G., Jr.; Indelicato, P.; de Billy, L.; Lindroth, E.; Anton, J. X-ray transition energies: New approach to a comprehensive evaluation. Rev. Mod. Phys. 2003, 75, 35–99. [Google Scholar] [CrossRef] [Scilit]
  20. Luke, P.N.; Goulding, F.S.; Madden, N.W.; Pehl, R.H. Low capacitance large volume shaped-field germanium detector. IEEE Trans. Nucl. Sci. 1989, 36, 926–930. [Google Scholar] [CrossRef] [Scilit]
  21. Barton, P.; Amman, M.; Martin, R.; Vetter, K. Ultra-low noise mechanically cooled germanium detector. Nucl. Instrum. Methods Phys. Res. Sect. A 2016, 812, 17–23. [Google Scholar] [CrossRef] [Scilit]
  22. Moxtek. Available online: https://www.moxtek.com/ (accessed on 6 October 2026).
  23. Amman, M.; Luke, P.N. Three-dimensional position sensing and field shaping in orthogonal-strip germanium gamma-ray detectors. Nucl. Instrum. Methods Phys. Res. Sect. A 2000, 452, 155–166. [Google Scholar] [CrossRef] [Scilit]
  24. Panth, R.; Liu, J.; Abt, I.; Liu, X.; Schulz, O.; Wei, W.-Z.; Mei, H.; Mei, D.-M.; Wang, G.-J. Characterization of high-purity germanium detectors with amorphous germanium contacts in cryogenic liquids. Eur. Phys. J. C 2020, 80, 667. [Google Scholar] [CrossRef] [Scilit]
  25. Abt, I.; Fischer, F.; Hagemann, F.; Hauertmann, L.; Schulz, O.; Schuster, M.; Zsigmond, A.J. Simulation of semiconductor detectors in 3D with SolidStateDetectors.jl. J. Instrum. 2021, 16, P08007. [Google Scholar] [CrossRef] [Scilit]
  26. Celani, A.; Bombelli, L.; Fiorini, C.; Frizzi, T.; Nava, R.; Pavlick, J.; Kastner, M.; Morichi, M.; Roberson, B.; Zakrzewski, B.; et al. VERDI: A versatile readout ASIC for radiation detectors. In Proceedings of the 2010 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), Knoxville, TN, USA, 30 October–6 November 2010; pp. 1382–1385. [Google Scholar] [CrossRef] [Scilit]
  27. Goulding, F.S.; Landis, D.A. Signal processing for semiconductor detectors. IEEE Trans. Nucl. Sci. 1982, 29, 1125–1141. [Google Scholar] [CrossRef] [Scilit]
  28. Saxena, S.; Hawari, A.I. Digital pulse deconvolution with adaptive shaping for real-time high-resolution high-throughput gamma spectroscopy. Nucl. Instrum. Methods Phys. Res. Sect. A 2020, 954, 161288. [Google Scholar] [CrossRef] [Scilit]
  29. Tartoni, N. Detector Group. In Diamond Annual Review 2018; Online Technical Report; Diamond Light Source: Oxfordshire, UK, 2018; Available online: https://www.diamond.ac.uk/Home/Corporate-Literature/Annual-Review/Review2018/Detector-Group.html (accessed on 6 November 2026).
  30. Jordanov, V.T.; Knoll, G.F. Digital synthesis of pulse shapes in real time for high resolution radiation spectroscopy. Nucl. Instrum. Methods Phys. Res. Sect. A 1994, 345, 337–345. [Google Scholar] [CrossRef] [Scilit]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.