Abstract
Background: Prosthesis mass, inertia, and alignment can alter mechanical loading during gait, but their relative influence may depend on how loading is quantified. This study examined their effects on absolute and relative joint reaction force outcomes in a transfemoral prosthesis model. Methods: A deterministic 3 × 3 × 3 full-factorial design generated 27 configurations from three mass, inertia, and mediolateral alignment levels, driven by identical prescribed kinematics and external loads in OpenSim. Resultant reactions were evaluated at the socket–prosthesis junction and prosthetic ankle using peak resultant reaction force, force–time integral (FTI), socket-to-ankle FTI ratio, and a secondary Peak Reaction Ratio. Factorial decomposition and range-normalized sensitivity analysis quantified parameter effects. Results: Within this deterministic computational framework, prosthesis mass was the dominant contributor to variation in the absolute loading outcomes, representing 90.28% and 94.06% of socket and ankle peak-force variation and 94.26% and 95.63% of their FTI variation, respectively. The FTI ratio showed a more distributed sensitivity to mass (53.55%), inertia (28.90%), and alignment (13.41%), whereas the Peak Reaction Ratio was primarily influenced by inertia (51.99%) and mass (45.85%). Peak timing separation ranged from 2.7% to 48.0% of the normalized gait cycle. Conclusions: Within the tested parameter ranges and fixed reference gait, prosthesis mass was the dominant contributor to absolute reaction-force variation in the rigid-body model, whereas relative socket-to-ankle outcomes showed greater contributions from inertia and alignment. These findings represent model-level sensitivities and should not be interpreted as direct clinical fitting recommendations.
1. Introduction
Lower-limb prosthesis performance depends not only on restoring locomotion but also on how mechanical loads are transmitted through the prosthetic system. In individuals with transfemoral amputation, the socket forms the primary mechanical interface between the residual limb and the prosthesis, and its fit, comfort, and load-distribution characteristics strongly influence prosthetic function [1,2]. Objective characterization of socket mechanics remains challenging. Pressure mapping, shear sensing, gait analysis, and imaging can provide quantitative information, yet clinical fitting still relies substantially on subjective feedback and heterogeneous measurement approaches [1,2,3,4].
The mechanical environment of a prosthetic limb cannot be described by interface pressure or peak force alone. Paternò et al. quantified interface–pressure distributions in transfemoral sockets, whereas Devin et al. examined the dynamic behavior of the socket interface during walking [3,4]. Using finite-element analysis, Henao et al. demonstrated that gait-cycle loading alters stress distribution at the residual-limb/socket interface [5], while Schwarze et al. evaluated prosthesis–socket loading using a validated multibody simulation approach [6]. Prosthesis mass and its distribution have also been shown to affect gait mechanics: Ikeda et al. reported changes in metabolic and spatiotemporal gait characteristics with added mass placement, and Hekmatfard et al. showed effects of mass distribution on spatiotemporal parameters and knee kinematics [7,8]. Prosthetic alignment represents another important mechanical factor. Zhang et al. reported alignment-dependent changes in gait and biomechanical loading in individuals with transfemoral amputation, while Groeneveld et al. demonstrated alignment-related changes in lower-limb kinetics [9,10]. Recent numerical work has also highlighted the importance of gait-related loading for the mechanical design and safety of osseointegrated prosthetic components [11]. These findings show that prosthetic loading is influenced by several interacting mechanical factors and cannot be fully characterized using a single loading variable. Importantly, changes in absolute reaction-force magnitude do not necessarily describe how loading is redistributed across different levels of the prosthetic system. A parameter that produces only a modest change in absolute force may still alter the mechanical relationship between proximal and distal loading locations. This distinction provides a rationale for evaluating absolute and relative loading measures as complementary outcomes, since they may convey different aspects of prosthetic mechanical behavior.
Musculoskeletal modeling provides a controlled framework for separating these effects. OpenSim-based models are widely used to simulate gait, estimate internal joint loading, and investigate human-device interaction [12,13]. More recent models have extended these approaches to transfemoral amputation and bone-anchored prostheses [14,15]. Such models are particularly useful for deterministic sensitivity analyses, in which model parameters can be systematically varied while kinematics and external loading are held constant. Under these conditions, joint reaction forces should be interpreted as model-level rigid-body reactions rather than direct estimates of physiological socket interface pressure. This controlled framework is particularly useful for determining whether the mechanical influence of a prosthesis parameter depends on the outcome used to characterize loading.
The present study builds on our previous isolated-parameter analysis [16]. The same baseline musculoskeletal modeling framework, model geometry, and prosthesis parameter definitions were retained to provide a common mechanical basis, while the present study used a different participant and a new reference gait dataset. The current work generated a complete 3 × 3 × 3 factorial set of 27 prosthesis configurations and recomputed the corresponding inverse-dynamics and Joint Reaction outputs for all combined parameter conditions.
In the previous study, prosthesis mass, rotational inertia, and mediolateral alignment were varied one at a time, and prosthesis mass was identified as the strongest determinant of absolute reaction-force magnitude when each parameter was perturbed in isolation [16]. That univariate approach could not quantify interaction effects or determine whether the relative importance of the three parameters changes across different loading metrics. The present analysis therefore extends the earlier work by quantifying main and interaction effects simultaneously, introducing socket-to-ankle FTI and Peak Reaction Ratios, and applying factorial and range-normalized sensitivity analyses. This distinction is important because conclusions about the mechanical relevance of a prosthesis parameter may depend on whether loading is characterized by peak magnitude, cumulative loading, or a relative measure between prosthetic levels.
The present study addressed these gaps by (i) varying all three parameters simultaneously in a complete 3 × 3 × 3 factorial design, allowing main effects and interactions to be partitioned within a common sensitivity space; (ii) introducing relative socket-to-ankle outcome measures (FTI ratio and Peak Reaction Ratio) that were not evaluated in [16]; and (iii) applying a range-normalized sensitivity analysis to compare parameter influence independently of the perturbation scale assigned to each factor. The scientific contribution therefore extends beyond identifying which individual parameter produces the largest change in reaction force. It evaluates whether conclusions about parameter influence remain consistent when loading is characterized using different mechanical outcomes.
The aim of this study was to determine how prosthesis mass, inertia, and mediolateral alignment jointly influence mechanical loading in a transfemoral prosthesis model, with particular emphasis on whether absolute and relative loading outcomes exhibit the same parameter sensitivity.
2. Materials and Methods
2.1. Study Design and Modeling Rationale
This study used a controlled computational framework to quantify the mechanical sensitivity of a transfemoral prosthesis model to changes in prosthesis mass, rotational inertia, and mediolateral alignment. A three-factor, three-level full-factorial design was implemented, yielding 27 prosthesis configurations.
To isolate the direct mechanical effects of these parameter changes, all configurations were driven by the same prescribed reference kinematics and external loads. The analysis was therefore conducted as a deterministic, fixed-kinematics sensitivity study, without introducing configuration-specific gait adaptation, neuromuscular compensation, or inter-subject variability.
A single experimentally recorded gait cycle from one healthy adult was used as the common reference input. The gait data were treated as a controlled mechanical input rather than as a representative sample of transfemoral amputee gait. Accordingly, the model provided a standardized basis for comparing parameter-dependent mechanical responses, without aiming to predict patient-specific or population-level loading.
2.2. Reference Gait Data and Experimental Setup
Reference gait data were collected from one healthy adult participant (male, age 25 years, height 176 cm, body mass 69 kg). Eligibility criteria included age 18–60 years, independent walking without assistive devices, and absence of neurological, orthopedic, balance, or gait-affecting disorders. Individuals with health conditions affecting gait or preventing safe participation, or who were unable to provide informed consent, were excluded. Reflective markers were placed according to the Rizzoli Orthopedic Institute marker protocol. Marker trajectories were recorded using a 13-camera Qualisys Oqus 7 motion-capture system at 120 Hz (Qualisys AB, Gothenburg, Sweden), while ground reaction forces were recorded simultaneously from two AMTI force plates (Advanced Mechanical Technology, Inc., Watertown, MA, USA) at 1200 Hz. Both data streams were synchronized during acquisition using Qualisys Track Manager (QTM, version 2025.2; Qualisys AB, Gothenburg, Sweden). Marker trajectories were visually inspected for missing samples prior to processing. The experimental gait-data acquisition was approved by the Kaunas Regional Biomedical Research Ethics Committee (approval No. BE-2-13/2026, 27 January 2026). Written informed consent was obtained from the participant before data collection.
Marker trajectories and force-plate signals were then low-pass filtered using zero-phase Butterworth filters with cutoff frequencies of 6 Hz and 15 Hz, respectively, consistent with filtering procedures previously applied in gait biomechanics studies [17,18]. Gait events were identified from the filtered vertical ground reaction force using a 20 N threshold. A representative gait cycle with complete marker trajectories and clean force-plate contact was selected and used as the common reference input for all prosthesis configurations (Figure S2). Inverse kinematics was subsequently performed using the filtered marker trajectories. Time-dependent outcomes were normalized to 0–100% of the gait cycle after the biomechanical analyses, for visualization only.
2.3. Musculoskeletal Model and Prosthesis Implementation
A generic lower-limb musculoskeletal model (Gait2354, OpenSim 4.5; Stanford University, Stanford, CA, USA) was used as the baseline framework. The model was scaled to the participant’s anthropometry using the experimental marker data before inverse kinematics was performed. A left transfemoral prosthesis was implemented by replacing the biological distal limb with rigid prosthetic components representing the socket/prosthetic shank assembly and prosthetic foot. The prosthesis implementation followed the modeling framework used in our previous study [16] and was consistent with published OpenSim approaches for representing transfemoral prosthetic limbs [14,15]. The prosthetic knee was modeled as a single-degree-of-freedom pin joint connecting the femur to the prosthetic shank and permitting flexion-extension, whereas the prosthetic ankle was modeled as a weld joint connecting the prosthetic shank to the prosthetic foot. The rigid ankle representation corresponded to a passive prosthetic foot without articulated ankle motion. Detailed joint definitions, frame locations, and Joint Reaction settings are provided in Supplementary Table S1.
Markers associated with the prosthesis were assigned to the corresponding prosthetic rigid bodies, while anatomical markers remained associated with the biological segments. The modified prosthesis model and the full-factorial parameter perturbation scheme are illustrated in Figure 1. The global OpenSim coordinate system was defined with +X directed anteriorly, +Y superiorly, and +Z toward the subject’s right side. For the left prosthetic limb, +Z therefore corresponded to the medial direction and -Z to the lateral direction. A local coordinate system was attached to the prosthetic segment and moved with the segment during gait.
Figure 1.
Modified transfemoral model and full-factorial prosthesis parameter perturbations. (A) Model configuration, reaction-force locations, and global and local coordinate systems. (B) Mass (0.75×, 1.00×, 1.25×), rotational inertia (0.50×, 1.00×, 1.50×), and mediolateral alignment (−10 mm lateral, 0 mm, +10 mm medial) levels used to generate 27 configurations. Rotational inertia scaling was applied to the principal moments of inertia independently of mass. The mediolateral translation arrows are schematic, with the actual displacement applied along the global Z-axis for the left prosthetic limb.
Inverse kinematics was performed once using the experimental marker trajectories to obtain the reference gait kinematics. The resulting joint-coordinate trajectories were prescribed identically to all 27 prosthesis configurations. Inverse-kinematics tracking quality was quantified using OpenSim marker-error outputs and reconstructed model-marker positions. For the modified prosthetic model, the mean frame-wise RMS marker error was 35.52 mm, with pelvis marker RMS errors ranging from 7.2 to 8.7 mm and prosthetic-side thigh, shank, and foot marker RMS errors ranging from 5.6 to 20.0 mm. Marker weights were 10 for the bilateral ASIS, sacral, heel, and toe-tip markers; 1 for the remaining trunk and lower-limb markers; 0.5 for the acromion markers; and 0.1 for the head marker. The bilateral subtalar and metatarsophalangeal coordinates were strongly constrained toward their model default values using coordinate-task weights of 1000. The principal applied model scale factors were 1.025 for the pelvis, 1.147 for both femora, 0.989 for both tibiae, and 1.077 for the torso; the prosthetic shank and foot geometries were not additionally scaled. The original scaled model had a mean frame-wise RMS marker error of 37.19 mm under the same reference motion, indicating that the modified prosthetic model did not degrade overall marker-tracking performance.
For each configuration, inverse dynamics was performed using the prescribed kinematics, model inertial properties, and measured ground reaction forces to quantify the net generalized forces and moments required by the prescribed motion. These inverse-dynamics generalized forces were used for dynamic-consistency assessment but were not interpreted as individual muscle or actuator forces. Joint Reaction analysis was subsequently performed using the same prescribed kinematics, model inertial properties, and measured external loads to obtain resultant reactions at the socket–prosthesis junction and prosthetic ankle, expressed in the ground reference frame. No muscle-force estimation, static optimization, neuromuscular optimization, reserve actuators, or other configuration-specific actuator forces were included. Consequently, the reported reactions represent rigid-body model-level joint reactions in the absence of explicit biological muscle-force contributions.
2.4. Prosthesis Parameter Perturbations and Outcome Analysis
The baseline distal prosthesis mass was 3.50 kg, comprising a 2.50 kg prosthetic shank and a 1.00 kg prosthetic foot. The corresponding center-of-mass coordinates were [0, −0.150, 0] m for the prosthetic shank and [0.050, −0.020, 0] m for the prosthetic foot. Baseline principal moments of inertia were 0.02440, 0.00488, and 0.02440 kg·m2 for the prosthetic shank and 0.00813, 0.00163, and 0.00813 kg·m2 for the prosthetic foot. Mass scaling factors of 0.75, 1.00, and 1.25 were applied to both segments, while rotational inertia was independently scaled by factors of 0.50, 1.00, and 1.50. These symmetric low, baseline, and high levels were selected to provide a controlled sensitivity range around the baseline configuration. Mass and inertia were varied independently to separate their mechanical effects within the factorial design. Consequently, not all combinations were intended to represent physically realizable commercial prostheses. All principal moments remained positive and satisfied the triangle inequalities after scaling, while the off-diagonal products of inertia remained zero throughout the analysis. Mediolateral alignment was altered by translating the complete distal prosthetic assembly (prosthetic shank and foot) by −10, 0, or +10 mm relative to the socket–prosthesis junction. The ±10 mm perturbation was selected to provide a symmetric alignment range around the baseline and was consistent with the 10 mm medial and lateral socket translations previously examined in individuals with transfemoral amputation [9]. The segment center-of-mass locations, prosthetic foot geometry, prescribed kinematics, and ground-reaction-force application points were unchanged across alignment conditions. Combining the three levels of each factor resulted in 27 unique configurations. As a secondary robustness analysis, a coupled mass-inertia subset was evaluated using the existing M0.75-I0.50, M1.00-I1.00, and M1.25-I1.50 configurations across all three alignment levels, representing monotonically increasing mass and rotational inertia.
Pelvis residual forces and moments were retained as indicators of dynamic consistency between the prescribed kinematics, model inertial properties, and experimentally measured external loads. Across the 27 configurations, RMS residual force ranged from 0.071 to 0.078 BW and peak residual force from 0.216 to 0.243 BW. Residual moments were normalized to body weight and body height (BW·BH); RMS residual moment ranged from 0.019 to 0.021 BW·BH, and peak residual moment ranged from 0.036 to 0.039 BW·BH. For the baseline configuration, additional inspection showed that the largest residual-force component occurred at the beginning of the analyzed gait cycle (0% gait cycle), whereas the peak socket reaction occurred later, at 14.7% of the gait cycle. The peak residual moment occurred substantially later in the cycle, at approximately 83% of the gait cycle.
Residuals were interpreted as indicators of dynamic inconsistency and not as physical loads acting at the prosthesis. They arise in inverse dynamics because the measured external loads and prescribed model kinematics do not satisfy the equations of motion exactly. Accordingly, elevated residuals reduce confidence in the absolute magnitude of calculated internal reactions, although the same experimental kinematics and external-load dataset were applied consistently to all 27 configurations. Published verification guidelines recommend residual forces below approximately 5% of the net external force and residual moments below approximately 1% of the net external force multiplied by center-of-mass height [19]. The residual moments in the present analysis were therefore of a more acceptable magnitude, whereas the residual forces exceeded the recommended criterion and were treated as a limitation when interpreting absolute joint-reaction magnitudes.
Reaction forces were evaluated at the socket–prosthesis junction and prosthetic ankle. Post-processing and quantitative analysis were performed in MATLAB R2023b (MathWorks, Natick, MA, USA). The OpenSim output files were imported and processed to calculate resultant reaction-force magnitudes, peak forces, force–time integrals, socket-to-ankle loading ratios, peak timing separation, and factorial sensitivity metrics. The same post-processing procedure was applied consistently across all 27 configurations.
The resultant reaction-force magnitude was calculated from the three orthogonal force components as:
where Fx, Fy, and Fz are the orthogonal joint reaction-force components expressed in the global OpenSim ground reference frame for both the socket–prosthesis junction and prosthetic ankle. The global axes were defined as +X anterior, +Y superior, and +Z toward the subject’s right side.
Resultant magnitude was selected as the primary force outcome to provide a coordinate-independent measure of overall mechanical loading across configurations. Directional force components were not interpreted as direct socket interface shear or pressure because the model did not represent the distributed residual-limb/socket contact interface. Forces were normalized to the participant’s body weight (BW), calculated from the fixed reference body mass of 69 kg (BW = 69 × 9.81 N), and the same normalization factor was used for all configurations. Therefore, changes in prosthesis mass did not alter the reference body weight used for normalization.
Absolute loading was characterized using peak resultant reaction force and the force–time integral (FTI):
where FR(t) is the resultant reaction force in newtons and BW is the fixed reference body weight (69 × 9.81 N). The ratio represents the dimensionless body-weight-normalized resultant force. The normalized force–time integral is reported in BW·s, whereas integration of the unnormalized force FR(t) yields N·s. FTI was calculated over the complete analyzed gait cycle (0.392–1.592 s; duration: 1.200 s) using the trapezoidal rule and the original time vector (76 samples at 0.016 s intervals). Time normalization to 0–100% of the gait cycle was applied only for visualization.
Relative socket-to-ankle loading was primarily characterized using the FTI ratio:
where FTIsocket and FTIankle represent the force–time integrals of the resultant reaction forces at the socket–prosthesis junction and prosthetic ankle, respectively, calculated over the same gait cycle.
The Peak Reaction Ratio was defined as:
where Fsocket,max and Fankle,max represent the independently occurring maximum resultant reaction forces at the socket–prosthesis junction and prosthetic ankle, respectively. Because these maxima were not necessarily synchronous, the Peak Reaction Ratio was interpreted as a configuration-level relationship between peak magnitudes. Both the Peak Reaction Ratio and FTI ratio were used only as comparative indices of reaction magnitude at two distinct model locations. Neither metric represents a fraction of force transmitted from the prosthetic ankle to the proximal prosthesis connection, and neither should be interpreted as a measure of mechanical load transfer or attenuation. Peak timing separation was evaluated separately. To assess the robustness of outcome extraction to temporal sampling density, peak and FTI outcomes were recalculated after retaining every second and every third sample of the original reaction-force waveforms.
Parameter influence was quantified using deterministic factorial decomposition across the complete 3 × 3 × 3 design. For each outcome, the response was represented as:
where (M), (I), and (A) denote prosthesis mass, rotational inertia, and mediolateral alignment, respectively. The contribution of each main effect or interaction was calculated as:
where SST is the total sum of squares across the 27 simulated configurations. These percentages represent shares of variation within the investigated parameter space. Main effects, two-way interactions, and the three-way interaction were retained separately. Because the deterministic design was unreplicated, no experimental-error term or inferential p-values were estimated. A complementary range-normalized sensitivity analysis was performed because factorial contribution estimates depend on the perturbation ranges assigned to each parameter. For each factor, outcome values were averaged across all combinations of the other two factors at the low, baseline, and high levels. Sensitivities were calculated for standardized +10% changes in mass or inertia and +10 mm changes in mediolateral alignment as follows:
where Y-high, Y-baseline, and Y-low denote the corresponding factor-level mean outcomes; f-range is the full investigated factor range; and f-std is the standardized increment. For example, for socket peak reaction force, the low-, baseline-, and high-mass level means were 10.012, 12.278, and 14.508 BW, respectively; scaling the full 0.75–1.25 mass-factor range to a +10% mass increment yielded a 7.32% relative response.
Y{ijk} = μ + Mi + Ij + Ak + (MI)ij + (MA)ik + (IA)jk + (MIA)ijk
3. Results
3.1. Reaction-Force Patterns Across the Gait Cycle
Resultant reaction-force profiles at the socket–prosthesis junction and prosthetic ankle showed distinct loading patterns across the gait cycle (Figure 2). Across the 27 prosthesis configurations, peak socket reaction forces ranged from 9.37 to 15.47 BW, whereas peak ankle reaction forces ranged from 3.45 to 4.57 BW. Socket-level reactions were consistently greater than ankle reactions, with the largest differences occurring during stance. For the baseline configuration, the peak resultant reaction was 12.45 BW at the socket–prosthesis junction and 4.00 BW at the prosthetic ankle, occurring at 14.7% and 61.3% of the gait cycle, respectively. Component-wise baseline force histories in both newtons and body-weight-normalized units are provided in Figure S3.
Figure 2.
Resultant joint reaction force (JRF) profiles at the socket–prosthesis junction (A) and prosthetic ankle (B). Low-load and high-load configurations were selected based on the minimum (9.37 BW) and maximum (15.47 BW) proximal peak resultant reaction forces across all 27 configurations. The low-load configuration used mass and inertia factors of 0.75 and 0.50, respectively, with −10 mm alignment; the high-load configuration used factors of 1.25 and 1.50, with +10 mm alignment. The baseline configuration used factors of 1.00 for both mass and inertia, with 0 mm alignment. Forces are normalized to body weight (BW), and the gait cycle is expressed from 0 to 100%. Filled circles indicate independently occurring peak resultant reaction forces.
A secondary coupled mass-inertia analysis showed the same monotonic loading trend. Averaged across the three alignment levels, proximal peak reaction increased from 9.45 BW in the M0.75-I0.50 condition to 12.36 BW at baseline and 15.24 BW in the M1.25-I1.50 condition. Corresponding ankle peak reactions were 3.63, 3.99, and 4.37 BW, respectively. Proximal and ankle FTI values likewise increased from 9.23 to 13.08 BW·s and from 2.98 to 3.92 BW·s, respectively.
The experimental vertical ground reaction force profiles used as the reference external loading exhibited a characteristic double-peaked stance pattern for both limbs (Figure S2). Peak vertical ground reaction force was 1.19 BW for the right limb and 1.20 BW for the left limb.
Although reaction-force magnitude varied between configurations, the overall waveform pattern remained comparable because the same reference kinematics and external loads were prescribed across all models. The socket and ankle maxima did not necessarily occur at the same instant within the gait cycle. The separation between their respective peak timings ranged from 2.7% to 48.0% of the gait cycle. Accordingly, socket and ankle peak values were treated as independently occurring maxima in the subsequent analyses.
3.2. Effects of Prosthesis Parameters on Absolute Loading
Absolute loading varied systematically across the 27 prosthesis configurations. The force–time integral ranged from 9.16 to 13.16 BW·s at the socket–prosthesis junction and from 2.94 to 4.01 BW·s at the prosthetic ankle.
Factorial decomposition showed that prosthesis mass was the dominant contributor to variation in all four absolute loading outcomes (Figure 3). Mass represented 90.28% of the variation in peak socket reaction force and 94.06% in peak ankle reaction force. For time-integrated loading, mass represented 94.26% of the variation in socket FTI and 95.63% in ankle FTI. Contributions from rotational inertia, mediolateral alignment, and interaction terms were comparatively small for these absolute outcomes. Interaction effects were generally small (Supplementary Table S2).
Figure 3.
Relative contributions of prosthesis mass, rotational inertia, mediolateral alignment, and their interactions to variation in six joint reaction force (JRF) outcomes across the 27-model full-factorial design. Contributions are expressed as percentages of the total sum of squares. Absolute outcomes include peak resultant reaction force and force–time integral (FTI) at the socket–prosthesis junction and prosthetic ankle; relative outcomes include the FTI ratio and Peak Reaction Ratio.
The range-normalized sensitivity analysis showed a consistent pattern (Figure 4). A +10% increase in prosthesis mass was associated with the largest relative change in each absolute loading outcome, corresponding to approximately 7.3% for socket peak reaction force, 4.4% for ankle peak reaction force, 5.6% for socket FTI, and 4.8% for ankle FTI. Equivalent standardized changes in rotational inertia and mediolateral alignment produced smaller changes in the corresponding outcomes. Thus, within the investigated parameter ranges, both the factorial decomposition and range-normalized analysis identified prosthesis mass as the parameter most strongly associated with absolute reaction-force magnitude and time-integrated loading.
Figure 4.
Range-normalized sensitivity of joint reaction force (JRF) outcomes to prosthesis mass, rotational inertia, and mediolateral alignment. (A) Peak resultant reaction force and force–time integral (FTI) at the socket–prosthesis junction and prosthetic ankle. (B) Socket-to-ankle FTI ratio and Peak Reaction Ratio. Values represent signed relative percentage changes per +10% change in mass or inertia and per +10 mm change in mediolateral alignment.
The extracted loading outcomes were insensitive to temporal downsampling. Across all 27 configurations, maximum deviations from the full-resolution results remained below 2% for peak reaction forces and below 0.05% for FTI outcomes when every second or every third sample was retained. This check assessed the robustness of outcome extraction to temporal sampling density and did not evaluate uncertainty in the underlying biomechanical model.
A directional check around the baseline showed broadly similar positive and negative mass sensitivities for the absolute loading outcomes, whereas the smaller inertia- and alignment-related responses showed somewhat greater asymmetry.
3.3. Relative Socket-to-Ankle Loading and Parameter Sensitivity
Relative socket-to-ankle loading showed a different pattern of parameter dependence from the absolute loading outcomes. For the FTI ratio, prosthesis mass represented 53.55% of the variation, followed by rotational inertia (28.90%) and mediolateral alignment (13.41%) (Figure 3). In contrast, variation in the Peak Reaction Ratio was primarily associated with rotational inertia (51.99%) and mass (45.85%), with a smaller contribution from alignment (0.34%) and interaction effects. Thus, the relative importance of the prosthesis parameters depended on whether loading was characterized using time-integrated or peak-based measures. Interaction effects were generally small. For the absolute loading outcomes, no individual interaction accounted for more than 0.38% of the total variation. For the relative measures, the largest two-way interaction was mass × inertia for the FTI ratio (1.97%), followed by inertia × alignment (1.31%). For the Peak Reaction Ratio, the three-way interaction contributed 0.74%, while the mass × inertia and inertia × alignment interactions contributed 0.56% and 0.43%, respectively. Directional interaction contrasts were modest: for the FTI ratio, the inertia × alignment interaction ranged from −0.52% at inertia 0.5× with −10 mm alignment to +0.49% at inertia 0.5× with +10 mm alignment. For the Peak Reaction Ratio, the inertia × alignment contrasts ranged from −1.28% at inertia 1.5× with +10 mm alignment to +0.81% at inertia 1.0× with +10 mm alignment. Overall, these interactions introduced modest non-additivity without changing the parameter ranking.
The range-normalized sensitivity analysis further demonstrated this metric dependence (Figure 4B). For the FTI ratio, standardized perturbations produced relative changes of approximately +0.84% for mass, +0.31% for inertia, and −1.06% for a +10 mm mediolateral alignment shift. The Peak Reaction Ratio showed greater sensitivity to mass (+2.95%) and inertia (+1.54%), while the corresponding alignment effect was smaller (+0.62%). Unlike the absolute loading outcomes, no single parameter consistently dominated both relative measures.
Across the 27 configurations, the FTI ratio and Peak Reaction Ratio showed a strong positive linear relationship (R2 = 0.753), as determined by ordinary least-squares linear regression with an intercept (Figure S1). However, their differing factorial contributions and normalized sensitivities indicate that they capture distinct aspects of the socket-to-ankle loading relationship. Given the temporal separation between socket and ankle maxima reported in Section 3.1, the Peak Reaction Ratio was retained as a secondary descriptive measure. Neither the Peak Reaction Ratio nor the FTI ratio was interpreted as a fraction of load transmitted between the two modeled locations.
4. Discussion
Previous prosthesis studies have primarily evaluated the mechanical effects of individual design or alignment parameters separately, including changes in prosthetic alignment, component configuration, and isolated perturbations of mass, inertia, and alignment [9,10,16,20]. Consequently, the relative contributions of these parameters when varied simultaneously, and whether their influence depends on the selected loading metric, remain insufficiently characterized. Previous experimental and biomechanical studies have demonstrated that prosthetic alignment and component configuration can influence lower-limb kinetics, gait mechanics, and limb clearance [10,20,21]. The present full-factorial analysis addressed this gap by quantifying the independent and interacting effects of mass, rotational inertia, and mediolateral alignment within a common fixed-kinematics framework.
The principal finding was that the relative contribution of each prosthesis parameter to outcome variation depended strongly on how loading was quantified. Within the investigated parameter ranges, prosthesis mass was the dominant contributor to the absolute reaction-force outcomes, representing 90.28–95.63% of the variation in peak reaction force and force–time integral at the socket–prosthesis junction and prosthetic ankle. In contrast, relative socket-to-ankle measures showed a more distributed sensitivity: the FTI ratio was influenced by mass, inertia, and alignment, whereas the Peak Reaction Ratio depended primarily on inertia and mass. These findings suggest that peak magnitude, cumulative loading, and relative loading ratios capture different aspects of prosthetic mechanical behavior. Importantly, the factorial contribution percentages are conditional on the perturbation ranges assigned to each parameter and should not be interpreted as universal rankings of biomechanical importance. The complementary range-normalized analysis nevertheless showed the same qualitative predominance of prosthesis mass for the absolute loading outcomes within the investigated parameter domain.
The strong mass dependence of the absolute outcomes is mechanically consistent with the fixed-kinematics design. Under identical prescribed motion and external loading, increasing prosthesis mass increases the gravitational and translational inertial forces that must be balanced by the model. Changes in rotational inertia primarily modify the rotational resistance of the prosthetic segments during acceleration and deceleration, whereas mediolateral alignment shifts alter the geometric relationship between the prosthetic segments and the resulting reaction forces. Because the same joint trajectories and external loads were imposed on all configurations, these results reflect parameter-dependent mechanical responses under standardized conditions. These findings should not be interpreted to mean that mass would dominate loading to the same extent during unconstrained gait, where individuals may adapt walking speed, segment motion, stance timing, and other movement strategies in response to prosthesis properties [22].
A different pattern emerged for the relative socket-to-ankle measures. For the FTI ratio, mass remained the largest contributor, but inertia and mediolateral alignment played a substantially greater role than they did for the absolute loading outcomes. The Peak Reaction Ratio showed an even stronger dependence on inertia. The interaction terms were comparatively small, indicating that the observed parameter dependence was driven predominantly by main effects. The mass × inertia (1.97%) and inertia × alignment (1.31%) interactions for the FTI ratio reflected modest non-additivity, with the latter showing opposite-signed interaction contrasts across alignment conditions. The Peak Reaction Ratio also showed directional asymmetry, although its interaction contributions were small. These effects did not change the overall interpretation of the relative loading metrics. Mechanically, these differences arise because relative measures depend on the responses at both modeled locations. Changes in rotational inertia modify the moment required to accelerate and decelerate the prosthetic segments and can therefore alter proximal and distal reactions to different extents. Similarly, mediolateral alignment changes the geometry of the prosthetic chain and the moment arms through which external and inertial forces act, which may affect proximal and distal reaction magnitudes differently even when the absolute resultant force changes only modestly. Previous studies in individuals with transfemoral amputation have similarly shown that relatively small translational or angular alignment changes can modify gait kinetics and ground-reaction loading, although the magnitude and direction of these effects vary across outcome measures and individuals [9,10].
The distinction between the FTI ratio and Peak Reaction Ratio was further supported by their different sensitivity profiles. Although the two measures were positively related across the 27 configurations, they were not equivalent. Socket and ankle force maxima were separated by 2.7–48.0% of the gait cycle, showing that the two peak values could arise during substantially different phases of gait. Consequently, their ratio reflects the relationship between the largest reaction magnitudes within a configuration, but not the instantaneous transmission of force from the ankle to the socket. The FTI ratio avoids this specific timing limitation by comparing cumulative loading over the same gait cycle, but it likewise represents a distinct mechanical quantity. More broadly, this supports the use of multiple complementary loading metrics in prosthetic biomechanics, since peak, cumulative, pressure, and shear measures may reflect different aspects of the residual-limb/socket mechanical environment [1,2,3,4,5,23,24].
The magnitude of the model-level socket reactions requires cautious interpretation. Across the investigated configurations, peak socket reactions ranged from 9.37 to 15.47 BW, compared with 3.45–4.57 BW at the prosthetic ankle. In the baseline configuration, the peak resultant reaction was 8.43 kN (12.45 BW) proximally and 2.71 kN (4.00 BW) at the prosthetic ankle, while the measured vertical ground reaction force peaked at approximately 1.20 BW. Component-wise inspection showed that the elevated proximal resultant was not attributable to a single isolated force component. These reactions therefore represent net rigid-body joint reactions required to satisfy the prescribed motion, segment inertial properties, and external loading, and should not be interpreted as direct physiological socket–interface forces.
The model did not include muscle-force estimation, distributed residual-limb/socket contact, or prosthetic component compliance. These omissions may affect both absolute reaction magnitudes and the apparent sensitivity to prosthesis mass and inertia. Confidence in the absolute values is further limited by the elevated pelvis residual forces, although the largest baseline residual force did not coincide with the proximal reaction peak. Published experimental and validated modeling studies report substantially lower directly measured transfemoral prosthetic interface loads, including peak forces of approximately 1.6 BW [6,25]. Accordingly, the present reactions are best interpreted as comparative model-level quantities within the fixed-kinematics framework, not as experimentally validated estimates of in vivo socket loading. Additional limitations include the use of a single healthy adult’s reference gait and the independent variation in mass and rotational inertia, which produced some factorial combinations that may not represent physically realizable prosthetic designs. The magnitude of the observed effects and the relative ranking of the investigated parameters may therefore depend on the specific gait pattern and loading conditions used. Furthermore, temporal downsampling assessed only the numerical robustness of outcome extraction, not uncertainty in the biomechanical model, since uncertainties related to marker tracking, segment mass properties, force-plate processing, joint definitions, and other modeling assumptions were not systematically propagated.
The present study extends our previous isolated-parameter analysis [16] by retaining the same baseline modeling framework while using a different participant and a new reference gait dataset. The full-factorial design enabled the simultaneous quantification of main effects and interactions across multiple loading metrics. In particular, the metric-dependent contributions of inertia and alignment to the FTI ratio and Peak Reaction Ratio provide insights beyond those obtainable from isolated parameter perturbations. Together, the factorial and range-normalized analyses demonstrate that parameter importance depends on the selected mechanical outcome.
A key strength of this study is that prosthesis mass, rotational inertia, and mediolateral alignment were evaluated simultaneously within the same mechanical framework, allowing their effects to be compared directly across several loading outcomes. This provides a more complete picture than evaluating each parameter separately. From a practical perspective, the results suggest that prosthesis optimization may depend on the outcome of interest: reducing absolute loading may require a different design priority than modifying the relative distribution of loading between the socket and distal prosthetic components. These findings may therefore help inform prosthesis design, component selection, and alignment strategies by identifying which mechanical parameters are most influential for specific loading outcomes. However, because the analysis was based on a deterministic model with prescribed kinematics and external loads, the findings should be used to guide further mechanistic investigation, not as direct clinical fitting recommendations.
Recent studies further emphasize the importance of considering prosthetic loading beyond rigid-body joint reactions. Paulus et al. demonstrated that transfemoral socket design can influence dynamic gait stability and joint loading [26], while Armitage et al. highlighted the challenges of relating socket interface pressure measurements to clinical outcomes [27]. Recent finite-element investigations have also shown that socket material properties and interface conditions influence local tissue stresses and structural responses [28,29]. Furthermore, Jahangir et al. demonstrated that uncertainties in musculoskeletal modeling inputs can substantially affect predicted mechanical outcomes [30]. Together, these findings support the need for subject-specific modeling, direct interface measurements, and independent validation when extending the present model-level sensitivity results toward clinical applications.
Future work should test whether these parameter-dependent patterns persist when gait adaptation, muscle forces, compliant prosthetic components, different gait speeds, alternative boundary and loading conditions and subject-specific residual-limb/socket mechanics are introduced, ideally validated experimentally in individuals with transfemoral amputation using instrumented prostheses or socket interface pressure and shear sensors. Independent computational approaches, including finite element analysis, could provide complementary verification of local stress and deformation behavior and help assess the robustness of the present rigid-body findings. Combining peak, time-integrated, and location-specific measures across a larger, more representative cohort would provide a stronger basis for linking computational sensitivity analyses to clinically meaningful prosthetic fitting and design decisions.
5. Conclusions
Prosthesis mass was the dominant determinant of absolute reaction-force magnitude and cumulative loading within the investigated parameter ranges, representing 90.28–95.63% of the variation in peak and FTI outcomes. Relative socket-to-ankle measures showed a more distributed sensitivity, with greater contributions from rotational inertia and mediolateral alignment. These findings describe deterministic within-model mechanical sensitivity under a fixed reference gait and a rigid-body prosthesis representation and do not account for muscle-force estimation or distributed residual-limb/socket contact. They should therefore not be interpreted as universal rankings of biomechanical importance or as direct clinical fitting recommendations. Future work should evaluate these relationships under more physiologically representative conditions and with independent experimental or finite-element validation.
Supplementary Materials
The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/bioengineering13121426/s1. Figure S1: Relationship between the Peak Reaction Ratio and socket-to-ankle force-time integral (FTI) ratio across the 27 prosthesis configurations. Colors indicate prosthesis mass scaling factors (0.75×, 1.00×, and 1.25×). The solid line represents the ordinary least-squares linear regression fit with an intercept (R2 = 0.753). Figure S2: Vertical ground reaction force profiles of the left and right limbs during representative stance phases from the reference gait trial. Forces are normalized to body weight (BW), and each stance phase is time-normalized to 0–100%. Figure S3. Component-wise joint reaction forces for the baseline prosthesis configuration at the socket–prosthesis junction and prosthetic ankle. Global X-, Y-, and Z-components and resultant force are shown in newtons (A,B) and body-weight-normalized units (C,D). Open circles indicate resultant peak forces of 8.43 kN (12.45 BW) proximally and 2.71 kN (4.00 BW) at the ankle. Table S1: OpenSim joint definitions and Joint Reaction settings used for the reported proximal prosthesis and prosthetic ankle reactions. Table S2: Shares of total variation (%) attributable to prosthesis mass, rotational inertia, mediolateral alignment, and their main and interaction effects across the 27 simulated configurations. Table S3: Prosthesis parameter settings and six primary loading outcomes for all 27 simulated configurations.
Author Contributions
Conceptualization, D.D. and A.D.; methodology, D.D.; formal analysis, L.P.; investigation, D.D.; resources, D.D., A.D., A.Ž. and L.P.; data curation, A.D. and L.P.; writing—original draft preparation, D.D.; writing—review and editing, D.D., D.J., A.Ž. and A.D.; visualization, D.D.; supervision, A.D. All authors have read and agreed to the published version of the manuscript.
Funding
This research has received funding from the Research Council of Lithuania (LMTLT), agreement No. S-ITP-26-2.
Institutional Review Board Statement
The study was approved by the Kaunas Regional Biomedical Research Ethics Committee (approval No. BE-2-13/2026, 27 January 2026). All procedures were conducted in accordance with the Declaration of Helsinki.
Informed Consent Statement
Written informed consent was obtained from the participant prior to data collection.
Data Availability Statement
The numerical inputs and per-configuration outcome data supporting this study are provided in the Supplementary Materials, including the parameter settings and six primary outcomes for all 27 simulated configurations. Additional model and analysis files are available from the corresponding author upon reasonable request.
Conflicts of Interest
The authors declare no conflicts of interest.
References
- Tiesler, L.M.; Edel, M.; Wang, F.; Pieroh, P.; Drossel, W.-G.; Carabello, A.; Zipplies, D.; Heyde, C.-E.; Schleifenbaum, S. Objective measurement methods for the evaluation of socket comfort in patients with transfemoral amputation: A systematic review. Front. Bioeng. Biotechnol. 2025, 13, 1576729. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Young, P.R.; Hebert, J.S.; Marasco, P.D.; Carey, J.P.; Schofield, J.S. Advances in the measurement of prosthetic socket interface mechanics: A review of technology, techniques, and a 20-year update. Expert Rev. Med. Devices 2023, 20, 729–739. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Paternò, L.; Truppa, L.; Ibrahimi, M.; Rosini, E.; Gruppioni, E.; Ricotti, L.; Menciassi, A. Quantitative analysis of interface pressures in transfemoral prosthetic sockets. Prosthet. Orthot. Int. 2024, 48, 176–183. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Devin, K.; Tang, J.; Moser, D.; Jiang, L. Dynamic socket interface mechanics for a transfemoral amputee during walking. Gait Posture 2023, 106, S47–S48. [Google Scholar] [CrossRef] [Scilit]
- Henao, S.C.; Orozco, C.; Ramírez, J. Influence of gait cycle loads on stress distribution at the residual limb/socket interface of transfemoral amputees: A finite element analysis. Sci. Rep. 2020, 10, 4985. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Schwarze, M.; Hurschler, C.; Seehaus, F.; Oehler, S.; Welke, B. Loads on the prosthesis-socket interface of above-knee amputees during normal gait: Validation of a multi-body simulation. J. Biomech. 2013, 46, 1201–1206. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ikeda, A.J.; Hurst, E.J.; Simon, A.M.; Finucane, S.B.; Hoppe-Ludwig, S.; Hargrove, L.J. The impact of added mass placement on metabolic and temporal-spatial characteristics of transfemoral prosthetic gait. Gait Posture 2022, 98, 240–247. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Hekmatfard, M.; Farahmand, F.; Ebrahimi, I. Effects of prosthetic mass distribution on the spatiotemporal characteristics and knee kinematics of transfemoral amputee locomotion. Gait Posture 2013, 37, 78–81. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Zhang, T.; Bai, X.; Liu, F.; Fan, Y. Effect of prosthetic alignment on gait and biomechanical loading in individuals with transfemoral amputation: A preliminary study. Gait Posture 2019, 71, 219–226. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Groeneveld, A.M.G.; Jonkergouw, N.; Bruijn, S.M.; Houdijk, H.; Kooiman, V.G.; Leijendekkers, R.A.; Prins, M.R. The effect of prosthetic alignment on lower limb kinetics in people with a transtibial bone-anchored prosthesis: An experimental within-subject study. Gait Posture 2025, 117, 274–282. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Hammoody, S.S.; Resan, K.K.; Muhammad, A.K.; Njim, E.K.; Madan, R. Innovative fracture-controlled safety device for osseo-integration-based prosthetic limbs. J. Med. Eng. Technol. 2026, 50, 415–427. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Seth, A.; Hicks, J.L.; Uchida, T.K.; Habib, A.; Dembia, C.L.; Dunne, J.J.; Ong, C.F.; DeMers, M.S.; Rajagopal, A.; Millard, M.; et al. OpenSim: Simulating musculoskeletal dynamics and neuromuscular control to study human and animal movement. PLoS Comput. Biol. 2018, 14, e1006223. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Rajagopal, A.; Dembia, C.L.; DeMers, M.S.; Delp, D.D.; Hicks, J.L.; Delp, S.L. Full-body musculoskeletal model for muscle-driven simulation of human gait. IEEE Trans. Biomed. Eng. 2016, 63, 2068–2079. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Raveendranathan, V.; Kooiman, V.G.M.; Carloni, R. Musculoskeletal model of osseointegrated transfemoral amputees in OpenSim. PLoS ONE 2023, 18, e0288864. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Carloni, R.; Luinge, H.; Raveendranathan, V. The gait1415+2 OpenSim musculoskeletal model of transfemoral amputees with a generic bone-anchored prosthesis. Med. Eng. Phys. 2024, 123, 104091. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Daublys, D.; Janosky, J.; Puodžiukynas, L.; Domeika, A. Mechanistic analysis of joint reaction forces to lower-limb prosthesis mass, inertia, and alignment. Prosthesis 2026, 8, 37. [Google Scholar] [CrossRef] [Scilit]
- Hood, S.; Ishmael, M.K.; Gunnell, A.; Foreman, K.B.; Lenzi, T. A kinematic and kinetic dataset of 18 above-knee amputees walking at various speeds. Sci. Data 2020, 7, 150. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Schreiber, C.; Moissenet, F. A multimodal dataset of human gait at different walking speeds established on injury-free adult participants. Sci. Data 2019, 6, 111. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Hicks, J.L.; Uchida, T.K.; Seth, A.; Rajagopal, A.; Delp, S.L. Is my model good enough? Best practices for verification and validation of musculoskeletal models and simulations of movement. J. Biomech. Eng. 2015, 137, 020905. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Köhler, T.M.; Pobatschnig, B.; Bellmann, M. Influencing Factors on Swing Phase Ground Clearance in Transfemoral Prostheses—A Biomechanical Study. Gait Posture 2025, 122, 389–397. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ravari, R.; Rehani, M.; Hebert, J.S. Biomechanical Characteristics of Transfemoral Bone-Anchored Prostheses during Gait: A Review of Literature. Prosthet. Orthot. Int. 2024, 48, 412–421. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Olaya-Mira, N.; Gómez-Hernández, L.M.; Viloria-Barragán, C.; Soto-Cardona, I.C. Methods to assess lower limb prosthetic adaptation: A systematic review. J. Neuroeng. Rehabil. 2025, 22, 100. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Herrin, K.R.; Kwak, S.T.; Rock, C.G.; Chang, Y.H. Gait quality in prosthesis users is reflected by force-based metrics when learning to walk on a new research-grade powered prosthesis. Front. Rehabil. Sci. 2024, 5, 1339856. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ibarra Aguila, S.; Sánchez, G.J.; Sauvain, E.E.; Alemon, B.; Fuentes-Aguilar, R.Q.; Huegel, J.C. Interface pressure system to compare the functional performance of prosthetic sockets during the gait in people with trans-tibial amputation. Sensors 2020, 20, 7043. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Frossard, L.; Laux, S.; Geada, M.; Heym, P.P.; Lechler, K. Load applied on osseointegrated implant by transfemoral bone-anchored prostheses fitted with state-of-the-art prosthetic components. Clin. Biomech. 2021, 89, 105457. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Paulus, P.; Gale, T.; Mroz, K.; Yatsenko, Y.; McKernan, G.; Buffat, D.; Fiedler, G.; Anderst, W. Influence of socket design on dynamic gait stability and joint loading in individuals with transfemoral amputations. JPO J. Prosthet. Orthot. 2026. advance online publication. [Google Scholar] [CrossRef] [Scilit]
- Armitage, L.; Cho, K.; Sparling, T.L.; Sleeman, T.; Alici, G.; Sreenivasa, M. The use of residual lower limb-prosthetic socket interface pressure measurements to inform clinical decisions: A systematic review. J. Biomech. 2026, 199, 113187. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Liu, F.; Marinopoulos, T.; Song, Q.; Yan, J.; Fong, D.T.P.; Li, S. An evaluated gait-driven finite element framework for transfemoral limb-socket interface biomechanics and socket design evaluation. J. Neuroeng. Rehabil. 2026. advance online publication. [Google Scholar] [CrossRef] [Scilit]
- Liu, F.; Fong, D.T.P.; Li, S. Balancing comfort and safety in transfemoral prosthetic sockets through finite-element modelling of interface mechanics. Procedia Struct. Integr. 2026, 77, 215–220. [Google Scholar] [CrossRef] [Scilit]
- Jahangir, S.; Bosch, W.; Esrafilian, A.; Mononen, M.E.; Tanska, P.; Stenroth, L.; Henriksen, M.; Alkjær, T.; Korhonen, R.K. Effect of uncertainties in musculoskeletal modeling inputs on sensitivity of knee joint finite element simulations. Med. Eng. Phys. 2025, 138, 104313. [Google Scholar] [CrossRef] [Scilit] [PubMed]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.