Research on Laser Automatic Phase−Locking Technology for Atomic Interferometric Gravity Gradient Measurement

,
and
1
School of Instrumentation and Optoelectronic Engineering, Beihang University, Beijing 100191, China
2
AVIC Changcheng Institute of Metrology and Measurement, Beijing 100095, China
*
Author to whom correspondence should be addressed.

Abstract

Atomic interferometric gravity gradient measurement enables atomic interference by manipulating atoms with lasers of specific frequencies. Thus, the frequency and phase−locking performance of the laser system exerts a significant impact on key experimental parameters, including the loading rate and ultimate cooling temperature of atomic clouds, the state selection efficiency of Raman transitions, the contrast of atomic interference fringes, and the level of detection noise. As atomic interferometric gravity gradient measurement transitions from static laboratory measurements to mobile field operations, conventional laser frequency and phase−locking methods struggle to meet the demand for rapid re−locking after device movement and cannot achieve timely system recovery in the event of laser unlocks. This work proposes an automatic laser frequency and phase−locking system that can detect real−time deviations in laser frequency and phase and implement rapid and precise corrections. Meanwhile, by utilizing the reference signal source in the optical phase−locked loop, the system realizes laser frequency hopping to satisfy the diverse laser frequency requirements across all stages of atomic interferometric gravity gradient measurement.

1. Introduction

Atomic interferometry technology, with its ultra−high sensitivity to gravitational fields and measurement accuracy, has emerged as the most promising technical solution in the field of gravitational acceleration and gravity gradient measurement [1]. The atomic interference gravity gradiometer based on this technology is internationally recognized as a novel principal gradiometer with immense development potential, owing to its advantages of high precision, zero drift, excellent traceability, and lack of mechanical wear [2,3].
The atomic interference gravity gradient measurement apparatus employs lasers to manipulate and detect atoms [4,5,6]. The performance of the laser system exerts a crucial influence on various parameters that determine the ultimate measurement accuracy, such as the loading rate and minimum cooling temperature of atomic ensembles, the state selection efficiency of Raman transitions, the contrast of atomic interference fringes, and the magnitude of detection noise [7]. To ensure the stability of the atomic interference process and measurement accuracy [8], traditional atomic interference gravity gradiometers rely on manual judgment to identify frequency and phase−lock loss. This requires substantial labor input to maintain system operation during long term field tests [9]. Meanwhile, their optical systems are typically designed with a large−scale, fixed layout, which severely restricts their application in various mobile measurement scenarios [10,11,12].
In existing atomic interferometry gravity gradiometry devices, the optical phase−locked loop (OPLL) has become the mainstream preferred scheme for generating coherent laser light in atomic physics experiments owing to its advantages such as a wide locking frequency range, low phase noise, pure output laser spectrum, and absence of additional sidebands. In an outdoor movable measurement environment, the lack of laser phase locking will directly lead to severe distortion of the atomic interference signal, making it impossible to effectively extract the gravity gradient measurement results via ellipse fitting. However, in current research on phase locking, no public achievements on automatic phase locking dedicated to atomic interferometry gravity gradiometers have been reported, and existing methods still rely on manual adjustment, which is not only cumbersome but also requires rescanning and relocking after each loss of lock, resulting in long system startup times and failure to achieve long−term stable operation; furthermore, special personnel are required to monitor the locking status in real time, and neither locking time consumption nor the recovery timing can be guaranteed.
Aiming at the core contradiction that high precision and excellent mobility are mutually exclusive in traditional atomic interferometric gravity gradiometers, the miniaturized, integrated, and automated design of both the core optical system and the frequency–phase−locking system constitutes the key to overcoming the mobility bottleneck in atomic interferometric gravity gradient measurement. Focusing on the optimized design of the core optical system, this paper proposes an auto−locked laser system with synchronized frequency and phase regulation, tailored specifically for atomic interferometric gravity gradient measurement. This system significantly reduces the volume and weight of the core optical module, thereby laying a solid technical foundation for the development of mobile atomic interferometric gravity gradiometers.
Compared with traditional laboratory−based atomic interferometric gravity gradient measurement experiments, mobile atomic interferometric gravity gradient measurement imposes new requirements on the instrument. In the process of gravity gradient measurement via atomic interferometry in the laboratory, frequency locking and phase locking need to be performed on two laser beams separately. The locking method relies on manually turning on the laser scanning voltage, and then locking the voltage according to the zero point of the frequency−locking or phase−locking error signal, thereby locking the frequency and phase of the laser. In mobile gravity gradient measurement, it is necessary to frequently change measurement positions to obtain more gravity gradient information in the region. The traditional manual frequency and phase−locking method cannot meet the requirement of quickly locking the laser frequency and phase after each movement. Meanwhile, during the measurement process, if a laser unlocking fault occurs, the traditional frequency and phase−locking method cannot restore the system in a timely manner, leading to an increase in bad data points that affect the measurement results.
The automatic laser frequency and phase−locking technique studied in this paper adopts a digital closed−loop control architecture to lock and correct laser frequency and phase deviations in real time. Meanwhile, taking advantage of the fast closed−loop capability of the digital laser phase−locking system [13,14,15], changing the reference frequency of the signal source in the optical phase−locked loop can realize different laser frequency jumps required in atomic interferometric gravity gradient measurement experiments, which simplifies the optical system of the entire gravity gradiometer and facilitates mobile measurement.

2. Materials and Methods

The process of completing an atomic interferometric gravity gradient measurement involves the following key steps: First, atoms are loaded into a magneto−optical trap (MOT) and cooled. Subsequently, during the atomic free fall process, the atoms are prepared into the initial state required for the experiment. Next, laser manipulation of the atoms is performed to complete the atomic interferometry experiment. Finally, atomic interference fringes are obtained through final−state detection. A coherent laser is indispensable throughout the entire process, with stringent requirements imposed on laser parameters at different stages of atomic interferometry. Furthermore, noise introduced by the laser phase constitutes one of the primary noise sources in gravitational acceleration measurement via atomic interferometry. Consequently, the quality of frequency and phase−locking for the laser system plays a decisive role in the final results of atomic interferometry.

2.1. Laser Frequency Design for Atomic Interferometry Based on Raman Light Blue Detuning

In the atom interferometric gravity gradiometry system, the Raman lasers refer to two coherent laser beams with a frequency difference of approximately 6.8 GHz and large red detuning relative to the excited state of the atomic D2 line. These lasers are employed for atomic state selection and matter−wave interference, and the phase of the Raman light directly affects the phase of the atom interferometer. In this study, a blue detuning scheme for Raman light is adopted, where the Raman laser frequency is set at a blue−detuned position relative to the resonance frequency corresponding to the ground−state−to−excited−state transition of 85Rb atoms with F’ = 3. This design significantly simplifies the preparation procedure of the Raman laser. By contrast, the red detuning scheme for Raman lasers is predominantly employed in traditional atomic interferometry experiments, with the laser frequency typically red−detuned relative to the resonance frequency of the ground−state−to−excited−state transition with F’ = 3. Moreover, to ensure that the detuning of the Raman laser from the nearest atomic resonance energy level exceeds 500 MHz, the frequency difference between the repumping laser and the cooling laser must be maintained at approximately 1 GHz. This conventional design not only complicates the preparation procedure of the Raman laser but also introduces substantial laser transmission loss, thereby reducing the optical power utilization efficiency of the system.
When designing the frequency locking point, it is necessary to take into account the frequency requirements of different stages in the atomic interference process. There exist significant differences in the frequency detuning and power parameters of cooling light between the MOT preparation stage and the polarization gradient cooling (PGC) stage: For MOT construction, the frequency detuning of the cooling light should be controlled at approximately three times the natural linewidth of the atoms (∆ν ≈ 3Γ) [16,17,18], and the optical power should be set to 3–5 times the saturation power. In contrast, during the polarization gradient cooling process, to achieve efficient deep cooling, the frequency detuning of the cooling light must exceed ten times the natural linewidth (∆ν > 10Γ), while the optical power needs to be reduced to around 0.5 times the saturation power [19].
There exists an essential distinction in the frequency tuning ranges between the Raman laser and the cooling laser: The Raman laser must satisfy the far detuning condition, with its frequency detuning from the nearest atomic resonance energy level typically required to exceed 500 MHz, so as to suppress the interference of resonant scattering on the interferometry process. In contrast, the cooling laser needs to operate near the atomic resonance frequency, achieving efficient atomic deceleration and cooling via near−resonant interaction. In addition, the Raman laser should feature ultra−low phase noise to ensure the coherence of quantum state evolution during atomic interferometry [20]. Meanwhile, to compensate for the influence of Doppler shift induced by atomic motion on the Raman resonance condition, precise frequency chirp modulation must be applied to the Raman laser [21,22,23].

2.2. Design of Laser Frequency–Phase−Locking Scheme

In this study a master–slave laser frequency−locking scheme was adopted. Specifically, the frequency of the master laser was stabilized using the modulation transfer spectroscopy (MTS) technique [24,25]. The output frequency of the slave laser was locked to that of the master laser via an OPLL system [26,27]. By adjusting the frequency of the OPLL reference signal source, the output frequency of the slave laser can be tuned to meet the frequency variation requirements for cooling light/polarization gradient cooling as well as the frequency chirp requirements for Raman laser.
The overall scheme for realizing atomic interferometric gravity gradient measurement using automatic laser frequency and phase locking is shown in Figure 1. The laser outputs two beams, namely the master laser and the slave laser. The master laser is split and then fed into the integrated automatic laser frequency−locking module designed in this study to achieve automatic frequency locking. After the frequency of the master laser is locked, the slave laser is automatically locked to a preset frequency point by using an optical phase−locked loop. The frequency and phase of the two laser beams are then controlled and stabilized by the digital system. By utilizing the loop characteristics of the phase−locked slave laser, the optical phase−locked loop is adjusted to realize the required frequency switching of the lasers during the atomic interference process, thereby enabling atomic interferometric gravity gradient measurement.
Figure 1. Overall scheme of atomic interferometric gravity gradient measurement using automatic laser frequency–phase locking.

2.2.1. Laser Frequency Locking

During the design of the frequency−locking scheme for the master laser, the laser is tuned to a frequency near to that of the repump laser. A portion of the master laser beam is separated by a beam splitter dedicated to frequency stabilization. After being frequency−shifted by a fiber−based electro−optic modulator (EOM) with a frequency shift of approximately 4 GHz, the −1st−order sideband is locked to the resonance peak of the 85Rb D2 line corresponding to the transition F = 3 → F′ = 4 using modulation transfer spectroscopy (MTS). At this point, the 0th−order component of the EOM represents the main output frequency of the laser, which is blue−detuned by 400 MHz relative to the resonant transition frequency F = 1 → F′ = 2 of the 87Rb D2 line.
Since only two error signals are available in the MTS spectrum, corresponding to 87Rb F = 2 → F′ = 3 and 85Rb F = 3 → F′ = 4 (with essentially no error signal near the repump laser frequency), and considering that the error signal peak of 85Rb exhibits a larger amplitude, the above frequency−shifted locking scheme is adopted in this work.
As illustrated in Figure 2, within the automatic frequency−locking module, the master laser injects laser light into the MTS optical path module. After undergoing modulation and demodulation, the optical signal is fed into the signal acquisition module, where it is collected by the analog−to−digital converter. Meanwhile, the error signal is transmitted to both the error detection module and the oscilloscope. The error detection module performs closed−loop control on the acquired feedback error via the PID algorithm and then feeds the regulated signal back to the laser. The error adjustment function of the PID algorithm is utilized to ensure the stability of the output laser frequency of the laser.
Figure 2. The operational principle and signal flow of automatic frequency locking for the master laser.

2.2.2. Laser Phase Locking

The slave laser is locked to the reference signal source with the master laser by means of the OPLL method. Changing the frequency of the reference signal source can conveniently vary the frequency of the slave laser.
As illustrated in Figure 3, for the phase−locking module, when a specific frequency offset exists between the laser to be locked and the reference laser, the coherent mixing of these two laser beams enables the photodetector to capture the corresponding beat frequency signal. Subsequently, the phase detector performs frequency and phase discrimination by comparing this beat frequency signal with a pre−calibrated reference frequency signal that is set to the desired offset frequency. This process yields a phase discrimination error voltage signal proportional to the phase difference between the two lasers. After being processed by the loop filter, this voltage signal is fed back to the laser to be locked for frequency tuning. As a result, the frequency of the target laser is dynamically synchronized with that of the reference laser, thus achieving stable offset phase locking between the two laser sources.
Figure 3. The operational principle and signal flow of automatic phase locking for the slave laser.
By virtue of the rapid feedback effect of automatic laser phase locking, the frequency of the slave laser can be adjusted by changing the frequency of the reference signal source in the OPLL, in combination with the requirements for miniaturization and portability of the optical path system and radio frequency control circuit in mobile gravity gradient measurement.
Frequency hopping illustrated in Figure 4. All laser frequency adjustments during atomic trapping and manipulation experiments are achieved by varying the frequency of the reference signal source in the OPLL. At the same time, the fiber AOM is capable of being configured for diffracted light output with a designated frequency, thus leading to a substantial reduction in the complexity of the RF driving module that provides the AOM driving frequency.
Figure 4. Frequency evolution of master/slave lasers and OPLL reference signal source during frequency switching.

2.3. Implementation of Automatic Frequency Locking of Lasers

For frequency locking of the main laser, modulation transfer spectroscopy with superior frequency−locking stability was employed. The digital control system for manipulating digital signals is required to possess three core functions: signal acquisition, data calculation, and control signal output. Specifically, the acquisition module is responsible for identifying the optical signals collected by the PD; the calculation module undertakes the search and judgment of the error zero point after demodulation; and the output module is dedicated to regulating the control terminal of the laser. The photograph of the assembled automatic laser frequency locking module is shown in Figure 5. The optical path and digital control system for automatic frequency locking are integrated into a compact box with dimensions of 14 cm × 7 cm × 23.5 cm.
Figure 5. Integrated automatic frequency−locking module for master laser.
The scanning process of the digital system is divided into two phases: coarse scanning and fine scanning. Coarse scanning refers to the wide−range voltage scanning driven by the digital control system, whose primary function is to locate the zero point of the error function. After the zero point of the error signal is identified via coarse scanning, fine scanning is executed based on the located position. Fine scanning is a narrow−range scanning process that refines the voltage value corresponding to the zero point of the error signal, thereby providing more precise voltage information for the subsequent PID closed−loop control. On the one hand, wide−range coarse scanning ensures rapid localization of the zero point; on the other hand, fine scanning effectively avoids the distortion of the error function caused by an excessively large scanning voltage step, enabling accurate determination of the exact position of the zero point. In addition, the two−stage scanning strategy contributes to saving the microcontroller memory space.
Through three rounds of scanning with gradually refined precision from coarse to fine, the zero point of the MTS error signal is accurately identified, and the master laser can be locked at the designated wavelength within 80 ms. During the entire automatic frequency locking process, the variations in the scanning voltage and the MTS error signal are shown in Figure 6.
Figure 6. Determination of the master laser locking point and subsequent laser frequency locking through coarse−to−fine control voltage scanning.

2.4. Implementation of Automatic Laser Phase Locking

Once frequency locking of the master laser is accomplished, beat frequency detection is carried out between the master and slave laser beams. The differential signal generated by the beat frequency is processed by frequency and phase discrimination, and then injected into the PID controller as the phase−locking error signal of the laser. The servo controller collects the frequency discrimination error signal produced after demodulation and adjusts the control voltage of the laser to drive the output laser to approach the target frequency. It is worth noting that the phase−locking error signal is derived from frequency and phase discrimination of the differential frequency signal. However, high−frequency signal detection also imposes stringent requirements on signal detectors and subsequent processing circuits. Therefore, frequency modulation is required for the obtained 6.834 GHz differential frequency signal. In the experiment, the 6.834 GHz differential frequency signal is mixed with a 6.784 GHz signal, and the resulting mixed signal is subjected to frequency and phase discrimination against the 100 MHz reference signal generated by a direct digital synthesizer. The assembled automatic phase−locking module is shown in Figure 7. The entire system is integrated into a box with dimensions of 14 cm × 9 cm × 23.5 cm.
Figure 7. Integrated automatic phase−locking module for slave laser.
The scanning process for the phase−locking error signal is identical to that for frequency locking. The output voltage is uniformly swept from −5 V to +5 V; after the voltage stabilizes, the corresponding error signal voltage value is recorded. Similarly, 200 data points are sampled and stored in two one−dimensional arrays, and there is no need to detect the relative positions of the signal peaks and troughs. The PID controller is then activated to achieve the goal of frequency stabilization.
Since the automatic phase−locking module is not only used for slave laser locking but also required to perform frequency hopping to generate laser outputs for different stages, it imposes high requirements on the bandwidth of the PID controller. Therefore, a dual−loop PID control architecture is adopted. Specifically, the inner loop is a fast loop, typically implemented by an analog PID controller to enable rapid dynamic response; the outer loop is a slow loop, controlled by a microcontroller, which is primarily designed to avoid large−scale frequency shifts caused by system drifts.
Figure 8 displays the error signal curves with the fast loop deactivated (left) and activated (right) after the initiation of scanning. It can be observed that the position corresponding to the zero point of the central error signal remains unchanged, which has a negligible impact on the final lock point controlled by the digital system. Therefore, the fast loop can be directly activated for scanning and lock point determination. This method effectively eliminates the phase noise interference introduced by PID switches or relays in the circuit.
Figure 8. Optical phase−locked loop error signal under control voltage scanning: (a) fast loop off; (b) fast loop on.
In the process of automatic laser phase locking using a digital control system, identifying the slave laser locking point represents a key challenge in the research field of automatic phase locking. To tackle this problem, a characteristic analysis of the phase−locking error curve must first be performed. Based on the error signals presented in Figure 8, the slave laser locking point exhibits the following characteristics: first, the overall profile of the error signal is approximately symmetric; second, the voltage difference near the zero point of the error signal is a fixed value, which is independent of the scanning voltage and determined solely by the optical path configuration; finally, relatively flat high−level signals are observed on both sides of the zero point of the error signal.
Based on the characteristics of the phase−locking error signal, the operational principle of the phase−locking error zero−point search function implemented in the digital control system is as follows: First, activate the scanning voltage. After the error signal is generated, each scanning voltage value corresponds to a specific error signal value according to the data collected by the digital control system. Given the distinct jump characteristics of the maximum and minimum values of the error signal curve relative to the zero point, the scanning voltage interval corresponding to the zero point of the error signal can be rapidly identified upon completion of scanning and data analysis.
Second, accurately calculate the voltage signals of the error signal curve. After locating the interval corresponding to the zero point of the error signal in the first step, calculate the voltage difference between every two consecutive data points of the error signal sequentially, denoted as ∆V = |Vi − Vi−1| (where i = 2, 3, …, n). The multiple voltage differences corresponding to the central lock−waiting zero−point interval of the error signal are fixed values, whereas other voltage differences deviate significantly from this fixed value. The voltage interval [Vi−1, Vi] is defined as the slave laser locking point interval when multiple ∆V values with minimal variation are identified.
Third, once the lock−waiting voltage interval [Vi−1, Vi] of the error signal is determined, the scanning voltage signal corresponding to the midpoint of this interval is designated as the slave laser locking point. The PID loop is then activated immediately to lock onto the zero point of the error signal. During the entire automatic phase−locking process, the voltage control signal input to the laser and the error signal of the OPLL are shown in Figure 9.
Figure 9. Determination of the OPLL error signal zero point and subsequent laser phase locking through three−step coarse−to−fine control voltage scanning.
In practical operation, the locking process is divided into three scanning phases, where the scanning voltage value corresponding to the zero point of the error signal is determined progressively with refined precision, followed by the activation of closed−loop locking.
To prevent locking errors, a verification step is required for the identified target lock point. The verification criteria are as follows: first, as can be observed from the complete error signal profile (Figure 8), the voltage difference intervals on both sides of the error signal zero point are essentially free from any interference. This represents the most distinctive feature that differentiates the lock−waiting error signal interval from other local minimum intervals. Therefore, the number of data points within the identified lock−waiting interval should be greater than 30. Second, abrupt voltage differences should exist simultaneously in both the left and right segments of the zero−point interval after locking.
Once frequency and phase locking of the slave laser is achieved, changing the frequency of the reference signal source enables the phase−locked loop to unlock and then rapidly relock in a very short time. The amplitude of the frequency change of the reference signal source determines the rapid relocking time after unlocking. The phase−locked loop relocking time under the maximum frequency jump is employed to characterize the fast frequency−switching performance of the overall system. The reference frequency of the phase−locked loop is changed from 6.761 GHz to 6.840 GHz (corresponding to the laser frequency jump during polarization gradient cooling); one branch of the voltage feedback control signal output from the automatic phase−locking module to the laser and the error signal are connected to an oscilloscope to record the time when the error signal jumps and quickly returns to the original value, as shown in Figure 10. The automatic phase−locked loop can achieve rapid relocking within 200 μs. For the frequency change corresponding to polarization gradient cooling in atomic interference experiments, this extremely short unlocking time achieves fast switching of the slave laser frequency without causing additional adverse effects.
Figure 10. Fast relocking of the loop within 200 μs during frequency hopping of the OPLL reference signal source.
Based on the characteristics of this fast feedback locking mechanism, rapid control of locked frequency hopping can be achieved during laser manipulation of atomic interferometry. By coordinating with the laser activation timing, fast switching between functional lasers for different purposes can be realized.

3. Results

After optimizing the laser automatic frequency and phase−locking system, we applied it to the atom interferometric gravity gradiometer and established a complete experimental verification platform. On this basis, two typical procedures in the atom interferometric gravity gradient measurement experiment were carried out: atomic trapping and obtaining the gravity gradient ellipse fitting signal through atom interference.
In the atom interferometric gravity gradient experiment, trapping atoms with a magneto−optical trap, performing polarization gradient cooling, and then detecting the falling fluorescence signal represent crucial steps for experimental implementation. This includes locking the laser frequency and phase separately, followed by rapid frequency switching via the fast−loop effect of the automatic phase−locking system to achieve polarization gradient cooling. Finally, the falling signal of the atomic ensemble is detected. By recording the signal of the falling atomic cloud with a photodetector, the atomic fluorescence signal curve shown in Figure 11 can be obtained:
Figure 11. Atomic cloud time−of−flight fluorescence signal: (a) upper interferometer; (b) lower interferometer.
Calculations show that both the upper and lower interferometers of the gradiometer can trap 3.08 × 107 atoms, with the temperature of the atomic cloud reaching 3.48 μK.
Following the experimental implementation of atomic cloud falling signal detection, we performed state selection and interference on the falling atoms. Ellipse fitting was carried out for the two interferometers to obtain the ellipse fitting results required for atomic interferometric gravity gradient measurement, as shown in Figure 12. After eliminating systematic offsets, the measured gravity gradient value was calculated to be 2976.8 E.
Figure 12. Atomic interferometric gravity gradient ellipse fitting achieved by automatic phase−locking frequency hopping.

4. Discussion

In this work, we developed an automatic laser frequency and phase−locking system suitable for mobile atom interferometric gravity gradient measurements, aiming to improve the measurement efficiency and long−term operational stability of the system. The system employs a digital control system to scan laser error signals and, based on a digital closed−loop control architecture, detects and corrects laser frequency and phase deviations in real time, thus achieving fast and precise locking. After system assembly and optimization, the frequency−locking and phase−locking subsystems can be integrated into metallic boxes with dimensions of 14 cm × 7 cm × 23.5 cm and 14 cm × 9 cm × 23.5 cm, respectively. Performance tests demonstrate that both the frequency locking of the master laser and the phase locking of the slave laser can be automatically and rapidly achieved within 80 ms. By taking advantage of the fast relocking characteristic of the optical phase−locked loop, the system enables rapid switching of the slave laser frequency. In the atom interferometric gravity gradient measurement experiment, the relocking time under the maximum frequency jump during the polarization gradient cooling stage can be controlled within 200 μs. Using this automatic laser frequency and phase−locking system, we obtained 3.08 × 107 atoms in a single interferometer region, with the atomic cloud temperature reduced to 3.48 μK. Finally, atom interferometric gravity gradient measurement was successfully implemented.
The main merit of this study is that it significantly improves the measurement efficiency of mobile atomic interferometric gravity gradiometers. We propose an automatic frequency and phase−locking method for mobile atomic interferometric gravity gradient measurement, which features both fast locking and long term monitoring capability. This method not only enhances the continuity and stability of the laser output during measurements, but also helps to simplify the operation process and improve the practicality of the whole system.
Nevertheless, the present study still has certain limitations. The optical setup for frequency locking and the circuit design for phase locking is not sufficiently integrated, which restricts further miniaturization and integration of the entire system. In future work, more compact and integrated optical and electronic schemes will be explored to promote the practical application and field deployment of mobile atomic interferometric gravity gradient measurement systems.

Author Contributions

Conceptualization, J.W. and J.B.; methodology, J.B.; software, J.W.; validation, J.W., J.B. and B.H.; data curation, J.W.; writing—original draft preparation, J.W.; writing—review and editing, B.H.; supervision, B.H.; project administration, J.B.; funding acquisition, J.B. All authors have read and agreed to the published version of the manuscript.

Funding

This study was supported by self-funded research funds.

Data Availability Statement

The raw data supporting the conclusions of this article will be made available by the authors on request.

Acknowledgments

The authors would like to express their sincere gratitude to Bangcheng Han for his constructive suggestions on the experimental setup of this research. Their thanks also go to Jinhai Bai for his valuable assistance in shaping the overall research framework and designing the experimental protocols.

Conflicts of Interest

The authors declare no conflicts of interest.

Abbreviations

The following abbreviations are used in this manuscript:
MOTMagneto−optical trap
MTSModulation transfer spectroscopy
OPLLOptical phase−locked loop
AOMAcousto−Optic Modulator
PGCPolarization gradient cooling

References

  1. Peters, A.; Chung, K.Y.; Chu, S. High−precision gravity measurements using atom interferometry. Metrologia 2001, 38, 25. [Google Scholar] [CrossRef] [Scilit]
  2. Mcguirk, J.M.; Foster, G.T.; Fixler, J.B.; Snadden, M.J.; Kasevich, M.A. Sensitive absolute−gravity gradiometry using atom interferometry. Phys. Rev. A 2002, 65, 033608. [Google Scholar] [CrossRef] [Scilit]
  3. Luo, Q.; Zhang, H.; Zhang, K.; Xiao, W.; Wang, Q.; Hu, Z.; Zhou, M.; Shao, J.; Chen, W. A compact laser system for a portable atom interferometry gravimete. Rev. Sci. Instrum. 2019, 90, 043104. [Google Scholar] [CrossRef] [Scilit]
  4. Zhu, H.; Huang, P.; Gao, B.; Chen, X.; Tang, B.; Zhou, Y.; Xiong, Z.; Lin, Q.; Hu, Y.; Luo, J. Miniaturized optical system for high−precision mobile atomic gravimeters. Opt. Express 2024, 32, 10. [Google Scholar] [CrossRef] [Scilit]
  5. Wu, X.; Pagel, Z.; Malek, B.S.; Holmes, M.; Hauth, N.; Oelpa, C.; Guldenschuh, C.; Wodey, E.; Schkolnik, C.; Müller, H. Gravity surveys using a mobile atom interferometer. Sci. Adv. 2019, 5, eaax0800. [Google Scholar] [CrossRef] [Scilit]
  6. Lellouch, S.; Bongs, K.; Holynski, M. Using atom interferometry to measure gravity. Contemp. Phys. A Rev. Phys. Assoc. Technol. 2022, 63, 138–155. [Google Scholar] [CrossRef] [Scilit]
  7. Hamilton, P.; Jaffe, M.; Brown, J.M.; Estey, B.; Müller, H. Atom interferometry in an optical cavity. Phys. Rev. Lett. 2015, 114, 100405. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  8. Huang, P.W.; Tang, B.; Chen, X.; Hu, Z.; Zhou, M.; Xiao, W.; Zhang, H.; Luo, Q. Accuracy and stability evaluation of the 85Rb atom gravimeter WAG−H5−1 at the 2017 International Comparison of Absolute Gravimeters. Metrologia 2019, 56, 045012. [Google Scholar] [CrossRef] [Scilit]
  9. Chen, Y.; Wang, W.; He, Y.; Sun, B.; Ma, Y.; Zhang, X.; Liu, L. Fiber−Optic Sagnac Interferometry for Gravity Gradient Measurements. IEEE Trans. Instrum. Meas. 2023, 72, 7000411. [Google Scholar] [CrossRef] [Scilit]
  10. D’Amico, G.; Rosi, G.; Zhan, S.; Sorrentino, F.; Tino, G.M. Canceling the Gravity Gradient Phase Shift in Atom Interferometry. Phys. Rev. Lett. 2017, 119, 253201. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  11. Hughes, S.A.; Thorne, K.S. Seismic gravity−gradient noise in interferometric gravitational−wave detectors. Phys. Rev. D 1998, 58, 603–609. [Google Scholar] [CrossRef] [Scilit]
  12. Sorrentino, F.; Lien, Y.H.; Rosi, G.; Tino, G.M. Sensitive gravity−gradiometry with atom interferometry: Progress towards an improved determination of the gravitational constant. New J. Phys. 2010, 12, 095009. [Google Scholar] [CrossRef] [Scilit]
  13. Hechenblaikner, G.; Wand, V.; Kersten, M.; Müller, J.; Clausnitzer, T.; Klar, H. Digital Laser Frequency Control and Phase−Stabilization Loops in a High Precision Space−Borne Metrology System. IEEE J. Quantum Electron. 2011, 47, 651–660. [Google Scholar] [CrossRef] [Scilit]
  14. Liu, Y.; Tong, S.; Chang, S.; Song, Y.; Dong, Y.; Zhao, X.; An, Z.; Yu, F. Design of a Phase Sensor Applied in the Optical Phase−Locked Loop Based on a High−Speed Coherent Laser Communication System. IEEE Access 2018, 6, 22131–22139. [Google Scholar] [CrossRef] [Scilit]
  15. Xu, Z.; Zhang, X.; Huang, K.; Wang, B.; Zhao, X.; Xiao, W. A digital optical phase−locked loop for diode lasers based on field programmable gate array. Rev. Sci. Instrum. 2012, 83, 181. [Google Scholar] [CrossRef] [Scilit]
  16. Li, A. Research Prospects for the Optimization of Magneto−Optical Trap Parameters for Cold Atom Interferometers. Appl. Sci. 2024, 14, 7062. [Google Scholar] [CrossRef] [Scilit]
  17. Veshapidze, G.; Bang, J.Y.; Fehrenbach, C.W.; Nguyen, H.; DePaola, B.D. Model−free measurement of the excited−state fraction in a 85Rb magneto−optical trap. Phys. Rev. A 2015, 91, 053423. [Google Scholar] [CrossRef] [Scilit]
  18. Kurtsiefer, S.C. Polarization gradient cooling of single atoms in optical dipole traps. Phys. Rev. A 2017, 96, 033406. [Google Scholar] [CrossRef] [Scilit]
  19. Birkl, G.; Yeazell, J.A.; Rückerl, R.; Raab, E. Polarization Gradient Cooling of Trapped Ions. Europhys. Lett. 1994, 27, 197–202. [Google Scholar] [CrossRef] [Scilit]
  20. Wang, Y.; Yan, S.; Xu, D.; Li, R.; Wang, K.; Zhan, M. Measuring the phase noise of Raman lasers with an atom−based method. Rev. Sci. Instrum. 2021, 92, 033203. [Google Scholar] [CrossRef] [Scilit]
  21. Wang, G.; Wang, E.; Wang, Y.; Liu, J.; Zhang, X.; Zhang, H.; Yang, M.; Zhu, L. A fiber−integrated and power−regulated single−sideband modualted Raman laser source for atom interferometry. Opt. Quantum Electron. 2025, 57, 37. [Google Scholar] [CrossRef] [Scilit]
  22. Lee, K.−S.; Kim, J.; Lee, S.B.; Park, S.E.; Kwon, T.Y. A phase−modulated laser system of ultra−low phase noise for compact atom interferometers. J. Korean Phys. Soc. 2015, 67, 318–322. [Google Scholar] [CrossRef] [Scilit]
  23. Wang, K.; Yao, Z.; Li, R.; Lu, S.; Chen, X.; Wang, J.; Zhan, M. Hybrid wide−band, low−phase−noise scheme for raman lasers in atom interferometry by integrating an acousto−optic modulator and a feedback loop. Appl. Opt. 2016, 55, 989. [Google Scholar] [CrossRef] [Scilit]
  24. McCarron, D.J.; King, S.A.; Cornish, S.L. Modulation transfer spectroscopy in atomic rubidium. Meas. Sci. Technol. 2008, 19, 252–253. [Google Scholar] [CrossRef] [Scilit]
  25. Preuschoff, T.; Schlosser, M.; Birkl, G. Optimization strategies for modulation transfer spectroscopy applied to laser stabilization. Opt. Express 2018, 26, 24010–24019. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  26. Ristic, S.; Bhardwaj, A.; Rodwell, M.J.; Coldren, L.A.; Johansson, L.A. An optical phase−locked loop photonic integrated circuit. J. Light. Technol. 2010, 28, 526–538. [Google Scholar] [CrossRef] [Scilit]
  27. Gliese, U.; Nielsen, N.T.; Bruun, M.; Christensen, E.; Stubkjaer, K.E.; Lindgren, S.; Broberg, B. A wideband heterodyne optical phase−locked loop for generation of 3−18 ghz microwave carriers. IEEE Photonics Technol. Lett. 1992, 4, 936–938. [Google Scholar] [CrossRef] [Scilit] [PubMed]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Article Metrics

Citations

Article Access Statistics

Multiple requests from the same IP address are counted as one view.