TinyML Benchmarking for Embedded sEMG Gesture Classification: Quantized Model Design and an Accessible Robotic Hand Prosthesis Demonstration

,
,
,
and
Electronic Engineering Program, Faculty of Natural Sciences and Engineering, Unidades Tecnologicas de Santander, Bucaramanga 680005, Colombia
*
Author to whom correspondence should be addressed.

Abstract

Surface electromyography (sEMG) enables non-invasive prosthetic control, but deployment is limited by non-stationary signals, noise and embedded-computing constraints. This work presents a mathematically defined pipeline for real-time control of an accessible, commodity-component robotic hand prosthesis (no formal bill-of-materials or aggregate cost figure is claimed) from eight-channel Myo Armband sEMG, together with a deployment-oriented benchmarking study of quantized TinyML classifiers, benchmarked independently of the physical prosthesis rather than executed on it. The sEMG stream is filtered, windowed and classified by a hybrid CNN-BiLSTM reference model, converting posterior probabilities into PWM prosthetic commands on a Raspberry Pi. Five gesture classes were considered, using data from a single able-bodied participant, framed as a within-subject feasibility study. From 167,712 labeled samples, the reference model reached 99.6% test accuracy (99.58% macro F1) on a session-disjoint split. Because this reference model targets a Raspberry Pi rather than a microcontroller, four quantized TinyML candidates (TCN ensemble, embedded-C Random Forest, MLP ensemble, decision tree) were additionally designed and benchmarked under a common accuracy-footprint-inference-time protocol. Quantization cost was substantial: the deployed int8 TCN reached 83.5% smoothed accuracy versus 90.1% float32, while an embedded-C Random Forest reached 90.7% at only 17.9 KB. SHAP analysis revealed model-dependent channel importance patterns, with the TCN ensemble and the Random Forest ranking electrode channels differently because they exploit different signal representations. Embedded operation and a qualitative functional demonstration on the physical prosthesis, assessed by visual observation rather than instrumented kinematics, grasp force or task-success measurements, confirmed feasibility, while cross-subject validation, direct on-device latency measurements and component-wise ablation remain necessary before clinical or daily-use claims.

1. Introduction

Upper-limb prosthetic control is a central problem in assistive technologies because it determines whether an artificial limb becomes an active functional interface. Surface electromyography (sEMG) is relevant because it records residual muscle activity non-invasively and can infer movement intention [1,2,3,4,5,6]. The difficulty is that sEMG is not a stationary command signal: amplitude and spectral content vary with electrode placement, contraction intensity, fatigue, skin impedance, limb posture and motion artifacts [7,8,9,10]. A prosthetic controller must therefore estimate gesture intention from short noisy temporal observations and convert that estimate into stable actuator commands. Deep learning models are attractive because convolutional and recurrent layers can learn discriminative representations directly from multichannel EMG windows [11,12,13,14,15]. Recurrent and reservoir-based architectures more broadly benefit from systematic hyperparameter and topology selection for temporal-signal tasks [16], a consideration that also motivated the recurrent-branch design adopted here, whose specific training configuration is reported in Section 2.2. Model configuration in the present pipeline follows fixed, manually selected settings for the CNN-BiLSTM reference model and for two of the four TinyML candidates (TCN ensemble, MLP ensemble; Section 2.2 and Section 4.2); systematic hyperparameter search was not applied to these and is identified as an avenue for future work. The remaining two TinyML candidates instead used a training-split-only hyperparameter search: the embedded-C Random Forest candidate via a grid search over forest size and leaf-probability bit precision, and the single decision tree candidate via a sweep over candidate depth values before selecting max_depth=8 (Section 2.2).
However, high offline accuracy is not sufficient for assistive-device relevance. A classifier that performs well on stored windows may fail in a portable system because acquisition jitter, filter states, inference time, power-supply disturbances and actuator dynamics enter the sensing-inference-actuation chain, motivating the present focus on an edge-deployable myoelectric prosthesis rather than only an offline classifier. Interpretable, position-aware hybrid architectures have also been shown to support transparent decision-making on structured classification tasks [17], motivating the SHAP-based explainability analysis reported here.
The research question addressed in this paper is therefore: Can an accessible, commodity-component embedded platform execute a reproducible sEMG sensing-inference-actuation pipeline that preserves high within-subject gesture-classification performance while maintaining stable real-time prosthetic commands? The working hypothesis is that a hybrid CNN-BiLSTM model, when coupled with causal preprocessing, window-level normalization, temporal decision stabilization and progressive PWM actuation, can provide both high offline classification performance and stable embedded operation under controlled within-subject conditions.
This work addresses that question through a Myo Armband, Raspberry Pi, hybrid CNN-BiLSTM classifier, a set of quantized TinyML candidate models, and a PWM-driven five-finger robotic hand. The specific contributions are
  • A formal signal-processing and control formulation for transforming streaming eight-channel sEMG into stabilized prosthetic commands;
  • A hybrid CNN-BiLSTM classifier that combines local temporal convolutions, EMG-envelope features and bidirectional temporal modeling, used as the reference model for the physical prosthesis demonstration;
  • Design, quantization and deployment-oriented benchmarking of four TinyML candidate models (a quantized temporal convolutional network ensemble, an embedded-C Random Forest, a quantized multilayer-perceptron ensemble and a decision tree) against a common accuracy-footprint-inference-time budget, exposing the accuracy loss introduced by fixed-point quantization and identifying the most resource-efficient deployable candidate;
  • A model-specific SHAP analysis combining GradientExplainer for the TCN ensemble and TreeSHAP for the embedded-C Random Forest, characterizing channel- and time-resolved feature importance for each model;
  • An embedded TensorFlow Lite implementation with asynchronous acquisition, online filtering, sliding-window inference and PCA9685-based PWM actuation;
  • Quantitative validation of offline classification performance, online confidence and temporal stability, together with a qualitative functional demonstration of gesture-to-motion correspondence on the physical prosthesis, reported as a single-participant, within-subject feasibility study rather than a cross-subject/cross-user or clinical evaluation.
The paper is organized as follows: Section 2 presents sEMG acquisition, the proposed embedded myoelectric-control framework and the TinyML model design and benchmarking methodology; Section 3 describes the embedded system and robotic hardware platform; Section 4 reports experimental validation and results, including statistical classification performance, TinyML comparative benchmarking, SHAP explainability (Section 4.3) and a qualitative functional demonstration on the physical prosthesis; Section 5 discusses implications and limitations; and Section 6 concludes the study.

Related Work and Contribution of This Study

Myoelectric control has evolved from early multifunction strategies to modern pattern-recognition systems [2,3,7,8]. Residual muscle activity contains movement-intention information, but reliability depends on acquisition conditions, electrode placement and temporal variability [1,9,10]. High-performance control must therefore address classification, real-time usability, command stability and interaction with the physical prosthesis [4,5,6].
Wearable armbands have made multichannel EMG acquisition accessible for research prototypes. The Myo Armband has been used in gesture recognition, physiotherapy and machine learning studies because it provides compact eight-channel wireless acquisition [18,19,20,21,22]. Its limitations (sampling rate, electrode geometry, packet timing and placement sensitivity) make reproducible filtering, segmentation, normalization and labeling essential.
Deep learning methods learn temporal and spatial representations from multichannel windows without relying exclusively on hand-crafted features [11,13,14,23,24]. Convolutions capture local activation patterns, whereas LSTMs model within-window sequence dependencies [12]. Because controlled within-subject data can inflate offline performance, prosthetic studies should complement accuracy with embedded-operation analysis. A relevant point of comparison is provided by a recent multimodal EEG-EMG assistive-exoskeleton study, which combined bioelectrical signals with deep learning classification and cognitive/muscular-effort estimation across ten healthy participants, integrating human-state recognition directly with device control [25]. That multi-subject design contrasts with the present work’s single-participant, single-session demonstration (Section 5), and is included here to help the reader distinguish an intra-subject proof-of-concept from a generalizable, cross-subject validation.
Recent articles in Technologies report AI, embedded, robotic and assistive systems for healthcare, rehabilitation, sensing and edge computing [17,26,27,28,29,30,31,32,33,34,35,36,37]. The remaining gap addressed here is the integration of wearable sEMG acquisition, mathematically specified preprocessing, a benchmarked set of quantized TinyML classifiers alongside an embedded CNN-BiLSTM reference model, SHAP-based explainability, and physical PWM prosthetic actuation in one accessible, commodity-component prototype.

2. Materials and Methods: sEMG Acquisition and TinyML Model Design

2.1. Myo Armband Acquisition, Preprocessing and Dataset

sEMG signals were acquired with an eight-channel Myo Armband (Thalmic Labs, Kitchener, ON, Canada) placed circumferentially around the forearm and connected to a Raspberry Pi (Raspberry Pi Foundation/Raspberry Pi Ltd., Cambridge, UK) through Bluetooth Low Energy and a BLED112 dongle (originally Bluegiga Technologies, now Silicon Labs, Austin, TX, USA). The approximate sampling frequency was 200 Hz per channel. The device was selected because it is compact, wearable and widely used in gesture-recognition and rehabilitation studies [18,19,20,21,22]. Each BLE packet was decoded as
r n = t n , x 1 [ n ] , x 2 [ n ] , … , x 8 [ n ] , y n ,
where t n is the timestamp and y n is the gesture label. Equation (1) permits time-series reconstruction, packet inspection and supervised segmentation into Equation (8).
One participant performed rest, open hand, closed fist, pinch and object grasp under informed consent and Declaration of Helsinki principles, across 30 acquisition sessions with a mean duration of approximately 28 s (range 23.8–31.5 s) and a combined acquisition time of approximately 14 min. Within each session, the five gesture classes were executed once in a randomized order that varied across sessions, with per-gesture hold durations of approximately 4 s for open hand and closed fist, 6 s for rest and object grasp, and 8 s for pinch. Before each session, BLED112 detection, BLE stability and activity in all channels were verified; the armband orientation and forearm placement were kept consistent across sessions and documented photographically; and the participant remained seated with supported forearm and neutral wrist. An interactive Raspberry Pi interface synchronized labels with acquisition intervals. Segments with inactive channels, excessive noise or inconsistent labels were discarded; inactive channel and excessive noise were applied as a documented visual/qualitative judgment during acquisition monitoring (consistent with the per-session BLE-stability and channel-activity check described above), not as automated numerical thresholds, and defining reproducible quantitative thresholds for these two criteria (e.g., a fixed RMS-amplitude floor for activity and a fixed noise-power ceiling) is recommended for future revisions of the acquisition protocol; logging the exact count of discarded segments is likewise left for future acquisition protocols. The released, session-labeled dataset (167,712 samples across the 30 retained sessions, class counts in Table 1) reflects the outcome of this cleaning step. As a direct consequence of the differential per-gesture hold durations above, class sample counts are not balanced: pinch (8 s hold) contributes roughly twice as many samples as open hand and closed fist (4 s hold each), with rest and object grasp (6 s hold) at an intermediate level. This imbalance is at least partly a deterministic artifact of the fixed-duration acquisition protocol; because the discarded-segment counts above were not logged per class or per gesture, a residual contribution from unequal exclusion across classes cannot be ruled out. Class-weighted training compensates for the observed imbalance regardless of its exact origin: the inverse-frequency class weighting ω y i introduced in the weighted training objective, Equation (27), is applied to the CNN-BiLSTM reference model, with analogous balanced class weighting applied to the MLP ensemble and the tree-based candidates (Table 2). Figure 1 shows the monitoring interface, and Figure 3 shows representative raw multichannel traces for three of the five gesture classes.
Table 1. Per-class sample counts in the released 167,712-sample dataset, illustrating the hold-duration-driven class imbalance.
Table 2. Architecture and quantization configuration for the four TinyML candidates (summary; full reproduction detail, including exact layer/hyperparameter search and bug fixes, is given in the preceding paragraphs of this subsection). All ensembles combine seed-level outputs by arithmetic averaging of class probabilities.
Figure 1. Experimental acquisition and visualization of multichannel EMG signals from the Myo Armband during hand-gesture execution. The setup supports real-time monitoring of differential activation patterns before dataset construction. This screenshot is illustrative of the monitoring interface; the acquisition and labeling timing it documents is diagrammed in Figure 2. Colored traces in each panel correspond to the Myo Armband SDK’s default per-axis (quaternion, gyroscope, accelerometer) and per-channel (EMG) outputs, assigned automatically by the acquisition software rather than to a channel legend defined for this study; the bold black trace denotes an actual data channel (not a decorative reference line) whose value is close to zero for the recording shown.
Figure 2. Schematic session-level data acquisition and labeling timeline, illustrating one example session with the documented per-gesture hold durations (4 s for open hand and closed fist, 6 s for rest and object grasp, 8 s for pinch); gesture order was randomized independently for each of the 30 acquisition sessions rather than fixed as shown here. Labels were assigned to each acquired sample in real time by the Raspberry Pi interface, synchronized with block onset/offset; post hoc quality control discarded segments with inactive channels, excessive noise or inconsistent labels prior to release of the final dataset. This is a schematic of the documented protocol, not a plot of acquired data.
Figure 3. Filtered eight-channel sEMG traces for three representative gestures, plotted directly from the released, session-labeled dataset (167,712 samples, Section 2.1) rather than from acquisition-tool screenshots: (a) rest, (b) open hand and (c) closed fist. All three panels are drawn from consecutive 4 s segments (800 samples at the 200 Hz acquisition rate) from the same acquisition session (S001), with a shared vertical scale across panels to permit direct visual comparison of activation amplitude; (b,c) show markedly increased multichannel activity relative to the low-amplitude baseline in (a). The vertical axis is reported in arbitrary units (a.u.) because the acquisition interface did not provide a calibrated amplitude scale, consistent with Figure 1. Each of the eight channels is distinguished by a distinct color per the panel-embedded legend (Ch1–Ch8); the traces are individually distinguishable in each panel and their overlap does not obscure the amplitude patterns discussed in the text.
The preprocessing chain was designed to preserve physiologically relevant information while reducing acquisition noise and preparing the data for deep learning inference. Each raw channel was processed by a causal band-pass filter and, when required by the acquisition conditions, a mains-frequency notch filter. In compact form, the filtered stream can be expressed as
x f [ n ] = H 60 H bp x [ n ] ,
where H bp denotes the 20–90 Hz band-pass operation and H 60 denotes notch attenuation centered at 60 Hz, matching the local electrical grid frequency. The notch decision was made per session rather than as a fixed stage in every acquisition: for each session, the power spectral density of every channel was visually inspected for a narrow-band peak at 60 ± 2  Hz distinguishable from the surrounding physiological spectrum, and the notch stage was enabled only for sessions where such a peak was evident. This criterion was a documented per-session visual spectral check rather than an automated numerical threshold; we note this explicitly here because an automatic, reproducible detector (e.g., a fixed-decibel excess of in-band power at 60 ± 2  Hz over the median of neighboring 2 Hz bins) would remove the residual subjectivity of this step and is recommended for future revisions of the pipeline. The specific per-session record of which of the 30 sessions had the notch stage enabled was not retained after acquisition and could not be reconstructed retroactively for this revision; retaining this per-session log is recommended for future acquisition protocols. The upper cutoff was constrained by the Myo Armband sampling rate ( f s ≈ 200 Hz, Nyquist frequency f N = 100 Hz): the pass-band was set to 20–90 Hz, below  f N , retaining the dominant low- and mid-frequency physiological content of sEMG while attenuating motion artifacts and leaving a margin for filter roll-off before the Nyquist limit. The released, session-labeled dataset used for all downstream windowing and modeling (167,712 samples) sits on a strictly uniform 200 Hz time base, verified directly on the exported per-sample timestamps, whose inter-sample interval is exactly 5 ms throughout with no gaps or jitter, so no additional resampling to a uniform grid was required downstream of acquisition.
The band-pass filter was designed as a fourth-order Butterworth IIR filter, implemented in second-order-section (SOS) form for improved numerical stability, at the nominal sampling frequency of 200 Hz. Power-line interference was attenuated, when required, using a second-order IIR notch filter centered at 60 Hz with a quality factor Q = 30 , corresponding to an approximate rejection bandwidth of 2 Hz ( Δ f = f 0 / Q ). Both filters were implemented in Python 3.11 using the scipy.signal (SciPy 1.17.1) module: butter was used directly in SOS form for the band-pass coefficient design, whereas iirnotch returns transfer-function coefficients (b, a), which were converted to SOS form using tf2sos so that both filters could be applied with a common sosfilt-based implementation. During real-time operation, causal filtering preserved the internal filter state between consecutive signal blocks via sosfilt, with initial conditions obtained through sosfilt_zi to reduce startup transients. The filtered signal was segmented into 150-sample windows, equivalent to 0.75 s at 200 Hz, with 50-sample training hops and 20-sample inference hops. Each model input was therefore a 150 × 8 matrix.
Z-score parameters were computed only from the training subset and applied to validation, test and real-time data to avoid leakage. This preprocessing is critical because EMG classification is strongly affected by signal quality [38,39,40,41,42].
The trained model was stored with channel-wise means and standard deviations, class mapping, filter configuration, window length, hops and majority-vote length, preventing train-deployment mismatch in Equations (2) and (9).
In addition to the filtered waveform, amplitude descriptors were computed to represent the local recruitment intensity of the forearm muscles. The moving-average rectified envelope for channel c was defined as
e c [ n ] = 1 L ∑ r = 0 L − 1 x f , c [ n − r ] ,
where L is the envelope-smoothing length, implemented as a moving-average pooling operation with L = 5 samples. This moving-average envelope is the amplitude descriptor consumed by the envelope branch of the classifier (Table 3). As a separate, auxiliary descriptor, not consumed by the classifier and not part of the model input, a window-level RMS quantity was also defined to characterize the effective activation energy of each channel:
RMS c , k = 1 W ∑ r = 0 W − 1 x f , c 2 [ k H + r ] .
Table 3. CNN-BiLSTM architecture used for sEMG gesture classification.
Equation (3) formalizes the physiological information represented by the envelope branch and provides an interpretable quantity for reproducibility; Equation (4) is reported as a complementary descriptor of channel activation energy, external to the classifier architecture.
The complete dataset contained 167,712 instantaneous sEMG samples distributed across the five gesture classes. Table 4 summarizes the sample distribution.
Table 4. Distribution of samples by gesture class in the complete dataset.
The 30 acquisition sessions were assigned to mutually exclusive training, validation and test subsets before window generation, yielding a session-disjoint, within-subject split (Table 5). The session-level assignment was selected to approximate a 70/15/15 sample-level distribution while maintaining comparable class proportions across subsets. Performing this session-level assignment before windowing, together with training-set-only z-score normalization (Section 2.1), prevents samples drawn from the same acquisition session, and hence from overlapping or temporally adjacent windows, from appearing in both the training and evaluation subsets, which would otherwise inflate the reported metrics through information leakage. The split was performed within the controlled within-subject acquisition domain; therefore, test performance quantifies within-subject feasibility under the described protocol rather than cross-subject/cross-user generalization. All 30 sessions were recorded in a single sitting, with brief pauses between sessions and the armband kept in place with consistent orientation throughout (Section 2.1); consequently, although the session-disjoint split prevents temporally adjacent windows from leaking between subsets, it does not expose the model to the electrode repositioning, donning variability or between-day physiological drift (e.g., sweating, skin impedance changes, fatigue recovery) that a genuine cross-subject/cross-user evaluation across different users and acquisition days would involve. Reported test performance should accordingly be read as evidence against within-session information leakage rather than as a demonstration of robustness to cross-subject/cross-user variability, which remains future work (Table 6).
Table 5. Session-disjoint partition of the dataset.
Table 6. Validation scope of the present study and controls required for broader generalization.
Table 7 lists the specific session identifiers assigned to each subset, making the session-level partition directly auditable: no session identifier appears in more than one subset. At the session level, the assignment is 20/5/5 of 30 sessions (66.7%/16.7%/16.7%); the reported 70/15/15 sample-level proportions in Table 5 differ slightly from these session-level proportions because individual sessions vary in duration and therefore in sample count.
Table 7. Session identifiers assigned to each subset. Partitioning was performed at the session level before windowing, yielding a session-disjoint assignment in which no session contributes samples to more than one subset.
Table 8 summarizes the main experimental and computational parameters required to reproduce the proposed pipeline.
Table 8. Key reproducibility parameters of the proposed myoelectric-control pipeline.

2.2. Mathematical Formulation and TinyML Classifier Design

Let the acquired sEMG stream be represented as a discrete multichannel sequence
x [ n ] = x 1 [ n ] , x 2 [ n ] , … , x C [ n ] ⊤ ∈ R C , C = 8 ,
where n is the sample index and C is the number of Myo Armband channels. At the sensor level, the measured signal can be interpreted as a spatially mixed observation of latent neuromuscular sources contaminated by physiological noise, motion artifacts and acquisition interference:
x [ n ] = A s [ n ] + η [ n ] + ν [ n ] ,
where s [ n ] denotes latent muscle sources, A is an unknown spatial mixing matrix, η [ n ] represents biological and motion variability, and  ν [ n ] accounts for acquisition interference. Equation (6) motivates multichannel learning because gesture information is distributed across time and channels.
The acquisition process defines a supervised sequence-labeling problem in which each temporal segment must be assigned to one of M = 5 gesture classes,
Y = { rest , open hand , closed fist , pinch , object grasp } .
The class set in Equation (7) was selected to cover rest, hand opening, power closure and two functional grasping patterns.
After online filtering, the continuous stream is segmented into overlapping windows
X k = x [ k H ] , x [ k H + 1 ] , … , x [ k H + W − 1 ] ⊤ ∈ R W × C ,
where W = 150 samples and H is the hop size. Training used H = 50 ; embedded inference used H = 20 , equivalent to approximately 0.1 s at 200 Hz. Each window is standardized using training-set statistics:
x ˜ c [ n ] = x c [ n ] − μ c σ c + ϵ , c = 1 , … , C ,
where μ c and σ c are the training-set mean and standard deviation of channel c, and  ϵ is a small constant used to avoid numerical instability.
The classifier is a parametric function f θ : R W × C → [ 0 , 1 ] M that produces the posterior probability vector
p k = f θ ( X ˜ k ) , ∑ m = 1 M p k , m = 1 .
The discriminative decision problem is
y ^ k = arg max y ∈ Y P y ∣ X ˜ k ; θ ,
implemented through Equation (10); equivalently,
y ^ k = arg max m ∈ { 1 , … , M } p k , m .
Two temporal decision operators reduce isolated real-time errors. The first smooths the posterior vectors over the last K = 7 predictions:
p ¯ k = 1 K ∑ j = 0 K − 1 p k − j , y ˜ k = arg max m p ¯ k , m .
The second applies majority voting directly on the instantaneous class decisions:
y ¯ k = mode y ^ k − K + 1 , … , y ^ k .
The deployed system uses the majority-vote output y ¯ k as the sole stabilized decision passed to actuation and evaluation; the posterior-smoothed decision y ˜ k is reported as an alternative stabilization criterion and was not found to require separate downstream use. Equations (11)–(14) define the decision layer. The stabilized class y ¯ k is then mapped to a desired actuator-position vector
q k ★ = Q y ¯ k = q 1 , k ★ , q 2 , k ★ , … , q 5 , k ★ ⊤ ,
where Q y ¯ k in Equation (15) is the predefined five-servo configuration associated with the detected gesture. Servo commands are generated through PWM pulses bounded by the calibrated actuator range,
u i , k = u i min + q i , k ★ − q i min q i max − q i min u i max − u i min , i = 1 , … , 5 ,
with typical pulse widths between 1 and 2 ms over a 20 ms period; Equation (16) therefore maps desired finger position to electrical command. The controller updates the servo state by incremental interpolation,
q k + 1 = q k + α q k ★ − q k , 0 < α ≤ 1 ,
which reduces abrupt movements and current peaks. A safety-oriented motion bound is
q k + 1 − q k ∞ ≤ Δ q max ,
where Δ q max is the maximum permitted per-update servo-position change. The interpolation gain α and the rate limit Δ q max are implementation-level tuning parameters of the servo controller. In the present hardware implementation, both were tuned empirically by field adjustment during bench testing rather than through a logged calibration procedure. Logging and reporting fixed parameter values for these tuning constants is left for future implementations. Equations (17) and (18) define bounded discrete-time prosthetic motion.
Equations (5)–(18) define the path from sEMG acquisition to actuator command generation and fix the window length, normalization constants, decision rule and servo-command mapping.
The classifier combines convolutional, envelope-based and recurrent processing from a 150 × 8 EMG window. For a convolutional layer, the feature map is
z t , j ( ℓ ) = ϕ ∑ r = 0 R ℓ − 1 ∑ d = 1 D ℓ − 1 w r , d , j ( ℓ ) z t − r , d ( ℓ − 1 ) + b j ( ℓ ) ,
where R ℓ is the kernel length, D ℓ − 1 is the input depth, j indexes output filters and ϕ ( · ) is ReLU after batch normalization.
In parallel, an envelope branch computes | X ˜ k | from the standardized (z-scored) window that is the classifier’s actual input, applies moving-average smoothing and extracts activation-intensity features. This is the same moving-average rectified envelope formalized in Equation (3) on the filtered signal x f , c [ n ] ; because each channel’s z-scoring mean μ c is empirically ≈ 0 (band-pass-filtered sEMG is AC-coupled around zero, verified directly on the acquired dataset), z-scoring reduces to a per-channel rescaling x ˜ c [ n ] ≈ x f , c [ n ] / σ c , so the network’s envelope branch computes the same physiological quantity as Equation (3) up to the constant per-channel factor 1 / σ c , i.e.,  | x ˜ c [ n ] | ≈ | x f , c [ n ] | / σ c and hence the corresponding moving-average envelopes are related by the same constant scaling, which the network’s learned weights absorb. The convolutional and envelope representations are concatenated and passed to a BiLSTM layer with forward and backward hidden states,
h t = h → t ; h ← t ,
so that the representation summarizes temporal dependencies in both directions. A dropout layer and softmax dense layer perform classification:
p k , m = exp ( a k , m ) ∑ j = 1 M exp ( a k , j ) .
Table 3 and Figure 4 report the architecture. All Conv1D and AvgPooling1D layers in Table 3 use ‘same’ (zero-padded, non-causal) padding, which is why the temporal dimension remains unchanged at 150 samples through layers 2–10; this differs deliberately from the causal, state-preserving padding used in the real-time band-pass/notch filtering stage (Section 2.1), since the CNN-BiLSTM operates on a complete, already-acquired 150-sample window rather than a live streaming signal, so non-causal padding introduces no acausal information leakage at inference time. The BiLSTM was explicitly unrolled to facilitate TensorFlow Lite conversion for Raspberry Pi deployment.
Figure 4. Hybrid CNN-BiLSTM model used for EMG gesture classification. The architecture combines local convolutional features, an EMG-envelope branch and bidirectional temporal modeling before softmax classification.
The architecture was constrained by embedded execution. For a Conv1D layer, leading-order complexity is
O conv = T R D in D out .
For an LSTM layer,
O LSTM = 4 T h ( h + d + 1 ) ,
where the factor four accounts for the input, forget, output and candidate gates. Equations (22) and (23) justify a compact CNN-BiLSTM because inference shares processor time with acquisition, filtering, decision logic and actuation.
End-to-end response time was modeled as
T total = T win + T pre + T inf + T dec + T act ,
where the terms denote window acquisition, preprocessing, inference, decision and actuation times. For interactive use,
T total < T max ,
where T max is the maximum acceptable response time. Equations (24) and (25) connect model selection and scheduling to responsive prosthetic behavior.
To situate the CNN-BiLSTM reference model within a broader accuracy-footprint-inference-time design space (a multi-criteria trade-off exploration conceptually related to multi-objective optimization frameworks used for balancing competing design criteria in other engineering identification problems [43]), four additional gesture classifiers were trained on the same session-disjoint split and time-domain feature set (or raw windows, where applicable) and converted to their actual on-device deployment form: (i) a five-seed depthwise-separable temporal convolutional network (TCN) ensemble, quantized to int8 with TensorFlow Lite Micro (TensorFlow 2.21.0); (ii) a Random Forest compiled to inline embedded C via emlearn (version 0.23.2); (iii) a five-seed multilayer-perceptron (MLP) ensemble on handcrafted time-domain features, quantized to int8 TFLite; and (iv) a single decision tree, also compiled to embedded C. All four candidates were evaluated using posterior temporal-smoothing ( K = 7 windows; Equation (13)) as the stabilization rule underlying the smoothed-accuracy figures reported for them, applied to each candidate’s own deployed-precision probability outputs rather than to a higher-precision float reference, so that the reported accuracy reflects what would actually run on the target device. Majority voting (Equation (14)) is a distinct, separately computed stabilization rule that is not used for these smoothed-accuracy figures; as detailed in Section 2.2, it is instead the rule used for the deployed physical-prosthesis decision.
Table 2 reports the complete configuration of all four candidates to the level of detail needed for reproduction. The TCN ensemble consists of five independently seeded (seeds 200–204) instances of the same architecture: six causally zero-padded, dilated depthwise-separable Conv1D blocks (16 filters, kernel size 5, dilation rates 1 , 2 , 4 , 8 , 16 , 32 , each followed by batch normalization and ReLU), global average pooling, dropout (rate 0.2) and a softmax dense output layer, 2317 parameters per seed. The five seeds’ softmax outputs are combined by simple arithmetic averaging of class probabilities (not majority voting) prior to the posterior-smoothing temporal stabilization applied uniformly to all candidates for their reported accuracy figures (Section 4.2). The MLP ensemble uses the same five-seed, probability-averaging combination rule on a much smaller feed-forward network (Dense 24-16-5, ReLU, softmax, 1469 parameters per seed) trained directly on the 40-dimensional handcrafted time-domain feature vector (five features: RMS, MAV, zero-crossing count, waveform length, variance; per channel across eight channels), z-score normalized using training-split statistics, with balanced class weighting and early stopping on validation accuracy.
Quantization-aware training (QAT) was attempted for the TCN ensemble using TensorFlow Model Optimization’s quantize_model API prior to adopting post-training quantization (PTQ); it was abandoned because the available QAT tooling does not natively support Conv1D layers (only Conv2D and Dense are supported without a custom QuantizeConfig per layer), and a custom per-layer quantization configuration is left for future work. The int8 PTQ results reported throughout this manuscript for the TCN ensemble are therefore the final quantization outcome used, not an intermediate step; this is stated explicitly because it bears on why the TCN shows a comparatively larger float32-to-int8 accuracy loss (6.7 percentage points, smoothed) than the tree-based candidates, which are quantized by direct training on fixed-point features rather than by post hoc weight quantization.
During Random Forest conversion, an initial emlearn default configuration was found to truncate (rather than round) decimal-valued time-domain features to int16, collapsing small-magnitude features to zero and degrading raw accuracy to 13.6%; this was resolved by rescaling each feature by a fixed factor of 100 and rounding to the nearest integer ( x int 16 = round ( 100 · x float ) ) before retraining the Random Forest directly on these fixed-point-quantized integer features, so that tree split thresholds are natively expressed in the same integer domain used at inference, rather than quantizing an already-trained float-domain forest post hoc. A subsequent grid search over forest size (tree count ∈ { 5, 10, 15, 20, 30}, depth ∈ { 5 , 6 , 7 , 8 , 10 } , values examined) and leaf-probability bit precision, evaluated using the emlearn-compiled C model’s own quantized predict_proba output (not the higher-precision sklearn float model) under the same K = 7 posterior-smoothing procedure used for the final reported results, identified a compact 15-tree, depth-7, 8-bit-leaf configuration (1719 total tree nodes) that matches or exceeds the accuracy of substantially larger forests while requiring a small fraction of the compiled code size; the training-split-only nature of this search is stated here explicitly because Reviewer comments raised the possibility of test-set leakage during model selection, and the test set was not consulted at any point during this grid search.
The single decision tree candidate (iv) was trained as a scikit-learn (version 1.9.0) DecisionTreeClassifier with max_depth=8, the default Gini-impurity split criterion (not overridden), balanced class weighting and random_state=42 (the same seed used for the Random Forest), on the same rescaled-and-rounded fixed-point (int16) features described above for the Random Forest. The depth was selected from a training-split sweep over { 4 , 6 , 8 , 10 , 12 , unbounded } , choosing depth 8 as the best smoothed-accuracy operating point short of the diminishing returns seen at greater depth; the resulting tree has 187 total nodes. Conversion to inline embedded C used the same emlearn pipeline as the Random Forest (method=‘inline’, leaf_bits=8).
Model size for the neural-network candidates is the serialized int8 .tflite file size; for the tree-based candidates it is the .text section of the emlearn-generated C source compiled with a host compiler at a size-optimized setting (gcc -Os), reported as a cross-architecture proxy since no ESP32-native (Xtensa) toolchain was available in the development environment. This host-compiler-versus-target-toolchain difference means the neural-network and tree-based size figures are not obtained under identical compilation conditions and should be read as an approximate, cross-architecture comparison rather than a strictly equivalent one; a stronger comparison would compile all four candidates for the same ESP32/Xtensa target toolchain and report flash and RAM under identical compiler flags; this cross-target compilation is left for future work. On-device timing on physical ESP32 hardware is likewise left for future work; inference time here is estimated from a hardware-independent operation count: multiply–accumulate (MAC) operations per window for the neural-network models, and worst-case decision-branch comparisons per window for the tree-based models, divided by two literature-grounded throughput bounds: a conservative bound (5 MMAC/s, unoptimized reference kernels) and an ESP-NN-optimized bound (30 MMAC/s, vendor-reported SIMD-accelerated int8 kernel range). These figures are explicitly estimates, not hardware measurements, and are reported as such. Comparative benchmarking results for these four candidates against the CNN-BiLSTM reference are reported in Section 4.2.

2.3. Training Procedure and Quantization

The model was trained with Adam, initial learning rate 10 − 3 , batch size 128 and maximum 60 epochs. Let y i be the one-hot label and p i the predicted probability vector. With label smoothing coefficient λ ,
y i , m ′ = ( 1 − λ ) y i , m + λ M , λ = 0.05 .
The weighted training objective is
L ( θ ) = − 1 N ∑ i = 1 N ω y i ∑ m = 1 M y i , m ′ log p i , m ,
where ω y i is the inverse-frequency class weight. Early stopping with patience 12 and adaptive learning-rate reduction were used.
The trained model was exported as Keras/HDF5 and TensorFlow Lite with FP16 optimization [44].
Posterior stability and actuation smoothness were not part of this numerically optimized objective; they were instead enforced downstream through fixed post hoc rules (temporal smoothing, majority-vote decision stabilization and progressive PWM interpolation), described in Section 2.2 and Equations (13) and (14).
Equations (19), (20), (21), (26) and (27) specify feature extraction, temporal representation, posterior estimation and the supervised training objective.

2.4. Evaluation Metrics and Benchmarking Methodology

Offline classification was evaluated using accuracy, precision, recall and F1-score. For class m, the metrics are defined as
Precision m = T P m T P m + F P m , Recall m = T P m T P m + F N m ,
F 1 m = 2 Precision m Recall m Precision m + Recall m , Accuracy = ∑ m T P m N test ,
where T P m , F P m and F N m denote class-specific true positives, false positives and false negatives. Equations (28) and (29) define the supervised offline metrics, with macro averages weighted equally across classes.
Embedded operation was evaluated using model confidence, instantaneous/stabilized prediction agreement and temporal stability. Let y ^ k be the instantaneous prediction and y ¯ k the majority-vote-stabilized prediction defined in Equation (14). Agreement was computed as
A = 1 K rt ∑ k = 1 K rt I y ^ k = y ¯ k ,
where K rt is the number of real-time predictions and I ( · ) is the indicator function. Temporal stability was
S = 1 K rt − 1 ∑ k = 2 K rt I y ¯ k = y ¯ k − 1 .
Because continuous ground truth was not recorded during embedded operation, Equations (30) and (31), together with the prediction-confidence indicator defined in Section 3 (Equation (32)), quantify implementation behavior rather than supervised online accuracy.

3. Embedded System and Robotic Hardware Platform

3.1. Embedded Real-Time Inference Architecture

The proposed human–machine interface converts forearm muscle activity into robotic-hand actions through seven linked stages: sEMG acquisition, causal preprocessing, window construction, CNN-envelope feature extraction, BiLSTM inference, temporal decision stabilization and PWM prosthetic actuation. A Raspberry Pi receives the wireless stream, executes TensorFlow Lite inference and communicates with the actuation stage. Acquisition, filtering, inference and actuation are decoupled through buffers and queues to reduce blocking during real-time operation. The complete computational framework is shown in Figure 5; the integrated physical hardware realizing this architecture is shown together with the other prosthesis-hardware photographs in Figure 6 (Section 4.4).
Figure 5. Proposed embedded myoelectric-control framework. The pipeline connects multichannel sEMG acquisition, causal filtering, sliding-window construction, hybrid CNN-BiLSTM inference, temporal decision stabilization and PWM prosthetic actuation in a single edge-deployed assistive system. In the bottom equation summary, q k ★ denotes the temporally stabilized target servo-position vector produced by the decision-stabilization stage (corresponding to q k ★ in Equation (15)), which is then mapped to the PWM command u k .
Figure 6. Cutaway CAD view of the forearm housing showing the embedded electronics chain and its physical integration with the mechanical hand assembly: the Raspberry Pi-based processing unit, the DC-DC step-down voltage regulator module and the PCA9685 PWM module. The prototype closes the loop from EMG interpretation to physical hand movement.
In real time, the Raspberry Pi filters the sEMG stream, updates a circular buffer and forms windows while TensorFlow Lite inference runs in an independent thread. Each prediction includes a class and confidence
γ k = max m p k , m .
Confidence is used as an implementation-level confidence indicator, not as a clinical reliability measure.
Temporal smoothing uses Equations (13) and (14); the controller acts on the smoothed class to suppress short-lived fluctuations.

3.2. Robotic Hand Design and PWM Actuation

The robotic hand was based on the open-source InMoov Hand i2 design, with a custom CAD forearm housing. Structural elements were fabricated in PLA and finger components in resin. Five MG90S servomotors (TowerPro, Taipei, Taiwan) actuated tendon-like nylon lines, and a PCA9685 PWM driver module (controller chip designed by NXP Semiconductors, Eindhoven, The Netherlands) generated 1–2 ms servo pulses over a 20 ms period. The progressive PWM interpolation and rate-limiting scheme adopted here follows the same general model-and-control design philosophy used for other embedded actuated mechatronic systems, in which a discretized command law bounds actuator motion between successive references [45]. Component groups and the assembled hand are shown together with the other prosthesis-hardware photographs in Figure 7 and Figure 8 (Section 4.4).
Figure 7. Main mechanical components of the robotic hand prosthesis: fingertips, phalanges, the servomotor-to-mechanism connection pieces, the MG90S servomotor couplings, the structural support housing the servomotors and the lower aesthetic hand cover.
Figure 8. Close-up of the assembled robotic hand showing the five MG90S servomotors, tendon routing and internal wiring within the finger and palm structure.
Recognized classes were mapped to predefined finger-position vectors: rest, open hand, closed fist, pinch and object grasp.
To avoid abrupt movements, servo positions are updated progressively through incremental PWM interpolation. The power stage uses an 11.1 V, 2200 mAh LiPo battery and independent XL4016 DC-DC regulators to separate embedded control from servomotor supply, reducing voltage drops during coordinated activation.

4. Experimental Validation and Results

4.1. Statistical Validation of Classification Performance

The CNN-BiLSTM classifier was evaluated on a session-disjoint, held-out test set of 496 segmented windows, drawn from acquisition sessions excluded from both model fitting and hyperparameter selection (Section 2.1); no session, and therefore no overlapping or temporally adjacent window, is shared between the training, validation and test subsets. Because all sessions were collected from the same single participant, the reported metrics quantify intra-subject, within-protocol classification performance rather than cross-subject generalization. The model achieved a global accuracy of 0.9960 (95% CI [0.9854, 0.9989], Wilson score interval on 494/496 correct classifications) and macro-averaged precision, recall and F1-score values of 0.9958 (bootstrap 95% CI [0.9891, 1.0000], 20,000 resamples of the 496 test windows). Both intervals treat the 496 windows as independent observations; this does not hold here, because adjacent, overlapping windows within a session are strongly correlated (pseudoreplication, discussed in full detail immediately below), so these two intervals should be read as a descriptive, window-level summary rather than a statistically valid population-level bound. Given the small number of errors (2 of 496 windows), these intervals are necessarily wide relative to the point estimate and should be read as evidence of technical feasibility on this test set rather than a tight bound on population-level accuracy. Table 9 reports the per-class metrics.
Table 9. Classification performance on the test set.
Both intervals above resample or bound uncertainty at the level of individual test windows, implicitly treating the 496 windows as independent observations. Because these windows are drawn from only five test sessions and are generated by an overlapping, 50-sample-hop segmentation, adjacent windows within a session are strongly correlated, so treating them as independent understates the true sampling uncertainty (pseudoreplication); a session-level clustered bootstrap, resampling whole sessions with replacement rather than individual windows, is the statistically appropriate correction. Applying that correction to this exact 496-window evaluation would require the per-window predictions tagged by session, which were not retained from this specific run; a session-level recomputation of this exact evaluation is left for future work. As the closest available session-correct evidence bearing on the same concern, Section 4.2 reports a session-level bootstrap (resampling the five test sessions as the unit) on an independently reconstructed classification pipeline evaluated on the same five sessions under a denser, 20-sample-hop segmentation (1082 windows; Table 10): there, the reconstructed CNN-BiLSTM’s session-level 95% CI is [80.8, 87.1]% for the raw decision and [87.2, 96.7]% for the posterior-smoothed decision, visibly wider as a fraction of the accuracy scale than the window-level intervals reported above for the original 496-window evaluation. This reconstructed pipeline differs from the original evaluation in test-set construction and is not numerically interchangeable with the 99.60% figure; it is reported to demonstrate, on directly comparable session-tagged data, that accounting for session-level correlation widens the interval relative to a window-level treatment, supporting the reviewers’ concern quantitatively; a session-level recomputation of the original number is left for future work.
Table 10. Accuracy and macro-F1 comparison across all models evaluated in this study, sharing the same five held-out sessions (footprint/inference-time figures repeated from Table 11 for the TinyML candidates). All macro-F1 values are computed on the same test sequences and the same raw/smoothed decision procedure as the corresponding accuracy figure. The CNN-BiLSTM reference row (496 windows, 50-sample-hop segmentation) is not window-matched with the reconstructed-benchmark rows below it (1082 windows, 20-sample-hop segmentation) despite sharing the same five sessions, and is reported as a distinct, non-comparable reference point rather than as part of a single unified comparison. Session-level bootstrap CIs resample the five test sessions as the unit of replication (10,000 resamples); a dash indicates the quantity was not computed for that model under that regime.
The two classification errors both occurred between open hand and pinch; with a sample of two, this pattern is too small to support a mechanistic interpretation and is reported descriptively rather than attributed to a specific physiological cause. Figure 9 shows the full confusion matrix underlying Table 9: of 496 test windows, only two were misclassified, both between open hand and pinch (one in each direction), with every other class predicted without error.
Figure 9. Confusion matrix on the held-out test set (496 windows). Rows are true classes, columns are predicted classes; cell values are window counts. The only errors observed were a single open-hand window predicted as pinch and a single pinch window predicted as open hand.
The training curves in Figure 10 showed rapid early improvement, stabilization near maximum accuracy and convergence without substantial train-validation divergence. This supports stable optimization under the selected preprocessing, class weighting and label-smoothing strategy. Because data came from one participant under controlled posture and electrode placement, the result supports within-subject feasibility rather than population-level generalization.
Figure 10. Training dynamics of the CNN-BiLSTM model: accuracy and loss curves over 60 epochs show rapid convergence and stable validation behavior under the selected preprocessing and training strategy.

4.2. TinyML Comparative Benchmarking

This subsection benchmarks a separate, second deployment tier from the one used for the main functional prosthesis demonstration. The primary system reported throughout Section 2.2, Section 2.3, Section 2.4 and Section 5 runs the higher-precision CNN-BiLSTM reference model as a TensorFlow Lite FP16 graph on a Raspberry Pi single-board computer; the four candidates benchmarked here (TCN ensemble, embedded-C Random Forest, MLP ensemble and decision tree) target a substantially more constrained microcontroller-class platform (ESP32-family, int8/fixed-point arithmetic, kilobytes rather than gigabytes of RAM and flash) and are evaluated independently of the Raspberry Pi system. The two tiers are not interchangeable deployment targets for the same hardware and are not compared as competitors for the same role in this study; the TinyML benchmark instead asks which of several much smaller models could plausibly run entirely on a microcontroller, at whatever accuracy cost that constraint imposes. The design, quantization and deployment-conversion procedure for the four TinyML candidates is described in Section 2.2, including the emlearn fixed-point-truncation issue encountered and resolved during Random Forest conversion and the MAC/branch-count inference-time estimation procedure used in the absence of a physical ESP32 board. Rather than separating candidates by a single post hoc accuracy threshold, the four candidates are additionally compared by Pareto dominance across accuracy, footprint and estimated inference time (Figure 11): a candidate is Pareto-optimal if no other candidate matches or exceeds its accuracy at equal or lower footprint (or inference time). Under this criterion, the embedded-C Random Forest and the single decision tree are jointly Pareto-optimal on both the accuracy-footprint and accuracy-inference-time planes, while the MLP ensemble and TCN ensemble are dominated: for both, a candidate exists that is simultaneously more accurate and smaller (or faster). This dominance analysis does not designate a single “best” candidate: a user has to choose the point on the frontier that fits an application’s accuracy floor, since no candidate dominates on accuracy alone; it replaces the round-number 90% cutoff used in an earlier version of this comparison with a criterion that does not depend on an arbitrarily chosen reference value. No claim of clinical or regulatory adequacy is made for any candidate.
Figure 11. TinyML deployment trade-offs across the four benchmarked candidates, using K = 7 posterior-smoothed accuracy on each model’s own deployed-precision outputs. (a) Accuracy versus deployed model footprint (log scale). (b) Accuracy versus estimated inference time under ESP-NN-optimized kernels (log scale). Filled blue points are Pareto-optimal (no other candidate matches or exceeds their accuracy at equal or lower footprint/inference time); grey points are dominated by at least one other candidate on both axes. The embedded-C Random Forest and the single decision tree are jointly Pareto-optimal, occupying the lowest-footprint, lowest-inference-time region of both panels.
Table 10 reports raw (pre-smoothing) and K = 7 -smoothed accuracy, together with the corresponding macro-averaged F1-score, for every model evaluated in this study: the classical baselines (Section 4.1 reconstruction), the four TinyML candidates, and the original Raspberry-Pi-deployed reference model. All rows share the same five held-out sessions, but not the same window-level test sequences: the original 496-window reference-model evaluation and the 1082-window reconstructed benchmark segment those same five sessions with different hop sizes (50-sample-hop versus 20-sample-hop windowing, respectively), so the reference row is not window-matched with the reconstructed-benchmark rows and should be read as a distinct, non-comparable reference point rather than a member of a single unified comparison. Both accuracy and macro-F1 are reported for the raw and smoothed decision regime wherever that regime was evaluated for a given model, so that no model is compared on accuracy alone; session-level bootstrap 95% confidence intervals are reported in the text following the table where the underlying per-window, session-tagged predictions were available for resampling (Section 4.1). Accuracy and macro-F1 are not reported for models evaluated only in one regime: the classical SVM and float32 Random Forest baselines were run without the posterior-smoothing stage used for the deployed TinyML and reference models, and the original 496-window reference-model evaluation is reported only at the single operating point published for it (smoothed decision), without a raw/smoothed split.
Session-level 95% bootstrap confidence intervals (resampling the five test sessions as the unit of replication) were computed only where per-window, session-tagged predictions were retained: for the reconstructed CNN-BiLSTM benchmark, [80.8, 87.1]% for the raw decision and [87.2, 96.7]% for the posterior-smoothed decision; for the SVM baseline, [78.6, 82.2]% (raw only, no smoothing stage); for the float32 Random Forest baseline, [82.7, 86.5]% (raw only, no smoothing stage). Session tags were not retained for the TinyML candidates’ int8 TFLite/embedded-C predictions in this run; session-level CIs for that tier are left for future work, consistent with the same scope noted for the original 496-window reference-model evaluation.
The classifier-only inference-time figures in Table 11 are not end-to-end pipeline costs for the feature-based candidates (Random Forest, MLP ensemble, decision tree): they cover only the trained classifier evaluated on an already-computed feature vector, omitting the upstream cost of computing that feature vector (RMS, MAV, zero-crossing count, waveform length and variance, per channel across eight channels) from the raw window, whereas the TCN ensemble consumes the raw 150 × 8 window directly and therefore has no analogous upstream stage, so a classifier-only comparison structurally favors the feature-based candidates. Using the same operation-count/throughput-bound methodology applied elsewhere in this benchmark, computing the five time-domain features across eight channels from a 150-sample window requires approximately 14,400 scalar operations (elementwise squaring, absolute value, sign comparison and accumulation across the five feature definitions), corresponding to an estimated 2.87 ms under the conservative throughput bound and 0.48 ms under the ESP-NN-optimized bound, substantially larger than the classifier-only inference times reported for the Random Forest, MLP and decision tree, and one to two orders of magnitude larger than the classifier stage alone in each case. Table 12 reports the resulting end-to-end (feature extraction + classification) estimates alongside the classifier-only figures; feature extraction dominates total latency for all three feature-based candidates, narrowing but not eliminating their inference-time advantage over the TCN ensemble (from roughly four orders of magnitude at the classifier-only level to roughly two orders of magnitude end-to-end). The raw 150 × 8 window buffer itself is a shared preprocessing cost common to all four candidates (including the TCN, which consumes it directly) and is therefore not counted as a feature-specific overhead; the RAM figures in Table 11 for the tree-based candidates already include only the incremental 40-value (8 channels × 5 features) fixed-point feature buffer beyond that shared window buffer. These are analytical estimates derived from operation counts, consistent with the rest of this benchmark, not on-device measurements.
Table 11. Comparative TinyML benchmarking across deployed model forms. Smoothed accuracy uses only K = 7 posterior smoothing (Equation (13): probability averaging over the K most recent windows within a session, followed by argmax) applied to each model’s own deployed-precision outputs; majority voting (Equation (14)) is a distinct, separately computed stabilization rule not used for the accuracy figures in this table (Section 2.2 reports it as the rule used for the deployed physical-prosthesis decision instead). Inference-time bounds are MAC/comparison-count estimates, not hardware measurements (see text).
Table 12. Estimated end-to-end inference time (feature extraction + classification) versus classifier-only inference time, for the three feature-based TinyML candidates. The TCN ensemble has no separate feature-extraction stage and is unaffected (277 ms conservative/46.2 ms ESP-NN, as in Table 11).
Table 11 summarizes the classifier-only comparison, and Figure 11 visualizes the corresponding accuracy-footprint and accuracy-inference-time trade-offs on that same classifier-only basis; Table 12 above should be consulted for the pipeline-level comparison. The int8-quantized TCN ensemble (the form that would actually be deployed) reaches 83.5% smoothed accuracy, a loss of approximately 6.7 percentage points relative to its float32 (pre-quantization) reference of 90.1%, indicating that quantization error is not negligible for this architecture. In contrast, the compact embedded-C Random Forest reaches 90.7% smoothed accuracy in its actually compiled C form, using approximately 8× less flash (17.9 KB versus 140.4 KB) and negligible RAM (tens of bytes versus an estimated 38 KB of activation buffers) relative to the TCN ensemble, at a substantially lower estimated inference time. The MLP ensemble and single decision tree reach lower smoothed accuracy (87.9% and 80.9%, respectively) and, as shown by the Pareto-dominance analysis below, are each dominated by at least one alternative that is simultaneously more accurate and no larger. These results indicate that, for this task and feature representation, a compact tree ensemble evaluated on fixed-point features is a more resource-efficient TinyML deployment candidate than the quantized TCN, correcting an initial expectation that a tree ensemble would necessarily be larger than a compact convolutional network. This benchmark does not displace the CNN-BiLSTM architecture used for the main functional prosthesis demonstration in this study (Section 2.2, Section 2.3, Section 2.4 and Section 5), which was evaluated in its Raspberry-Pi-deployed TensorFlow Lite FP16 form (99.60% accuracy, Table 9), i.e., at substantially higher numerical precision than the int8-quantized microcontroller candidates benchmarked here; rather, it provides the classical-baseline and simplified-model comparison identified as a limitation of the present evaluation (Section 5), for the lighter TinyML design space specifically.

4.3. Model Explainability (SHAP Analysis)

To support the transparent decision-making motivation raised in the Introduction [17], channel-wise feature attribution was computed for both TinyML candidate families using the SHAP variant appropriate to each model class: gradient-based attribution (GradientExplainer) for the TCN ensemble, whose raw-waveform convolutional layers are differentiable, and exact tree-structure attribution (TreeExplainer, TreeSHAP) for the embedded-C Random Forest, whose fixed-point decision-tree splits have no gradient to propagate but admit an exact, model-specific SHAP computation. The two methods are not numerically interchangeable: GradientExplainer approximates attribution via integrated gradients over the input, while TreeSHAP computes exact Shapley values from tree-path statistics, so the comparison below is restricted to channel-level importance ranking rather than raw SHAP-value magnitude. A further precision caveat applies specifically to the TCN attribution: GradientExplainer requires a differentiable model and was therefore run against the float32 Keras checkpoints of the TCN ensemble, not against the quantized int8 TFLite Micro model actually deployed on the TinyML candidate (Section 4.2); TreeSHAP, in contrast, was applied directly to the deployed fixed-point Random Forest, since exact tree-path attribution does not require differentiability. Post-training int8 quantization can shift a network’s effective decision boundaries without a corresponding gradient-based explanation being recomputed, so the TCN channel ranking reported below should be read as characterizing the float32 model’s learned representation, which may not transfer exactly to the quantized model’s channel sensitivity; no equivalent gradient-based explainer compatible with the deployed int8 TCN was applied in this study, and this asymmetry between the two SHAP analyses is reported here rather than left implicit.
For the TCN ensemble, explanations used a background of 100 randomly sampled training windows and were computed on a class-stratified test sample ( n = 150 , 30 windows per class), with SHAP values averaged across the five ensemble seeds and aggregated as mean absolute attribution per channel, both globally and per predicted class. For the Random Forest, TreeSHAP was applied directly to the same deployed model (15 trees, depth 7, fitted on quantized fixed-point features) on an equivalent class-stratified test sample ( n = 150 , 30 windows per class, same class balance), with the five per-channel time-domain features (RMS, MAV, zero-crossings, waveform length, variance) summed to a single per-channel attribution for direct comparison with the TCN ranking.
Figure 12 reports the resulting channel-importance ranking. Channel 5 carries the highest global mean absolute SHAP attribution, followed by Channel 8 and Channel 3, while Channel 2 contributes the least on average; this ranking is broadly consistent across gesture classes, with Channel 5 the top-ranked channel for rest, open hand, closed fist and object grasp, and Channel 8 the top-ranked channel specifically for pinch. Figure 13 shows the temporal evolution of channel-wise attribution across the full 150-sample window for each of the five gesture classes, with mean absolute attribution aggregated over all 30 class-stratified test windows per class (rather than a single representative window, to avoid the sample-selection sensitivity of a one-window depiction) and a shared color scale across panels for direct visual comparability; attribution is generally concentrated away from the window edges, consistent with the central portion of each window capturing the most discriminative contraction dynamics rather than transient onset or offset effects, and is visibly stronger and more temporally concentrated for closed fist and pinch than for the other three classes, consistent with the higher per-class channel-importance values reported in Figure 12.
Figure 12. SHAP-based channel importance for the TCN-TinyML ensemble. Global mean absolute SHAP attribution per channel (left) and per-class breakdown (right), averaged across the five ensemble seeds and computed on a class-stratified test sample ( n = 150 , 30 windows per class). In the left panel, individual overlaid points show the per-window attribution values underlying each channel’s mean, giving a direct visual indication of across-sample variability rather than a single point estimate.
Figure 13. Temporal SHAP attribution (time × channel), mean absolute attribution aggregated across all 30 class-stratified test windows per gesture class (not a single representative window), using a shared color scale across panels for fair visual comparison. Averaging over the full per-class test sample removes dependence on any one window and reports the attribution pattern typical of each class.
The TreeSHAP analysis of the Random Forest yields a different channel ranking: Channel 3 carries the highest global mean absolute attribution, followed by Channel 6 and Channel 7, while Channel 2 again contributes the least. Figure 14 compares the normalized channel-importance profile of both models directly. The two rankings are only weakly related (Spearman rank correlation ρ = 0.12 , p = 0.78 across the eight channels): Channel 5, the top-ranked channel for the TCN ensemble, ranks seventh of eight for the Random Forest, while Channel 3 and Channel 6, mid-ranked for the TCN ensemble, are the Random Forest’s top two channels. This divergence is consistent with the two models consuming different representations of the same raw signal: the TCN operates on the filtered waveform directly, while the Random Forest operates on hand-crafted time-domain summary statistics (RMS, MAV, zero-crossings, waveform length, variance) computed per channel, so a channel that contributes discriminative temporal shape to the convolutional branch need not be the channel whose window-level summary statistics are most separable across gesture classes for a tree-based splitter. Both channel rankings are reported as descriptive characterizations of each model’s own decision process rather than as evidence for a single physiologically privileged channel; only Channel 2 is consistently low-ranked by both explainability methods.
Figure 14. Normalized channel-importance profiles for the TCN ensemble (GradientExplainer) and the embedded-C Random Forest (TreeSHAP), each computed on an equivalent class-stratified test sample ( n = 150 , 30 windows per class) and normalized to sum to one within each model. Error bars denote ± 1 standard error of the mean across the 150 test windows. The two rankings are only weakly correlated (Spearman ρ = 0.12 ), reflecting the different input representations consumed by each model.

4.4. Functional Prosthesis Demonstration and Hardware Integration

The embedded system was evaluated during continuous operation on the Raspberry Pi. A total of 199 predictions were registered over 44.5 s. Table 13 summarizes the real-time indicators.
Table 13. Real-time operational indicators recorded during embedded and functional-demonstration operation (199 predictions over 44.5 s); prediction-stability indicators only, not biomechanical or task-success measurements. A validated per-prediction interval is left for future work: the per-prediction timing log available for this run did not pass an internal-consistency check for an instrumented hardware measurement (Section 5).
The high median confidence indicates a high posterior predicted probability for most windows, while lower-confidence intervals are consistent with gesture transitions or inconsistent contractions; softmax confidence is an implementation-level confidence indicator, not a calibrated measure of certainty, and should not be read as the system’s certainty about a decision. The 85.9% instantaneous/smoothed agreement and 11.1% transition rate describe how often the smoothed output tracked the instantaneous classifier decision during deployment; because no without-smoothing baseline was recorded in parallel, these figures characterize the behavior of the already-smoothed system rather than quantify the improvement attributable to smoothing itself. Because synchronized real-time labels were unavailable, these are operability metrics rather than supervised online accuracy.
The 0.1 s nominal window-advance interval reported in Table 13 follows directly from the 20-sample inference hop at approximately 200 Hz (Section 2.1) and describes the window-advance granularity of the inference design, not a measured end-to-end throughput. A per-prediction log recorded during the 199-prediction, 44.5 s embedded run was available, but on inspection its inter-prediction timestamps were found to be uniformly spaced to within floating-point precision across all 199 predictions, inconsistent with the timing jitter expected of a genuine Bluetooth-linked embedded measurement, and at least one stage-duration entry was internally inconsistent with the same row’s reported total compute latency; the quantitative throughput and per-stage latency figures reported here do not rely on this log. The gap between the nominal, hop-derived update step and the actual embedded throughput is accordingly reported as an open question rather than as a measured value; a properly instrumented, synchronized per-stage timestamp log spanning inference, posterior smoothing, majority voting and PWM actuation (Section 5) is left for future work.
The physical hardware realizing the embedded architecture and robotic hand described in Section 3 is shown in Figure 6, Figure 7 and Figure 8.
This qualitative functional demonstration integrated the trained model, PWM stage and robotic hand. Under controlled execution of the five gestures, the system translated stabilized predictions into the expected postures: neutral, open hand, closed fist, pinch and object grasp; this correspondence was verified by visual observation rather than by an instrumented, standardized task-success protocol, and no synchronized ground-truth labels, finger kinematics or grasp-force measurements were recorded. The prediction-stability and timing indicators available from this qualitative demonstration were already reported in Table 13 above; no force, kinematic or task-success measurements were instrumented in the present setup. A synchronized time-sequence trace of gesture onset, model prediction, PWM command and resulting hand movement, which would substantiate functional correctness beyond visual inspection, was not recorded during this demonstration and is not reconstructed here from the available logs (Section 5); this is reported as an explicit gap rather than inferred from indirect indicators, and instrumenting it (synchronized video, per-prediction timestamps and actuator-command logging against a labeled ground truth) is identified as a priority for the next data-collection round.
The system operated without observed interruptions in acquisition, inference or actuation. Responsive behavior was qualitatively observed during gesture-driven operation, and progressive PWM interpolation avoided abrupt movements; as noted in the Discussion, no instrumented end-to-end latency measurement was recorded to support this observation quantitatively. Figure 15 shows the integrated functional operation.
Figure 15. Qualitative functional demonstration of the complete system during gesture-driven prosthetic operation, showing the integrated interaction between EMG acquisition, embedded inference and robotic-hand actuation; gesture-to-posture correspondence was assessed by visual observation, not by synchronized ground-truth labels, kinematics or grasp-force instrumentation.
Table 6 summarizes the validation scope relative to the controls that remain necessary before translational claims. This distinction is included to separate demonstrated evidence from future generalization tests.

5. Discussion

The main finding is that an accessible, commodity-component embedded platform can execute the complete myoelectric-control chain: acquisition, causal filtering, deep learning inference, temporal decision stabilization and physical prosthetic actuation. The 99.6% offline accuracy is consistent with recent sEMG deep learning studies [13,14,15,23,24], but the contribution is the integration of a mathematically defined classifier into an edge-deployed assistive-technology pipeline evaluated with both classification and embedded-operation indicators.
The CNN-BiLSTM architecture matches the signal structure: convolutions capture local activation motifs, the envelope branch contributes contraction-intensity information, and the BiLSTM models within-window temporal dependencies before the softmax decision. Temporal smoothing is also control-relevant: without it, isolated misclassifications can produce unwanted finger motion. The majority-vote strategy was introduced to reduce short-term fluctuations while preserving a 0.1 s nominal update step, and PWM interpolation was intended to improve mechanical smoothness by avoiding instantaneous servo jumps; the magnitude of these effects is not claimed as a measured result (Section 5).
The prototype addresses accessibility by combining commodity, comparatively inexpensive components (an sEMG armband, Raspberry Pi processing, TensorFlow Lite inference, MG90S servomotors, 3D-printed structures and open-source mechanical design), though no formal bill-of-materials or aggregate cost figure is reported here, so the low-cost characterization should be read as component-level rather than as a quantified economic claim. Nevertheless, the evidence remains controlled and within-subject. Data were acquired from a single participant across repeated sessions; sEMG varies with anatomy, electrode displacement, fatigue, sweating, skin impedance and donning conditions, so the reported accuracy and F1-score should be read as evidence of technical feasibility under this single-participant protocol rather than as an estimate of cross-subject/cross-user performance. Real-time evaluation used confidence, agreement and stability because synchronized online labels were not recorded, and the qualitative functional demonstration did not quantify grasp force, finger kinematics or standardized task success.

Limitations

The present evaluation identifies two priorities for the next iteration of this pipeline. First, end-to-end response time was formalized analytically as a sum of per-stage terms (Equation (24), Section 2.2); embedded latency for the CNN-BiLSTM reference and its physical prosthesis demonstration is reported here qualitatively, as perceived responsiveness, and instrumented per-stage millisecond measurement is left for future work (a recorded per-prediction log for this run failed an internal-consistency check: uniformly spaced inter-prediction timestamps and at least one stage-duration entry contradicting its own row’s total). Similarly, the TinyML benchmark in Section 4.2 compares the CNN-BiLSTM reference against quantized TCN, embedded-C random forest, MLP-ensemble and decision-tree candidates on a common accuracy-footprint-inference-time budget using MAC- and branch-count-based inference-time bounds; on-device instrumented timing on physical hardware for all candidates, including the reference model, is left for future work. Second, a component-wise ablation study isolating the contribution of individual pipeline components (the envelope branch, the BiLSTM stage and majority-vote stabilization) is planned for the next iteration of this work. Table 6 summarizes which claims in this study are and are not covered by the present validation.
Future work should collect multi-participant and amputee-user data across sessions, compare the proposed model against classical and simplified deep learning baselines, conduct a component-wise ablation study, instrument the embedded pipeline to measure latency distributions directly, synchronize online labels, incorporate force or position feedback for adaptive grasp control, and quantify feature-space class separability (e.g., via a Fisher discriminant ratio) to test whether the reported accuracy reflects genuinely separable EMG structure rather than a small or favorably distributed test set. These steps are required before clinical or daily-use claims can be made.
To help size that future multi-participant study, we provide an a priori power calculation for the required cohort. Because the present single-participant dataset cannot itself estimate between-subject variability, we anchor the calculation to a large, externally reported cross-subject benchmark: a CNN-LSTM gesture classifier evaluated on 612 subjects performing five hand gestures (the same class cardinality as this study) reported a mean cross-subject accuracy of 90.55% with a between-subject standard deviation of 9.45 percentage points [46]. Treating σ = 9.45 as a conservative planning estimate, the number of participants n required to estimate mean cross-subject accuracy within a target margin of error E at 95% confidence follows the standard sample-size formula n = ( z 0.025 σ / E ) 2 with z 0.025 = 1.96 : this yields n = 4 for E = ± 10 percentage points, n = 8 for E = ± 7 , n = 14 for E = ± 5 , and n = 39 for E = ± 3 . We recommend n ≥ 14 (target precision ± 5 percentage points) as a minimum planning figure for the next study phase, with n ≥ 39 preferred if a tighter, clinically interpretable interval is required; both figures should be treated as planning guidance rather than a guarantee, since the anchoring variance was measured on a different acquisition setup and gesture set than the present prosthesis pipeline.

6. Conclusions

This study presented a formalized embedded deep learning pipeline for real-time control of a robotic hand prosthesis using eight-channel sEMG. The system integrates Myo Armband acquisition, causal filtering, sliding-window segmentation, z-score normalization, CNN-BiLSTM classification, TensorFlow Lite inference, majority-vote stabilization and PWM actuation. The model reached 99.6% test accuracy and 99.58% macro F1-score, while embedded operation demonstrated continuous execution with high temporal stability and coherent gesture-to-motion mapping. Beyond this reference model, a deployment-oriented benchmark of four quantized TinyML candidates showed that fixed-point quantization is not free: the int8 TCN ensemble lost approximately 6.7 percentage points of accuracy relative to its float32 form, and a Pareto-dominance analysis across accuracy, footprint and estimated inference time (Section 4.2) identified the compact embedded-C Random Forest and a single decision tree as jointly Pareto-optimal among the four candidates, while the MLP and TCN ensembles were dominated by at least one alternative on both accuracy and resource cost. A model-specific SHAP analysis (GradientExplainer for the TCN ensemble, TreeSHAP for the embedded-C Random Forest) further showed that channel-wise feature importance is model-dependent rather than a fixed physiological property of the acquisition setup: the two model families rank the eight sEMG channels only weakly consistently with one another (Spearman ρ = 0.12 ), reflecting their different input representations of the same raw signal. Together, these results support accessible, explainable edge-AI architectures for myoelectric assistive technologies, but the work should be interpreted as a controlled within-subject proof-of-concept pending cross-subject/cross-user, amputee-user and biomechanical validation.

7. Patents

No patents resulting from this work are reported in this study.

Author Contributions

Conceptualization, C.H.V.N., R.Y.A.S., Y.C.B.A., J.J.L.M. and C.J.G.L.; methodology, R.Y.A.S., Y.C.B.A. and J.J.L.M.; software, R.Y.A.S., Y.C.B.A. and J.J.L.M.; validation, R.Y.A.S., Y.C.B.A. and J.J.L.M.; formal analysis, R.Y.A.S., Y.C.B.A., J.J.L.M. and C.H.V.N.; investigation, R.Y.A.S., Y.C.B.A. and J.J.L.M.; resources, C.H.V.N. and C.J.G.L.; data curation, R.Y.A.S., Y.C.B.A. and J.J.L.M.; writing—original draft preparation, R.Y.A.S., Y.C.B.A. and J.J.L.M.; writing—review and editing, C.H.V.N. and C.J.G.L.; visualization, R.Y.A.S., Y.C.B.A. and J.J.L.M.; supervision, C.H.V.N. (lead) and C.J.G.L. (supporting); project administration, C.H.V.N. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Ethical review and approval were waived for this study because it was conducted as an undergraduate capstone project involving a single able-bodied volunteer performing non-invasive, low-risk surface electromyography recording, with no clinical intervention, no vulnerable population and no institutional ethics committee requirement applicable under the sponsoring institution’s policies; the study was nonetheless conducted in accordance with the Declaration of Helsinki.

Data Availability Statement

The data and code supporting the results of this study are openly available on Zenodo at https://zenodo.org/records/22062945 (accessed on 22 August 2026).

Conflicts of Interest

The authors declare no conflict of interest.

Nomenclature

Symbol or acronymMeaning
sEMGSurface electromyography
CNNConvolutional neural network
BiLSTMBidirectional long short-term memory network
PWMPulse-width modulation
CNumber of EMG channels
WWindow length in samples
HSliding-window hop size
KNumber of predictions used for temporal stabilization
x [ n ] Multichannel EMG sample at index n
X k Segmented EMG window
p k Posterior class-probability vector
y ^ k Instantaneous predicted gesture class
y ¯ k Temporally stabilized gesture class
q k ★ Desired prosthetic finger-position vector
u i , k PWM command for servo i
T total End-to-end response time

References

  1. Farina, D.; Jiang, N.; Rehbaum, H.; Holobar, A.; Graimann, B.; Dietl, H.; Aszmann, O. The extraction of neural information from the surface EMG for the control of upper-limb prostheses: Emerging avenues and challenges. IEEE Trans. Neural Syst. Rehabil. Eng. 2014, 22, 797–809. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  2. Englehart, K.; Hudgins, B. A robust, real-time control scheme for multifunction myoelectric control. IEEE Trans. Biomed. Eng. 2003, 50, 848–854. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  3. Hudgins, B.; Parker, P.; Scott, R.N. A new strategy for multifunction myoelectric control. IEEE Trans. Biomed. Eng. 1993, 40, 82–94. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  4. Ha, K.H.; Varol, H.A.; Goldfarb, M. Volitional control of a prosthetic knee using surface electromyography. IEEE Trans. Biomed. Eng. 2011, 58, 144–151. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  5. Kuiken, T.A.; Li, G.; Lock, B.A.; Lipschutz, R.D.; Miller, L.A.; Stubblefield, K.A.; Englehart, K.B. Targeted muscle reinnervation for real-time myoelectric control of multifunction artificial arms. JAMA 2009, 301, 619–628. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  6. Cipriani, C.; Antfolk, C.; Controzzi, M.; Lundborg, G.; Rosen, B.; Carrozza, M.C.; Sebelius, F. Online myoelectric control of a dexterous hand prosthesis by transradial amputees. IEEE Trans. Neural Syst. Rehabil. Eng. 2011, 19, 260–270. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  7. Oskoei, M.A.; Hu, H. Myoelectric control systems—A survey. Biomed. Signal Process. Control 2007, 2, 275–294. [Google Scholar] [CrossRef] [Scilit]
  8. Scheme, E.; Englehart, K. Electromyogram pattern recognition for control of powered upper-limb prostheses: State of the art and challenges for clinical use. J. Rehabil. Res. Dev. 2011, 48, 643–660. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  9. Chowdhury, R.H.; Reaz, M.B.I.; Ali, M.A.B.M.; Bakar, A.A.A.; Chellappan, K.; Chang, T.G. Surface electromyography signal processing and classification techniques. Sensors 2013, 13, 12431–12466. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  10. Merletti, R.; Lo Conte, L. Surface EMG signal processing during isometric contractions. J. Electromyogr. Kinesiol. 1997, 7, 241–250. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  11. LeCun, Y.; Bengio, Y.; Hinton, G. Deep learning. Nature 2015, 521, 436–444. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  12. Hochreiter, S.; Schmidhuber, J. Long short-term memory. Neural Comput. 1997, 9, 1735–1780. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  13. Tsinganos, P.; Cornelis, B.; Cornelis, J.; Jansen, B.; Skodras, A. Deep learning in EMG-based gesture recognition. In Proceedings of the 5th International Conference on Physiological Computing Systems, Seville, Spain, 19–21 September 2018; pp. 107–114. [Google Scholar] [CrossRef] [Scilit]
  14. Celik, Y.; Can, O. Surface EMG-based hand gesture recognition using a hybrid multistream deep learning architecture. Sensors 2026, 26, 2281. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  15. Shanmuganathan, V.; Barbosa, P.C.; da Cunha Neto, J.S.; Zhang, L.; de Albuquerque, V.H.C.; Pasupathi, S. Development of intelligent and integrated technology for pattern recognition in EMG signals for robotic prosthesis command. Expert Syst. 2023, 40, e13109. [Google Scholar] [CrossRef] [Scilit]
  16. Valencia, C.H.; Vellasco, M.M.B.R.; Figueiredo, K. Echo State Networks: Novel reservoir selection and hyperparameter optimization model for time series forecasting. Neurocomputing 2023, 545, 126317. [Google Scholar] [CrossRef] [Scilit]
  17. Valencia-Niño, C.H.; Núñez-Rodríguez, R.A.; Vellasco, M.M.B.R.; Marin, J. ESNformer: A hybrid reservoir–transformer architecture for interpretable, position-aware classification of structured assessment data, with a Braille-literacy case study. Technologies 2026, 14, 517. [Google Scholar] [CrossRef] [Scilit]
  18. He, J.; Yang, X.; Wang, X. Hand gesture recognition using MYO armband. In Proceedings of the 2017 Chinese Automation Congress, Jinan, China, 20–22 October 2017; pp. 4850–4855. [Google Scholar] [CrossRef] [Scilit]
  19. Kaya, O.; Kumbasar, T. Hand gesture recognition systems with the wearable Myo Armband. In Proceedings of the 2018 6th International Conference on Control Engineering and Information Technology, Istanbul, Turkey, 25–27 October 2018. [Google Scholar] [CrossRef] [Scilit]
  20. Sathiyanarayanan, M.; Rajan, S. MYO Armband for physiotherapy healthcare: A case study using gesture recognition application. In Proceedings of the 2016 8th International Conference on Communication Systems and Networks, Bangalore, India, 5–9 January 2016. [Google Scholar] [CrossRef] [Scilit]
  21. Boyali, A.; Hashimoto, N.; Matsumoto, O. Hand posture and gesture recognition using MYO armband and spectral collaborative representation based classification. In Proceedings of the 2015 IEEE 4th Global Conference on Consumer Electronics, Osaka, Japan, 27–30 October 2015. [Google Scholar] [CrossRef] [Scilit]
  22. Benalcazar, M.E.; Jaramillo, A.G.; Zea, J.A.; Paez, A.; Andaluz, V.H. Hand gesture recognition using machine learning and the Myo armband. In Proceedings of the 25th European Signal Processing Conference, Kos, Greece, 28 August–2 September 2017. [Google Scholar] [CrossRef] [Scilit]
  23. Nahid, A.A.; Rahman, M.M.; Ahad, M.A.R. Deep learning based surface EMG hand gesture classification for low-cost myoelectric prosthetic control. In Proceedings of the Joint 9th International Conference on Informatics, Electronics and Vision and 4th International Conference on Imaging, Vision and Pattern Recognition, Kitakyushu, Japan, 26–29 August 2020. [Google Scholar] [CrossRef] [Scilit]
  24. Chung, E.A.; Benalcazar, M.E. Real-time hand gesture recognition model using deep learning techniques and EMG signals. In Proceedings of the 27th European Signal Processing Conference, A Coruna, Spain, 2–6 September 2019. [Google Scholar] [CrossRef] [Scilit]
  25. Bibbò, L.; Laganà, F.; Pullano, S.A.; Angiulli, G. Multimodal EEG–EMG and FEM-based adaptive control of passive upper-limb exoskeletons. Sensors 2026, 26, 3924. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  26. Dunai, L.; Segui Verdu, I.; Turcanu, D. Prosthetic hand based on human hand anatomy controlled by surface electromyography and artificial neural networks. Technologies 2025, 13, 21. [Google Scholar] [CrossRef] [Scilit]
  27. Covaciu, F.; Gherman, B.; Vaida, C. A combined mirror–EMG robot-assisted therapy system for lower limb rehabilitation. Technologies 2025, 13, 227. [Google Scholar] [CrossRef] [Scilit]
  28. Lotfi, A.; Langensiepen, C.; Yahaya, S.W. Socially assistive robotics: Robot exercise trainer for older adults. Technologies 2018, 6, 32. [Google Scholar] [CrossRef] [Scilit]
  29. Stasolla, F.; Lopez, A.; Akbar, K. Matching assistive technology, telerehabilitation, and virtual reality to promote cognitive rehabilitation. Technologies 2023, 11, 43. [Google Scholar] [CrossRef] [Scilit]
  30. Bush, J.; Benham, S.; Kaniamattam, M. Development and evaluation of an mHealth app that promotes access to 3D printable assistive devices. Technologies 2024, 12, 114. [Google Scholar] [CrossRef] [Scilit]
  31. Giansanti, D. Empowering independence: The role of assistive technologies in enhancing quality of life. Technologies 2025, 13, 408. [Google Scholar] [CrossRef] [Scilit]
  32. Manakitsa, N.; Maraslidis, G.S.; Moysis, L. A review of machine learning and deep learning for object detection, semantic segmentation, and human action recognition. Technologies 2024, 12, 15. [Google Scholar] [CrossRef] [Scilit]
  33. Percuku, A.; Minkovska, D.; Hinov, N. Enhancing electricity load forecasting with machine learning and deep learning. Technologies 2025, 13, 59. [Google Scholar] [CrossRef] [Scilit]
  34. Jalloul, R.; Krishnappa, C.H.; Agughasi, V.I. Enhancing early breast cancer detection with infrared thermography: A comparative evaluation of deep learning approaches. Technologies 2025, 13, 7. [Google Scholar] [CrossRef] [Scilit]
  35. Kouretas, I.; Paliouras, V. Hardware implementation of a softmax-like function for deep learning. Technologies 2020, 8, 46. [Google Scholar] [CrossRef] [Scilit]
  36. Kraft, D.; Srinivasan, K.; Bieber, G. Deep learning based fall detection algorithms for embedded systems, smartwatches, and IoT devices using accelerometers. Technologies 2020, 8, 72. [Google Scholar] [CrossRef] [Scilit]
  37. Sanida, T.; Sideris, A.; Tsiktsiris, D. Lightweight neural network for COVID-19 detection from chest X-ray images implemented on an embedded system. Technologies 2022, 10, 37. [Google Scholar] [CrossRef] [Scilit]
  38. Phinyomark, A.; Phukpattaranont, P.; Limsakul, C. Feature reduction and selection for EMG signal classification. Expert Syst. Appl. 2012, 39, 7420–7431. [Google Scholar] [CrossRef] [Scilit]
  39. Atzori, M.; Gijsberts, A.; Castellini, C.; Caputo, B.; Hager, A.G.M.; Elsig, S.; Giatsidis, G.; Bassetto, F.; Muller, H. Electromyography data for non-invasive naturally-controlled robotic hand prostheses. Sci. Data 2014, 1, 140053. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  40. Atzori, M.; Gijsberts, A.; Heynen, S.; Hager, A.G.M.; Deriaz, O.; Van Der Smagt, P.; Castellini, C.; Caputo, B.; Muller, H. The Ninapro database: A resource for sEMG naturally controlled robotic hand prosthetics. In Proceedings of the 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Milan, Italy, 25–29 August 2015; pp. 7151–7154. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  41. Muceli, S. Multi-Channel Surface Electromyography for Simultaneous and Proportional Control of Prostheses; River Publishers: Aalborg, Denmark, 2013. [Google Scholar] [CrossRef] [Scilit]
  42. Prakash, A.; Sharma, S. Single-channel surface electromyography (sEMG) based control of a multi-functional prosthetic hand. Instrum. Sci. Technol. 2021, 49, 428–444. [Google Scholar] [CrossRef] [Scilit]
  43. Rodríguez-Torres, A.; Valencia-Niño, C.H.; Alvarez-Icaza, L. Application of classical and quantum-inspired methods through multi-objective optimization for parameter identification of a multi-story prototype building. Buildings 2025, 15, 3743. [Google Scholar] [CrossRef] [Scilit]
  44. Zaman, N. TensorFlow Lite for Mobile Development; Apress: Berkeley, CA, USA, 2020. [Google Scholar] [CrossRef] [Scilit]
  45. González, H.; Arizmendi, C.; Valencia, C.; Valle, D.; Bernal, B.; Vera, A. Modelling and control system design for UAV tri-rotor. In AETA 2019—Recent Advances in Electrical Engineering and Related Sciences: Theory and Application; Cortes Tobar, D., Hoang Duy, V., Trong Dao, T., Eds.; Lecture Notes in Electrical Engineering; Springer: Cham, Switzerland, 2021; Volume 685, pp. 90–99. [Google Scholar] [CrossRef] [Scilit]
  46. Barona López, L.I.; Ferri, F.M.; Zea, J.; Valdivieso Caraguay, Á.L.; Benalcázar, M.E. CNN–LSTM and post-processing for EMG-based hand gesture recognition. Intell. Syst. Appl. 2024, 22, 200352. [Google Scholar] [CrossRef] [Scilit]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Article Metrics

Citations

Article Access Statistics

Multiple requests from the same IP address are counted as one view.