Abstract
Biliary stents restore drainage in malignant biliary obstruction, but recurrent dysfunction creates a need to monitor the patency of the stent. We developed a pressure-sensing smart biliary stent by integrating a flexible contact-resistive sensor, comprising opposing Au mesh electrodes on liquid crystal polymer films, into a commercial self-expandable metal stent connected to an external monitoring system. Under loading through a compliant pad matched to the 1 mm × 2 mm sensing window (n = 4 sensors), the normalized resistance decreased approximately linearly up to 2.5 kPa and approached a plateau at approximately 3.0 kPa, with a sensitivity of −37.0 ± 0.4% kPa−1 in the linear region, a pressure hysteresis of 4.1 ± 1.5% of full scale, and, in a representative sensor, a variation within ±2.1% of the mean over five loading–unloading cycles. Between 30 and 45 °C, measured in the solution at the stent, the step means of ΔR/R0 remained within 0.10% of zero (n = 4 sensors). In recirculating model bile inoculated with Escherichia coli, the sensor output in the run with a complete recording began to change after approximately 400 min and reached the plateau of the response after approximately 840 min, whereas in sterile model bile the mean output of four sensors remained within 0.15% of baseline over 14 h. In a porcine biliary stricture model, the device was deployed endoscopically in one animal, its wires were externalized through a percutaneous endoscopic gastrostomy tube, and the monitor, zeroed in vivo after connection, displayed a change in readout after start-up, which was not verified against a reference pressure. The connection wire broke inside the percutaneous endoscopic gastrostomy tube within 1 week and had also broken at several steps of the procedure when it was pulled; in bench tests, it fractured at a monotonic tensile load of 1.878 N. As an initial feasibility study toward clinical application, these results indicate that sensor integration, endoscopic delivery, wire routing, and external readout can be achieved, identify the procedural steps at which the sensor and its wires failed, and define the extension of the measurement range and the protection of the connection that are required before occlusion monitoring can be evaluated in vivo.
1. Introduction
Malignant biliary obstruction (MBO) occurs in cholangiocarcinoma and other pancreatobiliary malignancies and can compromise bile drainage [1]. Endoscopic placement of self-expandable metal stents (SEMSs) is an established approach to drainage in selected patients with malignant extrahepatic obstruction [2]. Improvements in symptoms and quality of life have been reported after endoscopic stenting [3]. A meta-analysis of randomized trials has also found longer patency and fewer reinterventions with SEMSs than with plastic stents in malignant distal obstruction [4]. Nevertheless, subsequent dysfunction of the stent can interrupt drainage, creating a need to assess the condition of the implanted stent during follow-up.
Biliary stent dysfunction can result from various mechanisms, including tumor ingrowth or overgrowth, stent migration, and the accumulation of sludge or debris. In a meta-analysis of fully covered and partially covered SEMSs for distal malignant biliary obstruction, the rates of tumor ingrowth and stent migration differed slightly between the two designs [5]. Among these mechanisms, stent migration was more frequent with fully covered than with uncovered SEMSs in a large cohort, whereas occlusion was more frequent with uncovered SEMSs [6]. Accordingly, the TOKYO criteria define recurrent biliary obstruction as encompassing both stent occlusion and migration [7]. In addition, a review and analyses of occluded biliary stents have described the involvement of bacteria, sludge, and polymicrobial biofilms in stent occlusion [8,9,10]. Given these diverse mechanisms, changes detected by a stent-based monitoring system should be interpreted cautiously, as alterations in local conditions do not necessarily identify the underlying cause of stent dysfunction.
International consensus recommendations describe on-demand reintervention for malignant distal biliary strictures treated with metal stents and note the limited evidence for an optimal strategy to detect recurrent obstruction early [11]. Clinical assessment, liver function tests, and imaging remain important in evaluating suspected dysfunction. An implanted sensor could potentially add information between these assessments, but its signal must be validated against independently established stent failure.
Pressure provides functional information that complements anatomical imaging. Manometric studies during percutaneous biliary procedures have examined pressure under controlled perfusion in bile ducts with strictures [12]. More recently, measured intrabiliary pressure was associated with the level of extrahepatic obstruction and baseline serum bilirubin, but did not predict cholangitis or the resolution of jaundice after drainage [13]. These findings support investigating pressure as a monitoring variable, but a pressure change alone is not a validated diagnostic criterion for stent occlusion.
Sensor-based monitoring of implanted stents has been reviewed for in-stent restenosis in the vascular field [14]. For biliary stents, a magnetoelastic resonant sensor, in which a shift in resonant frequency is intended to indicate material accumulation, was integrated into a plastic stent, and wireless signals were obtained in vivo after endoscopic implantation in a pig [15]. Materials and surface modifications have also been investigated to address restenosis and deposition on biliary stents [16]. These developments motivate complementary approaches that provide information about the function of an implanted stent. For pressure sensing, the main challenges include fitting the sensing element into the stent, maintaining fluid isolation, establishing an electrical connection after deployment, and preserving that connection during follow-up.
Flexible pressure sensors have been developed using resistive [17] and capacitive [18] transduction mechanisms. A wireless stent-based system with printed soft pressure sensors read out by inductive coupling [19], a biodegradable wireless passive pressure sensor made from silk protein [20], and an injectable bioresorbable sensor for intracranial pressure and temperature measurements [21] demonstrate different approaches to implantable sensing. Protective barrier layers have further been used to isolate bioresorbable pressure sensors from biofluids and to control their operational lifetime [22]. Applying such concepts to the biliary environment requires the same validation after integration with a deformable stent and under sustained exposure to bile.
Herein, we present a wired pressure-sensing smart biliary stent and evaluate the complete route from sensor integration to external readout. We incorporated a flexible contact-resistive sensor into a commercial SEMS and characterized its sensitivity, hysteresis, linearity, and repeatability under loading matched to the sensing window. We then measured its temperature dependence in model bile under flow and examined its response in a recirculating bile flow model with and without bacteria. To assess endoscopic deployment, percutaneous endoscopic gastrostomy (PEG)-based wire externalization, and measurement using a pressure monitoring system developed for this study, we used a porcine biliary stricture model. Finally, because the connection wire broke during follow-up, we examined its failure under tensile loading, the condition under which it broke during the procedures. This study was intended as an initial evaluation of feasibility toward clinical application, that is, of whether each step from sensor integration to external readout can be performed and where it fails, rather than as a validation of occlusion monitoring.
2. Materials and Methods
2.1. Pressure Sensor Design and Characterization
The sensor comprised two opposing Au mesh electrodes, each patterned on a 25 μm liquid crystal polymer (LCP) film (VECSTAR CTQ-25; Kuraray, Tokyo, Japan) (Figure 1a,b). LCP has previously been investigated as a substrate and encapsulation material for implantable electronics [23]. Au electrodes (100 nm thick; nominal line width, 20 μm; pitch, 200 μm) were deposited by thermal evaporation and patterned by photolithography and lift-off, giving an active area of 2 mm2. The two LCP films were bonded face to face using a frame-shaped double-sided adhesive tape (93020LE; 3M, St. Paul, MN, USA) approximately 200 μm thick, which defined the film separation at the perimeter and left the active area unsupported. The outer edges were then sealed with cyanoacrylate adhesive (LOCTITE 4061; Henkel, Düsseldorf, Germany). Ideally, the opposing meshes would remain separated under no load, yielding an open circuit. Because the films were bonded without pre-tension, the unsupported central regions came into partial contact, producing a finite unloaded resistance, R0. Applied pressure increased the contact area between the meshes and reduced the contact resistance. Before characterization, the device was repeatedly loaded and unloaded to condition the response.
The electrical response was described using a lumped contact-resistance model:
Here, Rtotal is the total resistance, Relectrode is the electrode resistance, and Rcontact(P) is the pressure-dependent contact resistance. Electrical contact theory relates the constriction of current through interfacial contact regions to the mechanical contact state [24]. This relationship provides a physical rationale for the proposed decrease in resistance as loading increases contact between the meshes. The lumped model is a qualitative description of transduction and does not by itself define a resistance-to-pressure calibration or demonstrate contact formation in the assembled sensor. Because partial contact was present before loading, the unloaded contact resistance was finite, and the measured response reflects growth of an existing contact area rather than closure of an open gap.
Finite-element analysis (FEA) was performed in COMSOL Multiphysics 5.4 (COMSOL AB, Stockholm, Sweden) for the 25 μm LCP membrane and 100 nm Au meshes under uniform pressures of 1, 2, and 4 kPa. The LCP film was modeled as a linear elastic material with a density of 1400 kg/m3, a Poisson’s ratio of 0.34, and a Young’s modulus of 11 GPa. Au material properties were taken from the COMSOL material library. A fixed constraint was applied along the bonded perimeter, and the pressure load was applied uniformly to the exposed membrane surface. The first principal strain was used to describe the calculated mechanical response (Figure 1c). For characterization of the bare sensor, pressure was applied using a CT3 texture analyzer (AMETEK Brookfield, Middleboro, MA, USA). To confine the load to the sensing window (1 mm × 2 mm) and distribute it over the window, a polydimethylsiloxane (PDMS) pad with a face of 1 mm × 2 mm and a thickness of 1 mm (SYLGARD 184; Dow Corning, Midland, MI, USA; base to curing agent ratio, 10:1) was attached to the tip of a cylindrical stainless-steel probe 2 mm in diameter (TA39) with a double-sided tape that has a silicone adhesive on one side and an acrylic adhesive on the other (Chukoh Chemical Industries, Tokyo, Japan), so that only the pad contacted the sensor. The pad was aligned with the window, and the applied pressure was calculated as the force recorded by the load cell divided by the area of the pad face (2 mm2). The pad was intended to approximate uniform loading; because the distribution of contact pressure under the pad was not measured, this value is referred to as the nominal applied pressure. The electrical response of each of the four fabricated sensors was characterized over an applied-pressure range of 0–4 kPa in steps of 0.5 kPa during loading and unloading and expressed as ΔR/R0, where ΔR = R − R0 and R0 is the resistance at the unloaded reference condition. Loading through the pad differs from uniform fluid pressure, and the relationship between the nominal applied pressure and fluid pressure was not measured.
2.2. Stent Integration, External Readout, and Zeroing
Commercial self-expandable Nitinol stents (Taewoong Medical, Gimpo, Republic of Korea) with nominal diameters of 8–10 mm and lengths of 60–80 mm were used. The recoverable deformation of superelastic Nitinol is an established basis for self-expanding medical devices [25]. The pressure sensor was attached to the central inner surface of the stent using cyanoacrylate adhesive (Figure 1g). Insulated copper wires connected the sensor to external readout equipment. The polyurethane-coated wire (SX0090-PURP-050; Scientific Wire Company, Great Dunmow, UK) had a nominal diameter of 0.09 mm and a length of 2 m. A helical route along the stent frame was selected to accommodate expansion and contraction during delivery and deployment.
Figure 1.
Sensor architecture, characterization, readout, and integration with a biliary stent. (a) Exploded schematic of the opposing Au mesh electrodes and frame-shaped adhesive spacer, with optical images of the patterned electrode. Scale bars: 2 mm in the overview and 200 μm in the magnified image. (b) Schematic of pressure-induced contact between the Au meshes and the corresponding resistance model; in the fabricated sensor, partial contact between the central mesh regions produced a finite unloaded resistance, R0. (c) Calculated first principal strain at applied pressures of 1, 2, and 4 kPa. (d) ΔR/R0 of four sensors as a function of the nominal applied pressure (force divided by the 2 mm2 face of the PDMS pad) during loading in steps of 0.5 kPa; the legend gives R0 of each sensor, and the inset shows loading through the PDMS pad. The dashed vertical line marks the common bile duct pressure reported in pigs (0.81 ± 0.06 kPa) [26], and the shaded band the intrabiliary pressure measured in patients with extrahepatic biliary obstruction (9–28 mmHg, 1.2–3.7 kPa) [13], with the dotted line at its mean (18.4 mmHg, 2.5 kPa); because loading through the pad differs from fluid pressure, this comparison is approximate. (e) ΔR/R0 during five consecutive loading (solid lines) and unloading (dashed lines) cycles of one representative sensor (S2; R0 = 5.3 kΩ). (f) Sensor resistance (left axis; light blue, individual sensors; black, mean ± SD, n = 4) and divider output Vout (right axis; mean ± SD, n = 4) as functions of the nominal applied pressure; Vout was calculated from the measured resistance with VCC = 3.3 V and Rm = 1 kΩ. Inset: voltage-divider readout used in the flow and porcine experiments, with Vout measured across Rm. (g) Photographs of the covered SEMS, sensor attachment, and delivery assembly; scale bar: 2 mm. In (a), black arrows indicate the components, and the blue dotted box and blue arrows indicate the magnified regions. In (b), the blue and red block arrows indicate the applied pressure before and after the pressure increase, respectively, and the blue dashed box indicates the contact region between the meshes; Relectrode is shown in red and Rcontact(P) in blue. In (c), the color scale indicates the first principal strain. In (e), the arrows indicate the loading and unloading directions. In (g), the blue arrow indicates the attachment of the sensor to the covered SEMS, the yellow dotted lines indicate the sensor location and the magnified view, and the red dotted arrows indicate the PU-coated wires.
An external pressure monitoring system was developed for this study to acquire the electrical output of the resistive sensor and present a graphical and numerical readout on an integrated display. The sensor was read in a voltage-divider configuration (Figure 1f, inset). The sensor was connected in series with a 1 kΩ reference resistor, Rm, and the series pair was driven by a constant supply voltage, VCC = 3.3 V (voltage mode). The divider output, Vout = VCC·Rm/(Rsensor + Rm), was digitized by the built-in 12-bit analog-to-digital converter of an Arduino Nano ESP32 board (ESP32-S3; codes 0–4095). The board read the output every 100 ms (nominal 10 Hz), and the readings in each 1 s interval were averaged to record one value per second. Each 1 s value is a time average of consecutive readings from the same sensor and does not represent an independent replicate. The sensor resistance was calculated as Rsensor = Rm(VCC/Vout − 1). Because Rsensor decreased with pressure, Vout increased with pressure. With Rm = 1 kΩ, which lies between R0 (5.3–9.5 kΩ) and the loaded resistance at the upper end of the linear region (approximately 0.5 kΩ), the fractional sensitivity of the divider, Rm·Rsensor/(Rsensor + Rm)2, is greatest near Rsensor ≈ Rm, in the upper part of the range; this partly offsets the flattening of the sensor response, whereas at low pressures, where Rsensor ≫ Rm, Vout remains small (approximately 0.31–0.52 V without load). This circuit was used in the flow and porcine experiments; in the bench characterization, the resistance was measured with a separate instrument (Section 2.3). The unloaded resistance of each sensor was measured at room temperature after fabrication (Figure 1d; Table 1). Zeroing was performed in software, without hardware offset adjustment: when a recording was started, the monitoring program took the mean resistance during the first 30 s as R0 and expressed subsequent readings as ΔR/R0 relative to this value. No further offset correction was applied during a recording. The monitoring system displayed the output as a pressure converted from ΔR/R0 with a bench calibration. In the flow experiment, the output was recorded as a pressure calculated from ΔR/R0 with the calibration available at the time of the experiment. Because that calibration was replaced by the one described in Section 2.3, the recorded values were first converted back to ΔR/R0 with the same calibration relation that had been used to calculate them; this inverse conversion restores the measured change in resistance and does not depend on the accuracy of the earlier calibration. The result of the flow experiment is therefore reported as ΔR/R0, the quantity measured by the sensor. The corresponding nominal pressure was then assigned from the mean relation between ΔR/R0 and pressure of the four sensors measured in Section 2.3 (Figure 1d). It is shown as a reference scale and not as a measured pressure, because the sensor used in the flow experiment was not among the four calibrated sensors and because the calibration was obtained by loading the bare sensor through a pad and not under fluid pressure; values at the plateau indicate only that the pressure was at or above Pmax.
2.3. Loading–Unloading Characterization
Each of the four sensors was loaded through the PDMS pad described in Section 2.1 from 0 to 4 kPa and unloaded to 0 kPa in steps of 0.5 kPa at room temperature, and five consecutive loading–unloading cycles were recorded to evaluate repeatability (Figure 1e). The resistance was measured with a digital multimeter (Fluke 87V; Fluke Corporation, Everett, WA, USA), and the corresponding divider output was calculated as Vout = VCC·Rm/(Rsensor + Rm) with VCC = 3.3 V and Rm = 1 kΩ. Because the pad and the sensor were loaded in series, the pressure at each step was calculated from the force recorded by the load cell rather than from the set point.
2.4. In Vitro Flow Experiment
The flow model comprised a reservoir, a peristaltic pump (BT101L with a DG10-4 pump head; Lead Fluid, Baoding, China), transparent silicone tubing with an inner diameter of 8 mm, and the sensor-integrated stent (Figure 2). This in vitro biliary flow model was designed to induce biofilm formation with Escherichia coli and to examine whether the resulting deposits obstruct the stent and, if so, how the sensor output changes. The initial event in biliary stent clogging has been described as the development of an adherent bacterial biofilm [27], and E. coli is among the organisms most frequently detected on biliary stents [28]. A bile-mimicking solution was prepared by dissolving porcine bile extract (B8631; Sigma-Aldrich, St. Louis, MO, USA) in deionized (DI) water at 5 wt%. The solution was recirculated from an open reservoir by the peristaltic pump at 0.5–0.7 mL/min, and the reservoir was maintained at 37 °C; the temperature at the stent was not measured.
Figure 2.
Experimental setup for the in vitro flow experiment. The system contains a warmed reservoir of 5 wt% porcine bile solution, a peristaltic pump, silicone tubing containing the stent, and electronic circuitry for sensor readout. Solution was recirculated at 0.5–0.7 mL/min, and the experimental temperature was controlled at 37 °C. The photograph identifies the reservoir warming arrangement, pump, monitoring circuitry, and stent-containing tubing.
Escherichia coli was introduced into the bile-mimicking solution at an inoculum of 1 × 108 colony-forming units (CFU). Sensor output was recorded continuously for 4500 min during circulation, and the tubing and retrieved stent were examined visually after the experiment. The inoculated flow experiment was performed three times, and a complete recording of the sensor output was obtained in one run, which is reported here. The sterile run of Section 2.5 used the same loop, flow rate, and solution, with the four sensors of Section 2.3 and without inoculation; it was performed after the inoculated runs and with the temperature at the stent varied between 30 and 45 °C, and therefore indicates the baseline behavior of the system without bacteria rather than a concurrent control.
2.5. Evaluation of Temperature Dependence and Baseline Stability
Because the cavity between the films was sealed at its edges, the response could depend on temperature. To evaluate this dependence under the conditions of the flow experiment, the four sensors characterized in Section 2.3 were each mounted on a stent and installed in the flow model of Section 2.4. Sterile 5 wt% model bile without bacteria was circulated at 0.694 mL/min. The temperature of the solution was measured with the probe of a waterproof two-channel digital thermometer (T9234T; resolution, 0.1 °C; DAIHAN Scientific, Wonju, Republic of Korea), the same instrument used in the flow experiments, whose probe was inserted into the silicone tubing (inner diameter, 8 mm) that contained the stent, adjacent to the stent, so that the temperature of the solution was measured near the sensor. The solution was heated so that this measured temperature reached 30, 35, 40, 45, 40, 35, and 30 °C in sequence, and each step lasted 2 h (seven steps; 14 h in total), so that the response was recorded during both heating and cooling. The sensor output at each step was taken once the measured temperature had reached the target value. ΔR/R0 of each sensor was referenced to the R0 taken during the first 30 s of the recording at 30 °C, and the zero was not reset between temperatures. The mean and standard deviation of ΔR/R0 were calculated for each step from the time at which the measured temperature reached the target value, and the temperature coefficient was obtained by linear regression of the step means against temperature. Thermal hysteresis was taken as the largest difference between the step means at the same temperature during heating and cooling.
2.6. Porcine Biliary Stricture Model
Swine provide a relevant setting for surgical and device development because of anatomical and physiological similarities to humans [29], although species-specific differences must be considered. A total of 2 pigs weighing 30 ± 5 kg were used, and the experiments were approved by the Institutional Animal Care and Use Committee (IACUC; approval number P2023-016-A1C0). Animals were maintained under a 12 h light/dark cycle with ad libitum feeding and acclimatized for at least 7 days before experimentation. The number of animals entering each procedural stage, together with any exclusions and deaths and their causes, is reported according to the ARRIVE 2.0 guidelines [30] (Section 3.5).
Procedures were performed under general anesthesia with xylazine premedication, alfaxalone induction, and isoflurane maintenance. Heart rate, oxygen saturation, respiratory rate, and body temperature were monitored. Intravenous gentamicin and cefazolin were administered peri-procedurally, followed by oral cefaclor.
Endobiliary radiofrequency ablation (RFA) was used to induce a localized biliary stricture, following the general approach described in porcine models of RFA [31,32]. Under endoscopic retrograde cholangiopancreatography (ERCP) guidance, the RFA catheter was positioned in the distal common bile duct. The applied settings were 7–10 W, a target temperature of 80 °C, and a duration of 60–90 s. Follow-up cholangiography was performed 2 weeks after RFA to assess ductal narrowing and proximal dilation. One pig died after PEG placement, after attempts to deploy the sensor-integrated stent had failed and deployment had been postponed; the cause of death was not determined, but it was presumed to be cholangitis caused by the prolonged unrelieved biliary stricture (Section 3.5). Consequently, the subsequent in vivo evaluation of the smart biliary stent was performed in the remaining pig.
2.7. PEG Access for Wire Externalization
PEG provided a conduit for externalizing the sensor wires. The procedure used endoscopic gastric access and the pull-through technique, which a current guideline recommends as the standard method for PEG placement [33]. After confirmation of the biliary stricture, endoscopic visualization and transillumination were used to select the percutaneous insertion site. A 20 Fr PEG tube was placed and secured by an internal bumper and an external retainer. Because the wires were to be externalized through this tube, deployment of the sensor-integrated stent was scheduled after the PEG tract had matured (approximately 4 weeks after PEG placement).
Figure 3 illustrates the intended connection pathway from the stent-integrated sensor in the bile duct, through the duodenum and stomach, and out through the PEG tube to the external pressure monitoring system. The inset shows tube fixation across the gastric and abdominal walls. The schematic depicts the route and component arrangement; the measurement procedure is described in Section 2.8.
Figure 3.
Schematic of PEG-mediated wire externalization and connection to the pressure monitoring system. Sensor wires extend from the smart biliary stent through the duodenum and stomach and exit through a 20 Fr PEG tube for connection to the external monitor. The inset depicts fixation of the PEG tube across the gastric and abdominal walls using an internal bumper and an external retainer. The monitor display is illustrative, and the anatomy and device components are not shown to scale. The illustration was generated with the assistance of a generative artificial intelligence tool (ChatGPT (GPT-6; OpenAI, San Francisco, CA, USA); OpenAI, San Francisco, CA, USA) and edited by the authors; it is a conceptual schematic and does not accurately depict porcine anatomy.
2.8. Stent Deployment and In Vivo Readout
A side-viewing endoscope was used to access the duodenal papilla, and a guidewire was advanced beyond the stricture. Contrast was injected to identify the stricture. The sensor-integrated stent was advanced on its delivery catheter and deployed across the narrowed segment under fluoroscopic guidance. The stent position was assessed by post-deployment imaging.
After deployment, the sensor-connected wires were accessed in the stomach using a gastroscope, grasped with endoscopic forceps, and drawn out through the PEG tube. The resistance of the sensor was first measured through the externalized wires with the digital multimeter to confirm electrical continuity, and the wires were then connected to the pressure monitoring system developed for this study, which acquired and displayed the sensor output during the porcine experiment. The unit provided a real-time display without data storage, and readings were documented from the display. External duodenal compression was applied as a qualitative functional check of the sensor response after deployment.
Post-procedural observations included survival, feeding, vital signs, behavior, and visible complications at the access site. The condition of the wire connection was assessed at follow-up, and these observations were used to evaluate procedural feasibility and connection integrity. Necropsy was performed at the end of follow-up.
2.9. Tensile Testing of the Connection Wire
In most of the wire breakages during the procedures (Section 3.5), the wire broke while it was being pulled, and tension on the wire was also expected in the animal between its fixation at the cap of the PEG tube and the stent. To evaluate this loading condition, the insulated connection wire was loaded in monotonic tension until fracture on the bench using a Brookfield CT3 texture analyzer, and load was recorded in newtons (N) as a function of displacement.
2.10. Data Analysis
Segment sensitivity was calculated as S = Δ(ΔR/R0)/ΔP over three segments: from 0 to Plin, from Plin to Pmax, and from Pmax to 4 kPa, where Plin (2.5 kPa) is the upper end of the approximately linear region and Pmax (3.0 kPa) the pressure above which ΔR/R0 changed little. The output sensitivity of the divider was calculated as SV = ΔVout/(VCC·ΔP) (mV V−1 kPa−1), and the resistance sensitivity as ΔR/ΔP (Ω kPa−1, equivalent to mV mA−1 kPa−1 under constant-current excitation), for the same segments. Unlike S, SV depends on Rm as well as on the sensor, because dVout/dRsensor = −VCC·Rm/(Rsensor + Rm)2. Full scale (FS) was defined as the output span between 0 and Pmax. Pressure hysteresis was the largest difference between the loading and unloading outputs at the same pressure, expressed as a percentage of FS. Linearity was evaluated over 0–Plin as the coefficient of determination of a straight-line fit and the largest deviation from that line, expressed as a percentage of the span of this region. Repeatability was expressed as the largest deviation of ΔR/R0 from its five-cycle mean at the same pressure, and reproducibility as the coefficient of variation in ΔR/R0 across the four sensors at each pressure. The zero-temperature coefficient was the slope of a linear regression of the step means of ΔR/R0 on temperature during heating, and thermal hysteresis was the largest difference between the step means at the same temperature during heating and cooling. Values are reported as mean ± SD with the number of sensors or runs; given the small numbers, no hypothesis tests were performed.
3. Results
3.1. Sensor Response and Stent Integration
Figure 1a,b,g shows the sensor architecture and the assembled device. The calculated first principal strain increased with the applied pressure up to 4 kPa (Figure 1c), indicating a pressure-dependent mechanical response of the modeled structure.
Measured resistance decreased as applied pressure increased in all four sensors (Figure 1d). ΔR/R0 decreased approximately linearly up to 2.5 kPa, reaching −92.4 ± 1.1% (mean ± SD, n = 4 sensors), and changed little between 3.0 and 4.0 kPa (−2.3 ± 0.4% kPa−1); at 4.0 kPa, the resistance was 10–20 Ω. For the current sensor dimensions, we therefore took 3.0 kPa (Pmax) as the upper limit of the sensing range. This flattening of the response matters when converting resistance to pressure, because a large change in pressure at the upper end of the range produces only a small change in electrical output.
Figure 1g shows the incorporation of the sensor and its thin-wire connection into the stent and delivery assembly. The unloaded resistance R0 of the four fabricated sensors was 7.2, 5.3, 9.5, and 5.9 kΩ at room temperature (7.0 ± 1.9 kΩ, mean ± SD; coefficient of variation, 27%; Figure 1d). Because R0 was determined by the extent of the initial partial contact between the meshes, it varied between devices; each sensor was therefore zeroed with its own R0, and the response is expressed as ΔR/R0.
3.2. Sensitivity, Hysteresis, Linearity, and Repeatability
In all four sensors, ΔR/R0 decreased approximately linearly with pressure up to Plin = 2.5 kPa (R2 ≥ 0.999 in each sensor; largest deviation from the fitted line, 0.9–1.6% of the span) and reached the plateau at Pmax = 3.0 kPa (Figure 1d). The sensitivity was −37.0 ± 0.4, −10.1 ± 1.4, and −2.3 ± 0.4% kPa−1 in the three segments (mean ± SD, n = 4); the corresponding output sensitivities of the divider were 212 ± 18, 381 ± 40, and 135 ± 45 mV V−1 kPa−1, and the resistance sensitivity in the linear region was 2.58 ± 0.68 kΩ kPa−1, equivalent to 2.58 ± 0.68 mV mA−1 Pa−1 under constant-current excitation (Figure 1f; Table 1).
Pressure hysteresis was 4.1 ± 1.5% FS (range, 2.6–5.7%; n = 4), with the largest difference between loading and unloading at 1.0 kPa in all four sensors. During unloading, ΔR/R0 remained lower than during loading at the same pressure (Figure 1e), which is consistent with adhesion of the contacts formed during loading. Over the five cycles of the representative sensor shown in Figure 1e, ΔR/R0 at each pressure remained within ±2.1% of the five-cycle mean, with the largest deviation at 1.0 kPa, and the hysteresis of the individual cycles was 4.5 ± 0.4% FS (Figure 1e). The coefficient of variation in ΔR/R0 across sensors decreased from 6.5% at 0.5 kPa to 3.2% at 1.0 kPa and 1.1% at Plin.
3.3. Sensor Output During the In Vitro Flow Experiment
In the inoculated run with a complete recording, ΔR/R0 initially remained near its baseline and then decreased to the plateau of the response, approximately −98% (Figure 4a). The output began to change after approximately 400 min, passed the value corresponding to Plin after approximately 630 min, and reached the plateau after 840 min. The output then remained at the plateau until the end of the 4500 min recording. This flat segment indicates that the load on the sensor was at or above the upper end of the measurement range, not a stable value.
Figure 4.
Sensor output, visible deposits, and temperature dependence in the in vitro flow model. (a) ΔR/R0 as a function of time in the inoculated run with a complete recording (one of three runs performed); the left axis is reversed so that an increase in load is plotted upward. The right axis gives the corresponding nominal pressure from the mean calibration of Figure 1d as a reference scale only (Section 2.2); values at the plateau indicate a pressure at or above Pmax. The output departs from the initial baseline after approximately 400 min and reaches the plateau after 840 min, where it remains until the end of the 4500 min recording. (b) Photographs of the tubing assembly before and after the experiment, including an end view of the post-experiment tubing. (c) Front and side views of the retrieved stent showing visible deposits. The photographs document gross appearance and do not determine deposit composition. (d) ΔR/R0 of four sensors during circulation of sterile 5 wt% model bile at 0.694 mL/min (mean ± SD of the four sensors; right axis) and the temperature of the solution measured adjacent to the stent (left axis), which was held at 30, 35, 40, 45, 40, 35, and 30 °C for 2 h each; ΔR/R0 was referenced to R0 taken during the first 30 s at 30 °C without re-zeroing. (e) Step mean of ΔR/R0 at each temperature during heating (red, solid line) and cooling (blue, dashed line) (mean ± SD, n = 4); the 45 °C step is shared by both sequences.
In sterile model bile (Section 2.5), the mean ΔR/R0 of the four sensors remained within 0.15% of its initial value over 14 h and did not approach the plateau (Figure 4d), whereas the inoculated run reached the plateau after 840 min. Because the sterile run was performed after the inoculated runs, it indicates the baseline behavior of the system rather than a concurrent control.
The tubing changed visibly during the inoculated runs, and deposits were present in the post-experiment assembly and on the retrieved stent (Figure 4b,c). The change in output coincided with deposition on and around the sensor. These observations do not establish the degree of luminal obstruction, and the readout curve cannot assign distinct stages of bacterial adhesion and biofilm maturation.
3.4. Temperature Dependence and Baseline Stability
Between 30 and 45 °C, the step means of ΔR/R0 remained within 0.10% of zero, and the slope of the step means against temperature during heating was −0.003% °C−1 (n = 4 sensors; Figure 4d,e). These changes are comparable to or smaller than the standard deviation between the four sensors and correspond to less than 3 Pa with the sensitivity in the linear region. The standard deviation between sensors increased during the run, from 0.06% in the first step to 0.29% in the last step, whereas their mean remained near zero. The step means during cooling differed from those during heating at the same temperature by at most 0.04% (thermal hysteresis), and ΔR/R0 at 30 °C after the cycle differed from that at the start by 0.03%. Over the 14 h period, the mean ΔR/R0 remained within 0.15% of its initial value. For comparison, the pressure of gas in a rigid sealed cavity would increase by approximately 5 kPa over the same temperature range, which would exceed the measurement range of the sensor. The temperature dependence within the physiological range (37–39 °C) was therefore small relative to the pressure range of the sensor.
3.5. Animal Allocation, Biliary Stricture, and PEG Access
RFA and PEG placement were completed in both pigs (Table 2). PEG placement, a procedure routinely performed in clinical practice [33], was technically successful in both animals, and no complication of the procedure itself, such as bleeding, perforation, or infection at the insertion site, was observed. Because the sensor wires were to be externalized through the PEG tube, stent deployment had to wait until the PEG tract had matured, and during this period the RFA-induced stricture remained undrained. Deployment was further delayed because the first attempts failed when the connection wire broke (see below), and the procedure had to be postponed to a later session. In one pig, deployment was attempted without success, and the animal died while deployment was postponed. The cause of death was not determined; we presume that it was cholangitis caused by the prolonged unrelieved biliary stricture and not a complication of PEG placement itself, although other causes cannot be excluded. In the remaining pig, the sensor-integrated stent was deployed and the in vivo sensor evaluation was performed. Deployment attempts had to be repeated in both animals, because the connection wire broke before a readout could be obtained at four different steps: (i) while the sensor-integrated stent was loaded into the delivery catheter; (ii) inside the catheter while the catheter was withdrawn after the stent had been deployed in the bile duct; (iii) while the wires were grasped with endoscopic forceps and drawn out through the stomach and PEG tube; and (iv) while the externalized wires were connected to the monitoring system, which had to be done close to the skin of the animal in the operating room because the external length of the wire was insufficient. In the successful attempt, the resistance of the sensor was first measured externally with a digital multimeter, confirming electrical continuity of the implanted sensor, and a readout with the monitoring system connected was then obtained, after which the animal was followed for 1 week.
Table 1.
Characteristics of the sensor before integration into the stent.
Table 2.
Outcome of each animal.
ERCP-guided RFA was followed by localized narrowing of the distal common bile duct and proximal ductal dilation at the 2-week assessment (Figure 5). Figure 5 shows biliary access before ablation, catheter placement during RFA, and the subsequent ductal change. These findings provided a narrowed segment for testing stent deployment.
Figure 5.
ERCP-guided RFA and subsequent biliary narrowing in the porcine model. (a) Fluoroscopic image of biliary access before ablation. (b) Placement of the RFA catheter in the distal common bile duct. The applied settings were 7–10 W, a target temperature of 80 °C, and a duration of 60–90 s. (c) Follow-up cholangiogram 2 weeks after RFA showing localized narrowing (arrow) and proximal ductal dilation.
As shown in Figure 6, PEG placement established access between the stomach and the external abdominal wall. Fluoroscopic imaging showed the tube arrangement, and the external photograph documented the access site. Subsequently, this route was used to bring the sensor wires to the external monitoring equipment.
Figure 6.
Establishment of PEG access in the porcine model. (a) Fluoroscopic image obtained after PEG placement. (b) External view of the 20 Fr PEG tube at the abdominal wall. The tube provided the route used for subsequent sensor wire externalization.
3.6. Stent Deployment, Sensor Readout, and Connection Integrity
The sensor-integrated stent was deployed across the biliary stricture under fluoroscopic guidance (Figure 7a). The sensor wires were routed through the stomach and externalized through the PEG tube (Figure 7b), and the monitoring system was started after the wires had been connected. Because the program takes the mean resistance during the first 30 s after start-up as R0 (Section 2.2), the zero was set in vivo with the stent deployed, and the readout started at approximately 0 Pa. The readout therefore represents the change from this in vivo baseline; loads already present at start-up, such as those from attachment of the sensor, compression of the stent at the stricture, and wire tension, were included in R0 and are not reflected in the readout. The display showed 120 Pa, which remained stable while the animal was under anesthesia (Figure 7c), confirming that the signal from the implanted sensor could be read out electrically through the external connection. The recording was intended to follow how the readout changed over time, but the wire broke before a later reading could be obtained. Because the monitoring unit had no data-storage function, the value was documented by reading the display rather than from a stored time series, and it was not verified against an independent reference-pressure measurement. The displayed value was obtained with the calibration programmed in the monitoring unit and was not re-expressed with the calibration of Section 2.3. It therefore indicates a change in sensor output after start-up and cannot be interpreted as intrabiliary pressure.
Figure 7.
Stent deployment, external sensor readout, and separate bench tensile testing of the connection wire. (a) Post-deployment fluoroscopic image showing the stent and the indicated sensor location. (b) PEG route used for wire externalization. (c) Pressure monitoring system developed for this study, displaying a readout of 120 Pa after deployment; the readout was zeroed in vivo after connection, no data were stored, and the value was not verified against a reference pressure; the inset highlights the wire connection terminals and monitor display. (d) Load–displacement curve of the connection wire in monotonic tension, measured with a Brookfield CT3 texture analyzer. The maximum tensile load recorded before fracture was 1.878 N.
The displayed readout changed during external duodenal compression. This qualitative response provided a functional check after placement but was not used to quantify the applied pressure or dynamic measurement accuracy.
The pig that underwent stent deployment survived for more than 1 week, with no reported abnormalities in dietary intake, vital signs, or behavior during follow-up. We observed no bleeding, gastrointestinal perforation, or infection related to the procedure at the PEG or wire insertion site during the reported follow-up. Necropsy showed a tissue tract with adhesions along the PEG route. These observations were short-term and procedural.
Before the animal recovered from anesthesia, the monitoring unit was attached to the abdomen, the wires were fixed to avoid tension, and the unit and wires were covered with gauze and secured with extensive taping to prevent the animal from damaging them. When the device was checked under anesthesia at the 1-week follow-up, the monitoring unit remained attached to the abdomen, but the wire had broken inside the PEG tube, near the cap that closed the external end of the tube and held the wire, which ended the readout. Survival and the absence of observed major procedural complications therefore should be considered separately from the duration for which the electrical connection remained functional.
3.7. Tensile Test of the Connection Wire
In monotonic tension, the wire fractured at a maximum load of 1.878 N (Figure 7d).
4. Discussion
In this study, we integrated a flexible pressure-sensitive element, a biliary stent, an external wire route, and a dedicated monitoring unit into one experimental platform and evaluated each link of the chain from transduction to readout. Loading matched to the sensing window defined the sensitivity, hysteresis, linearity, and repeatability of the sensor, and the temperature experiment showed that the readout changed little between 30 and 45 °C. In the flow model, the output changed in the inoculated run with a complete recording, whereas it remained near baseline in a subsequent sterile run. The porcine experiment demonstrated stent delivery across a stricture, PEG-based wire externalization, and post-deployment readout, and tensile overload of the wire inside the PEG tube is a possible cause of the connection failure detected at follow-up.
The contribution of this study is the evaluation of the integrated sensing and access arrangement. An earlier study of a magnetoelastic biliary stent sensor targeted accumulation on the stent [15], whereas the present device uses a contact-resistive element to respond to local mechanical loading. The current results do not establish superiority over other sensing modalities. Rather, they show which parts of the wired pressure-sensing approach can be implemented and which require further validation.
The measurement range must be interpreted against physiological biliary pressures. Under loading through the pad, the response was approximately linear up to Plin (2.5 kPa) and approached a plateau at Pmax (3.0 kPa). Normal biliary pressure in humans is approximately 3.0 ± 2.5 mmHg (≈0.40 ± 0.33 kPa) [34], whereas intrabiliary pressure measured with an external transducer during percutaneous transhepatic biliary drainage in patients with extrahepatic biliary obstruction averaged 18.4 ± 3.9 mmHg (≈2.5 kPa) and ranged from 9 to 28 mmHg (≈1.2–3.7 kPa) [13]. Because these values were obtained with an external transducer in undrained ducts, they may not correspond directly to the load sensed at the inner surface of an indwelling stent. The porcine common bile duct pressure has been reported as 0.81 ± 0.06 kPa [26]. Within its linear region, the current range covers the reported normal range, the porcine baseline, and the mean pressure of the obstructed ducts, but the plateau at 3.0 kPa lies below the upper end of the obstructed range (3.0–3.7 kPa). Because the calibration was obtained by loading through a pad, these comparisons are approximate. A readout that reaches the plateau indicates that the local pressure has reached or exceeded this value, but it cannot quantify the degree of obstruction. Extending the range to at least approximately 4 kPa, with a linear region covering the obstructed range, would require redesign of the sensing element, for example its active area, membrane stiffness, or spacer geometry, and the redesigned element would need to be calibrated against a reference pressure.
The temperature dependence was much smaller than that expected from the gas enclosed in the cavity. The cavity was sealed at its edges, and no ingress of liquid was observed: model bile is an ionic conductor, and liquid filling the 200 μm gap over the 2 mm2 window would have formed a parallel conduction path of the order of 102–103 Ω (assuming a conductivity of the order of 1 S m−1), far below R0, whereas no such decrease occurred in any of the four sensors during the 14 h sterile experiment. For a rigid sealed cavity, the ideal gas law gives an increase in internal pressure of approximately 5 kPa between 30 and 45 °C, which would act to separate the films, yet the step means of ΔR/R0 remained within 0.10% of zero. One possible explanation is the regions in which the films adhered to each other during manual assembly and which produce the finite R0. Once adhered, these contacts resist separation by the internal pressure, whereas external pressure readily extends the contact around them. Because the electrodes are micropatterned meshes, small differences in the extent of these adhered regions produce large differences in resistance, which is consistent with the variation in R0 between devices (coefficient of variation, 27%). Because the reference of the sensor is the gas sealed at fabrication, an increase in atmospheric pressure acts on the sensor as an applied pressure. A vented reference or simultaneous measurement of atmospheric pressure would be required for quantitative in vivo measurement.
In the flow experiment, the sterile run indicates the baseline behavior of the sensor and the loop without bacteria. Over this 14 h run, the mean ΔR/R0 of the four sensors remained within 0.15% of its initial value, whereas the inoculated run reached the plateau after 840 min. Because the sterile run was performed afterwards, with different sensors and at varying temperature, it is not a concurrent control. The cause of the change in output was not established. Flow alone produces a negligible pressure in the tubing at this flow rate (of the order of 0.1 Pa per metre for laminar flow), and the reservoir was open to the atmosphere. In this model, deposits that narrow the lumen increase the flow resistance, and because the peristaltic pump maintains the flow rate, the pressure upstream of the deposits rises; visible deposits were present in the tubing and on the stent after the experiment (Figure 4b,c), although the degree of luminal narrowing was not measured. This is the expected cause of the change in output, although direct loading of the sensor by deposits on its surface could also contribute, and the two were not distinguished because the pressure and flow in the loop were not measured. The interval to the plateau is not a measurement of sensor response time, a confirmed time to complete stent occlusion, or a defined biofilm maturation period. Bacterial participation and polymicrobial biofilms have been demonstrated in retrieved biliary stents [8,9,10], and biofilms comprise surface-associated microbial communities within an extracellular matrix [35]. In vitro occlusion models that combine bacteria with other constituents of sludge have reproduced occlusion in plastic stents within days [36], indicating that the composition of the model fluid determines the time scale and that the present interval should not be extrapolated to patients.
The hydraulic and chemical properties of the model also constrain interpretation. Bile is a complex aqueous secretion whose composition is modified along the biliary system [37], and bile acid profiles differ among mammalian species, with hyocholic acid species predominating in pigs but not in humans [38]. A 5 wt% solution prepared from porcine bile powder is therefore only a model of bile, and we did not establish its equivalence to native human bile. Simulations of bile flow through biliary stents have shown that the pressure change depends on bile viscosity and stent design [39]. Measurements of the composition and viscosity of the model fluid would further support interpretation of the recorded trend.
The stricture induced by RFA provided an accessible preclinical setting for endoscopic deployment. Published studies of RFA in pigs support this approach for creating benign biliary narrowing [31,32]. Such a model does not reproduce malignant tissue invasion or the full course of recurrent stent dysfunction. Consequently, successful deployment and readout in this setting establish procedural feasibility, while in vivo detection of developing occlusion of the stent remains to be demonstrated.
The procedural failures show that the connection wire was the weakest element at every step of the procedure, not only during follow-up. The wire broke while the stent was loaded into the delivery catheter, while the catheter was withdrawn after deployment, while the wires were retrieved endoscopically, and while they were connected to the monitoring system. In most cases, the wire broke while it was being pulled, and each of these steps applies tension or friction to a 0.09 mm wire whose fracture load was 1.878 N in the bench test; this failure mode was the reason for characterizing the wire in tension. Successful clinical translation therefore requires a delivery system in which the wire is protected and not loaded during catheter loading and withdrawal, for example by a dedicated lumen or sheath; a wire retrieval method that limits the force applied with forceps; sufficient external wire length with a pre-attached connector so that connection does not require handling at the skin; and a stronger or strain-relieved conductor. These requirements were derived from the observed failures and were not tested in this study. The failures also had a consequence for the animals: they delayed stent placement, which prolonged the period during which the induced stricture remained undrained, and one animal died during this period, presumably of cholangitis. Future protocols should therefore not leave an induced stricture undrained while the monitoring connection is being established, for example by deploying the sensor-integrated stent in the same session in which the stricture is confirmed, or by placing a conventional stent for drainage if deployment is delayed, with predefined humane endpoints.
PEG access enabled the implanted sensor to be connected to external equipment, allowing prototype readout without a fully implanted telemetry system. The clinical literature documents both early and delayed PEG complications and emphasizes appropriate fixation and ongoing tube care [40,41]. These findings concern enteral access and do not establish the safety of routing sensor wires through a PEG tube. Our short-term observations support procedural feasibility, whereas the suitability of this route for monitoring depends on sustained connection integrity as well as the condition of the access site.
Tensile overload is a possible cause of the wire breakage. The wire broke inside the PEG tube, near the cap that closed the external end of the tube. Because the cap held the wire at the abdominal wall while the other end was anchored to the stent in the bile duct, the segment between these two fixed points could be stretched when their separation changed, for example with gastric and duodenal motility, respiration, or movement of the animal. The wire also broke at several steps of the procedure when it was pulled (Section 3.5), which shows that its fracture load of 1.878 N is readily exceeded by the tension applied during handling. The loading history in the animal was not measured, and repeated bending at the gastric and abdominal walls could also have damaged the wire, but was not evaluated; fatigue and fracture of conductor wires under repeated body movement are recognized failure modes of implanted leads [42]. The connection should therefore be designed so that the wire is not the weakest element and is not held taut between fixed points: slack or a coiled section within the PEG tube to accommodate changes in length, strain relief at the cap instead of rigid fixation, and a conductor with a higher fracture load. Strain-relief geometries, multistrand conductors, and polymer encapsulation can further improve the durability of the wire [42,43,44].
Packaging and readout architecture will also influence monitoring duration. Protective barriers have extended the operation of other implantable pressure sensors and flexible electronic systems [22,45], but performance in bile must be evaluated for the actual materials and assembly used here. LCP implant studies provide a basis for investigating this substrate [23], while a recent review of surgical adhesives describes the rapid bonding of cyanoacrylates together with concerns regarding their degradation products and tissue response [46]. These studies do not establish the biocompatibility or durability of the complete assembly used here. Identification of the materials, integrity of the seals, and performance after bile exposure must still be evaluated. Wireless, batteryless sensing offers a potential route to reduce dependence on external wires [19,47]. Such approaches would require assessment of size, power demand, transmission reliability, and encapsulation.
Our study has several limitations. The measurement range of approximately 3 kPa does not cover the upper end of obstructed biliary pressures, and the calibration was obtained by loading the bare sensor through a pad rather than under fluid pressure after mounting in the stent. The temperature was measured in the solution adjacent to the stent rather than at the sensor itself. In vivo, the readout was neither recorded nor compared with a reference pressure; because the wire broke before the next check, the change in output during follow-up could not be determined, and on-board storage of the data in the monitoring unit is required for future studies. The sensor output in the inoculated flow model was obtained from a single complete recording, so its variability between runs is unknown. In addition, although two pigs were initially included, one died after PEG placement, presumably of cholangitis caused by the unrelieved stricture, after unsuccessful attempts at stent deployment, leaving only one animal available for in vivo device evaluation, with a follow-up of 1 week. This small sample size limits the assessment of procedural safety, device performance, and reproducibility, and no independently verified occlusion endpoint was obtained. A subsequent study should combine an extended-range sensor calibrated under fluid pressure against a reference, in vivo data logging with a reference pressure measurement, a delivery system and connection that protect the wire from tension, and independent assessment of stent patency. Occlusion, migration, and other complications should be distinguished using explicit outcome definitions, consistent with the TOKYO reporting criteria [7]. Such a design would permit assessment of whether a pressure trend precedes verified dysfunction and whether it adds information to clinical follow-up. This distinction is necessary because intrabiliary pressure does not predict every consequence of obstruction [13].
5. Conclusions
In this study, we developed an integrated pressure-sensing smart biliary stent by combining a flexible Au mesh contact-resistive sensor with an external pressure monitoring system. Under loading matched to the sensing window, the sensor responded approximately linearly up to 2.5 kPa and approached a plateau at approximately 3.0 kPa, with a pressure hysteresis of 4.1 ± 1.5% FS, and its output changed little when the temperature of the solution at the stent was varied between 30 and 45 °C. In the flow model, the output changed in bile inoculated with bacteria and remained near baseline in a subsequent sterile run. In the porcine experiment, endoscopic deployment, PEG-mediated wire externalization, and post-deployment readout were achieved in one animal.
Tensile overload of the wire held between the cap of the PEG tube and the stent is a possible cause of the wire breakage found at the 1-week follow-up, although the cause was not established. Together with the sensing limit of approximately 3 kPa, which lies below the upper end of the reported range of obstructed biliary pressures, these findings define the next priorities: redesign of the sensing element to extend the range, a connection that accommodates changes in length without loading the wire, calibration under fluid pressure against a reference, and controlled evaluation against independently confirmed stent occlusion. Protection of the sensor and wires during catheter delivery and wire retrieval is a further priority, because these steps caused failures before readout. Thus, this study provides initial evidence of the feasibility of a wired pressure-sensing biliary stent toward clinical application and identifies the technical requirements that must be met before its clinical value can be evaluated.
Author Contributions
Conceptualization, J.L. and Y.W.K.; methodology, J.L., S.Y.H., S.M.H. and Y.W.K.; software, Y.W.K.; validation, S.Y.H., D.C.J. and Y.W.K.; formal analysis, Y.W.K., S.Y.H. and K.Y.; investigation, J.L., S.Y.H., S.M.H., D.C.J. and J.W.L.; resources, J.L. and Y.W.K.; data curation, S.Y.H. and S.M.H.; writing—original draft preparation, J.L. and Y.W.K.; writing—review and editing, J.L., S.Y.H. and Y.W.K.; visualization, Y.W.K. and K.Y.; supervision, Y.W.K.; project administration, J.L.; funding acquisition, J.L. and Y.W.K. All authors have read and agreed to the published version of the manuscript.
Funding
This research was supported by the Bio&Medical Technology Development Program of the National Research Foundation (NRF) funded by the Korean government (MSIT) (No. RS-2023-00223764).
Institutional Review Board Statement
The animal study protocol was approved by the Institutional Animal Care and Use Committee (IACUC) (approval number: P2023-016-A1C0).
Informed Consent Statement
Not applicable.
Data Availability Statement
The raw data supporting the conclusions of this article will be made available by the authors on request.
Acknowledgments
During the preparation of this manuscript, the authors used ChatGPT (GPT-6; OpenAI, San Francisco, CA, USA) to generate the schematic illustration in Figure 3. The authors have reviewed and edited the output and take full responsibility for the content of this publication.
Conflicts of Interest
The authors declare no conflicts of interest.
Abbreviations
The following abbreviations are used in this manuscript:
| ADC | Analog-to-digital converter |
| CFU | Colony-forming units |
| CV | Coefficient of variation |
| DI | Deionized |
| ERCP | Endoscopic retrograde cholangiopancreatography |
| FEA | Finite-element analysis |
| Fr | French (unit of measurement for medical devices) |
| FS | Full scale |
| IACUC | Institutional Animal Care and Use Committee |
| LCP | Liquid crystal polymer |
| MBO | Malignant biliary obstruction |
| PDMS | Polydimethylsiloxane |
| PEG | Percutaneous endoscopic gastrostomy |
| RFA | Radiofrequency ablation |
| SD | Standard deviation |
| SEMS | Self-expandable metal stents |
References
- Banales, J.M.; Marin, J.J.G.; Lamarca, A.; Rodrigues, P.M.; Khan, S.A.; Roberts, L.R.; Cardinale, V.; Carpino, G.; Andersen, J.B.; Braconi, C.; et al. Cholangiocarcinoma 2020: The next horizon in mechanisms and management. Nat. Rev. Gastroenterol. Hepatol. 2020, 17, 557–588. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Elmunzer, B.J.; Maranki, J.L.; Gómez, V.; Tavakkoli, A.; Sauer, B.G.; Limketkai, B.N.; Brennan, E.A.; Attridge, E.M.; Brigham, T.J.; Wang, A.Y. ACG clinical guideline: Diagnosis and management of biliary strictures. Am. J. Gastroenterol. 2023, 118, 405–426. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Barkay, O.; Mosler, P.; Schmitt, C.M.; Lehman, G.A.; Frakes, J.T.; Johanson, J.F.; Qaseem, T.; Howell, D.A.; Sherman, S. Effect of endoscopic stenting of malignant bile duct obstruction on quality of life. J. Clin. Gastroenterol. 2013, 47, 526–531. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Scatimburgo, M.V.C.V.; Ribeiro, I.B.; de Moura, D.T.H.; Sagae, V.M.T.; Hirsch, B.S.; Boghossian, M.B.; McCarty, T.R.; dos Santos, M.E.L.; Franzini, T.A.P.; Bernardo, W.M.; et al. Biliary drainage in inoperable malignant biliary distal obstruction: A systematic review and meta-analysis. World J. Gastrointest. Surg. 2021, 13, 493–506. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Vanella, G.; Coluccio, C.; Cucchetti, A.; Leone, R.; Dell’Anna, G.; Giuffrida, P.; Abbatiello, C.; Binda, C.; Fabbri, C.; Arcidiacono, P.G. Fully covered versus partially covered self-expandable metal stents for palliation of distal malignant biliary obstruction: A systematic review and meta-analysis. Gastrointest. Endosc. 2024, 99, 314–322.e19. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ghazi, R.; AbiMansour, J.P.; Mahmoud, T.; Martin, J.A.; Law, R.J.; Levy, M.J.; Abu Dayyeh, B.K.; Storm, A.C.; Petersen, B.T.; Chandrasekhara, V. Uncovered versus fully covered self-expandable metal stents for the management of distal malignant biliary obstruction. Gastrointest. Endosc. 2023, 98, 577–584.e4. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Isayama, H.; Hamada, T.; Fujisawa, T.; Fukasawa, M.; Hara, K.; Irisawa, A.; Ishii, S.; Ito, K.; Itoi, T.; Kanno, Y.; et al. TOKYO criteria 2024 for the assessment of clinical outcomes of endoscopic biliary drainage. Dig. Endosc. 2024, 36, 1195–1210. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Dinescu, B.; Voiosu, T.; Benguş, A.; Mateescu, R.B.; Voiosu, M.R.; Voiosu, A. The perfect biliary plastic stent: The search goes on. Ann. Gastroenterol. 2023, 36, 490–496. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Flumignan, V.K.; Sircili, M.P.; Germano, L.G.; Souza, A.V.S.; Silva, N.F.; Fukumasu, N.K.; Anjos, R.M.; Otoch, J.P.; Artifon, E.L.A. Characterization of bacterial biofilm composition in occluded plastic biliary stents. Clin. Exp. Gastroenterol. 2025, 18, 179–189. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Cacaci, M.; De Maio, F.; Matteo, M.V.; Posteraro, B.; Di Vito, M.; Menchinelli, G.; Tringali, A.; Monzo, F.R.; Torelli, R.; Costamagna, G.; et al. Pilot study on cultural and metagenomic analysis of bile and biliary stents lead to unveiling the key players in stent occlusion. Sci. Rep. 2024, 14, 3344. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Nakai, Y.; Isayama, H.; Wang, H.P.; Rerknimitr, R.; Khor, C.; Yasuda, I.; Kogure, H.; Moon, J.H.; Lau, J.; Lakhtakia, S.; et al. International consensus statements for endoscopic management of distal biliary stricture. J. Gastroenterol. Hepatol. 2020, 35, 967–979. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Kumar, S.; Vignesh, S.; Boruah, D.K.; Gupta, A.; Yadav, R.R.; Kapoor, V.K.; Behari, A.; Sharma, S. The utility of biliary manometry in assessing early catheter removal after percutaneous balloon dilatation of hepaticojejunostomy strictures. Cureus 2022, 14, e22761. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Saxena, D.; Sasturkar, S.V.; Mukund, A.; Patidar, Y.; Choudhury, A.K.; Kilambi, R.; Kale, P. Intrabiliary pressure in the pathophysiology of extra hepatic biliary obstruction. Hepatol. Forum 2024, 5, 198–203. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Yi, Y.; Wang, B.; Li, C. Sensors-based monitoring and treatment approaches for in-stent restenosis. J. Biomed. Mater. Res. Part B Appl. Biomater. 2023, 111, 490–498. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Nambisan, R.M.; Green, S.R.; Kwon, R.S.; Elta, G.H.; Gianchandani, Y.B. A microsystem for in vivo wireless monitoring of plastic biliary stents using magnetoelastic sensors. Microsyst. Nanoeng. 2024, 10, 159. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Li, Y.; Yuan, K.; Deng, C.; Tang, H.; Wang, J.; Dai, X.; Zhang, B.; Sun, Z.; Ren, G.; Zhang, H.; et al. Biliary stents for active materials and surface modification: Recent advances and future perspectives. Bioact. Mater. 2024, 42, 587–612. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Hu, J.; Dun, G.; Geng, X.; Chen, J.; Wu, X.; Ren, T.-L. Recent progress in flexible micro-pressure sensors for wearable health monitoring. Nanoscale Adv. 2023, 5, 3131–3145. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Mishra, R.B.; El-Atab, N.; Hussain, A.M.; Hussain, M.M. Recent progress on flexible capacitive pressure sensors: From design and materials to applications. Adv. Mater. Technol. 2021, 6, 2001023. [Google Scholar] [CrossRef] [Scilit]
- Herbert, R.; Lim, H.-R.; Rigo, B.; Yeo, W.-H. Fully implantable wireless batteryless vascular electronics with printed soft sensors for multiplex sensing of hemodynamics. Sci. Adv. 2022, 8, eabm1175. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Li, W.; Liu, A.; Wang, Y.; Qu, K.; Wen, H.; Zhao, J.; Shi, Y.; Wang, H.; Ye, M.; Guo, W. Implantable and degradable wireless passive protein-based tactile sensor for intracranial dynamic pressure detection. Electronics 2023, 12, 2466. [Google Scholar] [CrossRef] [Scilit]
- Tang, H.; Yang, Y.; Liu, Z.; Li, W.; Zhang, Y.; Huang, Y.; Kang, T.; Yu, Y.; Li, N.; Tian, Y.; et al. Injectable ultrasonic sensor for wireless monitoring of intracranial signals. Nature 2024, 630, 84–90. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Nie, Z.; Kwak, J.W.; Han, M.; Rogers, J.A. Mechanically active materials and devices for bio-interfaced pressure sensors—A review. Adv. Mater. 2024, 36, 2205609. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ahn, S.-H.; Koh, C.S.; Park, M.; Jun, S.B.; Chang, J.W.; Kim, S.J.; Jung, H.H.; Jeong, J. Liquid crystal polymer-based miniaturized fully implantable deep brain stimulator. Polymers 2023, 15, 4439. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Persson, B.N.J. On the electric contact resistance. Tribol. Lett. 2022, 70, 88. [Google Scholar] [CrossRef] [Scilit]
- Vanaei, S.; Hashemi, M.; Solouk, A.; Asghari Ilani, M.; Amili, O.; Hefzy, M.S.; Tang, Y.; Elahinia, M. Manufacturing, processing, and characterization of self-expanding metallic stents: A comprehensive review. Bioengineering 2024, 11, 983. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Duch, B.U.; Andersen, H.; Gregersen, H. Mechanical properties of the porcine bile duct wall. Biomed. Eng. Online 2004, 3, 23. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Yamabe, A.; Irisawa, A.; Kunogi, Y.; Kashima, K.; Nagashima, K.; Minaguchi, T.; Yamamiya, A.; Izawa, N.; Takimoto, Y.; Hoshi, K.; et al. Development of biliary stent applying the antibacterial activity of silver: A literature review. Bio-Med. Mater. Eng. 2021, 32, 63–71. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Blanco-Míguez, A.; Carloni, S.; Cardenas, C.; Conca Dioguardi, C.; Lambroia, L.; Capretti, G.; Nappo, G.; Fugazza, A.; Capogreco, A.; Armanini, F.; et al. Microbial composition associated with biliary stents in patients undergoing pancreatic resection for cancer. npj Biofilms Microbiomes 2024, 10, 35. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lunney, J.K.; Van Goor, A.; Walker, K.E.; Hailstock, T.; Franklin, J.; Dai, C. Importance of the pig as a human biomedical model. Sci. Transl. Med. 2021, 13, eabd5758. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Percie du Sert, N.; Hurst, V.; Ahluwalia, A.; Alam, S.; Avey, M.T.; Baker, M.; Browne, W.J.; Clark, A.; Cuthill, I.C.; Dirnagl, U.; et al. The ARRIVE guidelines 2.0: Updated guidelines for reporting animal research. PLoS Biol. 2020, 18, e3000410. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Kwon, C.-I.; Jang, S.I.; Jeong, S.; Kim, G.; Lee, T.H.; Cho, J.H.; Nahm, J.H.; Sung, M.J.; Ko, K.H. Optimal reproduction of a porcine benign biliary stricture model using endobiliary radiofrequency ablation. Sci. Rep. 2022, 12, 12046. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Park, J.K.; Yang, J.-I.; Park, J.K.; Lee, K.H.; Lee, J.K.; Lee, K.T. Feasibility of in vivo swine models using guide wire-assisted intraductal radiofrequency ablation for benign biliary stricture. Sci. Rep. 2023, 13, 7185. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Gkolfakis, P.; Arvanitakis, M.; Despott, E.J.; Ballarin, A.; Beyna, T.; Boeykens, K.; Elbe, P.; Gisbertz, I.; Hoyois, A.; Mosteanu, O.; et al. Endoscopic management of enteral tubes in adult patients—Part 2: Peri- and post-procedural management. European Society of Gastrointestinal Endoscopy (ESGE) Guideline. Endoscopy 2021, 53, 178–195. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Carr-Locke, D.L.; Gregg, J.A. Endoscopic manometry of pancreatic and biliary sphincter zones in man: Basal results in healthy volunteers. Dig. Dis. Sci. 1981, 26, 7–15. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Gruszecka, J.; Filip, R. Bacterial biofilms—A threat to biliary stents, understanding their formation, clinical consequences and management. Medicina 2025, 61, 512. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Surwase, S.; Balakrishnan, H.; Acharya, S.K.; Makharia, G.K.; Kumaraswamy, G.; Prasad, B.L.V. Accelerated in vitro model for occlusion of biliary stents: Investigating the role played by dietary fibre. BMJ Innov. 2018, 4, 39–45. [Google Scholar] [CrossRef] [Scilit]
- Cai, X.; Tacke, F.; Guillot, A.; Liu, H. Cholangiokines: Undervalued modulators in the hepatic microenvironment. Front. Immunol. 2023, 14, 1192840. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Sun, J.; Li, M.Y.; Zhou, H.; Chong, J.; Zhang, J.W.; Yu, B.; Chen, D.W.; Ge, L.P. Importance of gut microbiota for bile acid composition and concentration in pigs. Front. Anim. Sci. 2022, 3, 951840. [Google Scholar] [CrossRef] [Scilit]
- Cordista, V.; Patel, S.; Lawson, R.; Lee, G.; Verheyen, M.; Westbrook, A.; Shelton, N.; Sapkota, P.; Zabala Valencia, I.; Gaddam, C.; et al. Towards a customizable, SLA 3D-printed biliary stent: Optimizing a commercially available resin and predicting stent behavior with accurate in silico testing. Polymers 2024, 16, 1978. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Tae, C.H.; Lee, J.Y.; Joo, M.K.; Park, C.H.; Gong, E.J.; Shin, C.M.; Lim, H.; Choi, H.S.; Choi, M.; Kim, S.H.; et al. Clinical practice guidelines for percutaneous endoscopic gastrostomy. Clin. Endosc. 2023, 56, 391–408. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Boeykens, K.; Duysburgh, I. Prevention and management of major complications in percutaneous endoscopic gastrostomy. BMJ Open Gastroenterol. 2021, 8, e000628. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Whitestone, J.; Salih, A.; Goswami, T. Investigation of a deep brain stimulator (DBS) system. Bioengineering 2023, 10, 1160. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Li, Y.; Veronica, A.; Ma, J.; Nyein, H.Y.Y. Materials, structure, and interface of stretchable interconnects for wearable bioelectronics. Adv. Mater. 2025, 37, 2408456. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lam, D.V.; Yang, K.; Liu, D.X.; Lauricella, A.; Gao, Y.; Fielding, E.S.; Golobish, K.; Nieuwoudt, S.; Weber, D.J.; Fisher, L.E.; et al. Evaluation of gold helical microwire structure electrode for long-term rodent nerve stimulation. J. Neural Eng. 2025, 22, 036042. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Mariello, M.; Kim, K.; Wu, K.; Lacour, S.P.; Leterrier, Y. Recent advances in encapsulation of flexible bioelectronic implants: Materials, technologies, and characterization methods. Adv. Mater. 2022, 34, 2201129. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Boursier, M.; Bayon, Y.; Negrell, C.; Pinaud, J.; Caillol, S. Biocompatible glues: Recent progress and emerging frontiers in surgical adhesion. Polymers 2025, 17, 1749. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Kim, H.; Rigo, B.; Wong, G.; Lee, Y.J.; Yeo, W.-H. Advances in wireless, batteryless, implantable electronics for real-time, continuous physiological monitoring. Nano-Micro Lett. 2024, 16, 52. [Google Scholar] [CrossRef] [Scilit] [PubMed]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.