Topological Optimization of Subperiosteal Implants Under Three Fixation Strategies for Repairing a Patient-Specific Maxilla with Periodontitis: A Finite Element Analysis

,
,
,
and
1
Biomechanics Laboratory, School of Biological Science & Medical Engineering, Southeast University, Nanjing 211189, China
2
Department of Oral Surgery, Shanghai Ninth People’s Hospital, Shanghai Jiao Tong University School of Medicine, China National Center for Stomatology, National Clinical Research Center for Oral Diseases, Shanghai Key Laboratory of Stomatology, Shanghai 200011, China
*
Authors to whom correspondence should be addressed.
J. Funct. Biomater.2026, 17(12), 610;https://doi.org/10.3390/jfb17120610 
(registering DOI)
This article belongs to the Special Issue Biomechanical Studies and Biomaterials in Dentistry (3rd Edition)

Abstract

Subperiosteal implants are a popular treatment option for patients with destruction of the soft tissue and alveolar bone, but selecting a suitable fixation strategy and lightweighting the implants to achieve better mechanical stability remain critical clinical considerations. Therefore, this study comprehensively compared the mechanical performance of Ti6Al4V subperiosteal implants under three fixation strategies, i.e., buccal–palatal (B-P), buccal (B), and palatal (P) fixation, and further applied topological optimization (TO) to reduce the implant volume while maintaining its mechanical strength. A maxillary model was first reconstructed from CT images of a patient with maxillary periodontitis. Three cases were then considered, each including three original (or pre-TO) Ti6Al4V subperiosteal implants designed for the B-P, B, and P fixation, respectively. Subsequently, TO was applied to lightweight the implants by constraining the volume to 60% of the original implant volume and the maximum stress to the yield strength of Ti6Al4V (795 MPa). Finally, finite element (FE) analysis was employed to assess the mechanical performance of the implants, screws, and maxilla under chewing load using stress and displacement indices for pre- and post-TO models. The results showed that the pre-TO implants under different fixation strategies exhibited distinct mechanical performances, and each strategy had an advantage over the other two in different indices and cases. However, in all three cases, the post-TO implants under the B-P fixation were found to have the lowest volume-averaged stress, 95th percentile stress, and maximum displacement among the three strategies, and comparable maximum stress to the P fixation. This suggests that the B-P fixation strategy can achieve a better mechanical stability, and that TO can be used to optimize implant volume and thus reduce the risk of implant exposure due to its size and inadequate soft tissue volume.

1. Introduction

The maxilla is the largest bone of the midface and forms the skeletal framework of the face together with the zygomatic bone. It plays an important role in supporting the cranial base and transmitting masticatory forces [1]. Periodontitis is a serious, irreversible gum infection that damages the soft tissue and causes alveolar bone resorption, and thus leads to tooth loss and reduced quality of life. According to a report on periodontitis management [2], the severe form of the disease has a global prevalence of 11%. Therefore, maxillary reconstruction due to periodontitis is challenging and in great demand.
Traditional implant techniques require adequate maxillary bone volume and quality, and this greatly limits their clinical application [3], particularly in patients with severe bone loss. When bone volume is insufficient, additional bone augmentation surgery is required before implantation [4], further increasing the complexity of the maxillary reconstruction. Moreover, elderly patients with systemic comorbidity may be at higher risk of bone resorption due to implant failure, and this poses a considerable technical challenge to conventional treatment [5]. To overcome these limitations, the subperiosteal implant technique has been employed for patients unsuitable for traditional implant or bone augmentation procedures due to insufficient bone volume. This approach circumvents the strict bone volume requirement and has become an appealing maxillary reconstruction strategy with higher safety, reduced invasiveness, and broader indications.
Recent studies indicate that European research groups have played a pivotal role in advancing modern subperiosteal implant technology by developing patient-specific solutions for severe maxillofacial atrophy [6,7]. In this context, customized subperiosteal implants are fabricated by combining CAD/CAM and additive manufacturing, which are widely used in current clinical practice. Moreover, clinical studies have already shown that modern subperiosteal implants significantly improve patient outcomes and shorten treatment duration [8], and most subperiosteal implants are unilaterally fixed on either the buccal (B) or palatal (P) side [9,10]. Although unilateral fixation has demonstrated clinical feasibility, biomechanical studies suggest that it may induce local stress concentration and nonuniform stress distribution across the implant, particularly under complex masticatory loads [11]. Such performance compromises the long-term mechanical stability of the implant and increases its failure risk. Therefore, buccal–palatal (B-P) fixation is proposed to reduce local peak stress and global nonuniformity in stress distribution. To clarify the advantage of B-P fixation, the mechanical performance of the implants under the three fixation strategies, i.e., B, P, and B-P, must be compared.
Theoretically, the B-P fixation can reduce stress, thereby allowing a corresponding reduction in implant material usage. Such a reduction makes the implant lighter and is consistent with clinical needs, as smaller implant volume may facilitate surgical maneuvers and reduce the risks of wound dehiscence or implant exposure [12], which arise from inadequate soft tissue volume [13] and excessive implant size [14]. Therefore, optimizing the mechanical performance of subperiosteal implants with smaller material usage (or implant size) is also a key consideration. In this regard, finite element (FE) analysis, as a well-developed numerical method, can simulate the mechanical response of complex anatomical structures under physiological loads [15] and has been widely used in oral or maxillofacial biomechanics [16]. Based on the FE platform, topological optimization (TO) is an effective approach for redistributing material [17] under given constraints to meet specific objectives, typically by identifying low-stress regions and removing redundant material to achieve a lightweight design [18]. Accordingly, we employed TO to achieve a lightweight design of the subperiosteal implant.
This study has two main novelties or objectives: (1) comparing the mechanical performance of implants under three fixation strategies and (2) reducing implant material usage through TO. To achieve these objectives, a maxillary model was first reconstructed from CT image sequences of a specific patient with severe periodontitis that had caused tooth loosening. To generalize the conclusions, three cases were considered, and for each case, three subperiosteal implants for the B-P, B, and P fixation were designed, respectively. Subsequently, FE analysis was performed to evaluate their biomechanical performance, followed by a TO procedure to reduce the implant material volume.

2. Materials and Methods

2.1. Reconstruction of Maxillary Model

In this study, a 48-year-old female patient with maxillary periodontitis was included and underwent CT scanning at the Department of Oncology, Shanghai Ninth People’s Hospital, Shanghai Jiao Tong University School of Medicine. To evaluate the biomechanical performance of the implants under the three fixation strategies, a maxillary model was first constructed from CT image sequences (slice thickness: 0.625 mm) using Mimics 20.0 (Materialise Inc., Leuven, Belgium). The reconstructed model was then refined and optimized in Geomagic Studio 2013 (Geomagic Inc., Morrisville, NC, USA). Because cortical bone thickness varied spatially within the maxilla, accurate three-dimensional reconstruction of the cortical and cancellous regions was not performed. Instead, the outer surface of the maxilla was uniformly offset inward by an average thickness of 2.0 mm [19,20] to define the cortical bone region (gray in Figure 1), with the enclosed core region designated as cancellous bone (red in Figure 1). This study was approved by the hospital’s ethics committee, and written informed consent was obtained.
Figure 1. Reconstruction of a maxilla. (A) Coronal, (B) transverse, (C) sagittal views of the patient, and (D) geometric model. The cortical and cancellous regions were indicated in gray and red, respectively.

2.2. Maxillary Subperiosteal Implants and Three Fixation Strategies

To match the maxillary contour, the buccal, labial, and palatal bone surfaces were extracted. Three cases were studied. For each case, three implants for the buccal–palatal (B-P), buccal (B) and palatal (P) fixation were originally designed under the guidance of clinicians and the implant manufacturer (Gaofeng Medical Equipment (Wuxi) Co., Ltd., Wuxi, China). Accordingly, nine implants were modeled, each containing four abutments based on anatomical features and reconstruction requirements. Then, the screw-hole positions were determined, and the implants and screws were modeled as Ti6Al4V. The positions and numbers (SN) of the screws with a diameter of 3.3 mm were shown in Figure 2.
Figure 2. Original subperiosteal implants for B-P,B and P fixation in three cases. (A) The implant–maxilla system, (B) the implants. SN indicates the screw number, and arrows indicate the bar.

2.3. Static Finite Element Simulation

Masticatory movements are driven by facial muscles, and the resultant load on the maxilla arises from the complex occlusal contact between the maxillary and mandibular dentitions. During the simulation, the occlusal force was not applied obliquely but was simplified as a static upward force perpendicular to the implant abutments (Figure 3). This simplification was adopted for two reasons: (1) its reasonableness or acceptability was confirmed by the clinicians and (2) it has been widely adopted in previous FE analyses [21]. Finally, nine models were constructed (e.g., the B fixation model in Figure 3).
Figure 3. The loads applied to the implant abutments and the boundary conditions set for the FE model. The red arrows indicate the applied static loads.
The four abutments were labeled as S1, S2, S3, and S4 (Figure 3). All components, including the maxilla, implants, and screws, were discretized using tetrahedral elements. Static FE analysis was performed using ANSYS 2023R1 (ANSYS Inc., Canonsburg, PA, USA). The applied load was determined based on age-related biting force [22,23]. A total load of 270 N was adopted as a conservative and physiologically relevant estimate for the female subject and was distributed on the four abutments to simulate masticatory loads. The load distribution between anterior (S2 + S3) and posterior (S1 + S4) abutments was defined as 20% (54 N) and 80% (216 N), respectively, consistent with clinical observations [24,25], thereby ensuring physiological reasonableness. Then, the force acting on each abutment was determined by equally dividing the load assigned to the corresponding anterior or posterior abutments. All materials were assumed to be homogeneous, isotropic, and linear-elastic, with their properties listed in Table 1. Notably, the assumption of the cortical and cancellous bone mainly affects different stress distributions or stress-concentration positions within the maxilla. However, unlike the maxilla, the implant may not be strongly influenced, owing to their modulus difference (15 GPa vs. 110 GPa).
Table 1. Material properties and load distribution.
Regarding boundary conditions, the top surface of the maxillary models was constrained in all degrees of freedom, which is a common condition in FE studies. For contact between different components, penalty-based frictional contact with a coefficient of 0.2 was assigned between the implant and maxilla [26]. The interface between cortical and cancellous bone was defined using tie constraint to ensure continuity. Similarly, the contact pairs between the screws and implant (or maxilla) were set as tie constraints [21]. Since the tie constraint reflects a fully bonded condition, it mainly represents osseointegration after surgery. These boundary conditions were also shown in Figure 3.
Prior to FE analysis, a mesh convergence was studied to ensure simulation accuracy. The difference in maximum von Mises stress between successive mesh refinements was maintained below 5%. Based on the convergence study, the mesh sizes were set as follows: 0.4 mm for the screws and implant, and 2.0 mm for the cortical and cancellous bones.

2.4. Implant Topological Optimization (TO)

2.4.1. Density-Based TO

Among the TO approaches, the density-based method, also known as the material distribution method [27], is the most widely used. In this method, the design domain is discretized, and the relative density of each element is defined as the design variable [28]:
0 < ρ e ≤ 1
where ρ e represents the relative density of element e, with ρ e = 1   indicating fully retained material and ρ e → 0 indicating material removal (void).
Herein, to balance implant integrity with lightweight design, the implant element stress was constrained to prevent local material failure when reducing structural volume [29]. Specifically, considering a maximum volume reduction of 28% for the Ti6Al4V subperiosteal implant and a post-TO maximum stress (395.03 MPa) below the yield stress of Ti6Al4V (795 MPa) [18], we further reduced the volume percentage by an extra 12%. Accordingly, a volume constraint of 60% of the original implants volume (Figure 2) was imposed to avoid excessive material removal. The TO algorithm is formulated as
  min ρ V ρ = ∑ e = 1 N ρ e v e
s.t.
K ρ U = F
V ρ V 0 ≥ 0.6
σ e vM ≤ σ min ,   e = 1 , … , N
0 < ρ min ≤ ρ e ≤ 1 ,   e = 1 , … , N
where V ρ is the implant volume, ve is the volume of element e, K ρ is the global stiffness matrix, U is the global displacement vector, F is the applied load vector, V0 is the initial implant volume, σ e vM is the von Mises stress of element e, σ min is the yield stress of Ti6Al4V, N is the number of elements, and ρ min is the lower bound of density introduced to avoid numerical singularity of the stiffness matrix [30]. Subsequently, the resulting structural response was used to update the material distribution during the TO.

2.4.2. SIMP Interpolation Scheme

To drive the solution toward a discrete 0–1 design, the solid isotropic material with penalization (SIMP) approach is employed to avoid meaningless intermediate densities [31]. The material parameter is defined as
  E ρ e = ρ e p E 0
where E0 is the Young’s modulus of Ti6Al4V, and p is the penalization factor. Herein, E0 was set to 110 GPa, and p was chosen as p = 3 [32]. The penalization scheme promoted a distinct and manufacturable topology by suppressing intermediate-density elements and encouraging a clear solid–void distribution across the implant.

2.4.3. Sensitivity Filtering

TO is susceptible to numerical instabilities, including checkerboard patterns and mesh-dependent solutions [33]. To address these instabilities, a sensitivity filter originally proposed by Sigmund was applied [34], and it modifies element sensitivities by averaging them over neighboring elements within a specific filter radius [35]. Here, a filter radius of 1.5 times the average element size was adopted [36]. The filtered sensitivities were subsequently used to update the element density distribution during the iterative process.

2.4.4. Optimized Subperiosteal Implant

Following the TO settings in Section 2.4.1, Section 2.4.2 and Section 2.4.3, the optimization was started and terminated when the difference in the objective function between consecutive iterations fell below 0.1%. Notably, the optimized implants were not applicable owing to their rough surfaces (Figure S1 in the Supplementary Materials). Thus, the post-TO implants were finally obtained by smoothing their rough surfaces. Then, the volumes and reduction percentages of the post-TO implants were listed in Table 2. Apparently, the implant volume was reduced by thinning bars (red arrows in Figure 2 vs. Figure 4). In particular, one screw hole was deleted under the B-P fixation in case 1 due to its lower stress according to the TO algorithm, and the corresponding screw at the hole was also removed.
Table 2. Comparison of pre-TO and post-TO implant volumes in all cases.
Figure 4. The post-TO subperiosteal implants for different fixations in three cases. Arrows indicate the thinning bars with respect to Figure 2.

2.5. Performance Evaluation

To compare the mechanical performance of the pre- and post-TO implants under the three fixation strategies, we first examined the stress and displacement distributions across the implants, screws, and maxilla, which intuitively indicated their highly stressed or deformed regions. Four metrics were then presented: (1) the maximum stress, which assessed whether the yield strength of Ti6Al4V was exceeded; (2) the volume-averaged stress, which represented the global stress level; (3) the element count-based 95th percentile stress, which excluded extreme values and reflected the stress level of the majority of the components; and (4) the maximum displacement, which quantified the deformation. Specifically, the volume-averaged stress was calculated as
  σ ¯ v = ∑ e = 1 N σ e v e ∑ e = 1 N v e
where σ e is the stress of element e.

3. Results

3.1. The Stress Distribution of the Pre- and Post-TO Models

3.1.1. The Stress Distribution of Implants

Figure 5 compares von Mises stress distributions of the pre- and post-TO implants under the three fixation strategies in three cases. As expected, the maximum stress mainly occurs on the abutments or screw holes. Moreover, the bars of all implants are the main elements that bear the occlusal load due to the bending effect. Compared with the pre-TO implants, the highly stressed regions in the post-TO counterparts are enlarged, specifically for the B fixation in all cases. In particular, the maximum von Mises stresses of most implants are increased due to the optimization, except for the implants under the B-P and B fixation in case 1 and the implant under the P fixation in case 2 (squared implants in Figure 5). This is attributed to the implant volume reduction, which decreased the bar thickness and implant stiffness and thus led to higher stress levels.
Figure 5. Comparison of the stress distributions of pre-TO and post-TO implants under the three fixation strategies in three cases. Squared figures represent the decreasing maximum stress of the post-TO implant with respect to their pre-TO counterparts.

3.1.2. The Stress Distribution of Screws

Figure 6 presents a comparison of the von Mises stress distributions of the pre- and post-TO screws under the three fixation strategies in three cases. For all screws, the highly stressed regions are located at the connections between the screw and nut due to stress transfer from the implant to the screw. Different from the implant, compared with the pre-TO screw, the maximum von Mises stress of post-TO screws is decreased, except for the screws under the B-P and B fixation in case 2 and those under the B-P and P fixation in case 3 (squared screws in Figure 6). Moreover, the maximum stress of the B-P and B screws in case 2 were slightly increased compared with those of the B-P and P screws in case 3. The general stress decrease of the screws is due to the softening effect of the thinned post-TO implants, which could not effectively transfer the load from the implants to the screws.
Figure 6. Comparison of the stress distributions of pre-TO and post-TO screws under the three fixation strategies in three cases. Squared figures represent the increasing maximum stress of the post-TO screws with respect to their pre-TO counterparts.

3.1.3. The Stress Distribution of Maxillae

Figure 7 compares the stress distributions of the pre- and post-TO maxilla under the three fixation strategies in three cases. For all maxillae, the maximum stress is located at the screw holes or the implant–bone interface (e.g., the post-TO maxillae under the P fixation in case 1 and case 2). Compared with the pre-TO maxillae, the highly stressed regions in the post-TO counterparts are generally enlarged. The maximum stress increases except those under the B-P and B fixation in case 1 and case 3 (squared maxillae in Figure 7). Moreover, the decrease in the maxillary stress under the B-P and B fixations in case 1 is much more pronounced (e.g., from 106.85 to 69.48 MPa under B-P fixation) than that in case 3 (e.g., from 78.72 to 75.65 MPa under B-P fixation).
Figure 7. Comparison of the stress distributions of pre-TO and post-TO maxillae under the three fixation strategies in three cases. Squared figures represent the decreasing maximum stress of the post-TO maxillae with respect to their pre-TO counterparts.

3.2. The Displacement Distribution of the Pre- and Post-TO Models

3.2.1. The Displacement Distribution of Implants

Figure 8 compares the displacement distributions of the pre- and post-TO implants under the three fixation strategies in three cases. As expected, the maximum displacements of all implants almost always occur at the abutment on which the load was applied. Most maximum displacements of implants increase after TO except that under the B fixation in case 3 (squared implants in Figure 8). As stated for the stress distributions in Section 3.1.1 and Section 3.1.2, the optimization reduced the implant bar thickness and implant stiffness, which softened the implants and thus improved their deformation capacities.
Figure 8. Comparison of the displacement distributions of pre-TO and post-TO implants under the three fixation strategies in three cases. Squared figures represent the decreasing maximum displacement of the post-TO implants with respect to their pre-TO counterparts.

3.2.2. The Displacement Distribution of Screws

Figure 9 compares the displacement distributions of the pre- and post-TO screws under the three fixation strategies in three cases. Like the implants, most maximum displacements of the post-TO screws increase compared with those of the pre-TO screws, except for the screws under the B fixation in cases 1 and 2 (squared screws in Figure 9).
Figure 9. Comparison of the displacement distributions of pre-TO and post-TO screws under the three fixation strategies in three cases. Squared figures represent the decreasing maximum displacement of the post-TO screws with respect to their pre-TO counterparts.

3.2.3. The Displacement Distribution of Maxillae

Figure 10 compares the displacement distributions of the pre- and post-TO maxillae under the three fixation strategies in three cases. Like the implants and screws, most maximum displacements of maxillae increase after TO, except for the maxilla under the B fixation in cases 1 and 2 (squared maxillae in Figure 10), which is consistent with that of the screws (Figure 9) due to the tie contact between the screw and maxilla.
Figure 10. Comparison of the displacement distributions of pre-TO and post-TO maxillae under the three fixation strategies in three cases. Squared figures represent the decreasing maximum displacement of the post-TO maxilla with respect to their pre-TO counterparts.

3.3. The Maximum Stress of the Pre- and Post-TO Models

To clearly compare the pre- and post-TO maximum von Mises stresses of the implants, screws and maxilla, their values are summarized in Table S1 of the Supplementary Materials and plotted in Figure 11. Generally, the maximum stresses of each component under the three fixation strategies are comparable in the post-TO models, except for the maxilla stress under the P fixation in case 1 and under the B-P and P fixation in case 2 (stars in Figure 11B,D), which are much higher than their counterparts in the other fixations. In case 1, the stresses of the three post-TO components under the B-P and B fixation are comparable (Figure 11B). The optimization is most effective for the B fixation, where the maximum stresses are substantially reduced by 36%, 56% and 84% for the implant, maxilla and screw, respectively; likewise, the B-P fixation also exhibits a stress decrease, and the maximum stresses are reduced by 9%, 35% and 35% for the implant, maxilla and screw, respectively (Table S2). The slight decrease under the B-P fixation may be attributed to the smaller volume reduction compared to the B fixation (30% vs. 35% in Table 2). In case 2, the maximum stresses of the post-TO implant and screw are still minimal among the three fixations (Figure 11D), even though their stresses are increased by 6% and 10% with respect to the pre-TO counterparts (Table S2). Whereas, in case 3, the maximum stresses of the post-TO implant and screw under the B-P fixation are slightly greater than their counterparts under the B fixation (Figure 11F).
Figure 11. Comparison of the maximum von Mises stress under the three fixation strategies in three cases. (A) Pre-TO and (B) post-TO models in case 1; (C) Pre-TO and (D) post-TO models in case 2; (E) Pre-TO and (F) post-TO models in case 3. Stars represent the high stress in the maxilla.

3.4. The Volume-Averaged Stress of the Pre- and Post-TO Models

To eliminate the potential singularity effect of the maximum stress, the volume-averaged stress, which reflects the global stress level of the implant, are listed in Table S1 of the Supplementary Materials and plotted in Figure 12. Generally, the average stress levels of each component under the three fixations are comparable in the pre-TO (or post-TO) models except the pre-TO screw stress under the B-P fixation in case 1 and under the P fixation in case 2 (stars in Figure 12A,C). In case 1, the stress of the post-TO implant under the B-P fixation is the smallest among the three fixations, and the stresses of the maxilla and screw under the B-P fixation are comparable to the B fixation (Figure 12B). Specifically, the optimization is more effective for the implant and screw under the B-P fixation, where the volume-averaged stresses are reduced by 13% and 62%, respectively, and for the maxilla and screw under the B fixation, where the volume-averaged stresses are reduced by 31% and 10%, respectively (Table S3). Similar to case 1, in case 2, the average stress of the implant under the B-P fixation is the smallest; moreover, the stresses of the maxilla and screw under the B-P fixation are comparable to their counterparts under the P fixation (Figure 12D). However, in case 3, the maximum stress of the post-TO implant under the B-P fixation is slightly greater than its counterpart under the P fixation but much less than that under the B fixation (Figure 12F).
Figure 12. Comparison of the volume-averaged stress under the three fixation strategies in three cases. (A) Pre-TO and (B) post-TO models in case 1; (C) pre-TO and (D) post-TO models in case 2; (E) pre-TO and (F) post-TO models in case 3. Stars represent the high stress in the screws.

3.5. The 95th Percentile Stress of the Pre- and Post-TO Models

To more effectively evaluate the three post-TO implants, we also compared their 95th percentile stresses, which are listed in Table S1 of the Supplementary Materials. According to the stress statistics presented in Figures S2–S19 of the Supplementary Materials, they are shown in Figure 13. Similar to the average stress, the 95th percentile stresses of each component under the three fixations are comparable in the pre-TO (or post-TO) models, except for the pre-TO screw stress under the B-P fixation in case 1 and under the P fixation in case 2 (stars in Figure 13A,C). In case 1, the stress of the post-TO implant under the B-P fixation is the smallest among the three fixations, and the stresses of the maxilla and screw under the B-P fixation are comparable to the P fixation (Figure 13B). The optimization is more effective for the implant and screw under the B-P fixation, where the 95th percentile stresses are reduced by 5% and 52%, respectively, and for the maxilla under the B fixation, where the 95th percentile stress is reduced by 16% (Table S4). Similarly, in case 2, the 95th percentile stress of the implant under the B-P fixation is the smallest among the three fixations; moreover, the stresses of the maxilla and screw under the B-P fixation are comparable to their counterparts under the P fixation (Figure 13D). In case 3, the maximum stress of the post-TO implant under the B-P fixation is greater than its counterpart under the P fixation but much less than the B fixation (Figure 13F).
Figure 13. Comparison of the 95th percentile stress under the three fixation strategies in three cases. (A) Pre-TO and (B) post-TO models in case 1; (C) Pre-TO and (D) post-TO models in case 2; (E) Pre-TO and (F) post-TO models in case 3. Stars represent the high stress in the screws.

3.6. The Maximum Displacement of the Pre- and Post-TO Models

To more comprehensively evaluate the three post-TO implants, we compared their maximum displacements, which are also listed in Table S1 of the Supplementary Materials and plotted in Figure 14. Generally, the maximum displacement increases in the post-TO model compared to the pre-TO counterparts. Moreover, the maximum displacement of the maxilla shows extreme values, where the maximum displacement increases by 1363% (star in Figure 14B) and 652% (star in Figure 14D) under the P fixation in case 1 and case 2 (Table S5). In case 1, the maximum displacement of the post-TO implant under the B-P fixation is the smallest among the three fixations, and the stress of the screw under the fixation is comparable to that under the P fixation (Figure 14B). In case 2, the maximum displacement of the implant under the B-P fixation is also the smallest but comparable to that under the P fixation, and both are smaller than that under the B fixation (Figure 14D). In case 3, the displacements of the post-TO implant under the B-P and P fixation are comparable but less than that under the B fixation, and the displacements of the maxilla and screws are comparable among the three fixations (Figure 14F).
Figure 14. Comparison of the maximum displacement under the three fixation strategies in three cases. (A) Pre-TO and (B) post-TO models in case 1; (C)p-TO and (D) post-TO models in case 2; (E) pre-TO and (F) post-TO models in case 3. Stars represent the high maximum displacement in the maxilla.

4. Discussion

This study comprehensively compared the mechanical responses of three customized subperiosteal implants based on two unilateral (B and P) and one bilateral (B-P) fixation strategy employed in clinical practice. Moreover, the density-based TO procedure with a minimum volume constraint was applied to all implants to reduce the their volumes. Four key mechanical indices were used to evaluate the three fixation strategies, including the maximum stress, volume-averaged stress, 95th percentile stress, and maximum displacement of the implant, screws, and maxilla. Although the maximum stress reflects the stress concentration, which is related to implant longevity and patient outcome, it is highly sensitive to mesh quality in the linear-elastic FE analysis and may exhibit stress singularity. Consequently, the volume-averaged stress and the 95th percentile stress were here introduced: the former characterized the global stress level of the implant, while the latter excluded extreme values (e.g., stress singularities) beyond the 95th percentile.
Overall, among the pre-TO implants under the three fixation strategies in three cases, each showed its own advantages for different indices. For instance, the B-P fixation yielded the lowest maximum stresses of implant in case 3 (Figure 11C), while the P fixation produced the smallest 95th percentile stress of implant in case 1 (Figure 13A). Compared to the pre-TO maximum stress, volume-averaged stress, 95th percentile stress, and maximum displacement of the implant, screws, and maxilla under the three fixation strategies in three cases, the post-TO counterparts either increased or decreased (Tables S2–S5). However, the post-TO performance of the implant under the B-P fixation was better than (Figure 11D, Figure 12B,D, Figure 13B,D and Figure 14B,F) or comparable (Figure 11B,F, Figure 12F, Figure 13F and Figure 14D) to the other two fixations. This indicated that the bilateral B-P fixation strategy by topology optimization could provide the best mechanical stability in most cases. This is consistent with a systematic review [37], which reported that dual-fixation implants offer better biomechanical stability than single-fixation designs.
Regarding the maximum stress, it was well below the yield stress of Ti6Al4V (795 MPa) regardless of the fixation strategies or TO procedure. Specifically, the highest pre-TO and post-TO maximum stresses of implants both occurred under the P fixation and were 191.62 MPa in case 2 (Figure 11C) and 214.08 MPa in case 3 (Figure 11F), and these values were comparable to the reported maximum stress 178.9 MPa of an implant under the B fixation [10] and stress ranges of 131–206 MPa for monoblock implants and 124–178 MPa for dual implants [38]. Moreover, a validation study of a novel subperiosteal implant design reported that Ti6Al4V implants did not fail under a static load of 450 N or cyclic loading of 150 N for five million cycles, with an infinite life limit stress of 250 MPa [12]. In this regard, although the highest maximum stress of 230.72 MPa (Figure 11A) in the present study occurred in the screws under the B fixation with 270 N load (less than the static load of 450 N [12]), this value remained below the reported infinite life limit stress of 250 MPa. This apparently indicates that the implant would not undergo plastic deformation or fatigue fracture. However, under the P fixation in case 1 and case 2, the post-TO maxillary stresses were greatly increased beyond 200 MPa (stars in Figure 11B,D), which represented high stress concentration and probably caused damage to the maxilla, and thus should be avoided. Moreover, TO effectively reduced the local stress under the three fixation strategies in case 1, particularly the screw stress under the P-fixation, which was reduced by 84% (Table S2). These analyses showed that the B-P and B fixation strategies are more suitable from the stand of the low stress concentration (Figure 11B,D,F).
Unlike the maximum stress, the volume-averaged stress represents the global stress level and is rarely analyzed in literature. Herein, although the stress levels of the implants in pre-TO (or post-TO) models were comparable, the stresses of the post-TO implants under the B-P and B fixation were only reduced relative to their pre-TO counterparts in case 1 (Figure 12A vs. Figure 12B), in contrast to the other fixations in the three cases. The screw stress under the B-P fixation was greatly decreased by 62% in case 1 (Table S2), indicating that the post-TO implant influenced the stress redistribution or transfer between the implant, screw and maxilla. Moreover, the B-P fixation yielded the lowest average stresses among the three fixations in case 1 and case 2 (Figure 12B,D), and slightly higher stress to the P fixation in case 3 (Figure 12F), and this proved the advantage of the B-P fixation.
The 95th percentile stress, which excludes the singular stress associated with the maximum stress in linear-elastic FE analysis. The comparison among the three fixations in the pre- or post-TO models follows a similar trend to that of the volume-averaged stress. Specifically, compared with the B and P fixations in case 1 (Figure 13A vs. Figure 13B), only the post-TO implant under the B-P fixation exhibited a reduction relative to its pre-TO counterpart, and the corresponding screw stress was greatly reduced owing to change of implant structure. Moreover, the B-P fixation yielded the lowest average stresses among the three fixations in case 1 and case 2 (Figure 13B,D), and slightly higher stress than the P fixation in case 3 (Figure 13F), and this again confirms the advantage of the B-P fixation.
Regarding the maximum displacement, the implant stiffness was reduced owing to TO, and thus the maximum displacements were increased. Although the current post-TO displacement range in the order of 10−3–10−2 mm is much greater than the reported range in the order of 10−5–10−3 mm [39], it was still lower than the maximum displacement constraint of 0.1 mm in a clinical study [10], which clearly indicated that the displacement was allowable in clinical practice. Moreover, a greater displacement of 0.58 mm was acceptable in another work [40]. For the maxilla, the excessive post-TO displacement is around 0.1 mm (Figure 14B,D,F), but the displacement is completely acceptable, as the fracture initiation critical displacement occurred in the mandibular bone was tested as 3.1 ± 0.6 mm [41]. Therefore, if the implant maximum displacement was considered without the maxilla, the post-TO maximum displacement of the implant under the B-P fixation was always the smallest in three cases. This again demonstrated that the B-P fixation was the optimal strategy to enhance the mechanical stability of the subperiosteal implant.
In summary, the post-TO implant under the B-P fixation strategy performed best due to its lowest volume-averaged stress, 95th percentile stress, and maximum displacement among the three fixations, and comparable maximum stress to that under the B fixation across the three cases. Indeed, when extrapolating the finding to patients with other oral diseases or anatomical conditions, we must be careful about the maxillary morphology and bone quality. For instance, for patients with severe maxillary atrophy, the bone quality is insufficient for the B-P fixation or even for the unilateral B or P fixation, which need sufficient bone volume to fix the implant. In these scenarios, maxillary morphology and bone quality may determine the selection of the fixation strategy and implant structure (Figure 15), and thus influence the mechanical stability. Indeed, besides the current fixation strategy, the mechanical stability depends on other factors as well. For example, the torque of tightening screws greater than 25 Nm was proven to be more beneficial for early (<12 weeks) mechanical stability in a clinical study [42]. This suggests that, not limited to the implant design, integrating multiple engineering and clinical factors will be more reliable to improve the implant mechanical stability. In addition, the TO approach is capable of generating patient-specific material distribution in accordance with prescribed anatomical, loading, and mechanical constraints, thereby offering a useful framework for developing personalized subperiosteal implants while satisfying predefined mechanical constraints (e.g., stiffness).
Figure 15. Selection of clinical fixation strategies for individual patients. (A) B-P fixation, (B) B fixation, (C) P fixation. The yellow arrows indicate the screws located on the B or P sides.
Several limitations should be acknowledged. First, the linear-elastic FE analysis was conducted under static loading condition, and the nonlinearity of the dynamic cyclic loads during mastication should be carefully incorporated. Second, the chewing activity is often complex, and the loads acting on the abutment are oblique [39]; the oblique loads should also be included to simulate a more realistic environment. Third, the image heterogeneity of bone was not treated, and a gray-mapped Young’s modulus should be incorporated into the simulation to have more accurate results [21]. Fourth, the patient number is limited, and the sample size should be increased. In addition, animal experiments or clinical follow-up will be useful to validate the efficacy of the optimized designs.

5. Conclusions

This study numerically compared the biomechanical performances of patient-specific subperiosteal implants under three fixation strategies across three cases, and re-designed them through a density-based TO method. The key findings are that different fixation strategies affected the stress distribution in the implants, screws, and maxilla. The pre-TO implants exhibited distinct mechanical responses, and each fixation strategy had its own advantages in different stress indices and cases. However, the post-TO implant under the B-P fixation generally showed the most favorable performance owing to its lower mechanical index, indicating better mechanical stability. Overall, both the fixation strategy and the subperiosteal implant design for the maxillary reconstruction should be carefully selected, and TO may be further employed to optimize the implant by reducing the material volume while maintaining acceptable mechanical performance.

Supplementary Materials

The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/jfb17120610/s1, Figure S1. Structures of the optimized subperiosteal implants with rough surfaces. (a) Implants with B-P fixation; (b) implants with B fixation; (c) implants with P fixation. Figure S2. Stress-based element counts under the pre-TO B-P fixation in case 1. Figure S3. Stress-based element counts under the post-TO B-P fixation in case 1. Figure S4. Stress-based element counts under the pre-TO B fixation in case 1. Figure S5. Stress-based element counts under the post-TO B fixation in case 1. Figure S6. Stress-based element counts under the pre-TO P fixation in case 1. Figure S7. Stress-based element counts under the post-TO P fixation in case 1. Figure S8. Stress-based element counts under the pre-TO B-P fixation in case 2. Figure S9. Stress-based element counts under the post-TO B-P fixation in case 2. Figure S10. Stress-based element counts under the pre-TO B fixation in case 2. Figure S11. Stress-based element counts under the post-TO B fixation in case 2. Figure S12. Stress-based element counts under the pre-TO P fixation in case 2. Figure S13. Stress-based element counts under the post-TO P fixation in case 2. Figure S14. Stress-based element counts under the pre-TO B-P fixation in case 3. Figure S15. Stress-based element counts under the post-TO B-P fixation in case 3. Figure S16. Stress-based element counts under the pre-TO B fixation in case 3. Figure S17. Stress-based element counts under the post-TO B fixation in case 3. Figure S18. Stress-based element counts under the pre-TO P fixation in case 3. Figure S19. Stress-based element counts under the post-TO P fixation in case 3. Table S1. The maximum stress, volume-averaged stress, 95th percentile stress and maximum displacement between the pre-TO and post-TO models under the three fixation strategies in three cases. Table S2. Variation percentage of the maximum stress between the pre-TO and post-TO models. Table S3. Variation percentage of the volume-averaged stress between the pre-TO and post-TO models. Table S4. Variation percentage of the 95th percentile stress between the pre-TO and post-TO models. Table S5. Variation percentage of the maximum displacement between the pre-TO and post-TO models.

Author Contributions

Conceptualization, Q.C. and D.Z.; methodology, Q.C. and S.Z.; software, S.Z.; formal analysis, H.L. and J.G.; investigation, S.Z. and J.G.; resources, H.L.; data curation, S.Z.; writing—original draft preparation, Q.C. and S.Z.; writing—review and editing, Q.C.; supervision, Q.C. and D.Z.; project administration, Q.C. and D.Z.; funding acquisition, Q.C. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by the National Key Research and Development Program of China (No. 2024YFC2418600), and the financial support by the National Natural Science Foundation of China (No. 12672395).

Institutional Review Board Statement

The study was conducted in accordance with the Declaration of Helsinki and approved by the Ethics Committee of Shanghai Ninth People’s Hospital OF Shanghai Jiao Tong University School of Medicine (protocol code: SH9H2024-T509-1, 9 January 2025).

Data Availability Statement

The data supporting this study’s findings are available from the corresponding authors upon reasonable request.

Acknowledgments

The authors thank the School of Civil Engineering of Southeast University for the commercial software ANSYS 2023R1. Huabing Luo from Gaofeng Medical Equipment (Wuxi) Co., Ltd. and Yuanbin Hu from the Department of Orthopaedics, Yangzhou Hospital of TCM were also acknowledged for their useful discussions during the investigation and analysis.

Conflicts of Interest

The authors declare no conflicts of interest.

Abbreviations

The following abbreviations are used in this manuscript:
B-PBuccal–Palatal
BBuccal
PPalatal
TOTopology Optimization

References

  1. Brown, J.S.; Shaw, R.J. Reconstruction of the Maxilla and Midface: Introducing a New Classification. Lancet Oncol. 2010, 11, 1001–1008. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  2. Kwon, T.H.; Lamster, I.B.; Levin, L. Current concepts in the management of periodontitis. Int. Dent. J. 2021, 71, 462–476. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  3. Bassir, S.H.; Kholy, K.E.; Chen, C.Y.; Lee, K.H.; Intini, G. Outcome of Early Dental Implant Placement versus Other Dental Implant Placement Protocols: A Systematic Review and Meta-Analysis. J. Periodontol. 2019, 90, 493–506. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  4. Sakkas, A.; Wilde, F.; Heufelder, M.; Winter, K.; Schramm, A. Autogenous Bone Grafts in Oral Implantology—Is It Still a “Gold Standard”? A Consecutive Review of 279 Patients with 456 Clinical Procedures. Int. J. Implant Dent. 2017, 3, 17. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  5. Chrcanovic, B.R.; Albrektsson, T.; Wennerberg, A. Bone Quality and Quantity and Dental Implant Failure: A Systematic Review and Meta-Analysis. Int. J. Prosthodont. 2017, 30, 219–237. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  6. Anitua, E.; Eguia, A.; Staudigl, C.; Alkhraisat, M.H. Clinical Performance of Additively Manufactured Subperiosteal Implants: A Systematic Review. Int. J. Implant Dent. 2024, 10, 18. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  7. Vaira, L.A.; Biglio, A.; Salzano, G.; Lechien, J.R.; Riu, G.D. Additively Manufactured Custom-Made Subperiosteal Implant Rehabilitation for Severely Atrophic Maxillary Molar Area: A Technical Note. J. Stomatol. Oral Maxillofac. Surg. 2024, 125, 4. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  8. Pellegrino, G.; Karaban, M.; Barausse, C.; Guidice, A.; Antonelli, A.; Pistilli, R.; Felice, P. Indications and Complications of Subperiosteal Implants: Literature Review and Case Series. Dent. J. 2025, 13, 19. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  9. Demir, B.; Caglar, I. Biomechanical evaluation of prosthetic framework materials in subperiosteal implants: A finite element analysis. BMC Oral Health 2025, 25, 1929. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  10. Mommaerts, M.Y. Evolutionary steps in the design and biofunctionalization of the additively manufactured sub-periosteal jaw implant ‘AMSJI’for the maxilla. Int. J. Oral. Maxill. Surg. 2019, 48, 108–114. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  11. Acar, G.; Ari, I.; Tosun, E. Biomechanical Evaluation of Implant Options for Unilateral Maxillary Defects: A Finite Element Analysis. BMC Oral Health 2024, 24, 14. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  12. Vanaclocha, V.; Atienza, C.; Vanaclocha, A.; Peñuelas, A.; Gómez-Herrero, J.; Pérez-Carrió, F.; Diego-Leyda, J.A.; Sáiz-Sapena, N.; Vanaclocha, L. New Subperiosteal Dental Implant Design with Finite Element Analysis and Mechanical Validation: A Design Validation Study. Materials 2025, 18, 20. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  13. Thoma, D.S.; Gil, A.; Hämmerle, C.H.; Jung, R.E. Management and prevention of soft tissue complications in implant dentistry. Periodontol. 2000 2022, 88, 116–129. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  14. Kim, Y.K.; Yun, P.Y. Risk factors for wound dehiscence after guided bone regeneration in dental implant surgery. Maxillofac. Plast. Reconstr. Surg. 2014, 36, 116. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  15. Welch-Phillips, A.; Gibbons, D.; Ahern, D.P.; Butler, J.S. What Is Finite Element Analysis? Clin. Spine Surg. 2020, 33, 323–324. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  16. Dogru, S.C.; Cansiz, E.; Arslan, Y.Z. A Review of Finite Element Applications in Oral and Maxillofacial Biomechanics. J. Mech. Med. Biol. 2018, 18, 26. [Google Scholar] [CrossRef] [Scilit]
  17. Leung, M.C.A.; Smeets, L.C.; Laeven, P.F.; Tuijthof, P.G.J.; Kessler, P.P.A. Topology Optimization of a Mandibular Reconstruction Plate and Biomechanical Validation. Tissue Eng. Part A 2022, 28, S172. [Google Scholar]
  18. Carnicero, A.; Peláez, A.; Restoy-Lozano, A.; Jacquott, I.; Perera, R. Improvement of an Additively Manufactured Subperiosteal Implant Structure Design by Finite Elements Based Topological Optimization. Sci. Rep. 2021, 11, 15390. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  19. Sarrafpour, B.; Swain, M.; Li, Q.; Zoellner, H. Tooth Eruption Results from Bone Remodelling Driven by Bite Forces Sensed by Soft Tissue Dental Follicles: A Finite Element Analysis. PLoS ONE 2013, 8, 18. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  20. Li, Y.; Hu, Y.; Chen, H.; Meng, X.; Chen, D.; Gu, H.; Chen, Q.; Mu, Z.; Li, Z. A novel conceptual design of a biomimetic oral implant and its biomechanical effect on the repairment of a large mandibular defect. Med. Nov. Technol. Devices 2022, 15, 100147. [Google Scholar] [CrossRef] [Scilit]
  21. Wan, B.; Wang, Y.; Dunn, M.; Manzie, T.; Leinkram, D.; Sarkis, L.; Wykes, J.; Aung, Y.M.; Gillett, Z.; Clark, J. Patient-Specific Sensitivity of Osteotomy Interface Modelling in Mandibular Reconstruction. Oral. Oncol. 2026, 181, 108103. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  22. Takaki, P.; Vieira, M.; Bommarito, S. Maximum Bite Force Analysis in Different Age Groups. Int. Arch. Otorhinolaryngol. 2014, 18, 272–276. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  23. Gao, H.; Li, X.; Wang, C.; Ji, P.; Wang, C. Mechanobiologically Optimization of a 3D Titanium-Mesh Implant for Mandibular Large Defect: A Simulated Study. Mater. Sci. Eng. C 2019, 104, 9. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  24. Qadeer, S.; Yang, L.; Sarinnaphakorn, L.; Kerstein, R.B. Comparison of Closure Occlusal Force Parameters in Post-Orthodontic and Non-Orthodontic Subjects Using T-Scan® III DMD Occlusal Analysis. Cranio® 2016, 34, 395–401. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  25. Turkistani, K.A.; Alkayyal, M.A.; Abbassy, M.A.; Al-Dharrab, A.A.; Zahran, M.H.; Melis, M.; Zawawi, K.H. Comparison of Occlusal Bite Force Distribution in Subjects with Different Occlusal Characteristics. Cranio® 2023, 41, 204–211. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  26. Maintz, M.; Msallem, B.; de Wild, M.; Seiler, D.; Herrmann, S.; Feiler, S.; Sharma, N.; Dalcanale, F.; Cattin, P.; Thieringer, F.M. Parameter Optimization in a Finite Element Mandibular Fracture Fixation Model Using the Design of Experiments Approach. J. Mech. Behav. Biomed. Mater. 2023, 144, 11. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  27. Rozvany, G.I.N. Aims, Scope, Methods, History and Unified Terminology of Computer-Aided Topology Optimization in Structural Mechanics. Struct. Multidiscip. Optim. 2001, 21, 90–108. [Google Scholar] [CrossRef] [Scilit]
  28. Helvaci, D.O.; Yaman, U. Tiling of Microstructures According to the Density Values of SIMP Topology Optimization. In Proceedings of the 30th International Conference on Flexible Automation and Intelligent Manufacturing (FAIM), Athens, Greece, 15–17 June 2021; Elsevier Science BV: Amsterdam, The Netherlands, 2021. [Google Scholar]
  29. Yang, D.; Liu, H.; Zhang, W.; Li, S. Stress-Constrained Topology Optimization Based on Maximum Stress Measures. Comput. Struct. 2018, 198, 23–39. [Google Scholar] [CrossRef] [Scilit]
  30. Bruns, T.E. Zero Density Lower Bounds in Topology Optimization. Comput. Methods Appl. Mech. Eng. 2006, 196, 566–578. [Google Scholar] [CrossRef] [Scilit]
  31. Sigmund, O.; Maute, K. Topology Optimization Approaches: A Comparative Review. Struct. Multidiscip. Optim. 2013, 48, 1031–1055. [Google Scholar]
  32. Xue, H.J.; Yu, H.; Zhang, X.; Quan, Q. A Novel Method for Structural Lightweight Design with Topology Optimization. Energies 2021, 14, 18. [Google Scholar] [CrossRef] [Scilit]
  33. Sigmund, O.; Petersson, J. Numerical Instabilities in Topology Optimization: A Survey on Procedures Dealing with Checkerboards, Mesh-Dependencies and Local Minima. Struct. Optim. 1998, 16, 68–75. [Google Scholar] [CrossRef] [Scilit]
  34. Sigmund, O. On the Design of Compliant Mechanisms Using Topology Optimization. Mech. Struct. Mach. 1997, 25, 493–524. [Google Scholar] [CrossRef] [Scilit]
  35. Sigmund, O. A 99 Line Topology Optimization Code Written in MATLAB. Struct. Multidiscip. Optim. 2001, 21, 120–127. [Google Scholar] [CrossRef] [Scilit]
  36. Andreassen, E.; Clausen, A.; Schevenels, M.; Lazarov, B.S.; Sigmund, O. Efficient Topology Optimization in MATLAB Using 88 Lines of Code. Struct. Multidiscip. Optim. 2011, 43, 1–16. [Google Scholar] [CrossRef] [Scilit]
  37. Ronsivalle, V.; Giudice, P.L.; Santonocito, S.; Bocchieri, S.; Giudice, R.; Battaglia, S.; Crimi, S.; Bianchi, A.; Cicciù, M. Finite Element Method of Subperiosteal Implants: A Systematic Review on Biomechanical Performance and Stress Distribution. J. Cranio-Maxillofac. Surg. 2025, 53, 1275–1282. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  38. Ayhan, M.; Cankaya, A.B. Custom-Made Subperiosteal Implants: A Finite Element Analysis on Monoblock and Dual Implant Systems in Atrophic Maxilla. Int. J. Med. Sci. 2023, 20, 1755–1762. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  39. Zielinski, R.; Sowinski, J.; Piechaczek, M.; Okulski, J.; Kozakiewicz, M. Finite element analysis of subperiosteal implants in edentulism—On the basis of the MaI Implant® by Integra Implants®. Materials 2023, 16, 7466. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  40. Kasapoglu, M.B.; Az, Z.A.A. Comparison of Load Distribution in Subperiosteal Implants and Mini Plates in Orthognathic Surgery. Int. J. Med. Sci. 2025, 22, 2257. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  41. Zhao, S.G.; Jiang, W.B.; Liang, X.Y.; Zhang, Y.F.; Zhang, J.Z.; Li, R.Q.; Qin, S.; Quan, Y.Y.; Ding, K.; Li, Y.B.; et al. Biomechanical features and fracture characteristics of mandible based on digital image correlation technique. BMC Oral Health 2025, 25, 1987. [Google Scholar]
  42. Vergoullis, I.; Horton, L.; Lampropoulos, P.; Soldatos, N. Clinical Evaluation of Torque Protocols on the Stability of 3D-Printed Anatomical Healing Abutments. Appl. Sci. 2026, 16, 6657. [Google Scholar] [CrossRef] [Scilit]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.