Influence of Corneal Hysteresis on Agreement Between Corneal-Compensated, Non-Contact, and Rebound Tonometry

,
and
1
Ophthalmology Department, George Emil Palade University of Medicine, Pharmacy, Science, and Technology of Targu Mures, 540058 Targu Mures, Romania
2
Ophthalmology Department, Mures County Clinical Hospital, 540072 Targu Mures, Romania
*
Author to whom correspondence should be addressed.
J. Clin. Med.2026, 15(23), 9271;https://doi.org/10.3390/jcm15239271 
(registering DOI)
This article belongs to the Section Ophthalmology

Abstract

Background/Objectives: Corneal hysteresis (CH) may provide information beyond central corneal thickness (CCT) when examining differences between tonometric measurements. This study assessed associations between CH and inter-tonometer differences after adjustment for CCT and age. Methods: This cross-sectional study included 134 eyes from 67 patients attending follow-up one month after cataract surgery in one eye. Both eyes were examined using non-contact tonometry (NCT), iCare Pro, and the Ocular Response Analyzer (ORA). CCT was measured using the Topcon CT-1P. Agreement was assessed using Bland–Altman analysis, including proportional bias and regression-based limits of agreement. Multivariable models and comparisons across three CH groups used patient-clustered robust standard errors. Complete observations were used separately for each analysis. Results: Mean biases for IOPcc–NCT, IOPcc–iCare, and NCT–iCare were 0.15 mmHg (95% LoA, −5.15, 5.44), 1.68 mmHg (−5.41, 8.77), and 1.18 mmHg (−4.42, 6.78), respectively. Significant proportional bias was observed only for IOPcc–NCT. After adjustment for CCT and age, each 1-mmHg decrease in CH was associated with increases of 1.56, 2.23, and 0.39 mmHg in the respective differences (p < 0.001, p < 0.001, and p = 0.042). In the low CH group, IOPcc exceeded NCT and iCare by averages of 2.50 and 5.51 mmHg, respectively. Conclusions: Lower CH was associated with more positive inter-tonometer differences, particularly those involving IOPcc. Wide limits of agreement do not support assuming interchangeability. While mean differences were small in the overall cohort, they became clinically relevant in eyes with low CH. These findings suggest that considering CH in addition to CCT may support a more accurate interpretation of IOP measurements in biomechanically vulnerable eyes.

1. Introduction

Accurate intraocular pressure (IOP) measurement is central to glaucoma diagnosis, risk assessment, and therapeutic monitoring. Decisions regarding treatment initiation, escalation, and evaluation of disease control are frequently based on relatively small changes in measured IOP. Consequently, even modest systematic differences between tonometric methods may influence clinical interpretation, especially when measurements obtained with different devices are compared over time.
Although Goldmann applanation tonometry remains the clinical reference standard, non-contact tonometry (NCT) and rebound tonometry are widely used in routine practice because of their practicality and suitability for different clinical settings and patient populations [1]. In routine practice, values obtained with these instruments may be interpreted as broadly comparable. However, agreement at the population level does not necessarily imply interchangeability in individual eyes, and clinically meaningful discrepancies may occur particularly in eyes with altered corneal properties [2,3,4,5,6].
Tonometric techniques estimate IOP indirectly by analysing the corneal response to an externally applied force or transient mechanical interaction. The measured value therefore depends not only on the pressure within the eye, but also on how the cornea responds to the specific measurement principle of each device. This response is determined by the interaction among IOP, corneal geometry, intrinsic biomechanical properties, and the magnitude and temporal characteristics of the applied force [3]. Because different tonometers interact with the cornea in different ways, they may not be equally affected by variations in corneal behaviour. Thus, two eyes with the same true IOP may yield different measured values if their corneas differ in thickness, rigidity and viscoelasticity. Similarly, the same eye may produce different IOP estimates when examined with devices based on different measurement principles.
The influence of the cornea on tonometry has been interpreted primarily in relation to central corneal thickness (CCT). Nevertheless, thickness represents only one aspect of corneal behaviour. Corneas with similar CCT may have substantially different deformation responses because of differences in tissue composition, collagen architecture, viscoelasticity and age-related or ocular pathology changes. Despite increasing recognition of these factors, corneal biomechanics remain less integrated into clinical assessment. In particular, limited evidence is available regarding the extent to which biomechanical variability explains disagreement among tonometric methods commonly used.
Corneal hysteresis (CH), measured with the Ocular Response Analyzer (ORA), reflects the viscoelastic capacity of the cornea to absorb and dissipate mechanical energy. The ORA also provides corneal-compensated intraocular pressure (IOPcc), an estimate designed to reduce the influence of corneal biomechanical characteristics on IOP measurement [1,2]. CH has increasingly been recognized as a clinically relevant biomarker, both for interpreting IOP and for assessing susceptibility to glaucomatous damage [2,3,6,7]. However, the extent to which CH accounts for discrepancies between IOPcc and measurements obtained using conventional tonometric methods remains insufficiently understood.
Previous studies comparing ORA-derived measurements, non-contact tonometry, rebound tonometry, and Goldmann applanation tonometry have reported variable levels of agreement, with both overestimation and underestimation depending on the device, population, and ocular characteristics [7,8,9,10,11]. Most investigations have focused on average inter-device agreement across entire study populations. This approach may obscure clinically important differences in subgroups with reduced corneal biomechanical resilience.
The aim of this study was to evaluate the influence of corneal hysteresis on the differences between IOPcc, NCT, and rebound tonometry; to determine the independent contribution of CH after adjustment for CCT and age; and to assess inter-method discrepancies across CH strata, with particular emphasis on eyes within the lowest CH tertile.

2. Materials and Methods

2.1. Study Design and Participants

This cross-sectional study included 134 eyes from 67 patients attending the one-month follow-up visit after cataract surgery in one eye at an ophthalmology clinic in Targu-Mures, Romania. Both eyes were examined. The fellow eye had either not undergone cataract surgery or had undergone surgery previously.
Exclusion criteria included ocular infection, central corneal opacities, keratoconus, and previous corneal or refractive surgery. Demographic data, intraocular pressure (IOP) measurements, corneal biomechanical parameters, and ocular biometric characteristics were collected for all participants.

2.2. Tonometric and Biomechanical Measurements

IOP measurements were obtained using three different tonometric methods: NCT was performed with an air puff tonometer (Topcon CT-1P, Topcon Corporation, Tokyo, Japan); rebound tonometry using iCare Pro (iCare Finland Oy, Helsinki, Finland); and the Ocular Response Analyzer (ORA; Reichert Technologies, Depew, NY, USA). All tonometric measurements were performed by a single examiner with patients seated. Instruments were used in a fixed, non-randomized sequence: iCare, followed by NCT and then ORA, with an interval of 3–5 min between examinations with successive instruments. Results across instruments were reviewed after completion of the measurement sequence. Three consecutive measurements were initially obtained with each instrument. If the inter-reading difference exceeded 2 mmHg, the entire set of three measurements was repeated, and the mean of the final accepted set was used for analysis. In most cases, the initial three measurements were sufficient. For iCare and NCT, only valid readings without instrument-reported errors were accepted. For ORA, a predefined minimum waveform score of 7 was required.
All measurements were obtained at approximately the same time of day to minimize the potential influence of diurnal variation in intraocular pressure.
Central corneal thickness was measured using the pachymetry function of the Topcon CT-1P non-contact tonometer (Topcon Corporation, Tokyo, Japan) during the same examination session as NCT. The average of three final measurements was taken for analysis.

2.3. Statistical Analysis

Continuous variables were expressed as mean ± standard deviation (SD). Ranges were additionally reported for tonometric parameters.

2.3.1. Agreement Analyses

Agreement was assessed using Bland–Altman analysis, with mean bias and descriptive 95% limits of agreement (LoA; mean difference ± 1.96 SD). Proportional bias was assessed by regressing paired differences on paired means, using patient-clustered robust standard errors for inference.
Regression-based 95% LoA were calculated as the predicted difference ± 1.96 residual SD, assuming constant residual dispersion after exploratory assessment. Pointwise 95% confidence intervals were estimated using 5000 patient-cluster bootstrap resamples. Estimates were restricted to the observed paired-mean IOP range. No clinically acceptable margin of disagreement was prespecified.

2.3.2. Multivariable Regression Analysis

Multivariable linear regression was used to assess factors associated with inter-method differences. Separate ordinary least squares models with an intercept were fitted for IOPcc–NCT, IOPcc–iCare, and NCT–iCare, with CH, CCT, and age included simultaneously as predictors. Patient-clustered robust standard errors were used in all models. Regression coefficients (β), 95% confidence intervals (CIs), two-sided p-values, and model R2 values were reported.
To assess the potential influence of recent cataract surgery on inter-method tonometric differences, additional sensitivity analyses were performed by incorporating surgical status (operated versus fellow eye) as a covariate in the multivariable linear regression models, alongside corneal hysteresis (CH), central corneal thickness (CCT), and age. Interaction terms between CH and surgical status were subsequently introduced to evaluate whether the association between CH and inter-method differences varied according to surgical status. Patient-level cluster-robust standard errors were used to account for the correlation between fellow eyes.

2.3.3. Corneal Biomechanical Stratification

To further evaluate the association between corneal biomechanics and inter-tonometer agreement, eyes were stratified into three CH groups: low (8.3–9.7 mmHg), intermediate (9.9–11.5 mmHg), and high (11.6–15.9 mmHg). Mean inter-method differences (IOPcc–NCT, IOPcc–iCare, and NCT–iCare) were compared across these groups using linear regression with CH group entered as a categorical predictor. Patient-clustered robust standard errors were used to account for dependence between fellow eyes. Overall differences across groups were assessed using Wald F tests, and pairwise comparisons were performed with Bonferroni adjustment. Group means and their 95% confidence intervals were estimated from these models.
Statistical significance was defined as a two-sided p-value < 0.05. All analyses used complete observations for the variables required by each analysis. Consequently, the numbers of eyes and patients varied across analyses according to data availability.
Analyses used complete observations for the variables required by each analysis, without imputation. Pairwise agreement analyses required both corresponding tonometric measurements, whereas CH-group analyses additionally required CH and multivariable models additionally required all included predictors. Consequently, sample sizes differed across comparisons and models. The numbers of eyes and unique patients were reported alongside the results, and missing-data patterns were summarized.
All analyses were performed using Stata Statistical Software 17 (StataCorp LLC, College Station, TX, USA).

2.4. Ethics Approval

The study was conducted in accordance with the Declaration of Helsinki and approved by the ethics committee of the hospital where the study took place (2374/18.03.2026). Written informed consent was obtained from all subjects before procedures.

2.5. AI Statement

Generative AI (ChatGPT-6.2, Version 26.1120.41243) was used for language editing, text formatting, reformulation, literature searching, and assistance with generating and refining Stata commands and graphics. All statistical analyses were performed by the authors, and all results, interpretations, and integration with the current state of the art were independently reviewed and verified by the authors.

3. Results

A total of 134 eyes were included in the analysis. The mean age was 73.2 ± 5.4 years; baseline demographic, ocular biometric, biomechanical, and tonometric characteristics are summarized in Table 1. Four of the 67 participants (6.0%) had a diagnosis of glaucoma and were receiving topical IOP-lowering medication.
Table 1. Descriptive characteristics of the study population. Data are presented as mean ± standard deviation; ranges are shown in parentheses.
All three tonometric measurements were available for 116 of the 134 eyes. Eleven eyes had IOPcc and iCare measurements only, four had IOPcc and NCT only, two had NCT and iCare only, and one had iCare only. Consequently, the pair-specific analyses included different subsets of eyes, as reported in Table 2.
Table 2. Regression-based Bland–Altman analysis of agreement between tonometric methods. SD, standard deviation; LoA, limits of agreement; CI, confidence interval.

3.1. Bland–Altman Agreement Analysis

The mean IOPcc–NCT bias was small (+0.15 mmHg), but the 95% limits of agreement ranged from −5.15 to 5.44 mmHg, indicating substantial variation in inter-method differences across individual eyes. Significant proportional bias was observed only for IOPcc–NCT, with the inter-method difference increasing as the paired mean IOP increased. No statistically significant proportional bias was detected for IOPcc–iCare or NCT–iCare (Table 2, Figure 1).
Figure 1. Bland–Altman plots with fitted regression lines for IOPcc–NCT (A), IOPcc–iCare (B), and NCT–iCare (C).
For IOPcc–NCT, regression-based 95% limits of agreement shifted towards more positive differences as the paired mean IOP increased. At paired mean IOP values of 15 and 20 mmHg, the estimated limits were −5.23 to 4.56 mmHg and −3.53 to 6.26 mmHg, respectively, indicating substantial individual-eye variability at both pressure levels (Supplementary Figure S1).

3.2. Multivariable Linear Regression

Multivariable linear regression analysis showed that CH was independently associated with all three inter-tonometer differences (Table 3). After adjustment for CCT and age, each 1-mmHg decrease in CH was associated with a 1.56-mmHg increase in the IOPcc–NCT difference, a 2.23-mmHg increase in the IOPcc–iCare difference, and a 0.39-mmHg increase in the NCT–iCare difference.
Table 3. Multivariable linear regression analysis of factors associated with inter-tonometer differences. Values represent unstandardized regression coefficients (β) with 95% confidence intervals.
CCT remained independently associated with the IOPcc–NCT and NCT–iCare differences. Greater CCT was associated with a larger IOPcc–NCT difference and a smaller NCT–iCare difference, whereas no significant association was detected for IOPcc–iCare. Age was not significantly associated with any inter-tonometer difference. The models explained 44.0%, 52.3%, and 14.5% of the variance in the IOPcc–NCT, IOPcc–iCare, and NCT–iCare differences, respectively.
Surgical status was not independently associated with either tonometric difference (p = 0.46 and p = 0.18, respectively). Interaction analyses showed no statistically significant modification of the CH–bias association by surgical status for either IOPcc − NCT (β = 0.37; 95% CI: −0.29, 1.03; p = 0.26) or IOPcc − iCare (β = 0.29; 95% CI: −0.66, 1.25; p = 0.53). After adjustment for CH, CCT, age, and surgical status, CH remained significantly associated with both IOPcc − NCT (β = −1.55; 95% CI: −1.88, −1.23; p < 0.001) and IOPcc − iCare (β = −2.21; 95% CI: −2.66, −1.76; p < 0.001).

3.3. Inter-Tonometer Differences According to Corneal Hysteresis

Inter-tonometer differences varied across CH groups (Figure 2). Differences involving IOPcc decreased progressively with increasing CH, with significant differences between all pairs of groups. In the low CH group, IOPcc exceeded NCT and iCare by averages of 2.50 and 5.51 mmHg, respectively, whereas the corresponding differences were negative in the high CH group. The NCT–iCare difference also decreased with increasing CH, although only the comparison between the low and high CH groups remained statistically significant after Bonferroni adjustment.
Figure 2. Inter-tonometer differences across corneal hysteresis (CH) groups. Points represent mean inter-tonometer differences; error bars indicate 95% confidence intervals.

4. Discussion

The present study identified CH-associated variation in inter-tonometer differences after adjustment for CCT and age. Lower CH was associated with more positive IOPcc–NCT and IOPcc–iCare differences. In the low CH group, IOPcc exceeded iCare and NCT by averages of 5.5 and 2.5 mmHg, respectively. This group represents the lower CH range within the cohort rather than a validated clinical category of biomechanical vulnerability. Despite the small overall mean bias between IOPcc and NCT, the wide limits of agreement and significant proportional bias caution against assuming interchangeability at the individual-eye level.
For IOPcc–NCT, the overall 95% limits of agreement ranged from −5.15 to 5.44 mmHg. Thus, the near-zero average difference concealed substantial variation between individual eyes. Regression-based limits further illustrated that the expected difference shifted towards more positive values as paired mean IOP increased. Few observations were available at higher pressures, limiting the precision of these estimates. No clinically acceptable margin of disagreement was prespecified; therefore, clinical acceptability was not established.
Previous studies have reported inconsistent agreement between ORA, rebound tonometry, and non-contact tonometry [7,8,9,10,11]. Our findings indicate that inter-tonometer differences are associated with CH after adjustment for CCT and age. However, associations involving IOPcc may partly reflect the shared measurement information from which CH and IOPcc are derived. CH was also associated with NCT–iCare, although this association was modest and does not establish a physiological mechanism. These findings should consequently be interpreted as measurement associations rather than evidence of a causal biomechanical effect.
The observed differences do not establish which instrument more accurately estimates true IOP. In particular, higher IOPcc readings in eyes with low CH should not be interpreted as evidence that NCT or iCare underestimates true pressure. Although IOPcc was developed as a biomechanically compensated pressure estimate [12,13,14,15,16], this does not establish its superior accuracy in the present population.
CH may complement glaucoma assessment by providing information beyond corneal thickness and measured IOP. Previous studies have associated lower CH with glaucoma development and visual field progression independently of CCT [17,18,19,20,21]. CH has also been proposed as an indirect indicator of ocular biomechanical properties that may influence susceptibility to pressure-related optic nerve injury [21,22,23,24]. These observations provide a rationale for investigating CH as contextual information within a broader assessment of individual risk.
The present findings suggest a potential role for CH when interpreting measurements obtained with different tonometers during follow-up. In eyes resembling those in the low CH group, switching from NCT or iCare to IOPcc could produce an apparent pressure increase, whereas switching in the opposite direction could suggest a reduction without a corresponding physiological change. Awareness of inter-device differences may therefore assist interpretation of target IOP attainment or treatment response. Whenever possible, longitudinal comparisons should use the same tonometric method; when instruments change, the potential contribution of the measurement method should be considered. The group mean differences should not be used as fixed correction factors because individual discrepancies varied substantially.
IOP and CH should be interpreted alongside structural and functional findings. Menna et al. [25] reported increased measured RNFL thickness and improved OCT image quality after cataract surgery in eyes with and without preperimetric glaucoma, illustrating the influence of media opacity on structural measurements. This provides complementary context for longitudinal assessment: measurement-related influences should be considered before apparent tonometric or structural changes are attributed to disease progression or treatment response.
A strength of this study is the assessment of CH alongside CCT when examining inter-tonometer differences. By considering both parameters, together with age, the analysis extends beyond corneal thickness alone and characterizes CH-associated variation in measurements obtained with different instruments.
Several limitations should be acknowledged. The relatively modest sample consisted predominantly of older adults attending the one-month follow-up after cataract surgery in one eye, while fellow eyes were either unoperated or previously operated. Heterogeneity in surgical status and postoperative interval may have influenced the measurements and limits generalizability to other populations. The limited number of participants with glaucoma also precluded meaningful glaucoma-specific subgroup analyses.
The absence of Goldmann applanation tonometry limits comparison with the established clinical reference method, while the lack of direct manometric validation prevents conclusions regarding the accuracy of any device relative to true IOP. Furthermore, associations between CH and differences involving IOPcc may partly reflect shared ORA measurement characteristics. Prospective longitudinal studies are needed to determine whether CH-informed interpretation of inter-tonometer differences improves IOP monitoring or clinical decision-making.

Supplementary Materials

The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/jcm15239271/s1, Figure S1: Regression-based Bland–Altman limits of agreement for IOPcc–NCT, IOPcc–iCare, and NCT–iCare.

Author Contributions

F.V., E.A.M., and K.U.H. contributed to the conceptualization and study design. E.A.M. performed data collection, data analysis, and preparation of the original manuscript draft. F.V. and K.U.H. contributed to manuscript review and editing and provided supervision. All authors have read and agreed to the published version of the manuscript.

Funding

This work was supported by the George Emil Palade University of Medicine, Pharmacy, Science and Technology of Targu Mures, Research Grant number 795/3/22.01.2025.

Institutional Review Board Statement

The study was conducted in accordance with the Declaration of Helsinki and approved by the ethics committee of Mureș County Hospital (protocol code 2374; date of approval: 18 March 2026).

Data Availability Statement

Anonymized individual-level data and analytic materials underlying the findings of this study will be made available upon reasonable request to the corresponding author, subject to institutional approval and applicable ethical and data protection requirements.

Acknowledgments

This work was supported by the George Emil Palade University of Medicine, Pharmacy, Science and Technology of Targu Mures, Research Grant number 795/3/22.01.2025.

Conflicts of Interest

The authors declare no conflicts of interest.

Abbreviations

The following abbreviations are used in this manuscript:
CCTcentral corneal thickness
CHCorneal hysteresis
CRFCorneal resistance factor
IOPIntraocular pressure
IOPccCorneal compensated intraocular pressure
IOPgGoldmann-correlated intraocular pressure
NCTNon-contact tonometry
ORAOcular response analyzer

References

  1. Silva, F.; Lira, M. Intraocular pressure measurement: A review. Surv. Ophthalmol. 2022, 67, 1319–1331. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  2. Komninou, M.A.; Seiler, T.G.; Enzmann, V. Corneal biomechanics and diagnostics: A review. Int. Ophthalmol. 2024, 44, 132. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  3. Ariza-Gracia, M.Á.; Zurita, J.F.; Piñero, D.P.; Rodriguez-Matas, J.F.; Calvo, B. Coupled biomechanical response of the cornea assessed by non-contact tonometry: A simulation study. PLoS ONE 2015, 10, e0121486. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  4. Greene, P.R.; Sergienko, N.M.; Wang, S.K. Review: Measurement techniques for intraocular pressure. Optom. Open Access 2016, 1, 118. [Google Scholar] [CrossRef]
  5. Aoki, S.; Asaoka, R.; Fujino, Y.; Nakakura, S.; Murata, H.; Kiuchi, Y. Comparing corneal biomechanic changes among solo cataract surgery, microhook ab interno trabeculotomy and iStent implantation. Sci. Rep. 2023, 13, 19148. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  6. Kaushik, S.; Pandav, S.S.; Banger, A.; Aggarwal, K.; Gupta, A. Relationship between corneal biomechanical properties, central corneal thickness, and intraocular pressure across the spectrum of glaucoma. Am. J. Ophthalmol. 2012, 153, 840–849.e2. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  7. Brown, L.; Foulsham, W.; Pronin, S.; Tatham, A.J. The influence of corneal biomechanical properties on intraocular pressure measurements using a rebound self-tonometer. J. Glaucoma 2018, 27, 511–518. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  8. Vandewalle, E.; Vandenbroeck, S.; Stalmans, I.; Zeyen, T. Comparison of ICare, dynamic contour tonometer, and ocular response analyzer with Goldmann applanation tonometer in patients with glaucoma. Eur. J. Ophthalmol. 2009, 19, 783–789. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  9. Gillan, W.D.H. Intra-ocular pressure measurements using the Ocular Response Analyser and ICare tonometer: A comparison. Afr. Vis. Eye Health 2015, 74, a29. [Google Scholar] [CrossRef] [Scilit]
  10. Pluháček, F.; Unzeitigová, A.; Marešová, K.; Rybář, J. Influence of cornea on intraocular pressure measurement by ICARE PRO and ORA. Czech Slovak Ophtalmol. 2019, 75, 111–118. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  11. Wang, D.; Huang, W.; Li, Y.; Zheng, Y.; Foster, P.J.; Congdon, N.; He, M. Intraocular pressure, central corneal thickness, and glaucoma in chinese adults: The liwan eye study. Am. J. Ophthalmol. 2011, 152, 454–462.e1. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  12. Medeiros, F.A.; Weinreb, R.N. Evaluation of the influence of corneal biomechanical properties on intraocular pressure measurements using the ocular response analyzer. J. Glaucoma 2006, 15, 364–370. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  13. Annette, H.; Kristina, L.; Bernd, S.; Mark-Oliver, F.; Wolfgang, W. Effect of central corneal thickness and corneal hysteresis on tonometry as measured by dynamic contour tonometry, ocular response analyzer, and Goldmann tonometry in glaucomatous eyes. J. Glaucoma 2008, 17, 361–365. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  14. Oncel, B.; Dinc, U.A.; Orge, F.; Yalvaç, B.I. Comparison of IOP measurement by ocular response analyzer, dynamic contour, Goldmann applanation, and noncontact tonometry. Eur. J. Ophthalmol. 2009, 19, 936–941. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  15. Chui, W.S.; Lam, A.; Chen, D.; Chiu, R. The influence of corneal properties on rebound tonometry. Ophthalmology 2008, 115, 80–84. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  16. Carbonaro, F.; Andrew, T.; Mackey, D.A.; Spector, T.D.; Hammond, C.J. Comparison of three methods of intraocular pressure measurement and their relation to central corneal thickness. Eye 2010, 24, 1165–1170. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  17. Medeiros, F.A.; Meira-Freitas, D.; Lisboa, R.; Kuang, T.-M.; Zangwill, L.M.; Weinreb, R.N. Corneal hysteresis as a risk factor for glaucoma progression: A prospective longitudinal study. Ophthalmology 2013, 120, 1533–1540. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  18. Zhang, C.; Tatham, A.J.; Abe, R.Y.; Diniz-Filho, A.; Zangwill, L.M.; Weinreb, R.N.; Medeiros, F.A. Corneal hysteresis and progressive retinal nerve fiber layer loss in glaucoma. Am. J. Ophthalmol. 2016, 166, 29–36. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  19. Sit, A.J.; Chen, T.C.; Takusagawa, H.L.; Rosdahl, J.A.; Hoguet, A.; Chopra, V.; Richter, G.M.; Ou, Y.; Kim, S.J.; WuDunn, D. Corneal hysteresis for the diagnosis of glaucoma and assessment of progression risk: A report by the American Academy of Ophthalmology. Ophthalmology 2023, 130, 433–442. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  20. Susanna, C.N.; Diniz-Filho, A.; Daga, F.B.; Susanna, B.N.; Zhu, F.; Ogata, N.G.; Medeiros, F.A. A prospective longitudinal study to investigate corneal hysteresis as a risk factor for predicting development of glaucoma. Am. J. Ophthalmol. 2018, 187, 148–152. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  21. Murtagh, P.; O’Brien, C. Corneal hysteresis, intraocular pressure, and progression of glaucoma: Time for a “hyst-oric” change in clinical practice? J. Clin. Med. 2022, 11, 2895. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  22. Takeda, Y.; Hanyuda, A.; Takahashi, N.; Omodaka, K.; Ninomiya, T.; Kiyota, N.; Tsuda, S.; Yokoyama, Y.; Himori, N.; Takagi, A.; et al. Relationship between corneal hysteresis and the site of damage to peripapillary retinal nerve fibre layer thickness in open-angle glaucoma. Sci. Rep. 2024, 14, 26329. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  23. Wong, B.J.; Moghimi, S.; Zangwill, L.M.; Christopher, M.; Belghith, A.; Ekici, E.; Bowd, C.; Fazio, M.A.; Girkin, C.A.; Weinreb, R.N. Relationship of corneal hysteresis and anterior lamina cribrosa displacement in glaucoma. Am. J. Ophthalmol. 2020, 212, 134–143. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  24. Lee, K.M.; Kim, T.W.; Lee, E.J.; Girard, M.J.A.; Mari, J.M.; Weinreb, R.N. Association of corneal hysteresis with lamina cribrosa curvature in primary open-angle glaucoma. Investig. Ophthalmol. Vis. Sci. 2019, 60, 4171–4177. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  25. Menna, F.; De Luca, L.; Calabrò, M.; Meduri, A.; Lupo, S.; Vingolo, E.M. Retinal Nerve Fiber Layer Changes Following Cataract Surgery in Patients with and Without Preperimetric Glaucoma. J. Clin. Med. 2025, 14, 7255. [Google Scholar] [CrossRef] [Scilit] [PubMed]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.