Laser Efficiency and Laser Safety: Holmium YAG vs. Thulium Fiber Laser

,
,
and
1
Urology Department, Hospital Clínic de Barcelona, Villarroel 170, 08036 Barcelona, Spain
2
GRC Urolithiasis No. 20, Tenon Hospital, Sorbonne University, F-75020 Paris, France
3
Department of Urology AP-HP, Tenon Hospital, Sorbonne University, F-75020 Paris, France
4
Urology Department, University Hospital Southampton NHS Trust, Southampton SO16 6YD, UK
This article belongs to the Section Nephrology & Urology

Abstract

(1) Objective: To support the efficacy and safety of a range of thulium fiber laser (TFL) pre-set parameters for laser lithotripsy: the efficiency is compared against the Holmium:YAG (Ho:YAG) laser in the hands of juniors and experienced urologists using an in vitro ureteral model; the ureteral damage of both lasers is evaluated in an in vivo porcine model. (2) Materials and Methods: Ho:YAG laser technology and TFL technology, with a 200 µm core-diameter laser fibers in an in vitro saline ureteral model were used. Each participant performed 12 laser sessions. Each session included a 3-min lasering of stone phantoms (Begostone) with each laser technology in six different pre-settings retained from the Coloplast TFL Drive user interface pre-settings, for stone dusting: 0.5 J/10 Hz, 0.5 J/20 Hz, 0.7 J/10 Hz, 0.7 J/20 Hz, 1 J/12 Hz and 1 J/20 Hz. Both lasers were also used in three in vivo porcine models, lasering up to 20 W and 12 W in the renal pelvis and the ureter, respectively. Temperature was continuously recorded. After 3 weeks, a second look was done to verify the integrity of the ureters and kidney and an anatomopathological analysis was performed. (3) Results: Regarding laser lithotripsy efficiency, after 3 min of continuous lasering, the overall ablation rate (AR) percentage was 27% greater with the TFL technology (p < 0.0001). The energy per ablated mass [J/mg] was 24% lower when using the TFL (p < 0.0001). While junior urologists performed worse than seniors in all tests, they performed better when using the TFL than Ho:YAG technology (36% more AR and 36% fewer J/mg). In the in vivo porcine model, no urothelial damage was observed for both laser technologies, neither endoscopically during lasering, three weeks later, nor in the pathological test. (4) Conclusions: By using Coloplast TFL Drive GUI pre-set, TFL lithotripsy efficiency is higher than Ho:YAG laser, even in unexperienced hands. Concerning urothelial damage, both laser technologies with low power present no lesions.

Article Metrics

Citations

Article Access Statistics

Multiple requests from the same IP address are counted as one view.