Abstract
Localized extraoral soft-tissue tracking typically relies on large-scale facial stereophotogrammetry or low-cost smartphone reconstructions, which often lack the point-cloud density necessary for fine perioral details. This retrospective proof-of-concept study introduces a protocol transferring the high-resolution acquisition capabilities of a dental intraoral scanner (IOS; TRIOS 3, 3Shape) to an extraoral field for localized labial projection assessment, contextualized within a narrative literature background. Digital scans from five female patients who received hyaluronic acid filler (0.8 mL) were retrospectively analyzed at baseline (T0) and 30 days post-treatment (T1). A muscle deconditioning protocol was applied, and acquisitions were extended to stable subnasal structures to obtain a 3-point digital alignment. Sagittal linear cross-sections and colorimetric surface deviation maps were evaluated. Digital superimpositions captured distinct post-injection tissue displacements, with upper lip mean linear projection increases of 1.229 mm (median), 1.099 mm (right hemilip), and 1.048 mm (left hemilip) (nominal p < 0.05, with right hemilip retaining significance after Bonferroni correction; Cohen’s dz: 1.53–2.47). Utilizing an intraoral scanner extraorally represents a feasible chairside option by leveraging pre-existing clinical assets for localized lip projection tracking, although constrained by the pilot sample size and warranting larger prospective validation.
1. Introduction
The growing interest of the dental world in facial aesthetic medicine has increasingly led to the integration of facial aesthetic parameters in the customized planning of a treatment plan [1,2]. Among these, the labial and perioral region is the one that is most influenced and influences variations inherent in the aesthetics of the smile [3,4]. The digitalization of dental practices that has characterized the last decades has meant that, today, a large number of dentists have access to an intraoral scanner [5,6]. Intraoral scanners capture three-dimensional surface topography through advanced optical principles—most notably confocal microscopy, active wavefront sampling, or structured-light triangulation. In confocal systems such as the TRIOS 3 (3Shape, Copenhagen, Denmark), parallel confocal imaging projects structured light patterns onto the target substrate without powder coating, capturing in-focus optical slices across a focal depth of field (DOF) of approximately 12–18 mm and an optimal working distance ranging between 5 and 20 mm from the optical window. With a compact field of view (FOV) measuring 17 × 20 mm, the system records consecutive high-frequency image frames that are stitched in real time into dense point clouds with high native precision and truthfulness (<15 µm). Truthfulness refers to the proximity of a measurement to the true physical dimension, whereas precision indicates the degree of reproducibility among repeated scans [7]. The resulting digital datasets can be exported as open STL (Standard Tessellation Language) surface meshes or texture-mapped PLY and OBJ files displaying photorealistic color. While the acquisition cost of an intraoral scanner is considerably higher than that of mobile devices, modern dental clinics already possess these devices as amortized, pre-existing clinical assets. Repurposing an available chairside scanner for extraoral soft-tissue tracking maximizes capital asset efficiency and eliminates the recurring subscriptions and hardware purchases required for dedicated 3D facial imaging systems.
Recent literature focusing on extraoral digitization underscores a shifting paradigm in maxillofacial and aesthetic soft-tissue monitoring [5,8]. Historically, extraoral digitizations relied on bulky stereophotogrammetry or structured-light arrays; however, several studies have validated the extraoral application of intraoral scanners for capturing localized facial features, particularly in the fabrication of facial prostheses (such as nasal, auricular, and orbital prosthetics) and the recording of cleft defects [8,9,10]. In these prosthetic workflows, the high point-cloud density of intraoral optical tips allowed precise registration of mucosal margins and complex defect anatomies without the tissue compression associated with conventional elastomeric impressions. Recent investigations have sought to establish consistent handpiece scanning strategies over cutaneous surfaces, although operator movement and soft-tissue deformability continue to present registration challenges [11,12]. While full-face scanning is ideal for macro-structural cranial planning, localized cosmetic adjustments—such as labial projection changes—benefit substantially from the microscopic detail and localized mesh density that adjusted intraoral optical systems provide [5,11].
The primary aim of this article is to retrospectively evaluate a proof-of-concept clinical methodology for recording the labial and perioral region using a widely available intraoral scanner (3Shape® TRIOS3). Secondarily, to contextualize this protocol within current digital workflows, this study integrates a narrative review of contemporary 3D facial soft-tissue acquisition technologies, establishing a clear comparative framework for short- and long-term aesthetic monitoring.
2. Materials and Methods
2.1. Retrospective Case Selection
This study was designed as a retrospective review of clinical cases. Medical records and digital scans from a cohort of five female patients (aged 25–35 years) who had independently sought and received lip filler treatments were retrospectively analyzed. The study protocol conformed to the ethical guidelines of the 1964 Declaration of Helsinki and its later amendments, as well as European General Data Protection Regulation (GDPR) requirements. Under Italian national legislation governing observational research (Italian Medicines Agency [AIFA] Guidelines on Observational Studies, 20 March 2008) and internal institutional policies of the University of L’Aquila, formal prior approval by an Ethics Committee is waived for non-interventional, retrospective audits analyzing fully anonymized clinical records and routine diagnostic imaging collected during standard care. All patients provided written informed consent authorizing the clinical treatment, the acquisition of digital models, and the retrospective use of their anonymized data for scientific research and publication. Strict inclusion criteria for this retrospective chart review were established to isolate anatomical variables, requiring no history of volumetric or rejuvenation lip filler treatments within the preceding 12 months, a physiological lip competence at rest (rest position), and a shared therapeutic goal of increasing lip volume and enhancing vermilion border definition.
2.2. Postural Standardization and Control of Facial Mimetic Activity
Given that perioral soft tissues exhibit significant volumetric and positional variations even with minimal involuntary contraction of the orbicularis oris muscle, the reproducibility of the rest position across different time points—specifically at pre-treatment (T0) and 30 days post-treatment (T1)—was secured through a rigorous conditioning protocol applied as part of the routine clinical scanning procedure. First, cervical stabilization was achieved by performing each scanning session with the patient seated in a dental chair equipped with a locked, fixed headrest to inhibit any neck extension, flexion, or rotation. Second, to eliminate residual muscle tone or conscious labial hyperactivity, a method of post-deglutition muscle deconditioning was applied. Patients were instructed to swallow and immediately thereafter completely relax their lips, maintaining the tissues in a gentle, passive interlabial contact while avoiding any active clenching or compression pressure.
2.3. Modified 3D Intraoral Scanning Protocol
Three-dimensional digital acquisitions at T0 and T1 were performed using a high-resolution intraoral scanner (TRIOS 3, 3Shape, Copenhagen, Denmark). Prior to each clinical session, the scanner underwent optical calibration using the manufacturer’s dedicated calibration tip and 3D calibration target, followed by color balance calibration. The scanner operates via parallel confocal microscopy with a 17 × 20 mm FOV and native resolution < 15 µm. To enable the scanning of extraoral skin and lips, the software’s native artificial intelligence filter (Scan Assist Technology) was deactivated. Scan Assist is natively calibrated to eliminate soft tissues (such as cheeks, tongue, and lips) to isolate dental arches; its deactivation permits continuous raw surface mesh generation across mucocutaneous contours without automated clipping. During acquisition, the handpiece was maintained perpendicular to the cutaneo-mucosal surface at a stand-off distance of 10–15 mm. The scan started at the right labial commissure, advanced toward the midline using smooth vertical sweeps, and was extended superiorly to capture the philtrum, the subnasale, the bilateral alar bases, and the nasal tip, before completing the contralateral hemilip. Capturing these subnasal and nasal cartilaginous structures is essential, as they serve as rigid, stable anatomical reference areas unaffected by filler-induced soft-tissue expansion, enabling accurate subsequent digital superimposition.
2.4. Clinical Injection Procedure
To eliminate operator-dependent variability, both the optical scanning phases and the clinical injection procedures were performed by a single operator. The treatment involved the administration of a medium cross-linked hyaluronic acid (STYLAGE S, Vivacy), formulated with a technology that promotes biomimetic tissue integration under both static and dynamic conditions. The filler was injected in a strictly homogeneous quantity of 0.8 mL across all patients, utilizing a vertical vector technique (Russian Lips Technique) to enhance the projection and height of the lip body while minimizing artificial eversion. Specifically, the product was delivered using a 30G × 4 mm needle, targeting the superficial lamina propria of the vermilion border in serial retrograde micro-traces, maintaining a superficial injection plane to minimize the risk of accidental cannulation or compression of the deep superior and inferior labial arteries.
All clinical injection procedures and digital scans were performed exclusively as part of routine medical and aesthetic care, independent of the subsequent retrospective analysis.
2.5. Digital Alignment and Linear Projection Analysis
The Standard Triangle Language (STL) surface meshes exported at T0 and T1 were imported into Dental System (3Shape). Digital superimposition of baseline and follow-up models was performed using a point-based alignment algorithm, placing reference points exclusively on stable subnasal landmarks: the nasal tip, the subnasale, and the right and left alar base margins. Quantitative analysis of soft-tissue projection was performed using two complementary approaches:
For sagittal linear cross-sectional analysis, three parallel sagittal planes were established: (1) the median sagittal plane, positioned along the philtral midline; (2) the right hemilip sagittal plane; and (3) the left hemilip sagittal plane. The midline of each hemilip was defined on the horizontal transverse axis at a position precisely equidistant between the philtral midline and the respective labial commissure (cheilion) at rest. To avoid angular inclination bias, all three sagittal slicing planes were oriented mutually parallel and strictly perpendicular to the coronal plane (defined by the facial midline and bichelion axis) and the axial plane. Along each sagittal section, linear projection was measured along an axis perpendicular to the overlapping cutaneo-mucosal profiles at the point of maximum labial prominence for both the upper and lower lips.
Concurrently, surface deviation color maps were generated using Exocad (Exocad GmbH, Darmstadt, Germany). The distance calculation displayed outward tissue displacement within a 0 to +2 mm spectrum, where 0 mm (blue) indicated no change and +2 mm (red) represented maximum projection gain. Negative deviation values (<0 mm) were filtered out in this display setting to isolate positive tissue projection; however, it should be recognized that this unidirectional thresholding restricts the map from depicting localized tissue compression or dorsal inward displacement.
2.6. Statistical Analysis
Quantitative data for linear labial projection changes (ΔT1 − T0) were compiled across the six anatomical sections (median, right hemilip, and left hemilip for both upper and lower lips). Descriptive statistics were expressed as mean, standard deviation (SD), and 95% confidence intervals (95% CI). Given the small sample size (n = 5), distribution normality was explored using the Shapiro–Wilk test; however, recognizing that normality tests have limited power in small samples, both parametric paired Student’s t-tests and non-parametric Wilcoxon signed-rank tests were conducted. To protect against type I error inflation across the six anatomical comparisons, Bonferroni correction was applied, establishing an adjusted statistical significance threshold of α = 0.05/6 = 0.0083. Standardized effect sizes were calculated as Cohen’s dz () alongside their corresponding 95% confidence intervals to evaluate clinical magnitude. Subjective GAIS scores were summarized as frequency counts and percentages as descriptive clinical metrics. Statistical analyses were performed using SPSS version 28.0 (IBM Corp., Armonk, NY, USA).
3. Results
The registration technique performed according to the modified intraoral scanning protocol yielded a reproducible digital alignment of the 3D surface datasets acquired at pre-treatment (T0) and 30 days post-treatment (T1). This alignment provided a stable reference for evaluating post-injection linear projection changes across the perioral soft tissues based on the rigid anatomical reference points established on the nose (Figure 1A). The resulting 3D rendering demonstrated close surface registration across unaltered nasal areas, serving as a baseline for subsequent topographic evaluations (Figure 1B).
Figure 1.
Three-dimensional model alignment and registration workflow. (A) Identification of unaltered anatomical landmarks using the 3-point alignment function on the rigid subnasal structures (subnasale, alar base margins, and nasal tip) to register baseline (T0) and post-treatment (T1) datasets; the red and green paired markers indicate the homologous reference points placed on the pre- and post-treatment meshes for digital registration. (B) Final 3D color-textured rendering resulting from the superimposition of baseline and 30-day post-treatment scans of Patient I.
Quantitative analysis of the soft-tissue changes was carried out using linear measurements perpendicular to the overlapping labial curves, evaluated across three distinct sagittal cross-sections comprising the median portion of the philtrum (Figure 2A), the right hemilip midline (Figure 2B), and the left hemilip midline (Figure 2C). These sagittal cross-sections facilitated the mathematical calculation of the tangential distance at the specific coordinates of maximum cutaneo-mucosal prominence for both the upper and lower lips, capturing the largest anatomical portion present within each section through the dedicated 2D cross-sectional software interface (Figure 2D).
Figure 2.
Sagittal cross-sectional segmentation and linear measurement methodology. 3D visual planning interfaces demonstrating the orientation of the reference planes for data extraction; the blue circular outline with green shaded disk indicates the interactive virtual sagittal slicing plane used to generate cross-sections: (A) median plane section centered at the upper lip philtrum; (B) sagittal section along the midline of the right hemilip; (C) sagittal section along the midline of the left hemilip. (D) Representative two-dimensional cross-sectional software interface showing the linear measurement taken tangentially to the points of maximum prominence perpendicular to the overlapping labial curves to calculate the absolute projection increase in millimeters.
The data derived from these linear measurements revealed an increase in labial projection at T1 compared to baseline (T0). For the upper lip, the mean linear increases were 1.229 ± 0.741 mm at the median philtral section, 1.099 ± 0.480 mm at the right hemilip, and 1.048 ± 0.535 mm at the left hemilip. For the lower lip, the corresponding mean values were 0.797 ± 0.523 mm at the median section, 0.737 ± 0.344 mm at the right hemilip, and 1.207 ± 0.488 mm at the left hemilip (Table 1).
Table 1.
Sagittal linear measurement values across the study sample. Absolute values and calculated average measurements (expressed in millimeters) captured across the three reference sagittal sections (median section, right hemilip section, and left hemilip section) for both the upper and lower lips of all five treated patients at 30 days post-treatment (T1-T0).
These quantitative measurements were fully corroborated by the colorimetric surface deviation analysis, which illustrated a highly consistent pattern of projection changes ranging between 0 mm and 2 mm (Figure 3). The color mappings effectively converted the localized physical displacement of the lips into a visual gradient for all monitored subjects, demonstrating a precise and reproducible correlation between the peak mechanical projection measured in millimeters and the warmer red tone regions of the chromatic scale across the remaining pool of patients in the study cohort (Figure 4).
Figure 3.
(A,B) Colorimetric surface deviation mapping of Patient I. Three-dimensional topographic representation of the linear variations in labial projection between baseline (T0) and 30 days post-treatment (T1). The color transition scale dictates a predefined tolerance range extending from 0 mm (blue tones, indicating absence of variation) up to 2 mm (red tones, representing the maximum localized linear increase in projection).
Figure 4.
Multi-patient colorimetric analysis of volumetric lips enhancement. Comprehensive evaluation of the study cohort demonstrating the consistency and reproducibility of the treatment protocol across the remaining subjects: (A) Patient II, (B) Patient III, (C) Patient IV, and (D) Patient V. Chromatic variations illustrate localized linear soft-tissue projection changes along a predefined scale, where blue tones represent absence of variation (0 mm) and warmer yellow-to-red tones indicate maximum positive projection increase (up to +2 mm).
Clinical assessment of aesthetic outcomes provided descriptive qualitative documentation alongside objective digital measurements. The assessing clinician scored all five cases with a value of 2 on the Global Aesthetic Improvement Scale (GAIS), indicating that the perioral area was rated as “Improved” compared to baseline (Table 2). Patient self-assessment further documented satisfaction, with three patients rating their aesthetic results as 1 (“Much Improved”) and two patients assigning a value of 2 (“Improved”) (Table 3). While these qualitative scores concur with observed clinical changes, they represent independent descriptive ratings rather than formal statistical correlation with linear metrics.
Table 2.
Investigator-reported aesthetic assessment via Global Aesthetic Improvement Scale (GAIS). Frequency distribution of the aesthetic improvement scores assigned independently by the dental clinician at the 30-day follow-up (T1), utilizing the standardized 4-point qualitative index.
Table 3.
Patient-reported aesthetic satisfaction via Global Aesthetic Improvement Scale (GAIS). Subjective evaluation and frequency distribution of self-reported satisfaction scores recorded by the five female patients at the 30-day post-treatment follow-up (T1).
Statistical analysis confirmed that all evaluated anatomical regions exhibited a statistically significant mean projection increase from T0 to T1 at the unadjusted level (paired t-test, nominal p-values ranging from 0.0052 to 0.0270). The Shapiro–Wilk test yielded p > 0.05 across all sections, although test power was limited by the sample size (n = 5). When applying non-parametric Wilcoxon signed-rank tests, all differences were uniformly positive (W = 15, exact two-tailed p = 0.0625, which represents the minimal achievable p-value for n = 5). To account for multiple comparisons across the six sections, a Bonferroni correction was applied, establishing a significance threshold of α = 0.05/6 = 0.0083. Under this threshold, projection increases remained statistically significant in the right hemilip upper lip (p = 0.0069) and the left hemilip lower lip (p = 0.0052), whereas changes in other sections should be interpreted as nominal gains. Standardized effect sizes via Cohen’s dz ranged from 1.53 to 2.47, with wide 95% confidence intervals reflecting sample size constraints (Table 4).
Table 4.
Inferential statistical analysis of linear projection variations (T1-T0). Statistical metrics include mean change, standard deviation (SD), 95% confidence intervals (CI), paired t-test values, Wilcoxon signed-rank test statistics, Cohen’s dz with 95% CI. * Indicates nominal statistical significance (p < 0.05); † indicates significance retained after Bonferroni correction (α = 0.0083).
4. Discussion
To contextualize the findings of this proof-of-concept study, a narrative review of the current literature regarding three-dimensional (3D) facial acquisition systems is presented. Considering the most common and certified 3D acquisition methods, it is well established in the literature that volumetric rendering of the labial and perioral region is historically extrapolated from large-scale volumetric acquisitions of the entire face [1,9,10,12]. Several studies compare highly accredited, dedicated stereophotogrammetry systems—such as the Vectra M3 or 3dMDface technology, widely considered the gold standards for macro-facial analysis due to their instantaneous, single-capture mechanism [12,13]—with the diagnostic accuracy of 3D facial reconstructions derived from contemporary smartphone applications, which are currently available to clinical practitioners at a substantially reduced cost [10,11,12,14,15].
While images obtained from specific, professional 3D photography systems remain significantly superior in terms of full-face accuracy [15,16], recent findings by Karabiber et al. [11] and Hartmann et al. [14], alongside comparative studies by Revilla-León et al. [17] and D’Ettorre et al. [18], demonstrate that smartphone-based renderings frequently serve as a clinically acceptable alternative for capturing facial soft tissues when assisting in general aesthetic diagnoses, orthodontic planning, or maxillofacial assessments [1,11,13,14,15,19]. However, they inherently lack the point-cloud density necessary to track fine mucosal shifts and micro-volumetric labial updates [17] and are highly vulnerable to artifacts caused by hand instability and patient head movements during the increased acquisition time [18]. A key technological distinction must be made regarding localized resolution and topography truthfulness. The exceptionally high degree of accuracy inherent to modern intraoral scanners (IOS) allows these devices to replicate fine surface topography details down to the micron level [1,3,5,12]. Recent breakthroughs in cloud-point processing algorithms have further minimized surface reconstruction noise, optimizing the baseline rendering of delicate mucocutaneous transitions [20]. Therefore, transferring the application of an IOS from a strictly intraoral environment to an extraoral field presents an innovative opportunity to generate ultra-high-resolution 3D datasets, rendering them uniquely ideal for localized analytical, prospective, and longitudinal soft-tissue tracking [1,5].
To provide a clear overview of the current technological landscape in facial aesthetic monitoring, Table 5 summarizes the direct comparison between full-face stereophotogrammetry, smartphone applications, and the extraoral application of intraoral scanners based on key clinical parameters.
Table 5.
Comprehensive Comparison of 3D Soft-Tissue Acquisition Technologies.
While full-face stereophotogrammetry systems like the Vectra M3 or the 3dMDface system remain the recognized gold standards for comprehensive facial analysis [13,21], they exhibit clear practical drawbacks within routine clinical practice, including substantial economic costs, a large physical footprint requiring dedicated clinic space, and a complete lack of portability [8,12,14,18]. As highlighted by Hansson et al. and Hong et al. in their respective scientific validations, these complex device arrays present clear boundaries for wide clinical deployment. In sharp contrast, utilizing an intraoral scanner for extraoral perioral monitoring introduces a highly accessible, cost-effective, chairside alternative [1,12]. Given the widespread digitization of modern dental practices [1,12], most clinicians already possess a high-definition IOS within their immediate operative armamentarium. Expanding its clinical utility to include localized extraoral mapping optimizes existing digital assets, simplifies the administrative digital workflow, and enhances direct patient communication, enabling objective follow-up tracking without requiring additional equipment or diagnostic camera investments [1,12].
The quantitative findings derived from this pilot cohort explicitly underscore the clinical viability and measurement sensitivity of this modified digital workflow. Following a standardized clinical delivery of a tissue-mimetic, medium cross-linked hyaluronic acid filler (STYLAGE S, Vivacy) [1], the digital superimposition successfully captured distinct micro-volumetric variations at 30 days post-treatment (T1) [1]. This localized tissue expansion was verified across all three reference sagittal sections, showing a clear increase in mean labial projection for the upper lip (1.229 mm at the median section, 1.099 mm at the right hemilip, and 1.048 mm at the left hemilip) relative to the lower lip baseline comparisons (0.797 mm at the median section, 0.737 mm at the right hemilip, and 1.207 mm at the left hemilip) [1]. These definitive metric calculations are further sustained by the colorimetric surface deviation maps generated in EXOCAD® (Figure 3 and Figure 4A–D), where the continuous chromatic transition from blue to red objectively maps the spatial displacement of the tissues. This topographic representation aligns seamlessly with the high levels of patient self-reported satisfaction and investigator-reported aesthetic improvement captured via the qualitative GAIS index (Table 2 and Table 3). The linear projection gains observed in this cohort (mean upper lip projection increase of 1.05 to 1.23 mm; lower lip 0.74 to 1.21 mm at 30 days) are in close agreement with previously published 3D stereophotogrammetric and sonographic evaluations of hyaluronic acid lip augmentation. Specifically, Rho et al. [9], evaluating the lip lifting and volumizing efficacy of 1.0 mL hyaluronic acid injections using stereophotogrammetry (Vectra H1), reported anterior sagittal projection gains ranging from 1.02 to 1.38 mm at 4 to 8 weeks. Similarly, Nikolis et al. [10] reported 30-day projection gains of 1.1 to 1.3 mm for the upper lip following medium-viscosity cohesive hydrogel delivery. Kim [20] observed average sagittal projection increments between 0.8 and 1.2 mm utilizing 3D facial imaging and high-frequency ultrasound. While large effect sizes were obtained in our sample (Cohen’s dz: 1.53–2.47), these metrics must be interpreted cautiously; in a pilot cohort of five patients, effect sizes inherently present wide confidence intervals and do not by themselves establish statistical power or replace formal reliability testing. Furthermore, contemporary metrological validations utilizing advanced industrial inspection software confirm that localized distance distributions within such small sample cohorts remain highly descriptive of authentic soft-tissue updates [22,23].
From a rheological and clinical perspective, the behavior of the injected material plays an essential role in the accuracy of the 3D surface mapping. Hyaluronic acid hydrogels characterized by intermediate cross-linking density and particle-type structures possess a high elastic modulus (G’) and strong cohesiveness, which maximize their overall lifting capacity and resistance to mechanical deformation under static and dynamic conditions [8,11]. Unlike non-particle gels that exhibit highly fluid tissue integration and potential cross-border spreading over time, the selected cohesive network layout maintains a stable, sharply defined projection along the vermilion border and Cupid’s bow [8,11]. This structural stability minimizes superficial blurring and morphological migration, directly matching the high geometric trueness of the intraoral optical scanning device, which requires a stable substrate to minimize tracking errors during frame capture [3,11]. This technical synchronization is heavily emphasized in smile design paradigms, where the maintenance of specific vertical parameters and incisal design directly determines overall facial harmony [1,3,4,5]. Recent metrological studies using surface deviation algorithms confirm that the integration of cohesive fillers prevents the peripheral dispersion of data points, ensuring that post-injection alignment reflects true localized structural expansion rather than material migration [24]. This long-term structural resilience is supported by recent 3D sonographic literature, which confirms the absence of horizontal material displacement when particle-type hydrogels are delivered under proper hydraulic pressure [25].
Furthermore, our study addresses a critical challenge frequently noted in facial soft-tissue literature: the distortion of surface alignment algorithms due to post-treatment morphologic changes and involuntary mimetic activity [7,10,21]. The perioral area is highly mobile and prone to significant muscular distortion during animation [16,19]. Quantitative dynamic assessments of perioral surface stretch and strain indicate that involuntary muscle contractions can significantly warp surface-to-surface alignment results [10]. Recent biomechanical trials tracking perioral stretch kinematics show that even subconscious micro-movements of the orbicularis oris complex can compromise the structural alignment of soft tissues [26]. To eliminate this major confounding factor, our protocol implemented a strict post-deglutition muscle deconditioning procedure to achieve a fully reproducible rest position before each scan session. Additionally, our scanning trajectory deliberately extended superiorly to encompass unchanging, rigid anatomical landmarks, specifically the philtrum, the subnasale, the bilateral alar base margins, and the nasal tip. Because these specific nasal structures are completely unaffected by the sub-mucosal or intra-muscular delivery of labial hyaluronic acid, they provide an unyielding geometric anchor. Utilizing these zones for the guided 3-point alignment function (Figure 1) guarantees that the structural superimposition of T0 and T1 datasets remains mathematically unwarped by the treatment itself, successfully isolating the true cutaneo-mucosal displacement and projection of the lips. Recent investigations into 3D facial mesh alignment validate that excluding areas of acute clinical intervention from the primary best-fit registration area prevents registration bias, leading to a massive increase in topological trueness [9]. Advanced spatial analyses of facial mesh orientation confirm that rigid anchoring on fixed midface elements provides an unyielding reference system that prevents rotational registration errors [27].
Looking toward future clinical directions, the literature suggests that the integration of deep learning networks into extraoral IOS exports will soon automate the subtraction of dynamic muscle deformation in real time [6]. Recent neural network architectures applied to maxillofacial computing have demonstrated high efficiency in automatically predicting soft-tissue volumetric alterations, filtering out superficial operator artifacts [28]. Furthermore, combining ultra-high-resolution PLY files with face-scanning smartphone apps via hybrid alignment software could bridge the gap between macro-facial frames and micro-topographical details [10,29]. This data synchronization opens innovative frontiers for generating patient-specific, 3D-printed templates for injectable delivery, shifting lip aesthetic medicine from an empirical discipline to a mathematically guided, predictable clinical workflow [14,24].
Limitations
In the spirit of scientific transparency, several technical limitations inherent to using an IOS for extraoral surface capture must be formally acknowledged. First, intraoral scanners are natively engineered and calibrated to capture rigid, non-yielding dental structures under controlled moisture conditions [3,7]. This specialized substrate interaction has been shown in recent optical trials to lower the overall speed of data acquisition compared to standard tooth surfaces, potentially compounding operator-dependent variation [30]. As recently demonstrated by Dutton et al. and Renne et al., extraoral soft tissues possess vastly different light-reflection properties, and their pliable, soft nature makes them highly vulnerable to minor physical deformations if the scanner tip accidentally contacts the cutaneous or mucosal skin surface during the vertical up-down scanning trajectory [3,7,30]. Recent spectrophotometric updates emphasize that skin melanin content and mucosal moisture levels cause specific subsurface light scattering, altering the sensor absorption of standard intraoral hardware [31].
Second, the field of view (FOV) of a standard IOS is highly restricted—measuring just 17 × 20 mm in the case of the TRIOS 3—which demands a significantly longer acquisition time compared to the instantaneous, single-capture mechanism of multi-camera stereophotogrammetry arrays [8,12,14]. As analyzed by Son et al. [32] and Revilla-León et al. [19], this prolonged scanning time introduces a higher risk of “stitching errors,” where minor computational mismatches occur during the real-time stitching of overlapping image frames, especially if the patient exhibits subtle head movements or involuntary mimetic twitching during the extraoral procedure [3,8,17,19]. This limitation is particularly evident during full-arch extraoral extensions, where spatial accumulation errors can manifest along the transitional curves of the facial profile [32]. Mathematical models evaluating iterative frame registration show that spatial accumulation errors tend to amplify as the handpiece crosses non-reflecting perioral borders [33]. The extended acquisition duration inherently increases patient fatigue, heightening the risk of subconscious micro-contractions of the orbicularis oris muscle that may compromise surface registration precision despite cervical headrest stabilization and deglutition deconditioning.
Third, specific metrological and reliability assessments could not be performed within this retrospective proof-of-concept dataset: (a) test–retest precision from repeated, consecutive scans acquired on the same patient at the exact same time point was not evaluated, precluding direct calculation of scan-rescan measurement error; (b) intra-examiner and inter-examiner reliability (e.g., intraclass correlation coefficients for cross-sectional plane alignment and landmark digitization) were not quantified; and (c) concurrent validation against an established reference standard, such as 3D stereophotogrammetry (e.g., Vectra or 3dMD) or direct physical caliper anthropometry, was not performed. Consequently, claims of high absolute geometric accuracy must be interpreted with caution. Additionally, the same clinician performed the filler injection, optical scanning, cross-sectional measurements, and GAIS scoring; the absence of blinded, independent evaluators introduces a potential risk of observer and confirmation bias that must be recognized.
Finally, the present study is fundamentally constrained by its small sample size (n = 5) and retrospective design. Although large, standardized effect sizes were observed (Cohen’s dz: 1.53–2.47), they are accompanied by wide 95% confidence intervals and cannot substitute for formal statistical power or establish separation from potential measurement error. As highlighted in metrological evaluations by De Menezes, Rosati, and Sforza et al. [23], localized surface deviation analyses require rigorous rigid anchoring to remain descriptive of authentic soft-tissue updates. Anchoring on stable subnasal structures (subnasale, alar base margins, and nasal tip) provided a feasible alignment strategy; however, following this preliminary proof-of-concept, further prospective clinical trials utilizing larger, more diverse patient populations, blinded multiple raters, repeated scan-rescan protocols, and direct stereophotogrammetric validation are mandatory to fully establish the clinical reproducibility, topographic stability, and scalability of this extraoral digital technique [8,11,14].
5. Conclusions
This retrospective proof-of-concept study demonstrates the preliminary feasibility of utilizing a standard dental intraoral scanner for localized extraoral imaging to assess linear projection changes in the lips following hyaluronic acid injection. By deactivating soft-tissue removal filters and anchoring digital registration on stable subnasal structures, the protocol provides an accessible method to track localized soft-tissue shifts using equipment already present in dental practices. However, given the small sample size (n = 5), 30-day timeframe, and single-operator design, these findings should not be extrapolated to establish long-term clinical monitoring or broad diagnostic efficacy. Further prospective investigations incorporating larger cohorts, independent blinded raters, repeated-scan precision testing, and direct validation against established stereophotogrammetric systems are required before this technique can be recommended for routine clinical adoption.
Author Contributions
Conceptualization, F.C., G.B. and M.D.; methodology, F.C., G.B. and M.D.; validation, F.C., A.G. and G.C.; formal analysis, F.C., A.C. and M.D.; investigation, F.C.; data curation, F.C., A.G., G.C., F.D.A. and A.C.; writing—original draft preparation, F.C., G.C., A.C. and G.B.; writing—review and editing, A.G., F.D.A. and M.D.; supervision, M.D. All authors have read and agreed to the published version of the manuscript.
Funding
This research received no external funding.
Institutional Review Board Statement
This study was conducted as a retrospective, non-interventional analysis of pre-existing clinical records and digital scans obtained during routine patient care. No additional clinical procedures or patient recruitment were undertaken for research purposes. Given the retrospective nature of the research and the absence of any additional research-related interventions involving human subjects, prior Ethics Committee approval was considered not required in accordance with national legislation (Italian Medicines Agency [AIFA] Guidelines on Observational Studies, 20 March 2008, http://www.omceoge.org/newsdoc/AIFA_20_3_2008.pdf) (accessed on 8 September 2026).
Informed Consent Statement
Written informed consent for the clinical procedure, the routine photographic/digital documentation, and the subsequent use of anonymized clinical data for research and publication purposes was obtained from all patients prior to their treatments.
Data Availability Statement
The datasets generated and analyzed during the current study are not publicly available due to patient privacy and ethical restrictions under the General Data Protection Regulation (GDPR) but are available from the corresponding author upon reasonable request.
Acknowledgments
The authors would like to express their sincere gratitude to Massimo Frascaria for his valuable assistance in image processing and management.
Conflicts of Interest
The authors declare no conflicts of interest.
References
- Machado, A.W. 10 commandments of smile esthetics. Dent. Press J. Orthod. 2014, 19, 136–157. [Google Scholar] [CrossRef] [Scilit] [PubMed] [PubMed Central]
- Blatz, M.B.; Chiche, G.; Bahat, O.; Roblee, R.; Coachman, C.; Heymann, H.O. Evolution of Aesthetic Dentistry. J. Dent. Res. 2019, 98, 1294–1304. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Fallahi, H.R.; Keyhan, S.O.; Bohluli, B.; Cheshmi, B.; Jafari, P. Lip Lift Techniques in Smile Design. Dent. Clin. North Am. 2022, 66, 443–457. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Polo, M. Botulinum Toxin and Smile Design. Dent. Clin. North Am. 2022, 66, 419–429. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Eggmann, F.; Blatz, M.B. Recent Advances in Intraoral Scanners. J. Dent. Res. 2024, 103, 1349–1357. [Google Scholar] [CrossRef] [Scilit] [PubMed] [PubMed Central]
- Carrillo-Perez, F.; Pecho, O.E.; Morales, J.C.; Paravina, R.D.; Della Bona, A.; Ghinea, R.; Pulgar, R.; Pérez, M.d.M.; Herrera, L.J. Applications of artificial intelligence in dentistry: A comprehensive review. J. Esthet. Restor. Dent. 2022, 34, 259–280. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lee, S.J.; Kim, S.W.; Lee, J.J.; Cheong, C.W. Comparison of Intraoral and Extraoral Digital Scanners: Evaluation of Surface Topography and Precision. Dent. J. 2020, 8, 52. [Google Scholar] [CrossRef] [Scilit] [PubMed] [PubMed Central]
- Ortensi, L.; Rosa, G.R.M.; Ciletta, S.; Grande, F.; Pedullà, E. Evaluation of the accuracy of digital impressions with different scanning strategies: An in vitro study. J. Dent. 2024, 151, 105433. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Rho, N.K.; Goo, B.L.; Youn, S.J.; Won, C.H.; Han, K.H. Lip Lifting Efficacy of Hyaluronic Acid Filler Injections: A Quantitative Assessment Using 3-Dimensional Photography. J. Clin. Med. 2022, 11, 4554. [Google Scholar] [CrossRef] [Scilit] [PubMed] [PubMed Central]
- Nikolis, A.; Bertucci, V.; Solish, N.; Lane, V.; Objective, N.A.A. Quantitative Assessment of Flexible Hyaluronic Acid Fillers in Lip and Perioral Enhancement. Dermatol. Surg. 2021, 47, e168–e173. [Google Scholar] [CrossRef] [Scilit] [PubMed] [PubMed Central]
- Karabiber, G.; Yılmaz, H.N.; Yıldırım, G. Are different photogrammetry applications on smartphones sufficiently reliable? Korean J. Orthod. 2025, 55, 37–47. [Google Scholar] [CrossRef] [Scilit] [PubMed] [PubMed Central]
- de Menezes, M.; Rosati, R.; Allievi, C.; Sforza, C. A photographic system for the three-dimensional study of facial morphology. Angle Orthod. 2009, 79, 1070–1077. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Hong, C.; Choi, K.; Kachroo, Y.; Kwon, T.; Nguyen, A.; McComb, R.; Moon, W. Evaluation of the 3dMDface system as a tool for soft tissue analysis. Orthod. Craniofac Res. 2017, 1, 119–124. [Google Scholar] [CrossRef] [Scilit] [PubMed] [PubMed Central]
- Luo, Y.; Zhao, M.; Lu, J. Accuracy of Smartphone-Based Three-Dimensional Facial Scanning System: A Systematic Review. Aesthetic Plast. Surg. 2024, 48, 4500–4512. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Hartmann, R.; Weiherer, M.; Nieberle, F.; Palm, C.; Brébant, V.; Prantl, L.; Lamby, P.; Reichert, T.E.; Taxis, J.; Ettl, T. Evaluating smartphone-based 3D imaging techniques for clinical application in oral and maxillofacial surgery: A comparative study with the vectra M5. Oral Maxillofac. Surg. 2025, 29, 29. [Google Scholar] [CrossRef] [Scilit] [PubMed] [PubMed Central]
- Singh, P.; Hsung, R.T.; Ajmera, D.H.; Said, N.A.; Leung, Y.Y.; McGrath, C.; Gu, M. Smartphone-generated 3D facial images: Reliable for routine assessment of the oronasal region of patients with cleft or mere convenience? A validation study. BMC Oral Health 2024, 24, 1517. [Google Scholar] [CrossRef] [Scilit] [PubMed] [PubMed Central]
- Cooper, H.; Gray, T.; Fronek, L.; Witfill, K. Lip Augmentation with Hyaluronic Acid Fillers: A Review of Considerations and Techniques. J. Drugs Dermatol. 2023, 22, 23–29. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Revilla-León, M.; Kois, D.E.; Kois, J.C. A guide for maximizing the accuracy of intraoral digital scans. Part 1: Operator factors. J. Esthet. Restor. Dent. 2023, 35, 230–240. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- D’Ettorre, G.; Farronato, M.; Candida, E.; Quinzi, V.; Grippaudo, C. A comparison between stereophotogrammetry and smartphone structured light technology for three-dimensional face scanning. Angle Orthod. 2022, 92, 358–363. [Google Scholar] [CrossRef] [Scilit] [PubMed] [PubMed Central]
- Nehru, A.; Nagesh, S. Smartphone-based 3D facial scanning applications in orthodontic diagnosis: A cross-sectional study. J. Dent. Res. Dent. Clin. Dent. Prospect. 2025, 19, 97–103. [Google Scholar] [CrossRef] [Scilit] [PubMed] [PubMed Central]
- Kim, J.S. Changes in Volume of Lips in 3-Dimensional Analysis and Projection of Lips in Sonography After Injection of Particle-Type Hyaluronic Acid Filler Utilizing a 9-Point Injection Technique. Aesthet. Surg. J. Open Forum 2024, 6, ojae076. [Google Scholar] [CrossRef] [Scilit] [PubMed] [PubMed Central]
- Hansson, S.; Östlund, E.; Bazargani, F. The Vectra M3 3-dimensional digital stereophotogrammetry system: A reliable technique for detecting chin asymmetry. Imaging Sci. Dent. Erratum in Imaging Sci. Dent. 2023, 53, 269. 2022, 53, 43–51. [Google Scholar] [CrossRef] [Scilit] [PubMed] [PubMed Central]
- de Menezes, M.; Rosati, R.; Ferrario, V.F.; Sforza, C. Accuracy and reproducibility of a 3-dimensional stereophotogrammetric imaging system. J. Oral Maxillofac. Surg. 2010, 68, 2129–2135. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Germani, M.; Miranda de Souza Almeida, C.C.; Muñoz-Lora, V.R.M. Comparison of 2 Fillers for Lip Injection-A Randomized-Controlled Clinical Trial Assessed by 3D Imaging. Aesthetic Surg. J. Open Forum 2024, 6, ojae003. [Google Scholar] [CrossRef] [Scilit] [PubMed] [PubMed Central]
- Cotofana, S.; Pretterklieber, B.; Lucius, R.; Frank, K.; Haas, M.; Schenck, T.L.; Gleiser, C.; Weyers, I.; Wedel, T.; Pretterklieber, M. Distribution Pattern of the Superior and Inferior Labial Arteries: Impact for Safe Upper and Lower Lip Augmentation Procedures. Plast. Reconstr. Surg. 2017, 139, 1075–1082. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Sidequersky, F.V.; Mapelli, A.; Annoni, I.; Zago, M.; De Felício, C.M.; Sforza, C. Three-dimensional motion analysis of facial movement during verbal and nonverbal expressions in healthy subjects. Clin. Anat. 2016, 29, 991–997. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Wampfler, J.J.; Gkantidis, N. Superimposition of serial 3-dimensional facial photographs to assess changes over time: A systematic review. Am. J. Orthod. Dentofac. Orthop. 2022, 161, 182–197.e2. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Schwendicke, F.; Samek, W.; Krois, J. Artificial Intelligence in Dentistry: Chances and Challenges. J. Dent. Res. 2020, 99, 769–774. [Google Scholar] [CrossRef] [Scilit] [PubMed] [PubMed Central]
- Muñoz, G.; Brito, L.; Olate, S. Photographic Parameters in Three-Dimensional Facial Image Acquisition. A Scoping Rev. J. Craniofacial Surg. 2024, 35, e376–e380. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Dutton, E.; Ludlow, M.; Mennito, A.; Kelly, A.; Evans, Z.; Culp, A.; Kessler, R.; Renne, W. The effect different substrates have on the trueness and precision of eight different intraoral scanners. J. Esthet. Restor. Dent. 2020, 32, 204–218. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Revilla-León, M.; Jiang, P.; Sadeghpour, M.; Piedra-Cascón, W.; Zandinejad, A.; Özcan, M.; Krishnamurthy, V.R. Intraoral digital scans: Part 2-influence of ambient scanning light conditions on the mesh quality of different intraoral scanners. J. Prosthet. Dent. 2020, 124, 575–580. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Son, K.; Jin, M.U.; Lee, K.B. Feasibility of using an intraoral scanner for a complete arch digital scan, part 2: A comparison of scan strategies. J. Prosthet. Dent. 2023, 129, 341–349. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Amornvit, P.; Sanohkan, S. The Accuracy of Digital Face Scans Obtained from 3D Scanners: An In Vitro Study. Int. J. Environ. Res. Public Health 2019, 16, 5061. [Google Scholar] [CrossRef] [Scilit] [PubMed] [PubMed Central]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |