Abstract
The evaluation of peripheral pulmonary lesions (PPLs) is a persistent clinical challenge, made more pressing by the expansion of low-dose CT lung cancer screening. Conventional and earlier navigational bronchoscopy, including electromagnetic navigation, historically achieved diagnostic yields of approximately 60–70%, with CT-to-body divergence contributing to nondiagnostic sampling. This narrative review, prepared using the Scale for the Assessment of Narrative Review Articles (SANRA), summarizes robotic-assisted bronchoscopy (RAB) platform technology, procedural optimization, diagnostic performance, safety, molecular tissue adequacy, and emerging applications. A structured search was updated through September 2026; this is not a systematic review. Using the strict definition recommended by the American Thoracic Society/American College of Chest Physicians (ATS/ACCP), the most comprehensive meta-analysis reported a pooled RAB diagnostic yield of 69.6%, while the prospective multicenter TARGET study reported an adjudicated strict yield of 61.6%. Higher pooled estimates of approximately 84–87% generally use intermediate definitions that the ATS/ACCP statement recommends against as primary diagnostic-yield measures. RAB pneumothorax rates are approximately 2% in pooled analyses, compared with approximately 20–25% for CT-guided transthoracic needle biopsy. In the randomized RELIANT trial, RAB was non-inferior, but not superior, to electromagnetic navigation bronchoscopy; in VERITAS, navigational bronchoscopy as a category was non-inferior to transthoracic needle biopsy for 12-month diagnostic accuracy, with fewer pneumothoraces. RAB is a useful option for selected PPLs, but strict-definition performance is more modest than frequently quoted intermediate yields. Outcomes depend on lesion characteristics, sampling strategy, operator experience, center volume, and imaging infrastructure.
Keywords:
robotic-assisted bronchoscopy; peripheral pulmonary lesions; lung nodule; Ion endoluminal system; monarch platform; galaxy system; cone-beam CT; diagnostic yield; cryobiopsy; CT-to-body divergence; navigational bronchoscopy; lung cancer screening; transbronchial ablation; molecular profiling; interventional pulmonology 1. Introduction
Robotic-assisted bronchoscopy (RAB) has emerged as an important advancement in the diagnostic evaluation of peripheral pulmonary lesions (PPLs), addressing fundamental limitations of conventional bronchoscopy and earlier navigational platforms. The timely and accurate tissue diagnosis of PPLs is central to lung cancer management, particularly following the 2021 US Preventive Services Task Force expansion of low-dose CT screening eligibility, which lowered the age threshold to 50 years and the qualifying smoking history to 20 pack-years, substantially increasing the estimated eligible screening population [1].
Historically, the diagnostic evaluation of peripheral lung nodules relied on two primary modalities. CT-guided transthoracic needle biopsy (TTNB) achieves diagnostic yields of 80–90% for accessible peripheral lesions but carries pneumothorax rates of 20–25% and chest-tube requirements of 7–8% [2]. Conventional flexible bronchoscopy offers a favorable safety profile but has limited peripheral yield; pooled and meta-analytic data show substantially reduced yields for small (<2 cm) peripheral lesions [3,4], and the AQuIRE registry reported a yield of 63.7% for nonguided bronchoscopy [5]. Electromagnetic navigation bronchoscopy (ENB) improved peripheral navigational access, yet data from the prospective NAVIGATE study demonstrated a diagnostic yield of only 67.8% at 24 months [6]. This ceiling has been attributed primarily to CT-to-body divergence (CTBD), the systematic displacement of lesion position between the static preprocedural CT scan and the dynamic intraoperative pulmonary environment [7].
Although PPLs are the principal focus of this review, small or anatomically difficult central and mid-zone lesions may also be challenging and are not categorically excluded; lesion size and accessibility, rather than location alone, determine procedural difficulty.
RAB integrates robotic catheter control with advanced navigation technologies—shape-sensing fiber-optic localization or electromagnetic guidance—and compatibility with intraprocedural imaging adjuncts, principally cone-beam computed tomography (CBCT), to improve lesion localization and procedural stability. Early prospective studies and meta-analyses suggest improved diagnostic performance compared with conventional bronchoscopic techniques [8,9,10,11,12,13] and a favorable safety profile compared with transthoracic approaches, particularly in patients at elevated pneumothorax risk. However, outcomes vary meaningfully with lesion characteristics, operator experience, institutional volume, and availability of adjunctive imaging. Rigorous interpretation of the evidence therefore requires careful attention to diagnostic-yield definitions, study design, and procedural heterogeneity [8,9,10].
This review provides an evidence-based synthesis of RAB encompassing platform technology, preprocedural planning, diagnostic performance, adjunctive imaging strategies, biopsy optimization, safety, comparative effectiveness, advanced applications, training, cost-effectiveness, and future directions. The literature search strategy, inclusion criteria, and synthesis approach are described in Section 2. As lung cancer screening expands, optimizing minimally invasive diagnostic strategies for peripheral lesions remains a priority.
2. Methods: Literature Search Strategy and Study Selection
This narrative review was prepared using the Scale for the Assessment of Narrative Review Articles (SANRA; checklist in Supplementary Table S1), which supports transparent reporting and critical appraisal of narrative reviews but does not replace systematic-review methods or formal risk-of-bias assessment. No protocol was registered because this is a narrative rather than a systematic review. The search strategy, inclusion criteria, and synthesis approach were specified before literature retrieval, and the search was updated to include relevant publications available through September 2026.
Because this narrative review does not apply formal evidence grading, the patient-selection and ventilatory-strategy tables in Section 5 deliberately avoid recommendation grading. Instead, they separate the clinical scenario or procedural strategy from the certainty of the supporting evidence, described qualitatively as moderate, low, or very low according to study design and directness. Most supporting evidence is observational or indirect, and entries resting on expert consensus or mechanistic reasoning are identified as such. These certainty descriptors reflect confidence in the underlying evidence only and do not constitute graded recommendation.
2.1. Search Strategy
A structured literature search was conducted across four electronic databases—PubMed/MEDLINE, Embase, Web of Science, and the Cochrane Library—for articles published from January 2018 (the year the first RAB platform received FDA clearance) through September 2026. Controlled-vocabulary terms (MeSH/Emtree) and free-text keywords were combined using Boolean operators (AND/OR). Primary terms included “robotic bronchoscopy,” “robotic-assisted bronchoscopy,” “Ion endoluminal system,” “Monarch platform,” “Galaxy system,” “navigational bronchoscopy,” and “shape-sensing bronchoscopy.” These were combined with secondary terms: “peripheral pulmonary lesion,” “pulmonary nodule,” “diagnostic yield,” “cone-beam CT,” “CBCT,” “CT-to-body divergence,” “cryobiopsy,” “transbronchial cryobiopsy,” “next-generation sequencing,” “molecular profiling,” “lung cancer screening,” “transbronchial ablation,” “learning curve,” and “cost-effectiveness.” A supplementary search of ClinicalTrials.gov and the WHO International Clinical Trials Registry Platform (ICTRP) identified ongoing and recently completed trials. Conference abstracts from the American Thoracic Society (ATS; 2022–2025), CHEST (2022–2025), the European Respiratory Society (ERS; 2022–2025), and the American Association for Bronchology and Interventional Pulmonology (AABIP; 2022–2025) were manually screened to capture relevant data available only in abstract form, including the ERS 2025 randomized controlled trial by Steinack and colleagues and the October 2025 FDA clearance of the Ion artificial intelligence navigational system.
2.2. Inclusion and Exclusion Criteria
Studies were included if they evaluated robotic-assisted bronchoscopy platforms (Ion, Monarch, or Galaxy) in human subjects and reported diagnostic yield, safety outcomes, molecular tissue adequacy, or clinical effectiveness endpoints. Conference abstracts presented at major society meetings were included when no full-text publication was available and the data materially contributed to the evidence synthesis; such abstracts are identified as non-peer-reviewed where cited. Historical context regarding the evolution of navigational bronchoscopy (ENB, radial endobronchial ultrasound [R-EBUS], and conventional bronchoscopy) was drawn from landmark studies, meta-analyses, and registry datasets without restriction to the 2018–2026 window. Preclinical animal studies were excluded from the evidence synthesis but are cited where they provide the only available safety data for an emerging therapeutic application, such as transbronchial ablation. Cadaveric proof-of-concept studies were retained only when clearly identified and when they materially informed platform feasibility or comparative technical performance. Other exclusions were (1) case reports with fewer than five subjects, unless the reported application was novel with no other available human data, and (2) editorials or letters without original data.
2.3. Study Selection and Data Synthesis
Literature screening was performed by the author in two stages: title and abstract screening followed by full-text review of potentially eligible articles. A narrative synthesis was used rather than formal meta-analytic pooling; quantitative estimates from published meta-analyses are cited with attention to definition, design, and heterogeneity. Evidence included meta-analyses, prospective multicenter studies, randomized trials, registries, retrospective cohorts, conference abstracts, guidelines, consensus statements, and expert reviews. Because screening and study selection were conducted by a single reviewer without duplicate adjudication or formal risk-of-bias assessment, selection and confirmation bias remain possible. SANRA was used to improve reporting quality, not to imply systematic-review-level bias control.
3. Historical Context: The Evolution of Peripheral Bronchoscopy
Throughout this review, “conventional (nonguided) flexible bronchoscopy” refers to flexible bronchoscopy performed with standard-diameter scopes under fluoroscopic guidance only, without electromagnetic or shape-sensing navigation. This is explicitly distinct from thin and ultrathin bronchoscopy and from radial-EBUS–assisted sampling, each of which is treated as a separate modality where cited. Unless otherwise specified, all subsequent references to “conventional bronchoscopy” carry this definition.
Early bronchoscopy, pioneered by Gustav Killian in 1897 for foreign body retrieval, used rigid instruments that provided central airway access only (Figure 1). The development of flexible fiberoptic bronchoscopy by Shigeto Ikeda in 1966 extended reach into segmental and subsegmental airways, enabling bronchial washings, brushings, and transbronchial forceps biopsies. However, diagnostic yield for subcentimeter or peripheral lesions remained severely limited by the inability to navigate beyond the bronchoscopist’s direct field of view.
Figure 1.
Timeline of bronchoscopy technological evolution: from Killian’s 1897 rigid bronchoscopy through flexible fiberoptic bronchoscopy (Ikeda, 1966), radial endobronchial ultrasound (1992), electromagnetic navigation bronchoscopy (2004) and the NAVIGATE 24-month results [6], FDA clearance of Monarch and Ion (2018–2019) [14,15], the Galaxy System with integrated Tool-in-Lesion Tomosynthesis (TiLT) (2023) [16], and FDA clearance of AI-integrated Ion navigation and publication of the VERITAS trial (2025) [17,18]. The ascending line is a schematic depiction of technological evolution only; vertical position does not represent diagnostic yield, and the figure does not establish monotonic improvement in yield across eras. AI = artificial intelligence; ENB = electromagnetic navigation bronchoscopy; R-EBUS = radial endobronchial ultrasound.
Fluoroscopy-guided transbronchial biopsy was the first attempt to extend diagnostic bronchoscopy to radiographically visible but endoscopically inaccessible lesions. Systematic reviews demonstrated reasonable sensitivity (~88%) for central lesions but only ~34% yield for peripheral lesions under 2 cm [19]. R-EBUS, first described in 1992, added real-time confirmation of probe position within or adjacent to peripheral lesions, modestly improving the pooled yield to approximately 71% [4].
Electromagnetic navigation bronchoscopy (ENB), commercialized by the SuperDimension platform in 2004, was the watershed technology bringing navigation-assisted peripheral bronchoscopy into mainstream practice. ENB uses an electromagnetic field generator to track a sensor probe within a registered CT-derived airway map. Real-world performance was constrained by CTBD: Pritchett and colleagues characterized this displacement as typically 5–15 mm, sufficient to cause the navigation system to confirm position within the lesion on its virtual map while the biopsy tool is displaced from the actual lesion in real space [7]. The NAVIGATE trial, enrolling 1215 patients across 37 centers, reported a 24-month diagnostic yield of only 67.8% [6].
Ultrathin bronchoscopes (outer diameter 2.8 mm) improved access to more distal airways but did not address CTBD. The field recognized that a step-change was needed—one providing structural stability, precise distal articulation, real-time lesion localization beyond navigation mapping alone, and reliable tissue acquisition. This recognition set the stage for the introduction of robotic platforms in the late 2010s.
4. Robotic-Assisted Bronchoscopy Platforms: Technology and Engineering
Three robotic-assisted bronchoscopy systems are currently FDA-cleared and in clinical use (Table 1). While all three share the overarching goal of improving navigational precision and catheter stability for PPL biopsy, they differ substantially in technical architecture, navigation methodology, scope design, and imaging integration. Recent reviews summarize these platform-level distinctions and evolving clinical applications [20].
Table 1.
Comparative features of FDA-cleared robotic-assisted bronchoscopy platforms.
4.1. Ion™ Endoluminal System (Intuitive Surgical)
The Ion endoluminal system, FDA-cleared in February 2019, is widely deployed in clinical practice [14]. Ion’s most distinctive feature is its proprietary shape-sensing technology: a fiber-optic network with Bragg gratings embedded within the catheter wall that measures the catheter’s three-dimensional shape and precise position hundreds of times per second, enabling real-time location awareness entirely independent of electromagnetic fields. This eliminates interference from metallic equipment and obviates the need for external electromagnetic field generators.
The Ion catheter has a 3.5 mm outer diameter, enabling navigation to distal airways, commonly around the 6th–7th generation, though reachable depth varies with airway caliber and patient anatomy and more distal access is achievable. The fully articulating distal tip provides 180-degree deflection in any plane. All three platforms are additionally compatible with R-EBUS, fluoroscopy, CBCT or digital tomosynthesis, and the 1.1 mm cryoprobe, and the Ion catheter provides notable positional stability during sampling. Pre-procedural planning uses PlanPoint software, which automatically segments the target nodule from CT imaging and generates navigation pathways with vascular proxies defining anatomic danger zones. In October 2025, Intuitive received FDA clearance for AI integration across Ion’s complete navigational workflow—the first AI-embedded bronchoscopy navigation system—a development expected to further reduce learning curve variability [18].
4.2. Monarch™ Platform (Johnson & Johnson/Auris Health)
The Monarch Platform, FDA-cleared in March 2018, was the first RAB system to achieve commercial availability [15]. Monarch combines electromagnetic navigation with optical and robotic-kinematic position data and uses a distinctive telescoping dual-scope design: an outer guiding sheath (6.0 mm outer diameter) and an inner steerable bronchoscope (4.2 mm outer diameter) with a built-in camera, allowing continuous white-light visualization of the peripheral airway during navigation and biopsy. The BENEFIT feasibility study (n = 54) reported R-EBUS-confirmed lesion localization in 96.2% of cases [15]. The TARGET trial was a prospective, multicenter, single-arm study of 679 patients (median lesion size 1.85 cm) that reported an adjudicated strict diagnostic yield of 61.6% using ATS/ACCP criteria, a sensitivity for malignancy of 78.8%, and an adverse-event rate of 3.8% [21].
4.3. Galaxy System™ (Noah Medical)
The Galaxy System, FDA-cleared in March 2023, is the newest entrant and incorporates several distinctive innovations [16]. Its most notable feature is integrated Tool-in-Lesion Tomosynthesis (TiLT+), a proprietary digital tomosynthesis system that provides intraoperative three-dimensional imaging for real-time confirmation of biopsy-tool position within the target lesion, addressing CTBD without requiring a separate CBCT unit. Galaxy uses electromagnetic guidance to reach within approximately 2 cm of the lesion and then activates TiLT for final localization with an augmented fluoroscopy overlay. The Galaxy bronchoscope is single-use and disposable, eliminating reprocessing logistics; comparative per-case cost data remain limited. The first-in-human FRONTIER study (19 nodules in 18 patients; mean lesion size 2.0 cm) reported 100% TiLT-confirmed tool-in-lesion localization, with diagnostic yields of 89.5% (strict) and 94.7% (intermediate) [16]. A multicenter trial (NCT06056128) is ongoing.
5. Preprocedural Planning, Patient Selection, and Procedural Setup
5.1. CT Imaging Requirements
Optimal RAB outcomes require high-quality thin-slice (≤1 mm collimation) chest CT in breath-hold. Thicker reconstructions significantly impair automated airway segmentation accuracy. The CT is imported into platform-specific planning software (PlanPoint for Ion; Monarch Planning Software; Galaxy Planning Software), which segments the bronchial tree from the trachea through sub-subsegmental airways. Key planning parameters include: depth of airway pathway (generations to target), bronchus sign presence, lesion-to-pleura distance, vascular proximity, and anticipated need for CBCT adjunct. The bronchus sign—defined as a bronchus leading directly into or through the lesion on CT—remains an important predictor of diagnostic success, though its impact is attenuated by CBCT-guided tool-in-lesion confirmation [8,9].
5.2. Patient Selection
Patient selection for RAB should be individualized based on lesion characteristics, patient risk profile, availability of institutional expertise, and whether the clinical question requires tissue for diagnosis, molecular profiling, or both. Table 2 summarizes evidence-based indications, preferred scenarios, and contraindications in full detail. The following provides a structured clinical framework.
Table 2.
Clinical factors influencing patient selection for robotic-assisted bronchoscopy.
5.2.1. Primary Indications
RAB may be considered for (1) PPLs, most commonly in the outer two-thirds of the lung though not categorically limited to this region, where conventional flexible bronchoscopy has limited diagnostic yield (14% for peripheral lesions <2 cm; 31% for inner two-thirds lesions [22]); (2) intermediate- to high-probability pulmonary nodules, defined by Lung-RADS category 4A/4B [26] or by an intermediate-to-high malignancy probability on the Mayo model (5–65%), that require tissue diagnosis per American College of Chest Physicians (ACCP) and National Comprehensive Cancer Network (NCCN) guidelines [24,25]; (3) suspected or confirmed non-small cell lung cancer (NSCLC) requiring comprehensive molecular profiling, including a next-generation sequencing (NGS) panel, programmed death-ligand 1 (PD-L1) immunohistochemistry (IHC), ALK/ROS1 fluorescence in situ hybridization (FISH), and tumor mutational burden (TMB), particularly when liquid biopsy is insufficient or preserved tissue architecture is advantageous for IHC interpretation [27,28]; and (4) peripheral lesions with concurrent fluorodeoxyglucose (FDG)-avid mediastinal or hilar lymphadenopathy, where combined peripheral biopsy and endobronchial ultrasound-guided transbronchial needle aspiration (EBUS-TBNA) staging in a single procedure is clinically appropriate per ACCP/NCCN guidelines [24,25], including selected station 5/6 approaches reported with robotic bronchoscopy [29,30,31].
5.2.2. Potential Advantages over CT-Guided Transthoracic Needle Biopsy
RAB may be favored over CT-TTNB in selected patients when the transthoracic approach carries elevated risk or is technically limited: (1) emphysema or bullous lung disease, in which CT-TTNB pneumothorax risk is 20–25% overall and substantially higher when emphysematous parenchyma overlies the biopsy path [2], whereas RAB avoids pleural traversal and has a pooled pneumothorax rate of approximately 2.0% [8,10]; (2) a single functional lung or severely impaired contralateral pulmonary reserve, in which a pneumothorax could have major clinical consequences; (3) thrombocytopenia (platelet count <50,000/µL), coagulopathy (international normalized ratio [INR] >1.5), or therapeutic anticoagulation that cannot be safely interrupted, in which TTNB hemorrhage risk is elevated [2]; (4) deep-seated lesions requiring long parenchymal traversal (>3 cm), vascular proximity, or prior ipsilateral pleurodesis limiting lung mobility [2]; and (5) bilateral synchronous nodules requiring same-session evaluation, for which bilateral percutaneous sampling is generally avoided because of bilateral pneumothorax risk [32,33].
5.2.3. Lung Cancer Screening Population
Low-dose CT (LDCT)-detected nodules classified as Lung-RADS 4A or 4B [26] represent a growing indication for RAB. The 2021 US Preventive Services Task Force (USPSTF) expansion of screening eligibility to age 50 years and 20 pack-years substantially increased the eligible population [1], generating a large volume of screen-detected nodules that require risk stratification and, in many cases, tissue evaluation. Pooled RAB diagnostic yield is approximately 78% for lesions <2 cm, the size range of many LDCT-detected nodules, although this estimate uses non-strict definitions [9]. The favorable safety profile of RAB may be particularly relevant in this population of current and former smokers, in whom emphysema is common.
5.2.4. Relative and Absolute Contraindications
Relative contraindications include very small lesions (<6–8 mm), which are lower-yield but not automatically excluded; CBCT-assisted series have reported yields of approximately 66.6% for lesions ≤10 mm, and the decision should depend on whether a biopsy result would change management [36,37,38]. Pure ground-glass opacities may have lower cellular yield, and Fleischner Society or Lung-RADS [26] surveillance recommendations should guide whether tissue sampling is indicated. Inability to tolerate general anesthesia is another relative contraindication, although moderate sedation may be considered for selected cases with favorable anatomy at experienced centers. Absolute contraindications include active bronchospasm or status asthmaticus and hemodynamic instability or acute respiratory failure requiring escalating support. Uncorrectable severe coagulopathy contraindicates cryobiopsy specifically; standard fine-needle aspiration (FNA) or forceps biopsy may remain feasible after appropriate correction.
5.3. Anesthetic and Ventilatory Optimization
RAB is generally performed under general anesthesia with endotracheal intubation and neuromuscular blockade at high-volume centers, distinguishing it from conventional bronchoscopy under moderate sedation. General anesthesia provides airway control and limits patient movement during navigation, allows controlled ventilation and PEEP to reduce intraoperative atelectasis, and accommodates the 45–90 min duration of complex cases [41]. The I-LOCATE trial established that significant atelectasis develops as early as 30 min post-intubation, a critical time benchmark for procedural planning [42]. Comprehensive ventilatory strategies to optimize RAB outcomes are detailed in Table 3.
Table 3.
Anesthetic and ventilatory strategies to optimize procedural success in robotic-assisted bronchoscopy.
6. Diagnostic Yield: Evidence Base and Key Determinants
Diagnostic yield is the primary outcome metric for PPL bronchoscopy, but comparisons require attention to the definition used. The evidence below therefore focuses on RAB meta-analyses, prospective trials, adjunctive imaging, and sampling technique; the historical ~70% ceiling and the CTBD mechanism are summarized in Section 3.
6.1. Meta-Analytic Evidence for RAB
The most defensible headline estimates are those using the strict definition recommended by the American Thoracic Society (ATS)/ACCP (Figure 2). Li and colleagues [10] reported a pooled strict RAB diagnostic yield of 69.6% (95% CI 61.8–76.8%) across 27 cohort studies, with pneumothorax in 2.0% and chest-tube placement in 0.5%. TARGET, the largest prospective multicenter single-platform RAB study, reported an adjudicated strict yield of 61.6% [21]. Higher estimates should be interpreted according to their definitions: Ali and colleagues [8] reported an intermediate yield of 84.3% (95% CI 81.1–87.2%), and Li and colleagues reported 86.6% using an intermediate definition [10]. The ATS/ACCP statement recommends against intermediate and liberal definitions because they incorporate subsequent clinical information or follow-up into the index-test result [45]. Zhang and colleagues [9] reported a pooled yield of 80.4%, with higher yields in studies using cryobiopsy, but heterogeneity in study design and yield definitions limits cross-study comparison. A 2026 clinical review in the New England Journal of Medicine similarly noted that TARGET’s approximately 62% strict yield was not clearly better than that of other guided bronchoscopy techniques and characterized the evidence for real-time localization adjuncts as mixed [46]. Key prospective studies, randomized trials, and meta-analyses are summarized in Table 4, and additional prospective shape-sensing RAB series are summarized in Table 5.
Figure 2.
Diagnostic yield by bronchoscopic modality across selected registries, meta-analyses, and trials [4,5,6,8,9,10,40]. Values are not directly comparable because study design, lesion characteristics, and yield definitions differ. The ATS/ACCP statement recommends a strict definition and recommends against intermediate or liberal definitions as primary yield measures [45]. The ERS 2025 RAB + CBCT result is preliminary, non-peer-reviewed conference-abstract evidence and is shown in a separate color. ATS/ACCP = American Thoracic Society/American College of Chest Physicians; CBCT = cone-beam CT; ENB = electromagnetic navigation bronchoscopy; ERS = European Respiratory Society; RAB = robotic-assisted bronchoscopy; R-EBUS = radial endobronchial ultrasound.
Table 4.
Selected studies relevant to robotic-assisted bronchoscopy, including historical comparators, prospective studies, randomized controlled trials, and meta-analyses cited in this review.
Table 5.
Additional shape-sensing robotic-assisted bronchoscopy (ssRAB) series for peripheral pulmonary lesions, identified from expert consensus documents [34,35] and subsequent publications.
6.2. Pivotal Prospective Trials
Among platform-specific prospective data, TARGET [21] (described in Section 4.2) reported an adjudicated strict diagnostic yield of 61.6% using ATS/ACCP criteria in 679 patients. The single-center RELIANT cluster-randomized non-inferiority trial [49] compared RAB directly with digital-tomosynthesis-corrected ENB in 411 patients; diagnostic yield, defined as the proportion of cases yielding lesional tissue, was 77.8% with RAB and 75.5% with ENB, meeting the prespecified non-inferiority criterion (p = 0.007) without demonstrating superiority. VERITAS [17], a multicenter randomized non-inferiority trial of navigational bronchoscopy broadly rather than RAB alone, enrolled 234 patients with nodules 10–30 mm and reported 12-month diagnostic accuracy of 79.0% with navigational bronchoscopy versus 73.6% with CT-guided TTNB. Pneumothorax occurred in 3.3% versus 28.3%, and chest-tube placement or hospital admission in 0.8% versus 11.5%, respectively. These trials support bronchoscopic-first strategies in appropriately selected patients but do not establish superiority of RAB over ENB or CT-guided biopsy.
6.3. Diagnostic Yield Definitions: ATS/ACCP 2024 Consensus
Diagnostic yield is the proportion of bronchoscopic procedures that produce a specific diagnosis from the index procedure under a prespecified definition. Diagnostic accuracy additionally requires verification against a reference standard over follow-up, as in VERITAS. The 2024 ATS/ACCP Research Statement [45] recommends a strict diagnostic-yield definition based on a specific malignant or benign diagnosis established by the bronchoscopic sample and recommends against so-called intermediate or liberal definitions, which can incorporate follow-up or management decisions into the index-test result. Accordingly, this review treats strict yield as the primary benchmark and presents intermediate values only as secondary historical estimates. TARGET’s adjudicated strict yield was 61.6% [21].
6.4. Key Predictors of Diagnostic Success
Predictors of higher RAB diagnostic yield identified across multiple studies and meta-analyses include lesion size greater than 20 mm, presence of a CT bronchus sign, a concentric R-EBUS view confirming tool position within the lesion, CBCT-confirmed tool-in-lesion, and center or operator experience [8,9,10,59]. Predictors of lower yield include pure ground-glass or subsolid density, absence of a bronchus sign, upper-lobe location, lesion depth beyond the sixth airway generation, and lesion size below 10 mm [8,9,10]. CBCT-assisted Ion studies have reported yields of approximately 66.6% for lesions ≤10 mm [36,37,38].
7. Adjunctive Imaging: Overcoming CT-to-Body Divergence
CT-to-body divergence (CTBD) represents a fundamental limitation of navigation-based bronchoscopy, reflecting the spatial discrepancy between pre-procedural CT imaging—acquired in a breath-holding, awake, supine patient—and intraoperative lung anatomy under general anesthesia with controlled mechanical ventilation. Factors contributing to CTBD include atelectasis onset (significant atelectasis develops within approximately 30 min of intubation per the I-LOCATE trial [42]), gravitational redistribution of lung tissue, altered lung volumes under positive pressure ventilation, and diaphragmatic displacement. Resulting lesion displacement is typically 5–15 mm in magnitude per Pritchett and colleagues [7], sufficient to cause systematic biopsy of peritumoral or atelectatic tissue while the actual lesion remains unsampled, even when the navigation system confirms apparent tool-in-lesion on the virtual map. CTBD is the primary explanatory variable for the persistent gap between navigational success rates (often >95%) and actual diagnostic tissue acquisition in ENB and early RAB series.
7.1. Cone-Beam CT: Fixed and Mobile Systems
CBCT provides intraoperative three-dimensional volumetric imaging that can show biopsy-tool position relative to the lesion independently of the preprocedural CT-derived navigation map (Figure 3). Both fixed and mobile systems, including the Cios Spin (Siemens Healthineers, Erlangen, Germany), have been evaluated. In the prospective Bashour study [36] (n = 67; lesions 9–30 mm), tool-in-lesion confirmation increased from 34.3% before mobile CBCT to 98.6% after imaging and repositioning (absolute improvement 64.3 percentage points; p < 0.0001). This finding demonstrates improved localization but should not automatically be interpreted as an equivalent increase in diagnostic yield. Evidence for a yield benefit is mixed and may depend on platform, workflow, operator experience, and case selection. In a 2026 retrospective Monarch study, strict diagnostic yield was 70.9% without mobile CBCT and 71.7% with mobile CBCT (p = 0.976), while radiation dose increased from 7.4 to 285.9 mGy [50]. CBCT availability is also limited by capital cost, workflow complexity, radiation exposure, and infrastructure requirements.
Figure 3.
Schematic illustration of CTBD and mobile cone-beam CT correction during robotic-assisted bronchoscopy. In the study by Bashour and colleagues, tool-in-lesion confirmation increased from 34.3% to 98.6% after mobile CBCT imaging and repositioning [36]. CBCT = cone-beam CT; CTBD = CT-to-body divergence; GA = general anesthesia; mCBCT = mobile CBCT.
7.2. Randomized Evidence for RAB + CBCT
A randomized comparison of RAB + CBCT with ultrathin bronchoscopy under two-dimensional fluoroscopy was presented at the 2025 ERS Congress by Steinack and colleagues [40]. As of the September 2026 search update, these data remained available only as a non-peer-reviewed conference abstract. The abstract reported a diagnostic yield of 84.6% with RAB + CBCT versus 23.1% with the comparator (78 patients; 127 PPLs; median lesion size 11 mm; absolute difference 61.5 percentage points; 95% CI 44.1–78.9; p < 0.001). Overall, 85% of lesions lacked a CT bronchus sign and 27.6% were pure ground-glass opacities. Because the report is preliminary, single-center, and not peer-reviewed, it should not be weighted equivalently to published multicenter trials or used as a primary headline estimate.
7.3. Radial EBUS, Digital Tomosynthesis, and Augmented Fluoroscopy
Radial EBUS (R-EBUS) remains the most widely available intraoperative lesion-localization adjunct, providing real-time 360-degree cross-sectional ultrasound confirmation that the probe is within or adjacent to the target lesion. A concentric R-EBUS view is consistently associated with higher diagnostic yield across studies and meta-analyses [4]. Digital tomosynthesis (TiLT+ technology, Galaxy System; Illumisite, Medtronic, Minneapolis, MN, USA), offers an alternative approach to CTBD correction using standard C-arm fluoroscopic data, without the infrastructure requirements of CBCT [16]. Augmented fluoroscopy overlays the lesion’s intraoperative position onto real-time fluoroscopy and may permit tool repositioning. Multimodal localization is associated with high yields in experienced-center series, but the incremental diagnostic benefit of CBCT is not uniform across platforms or centers [36,37,38,50].
8. Biopsy Technique Optimization and Molecular Adequacy
8.1. Rationale for Cryobiopsy
Conventional biopsy tools passed through the 2.0 mm working channel of the Ion catheter, such as 21/22-gauge FNA needles and standard forceps, have inherent limitations (Table 6). FNA yields small, often fragmented cytological specimens that lack architectural detail, and forceps biopsies are susceptible to crush artifact and may inadequately sample eccentric lesions that lie adjacent to, rather than around, the airway. These tool limitations may contribute to nondiagnostic sampling. Cryobiopsy exploits the Joule–Thomson effect: rapid expansion of compressed gas at the probe tip cools it to approximately −79 °C with CO2 (or lower with N2O), freezing adjacent tissue within seconds. Unlike forceps, the 1.1 mm cryoprobe acquires tissue circumferentially, which is potentially advantageous for eccentric lesions that abut but do not surround the airway [27,60].
Table 6.
Biopsy modalities and sampling techniques in robotic-assisted bronchoscopy.
8.2. Clinical Evidence
Oberg and colleagues [27] retrospectively analyzed 112 patients with 120 PPLs sampled during Ion RAB with sequential FNA, forceps, and 1.1 mm cryoprobe biopsy. Overall yield was 90%, 18% of diagnoses were made exclusively by cryobiopsy, and all cryobiopsy samples submitted for molecular testing were adequate. A 2024 retrospective cohort comparing cryobiopsy with FNA also reported favorable cryobiopsy performance in small nodules [62]. Randomized evidence has since strengthened the broader bronchoscopic cryobiopsy literature. FROSTBITE-2 [63], a multicenter trial that randomized 500 patients with lung nodules or masses, lung transplants, or diffuse parenchymal lung disease, reported an overall diagnostic yield of 88.6% with the 1.1 mm cryoprobe versus 78.8% with forceps, and 83.2% versus 70.1% in the nodule-or-mass subgroup. FROSTBITE-2 did not identify a safety disadvantage with the cryoprobe: pneumothorax requiring chest-tube placement occurred in four patients in the forceps arm and none in the cryoprobe arm, and neither arm had grade 3–4 bleeding. In the PPL-specific Cryo-RCT [60], cryobiopsy improved histology-based yield overall (83% vs. 75%) and for adjacent or eccentric R-EBUS views (77% vs. 65%), but moderate bleeding (38% vs. 19%) and pneumothorax (2% vs. <1%) were more frequent. Safety findings therefore differ between the two trials. Neither trial was restricted to robotic bronchoscopy, and neither evaluated a complete RAB + CBCT strategy.
8.3. Molecular Adequacy for Next-Generation Sequencing
Contemporary management of lung adenocarcinoma requires not only a tissue diagnosis but also comprehensive molecular profiling, including broad-panel NGS, PD-L1 IHC, and companion diagnostic assays. Inadequate tissue quantity or quality may prevent patients from receiving effective targeted therapies. The American Association for Bronchology and Interventional Pulmonology–International Association for the Study of Lung Cancer (AABIP-IASLC) clinical practice guideline [64] concluded that guided bronchoscopy achieves tissue adequacy for comprehensive biomarker testing comparable to that of CT-guided percutaneous biopsy (80–100% vs. 72.3–96%). Whether adding CBCT and cryobiopsy to RAB further improves adequacy has not yet been established.
In a Mayo Clinic study, Yu Lee-Mateus and colleagues [61] found that 72% of 78 RAB samples met preanalytic criteria for molecular or PD-L1 testing; NGS adequacy was 48.6% for smears, 61.4% for touch preparations, and 29.2% for biopsies. In a Memorial Sloan Kettering series, Connolly and colleagues [28] reported that 84% of 128 RAB samples were adequate for molecular testing, with success rates of 96% (25/26) for hybrid-capture NGS, 94% (49/52) for polymerase chain reaction (PCR)-based testing, and 91% for PD-L1 IHC. Many experienced centers therefore use multimodal sampling, combining FNA, forceps biopsy, and cryobiopsy, with rapid on-site evaluation (ROSE), to maximize diagnostic yield and molecular adequacy, although no combination has been shown to be superior in a randomized trial.
9. Safety Profile and Complication Management
RAB’s safety profile is an important clinical advantage relative to CT-guided TTNB (Table 7) [8,10,21]. Across meta-analyses, pooled pneumothorax rates are approximately 2.0–2.3%, with chest-tube placement in 0.5–1.2%, compared with approximately 20–25% pneumothorax and 7–8% chest-tube placement in broader TTNB datasets [2]. In VERITAS, the randomized TTNB pneumothorax rate was higher at 28.3%, with chest-tube placement or admission in 11.5%, compared with 3.3% and 0.8%, respectively, after navigational bronchoscopy [17]. The overall pooled RAB complication rate in the 2025 meta-analysis was 3.0% [10], and TARGET reported a 3.8% adverse-event rate with no procedure-related deaths [21]. Cryobiopsy can increase tissue yield but may carry a bleeding and pneumothorax trade-off, as observed in Cryo-RCT [60].
Table 7.
Complications and safety profile of robotic-assisted bronchoscopy.
Pneumothorax risk is increased by emphysematous lung parenchyma, upper lobe location, pleural proximity, multiple biopsy passes, and periprocedural movement. In patients with severe emphysema—precisely those at highest TTNB risk—RAB provides a compelling alternative enabling tissue acquisition with acceptable complication probability. The emerging literature suggests CT-guided TTNB may carry a risk of pleural track tumor seeding in confirmed malignancies—a theoretical risk not associated with airway-based bronchoscopic approaches. Rare complications have also been reported and warrant awareness, including pneumatocele formation following RAB [65], pneumomediastinum after transbronchial cryobiopsy [66], and systemic or cerebral arterial air embolism complicating robotic-guided lung biopsy [67]; although individually uncommon and described in case reports, these underscore the importance of meticulous technique and vigilance for air embolism under positive-pressure ventilation.
Radiation exposure from CBCT depends on the equipment, number of spins, acquisition protocol, and procedural workflow. Dose-optimization protocols should be used, and spins should be limited to those needed for localization or repositioning. Radiation should not be characterized as uniformly modest: in a 2026 Monarch comparison, median radiation dose was 285.9 mGy with mobile CBCT versus 7.4 mGy without it, approximately a 40-fold difference, without a significant diagnostic-yield improvement [50].
10. Comparative Effectiveness: RAB Versus Competing Modalities
10.1. RAB vs. CT-Guided Transthoracic Needle Biopsy
CT-guided TTNB achieves high diagnostic performance for accessible peripheral lesions but carries substantially greater pneumothorax risk. In VERITAS, a randomized non-inferiority trial [17], navigational bronchoscopy achieved 12-month diagnostic accuracy of 79.0% versus 73.6% with TTNB (absolute difference 5.4 percentage points; 95% CI −6.5 to 17.2; p = 0.003 for non-inferiority). Pneumothorax occurred in 3.3% versus 28.3%, and chest-tube placement or hospital admission occurred in 0.8% versus 11.5%. VERITAS evaluated the broader category of navigational bronchoscopy rather than RAB specifically and was conducted at expert centers, limiting direct extrapolation to every RAB program. A propensity score-matched study by Yu Lee-Mateus and colleagues [68] found that patients evaluated with RAB were more likely to receive an early-stage lung cancer diagnosis than those evaluated with TTNB (OR 3.02; 95% CI 1.83–5.04; p < 0.001), although this observational association should not be interpreted as proof that RAB caused stage migration. For subsolid nodules, Fernandez-Bussy and colleagues [69] reported comparable diagnostic yields with fewer complications in the RAB group.
10.2. RAB vs. Electromagnetic Navigation Bronchoscopy
In RELIANT [49], RAB was non-inferior but not superior to digital-tomosynthesis-corrected ENB (diagnostic yield, defined as lesional tissue obtained, 77.8% vs. 75.5%; Section 6.2). Subgroup findings, including those for smaller lesions and lesions without a CT bronchus sign, should be considered exploratory and should not be used to infer a causal advantage from robotic maneuverability. The imaging context also matters: the ENB comparator incorporated digital tomosynthesis correction, whereas CBCT images in the RAB arm were not transmitted to the navigation platform. A propensity score-matched analysis by McNierney and colleagues [70] found comparable yields between Monarch and Ion, suggesting that adjunctive imaging, sampling strategy, case selection, and operator experience may be at least as important as platform choice.
10.3. Clinical Integration and Modality Selection
Selection of the appropriate diagnostic modality for peripheral pulmonary lesions should be individualized based on lesion characteristics, patient risk profile, and institutional resources and expertise. CT-guided TTNB remains an effective and appropriate option for pleural-based or easily accessible peripheral lesions, particularly in settings where advanced bronchoscopic technologies are not available or where patient anatomy renders bronchoscopic access impractical. Continued imaging surveillance rather than biopsy is appropriate when a histologic result would not change management (e.g., very small or pure ground-glass lesions per Fleischner/Lung-RADS), and surgical or single-anesthetic diagnostic-resection pathways may be preferable for high-probability early-stage lesions in operable candidates.
Bronchoscopic approaches, including RAB, may be favored when (1) concurrent mediastinal staging is indicated, enabling diagnosis and staging in a single procedure per ACCP/NCCN guidelines [24,25]; (2) CT-guided biopsy carries elevated pneumothorax risk because of overlying emphysema, bullous disease, or deep parenchymal traversal [2]; (3) bilateral synchronous nodules require same-session evaluation [32,33]; (4) coagulopathy contraindicates the transthoracic approach; or (5) combined diagnosis and resection in a single anesthetic event is being considered [51].
In practice, RAB may be combined with radial EBUS, CBCT where available, and complementary biopsy tools such as FNA, forceps, and cryoprobe sampling. Experienced-center series using these multimodal workflows report high yields and molecular adequacy, but no randomized trial has established one combination as universally superior. Programs should select a reproducible strategy based on local imaging resources, pathology support, lesion characteristics, operator experience, and safety protocols.
11. Advanced and Emerging Applications
11.1. Bilateral Same-Session Procedures
RAB enables bilateral same-session lung-nodule sampling, accessing lesions in both lungs during one anesthetic event, a capability generally avoided with TTNB because of bilateral pneumothorax risk [32,33]. Chrissian and colleagues [32] described sampling multiple synchronous nodules during same-session RAB, and Fernandez-Bussy and colleagues [33] evaluated diagnosis of multiple pulmonary nodules, including procedures combined with mediastinal staging. These reports support technical feasibility, but comparative outcome data remain limited.
11.2. Aortopulmonary Window and Extended Mediastinal Access
Ion’s fully articulating distal tip can permit navigation into left upper-lobe subsegmental bronchi for CBCT-guided TBNA of the aortopulmonary window (station 5), a region typically inaccessible to standard linear EBUS because of aortic-arch interposition [29]. Published reports also describe robotic access to station 6 and other challenging mediastinal locations [29,30,31]. These reports suggest a potential extension of bronchoscopic nodal staging, although the evidence remains limited to case reports and early technical experience.
11.3. Single Anesthetic Diagnosis and Resection
The single-anesthetic bronchoscopy and resection (SABR) paradigm—in which RAB biopsy with intraoperative frozen pathology confirmation is immediately followed by robotic-assisted thoracoscopic resection in the same anesthetic event—has compelling time-to-treatment and cost implications. Weiser and colleagues [51] compared 40 SABR patients to 30 staggered-procedure controls: significant reduction in time from clinic to operating room, lower total costs, and comparable intraoperative and postoperative outcomes. This model eliminates the diagnostic-to-resection delay that adversely affects lung cancer survival and is particularly attractive for patients with small peripheral nodules at high clinical probability for early-stage malignancy.
11.4. Fiducial Marker Placement and Preoperative Localization
RAB provides a highly precise platform for intrapulmonary fiducial marker deployment to guide SBRT targeting or surgical localization. Benn and colleagues [71] described indocyanine green-soaked fiducial marker deployment via RAB prior to VATS for nodule localization—demonstrating precise and reliable airway-based marker placement with pneumothorax risk substantially below CT-guided percutaneous fiducial placement.
11.5. Transbronchial Ablation: Therapeutic Horizons
The convergence of precise robotic navigation with miniaturized energy delivery enables bronchoscopic therapeutic ablation of early-stage peripheral malignancies. Microwave ablation (MWA) via bronchoscopic delivery is the most advanced under investigation. De Leon and colleagues [72] demonstrated the safety of bronchoscopic MWA in swine peripheral lung using RAB; ongoing clinical trials evaluate this for curative-intent treatment of stage IA lung cancer in patients ineligible for surgery or SBRT, as well as bridging therapy. Diffusing alpha-emitter radiation therapy (DaRT), cryoablation, and vapor thermal ablation are additional modalities in early bronchoscopic therapeutic investigation. If validated, RAB could transform from a diagnostic tool into a comprehensive one-stop interventional hub for early-stage lung cancer.
12. Learning Curve, Training, and Competency
Skill acquisition in RAB varies among operators. In a learning-curve analysis from Memorial Sloan Kettering, Bott and colleagues [52] analyzed RAB procedures performed by 9 proceduralists (interventional pulmonologists and thoracic surgeons) over their first 50 cases each. Median procedure time decreased from 62 min during the first 10 cases to 39 min after case 40 (p < 0.001). Approximately half of the proceduralists achieved technical proficiency within 25 lesions, and operators increasingly targeted more challenging lesions after the initial phase.
The AABIP-IASLC guideline [64] recommends ongoing program-level quality monitoring, including strict diagnostic yield, molecular adequacy, complications, and time-to-treatment. Intuitive’s FDA clearance for AI-integrated navigation [18] may provide navigational decision support, but its effect on learning-curve variability and clinical outcomes remains unproven.
13. Cost-Effectiveness and Healthcare Value
A decision-analytic model [73] quantified the economic value of improved bronchoscopy sensitivity for malignancy, demonstrating that each percentage-point increase in diagnostic sensitivity generates substantial downstream health and cost savings by reducing non-diagnostic procedures, repeat invasive testing, and delayed treatment initiation—supporting the economic case for RAB even at higher per-procedure cost.
The comparative cost-effectiveness of RAB vs. TTNB is particularly favorable when accounting for downstream costs of pneumothorax management (disproportionately high in emphysematous patients), repeat procedures following non-diagnostic TTNB, and the system cost of delayed diagnosis and later-stage treatment. The SABR paradigm generates additional savings by eliminating a separate diagnostic admission and compressing the pre-treatment diagnostic interval [51]. Capital investment is substantial: approximately $500,000–$700,000 for the robotic platform and $250,000–$350,000 for mobile CBCT, although pricing varies by vendor, configuration, and site. Reimbursement under Medicare and most commercial payers is available under navigational bronchoscopy CPT codes when clinical indications are met. The downstream revenue generated by RAB programs—through medical oncology, thoracic surgery, radiation oncology, and molecular diagnostic referrals—substantially exceeds procedural reimbursement alone and should be incorporated into institutional business case analyses.
14. Future Directions and Emerging Technologies
14.1. Artificial Intelligence Integration
AI is poised to transform multiple RAB workflow facets. In October 2025, Intuitive received FDA clearance for AI integration across Ion’s complete navigational workflow—the first AI-embedded bronchoscopy navigation system [18]. A randomized controlled trial by Agbontaen and colleagues [74] demonstrated AI-guided bronchoscopy instruction superior to expert human instruction for critical care physicians—illustrating AI’s potential to reduce learning curve variability and democratize RAB training broadly. Future applications include automated CT airway segmentation, AI-assisted nodule risk stratification, real-time AI-guided navigation path optimization, intraoperative machine learning-based endobronchial lesion detection, and AI-augmented ROSE cytology interpretation.
14.2. Optical Biopsy: Needle-Based Confocal Laser Endomicroscopy
Needle-based confocal laser endomicroscopy (nCLE) represents an emerging optical biopsy technology providing real-time histologic information via fluorescence-based needle probes during RAB. Manley and colleagues [75] demonstrated that nCLE during RAB could identify diagnostic optical patterns correlating with malignancy, potentially enabling rapid on-site optical characterization before formal tissue sampling—a ‘real-time virtual biopsy’ capability that may further reduce non-diagnostic procedure rates.
14.3. Next-Generation Platform Development
Ongoing platform development targets: sub-3 mm outer diameter catheters for deeper airway access; integrated real-time CBCT without separate units; closed-loop robotic feedback for automated biopsy tool advancement; single-use disposable designs reducing per-case cost and sterilization logistics; and expanded therapeutic capabilities including miniaturized ablation systems, brachytherapy seed placement, and direct intratumoral immunotherapy delivery. The Galaxy System’s integrated TiLT represents a first step toward self-contained CTBD correction without separate CBCT infrastructure, which may eventually expand access to intraoperative localization beyond major academic centers.
14.4. Geographic Access and Equity
Realizing RAB’s population-level impact requires addressing geographic access inequities. RAB is currently concentrated at tertiary academic centers and large community cancer programs in metropolitan areas and remains unavailable to many rural and geographically isolated populations. The Bonn University Group’s first 50-case Ion series [76] and reports from European and Asian centers document broader adoption. Real-time remote expert supervision (tele-proctoring) to support credentialing and quality at lower-volume or rural centers, regional center-of-excellence credentialing, and structured outreach programs are potential strategies to extend RAB access.
15. Integration with Lung Cancer Screening Programs
The expansion of LDCT screening is increasing the number of PPLs requiring risk stratification and tissue evaluation. Lung-RADS v2022 provides a standardized framework for managing screen-detected nodules [26]. As screening programs expand, demand for tissue sampling of Lung-RADS 4A and 4B nodules may strain existing diagnostic capacity, including TTNB services, and bronchoscopic approaches such as RAB may help meet this demand [1,26].
RAB may be particularly useful for screening-detected lesions when concurrent mediastinal staging, bronchoscopic access to multiple nodules, or avoidance of transthoracic pneumothorax risk is important. It should not, however, be considered the automatic procedural modality for every screening-detected PPL: TARGET’s strict yield was 61.6%, and modality selection should reflect lesion anatomy, malignancy probability, patient risk, and local expertise [21,46]. Implementation barriers include workforce limitations, capital and infrastructure requirements, geographic inequity, and evolving reimbursement. Programs should monitor strict diagnostic yield, molecular adequacy, complications, and time-to-treatment.
16. Limitations, Controversies, and Evidence Gaps
Despite the advances documented throughout this review, several limitations should inform clinical decision-making and future research. Many studies originate from high-volume centers with access to advanced imaging, which may limit generalizability to broader practice settings.
16.1. Study Design and Evidence Quality
Most RAB evidence derives from observational studies, single-arm prospective cohorts, and retrospective series. VERITAS [17] established non-inferiority of navigational bronchoscopy to TTNB for 12-month diagnostic accuracy, with fewer pneumothoraces, but evaluated navigational bronchoscopy broadly rather than RAB alone and was conducted at expert centers. RELIANT [49] established non-inferiority, not superiority, of RAB to ENB. The ERS 2025 RAB + CBCT comparison remains a non-peer-reviewed conference abstract [40]. In addition, literature screening for this narrative review was conducted by a single author without duplicate study selection or formal risk-of-bias assessment, creating potential selection and confirmation bias. SANRA improves reporting transparency but does not eliminate these limitations.
16.2. Heterogeneity in Diagnostic Yield Reporting
Substantial heterogeneity exists in diagnostic-yield definitions, procedural techniques, patient populations, and adjunctive imaging. The range from TARGET’s 61.6% adjudicated strict yield to much higher intermediate or single-center estimates reflects methodological differences as much as technical performance. The ATS/ACCP statement recommends strict yield and recommends against intermediate or liberal definitions [45]. Many studies predate that consensus, and adherence remains incomplete. The Ali meta-analysis reported substantial heterogeneity (I2 = 65.6%) [8]. Contemporary interpretation should therefore prioritize strict estimates, report intermediate values only with explicit labeling, and avoid pooling unlike definitions. CBCT improves tool-in-lesion confirmation, but diagnostic-yield benefit remains mixed and platform-dependent [36,46,50].
16.3. Operator Experience, Institutional Volume, and Generalizability
RAB outcomes are influenced by operator experience and institutional procedural volume, raising concerns about generalizability to community and lower-volume settings. In the learning-curve analysis by Bott and colleagues [52], approximately half of proceduralists achieved proficiency within 25 lesions, and diagnostic yield at this high-volume academic center ranged from 58% to 83% across operators; other learning-curve studies have reported similar operator-dependent improvement [77,78,79]. Most published series still come from experienced centers, so the performance of RAB in lower-volume settings remains incompletely defined.
16.4. Infrastructure Requirements and Access Inequity
General anesthesia, advanced bronchoscopy infrastructure, and access to intraprocedural imaging create substantial and unevenly distributed barriers. Fixed CBCT requires a purpose-built hybrid suite, while mobile CBCT adds capital cost, workflow demands, and radiation exposure. Rural and geographically isolated systems may have access to a robotic platform without CBCT or specialized pathology support. Because published high-yield multimodal series frequently originate from expert centers, their results may not be reproducible in lower-resource settings, and CBCT should not be assumed to improve diagnostic yield uniformly [46,50].
16.5. Emerging Technologies: Limited Evidence Base
Randomized trials now support improved diagnostic yield with the 1.1 mm cryoprobe [60,63], including a dedicated PPL trial [60]. Safety findings are not uniform: FROSTBITE-2 reported no pneumothorax requiring chest-tube placement in the cryoprobe arm and no grade 3–4 bleeding in either arm [63], whereas Cryo-RCT found more moderate bleeding and pneumothorax with cryobiopsy [60]. Neither trial was restricted to robotic bronchoscopy, and neither evaluated a complete RAB + CBCT workflow. Evidence supporting multimodal RAB + CBCT + cryobiopsy therefore remains largely observational. AI-assisted Ion navigation and transbronchial ablation also remain emerging technologies without sufficient comparative outcome evidence to support claims of clinical superiority.
16.6. Cost-Effectiveness Data Gaps
Formal cost-effectiveness analyses comparing RAB with competing modalities remain limited and largely institution-specific. The decision-analytic model by Ost and colleagues [73] provides a useful framework but relies on assumptions about improvements in diagnostic sensitivity that may not be universally reproducible. The single-anesthetic biopsy and resection cost data reported by Weiser and colleagues [51] are derived from a small single-center series and require multicenter validation before they inform institutional investment decisions. In addition, current CPT coding does not distinguish RAB performed with versus without CBCT, which may discourage investment in higher-capital imaging approaches.
17. Conclusions
RAB is an important advance in the evaluation of PPLs, offering navigational reach, catheter stability, and compatibility with multimodal localization and tissue acquisition. Its diagnostic performance should be described using strict ATS/ACCP definitions rather than headline intermediate estimates. CBCT reliably improves tool-in-lesion confirmation, and cryobiopsy can improve tissue acquisition, but neither adjunct provides a uniform diagnostic-yield benefit across all platforms, centers, or patients [36,45,50,60,63].
Using the recommended strict definition, the most comprehensive meta-analysis reported a pooled RAB yield of 69.6%, and TARGET reported 61.6% [10,21]. Intermediate pooled estimates of approximately 84–87% are substantially higher but are not the preferred benchmark, because the ATS/ACCP statement recommends against intermediate and liberal definitions [8,10,45]. VERITAS showed that navigational bronchoscopy as a category was non-inferior to TTNB for 12-month diagnostic accuracy, with fewer pneumothoraces [17]; it did not establish RAB-specific superiority.
Outcomes remain dependent on lesion characteristics, sampling tools, operator experience, procedural volume, and imaging infrastructure. TARGET and a 2026 clinical review of pulmonary nodules provide an important counterweight to high-yield single-center reports: strict multicenter performance may be similar to other guided bronchoscopic approaches, and localization adjuncts have produced mixed yield results [21,46]. Future studies should use strict yield, distinguish yield from diagnostic accuracy, and prospectively evaluate the incremental value, radiation exposure, and safety of adjunctive imaging and cryobiopsy.
Beyond diagnosis, RAB is expanding the bronchoscopic frontier into same-session bilateral sampling, combined peripheral biopsy and mediastinal staging, fiducial marker placement, single-anesthetic biopsy-and-resection workflows, and investigational transbronchial therapeutic ablation. RAB is likely to play an increasing role within multidisciplinary lung cancer pathways as evidence continues to evolve.
Supplementary Materials
The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/diagnostics16193151/s1, Table S1: SANRA checklist.
Funding
This research received no external funding.
Institutional Review Board Statement
Not applicable.
Informed Consent Statement
Not applicable.
Data Availability Statement
No new data were created or analyzed in this study. Data sharing is not applicable to this article.
Conflicts of Interest
The author declares no conflicts of interest.
References
- U.S. Preventive Services Task Force. Screening for Lung Cancer: US Preventive Services Task Force Recommendation Statement. JAMA 2021, 325, 962–970. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Heerink, W.J.; de Bock, G.H.; de Jonge, G.J.; Groen, H.J.; Vliegenthart, R.; Oudkerk, M. Complication rates of CT-guided transthoracic lung biopsy: Meta-analysis. Eur. Radiol. 2017, 27, 138–148. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Memoli, J.S.W.; Nietert, P.J.; Silvestri, G.A. Meta-analysis of guided bronchoscopy for the evaluation of the pulmonary nodule. Chest 2012, 142, 385–393. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ali, M.S.; Trick, W.; Mba, B.I.; Mohananey, D.; Sethi, J.; Musani, A.I. Radial endobronchial ultrasound for the diagnosis of peripheral pulmonary lesions: A systematic review and meta-analysis. Respirology 2017, 22, 443–453. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ost, D.E.; Ernst, A.; Lei, X.; Kovitz, K.L.; Benzaquen, S.; Diaz-Mendoza, J.; Greenhill, S.; Toth, J.; Feller-Kopman, D.; Puchalski, J.; et al. Diagnostic yield and complications of bronchoscopy for peripheral lung lesions: Results of the AQuIRE registry. Am. J. Respir. Crit. Care Med. 2016, 193, 68–77. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Folch, E.E.; Bowling, M.R.; Pritchett, M.A.; Murgu, S.D.; Nead, M.A.; Flandes, J.; Krimsky, W.S.; Mahajan, A.K.; LeMense, G.P.; Murillo, B.A.; et al. NAVIGATE 24-month results: Electromagnetic navigation bronchoscopy for pulmonary lesions at 37 centers. J. Thorac. Oncol. 2022, 17, 519–531. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Pritchett, M.A.; Bhadra, K.; Calcutt, M.; Folch, E. Virtual or reality: Divergence between preprocedural CT scans and lung anatomy during guided bronchoscopy. J. Thorac. Dis. 2020, 12, 1595–1611, Erratum in J. Thorac. Dis. 2020, 12, 4593. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ali, M.S.; Ghori, U.K.; Wayne, M.T.; Shostak, E.; De Cardenas, J. Diagnostic Performance and Safety Profile of Robotic-Assisted Bronchoscopy: A Systematic Review and Meta-Analysis. Ann. Am. Thorac. Soc. 2023, 20, 1801–1812. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Zhang, C.; Xie, F.; Li, R.; Cui, N.; Herth, F.J.F.; Sun, J. Robotic-assisted bronchoscopy for the diagnosis of peripheral pulmonary lesions: A systematic review and meta-analysis. Thorac. Cancer 2024, 15, 505–512. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Li, X.; Bai, J.; Zhou, X.; Wang, T.; Zhang, Y.; Hu, Y. Diagnostic performance and safety for robotic-assisted bronchoscopy in pulmonary nodules: A systematic review and meta-analysis. Int. J. Surg. 2025, 111, 4020–4032. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Balasubramanian, P.; Abia-Trujillo, D.; Barrios-Ruiz, A.; Garza-Salas, A.; Koratala, A.; Chandra, N.C.; Yu Lee-Mateus, A.; Labarca, G.; Fernandez-Bussy, S. Diagnostic yield and safety of diagnostic techniques for pulmonary lesions: Systematic review, meta-analysis and network meta-analysis. Eur. Respir. Rev. 2024, 33, 240046. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Chang, A.T.; Chan, J.W.; Siu, I.C.; Liu, W.; Lau, R.W.; Ng, C.S. Robotic-assisted bronchoscopy: Advancing lung cancer management. Front. Surg. 2025, 12, 1566902. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Pyarali, F.F.; Hakami-Majd, N.; Sabbahi, W.; Chaux, G. Robotic-assisted navigation bronchoscopy: A meta-analysis of diagnostic yield and complications. J. Bronchol. Interv. Pulmonol. 2024, 31, 70–81. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Simoff, M.J.; Pritchett, M.A.; Reisenauer, J.S.; Ost, D.E.; Majid, A.; Keyes, C.; Casal, R.F.; Parikh, M.S.; Diaz-Mendoza, J.; Fernandez-Bussy, S.; et al. Shape-sensing robotic-assisted bronchoscopy for pulmonary nodules: Initial multicenter experience using the Ion Endoluminal System. BMC Pulm. Med. 2021, 21, 322. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Chen, A.C.; Pastis, N.J., Jr.; Mahajan, A.K.; Khandhar, S.J.; Simoff, M.J.; Machuzak, M.S.; Cicenia, J.; Gildea, T.R.; Silvestri, G.A. Robotic bronchoscopy for peripheral pulmonary lesions: A multicenter pilot and feasibility study (BENEFIT). Chest 2021, 159, 845–852. [Google Scholar] [PubMed]
- Saghaie, T.; Williamson, J.; Philips, M.; Kafili, D.; Sundar, S.; Hogarth, D.K.; Ing, A. FRONTIER Study: First human trial of Galaxy System with integrated TiLT technology. J. Bronchol. Interv. Pulmonol. 2024, 31, e01025. [Google Scholar]
- Lentz, R.J.; Frederick-Dyer, K.; Planz, V.B.; Koyama, T.; Aboudara, M.C.; Avasarala, S.K.; Casey, J.D.; Cheng, G.Z.; D’Haese, P.F.; Duke, J.D.; et al. Navigational Bronchoscopy or Transthoracic Needle Biopsy for Lung Nodules. N. Engl. J. Med. 2025, 392, 2100–2112. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Intuitive Surgical. FDA Clearance for AI-Integrated Ion Navigational Workflow. Globe Newswire, 8 October 2025.
- Rivera, M.P.; Mehta, A.C.; Wahidi, M.M. Establishing the diagnosis of lung cancer: Diagnosis and management of lung cancer, 3rd ed: American College of Chest Physicians evidence-based clinical practice guidelines. Chest 2013, 143, e142S–e165S. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Zhang, X.-B.; Hogarth, D.K. Multimodal innovations and clinical applications of robotic-assisted bronchoscopy in pulmonary nodule diagnosis: A review of recent advances. BMC Pulm. Med. 2026, 26, 56. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Murgu, S.; Chen, A.C.; Gilbert, C.R.; Sterman, D.H.; Pederson, D.; Rafeq, S.; Laxmanan, B.; Schwiers, M.L.; Connelly, J.; Benz, H.L.; et al. A Prospective, Multicenter Evaluation of Safety and Diagnostic Outcomes with Robotic-Assisted Bronchoscopy: Results of the TARGET Trial. Chest 2025, 168, 539–555. [Google Scholar] [CrossRef] [Scilit]
- Baaklini, W.A.; Reinoso, M.A.; Gorin, A.B.; Sharafkaneh, A.; Manian, P. Diagnostic yield of fiberoptic bronchoscopy in evaluating solitary pulmonary nodules. Chest 2000, 117, 1049–1054. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Kalchiem-Dekel, O.; Connolly, J.G.; Lin, I.H.; Husta, B.C.; Adusumilli, P.S.; Beattie, J.A.; Buonocore, D.J.; Dycoco, J.; Fuentes, P.; Jones, D.R.; et al. Shape-Sensing Robotic-Assisted Bronchoscopy in the Diagnosis of Pulmonary Parenchymal Lesions. Chest 2022, 161, 572–582. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Gould, M.K.; Donington, J.; Lynch, W.R.; Mazzone, P.J.; Midthun, D.E.; Naidich, D.P.; Wiener, R.S. Evaluation of individuals with pulmonary nodules: When is it lung cancer? Chest 2013, 143, e93S–e120S. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Riely, G.J.; Wood, D.E.; Ettinger, D.S.; Aisner, D.L.; Akerley, W.; Bauman, J.R.; Bharat, A.; Bruno, D.S.; Chang, J.Y.; Chirieac, L.R.; et al. Non-Small Cell Lung Cancer, Version 4.2024, NCCN Clinical Practice Guidelines in Oncology. J. Natl. Compr. Canc. Netw. 2024, 22, 249–274. [Google Scholar] [PubMed]
- Christensen, J.; Prosper, A.E.; Wu, C.C.; Chung, J.; Lee, E.; Elicker, B.; Hunsaker, A.R.; Petranovic, M.; Sandler, K.L.; Stiles, B.; et al. ACR Lung-RADS v2022: Assessment Categories and Management Recommendations. J. Am. Coll. Radiol. 2024, 21, 473–488. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Oberg, C.L.; Lau, R.P.; Folch, E.E.; He, T.; Ronaghi, R.; Susanto, I.; Channick, C.; Tome, R.G.; Oh, S. Novel Robotic-Assisted Cryobiopsy for Peripheral Pulmonary Lesions. Lung 2022, 200, 737–745. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Connolly, J.G.; Kalchiem-Dekel, O.; Tan, K.S.; Dycoco, J.; Chawla, M.; Rocco, G.; Park, B.J.; Lee, R.P.; Beattie, J.A.; Solomon, S.B.; et al. Feasibility of shape-sensing robotic-assisted bronchoscopy for biomarker identification in thoracic malignancies. J. Thorac. Cardiovasc. Surg. 2023, 166, 231–240. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Vaca-Cartagena, B.F.; Valdes-Camacho, S.; Ibrahim, M.I.; Barrios-Ruiz, A.; Yu Lee-Mateus, A.; Funes-Ferrada, R.; Chadha, R.M.; Hazelett, B.N.; Robertson, K.S.; Abia-Trujillo, D.; et al. Staging aortopulmonary lymph nodes with robotic-assisted bronchoscopy: A case report and literature review. Respirol. Case Rep. 2025, 13, e70109. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Wisa, D.M.; Raoof, S.; Husta, B.C.; Lee, R.P.; Bott, M.; Kalchiem-Dekel, O.; Chawla, M. Accessing aortopulmonary lymph nodes via robotic-assisted bronchoscopy: A new window of opportunity. Chest 2023, 164, A5238–A5239. [Google Scholar] [CrossRef] [Scilit]
- Abouzgheib, W.; Ambrogi, C.; Chai, M. Unlocking the potential of robotic-assisted bronchoscopy: Overcoming challenging anatomy and locations. Ther. Adv. Respir. Dis. 2024, 18, 17534666241259369. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Chrissian, A.A.; Khosa, J.; Daher, N.; Heimbaugh, C.; Mirshahidi, H.; Leung, A.; Zaheer, S.; Bush, D.; Ho, E. Diagnostic Utility of Sampling Multiple Synchronous Pulmonary Nodules During Same-Session Robotic-Assisted Bronchoscopy. J. Bronchol. Interv. Pulmonol. 2025, 32, e1029. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Fernandez-Bussy, S.; Garza-Salas, A.; Barrios-Ruiz, A.; Balasubramanian, P.; Yu Lee-Mateus, A.; Koirala, T.; Hazelett, B.N.; Chadha, R.; Patel, N.M.; Reisenauer, J.S.; et al. Shape-sensing robotic-assisted bronchoscopy in the multiple pulmonary nodule diagnosis during a single anesthetic event. Respiration 2024, 103, 397–405. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Xie, F.; Zhang, C.; Li, C.; Liu, D.; Zhang, Q.; Mu, C.; Hou, G.; Li, M.; Tang, C.; Xu, F.; et al. Chinese expert consensus on shape-sensing robotic-assisted bronchoscopy (ssRAB) in the management of peripheral pulmonary lesions. Transl. Lung Cancer Res. 2025, 14, 1500–1515. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Brock, J.M.; Fernandez-Bussy, S.; Lee, A.Y.; Chadha, R.M.; Oberg, C.; Ng, C.S.; Chen, E.; Folch, E.; Zhang, H.; Reisenauer, J.S.; et al. World expert consensus recommendations on shape-sensing robotic-assisted bronchoscopy for the diagnosis of peripheral pulmonary lesions. Respiration 2026. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Bashour, S.I.; Khan, A.; Song, J.; Chintalapani, G.; Kleinszig, G.; Sabath, B.F.; Lin, J.; Grosu, H.B.; Jimenez, C.A.; Eapen, G.A.; et al. Improving shape-sensing robotic-assisted bronchoscopy outcomes with mobile cone-beam computed tomography guidance. Diagnostics 2024, 14, 1955. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Husta, B.C.; Menon, A.; Bergemann, R.; Lin, I.H.; Schmitz, J.; Rakočević, R.; Nadig, T.R.; Adusumilli, P.S.; Beattie, J.A.; Lee, R.P.; et al. The incremental contribution of mobile cone-beam computed tomography to the tool-lesion relationship during shape-sensing robotic-assisted bronchoscopy. ERJ Open Res. 2024, 10, 00993-2023. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Abia-Trujillo, D.; Folch, E.E.; Yu Lee-Mateus, A.; Balasubramanian, P.; Kheir, F.; Keyes, C.M.; Villalobos, R.; Chadha, R.M.; Hazelett, B.N.; Fernandez-Bussy, S.; et al. Mobile cone-beam computed tomography complementing shape-sensing robotic-assisted bronchoscopy in the small pulmonary nodule sampling: A multicentre experience. Respirology 2024, 29, 324–332. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Chan, L.T.; Lau, K.K.W.; Orton, C.M.; Temov, K.; Tana, A.; Baboolal, I.; Karir, A.; Agaoglu, E.; Garner, J.; Kalyal, A.; et al. Tool-in-lesion verification of shape-sensing robotic-assisted bronchoscopy with cone-beam CT in sampling peripheral pulmonary nodules. Thorax 2026, 81, 571–580. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Steinack, C.; Baumgartner, P.; Monsch, G.M.; Schneiter, D.; Oh, S.; Samara, E.T.; Opitz, I.; Clarenbach, C.; Ulrich, S.; Kohler, M.; et al. Robotic-assisted bronchoscopy with integrated cone-beam CT vs conventional bronchoscopy for diagnosing peripheral pulmonary lesions: An open-label randomized controlled trial. Eur. Respir. J. 2025, 66, OA1178. [Google Scholar] [CrossRef] [Scilit]
- Salahuddin, M.; Sarkiss, M.; Sagar, A.E.S.; Vlahos, I.; Chang, C.H.; Shah, A.; Sabath, B.F.; Lin, J.; Song, J.; Moon, T.; et al. Ventilatory strategy to prevent atelectasis during bronchoscopy under general anesthesia (VESPA): A multicenter randomized controlled trial. Chest 2022, 162, 1393–1401. [Google Scholar] [PubMed]
- Sagar, A.E.S.; Sabath, B.F.; Eapen, G.A.; Song, J.; Marcoux, M.; Sarkiss, M.; Arain, M.H.; Grosu, H.B.; Ost, D.E.; Jimenez, C.A.; et al. Incidence and location of atelectasis developed during bronchoscopy under general anesthesia: The I-LOCATE trial. Chest 2020, 158, 2658–2666. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Reisenauer, J.; Duke, J.D.; Kern, R.; Fernandez-Bussy, S.; Edell, E. Combining shape-sensing robotic bronchoscopy with mobile three-dimensional imaging to verify tool-in-lesion: A pilot study. Mayo Clin. Proc. Innov. Qual. Outcomes 2022, 6, 177–185. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ratwani, A.P.; Duke, J.D.; Paez, R.; Bridwell, G.; Shojaee, S.; Lentz, R.J.; Eagle, S.; Bhadra, K.; Maldonado, F. Sustained breath-holds in bronchoscopy: A pause worth taking? A narrative review. J. Thorac. Dis. 2025, 17, 10541–10550. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Gonzalez, A.V.; Silvestri, G.A.; Korevaar, D.A.; Gesthalter, Y.B.; Almeida, N.D.; Chen, A.; Gilbert, C.R.; Illei, P.B.; Navani, N.; Pasquinelli, M.M.; et al. Assessment of Advanced Diagnostic Bronchoscopy Outcomes for Peripheral Lung Lesions: An Official ATS/ACCP Research Statement. Am. J. Respir. Crit. Care Med. 2024, 209, 634–646. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Callister, M.E.J.; Silvestri, G.A. Pulmonary Nodules. N. Engl. J. Med. 2026, 395, 894–905. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Yarmus, L.; Akulian, J.; Wahidi, M.; Chen, A.; Steltz, J.P.; Solomon, S.L.; Yu, D.; Maldonado, F.; Cardenas-Garcia, J.; Molena, D.; et al. Interventional Pulmonary Outcomes Group. A prospective randomized comparative study of three guided bronchoscopic approaches for investigating pulmonary nodules: The PRECISION-1 study. Chest 2020, 157, 694–701. [Google Scholar] [CrossRef] [Scilit]
- Chen, A.C.; Gillespie, C.T. Robotic Endoscopic Airway Challenge: REACH assessment. Ann. Thorac. Surg. 2018, 106, 293–297. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Paez, R.; Lentz, R.J.; Duke, J.D.; Siemann, J.K.; Salmon, C.; Dahlberg, G.J.; Ratwani, A.P.; Casey, J.D.; Chen, H.; Chen, S.C.; et al. Robotic versus Electromagnetic Bronchoscopy for Peripheral Pulmonary Lesions: A Randomized Trial (RELIANT). Am. J. Respir. Crit. Care Med. 2025, 211, 1644–1651. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Chen, J.; McNierney, D.; Zein, J.G.; Vaszar, L.T.; Swanson, K.L.; Azadeh, N.; Sakata, K.K. Diagnostic performance of Monarch robotic bronchoscopy with and without mobile cone-beam CT support. Lung Cancer 2026, 213, 108910. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Weiser, L.; Perez, C.; Watson, J.J.; Knabe, K.; Razavi, A.; Krishna, V.; Fuller, C.; Soukiasian, S.; Brownlee, A.R.; Soukiasian, H.J.; et al. Combining robotic navigational bronchoscopy and lung resection into a single anesthetic event: Cost-effectiveness, wait times, and outcomes. Ann. Thorac. Surg. Short. Rep. 2025, 3, 324–328. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Bott, M.J.; Toumbacaris, N.; Tan, K.S.; Husta, B.C.; Medina, B.D.; Adusumilli, P.S.; Beattie, J.A.; Lee, R.P.; Park, B.J.; Dycoco, J.; et al. Characterizing a learning curve for robotic-assisted bronchoscopy: Analysis of skills acquisition in a high-volume academic center. J. Thorac. Cardiovasc. Surg. 2025, 169, 269–278. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Fielding, D.I.; Bashirzadeh, F.; Son, J.H.; Todman, M.; Chin, A.; Tan, L.; Steinke, K.; Windsor, M.N.; Sung, A.W. First human use of a new robotic-assisted fiber optic sensing navigation system for small peripheral pulmonary nodules. Respiration 2019, 98, 142–150. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Benn, B.S.; Romero, A.O.; Lum, M.; Krishna, G. Robotic-assisted navigation bronchoscopy as a paradigm shift in peripheral lung access. Lung 2021, 199, 177–186. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Reisenauer, J.; Simoff, M.J.; Pritchett, M.A.; Ost, D.E.; Majid, A.; Keyes, C.; Casal, R.F.; Parikh, M.S.; Diaz-Mendoza, J.; Fernandez-Bussy, S.; et al. Ion: Technology and techniques for shape-sensing robotic-assisted bronchoscopy. Ann. Thorac. Surg. 2022, 113, 308–315. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Xie, F.; Zhang, Q.; Mu, C.; Zhang, Q.; Yang, H.; Mao, J.; Simoff, M.J.; Huang, J.A.; Zhang, X.; Sun, J. Shape-sensing robotic-assisted bronchoscopy (SS-RAB) in sampling peripheral pulmonary nodules: A prospective, multicenter clinical feasibility study in China. J. Bronchol. Interv. Pulmonol. 2024, 31, e0981. [Google Scholar]
- Brownlee, A.R.; Perez, C.; Weiser, L.; Yu, W.; Bolster, D.; Knabe, K.; Soukiasian, S.; Chaux, G.; Rocco, R.; Soukiasian, H.J. 1121 shape-sensing robotic-assisted bronchoscopic biopsies: Diagnostic yield and surgical implications. Ann. Thorac. Surg. 2025, 120, 928–936. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Brock, J.M.; Dittrich, A.S.; Kontogianni, K.; Heussel, C.P.; Klotz, L.V.; Winter, H.; Schellenberg, M.; Keppler, U.; Herth, F. First European results of shape-sensing robotic-assisted bronchoscopy. Respiration 2026, 105, 599–607. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Chrissian, A.A.; Khosa, J.; Duchman, B.; Ho, E.; Cheng, G. Airway registration approach and other determinants of procedure efficiency during shape-sensing robotic-assisted bronchoscopy. J. Bronchol. Interv. Pulmonol. 2025, 32, e01028. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Matsumoto, Y.; Nakai, T.; Park, D.; Lin, C.K.; Furuse, H.; Kawahara, T.; Kho, S.S.; Tanaka, H.; Yamada, K.; Watanabe, T.; et al. Cryobiopsy versus conventional bronchoscopic sampling for peripheral pulmonary lesions (Cryo-RCT): An open-label, parallel-group, randomised trial. Lancet Respir. Med. 2026. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lee-Mateus, A.Y.; Sawal, N.; Hartley, C.; Edell, E.; Vierkant, R.A.; Reisenauer, J. Efficacy of robotic bronchoscopy for molecular marker analysis in primary lung cancer. Clin. Lung Cancer 2024, 25, e11–e17. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Abia-Trujillo, D.; Funes-Ferrada, R.; Lee-Mateus, A.Y.; Barrios-Ruiz, A.; Khoor, A.; Patel, N.M.; Hazelett, B.N.; Robertson, K.S.; Fernandez-Bussy, S. Cryobiopsy versus fine-needle aspiration for shape-sensing robotic-assisted sampling of small lung nodules. Lung Cancer 2024, 196, 107967. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Thiboutot, J.; Kapp, C.M.; Illei, P.; Shofer, S.; Gilbert, C.R.; DiBardino, D.; DeMaio, A.; Sethi, S.; Wahidi, M.M.; Benn, B.S.; et al. Cryobiopsy vs Forceps for Bronchoscopic Lung Biopsy: The FROSTBITE-2 Randomized Clinical Trial. JAMA 2026, 335, 2038–2045. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Chaddha, U.; Agrawal, A.; Ghori, U. Safety and sample adequacy for comprehensive biomarker testing of bronchoscopic biopsies: AABIP-IASLC Clinical Practice Guideline. J. Thorac. Oncol. 2025, 20, 1237–1256. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Balavenkataraman, A.; Garza Salas, A.; Barrios Ruiz, A.; Balasubramanian, P.; Fernandez-Bussy, S.; Abia Trujillo, D. Early pneumatocele identification post-robotic-assisted bronchoscopy cryobiopsy: A case series experience. Respiration 2024, 103, 275–279. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- DiBardino, D.M.; Haas, A.R.; Lanfranco, A.R.; Litzky, L.A.; Sterman, D.; Bessich, J.L. High complication rate after introduction of transbronchial cryobiopsy into clinical practice at an academic medical center. Ann. Am. Thorac. Soc. 2017, 14, 851–857. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Assaad, M.; Kollipara, V.; Abi Hatem, N. Cerebral arterial air embolism complicating robotic guided lung biopsy: A call for vigilance. Respir. Med. Case Rep. 2025, 58, 102320. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Yu Lee-Mateus, A.; Reisenauer, J.; Garcia-Saucedo, J.C.; Abia-Trujillo, D.; Buckarma, E.H.; Edell, E.S.; Grage, R.A.; Bowman, A.W.; Labarca, G.; Johnson, M.M.; et al. Robotic-assisted bronchoscopy versus CT-guided transthoracic biopsy for diagnosis of pulmonary nodules. Respirology 2023, 28, 66–73. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Fernandez-Bussy, S.; Yu Lee-Mateus, A.; Reisenauer, J.; Balasubramanian, P.; Barrios-Ruiz, A.; Garza-Salas, A.; Chandra, N.C.; Koratala, A.; Nadrous, A.; Edell, E.S.; et al. Shape-sensing robotic-assisted bronchoscopy versus CT-guided transthoracic biopsy for the evaluation of subsolid pulmonary nodules. Respiration 2024, 103, 280–288. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- McNierney, D.; Chen, J.; Zein, J.G.; Vaszar, L.; Swanson, K.; Azadeh, N.; Sakata, K.K. Comparing Monarch versus Ion Robotic-Assisted Bronchoscopy Platforms: A Propensity Score-Matched Analysis. J. Bronchol. Interv. Pulmonol. 2025, 32, e01019. [Google Scholar]
- Benn, B.S.; Bawaadam, H.; Colwell, E.M.; Peterson, M.D.; Tisol, W.B.; Niroula, A.; Jaber, W.S.; Khullar, O.V.; Daymude, K.; Phan, C.T.; et al. Spot on: Indocyanine green-soaked fiducial markers for lung nodules prior to thoracic surgery. Chest Pulm. 2025, 3, 100131. [Google Scholar] [PubMed]
- De Leon, H.; Royalty, K.; Mingione, L.; Jaekel, D.; Periyasamy, S.; Wilson, D.; Laeseke, P.; Stoffregen, W.C.; Muench, T.; Matonick, J.P.; et al. Device safety assessment of bronchoscopic microwave ablation of normal swine peripheral lung using robotic-assisted bronchoscopy. Int. J. Hyperth. 2023, 40, 2187743. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ost, D.E.; Maldonado, F.; Shafrin, J.; Kim, J.; Marin, M.A.; Amos, T.B.; Hertz, D.S.; Kalsekar, I.; Vachani, A. Economic value of bronchoscopy technologies that improves sensitivity for malignancy for peripheral pulmonary lesions. Ann. Am. Thorac. Soc. 2024, 21, 1759–1769. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Agbontaen, K.O.; Cold, K.M.; Woods, D.; Grover, V.; Aboumarie, H.S.; Kaul, S.; Konge, L.; Singh, S. Artificial Intelligence-Guided Bronchoscopy is Superior to Human Expert Instruction: A Randomized Controlled Trial. Crit. Care Med. 2025, 53, e1105. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Manley, C.J.; Kramer, T.; Kumar, R.; Gong, Y.; Ehya, H.; Ross, E.; Bonta, P.I.; Annema, J.T. Robotic bronchoscopic needle-based confocal laser endomicroscopy to diagnose peripheral lung nodules. Respirology 2023, 28, 475–483. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Zalepugas, D.; Skowasch, D.; Feodorovici, P.; Bedetti, B.; Schnorr, P.; Pizarro, C.; Tischler, V.; Arensmeyer, J.; Kuetting, D.; Schmidt, J.; et al. First 50 Cases with the ION Robotic-Assisted Navigational Bronchoscopy System in Routine Clinical Use in Germany: The Bonn Experience. J. Clin. Med. 2025, 14, 6155. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Brock, J.M.; Dittrich, A.S.; Kontogianni, K.; Heussel, C.P.; Klotz, L.V.; Winter, H.; Schellenberg, M.; Herth, F.J. First European experience of shape-sensing robotic-assisted bronchoscopy: Learning curve analysis. Respiration 2025, 104, 736–749. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Guarize, J.; Bertolaccini, L.; Bardoni, C.; Donghi, S.M.; Spaggiari, L. Characterizing the learning curve of ION robotic bronchoscopy: A CUSUM-based analysis of diagnostic yield. J. Bronchol. Interv. Pulmonol. 2026, 33, e1043. [Google Scholar]
- Xie, F.; Zhang, Q.; Liu, S.; Yan, L.; Zhou, Y.; Sun, J. Learning curve of a robotic-assisted bronchoscopy system in sampling peripheral pulmonary nodules. Chin. Med. J. 2022, 135, 2753–2755. [Google Scholar] [CrossRef] [Scilit] [PubMed]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the author. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.