Impact Attenuation Performance of a Lightweight, Ventilated Two-Zone Natural Rubber Hip Protector: Experimental Testing and Finite Element Optimization

,
,
,
,
and
1
Department of Mechanical and Mechatronics Engineering, Faculty of Engineering, Prince of Songkla University, Hat Yai 90110, Songkhla, Thailand
2
Department of Orthopedics, Faculty of Medicine, Prince of Songkla University, Hat Yai 90110, Songkhla, Thailand
*
Author to whom correspondence should be addressed.

Abstract

Hip fractures caused by sideways falls impose a major health and economic burden on older adults, yet hip protectors are often rejected because they are thick, heavy, and uncomfortable. This study used finite element (FE) modeling and drop-tower testing to develop a lightweight two-zone natural rubber hip protector (TZNRHP), comprising a stiff solid-rubber inner zone over the greater trochanter and a closed-cell rubber-foam outer zone. A hyperelastic–viscoelastic FE model was validated against drop-tower tests on a surrogate hip (4.75 kg impactor, 0.52 m drop), and protectors 9–15 mm thick with six ventilation-hole patterns were evaluated (n = 5 repeated impacts; one-way ANOVA with Tukey’s test). The FE model predicted peak forces within 5.4% for all validation conditions. Thickness significantly affected peak force (p < 0.001, η2 = 0.99), and the 11 mm protector was the thinnest design significantly below the 3100 N fracture criterion, reducing the peak force from 6232 ± 160 N (unprotected) to 2785 ± 79 N (55.3% attenuation) while being 28.3% lighter than the original 15 mm design. For a female-specific threshold of 2800 N, the 12 mm design was required. Thirteen 5 mm ventilation holes (8.4% open area) raised the peak force by only 2.6% (2856 ± 73 N; not significantly different from the least-perforated pattern), and the 11 mm protector reduced the maximum femoral von Mises stress from 9.2 to 5.83 MPa. The optimized TZNRHP is a promising lightweight, ventilated hip protector for older adults.

1. Introduction

Hip fractures are among the most serious injuries in older adults, and their incidence is rising as populations age [1,2]. The global number of hip fractures is projected to nearly double between 2018 and 2050 [3]. People with osteoporosis are at particular risk because of reduced bone mineral density and femoral strength [4,5,6]. More than 90% of hip fractures result from falls, most often sideways falls onto the greater trochanter [7], and one-year mortality is approximately 22.8%, rising markedly with age [8]. Surgical treatment is costly [9]: approximately USD 50,508 per intertrochanteric fracture patient (USD 5.96 billion annually) in the United States [10] and USD 5013 per hemiarthroplasty for osteoporotic femoral neck fracture in Thailand [11]. Preventive measures that act at the moment of a fall, such as wearable hip protectors, have therefore attracted considerable attention [12].
Hip protectors attenuate or redirect the impact energy delivered to the greater trochanter during a fall [12], and nursing-home residents who fell while wearing them had a lower risk of hip fracture than those who fell unprotected [13]. A Cochrane review concluded that hip protectors probably reduce the risk of hip fracture slightly in older people in nursing-care settings, but that their effectiveness is limited by poor acceptance and adherence [14]. Commercial designs are classified as soft (energy-absorbing) or hard-shell (energy-shunting) and are typically worn as pads in pockets of special undergarments or attached over the hips [15,16]. The pads, commonly made of plastics, polyurethane or polyethylene foams, or other elastomers, either absorb and dissipate impact energy or divert the load from the greater trochanter into the surrounding soft tissue [16,17].
To justify the need for an improved hip protector, the problem can be framed as a hierarchy that links the clinical outcome to the design requirement (Table 1). At the highest level (Level 1), falls are the leading cause of injury death in older adults: in the United States, the age-adjusted fall death rate among adults aged 65 years or older rose from 64.7 to 78.4 per 100,000 between 2018 and 2024 [18], and approximately 22.8% of patients die within one year of a hip fracture [8]. At Level 2, the hip fracture rate depends on how often a fall breaks the hip. In a cohort of 2377 video-captured falls in long-term care, 1.26% of falls caused a hip fracture; the risk was 5.5 times higher when the faller landed sideways and 3.4 times higher when the hip struck the ground but was 55% lower (relative risk 0.45) when a hip protector was worn [19]. Sideways falls account for 63–69% of fall-related hip fractures [20]. Approximately three-quarters of hip fractures occur in women [21], whose femoral strength in sideways-fall loading is lower than that of men (2800 ± 670 N in women and 3860 ± 940 N in men who sustained a hip fracture) [22]. At Level 3, one in four older adults falls each year [18]; exercise programs reduce the rate of falls by about 23% [23], and training in safe falling techniques can reduce hip impact forces [24], but falls cannot be eliminated, particularly among frail older adults. Fall prevention and impact protection are therefore complementary. This study addresses Level 4, the device level: in the relatively rare but critical falls that land directly on the greater trochanter, a protector must reduce the femoral force below the fracture threshold, and it must be thin, light, and breathable enough to be worn consistently. A successful protector should therefore (i) keep the peak force significantly below the femoral fracture force of the target population in a severe sideways-fall test and (ii) be clearly thinner and lighter than current pads and be ventilated, so that adherence can improve.
Table 1. Problem hierarchy for fall-related hip fractures and the level addressed by this study.
Protective performance depends on pad geometry and material properties. Laing et al. [15] found that force attenuation increased with pad thickness across 26 commercial hip protectors, whereas van Schoor et al. [16] showed that the overlying soft-tissue thickness strongly influences the protection provided by both hard and soft pads. For rubber-based pads, the transmitted force is further governed by the compound properties, which can be tailored through formulation [28]. Wearer comfort—determined by pad position, shape, contact area, thickness, density, weight, flexibility, and breathability—also strongly affects user acceptance [27,29,30,31].
Recent studies have explored new materials and architectures to improve this protection–comfort trade-off. Yum et al. [32] combined a shear-thickening polymer with a polyurethane foam layer and obtained 42.7% force attenuation at a low impact energy (25.1 J) with a pad thinner than the average commercial soft pad. Three-dimensional-printed hip protectors produced by fused deposition modeling have been optimized using response surface methodology [33], and studies of airbag-based protectors have shown that thicker, unpressurized torus-shaped airbags absorb impact most efficiently [34]. Studies of other polymer and hybrid material systems provide further design insight: bio-inspired interface–structure synergy can unlock synchronized toughening in stiffness-mismatched hybrid materials [35], a concept analogous to combining a stiff and a compliant zone in a single pad; porosity strongly governs the mechanical properties of porous materials, as shown by combined fabrication and FE analysis [36], which is relevant to both foamed rubber and perforated pads; nanofillers can tailor the mechanical and thermal properties of polyurethane-based elastomers [37]; and rubber-like damping materials degrade under variable amplitude–frequency dynamic loading [38], underlining the importance of durability under repeated impacts.
The finite element method (FEM) is an effective tool for hip protector design [39]. It enables systematic evaluation of design variables such as pad thickness, material properties, and ventilation-hole geometry at a far lower cost than extensive testing, and validated models reveal stress and deformation fields in the pad, soft tissue, and bone that are difficult to measure experimentally [40,41], thereby supporting rational design optimization [39]. Recent FE studies driven by experimental data have quantified how fall direction and hip protectors influence femoral fracture risk [42] and have evaluated hip protector performance under sideways-fall impact [43].
Nevertheless, hip protectors remain underused because of discomfort, bulk, weight, heat retention, and difficulty in putting them on [2,12,27]; thinner, lighter, and more breathable pads that maintain protection are therefore needed. Natural rubber (NR), a major agricultural product of Thailand, offers high resilience and tunable stiffness and can be processed as a solid or a closed-cell foam, making it attractive for impact protection [28,39]. To our knowledge, no study has combined a two-zone NR architecture, a calibrated hyperelastic–viscoelastic FE model, thickness minimization, and ventilation-hole optimization. This study therefore aimed to develop and validate a two-zone natural rubber hip protector (TZNRHP) that balances impact attenuation, weight, and ventilation by (i) developing and validating an FE model of a surrogate-hip drop-tower test, (ii) determining the minimum thickness that keeps the peak force significantly below a reference fracture threshold, and (iii) identifying a ventilation pattern that provides substantial open area without compromising protection.

2. Materials and Methods

2.1. Design of the Two-Zone Natural Rubber Hip Protector

The TZNRHP was designed to be protective, lightweight, compact, and anatomically conforming (Figure 1). Its curved geometry follows the contour of the lateral hip and overlying soft tissue, with maximum in-plane dimensions of 112 mm × 85 mm. The inner zone is a circular insert of higher-stiffness, non-foamed NR positioned over the impact centroid of the greater trochanter to resist localized loading (diameter 56 mm; thickness 15 mm; surface area 24.63 cm2; volume 36.94 cm3; density 1.196 g/cm3; mass 44.2 g). The surrounding outer zone (thickness 15 mm; mass 65 g) is made of closed-cell NR foam to absorb impact energy and distribute the load over a broader area.
Figure 1. Design of the TZNRHP, showing the inner and outer zones, the pad curvature, and its placement over the surrogate femur and silicone soft tissue.

2.2. Materials

Natural rubber (STR 5L) was supplied by Chana Latex Co., Ltd. (Songkhla, Thailand); Supercell DPT (dinitrosopentamethylenetetramine blowing agent) and Promotor-NP by A.F. Goodrich Chemicals Co., Ltd. (Samut Prakan, Thailand); and all other ingredients by Kitpaiboon Chemical Partnership Ltd. (Bangkok, Thailand). All were used as received; contents are given in parts per hundred rubber (phr). Both compounds contained 100 phr NR, 4 phr zinc oxide, 1.5 phr stearic acid, 1 phr Wingstay L (antioxidant), and 2 phr sulfur. The outer-zone (foam) compound additionally contained 3 phr Supercell DPT, 1.5 phr Promotor-NP, 0.5 phr diphenylguanidine (DPG), 0.5 phr N-cyclohexyl-2-benzothiazole sulfenamide (CBS), 50 phr china clay, 30 phr paraffinic oil, and 1 phr red pigment, whereas the inner-zone (solid) compound contained 1 phr 2-mercaptobenzothiazole (MBT), 2 phr paraffinic oil, and 0.5 phr carbon black.

2.3. Fabrication

The TZNRHP was fabricated in four stages (Figure 2): (1) the two compounds were mixed separately on a two-roll mill; (2) the non-foamed inner-zone insert was compression-molded and cured at 150 °C for 10 min; (3) the outer-zone foam compound was weighed and portioned; and (4) the insert and foam compound were compression-molded together in the curved mold at 125 °C for 25 min, during which the outer zone foamed, vulcanized, and bonded to the insert. The protectors were cooled to room temperature before testing.
Figure 2. Fabrication of the TZNRHP: compounding on a two-roll mill, molding of the inner-zone insert, preparation of the outer-zone foam compound, and final molding and bonding of the two zones.

2.4. Design Optimization

2.4.1. Thickness Optimization

The initial design was 15 mm thick. Protectors 9–15 mm thick (in 1 mm steps) were compression-molded using calibrated steel spacers (±0.05 mm), weighed (resolution 0.01 g), and drop-tested (n = 5; Section 2.5). The protection criterion was a mean peak force significantly below 3100 N (Section 2.7), the average femoral fracture force reported for older adults and a common design benchmark [25,26]; the thinnest protector meeting it was considered optimal. Because the femoral fracture force depends on bone mineral density, femoral geometry, age, and impact direction [25,26,44], with femurs from elderly donors failing at roughly half the load of those from young donors [44], a single threshold cannot represent all older adults. The 3100 N value was therefore used as a reference benchmark for comparing designs. This benchmark is conservative for men (femoral strength of 3860 ± 940 N in men with incident hip fracture [22]) but not for women, who sustain approximately three-quarters of hip fractures [21] and whose femoral strength is lower (2800 ± 670 N in a population-based FE study [22] and 2771 ± 1136 N averaged across published studies [20]). The selection was therefore also evaluated against thresholds of 2800 N (mean femoral strength of women with hip fracture) and 2000 N (approximately one standard deviation lower, representing women with severe osteoporosis; Section 3.2.2).

2.4.2. Ventilation-Hole Design

To improve breathability, six patterns of 5 mm diameter through-holes (H1–H6; 2–24 holes) were applied to the 11 mm protector (Figure 3), arranged in concentric elliptical rings that leave the central high-stress region over the greater trochanter intact. Open area was defined as the total hole area divided by the pad planform area, multiplied by 100. The open area of all six patterns was calculated using the same planform area. Each pattern was evaluated by FE simulation and drop-tower testing (n = 5) and was acceptable if its mean peak force remained significantly below 3100 N. The recommended pattern was the one beyond which any further increase in open area produced a disproportionately large increase in peak force, while an adequate safety margin was retained.
Figure 3. Ventilation-hole patterns H1–H6 (5 mm diameter through-holes) evaluated on the 11 mm TZNRHP.

2.5. Drop-Tower Impact Testing

Impact tests were performed using a custom drop tower (Figure 4A) to simulate a sideways fall onto the greater trochanter. No ASTM or ISO standard currently exists for the biomechanical impact testing of hip protectors; the test configuration was therefore based on the recommendations of the international consensus statement on hip protector biomechanical testing [17] and on previously published protocols [16,39], whereas the rubber and silicone materials were characterized according to ASTM D575-91 and ASTM D6147 (Section 2.6.2). A 4.75 kg steel impactor with a diameter of 75 mm was released from a height of 0.52 m, corresponding to a theoretical impact velocity of 3.19 m/s (v = (2gh)1/2). The surrogate hip assembly comprised an aluminum femur, an 18 mm thick silicone soft-tissue simulant, the protector, and a hardwood iliac crest support and was mounted on a piezoelectric load cell (Kistler 9321B; 0–50 kN, Kistler Group, Winterthur, Switzerland) [45]. Force data were recorded at a sampling rate of 50 kHz using an NI cDAQ-9178 system. For each condition, a single protector specimen was impacted five times at the same location (pseudo-replication) at 23 ± 2 °C, with a recovery interval of 20 min between impacts to allow viscoelastic recovery of the rubber. Because less than 0.3% of the modulus relaxes during an impact (Section 2.6.2), little viscoelastic deformation accumulates per impact. The conditions included two unprotected baselines: the bare femur and the femur covered with the soft-tissue simulant. The five repeated impacts therefore quantify impact-to-impact (test) repeatability rather than specimen-to-specimen manufacturing variability. No systematic change in peak force was observed across the five consecutive impacts, indicating that the specimens were not progressively damaged by repeated testing.
Figure 4. Impact test configuration: (A) experimental drop-tower apparatus and (B) the corresponding finite element model.
This configuration represents a severe sideways fall with direct impact on the greater trochanter—the type of fall associated with the highest risk of hip fracture [19]—rather than a typical fall. The impact velocity of 3.19 m/s is close to the mean hip impact velocity of an unexpected sideways fall (3.0 ± 1.0 m/s), and the unprotected peak force with the soft-tissue simulant (6232 N) lies within the range of femoral impact forces predicted for a fall from standing height (4050–6420 N), as summarized in a review of sideways-fall impact forces [20]. These forces exceed the mean femoral strength in sideways-fall loading reported in the same review (4027 ± 1464 N in men and 2771 ± 1136 N in women) [20], so the test evaluates whether a protector can prevent a fracture in the falls that would otherwise break the hip. Peak hip forces in real falls vary widely with fall height, direction, body mass, and soft-tissue thickness, and in low-severity falls (impact velocity ≈ 1.0 m/s) the force is only about 1–2.5 kN [20], which a protector would attenuate further. The proportion of real falls that reach this force level is not known and was not addressed in this study.

2.6. Finite Element Modeling

2.6.1. Geometry and Mesh

The eight components of the surrogate hip assembly (Figure 4B) were modeled in SolidWorks version 2020 (Dassault Systèmes from AppliCAD Public Co., Ltd., Bangkok, Thailand) and analyzed in MSC Marc/Mentat version 2020 (MSC Software, Sigma Solutions Co., Ltd., Bangkok, Thailand), with the TZNRHP parameterized by thickness (9–15 mm) and ventilation pattern (H1–H6). Fully integrated four-node tetrahedral elements were used, with the Herrmann (mixed displacement–pressure) formulation for the NR zones and silicone to avoid volumetric locking. Four progressively refined meshes were compared (Table 2).
Table 2. Mesh convergence analysis of the FE model (11 mm TZNRHP).
The fine mesh (78,600 elements) was selected because its peak force differed from that of the medium mesh by only 0.85% (criterion < 2%), whereas further refinement changed it by 0.32% at a 78% higher computational cost.

2.6.2. Material Models

Both NR zones were modeled as hyperelastic materials using the Mooney–Rivlin strain energy density function. Uniaxial compressive stress–strain data were obtained according to ASTM D575-91 using cylindrical specimens (diameter 29 mm, height 12.5 mm) at a crosshead speed of 12 mm/min. The three-parameter Mooney–Rivlin model (Equation (1)) was fitted to the data by nonlinear least squares in MSC Marc version 2020, and the fit quality was evaluated using the root-mean-square error (RMSE; Equation (2)):
W = C 10 I 1 − 3 + C 01 I 2 − 3 + C 11 I 1 − 3 I 2 − 3
where W is the strain energy density; I1 and I2 are the first and second invariants of the right Cauchy–Green deformation tensor; and C10, C01, and C11 are the material constants of the three-parameter Mooney–Rivlin model.
RMSE = 1 n ∑ i = 1 n σ FEM , i − σ Exp , i 2
where σFEA,i and σExp,i are the fitted and measured stresses at data point i, and N is the number of data points. The fit achieved an RMSE of 3.16%. The compression tests reached compressive strains of up to 50%, which include the maximum compressive strain predicted in the pads during impact (approximately 50%), so that the large-deformation response was fitted within the strain range used in the simulations. The three-parameter form was selected because its C11 coupling term allows the nonlinear stiffening of the compressive stress–strain curves at large strains to be represented with only three constants, whereas higher-order Mooney–Rivlin forms with additional terms (e.g., C20 and C30) increase the risk of unstable material behavior outside the range of the test data. The three fitted constants (C10, C01, and C11) of each material are reported in Table 3.
Table 3. Material properties of the components of the drop-tower test model.
The silicone soft-tissue simulant (Ecoflex 00-14) was characterized in the same manner, and the fitted constants of all three hyperelastic materials are listed in Table 3, together with the densities of the solid NR (1196 kg/m3), NR foam (770 kg/m3), and silicone (1040 kg/m3) used in the dynamic analysis. The viscoelastic behavior of both NR zones and the silicone was characterized by compressive stress-relaxation tests at 50% strain according to ASTM D6147, and the relaxation curves were fitted with a five-term Prony series (Equation (3)):
G ( t ) = G ∞ + ∑ i = 1 N G i e − t / τ i
where G(t) is the shear relaxation modulus, G∞ is the long-term (equilibrium) shear modulus, and Gi and τi are the i-th dimensionless Prony coefficient and relaxation time (s), respectively. The fitted parameters (Table 4) gave an RMSE below 3.72% for all three materials, which is adequate for dynamic impact simulation. The relaxation times span a wide range because the series was fitted to relaxation data recorded over a long period. The first terms for the inner zone (τ1 = 5.09 × 10−32 s) and the silicone (τ1 = 4.22 × 10−39 s) are numerical artifacts of the unconstrained least-squares fit and have no physical meaning; because their coefficients are negligible (G1 = 1.68 × 10−5 and 1.88 × 10−7) and their relaxation times are far below the minimum time increment (10−5 s), they act only as part of the instantaneous modulus and do not affect the solution. Conversely, the remaining terms (τi ≥ 2.6 s) are much longer than the impact duration: evaluating Equation (3) shows that less than 0.3% of the modulus relaxes within the 0.10 s simulation and only 0.02–0.04% up to the time of peak force (~0.015 s). The response during impact is therefore governed mainly by the instantaneous hyperelastic stiffness, whereas the long-term terms describe the recovery of the materials between impacts. Millisecond-scale rate effects are not explicitly represented by this Prony series, as discussed in Section 3.1. The remaining components were modeled as linear-elastic, isotropic materials (Table 3): the impactor and other steel parts as AISI 1045 carbon steel (South Steel Co., Ltd., Songkhla, Thailand), the femur as 6061-T6 aluminum (United Aluminium Industry Co., Ltd., Prachin Buri, Thailand), and the iliac crest support as hardwood (Figure 4B).
Table 4. Prony series parameters of the viscoelastic materials.

2.6.3. Boundary and Contact Conditions

The impactor was given an initial velocity of 3.19 m/s normal to the pad (x-direction) and was subjected to gravitational acceleration (9.81 m/s2); all of its other degrees of freedom were free (Figure 4B). The y- and z-direction constraints on the TZNRHP and silicone layer were applied only to the nodes on their outer peripheral edges, representing the fixation of these layers on the surrogate hip in the experiment and preventing rigid-body sliding in the implicit analysis; all other nodes, including those in the impact region, were free to move laterally, so the lateral bulging of the rubber and silicone under compression was not suppressed. The base and load cell were fully fixed, whereas the iliac crest support and femur were constrained only through contact. The inner and outer NR zones were connected by glued (tied) contact, which permitted neither separation nor sliding and represents the co-vulcanized bond formed during molding (Section 2.3); no delamination of the two zones was observed after testing. All other interfaces (impactor–protector, protector–silicone, silicone–femur, and femur–iliac crest support) used penalty-based touching contact, which allows separation and sliding, except the glued load cell–femur interface.

2.6.4. Analysis Settings

Dynamic finite element analysis was performed using an implicit time-integration scheme in MSC Marc/Mentat 2020 with the impactor positioned immediately above the specimen. The 0.10 s simulation comprised five stages: approach and initial contact (0–0.005 s), impact loading (0.005–0.015 s), maximum compression (0.015–0.025 s), unloading and rebound (0.025–0.050 s), and post-impact recovery (0.050–0.100 s). Automatic time stepping (minimum increment 1 × 10−5 s) was used, with smaller increments during loading and maximum compression to capture the rapidly varying contact force. The predicted impact force was taken as the load-cell reaction force for comparison with the experiments.

2.7. Data Analysis

Results are reported as mean ± standard deviation (SD; n = 5). FE accuracy was quantified as Error (%) = |FFEM − Fexp|/Fexp × 100, and impact attenuation was calculated relative to the unprotected condition with soft-tissue simulant (Equation (4)):
Percentage   of   peak   force   reduction   ( % )   =   1 − F P F N   *   100
where FN and FP are the peak forces (N) without and with the TZNRHP. Weight reduction was calculated relative to the 15 mm protector (thickness study) and pattern H1 (ventilation study). The effects of thickness (seven levels) and ventilation pattern (six levels) on the experimental peak force were tested separately by one-way ANOVA, with Welch’s ANOVA, which does not assume equal variances, as a robustness check, followed by Tukey’s HSD post hoc test (α = 0.05). Homogeneity of variance was assessed with Bartlett’s test before the ANOVA. Effect sizes were expressed as η2 and ω2, and 95% confidence intervals (CIs) were calculated for each group mean. Whether each mean was below the 3100 N criterion was tested by a one-sided one-sample t-test (H1: mean < 3100 N; df = 4) using Minitab software (Version 17, Minitab, LLC, State College, PA, USA). Unprotected baselines were excluded from the statistical analysis. Because the five replicates for each condition consisted of repeated impacts on one specimen, these tests compare conditions relative to the impact-to-impact variability of the test system; they do not include between-specimen manufacturing variability, and the p-values should be interpreted accordingly.

3. Results and Discussion

3.1. Validation of the Finite Element Model

The FE predictions agreed closely with the experiments (Figure 5; Table 5), with peak-force errors of 1.96%, 2.18%, and 5.36% for the bare femur, the femur with soft-tissue simulant, and the 15 mm TZNRHP, respectively, and reproduced the lower, broader force peak produced by the protector. The slightly larger error for the protected case likely reflects the strain-rate-dependent nonlinearity of the NR foam, which the quasi-static hyperelastic fit with a Prony series only approximates. Because the Mooney–Rivlin constants were obtained at a low strain rate and the Prony terms relax very little within the impact duration (Section 2.6.2), the model does not explicitly represent the higher stiffness and hysteresis of the rubber at millisecond-scale strain rates (an initial nominal strain rate of approximately 210–350 s−1, i.e., 3.19 m/s divided by the 9–15 mm pad thickness, compared with approximately 0.016 s−1 in the compression tests). The model’s adequacy for impact conditions was therefore established by direct validation against the drop-tower tests, with errors of 0.4–5.4% for the 10–15 mm protectors (Table 6). Calibration with high-strain-rate data (e.g., drop-weight or split-Hopkinson compression tests) would further improve the predictions, particularly for very thin pads. Across all 15 validation conditions (two unprotected baselines, seven thicknesses, and six ventilation patterns), the mean absolute error was 3.1% (2.6% excluding the 9 mm pad), and every FE prediction except that for the 9 mm pad lay within two standard deviations of the experimental mean. Because the silicone-only condition (2.18% error) involves no rubber pad, it validates the silicone model independently of the NR models. Similar errors have been reported in recent FE studies of hip impact [42,46], supporting the use of the model as a design tool.
Figure 5. Experimental and FE-predicted force–time histories for (A) the bare femur, (B) the femur with an 18 mm soft-tissue simulant, and (C) the femur protected by the 15 mm TZNRHP.
Table 5. Validation of the FE model: experimental and predicted peak impact forces under different impact conditions.
Table 6. Effect of TZNRHP thickness on weight and peak impact force (mean ± SD, n = 5).

3.2. Effect of Thickness on Impact Attenuation and Weight

3.2.1. Peak Impact Force and Weight

Increasing the thickness from 9 to 15 mm reduced the experimental peak force from 4199 ± 121 N to 1513 ± 78 N, raising attenuation from 32.6% to 75.7% (Table 6). This effect was highly significant (one-way ANOVA: F(6, 28) = 629.78, p < 0.001, η2 = 0.99; Welch’s F(6, 12.4) = 488.96, p < 0.001; Table 7), and Tukey’s test showed that every thickness differed significantly from every other. Bartlett’s test indicated homogeneous variances among the thickness groups (χ2(6) = 3.04, p = 0.80) and among the ventilation patterns (χ2(5) = 0.37, p = 0.996), supporting the ANOVA assumptions, and Welch’s ANOVA led to the same conclusions. The coefficient of variation of the five repeated impacts was 1.7–5.3% across all conditions, indicating good test repeatability, and the smallest difference between adjacent thickness levels (212 N, 12 vs. 13 mm) was approximately 2.5 times the pooled within-condition SD (86 N). The FE model reproduced the trend with errors of 0.4–5.4% for 10–15 mm; the larger error at 9 mm (10.9%) likely reflects the severe compression of the thinnest pad, which amplifies material-model uncertainty at large strains. A thicker pad provides a longer deformation distance and more material for energy absorption, prolonging the impact and lowering the peak force, consistent with previous reports [15,42,43].
Table 7. One-way ANOVA of the experimental peak impact force for the effects of protector thickness and ventilation-hole pattern.
The 9 and 10 mm protectors exceeded the 3100 N criterion (one-sided p > 0.999), whereas all protectors of 11–15 mm were significantly below it (p < 0.001). Greater thickness improved attenuation but increased weight from 78.3 g (11 mm) to 109.2 g (15 mm). The 11 mm protector reduced the peak force to 2785 ± 79 N (95% CI 2687–2883 N; 55.3% attenuation), 10.2% below the criterion (t = 8.92), while being 28.3% lighter than the 15 mm protector.

3.2.2. Selection of the Optimal Thickness

As shown in Figure 6, the peak force decreased consistently with increasing thickness, whereas the pad weight increased approximately linearly from 59.3 g (9 mm) to 109.2 g (15 mm). The 11 mm protector was the thinnest configuration for which both the experimental (2785 N) and FE (2876 N) peak forces fell below 3100 N. Thicker protectors provided additional attenuation but added weight and bulk, which are known to reduce comfort and adherence [2,27]; similar trade-offs have been reported for other hip protector designs [33,42]. The 11 mm thickness was therefore selected as the best balance of protection, weight, and wearability and was used in the ventilation-hole study. Because the fracture threshold varies with bone geometry, bone density, and age (Section 2.4.1), the sensitivity of this selection was examined. For the 2800 N threshold representative of women, the 11 mm protector (2785 ± 79 N) was not significantly below the criterion (one-sided p = 0.35), whereas the 12 mm protector was (2415 ± 57 N; p < 0.001); for the 2000 N threshold, only the 15 mm protector satisfied it (1513 ± 78 N; p < 0.001; 14 mm, p = 0.17). The 11 mm design is therefore optimal for the 3100 N benchmark, which is conservative for men, whereas the 12 mm design (86.1 g, still 21.2% lighter than the original 15 mm design) is recommended for women and the 15 mm design for women with severe osteoporosis. Because the H3 pattern increased the peak force of the 11 mm pad by only 2.6%, a ventilated 12 mm pad is expected to remain well below 2800 N, although this should be confirmed experimentally.
Figure 6. Effect of protector thickness on peak impact force and weight: experimental and FE results. The dashed line indicates the 3100 N reference fracture-force criterion [26].

3.3. Stress Distribution and Strain Energy Density in the Femur and Soft Tissue

Figure 7 shows the FE-predicted von Mises stress in the femur and the strain energy density in the soft-tissue simulant at peak impact. The maximum femoral stress fell from 9.2 MPa to 7.70, 5.83, 5.21, and 3.94 MPa for the 9, 11, 13, and 15 mm TZNRHP, respectively, and the high-stress region shrank with increasing thickness; the 11 mm TZNRHP thus reduced the peak stress by approximately 37%. The maximum soft-tissue strain energy density decreased from 1.20 mJ/mm3 (9 mm) to 0.53 mJ/mm3 (15 mm), indicating that more energy was absorbed by the protector rather than transmitted to the tissue. Lower stress concentration in the proximal femur reduces mechanical demand and is associated with greater fracture resistance [25,40,41,47]. These stresses should be interpreted comparatively because the surrogate femur is made of 6061-T6 aluminum (E = 72.4 GPa), which is about four times stiffer than human cortical bone (E ≈ 17–18 GPa) and has a simplified geometry. Human cortical bone has longitudinal ultimate strengths of approximately 190 MPa in compression and 130 MPa in tension [48], whereas the trabecular bone of the proximal femur yields at much lower stresses, typically of the order of several megapascals or less depending on density and anatomic site [49]. In a human femur, the load is carried by a thin cortical shell in the femoral neck and trochanteric region together with trabecular bone, so absolute stresses under the same load would differ from, and locally exceed, those predicted for the aluminum surrogate. The relative stress reduction of approximately 37% achieved by the 11 mm TZNRHP is therefore the meaningful result because a lower transmitted force reduces the stresses in the proximal femur approximately proportionally in the elastic range, regardless of the bone material [40,41]. The 15 mm TZNRHP reduced the stress further, but at the cost of greater weight and bulk; the 11 mm design therefore remains the best balance for the 3100 N benchmark.
Figure 7. FE-predicted von Mises stress in the femur at peak impact (left) and total strain energy density in the soft-tissue simulant (right). Rows from top to bottom correspond to TZNRHP thicknesses of 9, 11, 13, and 15 mm.

3.4. Effect of Ventilation Holes

Peak force increased with open area (Figure 8), reflecting the loss of pad stiffness and energy-absorbing volume, and the pattern effect was significant (F(5, 24) = 11.24, p < 0.001, η2 = 0.70, ω2 = 0.63; Welch’s F(5, 11.2) = 9.46, p = 0.001; Table 7). From H1 to H3, open area rose from 1.3% to 8.4% while peak force increased by only 2.6% (2785 to 2856 N; approximately 10 N per percentage point), and H2 and H3 did not differ significantly from H1 (Table 8). Beyond H3, this penalty rose to approximately 16–38 N per percentage point, and H5 and H6 were significantly higher than H1–H3. Although all patterns remained significantly below 3100 N, the margin narrowed for H6 (3039 ± 60 N; p = 0.043), whose 95% CI (2965–3113 N) included the criterion, and the FE prediction for H5 (3080 N) approached it. H3 (13 holes) was therefore recommended, providing 8.4% open area with no significant loss of protection and a 7.9% safety margin (95% CI 2765–2947 N). This agrees with reports that excessive perforation can compromise impact attenuation [27,30,31].
Figure 8. Relationship between open area and experimental peak impact force for the 11 mm TZNRHP, used to select the recommended pattern (H3).
Table 8. Effect of ventilation-hole pattern on weight and peak impact force for the 11 mm TZNRHP (mean ± SD, n = 5).

3.5. Comparison with Previous Studies

Table 9 compares the proposed protector with representative designs. The 11 mm H3 protector reduced the peak force to 2856 ± 73 N (54.2% attenuation) at 77.3 g, exceeding the 2.5–40.0% attenuation of 26 commercial protectors [15] while remaining thin and light. Hard shells achieved lower peak forces (804–1298 N) through energy shunting, whereas soft foam pads attenuated the force by 42–62%, comparable to the present design [16]. Because impactor mass, velocity, surrogate geometry, and soft-tissue thickness differed among studies, these comparisons are qualitative.
Table 9. Comparison of the biomechanical performance and design characteristics of representative hip protectors.

3.6. Limitations

This study has several limitations. First, the surrogate hip (aluminum bone, silicone soft tissue, hardwood iliac crest support, and fixed base) does not reproduce the geometry, heterogeneity, and fracture behavior of the human proximal femur and pelvis, or the pelvic stiffness and mass during a real fall [50]. Second, only one impact direction, impact energy, and soft-tissue thickness (18 mm) were tested, whereas fall kinematics and soft-tissue thickness vary widely among older adults [16,17]. Third, femoral stresses were computed for an aluminum surrogate and are comparative only. Fourth, ventilation was assessed only by open area, without air-permeability or thermal-comfort measurements. Fifth, the five replicates per group were repeated impacts on a single specimen (pseudo-replication), which does not capture specimen-to-specimen manufacturing variability and limits the power to detect small differences between neighboring ventilation patterns; future studies should test several independently manufactured specimens per design. Sixth, the hyperelastic constants were obtained from quasi-static tests, so millisecond-scale strain-rate effects were not explicitly modeled, and the lateral edge constraints applied to the pad and silicone layer may slightly stiffen the predicted response. Seventh, the 3100 N reference threshold is conservative for men but not for women, whose femoral strength is lower [22]; sex-specific thresholds were therefore examined as a sensitivity analysis. Eighth, the drop-tower test represents a single severe sideways fall; the distribution of hip impact forces in real falls and hence the proportion of falls in which the protector would prevent a fracture remains to be determined [19,20]. Finally, long-term durability and user acceptance were not evaluated; future work should include subject-specific models, multiple fall scenarios, and wear trials with older adults.

4. Conclusions

Hip protectors can prevent fall-related hip fractures only if they are worn, and their bulk, weight, and poor breathability are major reasons for non-adherence. This study showed that a two-zone natural rubber hip protector (TZNRHP), designed with a validated hyperelastic–viscoelastic FE model and drop-tower testing, can be made substantially thinner and lighter and can be ventilated while maintaining impact protection. The main conclusions are as follows:
(1) The FE model predicted peak forces within 0.4–5.4% for the 10–15 mm protectors (10.9% for 9 mm), confirming that the model can serve as an efficient design tool that reduces the number of prototypes and physical tests.
(2) Thickness had a strong effect on peak force (η2 = 0.99). The 11 mm protector was the thinnest design significantly below the 3100 N reference fracture force, reducing the peak force from 6232 ± 160 N to 2785 ± 79 N (55.3% attenuation) and the maximum femoral stress by approximately 37%, while being 28.3% lighter (78.3 g) than the original 15 mm design.
(3) Thirteen 5 mm ventilation holes (H3; 8.4% open area) increased the peak force by only 2.6% (2856 ± 73 N), a value that was not significantly different from that of the least-perforated pattern and was 7.9% below the criterion, whereas larger open areas caused disproportionate losses in protection.
The importance of these findings is that they provide quantitative design guidelines—a minimum thickness and an upper limit on ventilation—for lightweight, breathable soft hip protectors and show that locally produced natural rubber can be engineered into a low-cost protector that combines a stiff, load-spreading core with an energy-absorbing foam. Such a design may improve user acceptance and adherence, which are the main barriers to the effectiveness of hip protectors in older adults. Because women sustain approximately three-quarters of hip fractures and have lower femoral strength, a 12 mm pad is recommended for women and a 15 mm pad for individuals with severe osteoporosis. Future work should include independently manufactured specimens, high-strain-rate material characterization, multiple fall scenarios and soft-tissue thicknesses, subject-specific bone models, air-permeability and thermal-comfort measurements, and wear trials with older adults.
In terms of the problem hierarchy in Table 1, this study addresses the device level (Level 4). It shows that, in a severe laboratory sideways fall, a thin, light, and ventilated natural rubber pad can reduce the femoral force below the fracture threshold of older adults. It does not yet show a reduction in hip fractures (Level 2) or in fall-related deaths (Level 1). Reaching these levels will require wear trials to confirm that the thinner, ventilated design improves adherence, followed by clinical studies in high-risk settings such as long-term care, where both the importance of hip impact and the benefit of wearing hip protectors have been documented [19]. Because hip protectors act only once a fall has occurred, they should be combined with fall-prevention programs such as balance and strength exercises [23].

Author Contributions

T.P.: conceptualization, methodology, software, validation, formal analysis, investigation, data curation, writing—original draft, and visualization. W.T.: conceptualization, methodology. S.S.: methodology, software, validation, and writing—review and editing. C.D.: conceptualization, methodology, validation, writing—review and editing, supervision, project administration, and funding acquisition. B.T.: conceptualization and resources. V.Y.: conceptualization and resources. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by the Agricultural Research Development Agency (Public Organization), Thailand, grant number CRP5707011210.

Institutional Review Board Statement

Not applicable.

Data Availability Statement

The data presented in this study are included in the article. Further inquiries can be directed to the corresponding author.

Acknowledgments

The authors thank the Department of Mechanical and Mechatronics Engineering, Faculty of Engineering, and the Department of Orthopedics, Faculty of Medicine, Prince of Songkla University, Thailand, for providing research facilities and equipment. T.P. gratefully acknowledges the Prince of Songkla University Graduate Scholarship for supporting his doctoral studies. During manuscript preparation, the authors used ChatGPT (OpenAI; GPT-6.1 Astra) to assist with drafting and revising portions of the text and improving language, readability, and organization. The authors also used the image generation and editing tool available through ChatGPT (OpenAI; GPT-6.1 Sol) to refine the graphical abstract’s layout, typography, legibility, and dimensions. The prompts provided to the image tool included: “Improve all text and NUMERALS with crisp high-resolution sans serif typography, consistent font sizes, careful alignment and spacing, readable subscripts/symbols”; “Do not invent FE numeric results or mesh details: preserve that scientific panel faithfully”; and “Resize/reflow this exact image into a landscape graphical abstract, target WIDTH 1100 pixels HEIGHT 560 pixels, ratio 55:28. Keep ALL content, numbers and wording exactly as in selected source, all three columns and all illustrations, including original FE mesh and its color scale numerals.” The authors reviewed and edited the AI-assisted text and graphical output and take full responsibility for the content of this publication.

Conflicts of Interest

The authors declare no conflicts of interest. The funder had no role in the design of this study; in the collection, analyses, or interpretation of data; in the writing of the manuscript; or in the decision to publish the results.

References

  1. Kannus, P.; Niemi, S.; Parkkari, J.; Palvanen, M.; Heinonen, A.; Sievänen, H.; Järvinen, T.; Khan, K.; Järvinen, M. Why is the age-standardized incidence of low-trauma fractures rising in many elderly populations? J. Bone Miner. Res. 2002, 17, 1363–1367. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  2. Meyer, G.; Mühlhauser, I. Hip protectors in the elderly: Lack of effectiveness or just suboptimal implementation? Eur. Rev. Aging Phys. Act. 2006, 3, 85–90. [Google Scholar] [CrossRef] [Scilit]
  3. Sing, C.W.; Lin, T.-C.; Bartholomew, S.; Bell, J.S.; Bennett, C.; Beyene, K.; Bosco-Levy, P.; Bradbury, B.D.; Chan, A.H.Y.; Chandran, M.; et al. Global epidemiology of hip fractures: Secular trends in incidence rate, post-fracture treatment, and all-cause mortality. J. Bone Miner. Res. 2023, 38, 1064–1075. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  4. Wei, T.S.; Hu, C.H.; Wang, S.H.; Hwang, K.L. Fall characteristics, functional mobility and bone mineral density as risk factors of hip fracture in the community-dwelling ambulatory elderly. Osteoporos. Int. 2001, 12, 1050–1055. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  5. Geusens, P.; Autier, P.; Boonen, S.; Vanhoof, J.; Declerck, K.; Raus, J. The relationship among history of falls, osteoporosis, and fractures in postmenopausal women. Arch. Phys. Med. Rehabil. 2002, 83, 903–906. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  6. Taylor, B.C.; Schreiner, P.J.; Stone, K.L.; Fink, H.A.; Cummings, S.R.; Nevitt, M.C.; Bowman, P.J.; Ensrud, K.E. Long-term prediction of incident hip fracture risk in elderly white women: Study of osteoporotic fractures. J. Am. Geriatr. Soc. 2004, 52, 1479–1486. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  7. Stevens, J.A.; Olson, S. Reducing falls and resulting hip fractures among older women. Home Care Provid. 2000, 5, 134–141. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  8. Lo, J.C.; Srinivasan, S.; Chandra, M.; Patton, M.; Budayr, A.; Liu, L.H.; Lau, G.; Grimsrud, C.D. Trends in mortality following hip fracture in older women. Am. J. Manag. Care 2015, 21, 206–214. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  9. Burge, R.; Dawson-Hughes, B.; Solomon, D.H.; Wong, J.B.; King, A.; Tosteson, A. Incidence and economic burden of osteoporosis-related fractures in the United States, 2005–2025. J. Bone Miner. Res. 2007, 22, 465–475. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  10. Adeyemi, A.; Delhougne, G. Incidence and economic burden of intertrochanteric fracture: A Medicare claims database analysis. JBJS Open Access 2019, 4, e0045. [Google Scholar] [PubMed]
  11. Vanitcharoenkul, E.; Kitcharanant, N.; Maneeon, S.; Chotiyarnwong, P.; Unnanuntana, A. In-hospital costs of hemiarthroplasty in patients with osteoporotic femoral neck fracture at Faculty of Medicine Siriraj Hospital. J. Med. Assoc. Thai 2023, 106, 106–114. [Google Scholar] [CrossRef] [Scilit]
  12. Parker, M.J.; Gillespie, L.D.; Gillespie, W.J. Hip protectors for preventing hip fractures in the elderly. Cochrane Database Syst. Rev. 2004, 3, CD001255. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  13. Forsén, L.; Søgaard, A.J.; Sandvig, S.; Schuller, A.; Røed, U.; Arstad, C. Risk of hip fracture in protected and unprotected falls in nursing homes in Norway. Inj. Prev. 2004, 10, 16–20. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  14. Santesso, N.; Carrasco-Labra, A.; Brignardello-Petersen, R. Hip protectors for preventing hip fractures in older people. Cochrane Database Syst. Rev. 2014, 3, CD001255. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  15. Laing, A.C.; Feldman, F.; Jalili, M.; Tsai, C.M.J.; Robinovitch, S.N. The effects of pad geometry and material properties on the biomechanical effectiveness of 26 commercially available hip protectors. J. Biomech. 2011, 44, 2627–2635. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  16. van Schoor, N.M.; van der Veen, A.J.; Schaap, L.A.; Smit, T.H.; Lips, P. Biomechanical comparison of hard and soft hip protectors, and the influence of soft tissue. Bone 2006, 39, 401–407. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  17. Robinovitch, S.N.; Evans, S.L.; Minns, J.; Laing, A.C.; Kannus, P.; Cripton, P.A.; Derler, S.; Birge, S.J.; Plant, D.; Cameron, I.D.; et al. Hip protectors: Recommendations for biomechanical testing—An international consensus statement (part I). Osteoporos. Int. 2009, 20, 1977–1988. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  18. Centers for Disease Control and Prevention. Older Adult Falls Data. Available online: https://www.cdc.gov/falls/data-research/index.html (accessed on 16 November 2026).
  19. Yang, Y.; Komisar, V.; Shishov, N.; Lo, B.; Korall, A.M.B.; Feldman, F.; Robinovitch, S.N. The effect of fall biomechanics on risk for hip fracture in older adults: A cohort study of video-captured falls in long-term care. J. Bone Miner. Res. 2020, 35, 1914–1922. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  20. Nasiri Sarvi, M.; Luo, Y. Sideways fall-induced impact force and its effect on hip fracture risk: A review. Osteoporos. Int. 2017, 28, 2759–2780. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  21. American Academy of Orthopaedic Surgeons. OrthoInfo: Falls. Available online: https://www.orthoinfo.org/falls (accessed on 16 November 2026).
  22. Kopperdahl, D.L.; Aspelund, T.; Hoffmann, P.F.; Sigurdsson, S.; Siggeirsdottir, K.; Harris, T.B.; Gudnason, V.; Keaveny, T.M. Assessment of incident spine and hip fractures in women and men using finite element analysis of CT scans. J. Bone Miner. Res. 2014, 29, 570–580. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  23. Sherrington, C.; Fairhall, N.J.; Wallbank, G.K.; Tiedemann, A.; Michaleff, Z.A.; Howard, K.; Clemson, L.; Hopewell, S.; Lamb, S.E. Exercise for preventing falls in older people living in the community. Cochrane Database Syst. Rev. 2019, 1, CD012424. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  24. Groen, B.E.; Weerdesteyn, V.; Duysens, J. Martial arts fall techniques decrease the impact forces at the hip during sideways falling. J. Biomech. 2007, 40, 458–462. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  25. Pinilla, T.P.; Boardman, K.C.; Bouxsein, M.L.; Myers, E.R.; Hayes, W.C. Impact direction from a fall influences the failure load of the proximal femur as much as age-related bone loss. Calcif. Tissue Int. 1996, 58, 231–235. [Google Scholar] [CrossRef] [PubMed]
  26. Cheng, X.G.; Lowet, G.; Boonen, S.; Nicholson, P.H.F.; Brys, P.; Nijs, J.; Dequeker, J. Assessment of the strength of proximal femur in vitro: Relationship to femoral bone mineral density and femoral geometry. Bone 1997, 20, 213–218. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  27. Post, E.; Komisar, V.; Sims-Gould, J.; Korall, A.M.B.; Feldman, F.; Robinovitch, S.N. Development of a stick-on hip protector: A multiple methods study to improve hip protector design for older adults in the acute care environment. J. Rehabil. Assist. Technol. Eng. 2019, 6, 2055668319877314. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  28. Srewaradachpisal, S.; Dechwayukul, C.; Chatpun, S.; Spontak, R.J.; Thongruang, W. Optimization of the rubber formulation for footwear applications from the response surface method. Polymers 2020, 12, 2032. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  29. Holzer, L.A.; von Skrbensky, G.; Holzer, G. Mechanical testing of different hip protectors according to a European standard. Injury 2009, 40, 1172–1175. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  30. Park, J.H.; Jung, H.K.; Lee, J.R. Development and evaluation of fall impact protection pads using additive manufacturing. Materials 2019, 12, 3440. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  31. Wardiningsih, W.; Troynikov, O. An evaluation of force attenuation, comfort properties and density of materials for hip protective pads. J. Eng. Fibers Fabr. 2019, 14, 1558925019853955. [Google Scholar] [CrossRef] [Scilit]
  32. Yum, H.; Kim, Y.; Kim, B.; Lee, Y.; Lee, T. Force attenuation properties of shear-thickening polymer (STP)-based hip protector in simulated sideways fall. J. Mech. Sci. Technol. 2020, 34, 5351–5357. [Google Scholar] [CrossRef] [Scilit]
  33. Yahaya, M.S.; Ridzwan, M.I.Z. Optimization of the impact attenuation capability of three-dimensional printed hip protector produced by fused deposition modeling using response surface methodology. Polymers 2023, 15, 4388. [Google Scholar]
  34. Hongsakun, T.; Tangtrakulwanich, B.; Vittayaphadung, N.; Thongruang, W.; Srewaradachpisal, S. The effect of airbag design on impact attenuation for hip protection. Eng. Technol. Appl. Sci. Res. 2025, 15, 24236–24245. [Google Scholar] [CrossRef] [Scilit]
  35. Dai, L.; Fan, H.; Wang, Y.; Han, Q.; Wang, H.; Yang, D.; Hua, L.; Chen, Y. Bio-inspired interface-structure synergy unlocking synchronized helicoidal toughening in stiffness-mismatched hybrid materials. Adv. Compos. Hybrid. Mater. 2026, 9, 341. [Google Scholar] [CrossRef] [Scilit]
  36. Fada, R.; Shahgholi, M.; Azimi, R.; Farhadi Babadi, N. Estimation of porosity effect on mechanical properties in calcium phosphate cement reinforced by strontium nitrate nanoparticles: Fabrication and FEM analysis. Arab. J. Sci. Eng. 2024, 49, 1815–1825. [Google Scholar] [CrossRef] [Scilit]
  37. Heydari, S.F.; Shahgholi, M.; Karimipour, A.; Salehi, M.; Galehdari, S.A. The effects of graphene oxide nanoparticles on the mechanical and thermal properties of polyurethane/polycaprolactone nanocomposites; a molecular dynamics approach. Results Eng. 2024, 24, 102933. [Google Scholar] [CrossRef] [Scilit]
  38. Wang, Q.; Hong, Z.; Shen, L.; Shi, L.; Li, L.; Zheng, Q.; Ren, Z. Multiscale analysis of dynamic degradation and fatigue life prediction of slender-strip entangled metallic pseudo-rubber under variable amplitude-frequency loading. Nonlinear Dyn. 2026, 114, 778. [Google Scholar] [CrossRef] [Scilit]
  39. Srewaradachpisal, S.; Thongruang, W.; Dechwayukul, C.; Tangtrakulwanich, B.; Wongsiri, S. Hip protectors: Comparative study of FEM simulation and testing. J. Res. Appl. Mech. Eng. 2018, 3, 5. [Google Scholar]
  40. Dall’Ara, E.; Luisier, B.; Schmidt, R.; Kainberger, F.; Zysset, P.; Pahr, D. A nonlinear QCT-based finite element model validation study for the human femur tested in two configurations in vitro. Bone 2013, 52, 27–38. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  41. Zysset, P.K.; Dall’Ara, E.; Varga, P.; Pahr, D.H. Finite element analysis for prediction of bone strength. BoneKEy Rep. 2013, 2, 386. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  42. Galliker, E.S.; Laing, A.C.; Ferguson, S.J.; Helgason, B.; Fleps, I. The influence of fall direction and hip protector on fracture risk: FE model predictions driven by experimental data. Ann. Biomed. Eng. 2022, 50, 278–290. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  43. Kaeonarong, P.; Tangtrakulwanich, B.; Petdee, T.; Yuenyongviwat, V.; Thongruang, W.; Srewaradachpisal, S. Experimental study on the design and modeling of honeycomb structures for impact force attenuation in hip protectors. Ain Shams Eng. J. 2026, 17, 103986. [Google Scholar] [CrossRef] [Scilit]
  44. Courtney, A.C.; Wachtel, E.F.; Myers, E.R.; Hayes, W.C. Age-related reductions in the strength of the femur tested in a fall-loading configuration. J. Bone Jt. Surg. Am. 1995, 77, 387–395. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  45. Metz, R. Impact and drop testing with ICP force sensors. Pulse 2006, 2, 2. [Google Scholar]
  46. Aoshima, Y.; Murakami, S.; Mizuno, K.; Otaka, Y.; Yamada, M.; Jinzaki, M. Analysis of loading to the hip joint in fall using a whole-body finite element model. J. Biomech. 2022, 142, 111262. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  47. Ali Banijamali, S.M.; Oftadeh, R.; Nazarian, A.; Goebel, R.; Vaziri, A.; Nayeb-Hashemi, H. Effects of different loading patterns on the trabecular bone morphology of the proximal femur using adaptive bone remodeling. J. Biomech. Eng. 2015, 137, 011011. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  48. Reilly, D.T.; Burstein, A.H. The elastic and ultimate properties of compact bone tissue. J. Biomech. 1975, 8, 393–405. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  49. Morgan, E.F.; Keaveny, T.M. Dependence of yield strain of human trabecular bone on anatomic site. J. Biomech. 2001, 34, 569–577. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  50. Laing, A.C.; Robinovitch, S.N. Characterizing the effective stiffness of the pelvis during sideways falls on the hip. J. Biomech. 2010, 43, 1898–1904. [Google Scholar] [CrossRef] [Scilit] [PubMed]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.