Abstract
Perfluoroalkoxy polymer (PFA) is suitable for chemically resistant microfluidic devices, but complete cavity filling does not necessarily ensure accurate replication of microscale channels. In this study, the filling transition and microchannel replication of injection-molded high-flow PFA substrates were investigated using a three-factor, three-level L9 orthogonal array. Nozzle temperature, mold temperature, and holding pressure were varied, and macroscopic filling was evaluated using projected fill-area ratio and part mass. The actual mold-rib geometry was measured by three-dimensional surface profilometry, and molded channels with nominal widths of 100, 300, and 500 μm were evaluated from optical cross-sectional images. Complete filling was achieved under five of the nine molding conditions, with projected fill-area ratios of 95.74–96.77% and part masses of 1.749–1.757 g. Nozzle temperature had the largest effect on macroscopic filling, followed by mold temperature and holding pressure. Under complete-filling conditions, the opening-width, bottom-width, and depth replication ratios were 77.1–113.2%, 89.6–99.0%, and 83.7–96.5%, respectively. These results show that macroscopic filling and local microchannel replication should be evaluated separately when defining the molding window for PFA microfluidic substrates.
1. Introduction
Microfluidic devices are widely used for chemical analysis, biological assays, and small-volume fluid handling because they provide controlled transport within microscale channels [1,2]. Polymer-based fabrication is attractive for producing these devices because complex structures can be replicated at relatively low manufacturing cost. Polydimethylsiloxane (PDMS) is commonly used for prototyping, but swelling in organic solvents and absorption of hydrophobic compounds can limit its use in chemically demanding environments [3,4,5]. Fluoropolymers provide an alternative because of their chemical resistance and thermal stability. Among them, perfluoroalkoxy polymer (PFA) is melt-processable and can therefore be formed into structured components using thermoplastic processing methods [6,7,8,9].
Recent microfluidic studies have further demonstrated that channel geometry and flow-path design can directly affect mixing, transport, and analytical performance. For example, wave-shaped microfluidic channels have been used to enhance premixing and improve the repeatability of point-of-care diagnostic assays [10]. These applications highlight the importance of accurately reproducing the intended microchannel geometry because dimensional deviations can alter the hydraulic and functional behavior of the final device. Injection molding is suitable for repeated production of thermoplastic microfluidic substrates [11,12]. However, replication of microscale features is affected by both material flow and heat transfer during filling. Previous micro-injection molding studies have shown that nozzle temperature, mold temperature, injection conditions, and feature geometry affect the filling of microstructured cavities [13,14,15]. Local mold heating can improve microfeature filling by delaying solidification near the mold surface [16]. Studies on injection-molded microfluidic channels have also reported dimensional changes in channel width and depth after filling, cooling, and demolding [14,17]. These results indicate that complete filling of the macroscopic cavity does not necessarily represent accurate replication of the local channel geometry.
Recent work has also demonstrated the feasibility of micro-injection molding for producing thermoplastic substrates with microscale channels. Stampone et al. used material-jetted mold inserts to replicate PMMA thin plates containing a 250 μm straight microchannel and showed that mold-insert geometry and durability are important considerations for reliable microfeature replication [18]. These findings further support the need to evaluate the actual replicated channel geometry rather than relying solely on macroscopic filling or nominal mold dimensions.
This issue is relevant to PFA because its processing temperature is considerably higher than those of commonly used amorphous thermoplastics. The AP-201 grade used in this study is classified as a high-flow PFA for injection molding and has a melting peak temperature of approximately 303 °C [19]. High melt flow facilitates cavity filling, but it does not define the processing conditions required for microscale replication. The polymer must remain sufficiently flowable to fill the main cavity and the regions surrounding the mold ribs before solidification. The final channel geometry can also differ from the nominal design because the actual machined mold geometry contains dimensional deviations and tapered sidewalls. Therefore, replication should be evaluated using the measured mold geometry rather than the nominal CAD dimensions [20,21]. In particular, the base and top widths of a protruding mold rib correspond to the opening and bottom widths of the molded channel, respectively. However, systematic experimental data linking macroscopic cavity filling to local microchannel replication in injection-molded PFA substrates remain limited, particularly when replication is evaluated against the measured mold geometry rather than the nominal design.
In this study, injection-molded PFA microfluidic substrates, intended for subsequent chip assembly, were evaluated in terms of filling behavior and microchannel replication. Nozzle temperature, mold temperature, and holding pressure were varied using a three-factor, three-level L9 orthogonal array. Macroscopic filling was evaluated using projected fill-area ratio, part mass, and filling state. The actual top width, base width, and height of the mold ribs were measured by three-dimensional surface profilometry. Microchannel opening width, bottom width, and depth were then measured from cross-sectional optical images for nominal channel widths of 100, 300, and 500 μm under complete-filling conditions. The corresponding dimensions of the mold ribs and molded channels were used to calculate dimensional replication ratios. The study therefore distinguishes the processing conditions required for complete macroscopic filling from the dimensional replication achieved at the microchannel scale.
2. Materials and Methods
2.1. Materials and Mold Insert
A commercial high-flow perfluoroalkoxy polymer (PFA), NEOFLON AP-201 (Daikin Industries, Ltd., Osaka, Japan), was used as the molding material. The material had a melt flow rate of 30.8 g/10 min and a melting peak temperature of 303 °C [19]. The PFA pellets were dried at 120 °C for 4 h before injection molding. The main material properties and fixed molding conditions are summarized in Table 1.
Table 1.
Material properties, injection-molding equipment, mold configuration, and fixed processing conditions used for fabrication of the PFA microfluidic substrates.
A 30 mm × 30 mm mold insert was designed to form five parallel microchannels in the molded substrate. The mold insert was made of Alloy 625 and fabricated by micro-milling. After machining, the micro-rib surfaces were polished to remove visible machining marks. No additional surface coating or treatment was applied. Five raised ribs with nominal widths of 100, 200, 300, 400, and 500 μm were arranged at a pitch of 5 mm. The nominal rib width refers to the designed top width of each raised rib. The nominal rib height was fixed at 100 μm. Circular inlet and outlet features with a diameter of 2 mm were located at both ends of each rib. The distance between the inlet and outlet centers was 22 mm.
The insert was mounted in a two-plate, single-cavity mold. A side gate was located at the center of the lower side of the cavity, and the polymer melt flowed upward across the rib array. Cartridge heaters and a temperature controller were used to control the mold temperature. The mold insert design, molding configuration, and dimensional definitions used for the replication analysis are shown in Figure 1.
Figure 1.
Mold design, injection-molding configuration, and dimensional definitions used for microchannel replication analysis. (a) Isometric and top views of the mold insert showing five raised micro-ribs with nominal widths of 100, 200, 300, 400, and 500 μm. (b) Injection-molding setup showing the mold insert, side gate, heating system, and melt-flow direction. (c) Definitions of the mold-rib dimensions Wt,i, Wb,i, and Hi, and the corresponding molded-channel dimensions Wo,m, Wb,m, and Dm.
2.2. Injection Molding Conditions and Experimental Design
Injection molding was performed using an electric injection molding machine WIZ80E (LS Mtron Ltd., Anyang, Republic of Korea). The injection speed was fixed at 50 mm/s, and the cooling time was fixed at 17s for all experimental conditions. These parameters were maintained constant so that the present L9 design could focus on the effects of nozzle temperature setting (hereafter referred to as nozzle temperature), mold temperature, and holding pressure. The transition from velocity-controlled filling to pressure-controlled holding was set at an injection pressure of 33.9 MPa. The cushion position was maintained at 9.2 mm for all molding conditions. The holding time and cooling time were fixed at 1 and 17 s, respectively. No mold-release agent was used, and the molded parts were removed using a robotic handling system.
Three processing parameters were selected as control factors: nozzle temperature, mold temperature, and holding pressure. The nozzle temperature was varied at 370, 390, and 410 °C. These values represent machine temperature setpoints rather than directly measured melt temperatures. The mold temperatures were 160, 180, and 200 °C, and the holding pressure was varied at 100, 130, and 160 MPa according to the L9 experimental design. The factor levels are summarized in Table 2.
Table 2.
Control factors and factor levels used in the L9 orthogonal-array design for evaluating the effects of nozzle temperature, mold temperature, and holding pressure.
For the highest-temperature condition, the barrel-temperature profile was set to 410/380/370/360 °C from the nozzle side to the hopper side. The manufacturer recommends a cylinder-temperature range of approximately 350–390 °C for the investigated PFA grade. In the present study, the nozzle temperature was intentionally increased to 410 °C to enhance melt filling of the microscale channel features, while the upstream barrel zones were maintained within or close to the recommended processing range.
The molding cycle time was approximately 20 s, consisting of an injection stage of approximately 3 s and a cooling stage of 17 s. The actual material residence time in the barrel was not directly measured. No visible discoloration, carbonized material, or other apparent signs of thermal degradation were observed under the 410 °C nozzle-temperature condition.
A three-factor, three-level L9 orthogonal array was used to define the molding conditions [22,23]. Nine parameter combinations were tested, as summarized in Table 3. After each change in processing conditions, 10 molding shots were performed and discarded to allow the molding process to stabilize. Subsequently, 10 additional specimens were produced under the stabilized condition, and three specimens were randomly selected from these 10 specimens for quantitative evaluation. All other molding parameters were maintained at the fixed values listed in Table 1.
Table 3.
L9 molding conditions and measured macroscopic filling characteristics of the PFA microfluidic substrates. Projected fill-area ratio and part mass are reported as mean ± standard deviation for three independently molded specimens.
2.3. Evaluation of Macroscopic Filling Behavior
Macroscopic filling was evaluated using the projected fill-area ratio, part mass, and visual filling state. Front-view images of the molded substrates were acquired under the same imaging conditions and analyzed using ImageJ (Ver. 1.54t) [24]. The PFA specimen was separated from the background using the automatic threshold function, and the same thresholding procedure was applied consistently to all runs. The projected filled area was determined from the segmented pixel area.
The projected fill-area ratio, FA, was calculated from the segmented pixel area.
where Af is the projected filled area, and Ac is the nominal projected cavity area. The projected fill-area ratio, FA, was calculated relative to the nominal 30 mm × 30 mm cavity area, Ac = 900 mm2. Thus, the projected fill-area ratio represents the two-dimensional projected area of the molded specimen relative to the nominal cavity area and should not be interpreted as a volumetric cavity-filling fraction.
The mass of each molded substrate was measured after the sprue and runner were removed at the same gate position. A balance with a resolution of 0.001 g was used for the measurement. Three independently molded specimens were evaluated for each L9 condition, and the results are reported as mean ± standard deviation (n = 3).
The filling state was classified as short shot, partial filling, or complete filling by visual inspection of the external cavity geometry. Complete filling was defined as formation of the entire external 30 mm × 30 mm substrate outline without any visible unfilled region. This visual criterion was applied independently of the calculated projected fill-area ratio. The projected fill-area ratio was mainly used to quantify the degree of incomplete filling, while part mass was used as an additional measure of the amount of polymer filled into the cavity.
2.4. Three-Dimensional Characterization of the Mold Insert
The actual geometry of the mold-insert ribs was measured using a three-dimensional surface profilometer OLS4100 (Olympus, Tokyo, Japan). Three representative ribs, CH1, CH3, and CH5, with nominal widths of 100, 300, and 500 μm, respectively, were characterized. One representative transverse profile was evaluated for each selected rib, and the rib top width Wt,i, base width Wb,i, and height Hi were determined from the measured profiles, as defined in Figure 1c. Repeated transverse profiles of the same rib were not evaluated; therefore, the measured dimensions were used as reference values for the selected rib locations rather than as estimates of dimensional variation along the entire rib.
The measured mold geometry, rather than the nominal CAD dimensions, was used as the reference for the replication analysis because the actual ribs contained sidewall taper and dimensional deviations introduced during precision machining [20,21,25]. The measured rib dimensions are summarized in Table 4.
Table 4.
Measured dimensions of the mold-insert micro-ribs used as the reference geometry for microchannel replication analysis. The rib top width, rib base width, and rib height were obtained by three-dimensional surface profilometry.
For a positive mold rib forming a negative channel, the corresponding dimensions were defined as follows:
where Wo,m, Wb,m, and Dm are the opening width, bottom width, and depth of the molded channel, respectively. These corresponding dimensions were used to calculate the replication ratios described in Section 2.6. For reference, the response table for the L9 level-mean analysis is presented in Table 5.
Wb,i ↔ Wo,m,
Wt,i ↔ Wb,m,
and Hi ↔ Dm,
Table 5.
Response table for the effects of nozzle temperature, mold temperature, and holding pressure on projected fill-area ratio and part mass. Factor influence was evaluated using the difference between the maximum and minimum level means, Δ.
2.5. Cross-Sectional Measurement of Molded Microchannels
Microchannel replication was evaluated for the complete-filling conditions, Runs 5–9. Three representative channels, CH1, CH3, and CH5, with nominal widths of 100, 300, and 500 μm, respectively, were selected for dimensional measurement. Three independently molded specimens were evaluated for each molding condition. Each specimen was sectioned perpendicular to the channel direction at approximately 24 mm from the gate-side reference edge, corresponding to about 80% of the 30 mm specimen length along the melt-flow direction. This downstream position was intentionally selected to evaluate local replication in a region relatively far from the gate, where the melt experiences a longer flow path before reaching the measurement location. The specimens were cut using a razor blade (DORCO), and no additional polishing was performed before optical observation. Mechanical sectioning may introduce local deformation, burrs, or edge distortion, which can contribute to uncertainty in the measured cross-sectional dimensions. Therefore, the reported dimensions represent local downstream replication and should not be interpreted as dimensional uniformity along the entire channel length. The channel cross sections were observed using an optical microscope Axio Vert. A1 (Carl Zeiss Microscopy GmbH, Jena, Germany). Under the imaging conditions used in this study, the effective pixel size was approximately 0.07–0.15 μm/pixel. The images were calibrated using the microscope scale information before dimensional measurement.
The channel opening width Wo,m was defined as the distance between the two upper channel edges, whereas the bottom width Wb,m was measured between the two lower corner positions where the sidewalls transitioned to the channel bottom. The channel depth Dm was measured vertically from the channel opening level to the channel bottom.
The molded PFA specimens were not measured using the three-dimensional surface profiler because stable surface detection was difficult due to optical transmission through the material. Therefore, the molded-channel dimensions were obtained from optical cross-sectional images, whereas the mold-insert geometry was measured by three-dimensional surface profilometry. Three independently molded specimens were evaluated for each molding condition, and the results are presented as mean ± standard deviation to indicate specimen-to-specimen repeatability. Although the image resolution permitted submicrometer pixel-level discrimination, the dimensional values were reported to the nearest 1 μm to account conservatively for practical uncertainty associated with razor-blade sectioning and optical edge identification. The measured channel dimensions and replication ratios are summarized in Table 6.
Table 6.
Measured dimensions and replication ratios of injection-molded PFA microchannels under complete-filling conditions. Channel opening width, bottom width, and depth are reported as mean ± standard deviation for nominal channel widths of 100, 300, and 500 μm, together with the corresponding replication ratios based on the measured mold-insert geometry.
2.6. Dimensional Replication Analysis
Dimensional replication was quantified by comparing the molded-channel dimensions with the corresponding measured mold-rib dimensions. The opening-width replication ratio was calculated as
where Wo,m is the molded-channel opening width and Wb,i is the measured mold-rib base width.
The bottom-width replication ratio was calculated as
where Wb,m is the molded-channel bottom width and Wt,i is the measured mold-rib top width.
The depth replication ratio was calculated as
A replication ratio of 100% represents dimensional agreement between the molded channel and the corresponding mold-rib dimension.
For the additional comparison of dimensional accuracy among complete-filling conditions, the absolute dimensional errors were calculated as
A value of 0% therefore represents exact dimensional agreement with the corresponding measured mold-insert geometry.
The dimensional values and replication ratios are reported as mean ± standard deviation based on three specimens for each channel width and molding condition. The mean absolute dimensional error for each run was calculated using the nine measurements obtained from three channel widths and three specimens.
To quantitatively compare microchannel replication among the complete-filling conditions, one-way analysis of variance (ANOVA) was performed separately for each nominal channel width (100, 300, and 500 μm). Runs 5–9 were treated as comparison groups, and the opening-width, bottom-width, and depth replication ratios were analyzed separately. Statistical significance was defined as (p < 0.05). When a significant ANOVA result was obtained, Tukey’s honestly significant difference (HSD) test was used for post hoc pairwise comparisons. These analyses were used only to evaluate differences among the tested complete-filling conditions and not to estimate the independent effects of the processing parameters.
The effects of the molding parameters on the macroscopic filling responses were evaluated using the level means of the L9 orthogonal array. For each processing factor, the difference between the maximum and minimum level means was calculated as
where represents the mean response at factor level k. The relative influence of nozzle temperature, mold temperature, and holding pressure was ranked according to the corresponding values. The response table is given in Table 5.
The L9 orthogonal array was used to evaluate the main effects of nozzle temperature, mold temperature, and holding pressure. Because the present three-factor, three-level L9 design did not provide sufficient degrees of freedom to independently estimate interactions among the processing parameters, interaction effects were not analyzed. Therefore, the level-mean responses and factor rankings presented in this study should be interpreted as main-effect comparisons only. The microchannel replication data from Runs 5–9 were not subjected to an independent Taguchi main-effect analysis because these runs represent only the complete-filling subset of the original L9 array. The dimensional data were therefore used only for comparison of channel-size effects and condition-specific replication behavior.
3. Results and Discussion
3.1. Macroscopic Filling Transition of PFA Microfluidic Substrates
Figure 2 shows the PFA microfluidic substrates obtained under the nine L9 molding conditions. Runs 1 and 2 exhibited severe short shots, with the melt filling only the lower region of the cavity. The projected fill-area ratios were 27.87 ± 0.78% and 34.01 ± 0.55%, respectively. In Run 3, the filled region extended further toward the upper part of the cavity, and the fill-area ratio increased to 52.36 ± 1.34%. However, a large unfilled region remained, and the condition was classified as partial filling.
Figure 2.
Macroscopic filling behavior of injection-molded PFA microfluidic substrates under the nine L9 molding conditions. The processing conditions shown for each run are given in the order of nozzle temperature, mold temperature, and holding pressure. Scale bars: 10 mm.
A substantial increase in cavity filling was observed in Run 4. The projected fill-area ratio reached 90.43 ± 1.72%, and most of the external substrate geometry was formed. A remaining unfilled region was still visible near the end of the flow path, and Run 4 was therefore classified as partial filling. Complete filling was obtained from Run 5 onward. Runs 5–9 formed the full external geometry of the 30 mm × 30 mm substrate, with projected fill-area ratios ranging from 95.74 to 96.77%.
Within the tested L9 conditions, the transition from partial to complete filling was observed between Runs 4 and 5. Both runs used a nozzle temperature of 390 °C, while the mold temperature and holding pressure increased from 160 to 180 °C and from 130 to 160 MPa, respectively. Thus, the combination of mold temperature and holding pressure used in Run 5 was sufficient to achieve complete cavity filling at a nozzle temperature of 390 °C. However, the individual contribution of each parameter cannot be determined from a direct comparison of these two runs because both factors changed simultaneously. The effects of the three processing parameters are therefore evaluated using the level-mean analysis in Section 3.2 and Section 3.3.
The projected fill-area ratios of Runs 5–9 remained slightly below 100%, ranging from 95.74 to 96.77%, despite the absence of visible unfilled regions. This systematic deviation does not indicate approximately 4% incomplete cavity filling because complete filling was classified independently based on formation of the entire external substrate outline. After molding, dimensional contraction during cooling can reduce the projected specimen dimensions relative to the nominal 30 mm × 30 mm cavity area. In addition, small uncertainties associated with two-dimensional image segmentation can affect the detected specimen boundary. No substantial out-of-plane warpage was observed when the specimens were placed on a flat surface. Therefore, the projected fill-area ratio was more useful for quantifying incomplete filling in Runs 1–4 than for distinguishing among the complete-filling conditions. Once the external cavity geometry was fully formed, additional evaluation using part mass and local microchannel dimensions was required to identify differences among the complete-filling conditions.
3.2. Effects of Processing Parameters on Projected Fill-Area Ratio
Figure 3a shows the projected fill-area ratio under the nine L9 molding conditions, and Figure 3b presents the corresponding level-mean responses. The fill-area ratio increased markedly with increasing nozzle temperature. The level mean increased from 38.08% at 370 °C to 94.60% at 390 °C and reached 96.09% at 410 °C. The increase from 370 to 390 °C was much larger than that from 390 to 410 °C, indicating that the response approached a plateau after complete cavity filling was achieved.
Figure 3.
Projected fill-area ratio of the PFA microfluidic substrates and corresponding main effects of the processing parameters. (a) Mean projected fill-area ratio for Runs 1–9, with error bars representing the standard deviation of three specimens. (b) Level-mean responses for nozzle temperature, mold temperature, and holding pressure. The dashed line indicates the grand mean of 76.26%.
The mold temperature showed a smaller but monotonic effect. The level means increased from 71.51% at 160 °C to 75.64% at 180 °C and 81.62% at 200 °C. The holding-pressure response was less monotonic, with level means of 73.65, 73.40, and 81.73% at 100, 130, and 160 MPa, respectively. The corresponding Delta values were 58.00, 10.12, and 8.33 percentage points for nozzle temperature, mold temperature, and holding pressure. Therefore, nozzle temperature showed the largest level-mean range for projected fill-area ratio, followed by mold temperature and holding pressure.
The strong effect of nozzle temperature is consistent with previous micro-injection molding studies showing that nozzle temperature, mold temperature, and heat transfer at the polymer–mold interface affect microfeature filling [12,13,16,26]. In contrast, holding pressure is mainly applied after the primary filling stage and therefore has a smaller effect on the extent of macroscopic filling. These mechanisms were not measured directly in the present study, but they are consistent with the observed level-mean trends.
Overall, the projected fill-area ratio was effective for identifying the transition from incomplete to complete filling. However, once complete filling was reached, its sensitivity decreased because the response remained close to 96% for Runs 5–9.
3.3. Effect of Processing Parameters on Part Mass
Figure 4a shows the measured part mass under the nine L9 molding conditions. The mean part mass increased from 0.502 g in Run 1 to 1.634 g in Run 4 as the cavity filling progressed. Runs 5–9 showed similar part masses between 1.749 and 1.757 g, indicating that the mass approached a stable value after complete cavity filling was achieved.
Figure 4.
Part mass of the PFA microfluidic substrates and corresponding main effects of the processing parameters. (a) Mean part mass for Runs 1–9, with error bars representing the standard deviation of three specimens. (b) Level-mean responses for nozzle temperature, mold temperature, and holding pressure. The dashed line indicates the grand mean of 1.384 g.
The main effects shown in Figure 4b were consistent with those obtained from the projected fill-area ratio. The level mean increased from 0.685 g at a nozzle temperature of 370 °C to 1.715 g at 390 °C and 1.751 g at 410 °C. The corresponding means for mold temperature were 1.295, 1.373, and 1.483 g, while those for holding pressure were 1.337, 1.332, and 1.481 g. The Delta values were 1.066, 0.188, and 0.149 g, respectively. Nozzle temperature showed the largest level-mean range for part mass, followed by mold temperature and holding pressure.
The large increase in part mass between 370 and 390 °C is consistent with the filling transition observed in Figure 2. Higher melt and mold temperatures are expected to maintain melt flow for a longer period, allowing a larger volume of polymer to enter the cavity before solidification. The smaller effect of holding pressure is reasonable because this parameter mainly influences the post-filling stage rather than the initial propagation of the melt front.
The similar trends in projected fill-area ratio and part mass indicate that both responses capture the macroscopic filling transition. The use of part mass as an indicator of filling behavior is consistent with previous micro-injection molding studies in which part weight was used to evaluate filling quality and process stability [15,27]. However, the small variation in part mass among Runs 5–9 shows that this parameter has limited sensitivity once complete cavity filling has been achieved. Local microchannel replication therefore requires separate dimensional evaluation.
3.4. Cross-Sectional Characteristics of Replicated Microchannels
Figure 5 compares the measured mold-rib geometry with representative cross sections of the molded PFA microchannels. The three-dimensional profiles showed that the mold ribs had tapered sidewalls rather than ideal rectangular cross sections. For CH1, CH3, and CH5, the measured rib top widths were 95.6, 284.4, and 495.2 μm, while the corresponding base widths were 255.4, 442.8, and 625.8 μm. The rib heights were 100.2, 100.5, and 100.4 μm, respectively.
Figure 5.
Measured mold-insert geometry and representative cross-sectional images of injection-molded PFA microchannels. (a) Three-dimensional surface profiles of the mold-insert ribs for CH1, CH3, and CH5 with nominal widths of 100, 300, and 500 μm, respectively. The measured rib top width Wt,i, base width Wb,i, and height Hi are indicated. Representative molded-channel cross sections are shown for (b) Run 5, (c) Run 6, and (d) Run 9. The processing conditions are given as nozzle temperature/mold temperature/holding pressure. Scale bars in the optical images: 200 μm.
The large difference between the rib top and base widths confirms that the mold features had a tapered geometry. Therefore, the channel opening width was compared with the rib base width, while the channel bottom width was compared with the rib top width. This correspondence is important because direct comparison with the nominal design width would include both mold-machining deviation and molding-induced dimensional change.
Representative cross sections for Runs 5, 6, and 9 are shown in Figure 5b–d. All three conditions produced well-defined cross-sectional channel profiles at the selected downstream position for nominal widths of 100, 300, and 500 μm, but differences in opening width, bottom width, and depth remained. Similar dimensional changes in channel width and depth have been reported for injection-molded microfluidic structures after filling, cooling, and demolding [14,15]. For example, Run 5 showed opening widths of 239.3, 412.8, and 613.7 μm for the 100, 300, and 500 μm channels, respectively, with corresponding bottom widths of 91.1, 264.9, and 487.5 μm.
The depth direction showed larger differences among the complete-filling conditions. Table 6 shows that the depth replication ratio ranged from 83.7 to 96.5%, whereas the bottom-width ratio generally remained closer to the measured mold reference. These results indicate that the local channel geometry remained sensitive to molding conditions even after complete cavity filling was achieved. The measured mold geometry was therefore used as the reference for the quantitative replication analysis in Section 3.5.
3.5. Dimensional Replication Under Complete-Filling Conditions
Figure 6 compares the opening-width ratio RW,o, bottom-width ratio RW,b, and depth ratio RD of the PFA microchannels under the complete-filling conditions. The corresponding values are summarized in Table 6.
Figure 6.
Dimensional replication characteristics of injection-molded PFA microchannels under complete-filling conditions. (a) Opening-width ratio RW,o, (b) bottom-width ratio RW,b, and (c) depth ratio RD for nominal channel widths of 100, 300, and 500 μm. Each point represents the mean of three specimens, and error bars indicate the standard deviation. The dashed line represents 100% replication relative to the measured mold-insert geometry.
The opening-width ratio showed the largest variation among the three dimensional responses. For the 100 μm channel, RW,o ranged from 77.1 to 113.2%. The corresponding ranges were 89.0–105.6% for the 300 μm channel and 96.4–104.6% for the 500 μm channel. The 500 μm channel remained closer to the mold reference under most conditions, whereas the 100 μm channel showed larger condition-dependent variation. This trend is consistent with previous studies showing that relative dimensional deviations become more significant as microfeature size decreases [13,14,17]. Run 8 produced opening-width ratios above 100% for all three channel widths.
Run 8 showed opening-width replication ratios above 100% for all three channel sizes. These values should not be interpreted as direct evidence of physical over-replication of the mold geometry. The molded parts were removed by a robot immediately after mold opening, with no additional in-mold cooling or flat-fixture constraint, and the channel dimensions were subsequently measured from razor-cut, unpolished cross sections. Therefore, the final measured geometry may reflect not only replication during molding but also thermal contraction, elastic recovery or local deformation during demolding, unconstrained post-demolding deformation, sectioning-induced deformation or edge distortion, and uncertainty in optical edge identification. Because these effects were not independently quantified, the replication ratios above 100% are interpreted as apparent dimensional deviations relative to the measured mold reference rather than as confirmed over-replication. The bottom-width ratio was more stable than the opening-width ratio. Across Runs 5–9, RW,b remained between 89.6 and 99.0%. The 300 and 500 μm channels were generally reproduced within approximately 94–99% of the corresponding mold-rib top widths. In contrast, the 100 μm channel showed larger deviations and larger scatter under some conditions, particularly in Run 6.
The depth ratio showed a different trend. The measured RD values ranged from 83.7 to 96.5% and remained below 100% for all tested conditions. Run 9 showed relatively low depth ratios of 83.7, 87.3, and 84.2% for the 100, 300, and 500 μm channels, respectively. No consistent monotonic relationship between nominal channel width and depth replication was observed.
One-way ANOVA confirmed that the opening-width replication ratio differed significantly among Runs 5–9 for all three nominal channel widths, with p = 0.0019, 0.0290, and 0.0018 for the 100, 300, and 500 μm channels, respectively. Tukey’s HSD test indicated that Run 8 showed significantly higher opening-width replication than several of the other complete-filling conditions, particularly for the 100 and 500 μm channels. The bottom-width replication ratio differed significantly among the runs only for the 300 μm channel (p = 0.0181), whereas no significant differences were observed for the 100 and 500 μm channels. In contrast, the depth replication ratio did not differ significantly among Runs 5–9 for any of the three channel widths (p > 0.05).
These results indicate that condition-dependent differences were most consistently observed in the channel opening width. However, because Runs 5–9 constitute only the complete-filling subset of the original L9 design, the statistical differences should be interpreted as comparisons among the tested molding conditions rather than as independent effects of nozzle temperature, mold temperature, or holding pressure.
These results show that complete macroscopic filling did not ensure uniform replication of the microchannel cross section. Bottom width was replicated relatively consistently, whereas opening width and depth showed larger variations depending on channel size and molding condition. Because Runs 5–9 represent only the complete-filling subset of the original L9 design, the dimensional data were used for condition-specific and channel-size comparisons rather than for independent ranking of the three processing parameters. The replication ratios reported in this section were obtained from a single downstream cross-sectional position located at approximately 80% of the channel length. Therefore, these values represent local downstream replication characteristics and should not be interpreted as dimensional uniformity along the entire microchannel.
3.6. Relationship Between Macroscopic Filling and Microchannel Replication
The macroscopic filling results and the microchannel measurements showed different sensitivities to the molding conditions. Runs 5–9 were classified as complete-filling conditions, with projected fill-area ratios of 95.74–96.77% and part masses of 1.749–1.757 g. These small differences indicate that the macroscopic filling response had reached a stable region once the external cavity was fully formed.
In contrast, the dimensional replication results showed measurable differences among the same conditions. The opening-width ratio ranged from 77.1 to 113.2%, the bottom-width ratio from 89.6 to 99.0%, and the depth ratio from 83.7 to 96.5%. In particular, Run 9 maintained relatively high bottom-width replication but showed lower depth replication for all three channel widths.
Figure 7 further compares the absolute dimensional errors under the complete-filling conditions. The dominant dimensional error differed among the runs. Run 7 showed a relatively small bottom-width error but a larger depth error, while Run 9 showed the largest depth error. This confirms that complete filling of the main cavity does not guarantee the same replication accuracy for all dimensions of the microchannel cross section.
Figure 7.
Mean absolute dimensional errors of injection-molded PFA microchannels under complete-filling conditions. The opening-width error EW,o, bottom-width error EW,b, and depth error ED were calculated relative to the measured mold-insert geometry. Each value represents the mean of nine measurements obtained from three channel widths and three specimens.
The persistence of dimensional deviations after complete macroscopic filling can be interpreted in terms of the local thermal and pressure history around the micro-ribs. During filling, the PFA melt transfers heat rapidly to the mold surface, which can increase the local melt viscosity and promote the formation of a highly viscous or partially solidified layer near the polymer–mold interface. Macroscopic cavity filling can continue through the hotter bulk melt even when further deformation of the polymer around the microscale rib geometry becomes increasingly restricted. This effect can be more pronounced for smaller microfeatures because of their greater sensitivity to interfacial cooling and local flow resistance.
Nozzle and mold temperatures can influence this behavior by affecting the time during which the polymer remains sufficiently mobile near the mold surface. Holding pressure may also assist local replication by transmitting pressure to the microfeature region after the primary filling stage; however, its effectiveness is expected to decrease as local solidification progresses and pressure transmission becomes restricted. Therefore, the final microchannel dimensions likely reflect the combined effects of melt flow, interfacial heat transfer, local solidification, and pressure transmission. These mechanisms were not independently measured in the present study and are therefore proposed as physical interpretations of the observed replication behavior rather than as directly verified causal relationships.
The difference arises from the scale of the two evaluations. Projected fill-area ratio and part mass mainly represent filling of the 30mm × 30 mm cavity, whereas the replication ratios describe local geometry formed around the microscale mold ribs. Previous studies have similarly shown that process-level indicators and local microfeature quality should be evaluated separately when assessing micro-injection molded components [22,28]. Therefore, complete macroscopic filling should be regarded as the first requirement for PFA microfluidic substrate molding, followed by a separate assessment of local dimensional replication.
The mold-insert material and surface condition may also influence local microchannel replication. In the present study, the Alloy 625 insert was fabricated by micro-milling and subsequently polished to remove visible machining marks. The thermal conductivity of the insert can affect heat transfer from the PFA melt to the mold surface and consequently influence the development of a highly viscous or locally solidified interfacial layer. In addition, the surface condition of the polished micro-ribs may affect polymer–mold interfacial contact and demolding behavior. However, Alloy 625 and the polished surface condition were fixed throughout the experiments, and surface roughness was not independently quantified. Therefore, their individual contributions to the observed replication behavior could not be separated in the present study.
Other molding parameters that were held constant in the present study may also influence microchannel replication. In particular, injection speed can affect the melt-flow rate, shear history, and thermal state of the polymer as it enters the microstructured cavity, while cooling time can influence solidification, shrinkage, and the final dimensions after demolding. In the present experiments, injection speed and cooling time were fixed at 50 mm/s and 17 s, respectively, and their individual effects were therefore not independently evaluated. The replication behavior observed in this study should consequently be interpreted within these fixed processing conditions. Future work should examine these parameters together with the thermal and pressure-related factors considered here.
Demolding may also contribute to the final microchannel geometry. In the present molding process, the part was removed by a robot immediately after mold opening, with no additional in-mold cooling or subsequent flat-fixture constraint. The PFA substrate could therefore undergo thermal contraction, elastic recovery during release from the tapered mold ribs, and unconstrained deformation after demolding. Consequently, the geometry measured after demolding may differ from that present at the end of the holding and cooling stages. Because in-mold and post-demolding dimensions were not measured separately, the contribution of demolding-induced recovery and deformation could not be quantified and is considered a possible factor affecting the measured replication ratios.
The dimensional deviations observed in the molded microchannels may also have functional implications for subsequent microfluidic operation. Variations in channel opening width, bottom width, and depth can alter the effective flow cross section and may consequently affect hydraulic resistance, pressure drop, and flow distribution. However, direct fluidic performance was not evaluated in the present study. Therefore, the measured replication ratios should be interpreted as dimensional characteristics of the molded channels rather than as direct indicators of fluidic performance.
The present dimensional analysis was limited to Runs 5–9, which represent only the complete-filling subset of the original L9 design. Therefore, independent main effects of nozzle temperature, mold temperature, and holding pressure on microchannel replication could not be determined from these data. Accordingly, the dimensional results were interpreted only as condition-specific comparisons among the fully filled molding conditions and were not used to identify an optimal combination of processing parameters. In addition, the dimensional measurements were obtained from three independently molded specimens at a single downstream cross-sectional position located at approximately 80% of the channel length. Because local thermal and pressure histories may vary along the melt-flow direction, these measurements should be interpreted as local downstream replication characteristics rather than as dimensional uniformity along the entire channel. Further work should evaluate replication at multiple positions along the channel length and quantitatively relate the as-molded channel geometry to hydraulic resistance, pressure drop, flow distribution, and subsequent bonding performance.
Overall, the results indicate that a practical molding window for PFA microfluidic substrates should consider both cavity-filling stability and microchannel dimensional replication. Complete macroscopic filling is necessary, but it is not sufficient to ensure accurate reproduction of the local channel geometry.
4. Conclusions
The filling transition and microchannel replication of injection-molded high-flow PFA microfluidic substrates were investigated using a three-factor, three-level L9 orthogonal array. Nozzle temperature, mold temperature, and holding pressure were varied, and both macroscopic filling and local channel replication were evaluated. The following conclusions were obtained.
- A clear transition from incomplete to complete cavity filling was observed within the tested process window. The projected fill-area ratio increased from 27.87% in Run 1 to 90.43% in Run 4, while Runs 5–9 produced complete external cavity geometry with fill-area ratios of 95.74–96.77%.
- Among the main effects evaluated using the L9 design, nozzle temperature showed the largest level-mean range (Δ), followed by mold temperature and holding pressure. The level mean of the fill-area ratio increased from 38.08% at 370 °C to 94.60% at 390 °C, while the increase from 390 to 410 °C was small. A similar trend was observed for part mass. Interaction effects among the processing parameters were not independently evaluated in the present L9 design.
- The part mass increased with cavity filling and then stabilized at 1.749–1.757 g for Runs 5–9. This indicates that projected fill-area ratio and part mass are effective for identifying the macroscopic filling transition, but both responses have limited sensitivity after complete filling is achieved.
- Three-dimensional measurement of the mold insert showed that the micro-ribs had tapered cross sections. Therefore, channel opening width, bottom width, and depth were evaluated using the corresponding measured rib base width, top width, and height rather than the nominal CAD dimensions.
- Microchannel replication varied with channel size and among the tested complete-filling conditions. The opening-width ratio ranged from 77.1 to 113.2%, the bottom-width ratio from 89.6 to 99.0%, and the depth ratio from 83.7 to 96.5%. The bottom width was reproduced more consistently than the opening width and depth. Because Runs 5–9 represent only the complete-filling subset of the original L9 design, these results should be interpreted as comparative replication characteristics rather than as independent effects of the three processing parameters.
- Complete macroscopic cavity filling did not ensure accurate replication of the entire microchannel cross section. Therefore, a practical molding window for PFA microfluidic substrates should be defined using both macroscopic filling stability and local dimensional replication.
The present study was limited to the complete-filling subset of the L9 design for microchannel dimensional analysis; therefore, the independent effects of nozzle temperature, mold temperature, and holding pressure on microchannel replication could not be separately determined from these data. In addition, the dimensional measurements were obtained from three independently molded specimens at a single downstream cross-sectional position and should therefore be interpreted as local replication characteristics rather than as dimensional uniformity along the entire channel length. The present study also focused on dimensional replication and did not directly evaluate fluidic performance. Further work should evaluate replication at multiple positions along the channel length and quantitatively relate the as-molded geometry to hydraulic resistance, pressure drop, flow distribution, bonding performance, and overall microfluidic performance.
Author Contributions
Conceptualization, S.M.P. and K.J.C.; methodology, S.H.S., S.M.P. and K.J.C.; investigation, S.H.S. and K.J.C.; formal analysis, K.J.C.; supervision, S.M.P. and K.J.C.; writing—original draft preparation, S.H.S. and K.J.C.; writing—review and editing, K.J.C.; project administration, K.J.C.; funding acquisition, K.J.C. All authors have read and agreed to the published version of the manuscript.
Funding
This research was funded by the Ministry of Trade, Industry and Energy (MOTIE) of Korea through the Machinery and Equipment Industry Technology Development Program, grant number RS-2026-25610019, supervised by the Korea Planning & Evaluation Institute of Industrial Technology (KEIT).
Institutional Review Board Statement
Not applicable.
Informed Consent Statement
Not applicable.
Data Availability Statement
The original contributions presented in this study are included in the article material. Further inquiries can be directed to the corresponding authors.
Conflicts of Interest
The authors declare no conflicts of interest. The funders had no role in the design of the study; in the collection, analysis, or interpretation of data; in the writing of the manuscript; or in the decision to publish the results.
References
- Whitesides, G.M. The origins and the future of microfluidics. Nature 2006, 442, 368–373. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Sackmann, E.K.; Fulton, A.L.; Beebe, D.J. The present and future role of microfluidics in biomedical research. Nature 2014, 507, 181–189. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Duffy, D.C.; McDonald, J.C.; Schueller, O.J.A.; Whitesides, G.M. Rapid prototyping of microfluidic systems in poly(dimethylsiloxane). Anal. Chem. 1998, 70, 4974–4984. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lee, J.N.; Park, C.; Whitesides, G.M. Solvent compatibility of poly(dimethylsiloxane)-based microfluidic devices. Anal. Chem. 2003, 75, 6544–6554. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Toepke, M.W.; Beebe, D.J. PDMS absorption of small molecules and consequences in microfluidic applications. Lab Chip 2006, 6, 1484–1486. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ren, K.; Dai, W.; Zhou, J.; Su, J.; Wu, H. Whole-Teflon microfluidic chips. Proc. Natl. Acad. Sci. USA 2011, 108, 8162–8166. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Begolo, S.; Colas, G.; Viovy, J.-L.; Malaquin, L. New family of fluorinated polymer chips for droplet and organic solvent microfluidics. Lab Chip 2011, 11, 508–512. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Kim, J.-O.; Kim, H.; Ko, D.-H.; Min, K.-I.; Im, D.J.; Park, S.-Y.; Kim, D.-P. A monolithic and flexible fluoropolymer film microreactor for organic synthesis applications. Lab Chip 2014, 14, 4270–4276. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Nightingale, A.M.; Hassan, S.; Makris, K.; Bhuiyan, W.T.; Harvey, T.J.; Niu, X. Easily fabricated monolithic fluoropolymer chips for sensitive long-term absorbance measurement in droplet microfluidics. RSC Adv. 2020, 10, 30975–30981. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Yin, B.-F.; Wan, X.-H.; Yang, M.-Z.; Qian, C.-C.; Sohan, A.S.M.M.F. Wave-shaped microfluidic chip assisted point-of-care testing for accurate and rapid diagnosis of infections. Mil. Med. Res. 2022, 9, 8. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Mair, D.A.; Geiger, E.; Pisano, A.P.; Fréchet, J.M.J.; Svec, F. Injection molded microfluidic chips featuring integrated interconnects. Lab Chip 2006, 6, 1346–1354. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Attia, U.M.; Marson, S.; Alcock, J.R. Micro-injection moulding of polymer microfluidic devices. Microfluid. Nanofluid. 2009, 7, 1–28. [Google Scholar] [CrossRef] [Scilit]
- Sha, B.; Dimov, S.; Griffiths, C.; Packianather, M.S. Investigation of micro-injection moulding: Factors affecting the replication quality. J. Mater. Process. Technol. 2007, 183, 284–296. [Google Scholar] [CrossRef] [Scilit]
- Jiang, B.; Zhu, L.; Min, L.; Li, X.; Zhai, Z.; Drummer, D. Characterization of microchannel replicability of injection molded electrophoresis microfluidic chips. Polymers 2019, 11, 608. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Attia, U.M.; Alcock, J.R. An evaluation of process-parameter and part-geometry effects on the quality of filling in micro-injection moulding. Microsyst. Technol. 2009, 15, 1861–1872. [Google Scholar] [CrossRef] [Scilit]
- Chen, S.-C.; Lin, C.-Y.; Chang, J.-A.; Minh, P.S. Gas-assisted heating technology for high aspect ratio microstructure injection molding. Adv. Mech. Eng. 2013, 5, 282906. [Google Scholar] [CrossRef] [Scilit]
- Wang, Y.; Weng, C.; Fei, Z.; Sun, H.; Jiang, B. Enhancing structural replication of microfluidic chips: Parameter optimization and mold insert modification. Polym. Eng. Sci. 2024, 64, 2082–2095. [Google Scholar] [CrossRef] [Scilit]
- Stampone, B.; Giorleo, L.; Trotta, G. The behaviour of micro-injection moulding inserts produced with material jetting technology. Compos. Part B Eng. 2025, 288, 111913. [Google Scholar] [CrossRef] [Scilit]
- Daikin Industries Ltd. NEOFLON PFA AP-201 Technical Data Sheet, Version 02; Daikin Industries Ltd.: Osaka, Japan, 2019. [Google Scholar]
- Hansen, H.N.; Hocken, R.J.; Tosello, G. Replication of micro and nano surface geometries. CIRP Ann. 2011, 60, 695–714. [Google Scholar] [CrossRef] [Scilit]
- Bellantone, V.; Surace, R.; Fassi, I. Experimental uncertainty evaluation in optical measurements of micro-injection molded products. J. Manuf. Mater. Process. 2024, 8, 21. [Google Scholar] [CrossRef] [Scilit]
- Moayyedian, M.; Abhary, K.; Marian, R. Optimization of injection molding process based on fuzzy quality evaluation and Taguchi experimental design. CIRP J. Manuf. Sci. Technol. 2018, 21, 150–160. [Google Scholar] [CrossRef] [Scilit]
- Park, J.; Lee, B.-K. Taguchi analysis of effects of processing parameters on macro/micro-scale deformation behavior of injection-molded microfluidic chips. J. Korean Soc. Precis. Eng. 2025, 42, 381–389. [Google Scholar] [CrossRef] [Scilit]
- Schneider, C.A.; Rasband, W.S.; Eliceiri, K.W. NIH Image to ImageJ: 25 years of image analysis. Nat. Methods 2012, 9, 671–675. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Baruffi, F.; Parenti, P.; Cacciatore, F.; Annoni, M.; Tosello, G. On the application of replica molding technology for the indirect measurement of surface and geometry of micromilled components. Micromachines 2017, 8, 195. [Google Scholar] [CrossRef] [Scilit]
- Zhang, H.; Fang, F.; Gilchrist, M.D.; Zhang, N. Precision replication of micro features using micro injection moulding: Process simulation and validation. Mater. Des. 2019, 177, 107829. [Google Scholar] [CrossRef] [Scilit]
- Trotta, G.; Cacace, S.; Semeraro, Q. Optimizing process parameters in micro injection moulding considering the part weight and probability of flash formation. J. Manuf. Process. 2022, 79, 250–258. [Google Scholar] [CrossRef] [Scilit]
- Baruffi, F.; Calaon, M.; Tosello, G. Micro-injection moulding in-line quality assurance based on product and process fingerprints. Micromachines 2018, 9, 293. [Google Scholar] [CrossRef] [Scilit] [PubMed]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.