Abstract
The smart stent, often referred to as a sensor-integrated stent, extends the function of the stent scaffold from mechanical support to a wireless sensing platform. This has the potential to reduce the delays inherent in current surveillance modalities by enabling continuous or on-demand monitoring within the lumen. Research on stent-based monitoring has followed separate routes, which created a need for a structural overview of development trends. This paper reviews the smart stent for vascular and non-vascular applications across five interdependent subsystems: sensing, data and power transmission, metallization and insulation, sensor–stent integration, and material platform. Within each subsystem, we organize the relevant approaches according to their defining design feature, describing the governing principles, analyzing the effects of design parameters on performance, and identifying the factors that influence the interpretation of reported performance. Based on this analysis, we discuss the need to develop common criteria for measurement conditions and values, facilitating comparison across studies.
1. Introduction
Since the first coronary stent implantation by Puel and Sigwart in 1986 [1], stent technology has evolved dramatically. Bare-metal scaffolds evolved into drug-eluting stents that reduced restenosis rates from ~30% to <10% [2,3], followed by bioresorbable and polymer-free designs [4,5]. Today, stents are deployed not only in coronary, peripheral, and carotid arteries but also in urinary, gastrointestinal, and airway systems [6]. The clinical demand for these devices spans multiple disease burdens: cardiovascular diseases caused 19.8 million deaths worldwide in 2022 [7,8], while gastrointestinal and pulmonary conditions requiring stent intervention—including esophageal, biliary, and airway obstructions from malignancies or strictures—continue to rise with aging populations and increasing cancer prevalence [9,10]. Together, these indications sustain a global stent market of more than ten billion US dollars annually [11].
Current post-stent surveillance primarily relies on noninvasive imaging modalities such as coronary computed tomography angiography, stress echocardiography, and magnetic resonance imaging (MRI) to assess stent patency and detect complications [12,13,14,15]. In select cases, invasive catheter-based measurements including fractional flow reserve (FFR) and intravascular ultrasound (IVUS) provide more detailed hemodynamic and anatomical assessments [16,17]. However, these approaches present critical limitations. Imaging methods are typically performed at scheduled intervals or when symptoms arise, limiting their ability to detect early pathological changes [18,19]. Catheter-based assessments are also confined to the peri-procedural period because routine surveillance is not recommended due to procedural risks, cumulative radiation exposure, and cost [20,21]. Consequently, impaired patency is often detected with substantial delays between onset and identification. In the bare-metal-stent era, by six months after percutaneous coronary intervention (PCI), ~23% of patients had already developed restenosis (≥50% diameter stenosis), with more than half remaining asymptomatic [22,23]; even with drug-eluting stents, a substantial proportion of in-stent restenosis remains clinically silent until it becomes flow-limiting [24,25]. In contrast, stent thrombosis rarely occurs (1.3% of patients at 9 months) but typically presents as an acute event with a higher case-fatality rate (~22–45%) [26,27,28]. Conventional stents cannot address these limitations: they only provide mechanical support but remain blind to the internal environment [29].
A smart stent is defined as a stent-integrated system capable of continuous or on-demand sensing of physiological parameters and wireless transmission of the resulting data to an external receiver [30,31,32,33,34]. This converts the stent from a passive scaffold into an active sensing platform situated closest to where restenosis, thrombosis, and other complications originate [31,35]. Smart stents are designed to provide continuous, real-time, non-invasive luminal monitoring post-implantation for the early detection of in-stent restenosis (ISR) and other stent-related complications [31,34,36]. This capability would allow preemptive intervention rather than delayed response to advanced damage, ultimately transforming post-implantation care from reactive, symptom-based management to proactive, data-informed monitoring [24,37].
Several reviews of smart stents have been surveyed from the perspective of clinical application [38,39], grouped devices by monitoring target and degradability [35], or reviewed ISR monitoring and treatment approaches with their principles, performance and limitations [40,41]. Previous works state individual device design parameters and performance rather than handle the comprehensive smart stent structure within each engineering subsystem to the reported performance [35,40,41]. Therefore, in this paper, we review recent advances and challenges in wireless stent-integrated monitoring from a subsystem-level view. The merit of this perspective is that it addresses the smart stent as an integrated system, providing direction on how the device should be configured and advanced, and how its performance should be reported and compared. Section 2 first introduces the inductor–capacitor (LC) resonant architecture shared by most platforms and then examines the five subsystems in turn: the sensing element (Section 2.1), transmission of power and data (Section 2.2), metallization and insulation (Section 2.3), the integration scheme that joins sensor to scaffold (Section 2.4), and the material platform (Section 2.5). Section 3 discusses cross-cutting performance metrics, proposes common reporting criteria, and outlines future research directions.
2. Stent-Based Monitoring
Smart stents can support continuous surveillance of hemodynamic and structural variations at the intervention site, serving as a long-term wireless sensing platform deployable via minimally invasive procedures. Realizing this potential requires concurrent advances across multiple engineering domains and in manufacturing, because improvements in one domain may compromise performance in another (Figure 1). The sensing element must transduce physiological deviation into a measurable signal with sufficient sensitivity and selectivity while maintaining a minimal footprint compatible with catheter-based delivery. The wireless power and data transmission subsystem (hereafter, the wireless subsystem) must receive power from an external source and transfer data to it, both through lossy biological tissue. Frequency-dependent electromagnetic attenuation, dielectric loading, and motion-induced coupling variation within the tissue collectively degrade power delivery efficiency and signal quality as implantation depth increases [30,33,42,43]. Conformal coatings must simultaneously preserve electromagnetic performance and meet chronic implantation safety standards [30,44,45], fabrication processes must yield devices that survive the extreme mechanical deformations of crimping, catheter delivery, and balloon expansion [31,32,46], and the choice of stent material platform must balance mechanical support against electromagnetic transparency to ensure reliable wireless readout [36,47].
Figure 1.
Schematic illustration of the subsystem covered in this review. The wireless sensing platform integrates five interdependent engineering domains (Section 2.1, Section 2.2, Section 2.3, Section 2.4 and Section 2.5)—the physiological sensor (Section 2.1), wireless power and data link (Section 2.2), biocompatible conformal coating (Section 2.3), sensor–stent joint (Section 2.4), and structural scaffold (Section 2.5)—to enable continuous post-intervention surveillance via inductive readout.
The stentenna concept, in which the metallic stent scaffold itself serves as a resonant or radiating structure, provides the foundational architecture for wireless stent-based monitoring [48,49,50,51]. The metallic scaffold functions as an inductor (L) and a parasitic capacitance or integrated capacitive sensor serves as the capacitor (C), yielding a resonant frequency determined by f = 1/(2π√LC) that shifts in response to physiological changes modulating the sensor capacitance [48,49,50,51]. An external reader coil generates an alternating magnetic field that inductively couples with this implanted LC tank, producing a detectable dip in the reflection coefficient (S11) or a phase shift in the impedance spectrum at the resonant frequency, from which the measured variable is extracted [30,31,33,48,49,51]. The original stentenna concept emphasized the scaffold’s dual role as both mechanical support and wireless antenna [48,49,51], and subsequent developments extended this perspective to encompass stent structures incorporating discrete LC resonant systems independently mounted on or embedded within the stent [30,31]. This LC architecture is mainly realized by using the stent scaffold itself as the inductor and mounting or embedding a capacitive sensor onto a conventional stent. In this structure, the coil need not be confined to the sensor footprint and can extend along the entire stent, which can strengthen magnetic coupling with the reader coil and extend the readout range [33,52,53]. In a configuration with an attached discrete LC resonator, however, the stent itself need not be modified [54,55], so established stents can be used in their existing form, with their mechanical behavior and manufacturing processes unchanged [56]. The coil, capacitor geometry, substrate material, and transduction mechanism of the resonator can then be optimized independently of the scaffold to match the target measurand [34,56,57]. Impedance-based, piezoelectric, and electrochemical sensing are also being investigated for stent-based monitoring, although they remain less widely explored. The sensing mechanisms, communication architectures, signal quality preservation strategies, sensor–stent integration methods, and material platform alternatives that collectively constitute this overall design space are examined in the sections that follow.
In coronary arteries, the primary implantation site for smart stents [31,58,59], the stent is located approximately 5 cm from the anterior chest wall, on the beating heart, where it is exposed to pulsatile, high-pressure blood flow in a conductive coagulation-prone medium [60] and to approximately 105 cycles of radial distension per day [61]. These conditions require thin, hemocompatible layers [59], a wireless subsystem that tolerates blood-induced loss and cardiac motion [31,58], and battery-free operation suited to permanent implantation [59]. In the ureter, urine is transported intermittently by peristalsis, with a low baseline pressure and transient pressure peaks during contraction [62,63], and urinary solutes can precipitate on the stent surface, while bacteria adhere to and colonize the surface, contributing to encrustation and biofilm formation, respectively [64]. The sensing element must resolve small pressure changes, and the device must resist encrustation [64]. In the airway, the stent is located in a space filled with air and remains in direct contact with the surrounding airway wall. The lumen deforms during respiration, with forced expiration reducing its cross-sectional area by approximately 50% in healthy adults. [65,66]. Respiratory motion can cause repetitive friction between the stent and the airway wall, potentially contributing to tissue injury and granulation tissue formation [67,68]. Airway stent placement is also associated with bacterial colonization, mucus plugging, and stent migration [69]. Because stents impair the natural clearance of airway secretions, secretion layer status, stent patency, and stent position are key sensing targets [68,70,71,72]. In particular, given the secretion retention and mucoid impaction associated with airway stents, monitoring secretion accumulation is important for assessing stent patency and airway conditions [70,71,73].
2.1. Sensing
Because the stent scaffold provides a nominally fixed inductance—though in practice inductance shifts modestly with deployment geometry and surrounding medium—the measurand in LC-based stent sensors is encoded primarily in the capacitance, governed for a parallel-plate geometry by C = ε0εrA/d [30,33,46]. This expression defines a design space in which the structural design of LC stent sensors is derived from the combination of inter-electrode gap d, dielectric environment εr, and effective electrode area A. The interplay among these variables determines the trade-offs between transduction sensitivity, deployment survivability, and signal stability in the biological environment. Pressure-induced deflection of diaphragm structures reduces the inter-electrode gap [30,36,42,44,53,74,75] (Section 2.1.1); compressive strain-induced deformation of microstructured elastomers modulates d and the effective dielectric environment εr [33,58,76,77,78] (Section 2.1.2); medium ingress into dielectric regions alters permittivity εr and the effective electrode area A via local dielectric replacement [71,79] (Section 2.1.3); hydraulic amplification structures convert pressure-induced reduction in gap d into discrete changes in the effective electrode area A (Section 2.1.4); variation in the interdigitated electrode area A reflects the attenuation of pulsatile arterial wall strain associated with restenosis (Section 2.1.5). These mechanisms can be extended and integrated into diverse hybrid configurations. Impedance-based sensing (Section 2.1.6) lies outside this capacitance-modulation framework: rather than modulating the energy stored in a resonant circuit, it applies a frequency-swept excitation through electrodes in contact with the medium and measures the complex impedance, whose resistive (dissipative) and reactive (storage) components reflect the electrical properties of cells, tissue and blood. Piezoelectric sensing (Section 2.1.7) and electrochemical sensing (Section 2.1.8) are routes outside this framework: the former converts mechanical stress into charge, and the latter converts chemical species into a current or a potential. The sections that follow examine representative implementations across this design space (Figure 2).
Figure 2.
Schematic illustration of sensor implementation in smart stent. A sensing element implemented on the stent transduces physiological changes either via capacitance modulation (C = ε0εᵣA/d), through diaphragm deflection (Section 2.1.1), microstructured dielectric deformation (Section 2.1.2), medium ingress (Section 2.1.3), hydraulic amplification (Section 2.1.4), and interdigitated strain sensing (Section 2.1.5), or via alternative routes including impedance (Section 2.1.6), piezoelectric (Section 2.1.7), and electrochemical (Section 2.1.8) sensing.
2.1.1. Δd Modulation: Pressure-Based Intravascular Sensing
Gap modulation is the most reported transduction route in stent-mounted LC sensors. The typical sensor consists of a rigid base electrode, a sealed cavity, and a thin deflectable diaphragm, implemented either as a micromachined plate [42,44] or as a flexible polymer membrane [53,74]. The cavity defines the initial gap and the volume available for deformation [30,44,53,74]. Applied pressure deflects the diaphragm toward the fixed electrode, and because the deflection is proportional to the fourth power of its in-plane dimensions and inversely to the cube of its thickness, diaphragm geometry is the main design consideration in this sensor [30,42,44]. The available deflection range is limited by the cavity depth, which constrains the displacement of the diaphragm [30,53,74]. The reference cavity sets a second constraint, as the trapped air opposes deflection upon compression; therefore, increasing the sealed air volume or vacuum sealing the cavity can recover sensitivity [30,44].
The dead-end hole structure adds eight holes arranged around the perimeter of the cavity to increase the sealed air volume, enhancing diaphragm deflection 6–7-fold (Figure 3a). The electrode area removed by the holes has little effect on the capacitance change, whereas the pressure sensitivity of chips with dead-end holes rose to 100 ppm/mmHg from 48 ppm/mmHg for chips of the same batch without them [44]. Patterning the deflective plate has also been considered: in a simulation study, a square plate containing four straight slots yielded a three times steeper deflection–pressure slope than a simple square plate [42]. A coating deposited over the diaphragm restricts its deflection [30].
Figure 3.
Gap-distance (Δd) modulation in stent-integrated capacitive pressure sensing. (a) Cross-section of a stainless-steel capacitive pressure sensor chip [44]. Drawn based on Ref. [44]. (b) Resonance frequency of the stent in air, DI water and Dulbecco’s phosphate-buffered saline (DPBS) [53]. Reproduced with permission from Ref. [53]. Copyright 2023, American Chemical Society. (c) Layer composition of the pressure sensor with a printed dielectric, before and after applying pressure [33]. (d) Relative capacitance change against pressure for patterned and solid dielectrics [33]. (e) Pressure response at bending radii from flat to 0.25 mm [33]. Reproduced with permission from Ref. [33]. Copyright 2022, The Authors, published by AAAS. (f) In-stent restenosis model in which hyperplastic tissue accumulates inward [79]. Drawn based on Ref. [79]. (g) Schematic of hydraulic capacitive pressure sensor [46]. (h) Main operating principle of the hydraulic capacitive pressure sensor (Ctotal: total output capacitance): displacement amplification via the hydraulic ratio between the chamber and the channel [46]. (i) Measured stepwise capacitance change against liquid displacement and applied pressure [46]. Reproduced with permission from Ref. [46]. Copyright 2021, IEEE. (j) Assembled sliding-plate strain sensor with close-ups of the fingers without and with strain [59]. (k) Relative capacitance change against strain for the shifting of plates with and without strain [59]. Reproduced with permission from Ref. [59]. Copyright 2022, Elsevier.
Pressure is derived from the resonant frequency shift read at the external antenna, which is defined by the inductance of the on-stent antenna and by the total capacitance of the tank [30,36,75]. That total capacitance comprises the response of the sensor together with the capacitance added by the surrounding medium [30,36,74]. The pressure sensor chip embedded in the intelligent ureteral stent platform has an equivalent series resistance (ESR) of up to 150 Ω, which lowers the quality factor of the LC tank and thus the depth of the resonance dip [36,75]. A compensation capacitor connected in parallel with the sensor reduces the overall ESR of the capacitive elements, but the addition also raises the total capacitance of the tank and therefore lowers the frequency response to pressure [36,75]. Given this trade-off, the compensation capacitance was reduced from about 60 pF in the 2021 prototype to 35 pF in the 2023 design, and the resulting decrease in total capacitance was likely a major reason for the increase in response, for example, from about 1.0 to 2.4 kHz/mmHg in air [36,75].
In vivo studies have been reported for these sensor mechanisms: in a swine bypass-graft model, frequency shifts (e.g., −260 kHz) were detected when blood flow was restricted downstream or upstream of the device, and in a rat, whose femoral artery was too narrow for the crimped stent, the device was used as a 2D planar mesh into a stent and was held in contact with the outer surface of the artery using a tissue adhesive; it resonated at 136 MHz during diastole and 129 MHz during systole [30,53]. The baseline resonant frequency differs from device to device, reported in air at 81.9 MHz for a gold-coated stainless-steel stent expanded to 5 mm [30], at 448 and 67 MHz for two self-rolling polymer devices [53,74], and at 61.8 and 70.8 MHz through a wired interface for the intelligent ureteral stent [36,75]. Once a device is implanted, its baseline is fixed, so pressure is read as a shift from that device’s own baseline rather than from an absolute frequency. These values also decline in conductive media; for example, a polymer stent resonating at 67 MHz in air shifted to 22 MHz in deionized (DI) water and 18 MHz in saline [53] (Figure 3b). Because the shift produced by the medium (tens of MHz) is far larger than that produced by physiological pressure (tens of kHz), the medium in which a sensitivity value was obtained must be stated for the value to be interpretable.
2.1.2. Δd + Δε Dual Modulation: Microstructured Dielectric Sensors
Microstructured dielectric layers introduce a second transduction pathway that can amplify the capacitance response beyond what gap closure alone achieves [33,58,76]. When a patterned elastomer sits between two electrodes, applied pressure simultaneously reduces the inter-electrode gap and deforms each microstructure tip against the opposing electrode [33,58,76] (Figure 3c). As the contact footprint grows, regions that were previously air-filled gaps convert to solid polydimethylsiloxane (PDMS) contact, raising the effective permittivity of the dielectric [33,58,76]. Because both Δd and the contact-area-driven Δε increase capacitance in the same direction, the dual modulation is expected to produce greater sensitivity than either mechanism in isolation [33,58,76]. Direct experimental comparison supports this expectation: disconnected PDMS microstructures, which have interstitial space to deform under pressure, produced notably higher sensitivity and faster response than a solid PDMS film of comparable thickness, where the incompressible bulk resists both gap closure and contact area growth [33,76,77,78] (Figure 3d). A distinctive feature of this approach is that the contact-area growth is pressure-dependent, producing a nonlinear response: at low pressures, deformation produces a relatively large change in contact area, whereas at higher pressures, increasing elastic resistance limits deformation and reduces the incremental change in contact area per unit pressure. The location of this high-sensitivity window is determined by microstructure geometry and modulus: generic micropyramid implementations without structural tuning exhibit their highest sensitivity within a sub-kilopascal regime, well below arterial pressures [76,78], whereas the pyramid dimensions used in implantable pressure-sensor designs were adjusted to shift the transition into the clinically relevant arterial pressure range [33,58]. No closed-form analytical model linking microstructure geometry to the pressure–capacitance curve has been reported to date, leaving the selection of pyramid height, spacing, and modulus as an empirically guided process [33,58,76].
Aerosol-jet-printed PDMS dielectric lines between silver nanoparticle (AgNP) electrodes sustained 0.013 kPa−1 average sensitivity over 2500 cycles spanning 0–1000 mmHg in bench testing, with a pressure resolution of approximately 5 mmHg [33]. The thin printed layers conformed to a bending radius of 1.5 mm without measurable sensitivity loss and maintained pressure sensing down to a bending radius of 0.25 mm, approximately 20 times smaller than prior capacitive sensors, indicating that the transduction mechanism survives the curvature imposed by stent deployment [33] (Figure 3e). After balloon expansion at 14 atm for 30 s, the sensor exhibited a 3.5% baseline capacitance shift that stabilized within 60 s [33], suggesting that the microstructure undergoes limited plastic rearrangement during deployment but retains functional transduction. An alternative implementation using PDMS micropyramid dielectrics with far-field radar backscatter readout (Section 2.2) produced approximately 5 MHz frequency shift per 50–60 mmHg with 7% baseline drift over 1000 cycles in air [58]. Across both implementations, the absence of a predictive contact-mechanics model remains a significant barrier to systematic optimization of the Δd + Δε transduction pathway [33,58,76].
2.1.3. Δε + ΔA Coupled Modulation: Permittivity-Based Sensing Through Medium Ingress
Unlike conventional sensors that transduce mechanical stimuli, permittivity-based approaches exploit the surrounding luminal dielectric environment as a variable capacitive load to detect obstructions [71,79]. This capacitive modulation is coupled: as the obstructing medium accumulates inward, both the permittivity of the medium in contact with the stent’s electromagnetic field changes (Δε) and the cross-sectional area fraction occupied by each medium shifts (ΔA), so the capacitive loading reflects a simultaneous change in two parameters [71,79] (Figure 3f). Under the baseline tissue parameters adopted in [79], blood and neointimal tissue were assigned simplified relative permittivities of approximately 64 and 47, respectively—where blood properties remain frequency- and hematocrit-dependent, and the neointimal value was approximated from bulk vessel wall properties rather than measured directly on restenotic tissue [79,80,81]. A minimally modified clinical ring stent demonstrated approximately 15 MHz shift per 10% cross-sectional area occlusion in simulation, with the stent resonance falling between 457 and 515 MHz and the external interrogation coil tuned near 350 MHz [79]. The response saturates only once the neointimal layer exceeds approximately 2.8 mm in thickness, so the clinically significant threshold of 50% area occlusion—corresponding to 1.5 mm of hyperplasia—remains well within the monotonic region of the transfer curve [79]. For airway applications, a coplanar capacitor within a silicone stent tube detected mucus-like filling with a logarithmic capacitance response, but exhibited non-monotonic behavior with solid obstructions, indicating that heterogeneous obstruction geometry redistributes the electric field in ways that uniform-medium models cannot capture [71]. Neither implementation has demonstrated wireless in vivo readout.
The dominant limitation is sensitivity to biological variability of the dielectric environment. Without patient-specific calibration, false-negative rates as high as 35.65% were predicted in simulation for a ±4 unit variation in tissue permittivity [79], because the transduction depends on absolute dielectric values rather than mechanical displacement [71,79]. Actual neointimal tissue is a heterogeneous mixture of smooth muscle cells, collagen, and proteoglycan matrix whose aggregate permittivity has not been directly measured at the relevant frequencies [79,80]. Reducing this uncertainty to clinically acceptable levels would likely require either individualized baseline calibration or differential measurement schemes designed to cancel patient-specific offsets.
2.1.4. Δd → ΔA Cascade Modulation: Hydraulic Amplification
All preceding categories couple the physical stimulus directly to the capacitance-modulating element, bounding sensitivity to what a single-stage conversion can deliver. A cascade architecture addresses this by physically separating the pressure-sensing membrane from the capacitive readout electrodes and connecting them through a hydraulic intermediary (Figure 3g,h): membrane deflection displaces a high-permittivity liquid (εr ≈ 42) through a microchannel, and the advancing meniscus replaces a low-permittivity medium above successive coplanar electrode pairs, producing a discrete capacitance step at each crossing with reported amplification ratios on the order of 102 [46]. This physical-domain amplification is significant because it elevates the signal amplitude before the point in the transduction chain where parasitic losses, coupling attenuation, and medium-induced dissipation would otherwise compress a small capacitance change below the readout noise floor [46]. Among stent-sensor architectures reported to date, this is the only design that decouples the sensing and readout surfaces through a fluidic link [46]. The stepwise encoding provides a form of noise rejection in that sub-threshold pressure fluctuations produce no output until the meniscus crosses the next electrode boundary (Figure 3i), though the effectiveness of this rejection depends on the meniscus pinning stability and contact angle hysteresis. A representative implementation demonstrated 11.8 kHz/mmHg wireless sensitivity in a liquid medium, with each transduction stage described by closed-form analytical expressions [46].
The fluidic intermediary that enables amplification also defines the dominant barriers to adoption. The working liquid must maintain stable viscosity, wetting, and volume at body temperature over a monitoring period spanning months to years, a condition that has not been demonstrated beyond fabrication-stage characterization [46,82]. Material biocompatibility of the inductor and electromagnetic attenuation at the operating frequency are also considered for applicability [46].
2.1.5. ΔA Modulation: Strain-Based Restenosis Monitoring
Pressure-based sensors respond to the hemodynamic consequences of luminal narrowing, which become measurable only after the obstruction has reached functional significance. Capacitive strain sensors are designed to access an earlier pathological stage by detecting the progressive wall stiffening that is expected to precede flow-limiting stenosis, and they retain this access once tissue grows over the device, since strain transduction does not depend on interfacing with blood flow [83]. Reported implementations rely on a sliding mechanism between two overlapping plates [59,83] (Figure 3j,k). Applied strain displaces the plates in opposite directions, reducing the finger overlap area and thereby lowering the capacitance [59,83]. The sensor is connected across an inductive stent to form an LC circuit, which is read through the S11 minimum [59,83]. Because the modulation is geometric rather than dielectric, it escapes the gauge factor of approximately unity that bounds Poisson-effect capacitive strain sensors [59,83], and sensitivity is governed by the geometry of the overlapping fingers and of the plates that carry them [59]. A gauge factor of 3.7 was obtained at a finger width of 0.75 mm, increasing to 10.5 as the finger width decreased to 0.25 mm [59]. Sensitivity also scales with how far each plate extends beyond its fingers; plates extending 1.0, 1.5, and 2.0 mm beyond the fingers showed progressively larger relative capacitance change per unit strain [83]. Reducing either the printed conductive line width or the base thickness of the finger structure raises sensitivity, because both shrink the initial overlap area, so that a given sliding displacement removes a larger fraction of it [83]. The two implementations [59,83] originate from the same group and share the sliding-plate concept; the later sensor [83] produced an average capacitance change of 18% at 5% strain, compared with less than 10% for the earlier outer-mounted design [59] and was mounted on the inner stent surface to prevent the wrinkling and detachment of outer-mounted sensors when pushed through a guide catheter. In embedded silicone artery models, wall stiffening reduced the amplitude of the cyclic strain and hence of the wireless response, which fell monotonically across simulated stenosis levels: normalized resonance change decreased by 35% at 60% restenosis and 68% at 90% [59], while the fractional resonance change fell from 2.5% in a healthy artery to below 1% at 75% restenosis [83].
Most practical limitations of ΔA sensing arise from its dependence on the relative position of the two plates, which is set during assembly and can change during implantation. Baseline capacitance averaged 5 pF but ranged from 2 to 10 pF because the fingers were aligned and sealed by hand, and pressing them together with a laminating press during sealing reduced this variability [59]. The fingers are aligned at zero strain, where capacitance is maximal [83], so the sensor rests at the symmetric midpoint of its response with no margin against a shift opposite to the strain-induced sliding. Such a shift occurred in some devices during implantation into the artery model owing to the softness of the sensor [59]. Because of this shift in the opposite direction, strain moved the fingers back toward alignment, so capacitance increased with small strains instead of decreasing, and the wirelessly measured resonance shift was inverted as well [59]. Inner mounting prevented the wrinkling and detachment described above, although finger alignment after deployment was not reported [83]. It also places the sensor on the luminal side of the struts, and because the strain response was measured only after the device had been embedded in the silicone wall [83], how the sensor couples to wall strain before tissue grows over it has not been characterized. Neither study addressed how to detect or prevent signal inversion. We suggest assembling the plates with a deliberate initial offset in the direction of strain-induced sliding and forming a mechanical stop on one plate that limits travel back toward the midpoint. This would keep a shift during implantation from crossing the midpoint at the cost of a lower maximum detectable strain.
2.1.6. Impedance-Based Sensing
Most of the described LC-based sensing routes identify the biological change through shifts in a single scalar observable, the resonance frequency. Impedance-based detection instead uses frequency as an input: an alternating excitation is applied through electrodes in contact with the medium and swept across a band, and the impedance is measured as a function of frequency [43,45,84] (Figure 4). Therefore, the results are complex spectra across the swept band, with reported frequencies ranging from 1 kHz to 1 MHz. The electrical discrimination of vascular cell types associated with restenosis was investigated, where the post-confluence reactance characterization showed maximum declines of 78% and 32% at 10 kHz in porcine endothelial cells and human umbilical vein endothelial cells (HUVECs), respectively [43] (Figure 4a–c). The same measurement was implemented on a nitinol stent using microfabricated electrodes, which detected a smooth muscle cell monolayer and extended to clotted blood and to calcified and lipid regions of human carotid plaque [84] (Figure 4d,e). A telemetry package showing 89.5–91.8% agreement with the LCR meter enabled wireless readout and thereby facilitated miniaturization [84]. A subsequent study extended the sensing concept to the stent itself by selectively metallizing it into two electrode sets covering the intraluminal space, so that detection was distributed across the device rather than confined to an integrated sensor site. In an ex vivo experiment using a porcine aorta, the stent detected clot-induced occlusion through a 16 ± 3% increase in impedance [45] (Figure 4f).
Figure 4.
Swept-frequency impedance spectroscopy and its frequency-dependent limits for electrode-based detection of vascular cells, plaque and clot. (a) Impedance modulus and phase angle captured in EC media, obtained from swept measurements between 1 kHz and 100 kHz. Data represent the mean ± SEM (gold n = 10, platinum n = 16) and not for phase measurements; max gold SEM 0.771° and max platinum black SEM 1.076° [43]. (b,c) Mean reactance profiles at 10 kHz for porcine ECs and HUVECs, measured in cell-seeded and cell-free electrodes. Media change and ±SEM are represented by the red line and error bars [43]. Reproduced with permission from Ref. [43]. Copyright 2018, the Authors, published by PLOS. (d) Impedance analysis from 1 kHz to 1 MHz of calcified plaque and lipid core, classified as mild (left), moderate (middle) and severe (right) atherosclerosis [84]. (e) Impedance tested at 20 kHz measured under a 50,000 smooth muscle cell in two generations of miniaturized wireless impedance analyzer to LCR, N = 3, ±SE, ns = nonsignificant, one-way ANOVA, F value = 4.417, df = 8 [84]. Reproduced with permission from Ref. [84]. Copyright 2022, the Authors, published by Wiley-VCH GmbH. (f) The normalized area under the curve (AUC) of the impedance frequency sweep for three stents, each tested three times with and without clot surrounded by fresh porcine aorta and normalized to its unclotted measurement (control), clotting produced a 16 ± 3% increase in impedance (two-tailed one-sample t-test versus control; ** p = <0.001, p = 0.0090; n = 9) [45]. Reproduced with permission from Ref. [45]. Copyright 2024, the Authors, published by Wiley-VCH GmbH. (g) Variation in dielectric in real part (ε′) and imaginary part (ε″) [85]. Reproduced with permission from Ref. [85]. Copyright 2025, the Authors, published by The Royal Society of Chemistry. (h) Magnitude plot of blood impedance characterization for three sensors with different numerical conditions. The inset shows the plateau near the low-frequency cutoff represented in the dashed vertical line [86]. Reproduced with permission from Ref. [86]. Copyright 2020, the Authors, published by MDPI.
The implemented swept range carries information on cell shape and membrane integrity, whereas access to cytoplasmic and nuclear properties requires frequencies above 1 MHz; the capacitive barrier of the cell membrane is bypassed as frequency rises [85] (Figure 4g). Even within the 1 kHz–1 MHz range, the double-layer impedance induced in series with the sample impedance acts as a barrier for low-frequency measurements, hiding the first of the two plateaus that correspond to blood conductivity [86]. This barrier becomes negligible above the cutoff frequency that is determined by the electrode geometry [86]. When one blood sample was measured with three interdigitated sensors, only the sensor for which this plateau was visible returned a conductivity close to the literature value [86] (Figure 4h). Because the plateau appears in the spectrum itself, a swept measurement can indicate whether the acquisition range is valid and can yield a sample-specific dielectric value; this may offer a route to the patient-specific calibration identified in Section 2.1.3 for permittivity-based sensing, although the sensors in [86] were coplanar rather than facing each other across a lumen [45,71,79,86]. Exploiting this would require retaining the frequency structure, but the reported metric is the area under the impedance sweep, which reduces the band to a single value and obscures plateau presence [45,84]. Absolute impedance is not uniform across devices, so each stent requires its own baseline measurement after deployment, and while occlusion can be detected, its extent cannot be quantified [45,84]. Therefore, this detection method needs further investigation under flow conditions, and in particular a demonstration that the baseline remains stable as the electrodes become endothelialized [45,84].
2.1.7. Piezoelectrical Sensing
In piezoelectric sensing, mechanical stress from blood pressure or flow generates charge directly in the sensing film [32,87,88]. Unlike capacitive sensing, where pressure acts through a change in capacitance, this charge is produced directly by the stress. The film produces its own signal; pulsatile pressure or flow can be sensed without an external bias [32,87]. One piezoelectric stent was developed using a perforated polyvinylidene fluoride (PVDF) membrane with gold plating on both sides as electrodes, designed for integration into a stent graft for post-endovascular aneurysm repair (EVAR) surveillance [32]. This membrane harvested 0.23 mW from 14 MHz ultrasound, which was enough to drive the transmitter circuit (30 cm from the transducer) [32]. Wing-like structures patterned into the PVDF membrane (termed fins in the original study) served as flow sensors by converting induced surface stress into a voltage through the piezoelectric effect. Sensitivities of 0.16 mV/(mL/min) from 30 to 200 mL/min and 0.06 mV/(mL/min) from 201 to 980 mL/min were obtained (water flow in a tube) [32]. The voltage generated by the fins shifted the carrier frequency of the FM transmitter between 81 and 87.74 MHz, and the signal was detected by a spectrum analyzer at about 1 m [32].
In 2024, two studies from research teams with partially overlapping membership extended this design along different lines [87,88]. The first focused on the 3D print manufacturing technique [87]. The stent was 3D printed from PVDF-TrFE and poled during printing, so that its shape could be customized [87]. It showed a peak-to-peak voltage response of 0.943 mV/mmHg as the pressure in the aneurysm sac varied between 25 and 85 mmHg (oscilloscope readout, DI water, 80 BPM pulsatile flow) [87]. Wireless readout utilized the stent as an LC tank (coil-like stent with double-sided electrodes) through inductive coupling with an external coil [87]. Changes in dimensions induced by pressure altered mutual coupling, shifting the resonance frequency with a sensitivity of −2.2 kHz/mmHg [87]. The second adopted a different sensor integration and readout strategy [88]. A perforated PVDF film metallized with titanium and copper was rolled into a tube and mounted on a conventional stent graft as a modular sensor, with the tube itself serving as the resonator [88]. Resonance frequency was wirelessly detected through inductive coupling with an external spiral antenna, with a sensitivity of 687.5 Hz/mmHg (artificial blood, 3 cm of pork belly interposed between the stent and the antenna) [88].
A 2026 study from the same research groups addressed the short readout range of passive telemetry [89]. It incorporated an active transmission module into the smart stent system to support continuous, real-time data and long-range monitoring, thereby reducing the burden on patients and caregivers [89]. The voltage of a perforated PVDF tube was transmitted via a 433 MHz module on a flexible printed circuit board up to 100 cm in air [89]. The device showed an average diastolic pressure sensitivity of 3.35 mV/mmHg (in vitro AAA model, artificial blood) [89]. Type I–III endoleaks emulated in the same model were distinguished by their distinct changes in diastolic and peak-to-peak voltage. However, the device operated via an external power supply or battery, so its power supply remains to be addressed [89]. None of these piezoelectric stents has yet been tested in vivo [32,87,88,89].
2.1.8. Electrochemical Sensing
Electrochemical and biochemical sensors detect chemical species in the surrounding blood or vessel wall [90,91]. Unlike capacitive sensing, the signal reflects the chemical composition of the medium. Reactive oxygen species (ROS) and the resulting inflammation contribute to restenosis and in-stent thrombosis [92], so measuring such species at the stent could allow in situ monitoring of these complications. One study developed a stent-based wireless glucose monitoring system, in which the solenoidal stent body served as the antenna for near-field communication (NFC) [90]. A separate glucose tri-electrode converted glucose concentration into electrochemical current through the reduction of hydrogen peroxide, itself a ROS [90]. Interface electronics regulated and measured this current, and an NFC tag sent the data to an external NFC reader [90]. The system showed a sensitivity of 0.879 µA/mM for glucose detection [90]. Data was transmitted over a distance of up to 23.60 mm in air [90]. Only in vitro tests were reported [90]. Because the study was evaluated in vitro using a relatively simple glucose solution, potential interference from various electroactive species in blood and electrode surface changes caused by protein adsorption and platelet adhesion remain unassessed. In addition, blood flow can alter mass transport to the electrode, and RF attenuation in biological media may reduce the communication distance.
Another study targeting an inflammatory biomarker proposed a sensor attached to a stent to detect pentraxin-3 (PTX3) released by inflammatory cells in the arterial wall [91]. Diffusion modeling predicted a higher PTX3 concentration at the biosensor for vulnerable plaques than for stable plaques [91]. The study was limited to modeling, and neither the transduction method nor the communication link was addressed [91]. To our knowledge, electrochemical measurement of ROS, pH, or inflammatory markers on a stent has not yet been demonstrated experimentally. Nevertheless, independent biosensors could be incorporated into stent platforms to enable biochemical monitoring and functional coatings that detect intraluminal biochemical changes or respond to specific biochemical stimuli could provide additional sensing or controlled drug-release capabilities [93,94,95,96].
2.2. Transmission of Power and Data
This section addresses link sensitivity—the minimum detectable frequency shift at the external reader driven by the coupling geometry, tissue attenuation, quality factor, and readout method. Transmission is achieved by telemetry to extract frequency-encoded information from the stent and by wireless power transfer to supply sufficient energy to the implant, both operating across a lossy, dispersive biological channel. The majority of stent transmission mechanisms reported to date operate in one of two dominant electromagnetic regimes: near-field inductive coupling, in which the stent and reader exchange energy through mutual inductance at distances much shorter than the operating wavelength, or far-field radiative coupling, in which the stent functions as a transmitting or backscattering antenna. The subsections below treat near-field readout (Section 2.2.1), far-field operation (Section 2.2.2), and emerging multiplexed and active architectures (Section 2.2.3).
2.2.1. Near-Field Inductive Readout
Passive near-field inductive coupling remains the most widely adopted wireless readout strategy, with implementations spanning from the stentenna concept through the most recent designs. The reader drives a current through its coil and monitors the impedance or reflection coefficient as a function of frequency; at the implant’s resonant frequency, the coupled impedance produces a detectable perturbation—either a dip in |S11| or a zero-crossing in the impedance phase—from which the resonant frequency shift is resolved [30]. The depth of this perturbation depends on the coupling coefficient k, which for coaxial loops scales as approximately (r_stent/d)3 at distances exceeding the coil radii, imposing a steep penalty: doubling the separation reduces the coupled signal by approximately 8-fold. The quality factor of the implanted LC tank determines the sharpness of the resonance manifested in the reader’s impedance spectrum.
Q (Q = ωL/R, with R the dominant loss term in conductive media) traverses a wide envelope across the operating medium, ranging from a maximum of 39 in air to a minimum of 2.7 in physiological saline among reported gold-coated stainless-steel stent devices [30,31]. A low quality factor broadens and shallows the impedance dip [97]. Several strategies have been developed to extend the near-field readout range beyond this limitation. Reader–implant resonance matching (the technique of tuning the external coil to the implant’s resonant frequency) maximizes the fraction of source power delivered to the coupled system. With this approach, readout ranges of 5.5 cm in air and 3.5 cm in saline were achieved [33], an advance over earlier stent-based systems, which reached maximum distances of only 3 cm in air and 1 cm in blood [42,47,92].
This concept was further developed into resonance-coupled telemetry, in which the matched-frequency configuration maximizes the phase difference at the coupling point and a reduction in compensation capacitance raises the pressure-sensitive fraction from 11.5% to approximately 18%, deepening the S11 dip approximately 6-fold [36,75]. For the recently developed ureteral stent, an approximately 25-fold enhancement over the preceding 2021 design was reported under the artificial-urine condition [36,75]. An alternative approach exploits the temporal rather than spectral signature of the LC resonance: in transient resonance readout, a pulsed radio-frequency (RF) excitation is gated off before measurement, allowing the LC tank to ring down at its natural resonant frequency without direct feedthrough from the reader [97]. Separating excitation from measurement in time raised the readout range to 2.75 cm in air [97]. The resonant frequency is then estimated from the ring-down waveform, and the estimator (zero-crossing counting, spectral peak fitting, or damped-sinusoid fitting) determines both the resolution and the robustness to noise; systematic comparison of these estimators for stent applications has not yet been reported.
Passive LC sensors require no on-board power, but active multi-sensor systems demand continuous direct-current (DC) supply across the tissue barrier—a constraint governed by the specific absorption rate (SAR = σ|E|2/ρ, with σ the tissue conductivity and ρ the tissue density) of the surrounding medium [98]. Coupling modalities that impose electric fields in tissue generate disproportionately greater heating than magnetic-field links, in which tissue heating arises primarily from magnetically induced secondary electric fields. This asymmetry is illustrated by a comparison using a stent-based receiver 4 mm in diameter and 40 mm in length in muscle tissue [99]: in a finite element full-wave simulation, the inductive link at 4 MHz gave an SAR of 60 µW/kg for 500 mW of input power, from which it was estimated that 13 kW of input power would be required to reach the SAR threshold (1.6 W/kg averaged over 1 g of tissue), assuming a linear relationship, whereas the capacitive link, driven from 200 MHz to 400 MHz, was reported to be safe provided the input power was below 53 mW. This threshold differs from the local limit for the head and torso (2 W/kg averaged over any 10 g of tissue) in IEEE C95.1 (2019 edition) and the ICNIRP guidelines (2020 edition) [98,100]; the 1 g threshold was the IEEE C95.1 value for exposure of the public (editions before 2005) [98] and remains the FCC limit for general population/uncontrolled exposure [101]. Because both thresholds remain in use, the input limit varies with the threshold used in the calculation; for an endovascular aortic stent system at 915 MHz, the maximum allowable net input power was 3.89 mW under the 1 g threshold (1.6 W/kg) and 34.4 mW under the 10 g threshold (2 W/kg) [102]. At higher frequencies the absolute power margins compress further. For example, a serpentine endovascular aneurysm repair (EVAR) stent (operating at 868 MHz) achieved 4% wireless power transfer efficiency at 10 cm through heterogeneous tissue, with a rectifier converting received RF to DC at 85% efficiency [103]. Yet the SAR-limited transmit power is 15.17 mW (at 868 MHz) for this antenna (peak gain −12.1 dBi) [102,103]. Under this limit, the available DC power is at most 15.17 mW × 0.04 × 0.85 ≈ 0.52 mW, where 0.04 is the power transfer efficiency reported at 10 cm and 0.85 is the peak RF-to-DC efficiency reported at a 15 dBm input. This is sufficient for low-power sensor electronics but below the 1.5 mW continuous draw required for multi-sensor operation [104]. The far-field link budget is therefore constrained not only by antenna geometry or rectifier efficiency but also by the tissue safety envelope imposed on the upstream transmit power.
2.2.2. Far-Field Radiative and Backscatter Operation
When the stent scaffold becomes electrically large enough to sustain resonant current distributions, coupling can extend beyond mutual inductive interaction. For a conventional two-coil inductive link, the received power decreases approximately as d−6 because the magnetic dipole field decays as d−3. By contrast, in the radiative regime, the received power (equivalently, the incident power density) decreases approximately as d−2, so that the interrogation range is no longer limited by the steep near-field roll-off [105]. This transition has been observed experimentally in stent systems. Near-field implementations were read at 1.25–5.5 cm in air with steep signal decay [33,97]. These distances fell to 3.5 cm in saline and 3 cm through saline and tissue [33], and conductive media shortened the range by about 30% [97]. A tuned transient inductive readout system using pulsed excitation read a sensor coil with a ceramic capacitor at up to 8 cm in air, 7 cm in saline, and 6 cm when the coil was embedded in meat [54]. In vivo wireless readings from an inductive stent in a rabbit iliac artery were unreliable because of the small artery and its depth below the skin [33]. In contrast, far-field communication links have reached 50 cm in air [58]. In biological media, an EVAR stent antenna placed in a phantom transmitted to an external antenna over 0–2 m with received power consistent with the link-budget calculation, and received power at 4% efficiency at 10 cm and 1.6% at 15 cm [103]. These results are not directly comparable with near-field communication ranges measured in tissue. Where the stent itself acts as the radiating element, the resonant frequency is governed by its effective electrical length, which scales with the axial length of the scaffold and is further set by the winding topology and by the permittivity of the surrounding medium. Axial length dominates within a given design: lengthening the same stentenna from 20 to 40 mm shifts its band monotonically downward [79], and across platforms a coronary-scale stent of 18 mm resonates near 2 GHz in free space [106], a 30 mm modified clinical ring stent operates within the 401–406 MHz Medical Device Radiocommunications Service (MedRadio) and 433 MHz industrial, scientific, and medical (ISM) bands once embedded in tissue [79], and EVAR-scale aortic segments of about 120 mm fall within the 868 and 915 MHz ISM bands [102]. Changing only the connector topology at an identical 18 mm axial length moves the resonance from 2.07 to 2.97 GHz [106]. Platforms in which a lumped inductor and capacitor define the tank follow their circuit elements instead of the scaffold, as in the 100 mm EVAR graft read out at 2 and 12 MHz [104] and the 150 mm graft carrying a 100 mm planar inductor read out at 6–17 MHz [55].
Within the far-field regime, the radiation performance depends critically on the current path. Conventional multi-connector stents form meshed inter-ring loops that create closed current paths, distributing current irregularly and suppressing coherent radiation [102,106]. Reducing the connections to a single path enforces a half-wavelength resonant mode with coherent, omnidirectional radiation [106]. In EVAR geometries, a similar effect is achieved using an external feed wire [55,102]. A single-connector cobalt–chromium coronary stent achieved 74.5% radiation efficiency and 1.38 dBi gain at 2.07 GHz but this was measured in air and its performance inside the body remains to be examined [106]. In tissue-implanted conditions, EVAR stent antennas exhibited gains of −12 to −21 dBi (868–915 MHz), far lower than those measured in free space, reflecting absorption of radiated power in the surrounding tissue [102,103,106]. The progression from an earlier meshed-ring design to the serpentine geometry improved gain by 8.65 dB while simultaneously reducing peak SAR by a factor of 4.3, consistent with the serpentine geometry distributing current more uniformly along the stent length and reducing the peak current density that drives local tissue heating [102,103]. Monostatic radar backscatter circumvents SAR-limited onboard RF transmission through external interrogation, eliminating the need for implant-generated emission and relocating active signal generation to the reader [58]. In a 2 GHz coronary-scale implementation, this passive readout strategy achieved a readout range exceeding 50 cm in air through background subtraction and time gating, with quality factor maintained at 8.9 [58]. Whether comparable ranges can be maintained despite the additional 20–30 dB loss introduced by several centimeters of tissue has not yet been reported for this platform [58]. Furthermore, increasing the reader’s transmission power to compensate for tissue attenuation is constrained by the SAR limit discussed in Section 2.2.1, as the incident field also deposits energy in the tissue.
2.2.3. Multiplexed and Active Architectures
Most of the near-field and far-field architectures discussed above share a common limitation: each implant reports one scalar measurement, such as the resonant frequency of an LC resonator from one anatomical location. Multiplexed architectures and active architectures address this limitation. In an early multiplexed and active implementation for EVAR, 16 capacitive pressure sensors were mounted on a stent graft exterior, digitized and multiplexed through a wireless data link with inductive power at 2 MHz and data transmission at 12 MHz, achieving approximately 0.5 mmHg resolution at 1.5 mW consumption [104]. More recently, in an active single-site implementation for airway stents, spatial tracking of stent position has been shown using a 3-axis Hall-effect sensor that detects the quasi-static magnetic field from an external rotating permanent magnet, achieving translational tracking accuracy below 2% relative error at 3.5 cm [72]. In a passive multiplexed implementation, a recent dual-sensor smart stent addressed a different robustness axis: tolerance to reader–implant misalignment. The wireless readout maintained reliable pressure tracking under vertical displacements up to 10 mm, lateral offsets up to 6 mm, and rotational deviations up to 80° [34]. These implementations indicate potential for stent telemetry to move beyond single-point readout through spatially and functionally multiplexed and active architectures.
Section 2.1 and Section 2.2 show that the sensing mechanism and the telemetry architecture are two key determinants of smart stent performance. Table 1 shows the sensing mechanisms and wireless telemetry performance of the smart stents reviewed in these sections. Regarding sensing, sensitivity is reported in different metrics across studies, which makes direct comparison of sensor performance difficult. Regarding power and data transmission, implementations are based on passive near-field inductive readout, far-field backscatter, or active architectures. Moreover, in vivo validation in animal models has been reported in a few studies, and further animal studies are needed to confirm the performance of these devices under physiological conditions.
Table 1.
Summary of sensing mechanisms and wireless telemetry performance of smart stents.
2.3. Metallization and Insulation
2.3.1. Electroplating and Encapsulation
Reliable LC resonant sensing inside a blood vessel requires preserving the quality factor (Q) of the implanted circuit in the presence of surrounding biological media, which induce capacitive shorting between adjacent loops and elevate series resistance. 316 L stainless steel stents exhibit series resistances of 22–30 Ω, imposing significant resistive loss that limits the achievable resonator Q [30,31]. Two complementary strategies have been pursued: metallization of the conductive scaffold to reduce series resistance and conformal dielectric insulation using polymer to suppress inter-strut parasitic capacitance, thereby elevating Q to levels sufficient for reliable wireless readout [30,31].
Two metallization routes have been explored since the earliest stentenna implementations, each exploiting distinct material properties to reduce conductor resistance while ensuring biocompatibility. Gold electroplating and copper electroplating with biocompatible conformal coating represent distinct design philosophies. The former relies on an inherently biocompatible conductor, whereas the latter pairs a high-conductivity yet cytotoxic metal with a protective polymer barrier. Both approaches have developed in parallel and have recently converged into hybrid architectures that combine their respective advantages [58].
Gold electroplating reduces stent resistance by exploiting the skin effect, which concentrates RF current near the conductor surface at the 30–100 MHz operating frequencies of helical antenna stents, so that most of the current flows in a sufficiently thick gold layer whose conductivity is more than an order of magnitude higher than that of bare 316 L stainless steel [30,31,51]. The gold layer is electroplated to a thickness of 15–20 µm on top of a thin gold strike layer that promotes adhesion to the stainless steel, and a custom rotational set-up is used during plating in a cyanide-based bath which makes deposition of material uniform over the entire surface [30,31]. In the first design, gold electroplating to a thickness of 15 µm, which exceeds the skin depth of gold at 50 MHz (≈10 µm), lowered the stent resistance 20-fold from 30 Ω to 1.5 Ω [30]. Combining gold electroplating with laser microwelding enabled the device to resolve pressure changes of 12.4 mmHg in water flow, which is expected to be sufficient to detect the ~19 mmHg reduction gradient associated with arterial pressure that corresponds to the clinical fractional flow reserve threshold of 0.80 [31,107]. Furthermore, gold is inherently biocompatible and corrosion-resistant, so it remains stable in the body without releasing harmful ions [30,31].
Copper electroplating provides comparable or superior bulk conductivity at substantially lower material cost and higher deposition rates. This material has the advantage of compatibility with standard microelectromechanical systems (MEMS) fabrication processes [46,51,74]. However, copper ions are cytotoxic when released into biological tissue, creating a need for conformal biocompatible coatings that simultaneously function as electrical insulation and a biological barrier [36,74]. Parylene-C deposition on the Cu layer has emerged as the standard solution, serving this dual electromagnetic and biological function [36,74,75]. Cu conductor thicknesses range from approximately 3 μm for thin-film MEMS coils to 100 μm for groove-filled antenna structures, with Parylene-C coatings of 1–25 μm depending on the required insulation performance and mechanical flexibility [36,74,75]. The self-rollable smart stent employed approximately 3 μm electroplated Cu coils with Parylene-C passivation, achieving cardiomyocyte viability above 95% and a hemolytic index of 0.13%—well within the ASTM F756-13 nonhemolytic criterion (<2%)—confirming that the Parylene-C barrier adequately prevents Cu ion release under chronic culture conditions [74]. For the ureteral stent platform, Cu was electroplated to 100 μm within laser-machined grooves on the polymeric stent wall, with 25 μm Parylene-C enabling operation in conductive artificial urine (8 mS/cm [36]; 9.4 mS/cm [75]). In the next study in the same group, bending down to a 15 cm radius of curvature—a level the authors, based on a urologist co-author’s interventional experience, considered representative of severe bending in typical ureteral stent use—maintained structural integrity with at most 1.1% resonant frequency variation [36]. When the stent was forced to bend below a 13 cm radius, however, the antenna’s Cu lead cracked, causing electrical failure. The authors noted that such bending would rarely be required in typical clinical situations during stent preparation, placement, or indwelling, and that the antenna’s design and materials could be optimized to tolerate greater deflection with a larger safety margin [36]. These Cu–Parylene-C systems have demonstrated that biocompatibility equivalent to Au-based designs can be achieved through coating engineering, expanding the available design space for conductor material selection, geometry, and thickness optimization.
Recent research has explored integrated strategies that combine both routes into a single multilayer architecture. This approach sequentially electroplates Cu (∼30 μm) and Au (∼10 μm), followed by Parylene-C encapsulation [58]. In this structure, the Cu layer serves as a conductive adhesion layer and a low-resistance bulk conductor; the Au overlayer ensures surface biocompatibility and corrosion resistance at the blood–tissue interface, and Parylene-C encapsulates the conductive films, providing electrical insulation and additional biocompatibility [58]. This trilayer stack reduced total stent resistance below 2 Ω and enabled far-field radar backscattering readout at 2 GHz over 50 cm. This working distance is approximately 20-fold greater than the 1–3 cm range typical of prior near-field inductive stent systems [53,58,83]. Part of this gain, however, derives from the change in readout regime (Section 2.2.2) rather than from the metallization alone, so the convergence of the two metallization philosophies should be credited with enabling, rather than by itself producing, performance levels inaccessible to either route alone [30,31,58].
2.3.2. Thickness Engineering of Encapsulation
Insulation design requires balancing two competing parameters: thicker Parylene-C layers more effectively suppress inter-loop capacitive shorting, while thinner layers preserve the mechanical compliance of the pressure sensor diaphragm. A two-stage selective coating strategy effectively decouples RF isolation from diaphragm stiffness: a thick first layer (20–30 μm) applied exclusively to the inductive stent before sensor integration to suppress capacitive shorting between its loops, followed by a thin second layer (2–3 μm) over the complete assembly to preserve diaphragm flexibility while providing biocompatible encapsulation [30]. Under this scheme, improved electrical insulation could enhance the resonant response in conductive biological media, although the phase dips were substantially reduced in 0.15 M PBS (Q = 2.7) and sheep blood (Q = 5). The resonant frequency decreased from 74.5 MHz in air to 63.5 MHz in DI water, 57.2 MHz in sheep blood, and 49.8 MHz in PBS [30]. Even with selective coating, Q in physiological media remains several-fold below Q in air; whether thicker or higher-permittivity-contrast insulation, or a different scaffold material (Section 2.5), is the more effective remedy has not been tested in a single controlled comparison. The electroplating and encapsulation strategies discussed in Section 2.3 are summarized in Table 2.
Table 2.
Summary of electroplating and encapsulation strategies in wireless sensing stents.
2.4. Sensor–Stent Integration
Securing a reliable mechanical and electrical interface between the capacitive sensor and the stent scaffold is essential for in vivo translation (Figure 5). Integration strategies fall into three parallel families—discrete bonding, monolithic co-fabrication, and conformal film-based transfer—each addressing in a different way the mechanical delamination and the electrical degradation that occur at the sensor–stent junction. The earliest implementations used conductive epoxy, which provided adequate bench-top performance yet could not withstand the radial compression forces of PCI—crimping forces exceeding 100 N followed by balloon expansion at 16 atm [31]. Filler-free laser microwelding with a fiber laser (~1070 nm; single pulse, 130 W peak power, 0.5 ms) replaced conductive epoxy with direct metal-to-metal joints (Figure 5a–c), achieving twice the shear strength (from approximately 50 N to over 100 N) and a 6.4-fold improvement in electrical conductance (resistance from 5.3 Ω to 0.83 Ω), and ≈80% of microwelded stents passed bench testing of PCI compatibility—crimping, sheath insertion, and balloon deployment—without mechanical or electrical failure [31,44].
Figure 5.
Joining strategies for integrating a sensor with a stent scaffold. (a–c) Photographs (a,c) and a scanning electron micrograph (SEM) (b) of the microwelded joint between the sensor chip and the antenna stent [31]. Reproduced under the terms of the CC BY license [31]. Copyright 2018, the Authors, published by Wiley-VCH. (d) SEM images of the fabricated sensor on a coil (copper wire) [46]. Reproduced with permission from Ref. [46]. Copyright 2021, IEEE. (e,f) LC pressure sensor integrated smart stent before (e) and after (f) self-rolling [74]. Reproduced with permission from Ref. [74]. Copyright 2022, The Royal Society of Chemistry. (g) Exploded schematic of the electrode sensor attached to the graft with a thermal adhesive; inset, photograph of the mounted sensor [55]. (h) Wider-field photograph of the stent graft with the mounted electrode sensor, showing the overall structure [55]. Reproduced with permission from Ref. [55]. Copyright 2025, the Authors, published by AAAS. (i) External mounting of the sensor on the outer surface of a pre-crimped PLA/PCL hybrid stent: connection holes over the PCL connector struts are filled with molten PCL to fuse the sensor to the stent, and a water-soluble PVA interface holds the sensor against the stent surface until the PVA dissolves after implantation. Schematic drawn based on the description in Ref. [34].
Subsequent designs pursued direct co-fabrication strategies that integrally fabricate the sensing and structural elements on a shared substrate and transform the resulting planar or linear structure into three-dimensional geometry, eliminating the need for a separate sensor attachment step. Various materials and structural platforms have been explored for the fabrication of these devices: copper wire-enabled standard MEMS capacitive sensor fabrication on a wound inductor coil [46] (Figure 5d), gold-plated polyvinylidene fluoride (PVDF) film served simultaneously as the stent scaffold, piezoelectric flow sensor, ultrasonic energy harvester, and circuit substrate through laser ablation and surface-mount device assembly [32], and photosensitive SU-8 provided full compatibility with wafer-scale MEMS lithography for LC sensor and stent mesh co-patterning [74] (Figure 5e,f). The critical distinction among these approaches lies in how the planar structure is converted into the cylindrical stent form. The wire-based sensor was wound manually around a 2.5 mm diameter tube after wafer release, producing a monolithic sensor–inductor coil with no discrete bonding interface [46]. Similarly, the PVDF platform was manually rolled into the cylindrical configuration after PDMS passivation (5 μm per side) at the cost of operator-dependent variability in the final geometry [32]. The self-rollable smart stent eliminated this manual step: removal of the SiO2 sacrificial layer triggered spontaneous rolling of the multilayer structure into a cylindrical configuration due to residual stress mismatch [74]. Thermo-compression bonding (150 °C, 850 gf/cm2, 30 s) joined the capacitor plates, and flexible hinge lengths (25–85 μm) controlled the final diameter from 1.66 ± 0.12 mm to 2.93 ± 0.22 mm with high reproducibility. These results indicate that the 2D-to-3D conversion can achieve the dimensional precision required for coronary deployment [74]. This self-rolling platform was subsequently extended by enlarging the inductor coil to cover the entire stent strut surface rather than a confined region, which doubled the sensing resolution and coupling distance and enabled the integration of two independent pressure sensors within a single stent for simultaneous blood pressure and blood flow monitoring [53].
A third integration philosophy, thin-film transfer-based integration, enables sensor incorporation onto clinically established stent platforms without requiring modification of the underlying metallic scaffold [55]. These approaches fabricate free-standing thin-film electronic structures that are subsequently transferred, bonded, or laminated onto the stent or graft surface. A thin-film capacitive pressure-sensing diaphragm array for intravascular applications demonstrated that eliminating the rigid substrate used in conventional micromachined pressure sensors enables an ultrathin device architecture with a total thickness of only 65 μm, while introducing only approximately 15 μm of additional profile when integrated onto intravascular structures [108]. The resulting structure was fabricated as a transferable thin film and bonded onto target intravascular surfaces using a thermoplastic polyimide adhesive layer [108]. A low-profile endoleak surveillance system based on a distributed flexible-PCB architecture demonstrated the feasibility of surface-integrated sensing under catheter-delivery constraints [104]. The system employed four flexible PCB strips, each carrying four pressure sensors with ASIC-based interfacing, mounted onto the exterior of the stent graft while remaining within the 2 mm height constraint imposed by the delivery catheter geometry [104]. More recently, a 2025 study demonstrated an ultrathin coplanar capacitive sensor (total thickness of 5 μm) transferred onto the outer surface of the stent graft using a specially formulated film-type thermal adhesive made of co-polyamide (Co-PA) [55]. In tests on stent fabric laminates, this Co-PA adhesive exhibited a low bending load of 1.5 N under compression and an adhesion force of 10 N in a 180° peel test; Co-PA also did not penetrate the inner wall of the stent graft or compromise the intrinsic waterproof properties of the graft fabric, and the sensor did not cause additional leakage between the stent and a vascular phantom [55] (Figure 5g,h). Although cyclic deformation, crimping, and hydrolytic stability evaluations suggested promising mechanical robustness for these surface-integrated systems, the long-term stability under chronic vascular loading conditions remains an important area for further investigation [55,104].
In addition to integration mechanisms—robust bonding, monolithic integration, or conformal film-based transfer—the physical placement of the sensing element within the stent structure must be considered. Early implementations commonly positioned the sensor on the exterior surface to simplify assembly and maximize exposure to the surrounding environment [31,54]. This configuration creates a vulnerability to shear-induced wrinkling and sensor detachment resulting from contact with the guide catheter during catheter-based delivery [83]. By relocating the sensor to the stent interior, it remains intact during catheter-based navigation, enabling successful deployment through both 8F and 7F guide catheters [83]. Interior placement, however, exposes the sensor directly to blood flow and may perturb local hemodynamics or provide a nidus for thrombus until it is endothelialized, a trade-off that has not yet been quantified. Another approach mounts the sensor externally onto a pre-crimped stent and retains it with a polyvinyl alcohol sacrificial layer that dissolves upon balloon expansion, decoupling the integration step from the radial strain of expansion itself [34] (Figure 5i). Both approaches improve catheter compatibility relative to the original exterior-on-expanded configuration while diverging in their treatment of the sensor–surface interface.
2.5. Material Strategies for Stent-Based Monitoring Systems
This section recasts material selection as a higher-order design decision that operates across the sensing (Section 2.1), communication (Section 2.2), and integration (Section 2.4) axes; selecting between metallic, polymeric, and hybrid scaffolds determines not only the achievable wireless performance envelope but also the anatomic niches, monitoring durations, and clinical indications a given platform can practically serve (Figure 6). Three observations motivate this synthesis. The same conductivity advantage that elevates metallic Q in dry conditions does not transfer cleanly to the conductive biological media where implants must operate. Polymer scaffolds, by virtue of their elastic compliance and patternability, open application niches that balloon-expandable metallic stents cannot easily enter. Bioresorbable materials introduce a temporal axis—finite functional lifetime—that the preceding sections did not address.
The metallization performance established in Section 2.3 positions metallic scaffolds as the most influential option for improving quality factor under non-physiological readout conditions, but the mechanical rigidity that retains shape after balloon expansion can restrict the deployment to anatomies in which permanent luminal support is clinically desirable [31,109,110]. Coronary applications, where pulsatile vessel-wall pressure demands the radial support that the high stiffness and yield strength of metallic alloys provide, define the natural anatomic niche of the metallic paradigm [111]. For luminal sites characterized by elastic compliance and active peristaltic or respiratory motion, including ureteral, airway, and ductal anatomies, balloon-expanded metallic mechanics are mismatched to the host tissue [6]. The substantial stiffness mismatch between metallic alloys and soft luminal tissue limits the conformability of rigid scaffolds to the intrinsic deformation of the vessel wall [109]. This anatomic mismatch—rather than any electromagnetic shortcoming—bounds the addressable application set for metallic platforms, motivating alternative scaffold paradigms regardless of any wireless performance advantage the metallic option might otherwise confer [6,44,109].
Polymer scaffolds, whose fabrication and metallization aspects are documented in Section 2.3 and Section 2.4, can preserve a higher quality factor than metallic stent inductors under conductive biological media [47,112]. The self-rollable SU-8 stent retained Q = 40.78 in air and Q = 14.80 in 17.28 mS/cm culture medium [74], approximately five-fold higher in the conductive medium than the Q = 2.7 reported for the gold-electroplated 316 L stainless-steel stentenna in 0.15 M PBS [30]. This advantage has been independently demonstrated on 3D-printed polycaprolactone (PCL) and PCL/polylactic acid (PLA) platforms, where polymer scaffolds produced minimal interference with the resonant circuit operation in contrast to the metal-based structures, in which the metallic loops introduce parasitic capacitive shorting under ionic loading [34,47,112] (Figure 6a). To determine whether this effect generalizes across all polymer scaffold geometries, the underlying mechanistic origin—distinguishing the absence of inter-loop parasitic coupling from variations in diaphragm compliance or insulation thickness—warrants systematic parametric isolation, ideally by mounting an identical LC sensor on metallic and polymeric scaffolds of the same geometry and measuring both in the same medium, as was partly done in [47,112].
Figure 6.
Material strategies across metallic, piezoelectric-polymer and bioresorbable scaffolds. (a) Polymer stent with an integrated pressure sensor, and the phase response for the antenna alone and with metal and polymer scaffolds [112]. Reproduced with permission from Ref. [112]. Copyright 2016, the Authors, published by MDPI. (b) Fabricated PVDF smart stent with surface-mounted electronics [32]. (c) System block diagram combining rectification, flow sensing and RF transmission on one membrane [32]. (d) Flow-sensor output voltage [32]. (e) Spectrum-analyzer output received about 1 m away [32]. Reproduced with permission from Ref. [32]. Copyright 2020, the Authors, published by IEEE. (f) Force–displacement curves of PCL, PLA and PCL/CNT scaffolds [57]. (g) Printed PCL/CNT and pristine PCL strut networks [57]. (h) Reflection coefficient of the LC sensor in PCL and PCL/CNT stents [57]. (i) Resonant-frequency tracking of pulsatile pressure against a reference sensor [57]. Reproduced with permission from Ref. [57]. Copyright 2025, American Chemical Society.
Polymer scaffolds can do more than provide an elastic substrate; they can constitute the sensing element themselves, eliminating the discrete capacitive transducers that the metallic paradigm requires [32,87]. Piezoelectric PVDF integrates flow transduction, energy harvesting, and RF transmission within a single architecture: a re-entrant honeycomb-perforated membrane with gold electrodes on its top and bottom surfaces [32] (Figure 6b,c). The structure exhibits a piezoelectric coefficient of d33 = 30 pC/N, generating 53–131 mV under flow (30–980 mL/min) and enabling frequency-modulated transmission (81–87 MHz) with ~1 m readout distance [32] (Figure 6d,e). A 3D-printed PVDF-TrFE variant (d33 = 9.967 ± 0.306 pC/N before rolling) demonstrated dual wired piezoelectric sensing at 0.943 mV/mmHg alongside wireless LC readout at −2.2 kHz/mmHg with R2 = 0.91 from a single tubular structure interrogated at 8.66 MHz in deionized water [87]. These implementations differ fundamentally from the polymer-as-substrate strategies catalogued in Section 2.3 and Section 2.4—where polymers served as conformal insulation (Parylene-C), fabrication substrates (SU-8, PVDF film), and adhesive interfaces (thermoplastic polyimide, co-polyamide thermoadhesive)—in that the intrinsic piezoelectricity of the polymer enables a transition from passive support functions to active transduction [32,87].
Biodegradable polymers provide a critical benefit because their finite functional window bounds the implant’s lifespan to prevent chronic tissue irritation, unlike permanent metallic or stable polymer scaffolds [113,114]. However, the use of biodegradable stents is constrained by mechanical strength, progressive loss of mechanical integrity during degradation, nonuniform degradation and a degradation-associated risk of restenosis. The intrinsic mechanical strength limitations of bioresorbable polymer scaffolds can restrict their applicability in load-bearing vascular environments without reinforcement strategies [57]. Carbon-nanotube (CNT) reinforcement of a 3D-printed PCL matrix at 1 wt% loading raised radial force to approximately 0.1 N/mm—a 7.7-fold enhancement over pristine PCL—thereby elevating the scaffold to the lower bound of clinical coronary specifications [57] (Figure 6f,g). This penalty is inherently electromagnetic: the conductive CNT network induces an approximately 70% reduction in S11 magnitude primarily driven by absorption losses [57] (Figure 6h,i). Whether this attenuation can be reconciled with clinically relevant wireless readout distances through the optimization of CNT concentration or distribution warrants further investigation. Beyond the initial strength limitation, mechanical integrity declines as degradation proceeds. Under thermally accelerated conditions (PBS, 50 °C), the Fantom Encore scaffold (commercial BRS, Reva Medical) exhibited reduced radial strength and stiffness, with work-to-fracture reduced to <10% of its initial value during the degradation phase (days 63–112), while increased enthalpy of fusion, indicative of increased relative crystallinity, coincided with reduced ductility and increased brittleness [115]. Degradation is nonuniform: in PLLA scaffolds, the core had lower crystallinity (38.3%) than the surface (61.6%) [116]. During static immersion in 37 °C PBS, core crystallinity decreased to 29.3% at day 110, while surface crystallinity increased to approximately 70% in the first 10 days and remained unchanged until day 90 [116]. The decrease in core crystallinity was attributed to more substantial microstructural disruption in the low crystalline region accelerated in the first 5 days after fluid contact [116].
The degradation of bioresorbable polymer scaffolds can increase surface roughness and alter chemistry, generating degradation byproducts [113,117]. These changes promote platelet adhesion and activation, increasing the risk of thrombus formation, which impairs endothelialization [114,117]. Furthermore, impaired vascular healing and persistent inflammatory responses drive excessive neointimal hyperplasia, leading to vascular restenosis [113,114]. Reported strategies to mitigate these effects include polymer scaffolds engineered for higher strength and thinner struts [57], magnesium-, iron-, and zinc-based bioresorbable metals [92,118,119,120,121], surface modification of polymer substrates, such as topographical patterning and biomolecule immobilization [122,123], and drug delivery designs that promote endothelial healing, such as polymer-free, abluminal, or pro-healing dual-therapy coatings [124,125]; localized drug delivery has also been integrated with sensing in a bioresorbable electronic stent [92]. These strategies are not discussed in detail in this review although research on bioresorbable stents (BRS) and drug-eluting stents (DES) largely affects smart stent development. The integration of these strategies into smart stents is illustrated in future directions (Section 3.5).
Across these paradigms, material selection determines a compound envelope spanning the anatomic, electromagnetic, mechanical, and temporal axes. Metallic scaffolds command high Q in non-physiological conditions and robust balloon-expandable mechanics, but their anatomic reach is largely restricted to sites that require permanent rigid scaffolding [44,48]. Polymer scaffolds extend applications into compliant anatomies and have shown favorable Q under conductive loading, although the latter may depend on device location, telemetry configuration, insulation thickness, and temperature as much as on the scaffold material itself [30,74], and their lower mechanical strength limits load-bearing use [57]. Piezoelectric polymers open a route to self-powered sensing [32,87], and bioresorbable scaffolds—reinforced where necessary by CNT or PCL/PLA hybrids [34,57]—provide temporary support followed by resorption [34,113,114]. Hybrid architectures that stack metallic conductors on polymeric scaffolds are the natural engineering response to these single-paradigm limits, but each added layer introduces an interface whose chronic reliability must be verified separately (Section 2.3 and Section 2.4). Material strategy should therefore be understood not as a choice between conductivity values but as a co-design problem encompassing target anatomy, intended monitoring duration, and the readout regime of Section 2.2.
3. Discussion
This review examined stent-based monitoring across five engineering subsystems: sensing mechanisms (Section 2.1), power and data transmission (Section 2.2), metallization and insulation (Section 2.3), sensor–scaffold integration (Section 2.4), and material strategies (Section 2.5). Performance trends across these subsystems and their associated trade-offs are examined in Section 3.1. Barriers to in vivo translation are examined in Section 3.2. Common reporting criteria at the subsystem level are proposed in Section 3.3. Clinical translation, including regulatory verification and MRI safety, is addressed in Section 3.4. Future research directions that follow from these observations are outlined in Section 3.5.
3.1. Cross-Cutting Performance Trends
Quality factors measured in air decrease significantly when transitioned to physiological saline and blood, with a corresponding loss of sensitivity [30,74]. At the lower bound of the reported range, the impedance dip becomes shallow enough that conventional impedance-phase tracking is restricted to short interrogation distances [30,97], although matching the reader frequency to the implant’s resonance has been shown to extend this range [33,36]. Polymer scaffolds have been reported to preserve considerably higher quality factors under conductive loading, in which the conductive surrounding medium dissipates energy from the resonator, than gold-electroplated stainless steel in media with comparable conductivity [30,74], and the advantage has been observed in printed polymer platforms [34,47,112]. However, the mechanistic origin of this performance has not yet been isolated from concurrent differences in diaphragm compliance and mechanical stiffness that accompany changes in scaffold material, so that a systematic parametric study is required before it can be generalized across polymer scaffold geometries.
Reported working distance has been extended from 1–3 cm to more than 50 cm in air, but this gain reflects a change in readout regime rather than an improvement within one: near-field inductive and transient-resonance readout remain limited to a few centimeters, whereas far-field operation reaches tens of centimeters in air [33,54,58,97]. However, in biological media, with near-field readout at 6 cm in meat [54] and, for a far-field EVAR stent antenna placed in a phantom, received power over 0–2 m that agreed with the link-budget calculation and 4% power transfer efficiency at 10 cm [103]. In either regime, the usable range is first limited less by antenna design than by the tissue safety envelope (Section 2.2.2). The power that can be delivered to the implant within this limit still falls short of the continuous draw required for active multi-sensor operation [103,104], indicating that further development of the power delivery chain remains necessary. Because passive LC platforms can be read only while an external reader is energizing them, monitoring is confined to the intervals in which a reader is present. Piezoelectric polymer architectures, which harvest energy from flow itself, could in principle remove this dependence on an external power link and allow sensing between reader sessions; however, their output (tens to hundreds of mV under flow) has not yet been shown to sustain a wireless data link in vivo [32,87].
Regarding metallization and integration, stratified material stacks have emerged as a prevalent trend, distributing electromagnetic, mechanical, and biological roles across separate layers in contrast to the single-function antenna stents of earlier designs [58,74]. However, these layers still interact through shared design variables, as discussed in the following paragraphs. Sensor–scaffold integration strategies, in turn, have diversified according to the deployment constraints that dominate each application [31,32,53,74,108]. The diversification of joining strategies is expected to widen the range of anatomies and delivery procedures to which a given sensing concept can be adapted, at the cost of additional interfaces whose chronic reliability requires individual verification [31,44]. Bioresorbable scaffolds provide monitoring only during the healing phase prior to resorption, ensuring the implant remains transient rather than permanent [57,113,114]. If the mechanical strength of resorbable polymers can be raised to clinical specifications, bioresorbable stents can eliminate the long-term biological burden of permanent implants after fulfilling their intended function.
The subsystems in Section 2 were reviewed separately, but they share design variables, so a choice made in one subsystem can affect the others. These couplings are summarized here in the order in which they are discussed in Section 2. In the sensing subsystem (Section 2.1), the thickness and stiffness of the sensing film determine the sensitivity. A thinner or softer film deflects more under a given pressure, but reduces the allowable strain during crimping and balloon expansion (Section 2.4). In the transmission subsystem (Section 2.2), the operating frequency is the most widely coupled variable. A higher frequency can allow smaller resonant structures and extend wireless communication distances. However, it also increases tissue losses and SAR, thereby reducing Q in vivo and limiting the permissible power. For dielectric sensing, frequency also influences the permittivity contrast between blood and tissue, which underlies the sensing response (Section 2.1.3). Resonance of the stent near the RF operating frequency of an MRI scanner raises additional safety concerns (Section 3.4).
In the metallization subsystem (Section 2.3), the conductor and its insulation act on Q through different mechanisms. The conductor determines the resistance of the resonator. A thicker or more conductive layer lowers this resistance and raises Q, but also changes the strut cross-section, which affects the inductance, profile, and stiffness of the stent (Section 2.4 and Section 2.5). The insulation mitigates a separate source of loss. Blood and tissue are conductive, so the surrounding medium dissipates energy from the resonator and lowers Q. A thicker insulation layer reduces this loading, but it also separates the sensing field from the medium, which weakens the sensitivity of dielectric sensing (Section 2.1.3).
In the integration subsystem (Section 2.4), each joint between the sensor and the scaffold is both a mechanical and an electrical interface. A joint that cracks during crimping or balloon expansion interrupts the circuit, while a resistive joint lowers Q (Section 2.2 and Section 2.3). The joint also determines how wall strain and pressure reach the sensor (Section 2.1). In the material subsystem (Section 2.5), the scaffold material determines both the electrical loading and the mechanical stiffness of the platform. Polymer scaffolds better maintain Q under conductive loading (Section 2.2), but they provide lower radial force. Their compliance also affects the deflection of attached diaphragms (Section 2.1). In bioresorbable scaffolds, these properties evolve as degradation proceeds, causing L, C, and Q to drift over the functional lifetime. Therefore, a gain in one subsystem should be evaluated alongside its associated trade-offs in the others (Section 3.3).
3.2. Barriers to In Vivo Translation
Most performance values reviewed in Section 2 were obtained in vitro, ex vivo, or in simulation. In vivo data come from a small number of animal studies. In a rat, the femoral artery was too narrow for a crimped stent, so the sensor was attached to the outer surface of the artery as a planar mesh [53]. In a rabbit iliac artery, catheter deployment was demonstrated, whereas wireless pressure sensing was tested after harvesting the artery because of the small vessel size and distance to the skin [33]. Resonance shifts were detected in swine bypass-graft models [30,31], and wireless readout at 5 cm was shown in a canine model [92]. However, sensing performance in stent-based systems has been followed only over days to weeks, and chronic stent implantation in large animals has not been reported among the studies reviewed here.
Bench conditions simplify several factors that affect the smart stent system after implantation. First, surface coverage by adsorbed plasma proteins, platelets, endothelium, and neointima alters the dielectric environment around the device, shifting the baseline signal [45,79,84,126,127,128]. It can also affect wall-strain transfer in strain sensors [83]. In ureteral stents, biofilm formation and encrustation can play a similar role [129]. Second, dynamic variations in biological media alter device response: blood electrical conductivity varies with hematocrit and temperature [130], higher fluid viscosity damps resonance and shifts the baseline frequency and quality factor [131], and local inflammatory mediators such as interleukin-1 increase in the stented vessel wall [132]. Third, the relative alignment between the reader and the stent fluctuates over time as cardiac and respiratory motion and vessel pulsation alter inductive coupling [30,33,42,43]. Measurements can also be affected by variations in reader placement. Fourth, signal attenuation through tissue, which differs across patients in thickness and composition [133], limits wireless readout. Particularly at deep anatomical sites such as the posterior abdominal aorta (4.6 to 16.7 cm below the skin [89]), implantation depths exceed most passive readout ranges demonstrated even in saline or air (Section 2.2.2). Fifth, long-term operational durability requires insulation layers, bonded joints, and working liquids to withstand cyclic pulsatile loading at body temperature [46,55,82,104,134], while active devices additionally require a sustainable power supply [89]. Sixth, mechanical survival during delivery and positional stability after deployment involve stresses partially reproduced by crimping and expansion tests [31,44,135].
Clinical translation therefore requires staged validation. Bench tests should first be conducted in blood under pulsatile flow at body temperature, and then perfused ex vivo vessels should be used to incorporate protein adsorption and pulsatile wall motion prior to animal studies. Acute studies in small animals can confirm initial wireless readout, whereas studies in large animals are necessary to evaluate wireless readout performance because their vessel dimensions and implantation depths more closely approximate human anatomy [133,134]. In addition, chronic studies are required to assess long-term device stability [134]. These longitudinal studies should track baseline drift during endothelialization and confirm device positioning via post-implantation imaging [126,134,136]. Sensor readings should also be benchmarked against independent reference measurements, such as catheter-based pressure measurements or intravascular imaging [134], and validated through histological examination after explantation [127,134]. Furthermore, calibration and readout compensation can isolate the physiological signal of interest from drift and coupling artifacts: post-implantation baseline recalibration can compensate for slow biological changes [128], and distance-independent readout compensation circuits can suppress variations in interrogation distance and coupling [137]. Finally, reporting animal model, implantation site, implantation depth, and baseline drift in a consistent format would enable direct comparison across in vivo studies (Section 3.3).
3.3. Toward Common Reporting Criteria
A further complication concerns the metrics themselves. The reporting of these metrics is inconsistent across the literature, and they are rarely obtained under common, comparable measurement conditions. In particular, resonant frequency, quality factor, sensitivity, and readout distance are each reported in different media (air, deionized water, saline of unstated concentration, culture medium, blood), so that the medium-induced shift—tens of MHz (Section 2.1.1)—dwarfs the physiological signal. This makes cross-platform performance comparisons meaningless. The definition of a standardized characterization protocol, together with the generation of a benchmark dataset reporting these metrics jointly under common measurement conditions, would therefore be of considerable value to the research community. It would allow meaningful comparison across studies for the first time.
3.4. Clinical Translation
Stents are generally Class III devices in the United States, the European Union, and China [138,139,140]. Adding a sensor to an approved stent is therefore likely to be treated as a new Class III device rather than a variant of the original, since FDA expects a new premarket approval when a modification needs both new preclinical and new clinical data [138,141]. Modular sensors attached to clinically established stent platforms [55,88] may allow data on the base stent to be reused [138]. The combined device nevertheless has to be tested as a finished product, and each added element brings its own verification burden, including additional coating characterization and integrity and durability testing [138]. New materials and coatings require long-term compatibility testing under blood-contact conditions [138,142]. Dissimilar metals in plated or welded structures call for corrosion testing, including assessment of galvanic corrosion [138,143]. Devices with batteries or active circuits also fall under the active implantable device framework, which adds electrical safety and electromagnetic compatibility requirements [144]. Coatings are assessed for integrity and particulate generation during simulated delivery and deployment [138]. The 2010 FDA guidance recommends accelerated durability testing under pulsatile flow and physiologic loading equivalent to ten years of real-time use [138].
MRI compatibility is an important consideration because stented patients may require MRI during follow-up [145]. Two major MRI-related risks for metallic implants are RF-induced heating and displacement force and torque from the static magnetic field, and conventional metallic stents are tested against them and typically labeled MR Conditional [138,145]. A smart stent differs because it is designed to resonate, and several LC stents in Table 1 resonate between 58 and 136 MHz [33,36,53,59,75], a range that spans the RF frequencies of 1.5 T and 3 T scanners (about 64 and 128 MHz) [145]. A resonance near the scanner frequency amplifies the RF field within the stent, adding resonator-induced heating to the conventional SAR-related heating [145]. Because the quality factor decreases in ionic media and the resonance frequency shifts in blood and tissue [41,55], neither the frequency separation from the scanner RF nor the resulting heating margin can be reliably established through bench design alone [145]. The static magnetic field can exert displacement forces and torque on magnetic materials. The magnetic properties of the antenna conductors, stent platform, and any battery or electronic components therefore need to be characterized, with force and torque testing performed where relevant [138,145]. Practical measures include placing the resonance away from the scanner frequencies [145] and evaluating MRI safety at both 1.5 T and 3.0 T, as recommended by the FDA for intravascular stents [138]. These MRI safety requirements necessitate prioritizing the frequency band, materials, and power source at the design stage, reducing the need for potential safety verification that arises from subsequent design changes [138,141].
3.5. Future Directions
Beyond the improvements of each subsystem, smart stents have potential development directions.
Multi-parameter sensing. A single sensor reading can be difficult to reliably distinguish in-stent restenosis from thrombosis, because both narrow the lumen and change the local pressure [33]. Measuring pressure together with flow, wall strain, or impedance in one scaffold could separate them, since neointimal tissue and thrombus differ in their mechanical and electrical properties. The main obstacle is space: each added sensor competes for the limited strut area, and in passive LC designs each sensor also needs its own resonance band. The next step is to identify, in animal models, the smallest set of parameters that separates the two conditions before adding more sensors.
Patient-specific scaffolds. 3D printing of polymeric scaffolds with integrated sensors could match the scaffold to the vessel geometry of each patient, which could reduce migration and malposition, and tailor the scaffold stiffness, resonance frequency, and pressure sensitivity for each patient [57,146]. However, printed polymer scaffolds still have lower radial strength than metallic stents, and printed conductors have higher resistance than electroplated metal conductors, resulting in a lower quality factor. Printed polymer scaffolds still have to improve their radial strength and conductivity, and the printing processes need reproducibility validation before patient-specific devices can be realized.
Closed-loop sensing and therapy. Drug-eluting stents already deliver therapy, but their release profiles are predetermined by the drug and coating design at manufacturing. Linking the sensor output to on-demand release, for example through wireless heating or electrically triggered release, could turn surveillance into intervention [92]. Unresolved issues include the safety of wireless actuation inside a vessel, the limited drug reservoir, and false triggering caused by noisy signals. The first commercial systems are expected to require clinician confirmation of sensor readings before triggering drug release.
AI-assisted signal interpretation. Artificial intelligence-assisted signal interpretation has the potential to effectively detect the clinically useful signal, leading to earlier and more reliable identification of stent-related complications [37]. Machine learning models trained on long-term recordings could separate these effects from true pathological changes and detect complications earlier than fixed thresholds. The main limitation is the lack of data, because long-term in vivo recordings of relevant physical, biochemical and sensor-performance parameters paired with angiographic or intravascular imaging are scarce. Building such paired datasets in large-animal studies is, in our view, the most important step for this direction.
Across these directions, three practical steps would accelerate clinical translation. First, common measurement conditions and reporting criteria, such as the testing medium, reader distance, and sensitivity units, are needed so that results from different studies can be compared. Second, long-term implantation studies in large animals are needed to quantify how tissue coverage and motion affect overall system performance. Third, the safety tests required for regulatory approval, such as MRI compatibility and coating durability should be considered from the early stages of device design.
4. Conclusions
In conclusion, our review indicates that smart stents can enable the continuous, localized surveillance of stented lesions, allowing complications such as in-stent restenosis and stent thrombosis to be detected earlier and narrowing the delay that limits current surveillance modalities. Several requirements must be met before stent-integrated sensing can support graded, rather than threshold-based, surveillance. The first is signal quality. The quality factor of the implanted resonator falls sharply in conductive media, and this loss must be contained through the choice of scaffold material, selective encapsulation of the inductor, or readout schemes that tolerate low Q. The second is power. The amount that can be delivered to the implant within tissue safety limits remains small, so future platforms will need to exploit the larger SAR margin of inductive links for power transfer, move the data link to backscatter so that the implant does not radiate, or harvest energy from flow directly. The third is stability over time. Chronic drift has been characterized only over days to weeks and must be validated under sustained in vivo loading. Finally, resolution must reach a clinically meaningful level, on the order of the pressure gradient that distinguishes a significant stenosis from a benign one. Progress on these requirements will be difficult to judge unless the field also agrees on how the underlying quantities are reported. At a minimum, resonant frequency, quality factor, sensitivity, and readout distance should be stated together with the medium, temperature, and reader configuration under which they were obtained. With these criteria in place, the next generation of stent-integrated sensors can be evaluated on a common basis, and the question of which architecture is best suited to a given clinical setting can begin to be answered with data rather than by inference across incompatible studies.
Author Contributions
Conceptualization, K.K. and S.N.; formal analysis, K.K. and S.N.; investigation, S.N.; resources, K.K.; writing—original draft preparation, K.K. and S.N.; writing—review and editing, K.K., J.-H.B. and S.N.; visualization, K.K. and S.N.; supervision, K.K.; project administration, K.K.; funding acquisition, K.K. All authors have read and agreed to the published version of the manuscript.
Funding
This research was funded by the National Research Foundation of Korea (RS-2026-25481951), the Korea Basic Science Institute (National Research Facilities and Equipment Center) grant funded by the Korean government (MSIT) (RS-2026-25498557) and the Hankuk University of Foreign Studies Research Fund (of 2026).
Institutional Review Board Statement
Not applicable.
Informed Consent Statement
Not applicable.
Data Availability Statement
No new data were created or analyzed in this study. Data sharing is not applicable to this article.
Acknowledgments
AI-assisted figure preparation. The schematic illustrations in Figure 1 and Figure 2 were prepared with the assistance of Claude (Opus 5.5; Anthropic). The model wrote Python code (Python 3.12) using the python-pptx library (version 1.0.2) that drew each illustration as an editable vector component in PowerPoint. No image-generation model was used, and no published figure was reproduced. The depicted structures were based on the text of the cited articles and were checked by the authors against those articles. The authors specified the content, layout, and color palette, corrected the drawings over several rounds, and assembled and edited the final figures.
Conflicts of Interest
The authors declare no conflicts of interest.
References
- Sigwart, U.; Puel, J.; Mirkovitch, V.; Joffre, F.; Kappenberger, L. Intravascular stents to prevent occlusion and restenosis after transluminal angioplasty. N. Engl. J. Med. 1987, 316, 701–706. [Google Scholar] [CrossRef] [Scilit]
- Moses, J.W.; Leon, M.B.; Popma, J.J.; Fitzgerald, P.J.; Holmes, D.R.; O’Shaughnessy, C.; Caputo, R.P.; Kereiakes, D.J.; Williams, D.O.; Teirstein, P.S.; et al. Sirolimus-eluting stents versus standard stents in patients with stenosis in a native coronary artery. N. Engl. J. Med. 2003, 349, 1315–1323. [Google Scholar] [CrossRef] [Scilit]
- Stone, G.W.; Ellis, S.G.; Cox, D.A.; Hermiller, J.; O’Shaughnessy, C.; Mann, J.T.; Turco, M.; Caputo, R.; Bergin, P.; Greenberg, J.; et al. A polymer-based, paclitaxel-eluting stent in patients with coronary artery disease. N. Engl. J. Med. 2004, 350, 221–231. [Google Scholar] [CrossRef] [Scilit]
- Stefanini, G.G.; Byrne, R.A.; Windecker, S.; Kastrati, A. State of the art: Coronary artery stents-past, present and future. EuroIntervention 2017, 13, 706–716. [Google Scholar] [CrossRef] [Scilit]
- Li, J.; Hu, X.; Chen, Y.; Fan, D.; Tan, C.; Yang, S.; Wu, H.; Wang, Y.; An, Q.; Xiao, Z.; et al. Review of recent progress in vascular stents: From conventional to functional vascular stents. Chin. Chem. Lett. 2025, 36, 110492. [Google Scholar] [CrossRef] [Scilit]
- Chen, M.; Yan, J.; Ding, Y. Recent advances in non-vascular stents for occlusive luminal disease treatment. J. Mater. Chem. B 2026, 14, 1167–1191. [Google Scholar] [CrossRef] [Scilit]
- World Health Organization. Cardiovascular Diseases (cvds). Available online: https://www.who.int/news-room/fact-sheets/detail/cardiovascular-diseases-(cvds) (accessed on 26 September 2026).
- Canfield, J.; Totary-Jain, H. 40 years of percutaneous coronary intervention: History and future directions. J. Pers. Med. 2018, 8, 33. [Google Scholar] [CrossRef] [Scilit]
- World Health Organization. Ageing and Health. Available online: https://www.who.int/news-room/fact-sheets/detail/ageing-and-health (accessed on 26 September 2026).
- Bray, F.; Laversanne, M.; Sung, H.; Ferlay, J.; Siegel, R.L.; Soerjomataram, I.; Jemal, A. Global cancer statistics 2022: Globocan estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA Cancer J. Clin. 2024, 74, 229–263. [Google Scholar] [CrossRef] [Scilit]
- Grand View Research. Stents Market (2025–2030). Available online: https://www.grandviewresearch.com/industry-analysis/stents-market-report (accessed on 26 September 2026).
- Abbara, S.; Blanke, P.; Maroules, C.D.; Cheezum, M.; Choi, A.D.; Han, B.K.; Marwan, M.; Naoum, C.; Norgaard, B.L.; Rubinshtein, R.; et al. Scct guidelines for the performance and acquisition of coronary computed tomographic angiography: A report of the society of cardiovascular computed tomography guidelines committee: Endorsed by the north american society for cardiovascular imaging (nasci). J. Cardiovasc. Comput. Tomogr. 2016, 10, 435–449. [Google Scholar]
- Nieman, K. Noninvasive stent imaging with msct. EuroIntervention 2009, 5, D107–D111. [Google Scholar]
- Ricci, F.; Khanji, M.Y.; Bisaccia, G.; Cipriani, A.; Di Cesare, A.; Ceriello, L.; Mantini, C.; Zimarino, M.; Fedorowski, A.; Gallina, S.; et al. Diagnostic and prognostic value of stress cardiovascular magnetic resonance imaging in patients with known or suspected coronary artery disease: A systematic review and meta-analysis. JAMA Cardiol. 2023, 8, 662–673. [Google Scholar] [CrossRef] [Scilit]
- Virani, S.S.; Newby, L.K.; Arnold, S.V.; Bittner, V.; Brewer, L.C.; Demeter, S.H.; Dixon, D.L.; Fearon, W.F.; Hess, B.; Johnson, H.M.; et al. 2023 aha/acc/accp/aspc/nla/pcna guideline for the management of patients with chronic coronary disease: A report of the american heart association/american college of cardiology joint committee on clinical practice guidelines. Circulation 2023, 148, e9–e119, Correction in Circulation 2023, 148, e148. Correction in Circulation 2023, 148, e186. [Google Scholar] [CrossRef] [Scilit]
- Jang, J.-S.; Song, Y.-J.; Kang, W.; Jin, H.-Y.; Seo, J.-S.; Yang, T.-H.; Kim, D.-K.; Cho, K.-I.; Kim, B.-H.; Park, Y.H.; et al. Intravascular ultrasound-guided implantation of drug-eluting stents to improve outcome. JACC Cardiovasc. Interv. 2014, 7, 233–243. [Google Scholar] [CrossRef] [Scilit]
- Koo, B.K.; Hu, X.; Kang, J.; Zhang, J.; Jiang, J.; Hahn, J.Y.; Nam, C.W.; Doh, J.H.; Lee, B.K.; Kim, W.; et al. Fractional flow reserve or intravascular ultrasonography to guide pci. N. Engl. J. Med. 2022, 387, 779–789. [Google Scholar] [CrossRef] [Scilit]
- Vanhoenacker, P.K.; Heijenbrok-Kal, M.H.; Van Heste, R.; Decramer, I.; Van Hoe, L.R.; Wijns, W.; Hunink, M.G. Diagnostic performance of multidetector ct angiography for assessment of coronary artery disease: Meta-analysis. Radiology 2007, 244, 419–428. [Google Scholar] [CrossRef] [Scilit]
- Schuetz, G.M.; Zacharopoulou, N.M.; Schlattmann, P.; Dewey, M. Meta-analysis: Noninvasive coronary angiography using computed tomography versus magnetic resonance imaging. Ann. Intern. Med. 2010, 152, 167–177. [Google Scholar] [CrossRef] [Scilit]
- Wennberg, E.; Abualsaud, A.O.; Eisenberg, M.J. Patient management following percutaneous coronary intervention. JACC Adv. 2025, 4, 101453. [Google Scholar] [CrossRef] [Scilit]
- Einstein, A.J.; Moser, K.W.; Thompson, R.C.; Cerqueira, M.D.; Henzlova, M.J. Radiation dose to patients from cardiac diagnostic imaging. Circulation 2007, 116, 1290–1305. [Google Scholar] [CrossRef] [Scilit]
- Zellweger, M.J.; Weinbacher, M.; Zutter, A.W.; Jeger, R.V.; Mueller-Brand, J.; Kaiser, C.; Buser, P.T.; Pfisterer, M.E. Long-term outcome of patients with silent versus symptomatic ischemia six months after percutaneous coronary intervention and stenting. J. Am. Coll. Cardiol. 2003, 42, 33–40. [Google Scholar] [CrossRef] [Scilit]
- Ruygrok, P.N.; Webster, M.W.; de Valk, V.; van Es, G.A.; Ormiston, J.A.; Morel, M.A.; Serruys, P.W. Clinical and angiographic factors associated with asymptomatic restenosis after percutaneous coronary intervention. Circulation 2001, 104, 2289–2294. [Google Scholar] [CrossRef] [Scilit]
- Massussi, M.; Drera, A.; Pancaldi, E.; Pezzola, E.; Tagazzini, L.; Fiorina, C.; Branca, L.; Chizzola, G.; Metra, M.; Curello, S.; et al. Eight-month angiographic outcomes and in-stent restenosis in patients undergoing percutaneous coronary intervention on unprotected left main coronary artery. Coron. Artery Dis. 2025, 37, 56–62. [Google Scholar] [CrossRef] [Scilit]
- Giustino, G.; Colombo, A.; Camaj, A.; Yasumura, K.; Mehran, R.; Stone, G.W.; Kini, A.; Sharma, S.K. Coronary in-stent restenosis. JACC 2022, 80, 348–372. [Google Scholar] [CrossRef] [Scilit]
- Iakovou, I.; Schmidt, T.; Bonizzoni, E.; Ge, L.; Sangiorgi, G.M.; Stankovic, G.; Airoldi, F.; Chieffo, A.; Montorfano, M.; Carlino, M.; et al. Incidence, predictors, and outcome of thrombosis after successful implantation of drug-eluting stents. JAMA 2005, 293, 2126–2130. [Google Scholar] [CrossRef] [Scilit]
- Moukarbel, G.V. Coronary stent thrombosis and mortality: Does the relationship stand the test of time? J. Am. Heart Assoc. 2022, 11, e025341. [Google Scholar] [CrossRef] [Scilit]
- Cutlip, D.E.; Windecker, S.; Mehran, R.; Boam, A.; Cohen, D.J.; van Es, G.-A.; Gabriel Steg, P.; Morel, M.-A.; Mauri, L.; Vranckx, P.; et al. Clinical end points in coronary stent trials. Circulation 2007, 115, 2344–2351. [Google Scholar] [CrossRef] [Scilit]
- Protty, M.B.; Dissanayake, T.; Jeffery, D.; Hailan, A.; Choudhury, A. Stent failure: The diagnosis and management of intracoronary stent restenosis. Expert Rev. Cardiovasc. Ther. 2023, 21, 501–506. [Google Scholar] [CrossRef] [Scilit]
- Chen, X.; Brox, D.; Assadsangabi, B.; Hsiang, Y.; Takahata, K. Intelligent telemetric stent for wireless monitoring of intravascular pressure and its in vivo testing. Biomed. Microdevices 2014, 16, 745–759. [Google Scholar] [CrossRef] [Scilit]
- Chen, X.; Assadsangabi, B.; Hsiang, Y.; Takahata, K. Enabling angioplasty-ready “smart” stents to detect in-stent restenosis and occlusion. Adv. Sci. 2018, 5, 1700560. [Google Scholar] [CrossRef] [Scilit]
- Islam, S.; Song, X.; Choi, E.T.; Kim, J.; Liu, H.; Kim, A. In vitro study on smart stent for autonomous post-endovascular aneurysm repair surveillance. IEEE Access 2020, 8, 96340–96346. [Google Scholar] [CrossRef] [Scilit]
- Herbert, R.; Lim, H.-R.; Rigo, B.; Yeo, W.-H. Fully implantable wireless batteryless vascular electronics with printed soft sensors for multiplex sensing of hemodynamics. Sci. Adv. 2022, 8, eabm1175. [Google Scholar] [CrossRef] [Scilit]
- Wei, J.; Shanmugasundaram, A.; Oyunbaatar, N.-E.; Wang, L.; Lee, D.-W. Biostable wireless sensor-integrated bioresorbable stent for real-time monitoring of vascular pressure and fractional flow reserve. Microsyst. Nanoeng. 2026, 12, 115. [Google Scholar] [CrossRef] [Scilit]
- Vishnu, J.; Manivasagam, G. Perspectives on smart stents with sensors: From conventional permanent to novel bioabsorbable smart stent technologies. Med. Devices Sens. 2020, 3, e10116. [Google Scholar] [CrossRef] [Scilit]
- Yousefi Darestani, M.R.; Lange, D.; Chew, B.H.; Takahata, K. Electromechanically functionalized ureteral stents for wireless obstruction monitoring. ACS Biomater. Sci. Eng. 2023, 9, 4392–4403. [Google Scholar] [CrossRef] [Scilit]
- Chen, G.; Kim, W.J.; Yang, Y.; Li, Y.R.; Tian, J.; Zhang, J.; Zhao, X.; Scott, K.; Defelice, L.G.; Liu, Z.; et al. Self-powered in-stent restenosis diagnosis via magnetoelastic stents. Nat. Cardiovasc. Res. 2026, 5, 155–167. [Google Scholar] [CrossRef] [Scilit]
- Molloy, A.; Beaumont, K.; Alyami, A.; Kirimi, M.; Hoare, D.; Mirzai, N.; Heidari, H.; Mitra, S.; Neale, S.L.; Mercer, J.R. Challenges to the development of the next generation of self-reporting cardiovascular implantable medical devices. IEEE Rev. Biomed. Eng. 2022, 15, 260–272. [Google Scholar] [CrossRef] [Scilit]
- Sharma, D.; Björninen, T. Stents, smart stents and their importance for wireless biomedical applications. IEEE J. Electromagn. RF Microw. Med. Biol. 2026, 10, 294–306. [Google Scholar] [CrossRef] [Scilit]
- Alghrairi, M.; Sulaiman, N.; Mutashar, S. Health care monitoring and treatment for coronary artery diseases: Challenges and issues. Sensors 2020, 20, 4303. [Google Scholar] [CrossRef] [Scilit]
- Yi, Y.; Wang, B.; Li, C. Sensors-based monitoring and treatment approaches for in-stent restenosis. J. Biomed. Mater. Res. Part B Appl. Biomater. 2023, 111, 490–498. [Google Scholar] [CrossRef] [Scilit]
- Iqbal, M.; Sultan, S.; Qasaimeh, M.A. Novel capacitive mems sensor for monitoring in-stent restenosis. In 2019 41st Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Berlin, Germany, 23–27 July 2019; IEEE: Piscataway, NJ, USA, 2019; pp. 368–371. [Google Scholar]
- Holland, I.; McCormick, C.; Connolly, P. Towards non-invasive characterisation of coronary stent re-endothelialisation–an in-vitro, electrical impedance study. PLoS ONE 2018, 13, e0206758. [Google Scholar] [CrossRef] [Scilit]
- Chen, X.; Brox, D.; Assadsangabi, B.; Mohamed Ali, M.S.; Takahata, K. A stainless-steel-based implantable pressure sensor chip and its integration by microwelding. Sens. Actuators A Phys. 2017, 257, 134–144. [Google Scholar] [CrossRef] [Scilit]
- Kirimi, M.T.; Hoare, D.; Holsgrove, M.; Czyzewski, J.; Mirzai, N.; Mercer, J.R.; Neale, S.L. Detection of blood clots using a whole stent as an active implantable biosensor. Adv. Sci. 2024, 11, e2304748. [Google Scholar] [CrossRef] [Scilit]
- Bulbul, A.; Kim, H. Pressure sensor embedded inductive coil toward a wireless pressure sensing stent. J. Microelectromech. Syst. 2021, 30, 224–233. [Google Scholar] [CrossRef] [Scilit]
- Park, J.; Kim, J.-K.; Kim, D.-S.; Shanmugasundaram, A.; Park, S.A.; Kang, S.; Kim, S.-H.; Jeong, M.H.; Lee, D.-W. Wireless pressure sensor integrated with a 3d printed polymer stent for smart health monitoring. Sens. Actuators B Chem. 2019, 280, 201–209. [Google Scholar] [CrossRef] [Scilit]
- Takahata, K.; DeHennis, A.; Wise, K.D.; Gianchandani, Y.B. Stentenna: A micromachined antenna stent for wireless monitoring of implantable microsensors. In Proceedings of the 25th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (IEEE Cat. No.03CH37439), Cancun, Mexico, 17–21 September 2003; IEEE: Piscataway, NJ, USA, 2003; Volume 3364, pp. 3360–3363. [Google Scholar]
- Takahata, K.; Gianchandani, Y.B. A planar approach for manufacturing cardiac stents: Design, fabrication, and mechanical evaluation. J. Microelectromech. Syst. 2004, 13, 933–939. [Google Scholar] [CrossRef] [Scilit]
- Takahata, K.; DeHennis, A.; Wise, K.D.; Gianchandani, Y.B. A wireless microsensor for monitoring flow and pressure in a blood vessel utilizing a dual-inductor antenna stent and two pressure sensors. In Proceedings of the 17th IEEE International Conference on Micro Electro Mechanical Systems. Maastricht MEMS 2004 Technical Digest, Maastricht, The Netherlands, 25–29 January 2004; IEEE: Piscataway, NJ, USA, 2004; pp. 216–219. [Google Scholar]
- Takahata, K.; Gianchandani, Y.B.; Wise, K.D. Micromachined antenna stents and cuffs for monitoring intraluminal pressure and flow. J. Microelectromech. Syst. 2006, 15, 1289–1298. [Google Scholar] [CrossRef] [Scilit]
- Huang, Q.A.; Dong, L.; Wang, L.F. LC passive wireless sensors toward a wireless sensing platform: Status, prospects, and challenges. J. Microelectromech. Syst. 2016, 25, 822–841. [Google Scholar] [CrossRef] [Scilit]
- Oyunbaatar, N.-E.; Kim, D.-S.; Shanmugasundaram, A.; Kim, S.-H.; Jeong, Y.-J.; Jo, J.; Kwon, K.; Choi, E.; Lee, D.-W. Implantable self-reporting stents for detecting in-stent restenosis and cardiac functional dynamics. ACS Sens. 2023, 8, 4542–4553. [Google Scholar] [CrossRef] [Scilit]
- Herbert, R.; Mishra, S.; Lim, H.R.; Yoo, H.; Yeo, W.H. Fully printed, wireless, stretchable implantable biosystem toward batteryless, real-time monitoring of cerebral aneurysm hemodynamics. Adv. Sci. 2019, 6, 1901034. [Google Scholar] [CrossRef] [Scilit]
- Park, S.Y.; Kim, S.H.; Baik, U.; Huh, S.J.; Avila, R.; Shin, H.H.; Yoo, H.; Shin, H.; Kim, J.-T.; Lim, Y.-H.; et al. A wireless, implantable sensor for continuous monitoring of blood leakage after endovascular aneurysm repair. Sci. Adv. 2025, 11, eady6148. [Google Scholar] [CrossRef] [Scilit]
- Shalabi, N.; Searles, K.; Herout, R.; Almutairi, K.; Chew, B.H.; Lange, D.; Takahata, K. Ureteral stent sleeve for early detection of hydronephrosis. Microsyst. Nanoeng. 2026, 12, 97. [Google Scholar] [CrossRef] [Scilit]
- Oyunbaatar, N.E.; Wei, J.; Wang, L.; Kim, S.H.; Lee, H.; Kwon, K.; Won, Y.; Lee, D.W. 3d-printed cnt-reinforced bioresorbable vascular scaffold with enhanced mechanical stability and integrated wireless pressure sensor for continuous hemodynamic monitoring. ACS Sens. 2025, 10, 5724–5735. [Google Scholar] [CrossRef] [Scilit]
- Bateman, A.; He, Y.; Cherono, C.; Lee, J.; Ghalichechian, N.; Yeo, W.-H. Implantable membrane sensors and long-range wireless electronics for continuous monitoring of stent edge restenosis. ACS Appl. Mater. Interfaces 2025, 17, 42781–42790. [Google Scholar] [CrossRef] [Scilit]
- Herbert, R.; Elsisy, M.; Rigo, B.; Lim, H.R.; Kim, H.; Choi, C.; Kim, S.; Ye, S.H.; Wagner, W.R.; Chun, Y.; et al. Fully implantable batteryless soft platforms with printed nanomaterial-based arterial stiffness sensors for wireless continuous monitoring of restenosis in real time. Nano Today 2022, 46, 101557. [Google Scholar] [CrossRef] [Scilit]
- Gradus-Pizlo, I.; Sawada, S.G.; Wright, D.; Segar, D.S.; Feigenbaum, H. Detection of subclinical coronary atherosclerosis using two-dimensional, high-resolution transthoracic echocardiography. J. Am. Coll. Cardiol. 2001, 37, 1422–1429. [Google Scholar] [CrossRef] [Scilit]
- American Heart Association. Tachycardia: Fast Heart Rate. Available online: https://www.heart.org/en/health-topics/arrhythmia/about-arrhythmia/tachycardia--fast-heart-rate (accessed on 15 November 2026).
- Roshani, H.; Dabhoiwala, N.F.; Dijkhuis, T.; Lamers, W.H. Intraluminal pressure changes in vivo in the middle and distal pig ureter during propagation of a peristaltic wave. Urology 2002, 59, 298–302. [Google Scholar] [CrossRef] [Scilit]
- Hill, W.G. Control of urinary drainage and voiding. Clin. J. Am. Soc. Nephrol. 2015, 10, 480–492. [Google Scholar] [CrossRef] [Scilit]
- Amado, P.; Zheng, S.; Lange, D.; Carugo, D.; Waters, S.L.; Obrist, D.; Burkhard, F.; Clavica, F. The interplay between bacterial biofilms, encrustation, and wall shear stress in ureteral stents: A review across scales. Front. Urol. 2024, 3, 1335414. [Google Scholar] [CrossRef] [Scilit]
- Boiselle, P.M.; O’Donnell, C.R.; Bankier, A.A.; Ernst, A.; Millet, M.E.; Potemkin, A.; Loring, S.H. Tracheal collapsibility in healthy volunteers during forced expiration: Assessment with multidetector CT. Radiology 2009, 252, 255–262. [Google Scholar] [CrossRef] [Scilit]
- Mitropoulos, A.; Song, W.-J.; Almaghlouth, F.; Kemp, S.; Polkey, M.; Hull, J.H. Detection and diagnosis of large airway collapse: A systematic review. ERJ Open Res. 2021, 7, 00055–02021. [Google Scholar] [CrossRef] [Scilit]
- Roodenburg, S.A.; Pouwels, S.D.; Slebos, D.-J. Airway granulation response to lung-implantable medical devices: A concise overview. Eur. Respir. Rev. 2021, 30, 210066. [Google Scholar] [CrossRef] [Scilit]
- Gupta, A.; Slebos, D.-J.; Pouwels, S.D. Lung implantable devices: The issue with granulation tissue. Breathe 2025, 21, 240243. [Google Scholar] [CrossRef] [Scilit]
- Madisi, N.Y.; Greenberg, D.; Balwan, A. Airway stent associated respiratory tract infections. AME Med. J. 2024, 9, 3. [Google Scholar] [CrossRef] [Scilit]
- Wang, Y.; Negron, C.; Khoshnaw, A.; Edwards, S.; Vu, H.; Quatela, J.; Park, N.; Maldonado, F.; Demarest, C.; Simon, V.; et al. Sensory artificial cilia for in situ monitoring of airway physiological properties. Proc. Natl. Acad. Sci. USA 2024, 121, e2412086121. [Google Scholar] [CrossRef] [Scilit]
- Ruiz, L.J.L.; Zhu, J.; Fitzgerald, L.; Quinn, D.; Lach, J. Capacitive sensing for monitoring stent patency in the central airway. In Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference; IEEE: New York, NY, USA, 2021; pp. 5441–5445. [Google Scholar]
- Ge, R.; Wang, Y.; Negron, C.; Fan, H.; Maldonado, F.; Demarest, C.T.; Simon, V.; Zhou, Y.; Dong, X. A wireless implantable sensory ring for continuous airway stent migration tracking. npj Flex. Electron. 2026, 10, 25. [Google Scholar] [CrossRef] [Scilit]
- Wang, Y.; Kim, M.; Fan, H.; Ge, R.; Wang, D.; Maldonado, F.; Demarest, C.; Zhou, Y.; Dong, X. A miniature magnetic switch unlocking multimodal, chip-free, and batteryless airway sensing. Sci. Adv. 2026, 12, eaed3998. [Google Scholar] [CrossRef] [Scilit]
- Kim, D.-S.; Oyunbaatar, N.-E.; Shanmugasundaram, A.; Jeong, Y.-J.; Park, J.; Lee, D.-W. Stress induced self-rollable smart-stent-based u-health platform for in-stent restenosis monitoring. Analyst 2022, 147, 4793–4803. [Google Scholar] [CrossRef] [Scilit]
- Yousefi Darestani, M.R.; Shalabi, N.; Lange, D.; Chew, B.H.; Takahata, K. Intelligent ureteral stent for early detection of hydronephrosis. Adv. Mater. Technol. 2021, 6, 2100652. [Google Scholar] [CrossRef] [Scilit]
- Lv, L.; Liu, T.; Jiang, T.; Li, J.; Zhang, J.; Zhou, Q.; Dhakal, R.; Li, X.; Li, Y.; Yao, Z. A highly sensitive flexible capacitive pressure sensor with hierarchical pyramid micro-structured pdms-based dielectric layer for health monitoring. Front. Bioeng. Biotechnol. 2023, 11, 1303142. [Google Scholar] [CrossRef] [Scilit]
- Nur, R.; Matsuhisa, N.; Jiang, Z.; Nayeem, M.O.G.; Yokota, T.; Someya, T. A highly sensitive capacitive-type strain sensor using wrinkled ultrathin gold films. Nano Lett. 2018, 18, 5610–5617. [Google Scholar] [CrossRef] [Scilit]
- Mannsfeld, S.C.B.; Tee, B.C.K.; Stoltenberg, R.M.; Chen, C.V.H.H.; Barman, S.; Muir, B.V.O.; Sokolov, A.N.; Reese, C.; Bao, Z. Highly sensitive flexible pressure sensors with microstructured rubber dielectric layers. Nat. Mater. 2010, 9, 859–864. [Google Scholar] [CrossRef] [Scilit]
- Ding, S.; Pichon, L.; Chen, Y. A low-cost microwave stentenna for in-stent restenosis detection. IEEE Trans. Antennas Propag. 2025, 73, 8681–8692. [Google Scholar] [CrossRef] [Scilit]
- Clare, J.; Ganly, J.; Bursill, C.A.; Sumer, H.; Kingshott, P.; de Haan, J.B. The mechanisms of restenosis and relevance to next generation stent design. Biomolecules 2022, 12, 430. [Google Scholar] [CrossRef] [Scilit]
- Gabriel, C.; Gabriel, S.; Corthout, E. The dielectric properties of biological tissues: I. Literature survey. Phys. Med. Biol. 1996, 41, 2231. [Google Scholar] [CrossRef] [Scilit]
- Pereiro, I.; Fomitcheva Khartchenko, A.; Petrini, L.; Kaigala, G.V. Nip the bubble in the bud: A guide to avoid gas nucleation in microfluidics. Lab Chip 2019, 19, 2296–2314. [Google Scholar] [CrossRef] [Scilit]
- Rigo, B.; Bateman, A.; Lee, J.; Kim, H.; Lee, Y.; Romero, L.; Jang, Y.C.; Herbert, R.; Yeo, W.-H. Soft implantable printed bioelectronic system for wireless continuous monitoring of restenosis. Biosens. Bioelectron. 2023, 241, 115650. [Google Scholar] [CrossRef] [Scilit]
- Hoare, D.; Tsiamis, A.; Marland, J.R.K.; Czyzewski, J.; Kirimi, M.T.; Holsgrove, M.; Russell, E.; Neale, S.L.; Mirzai, N.; Mitra, S.; et al. Predicting cardiovascular stent complications using self-reporting biosensors for noninvasive detection of disease. Adv. Sci. 2022, 9, 2105285. [Google Scholar] [CrossRef] [Scilit]
- Dadkhah Tehrani, F.; O’Toole, M.D.; Collins, D.J. Tutorial on impedance and dielectric spectroscopy for single-cell characterisation on microfluidic platforms: Theory, practice, and recent advances. Lab Chip 2025, 25, 837–855. [Google Scholar] [CrossRef] [Scilit]
- Claudel, J.; Ngo, T.-T.; Kourtiche, D.; Nadi, M. Interdigitated sensor optimization for blood sample analysis. Biosensors 2020, 10, 208. [Google Scholar] [CrossRef] [Scilit]
- Islam, S.; Tan, J.Y.; Mou, T.D.; Ray, S.; Liu, H.; Kim, J.; Kim, A. 3d-printable self-powered piezoelectric smart stent for wireless endoleaks sensing. In 2024 IEEE 37th International Conference on Micro Electro Mechanical Systems (MEMS), Austin, TX, USA, 21–25 January 2024; IEEE: Piscataway, NJ, USA, 2024; pp. 789–792. [Google Scholar]
- Tan, J.Y.; Kim, C.; Mou, T.D.; Kim, A.; Kim, J. Wireless pressure sensing smart stent for enhanced post-endovascular aneurysm repair (EVAR) surveillance. IEEE Access 2024, 12, 198123–198131. [Google Scholar] [CrossRef] [Scilit]
- Tan, J.Y.; Ray, S.; Liu, H.; Kim, A.; Kim, J. Wireless smart stent for real-time in-stent pressure monitoring and endoleak detection. IEEE Sens. J. 2026, 26, 19239–19250. [Google Scholar] [CrossRef] [Scilit]
- Zhao, Z.; Cai, H.; Wang, H.; Wang, L.; Wang, J.; Cao, X.; Chen, X. Stent-based wireless glucose monitoring system. In 2024 IEEE 19th International Conference on Nano/Micro Engineered and Molecular Systems (NEMS), Kyoto, Japan, 2–5 May 2024; IEEE: Piscataway, NJ, USA, 2024; pp. 1–4. [Google Scholar]
- Al-Zu’bi, M.M.; Mohan, A.S. Implantable biosensor interface platform for monitoring of atherosclerosis. IEEE Sens. Lett. 2020, 4, 5500204. [Google Scholar] [CrossRef] [Scilit]
- Son, D.; Lee, J.; Lee, D.J.; Ghaffari, R.; Yun, S.; Kim, S.J.; Lee, J.E.; Cho, H.R.; Yoon, S.; Yang, S.; et al. Bioresorbable electronic stent integrated with therapeutic nanoparticles for endovascular diseases. ACS Nano 2015, 9, 5937–5946. [Google Scholar] [CrossRef] [Scilit]
- Chen, Y.; Fu, K.X.; Cotton, R.; Ou, Z.; Kwak, J.W.; Chien, J.-C.; Kesler, V.; Nyein, H.Y.Y.; Eisenstein, M.; Soh, H.T. A biochemical sensor with continuous extended stability in vivo. Nat. Biomed. Eng. 2025, 9, 1517–1530. [Google Scholar] [CrossRef] [Scilit]
- Hu, C.; Wang, L.; Liu, S.; Sheng, X.; Yin, L. Recent development of implantable chemical sensors utilizing flexible and biodegradable materials for biomedical applications. ACS Nano 2024, 18, 3969–3995. [Google Scholar] [CrossRef] [Scilit]
- Zhuk, I.; Jariwala, F.; Attygalle, A.B.; Wu, Y.; Libera, M.R.; Sukhishvili, S.A. Self-defensive layer-by-layer films with bacteria-triggered antibiotic release. ACS Nano 2014, 8, 7733–7745. [Google Scholar] [CrossRef] [Scilit]
- Cado, G.; Aslam, R.; Séon, L.; Garnier, T.; Fabre, R.; Parat, A.; Chassepot, A.; Voegel, J.-C.; Senger, B.; Schneider, F.; et al. Self-defensive biomaterial coating against bacteria and yeasts: Polysaccharide multilayer film with embedded antimicrobial peptide. Adv. Funct. Mater. 2013, 23, 4801–4809. [Google Scholar] [CrossRef] [Scilit]
- Brox, D.S.; Chen, X.; Mirabbasi, S.; Takahata, K. Wireless telemetry of stainless-steel-based smart antenna stent using a transient resonance method. IEEE Antennas Wirel. Propag. Lett. 2016, 15, 754–757. [Google Scholar] [CrossRef] [Scilit]
- C95.1-2019; IEEE Standard for Safety Levels with Respect to Human Exposure to Electric, Magnetic, and Electromagnetic Fields, 0 hz to 300 ghz. IEEE: New York, NY, USA, 2019.
- Wiseman, G.; Redoute, J.M.; Ganesan, K.; Rind, G.S.; John, S.E.; Ronayne, S.M.; Opie, N.L.; Garrett, D.J.; Prawer, S. Near-field wireless power transfer to stent-based biomedical implants. IEEE J. Electromagn. RF Microw. Med. Biol. 2018, 2, 193–200. [Google Scholar] [CrossRef] [Scilit]
- International Commission on Non-Ionizing Radiation Protection (ICNIRP). Guidelines for limiting exposure to electromagnetic fields (100 khz to 300 ghz). Health Phys. 2020, 118, 483–524. [Google Scholar] [CrossRef] [Scilit]
- Federal Communications Commission. Radiofrequency Radiation Exposure Limits, 47 CFR § 1.1310. Available online: https://www.ecfr.gov/current/title-47/chapter-I/subchapter-A/part-1/subpart-I/section-1.1310 (accessed on 15 November 2026).
- Shah, S.A.A.; Lim, Y.H.; Yoo, H. A novel development of endovascular aortic stent system featuring promising antenna characteristics. IEEE Trans. Antennas Propag. 2022, 70, 2214–2222. [Google Scholar] [CrossRef] [Scilit]
- Shah, S.A.A.; Basir, A.; Lim, Y.H.; Yoo, H. A novel efficient wirelessly powered biotelemetric endovascular aortic stent antenna system. IEEE Trans. Antennas Propag. 2023, 71, 7132–7145. [Google Scholar] [CrossRef] [Scilit]
- John, B.; Spink, C.; Braunschweig, M.; Ranjan, R.; Schroeder, D.; Koops, A.; Woldt, G.; Rauh, I.; Leuzinger, A.; Adam, G.; et al. Telemetric system for monitoring of endoleak in abdominal aorta aneurysm using multiple pressure sensors integrated on a stent graft. In 2016 IEEE Biomedical Circuits and Systems Conference (BioCAS), Shanghai, China, 17–19 October 2016; IEEE: Piscataway, NJ, USA, 2016; pp. 384–387. [Google Scholar]
- Jackson, J.D. Classical Electrodynamics, 3rd ed.; Wiley: New York, NY, USA, 1998. [Google Scholar]
- Liu, C.-H.; Chen, S.-C.; Hsiao, H.-M. A single-connector stent antenna for intravascular monitoring applications. Sensors 2019, 19, 4616. [Google Scholar] [CrossRef] [Scilit]
- Vrints, C.; Andreotti, F.; Koskinas, K.C.; Rossello, X.; Adamo, M.; Ainslie, J.; Banning, A.P.; Budaj, A.; Buechel, R.R.; Chiariello, G.A.; et al. 2024 esc guidelines for the management of chronic coronary syndromes: Developed by the task force for the management of chronic coronary syndromes of the european society of cardiology (esc) endorsed by the european association for cardio-thoracic surgery (eacts). Eur. Heart J. 2024, 45, 3415–3537. [Google Scholar] [CrossRef] [Scilit]
- Starr, P.; Bartels, K.; Agrawal, C.M.; Bailey, S. A thin-film pressure transducer for implantable and intravascular blood pressure sensing. Sens. Actuators A Phys. 2016, 248, 38–45. [Google Scholar] [CrossRef] [Scilit]
- Lee, S.H.; Lee, H.J.; Lee, S.; Kim, D.-H.; Kim, H.J.; Sunwoo, S.-H. Intrinsically soft implantable electronics for long-term biosensing applications. Adv. Sens. Res. 2025, 4, 2500002. [Google Scholar] [CrossRef] [Scilit]
- Zhao, F.; Liu, F.; Gao, C.; Wang, G.; Zhang, Y.; Yu, F.; Tian, J.; Tan, K.; Zhang, R.; Liang, K.; et al. Interfacing exogenous stents with human coronary artery by self-assembled coating: Designs, functionalities and applications. NPG Asia Mater. 2024, 16, 28. [Google Scholar] [CrossRef] [Scilit]
- Vishnu, J.; Manivasagam, G.; Mantovani, D.; Udduttula, A.; Coathup, M.J.; Popat, K.C.; Ren, P.G.; Prashanth, K.G. Balloon expandable coronary stent materials: A systematic review focused on clinical success. In Vitro Models 2022, 1, 151–175, Correction in In Vitro Models 2022, 1, 177. [Google Scholar] [CrossRef] [Scilit]
- Park, J.; Kim, J.-K.; Patil, S.J.; Park, J.-K.; Park, S.; Lee, D.-W. A wireless pressure sensor integrated with a biodegradable polymer stent for biomedical applications. Sensors 2016, 16, 809. [Google Scholar] [CrossRef] [Scilit]
- Tesfamariam, B. Bioresorbable vascular scaffolds: Biodegradation, drug delivery and vascular remodeling. Pharmacol. Res. 2016, 107, 163–171. [Google Scholar] [CrossRef] [Scilit]
- Jinnouchi, H.; Torii, S.; Sakamoto, A.; Kolodgie, F.D.; Virmani, R.; Finn, A.V. Fully bioresorbable vascular scaffolds: Lessons learned and future directions. Nat. Rev. Cardiol. 2019, 16, 286–304. [Google Scholar] [CrossRef] [Scilit]
- Fiuza, C.; Polak-Kraśna, K.; Antonini, L.; Petrini, L.; Carroll, O.; Ronan, W.; Vaughan, T.J. An experimental investigation into the physical, thermal and mechanical degradation of a polymeric bioresorbable scaffold. J. Mech. Behav. Biomed. Mater. 2022, 125, 104955. [Google Scholar] [CrossRef] [Scilit]
- Wang, P.-J.; Ferralis, N.; Conway, C.; Grossman, J.C.; Edelman, E.R. Strain-induced accelerated asymmetric spatial degradation of polymeric vascular scaffolds. Proc. Natl. Acad. Sci. USA 2018, 115, 2640–2645. [Google Scholar] [CrossRef] [Scilit]
- Toong, D.W.Y.; Toh, H.W.; Ng, J.C.K.; Wong, P.E.H.; Leo, H.L.; Venkatraman, S.; Tan, L.P.; Ang, H.Y.; Huang, Y. Bioresorbable polymeric scaffold in cardiovascular applications. Int. J. Mol. Sci. 2020, 21, 3444. [Google Scholar] [CrossRef] [Scilit]
- Lin, W.; Qin, L.; Qi, H.; Zhang, D.; Zhang, G.; Gao, R.; Qiu, H.; Xia, Y.; Cao, P.; Wang, X.; et al. Long-term in vivo corrosion behavior, biocompatibility and bioresorption mechanism of a bioresorbable nitrided iron scaffold. Acta Biomater. 2017, 54, 454–468. [Google Scholar] [CrossRef] [Scilit]
- Shen, D.; Qi, H.; Lin, W.; Zhang, W.; Bian, D.; Shi, X.; Qin, L.; Zhang, G.; Fu, W.; Dou, K.; et al. PDLLA-Zn-nitrided Fe bioresorbable scaffold with 53-μm-thick metallic struts and tunable multistage biodegradation function. Sci. Adv. 2021, 7, eabf0614. [Google Scholar] [CrossRef] [Scilit]
- Yang, H.; Wang, C.; Liu, C.; Chen, H.; Wu, Y.; Han, J.; Jia, Z.; Lin, W.; Zhang, D.; Li, W.; et al. Evolution of the degradation mechanism of pure zinc stent in the one-year study of rabbit abdominal aorta model. Biomaterials 2017, 145, 92–105. [Google Scholar] [CrossRef] [Scilit]
- Zhou, C.; Li, H.-F.; Yin, Y.-X.; Shi, Z.-Z.; Li, T.; Feng, X.-Y.; Zhang, J.-W.; Song, C.-X.; Cui, X.-S.; Xu, K.-L.; et al. Long-term in vivo study of biodegradable Zn-Cu stent: A 2-year implantation evaluation in porcine coronary artery. Acta Biomater. 2019, 97, 657–670. [Google Scholar] [CrossRef] [Scilit]
- Pacharra, S.; Ortiz, R.; McMahon, S.; Wang, W.; Viebahn, R.; Salber, J.; Quintana, I. Surface patterning of a novel PEG-functionalized poly-L-lactide polymer to improve its biocompatibility: Applications to bioresorbable vascular stents. J. Biomed. Mater. Res. Part B Appl. Biomater. 2019, 107, 624–634. [Google Scholar] [CrossRef] [Scilit]
- Kang, S.; Chi, Y.J.; Cho, K.; Lee, H.J.; Koh, W.G. Dual surface modification of poly(L-lactide) scaffold achieved by thermal incorporation of aligned nanofiber and click immobilization of VEGF to enhance endothelialization and blood compatibility. Appl. Surf. Sci. 2022, 589, 152969. [Google Scholar] [CrossRef] [Scilit]
- Lansky, A.; Wijns, W.; Xu, B.; Kelbæk, H.; van Royen, N.; Zheng, M.; Morel, M.-A.; Knaapen, P.; Slagboom, T.; Johnson, T.W.; et al. Targeted therapy with a localised abluminal groove, low-dose sirolimus-eluting, biodegradable polymer coronary stent (TARGET All Comers): A multicentre, open-label, randomised non-inferiority trial. The Lancet 2018, 392, 1117–1126. [Google Scholar] [CrossRef] [Scilit]
- Haude, M.; Lee, S.W.; Worthley, S.G.; Silber, S.; Verheye, S.; Erbs, S.; Rosli, M.A.; Botelho, R.; Meredith, I.; Sim, K.H.; et al. The REMEDEE trial: A randomized comparison of a combination sirolimus-eluting endothelial progenitor cell capture stent with a paclitaxel-eluting stent. JACC Cardiovasc. Interv. 2013, 6, 334–343. [Google Scholar]
- Opie, N.L.; John, S.E.; Rind, G.S.; Ronayne, S.M.; Grayden, D.B.; Burkitt, A.N.; May, C.N.; O’Brien, T.J.; Oxley, T.J. Chronic impedance spectroscopy of an endovascular stent-electrode array. J. Neural Eng. 2016, 13, 046020. [Google Scholar] [CrossRef] [Scilit]
- Opie, N.L.; van der Nagel, N.R.; John, S.E.; Vessey, K.; Rind, G.S.; Ronayne, S.M.; Fletcher, E.L.; May, C.N.; O’Brien, T.J.; Oxley, T.J. Micro-CT and histological evaluation of an neural interface implanted within a blood vessel. IEEE Trans. Biomed. Eng. 2017, 64, 928–934. [Google Scholar] [CrossRef] [Scilit]
- Abdelkader, A.E.A.; Ilonze, O.J.; Guglin, M. Dysfunctional implantable pulmonary artery sensor device (CardioMEMS) in group 2 pulmonary hypertension. JACC Case Rep. 2025, 30, 102801. [Google Scholar] [CrossRef] [Scilit]
- Buhmann, M.T.; Abt, D.; Nolte, O.; Neu, T.R.; Strempel, S.; Albrich, W.C.; Betschart, P.; Zumstein, V.; Neels, A.; Maniura-Weber, K.; et al. Encrustations on ureteral stents from patients without urinary tract infection reveal distinct urotypes and a low bacterial load. Microbiome 2019, 7, 60. [Google Scholar] [CrossRef] [Scilit]
- Wolf, M.; Gulich, R.; Lunkenheimer, P.; Loidl, A. Broadband dielectric spectroscopy on human blood. Biochim. Biophys. Acta (BBA)-Gen. Subj. 2011, 1810, 727–740. [Google Scholar] [CrossRef] [Scilit]
- Nambisan, R.M.; Green, S.R.; Kwon, R.S.; Elta, G.H.; Gianchandani, Y.B. A microsystem for in vivo wireless monitoring of plastic biliary stents using magnetoelastic sensors. Microsyst. Nanoeng. 2024, 10, 159. [Google Scholar] [CrossRef] [Scilit]
- Morton, A.C.; Arnold, N.D.; Gunn, J.; Varcoe, R.; Francis, S.E.; Dower, S.K.; Crossman, D.C. Interleukin-1 receptor antagonist alters the response to vessel wall injury in a porcine coronary artery model. Cardiovasc. Res. 2005, 68, 493–501. [Google Scholar] [CrossRef] [Scilit]
- Zafar, H.; Neelam-Naganathan, D.; Middleton, J.T.; Binmahfooz, S.K.; Battersby, C.; Rogers, D.; Swift, A.J.; Rothman, A.M.K. Anatomical characterization of pulmonary artery and implications to pulmonary artery pressure monitor implantation. Sci. Rep. 2023, 13, 20528. [Google Scholar] [CrossRef] [Scilit]
- Cleven, N.J.; Isfort, P.; Penzkofer, T.; Woitok, A.; Hermanns-Sachweh, B.; Steinseifer, U.; Schmitz-Rode, T. Wireless blood pressure monitoring with a novel implantable device: Long-term in vivo results. Cardiovasc. Interv. Radiol. 2014, 37, 1580–1588. [Google Scholar] [CrossRef] [Scilit]
- Green, S.R.; Kwon, R.S.; Elta, G.H.; Gianchandani, Y.B. In vivo and in situ evaluation of a wireless magnetoelastic sensor array for plastic biliary stent monitoring. Biomed. Microdevices 2013, 15, 509–517. [Google Scholar] [CrossRef] [Scilit]
- Räber, L.; Mintz, G.S.; Koskinas, K.C.; Johnson, T.W.; Holm, N.R.; Onuma, Y.; Radu, M.D.; Joner, M.; Yu, B.; Jia, H.; et al. Clinical use of intracoronary imaging. Part 1: Guidance and optimization of coronary interventions. An expert consensus document of the European Association of Percutaneous Cardiovascular Interventions. Eur. Heart J. 2018, 39, 3281–3300, Correction in Eur. Heart J. 2019, 40, 308. [Google Scholar] [CrossRef] [Scilit]
- Baù, M.; Demori, M.; Ferrari, M.; Ferrari, V. Contactless readout of passive LC sensors with compensation circuit for distance-independent measurements. Proceedings 2018, 2, 842. [Google Scholar] [CrossRef] [Scilit]
- U.S. Food and Drug Administration. Non-Clinical Engineering Tests and Recommended Labeling for Intravascular Stents and Associated Delivery Systems: Guidance for Industry and FDA Staff. Available online: https://www.fda.gov/regulatory-information/search-fda-guidance-documents/non-clinical-engineering-tests-and-recommended-labeling-intravascular-stents-and-associated-delivery (accessed on 15 November 2026).
- European Parliament and Council of the European Union. Regulation (EU) 2017/745 of 5 April 2017 on Medical Devices, Amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 and Repealing Council Directives 90/385/EEC and 93/42/EEC. Available online: https://eur-lex.europa.eu/eli/reg/2017/745/oj (accessed on 15 November 2026).
- National Medical Products Administration. Rules for Classification of Medical Devices (Decree No. 15 of China Food and Drug Administration). Available online: https://english.nmpa.gov.cn/2019-10/11/c_415411.htm (accessed on 15 November 2026).
- U.S. Food and Drug Administration. Modifications to Devices Subject to Premarket Approval (PMA): The PMA Supplement Decision-Making Process: Guidance for Industry and FDA Staff. Available online: https://www.fda.gov/regulatory-information/search-fda-guidance-documents/modifications-devices-subject-premarket-approval-pma-pma-supplement-decision-making-process (accessed on 15 November 2026).
- U.S. Food and Drug Administration. Use of International Standard ISO 10993-1, “Biological Evaluation of Medical Devices-Part 1: Evaluation and Testing Within a Risk Management Process”: Guidance for Industry and Food and Drug Administration Staff. Available online: https://www.fda.gov/regulatory-information/search-fda-guidance-documents/use-international-standard-iso-10993-1-biological-evaluation-medical-devices-part-1-evaluation-and (accessed on 15 November 2026).
- U.S. Food and Drug Administration. Select Updates for Non-Clinical Engineering Tests and Recommended Labeling for Intravascular Stents and Associated Delivery Systems: Guidance for Industry and Food and Drug Administration Staff. Available online: https://www.fda.gov/regulatory-information/search-fda-guidance-documents/select-updates-non-clinical-engineering-tests-and-recommended-labeling-intravascular-stents-and (accessed on 15 November 2026).
- U.S. Food and Drug Administration. Electromagnetic Compatibility (EMC) of Medical Devices: Guidance for Industry and Food and Drug Administration Staff. Available online: https://www.fda.gov/regulatory-information/search-fda-guidance-documents/electromagnetic-compatibility-emc-medical-devices (accessed on 15 November 2026).
- U.S. Food and Drug Administration. Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment: Guidance for Industry and Food and Drug Administration Staff. Available online: https://www.fda.gov/regulatory-information/search-fda-guidance-documents/testing-and-labeling-medical-devices-safety-magnetic-resonance-mr-environment (accessed on 15 November 2026).
- Paunović, N.; Bao, Y.; Coulter, F.B.; Masania, K.; Geks, A.K.; Klein, K.; Rafsanjani, A.; Cadalbert, J.; Kronen, P.W.; Kleger, N.; et al. Digital light 3d printing of customized bioresorbable airway stents with elastomeric properties. Sci. Adv. 2021, 7, eabe9499. [Google Scholar] [CrossRef] [Scilit]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.