Abstract
Nanowell impedance sensors offer enhanced sensitivity, label-free detection, and real-time monitoring capabilities, making them ideal for precise biomolecule detection and analysis. Despite extensive research, further characterization is required to ensure their safe application in humans. Our study presents significant sensor characterizations, including comprehensive biocompatibility studies, to underscore their potential. We first demonstrated femtomolar (fM)-level detection of IL-6, TNF-α, CXCL5, VEGF, and HMGB1, with detection limits of 100 fM, 100 fM, 300 fM, 500 fM, and 400 fM, respectively. Additional characterizations differentiated negative control and antibody/target biomarker responses across eight frequencies, ranging from 1 MHz to 40 MHz. No significant differences in viable cell numbers or morphology were observed in RPE1 and HEPG2 cell lines exposed to the sensor compared with the control groups across various time intervals. Following successful acquisition of in vitro biocompatibility results, a comprehensive in vivo assessment using 10 mice demonstrated that wound closure at 21 days was not impeded by the presence of the sensor (p < 0.05), and no significant differences were observed in ELISA measurements of IL-6 and TNF-α (p < 0.05). These results demonstrate the nanowell impedance sensor’s exceptional sensitivity, broad biomarker detection capabilities, and verified biocompatibility, establishing a strong foundation for advancing its application.
1. Introduction
Biomarkers enhance the sensitivity and specificity of detecting exposures or risk factors, which is crucial for identifying disease onset, tracking its progression, and evaluating treatment effectiveness [1,2,3,4,5]. The systematic discovery of new biomarker proteins and the development of key assays enable the creation and validation of precise, reliable methods for detecting known biomarkers [6]. Affinity-based label-based biosensors typically use fluorescent labels to identify biomarkers, offering higher specificity compared to another type of affinity-based biosensors, label-free biosensors, in complex mixtures due to tagging [7]. However, they require additional steps and may interfere with real-time analysis. In contrast, label-free detection allows direct, real-time monitoring without altering the analyte’s properties through labeling, providing enhanced portability [7,8,9]. Label-free, electronic-based technologies also provide significant benefits, such as ease of miniaturization and minimal reagent use [10,11]. Optical detection leverages the precision of light-based measurements, while electronic detection streamlines device design by eliminating the reliance on complex optical setups [12,13]. Electrical biosensors, which measure biomarker interactions through electrical currents or voltages, are particularly advantageous due to their low cost, minimal power consumption, and ease of miniaturization [14,15,16]. These features make them ideal for applications requiring compact, affordable solutions, such as point-of-care diagnostics and clinical settings [17,18,19]. Impedance biosensors excel in providing label-free, real-time detection with streamlined device architecture, making them well-suited for integration into portable and user-friendly diagnostic systems [20,21]. These biosensors encounter challenges such as limited sensitivity, susceptibility to noise, and micrometer-scale spacing of sensing electrodes, which impact their sensitivity [22,23,24]. Hence, the geometry and design of the setup should feature a small distance between sensing electrodes, enable rapid detection of target biomarkers, and ensure effective signal extraction amidst noise. Additionally, it should support antibody immobilization while enabling the localized detection of target biomarkers, such as proteins.
Several key conditions must be satisfied before using sensors outside the laboratory or in human applications. One of these key factors is the sensor’s ability to perform sensitive measurements [25]. Detecting biomarkers at femtomolar levels significantly improves the precision of diagnostic tests by minimizing the chance of false negatives and enhancing overall accuracy [26]. This low-level detection is especially valuable for spotting diseases in their early stages, even before symptoms emerge, which enables prompt intervention and treatment [26]. The analysis of multiple biomarkers, rather than one, is essential for acquiring a comprehensive understanding and accurate diagnosis of certain diseases, such as cancer, Alzheimer’s disease, and cardiovascular conditions [27,28,29]. Monitoring multiple biomarkers at these minimal concentrations fosters a clearer and more precise understanding of disease progression or response to treatment, leading to better patient management [30]. Such sensitive detections have the potential to be used for more customized treatments based on individual biomarker profiles, potentially resulting in more personalized and effective therapeutic approaches [31]. Additional characterizations ensure the biosensor’s reliability and accuracy, preparing it for real-world applications [32]. Specifically, studying a sensor’s performance at higher frequencies broadens its applications by enabling wireless operation, which typically occurs at these frequencies [33]. This enhancement improves portability and supports real-time, remote monitoring, making the sensor more versatile in various settings [34,35].
It is essential to establish the biocompatibility of a biosensor through comprehensive testing before its use for measurement in humans [36]. Monitoring cell morphology and proliferation over time allows us to detect cytotoxic effects; viable cell counts complement these observations by revealing underlying issues even when cells appear normal [37]. Ensuring an in vitro biocompatibility study is successful allows for the advancement of in vivo studies using live animals [38]. These studies evaluate how well the biosensor integrates with biological tissue without causing rejection, confirm material compatibility to prevent harmful substances from leaching into wounds, and verify that a device maintains functionality over time without degradation [39,40]. Assessing the sensor’s impact on wound healing and inflammatory responses is important for evaluating tissue compatibility and potential adverse effects [41]. Safety assurance is also achieved by confirming that toxic or adverse reactions are not caused when the biosensor is in contact with biological tissues [42]. Evaluating healing rates and tissue responses, such as inflammation or swelling, determines if the sensor affects wound-healing speed and identifies any adverse reactions, with minimal responses indicating the sensor is well-tolerated [43]. Ultimately, these biocompatibility experiments provide visual and measurable evidence, strengthening a sensor’s validity for biomedical applications and offering insights into its long-term biocompatibility [44].
In the following sections, the fabrication process for a nanowell sensor containing a nanowell array with an overlapping area of two electrodes separated by a 40 nm gap is initially discussed, which enables sensitive detection of target biomarkers. This is followed by an overview of the nanowell array showing how the binding of target biomarkers in the nanowells changes electrical properties between the electrodes, enabling quantification of the biomarkers. Reagent preparation, specifications of the setup for performing real-time impedance measurements, and considerations for in vitro biocompatibility experiments, such as the required cell culture, as well as for in vivo studies, including live animal wound models and wound fluid collection, are also introduced. The full experimental procedure for detecting target biomarkers is outlined, along with an evaluation of the multiplexing performance of the nanowell impedance sensor for the sensitive detection of six key biomarkers. The subsequent section covers the physical characterization of the nanowell sensor. Validation of the sensor’s in vitro biocompatibility is discussed by assessing cellular parameters with a viable cell counter at distinct time intervals and comparing cellular morphology between the control group (without the sensor) and the test group (with the sensor). Finally, a comprehensive in vivo biocompatibility experiment is conducted using 10 mice to analyze wound closure over 21 days, comparing control and test animals. Additionally, rigorous ELISA testing on collected samples is performed to measure cytokine levels in both groups.
2. Materials and Methods
2.1. Overview of the Nanowell Impedance Sensor
The fabrication of nanowell impedance sensors involves a series of precise steps. Oxygen plasma is initially used to clean a fused silica wafer (University Wafer, South Boston, MA, USA). The initial electrode was patterned using standard photolithography techniques, and the electrode layer (Bottom electrode) was deposited through electron beam evaporation, followed by a lift-off process. This metal layer included a 5 nm chromium underlayer to enhance adhesion, with a 100 nm gold film on top. To protect the electrode, a 40 nm aluminum oxide layer was applied via atomic layer deposition (ALD). Another electrode (Top electrode), with the same structure, consisting of a 5 nm chromium underlayer and 100 nm gold film, was patterned using the same method. To prevent fouling, a 40 nm aluminum oxide layer was also deposited on this electrode. The sensing area was defined by spinning a layer of photoresist onto the wafer and patterning an array of nanowells at the electrode overlap using photolithography. The next process involved multiple wet-etching steps, including the removal of the top aluminum oxide layer using a buffered oxide etchant, followed by the removal of the gold and chromium layers using specialized etchants. Then, the bottom aluminum oxide layer was etched away from within the wells. Afterward, the photoresist was removed, and a mask was used to pattern a fresh layer of photoresist, allowing selective etching of alumina in areas outside the sensing region. This process resulted in the sensor, which consists of two overlapping electrodes. A 5-mm circular PDMS fluidic cell was bonded to the nanowell array, and the connection pads were wire-bonded to complete the setup. Figure 1 and Figure 2 show the fabrication process for the nanowell impedance sensor.
Figure 1.
Overview of the fabrication of the nanowell sensor.
Figure 2.
Nanowell array layout and fabrication, in which (a) is a schematic view of the nanowell array in the overlapping part of the two electrodes; (b) illustrates a wafer including multiple sensor patterns, which is then cut into smaller pieces for PDMS and wire bonding; and (c) shows an optical image of the nanowell array.
A nanowell impedance array operates similarly to traditional impedance-based biosensors but offers enhanced precision and sensitivity at the nanoscale. It consists of a pair of electrodes separated by a 40-nanometer gap that contains an array of nanowells. Each nanowell houses immobilized probe molecules, such as antibodies, specific to the target analyte. These nanowells are precisely positioned in the overlapping region of the electrodes, ensuring there is a conductive pathway between them. When a sample containing the target biomolecules—such as proteins—is introduced, these molecules selectively bind to the probes within the nanowells. This binding alters the local dielectric environment, affecting the impedance between the electrodes. As biomolecules bind to the immobilized probes, they modify the conductivity within the nanowells, resulting in a measurable change in impedance across the electrodes. Specifically, the binding of target molecules decreases conductivity because the bound molecules partially obstruct the conductive path. This impedance change is continuously monitored in real time. The exceptional sensitivity of the nanowell impedance sensor allows for precise quantification of the target analyte. The magnitude of the impedance change is directly proportional to the concentration of the target biomolecule in the sample.
2.2. Reagent Preparation and Antibody Immobilization
Antibodies and target biomarkers were suspended in phosphate-buffered saline (PBS). All antibodies, including Human IL-6 antibody (MAB206), Human HMGB1 antibody (MAB1690), Human VEGF antibody (MAB293), Human CXCL5 antibody (MAB254), Human TNF-α antibody (AB-210), and anti-cortisol, were suspended in PBS at a concentration of 0.2 mg/mL. The antibodies were physically adsorbed onto the sensor surface by injecting the corresponding solution into the sensor well and incubating for 10 min. The target biomarkers used in this study, including interleukin 6 (IL-6, 206-IL), tumor necrosis factor-alpha (TNF-α, 210-TA), high mobility group box 1 (HMGB1), C-X-C motif chemokine ligand 5 (CXCL5), vascular endothelial growth factor (VEGF), and cortisol, were suspended at concentrations ranging from 6 nM to 6 fM until the detection limit was reached. The target cortisol hormone and anti-cortisol were purchased from Fitzgerald Industries (Cortisol-BSA, 80-1434, Acton, MA, USA), and all other target biomarkers and antibodies were commercially obtained from R&D Systems (Minneapolis, MN, USA).
2.3. Real-Time Impedance Measurements
To confine the fluid atop the sensor, we constructed a millimeter-sized PDMS fluidic cell. Reagents were manually added to the well in sequence, and the complex impedance was monitored in real time. The impedance within the nanowells was measured using a multi-frequency lock-in amplifier (Zurich Instruments HF2IS, Zurich, Switzerland). An AC excitation source supplying 0.4 V at a frequency of 1 MHz was connected to one electrode, while the second electrode was connected to the input of the lock-in amplifier (with a gain of 1 k, bandwidth of 2 Hz, and sampling rate of 225 samples/s). Both the real and imaginary components of the impedance were acquired. The percentage change in voltage was measured for each sensor in real time. Figure 3 shows an overview of the real-time measurement of target biomarkers using a lock-in amplifier.
Figure 3.
Detection of trace target biomarkers in a nanowell impedance sensor using a lock-in amplifier.
2.4. Cell Culture and In Vitro Biocompatibility
Cell cultures: RPE1, a normal retinal pigment cell line used for quality control and drug development research, was grown in DMEM/F12 media. HEPG2, a normal hepatocyte cell line, was grown in MEM media. Both media contain 10% fetal bovine serum (FBS) and 1% antibiotics (PenStrip). The cell lines were cultured in a 37 °C incubator with 5% CO2. The cells were checked for growth and morphology using a bright-field microscope to pass them before they achieved 70–80% confluence on the Petri dishes used for culturing the cells. These cell lines were obtained from the American Type Culture Collection (ATCC, Manassas, VA, USA) and tested for mycoplasma using the MycoAlert Mycoplasma Detection Kit (Lonza, Walkersville, MD, USA) before starting any experiments [45,46].
5000 cells/well in 6-well plates were seeded in 3–5 mL of complete media. 24 h later, the sensor was completely immersed in the well for different time points to check the effect on proliferation. The cells were washed twice with 1× PBS and treated with 100 µL trypsin; an additional 900 µL of fresh complete media was used to collect the cells from the well. Cell viability was then determined using the Vi-CELL™ Series Cell Viability Analyzer (Beckman Coulter, Carlsbad, CA, USA). For microscopy analysis of morphological characterization, the sensor was removed to assess the morphology of the cells plated in that well at a particular time point using bright-field microscopy (Rutgers Central Lab Services Core Facility), with 10× and 20× magnification. The cells were then checked for the proliferation assay, as mentioned above.
2.5. Nanowell Impedance Sensor Biocompatibility and Wound Fluid Collection
All animal procedures were approved by the Rutgers Institutional Animal Care and Use Committee (IACUC). Wild-type male C57BL/6 mice (approximately 9 weeks old and 25 g) were purchased from Charles River Laboratories (Wilmington, MA, USA) and acclimatized to laboratory conditions for one week. Twenty-four hours before surgery, mice were anesthetized with isoflurane (Henry Schein, Melville, NY, USA), and their dorsal fur was removed by shaving followed by application of a depilatory cream (Nair). On the day of surgery, mice were anesthetized again, and the dorsal skin was disinfected by alternating applications of Betadine scrub and 70% ethanol, repeated three times. A 1.5 × 1.5 cm full-thickness excisional wound was then created on the dorsum of each mouse and covered with a transparent, semi-permeable Tegaderm dressing to protect the wound. For the biocompatibility assessment, nanowell impedance sensors were positioned directly over the wound, with the sensing surface facing the wound bed to ensure tissue contact, and secured using Tegaderm dressing. Sensor exposure was evaluated in separate animal groups for 1, 3, 6, and 24 h (NS-1h, NS-3h, NS-6h, and NS-24h, respectively). Following the designated exposure period, the sensors were removed, and the wounds were redressed with Tegaderm. Mice were then returned to their individual cages and monitored until complete wound closure. Wound healing was assessed through serial photographic documentation. At designated time points up to 10 days post-wounding, the Tegaderm dressing was removed every alternative day, and the wound surface was rinsed with 1 mL of phosphate-buffered saline (PBS). The collected wound fluid was stored frozen for subsequent analysis. Inflammatory responses were evaluated by quantifying IL-6 and TNF-α levels using ELISA kits (Invitrogen, Thermo Fisher Scientific, Waltham, MA, USA), allowing comparison of inflammatory biomarker levels between sensor-exposed and control groups.
3. Results
3.1. Target Biomarker Detection
The experimental procedure for detecting each biomarker includes six steps. The procedure for CXCL-5 is discussed below; the procedures for the other five biomarkers are similar. The experiment starts by filling the sensor with phosphate-buffered saline (PBS) buffer. The sensor is connected to a lock-in amplifier for real-time monitoring of impedance. The experiments indicate that the output voltage significantly increases from its original value upon adding PBS to the empty sensor (Figure 4a). Correspondingly, the impedance of the nanowell array changes in the opposite direction to the output voltage, decreasing significantly at this step. Since PBS has high conductivity, the impedance in the sensor decreases greatly, which is in line with expectations. As a result, the current input to the lock-in amplifier—and consequently, the output voltage—undergoes an instantaneous change. This instantaneous change refers to the baseline shift that occurs immediately after adding reagents to the sensor. A blank PBS sample, serving as a negative control, is added to the sensor, and there is a small instantaneous change in the positive direction in the baseline (Figure 4b). The instantaneous baseline change is much smaller than what was measured in the previous step. Then, CXCL-5 human antibody is added to the well, and antibodies are immobilized on the surface of the gold electrodes (Figure 4c).
Figure 4.
Six-step measurement process using nanowell sensor for target biomarker detection and negative control comparison. The process consists of the following steps: (a) PBS is applied to the dry sensor, (b) a blank PBS sample is introduced as a negative control, and (c) 5 µL of antibody solution is added to the sensor. Next, (d) the sample is removed from the PDMS fluidic cell and replaced with fresh PBS. Afterward, (e) another blank PBS sample is introduced as a negative control. Finally, (f) the target biomarker solution is applied to the sensor.
By adding this sample—in contrast to the negative control—after the instantaneous baseline change, there is a gradual decrease in the output voltage. When the antibodies are present on the electrode surfaces, they occlude the ionic current passing between the electrodes, causing a change in resistance. The gradual change observed for the antibody corresponds to an exponential shift in the baseline following the previously discussed instantaneous change. This behavior aligns with the time-dependent nature of antigen–antibody binding. Comparing this gradual change to that of the negative control helps eliminate the possibility of nonspecific binding to the sensor surface. The gradual change in the antibody step differs from the previous step, where the negative control sample gradually increased. Saturation occurs 10 to 15 min after adding the antibody solution to the well. Next, the excess antibodies are removed from the sensor, and PBS is added to the well (Figure 4d). The subsequent step involves adding another negative control on top of the solution already in the well, gradually increasing after an instantaneous change (Figure 4e). Finally, the target biomarker—human CXCL-5 protein—is added to the fluidic cell, which leads to a gradual exponential decrease in the output voltage, the same as in the third step relevant to antibody absorption (Figure 4f). The presence of the target biomarker on the electrode surface obstructs the ionic current between the electrodes, resulting in a time-dependent change in resistance that can be measured by monitoring the gradual change. Yet, this gradual change for the negative control is in the positive direction, which is different and distinguishable from that caused by the target biomarker.
3.2. Multiplexing Performance of the Nanowell Impedance Sensor
This section demonstrates the potential of the described label-free electronic detection method employing a nanowell array for the detection of multiple key biomarkers essential for diagnosing various diseases such as cancer and inflammatory conditions. The 6 biomarkers used for this study are Interleukin 6 (IL-6), TNF-α, High mobility group box 1 (HMGB1), C-X-C motif chemokine ligand 5 (CXCL5), Vascular endothelial growth factor (VEGF), and Cortisol. IL-6 is a multifaceted, alpha-helical cytokine that undergoes phosphorylation and variable glycosylation. It plays key roles in the acute phase response, inflammation, hematopoiesis, bone metabolism, and cancer progression and serves as a biomarker for diseases including rheumatoid arthritis, sepsis, and cancer [47]. TNF-α is a cytokine involved in inflammation and immune regulation, playing a key role in processes like cell death and the inflammatory response, and can act as a biomarker for several diseases such as rheumatoid arthritis and psoriasis [48]. HMGB1 is a nuclear protein that binds DNA, enhancing the accessibility of transcription factors. It also contributes to the production of TNF, IL-6, and interferon-γ. HMGB1 has garnered significant attention as a biomarker for traumatic brain injury (TBI), neuroinflammatory disorders, epileptogenesis, and cognitive impairments [49]. CXCL5, also referred to as epithelial-cell-derived neutrophil-activating peptide (ENA-78), is a pro-inflammatory chemokine in the CXC subfamily and acts as a biomarker for diseases such as lung cancer and rheumatoid arthritis [50]. VEGF is a potent mediator of both angiogenesis and vasculogenesis in the fetus and adult, playing a key role in normal processes like wound healing and in diseases such as cancer. Circulating VEGF levels are also associated with disease activity in autoimmune conditions like rheumatoid arthritis, multiple sclerosis, and systemic lupus erythematosus [51]. Cortisol, produced by the adrenal glands, regulates metabolism, immune response, and stress by raising blood sugar levels and affecting various physiological processes, and it serves as a biomarker for both psychological and physiological stress [52], with its concentration-dependent calibration previously demonstrated by our group [53]. The experimental procedure for all biomarkers followed the same approach described in Section 3.1. Each biomarker detection experiment was conducted independently, consistently showing a gradual decrease during the antibody step and a gradual increase during the negative control step across all experiments. Figure 5a compares the detection limits of 6 biomarkers, and Figure 5b–g show the output signal of the nanowell sensor. The results show detection limits for IL-6, TNF-α, CXCL5, VEGF, and HMGB1 at 100 fM, 100 fM, 300 fM, 500 fM, and 400 fM, respectively. The gradual decrease described in Section 3.1 is illustrated in Figure 5b–g. The detection limits are also summarized in Table 1.
Figure 5.
Multiplexing performance of the nanowell impedance sensor. (a) Detection limit of 6 biomarkers at the femtomolar level. Output voltage change for different biomarkers, including (b) IL-6, (c) HMGB1, (d) VEGF, (e) TNF-α, (f) CXCL-5, and (g) cortisol.
Table 1.
Detection Limit for different biomarkers.
3.3. Sensor Characterization with Variable PBS Volumes and Frequency Measurements
Adding reagents, particularly PBS, to a dry sensor has been observed to cause an immediate shift in the baseline. This effect warrants further investigation to determine the underlying cause of the change. Consequently, an experiment was designed to examine whether adding different amounts of PBS to the dry sensor results in varying changes in the baseline. Figure 6a illustrates how the instantaneous change in output varies with different volumes of PBS added to the dry sensor. It appears that as the amount of PBS added to the well increases, the output voltage also rises. However, beyond a certain volume (20 µL), no noticeable increase in output voltage is produced by further additions of PBS, suggesting that the effect plateaus after this volume is reached. The conclusion is that the observed instantaneous change in output is likely linked to variations in the solution resistance (R_sol), indicating that the relationship between PBS volume and output voltage is influenced by changes in the solution’s resistance properties.
Figure 6.
Sensor characterization with variable PBS volumes and frequency measurements. (a) Impact of PBS volume on baseline shifts. (b) The theoretical value of the sensor’s absolute impedance over a wide range of frequencies. Output voltage in (c) negative control step at 25 MHz, (d) target biomarker step at 25 MHz, (e) negative control step at 35 MHz, and (f) target biomarker step at 35 MHz.
Subsequent characterization involved calculating the sensor’s impedance across a wide frequency range, from 1 Hz to 1 GHz in MATLAB R2020a (MathWorks, Natick, MA, USA) (Figure 6b). This was performed using a lumped model, which includes Rsol for solution resistance, Cox for the capacitance between the two electrodes, and Cdl for the double-layer capacitance.
Additionally, to validate the nanowell array’s usability across a wide range of frequencies, a comprehensive experiment was conducted at eight frequencies ranging from 1 to 40 MHz. The specific frequencies tested were 1 MHz, 10 MHz, 15 MHz, 20 MHz, 25 MHz, 30 MHz, 35 MHz, and 40 MHz, providing a thorough assessment of the array’s performance at multiple intermediate points. Frequency-dependent measurements provide insights into the overall performance and reliability of the biosensor, aiding in understanding how the device behaves under different conditions and ensuring its robustness for practical applications. These measurements are essential for optimizing the design and functionality of the protein biosensor, ensuring accurate, reliable, and efficient detection of target proteins.
Measurements for the negative control and target protein steps at 25 MHz and 35 MHz are shown in Figure 6c,f. Similar to the full experiments discussed earlier with the sensor, the same pattern was observed at different frequencies for each step. These steps include adding the negative control (blank PBS sample), the addition of the antibody, and the addition of the corresponding target protein. A gradual decrease in the signal is observed with the addition of the antibody and the target protein, while a gradual increase is seen with the addition of the negative control. This trend over time demonstrates that the array of nanowells can effectively differentiate between proteins and antibodies across a range of frequencies.
3.4. In Vitro Biocompatibility Assessment of Nanowell Impedance Sensor
The biocompatibility of the nanowell impedance sensor was checked using two different human normal cells under in vitro conditions. RPE1, a normal retinal pigment epithelial cell used as a normal cell line for quality control and drug development research, is used in this study to evaluate the toxicity, if any, induced by the sensor and whether it impacted proliferation or exerted any other visible forms of toxicity in terms of morphological changes to the cells within one week. The cell proliferation was evaluated in a six-well plate, using more than 90–95% viable cells after checking their viability using a Vi-Cell XR© cell viability analyzer (Beckman Coulter) based upon the trypan blue dye exclusion method, as shown in Figure 7a. The sensor was inserted completely into the plate well after twenty-four hours of plating the cells in a well. It was ensured that the sensor was completely covered with the growth media in a well plate. By looking at the RPE1 cell proliferation, there was no significant difference found between the cell proliferation with and without the device, as shown in Figure 7b. There was also no change in the morphology of RPE1 cells found for 5 days, as shown by the microscopic images (at 10× and 20× magnification) with and without the sensor (Figure 7c), highlighting that the sensor is biocompatible with the RPE1 cells in vitro without any induced toxicity. Importantly, the sensor was found to be working properly after use in the RPE1 cell cultures in vitro for 5 days, as shown by the voltage measurement in Figure 7d. Similarly, we chose another normal epithelial cell line, HEPG2, a normal hepatocyte cell line, to investigate the biocompatibility of this sensor. Here, we also showed that HEPG2 cell proliferation was not affected by the addition of a sensor in the cell culture, as shown in Figure 7e, and there was also no visible change in the morphology of the HEPG2 cells when analyzed using the microscope (at 10× and 20× magnification), as shown in Figure 7f. Overall, these in vitro studies highlight the biocompatibility potential of this sensor to be used for humans in the future without any visible or induced toxicity, confirmed by using the two different normal cell lines of human origin.
Figure 7.
In vitro biocompatibility assessment of nanowell impedance sensor: (a) test group (sensor-containing plates) and control group (plates containing cell growth medium without sensors) within in vitro biocompatibility experiments, (b) proliferation assay of RPE1 (normal retinal) cells with sensor, (c) morphology studies of control group and test group over time using RPE1 cell line, (d) testing the functionality of sensors after the biocompatibility experiments, (e) proliferation assay of HEPG2 cells with sensor, and (f) morphology studies of control group and test group over time using HEPG2 cell line.
3.5. In Vivo Biocompatibility Assessment of Nanowell Impedance Sensor
Following confirmation of in vitro biocompatibility, the nanowell impedance sensor was evaluated in an in vivo mouse model of full-thickness wounds. Ten mice received wounds of identical dimensions and were divided into sensor-exposed and control groups, with the sensor placed over the wound in the treatment group and no sensor applied in the control group (Figure 8a and Figure 8b, respectively). To assess the effects of exposure duration, the sensor-exposed group was further divided into four subgroups, in which the wounds were exposed to the sensor for 1, 3, 6, or 24 h (NS-1h, NS-3h, NS-6h, and NS-24h, respectively). Wound closure was monitored over three weeks and compared with the control group. Notably, all groups exhibited comparable wound healing, with complete closure occurring at approximately 21 days (Figure 8c,d).
Figure 8.
In vivo biocompatibility assessment of nanowell impedance sensors. (a) An animal in the control group where a wound is created, with no other modifications. (b) An animal in the test group where a sensor is placed inside the wound after creation. (c) Representative images of skin wounds in different groups of mice after sensor application. Scale bar = 1 cm. (d) Quantified wound images showing the fraction of wound closure as a function of time. Data are represented as mean ± SEM (n = 1–3/group). Statistical significance was determined by ANOVA followed by post-hoc Tukey’s HSD. (e,f) ELISA measurements of IL-6 and TNF-α in wound fluid collected at different time points from the same group of mice, respectively.
To further investigate potential inflammatory responses, wound fluid was collected and analyzed for IL-6 and TNF-α using ELISA. No significant differences in the levels of either inflammatory biomarker were observed between the sensor-exposed and control groups during the first 10 days (Figure 8e,f). Together, the comparable wound closure rates and absence of significant changes in inflammatory markers indicate that the nanowell impedance sensor demonstrated favorable in vivo biocompatibility under the tested conditions, supporting its potential for further development toward in vivo applications. Following the biocompatibility assessment, the sensors were subjected to performance testing and demonstrated consistent and reliable operation.
4. Conclusions
The nanowell impedance sensor consists of two overlapping electrodes separated by a 40 nm aluminum oxide layer, with a sensing region that includes a 5-by-5 array of nanowells created through multiple etching steps through the electrodes. A conductive path is formed between the two electrodes when the nanowell is filled with a solution. Functionalizing the nanowell array with antibodies allows for real-time impedance monitoring between the electrodes, revealing changes in ionic resistance as the target biomarker binds to its corresponding antibody within the nanowells. This gradual change is measured within 10 min, reflecting the time-dependent nature of antigen–antibody binding. By comparing this change to that of the negative control, the possibility of non-specific binding to the sensor surface is eliminated. The sensor structure and measurement frequency contribute to achieving high sensitivity, enabling the detection of even slight impedance changes and allowing for a lower detection limit. Multiple experiments and characterizations were conducted to demonstrate the robust capabilities of nanowell sensors, an essential step before proceeding to human measurements. A repeatable experimental procedure was used to differentiate the signal of the negative control from that of the target biomarker or corresponding antibody, revealing distinct time-dependent changes confirming the target biomarkers’ successful binding in the sensor. This was conducted using a variety of biomarkers, namely, IL-6 and TNF-α as cytokines, CXCL5 as a chemokine, and cortisol as a hormone, demonstrating the sensor’s versatility across different biomarker types. Additionally, we achieved a low femtomolar detection level for these biomarkers, a sensitivity critical for detecting diseases at early stages and enabling timely intervention and treatment. For several human diseases, such as cancer, cardiovascular disease, Alzheimer’s disease, and autoimmune disorders, a combination of multiple biomarkers provides a better understanding of disease onset and progression. This study highlights the nanowell impedance sensor’s capability to detect multiple biomarkers, enhancing prospects for early diagnosis and therapeutic interventions. Additionally, we conducted further characterizations to enhance the reliability of these sensors and to better understand their physical behavior, as demonstrated by testing with varying PBS concentrations. We also conducted measurements across eight different frequencies, successfully distinguishing between the negative control and target biomarkers. These measurements expand the potential applications of these sensors, including integration with inductive coupling wireless systems that typically operate at higher frequencies in wearable and implantable biosensors, thereby enhancing portability and enabling continuous biomarker monitoring. Additionally, before proceeding with any human measurements, it is essential to confirm the sensor’s biocompatibility. We conducted a comprehensive two-step biocompatibility assessment, beginning with in vitro experiments using multiple cell cultures. After confirming successful results in vitro, we proceeded to in vivo experiments with 10 animals. In the in vitro phase, we demonstrated that cell morphology remained unaffected in the presence of the sensor over a 7-day period, monitoring morphology at different time points and comparing it with the control group, not containing the sensor. Simultaneously, viable cell counts were conducted to assess cell proliferation, providing an additional indication of the sensor’s biocompatibility. Following the in vitro biocompatibility experiments, the sensors were tested, showing normal performance, which indicated that the cell culture medium had no effect on sensor functionality. In the in vitro triplicate experiments, initial studies were conducted with RPE1 cells, confirming the sensor’s biocompatibility. To further validate these findings, additional tests were performed using HEPG2 normal cell lines and MDA-MB-468 cancer cell lines. Following successful in vitro confirmation of the nanowell sensors’ biocompatibility, the next step involved in vivo biocompatibility experiments with live animals. Wounds were created on 10 animals, with sensors placed in contact with the wounds of seven animals, while three others served as controls without sensors. The sensors in the test group remained in contact with the wounds for intervals of 1, 3, 6, and 24 h, and wound closure was monitored over a 21-day period. Observations showed that wound closure occurred at the same rate and time in both the test and control groups. The wound healing patterns were also consistent between the two groups, as confirmed by measurements of TNF-α and IL-6 levels, with no significant difference (p < 0.05).
Author Contributions
Conceptualization, H.R. and M.J.; methodology, H.R., S.K., D.P. and G.M.R.; investigation, H.R., S.K., G.M.R. and D.P.; visualization, H.R., S.K. and G.M.R.; validation, H.R., S.K., G.M.R., Z.M. and P.X.; supervision, M.J.; writing—original draft preparation, H.R., G.M.R. and S.K.; writing—review and editing, H.R., S.K., G.M.R., Z.M., P.X., F.B. and M.J. All authors have read and agreed to the published version of the manuscript.
Funding
This research was funded by the Defense Advanced Research Projects Agency (DARPA) and the Army Research Office (ARO), grant number W911NF-20-1-0295.
Institutional Review Board Statement
The animal study protocol was approved by the Institutional Review Board of Rutgers University (protocol code: PROTO202000083 and approved on 23 June 2020).
Informed Consent Statement
Not applicable.
Data Availability Statement
The data supporting the findings of this study are included within the article.
Acknowledgments
The views here are those of the authors and do not represent the official policies of DARPA, ARO, or the U.S. government. Central lab facilities from the Rutgers Cancer Institute are also acknowledged.
Conflicts of Interest
The authors declare no conflicts of interest.
References
- Das, S.; Dey, M.K.; Devireddy, R.; Gartia, M.R. Biomarkers in cancer detection, diagnosis, and prognosis. Sensors 2023, 24, 37. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Raji, H.; Tayyab, M.; Sui, J.; Mahmoodi, S.R.; Javanmard, M. Biosensors and machine learning for enhanced detection, stratification, and classification of cells: A review. Biomed. Microdevices 2022, 24, 26. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Raji, H.; Javanmard, M.; Xie, P.; Mahmoodi, S.R. Wireless Power-Up and Readout of Label-Free Electronic Detection of Protein Biomarkers. U.S. Patent Application 18/694,664; published on 28 November 2024,
- Meng, Z.; Raji, H.; Tayyab, M.; Javanmard, M. Cell phone microscopy enabled low-cost manufacturable colorimetric urine glucose test. Biomed. Microdevices 2023, 25, 43. [Google Scholar] [CrossRef] [Scilit]
- Meng, Z.; Tayyab, M.; Lin, Z.; Raji, H.; Javanmard, M. A computer vision enhanced smart phone platform for microfluidic urine glucometry. Analyst 2024, 149, 1719–1726. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ou, F.S.; Michiels, S.; Shyr, Y.; Adjei, A.A.; Oberg, A.L. Biomarker discovery and validation: Statistical considerations. J. Thorac. Oncol. 2021, 16, 537–545. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Syahir, A.; Usui, K.; Tomizaki, K.Y.; Kajikawa, K.; Mihara, H. Label and label-free detection techniques for protein microarrays. Microarrays 2015, 4, 228–244. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ramachandran, N.; Larson, D.N.; Stark, P.R.; Hainsworth, E.; LaBaer, J. Emerging tools for real-time label-free detection of interactions on functional protein microarrays. FEBS J. 2005, 272, 5412–5425. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Nikfarjam, A.; Raji, H.; Hashemi, M.M. Label-free impedance-based detection of encapsulated single cells in droplets in low cost and transparent microfluidic chip. J. Bioeng. Res. 2019, 1, 29–37. [Google Scholar]
- Raji, H.; Xie, P.; Mahmoodi, S.R.; Javanmard, M. Wireless power-up and readout of label-free electronic detection of protein biomarkers. In Proceedings of the 25th International Conference on Miniaturized Systems for Chemistry and Life Sciences, MicroTAS 2021; Chemical and Biological Microsystems Society: San Diego, CA, USA, 2021; pp. 823–824. [Google Scholar]
- Meng, Z.; Raji, H.; Kokabi, M.; Zou, D.; Chan, J.; Liu, Q.; Zhang, R.; Javanmard, M. Microfluidic Assays for CD4 T Lymphocyte Counting: A Review. Biosensors 2025, 15, 33. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Wei, Y.J.; Zhao, Y.N.; Zhang, X.; Wei, X.; Chen, M.L.; Chen, X.W. Biochemical analysis based on optical detection integrated microfluidic chip. TrAC Trends Anal. Chem. 2023, 158, 116865. [Google Scholar] [CrossRef] [Scilit]
- Zhang, H.; Chon, C.H.; Pan, X.; Li, D. Methods for counting particles in microfluidic applications. Microfluid. Nanofluidics 2009, 7, 739–749. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Sokolov, A.N.; Roberts, M.E.; Bao, Z. Fabrication of low-cost electronic biosensors. Mater. Today 2009, 12, 12–20. [Google Scholar] [CrossRef] [Scilit]
- Petralia, S.; Castagna, M.E.; Motta, D.; Conoci, S. Miniaturized electrically actuated microfluidic system for biosensor applications. BioNanoScience 2016, 6, 139–145. [Google Scholar] [CrossRef] [Scilit]
- Ghafar-Zadeh, E.; Forouhi, S.; Azadmousavi, T. Advances in Electronic Biosensors. In Advanced CMOS Biochips: Design and Fabrication; Springer: Dordrecht, The Netherlands, 2024; pp. 197–224. [Google Scholar]
- Ye, X.; Jiang, T.; Ma, Y.; To, D.; Wang, S.; Chen, J. A portable, low-cost and high-throughput electrochemical impedance spectroscopy device for point-of-care biomarker detection. Biosens. Bioelectron. X 2023, 13, 100301. [Google Scholar] [CrossRef] [Scilit]
- Selvam, A.P.; Wangzhou, A.; Jacobs, M.; Wu, T.; Mohan, C.; Prasad, S. Development and validation of an impedance biosensor for point-of-care detection of vascular cell adhesion molecule-1 toward lupus diagnostics. Future Sci. OA 2017, 3, FSO224. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Bamshad, A.; Nikfarjam, A.; Sabour, M.H.; Raji, H. Theoretical and numerical investigation of liquid-gas interface location of capillary driven flow during the time throughout circular microchannels. In Proceedings of the 2017 5th RSI International Conference on Robotics and Mechatronics (ICRoM), Tehran, Iran, 25–27 October 2017; IEEE: New York, NY, USA, 2017; pp. 432–437. [Google Scholar]
- Raji, H.; Xie, P.; Tayyab, M.; Meng, Z.; Mahmoodi, S.R.; Javanmard, M. Wireless power-up and readout from a label-free biosensor. Biomed. Microdevices 2025, 27, 2. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Raji, H.; Xie, P.; Mahmoodi, S.R.; Javanmard, M. Wireless Power-Up and Readout from A Label-Free Sensor RFID. In Proceedings of the 2023 22nd International Conference on Solid-State Sensors, Actuators and Microsystems (Transducers), Kyoto, Japan, 25–29 June 2023; IEEE: New York, NY, USA, 2023; pp. 1077–1080. [Google Scholar]
- Teke, M.; Sayıklı, Ç.; Canbaz, Ç.; Sezgintürk, M.K. A novel biosensing system using biological receptor for analysis of vascular endothelial growth factor. Int. J. Pept. Res. Ther. 2014, 20, 221–230. [Google Scholar] [CrossRef] [Scilit]
- Bernabini, C.; Holmes, D.; Morgan, H. Micro-impedance cytometry for detection and analysis of micron-sized particles and bacteria. Lab Chip 2011, 11, 407–412. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Hashemi, M.M.; Nikfarjam, A.; Raji, H. Novel fabrication of extremely high aspect ratio and straight nanogap and array nanogap electrodes. Microsyst. Technol. 2019, 25, 541–549. [Google Scholar] [CrossRef] [Scilit]
- Mehrotra, P. Biosensors and their applications–A review. J. Oral Biol. Craniofacial Res. 2016, 6, 153–159. [Google Scholar]
- Wang, S.; Deng, R.; Chen, Z.; Huang, L.; Song, Y.; Yuan, D.; Li, Y.; Liu, H.; Yang, F.; Fan, B.; et al. High-Performance Plasma Biomarker Panel for Alzheimer’s Disease Screening Using a Femtomolar-Level Label-Free Biosensing System. ACS Nano 2023, 18, 2117–2130. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Wu, L.; Qu, X. Cancer biomarker detection: Recent achievements and challenges. Chem. Soc. Rev. 2015, 44, 2963–2997. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Bellaver, B.; Ferrari-Souza, J.P.; Uglione da Ros, L.; Carter, S.F.; Rodriguez-Vieitez, E.; Nordberg, A.; Pellerin, L.; Rosa-Neto, P.; Leffa, D.T.; Zimmer, E.R. Astrocyte biomarkers in Alzheimer disease: A systematic review and meta-analysis. Neurology 2021, 96, e2944–e2955. [Google Scholar] [PubMed]
- Ouyang, M.; Tu, D.; Tong, L.; Sarwar, M.; Bhimaraj, A.; Li, C.; Cote, G.L.; Di Carlo, D. A review of biosensor technologies for blood biomarkers toward monitoring cardiovascular diseases at the point-of-care. Biosens. Bioelectron. 2021, 171, 112621. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Hui, D.; Mo, L.; Paiva, C.E. The importance of prognostication: Impact of prognostic predictions, disclosures, awareness, and acceptance on patient outcomes. Curr. Treat. Options Oncol. 2021, 22, 12. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Gambardella, V.; Tarazona, N.; Cejalvo, J.M.; Lombardi, P.; Huerta, M.; Roselló, S.; Fleitas, T.; Roda, D.; Cervantes, A. Personalized medicine: Recent progress in cancer therapy. Cancers 2020, 12, 1009. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Capatina, D.; Feier, B.; Hosu, O.; Tertis, M.; Cristea, C. Analytical methods for the characterization and diagnosis of infection with Pseudomonas aeruginosa: A critical review. Anal. Chim. Acta 2022, 1204, 339696. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lee, M.; Seo, S. Wearable wireless biosensor technology for monitoring cattle: A review. Animals 2021, 11, 2779. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Kim, H.J.; Hirayama, H.; Kim, S.; Han, K.J.; Zhang, R.; Choi, J.W. Review of near-field wireless power and communication for biomedical applications. IEEE Access 2017, 5, 21264–21285. [Google Scholar] [CrossRef] [Scilit]
- Raji, H.; Kumar, S.; Meng, Z.; Berthiaume, F.; Javanmard, M. Wireless power-up and readout of label-free nanosensors for in-vivo monitoring of protein concentrations in live animals. Langmuir 2025, 41, 25100–25107. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Soto, R.J.; Hall, J.R.; Brown, M.D.; Taylor, J.B.; Schoenfisch, M.H. In vivo chemical sensors: Role of biocompatibility on performance and utility. Anal. Chem. 2017, 89, 276–299. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Li, W.; Zhou, J.; Xu, Y. Study of the in vitro cytotoxicity testing of medical devices. Biomed. Rep. 2015, 3, 617–620. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Wang, X.; Gu, X.; Yuan, C.; Chen, S.; Zhang, P.; Zhang, T.; Yao, J.; Chen, F.; Chen, G. Evaluation of biocompatibility of polypyrrole in vitro and in vivo. J. Biomed. Mater. Res. Part A 2004, 68, 411–422. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Karakullukcu, A.B.; Taban, E.; Ojo, O.O. Biocompatibility of biomaterials and test methods: A review. Mater. Test. 2023, 65, 545–559. [Google Scholar] [CrossRef] [Scilit]
- Christensen, M.B.; Cui, X.T.; Rieth, L.; Warren, D.J. Biocompatibility of implanted devices, modulation, and repair in the nervous system. Front. Neurosci. 2024, 18, 1505912. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Wilkinson, H.N.; Hardman, M.J. Wound healing: Cellular mechanisms and pathological outcomes. Open Biol. 2020, 10, 200223. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Gad, S.C.; Gad-McDonald, S. Biomaterials, Medical Devices, and Combination Products: Biocompatibility Testing and Safety Assessment; CRC Press: Boca Raton, FL, USA, 2015. [Google Scholar]
- Xu, R.; Fang, Y.; Zhang, Z.; Cao, Y.; Yan, Y.; Gan, L.; Xu, J.; Zhou, G. Recent advances in biodegradable and biocompatible synthetic polymers used in skin wound healing. Materials 2023, 16, 5459. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Wang, Y.; Huang, Y.; Bai, H.; Wang, G.; Hu, X.; Kumar, S.; Min, R. Biocompatible and biodegradable polymer optical fiber for biomedical application: A review. Biosensors 2021, 11, 472. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Rather, G.M.; Lin, S.Y.; Lin, H.; Banach-Petrosky, W.; Hirshfield, K.M.; Lin, C.Y.; Johnson, M.D.; Szekely, Z.; Bertino, J.R. Activated matriptase as a target to treat breast cancer with a drug conjugate. Oncotarget 2018, 9, 25983. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Rather, G.M.; Anyanwu, M.; Minko, T.; Garbuzenko, O.; Szekely, Z.; Bertino, J.R. Anti-tumor effects of a penetratin peptide targeting transcription of E2F-1, 2 and 3a is enhanced when used in combination with pemetrexed or cisplatin. Cancers 2021, 13, 972. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Hirano, T. IL-6 in inflammation, autoimmunity and cancer. Int. Immunol. 2021, 33, 127–148. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Wcisło-Dziadecka, D.; Zbiciak-Nylec, M.; Brzezińska-Wcisło, L.; Mazurek, U. TNF-α in a molecularly targeted therapy of psoriasis and psoriatic arthritis. Postgrad. Med. J. 2016, 92, 172–178. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Paudel, Y.N.; Shaikh, M.F.; Chakraborti, A.; Kumari, Y.; Aledo-Serrano, Á.; Aleksovska, K.; Alvim, M.K.M.; Othman, I. HMGB1: A common biomarker and potential target for TBI, neuroinflammation, epilepsy, and cognitive dysfunction. Front. Neurosci. 2018, 12, 628. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Cheng, Y.; Ma, X.L.; Wei, Y.Q.; Wei, X.W. Potential roles and targeted therapy of the CXCLs/CXCR2 axis in cancer and inflammatory diseases. Biochim. Biophys. Acta (BBA) Rev. Cancer 2019, 1871, 289–312. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Carvalho, J.F.; Blank, M.; Shoenfeld, Y. Vascular endothelial growth factor (VEGF) in autoimmune diseases. J. Clin. Immunol. 2007, 27, 246–256. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Miller, N.; Asali, A.A.; Agassi-Zaitler, M.; Neumark, E.; Eisenberg, M.M.; Hadi, E.; Elbaz, M.; Pasternak, Y.; Fishman, A.; Biron-Shental, T. Physiological and psychological stress responses to labor and delivery as expressed by salivary cortisol: A prospective study. Am. J. Obstet. Gynecol. 2019, 221, 351.e1–351.e7. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Mahmoodi, S.R.; Xie, P.; Zachs, D.P.; Peterson, E.J.; Graham, R.S.; Kaiser, C.R.W.; Lim, H.H.; Allen, M.G.; Javanmard, M. Single-step label-free nanowell immunoassay accurately quantifies serum stress hormones within minutes. Sci. Adv. 2021, 7, eabf4401. [Google Scholar] [CrossRef] [Scilit] [PubMed]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.