Micro- and Nanoplastics in Maintenance Hemodialysis

,
,
,
,
,
,
,
and
1
Unit of Nephrology and Dialysis, Department of Clinical and Experimental Medicine, University of Messina, 98125 Messina, Italy
2
Department of Nephrology and Dialysis, G. Brotzu Hospital, 09121 Cagliari, Italy
3
PhD Program in Molecular and Clinical Medicine, University of Palermo, 90100 Palermo, Italy
4
Unit of Geriatric Medicine, Italian National Research Center on Aging (IRCCS INRCA), 87100 Cosenza, Italy

Abstract

This narrative review examines the evidence on renal toxicology, dialysis fluids, extracorporeal circuits, and membrane transport as it bears on micro- and nanoplastics exposure during maintenance dialysis. Each week, treatment brings blood into contact with 300 to 600 L of treated water and dialysate, with the polymeric surfaces of the extracorporeal circuit and, in online hemodiafiltration, with substitution fluid prepared online and infused into the circulation. Studies from the 1980s documented material shedding from silicone pump tubing, and microplastics have since been identified in dialysis-related fluids. Human particle biomonitoring in dialysis populations rests on two studies: nine incident patients sampled before and after a single session, with circulating counts of 30 to 100 particles/mL at a 20-micrometer detection floor, and 92 participants compared cross-sectionally, 31 of them on hemodialysis, with median counts near 10,000 particles/mL at a 0.1-micrometer floor. Both datasets are limited in size, and their figures differ by one to two orders of magnitude because size window, identification specificity and sample preparation differ between them, so neither can be set against the other, and neither establishes what dialysis itself contributes. Only one tubing study has resolved shed fragments below one micrometer, leaving the materials now in use poorly characterized at submicrometer resolution. In rodent and cell models, micro- and nanoplastics exposure produces oxidative stress, inflammation, and interstitial fibrosis in the kidney, at administered doses with no validated conversion to human internal dose. Sources were selected by editorial judgment without prespecified eligibility criteria, and no formal risk-of-bias appraisal was applied. Contamination-controlled, source-resolved mass-balance measurement would establish the contribution of dialysis to systemic micro- and nanoplastics burden.

1. Introduction

Global plastic production rose from two million tonnes in 1950 to 475 million tonnes in 2022, and baseline projections put it at 1200 million tonnes by 2060, with some 8000 million tonnes of plastic waste already accumulated in the environment and with less than a tenth of production recycled [1,2]. These polymers degrade slowly and progressively fragment into ever-smaller particles [1,3,4]. The Lancet Countdown considers the health impacts of plastics across their entire life cycle, from fossil-fuel extraction to production to use, resulting in more than 1.5 trillion United States dollars a year in health-related economic losses [2]. The Minderoo-Monaco Commission that preceded it catalogued thousands of chemical additives associated with plastic materials, including phthalates, bisphenols, per- and polyfluoroalkyl substances (PFAS), and brominated flame retardants, many of which are toxic [5]. For consistency, this review defines microplastics as particles between 1 µm and 5 mm, and nanoplastics as particles smaller than 1 µm; both definitions cover particles manufactured at that size and particles produced by the fragmentation of larger items [6]. Unless otherwise specified, the term particles denotes micro- and nanoplastic particles; particulate matter counted without polymer identification, and material of other composition such as cellulose fibers, pigments or silicone, carry an explicit label wherever they appear. Particles have been recovered from drinking water, table salt, seafood, household tap water, indoor and street air, and cosmetics, sources that reach humans by ingestion, inhalation, or dermal contact [5,7,8,9,10,11,12,13,14,15]. Presence in these sources indicates external contact; absorption across the intestinal, pulmonary, and cutaneous barriers is a separate step that these data do not measure, and a particle that is ingested or inhaled may transit the body without being taken up. Published estimates of human ingestion vary by several orders of magnitude because they depend on analytical methods and exposure assumptions, so quantitative estimates of intake remain uncertain, and estimates of uptake more so [9].
Particles entering the circulation may distribute to highly perfused organs, although biodistribution depends on particle characteristics and biological clearance. In whole blood from 22 healthy donors, polymers were quantified above the limit of quantification in 17, at a mean summed concentration of 1.6 µg/mL for particles at or above 700 nm [16]. Pyrolysis gas chromatography-mass spectrometry has identified and quantified polymer mass from human kidney, liver, and brain, with polyethylene predominating; the method fixes polymer identity and quantity but leaves particle number, size, and morphology undetermined [17].

Methodological Note

This article is a narrative review with a documented search strategy. It was neither conducted nor registered as a systematic review; no formal appraisal of risk of bias was applied, and the selection of sources reflects editorial judgment, with no prespecified eligibility criteria. PubMed/MEDLINE, Embase, Scopus and Web of Science were searched from inception to 15 August 2026 for English-language articles on micro- and nanoplastics in kidney disease and in extracorporeal treatment, and the bibliographies of the retrieved articles were screened together with the standards governing dialysis fluid quality. Preference went to human biomonitoring studies with spectroscopic or thermoanalytical polymer identification and declared contamination controls, and to controlled measurements of dialysis fluids, treatment water and circuit components; animal and cell work is cited for mechanism, its administered doses having no established equivalence to human internal exposure. Statements that a measurement was not found describe the result of these searches at the date of submission. Supplementary Table S1 records, for each human study drawn on, the sample size, the polymer identification method, the size floor and the contamination controls and blank reporting declared by the investigators; no formal quality score was derived from it. Polymers, particle sizes, analytical platforms, and reported units vary too widely across this literature for a quantitative synthesis, and the review is organized thematically.

2. Detection in Human Tissues and Biological Matrices: Methods, Size Windows, and Analytical Limitations

2.1. Analytical Methods and the Boundary Between Microplastic and Nanoplastic Measurements

The techniques used in human studies do not cover the same size range, and the distinction between microplastic and nanoplastic measurements runs through everything that follows. Micro-Raman spectroscopy identifies individual particles down to about 0.5 to 5 μm, micro-Fourier-transform infrared spectroscopy to 10 to 20 μm, and laser direct infrared imaging to about 20 μm, so these spectroscopic methods count and size microplastics and do not reach the nanoplastic range as defined here [18]. Scanning electron microscopy with energy-dispersive X-ray analysis extends counting and sizing to a stated floor of 0.1 μm, but it resolves elemental composition and does not identify a polymer [19]. Pyrolysis gas chromatography-mass spectrometry is the one technique in this literature that captures polymer below one micrometer, because it digests the sample and quantifies polymer mass with a lower size limit dictated by the pore size of the filter used. The lower size limit is then set by sample preparation: 700 nm in the blood study that used a filtration step [16], and 300 nm in a urine study that combined thermal desorption with pyrolysis [20]. The method returns neither particle number nor size distribution [17,18]. In the human studies assembled here, particle-resolved data therefore concern microplastics from about 1 µm upwards [19,21,22,23], and the nanoplastic fraction appears only as unsized polymer mass in blood, plaque, and tissue [16,17,24]. Any nanoplastic fraction named in these samples is inferred from that mass and was not measured as such.

2.2. Human Tissues and Biological Matrices

Micro-Raman spectroscopy of ten kidney specimens, taken from the healthy poles of tumor nephrectomies, and of ten urine samples from healthy donors identified 66 candidate microparticles, 23 from urine and 43 from kidney. Of these, 26 could be characterized as particles ranging from 1 to 29 μm in size, with polyethylene and polystyrene predominating, along with several pigments, a measurement that lies in the microplastic range throughout [21]. Urinary and kidney tissue particles measured, respectively, between 3 and 13 and between 1 and 29 μm. The size of these particles makes direct filtration across an intact glomerular barrier unlikely, and the route by which they reached renal tissue remains uncertain. The urine and kidney specimens were obtained from different individuals, so the two matrices do not represent a single particle balance. The authors described this as the first spectroscopic demonstration of particles in the renal parenchyma, although the small, single-center sample limits the generalizability of these findings [21]. Particles had been recovered from human urine before, their content varying with environment and geography [25,26]. Transplant material widened the human renal evidence: in a multimodal study of tissue from 28 deceased kidney donors, pyrolysis gas chromatography-mass spectrometry was combined with laser direct infrared imaging for counting and typing and scanning electron microscopy for morphology [22]. Particles were recovered from every tissue examined, kidney, adrenal gland, renal artery, renal vein, and ureter alike, with polyethylene and polyvinyl chloride dominant by mass, at 25.7 to 98.9 and 31.2 to 65.4 µg/g across tissues; renal tissue carried six polymer types, most particles measuring between 20 and 500 μm [22]. Particles at the upper end of that range are difficult to reconcile with delivery via the renal circulation, and the report provides no procedural blanks against which contamination during retrieval and processing could be ruled out [22]. Correlations between particle type and early recipient blood pressure were nominal and did not survive correction for multiple comparisons, so they are hypothesis-generating only, and a cross-sectional design measuring blood pressure at 15 days says nothing about longer-term outcomes [22]. Presence in human renal tissue has been reported in a small number of studies, none of which has established the route of entry or harm [27,28,29].
A scoping review examined human internal exposure across several organ systems, including the cardiovascular, digestive, endocrine, integumentary, lymphatic, respiratory, reproductive, and urinary systems [30]. Particles have been found in the placenta, first in the study that coined the term “plasticenta”, and then in meconium, infant feces, breast milk, and infant formula, which places exposure before birth [31,32]. Their detection in semen, the reported associations with semen quality, and a systematic review of reproductive outcomes indicate that particles are also present in the adult reproductive tract [33,34]. The size detection limits and the specificity of polymer identification differ between these matrices, so the counts reported in one do not convert into those reported in another [30].

2.3. Contamination Control, Blanks, and Interlaboratory Comparability

Micro- and nanoplastic particles in biological matrices occur at low concentrations, vary widely in size and shape, and carry no chemical signature that distinguishes a particle of environmental origin from one introduced during sampling or analysis. No tolerable human threshold has been set [29,35]. The dialysis environment is itself composed of the polymers under study, so tubing, connectors, fluid bags, and the membrane can all contaminate any sample drawn to measure particle content, and a positive finding may originate in the measurement process itself. In high-profile detection studies, the adequacy of blank controls and anti-contamination procedures is itself disputed, even more so when blood is sampled through plastic access and processed with plastic labware [36,37]. A Matters Arising section in the journal Nature Medicine raises the same objection to the pyrolysis gas chromatography-mass spectrometry measurements of polymer mass in human liver, kidney, and brain [17], citing limited contamination controls and the absence of validation steps as grounds for doubting the reported concentrations [38]. The original authors have replied, defending their digestion protocol and quality-control steps, so the methodological dispute remains open [39].
An interlaboratory comparison was conducted under the VAMAS prestandardization platform. In this work, 84 laboratories registered and 56 returned results using ISO-approved methods on pressed tablets of polyethylene and polyethylene terephthalate powders of 10 to 200 μm in a water-soluble matrix. The results put the identification limit of micro-Raman spectroscopy at 0.5 to 5 μm and that of micro-Fourier-transform infrared spectroscopy at 10 to 20 μm. The reliable detection limit of infrared methods, under the conditions tested, was generally above 20 μm [18]. The European monitoring requirement adopted for drinking water applies to particles above 20 μm, and thermoanalytical methods such as pyrolysis gas chromatography-mass spectrometry return polymer mass and identity without particle number, so a concentration from that route provides neither a count nor a size distribution [18]. Each method therefore imposes its own lower bound on the size distribution that a study can report. The organizers of that exercise judge microplastic analysis to be at an early stage of development and place full standardization roughly ten years away [18]. Biological matrices should take longer still, since digestion steps and organic background add sources of variance that the tablet matrix did not contain. Laboratory blanks pose a related problem. A reagent or procedural blank that returns no particles excludes contamination in the laboratory and says nothing about contamination introduced at sampling, during transport or in the clinical environment. Until the field standardizes across laboratories, particle counts and concentrations should not be compared from study to study without caution (Supplementary Table S1).
Publication bias further restricts what can be inferred from the assembled literature. Selective publication may favor studies reporting particle detection over null findings, and the studies that do appear come with detection limits and preparation protocols that shift from paper to paper, so the assembled literature resists pooling.

2.4. Comparability of Human Measurements Across Studies

The human blood measurements show the above concretely, since the reported figures do not lie on a common scale. Pyrolysis GC-MS put summed polymer mass in healthy donors at 1.6 μg/mL, for particles at or above 700 nm [16]. In the cross-sectional survey polyethylene alone reached 21.3 μg/mL, at a limit of detection, 4.24 μg/mL, that itself exceeds the earlier figure [19]. Counting gives the same picture: 30 to 100 particles/mL with a floor at 20 μm and a polymer spectrum confirmed for each particle [23], against medians near 10,000 with a stated floor of 0.1 μm, no filtration step, and identification by elemental composition in place of polymer identity [19]. A hundredfold separates the two counts, and more than tenfold the two masses, in matrices nominally the same. The divergence between the counts follows from the size window each method opens and from how specifically each identifies what it counts: a method that counts every carbonaceous object above 0.1 μm will return more objects than one that confirms a polymer spectrum above 20 μm. The divergence between the masses has sources of its own. A bulk polymer mass depends on extraction recovery, on calibration against reference polymers, on interference from the biological matrix, on which polymers the method was calibrated to cover, and on the background contamination subtracted or left in, and each of these moves the figure in a direction that is known in sign and unknown in size. A case for or against a raised burden therefore cannot be built by comparing studies and will require exposed and unexposed material to pass through the same laboratory under the same protocol and size floor. A related distinction applies within a single study. Energy-dispersive X-ray analysis resolves elemental composition and does not establish polymer identity, so counts from that route and polymer identities from pyrolysis on separate aliquots measure different objects, and the second does not validate the first [19]. Fluorescent staining supplied counts without spectroscopic confirmation in the apheresis series for the same reason [40].
These limits cut directly across the size range of mechanistic relevance. The 1 to 29 µm span recovered from kidney tissue by micro-Raman [21] begins near that technique’s own floor; the fibers recovered from dialysis fluids by infrared spectroscopy [41] lie above the infrared threshold; the summed polymer concentration measured in blood [16] is a mass with no particle count. The size range below the 5.5-nanometer hydrodynamic diameter associated with efficient renal clearance in the engineered-particle model [42] has not been addressed by the analytical approaches used in the human studies reviewed here. Differences among published studies may therefore reflect analytical methodology as much as biological variation.
In Table 1, each row pairs a limitation that recurs across the literature reviewed with the consequence it carries for interpretation. The references identify studies in which the limitation is documented, and they are not an exhaustive list of the studies it affects. Each entry describes the evidence as a body and does not appraise individual studies, whose design, polymer identification method, size floor, contamination controls, and blank reporting are recorded in Supplementary Table S1. Where one of these limitations bears on a result, the text refers to this table instead of restating it.
Table 1. Cross-cutting limitations of the evidence reviewed.

3. Toxicokinetics and the Hypothesis of Impaired Renal Elimination

No metabolic pathway has been established for most conventional polymer particles, even though the chemicals they carry are biotransformed. The fate of the polymer itself is therefore largely determined by distribution and elimination [49], and absorption, distribution, biotransformation, and elimination can be treated separately. The kidney receives close to a fifth of the cardiac output and concentrates the glomerular filtrate along its tubules. One possible framework is that particles reaching the kidney undergo filtration where size permits, tubular interaction, urinary excretion, and possible tissue retention [29]. A frequently cited size-dependent benchmark comes from work on engineered nanoparticles. Quantum dots injected intravenously into rodents required a final hydrodynamic diameter below 5.5 nm for rapid, efficient urinary excretion, and coatings that allowed serum protein adsorption increased the diameter to more than 15 nm, abolishing renal excretion [42]. That figure is three orders of magnitude below the 1–29 µm range recovered from human kidney tissue [21]. The size classes that recur through this literature do not constitute a single biological entity. Intact glomerular passage has been demonstrated for engineered nanocrystals with hydrodynamic diameters below 5.5 nm [42], but no equivalent measurement has been reported for polymeric particles; micrometer-sized particles found in kidney tissue would require an explanation other than direct intact glomerular filtration, assuming that particle identification and localization are valid [21]. Sizes larger again, up to the millimeter-scale fibers recovered from dialysis fluids [41], raise the separate question of what a dialysis membrane can pass [50]. Tubular uptake is supported by more consistent evidence: proximal tubular epithelial cells are the primary endocytic compartment of the nephron and take up particles via clathrin- and caveolin-mediated endocytosis, with larger particles entering through macropinocytosis or phagocytosis [51,52,53]. Human embryonic kidney and proximal tubular cell lines internalize submicron particles in a size-dependent way, and microfluidic kidney models show uptake with downstream signaling perturbation [53,54,55].
The protein corona that a particle acquires in the blood governs its interaction with cells. On silica and polystyrene nanoparticles, this layer assembles in under half a minute and binds close to three hundred plasma proteins, including complement components. Its qualitative composition remains stable even as the bound quantity continues to increase, and by that point it already conditions hemolysis, thrombocyte activation, cellular uptake, and endothelial cell death [56]. Because that composition follows from the surrounding plasma, the uremic milieu of a dialysis patient, often low in albumin and loaded with retention solutes, may build a corona unlike the one formed in healthy plasma. The consequences for particle handling have not been assessed in any study identified here. The proximal tubule, with its high metabolic demand and dense mitochondrial content, is particularly vulnerable to injury from the particles it internalizes [57,58]. Potential exposure during dialysis introduces additional medical contact pathways, such as exposure via extracorporeal circuits and infused fluids, the contributions of which to systemic absorption are unknown. Biotransformation, as far as it has been studied, acts on the sorbed and leached chemicals through receptor-regulated phase I and phase II pathways [59,60]. The polymer backbone is treated here as persistent because no metabolic clearance pathway has been established for it in the literature reviewed, a position that reflects the lack of such a pathway and leaves surface oxidation, fragmentation or enzymatic alteration of the polymer in vivo an open question. Human measurements have been reported in only a limited number of matrices using different analytical approaches, precluding direct comparison across studies. Summed polymer mass has been quantified at 1.6 µg/mL in blood [16] and at 21.7 µg/mg in carotid plaque [24], while kidney tissue has been characterized by particle count [21,22]. No measurement of the rate at which any of it clears from the body was identified.
End-stage kidney disease sharply reduces glomerular filtration, and in anuric patients, urine production ceases. As particles have been recovered from human urine, urine is a candidate disposal route. Loss of urinary excretion could therefore reduce one potential elimination pathway, provided that urinary clearance contributes meaningfully to whole-body elimination and is not offset by other routes. Neither condition has been examined in this population [21,35]. Reviews have proposed that failing kidneys clear fewer particles and that hemodialysis patients, in whom urine output is absent, may accumulate them systemically, a claim currently supported only by the recovery of particles from urine [35,61]. The proposition is biologically plausible; in the studies assembled here, particle balance has not been measured in the same patients across the loss of renal function, so the hypothesis rests on plausibility alone.
The only human comparison of urinary particle counts identified for this review set patients with stage 3 to 4 chronic kidney disease against hospital patients with no history of it, the hemodialysis patients enrolled alongside them having passed no urine and contributed blood alone [19]. Counts were higher in the patients, at medians of 1231 against 411 particles/mL, and the particles recovered from their urine were smaller, at median diameters of 4.0 against 5.9 μm; the investigators read the first as some degree of urinary elimination preserved despite a reduced glomerular filtration rate and the second as compatible with selective retention of the larger particles within the circulation or renal tissue. The blood-side counterpart of that comparison points the same way: particles circulating in the patients were marginally larger than in the comparison subjects, at medians of 4.5 against 3.9 μm, so the size argument draws on two compartments, not on urine alone; the differences are of the order of a micrometer, however, on interquartile ranges that overlap throughout [19]. The study measures a concentration in spot urine, and neither of the two conditions above concerns a concentration. Counts were reported per mL of spot urine, with no normalization to urinary creatinine or osmolality, although the same study normalized bisphenol A for dilution, and without urine volume, collection interval, or a paired plasma concentration, none of which the design collected; the interquartile range reached zero in both groups, so a substantial share of participants excreted no detectable particles. Urinary bisphenol A in the same samples ran about six times lower in the patients, which the investigators attribute to impaired renal elimination under comparable exposure, so the two analytes moved in opposite directions under the same dilution and the same reduction in filtration [19]. The transition on which the hypothesis turns, from residual diuresis to anuria, was not sampled. The analytical design qualifies the inference in two further ways: particles were counted and sized by scanning electron microscopy with energy-dispersive X-ray analysis, a method that determines elemental composition and so separates carbonaceous material from mineral material, but it does not identify a polymer. Polymer identity instead came from the pyrolysis of separate aliquots, so the counts and the polymer profile are not measurements of the same objects [19]. No estimated glomerular filtration rate, serum creatinine, or stage of chronic kidney disease was recorded, so the reduction in filtration on which the interpretation turns is not quantified in the participants in whom it is invoked [19].
Residual kidney function is an established determinant of survival in peritoneal dialysis. In a reanalysis of 601 patients of the CANUSA cohort, each 250 mL of daily urine volume was associated with a 36% lower relative risk of death (relative risk 0.64, 95% CI 0.51 to 0.80), whereas peritoneal creatinine clearance showed no association with survival [62,63], and the rate at which residual function declines predicted death and technique failure in a later cohort [64]. These observations concern peritoneal dialysis and do not concern particle handling; they emphasize the physiological importance of preserved urinary excretion. The contribution of residual diuresis to the systemic elimination of micro- and nanoplastics remains unknown.

4. Experimental Nephrotoxicity

Interpreting experimental nephrotoxicity is challenging due to substantial heterogeneity in polymer type, particle size, shape, surface chemistry, administered dose, route of administration, and duration, all of which can significantly influence observed outcomes.

4.1. Histology

Experimental rodent and avian studies have consistently reported structural renal alterations, cataloged across in vivo designs in a systematic review [65]. Mice given polystyrene particles from 80 nm up to several μm developed glomerular basement membrane thickening, podocyte injury with altered WT-1 expression, and fusion of podocyte foot processes [66]. Chronic exposure led to mesangial matrix expansion and endocapillary proliferation; at higher doses, the glomerular capillaries became congested, and the tuft became hypotrophic [66,67]. Tubular lesions appeared across several designs: tubular dilatation, loss of brush borders, a flattened epithelium, and dose-dependent degeneration of proximal and distal tubules [68,69]. Interstitial fibrosis, with collagen deposition and raised alpha-smooth muscle actin, was the common endpoint across these designs [27,66].

4.2. Oxidative and Endoplasmic Reticulum Stress

Juvenile rats given 1000-nanometer polystyrene microplastics by mouth at 2.0 mg/kg/day for 28 days grew more slowly, and the organ index of kidney, heart, and ovary declined in parallel. Renal histology showed lesions; blood urea nitrogen and creatinine levels rose, as did interleukin-1 beta, interleukin-6, and tumor necrosis factor alpha. Renal cell apoptosis increased on terminal deoxynucleotidyl transferase dUTP nick end labeling. The authors interpreted these findings as consistent with oxidative stress driving endoplasmic reticulum stress and inflammation [43]. Similar oxidative and mitochondrial disruptions, with increased reactive oxygen species, mitochondrial depolarization, and cell-cycle arrest, were observed in human proximal tubular cells, human embryonic kidney cells, and fish kidney cells [55,68,70]. One study combined human proximal tubular cells, which took up polystyrene microplastics at concentrations of 0.05 to 0.8 mg/mL, with oral gavage in male C57BL/6 mice [68]. The cells showed mitochondrial reactive oxygen species, endoplasmic reticulum stress, inflammation, and autophagy; the animals accumulated particles in the kidney and developed histopathological lesions with raised endoplasmic reticulum stress, inflammatory, and autophagy-related proteins, the last marked by higher LC3 and beclin 1 and by altered signaling through mitogen-activated protein kinase and the protein kinase B and mammalian target of rapamycin axis [68]. Endoplasmic reticulum stress and the unfolded protein response are implicated in kidney disease progression, sharing the maladaptive signaling and disordered autophagy documented in chronic kidney disease, diabetic nephropathy, and the fibrotic response [71].

4.3. Inflammation and Regulated Cell Death

Renal tissue from exposed animals showed activation of nuclear factor kappa B and increased secretion of tumor necrosis factor alpha and interleukins 1 beta and 6, with macrophages infiltrating the tubulointerstitium [66,67]. Experimental studies have implicated several regulated cell death pathways, among them necroptosis through the RIP1-RIP3-MLKL axis in avian kidney and ferroptosis driven by iron-dependent lipid peroxidation [61,72]. Part of this injury reaches the kidney by indirect routes: when mice were given polystyrene microplastics, a breached intestinal barrier raised urinary levels of the complement product C5a and increased renal C5aR expression, accompanied by changes resembling aspects of chronic kidney disease. Sealing the barrier with an antibiotic combination cleared both the injury and the complement changes, and the particles themselves had little direct effect on renal cells or on complement [73].
Polypropylene exposure has also been shown to alter the renal lipidomic profile [74]. Particle-driven reactive oxygen species, lysosomal disruption, and release of damage-associated molecular patterns contribute to the engagement of the NLRP3 inflammasome, which matures interleukin-1 beta and interleukin-18 and drives pyroptosis in acute kidney injury, diabetic kidney disease, and chronic renal fibrosis [75]. That fit is inferential, since activation of renal NLRP3 in situ by micro- and nanoplastics was not demonstrated in the work reviewed here. Human embryonic kidney cells showed the opposite result: at a cytotoxic concentration of polystyrene, the autophagy that appeared was accompanied by reduced NLRP3 expression and a cytokine profile that the authors interpreted as dampened inflammation [76].

4.4. Functional Biomarkers

Some designs report raised serum creatinine and blood urea nitrogen; others report lower serum creatinine alongside albuminuria and raised creatinine clearance, a pattern interpreted as renal hyperfiltration [27,43,73,76,77], so the functional readouts diverge. Male C57BL/6 mice given 1-micrometer polystyrene microplastics in drinking water at 0.1 or 1 mg/L for eight weeks, concentrations described by the investigators as environmentally relevant, showed no change in body weight or kidney coefficient, lower blood urea nitrogen at both doses, and unchanged serum creatinine and uric acid. Glomerular tuft area rose 13% at the lower dose and 19% at the higher, with collagen on Masson staining unchanged. Renal cortical transcriptomics returned several hundred differentially expressed genes at each dose, enriched in mitochondrial terms and in the oxidative phosphorylation pathway [44]. The investigators interpreted this as compensatory glomerular hypertrophy with enhanced filtration, at exposures still short of exhausting renal reserve, and attributed the contrary reports of raised urea and creatinine to the higher doses used elsewhere [44]. The direction of the functional response may therefore depend on dose, a possibility that no single dose–response experiment has yet tested.
Animal and human studies report exposure in units that do not convert into one another. The animal work gives an administered dose, in mg per kilogram per day or mg per liter of drinking water, and never the internal renal concentration reached [43,44]. The human measurements provide the following concentrations: 1.6 µg/mL of summed polymer in blood for particles at or above 700 nm [16] and 21.7 µg/mg of plaque in carotid atheroma [24]. Cellular designs sit at neither pole; some dose polymer at mg per mL [59,68] and others report effects at substantially lower nominal concentrations: round polystyrene microspheres of 3.54 μm impaired barrier integrity in human embryonic kidney cells at 300 ng/mL by depleting zonula occludens-2 and alpha-1-antitrypsin, while 3 ng/mL stayed non-cytotoxic and still shifted cytokine release [76]. A nominal concentration in culture medium is neither the dose delivered to cells nor a particle count, and it differs from a summed polymer mass in blood in terms of matrix, polymer, and size distributions. No validated conversion links an external or administered dose to the burden reached in human kidney tissue, so the exposures that injure these models cannot be placed either inside or outside the human range. Detection in human tissue documents exposure and does not by itself establish toxicity at the measured concentrations, and the mechanisms identified experimentally have been shown only at administered doses whose relation to human internal dose is unknown (Table 1).

5. Interactions Between Microplastics and Plastic-Associated Chemicals

Micro- and nanoplastics can adsorb organic pollutants and metals onto their surface, raising the possibility that particles may function as carriers of co-contaminants [78,79]. The chemical load a particle carries falls into two classes. Firstly, additives incorporated at manufacture, among them phthalate plasticizers, bisphenols, brominated flame retardants, antioxidants, ultraviolet stabilizers and per- and polyfluoroalkyl substances, leach from the polymer [5]. Secondly, hydrophobic organic contaminants that are sorbed from the environment onto the particle surface, among them polycyclic aromatic hydrocarbons, together with metals, travel with it [78,79]. Several compounds in both classes act as endocrine disruptors in the dialysis setting [80]. Sequential thermal desorption and pyrolysis in a single gas chromatography-mass spectrometry run, the so-called double-shot configuration, releases the additives and sorbed chemicals in a first heating step and pyrolyzes the polymer in a second, so that the organic load and the polymer identity are read from the same aliquot [81,82]; applied to human urine with a 300-nanometer preparation floor, it has quantified four polymers [20]. That configuration has not reached dialysis fluids, circuit components, or the blood of dialysis patients in the studies reviewed here, and it would give the particle-associated chemical load a direct measurement that co-exposure experiments cannot supply. The extent to which this “vector effect” contributes to internal exposure remains uncertain. The co-exposure experiments test the interaction between the particle and the chemical and leave open the contribution of the particulate route to total chemical exposure. The most direct measurements come from a renal co-exposure model. Human proximal tubular cells were exposed for 5 to 24 h to polyethylene microspheres of 1 to 4 μm at 0.2 mg/mL, to bisphenol A at 100 nmol/L, and to both together; particle uptake followed time-dependent kinetics, with 4.4% of cells holding microspheres at 5 h and 24.8% at 24 h [59]. Bisphenol A alone left viability intact; particles alone and the combination each reduced it by roughly 20%, so no interaction was evident for cell viability [59]. The interaction surfaced instead in the oxidative and inflammatory readouts, where co-exposure produced larger increases in oxidative stress and inflammatory markers than either exposure alone, with malondialdehyde rising to 2.6-fold and the chemokine receptors CCR-2 and CCR-5 increasing to between 6.7- and 8.4-fold above control [59]. The bisphenol A concentration was set to average human environmental exposure and the particle concentration to earlier experimental work, so the design tests interaction while leaving environmental realism open [59]. In intestinal and pulmonary cell models, phthalate esters combined with bisphenol A exceeded the effect of either alone, on top of the physical and chemical toxicity of irregular microfragments; those designs, though, ran no formal test of interaction or departure from additivity, so the combined effect appears here with no claim of synergy [83,84,85]. Plastic additives, among them phthalates, brominated flame retardants, alkylphenols, and per- and polyfluoroalkyl substances, inhibit solute-carrier uptake transporters and ATP-binding-cassette efflux pumps in vitro, though the review that gathered these findings warned that the low additive concentrations reached through environmental or dietary exposure limit their in vivo relevance [86]. Most human exposure to bisphenols, phthalates, and PFAS arises from multiple environmental and dietary sources independent of microplastic particles. Co-exposure studies therefore address biological interaction and leave the quantitative contribution of particles to total chemical dose unmeasured.
Human epidemiological evidence is strongest for PFAS, and these compounds are considered here because of their association with plastic manufacture; their transport by micro- or nanoplastics in vivo has not been shown. Analyses of the National Health and Nutrition Examination Survey from 2003 to 2018 found that serum concentrations of four compounds tracked inversely with estimated glomerular filtration rate beyond a fitted cut-off, which the investigators interpreted as arguing against reverse causation [87]. Follow-up of the Diabetes Prevention Program trial provided the prospective counterpart: baseline plasma concentrations were inversely related to filtration rate over 14 years, with the lifestyle intervention leaving the associations unchanged and participants hypertensive at baseline showing the stronger relation [88]. Data from pediatric populations point the other way: among 9660 children living near a chemical plant, an interquartile rise in measured perfluorooctanoic acid accompanied a fall in estimated glomerular filtration rate of 0.75 mL/min/1.73 m2 (95% CI −1.41 to −0.10; p = 0.02). The concentration predicted at enrollment from residential histories through a validated pharmacokinetic model, less vulnerable to reverse causation, showed no relation (−0.10; 95% CI −0.80 to 0.60; p = 0.78), so the authors read the cross-sectional association as a consequence of reduced kidney function [89]. These studies support association without establishing causation, and the conclusion reached depends on which exposure metric is used.

6. Systemic and Cardiovascular Implications

6.1. Human Cardiovascular Observations

In carotid endarterectomy patients, polyethylene was detected in the plaque of 150 of 257 who completed follow-up, at a mean of 21.7 µg/mg of plaque, and those with detectable particles in the atheroma ran a higher risk of the composite of myocardial infarction, stroke, or death from any cause over a mean 34 months (hazard ratio 4.53, 95% CI 2.00 to 10.27) [24]. The study was observational and cannot prove causation. The findings remain the subject of methodological debate, particularly regarding analytical contamination control and reproducibility [36,37]. Plaque-associated polymers could represent biologically active particles, markers of cumulative exposure, or analytical artifacts, and the available data do not distinguish between these possibilities. In one rodent model, circulating particles taken up by immune cells obstructed cerebral capillaries, with neurobehavioral sequelae reported alongside [90].

6.2. Aryl Hydrocarbon Receptor Signaling as a Potential Mechanistic Link

Particle-bound pollutants give a further point of contact with uremic biology, and experimental renal models provide one example of this interaction. In human proximal tubular cells, co-exposure to polyethylene microplastic and bisphenol A raised aryl hydrocarbon receptor transcript and protein above the level reached by either agent alone, the receptor nuclear at 5 h and perinuclear by 24 h, while heat shock protein 90, which holds the receptor in its cytoplasmic complex, fell by 40% under the same conditions [59]. Polycyclic aromatic hydrocarbons such as benzo[a]pyrene sorb to the hydrophobic surface of plastic particles [78,79] and are classic ligands of that same receptor [91]; the tryptophan-derived uremic solute indoxyl sulfate is itself a potent endogenous agonist at nanomolar concentrations [92] and injures podocytes through the receptor in vivo [93]. In a prospective series of 116 individuals with chronic kidney disease stages 3 to 5D, the receptor-activating potential of serum exceeded that of healthy controls, was associated with estimated glomerular filtration rate and indoxyl sulfate concentration, and clustered with cardiovascular events above its median. The target genes CYP1A1 and AHRR were upregulated in whole blood, and a single dialysis session lowered the activating potential, so treatment begins with the receptor system under agonist load and relieves part of that load [60]. These observations raise the possibility that particle-associated aromatic pollutants and retained uremic toxins converge on receptor activation in the vasculature and the residual kidney [94], a convergence proposed before as a route by which micro- and nanoplastics might potentiate uremic toxicity in advanced chronic kidney disease [35]. Each component of this pathway is supported individually; the complete sequence has not been demonstrated in patients. No study measuring pollutants sorbed to particles recovered from patients was identified, and extending this reasoning to the dialysis population depends on first showing that treatment introduces the particulate vector alongside the retained solute.

7. Potential Exposure from Maintenance Dialysis

7.1. Source Compartments: Dialysis Fluids and Water

Roughly 300 to 600 L of dialysate and treatment water pass through the dialyzer of a hemodialysis patient each week, separated from blood by the membrane [41]. The one head-to-head comparison of the two modalities examined 30 samples drawn from four dialysis solutions using Fourier-transform infrared spectroscopy and recovered 36 candidate particles, 35 of them fibers, from every solution tested [41]. Concentrations were statistically indistinguishable between hemodialysis and peritoneal dialysis fluids, at 0.29 ± 0.16 against 0.34 ± 0.02 particles/L, yet hemodialysis fluids carried significantly larger particles, at a mean of 1.31 ± 0.98 mm (range 0.15 to 4.41) against 0.64 ± 0.43 mm (range 0.11 to 1.32), with fiber lengths across both modalities spanning 0.11 to 4.41 mm. The modalities separate on particle size and coincide on count [41]. Contamination was present throughout, and compromised particles were excluded from the final dataset [41]. Weekly delivery through pre-packaged fluids yielded roughly 10 particles for hemodialysis and 14 for peritoneal dialysis, figures for which the underlying counts do not support precision; the polymer profile, dominated by ethylene-vinyl acetate combinations, polyethylene, and polyvinyl chloride, matched the packaging [41]. Two features of the sampling design constrain that estimate: the samples came from unopened containers of acid and bicarbonate concentrate, before any contact with the machine or water treatment, so the figure covers the packaged component alone; and the weekly arithmetic rests on the 33 L of concentrate a patient receives across three sessions, not the 300 to 600 L of dialysate that concentrate goes on to make. Half the recovered particles were cellulose, a natural polymer, and the authors caution that pre-packaged samples may not reflect online-prepared fluids [41]. Earlier work examined dialysis water instead, combining a literature review, a field investigation of a London dialysis unit, and a model of patient intake, and named treatment water as a candidate route in its own right [95]. Assuming 99% filtration efficiency, that study modeled weekly delivery as ranging from 0.0021 to 3768 particles, an interval whose width follows directly from the assumptions applied [95]. It located candidate release points along the treatment train, in storage tanks and in distribution downstream of purification, and raised the possibility that reverse osmosis membranes, synthetic polymers that age, act as a source as well as a barrier, identifying the quantification of particles in water after filtration as the outstanding gap [95]. The efficiency of reverse osmosis and ultrafiltration for the smallest micro- and nanoplastic particles has not been systematically characterized [41,95]. These fluid measurements transfer poorly between studies and into any estimate of systemic burden: one study counted 36 candidate particles in 30 aliquots by infrared spectroscopy [41], the other modeled a weekly load from assumed water volumes [95], and the two report quantities in units that do not convert, under the contamination and size-floor constraints collected in Table 1. Table 2 lists the polymers and plastic-associated chemicals reported across the dialysis-related studies. Size cuts across all of these considerations, since the fibers recovered from dialysis fluids were millimeter-scale, far above the micrometer particles found in kidney tissue, which leaves open the question of whether fluid-borne contamination and tissue-deposited particles belong to the same population.

7.2. Circuit-Derived Particles

Mechanical compression and friction of tubing in peristaltic roller pumps cause spallation, shedding submicron- and micrometer-scale fragments into the blood path, a mechanism also observed in automated peritoneal dialysis cyclers [96]. Release of this kind was measured on the dialysis circuit itself four decades before environmental plastics drew wide interest, and that earlier literature has since been little cited. Silicone pump-head tubing stressed under simulated hemodialysis showed, on scanning electron microscopy, inner surface abrasion after 5 h of normal use, extensive damage by 20 h and splitting and laceration by 96 h, while atomic absorption spectrometry of the circulated blood found polydimethylsiloxane concentrations climbing with duration of use [97]. Release rates were subsequently compared across silicone rubber, polyvinyl chloride, and a co-extruded polyurethane-coated polyvinyl chloride, driven by blood pumps of the kind used in hemodialysis units [98]. A separate study identified a modifiable machine setting as a determinant: reducing the occlusion force from 22 to 5.5 kiloponds (roughly 216 to 54 N) lowered silicone particle release by about 80% and reduced the silicone recovered from the recirculation fluid from 1.6 mg to below 0.23 mg [99]. The single modern extension extends particle sizing down to 0.6 μm and yields a mean spalled diameter of 0.83 ± 0.03 μm for polyvinyl chloride tubing, with polyvinyl chloride and polyurethane particles roughly twice that size, a range beyond the resolution of the particle counters used in the earlier studies [96].
Spalled material was subsequently recovered from human tissue, with histological consequences. Refractile silicone particles were identified in the livers of hemodialysis patients with chronic renal failure, alongside hepatic inflammation and fibrosis of varying degree, with granulomatous hepatitis in nine cases [100]. Retrospective review found the material in 18 of 38 liver biopsies from hemodialysis patients with clinical hepatic dysfunction; among 31 autopsies, 22 carried silicone in the liver and all of them carried it in the spleen, with marrow, lungs, and nodes involved in some. Its identity was confirmed by atomic absorption and electron microprobe analysis, and its source was traced to a segment of silicone tubing in the roller pump of the dialysis machine [100]. The inflammatory sequel was reproduced in rats loaded with 3 × 109 silicone or polyvinyl chloride particles over four weeks: their unstimulated peritoneal macrophages released interleukin-1 at a median of 21.8 units (range 10 to 36.7) and 94 units (range 36 to 336) per 106 cells, against 4.1 units (range 1.2 to 10.3) in controls [101]. Intravenous silicone, polyvinyl chloride, or polyurethane particles lodged in macrophage lysosomes of spleen, liver, and lung and drove spontaneous prostaglandin E2 release from 4.27 ± 0.85 ng/mL/24 h in controls to 51.9 ± 13.2 for silicone and 57.5 ± 10.6 for polyvinyl chloride [102]. Spleen macrophages from two dialysis patients with hepatosplenomegaly and pancytopenia held silicone particles on transmission electron microscopy and electron microprobe analysis and released more prostaglandin E2, thromboxane B2, and 6-keto-prostaglandin F1α than control cells [103]. Most of this literature concerns silicone pump segments that are no longer used routinely, having been replaced by polyvinyl chloride. Apart from the single tubing study that resolved spalled fragments down to 0.6 μm [96], the polymers now in use in extracorporeal circuits have not been characterized at submicron resolution under treatment conditions, and none has been examined below that floor; closing that gap is a priority. These historical studies concern medical-device polymers, a category distinct from environmental microplastics, and establish the principle that extracorporeal circuits can generate particulate material.
No spectroscopic measurement of the circuit’s output under pump-driven blood flow was identified in this review. The closest is the equipment arm of a dual-simulation study [23], where ultrapure water run through the dialysis machine returned particles identifiable by polymer type, an approach that samples the machine as a whole and attributes the release to no single component. Intravenous devices contribute a further source of particles to the blood path. Raman spectroscopy with energy-dispersive analysis recovered between 0.44 and 2.00 particles per unit from syringes, infusion sets, and indwelling venous needles, mostly polyethylene and polypropylene fragments [104]. Release varied by brand, specification, and use; detection ranged from none to every unit; repeated use markedly increased it; and the built-in filtration membranes both incompletely retained particles and shed fibers of their own [104]. Estimated annual delivery remained modest, roughly six particles per infusion set and four per syringe, well below estimates for dietary intake, though these devices deliver directly into the bloodstream and the dietary figures carry wide uncertainty of their own [104]. A dialysis circuit exposes blood to a far larger polymer surface than an infusion set, under pump-driven flow across four-hour sessions three times weekly, so these figures extrapolate poorly.

7.3. Plastic-Derived Chemicals

Bisphenol A and bisphenol S have been measured in hemodialyzers, and membrane composition shapes plasma bisphenol A during online hemodiafiltration [105,106]. Phthalates have been measured as well: di-2-ethylhexyl phthalate, the plasticizer of the polyvinyl chloride blood lines, leaches from those parts of the circuit. Measured across the dialyzer in 21 maintenance hemodialysis patients, a single four-hour session extracted 75.2 mg on average (range 44.3 to 197.1 mg), of which the patient retained between 3.6 and 59.6 mg. Extraction rate tracked serum lipid in the fifteen patients contributing to that analysis (r = 0.705, p < 0.05), the mark of a lipophilic plasticizer partitioning into blood, a process distinct from a particle shearing off a surface [107]. Exposure to the same plasticizer has also been linked to uremic pruritus [108]. One prospective series has tested whether the chemical exposure is modifiable. Twenty-five prevalent hemodialysis patients, each serving as their own control, were followed for six months across a change to bisphenol A-free dialyzers: of five bisphenol analogs measured in serum, only bisphenol AF fell appreciably, while primary DNA damage on the comet assay decreased and chromosomal damage on the micronucleus test did not [109]. The design cannot separate changes in the device from the passage of time; oxidized DNA bases were unchanged, and the endpoints are biomarkers, with no clinical outcome measured. It remains the only interventional evidence identified here that a plastic-associated exposure in dialysis responds to a change of device material. It also demonstrates, in this population, the within-patient design that the particulate question will need. Particle-associated and dissolved chemical exposures may overlap, though the incremental chemical dose attributable to particle carriage has not been quantified in dialysis patients [59].
Table 2. Polymers and plastic-associated chemicals reported in dialysis-related studies.

7.4. Membrane Permeability and the Backfiltration Pathway

In high-flux dialyzers, fluid moves from the dialysate into the blood compartment in substantial volume, and the magnitude of that flux depends on the pressure profile along the fiber bundle. Positive transmembrane pressure at the blood inlet drives filtration in the proximal segment; distally, the hydraulic pressure drop along the lumen and the oncotic pressure that rises as plasma protein concentration increases reverse the gradient, and dialysate returns across the membrane as backfiltration, with the two fluxes balancing when net ultrafiltration is held at zero. Scintigraphy in a high-flux dialyzer at a blood flow of 300 mL/min put the filtration–backfiltration rate at 23.1 mL/min, rising to 48.2 mL/min once the fiber inner diameter was narrowed and the blood-compartment pressure drop rose from 112 to 159 mm of mercury. Vitamin B12 and inulin clearance then climbed by more than 30%, with urea, creatinine, and phosphate clearance unchanged [111]. Estimates of the mechanism reach 60 mL/min, near 3.6 L/h, and the filtration–backfiltration couple is considered a major route by which larger compounds are cleared during standard high-flux hemodialysis, which is why dialysate quality is controlled stringently in this modality [45]. On that arithmetic, several liters of dialysate participate in internal backfiltration across the membrane during a session, though net fluid balance stays determined by the prescribed ultrafiltration, so any case for dialysate as a particulate source must be argued through this flux. Pore geometry sets the size of what that flux can carry. Backfiltered fluid is screened by the pore population that defines the membrane class, and the operative dimension is set once plasma protein deposits on the surface, a layer reported to approach completion after approximately 40 min of blood contact under the conditions studied, and so governing most of a four-hour session [46]. Size discrimination occurs in a thin inner skin, about 1 µm wide or less, at the blood–membrane interface; the remaining 20 to 50 μm of wall provide mechanical support and an adsorptive surface [45]. In the medium-cut-off membranes, among the most permeable classes in maintenance use, the effective pore radius under treatment ranges from 3.0 to 3.5 nm, and the protein layer restricts passage to molecules with a radius below 3.5 nm [46]. An aperture of that dimension is three to six orders of magnitude smaller than the fibers of 0.11 to 4.41 mm reported in dialysis fluids [41] and the particles of 20 to 50 μm reported in the blood of dialysis patients [23], once radius is compared with radius. An intact particle sieving across an undamaged membrane is therefore difficult to reconcile with these dimensions, at least for the sizes so far measured in any clinical sample, and, by the same logic, the membrane screens backfiltered dialysate at the point of entry. That logic depends on pore dimensions inferred from solute sieving, with no direct test of membrane permeability to particles, and it holds for an intact membrane under the conditions studied. This shifts the question to the blood side without resolving it, since the appearance of particles in blood across a session carries no information about their origin [23]. The three modalities differ in how the fluid is delivered, and those differences delineate distinct exposure scenarios without ranking them.
An experimental precedent exists for the question that the geometric argument leaves open, and it concerns bacterial products, with no particulate challenge. Transfer of pyrogenic material from dialysate to the blood side has been measured directly across high- and low-flux membranes in a saline model under both diffusive and convective conditions: with purified lipopolysaccharide as the challenge, transfer was substantial across one polyethersulfone membrane and undetectable across polysulfone and Helixone. With bacterial culture filtrates, the concentrations recovered behind polyamide S and polysulfone were around a tenth and a hundredth of those behind the polyethersulfone device [47]. A later design brought the measurement closer to treatment conditions by running full-sized low-flux, high-flux, medium-cut-off, and high-cut-off dialyzers on a clinical monitor at blood and dialysate flows of 400 and 500 mL/min, with the membrane precoated in human plasma and net ultrafiltration set to zero to maximize backfiltration. Endotoxin appeared on the blood side above the paired baseline in nine of twenty-four experiments, distributed one, one, three, and four across the four membrane classes; the logarithmic retention values did not separate, and none of the recovered solutions induced interleukin-1 beta in a monocytic bioassay [48]. Two points from that work inform the particle question: transmembrane passage of material of nanometric dimension is neither excluded by pore geometry nor determined by it, since retention tracked polymer chemistry more closely than nominal pore size, and the experimental architecture needed to settle it already exists in validated form, wanting only a characterized particle population in place of the bacterial filtrate.

7.5. Modality-Specific Transfer Pathways

In conventional hemodialysis, the membrane keeps dialysate and blood apart, so a particle originating in the treatment water or dialysate would have to survive purification, remain in the final fluid, then cross the membrane or be carried in by backfiltration before reaching the circulation, a chain still only partially measured. A particle shed from blood-side tubing, connectors, pump segments, or the dialyzer housing meets no such barrier. Cardiac surgery lends support to blood-side generation: in vitro, conventional cardiopulmonary bypass, and minimally invasive extracorporeal circuits released 60.4 ± 7.6 and 48.4 ± 31.3 microplastics/L/h, respectively, compared with 5.6 ± 10.4 in procedural blanks (whose standard deviation exceeded their mean), and 20 polymer types were recovered [110]. There was no dialyzer and no patient, so the experiment shows only that extracorporeal circuits shed particles, and delivery by dialysis remains unmeasured; the more informative finding is the absence of a significant difference between two architectures of unlike surface area and priming volume [110].
Online hemodiafiltration shares every blood-side surface of conventional hemodialysis and adds one further element: the infusion of substantial volumes of substitution fluid into the extracorporeal blood circuit, a distinct potential transfer pathway of unquantified size. The volume of fluid processed or infused is no proxy for particle exposure. Systemic dose depends on the particle concentration at the point of fluid entry into the patient, on size distribution and polymer identity, on how efficiently particles cross the relevant barriers, and on whatever the extracorporeal system simultaneously retains or removes. None of this has been measured for online-prepared substitution fluid. One comparison between modalities exists, and it is thin: six patients in a cross-sectional survey had higher circulating counts than the twenty-five treated with acetate-free biofiltration in the same unit, with interquartile ranges that overlap across most of their width. The blood was drawn before connection to the circuit, so it describes the burden a patient carries into a session and carries no information on what the session delivers [19]. Nothing reviewed here, including that observation, shows that online hemodiafiltration produces a greater systemic burden than conventional hemodialysis. The technique does confer established clinical benefit, with randomized trials showing lower all-cause mortality with high-volume postdilution online hemodiafiltration than with high-flux hemodialysis [112,113]. That benefit has been attributed mainly to improved clearance of middle molecules and reduced inflammation [114]; particulate exposure has not been measured in either modality.
Peritoneal dialysis presents a different configuration of exposure, in which the solution remains in prolonged contact with polymeric containers and tubing and is instilled into the peritoneal cavity, with no direct entry into the bloodstream, so any systemic exposure would depend on release into the solution, on the handling of that material by the peritoneal compartment, and on transperitoneal uptake afterwards. A murine model has begun to address that route, with microplastics infused into the peritoneal cavity followed by overexpression of growth differentiation factor 15 in the kidney and heart, reported so far only in abstract form [115]. Particle counts in packaged peritoneal solutions cannot, then, be set alongside counts in hemodialysis fluids as if both measured systemic dose.

7.6. The Dialyzer as a Source of Particles and as a Candidate Site of Retention

The release documented from the dialyzer is of soluble material, with no particulate counterpart. Polyvinylpyrrolidone, blended with polysulfone and polyethersulfone to confer the hydrophilicity that allows the membrane to wet and limits protein deposition, has been implicated in reactions observed under certain sterilization conditions and during bleach-based reprocessing, without direct measurement of its release from the membrane [45,50]. No measurement of particulate shedding from a dialyzer was identified, and the release documented in extracorporeal circuits comes from tubing squeezed by roller pumps [96,110]. The dialyzer holds polymers beyond the membrane, and their location within the device bears on the question [45]. Several thousand hollow fibers are wound into a bundle and set in a plastic housing, with a polyurethane potting compound encapsulating them at either end [45]. Polyurethane, therefore, meets blood at both ends of every device, and poly (methyl methacrylate) forms the membrane itself in one class of dialyzer [45]. Ethylene oxide used to sterilize the device settles into the same polyurethane filler, consistent with the accepted account of sensitization in the type A dialyzer reactions described when that sterilization method was in widespread use [50]. Both polymers were recovered from post-dialysis blood in the only clinical measurement available [23], though that study assigns no origin to its findings. The appearance of poly (methyl methacrylate) in blood would carry an origin only where the device in use is known, and the polymer is also common in laboratory and environmental settings. Polyurethane, however, also coats the co-extruded pump tubing described above [98], so its presence in blood suggests an upstream origin. A further detail bears on what a membrane would shed: the skin layer is a nodular lattice of closely packed polymer spheres 50 to 100 nm across [45], a scale below the size range addressed by the analytical studies identified in this review. Dialyzer reuse has been proposed to compromise membrane integrity and so raise particulate exposure, a mechanism still untested and relevant wherever reuse persists [50]. Protein adsorption at the blood–membrane interface depends on membrane properties and on the hydrodynamics near the surface, follows the Vroman sequence within minutes, and varies in extent with roughness, zeta potential, and wettability [50]; the resulting secondary membrane, the gel or protein cake, reduces effective permeability from the moment blood arrives and is mostly albumin, fibrinogen, and immunoglobulins [45]. Pore dimensions quoted across that literature mix membranes built for laboratory separation with those used clinically, and only the clinical ones are relevant to transfer in a patient. That review calls microplastic exposure in hemodialysis a theoretical risk not yet systematically explored [50]. The framing here is the membrane literature’s own: Ronco and Clark locate a modality’s biocompatibility at the level of the whole system, counting the blood tubing and every other biomaterial of the extracorporeal circuit alongside the membrane, as well as the fluid to which the patient is exposed [45].
The same membrane may also act on particles already circulating. Blood traverses a large polymer surface over four-hour sessions, and because a corona of plasma protein forms on a particle within seconds of entering blood and shapes its behavior from then on [56], a dialyzer that sheds particles is equally a candidate site of retention. No study of this was identified, and the obstacle is that a blood-side count cannot separate the possible outcomes: a particle leaving the circulation may lodge on the membrane or pass into spent dialysate, and a particle that seems to have arrived may never have been in the patient, the signature of contamination at sampling. Retention divides further into surface adsorption set by the corona and by membrane roughness, zeta potential, and wettability [50,56], and mechanical entrapment set by particle size against pore dimension [50], each depending on flow, priming, and particle size and neither of them distinguishable from the blood side.
That points to a research program built on a conceptual mass-balance framework, an accounting device still to be validated, in which the change in circulating burden across a session equals the inputs, from water and dialysate, from circuit components, and from infusion, minus what the dialyzer retains and what leaves by other routes. Figure 1 maps those compartments, and the points at which the balance would have to be sampled. Set out in this form, the framework exposes the principal sources of misattribution: a rise measured after a session may be true input, surface shedding, hemoconcentration, redistribution from tissue, analytical drift, or contamination at sampling, and a flat concentration may hide input and removal canceling out, which is the main reason to read the nine-patient observation cautiously [23]. Until inputs and removals are quantified within the same session, modality-specific exposure stays an open empirical question.
Figure 1. Candidate sources, barriers, and measurement points for a source-resolved mass balance across a hemodialysis session. Particles entering with treatment water, concentrates and packaging are screened by the water-treatment train and, on the dialysate side, by the membrane. Substitution fluid infused during online hemodiafiltration, and material shed by tubing, pump segment, connectors and dialyzer housing, reach the circulation without crossing a membrane. Closing the balance requires paired fluid-side and blood-side sampling within the same session, under identical procedural blanks and with correction for hemoconcentration; the filled circles mark the points at which measurement is needed. Solid arrows indicate inputs to the circulating burden; the dashed arrow indicates losses from it.

8. Regulatory Specification of Particulate Matter in Dialysis Fluids

The regulatory specification for dialysis fluids covers microbiological and dissolved chemical contamination but sets no limit on particulate polymer. ISO 23500-5 lays down the minimum chemical and microbiological quality requirements for dialysis fluids, with its scope explicitly extending to substitution fluid produced online for hemodiafiltration and hemofiltration, with colony-forming unit and endotoxin limits that tighten from standard to ultrapure fluid, and a requirement that online-prepared substitution fluid be sterile and nonpyrogenic [116]. The chemical requirements govern composition and the trace elements inherited from water and concentrates, and they oblige delivery systems to add no chemical contaminants to the final fluid [116]. No specification for subvisible particulate matter was identified in the sections of that document reviewed here. Device manufacture exhibits the same asymmetry, since hollow fibers and dialyzers are made in clean rooms to keep bacterial load and endotoxin levels low [45], so control at both stages is directed toward microbial contamination and leaves particulate contamination unquantified.
A different standard applies to packaged injections: USP general chapter 788, harmonized with European Pharmacopoeia 2.9.19 and Japanese Pharmacopoeia 6.07, allows large volume parenterals no more than 25 particles/mL at or above 10 μm and 3 particles/mL at or above 25 μm [117,118]. A packaged 250 mL bag therefore carries a numerical particle limit, while ISO 23500-5, whose scope covers the substitution fluid infused during a high-dose hemodiafiltration session at 23 L of convection or more [112,113], sets none [116]; the applicability of the pharmacopoeial monographs to online-prepared fluid is a regulatory question this review leaves open. The gap follows from how conformity is established: ISO 23500-5 states that the sterility of online-produced substitution fluid cannot be proven by sampling, so conformity rests on a validated system operating properly, with no analysis of the delivered fluid [116]. The bacteria- and endotoxin-retentive filters that fluid passes through are themselves polymer membranes, qualified for microbial and endotoxin retention alone [116].
The fibers recovered from dialysis fluids ran to a millimeter scale [41], and the particles that appeared in blood across a session were mostly 20 to 50 μm [23], both above the thresholds at which pharmacopoeial limits apply, though the instrument in the blood study saw nothing below 20 μm, so its distribution is capped from below at that figure. No measurement identified here establishes the particulate load delivered by online preparation, and light-obscuration methods of the kind USP 788 prescribes would count particles at these thresholds without assigning polymer identity, so a compliance test and a source attribution remain separate exercises. From the pharmaceutical side, a recent appraisal of infusion therapy reaches the same conclusion: the established test methods and regulatory limits for particulate contamination were written without particulate plastic in mind, and any assessment adequate to judge health risk will need a standardized approach that combines several detection methods [119].

9. Human Data in Dialysis Populations

The only study identified here to examine within-session change in circulating particle concentration combined an equipment arm, in which ultrapure water was passed through the dialysis machine, with a patient arm, in which whole blood was drawn immediately before and after treatment from nine incident patients who had had no more than two previous sessions, and identified particles by laser direct infrared imaging spectrometry [23]. Water run through the equipment carried chlorinated polyethylene, ethylene-vinyl acetate, polyamide, and polypropylene, and none was detected in the pre-machine water samples [23]. In blood, the median particle count rose across the session from 30 to 100 particles/mL. The investigators reported a Hodges–Lehmann estimate of the paired difference of 50.0 particles/mL (one-sided 95% lower confidence limit, 2.5 particles/mL), a matched-pairs rank-biserial correlation of 0.733, and a prespecified one-sided Wilcoxon signed-rank exact test yielding p = 0.027, together with a two-sided sensitivity analysis of their own yielding p = 0.054 [23].
Chlorinated polyethylene, ethylene-vinyl acetate, poly (methyl methacrylate), polytetrafluoroethylene, and polyurethane were detected in post-dialysis samples but not in the paired pre-dialysis samples, with most particles measuring 20 to 50 μm within the method’s measurable range, whose 20-micrometer floor coincides with the lower edge of that distribution [23]. No association with the total volume of blood processed was detected, in aggregate or by polymer type, and the study was too small to determine release kinetics or source [23]. The two arms overlap in chlorinated polyethylene, ethylene-vinyl acetate, and polypropylene, although polypropylene was already present in blood before treatment. The appearance in blood of polytetrafluoroethylene, polyurethane, and poly (methyl methacrylate) with no counterpart in the water is compatible with, but not specific for, a blood-side or sampling-related source, whose contribution remains unquantified.
The absence of a volume response admits more than one explanation, among them release during priming or connection, and procedural contamination during sampling. The rise itself rests on a one-sided test fixed in advance; a two-sided sensitivity analysis yielded p = 0.054, and correction for multiple comparisons across polymer types raised the total false discovery rate to 0.324, so the findings should be regarded as preliminary and requiring independent confirmation [23]. Nine patients, no control arm, and blood obtained within an environment containing many of the polymers under investigation limit what the observation can support.
Between January 2022 and November 2024, a university nephrology unit and a dialysis clinic in Italy enrolled 92 participants for a comparison between groups: 31 patients on maintenance hemodialysis for at least six months, 31 with stage 3 to 4 chronic kidney disease, and 30 hospital patients with no history of it, each sampled once [19]. Blood particle counts were higher in the two patient groups than in the comparison group, at medians of 9889, 9234, and 5755 particles/mL. Pyrolysis of the same samples showed polyethylene ahead of the other polymers by mass in every group, with polypropylene higher in the dialysis patients, which the investigators considered consistent with exposure to components of the dialysis system used at their center [19]. The polymer distribution qualifies that reading: polyethylene exceeded its limit of detection in all three groups, at 21.3 µg/mL in the comparison subjects against 26.9 and 25.7 in the patients; polypropylene reached 3.34 in the hemodialysis patients but also 1.80 in the comparison subjects and fell below detection in chronic kidney disease without dialysis [19]. The gradient is therefore not monotonic, and subjects with no dialysis exposure carried the polymer credited to the dialysis system. Both these masses and the particle counts are one to two orders of magnitude above the other human blood measurements assembled here [16,23], for the reasons of size window, identification specificity and sample preparation set out in Section 2.4.
In the dialysis group, blood came from the vascular access before connection to the circuit, so these counts describe the burden a patient carries into a session and say nothing about what happens across one. Counts also differed by dialysis vintage, treatment modality, and sex, though the investigators present these comparisons as exploratory and without hypothesis testing, and the study was not designed to establish independent associations [19]. The subgroups are small: six patients on online hemodiafiltration returned a median of 15,004 particles/mL, compared with 8664 in twenty-five on acetate-free biofiltration, with interquartile ranges that overlap across most of their width [19]. Counts fell with more recent initiation, from 10,701 among the eleven who began in 2017 or earlier to 6272 among the five who began in 2021. That direction is compatible with cumulative exposure, and equally with the differences in age, comorbidity and device generation that separate those groups [19]. The comparison between the sexes, the report’s declared focus, went in opposite directions across groups, which is difficult to attribute to any single mechanism at these group sizes [19]. Contamination control ran to glass and aluminum labware, filtered reagents, reagent blanks that returned no detectable particles, and spiked recoveries between 83% and 103%, which makes laboratory handling a less likely explanation for the difference between the groups; blanks of that kind still leave contamination that arises before analysis unaddressed [19]. The design cannot separate kidney failure from dialysis exposure or from the other differences between the groups, and the higher burden in the patients on dialysis admits all three. A registered biomonitoring study will measure particles in the blood of chronic kidney disease and hemodialysis patients and assess genotoxic and immunological effects [120].

10. Pediatric Dialysis

Children receiving maintenance dialysis accumulate more treatment years than adults, and they do so during growth and organ development. Microplastics figure among the planetary health challenges identified as bearing particularly on the health of children and adolescents [121], and the pediatric environmental health literature attributes their distinct vulnerability to environmental exposure to anatomical and functional immaturity, a higher metabolic rate, and a life expectancy over which latent effects can emerge [122]. Particle biomonitoring in this population was not found in the studies reviewed, and the design that this review proposes for adults, paired fluid-side and blood-side sampling within the same session under identical procedural blanks, would need age-appropriate sampling volumes in children; a pediatric arm belongs in any standardized biomonitoring framework.

11. Toxicodynamics in the Uremic Host

Chronic inflammation is prevalent in the dialysis population, contributes to erythropoiesis-stimulating agent resistance and to excess mortality, and persists in a subset of patients without an identifiable conventional cause [123]. Should dialysis prove to introduce particulate material into the circulation, it would be one candidate contributor to that unexplained inflammatory burden.
The pathways that particles engage in experimental systems are already under strain in uremia. The aryl hydrocarbon receptor is the best-documented instance (Section 6.2): the receptor system that already carries an agonist load in these patients is the same one that the particle and its sorbed chemical activate [59,60]. Complement overlaps in the same way, since particles injured the mouse kidney through the C5a-C5aR axis, with little direct effect on renal cells [73]. An older observation is specific to the dialysis population: macrophages loaded with particles from dialysis tubing secreted interleukin-1 [101], the cytokine whose maturation the NLRP3 inflammasome governs within the kidney [75], which puts a particle-driven inflammatory stimulus in dialysis patients on record decades before the present literature, albeit in reticuloendothelial macrophages, with no renal counterpart.
Particles may therefore act on pathways already activated, with the potential to affect the residual native kidney and a vasculature primed for injury [43,124]. The oxidative stress, sterile inflammation, and fibrogenesis they produce in experimental systems are among the drivers of progression in established chronic kidney disease, and should they extend to particles, reduced clearance in advanced disease would lengthen exposure to them and to their adsorbed contaminants [29,124]. The single human dataset linking particle burden to outcome comes from a vascular bed, with no renal counterpart [24], in a population whose cardiovascular risk overlaps with that of dialysis.

12. Study Designs and Research Priorities

Maintenance dialysis provides natural contrasts across modalities, membrane types, water-treatment configurations, and stages of residual renal function, and those contrasts could be exploited to investigate systemic particle exposure; the task is to match that clinical spectrum to designs that can apportion exposure to its sources and relate it to outcome [35,61]. Contrasting hemodialysis, hemodiafiltration, and peritoneal dialysis would test the relationship between the directness of fluid delivery and circulating particle burden, and an incident cohort of patients who retain residual diuresis and subsequently lose it would provide a within-person transition over which particle handling could be followed directly, testing the elimination hypothesis. Once membrane and water-treatment contrasts are set against measured burden, polymer choice and purification stringency become testable as modifiable determinants. Amplification of injury already underway in the uremic host calls for a fresh design again, one linking baseline particle burden to later cardiovascular events and inflammatory markers, including erythropoiesis-stimulating agent responsiveness.
Exposure metrics should be resolved by modality and membrane, and reporting should span particle counts, mass, size distribution, and polymer identity. The permeability of high-flux and hemodiafiltration membranes to submicron particles needs direct measurement, for which the circuits built to quantify endotoxin transfer supply a validated template [47,48], and so does the dialyzer’s capacity to remove particles already circulating. That capacity is difficult to measure in a patient, where the input is unknown and can be fixed only in a circuit spiked with a characterized particle population; the first design, then, is a closed loop in vitro into which particles of defined size, polymer, and count are introduced on the blood side. Recovery would be accounted for across compartments, the effluent blood, the spent dialysate, and the dialyzer itself after elution or digestion, and a balance that failed to close would identify the residual requiring explanation. Running the same loop without a dialyzer, and again without spiked particles, separates what tubing and pumps shed from what the membrane does and fixes the procedural floor against which any patient measurement is read. Adsorption and entrapment can then be distinguished by their dependencies: adsorption varies with priming and the protein corona and should saturate, whereas entrapment tracks particle size relative to pore dimensions and shifts the size distribution of what emerges. Depletion spread evenly across the size range would point to the first; depletion confined to the coarse end, with the fine fraction recoverable in dialysate, to the second. Water treatment requires study in its own right, the efficiency of reverse osmosis and ultrafiltration for the smallest particles calling for direct evaluation [41,95].
The polymer profile of packaged concentrates, which matched their containers, makes lower-shedding or polymer-reduced materials for concentrates and substitution fluids worth evaluating, alongside bisphenol-free, low-leaching device components; particulate specifications for fluids intended for direct infusion merit evaluation, since the chemical and microbiological requirements now in force omit them [41,80,116,125]. The 2024 edition of ISO 23500-5 entered systematic review in October 2025 and a working draft to replace it is under development, so a particulate specification could be proposed while the text is open for revision [126]. These priorities align with the wider movement towards environmentally sustainable nephrology [61,127]. The same polymers leave the unit in the effluent, several hundred liters per patient each week, discharged to sewage with no treatment step directed at particulate polymer, which makes the dialysis unit a candidate environmental source as well as a site of patient exposure; no measurement of particle concentrations in that effluent was identified [128]. Table 3 sets the available evidence against the matching gaps and the study designs that would close them.
Table 3. Current evidence and research priorities for micro- and nanoplastic exposure in maintenance dialysis.
Biomonitoring across a single session, blood drawn before and after treatment in the same patient, provides the most direct available assessment of within-session change and avoids the between-person variation that undermines case–control work; applied once now, in nine patients, it needs replication at larger scale, with a control arm and across modality and membrane, to establish the source of the signal, the water train or the circuit [23]. Measurement in the same patients at the initiation of dialysis, with paired fluid-side sampling and procedural blanks from the circuit and the sampling set, would directly address the contribution of treatment, a design that the investigators of the larger between-person comparison also propose [19]. Any such design would need to record renal function and prespecify adjustments for the demographic and clinical differences that cross-sectional comparisons could not address [19]. Therapeutic plasma exchange illustrates how much depends on this: across 174 procedures in 114 patients, circulating particles fell only where the starting count was high; at low starting counts the tubing and fluid bags of the apheresis set added particles of their own and counts could rise across the procedure, a reading tempered by an assay that used fluorescent staining without spectroscopic confirmation of polymer identity [40].

13. Implications for Regulators, Decisionmakers and Public Health

13.1. Policy Measures

The analysis above has consequences beyond the research agenda, and they fall to regulators, device manufacturers, water-quality authorities, and the bodies that run nephrology registries. A standardized, contamination-controlled biomonitoring framework for dialysis patients would fix the size floor, the polymer identification method and the reporting units in advance, require procedural blanks from the circuit, the sampling set and the laboratory to be reported with every patient value, and include a pediatric arm. Nephrology registries that already record modality, membrane, water-treatment configuration, and hard outcomes could host such a framework, and the registered biomonitoring study in chronic kidney disease and hemodialysis offers a starting point [19,23,120].
ISO 23500-5 sets chemical and microbiological requirements for dialysis fluid and for online-prepared substitution fluid and sets no particulate limit [116]. The pharmacopoeial chapters for large volume parenterals supply a model, counting particles at 10 and 25 μm by light obscuration [117,118], though a specification written for polymeric particles would need a lower size threshold and a polymer identification step that those chapters do not require. The 2024 edition entered systematic review in October 2025 and a committee draft is under development [126], which places the question before the responsible committee while the text is open. Pre-market assessment of dialyzers, blood tubing sets, pump segments and peritoneal dialysis containers could include a particulate shedding test under simulated treatment conditions at submicron resolution, with the blood- and dialysate-contacting polymers declared on the label, as a counterpart to the chemical characterization that bisphenol A-free dialyzer lines have undergone [105,109]. Water-treatment standards could define a performance requirement for reverse osmosis and ultrafiltration against submicron polymer particles, alongside the microbial and endotoxin retention for which these stages are qualified now [95,116]. Dialysis effluent, several hundred liters per patient each week discharged with no treatment step directed at particulate polymer, could be characterized for particle content as part of the environmental agenda that nephrology has begun to adopt [127,128].

13.2. Precautionary Measures Available to Dialysis Units

The evidence reviewed here does not justify a change in clinical practice: the studies establish neither net particle delivery by dialysis nor accumulation nor harm, and online hemodiafiltration, the modality that infuses the largest fluid volume, carries an established survival benefit [112,113]. Some measures carry little cost or risk and would make a unit’s own data interpretable if a measurement were later attempted. Occlusion force on roller pumps can be reviewed against the manufacturer’s specification, a setting that one historical experiment related to the quantity of silicone shed from the pump segment [99], with its relevance to polyvinyl chloride segments unmeasured. Low-shedding and bisphenol-free components can be preferred where the choice carries no clinical penalty [80,109,125]. Particulate contamination can be added to the monitoring of water treatment at the size thresholds that light-obscuration counting already resolves, with the result recorded as a count, since a compliance verdict presupposes a limit that does not yet exist [117,118]. Any local measurement should be reported with its procedural blanks. These steps are precautionary and research-enabling; they do not rest on demonstrated harm, and they should not be presented to patients as protection against one.

14. Conclusions

Among the established findings, spectroscopy has identified microplastic particles of 1 to 29 µm in human kidney tissue and urine, while pyrolysis quantifies polymer mass without sizing it, so the nanoplastic fraction in human samples remains inferred [16,17,21,22,24]. Fibers of 0.11 to 4.41 mm have been recovered from packaged concentrates, and treatment water is a candidate route in its own right [41,95]. Extracorporeal tubing sheds particles under roller-pump flow, with silicone later recovered from liver and spleen [97,98,99,100]. In experimental kidneys, particles produce oxidative stress, inflammation, and fibrosis at administered doses with no validated conversion to human internal dose [43,66,67,68,73]. Circulating counts were higher after a single session in nine incident patients, and higher in hemodialysis and in non-dialysis chronic kidney disease than in comparison subjects in a cross-sectional study of 92 [19,23].
The nine-patient study measured a change in concentration across a session; converting that change into net delivery would have required paired fluid-side and blood-side sampling under common blanks with correction for hemoconcentration, and without them a rise remains compatible with contamination, plasma volume reduction and redistribution as well as with input [23]. Pore dimensions derived from solute transport make the sieving of intact micrometer particles across an undamaged membrane difficult to reconcile with the geometry, and how much of a circulating load the dialyzer retains calls for a circuit in which the input is fixed [45,46]. One tubing study resolved spalled fragments to 0.6 µm, and the polymers now in routine use remain uncharacterized below that floor [96]. Urine has been proposed as a disposal route on the strength of particles recovered from it, and the clearance itself remains unquantified [19,21,35]. Carotid plaque supplies the one human dataset linking particle burden to a clinical outcome [24].
A closed-loop circuit spiked with a characterized particle population on the blood side, with recovery accounted for across effluent blood, spent dialysate and the dialyzer, would separate what the membrane retains from what the tubing sheds, on the template already validated for bacterial products [47,48]. A source-resolved mass balance in the same patients, with paired fluid-side sampling, identical blanks and correction for hemoconcentration, would establish net delivery, and carried across the initiation of dialysis it would follow the loss of residual diuresis [19,23]. One laboratory processing exposed and unexposed material under a single protocol and size floor would supply the comparability the literature now lacks [18]. Maintenance dialysis offers a setting in which recurrent contact between blood and polymer-containing pathways can be quantified at its sources, and, until those quantities exist, a higher post-session count records the analytical environment as much as the treatment.

Supplementary Materials

The following supporting information can be downloaded at https://www.mdpi.com/article/10.3390/jox16080233/s1, Table S1: Internal validity checklist of the human biomonitoring studies and of the dialysis-fluid and circuit measurements drawn on in the review.

Author Contributions

Conceptualization, G.G. and D.S.; methodology, G.G., M.F. and T.L.C.; investigation, G.G., M.F., L.L.C., L.S., T.L.C., A.L.S. and A.S.; writing—original draft preparation, G.G., M.F., L.L.C., L.S. and A.L.S.; writing—review and editing, G.G., M.F., L.L.C., L.S., T.L.C., A.L.S., A.C., A.S. and D.S.; supervision, A.C. and D.S. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Data Availability Statement

No new data were created or analyzed in this study. Data sharing is not applicable to this article.

Acknowledgments

During the preparation of this manuscript, the authors used Claude Fable 5.2 (Anthropic) to generate the initial version of Figure 1. The figure was subsequently revised and finalized by the authors using Microsoft PowerPoint (Microsoft 365, version 2611, build 20625.20108; Microsoft Corporation). Additionally, Nature Research Assistant (Springer Nature, beta version) was used to critically reassess the manuscript’s structure, scientific content, and overall coherence. All AI-assisted outputs were independently reviewed and critically evaluated by the authors, who retained full responsibility for all editorial and scientific decisions and for the accuracy, integrity, and content of the final publication. The visual elements included in the graphical abstract were generated using Google Gemini 4 Hydrogen (Google LLC). These elements were subsequently arranged, integrated, and refined by the authors in Microsoft PowerPoint (Microsoft Corporation) to create the final graphical abstract.

Conflicts of Interest

The authors declare no conflicts of interest.

Abbreviations

AHRR, aryl hydrocarbon receptor repressor; ATP, adenosine triphosphate; C5a, complement component 5a; C5aR, complement component 5a receptor; CCR-2, C-C motif chemokine receptor 2; CCR-5, C-C motif chemokine receptor 5; CI, confidence interval; CKD, chronic kidney disease; CV, cardiovascular; CYP1A1, cytochrome P450 family 1 subfamily A member 1; DNA, deoxyribonucleic acid; dUTP, deoxyuridine triphosphate; FTIR, Fourier-transform infrared spectroscopy; HD, hemodialysis; HDF, hemodiafiltration; ISO, International Organization for Standardization; LC3, microtubule-associated protein 1 light chain 3; LDIR, laser direct infrared imaging; MCO, medium cut-off; MLKL, mixed lineage kinase domain-like pseudokinase; NLRP3, NOD-like receptor family pyrin domain containing 3; PD, peritoneal dialysis; PFAS, per- and polyfluoroalkyl substances; RIP1, receptor-interacting protein kinase 1; RIP3, receptor-interacting protein kinase 3; RO, reverse osmosis; SEM-EDX, scanning electron microscopy with energy-dispersive X-ray analysis; USP, United States Pharmacopeia; VAMAS, Versailles Project on Advanced Materials and Standards; WT-1, Wilms tumor 1.

References

  1. Organisation for Economic Co-operation and Development (OECD). Global Plastics Outlook: Economic Drivers, Environmental Impacts and Policy Options; OECD Publishing: Paris, France, 2022. [Google Scholar] [CrossRef] [Scilit]
  2. Landrigan, P.J.; Dunlop, S.; Treskova, M.; Raps, H.; Symeonides, C.; Muncke, J.; Spring, M.; Stegeman, J.; Almroth, B.C.; Chiles, T.C.; et al. The Lancet Countdown on Health and Plastics. Lancet 2025, 406, 1044–1062. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  3. Rhodes, C.J. Plastic Pollution and Potential Solutions. Sci. Prog. 2018, 101, 207–260. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  4. Chamas, A.; Moon, H.; Zheng, J.; Qiu, Y.; Tabassum, T.; Jang, J.H.; Abu-Omar, M.; Scott, S.L.; Suh, S. Degradation Rates of Plastics in the Environment. ACS Sustain. Chem. Eng. 2020, 8, 3494–3511. [Google Scholar] [CrossRef] [Scilit]
  5. Landrigan, P.J.; Raps, H.; Cropper, M.; Bald, C.; Brunner, M.; Canonizado, E.M.; Charles, D.; Chiles, T.C.; Donohue, M.J.; Enck, J.; et al. The Minderoo-Monaco Commission on Plastics and Human Health. Ann. Glob. Health 2023, 89, 23. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  6. Frias, J.P.G.L.; Nash, R. Microplastics: Finding a Consensus on the Definition. Mar. Pollut. Bull. 2019, 138, 145–147. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  7. Prata, J.C.; Da Costa, J.P.; Lopes, I.; Duarte, A.C.; Rocha-Santos, T. Environmental Exposure to Microplastics: An Overview on Possible Human Health Effects. Sci. Total Environ. 2020, 702, 134455. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  8. Cox, K.D.; Covernton, G.A.; Davies, H.L.; Dower, J.F.; Juanes, F.; Dudas, S.E. Human Consumption of Microplastics. Environ. Sci. Technol. 2019, 53, 7068–7074. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  9. Senathirajah, K.; Attwood, S.; Bhagwat, G.; Carbery, M.; Wilson, S.; Palanisami, T. Estimation of the Mass of Microplastics Ingested—A Pivotal First Step towards Human Health Risk Assessment. J. Hazard. Mater. 2021, 404, 124004. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  10. Zhang, Q.; Xu, E.G.; Li, J.; Chen, Q.; Ma, L.; Zeng, E.Y.; Shi, H. A Review of Microplastics in Table Salt, Drinking Water, and Air: Direct Human Exposure. Environ. Sci. Technol. 2020, 54, 3740–3751. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  11. Vega-Herrera, A.; Llorca, M.; Borrell-Diaz, X.; Redondo-Hasselerharm, P.E.; Abad, E.; Villanueva, C.M.; Farré, M. Polymers of Micro(Nano) Plastic in Household Tap Water of the Barcelona Metropolitan Area. Water Res. 2022, 220, 118645. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  12. Abbasi, S.; Keshavarzi, B.; Moore, F.; Turner, A.; Kelly, F.J.; Dominguez, A.O.; Jaafarzadeh, N. Distribution and Potential Health Impacts of Microplastics and Microrubbers in Air and Street Dusts from Asaluyeh County, Iran. Environ. Pollut. 2019, 244, 153–164. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  13. Amato-Lourenço, L.F.; Dos Santos Galvão, L.; De Weger, L.A.; Hiemstra, P.S.; Vijver, M.G.; Mauad, T. An Emerging Class of Air Pollutants: Potential Effects of Microplastics to Respiratory Human Health? Sci. Total Environ. 2020, 749, 141676. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  14. Habib, R.Z.; Salim Abdoon, M.M.; Al Meqbaali, R.M.; Ghebremedhin, F.; Elkashlan, M.; Kittaneh, W.F.; Cherupurakal, N.; Mourad, A.-H.I.; Thiemann, T.; Al Kindi, R. Analysis of Microbeads in Cosmetic Products in the United Arab Emirates. Environ. Pollut. 2020, 258, 113831. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  15. Xu, J.-L.; Lin, X.; Wang, J.J.; Gowen, A.A. A Review of Potential Human Health Impacts of Micro- and Nanoplastics Exposure. Sci. Total Environ. 2022, 851, 158111. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  16. Leslie, H.A.; Van Velzen, M.J.M.; Brandsma, S.H.; Vethaak, A.D.; Garcia-Vallejo, J.J.; Lamoree, M.H. Discovery and Quantification of Plastic Particle Pollution in Human Blood. Environ. Int. 2022, 163, 107199. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  17. Nihart, A.J.; Garcia, M.A.; El Hayek, E.; Liu, R.; Olewine, M.; Kingston, J.D.; Castillo, E.F.; Gullapalli, R.R.; Howard, T.; Bleske, B.; et al. Bioaccumulation of Microplastics in Decedent Human Brains. Nat. Med. 2025, 31, 1114–1119. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  18. Ciornii, D.; Hodoroaba, V.-D.; Benismail, N.; Maltseva, A.; Ferrer, J.F.; Wang, J.; Parra, R.; Jézéquel, R.; Receveur, J.; Gabriel, D.; et al. Interlaboratory Comparison Reveals State of the Art in Microplastic Detection and Quantification Methods. Anal. Chem. 2025, 97, 8719–8728. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  19. Fiore, M.; Lo Cicero, L.; Pulvirenti, E.; Zanoli, L.; Palella, M.; Bivona, F.; Timperanza, C.; Longo, M.V.; Grasso, A.; Cristaldi, A.; et al. Concurrent Assessment of Micro- and Nanoplastics and Bisphenol A in Non-Dialysis Chronic Kidney Patients and Hemodialysis Patients: Results from a Cross-Sectional Human Biomonitoring Study Focusing on Sex-Related Patterns. Microplastics 2026, 5, 159. [Google Scholar] [CrossRef] [Scilit]
  20. Ji, S.; Wang, W.; Wang, Y.; Bai, H.; Li, Z.; Huo, Z.; Luo, K. Quantitative Detection of Micro- and Nanoplastics (≥300 nm) in Human Urine Using Double-Shot Py-GC/MS with Internal Standard Calibration. Toxics 2025, 13, 452. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  21. Massardo, S.; Verzola, D.; Alberti, S.; Caboni, C.; Santostefano, M.; Verrina, E.; Angeletti, A.; Lugani, F.; Ghiggeri, G.M.; Bruschi, M.; et al. MicroRaman Spectroscopy Detects the Presence of Microplastics in Human Urine and Kidney Tissue. Environ. Int. 2024, 184, 108444. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  22. Wang, T.; Li, N.; Zhao, Z.; Zhang, Y.; Lin, L.; Ding, Y.; Feng, R.; Yang, Z.; Du, W.; Zeng, Y. Microplastic Contamination in Kidney Transplant Donor Tissues and Its Association with Early Post-Transplant Blood Pressure Changes in Recipients: A Multi-Modal Analytical Study. J. Hazard. Mater. 2025, 497, 139728. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  23. Deng, Y.; Sun, T.; Long, Z.; Luo, X.; Geng, X.; Lin, Z.; Chen, C.; Li, X. Microplastic Entry into Bloodstream via Hemodialysis: A Dual-Simulation Clinical Study. Environ. Pollut. 2026, 398, 128085. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  24. Marfella, R.; Prattichizzo, F.; Sardu, C.; Fulgenzi, G.; Graciotti, L.; Spadoni, T.; D’Onofrio, N.; Scisciola, L.; La Grotta, R.; Frigé, C.; et al. Microplastics and Nanoplastics in Atheromas and Cardiovascular Events. N. Engl. J. Med. 2024, 390, 900–910. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  25. Pironti, C.; Notarstefano, V.; Ricciardi, M.; Motta, O.; Giorgini, E.; Montano, L. First Evidence of Microplastics in Human Urine, a Preliminary Study of Intake in the Human Body. Toxics 2023, 11, 40. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  26. Song, X.; Chen, T.; Chen, Z.; Du, L.; Qiu, X.; Zhang, Y.; Li, Y.; Zhu, Y.; Tan, Z.; Mo, Y.; et al. Micro(Nano)Plastics in Human Urine: A Surprising Contrast between Chongqing’s Urban and Rural Regions. Sci. Total Environ. 2024, 917, 170455. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  27. De Oliveira, R.B.; Pelepenko, L.E.; Masaro, D.A.; Lustosa, G.M.M.M.; De Oliveira, M.C.; Roza, N.A.V.; Marciano, M.A.; Dos Reis, L.M.; Kamel, S.; Louvet, L.; et al. Effects of Microplastics on the Kidneys: A Narrative Review. Kidney Int. 2024, 106, 400–407. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  28. La Porta, E.; Exacoustos, O.; Lugani, F.; Angeletti, A.; Chiarenza, D.S.; Bigatti, C.; Spinelli, S.; Kajana, X.; Garbarino, A.; Bruschi, M.; et al. Microplastics and Kidneys: An Update on the Evidence for Deposition of Plastic Microparticles in Human Organs, Tissues and Fluids and Renal Toxicity Concern. Int. J. Mol. Sci. 2023, 24, 14391. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  29. Tan, R.-Y.; She, Q.-Y.; Ma, Y.-C.; Liu, M.-H.; Li, L.-J.; Huang, L.-L.; Zhong, Y.-W.; Bi, H.-X. The Threat of Microplastics to Human Kidney Health: Mechanisms of Nephrotoxicity and Future Research Directions. Environ. Res. 2025, 283, 122124. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  30. Roslan, N.S.; Lee, Y.Y.; Ibrahim, Y.S.; Tuan Anuar, S.; Yusof, K.M.K.K.; Lai, L.A.; Brentnall, T. Detection of Microplastics in Human Tissues and Organs: A Scoping Review. J. Glob. Health 2024, 14, 04179. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  31. Ragusa, A.; Svelato, A.; Santacroce, C.; Catalano, P.; Notarstefano, V.; Carnevali, O.; Papa, F.; Rongioletti, M.C.A.; Baiocco, F.; Draghi, S.; et al. Plasticenta: First Evidence of Microplastics in Human Placenta. Environ. Int. 2021, 146, 106274. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  32. Liu, S.; Guo, J.; Liu, X.; Yang, R.; Wang, H.; Sun, Y.; Chen, B.; Dong, R. Detection of Various Microplastics in Placentas, Meconium, Infant Feces, Breastmilk and Infant Formula: A Pilot Prospective Study. Sci. Total Environ. 2023, 854, 158699. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  33. Hunt, K.; Davies, A.; Fraser, A.; Burden, C.; Howell, A.; Buckley, K.; Harding, S.; Bakhbakhi, D. Exposure to Microplastics and Human Reproductive Outcomes: A Systematic Review. BJOG 2024, 131, 675–683. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  34. Guo, Y.; Rong, M.; Fan, Y.; Teng, X.; Jin, L.; Zhao, Y. The Presence of Microplastics in Human Semen and Their Associations with Semen Quality. Toxics 2025, 13, 566. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  35. Augey, L.; Berdougo-Tritz, J.; Mbarek, A.; Jadoul, M.; Massy, Z.; Deray, G.; Liabeuf, S. Effects of Microplastics and Nanoplastics on the Kidneys. Nephrol. Dial. Transplant. 2026, 41, 1412–1420. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  36. Holda, M.K.; Batko, J. Microplastics and Nanoplastics in Atheromas. N. Engl. J. Med. 2024, 390, 1726. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  37. Pan, Y.; Xu, S.; Yang, X. Standardizing Methodologies to Study Microplastics and Nanoplastics in Cardiovascular Diseases. Trends Endocrinol. Metab. 2025, 36, 105–108. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  38. Monikh, F.A.; Materić, D.; Valsami-Jones, E.; Grossart, H.-P.; Altmann, K.; Holzinger, R.; Lynch, I.; Stubenrauch, J.; Peijnenburg, W. Challenges in Studying Microplastics in Human Brain. Nat. Med. 2025, 31, 4034–4035. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  39. Campen, M.J.; West, A.B.; Garcia, M.; Gullapalli, R.; El Hayek, E. Reply to: Challenges in Studying Microplastics in Human Brain. Nat. Med. 2025, 31, 4036–4037. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  40. Weinstein, R.; Yuksel, Z.S.; Anderson, C.; Grant, D.; Milius, P.; Waldron, K.; David, L.A.; Molina, C.; Ruhoy, I.S.; Kaufman, D.L.; et al. Can Plasma Exchange Be Used to Lower the Circulating Burden of Microplastics in Human Patients? J. Clin. Apher. 2026, 41, e70135. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  41. Kara, E.; Konur, K.; Şentürk Koca, Y.; Aytan, Ü. Unveiling Hidden Contaminants: A Systematic Quantification and Characterization of Microplastics in Hemodialysis and Peritoneal Dialysis Fluids. BMC Nephrol. 2025, 26, 359. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  42. Choi, H.S.; Liu, W.; Misra, P.; Tanaka, E.; Zimmer, J.P.; Itty Ipe, B.; Bawendi, M.G.; Frangioni, J.V. Renal Clearance of Quantum Dots. Nat. Biotechnol. 2007, 25, 1165–1170. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  43. Wang, W.; Guan, J.; Feng, Y.; Nie, L.; Xu, Y.; Xu, H.; Fu, F. Polystyrene Microplastics Induced Nephrotoxicity Associated with Oxidative Stress, Inflammation, and Endoplasmic Reticulum Stress in Juvenile Rats. Front. Nutr. 2022, 9, 1059660. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  44. Shen, T.; Zhang, W.; Wang, Y.; Li, H.; Wu, J.; Wang, Q.; Qin, L.; Zhang, L.; Liu, C.; Li, R. Effects of Microplastic (MP) Exposure at Environmentally Relevant Doses on the Structure, Function, and Transcriptome of the Kidney in Mice. Molecules 2023, 28, 7104. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  45. Ronco, C.; Clark, W.R. Haemodialysis Membranes. Nat. Rev. Nephrol. 2018, 14, 394–410. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  46. Boschetti-de-Fierro, A.; Voigt, M.; Storr, M.; Krause, B. MCO Membranes: Enhanced Selectivity in High-Flux Class. Sci. Rep. 2015, 5, 18448. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  47. Weber, V.; Linsberger, I.; Rossmanith, E.; Weber, C.; Falkenhagen, D. Pyrogen Transfer across High- and Low-flux Hemodialysis Membranes. Artif. Organs 2004, 28, 210–217. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  48. Schepers, E.; Glorieux, G.; Eloot, S.; Hulko, M.; Boschetti-de-Fierro, A.; Beck, W.; Krause, B.; Van Biesen, W. Assessment of the Association between Increasing Membrane Pore Size and Endotoxin Permeability Using a Novel Experimental Dialysis Simulation Set-Up. BMC Nephrol. 2018, 19, 1. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  49. Lombardi, M.; Bergesio, F. Micro- and Nanoplastics and the Kidney: Exposure Pathways, Toxicokinetics, and Pathophysiological Insights. G. Clin. Nefrol. Dial. 2026, 38, 78–81. [Google Scholar] [CrossRef] [Scilit]
  50. Ávila, E.; Sepúlveda, R.A.; Retamal, J.; Hachim, D. Biocompatibility in Hemodialysis: Artificial Membrane and Human Blood Interactions. BMC Nephrol. 2025, 26, 482. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  51. Liu, L.; Xu, K.; Zhang, B.; Ye, Y.; Zhang, Q.; Jiang, W. Cellular Internalization and Release of Polystyrene Microplastics and Nanoplastics. Sci. Total Environ. 2021, 779, 146523. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  52. Firdessa, R.; Oelschlaeger, T.A.; Moll, H. Identification of Multiple Cellular Uptake Pathways of Polystyrene Nanoparticles and Factors Affecting the Uptake: Relevance for Drug Delivery Systems. Eur. J. Cell Biol. 2014, 93, 323–337. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  53. Richards, C.J.; Burgers, T.C.Q.; Vlijm, R.; Roos, W.H.; Åberg, C. Rapid Internalization of Nanoparticles by Human Cells at the Single Particle Level. ACS Nano 2023, 17, 16517–16529. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  54. Xiao, M.; Li, X.; Zhang, X.; Duan, X.; Lin, H.; Liu, S.; Sui, G. Assessment of Cancer-Related Signaling Pathways in Responses to Polystyrene Nanoplastics via a Kidney-Testis Microfluidic Platform (KTP). Sci. Total Environ. 2023, 857, 159306. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  55. Goodman, K.E.; Hua, T.; Sang, Q.-X.A. Effects of Polystyrene Microplastics on Human Kidney and Liver Cell Morphology, Cellular Proliferation, and Metabolism. ACS Omega 2022, 7, 34136–34153. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  56. Tenzer, S.; Docter, D.; Kuharev, J.; Musyanovych, A.; Fetz, V.; Hecht, R.; Schlenk, F.; Fischer, D.; Kiouptsi, K.; Reinhardt, C.; et al. Rapid Formation of Plasma Protein Corona Critically Affects Nanoparticle Pathophysiology. Nat. Nanotechnol. 2013, 8, 772–781. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  57. Bonventre, J.V. Primary proximal tubule injury leads to epithelial cell cycle arrest, fibrosis, vascular rarefaction, and glomerulosclerosis. Kidney Int. Suppl. 2014, 4, 39–44. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  58. Bhargava, P.; Schnellmann, R.G. Mitochondrial Energetics in the Kidney. Nat. Rev. Nephrol. 2017, 13, 629–646. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  59. Verzola, D.; Rumeo, N.; Alberti, S.; Loiacono, F.; La Maestra, S.; Passalacqua, M.; Artini, C.; Russo, E.; Verrina, E.; Angeletti, A.; et al. Coexposure to Microplastic and Bisphenol A Exacerbates Damage to Human Kidney Proximal Tubular Cells. Heliyon 2024, 10, e39426. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  60. Dou, L.; Poitevin, S.; Sallée, M.; Addi, T.; Gondouin, B.; McKay, N.; Denison, M.S.; Jourde-Chiche, N.; Duval-Sabatier, A.; Cerini, C.; et al. Aryl Hydrocarbon Receptor Is Activated in Patients and Mice with Chronic Kidney Disease. Kidney Int. 2018, 93, 986–999. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  61. Lombardi, M.; Bergesio, F. Micro- and Nanoplastics and the Kidney: Human Evidence, Clinical Vulnerabilities, Dialysis-Related Exposure, and CKD. G. Clin. Nefrol. Dial. 2026, 38, 82–88. [Google Scholar] [CrossRef] [Scilit]
  62. Bargman, J.M.; Thorpe, K.E.; Churchill, D.N. Relative Contribution of Residual Renal Function and Peritoneal Clearance to Adequacy of Dialysis: A Reanalysis of the CANUSA Study. J. Am. Soc. Nephrol. 2001, 12, 2158–2162. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  63. Canada-USA (CANUSA) Peritoneal Dialysis Study Group. Adequacy of Dialysis and Nutrition in Continuous Peritoneal Dialysis: Association with Clinical Outcomes. J. Am. Soc. Nephrol. 1996, 7, 198–207. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  64. Liao, C.-T.; Chen, Y.-M.; Shiao, C.-C.; Hu, F.-C.; Huang, J.-W.; Kao, T.-W.; Chuang, H.-F.; Hung, K.-Y.; Wu, K.-D.; Tsai, T.-J. Rate of Decline of Residual Renal Function Is Associated with All-Cause Mortality and Technique Failure in Patients on Long-Term Peritoneal Dialysis. Nephrol. Dial. Transplant. 2009, 24, 2909–2914. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  65. Aditya, M.R.; Muhammad, A.R.; Adriansyah, V.; Samsu, N.; Sulistomo, H.W. Microplastic Exposure and Its Consequences for Renal and Urinary Health: Systematic Review of in Vivo Studies. All Life 2025, 18, 2457760. [Google Scholar] [CrossRef] [Scilit]
  66. Xiong, X.; Gao, L.; Chen, C.; Zhu, K.; Luo, P.; Li, L. The Microplastics Exposure Induce the Kidney Injury in Mice Revealed by RNA-Seq. Ecotoxicol. Environ. Saf. 2023, 256, 114821. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  67. Meng, X.; Zhang, J.; Wang, W.; Gonzalez-Gil, G.; Vrouwenvelder, J.S.; Li, Z. Effects of Nano- and Microplastics on Kidney: Physicochemical Properties, Bioaccumulation, Oxidative Stress and Immunoreaction. Chemosphere 2022, 288, 132631. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  68. Wang, Y.-L.; Lee, Y.-H.; Hsu, Y.-H.; Chiu, I.-J.; Huang, C.C.-Y.; Huang, C.-C.; Chia, Z.-C.; Lee, C.-P.; Lin, Y.-F.; Chiu, H.-W. The Kidney-Related Effects of Polystyrene Microplastics on Human Kidney Proximal Tubular Epithelial Cells HK-2 and Male C57BL/6 Mice. Environ. Health Perspect. 2021, 129, 057003. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  69. Ahmed, Y.H.; El-Naggar, M.E.; Rashad, M.M.; Youssef, A.M.; Galal, M.K.; Bashir, D.W. Screening for Polystyrene Nanoparticle Toxicity on Kidneys of Adult Male Albino Rats Using Histopathological, Biochemical, and Molecular Examination Results. Cell Tissue Res. 2022, 388, 149–165. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  70. Lu, H.; Hou, L.; Zhang, Y.; Guo, T.; Wang, Y.; Xing, M. Polystyrene Microplastics Mediate Cell Cycle Arrest, Apoptosis, and Autophagy in the G2/M Phase through ROS in Grass Carp Kidney Cells. Environ. Toxicol. 2024, 39, 1923–1935. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  71. Cybulsky, A.V. Endoplasmic Reticulum Stress, the Unfolded Protein Response and Autophagy in Kidney Diseases. Nat. Rev. Nephrol. 2017, 13, 681–696. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  72. Meng, X.; Yin, K.; Zhang, Y.; Wang, D.; Lu, H.; Hou, L.; Zhao, H.; Xing, M. Polystyrene Microplastics Induced Oxidative Stress, Inflammation and Necroptosis via NF-κB and RIP1/RIP3/MLKL Pathway in Chicken Kidney. Toxicology 2022, 478, 153296. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  73. Liang, Y.; Liu, D.; Zhan, J.; Liu, X.; Li, P.; Ma, X.; Hou, H.; Wang, P. Polystyrene Microplastics Induce Kidney Injury via Gut Barrier Dysfunction and C5a/C5aR Pathway Activation. Environ. Pollut. 2024, 342, 122909. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  74. Liu, M.; Wang, M.; Sun, X.; Mu, J.; Teng, T.; Jin, N.; Song, J.; Li, B.; Zhang, D. Polypropylene Microplastics Triggered Mouse Kidney Lipidome Reprogramming Combined with ROS Stress as Revealed by Lipidomics and Raman Biospectra. Chemosphere 2025, 370, 143926. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  75. Luo, Y.; Long, M.; Wu, X.; Zeng, L. Targeting the NLRP3 Inflammasome in Kidney Disease: Molecular Mechanisms, Pathogenic Roles, and Emerging Small-Molecule Therapeutics. Front. Immunol. 2025, 16, 1703560. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  76. Chen, Y.-C.; Chen, K.-F.; Lin, K.-Y.A.; Chen, J.-K.; Jiang, X.-Y.; Lin, C.-H. The Nephrotoxic Potential of Polystyrene Microplastics at Realistic Environmental Concentrations. J. Hazard. Mater. 2022, 427, 127871. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  77. Jiang, N.; Zheng, X.; Zhang, N.; Cao, Y. The Detrimental Effects of Microplastic Exposure on Kidney Function. Front. Med. 2025, 12, 1620733. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  78. Zhang, M.; Xu, L. Transport of micro- and nanoplastics in the environment: Trojan-Horse effect for organic contaminants. Crit. Rev. Environ. Sci. Technol. 2022, 52, 810–846. [Google Scholar] [CrossRef] [Scilit]
  79. Hu, L.; Zhao, Y.; Xu, H. Trojan Horse in the Intestine: A Review on the Biotoxicity of Microplastics Combined Environmental Contaminants. J. Hazard. Mater. 2022, 439, 129652. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  80. Cambien, G.; Dupuis, A.; Guihenneuc, J.; Bauwens, M.; Belmouaz, M.; Ayraud-Thevenot, S. Endocrine Disruptors in Dialysis Therapies: A Literature Review. Environ. Int. 2023, 178, 108100. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  81. Akoueson, F.; Chbib, C.; Monchy, S.; Paul-Pont, I.; Doyen, P.; Dehaut, A.; Duflos, G. Identification and Quantification of Plastic Additives Using Pyrolysis-GC/MS: A Review. Sci. Total Environ. 2021, 773, 145073. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  82. Okoffo, E.D.; Ribeiro, F.; O’Brien, J.W.; O’Brien, S.; Tscharke, B.J.; Gallen, M.; Samanipour, S.; Mueller, J.F.; Thomas, K.V. Identification and Quantification of Selected Plastics in Biosolids by Pressurized Liquid Extraction Combined with Double-Shot Pyrolysis Gas Chromatography–Mass Spectrometry. Sci. Total Environ. 2020, 715, 136924. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  83. Wang, Q.; Bai, J.; Ning, B.; Fan, L.; Sun, T.; Fang, Y.; Wu, J.; Li, S.; Duan, C.; Zhang, Y.; et al. Effects of Bisphenol A and Nanoscale and Microscale Polystyrene Plastic Exposure on Particle Uptake and Toxicity in Human Caco-2 Cells. Chemosphere 2020, 254, 126788. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  84. Shi, Q.; Tang, J.; Wang, L.; Liu, R.; Giesy, J.P. Combined Cytotoxicity of Polystyrene Nanoplastics and Phthalate Esters on Human Lung Epithelial A549 Cells and Its Mechanism. Ecotoxicol. Environ. Saf. 2021, 213, 112041. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  85. Choi, D.; Bang, J.; Kim, T.; Oh, Y.; Hwang, Y.; Hong, J. In Vitro Chemical and Physical Toxicities of Polystyrene Microfragments in Human-Derived Cells. J. Hazard. Mater. 2020, 400, 123308. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  86. Tastet, V.; Le Vée, M.; Bruyère, A.; Fardel, O. Interactions of Human Drug Transporters with Chemical Additives Present in Plastics: Potential Consequences for Toxicokinetics and Health. Environ. Pollut. 2023, 331, 121882. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  87. Moon, J. Perfluoroalkyl Substances (PFASs) Exposure and Kidney Damage: Causal Interpretation Using the US 2003–2018 National Health and Nutrition Examination Survey (NHANES) Datasets. Environ. Pollut. 2021, 288, 117707. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  88. Lin, P.-I.D.; Cardenas, A.; Hauser, R.; Gold, D.R.; Kleinman, K.P.; Hivert, M.-F.; Calafat, A.M.; Webster, T.F.; Horton, E.S.; Oken, E. Per- and Polyfluoroalkyl Substances and Kidney Function: Follow-up Results from the Diabetes Prevention Program Trial. Environ. Int. 2021, 148, 106375. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  89. Watkins, D.J.; Josson, J.; Elston, B.; Bartell, S.M.; Shin, H.-M.; Vieira, V.M.; Savitz, D.A.; Fletcher, T.; Wellenius, G.A. Exposure to Perfluoroalkyl Acids and Markers of Kidney Function among Children and Adolescents Living near a Chemical Plant. Environ. Health Perspect. 2013, 121, 625–630. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  90. Huang, H.; Hou, J.; Li, M.; Wei, F.; Liao, Y.; Xi, B. Microplastics in the Bloodstream Can Induce Cerebral Thrombosis by Causing Cell Obstruction and Lead to Neurobehavioral Abnormalities. Sci. Adv. 2025, 11, eadr8243. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  91. Goedtke, L.; Sprenger, H.; Hofmann, U.; Schmidt, F.F.; Hammer, H.S.; Zanger, U.M.; Poetz, O.; Seidel, A.; Braeuning, A.; Hessel-Pras, S. Polycyclic Aromatic Hydrocarbons Activate the Aryl Hydrocarbon Receptor and the Constitutive Androstane Receptor to Regulate Xenobiotic Metabolism in Human Liver Cells. Int. J. Mol. Sci. 2021, 22, 372. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  92. Schroeder, J.C.; DiNatale, B.C.; Murray, I.A.; Flaveny, C.A.; Liu, Q.; Laurenzana, E.M.; Lin, J.M.; Strom, S.C.; Omiecinski, C.J.; Amin, S.; et al. The Uremic Toxin 3-Indoxyl Sulfate Is a Potent Endogenous Agonist for the Human Aryl Hydrocarbon Receptor. Biochemistry 2010, 49, 393–400. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  93. Ichii, O.; Otsuka-Kanazawa, S.; Nakamura, T.; Ueno, M.; Kon, Y.; Chen, W.; Rosenberg, A.Z.; Kopp, J.B. Podocyte Injury Caused by Indoxyl Sulfate, a Uremic Toxin and Aryl-Hydrocarbon Receptor Ligand. PLoS ONE 2014, 9, e108448. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  94. Massy, Z.A.; Drueke, T.B. Role of Uremic Toxins in Vascular Disease—The End of Nihilism? Kidney Int. 2022, 101, 1100–1102. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  95. Passos, R.S.; Davenport, A.; Busquets, R.; Selden, C.; Silva, L.B.; Baptista, J.S.; Barceló, D.; Campos, L.C. Microplastics and Nanoplastics in Haemodialysis Waters: Emerging Threats to Be in Our Radar. Environ. Toxicol. Pharmacol. 2023, 102, 104253. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  96. Liu, Y.; Faria, M.; Leonard, E. Spallation of Small Particles From Peristaltic Pump Tube Segments. Artif. Organs 2017, 41, 672–677. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  97. Leong, A.S.-Y.; Gove, D.W. Pathological Findings in Silicone Spallation in Vitro Studies. Pathology 1983, 15, 189–192. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  98. Barron, D.; Harbottle, S.; Hoenich, N.A.; Morley, A.R.; Appleton, D.; McCabe, J.F. Particle Spallation Induced by Blood Pumps in Hemodialysis Tubing Sets. Artif. Organs 1986, 10, 226–235. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  99. Bommer, J.; Pernicka, E.; Kessler, J.; Ritz, E. Reduction of Silicone Particle Release during Haemodialysis. Proc. Eur. Dial. Transpl. Assoc. 1984, 21, 287–290. [Google Scholar]
  100. Leong, A.S.-Y.; Disney, A.P.S.; Gove, D.W. Spallation and Migration of Silicone from Blood-Pump Tubing in Patients on Hemodialysis. N. Engl. J. Med. 1982, 306, 135–140. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  101. Bommer, J.; Weinreich, T.; Lovett, D.H.; Bouillon, R.; Ritz, E.; Gemsa, D. Particles From Dialysis Tubing Stimulate Interleukin-1 Secretion by Macrophages. Nephrol. Dial. Transplant. 1990, 5, 208–213. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  102. Bommer, J.; Gemsa, D.; Waldherr, R.; Kessler, J.; Ritz, E. Plastic Filing from Dialysis Tubing Induces Prostanoid Release from Macrophages. Kidney Int. 1984, 26, 331–337. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  103. Bommer, J.; Gemsa, D.; Kessler, J.; Ritz, E. Evidence for Macrophage Activation in Dialysis Patients Exposed to Silicone Filing. Nephron 1985, 39, 395–397. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  104. Li, B.; Li, M.; Du, D.; Tang, B.; Yi, W.; He, M.; Liu, R.; Yu, H.; Yu, Y.; Zheng, J. Characteristics and Influencing Factors of Microplastics Entering Human Blood through Intravenous Injection. Environ. Int. 2025, 198, 109377. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  105. Haq, Z.; Wang, X.; Cheng, Q.; Dias, G.F.; Moore, C.; Piecha, D.; Kotanko, P.; Ho, C.-H.; Grobe, N. Bisphenol A and Bisphenol S in Hemodialyzers. Toxins 2023, 15, 465. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  106. Mas, S.; Bosch-Panadero, E.; Abaigar, P.; Camarero, V.; Mahillo, I.; Civantos, E.; Sanchez-Ospina, D.; Ruiz-Priego, A.; Egido, J.; Ortiz, A.; et al. Influence of Dialysis Membrane Composition on Plasma Bisphenol A Levels during Online Hemodiafiltration. PLoS ONE 2018, 13, e0193288. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  107. Faouzi, M.A.; Dine, T.; Gressier, B.; Kambia, K.; Luyckx, M.; Pagniez, D.; Brunet, C.; Cazin, M.; Belabed, A.; Cazin, J.C. Exposure of Hemodialysis Patients to Di-2-Ethylhexyl Phthalate. Int. J. Pharm. 1999, 180, 113–121. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  108. Mettang, T.; Thomas, S.; Kiefer, T.; Fischer, F.-P.; Kuhlmann, U.; Wodarz, R.; Rettenmeier, A.W. Uraemic Pruritus and Exposure to Di(2-Ethylhexyl)Phthalate (DEHP) in Haemodialysis Patients. Nephrol. Dial. Transplant. 1996, 11, 2439–2443. [Google Scholar] [CrossRef] [Scilit]
  109. Ruiz, C.E.; Vela, L.; Nadal, M.; González, N.; Marcos, R.; Hernández, A.; Pastor, S.; Coll, E. DNA Damage and Bisphenol Levels in Chronic Kidney Disease Patients Undergoing Hemodialysis. J. Xenobiotics 2025, 15, 167. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  110. Green, J.L.; Field, D.T.; Bennett, R.; Jenner, L.C.; Chapman, E.C.; Sadofsky, L.R.; Rotchell, J.M.; Loubani, M. Microplastics in Cardiopulmonary Bypass: Quantification and Characterization of Particles across Systems. Interdiscip. Cardiovasc. Thorac. Surg. 2025, 40, ivaf080. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  111. Ronco, C.; Brendolan, A.; Lupi, A.; Metry, G.; Levin, N.W. Effects of a Reduced Inner Diameter of Hollow Fibers in Hemodialyzers. Kidney Int. 2000, 58, 809–817. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  112. Maduell, F.; Moreso, F.; Pons, M.; Ramos, R.; Mora-Macià, J.; Carreras, J.; Soler, J.; Torres, F.; Campistol, J.M.; Martinez-Castelao, A. High-Efficiency Postdilution Online Hemodiafiltration Reduces All-Cause Mortality in Hemodialysis Patients. J. Am. Soc. Nephrol. 2013, 24, 487–497. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  113. Blankestijn, P.J.; Vernooij, R.W.M.; Hockham, C.; Strippoli, G.F.M.; Canaud, B.; Hegbrant, J.; Barth, C.; Covic, A.; Cromm, K.; Cucui, A.; et al. Effect of Hemodiafiltration or Hemodialysis on Mortality in Kidney Failure. N. Engl. J. Med. 2023, 389, 700–709. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  114. Rama, I.; Llaudó, I.; Fontova, P.; Cerezo, G.; Soto, C.; Javierre, C.; Hueso, M.; Montero, N.; Martínez-Castelao, A.; Torras, J.; et al. Online Haemodiafiltration Improves Inflammatory State in Dialysis Patients: A Longitudinal Study. PLoS ONE 2016, 11, e0164969. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  115. La Porta, E.; Verzola, D.; Esposito, P.; Viazzi, F.; Cravedi, P.; Verrina, E.E. Peritoneal Infusion of Microplastics Induces Over-Expression of GDF-15 in Kidney and Heart in Mice. Nephrol. Dial. Transplant. 2025, 40, gfaf116.0735. [Google Scholar] [CrossRef] [Scilit]
  116. ISO 23500-5:2024; Preparation and Quality Management of Fluids for Haemodialysis and Related Therapies—Part 5: Quality of Dialysis Fluid for Haemodialysis and Related Therapies. 2nd ed. International Organization for Standardization (ISO): Geneva, Switzerland, 2024.
  117. United States Pharmacopeia (USP). General Chapter <788> Particulate Matter in Injections. In USP-NF; United States Pharmacopeia: Rockville, MD, USA, 2025. [Google Scholar] [CrossRef] [Scilit]
  118. European Directorate for the Quality of Medicines & HealthCare (EDQM). Particulate Contamination: Sub-Visible Particles, General Chapter 2.9.19. In European Pharmacopoeia, 10th ed.; Supplement 10.3; EDQM, Council of Europe: Strasbourg, France, 2021. [Google Scholar]
  119. Von Pawlowski, M.; Saur, L.; Röthele, J.; Pylatiuk, C. Microplastic Particles and Infusion Therapy—Evidence, Implications, and Unanswered Questions. J. Hazard. Mater. Adv. 2026, 21, 101068. [Google Scholar] [CrossRef] [Scilit]
  120. Marcos, R. Biomonitoring of Internal Exposure to MNPLs (Micro and NanoPlastics), and Its Effects, in Blood of Patients with Chronic Kidney Disease (CKD); ClinicalTrials.gov: Bethesda, MD, USA, 2024. Available online: https://clinicaltrials.gov/study/NCT05155267 (accessed on 11 June 2026).
  121. Campillo i López, F. It Is Time for Planetary Paediatrics. An. Pediatría (Engl. Ed.) 2022, 96, 177–178. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  122. Ortega-García, J.A.; Tellerías, L.; Ferrís-Tortajada, J.; Boldo, E.; Campillo-López, F.; van den Hazel, P.; Cortes-Arancibia, S.; Ramis, R.; Gaioli, M.; Monroy-Torres, R.; et al. Threats, Challenges and Opportunities for Paediatric Environmental Health in Europe, Latin America and the Caribbean. An. Pediatría (Engl. Ed.) 2019, 90, 124.e1–124.e11. [Google Scholar] [CrossRef] [Scilit]
  123. Yan, Z.; Shao, T. Chronic Inflammation in Chronic Kidney Disease. Nephron 2024, 148, 143–151. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  124. Ruiz-Ortega, M.; Rayego-Mateos, S.; Lamas, S.; Ortiz, A.; Rodrigues-Diez, R.R. Targeting the Progression of Chronic Kidney Disease. Nat. Rev. Nephrol. 2020, 16, 269–288. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  125. Quiroga, B. Strategies to Protect Dialysis Patients against Bisphenol A. Biomolecules 2021, 11, 1375. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  126. ISO/CD 23500-5; Preparation and Quality Management of Fluids for Haemodialysis and Related Therapies—Part 5: Quality of Dialysis Fluid for Haemodialysis and Related Therapies. 3rd ed. Committee Draft, under development. International Organization for Standardization (ISO): Geneva, Switzerland, 2026. Available online: https://www.iso.org/standard/90345.html (accessed on 16 June 2026).
  127. Vanholder, R.; Agar, J.; Braks, M.; Gallego, D.; Gerritsen, K.G.F.; Harber, M.; Noruisiene, E.; Pancirova, J.; Piccoli, G.B.; Stamatialis, D.; et al. The European Green Deal and Nephrology: A Call for Action by the European Kidney Health Alliance. Nephrol. Dial. Transplant. 2023, 38, 1080–1088. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  128. Kanwal, A.; Han, J.; Xu, J.; Xue, X.; Akhmedova, N.; Gao, Z.; Fu, R. Microplastics in Dialysis Effluent: Environmental Chemistry Insights and Metal-Organic Frameworks-Based Remediation Prospects. Environ. Toxicol. Pharmacol. 2026, 124, 105033. [Google Scholar] [CrossRef] [Scilit] [PubMed]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Article Metrics

Citations

Article Access Statistics

Multiple requests from the same IP address are counted as one view.