Abstract

Background: Clinically locating the point of no rotation to determine the subtalar joint axis location by applying pressure on the plantar surface of the foot was described by Kirby in 1987 but was never validated. We sought to extend a previously validated mechanical model to cadaver feet and to examine the intratester and intertester reliability. Methods: Four testers with different levels of experience determined the subtalar joint axis location and moved the subtalar joint through its range of motion, capturing the movement using kinematic analysis. The comparison of the spatial subtalar joint axis location as determined by palpation between and within testers determined the intertester and intratester reliability. The helical axis method was performed to validate the model. Results: The intrarater reliability varied from a high of α = 0.96 to a low of α = 0.26 for the slope and was, in general, high (α = 0.78–0.95) for the intersection. The interrater reliability scored moderate to high, depending on the specific cadaver specimen. Concerning the exact location of the subtalar joint axis, no significant difference was found between the results determined by different testers and the helical axis method. Conclusions: The palpation technique as part of the subtalar joint axis location and rotational equilibrium theory proposed by Kirby is a reliable and valid clinical tool. Experience in performing the palpation technique has a positive influence on the accuracy of the results. In the context of evidence-based practice, this technique could be a standard tool in the examination of patients with lower-limb–related pathologic disorders.
In 1977, Root, Orien, and Weed designed a biomechanical protocol believing that if the morphologic foot variations in the population could be named and meticulously classified, the foot and its function might be better understood.[1,2] Recent literature has, however, shown that the different goniometric measurements necessary to complete this biomechanical protocol have poor intertester and fair intratester reliability.[3] Furthermore, the actual subtalar joint (STJ) neutral position has never been validated.[1,4,5] In search of another, more valid and reliable, biomechanical protocol to examine foot form and function, a handful of new and existing models were put forward as a possible framework to perform a biomechanical examination, such as the Kirby model.[6] The aim of this project was to investigate whether the STJ axis (STJA) palpation technique—defined by a slope and intersection—projected in the plantar surface of the foot is a valid and reliable examination tool to locate the STJA. This is important to ascertain as today’s evidencebased medical environment demands research underpinning the effectiveness of a chosen diagnosis and treatment protocol.[7,8]

Methods

Cadaver Specimens

Convenience sampling was used to acquire three cadaver specimens (two unfixed and one embalmed with a phenol mixture/solution) obtained via the human body donation program of the Anatomy Division Kulak of the University of Leuven, Kortrijk, Belgium (Table 1). None of the cadaver specimens showed musculoskeletal pathologic abnormalities at the lower limb, and there were no signs of surgical interventions or fixed deformities of the foot and ankle.
Table 1. Cadaver Specimen Characteristics.
Procedure
Before data were collected, the cadaver specimens were prepared for analysis. A 15-minute manual mobilization of the foot and ankle was performed at room temperature. Stretching of the triceps surae muscle and mobilization of the subtalar, midtarsal, and intermetatarsal joints was performed, ensuring minimal influence of rigor mortis on the palpation technique and the motion analysis performed. Bone pins with reflective markers were inserted into the cadaver foot as described in part I (Figure 1). A sixcamera setup was arranged to track the motion using a video-based motion analysis system (Vicon MCam 460, six cameras, 120 Hz; Vicon Motion Systems Ltd, Oxford, England).
Figure 1. Examples of bone pin models used in this research without (A) and with (B) the clusters marked.
Data for the intratester and intertester reliability were obtained by four different testers (except for KULAK617, with only three testers), each with a different level of experience in examining a human foot and performing the palpation technique. Each tester performed six trials, the first three trials using the thumb, as described in the literature, and the last three trials using a blunt punch pin (punch pin technique), as suggested by De Schepper et al [9] to reduce the discrepancy between thumb size and pen point. Two points of no rotation were determined on the calcaneus (Figure 2). The line through these points was hypothesized to be the projection of the STJA on the plantar surface of the foot. Once a tester determined the locations of both points, these were indicated with a marker pen on the plantar surface of the foot, which was covered with transparent foil. Then, an image (parallel to the plantar surface) was taken using a digital camera (Canon EOS 30D; Canon USA Inc, Melville, New York), the foil was removed, and the next tester performed the palpation test.
Figure 2. Demonstration of the palpation technique perfomed with the punch pin.

Data Processing

The kinematic data collected using the reflective markers and the motion analysis system were analyzed in Matlab (The MathWorks Inc, Natick, Massachusetts) and were transformed to a two-dimensional screendump picture. These pictures were placed next to the digital pictures taken during the palpation procedure and were recalibrated and transposed to sets defining an intersection on the xaxis and a slope or angle with the x-axis, making comparison and further statistical analysis possible (Figure 3). This data processing procedure was described in detail in part I.[10]
Figure 3. Determination of the intersection and slope parameters on the plantar surface of the foot.

Statistical Analysis

All of the statistical tests were performed with a statistical analysis software program (SPSS version 17.0; SPSS Inc, Chicago, Illinois). All of the data points (83 slopes and 83 intersections) were taken and tested for normality using a Kolmogorov-Smirnov test and showed no significant difference with the normal distribution (Table 2). Based on this result, parametric statistical analysis tests were performed. Statistical significance for this study was set at the 95% confidence level (P ≤ .05).
Table 2. Test for Normality.
The coefficient internal consistency of the slope and intersection of the STJA was analyzed with the Cronbach α, which is commonly used as an estimate of the reliability of a test performed by a sample of participants, in this case, four testers. The SEM and SEM% were calculated to determine the range of certainty around a tester’s reported score. The SEM was calculated using pooled SDs of the three measurements for thumb and punch pin data:
Japma 104 00365 i010
The smallest detectable difference (SDD) and SDD% were used to measure a statistically significant change for each individual tester as a measure of the ability to detect clinically important changes in a specific condition, eg, the location of the STJA:
Japma 104 00365 i011
Validity was evaluated using a parametric univariate analysis of variance (general linear model) with a Bonferroni post hoc test assuming equal variance among the different testers. The slope and intersection were set as the dependent variables, with the testers being the fixed factor and the cadaver specimens the random factor.

Results

Reliability Analysis

The intraclass coefficient (ICC) of the STJA location determined with the thumb method varied from a high of α = 0.96 (P = .004) to a low of α = 0.26 (P = .358) for the slope and was, in general, high (α = 0.78 [P = .091] to 0.95 [P =.009]) for the intersection (Table 3). The intratester reliability of the determined STJA location using the punch pin technique varied from a high of α = 0.96 (P = .006) to a low of a 0.46 (P = .306) for the slope and was very high (α = 0.99 [P = .003] to 0.97 [P = .003]) for the intersection (Table 4).
Table 3. Intratester Reliability Statistics for the Slope and Intersection of the STJA Location Determined with the Thumb.
Table 4. Intratester Reliability Statistics for the Slope and Intersection of the STJA Location Determined with the Punch Pin.
The intertester reliability using the thumb method scored moderate to high, depending on the specific cadaver specimen. For KULAK605, α = 0.83 (P = .034) for the slope combined with α = 0.97 (P < .001) for the intersection. The reliability for KULAK617 was set at α = 0.90 (P = .03) for the slope combined with α = 0.68 (P = .16) for the intersection. Similar results were found for KULAK624, with α = 0.72 (P = .09) for the slope combined with α = 0.93 (P =.004) for the intersection (Table 5).
Table 5. Intertester Reliability Statistics for the Slope and Intersection of the STJA Location Determined with the Thumb.
The intertester reliability using the punch pin technique scored only moderate. For KULAK605, α = 0.61 (P = .15) for the slope combined with α = 0.96 (P = .001) for the intersection. The reliability for KULAK617 was set at α = 0.63 (P = .18) for the slope combined with α = 0.72 (P = .13) for the intersection. The reliability for KULAK624 was set at α = 0.90 (P = .009) for the slope combined with α = 0.94 (P = .003) for the intersection (Table 6).
Table 6. Intertester Reliability Statistics for the Slope and Intersection of the STJA Location Determined with the Punch Pin.
The mean slopes and intersections (six trials) for testers 1 to 4 and the kinematic analysis (Vicon) are visualized in Figure 4.
Figure 4. Overview of the determined mean subtalar joint axis location by testers (T) 1, 2, 3, and 4 and the kinematic analysis (Vicon) for specimens KULAK605 (A), KULAK617 (B), and KULAK624 (C).

Validity Analysis

In general, there was no significant difference in the results determined by different testers and the kinematic analysis when leaving the effect of the variance of the different cadaver specimens out of the analysis (Table 7). The kinematic analysis results are labeled as ‘‘tester 5’’ in Table 8 and Table 9, where the Bonferroni post hoc results are listed for the slope and intersection, respectively. Tester 3 was the only tester determining a significantly different slope compared with the kinematic analysis. As shown in Table 9, testers 1, 3, and 4 determined a translation along the x-axis, which was not significantly different from the location determined by the kinematic analysis. Tester 2 was the only tester determining a significant different intersection on the x-axis with the kinematic analysis.
Table 7. Univariate General Linear Model Test of Between-Subject Effects with the Dependent Variables Being Slope.
Table 8. Bonferroni Post Hoc Test for the Slope Based on Observed Means.
Table 9. Bonferroni Post Hoc Test for the Intersection Based on Observed Means and Intersection Based on Observed Means.

Discussion

Methodological Considerations

The fact that cadaver specimens were used could be considered an influencing factor to the data collected; however, this paper did not aim to investigate the normal values of STJ motion but wanted to investigate the validity and accuracy of the Kirby palpation technique to locate the axis of rotation. Despite the fact that cadaver specimens were used rather than volunteers, the obtained orientation of the STJA relative to the sagittal plane corresponds with the existing literature, with a mean deviation of 12.38 and a range of 58 to 178.[11,12,13,14] This correspondence was because we selected cadaver specimens with good subtalar mobility and preconditioned the specimens via manipulation of the joints, allowing determination of the axis of rotation via palpation.
We decided to determine two points on the plantar surface of the calcaneus because the STJ is located under the calcaneus, and any further palpation in the midfoot and forefoot would be a “projection” of the STJA and, thus, possibly more prone to error. However, the fact that the points were only approximately 5 cm apart increased the chance of error on the slope determined. Also, the fact that only two points were used increased the possibility of error. This choice was made because two points define a straight line, and determination of a straight line through three or four points would be clinically more subjective.

Reliability and Validity of the Kirby Palpation Technique

Despite these limitations, the reliability data reported in this study were comparable with the results reported by Payne et al [15] and De Schepper et al.[9] The overall reliability of the Kirby palpation technique compared with the reliability of the STJ measurements described by Root was similar for the intratester reliability level, where it ranged from α = 0.74 to 0.90.[16] Based on the results provided by Elveru et al,[16] the ICCs for intertester reliability were only α = 0.25 for measuring the STJ neutral position, α = 0.32 for STJ inversion, and α = 0.17 for STJ eversion.
Figure 4 and Figure 5 illustrate the positions of the lines as determined by the different testers on the three different specimens. The four mean locations determined by each tester are clearly clustered around the location determined by the kinematic analysis. Figure 5 shows the six individual marked positions for all of the testers, including the kinematic analysis (Vicon).
Figure 5. Overview of the determined subtalar joint axis location by testers (T) 1, 2, 3, and 4 and the kinematic analysis (Vicon) for specimens KULAK605 (A), KULAK617 (B), and KULAK 624 (C).
When discussing the results of the statistical analysis performed on the slope and intersection, the F-ratio of the slope was less than 1.0 (F = 0.226), indicating that even without significant differences among testers, the individual testers vary considerably in their determination of the slope. The low F-ratio could be due to the fact that the two points determined on the plantar surface of the heel were only approximately 5 cm apart. The variation for the intersection was less and was indicated by an F-ratio of 2.870.
Based on the post hoc Bonferroni results for the slope and intersection, testers 1 and 4—the most experienced testers—were able to determine the STJA without a significant difference with the kinematic analysis for the slope and intersection. The STJA location obtained by tester 2 displayed a significant difference with the kinematically determined intersection (translation) on the x-axis, and the data obtained by tester 3 displayed a significant difference with the kinematically determined slope (orientation).

Thumb Versus Punch Pin Technique

The difference between using a punch pin and using the thumb in determining the STJA location was small when looking at the range of the data collected. The largest effect was seen on the intersection data, and less on the slope data. The SEM and SDD increased for the more experienced testers (negative effect) and decreased for the inexperienced testers (positive effect) when using a punch pin. This result could be explained by the fact that the experienced testers were accustomed to performing the test with their thumbs and are now confronted with a new ‘‘feeling,’’ which takes time to get familiar with. The inexperienced testers got the immediate advantage of the decrease in discrepancy between the thumb and the actual location of the axis, making it easier to accurately locate the axis of rotation.
Clinically, performing the STJA palpation technique could be prone to error. For example, when a tester does not push parallel to the midsagittal plane of the patient, points of no rotation could be determined from a different angle, introducing an error in the exact location in the transverse plane. The importance of meticulously performing this clinical technique with respect to the correct plantar parallel position of the foot, the correct direction of the pressure, and the correct amount of loading force applied with the thumb is crucial in the reproduction of the STJA location. The differences among testers, visible in Figure 4 and Figure 5, could be, in part, explained by the inaccurate execution of each parameter necessary to perform the STJA palpation technique. Besides the possible errors directly related to execution of the palpation technique, a further influencing factor for KULAK605 could be that the testers had to use an extra bandage on the foot to place the OpSite foil (Medline Industries Inc, Mundelein, Illinois) owing to excess foot humidity and loosening of epidermis.

Conclusions

This study found no significant difference (P ≤ .05) 1between the location of the STJA (orientation and translation) determined by the different testers and that obtained by the three-dimensional kinematic model. Regarding precision, good to excellent reliability (Cronbach α = 0.8) for determination of the location of the STJA on an intratester level could be accepted only for testers 1 and 2. For the intertester reliability, the result was unequivocal because the Cronbach α was not consistently greater than 0.8 for slope (orientation) and intersection (translation) in the three cadaver specimens. The results concerning the reliability of the palpation technique in this study confirmed the previous research projects.[9,14]
Based on these results, we conclude that clinical determination of the location of the STJA in the transverse plane using the Kirby palpation technique is a valid and reliable method. Although there was some variation in precision between observers/ testers, experience did seem to have a clear influence. This is an important step forward as it equips the podiatric physician with a valuable clinical tool, helping to improve insight into the biomechanics of the foot and facilitating further clinical decision making in foot orthosis prescription and design.

Funding

None reported.

Conflicts of Interest

None reported.

References

  1. LEE, W.E. Podiatric biomechanics: an historical appraisal and discussion of the Root model as a clinical system of approach in the present context of theoretical uncertainty. Clin Podiatr Med Surg 2001, 18, 555. [Google Scholar] [PubMed]
  2. ROOT ML, ORIEN WP, WEED JH: Normal and Abnormal Function of the Foot, Clinical Biomechanics Corp, Los Angeles, 1977.
  3. MENZ, H.B. Clinical hindfoot measurement: a critical review of the literature. The Foot 1995, 5, 57. [Google Scholar] [CrossRef] [Scilit]
  4. PAYNE, C.B. The past, present, and future of podiatric biomechanics. JAPMA 1998, 88, 53. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  5. PAYNE, C.B. Should the baby be thrown out with the bathwater? Australas J Podiatr Med 1997, 31, 73. [Google Scholar]
  6. KIRBY, K.A. Methods for determination of positional variations in the subtalar joint axis. JAPMA 1987, 77, 228. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  7. PAYNE, C.B.; BIRD, A.R. Teaching clinical biomechanics in the context of uncertainty. JAPMA 1999, 89, 525. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  8. KIRBY KA: “Evolution of Foot Orthoses in Sports,” in Athletic Footwear and Orthoses in Sports Medicine, edited by MB Werd, EL Knight, p 19, Springer Science, New York, 2010.
  9. DE SCHEPPER J, VAN ALSENOY K, RIJCKAERT J, ET AL: Intratest reliability in determining the subtalar joint axis using the palpation technique described by K. Kirby. JAPMA 102: 122, 2012.
  10. VAN ALSENOY KK, DE SCHEPPER J, SANTOS D, ET AL: The Subtalar Joint Axis Palpation Technique—Part 1. JAPMA 104: 238, 2014.
  11. LEWIS GS, COHEN TL, SEISLER AR, ET AL: In vivo tests of an improved method for functional location of the subtalar joint axis. J Biomech 42: 146, 2009.
  12. LEARDINI, A.; STAGNI, R.; O’CONNOR, J.J. Mobility of the subtalar joint in the intact ankle complex. J Biomech 2001, 34, 805. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  13. LUNDBERG, A.; SVENSSON, O.K. The axes of rotation of the talocalcaneal and talonavicular joints. The Foot 1993, 3, 65. [Google Scholar] [CrossRef] [Scilit]
  14. VAN LANGELAAN, E.J. A kinematical analysis of the tarsal joints: an X-ray photogrammetric study. Acta Orthop Scand Suppl 1983, 204, 1. [Google Scholar]
  15. PAYNE, C.; MUNTEANU, S.; MILLER, K. Position of the subtalar joint axis and resistance of the rearfoot to supination. JAPMA 2003, 93, 131. [Google Scholar] [CrossRef] [Scilit]
  16. ELVERU, R.A.; ROTHSTEIN, J.M.; LAMB, R.L. Goniometric reliability in a clinical setting: subtalar and ankle joint measurements. Phys Ther 1988, 68, 672. [Google Scholar] [CrossRef] [Scilit] [PubMed]