Real-Time SpO2 Estimation from Chest Reflectance Photoplethysmography: Algorithm Development and Clinical Validation Against Arterial SaO2

,
,
,
,
,
and
1
Department of Development, MEZOO Co., Ltd., 200, Gieopdosi-ro, Jijeong-myeon, Wonju-si 26354, Republic of Korea
2
Department of Anesthesiology and Pain Medicine, Asan Medical Center, University of Ulsan College of Medicine, Seoul 05505, Republic of Korea
3
Department of Clinical Pharmacology and Therapeutics, Asan Medical Center, University of Ulsan College of Medicine, Seoul 05505, Republic of Korea
*
Author to whom correspondence should be addressed.
This article belongs to the Special Issue Advances in Biosignal Sensing and Signal Processing

Abstract

Peripheral oxygen saturation (SpO2) estimation using photoplethysmography (PPG) is typically based on the pulsatile components of red and infrared (IR) PPG signals acquired from peripheral sites, such as the finger or earlobe. Although these sites provide strong PPG signals, they are less suitable for integrated monitoring with electrocardiography (ECG), including arrhythmia detection. Simultaneous acquisition of PPG and ECG from a chest-worn device may therefore enable continuous and integrated physiological monitoring. However, chest PPG is characterized by lower perfusion and greater susceptibility to respiratory and motion artifacts, making reliable extraction of pulsatile red and IR components challenging. In this study, we developed and clinically evaluated a real-time SpO2 estimation algorithm based on chest reflectance PPG. Green PPG, which provides a relatively distinct pulsatile waveform under low-perfusion conditions, was used for beat detection. A heart-rate-adaptive analysis window, normalized cross-correlation-based signal quality assessment, and least-squares estimation of the AC amplitude ratio were combined to calculate the red-to-IR ratio of ratios, from which SpO2 was estimated using a calibration equation. Algorithm performance was evaluated against reference arterial oxygen saturation (SaO2) measured using a blood gas analyzer. The proposed method achieved an accuracy root mean square of 2.91% relative to the reference SaO2. These results support the technical feasibility of chest reflectance PPG-based SpO2 estimation and its accuracy under controlled desaturation conditions.

1. Introduction

Recent advances in wearable technology have focused on the simultaneous measurement of multiple physiological parameters, including heart rate, respiration, oxygen saturation, and electrocardiogram (ECG), using a single wearable device [1,2]. Integrating multiple physiological measurements into a single device can reduce the burden of sensor attachment and user discomfort compared with measuring each parameter separately, while providing complementary physiological information. Therefore, such technologies have substantial potential for applications in daily health management, remote patient monitoring, and long-term continuous monitoring.
Among these physiological parameters, oxygen saturation is particularly relevant to continuous cardiopulmonary monitoring because pulse oximetry provides a noninvasive estimate of arterial oxygen saturation using photoplethysmography (PPG). PPG-based pulse oximetry is well suited to wearable implementation because of its noninvasive nature, compatibility with compact optical sensors, and ability to be integrated with other physiological measurements [3]. In particular, incorporating SpO2 estimation into a chest-worn platform that already acquires ECG and other cardiorespiratory signals may enable multimodal monitoring from a single anatomical site and reduce reliance on additional peripheral sensors.
Previous studies on PPG-based oxygen saturation measurement and commercial pulse oximeters have primarily focused on peripheral sites, such as the finger, earlobe, and wrist, where PPG signals can be acquired with relatively large amplitudes and high stability [4]. At these sites, peripheral oxygen saturation (SpO2) is conventionally estimated by calculating the ratio of ratios (R) from the pulsatile (AC) and direct-current (DC) components of red and IR PPG signals and subsequently applying an empirically derived calibration equation [4,5]. Unlike transmission-mode pulse oximetry, which requires light to pass through the tissue, reflectance-mode pulse oximetry allows the light source and photodetector to be positioned on the same surface. This configuration enables measurements at various anatomical sites, including the foot, forehead, wrist, finger-ring interface, and chest [6,7]. Accordingly, with the growing demand for wearable sensors, the range of applications for reflectance PPG-based oxygen saturation measurement has continued to expand. Wrist-worn reflectance pulse oximeters have demonstrated utility for monitoring oxygen saturation during sleep and assessing sleep apnea [8]. In addition, ring-type devices have shown the potential for high-sensitivity and low-power measurements through optimization of their optical configurations [9]. These developments have stimulated active research into reflectance-mode wearable devices with diverse form factors. Recent studies have also highlighted the chest as a promising site for multimodal physiological monitoring. Long et al. demonstrated photonic sensing of seismocardiographic signals [10], and Qiu et al. developed a multimodal chest patch for simultaneous ECG, heart-sound, and respiratory monitoring [11]. These studies further support the potential of chest-worn platforms for integrated cardiorespiratory monitoring.
However, peripheral-site measurements have limitations for long-term, continuous, and unobtrusive monitoring. In cold environments or under conditions of impaired peripheral circulation, reduced PPG signal amplitude may compromise measurement reliability [12]. In particular, finger-based oxygen saturation monitoring may restrict user activity during prolonged wear, while motion during daily activities may degrade signal quality [13]. In contrast, the chest is suitable for long-term monitoring using a wearable patch and has a relatively limited effect on user activity [14]. Moreover, because ECG and PPG signals can be acquired simultaneously from the chest using a single device, the chest may serve as an advantageous measurement site for the integrated monitoring of oxygen saturation and ECG [15].
Nevertheless, important technical challenges remain in estimating SpO2 using chest reflectance PPG. Compared with peripheral sites such as the finger and earlobe, the chest generally exhibits lower perfusion, resulting in smaller pulsatile components in the red and IR PPG signals. Chest PPG is also highly susceptible to motion artifacts caused by respiration and body movement. Consequently, stable estimation of the AC components using conventional peak- and valley-detection methods is difficult, which may reduce the accuracy of the R calculation and subsequent SpO2 estimation [15,16,17]. In particular, a previous study [15] demonstrated the feasibility of sternum-based reflectance pulse oximetry; however, that study primarily focused on offline analysis and feature extraction. Real-time motion artifact detection, signal quality assessment, and stabilization of SpO2 output, which are essential for practical wearable devices, were not sufficiently addressed.
In this study, we developed a real-time SpO2 estimation algorithm based on reflectance PPG signals acquired using a chest-worn device (HiCardi M350; MEZOO Co., Ltd., Wonju-si, Gangwon-do, Republic of Korea) and implemented the algorithm in the M350 device. We also conducted a controlled desaturation study involving human participants and evaluated the accuracy of the proposed algorithm by comparison with reference SaO2 values obtained using an arterial blood gas analyzer. The study was designed in accordance with ISO 80601-2-61:2017, which was applicable at the time of study design, and with reference to FDA guidance for pulse oximeters. A prespecified accuracy root mean square ( A rms ) criterion of 4.0% or less was used for interpretation of algorithm accuracy under the controlled study conditions. To compensate for the instability of red and IR peak and valley detection caused by low chest perfusion and respiratory and motion artifacts, the proposed algorithm integrates green PPG-based beat detection, a heart-rate-adaptive analysis window, a normalized cross-correlation-based signal quality index, least-squares-based AC component estimation, and real-time post-processing. In contrast to previous sternum-based reflectance PPG studies that primarily focused on offline analysis and the feasibility of feature extraction, the present study implemented a complete real-time SpO2 estimation pipeline for use in a practical wearable device.

2. Methods

2.1. Study Design

This study was conducted as a clinical investigation to develop a real-time method for estimating SpO2 from a chest-worn reflectance PPG device and to evaluate its accuracy. The clinical investigation was designed as a controlled desaturation study in accordance with ISO 80601-2-61:2017 and consisted of two stages with non-overlapping participant cohorts. Stage 1 involved five healthy adult volunteers and was used for algorithm development and finalization of the calibration model. The final calibration model was locked before Stage 2 and was not modified using Stage 2 data. Stage 2 involved 12 healthy adult volunteers and was used for pivotal accuracy evaluation. In total, 17 participants were enrolled across the two stages, and the measurement procedures were identical. Because both stages were conducted within the same overall clinical investigation, the Stage 2 evaluation should be regarded as an independent within-study accuracy evaluation rather than as external validation.

2.2. Measurement Devices

The HiCardi M350 (HiCardi M350; MEZOO Co., Ltd., Wonju-si, Gangwon-do, Republic of Korea) used in this study is a chest-worn patch-type patient monitoring device designed to simultaneously acquire green, red, and IR PPG signals, ECG, respiration, and accelerometer signals. After the electrodes had been attached, the device was affixed to the chest, as illustrated in Figure 1A. The sensor was positioned on the left anterior chest in alignment with the participant’s left shoulder line and secured to ensure stable contact of the electrodes and optical sensor with the skin (Figure 1B). The M350 acquires reflectance-mode PPG signals, with the red and IR PPG signals used for SpO2 estimation in this study. The green PPG, ECG, and accelerometer signals were used as auxiliary signals for beat detection, signal quality assessment, and motion artifact detection, respectively.
Figure 1. Measurement device and sensor placement. (A) Front and back views of the HiCardi M350 device and electrode attachment configuration. (B) Placement of the HiCardi M350 on the left anterior chest according to the left shoulder line. (C) Radiometer ABL90 FLEX PLUS blood gas and CO-oximetry analyzer used to determine the reference arterial oxygen saturation (model 393-092; Radiometer Medical ApS, Brønshøj, Denmark).
The SpO2 processing algorithm was implemented directly on the M350 using an nRF52832 microcontroller (Nordic Semiconductor ASA, Trondheim, Norway) with a 64-MHz ARM Cortex-M4F processor. At the PPG sampling rate of 125 Hz, corresponding to an 8-ms sampling interval, the measured worst-case execution time during SpO2 processing was approximately 1.6 ms, corresponding to 20.0% of the available sampling interval and leaving a timing margin of approximately 6.4 ms. The compiled firmware including the SpO2 processing functionality occupied 140.47 kB of flash memory and 52.33 kB of RAM, with incremental SpO2-related memory requirements of approximately 45.37 kB of flash memory and 22.80 kB of RAM. These measurements indicate that the SpO2 processing can be executed within the timing and memory constraints of the M350 embedded hardware.
Reference oxygen saturation was defined as SaO2 measured from arterial blood using an ABL90 FLEX PLUS blood gas and CO-oximetry analyzer (model 393-092; Radiometer Medical ApS, Brønshøj, Denmark), as shown in Figure 1C.

2.3. Participants

The final analysis included 12 healthy adult volunteers, comprising six men and six women. The mean age was 28.9 ± 4.5 years, the mean height was 171.7 ± 9.8 cm , and the mean body weight was 70.0 ± 11.6 kg . Before participation, all participants received a detailed explanation of the study objectives, measurement procedures, anticipated risks, and criteria for study termination, and provided written informed consent. The study was conducted in accordance with procedures approved by the Institutional Review Board (IRB No. 2025-1580). The inclusion criteria were as follows: men and women aged 20 to 49 years; carboxyhemoglobin < 3 % ; methemoglobin < 2 % ; total hemoglobin concentration > 10 g / dL ; voluntary agreement to participate in the clinical study with written informed consent; and the ability to understand the study instructions and participate throughout the entire study period.

2.4. Controlled Desaturation Protocol

To induce controlled hypoxemia, each participant’s nostrils were occluded with a nose clip, and the participant breathed through a mouthpiece connected to an open breathing circuit. A gas mixture consisting of medical air, N2, and CO2 was administered, and the inspired oxygen concentration was progressively reduced by adjusting the flow rates of nitrogen and medical air. CO2 was included to minimize respiratory alkalosis caused by hyperventilation during hypoxemia induction. Airway gases were continuously analyzed using a mass spectrometer, and the gas flow rates were adjusted to achieve and maintain each target oxygen saturation plateau.
A Nellcor OxiMax N-560 pulse oximeter sensor (N-560; Covidien LLC, Mansfield, MA, USA) was attached to the participant’s ring finger to monitor attainment and stability of the target plateaus in real time. The N-560 SpO2 measurements were used solely to guide adjustment of the gas mixture and were not used as reference values for the accuracy analysis. Reference oxygen saturation was defined exclusively as arterial SaO2 measured using the ABL90 FLEX PLUS blood gas and CO-oximetry analyzer (Figure 1C). Before arterial catheterization, Allen’s test was performed to confirm adequate collateral circulation. Collateral circulation was considered adequate when normal palm color returned within 10 s after release of ulnar artery compression. Following local anesthesia with 0.5 mL of lidocaine, a 22-gauge catheter was inserted into the radial artery for serial arterial blood sampling.
Hypoxemia was induced stepwise over an SpO2 range of 70–100%, with target plateaus of 94, 90, 85, 80, 75, and 70%. The desaturation procedure was performed twice, with 100% oxygen administered between rounds until oxygen saturation had sufficiently recovered. All procedures, including arterial blood sampling, were conducted under the supervision of trained medical personnel, including an anesthesiologist. Oxygen saturation, heart rate, vital signs, and symptoms associated with hypoxemia or arterial catheterization were continuously monitored. If clinically significant symptoms or vital-sign abnormalities occurred, the procedure was discontinued immediately, 100% oxygen was administered, and appropriate treatment was provided. All adverse events were documented and classified as mild, moderate, or severe according to their functional impact and the required medical intervention.
A total of 25 arterial blood samples, corresponding to approximately 50 mL of blood, were collected from each participant. These comprised one sample obtained during baseline normoxia (S1), 12 samples collected during the first desaturation round (S2–S13), and 12 samples collected during the second round (S14–S25). During each round, two arterial blood samples were obtained at each of the six target plateaus, as illustrated in Figure 2. Sampling was initiated only after both the breath-by-breath SaO2 estimated using the Severinghaus equation and the SpO2 measured using the N-560 had stabilized within the target range. The first arterial blood sample at each plateau was collected at least 30 s after the target plateau had been reached, and consecutive samples obtained at the same plateau were separated by an interval of at least 20 s.
Figure 2. Controlled desaturation protocol and arterial blood sampling schedule. Six target oxygen saturation plateaus (94, 90, 85, 80, 75, and 70%) were sequentially induced during each of two desaturation rounds, with recovery to normoxia between rounds. At each plateau, two arterial blood samples were collected: the first at least 30 s after stabilization and the second after an additional interval of at least 20 s. The blue line indicates the target oxygen saturation, and the red circles indicate arterial blood sampling time points.

2.5. Overall Processing Steps

The overall SpO2 estimation pipeline is illustrated in Figure 3. All processing stages were designed to minimize computational complexity and output latency, enabling real-time SpO2 monitoring. The M350 simultaneously acquired green, red, and IR PPG signals, together with ECG, accelerometer, and respiration signals. However, the respiration signal was not used in the proposed algorithm. The acquired signals underwent preprocessing, PPG and ECG peak detection, motion artifact detection, HR-adaptive windowing, and normalized cross-correlation-based signal quality assessment. For valid PPG beats, ensemble averaging and least-squares-based AC component estimation were applied to calculate the R. Beat-wise SpO2 values were then obtained using the calibration equation followed by post-processing.
Figure 3. Overall pipeline of the proposed chest reflectance PPG-based SpO2 estimation algorithm. Raw green, red, and IR PPG signals, together with accelerometer data, are first processed by DC offset calibration, filtering, and motion detection. Green PPG is used for peak detection, pulse-rate estimation, and RR-interval consistency checking. Beat-wise SpO2 estimation is then performed using an HR-adaptive analysis window, NCC-based signal quality assessment, beat ensemble averaging, least-squares AC estimation, and calibration of the ratio-of-ratios value. The final SpO2 output is generated after low-SQI output hold, median and moving-average filtering, and hypoxia-sensitive dual-track post-processing.

2.5.1. Signal Preprocessing

The green, red, and IR PPG signals and the ECG signal were acquired at a sampling rate of 125 Hz. All subsequent signal processing was performed at the original sampling rate without resampling. DC offset calibration was first applied to the raw PPG signals to correct channel-specific baselines. In chest reflectance PPG, the DC level of each wavelength channel may vary depending on sensor–skin contact, skin-surface curvature, sensor attachment pressure, ambient light, and differences in the optical paths between the LEDs and photodiode [17,18]. Because these channel-dependent DC offsets may affect AC component estimation and comparisons among wavelength channels, the baseline of each channel was corrected before subsequent signal processing.
Following DC offset calibration, sequential digital filtering was performed to obtain waveforms suitable for beat-wise signal analysis and SpO2 estimation. Previous studies have reported that band-pass filtering within the 0.5–8 Hz range effectively suppresses baseline wander and high-frequency components that are not relevant to PPG analysis, whereas moving-average filtering is useful for reducing short-term signal fluctuations [17,19]. In pulse oximetry, the filter bandwidth of the photodiode signal may affect not only noise suppression but also pulse-waveform morphology and ratio-of-ratios-based SpO2 estimation. Therefore, preprocessing is required to remove unwanted frequency components while preserving the pulsatile component [20].
In this study, a second-order finite impulse response (FIR) high-pass filter with a cutoff frequency of 0.8 Hz was applied to the PPG signals to suppress low-frequency drift caused by respiratory variation, changes in sensor–skin contact, and baseline wander. A second-order FIR band-stop filter centered at 60 Hz was subsequently applied to remove power-line interference [21]. A second-order FIR low-pass filter with a cutoff frequency of 8 Hz was then applied to attenuate noise at frequencies above the pulsatile component [19,21]. Finally, a 25-point moving-average filter was applied to reduce short-term signal fluctuations and provide stable input signals for subsequent beat-wise analysis and SpO2 estimation [19]. The ECG signal was processed using a 0.5–15 Hz band-pass filter to suppress baseline drift and high-frequency noise while preserving the cardiac waveform required for subsequent R-peak detection.

2.5.2. Motion Artifact Detection

In chest-worn reflectance PPG, body movement, respiratory chest-wall motion, changes in sensor attachment pressure, and variations in sensor–skin contact can distort the PPG waveform [3,22]. These disturbances can destabilize the pulsatile components of the red and IR PPG signals, thereby affecting the accuracy of peak and valley detection and AC component estimation. Moreover, failure to adequately detect movement artifacts in pulse oximetry recordings may cause transient decreases in SpO2 to be misinterpreted as hypoxemia, even in the absence of true oxygen desaturation [23]. Therefore, an accelerometer (ACC)-based motion artifact detection step was incorporated into the proposed algorithm to improve the reliability of SpO2 estimation.
Triaxial ACC signals ( a x , a y , and a z ) were acquired at a sampling rate of 25 Hz and processed at the original sampling rate without resampling. A second-order IIR band-pass filter, implemented in direct-form II transposed, was applied to the raw signal from each axis, with lower and upper cutoff frequencies of 0.5 Hz and 3 Hz, respectively. This filtering step was performed to suppress the gravitational component and low-frequency variations associated with gradual postural changes while retaining dynamic acceleration components caused by body movement, upper-body motion, and changes in sensor contact. The vector magnitude of the filtered triaxial ACC signals was then calculated and used as a single motion reference signal with reduced dependence on sensor orientation [24,25]:
A mag ( n ) = a x ( n ) 2 + a y ( n ) 2 + a z ( n ) 2 .
Here, a x ( n ) , a y ( n ) , and a z ( n ) represent the filtered acceleration signals along the three axes at the nth sample. The difference in acceleration magnitude between consecutive samples was calculated and used as a signed motion index reflecting rapid changes in movement [26]. When the motion index fell outside the range 20.06 to + 20.06 ADC counts for five consecutive samples, the motion flag was set and the corresponding interval was classified as containing motion artifacts; the flag was cleared after five consecutive samples within the range.
Intervals classified as containing motion artifacts were excluded from subsequent beat validation and SpO2 estimation. During real-time monitoring, updates of the SpO2 estimate were suspended while motion artifacts were detected, and the most recent valid SpO2 value was temporarily retained. This procedure was intended to reduce transient SpO2 output errors caused by body movement and changes in sensor contact.

2.5.3. PPG Peak Detection

PPG-based SpO2 estimation requires reliable segmentation of the red and IR PPG signals into individual cardiac cycles. In this study, the green PPG signal was used as the reference signal for defining the beat-wise analysis windows. Because green light has a shorter wavelength than red and IR light, green PPG is relatively more sensitive to pulsatile blood-volume changes in superficial vascular layers and is therefore advantageous for cardiac-cycle detection [27,28,29]. Green PPG has also been used as a reference waveform for beat-quality assessment and outlier rejection of red and IR PPG signals in a previous study of sternal reflectance pulse oximetry [15]. A recent study of 24 healthy participants also reported higher green-wavelength PPG signal quality than that of red and IR PPG at core-body sites, including the sternum, supporting the use of green PPG as a reference waveform for chest-based beat detection [30]. In contrast, although red and IR PPG signals are essential for SpO2 estimation because they reflect differences in light absorption between oxygenated and deoxygenated hemoglobin, their pulsatile components may have low amplitudes at low-perfusion reflectance measurement sites such as the chest. Consequently, peak and valley detection may become unstable because of respiratory and motion-related disturbances [15]. Therefore, the beat locations detected from the green PPG signal were used to align the red and IR PPG signals into corresponding cardiac cycles.
To quantitatively compare pulsatile signal amplitude across wavelengths, the perfusion index (PI) was calculated for green, red, and IR PPG signals in all 12 participants during the controlled desaturation protocol. Signals were analyzed in 10-s windows with 5-s overlap. For each channel, the DC component was defined as the mean of the DC-corrected signal, whereas the AC component was estimated from the 0.5–5 Hz band-pass-filtered signal as the 95th–5th percentile difference. PI was calculated as 100 × A C / D C . Windows containing signal saturation, low green-channel pulse-spectrum SNR, or PI values exceeding 20% were excluded. For each participant, the median PI across valid windows was calculated for each wavelength, and green-versus-red and green-versus-IR differences were evaluated using paired Wilcoxon signed-rank tests. This supplementary analysis was performed to provide quantitative support for the predefined use of green PPG as the reference channel for cardiac-cycle detection.
For beat detection, peaks and valleys in the green PPG signal were detected using an algorithm based on adaptive threshold detection (ADT) [31]. ADT identifies candidate peaks and valleys using a threshold that is dynamically updated according to changes in PPG amplitude. As illustrated in Figure 4, the threshold was designed to track the waveform, and points at which the direction of the signal changed were identified as candidate peaks or valleys. This approach is more robust to baseline and amplitude variations than fixed-threshold or simple local-extrema detection.
Figure 4. Adaptive threshold-based PPG peak and valley detection. The dynamic thresholds for V max and V min track the PPG waveform to identify candidate peaks and valleys. Candidate detections occurring within the refractory period are rejected to prevent physiologically implausible duplicate beat detections.
To enable real-time processing, the detection algorithm was implemented in a sample-by-sample manner using only the current input sample and state variables updated from the preceding beat. An adaptive refractory rule was applied to remove duplicate detections occurring at short intervals: rather than a fixed post-detection blanking interval, a candidate beat was rejected when the resulting R-R interval fell outside 0.50 to 1.50 times the running mean R-R interval, so that the effective refractory limit adapts to the prevailing pulse rate; before a running mean was established, absolute limits of 0.5 and 1.5 s (120 and 40 bpm) were applied. Each candidate location was corrected by searching for the actual local maximum or minimum within a limited surrounding interval. If no beat was detected for a predefined period, a back-search was performed within a limited interval based on the most recent valid pulse interval. Intervals classified as unreliable by motion artifact detection or the signal quality index (SQI) were excluded from the final SpO2 calculation. The resulting green PPG beat locations were used to define the beat-wise analysis windows for the red and IR PPG signals.

2.5.4. ECG R-Peak Detection

The ECG signal was used as an auxiliary signal to assess the temporal consistency of green PPG-based beat detection and to provide reference heart-rate information for defining the HR-adaptive analysis window. Motion-induced changes in blood volume at the measurement site can cause motion artifacts and pulse-detection errors in reflectance PPG [32], and ECG R-peaks have been used as reference fiducial points for evaluating PPG peak-detection performance [33]. Cardiac rhythm information derived from ECG has also been applied to the processing and reconstruction of PPG signals affected by motion artifacts [34]. Based on these findings, heart-rate information derived from ECG R-peaks was used as an auxiliary reference to assess the plausibility of PPG beat timing and to establish stable analysis windows.
For integration into the real-time beat-wise SpO2 estimation framework, an ECG R-peak detection algorithm was implemented based on an angle-based QRS detection method [35]. This method identifies candidate R-peak regions using the rapid slope changes characteristic of the QRS complex. For each incoming ECG sample, slope-based features and an adaptive threshold were sequentially updated, and regions exceeding the threshold were identified as R-peak candidates. The final R-peak location was then refined by searching for the maximum amplitude of the original ECG signal within a limited local interval surrounding each candidate. All operations were performed in a sample-by-sample manner using a fixed-length buffer to minimize computational burden and processing latency.
Instantaneous heart rate was calculated from the detected ECG R-peaks and used to assess the physiological plausibility and temporal consistency of the pulse intervals detected from the green PPG signal. ECG R-peaks were not directly used to estimate the AC or DC components of the red and IR PPG signals; rather, they served as an auxiliary reference for defining the HR-adaptive analysis window and identifying abnormal PPG beat intervals. This approach reduced the influence of low perfusion, respiratory modulation, and transient waveform distortion on PPG beat segmentation, thereby providing stable analysis intervals for real-time beat-wise SpO2 estimation.

2.5.5. HR-Adaptive Analysis Window

Reliable definition of the analysis interval corresponding to each pulse cycle is essential for PPG-based SpO2 estimation. In conventional pulse oximetry, the pulsatile AC components are typically estimated by directly detecting peaks and valleys in the red and IR PPG signals [4,5]. However, at low-perfusion reflectance measurement sites such as the chest, the pulsatile amplitudes of the red and IR PPG signals are relatively small, and the waveforms are readily distorted by respiratory motion and changes in sensor–skin contact. Consequently, reliable peak and valley detection is difficult [15,16,17]. Therefore, rather than directly detecting peaks in the red and IR PPG signals, the proposed method defined an analysis window for each pulse cycle using peaks detected from the green PPG signal, which exhibited a more distinct pulsatile component.
The interval between consecutive ECG R-peaks was defined as the R-R interval, whereas the interval between consecutive green PPG peaks was defined as the pulse interval. Whenever a new green PPG peak was detected, the pulse interval and pulse rate were calculated from the time difference relative to the preceding peak. A buffer containing recent pulse intervals was continuously updated and used to assess the temporal plausibility of each newly detected peak. If a pulse interval fell outside the predefined physiological range or was inconsistent with the recent pulse-interval pattern, the corresponding peak was classified as an artifact or false detection and excluded from subsequent SpO2 estimation.
Each valid green PPG peak was used as a reference point for constructing the HR-adaptive analysis window. Rather than using a fixed window length, the window was dynamically adjusted according to the current pulse rate and recent pulse intervals. A shorter window was applied when the pulse interval decreased, whereas a longer window was applied when the pulse interval increased, thereby ensuring that the corresponding red and IR PPG segments were reliably included within each pulse cycle. The window duration was set to 1.4 times the current pulse interval and constrained to 80–220 samples (0.64–1.76 s at 125 Hz) to prevent physiologically implausible window lengths.
Whereas a previous study applied a fixed analysis-window length based on the mean R-R interval over the entire recording session [15], the proposed method updated the window length according to real-time changes in the pulse interval. This approach accounted for changes in heart rate and transient variations in pulse interval while allowing the AC and DC components of the red and IR PPG signals to be estimated over corresponding pulse cycles. Within each analysis window, the DC component of each wavelength was calculated as the mean of the low-pass-filtered signal. AC component estimation and R calculation were performed only for windows that satisfied the signal-quality criteria.

2.5.6. Signal Quality Index Based on Normalized Cross-Correlation

A normalized cross-correlation (NCC)-based SQI was calculated to determine whether each PPG beat segmented using the HR-adaptive analysis window was suitable for SpO2 estimation [15,36,37]. In chest reflectance PPG, the red and IR waveforms may be transiently distorted by low perfusion, body movement, respiratory motion, or changes in sensor–skin contact. Therefore, the NCC-based SQI was used as a beat-wise quality-control metric.
The green PPG signal, which was relatively stable under low-perfusion conditions, was used as the reference waveform. Waveform similarity between the green PPG and the red or IR PPG signal was evaluated within the same beat window [38]. Because NCC quantifies morphological similarity while reducing the influence of differences in absolute amplitude among channels, it was used to determine whether beat morphology was preserved in each wavelength channel. The mean-centered waveforms within each beat window were defined as
x λ ( n ) = PPG λ ( n ) PPG λ ¯ ,
g k ( n ) = PPG green ( n + k ) PPG green ¯ .
Here, λ { red , IR } denotes the PPG channel, n denotes the sample index within the beat window, N denotes the number of samples in the window, and k denotes the lag used to compensate for small temporal offsets between the green PPG and the red or IR PPG signal. PPG λ ¯ and PPG green ¯ represent the corresponding mean values within the beat window. The NCC between the PPG beat at wavelength λ and the green PPG reference beat was then calculated. To account for small temporal offsets among the red, IR, and green PPG signals, NCC was evaluated over lags ranging from 8 to + 8 samples, corresponding to 64 to + 64 ms. The maximum NCC value within this range was used as the SQI for the corresponding beat:
NCC k , λ = n = 1 N x λ ( n ) g k ( n ) n = 1 N x λ 2 ( n ) n = 1 N g k 2 ( n ) ,
SQI λ = max k [ 8 , 8 ] NCC k , λ .
Only beats for which both the red and IR PPG SQI values were equal to or greater than the predefined threshold were classified as valid and included in ensemble averaging, AC component estimation, and R calculation. If the SQI of either channel was below the threshold, the corresponding beat was considered potentially distorted by body movement, respiratory motion, or unstable sensor contact and was excluded from subsequent processing.
Beats that did not satisfy the SQI criterion were not included in the accumulation of valid windows for ensemble averaging. During these periods, updates of the SpO2 estimate were suspended, and the most recent valid output was temporarily retained. If poor signal quality persisted beyond a predefined allowable duration, the accumulated valid windows were reset and the SpO2 output was marked as invalid. Once both the red and IR PPG signals again satisfied the SQI criterion, valid windows were accumulated anew and SpO2 estimation was resumed. This beat-wise quality-control procedure prevented distorted PPG segments from being included in the ratio-of-ratios calculation and improved the stability of real-time SpO2 output.

2.5.7. SpO2 Estimation Using Least-Squares Linear Transformation

SpO2 estimation was performed using only red and IR PPG beats that were segmented within HR-adaptive analysis windows and satisfied the NCC-based SQI criterion. Four consecutive valid beat windows were accumulated and ensemble-averaged to obtain representative red and IR PPG waveforms. This procedure was applied to suppress random noise and transient beat-to-beat waveform variability before estimating the AC amplitude ratio. The ensemble size was determined through a preliminary comparison of different numbers of beats, and a four-beat ensemble was selected as a compromise between noise reduction and temporal responsiveness.
To reduce the dependence of AC estimation on individual peak and valley locations, a least-squares-based linear transformation was applied [15,39]. In chest-worn reflectance PPG, local peaks and valleys may be distorted by low perfusion, respiratory motion, and variations in sensor–skin contact, thereby increasing the variability of peak-to-valley-based AC estimates. Accordingly, the relative amplitude relationship between the red and IR PPG waveforms was estimated using all samples within each ensemble-averaged window. As illustrated in Figure 5C, a linear regression of the red PPG amplitude on the corresponding IR PPG amplitude was performed, and the slope of the fitted line was used as the red-to-IR AC amplitude ratio.
Figure 5. Key steps in chest reflectance PPG-based SpO2 estimation. (A) Simultaneously recorded ECG and green, red, and IR PPG signals. Green PPG peaks define the HR-adaptive analysis windows, while ECG R-peaks provide reference cardiac timing. The green dashed line marks the green PPG peak. (B) Beat-aligned IR and red PPG waveforms extracted within each analysis window and combined by ensemble averaging. (C) Least-squares linear regression between the averaged IR and red PPG samples to estimate their relative pulsatile amplitude ratio for SpO2 calculation.
The ensemble-averaged red and IR PPG waveforms were denoted by PPG red ( n ) and PPG IR ( n ) , respectively. The linear relationship between the two waveforms, the least-squares parameter estimates, the wavelength-specific DC components, and the final ratio-of-ratios, R, were defined as follows:
PPG red ( n ) α 1 PPG IR ( n ) + α 2 ,
( α 1 * , α 2 * ) = arg min α 1 , α 2 n = 1 N PPG red ( n ) α 1 PPG IR ( n ) α 2 2 ,
A C red A C IR = α 1 * ,
D C red = 1 N n = 1 N PPG red ( n ) , D C IR = 1 N n = 1 N PPG IR ( n ) ,
R = A C red / D C red A C IR / D C IR = α 1 * D C red / D C IR .
Here, α 1 * represents the scale factor between the pulsatile components of the red and IR PPG waveforms and was therefore used as the red-to-IR AC amplitude ratio. The parameter α 2 * is the intercept term that accounts for differences in baseline level between the two waveforms, and N denotes the number of samples within the analysis window. The DC component at each wavelength was calculated as the mean value of the corresponding ensemble-averaged PPG waveform.
The resulting R value was subsequently converted to SpO2 using the calibration equation. By estimating the relative AC amplitude from the entire waveform rather than from individual peaks and valleys, the proposed approach was designed to reduce sensitivity to local peak distortion, valley drift, and residual noise. Restricting the analysis to SQI-qualified beats and applying four-beat ensemble averaging further limited the influence of low-quality beats and transient waveform variations on the estimation of R.

2.5.8. SpO2 Calibration Equation Estimation

A polynomial calibration equation was applied to convert the calculated R into SpO2. The relationship between R and SpO2 in pulse oximetry is not strictly linear, and the nonlinearity may become more pronounced at lower oxygen saturation levels. Accordingly, first-, second-, and third-order polynomial calibration models were fitted and compared. The goodness of fit of each model was evaluated using the fit root-mean-square error (fit RMSE), defined as the root-mean-square difference between the reference SaO2 measured using an arterial blood gas analyzer and the SpO2 predicted by applying the corresponding calibration equation to the R values obtained from the M350. Because fit RMSE generally decreases as the polynomial order increases, it was not used as the sole criterion for model selection.
To assess model generalizability, subject-level leave-one-subject-out (LOSO) cross-validation was additionally performed. In each iteration, the calibration equation was fitted using data from four of the five participants and evaluated using data from the remaining participant. This procedure was repeated until each participant had served once as the validation subject, and the mean and standard deviation of the subject-specific RMSE values were calculated for each polynomial order. The final calibration equation was selected by jointly considering the fit RMSE and LOSO RMSE, whether the model preserved a physiologically plausible monotonically decreasing relationship between R and SpO2 over the observed R range, the stability of the estimated coefficients under the small-sample condition, and model parsimony.

2.5.9. SpO2 Post-Processing

Post-processing was applied to the final SpO2 estimates to suppress residual noise and abnormal spikes. When the lower of the red and IR PPG SQI values was below 0.65, the corresponding SpO2 estimate was discarded, and the most recent valid output was retained using an SQI-freeze rule. No maximum holding duration was imposed on the retained SpO2 value in the current implementation; the most recent valid value therefore remained displayed until a subsequent SpO2 estimate satisfied the SQI criterion. The SQI threshold of 0.65 was empirically selected during algorithm development. Although previous studies used an NCC-based threshold of 0.70 [15,40], preliminary evaluation showed that this value resulted in excessive rejection of otherwise usable beats; therefore, 0.65 was fixed before validation and was not optimized using the validation dataset. A median filter and a moving-average filter were then applied sequentially, and the degree of smoothing was adjusted by adapting their window lengths according to the prevailing saturation level. This dual-track filtering strategy was designed to balance output stability against responsiveness to hypoxemic changes. Under normal conditions, both window lengths were set to 5 beats. When SpO2 remained at or below 88% for at least three consecutive beats, the condition was classified as hypoxemia and both window lengths were reduced to 3 beats, allowing more rapid tracking of changes in oxygen saturation; the default lengths were restored once SpO2 remained at or above 93% for five consecutive beats.
Because post-processing was performed on beat-wise SpO2 estimates, the temporal span of the filtering windows varied with heart rate rather than being directly determined by the 125-Hz PPG sampling rate. Consequently, the effective response time of the complete post-processing pipeline was expected to vary with heart rate and was not independently quantified in seconds in the present study. This adaptive post-processing strategy was designed to maintain output stability in chest-worn measurements affected by low perfusion and motion while preserving responsiveness to clinically important changes in SpO2.

2.6. Evaluation

The performance of the proposed SpO2 estimation algorithm was evaluated using paired samples obtained from the second-stage controlled desaturation study. The sample size was determined in accordance with ISO 80601-2-61:2017, which requires a minimum of 10 healthy participants and at least 200 valid data pairs distributed across the target saturation range. For each sample, the SpO2 estimated from the M350 was compared with the reference SaO2 measured using an arterial blood gas analyzer. The accuracy root mean square ( A rms ), which is commonly used to assess pulse oximeter accuracy, was calculated as the primary performance metric. The residual standard error ( S res ) was also calculated to quantify the dispersion of the estimates around the fitted relationship. The two metrics were defined as follows:
A rms = 1 n i = 1 n SpO 2 , M 350 , i SaO 2 , ref , i 2 ,
SpO 2 , fit , i = β ^ 0 + β ^ 1 SaO 2 , ref , i ,
S res = i = 1 n SpO 2 , M 350 , i SpO 2 , fit , i 2 n 2 .
Here, SpO 2 , M 350 , i denotes the SpO2 value estimated by the proposed algorithm, and SaO 2 , ref , i denotes the corresponding reference arterial oxygen saturation measured using an arterial blood gas analyzer. A linear regression model was fitted using reference SaO2 as the independent variable and M350-derived SpO2 as the dependent variable. The fitted value was calculated as SpO 2 , fit , i = β ^ 0 + β ^ 1 SaO 2 , ref , i . The residual standard error, S res , represents the dispersion of the M350-derived SpO2 values around the fitted regression line.
Because each participant contributed multiple paired measurements rather than a single independent observation, the point estimates of A rms and mean bias were calculated from the pooled set of valid data pairs. To account for within-participant dependence when estimating statistical uncertainty, 95% confidence intervals (CIs) for A rms and mean bias were obtained using participant-level cluster bootstrap resampling with 10,000 replicates. In each bootstrap replicate, participants were sampled with replacement, and all repeated measurements from each selected participant were retained together. The 2.5th and 97.5th percentiles of the bootstrap distributions were used as the lower and upper limits of the 95% CIs. In addition, participant was treated as a random factor for variance-component analysis. The variance of the paired SpO2–SaO2 differences was partitioned into between-participant ( σ u 2 ) and within-participant ( σ 2 ) components using a random-intercept mixed-effects model fitted by restricted maximum likelihood (REML). The intraclass correlation coefficient (ICC) was calculated as σ u 2 / ( σ u 2 + σ 2 ) . These analyses were performed both over the full reference saturation range and within each nominal saturation range.

3. Results

3.1. Comparison of PPG Perfusion Index Across Wavelengths

To quantitatively compare wavelength-specific pulsatile amplitude, participant-level median PI values were derived from valid 10-s windows during the controlled desaturation protocol. Across the 12 participants, the median PI was 0.563% [IQR, 0.420–0.780] for green PPG, compared with 0.166% [IQR, 0.133–0.217] for red PPG and 0.158% [IQR, 0.127–0.214] for IR PPG. Green PPG showed significantly higher PI than both red and IR PPG (paired Wilcoxon signed-rank tests, both p < 0.001 ), and the green-channel PI was higher than both other channels in all 12 participants. These findings provide quantitative support for using green PPG as the reference signal for cardiac-cycle detection at the chest.

3.2. Calibration Model Comparison and Selection

No prespecified adverse events occurred during the calibration study, and all participants completed the planned arterial blood sampling procedure. The calibration-development dataset comprised 94 paired ( R , SaO 2 ) observations obtained from the five participants enrolled in Stage 1. No Stage 2 data were used for calibration fitting or model selection. First-, second-, and third-order polynomial models were fitted to characterize the relationship between R and reference SaO2. The fitted polynomial calibration models are presented in Figure 6. The measured data showed an overall decrease in reference SaO2 with increasing R, with pronounced nonlinear curvature in the low-saturation region at R > 1.1 . This trend is consistent with the established nonlinear inverse relationship between R and oxygen saturation in pulse oximetry [41,42].
Figure 6. Polynomial calibration models relating R to reference SaO 2 . Red circles represent 94 paired measurements from five participants, black lines indicate the fitted calibration curves, and gray shaded areas indicate ± 4 percentage points around the fitted curves. The linear (A), quadratic (B), and cubic models (C) yielded fit RMSEs of 3.39%, 2.96%, and 2.65%, respectively. The quadratic model was selected as the final calibration function based on predictive performance, monotonicity, physiological plausibility, and model parsimony. Exact device-specific coefficients are not reported because they constitute proprietary calibration parameters of the investigational device.
The fit RMSEs for the linear, quadratic, and cubic models were 3.39%, 2.96%, and 2.65%, respectively. In LOSO cross-validation across the five participants, the corresponding mean RMSEs were 3.29 ± 1.48 % , 3.19 ± 1.30 % , and 2.87 ± 0.91 % , respectively. Thus, the cubic model yielded the lowest numerical error for both the fit and LOSO cross-validation.
However, model selection was not based solely on numerical error. Although the cubic model provided the lowest RMSE, it did not preserve a monotonically decreasing relationship over the full calibration range. The slope changed sign at approximately R = 0.72 and R = 1.37 , resulting in an increase in predicted SpO2 with increasing R when R < 0.72 or R > 1.37 . This behavior is inconsistent with the established inverse optical relationship between R and oxygen saturation expected in conventional red/infrared pulse oximetry [41,42,43]. In particular, because the region at R < 0.72 corresponds to normal or high oxygen saturation, the non-monotonic behavior in this range reduces the physical plausibility of the cubic calibration function despite its lower numerical error.
The linear model preserved the expected inverse relationship but showed higher fit and LOSO RMSEs than the quadratic model, indicating that the linear model did not adequately capture the nonlinear relationship between R and oxygen saturation. In contrast, the quadratic model maintained a monotonically decreasing relationship throughout the observed R range while achieving lower fit and LOSO RMSEs than the linear model. Thus, the linear model provided lower predictive performance than the quadratic model, whereas the cubic model achieved lower numerical error at the expense of physically implausible non-monotonic behavior. The quadratic model was the lowest-order model that both improved predictive accuracy relative to the linear model and preserved the expected monotonic inverse relationship throughout the observed calibration range. Therefore, based on the combined consideration of predictive accuracy, physical plausibility, monotonicity, and model parsimony, the quadratic model was selected as the final calibration function. This finalized model was subsequently carried forward unchanged to the Stage 2 pivotal accuracy evaluation, where it was applied to beat-wise SpO2 estimation.

3.3. Accuracy of the Proposed SpO2 Estimation Algorithm

The accuracy of the proposed SpO2 estimation algorithm was evaluated using paired samples obtained during the controlled desaturation study. A total of 300 arterial reference samples were collected from 12 participants, with 25 samples obtained from each participant. The first arterial sample (S1) from each participant was collected at baseline before the controlled desaturation procedure and was excluded according to the prespecified analysis criteria. After exclusion of these 12 baseline samples, 288 desaturation samples remained eligible for analysis. An additional 10 samples with reference SaO2 values outside the 73–97% range were excluded in accordance with the pooled-data trimming criteria specified in ISO 80601-2-61:2017. These comprised samples S14–S16 from participant 5; S2, S14, and S15 from participant 7; S14 from participant 9; and S2, S3, and S14 from participant 10. Consequently, 278 paired samples were included in the final accuracy analysis.
Physiological circulatory delay was considered when temporally aligning the M350-derived SpO2 time series with the reference SaO2 values obtained from arterial blood sampling. Changes in oxygen saturation measured by pulse oximetry may lag behind changes in arterial SaO2 because of the circulation time to the measurement site and the response characteristics of the measurement system [44,45]. To minimize potential temporal mismatch, each arterial SaO2 measurement was paired with the temporally nearest M350-derived SpO2 output at the corresponding blood-sampling time. The same temporal-pairing procedure was applied uniformly to all participants, without applying a fixed or participant-specific time shift.
Figure 7 shows the temporal changes in M350-derived SpO2 during controlled desaturation in four representative participants. The continuously estimated M350 SpO2 values followed the stepwise reductions in oxygen saturation, and the reference SaO2 values measured at each arterial blood sampling time are superimposed on the same time axis. Figure 8A shows the relationship between M350-derived SpO2 and reference SaO2 across all 278 paired samples. Each point represents one paired measurement, and the colors indicate individual participants. The black dashed line represents the identity line ( y = x ), whereas the blue solid line represents the linear regression line fitted to all paired samples. M350-derived SpO2 showed a strong linear association with reference SaO2 ( y = 1.174 x 15.542 , R 2 = 0.916 ), with a S res of 2.54% around the fitted regression line. Over the trimmed reference SaO2 range of 73–97%, the pooled A rms was 2.91%, satisfying the prespecified accuracy criterion of 4%. Figure 8B presents the participant-specific A rms values, with a mean across participants of 2.81%.
Figure 7. Representative SpO2 tracings for four participants (11, 2, 9, and 6). In each panel, the blue line denotes the continuous SpO2 measured by the M350 device (M350-SpO2) and the orange dots denote the reference arterial SaO2 obtained from intermittent blood sampling (Ref-SaO2), plotted against clock time (HH:MM:SS).
Figure 8. Accuracy of M350-derived SpO2 estimates against reference SaO2. (A) Scatter plot of M350 SpO2 versus reference SaO2 for all participants. Colors indicate individual participants, the black dashed line represents the identity line ( y = x ), and the blue solid line represents the linear regression fit, and the gray shaded region represents ± 4 percentage points around the identity line. (B) Participant-specific A rms values. The red dashed line indicates the 4% accuracy criterion, and the black dashed line indicates the mean A rms across participants (2.81%).
For the nominal saturation ranges of 70–80%, 80–90%, and 90–100%, the A rms values were 3.42%, 3.04%, and 2.13%, respectively, with participant-level cluster-bootstrap 95% confidence intervals (CIs) of 2.51–4.17%, 2.35–3.64%, and 1.64–2.66%, respectively (Table 1). The overall A rms was 2.91% (95% CI, 2.45–3.34%). The overall mean bias was 0.70 % (95% CI, 1.58 to 0.15%), whereas the range-specific mean biases were 2.44 % , 0.44 % , and 0.71%, respectively. These findings indicate that the M350 tended to underestimate reference SaO2 at lower saturation levels, whereas a slight overestimation was observed at higher saturation levels.
Table 1. Agreement between the M350 pulse oximeter and the reference arterial oxygen saturation, stratified by reference saturation range. Each column summarizes all pairs falling within that range, pooled across the 12 participants; the final column covers the full range after pooled-data trimming. Confidence intervals were obtained by participant-level (cluster) bootstrap resampling with 10,000 replicates, which accounts for the repeated measurements contributed by each participant.
The 95% CIs for the mean bias and A rms , accounting for repeated measurements within participants, are presented in Table 1. Over the full saturation range, the between-participant variance of the paired SpO2–SaO2 differences was 2.39 % 2 (SD, 1.55%), whereas the within-participant variance was 5.81 % 2 (SD, 2.41%). The corresponding ICC was 0.29, indicating that approximately 71% of the total variability arose within participants.
Table 2 summarizes these participant-specific results in detail. Each participant contributed 21–24 data pairs spanning a comparable portion of the reference range (73.1–97.0% overall). Per-participant A rms ranged from 1.37% to 4.28% and per-participant bias from 2.86 % to + 1.87 % . On average, 74.6% of each participant’s pairs agreed with the reference within ± 3 % (range 47.6–95.8%). Weighting each participant equally rather than pooling the data pairs yielded a mean bias of 0.70 ± 1.63 % and a mean A rms of 2.81 ± 0.80 % ; because A rms is a root-mean-square quantity, its participant-weighted analogue is the quadratic mean of the per-participant values, which is 2.91% and therefore identical to the pooled estimate.
Table 2. Agreement between the M350 pulse oximeter and the reference arterial oxygen saturation, stratified by participant. Each row summarizes the paired samples contributed by one participant after data trimming. The final two rows report the mean and standard deviation of these values across the 12 participants.

3.4. Bland–Altman Analysis

Agreement between M350-derived SpO2 and reference SaO2 was assessed using Bland–Altman analysis of the 278 paired samples included in the accuracy analysis. For each paired sample, the mean of the two measurements was plotted on the x-axis, and their difference ( M 350 SpO 2 reference SaO 2 ) was plotted on the y-axis (Figure 9). Across the overall analysis range, the mean bias was 0.70 % , and the standard deviation of the differences was 2.83%, yielding 95% limits of agreement from 6.24 % to 4.84 % . Thus, M350-derived SpO2 was, on average, 0.70 percentage points lower than reference SaO2. For the nominal saturation ranges of 70–80%, 80–90%, and 90–100%, the mean biases were 2.44 % , 0.44 % , and 0.71 % , respectively, with corresponding 95% limits of agreement of 7.16 % to 2.28 % , 6.37 % to 5.49 % , and 3.25 % to 4.68 % , respectively. These findings indicate that underestimation was more pronounced at lower saturation levels, whereas a slight overestimation was observed at higher saturation levels.
Figure 9. Bland–Altman analysis of agreement between M350-derived SpO2 and reference SaO2. Results are presented for the 70–80%, 80–90%, 90–100%, and overall 70–100% reference saturation ranges. Each point represents one paired measurement, with colors indicating individual participants. The x-axis represents the mean of the M350 SpO2 and reference SaO2, and the y-axis represents their difference (M350 SpO2 minus reference SaO2). Bias denotes the mean difference between the two measurements, LoA denotes the 95% limits of agreement calculated as the mean bias ± 1.96 standard deviations, and RMS denotes the root-mean-square error. The solid black line indicates the mean bias, the red dashed lines indicate the 95% limits of agreement, and the black dotted line indicates zero difference.

4. Discussion

In this study, PPG-based SpO2 estimation, which has conventionally been performed at peripheral sites such as the finger, earlobe, and wrist, was extended to chest-worn reflectance PPG and implemented as a real-time algorithm for a wearable device. The chest is well suited for the simultaneous acquisition of ECG, respiration, and motion signals using a single patch-type device. However, compared with peripheral sites, chest PPG is characterized by lower perfusion and greater susceptibility to respiratory and motion artifacts, which limits conventional approaches that estimate the AC components by directly detecting peaks and valleys in the red and IR PPG waveforms. To address these challenges, the proposed SpO2 estimation pipeline integrated green PPG-based beat detection, auxiliary validation using ECG R-peaks, HR-adaptive analysis windows, an NCC-based SQI, least-squares estimation of the AC amplitude ratio, and real-time post-processing.
The proposed algorithm was evaluated against reference SaO2 measured using an arterial blood gas analyzer during a controlled desaturation study. Over the trimmed reference SaO2 range of 73–97%, the overall A rms was 2.91%, satisfying the 4.0% accuracy criterion specified in ISO 80601-2-61:2017. The A rms values for the nominal saturation ranges of 70–80%, 80–90%, and 90–100% were 3.42%, 3.04%, and 2.13%, respectively. Although the estimation error increased at lower saturation levels, it remained within 4.0% in all three ranges. Bland–Altman analysis showed a small overall mean bias of 0.70 % , although underestimation was more pronounced at lower saturation levels. Importantly, the 95% limits of agreement ranged from 6.24 % to + 4.84 % , indicating that individual SpO2 estimates could deviate from arterial SaO2 by more than the overall A rms alone suggests. This combination of a small mean bias and relatively wide limits of agreement indicates limited systematic error but appreciable variability among individual measurements. The median absolute error was 1.65%, and 74.8% of all paired samples fell within ± 3 % of the reference. The dispersion was also not uniform across the measurement range, with the observed proportional bias indicating a saturation-dependent component of the measurement error. A rms should therefore be interpreted together with the limits of agreement, the distribution of individual measurement errors, and their dependence on saturation, rather than as the error expected for each individual measurement.
This saturation dependence was most pronounced in the 70–80% range, where the mean bias reached 2.44 % , and warrants particular consideration because measurement errors at low oxygen saturation can be clinically relevant. Although this underestimation would be unlikely to obscure the presence of hypoxemia, it could exaggerate its apparent severity and potentially influence threshold-based clinical decisions or confirmatory testing [46]. Because the 70–80% range is already well below commonly used oxygen-saturation targets [47], the observed negative bias would be unlikely to change the recognition of marked hypoxemia itself, although it may affect assessment of its severity. The reduced accuracy at low saturation is consistent with previous controlled-desaturation studies reporting greater bias or error dispersion below 80% oxygen saturation [48,49]. Further evaluation in patients with naturally occurring hypoxemia is therefore needed to determine the clinical impact of this saturation-dependent bias.
Taken together, these findings support the technical feasibility of chest reflectance PPG as an optical signal source for SpO2 estimation under controlled desaturation conditions when combined with appropriate signal quality assessment and beat-wise analysis. Notably, the proposed algorithm was not limited to offline analysis but was implemented in the HiCardi M350 and evaluated by direct comparison with arterial SaO2.

4.1. Technical Feasibility and Potential Clinical Implications

The principal implication of this study is the technical feasibility of integrating ECG and SpO2 monitoring using a single chest-worn device. Conventional pulse oximeters are commonly placed on the finger or earlobe, where relatively strong PPG signals can be obtained. However, prolonged monitoring using a finger probe may be limited by wearability and signal stability during physical activity. Relative movement between the sensor and the finger can introduce measurement errors, and motion artifact is a major limitation of continuous SpO2 monitoring [50,51]. In a study of patients with chronic obstructive pulmonary disease using a wearable finger pulse oximeter, the proportion of valid SpO2 data decreased during moderate-to-vigorous physical activity because of motion artifact [13]. In addition, device comfort and user preference may influence device selection and long-term adherence in SpO2 self-monitoring [52].
The chest is suitable for patch-based monitoring and is spatially compatible with conventional ECG acquisition. A single chest patch can therefore support long-term acquisition of ECG, heart rate, respiration-related signals, and motion data. Previous studies have demonstrated the feasibility of continuous ECG-based heart rate and respiratory rate monitoring using chest-worn devices [53,54]. The clinical potential of chest-patch vital sign monitoring has also been investigated under motion and controlled hypoxia conditions [14]. Furthermore, sternum-based reflectance PPG has been proposed as a potential measurement configuration for wearable pulse oximetry [15]. Collectively, these findings suggest that chest-worn patches may provide a suitable platform for future integration of ECG and PPG-derived SpO2 measurements, although their performance during unrestricted daily activity requires further evaluation.
Integrated continuous monitoring may be particularly useful in settings that require prolonged physiological surveillance, including general wards, perioperative care, and post-discharge monitoring. A systematic review of continuous vital sign monitoring using wearable wireless devices in hospitalized patients suggested that continuous monitoring may facilitate earlier detection of clinical deterioration than conventional intermittent measurements [55]. The feasibility of continuous vital sign monitoring in the home after hospital discharge has also been reported [56]. In addition, wearable measurements of heart rate, respiratory rate, SpO2, and body temperature have been evaluated in patients recovering from surgery, supporting their potential use in perioperative and postoperative care [57]. The chest-based SpO2 estimation algorithm developed in this study provides a technical basis for incorporating oxygen saturation into such multimodal continuous monitoring platforms. However, the accuracy demonstrated in the present study was limited to controlled desaturation conditions and does not establish clinical performance during prolonged ambulatory or free-living monitoring or in patients with cardiopulmonary disease.

4.2. Algorithmic Considerations for Chest Reflectance PPG

Reliable SpO2 estimation from chest reflectance PPG requires stable extraction of the relatively small and variable pulsatile components of the red and IR PPG signals under low-perfusion conditions. Conventional pulse oximetry typically calculates the AC/DC ratios by detecting peaks and valleys in the red and IR PPG waveforms [4,5]. At reflectance measurement sites such as the chest, however, low perfusion, respiratory motion, and changes in sensor contact can destabilize peak–valley detection in the red and IR signals [15,16]. Therefore, rather than detecting beat landmarks directly from the red and IR waveforms, the proposed algorithm used green PPG, which generally exhibits a more distinct pulsatile component, as the reference signal for beat segmentation [27,29]. The detected green PPG beat timing was used to define HR-adaptive analysis windows, thereby reducing beat misalignment during changes in heart rate and ensuring that the corresponding red and IR PPG segments were included within each cardiac cycle.
The red-to-IR AC amplitude ratio was estimated using a least-squares-based linear transformation that did not depend directly on individual peak and valley locations. Conventional ratio-of-ratios estimation assumes that the AC component at each wavelength can be extracted reliably [15,16,17]. Under low-perfusion or motion-contaminated conditions, however, peak distortion, valley drift, and baseline fluctuation may increase the variability of peak-to-valley-based AC estimates [58,59]. The proposed approach therefore used all samples within each analysis window to regressively estimate the relative amplitude relationship between the red and IR PPG waveforms. A similar linear transformation between red and IR PPG beats was previously proposed for sternum-based reflectance pulse oximetry to overcome the limitations of peak and valley extraction in respiration-contaminated PPG [15]. Accordingly, the least-squares approach used in the present study was intended to reduce sensitivity to local waveform extrema and provide a more stable estimate of the R in chest PPG.
Signal quality was assessed using an NCC-based SQI. Correlation and template-matching-based SQIs have been used to evaluate the preservation of PPG beat morphology and exclude artifact-contaminated beats [15,60,61]. In the proposed algorithm, the green PPG beat was used as the reference waveform, and its morphological similarity to the corresponding red and IR PPG beats was assessed. Only beats satisfying the SQI criterion were included in the SpO2 calculation. Because motion artifact may cause transient false desaturation or isolated spikes [23,50], the output was not updated during low-SQI periods; instead, the most recent valid SpO2 value was retained. Median and moving-average filtering were then applied sequentially to suppress isolated spikes and physiologically implausible abrupt changes. Conversely, when SpO2 persistently decreased into the hypoxemic range, a dual-track strategy reduced the filter length to improve responsiveness. This framework was designed to balance output stability under normal conditions with the ability to track clinically important desaturation events and represents a key component of the chest-worn SpO2 estimation pipeline.

4.3. Comparison with Prior Work

The present study extends previous work on chest reflectance PPG-based SpO2 estimation and pulse oximetry signal processing. Chan et al. measured ECG and multiwavelength PPG using a wearable patch biosensor placed over the sternum and demonstrated the feasibility of sternum-based reflectance pulse oximetry using a breath-hold protocol [15]. To address unstable peak and valley extraction caused by low perfusion and respiratory artifact, their method incorporated green PPG-based outlier rejection and a least-squares linear transformation between red and IR PPG beats to derive the R. The present study shares the general algorithmic principles of green PPG-based beat-level quality assessment and least-squares estimation of the AC amplitude ratio.
However, the present study differs in both the reference measurement and the implementation of the algorithm. Chan et al. [15] evaluated sternum-derived SpO2 against finger pulse oximeter SpO2, whereas the present study used SaO2 measured using an arterial blood gas analyzer during controlled desaturation. Because the chest and finger may differ in their temporal responses and hemodynamic characteristics, direct comparison with arterial SaO2 reduces the contribution of site-dependent differences and provides a more direct assessment of algorithm accuracy. In addition, the analysis by Chan et al. was primarily performed offline and included breath-hold interval selection, alignment between finger SpO2 and chest PPG, and comparison of calibration schemes. In contrast, the present study integrated green PPG-based beat detection, ECG-assisted validation, HR-adaptive analysis windows, an NCC-based SQI, accelerometer-based motion artifact detection, and real-time post-processing into a continuous pipeline implemented on a chest-worn device.
Alkhoury et al. compared signal-processing strategies for mitigating motion artifact in pulse oximetry [59]. Their approach used comb filtering tuned to ECG-derived heart rate to preserve the cardiac fundamental and harmonic components of the PPG signal while attenuating motion-related noise between these frequencies. The present study similarly used ECG as an auxiliary signal for PPG-based SpO2 estimation. However, rather than using ECG-derived heart rate to tune a frequency-domain filter, it was used to assess the physiological plausibility of green PPG-derived pulse intervals and to define HR-adaptive analysis windows. Thus, whereas the approach of Alkhoury et al. focused on frequency-domain purification of motion-contaminated PPG, the present study implemented an integrated real-time pipeline for SpO2 estimation from low-perfusion chest reflectance PPG and evaluated its accuracy against arterial SaO2.
The principal contribution of this study is therefore the demonstration of the technical feasibility of real-time SpO2 estimation from chest reflectance PPG together with an evaluation of its accuracy against arterial SaO2 under controlled desaturation conditions. The proposed method provides a practical framework for incorporating SpO2 estimation into multimodal chest-worn monitoring systems.

4.4. Limitations and Future Directions

This study has several limitations. First, the number of participants used for algorithm calibration and accuracy evaluation was limited, and all 17 participants enrolled across the two study stages were young, healthy adults evaluated under controlled desaturation conditions. Accordingly, the performance of the proposed algorithm in older adults and in individuals with impaired circulation or low perfusion remains unknown and requires further evaluation. In addition, the subject-level LOSO analysis performed within the five-participant calibration cohort should be regarded as preliminary evidence of model robustness and generalizability and cannot substitute for independent external validation in a larger cohort. Although the Stage 2 accuracy cohort was independent of the Stage 1 calibration-development cohort, both were evaluated within the same overall clinical investigation. Therefore, the reported accuracy represents an independent within-study evaluation and should not be interpreted as external validation of algorithm performance.
More broadly, the study population did not adequately represent the full range of skin pigmentation, body habitus, chest tissue characteristics, and perfusion conditions that may be encountered in clinical and wearable settings. These factors may influence optical absorption and scattering, optical path length, pulsatile signal amplitude, and PPG signal quality and therefore may affect SpO2 estimation performance. Skin-pigmentation-related bias is particularly relevant to pulse oximetry; previous studies have reported greater SpO2–SaO2 errors, overestimation during hypoxemia, and a higher prevalence of occult hypoxemia in individuals with darker skin pigmentation [62,63,64]. Because standardized pigmentation measures such as the individual typology angle (ITA) or Monk Skin Tone (MST) were not prospectively collected and the present cohort was too small for reliable subgroup comparisons, pigmentation-stratified analysis was not performed. Future validation studies should therefore include larger and more diverse populations, including older adults and individuals with impaired circulation or low perfusion, prospectively characterize skin pigmentation, and evaluate bias, A rms , limits of agreement, and error distributions across pigmentation groups, with additional analyses according to body habitus, chest tissue characteristics, and perfusion status where sufficient sample sizes are available.
Second, the study was conducted under controlled desaturation conditions with restricted movement; therefore, performance in the presence of substantial motion artifacts encountered in daily life was not fully evaluated. Accordingly, the present results should not be interpreted as validation of performance during unrestricted or free-living continuous monitoring. Walking, postural changes, changes in sleeping position, and variations in sensor contact pressure may substantially distort chest PPG waveforms. Although green PPG provided a relatively stable reference for beat detection, SpO2 estimation ultimately depends on reliable extraction of the pulsatile AC components from the red and IR PPG signals. Therefore, when the red or IR pulsatile components are markedly attenuated, particularly under low-perfusion conditions, the reliability of R estimation and subsequent SpO2 estimation may be compromised even when green PPG-based beat detection remains stable. In addition, the NCC-based SQI may misclassify common-mode artifacts as acceptable beats when the green, red, and IR PPG signals are distorted in a similar manner. Future studies should evaluate the algorithm during ambulatory and prolonged wear and investigate a multimodal SQI framework that integrates accelerometer-derived motion indices, perfusion indices, PPG morphology, and beat-to-beat consistency. Separate performance testing under motion and low-perfusion conditions will also be necessary for devices intended for continuous monitoring.
Third, the magnitude and direction of the estimation error varied with oxygen saturation. The overall mean bias and A rms were 0.70 % and 2.91%, respectively, whereas the corresponding values in the nominal 70–80% range were 2.44 % and 3.42%. The greater negative bias and wider dispersion observed at lower saturation levels warrant particular caution when interpreting performance in clinically significant hypoxemia. Moreover, although the nominal 70–80% range was represented in the dataset, observations near the lower boundary of the achieved SaO2 range were limited, restricting the strength of conclusions regarding performance at the most severe levels of desaturation. A significant proportional bias was also observed between the SpO2–SaO2 difference and mean saturation (slope = 0.20 , p < 0.001 ), indicating increasing underestimation at lower saturation levels. As discussed above, this saturation-dependent underestimation may affect interpretation of hypoxemia severity and therefore represents an important limitation of the current calibration. Reduced perfusion, respiratory baseline fluctuations, changes in sensor contact pressure, and variations in tissue optical path length may exert a greater influence on R under hypoxemic conditions. Although the proposed method incorporated nonlinear calibration and multiple signal-quality control steps, a single global calibration equation may not fully capture the R–SaO2 relationship across the entire saturation range, particularly at low saturation levels. Future studies should include more observations in the hypoxemic range and evaluate saturation-dependent or physiologically constrained calibration approaches in independent clinical datasets while maintaining computational feasibility for real-time implementation.
Fourth, the post-processing framework, which included SQI freezing, median filtering, moving-average filtering, and dual-track filtering, involves an inherent trade-off between output stability and temporal responsiveness. Stronger smoothing can suppress artifact-related fluctuations but may delay the detection of true desaturation or resaturation, whereas greater responsiveness may increase the influence of transient artifacts on the reported SpO2 value. In particular, the SQI-freezing rule imposed no maximum holding duration in the current implementation; therefore, the most recent valid SpO2 value could remain displayed until a subsequent estimate satisfied the SQI criterion. During prolonged signal-quality degradation, this could result in an outdated SpO2 value remaining displayed. Future implementations should incorporate a bounded holding interval after which the output is invalidated. In addition, residual temporal mismatch may remain despite nearest-time pairing because physiological circulatory delay can vary among individuals, particularly during rapid desaturation or resaturation. This may increase the apparent bias and dispersion of the SpO2–SaO2 differences. Future studies should optimize post-processing parameters across a range of desaturation and resaturation rates and realistic wearing conditions, while also evaluating the sensitivity of accuracy estimates to interindividual variability in circulatory delay. Adaptive post-processing methods that dynamically adjust filter characteristics according to signal quality, rate of change, and current saturation level should also be investigated. Desaturation-event detection delay, recovery time, and false-alarm rate should be quantitatively assessed.
Finally, the present study was designed according to ISO 80601-2-61:2017, which was the applicable edition at the time of study design. Accordingly, the reported 4.0% A rms criterion should be interpreted as the prespecified accuracy criterion used for the present study rather than as evidence of conformity with subsequently updated requirements. ISO 80601-2-61:2017 has since been superseded by ISO 80601-2-61:2026 [65]. In parallel, current regulatory recommendations have placed greater emphasis on evaluating pulse-oximeter performance across diverse skin pigmentation. In particular, the 2025 FDA draft guidance recommends standardized characterization of skin pigmentation using MST and ITA, broader representation across pigmentation groups, and reporting of accuracy and associated uncertainty both overall and across pigmentation strata [66]. Because the present study was not designed according to these updated standards and recommendations, its results should not be interpreted as demonstrating conformity with ISO 80601-2-61:2026 or with the proposed FDA performance framework. Future validation studies should therefore reflect the updated ISO standard and applicable regulatory recommendations.
In summary, the integration of green PPG-based beat segmentation, SQI-guided beat selection, least-squares estimation of the AC amplitude ratio, and real-time post-processing supported the technical feasibility of chest reflectance PPG-based SpO2 estimation and demonstrated its accuracy against arterial SaO2 under controlled desaturation conditions. Nevertheless, because the study involved a limited number of participants and was conducted under controlled desaturation conditions with restricted movement, the generalizability of the algorithm should be established in larger and more diverse populations, ambulatory and prolonged monitoring settings, datasets containing sufficient observations in the hypoxemic range, and independent external validation cohorts. Further improvement in low-saturation calibration and dynamic response, together with validation against updated standards and applicable regulatory recommendations, may facilitate the extension of this algorithm to continuous wearable monitoring systems that simultaneously provide ECG and SpO2.

5. Conclusions

In this study, a real-time SpO2 estimation algorithm integrating green PPG-based beat segmentation, signal quality assessment, least-squares estimation of the AC amplitude ratio, and adaptive post-processing was developed and implemented in the HiCardi M350 chest-worn device. When evaluated against arterial SaO2 during a controlled desaturation study, the algorithm achieved a pooled A rms of 2.91% over the reference SaO2 range of 73–97%, satisfying the prespecified accuracy criterion of 4.0%. These findings support the technical feasibility of chest reflectance PPG-based SpO2 estimation and demonstrate its accuracy under controlled desaturation conditions when combined with appropriate signal quality control and beat-wise analysis. The proposed approach may provide a basis for future development of multimodal chest-worn monitoring systems capable of providing simultaneous ECG and SpO2 measurements. However, because the present study was conducted in healthy volunteers under controlled desaturation and restricted-movement conditions, further validation in larger and more diverse populations, including patients with cardiopulmonary disease, and under ambulatory, prolonged-wear, and free-living conditions is required to establish its applicability to continuous real-world SpO2 monitoring.

Author Contributions

Conceptualization, W.L.; methodology, W.L., J.S., M.S. and S.C.; software, W.L., J.S. and M.S.; validation, W.L.; formal analysis, W.L.; investigation, M.-K.S., K.M.K. and G.-J.N.; data curation, W.L. and M.-K.S.; writing—original draft preparation, W.L. and S.C.; writing—review and editing, J.S., M.S., K.M.K., G.-J.N. and S.C.; supervision, S.C.; project administration, J.S., M.S. and S.C.; funding acquisition, S.C. All authors have read and agreed to the published version of the manuscript.

Funding

This work was supported by the Technological Innovation R&D Program (SCALEUP TIPS, Tech Investor Program for Scaleup) (RS-2023-00303821) funded by the Ministry of SMEs and Startups (MSS, Republic of Korea).

Institutional Review Board Statement

The study was conducted in accordance with the Declaration of Helsinki and approved by the Institutional Review Board of Asan Medical Center (IRB No. 2025-1580).

Data Availability Statement

The data presented in this study are available on request from the corresponding author. The data are not publicly available due to privacy restrictions applicable to clinical investigation data.

Acknowledgments

During the preparation of this manuscript, the authors used ChatGPT (GPT-5.6; OpenAI, San Francisco, CA, USA) for the purposes of preliminary literature identification and improvement of the clarity, readability, and linguistic quality of the text. The authors have reviewed and edited the output and take full responsibility for the content of this publication.

Conflicts of Interest

W.L., J.S., M.S., M.-K.S. and S.C. are employees of MEZOO Co., Ltd., the manufacturer of the HiCardi M350 device evaluated in this study. K.M.K. and G.-J.N. declare no conflicts of interest. The funders had no role in the design of the study; in the collection, analyses, or interpretation of data; in the writing of the manuscript; or in the decision to publish the results.

Abbreviations

The following abbreviations are used in this manuscript:
ACAlternating current (pulsatile) component
ACCAccelerometer
MAEMean absolute error
ADTAdaptive threshold detection
A rms Accuracy root mean square
DCDirect current component
ECGElectrocardiogram
FIRFinite impulse response
HRHeart rate
IRInfrared
ITAIndividual typology angle
LOSOLeave-one-subject-out
MSTMonk Skin Tone
S res Residual standard error
NCCNormalized cross-correlation
PPGPhotoplethysmography
RRatio of ratios
RMSERoot-mean-square error
SaO2Arterial oxygen saturation
SpO2Peripheral oxygen saturation
SQISignal quality index
LoALimits of agreement

References

  1. Kim, D.; Min, J.; Ko, S.H. Recent developments and future directions of wearable skin biosignal sensors. Adv. Sens. Res. 2024, 3, 2300118. [Google Scholar] [CrossRef] [Scilit]
  2. Charlton, P.H.; Kyriacou, P.A.; Mant, J.; Marozas, V.; Chowienczyk, P.; Alastruey, J. Wearable photoplethysmography for cardiovascular monitoring. Proc. IEEE 2022, 110, 355–381. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  3. Charlton, P.H.; Allen, J.; Bailón, R.; Baker, S.; Behar, J.A.; Chen, F.; Clifford, G.D.; Clifton, D.A.; Davies, H.J.; Ding, C.; et al. The 2023 wearable photoplethysmography roadmap. Physiol. Meas. 2023, 44, 111001. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  4. Nitzan, M.; Romem, A.; Koppel, R. Pulse oximetry: Fundamentals and technology update. Med. Devices Evid. Res. 2014, 7, 231–239. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  5. Aoyagi, T. Pulse oximetry: Its invention, theory, and future. J. Anesth. 2003, 17, 259–266. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  6. Lee, H.; Ko, H.; Lee, J. Reflectance pulse oximetry: Practical issues and limitations. ICT Express 2016, 2, 195–198. [Google Scholar] [CrossRef] [Scilit]
  7. Dcosta, J.V.; Ochoa, D.; Sanaur, S. Recent progress in flexible and wearable all organic photoplethysmography sensors for SpO2 monitoring. Adv. Sci. 2023, 10, 2302752. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  8. Jung, H.; Kim, D.; Lee, W.; Seo, H.; Seo, J.; Choi, J.; Joo, E.Y. Performance evaluation of a wrist-worn reflectance pulse oximeter during sleep. Sleep Health 2022, 8, 420–428. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  9. Joo, M.G.; Lim, D.H.; Park, K.K.; Baek, J.; Choi, J.M.; Baac, H.W. Reflection-boosted wearable ring-type pulse oximeters for SpO2 measurement with high sensitivity and low power consumption. Biosensors 2023, 13, 711. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  10. Long, Y.; Min, R.; Xiao, K.; Wang, Z.; Liu, L.; Sun, Y.; Li, X.; Li, Z.; Zalevsky, Z. Decoding subject-invariant emotional information from cardiac signals detected by photonic sensing system. Opto-Electron. Technol. 2025, 1, 250010. [Google Scholar] [CrossRef] [Scilit]
  11. Qiu, S.; Xiao, T.; Li, Y.; Yu, X.; Wu, S.; Zhang, Y.; Lin, Y.; Zhao, N. A multi-modal smart chest patch for real-time cardiopulmonary monitoring and anomaly detection. Sci. China Mater. 2025, 68, 4413–4422. [Google Scholar] [CrossRef] [Scilit]
  12. Khan, M.; Pretty, C.G.; Amies, A.C.; Elliott, R.; Chiew, Y.S.; Shaw, G.M.; Chase, J.G. Analysing the effects of cold, normal, and warm digits on transmittance pulse oximetry. Biomed. Signal Process. Control 2016, 26, 34–41. [Google Scholar] [CrossRef] [Scilit]
  13. Buekers, J.; Theunis, J.; De Boever, P.; Vaes, A.W.; Koopman, M.; Janssen, E.V.; Wouters, E.F.; Spruit, M.A.; Aerts, J.M. Wearable finger pulse oximetry for continuous oxygen saturation measurements during daily home routines of patients with chronic obstructive pulmonary disease (COPD) over one week: Observational study. JMIR mHealth uHealth 2019, 7, e12866. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  14. Morgado Areia, C.; Santos, M.; Vollam, S.; Pimentel, M.; Young, L.; Roman, C.; Ede, J.; Piper, P.; King, E.; Gustafson, O.; et al. A chest patch for continuous vital sign monitoring: Clinical validation study during movement and controlled hypoxia. J. Med. Internet Res. 2021, 23, e27547. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  15. Chan, M.; Ganti, V.G.; Heller, J.A.; Abdallah, C.A.; Etemadi, M.; Inan, O.T. Enabling continuous wearable reflectance pulse oximetry at the sternum. Biosensors 2021, 11, 521. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  16. Gehring, H.; Hornberger, C.; Matz, H.; Konecny, E.; Schmucker, P. The effects of motion artifact and low perfusion on the performance of a new generation of pulse oximeters in volunteers undergoing hypoxemia. Respir. Care 2002, 47, 48–60. [Google Scholar] [CrossRef] [Scilit]
  17. Park, J.; Seok, H.S.; Kim, S.S.; Shin, H. Photoplethysmogram analysis and applications: An integrative review. Front. Physiol. 2022, 12, 808451. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  18. May, J.M.; Mejía-Mejía, E.; Nomoni, M.; Budidha, K.; Choi, C.; Kyriacou, P.A. Effects of contact pressure in reflectance photoplethysmography in an in vitro tissue-vessel phantom. Sensors 2021, 21, 8421. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  19. Elgendi, M.; Norton, I.; Brearley, M.; Abbott, D.; Schuurmans, D. Systolic peak detection in acceleration photoplethysmograms measured from emergency responders in tropical conditions. PLoS ONE 2013, 8, e76585. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  20. Stuban, N.; Niwayama, M. Optimal filter bandwidth for pulse oximetry. Rev. Sci. Instrum. 2012, 83, 104708. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  21. Mejia-Mejia, E.; May, J.M.; Kyriacou, P.A. Effect of filtering of photoplethysmography signals in pulse rate variability analysis. In 2021 43rd Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC); IEEE: New York, NY, USA, 2021; pp. 5500–5503. [Google Scholar]
  22. Scardulla, F.; D’Acquisto, L.; Colombarini, R.; Hu, S.; Pasta, S.; Bellavia, D. A study on the effect of contact pressure during physical activity on photoplethysmographic heart rate measurements. Sensors 2020, 20, 5052. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  23. Poets, C.; Stebbens, V. Detection of movement artifact in recorded pulse oximeter saturation. Eur. J. Pediatr. 1997, 156, 808–811. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  24. Alzahrani, A.; Hu, S.; Azorin-Peris, V. A comparative study of physiological monitoring with a wearable opto-electronic patch sensor (OEPS) for motion reduction. Biosensors 2015, 5, 288–307. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  25. Moscato, S.; Lo Giudice, S.; Massaro, G.; Chiari, L. Wrist photoplethysmography signal quality assessment for reliable heart rate estimate and morphological analysis. Sensors 2022, 22, 5831. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  26. Nathan, V.; Jafari, R. Particle filtering and sensor fusion for robust heart rate monitoring using wearable sensors. IEEE J. Biomed. Health Inform. 2017, 22, 1834–1846. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  27. Maeda, Y.; Sekine, M.; Tamura, T.; Moriya, A.; Suzuki, T.; Kameyama, K. Comparison of reflected green light and infrared photoplethysmography. In Proceedings of the 2008 30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Vancouver, BC, Canada, 20–24 August 2008; pp. 2270–2272. [Google Scholar]
  28. Lee, J.; Matsumura, K.; Yamakoshi, K.i.; Rolfe, P.; Tanaka, S.; Yamakoshi, T. Comparison between red, green and blue light reflection photoplethysmography for heart rate monitoring during motion. In Proceedings of the 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Osaka, Japan, 3–7 July 2013; pp. 1724–1727. [Google Scholar]
  29. Moco, A.V.; Stuijk, S.; De Haan, G. New insights into the origin of remote PPG signals in visible light and infrared. Sci. Rep. 2018, 8, 8501. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  30. Nawar, A.; Kilic, O.S.; Rahman, F.N.; Chen, C.; Berkebile, J.A.; Chan, M.; Shah, A.J.; Inan, O.T. Elucidating the reactivity of reflectance-based core-body photoplethysmogram to posture and respiratory changes. IEEE Sens. J. 2025, 25, 35447–35459. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  31. Shin, H.S.; Lee, C.; Lee, M. Adaptive threshold method for the peak detection of photoplethysmographic waveform. Comput. Biol. Med. 2009, 39, 1145–1152. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  32. Maeda, Y.; Sekine, M.; Tamura, T. Relationship between measurement site and motion artifacts in wearable reflected photoplethysmography. J. Med. Syst. 2011, 35, 969–976. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  33. Lee, J.; Kim, M.; Park, H.K.; Kim, I.Y. Motion artifact reduction in wearable photoplethysmography based on multi-channel sensors with multiple wavelengths. Sensors 2020, 20, 1493. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  34. Masinelli, G.; Dell’Agnola, F.; Valdés, A.A.; Atienza, D. SPARE: A spectral peak recovery algorithm for PPG signals pulsewave reconstruction in multimodal wearable devices. Sensors 2021, 21, 2725. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  35. Song, M.H.; Cho, S.P.; Kim, W.; Lee, K.J. New real-time heartbeat detection method using the angle of a single-lead electrocardiogram. Comput. Biol. Med. 2015, 59, 73–79. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  36. Krasteva, V.; Jekova, I. QRS template matching for recognition of ventricular ectopic beats. Ann. Biomed. Eng. 2007, 35, 2065–2076. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  37. Hong, S.; Heo, J.; Park, K.S. Signal quality index based on template cross-correlation in multimodal biosignal chair for smart healthcare. Sensors 2021, 21, 7564. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  38. Karlen, W.; Kobayashi, K.; Ansermino, J.M.; Dumont, G.A. Photoplethysmogram signal quality estimation using repeated Gaussian filters and cross-correlation. Physiol. Meas. 2012, 33, 1617–1629. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  39. Han, H.; Kim, J. Artifacts in wearable photoplethysmographs during daily life motions and their reduction with least mean square based active noise cancellation method. Comput. Biol. Med. 2012, 42, 387–393. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  40. Shahbakhti, M.; Bosch, A.T.; Long, X.; van Dijk, J.P.; Thijs, R.D. Pulse-based photoplethysmography quality assessment improves wearable seizure detection performance. Epilepsia Open 2026, 11, 1149–1158. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  41. Chan, E.D.; Chan, M.M.; Chan, M.M. Pulse oximetry: Understanding its basic principles facilitates appreciation of its limitations. Respir. Med. 2013, 107, 789–799. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  42. Stubán, N.; Masatsugu, N. Non-invasive calibration method for pulse oximeters. Period. Polytech. Electr. Eng. (Arch.) 2008, 52, 91–94. [Google Scholar] [CrossRef] [Scilit]
  43. Blaney, G.; Sassaroli, A.; Fantini, S. Critical analysis of the relationship between arterial saturation and the ratio-of-ratios used in pulse oximetry. J. Biomed. Opt. 2024, 29, S33313. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  44. Severinghaus, J.W. Recent developments in pulse oximetry. Anesthesiology 1992, 76, 1018–1038. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  45. MacLeod, D.; Cortinez, L.; Keifer, J.; Cameron, D.; Wright, D.; White, W.; Moretti, E.; Radulescu, L.; Somma, J. The desaturation response time of finger pulse oximeters during mild hypothermia. Anaesthesia 2005, 60, 65–71. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  46. Shi, C.; Goodall, M.; Dumville, J.; Hill, J.; Norman, G.; Hamer, O.; Clegg, A.; Watkins, C.L.; Georgiou, G.; Hodkinson, A.; et al. The accuracy of pulse oximetry in measuring oxygen saturation by levels of skin pigmentation: A systematic review and meta-analysis. BMC Med. 2022, 20, 267. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  47. O’driscoll, B.; Howard, L.; Earis, J.; Mak, V.; Group, B.E.O.G.D. British Thoracic Society Guideline for oxygen use in adults in healthcare and emergency settings. BMJ Open Respir. Res. 2017, 4, e000170. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  48. Feiner, J.R.; Severinghaus, J.W.; Bickler, P.E. Dark skin decreases the accuracy of pulse oximeters at low oxygen saturation: The effects of oximeter probe type and gender. Anesth. Analg. 2007, 105, S18–S23. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  49. Lin, J.; Pan, S.; Ma, L.; Ye, J.; Liu, C.; Wang, F.; Zhang, X. Agreement and Accuracy of a Commercial Pulse Oximeter Compared with Arterial Oxygen Saturation During Controlled Hypoxemia in Healthy Volunteers. Med. Devices Evid. Res. 2026, 19, 626574. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  50. Langton, J.; Hanning, C. Effect of motion artefact on pulse oximeters: Evaluation of four instruments and finger probes. Br. J. Anaesth. 1990, 65, 564–570. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  51. Preejith, S.; Ravindran, A.S.; Hajare, R.; Joseph, J.; Sivaprakasam, M. A wrist worn SpO2 monitor with custom finger probe for motion artifact removal. In Proceedings of the 2016 38th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Orlando, FL, USA, 16–20 August 2016; pp. 5777–5780. [Google Scholar]
  52. Liu, Y.; Arnaert, A.; da Costa, D.; Sumbly, P.; Debe, Z.; Charbonneau, S. Experiences of patients with chronic obstructive pulmonary disease using the Apple Watch Series 6 versus the traditional finger pulse oximeter for home SpO2 self-monitoring: Qualitative study part 2. JMIR Aging 2023, 6, e41539. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  53. Wei, J.C.; van den Broek, T.J.; van Baardewijk, J.U.; van Stokkum, R.; Kamstra, R.J.; Rikken, L.; Gijsbertse, K.; Uzunbajakava, N.E.; van den Brink, W.J. Validation and user experience of a dry electrode based Health Patch for heart rate and respiration rate monitoring. Sci. Rep. 2024, 14, 23098. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  54. Baty, F.; Cvetkovic, D.; Boesch, M.; Bauer, F.; Adão Martins, N.R.; Rossi, R.M.; Schoch, O.D.; Annaheim, S.; Brutsche, M.H. Validation of a textile-based wearable measuring electrocardiogram and breathing frequency for sleep apnea monitoring. Sensors 2024, 24, 6229. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  55. Leenen, J.P.; Leerentveld, C.; van Dijk, J.D.; van Westreenen, H.L.; Schoonhoven, L.; Patijn, G.A. Current evidence for continuous vital signs monitoring by wearable wireless devices in hospitalized adults: Systematic review. J. Med. Internet Res. 2020, 22, e18636. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  56. Songthawornpong, N.; Vijayakumar, T.; Jensen, M.S.V.; Elvekjaer, M.; Sørensen, H.B.; Aasvang, E.K.; Meyhoff, C.S.; Eriksen, V.R. Continuous monitoring of vital signs after hospital discharge: A feasibility study. Patient Saf. 2023, 5, 53–63. [Google Scholar] [CrossRef] [Scilit]
  57. van Melzen, R.; Haveman, M.E.; Schuurmann, R.C.; van Amsterdam, K.; El Moumni, M.; Tabak, M.; Struys, M.M.; de Vries, J.P.P. Validity and reliability of wearable sensors for continuous postoperative vital signs monitoring in patients recovering from trauma surgery. Sensors 2024, 24, 6379. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  58. Hayes, M.J.; Smith, P.R. A new method for pulse oximetry possessing inherent insensitivity to artifact. IEEE Trans. Biomed. Eng. 2001, 48, 452–461. [Google Scholar] [CrossRef] [PubMed]
  59. Alkhoury, L.; Choi, J.w.; Wang, C.; Rajasekar, A.; Acharya, S.; Mahoney, S.; Shender, B.S.; Hrebien, L.; Kam, M. Heart-rate tuned comb filters for processing photoplethysmogram (PPG) signals in pulse oximetry. J. Clin. Monit. Comput. 2021, 35, 797–813. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  60. Sukor, J.A.; Redmond, S.; Lovell, N. Signal quality measures for pulse oximetry through waveform morphology analysis. Physiol. Meas. 2011, 32, 369–384. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  61. Sabeti, E.; Reamaroon, N.; Mathis, M.; Gryak, J.; Sjoding, M.; Najarian, K. Signal quality measure for pulsatile physiological signals using morphological features: Applications in reliability measure for pulse oximetry. Inform. Med. Unlocked 2019, 16, 100222. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  62. Bickler, P.E.; Feiner, J.R.; Severinghaus, J.W. Effects of skin pigmentation on pulse oximeter accuracy at low saturation. Anesthesiology 2005, 102, 715–719. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  63. Sjoding, M.W.; Dickson, R.P.; Iwashyna, T.J.; Gay, S.E.; Valley, T.S. Racial bias in pulse oximetry measurement. N. Engl. J. Med. 2020, 383, 2477–2478. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  64. Fawzy, A.; Ali, H.; Dziedzic, P.H.; Potu, N.; Calvillo, E.; Golden, S.H.; Iwashyna, T.J.; Suarez, J.I.; Hager, D.N.; Garibaldi, B.T. Skin pigmentation and pulse oximeter accuracy in the intensive care unit: A pilot prospective study. Am. J. Respir. Crit. Care Med. 2024, 210, 355–358. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  65. ISO 80601-2-61:2026; Medical Electrical Equipment—Part 2-61: Particular Requirements for Basic Safety and Essential Performance of Pulse Oximeter Equipment. International Organization for Standardization: Geneva, Switzerland, 2026.
  66. U.S. Food and Drug Administration. Pulse Oximeters for Medical Purposes—Non-Clinical and Clinical Performance Testing, Labeling, and Premarket Submission Recommendations; Draft Guidance for Industry and Food and Drug Administration Staff; U.S. Food and Drug Administration: Silver Spring, MD, USA, 2025.
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Article Metrics

Citations

Article Access Statistics

Multiple requests from the same IP address are counted as one view.