Development of Intraoperative Near-Infrared Fluorescence Imaging System Using a Dual-CMOS Single Camera

,
,
,
,
,
,
,
,
…
1
Intelligent Photonic IoT Research Center, Korea Photonics Technology Institute, Gwangju 61007, Korea
2
Department of Biomedical Science & Engineering, Gwangju Institute of Science and Technology, Gwangju 61005, Korea
3
Optical Precision Measurement Research Center, Korea Photonics Technology Institute, Gwangju 61007, Korea
4
Medical Laser Research Center, Dankook University, Cheonan 31116, Korea
This article belongs to the Collection Biomedical Imaging & Instrumentation

Abstract

We developed a single-camera-based near-infrared (NIR) fluorescence imaging device using indocyanine green (ICG) NIR fluorescence contrast agents for image-induced surgery. In general, a fluorescent imaging system that simultaneously provides color and NIR images uses two cameras, which is disadvantageous because it increases the imaging head of the system. Recently, a single-camera-based NIR optical imaging device with quantum efficiency partially extended to the NIR region was developed to overcome this drawback. The system used RGB_NIR filters for camera sensors to provide color and NIR images simultaneously; however, the sensitivity and resolution of the infrared images are reduced by 1/4, and the exposure time and gain cannot be set individually when acquiring color and NIR images. Thus, to overcome these shortcomings, this study developed a compact fluorescent imaging system that uses a single camera with two complementary metal–oxide semiconductor (CMOS) image sensors. Sensitivity and signal-to-background ratio were measured according to the concentrations of ICG solution, exposure time, and camera gain to evaluate the performance of the imaging system. Consequently, the clinical applicability of the system was confirmed through the toxicity analysis of the light source and in vivo testing.

Article Metrics

Citations

Article Access Statistics

Multiple requests from the same IP address are counted as one view.