Abstract
Post-acquisition processing settings can alter quantitative mass spectrometry results and, potentially, biomarker evaluation. We reprocessed the same acquired centroid micro-liquid chromatography–quadrupole time-of-flight mass spectrometry (micro-LC-QTOF-MS) data from 300 Japanese adults using mass extraction window settings of 10, 20, 35, 50, 100, and 200 ppm. Plasma mid-regional proadrenomedullin (MR-proADM) measurements were compared with the KRYPTOR immunoassay. A reduced estimated glomerular filtration rate (eGFR; <60 mL/min/1.73 m2) was present in 89 participants. Receiver operating characteristic analyses retained all participants, with below-quantification results tied at the lowest rank. QTOF areas under the curve were 0.395, 0.560, 0.672, 0.685, 0.688, and 0.595, respectively, compared with 0.758 for KRYPTOR. Age- and sex-adjusted odds ratios per concentration doubling (n = 273) were 2.34–2.83 at 35–200 ppm. Substantial concentration differences persisted between methods. Reprocessed calibrators showed that the 10-ppm setting retained only 24–59% of the analyte signal at ≤0.5 ng/mL, whereas the 200-ppm setting admitted co-extracted signals. Thus, discrimination depended on the extraction setting: intermediate settings outperformed the narrowest and widest settings, but every QTOF setting discriminated less well than the immunoassay, and the concentrations were not interchangeable. These exploratory findings emphasize processing-dependent biomarker evaluation and require independent analytical and clinical validation.
1. Introduction
Biomarker-based stratification plays an increasingly important role in biomedical research and clinical decision-making [1]. Advances in analytical technologies have enabled sensitive and high-throughput quantification of circulating biomarkers, thereby facilitating their use across a wide range of disease areas [2]. In many studies, however, analytical conditions tend to be treated as of secondary importance once sufficient analytical sensitivity is achieved. In practice, variability introduced by analytical parameters can propagate into downstream interpretation, directly influencing apparent biomarker performance and, in some cases, leading to inconsistent or irreproducible conclusions across studies [3,4]. A growing body of literature has highlighted challenges in biomarker reproducibility, particularly in studies employing high-dimensional analytical platforms such as mass spectrometry [5]. Even when identical biological markers and clinical endpoints are investigated, the reported effect sizes and discriminative performance metrics often differ substantially between studies [6]. Although biological heterogeneity is frequently explained as a primary factor for such discrepancies, analytical variability represents a less visible but equally important contributor [7]. Differences in instrumental configuration, data acquisition parameters, and signal processing strategies can alter quantitative distributions, which in turn may affect downstream statistical evaluation and classification performance. Receiver operating characteristic (ROC) analysis is widely used to assess the discriminative ability of biomarkers [8]. Although ROC curves are often interpreted as reflections of underlying biological differences, these metrics are inherently dependent on the quality and characteristics of analytical measurements [9]. Factors such as signal-to-noise ratio, spectral interference, and quantitative precision directly influence biomarker distributions and classification boundaries [10]. Consequently, ROC-based evaluation is not analytically agnostic, and differences in analytical methodology may translate into apparent differences in biomarker performance even when the underlying biological signal is unchanged.
Mid-regional proadrenomedullin (MR-proADM) is a stable surrogate marker of adrenomedullin release [11] and has attracted attention as a biomarker of vascular function [12], inflammation [13], and organ dysfunction [14]. It has been investigated in cardiovascular disease [15], community-acquired pneumonia [16], and renal dysfunction [17]. MR-proADM is commonly quantified using immunoassays [11,18], which provide a clinically established comparator for alternative measurements. Mass spectrometry offers structural selectivity and flexibility for peptide measurement [19,20]. Our group previously described a micro-liquid chromatography–quadrupole time-of-flight mass spectrometry (micro-LC-QTOF-MS) assay for plasma MR-proADM [21].
Instrumental resolving power, mass error, and the mass interval used to extract an ion chromatogram are distinct quantities. The mass extraction window is a post-acquisition processing parameter, not a measure of resolving power. Previous work has shown that quantitative high-resolution mass spectrometry results can depend on the extraction window, the use of centroid or continuum data, the extraction center, and software-specific processing [22]. The remaining question addressed here is how these processing choices relate to downstream biomarker discrimination in a common clinical dataset.
In this study, we compared MR-proADM concentration estimates obtained by reprocessing identical acquired raw data using six mass extraction windows. We examined concentration distributions, differences from the KRYPTOR immunoassay, and discrimination between participants with estimated glomerular filtration rate (eGFR) < 60 and ≥60 mL/min/1.73 m2. The objective was to characterize processing dependence within one analytical workflow and cohort, rather than to validate a new diagnostic test or determine a universally optimal extraction window.
2. Results
2.1. Participant Characteristics and Quantifiable Results
The study included 300 participants aged 39–74 years: 211 with eGFR ≥ 60 and 89 with eGFR < 60 mL/min/1.73 m2. The latter group was older and had lower eGFR, whereas the other characteristics reported in Table 1 did not differ significantly between groups. The median eGFR in the reduced-eGFR group was 55.1 mL/min/1.73 m2 (interquartile range, 52.4–57.9).
Table 1.
Participant characteristics stratified by renal function.
At 10 ppm, three results were recorded as zero to indicate concentrations below the limit of detection (LOD), and 23 additional positive numerical outputs were below the lower limit of quantification (LLOQ) of 0.05 ng/mL. One result at 20 ppm was also below the LLOQ; no such results occurred at 35, 50, 100, or 200 ppm. All 300 KRYPTOR values exceeded its reported LLOQ of 1.18 ng/mL. Consequently, concentration-based comparisons used 274 participants at 10 ppm, 299 at 20 ppm, and 300 at the other windows. The common quantifiable subset contained 273 participants (199 with eGFR ≥ 60 and 74 with eGFR < 60); the participant with a below-LLOQ result at 20 ppm differed from those at 10 ppm (Table S1). For this participant, the concentration estimates were 0.0837 ng/mL at 10 ppm and 0.0272 ng/mL at 20 ppm, and quantifiable values were obtained at the four wider settings; the analyte peak areas of this participant were nearly identical at 10 and 20 ppm (4.1 and 4.2 counts, with an internal-standard area of 594.5), and the different estimates arose from the window-specific calibration curves, whose intercepts were −0.0009 at 10 ppm and +0.0045 at 20 ppm (Section 2.6).
2.2. Concentration Distributions and Correlations
Natural-log-transformed concentrations showed extraction-window-dependent differences in their distributions (Figure 1). Cohen’s d, calculated as the reduced-eGFR minus comparison-group mean difference on the natural-log scale divided by the pooled standard deviation, was −0.20, 0.23, 0.61, 0.66, 0.67, and 0.36 at 10, 20, 35, 50, 100, and 200 ppm, respectively, and 0.94 for KRYPTOR. These estimates describe between-participant group separation among quantifiable results; they do not measure analytical repeatability.
Figure 1.
Distributions of quantifiable plasma MR-proADM concentrations by renal-function group and mass extraction window. Concentrations are shown as ln[concentration/(1 ng/mL)]. White circles and thick bars indicate medians and interquartile ranges. Cohen’s d was calculated on the natural-log scale as eGFR < 60 minus eGFR ≥ 60. Results below the LLOQ were excluded without concentration substitution. Sample sizes were 274 at 10 ppm (200/74 by renal-function group), 299 at 20 ppm (210/89), and 300 for the remaining methods (211/89). The ppm values denote TargetLynx XS processing settings, not instrumental resolving power.
Pearson correlations between the natural-log-transformed QTOF and KRYPTOR concentrations were 0.108, 0.403, 0.485, 0.500, 0.496, and 0.386 at 10, 20, 35, 50, 100, and 200 ppm, respectively (Figure 2). The largest correlation was observed at 50 ppm, but the maximum value was 0.500. Correlation was considered separately from agreement between concentrations.
Figure 2.
Associations between quantifiable QTOF and KRYPTOR MR-proADM concentrations. Panels correspond to the six mass extraction window settings. Both axes show natural-log-transformed concentrations, ln[concentration/(1 ng/mL)]. Points are grouped by eGFR ≥ 60 or <60 mL/min/1.73 m2. Lines show fitted linear relationships, and Pearson’s r is reported for each panel. The sample sizes are 274 at 10 ppm, 299 at 20 ppm, and 300 at each other window. These are method-specific quantifiable subsets; the regression lines and correlation coefficients do not establish measurement agreement.
2.3. Concentration Differences Between Methods
For quantifiable pairs, mean QTOF-minus-KRYPTOR differences were −1.988, −1.841, −1.723, −1.707, −1.666, and −1.369 ng/mL at 10, 20, 35, 50, 100, and 200 ppm, respectively. Mean individual percentage differences, calculated relative to KRYPTOR, were −86.9%, −80.5%, −75.4%, −74.7%, −72.9%, and −59.3% (Table S2). The bootstrap CIs of these mean differences are narrow (Table S2) because the same 300 participants were reprocessed at every setting and the differences are paired; they describe the precision of the mean difference in this cohort and do not indicate the analytical reproducibility of either method. Thus, substantial differences persisted at 35–100 ppm despite their comparatively favorable discrimination.
Geometric mean QTOF/KRYPTOR ratios ranged from 0.103 to 0.389. At 50 and 100 ppm, the ratios were 0.236 and 0.255, with approximate 95% ratio limits of agreement of 0.111–0.502 and 0.125–0.520, respectively. Log-ratio plots showed concentration-related trends (Figure S1), and the method-specific restriction to quantifiable results also limited interpretation. The limits of agreement are descriptive approximations, not validated clinical acceptance limits or proof of interchangeability [23]. KRYPTOR was not assumed to represent an error-free true concentration.
2.4. Discrimination of Reduced eGFR
The primary ROC analysis retained all 300 participants and used a common direction in which higher reporting scores indicated reduced eGFR. All results below the relevant LLOQ were treated as one lowest-ranked category within each method. The KRYPTOR area under the curve (AUC) was 0.758 (95% confidence interval [CI], 0.697–0.819). QTOF AUCs were 0.395, 0.560, 0.672, 0.685, 0.688, and 0.595 at 10, 20, 35, 50, 100, and 200 ppm, respectively (Figure 3 and Table 2). The 95% CIs of the six QTOF AUCs each spanned approximately 0.13–0.14 and overlapped almost completely among the 35-, 50-, and 100-ppm settings, so the ordering among these three settings is not established by these data. The 10-ppm AUC was below 0.5 under this common direction; its direction was not reversed to maximize the AUC. The direction (higher score positive for reduced eGFR) was specified before the analysis on the basis of the known increase in circulating MR-proADM with declining renal function [17] and was applied identically to KRYPTOR and to all six QTOF settings. An AUC below 0.5 therefore indicates that, at 10 ppm, the reporting scores of participants with reduced eGFR tended to rank below those of the comparison group. Two features of the 10-ppm output contributed. First, 15 of the 26 participants with below-LLOQ results, who shared the lowest tied rank, belonged to the reduced-eGFR group (15 of 89), compared with 11 of the 211 participants in the comparison group. Second, the ordering among quantifiable results was also weakly inverted (AUC 0.440 in the method-specific quantifiable pairs; Table S1, panel C), in line with the negative Cohen’s d at 10 ppm (Figure 1).
Figure 3.
Receiver operating characteristic curves for discrimination of eGFR <60 mL/min/1.73 m2. All 300 participants were retained. Within each method, below-LLOQ results were assigned one lowest tied reporting category, while quantifiable results retained their concentration ordering. Higher scores were defined as positive for every method. These curves therefore describe LLOQ-aware reporting scores, not estimated latent concentrations below the LLOQ. The diagonal indicates chance-level discrimination.
Table 2.
Discrimination of reduced eGFR and paired AUC differences.
AUC differences were defined as ΔAUC = AUC(KRYPTOR) − AUC(QTOF). All six differences were positive. After Holm correction of the six paired DeLong comparisons, the adjusted p values were <0.0001 at 10 and 20 ppm, 0.0287 at 35 ppm, 0.0538 at both 50 and 100 ppm, and <0.0001 at 200 ppm. The nonsignificant comparisons at 50 and 100 ppm do not establish equivalence or non-inferiority.
2.5. Adjusted Associations and Sensitivity Analyses
In the common quantifiable subset of 273 participants, the age- and sex-adjusted odds ratios per doubling of concentration were 17.73 (95% CI, 5.12–61.36) for KRYPTOR and 0.87 (0.66–1.15), 1.33 (0.93–1.90), 2.34 (1.46–3.74), 2.52 (1.55–4.10), 2.83 (1.69–4.74), and 2.55 (1.36–4.76) for QTOF at 10, 20, 35, 50, 100, and 200 ppm, respectively (Table S3 and Figure S2). After adjustment, the 95% CIs excluded 1 at 35, 50, 100, and 200 ppm, and adding concentration to the age-and-sex model remained significant after Holm correction (adjusted p = 0.0010, 0.0005, 0.0002, and 0.0084, respectively). In contrast, the CIs at 10 and 20 ppm included 1 (adjusted p = 0.3295 and 0.2266). These estimates represent adjusted associations in a selected quantifiable population, not age-adjusted single-marker AUCs.
The numerically higher QTOF AUCs at 35–100 ppm were also observed in sensitivity analyses, although statistical significance depended on participant eligibility. At 50 and 100 ppm, the Holm-adjusted p values were 0.0728 in the common quantifiable subset (n = 273), compared with 0.0538 in the primary analysis. An exploratory analysis retaining positive below-LLOQ numerical outputs (n = 297) gave 0.0398 for both comparisons; these numerical outputs were not treated as validated concentrations. ROC and covariate-adjustment sensitivity analyses are presented in Tables S1 and S3.
2.6. Window-Specific Calibration, Quality Control, and Chromatographic Behavior
To characterize the analytical behavior of each setting, the duplicate calibration standards (eight levels, 0.05–100 ng/mL), the analyte-free calibration blanks, and the QC samples of the study batch were reprocessed at the six settings (Table 3 and Table S4). The calibration slope (0.0937–0.0950) and r2 (0.9928–0.9941) were nearly identical across settings, because the slope is determined mainly by the higher calibrators, whose peak areas were retained almost completely at every setting (87–99% of the 200-ppm area at ≥2 ng/mL for 10 ppm and ≥94% for the other settings; Figure 4a and Figure S4). Calibrator replicates whose back-calculated concentration fell outside the acceptance tolerance were excluded from the fit by the software (three or four of the 16 replicates at each setting, mostly at 0.05–0.5 and 100 ng/mL; open symbols in Figure S4; Table S4); with all 16 replicates included, the slopes were 0.088–0.092 and r2 0.985–0.993, the lowest values being those of the 10-ppm setting. The intercept, however, increased from −0.0009 at 10 ppm to 0.0045–0.0086 at the wider settings, and the analyte-free blank showed a response of 10–58% of the LLOQ response at 20–200 ppm but none at 10 ppm. At the low end, the narrow settings lost a large and concentration-dependent fraction of the analyte signal: at 10 ppm, the peak area was 24%, 51%, and 59% of the 200-ppm area at 0.05, 0.2, and 0.5 ng/mL, respectively; at 20 ppm, 63%, 68%, and 88%; and at 35 ppm, 74%, 76%, and 92% (Figure 4a); the internal standard was extracted with the 35-ppm window at every setting (Section 4.2). Consequently, back-calculated accuracy at the two lowest calibrator levels was poor and inconsistent at all settings in this batch: one of the two 0.05 ng/mL calibrators back-calculated to 9–23% of nominal at 20–50 ppm and to below zero at 100 ppm, and both 0.2 ng/mL replicates were within ±20% of nominal only at 35 ppm (87% and 112%). The QC samples, which contained endogenous MR-proADM, gave accuracies of 93–103% (0.15 ng/mL spike), 94–95% (30 ng/mL), and 87–88% (75 ng/mL) relative to the endogenous plus spiked concentration, with CVs of 0.1–2.5% at 30 and 75 ng/mL and 2.8–18.6% at 0.15 ng/mL (n = 2 per level); the apparent endogenous concentration of the QC plasma rose from 0.46 ng/mL at 10 and 20 ppm to 0.53, 0.48, 0.66, and 0.86 ng/mL at 35, 50, 100, and 200 ppm.
Table 3.
Window-specific calibration and quality-control performance of the study batch after reprocessing at the six mass-window settings.
Figure 4.
Window-specific analytical behavior. (a) Peak area of the analyte in the duplicate calibration standards at each mass-window setting relative to the 200-ppm setting (mean of duplicates) versus nominal concentration; the internal standard was extracted with the 35-ppm window at every setting. (b) Extracted ion chromatograms of the MR-proADM product ion (m/z 833.0412) for one participant sample reprocessed at the six settings, with the concentration estimate and apex intensity at each setting; the internal-standard chromatogram (m/z 838.8527, divided by 10) is shown as a dashed line. (c) Centroid of the target isotopologue of the MR-proADM product ion in a single scan at the peak apex (5.244 min), with the half-widths of the extraction windows drawn around the target m/z; the observed centroid (833.0358, −6.5 ppm) lies outside the ±5 ppm window of the 10-ppm setting. (d) The same scan for the internal-standard product ion (observed 838.8541, +1.7 ppm), shown for reference; the internal standard was extracted at 35 ppm at every setting. The ±100 ppm half-width of the 200-ppm setting lies outside the plotted range of (c,d). (e) Analyte peak area of the 300 participant samples at each setting relative to the 35-ppm setting (median, interquartile range, and 5th–95th percentiles). (f) Fraction of the 200-ppm analyte area retained at 10 ppm versus the internal-standard area of the same injection for the 300 participant samples.
To characterize the analytical behavior of each setting, the duplicate calibration standards (eight levels, 0.05–100 ng/mL), the analyte-free calibration blanks, and the QC samples of the study batch were reprocessed at the six settings (Table 3 and Table S4). The calibration slope (0.0937–0.0950) and r2 (0.9928–0.9941) were nearly identical across settings, because the slope is determined mainly by the higher calibrators, whose peak areas were retained almost completely at every setting (87–99% of the 200-ppm area at ≥2 ng/mL for 10 ppm and ≥94% for the other settings; Figure 4a and Figure S4). Calibrator replicates whose back-calculated concentration fell outside the acceptance tolerance were excluded from the fit by the software (three or four of the 16 replicates at each setting, mostly at 0.05–0.5 and 100 ng/mL; open symbols in Figure S4; Table S4); with all 16 replicates included, the slopes were 0.088–0.092 and r2 0.985–0.993, the lowest values being those of the 10-ppm setting. The intercept, however, increased from −0.0009 at 10 ppm to 0.0045–0.0086 at the wider settings, and the analyte-free blank showed a response of 10–58% of the LLOQ response at 20–200 ppm but none at 10 ppm. At the low end, the narrow settings lost a large and concentration-dependent fraction of the analyte signal: at 10 ppm, the peak area was 24%, 51%, and 59% of the 200-ppm area at 0.05, 0.2, and 0.5 ng/mL, respectively; at 20 ppm, 63%, 68%, and 88%; and at 35 ppm, 74%, 76%, and 92% (Figure 4a); the internal standard was extracted with the 35-ppm window at every setting (Section 4.2). Consequently, back-calculated accuracy at the two lowest calibrator levels was poor and inconsistent at all settings in this batch: one of the two 0.05 ng/mL calibrators back-calculated to 9–23% of nominal at 20–50 ppm and to below zero at 100 ppm, and both 0.2 ng/mL replicates were within ±20% of nominal only at 35 ppm (87% and 112%). The QC samples, which contained endogenous MR-proADM, gave accuracies of 93–103% (0.15 ng/mL spike), 94–95% (30 ng/mL), and 87–88% (75 ng/mL) relative to the endogenous plus spiked concentration, with CVs of 0.1–2.5% at 30 and 75 ng/mL and 2.8–18.6% at 0.15 ng/mL (n = 2 per level); the apparent endogenous concentration of the QC plasma rose from 0.46 ng/mL at 10 and 20 ppm to 0.53, 0.48, 0.66, and 0.86 ng/mL at 35, 50, 100, and 200 ppm.
Extracted ion chromatograms of one participant sample (1.17–1.35 ng/mL across settings) showed the same analyte peak at all settings, with an apex intensity at 10 ppm of 89% of that at 200 ppm, whereas the 200-ppm setting additionally extracted two signals at 5.6 and 6.3 min that decreased progressively with narrower windows and were minor or absent at 10–35 ppm (Figure 4b). In a single product-ion scan at the apex of this peak, the six isotopologue centroids of the MR-proADM product ion deviated from their theoretical m/z by −1.6 to −8.2 ppm (mean −4.9 ppm); the centroid of the target isotopologue (m/z 833.0358, −6.5 ppm) lay outside the ±5 ppm window of the 10-ppm setting but inside all wider windows, whereas the internal-standard centroid (m/z 838.8541, +1.7 ppm) lay inside every window (Figure 4c,d).
Across the 300 participant samples, the analyte peak area relative to the 35-ppm (validated) setting had medians of 38% (interquartile range, 19–59%) at 10 ppm, 85% (67–93%) at 20 ppm, 104% (102–108%) at 50 ppm, 113% (107–122%) at 100 ppm, and 161% (133–196%) at 200 ppm (Figure 4e). The fraction of the 200-ppm area retained at 10 ppm (median 23%, interquartile range 11–34%) correlated with the internal-standard area of the same injection (Spearman ρ = 0.48, p < 0.0001) and with the absolute analyte area at 200 ppm (ρ = 0.42, p < 0.0001), but only weakly with the 200-ppm concentration estimate (ρ = 0.13, p = 0.025) (Figure 4f). The Spearman correlation between the concentration estimates at 10 ppm and at 200 ppm was 0.45, compared with 0.70 at 20 ppm and 0.74 at 35 ppm (n = 297 participants with a numerical output at 10 ppm; n = 300 otherwise). For the participant whose 20-ppm result fell below the LLOQ (Section 2.1), the analyte areas were 4.1, 4.2, 7.9, 15.5, 21.3, and 31.9 counts at 10, 20, 35, 50, 100, and 200 ppm with an internal-standard area of 594.5; the responses at 10 and 20 ppm (0.0069 and 0.0071) were nearly identical, and the lower 20-ppm estimate (0.027 versus 0.084 ng/mL) resulted from the positive intercept of the 20-ppm calibration curve (+0.0045 versus −0.0009 at 10 ppm) (Figure S3).
3. Discussion
This study examined how a post-acquisition mass extraction setting affects the evaluation of a peptide biomarker in a common set of plasma samples. All six QTOF results for each participant were derived from the same acquired centroid data. The observed changes therefore relate to the processing workflow rather than to separate acquisitions at different instrumental resolving powers. The principal finding is that extraction settings were associated with differences in MR-proADM concentration estimates, between-group separation, and discrimination of reduced eGFR.
The relationship between extraction window width and discrimination was non-monotonic. Settings of 35, 50, and 100 ppm yielded numerically higher AUCs than the narrower settings and the widest setting evaluated, but their AUCs remained below that of KRYPTOR. The absence of statistically significant differences at 50 and 100 ppm after multiple-comparison correction does not establish equivalent diagnostic performance. These settings should be regarded as comparatively favorable within this dataset, rather than a validated optimum or a substitute for the immunoassay. Moreover, the AUC CIs of the three intermediate settings overlapped almost completely, no hypothesis test compared one QTOF setting with another, and the study was not designed to rank settings; the comparison among settings is descriptive and exploratory, and the window-level AUC estimates carry substantial uncertainty.
Processing-related explanations are plausible but were not directly established. For centroid data, a narrow extraction window may omit a recorded analyte centroid when its measured m/z falls outside the extraction interval in a given scan. A wider window may include neighboring or background signals. The effects depend on the extraction center, mass error, data representation, and software implementation [22]. Narrowing a window does not itself introduce new neighboring ions into the accepted interval, so the lower-ppm findings cannot simply be attributed to increased interference, nor does re-extraction alter ion transmission, flight path, or the resolving power of the original acquisition. The contribution of signal exclusion or co-extracted ions would require direct inspection of chromatograms and spectra. Two further points bear on the expectation that the narrowest window should give the most selective, and hence the most accurate, result. First, the 10-ppm setting corresponds to ±5 ppm around the target m/z (Section 4.2), which is of the same order as the scan-to-scan mass error of centroided TOF data at modest ion counts [22]. The reprocessed calibrators showed the expected consequence: the 10-ppm setting retained only 24–59% of the analyte peak area at 0.05–0.5 ng/mL, 87–91% at 2–5 ng/mL, and 92–99% at 20–100 ng/mL (Section 2.6; Figure 4a). The calibration linearity was nevertheless almost identical across settings (slope 0.0937–0.0950, r2 0.9928–0.9941; Figure S4) because a weighted linear fit over 0.05–100 ng/mL is governed by the upper calibrators, whose signal was retained at every setting; the window dependence was confined to the intercept and to the accuracy of the lowest calibrators, that is, to the concentration range of the participant samples. Identical linearity therefore does not imply identical performance at low concentrations. The loss is thus concentrated at low ion counts, where ion statistics broaden the distribution of the centroid m/z across scans, so that individual scans fall outside a ±5 ppm window even when the mean mass error of the peak is small; the single-scan spectrum in Figure 4c, in which the target centroid lay 6.5 ppm below its theoretical value while the isotopologue centroids of the same scan scattered between −1.6 and −8.2 ppm, illustrates this directly. With a mean offset of about −5 ppm, a ±5 ppm window sits at the edge of the centroid distribution, which explains the strong dependence of the 10-ppm result on ion count; lock-mass correction was applied during acquisition (Section 4.2), and the behavior reflects the width of the window relative to the centroid distribution rather than drift of the mass axis. Because the internal standard was extracted with the fixed 35-ppm window at every setting (Section 4.2), narrowing the analyte window reduced the area ratio and the concentration estimate in proportion to the analyte signal lost, and in participant plasma, where the analyte (approximately 0.1–1 ng/mL) is measured at far lower ion counts than in the calibrators used for validation, the median area retained at 10 ppm was only 38% of that at 35 ppm (Figure 4e). Second, the magnitude of this loss depends on the absolute analyte ion count of each sample, not on its concentration alone. Internal-standard normalization corrects the area ratio for sample-specific ionization efficiency, but it cannot restore scans in which the analyte centroid was excluded; samples with a lower absolute analyte signal at a given concentration, for example, because of stronger matrix suppression or a lower instrument response for that injection, therefore lose proportionally more signal at narrow windows. In the participant samples, the fraction of the 200-ppm analyte area retained at 10 ppm correlated with the internal-standard area of the same injection (Spearman ρ = 0.48), a proxy for the absolute response of that injection, and with the absolute analyte area (ρ = 0.42), but only weakly with the concentration estimate (ρ = 0.13) (Figure 4f), which supports an ion-count-dependent loss. Narrow-window estimates are therefore sensitive to variation in absolute signal between injections that the internal standard would otherwise correct, and their ordering among participants can differ from that of the wider settings even though the same acquired data were processed, as the correlation of only 0.45 between the 10-ppm and 200-ppm estimates shows. At the other extreme, the 200-ppm setting (±100 ppm) admitted co-extracted signals near the analyte retention time in participant plasma (Figure 4b), raised the apparent endogenous concentration of the QC plasma almost twofold relative to 10 ppm, and increased the back-calculated accuracy of the 0.2 ng/mL calibrators to 125–142%, indicating inclusion of background and neighboring ions. The quantifier product ion, the precursor charge state, and the sample preparation were selected and validated in the previous report at the 35-ppm setting [21]; the accuracy (92.5–106.5%) and precision (CV < 5.8%) reported there refer to the method’s own calibrators and to that setting, and they do not address agreement with the immunoassay, which depends on the factors discussed below.
Concentration agreement and clinical discrimination were distinct outcomes. QTOF concentrations were substantially lower than KRYPTOR concentrations at all extraction windows, including 35–100 ppm. Neither the observed correlations nor the relatively similar AUCs demonstrate that the concentrations can be used interchangeably. The modest correlations with KRYPTOR (Pearson’s r ≤ 0.500) are also consistent with single measurements by each method in a cohort with a limited concentration range: the between-method correlation is bounded by the biological spread of concentrations relative to the combined analytical variation in the two methods, and a low correlation does not by itself identify which method is less precise. The immunoassay was used as a comparator, not as a true-value standard; the differences cannot be assigned entirely to an error in either method. The concentration-related trends in log-ratio plots and the use of truncated, quantifiable subsets further limit the interpretation of pooled limits of agreement [23].
Several factors could contribute to the systematically lower QTOF concentrations, and the present data cannot apportion the difference among them. First, the two methods are calibrated against different materials: the micro-LC-QTOF-MS method is calibrated with a synthetic human MR-proADM standard spiked into charcoal-treated albumin solution [21], whereas KRYPTOR is calibrated with manufacturer-supplied calibrators, and the commutability of the two calibration systems with native plasma MR-proADM has not been established. Second, the immunoassay is a sandwich assay directed against two epitopes within MR-proADM (45–92) [11,18] and may respond to additional circulating forms, such as proteolytic fragments or modified peptides that retain both epitopes, whereas the mass spectrometric method quantifies a single product ion of one charge state of the intact peptide; the previous report proposed the absence of cross-reactivity as a possible reason for the lower LC–MS values [21]. Third, recovery and matrix effects are unlikely to account for most of the difference for the spiked standard because the internal-standard-corrected recovery (89.6–114.2%) and matrix effect (95.3–111.7%) were close to 100% in the validation [21]; however, incomplete release of endogenous MR-proADM from plasma proteins during precipitation and extraction, which a spiked standard does not necessarily reproduce, cannot be excluded. Fourth, KRYPTOR results were converted from nmol/L to ng/mL using a molecular mass of 5114.76 g/mol (Section 4.3); a different assumed molecular mass would rescale all differences uniformly but would not change the concentration-related trends, the correlations, or the discrimination results, which depend on ordering. Resolving these contributions would require analysis of common reference materials by both methods and comparison of calibrator traceability, which were beyond the scope of this reprocessing study.
Low-concentration reporting was especially relevant at 10 ppm. Zero-coded results indicated values below the LOD, and positive outputs below 0.05 ng/mL were below the common QTOF LLOQ. Retaining their exact numerical ordering would imply more quantitative information than the reporting limits support. Our primary ROC analysis instead assigned all below-LLOQ results the same lowest rank, while concentration-based analyses excluded them. These approaches address different estimands: the former evaluates discrimination by the available reporting categories, whereas the latter describes quantifiable results in a selected population. The resulting changes in participant selection and ties must be considered when comparing windows. In addition, because each setting was processed with its own calibration curve, whose intercept differed between settings, and with automatic peak integration, the concentration estimate for an individual sample need not change monotonically with window width near the LLOQ, where the response is comparable to the intercept. The single participant whose 20-ppm result fell below the LLOQ while the 10-ppm and wider-window results were quantifiable (Section 2.1) illustrates this behavior. The window-specific calibration data also qualify the LLOQ: in the study batch, one of the two 0.05 ng/mL calibrators fell far outside the acceptance range at 20–100 ppm, including the validated 35-ppm setting, the analyte-free blank showed a response of up to 58% of the LLOQ response at the wider settings, and both 0.2 ng/mL calibrators met the ±20% criterion only at 35 ppm (Table 3). The LLOQ of 0.05 ng/mL established in the validation [21] was therefore not confirmed within this batch at any setting, and results between 0.05 and 0.2 ng/mL should be regarded as semi-quantitative. This does not alter the primary ROC analysis, in which all results below 0.05 ng/mL were assigned the same lowest rank and quantifiable results retained their ordering.
Age- and sex-adjusted models provided complementary evidence. Positive associations at 35–200 ppm remained in the common quantifiable subset after adjustment, suggesting that age and sex alone did not explain those associations. Nevertheless, these models do not remove potential confounding from all clinical factors, nor do they establish performance in participants excluded because of nonquantifiable results. The 10-ppm adjusted association did not differ clearly from the null in that subset and should not be interpreted as a biologically inverse relationship. The unadjusted single-marker AUCs and the adjusted odds ratios also answer different statistical questions; for example, the adjusted association at 200 ppm was comparable to those at 35–100 ppm, although its AUC was lower.
The findings support evaluating data-processing choices as part of biomarker method development, alongside established considerations in biomarker validation [5,24]. AUC depends on the ordering of measurements between groups [8,9]; a uniform positive rescaling of all concentrations would not change it. The AUC differences therefore involve changes in ordering, ties, or reporting-limit handling rather than only a shift in the concentration scale. Processing parameters should be chosen using analytical validation, including accuracy and signal retention at the low end of the calibration range rather than calibration linearity alone, and subsequently evaluated in independent clinical data, not selected solely to maximize AUC in the development cohort. The present contribution is the link between a processing setting and downstream biomarker assessment, rather than the discovery of a new physical resolution–sensitivity trade-off.
Several limitations should be considered. First, the study involved one biomarker, one QTOF platform, one processing workflow, and six discrete settings. It cannot establish a universally appropriate ppm range. The data were acquired in centroid mode, as in the validated method [21], and profile (continuum) data were not acquired; the window dependence of profile-mode extraction, in which the window truncates the continuous peak shape in every scan rather than excluding whole scans, could not be examined retrospectively, and no sample volume remained for re-acquisition. Confirmation on other instruments and with profile-mode acquisition is therefore needed before the magnitude of the effect is generalized, although the mechanism—exclusion of centroids whose scan-to-scan m/z dispersion exceeds the half-width of the window—is a property of centroid-mode processing rather than of a particular instrument. Second, the clinical comparison was based on an eGFR threshold in one adult cohort, predominantly representing modest reductions in eGFR rather than advanced renal dysfunction; chronicity and other criteria required for a clinical CKD diagnosis were not evaluated in this analysis. Third, each participant sample was measured once in a single batch. Between-participant concentration dispersion therefore does not establish analytical precision, and extraction conditions are not independent technical replicates. Duplicate quality-control (QC) measurements at three concentration levels and the reprocessed calibrators (Section 2.6) characterize window-specific accuracy and signal retention in one batch, but duplicate measurements do not establish precision, near-LLOQ performance in this batch was poor at every setting, and the chromatograms of one sample and the spectrum of one scan illustrate rather than quantify the mechanism. No sample volume remained for repeated measurements of selected samples in a separate batch, so the consistency of the window dependence across batches could not be assessed, and common reference materials could not be analyzed by both methods because the immunoassay was no longer available; together with the proprietary nature of the immunoassay calibrators, this prevents apportioning the systematic difference between the methods among calibration, recovery, and the measured molecular forms. The stored centroid data also do not support retrospective measurement of the original mass-peak full width at half maximum (FWHM). Fourth, confounding beyond age and sex, selection of quantifiable subsets, and the limited number of reduced-eGFR participants constrain inference. Finally, the same cohort was used to compare processing settings and assess discrimination; independent analytical and clinical validation remains necessary.
4. Materials and Methods
4.1. Participants and Study Design
In this study, we utilized the database of the Japan Multi-Institutional Collaborative Cohort (J-MICC) Study, containing cross-sectional data from the general adult population in Japan, and selected data obtained in the Kyoto area (J-MICC Kyoto Study) [25,26,27]. Participants were community-dwelling individuals receiving routine health check-ups. All participants provided written informed consent for the original cohort study, whose protocol was approved by the ethics committee of Kyoto Prefectural University of Medicine (RBMR-E-267-10).
The present study involved secondary use of archived plasma samples and data. We randomly selected 211 participants with eGFR ≥ 60 mL/min/1.73 m2 and 89 with eGFR < 60 mL/min/1.73 m2. Their observed age range was 39–74 years. The latter group is described as having reduced eGFR; the analysis did not establish the chronicity of the finding. Sex, age, body weight, height, aspartate aminotransferase, alanine aminotransferase, and serum creatinine were obtained from health examination records. Body mass index was calculated as weight (kg) divided by height squared (m2). Estimated GFR was calculated according to the Japanese Society of Nephrology guideline [28]. Plasma was stored at −80 °C until analysis. The 198 general-population participants whose samples were analyzed in our previous method report [21] are all included in the eGFR ≥ 60 group of the present study; the remaining 13 participants with eGFR ≥ 60 and all 89 participants with eGFR < 60 were not part of that report. The previous report established and validated the micro-LC-QTOF-MS method and applied it at a single processing setting; the present study adds the reduced-eGFR group, the reprocessing of every sample at six extraction windows, the comparison with the KRYPTOR immunoassay, and the evaluation of discrimination of reduced eGFR. The micro-LC-QTOF-MS data analyzed here were newly acquired for all 300 participants in a single analytical batch (Section 4.2) and are not the measurements reported previously.
This study was conducted in compliance with the Declaration of Helsinki and approved by the ethics committees of Meiji Pharmaceutical University (approval numbers: 202133) and Kyoto Prefectural University of Medicine (approval number: ERB-C-2266). The requirement for written informed consent was waived by the ethics committee, because the study was a secondary analysis of existing, anonymized samples and data. An opt-out procedure was implemented, and information about the study was publicly disclosed, allowing individuals the opportunity to refuse participation.
4.2. Micro-LC-QTOF-MS Measurement and Mass Extraction
Plasma MR-proADM was measured using the micro-LC-QTOF-MS method previously reported by our group [21]. Briefly, 100 µL of plasma was mixed with 10 µL of a working solution of the stable-isotope-labeled internal standard (13C25,15N5-MR-proADM; final concentration, 95 ng/mL) and 10 µL of solvent, deproteinized with 240 µL of methanol/acetonitrile (50/50, v/v), and centrifuged (13,000× g, 10 min, 10 °C); 250 µL of the supernatant was diluted with 750 µL of water, loaded onto an Oasis PRiME HLB µElution plate (Waters, Milford, MA, USA), washed with 200 µL of 20% methanol, eluted with 35 µL of methanol containing 2% formic acid, and diluted with 35 µL of 2% aqueous formic acid. Calibration samples (eight levels, 0.05–100 ng/mL) were prepared by spiking human MR-proADM standard (Phoenix Pharmaceuticals, Burlingame, CA, USA) into charcoal-treated 2% human serum albumin solution, and QC samples (0.15, 30, and 75 ng/mL) by spiking unadulterated plasma; both were processed in the same way as the participant samples. Ten microliters of extract were injected onto a trapping column (nanoEase M/Z HSS T3 Trap, 100 Å, 5 µm, 300 µm × 50 mm; Waters) and separated on an iKey HSS T3 separation device (100 Å, 1.8 µm, 150 µm × 100 mm; Waters) with an ACQUITY M-Class micro-LC system (Waters), using water (A) and acetonitrile (B), each containing 0.1% formic acid, at 2.5 µL/min: trapping at 1% B (20 µL/min, 1.5 min), a linear gradient from 1% to 85% B over 6.0 min, 85% B until 8.0 min, return to 1% B at 8.5 min, and re-equilibration until 13.5 min. Detection used a Xevo G2-XS QTOF mass spectrometer with an IonKey/MS interface (Waters) in positive electrospray ionization mode (spray voltage, 3.2 kV; cone voltage, 20 V; source temperature, 150 °C; desolvation temperature, 500 °C), operated in targeted TOF MS/MS mode with argon as the collision gas (collision energy, 22 V; scan time, 0.15 s). The precursor ions were m/z 731.5304 (7+) for MR-proADM and m/z 735.8264 (7+) for the internal standard, and the extracted ion chromatograms of the corresponding y40 (5+) product ions, m/z 833.0412 and m/z 838.8527, were used for quantification. Lock-mass correction was applied during acquisition using leucine enkephalin (m/z 556.2771) acquired every 30 s. Concentrations were calculated from the peak-area ratio of MR-proADM to the internal standard by weighted (1/x) linear regression in TargetLynx; calibrator replicates whose back-calculated concentration fell outside the acceptance tolerance (±15%; ±20% at the LLOQ) were excluded from the fit by the software, and the curve was refitted. In the previous report, the extracted ion chromatograms were generated with a mass window of 35 ppm [21]; the 35-ppm setting of the present study therefore corresponds to the originally validated processing conditions, and the validation results quoted below refer to that setting. The reported LLOQ was 0.05 ng/mL; the prior validation reported within-batch accuracy of 92.5–106.5% and a coefficient of variation below 5.8%. These validation results refer to the previously evaluated method and do not establish the corresponding performance separately at all six extraction windows.
Each participant sample was measured once (n = 1) by micro-LC-QTOF-MS, and all participant measurements were acquired within a single analytical batch. Low-, medium-, and high-concentration QC samples (0.15, 30, and 75 ng/mL) were each measured in duplicate (n = 2 per concentration level). The raw data were stored in centroid format. The six extraction settings were applied to the same acquired data for each sample; they were not separate instrument acquisitions or independent replicates. Randomization of acquisition order across extraction settings was therefore not applicable.
Raw data were reprocessed in TargetLynx XS within Waters MassLynx using the “Chromatogram mass window (PPM)” settings of 10, 20, 35, 50, 100, and 200 ppm. In this software, the entered value defines the total width of the mass window centered on the target m/z; a setting of 50 ppm therefore extracts ions within ±25 ppm of the target m/z, and the six settings correspond to ±5, ±10, ±17.5, ±25, ±50, and ±100 ppm (±0.0042 to ±0.083 Da at m/z 833.0412), with the 35-ppm setting corresponding to the validated method [21]. The target m/z, the chromatographic integration parameters, and all other processing settings were held constant, and only the mass window of the analyte was changed between reprocessing runs; the internal standard was extracted with the 35-ppm window at every setting so that the internal-standard area of each injection was identical across settings and the window dependence of the concentration estimates reflects the analyte extraction alone. Instrumental resolving power was not varied. Calibration standards and QC samples from the same batch were reprocessed together with the participant samples at each setting, and a separate calibration curve was constructed for each setting using the regression model of the previous report [21], so that each concentration estimate was derived from its own window-specific calibration. Chromatographic signals were integrated to obtain the corresponding concentration estimates. The QTOF LLOQ of 0.05 ng/mL was applied to all six outputs. No additional concentration rescaling was introduced in the statistical analysis.
For the window-specific evaluation, the calibration standards, the analyte-free calibration blanks, and the QC samples of the batch were reprocessed at each setting, and the calibration parameters, back-calculated calibrator concentrations, and QC results were taken from the TargetLynx output. Analyte and internal-standard peak areas at each setting were expressed relative to the 200-ppm setting for the same injection. QC accuracy was calculated relative to the sum of the endogenous concentration of the QC plasma (mean of the two unspiked QC injections at the same setting) and the spiked concentration. Extracted ion chromatograms of one participant sample at the six settings and the product-ion spectrum of a single scan at the peak apex were exported from MassLynx; isotopologue centroid m/z values were compared with theoretical values calculated from the target m/z, which corresponds to the most abundant (M + 2) isotopologue of the 5+ product ion, using an isotopologue spacing of 1.00335/5 Da. For the participant samples, analyte peak areas at each setting were expressed relative to the 35-ppm and 200-ppm settings of the same injection, and the association of the 10-ppm retention with the internal-standard area, the absolute analyte area, and the concentration estimate was described by Spearman rank correlation.
4.3. Comparator Immunoassay
Plasma MR-proADM concentrations were independently measured using the commercially available KRYPTOR time-resolved fluorescence immunoassay (Thermo Fisher Scientific Inc., Waltham, MA, USA) according to the manufacturer’s instructions [11,18]. The assay was used as a comparator, without assuming error-free true concentrations. The assay reports concentrations in nmol/L; these were converted to ng/mL using a molecular mass of 5114.76 g/mol for MR-proADM (45–92) so that the reported LLOQ of 0.23 nmol/L corresponds to 1.18 ng/mL. The intra-day and inter-day coefficients of variation are reported to be below 20%. All QTOF and KRYPTOR concentrations used here were expressed in ng/mL, and all KRYPTOR values exceeded 1.18 ng/mL.
4.4. Statistical Analysis
Participant characteristics are expressed as numbers (%) or medians (interquartile ranges). Sex was compared using a two-sided Fisher exact test, and age using a two-sided Mann–Whitney U test. The original analyses of the other characteristics used Welch’s t-test for parametric data and the Mann–Whitney U test for nonparametric data, with distributions assessed using the Shapiro–Wilk test.
Zero-coded QTOF results were classified as below LOD, not as measured zero concentrations. Positive outputs below 0.05 ng/mL were classified as below LLOQ. No concentration was substituted for either category. Concentration summaries, violin plots, correlations, and method-difference analyses used values at or above the relevant LLOQs. Violin plots and Pearson correlations used natural-log-transformed concentrations. Cohen’s d was calculated on that scale as the reduced-eGFR minus comparison-group mean difference divided by the pooled standard deviation. Method-specific sample sizes are reported because quantification eligibility differed between windows. The window-specific calibration, QC, chromatographic, and single-scan data (Section 2.6) were summarized descriptively, without statistical comparison between settings.
The primary ROC analysis included all 300 participants. Results below the LLOQ formed one lowest tied reporting category within each method; quantifiable results retained their concentration ordering. Reduced eGFR was the positive outcome, and higher scores were positive for every method. This direction was fixed before the analysis for all methods, on the basis of the expected increase in MR-proADM with declining renal function, and was not chosen or reversed according to the resulting AUC; AUCs below 0.5 were reported as obtained. Individual AUC confidence intervals used DeLong’s normal approximation. Paired, two-sided DeLong tests compared each QTOF AUC with KRYPTOR [29], and Holm correction was applied across the six comparisons. No hypothesis test compared one QTOF setting with another; differences among the six QTOF settings were described using point estimates and CIs only, and the study was not designed or powered to rank the settings. Differences were defined as ΔAUC = AUC(KRYPTOR) − AUC(QTOF). The 95% CIs for ΔAUC used 3000 percentile bootstrap resamples stratified by renal-function group, with all paired measurements for each selected participant resampled together (NumPy PCG64 seed 20261008). These CIs were not multiplicity-adjusted.
Method comparisons used paired differences (QTOF − KRYPTOR, ng/mL), mean individual percentage differences [100 × (QTOF − KRYPTOR)/KRYPTOR], and log ratios [23]. The geometric mean ratio was exp[mean ln(QTOF/KRYPTOR)]. Approximate 95% ratio limits of agreement were obtained by exponentiating the mean log ratio ±1.96 sample standard deviations. Separate 95% percentile bootstrap CIs for the summary estimates used 3000 group-stratified paired resamples. Limits of agreement were interpreted descriptively in view of concentration-related trends, distributional departures, and selection of quantifiable pairs.
Separate logistic regression models evaluated each method in the common quantifiable subset (n = 273). Predictors were log2[concentration/(1 ng/mL)], continuous age centered at 60 and scaled per 10 years, and male sex with female sex as the reference. Odds ratios per concentration doubling and nominal 95% Wald CIs were reported. The incremental association of concentration was tested against the age-and-sex model using a one-degree-of-freedom likelihood-ratio test, with Holm correction across seven methods. These models estimated adjusted associations rather than age-adjusted single-marker ROC curves.
Sensitivity analyses assessed ROC results in the common quantifiable subset and in method-specific quantifiable pairs. Exploratory analyses retained the original numerical ordering, including below-LLOQ outputs, or restricted the analysis to the common-positive subset; these were used only to assess sensitivity to the treatment of low results, not to validate their quantitative estimates. A further sensitivity analysis assigned below-LOD codes a lower rank than the other below-LLOQ outputs, an explicit additional ordering assumption. For adjusted associations, sensitivity analyses added a quadratic age term, used method-specific quantifiable subsets, or, in an exploratory model, retained positive below-LLOQ outputs in the 297 participants without zero-coded results. The additional sensitivity and covariate-adjustment analyses were exploratory and were not prospectively specified; the same data were used to compare settings and assess performance.
Descriptive analyses reported in the original dataset were performed in R (version 4.5.2). The concentration, ROC, and regression reanalyses were implemented in Python; numerical reproduction used Python 3.13.5, NumPy 2.3.5, SciPy 1.17.0, and scikit-learn 1.8.0. Two-sided p values below 0.05 were used to describe statistical significance, with the multiplicity procedures specified above. No equivalence or non-inferiority hypothesis was tested. Because the Python analysis code was developed with the assistance of a generative AI tool (see below), the analyses were independently recomputed in R (version 4.3.3; R Foundation for Statistical Computing, Vienna, Austria), using the pROC package (version 1.18.5) [30] for the ROC analyses and the stats package (glm) for the logistic regression models. The AUCs and their DeLong CIs, the paired DeLong tests and Holm-adjusted p values, the Cohen’s d values, the method-comparison statistics, and the age- and sex-adjusted odds ratios agreed with the Python implementation to the reported precision (differences ≤ 0.001 in AUC; bootstrap CIs within Monte Carlo error).
Generative AI was used in the preparation of this work. ChatGPT (GPT-6 Astra Pro; OpenAI, San Francisco, CA, USA) and Claude (Claude Fable 5.1; Anthropic, San Francisco, CA, USA) were used to assist with drafting and revising of the manuscript text; developing the statistical analysis code, including the R cross-validation script; and preparing figures and tables, including those in the Supplementary Materials, from the measured data. All numerical results, figures, and tables were computed from participant-level measurements; no data or experimental observations were generated, simulated, or modified by these tools. The authors reviewed and verified all AI-assisted text, code, and outputs against the underlying data.
5. Conclusions
Within a single micro-LC-QTOF-MS workflow, a single analytical batch, and an adult cohort, the mass extraction window affected MR-proADM concentration estimates and discrimination of reduced eGFR. Settings of 35, 50, and 100 ppm yielded numerically higher AUCs than the other QTOF settings examined, but substantial concentration differences from KRYPTOR remained. The results concern post-acquisition processing of one biomarker on one platform and do not establish a universal optimum, superior analytical precision, or quantitative interchangeability. Independent analytical and clinical validation is needed before extending these findings to other assays, populations, or clinical endpoints.
Supplementary Materials
The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/molecules31234332/s1, Table S1: Reporting limits and ROC sensitivity analyses; Table S2: Concentration comparisons between QTOF and KRYPTOR; Figure S1: Log-ratio method-comparison plots among quantifiable pairs; Table S3: Age- and sex-adjusted associations with reduced eGFR; Figure S2: Age- and sex-adjusted odds ratios per doubling of MR-proADM concentration in the common quantifiable subset; Table S4: Window-specific back-calculated calibrator accuracy, signal retention, and QC precision; Figure S3: Extracted ion chromatograms and window-specific results for the participant with a below-LLOQ result at 20 ppm only; Figure S4: Window-specific calibration curves of the study batch.
Author Contributions
Conceptualization, Y.S.; methodology, T.K., M.Y., J.N. and D.T.; software, Y.S.; validation, T.K., M.Y., J.N. and D.T.; formal analysis, Y.S., T.K. and J.N.; investigation, Y.S., T.K., M.Y., J.N., D.T. and A.O.; resources, Y.S.; data curation, Y.S.; writing—original draft preparation, Y.S.; writing—review and editing, Y.S., T.K., A.O. and K.O.; visualization, Y.S. and T.K.; supervision, K.O.; project administration, Y.S. All authors have read and agreed to the published version of the manuscript.
Funding
This study was supported by Grants-in-Aid for Scientific Research in Priority Areas of Cancer (No. 17015018) and Innovative Areas (No. 221S0001) and by the Japan Society for the Promotion of Science (JSPS) KAKENHI Grant (Nos. 16H06277 and 22H04923 [CoBiA]) from the Japanese Ministry of Education, Culture, Sports, Science and Technology. This work was also supported in part by funding for the BioBank Japan Project from the Ministry of Education, Culture, Sports, Science and Technology from April 2003 to March 2015 and the Japan Agency for Medical Research and Development since April 2015.
Institutional Review Board Statement
The original J-MICC Kyoto Study was approved by Kyoto Prefectural University of Medicine (RBMR-E-267-10). This secondary study was approved by Meiji Pharmaceutical University (202133) and Kyoto Prefectural University of Medicine (ERB-C-2266) and was conducted in accordance with the Declaration of Helsinki.
Informed Consent Statement
Written informed consent was obtained for the original cohort study. For this secondary analysis of existing anonymized samples and data, the requirement for additional written informed consent was waived by the ethics committee, and an opt-out procedure was implemented.
Data Availability Statement
Summary results are provided in the article and Supplementary Materials. Participant-level data from the J-MICC Kyoto Study cannot be made publicly available because of the conditions of the ethical approval and informed consent of the cohort. The de-identified per-participant concentration estimates at the six settings, the TargetLynx processing parameters for each extraction window, and the analysis code (Python scripts and the R cross-validation script) are available from the corresponding author upon reasonable request, subject to approval by the J-MICC Study.
Acknowledgments
During the preparation of this manuscript, the authors used ChatGPT (GPT-6 Astra Pro; OpenAI) and Claude (Claude Fable 5.1; Anthropic) for the purposes of drafting and revising the text, developing statistical analysis code, and preparing figures and tables from the measured data. The authors have reviewed and edited the output and take full responsibility for the content of this publication.
Conflicts of Interest
The authors declare no conflict of interest.
References
- Trusheim, M.R.; Berndt, E.R.; Douglas, F.L. Stratified medicine: Strategic and economic implications of combining drugs and clinical biomarkers. Nat. Rev. Drug Discov. 2007, 6, 287–293. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Van Gool, A.; Corrales, F.; Čolović, M.; Krstić, D.; Oliver-Martos, B.; Martínez-Cáceres, E.; Jakasa, I.; Gajski, G.; Brun, V.; Kyriacou, K.; et al. Analytical techniques for multiplex analysis of protein biomarkers. Expert Rev. Proteom. 2020, 17, 257–273. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Phua, S.X.; Lim, K.P.; Goh, W.W. Perspectives for better batch effect correction in mass-spectrometry-based proteomics. Comput. Struct. Biotechnol. J. 2022, 20, 4369–4375. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Yu, Y.; Zhang, N.; Mai, Y.; Ren, L.; Chen, Q.; Cao, Z.; Chen, Q.; Liu, Y.; Hou, W.; Yang, J.; et al. Correcting batch effects in large-scale multiomics studies using a reference-material-based ratio method. Genome Biol. 2023, 24, 201. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Nakayasu, E.S.; Gritsenko, M.; Piehowski, P.D.; Gao, Y.; Orton, D.J.; Schepmoes, A.A.; Fillmore, T.L.; Frohnert, B.I.; Rewers, M.; Krischer, J.P.; et al. Tutorial: Best practices and considerations for mass-spectrometry-based protein biomarker discovery and validation. Nat. Protoc. 2021, 16, 3737–3760. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ioannidis, J.P.; Panagiotou, O.A. Comparison of effect sizes associated with biomarkers reported in highly cited individual articles and in subsequent meta-analyses. JAMA 2011, 305, 2200–2210. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Yu, Y.; Mai, Y.; Zheng, Y.; Shi, L. Assessing and mitigating batch effects in large-scale omics studies. Genome Biol. 2024, 25, 254. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Zweig, M.H.; Campbell, G. Receiver-operating characteristic (ROC) plots: A fundamental evaluation tool in clinical medicine. Clin. Chem. 1993, 39, 561–577. [Google Scholar] [CrossRef] [Scilit]
- Obuchowski, N.A. Receiver operating characteristic curves and their use in radiology. Radiology 2003, 229, 3–8. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Sandberg, A.; Branca, R.M.; Lehtiö, J.; Forshed, J. Quantitative accuracy in mass spectrometry based proteomics of complex samples: The impact of labeling and precursor interference. J. Proteom. 2014, 96, 133–144. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Morgenthaler, N.G.; Struck, J.; Alonso, C.; Bergmann, A. Measurement of midregional proadrenomedullin in plasma with an immunoluminometric assay. Clin. Chem. 2005, 51, 1823–1829. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Koyama, T.; Kuriyama, N.; Suzuki, Y.; Saito, S.; Tanaka, R.; Iwao, M.; Tanaka, M.; Maki, T.; Itoh, H.; Ihara, M.; et al. Mid-regional pro-adrenomedullin is a novel biomarker for arterial stiffness as the criterion for vascular failure in a cross-sectional study. Sci. Rep. 2021, 11, 305. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Fialek, B.; De Roquetaillade, C.; Pruc, M.; Navolokina, A.; Chirico, F.; Ladny, J.R.; Peacock, F.W.; Szarpak, L. Systematic review with meta-analysis of mid-regional pro-adrenomedullin (MR-proadm) as a prognostic marker in Covid-19-hospitalized patients. Ann. Med. 2023, 55, 379–387. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Valenzuela-Sánchez, F.; Valenzuela-Méndez, B.; Rodríguez-Gutiérrez, J.F.; Estella-García, Á.; González-García, M. New role of biomarkers: Mid-regional pro-adrenomedullin, the biomarker of organ failure. Ann. Transl. Med. 2016, 4, 329. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Khan, S.Q.; O’Brien, R.J.; Struck, J.; Quinn, P.; Morgenthaler, N.; Squire, I.; Davies, J.; Bergmann, A.; Ng, L.L. Prognostic value of midregional pro-adrenomedullin in patients with acute myocardial infarction: The LAMP (Leicester Acute Myocardial Infarction Peptide) study. J. Am. Coll. Cardiol. 2007, 49, 1525–1532. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Christ-Crain, M.; Morgenthaler, N.G.; Stolz, D.; Müller, C.; Bingisser, R.; Harbarth, S.; Tamm, M.; Struck, J.; Bergmann, A.; Müller, B. Pro-adrenomedullin to predict severity and outcome in community-acquired pneumonia [ISRCTN04176397]. Crit. Care 2006, 10, R96. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Gouya, G.; Sturm, G.; Lamina, C.; Zitt, E.; Freistatter, O.; Struck, J.; Wolzt, M.; Knoll, F.; Lins, F.; Lhotta, K.; et al. The association of mid-regional pro-adrenomedullin and mid-regional pro-atrial natriuretic peptide with mortality in an incident dialysis cohort. PLoS ONE 2011, 6, e17803. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Caruhel, P.; Mazier, C.; Kunde, J.; Morgenthaler, N.G.; Darbouret, B. Homogeneous time-resolved fluoroimmunoassay for the measurement of midregional proadrenomedullin in plasma on the fully automated system B.R.A.H.M.S KRYPTOR. Clin. Biochem. 2009, 42, 725–728. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Vidova, V.; Spacil, Z. A review on mass spectrometry-based quantitative proteomics: Targeted and data independent acquisition. Anal. Chim. Acta 2017, 964, 7–23. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- van Bentum, M.; Selbach, M. An Introduction to Advanced Targeted Acquisition Methods. Mol. Cell. Proteom. 2021, 20, 100165. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Suzuki, Y.; Koyama, T.; Negami, J.; Toyama, D.; Oda, A.; Tanaka, R.; Iwao, M.; Tatsuta, R.; Ando, T.; Shin, T.; et al. Ultra-sensitive Quantification of Mid-regional Proadrenomedullin in Human Plasma Using Micro-liquid Chromatography Coupled to Quadrupole Time-of-Flight Mass Spectrometry. ACS Omega 2025, 10, 24214–24223. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Vereyken, L.; Dillen, L.; Vreeken, R.J.; Cuyckens, F. High-Resolution Mass Spectrometry Quantification: Impact of Differences in Data Processing of Centroid and Continuum Data. J. Am. Soc. Mass Spectrom. 2019, 30, 203–212. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Bland, J.M.; Altman, D.G. Measuring agreement in method comparison studies. Stat. Methods Med. Res. 1999, 8, 135–160. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Rifai, N.; Gillette, M.A.; Carr, S.A. Protein biomarker discovery and validation: The long and uncertain path to clinical utility. Nat. Biotechnol. 2006, 24, 971–983. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Hamajima, N. The Japan Multi-Institutional Collaborative Cohort Study (J-MICC Study) to detect gene-environment interactions for cancer. Asian Pac. J. Cancer Prev. 2007, 8, 317–323. [Google Scholar] [PubMed]
- Wakai, K.; Hamajima, N.; Okada, R.; Naito, M.; Morita, E.; Hishida, A.; Kawai, S.; Nishio, K.; Yin, G.; Asai, Y.; et al. Profile of participants and genotype distributions of 108 polymorphisms in a cross-sectional study of associations of genotypes with lifestyle and clinical factors: A project in the Japan Multi-Institutional Collaborative Cohort (J-MICC) Study. J. Epidemiol. 2011, 21, 223–235. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Takeuchi, K.; Naito, M.; Kawai, S.; Tsukamoto, M.; Kadomatsu, Y.; Kubo, Y.; Okada, R.; Nagayoshi, M.; Tamura, T.; Hishida, A.; et al. Study Profile of the Japan Multi-institutional Collaborative Cohort (J-MICC) Study. J. Epidemiol. 2021, 31, 660–668. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Japanese Society of Nephrology. Evidence-Based Clinical Practice Guideline for CKD 2018; Tokyo Igakusha: Tokyo, Japan, 2018. [Google Scholar]
- DeLong, E.R.; DeLong, D.M.; Clarke-Pearson, D.L. Comparing the areas under two or more correlated receiver operating characteristic curves: A nonparametric approach. Biometrics 1988, 44, 837–845. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Robin, X.; Turck, N.; Hainard, A.; Tiberti, N.; Lisacek, F.; Sanchez, J.-C.; Müller, M. pROC: An open-source package for R and S+ to analyze and compare ROC curves. BMC Bioinform. 2011, 12, 77. [Google Scholar] [CrossRef] [Scilit] [PubMed]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.