Reviewer 1 Report
Comments and Suggestions for AuthorsThe article presents the development of a mechatronic device capable of characterizing the kinematics of the knee joint through the recording of the Pivot Point Knee Joint (PPKJ). To carry out the design of this device, the authors scanned the leg of a participant, taking into account their anthropometric characteristics (height and weight). The developed device was tested on ten healthy volunteers without any lower-limb pathologies. Subsequently, based on the collected data, the authors formed two trajectory groups. For each group, the mean trajectory was determined, and a mathematical function was obtained through cubic polynomial regression. Additionally, the similarity between trajectories was analyzed using the cumulative Euclidean distance. Finally, the obtained kinematic data make it possible to design customized prostheses, orthoses, and exoskeletons, according to the results reported by the authors.
The specific observations are described below:
- A large part of the results section doesn’t belong there; it should be moved to other sections, or new ones should be created. Restructuring is needed.
- It is not clear where the potentiometers are located.
- The advantage of using the proportional mechanism is not clear.
- There is no comparison between the results obtained with the proposed mechanism and those obtained with preexisting systems. It is suggested to add experiments.
- Line 12: The word “storage” does not sound appropriate, as it causes confusion when reading. It is suggested to replace it with “recording.”
- Line 42: At the beginning of the paragraph, the phrase “It is essential to consider that...” could be changed to “It is important to note that...” for a more natural expression.
- Line 128: Similarly, in the phrase “to obtain a subsequent storage of these data,” it is recommended to restructure it as “to record these data for subsequent analysis,” which sounds more fluent and clearer.
- Table 1: In Item Number 9, the description “Ball joint positioner” ends with a period, which appears to be a typographical error.
- Figures 8 and 9: The figures are cropped at the bottom and not properly displayed. It is suggested to include improved images where all three axes can be clearly observed.
- Line 290: The sentence “It has reproducibility as its primary advantage” could be replaced by “Its main advantage is its reproducibility,” which sounds more natural in English.
- Line 301: The citations appear as [41], [42], [43], [44]; it is recommended to group them as [41–44].
- Line 317: In “48 and 59kg,” a space should be added between “59” and “kg,” resulting in “48 and 59 kg.”
- In general, the word “device” is repeated frequently throughout the manuscript. It could be alternated with terms such as “system” or “mechanism” to improve readability and avoid redundancy.
Author Response
Please find the details attached.
Author Response File:
Author Response.pdf
Reviewer 2 Report
Comments and Suggestions for AuthorsThis paper proposes a mechanism device based on a 60%-40%proportional mechanism for the non-invasive characterization of the instantaneous helical axis trajectory during knee flexion and extension. The topic demonstrates clear innovation and clinical application value. The research highlight lies in the design of a low-cost, non-invasive measurement system capable of acquiring individualized knee joint pivot point trajectories, providing valuable data support for the design of customized prostheses, and rehabilitation equipment. The methodology section is described in detail, the experimental design is reasonable, and reproducible trajectory data were successfully obtained from 10 healthy subjects, validating the feasibility of the device.
- It is recommended to clearly state the research purpose, methods, key results, and conclusion in the abstract. Currently, the method description is excessive, and the conclusion is not prominent enough.
- Keywords should be unified into noun forms, e.g., change "customized design "to" customized prosthesis design". Meanwhile, relative research, such as Rolling based locomotion on rough terrain for a wheeled quadruped using centroidal dynamics”, “Horizon-stability control for wheel-legged robot driving over unknow, rough terrain”, can be referred to in Introduction.
- It is recommended to use numerical numbering(e.g.,2.1,2.2)to enhance the sense of hierarchy. Currently, some subsections are not numbered.
- Figure 3-b Labeling Unclear: The links and joint numbers in the figure should be explained using arrows or a legend.
- Figure/Table Citations: All figures and tables in the text should be explicitly cited where first mentioned(e.g.," as shown in Figure 3a").
- Methods Section:The sensor calibration process and data sampling frequency should be detailed to enhance reproduce ability.
- The meaning of variables in formulas(1)and(2)should be annotated ,e.g., what angles,α,β,γ, represent.
- Compare with existing studies(e.g., the Nietert mechanism or Otto Bock devices)to highlight the advantages and limitations of the present device.
- Use either "PPKJ" or "KJPP" consistently throughout the text; avoid mixing them.
- It is recommended to supplement the discussion with specific application scenarios for expanding the use of this device in rehabilitation assessment, surgical planning, and personalized prosthesis fitting.
- Clearly state that the device is currently only applicable to specific height/weight ranges and propose future extensions to broader populations or pathological conditions.
- Some sentences and logical structures in the paper lack clarity and coherence.
- The author is requested to strictly standardize the format of all references according to MDPI's publication requirements.
- The methodology section should use the past tense, while the results and discussion sections should use the present tense.
- Trajectory Diagrams(Figures 8–9):Error bars or standard deviation shadows should be added to illustrate individual differences.
Author Response
Please find the details attached
Author Response File:
Author Response.pdf
Reviewer 3 Report
Comments and Suggestions for AuthorsMechatronic Device for Accurate Characterization of Knee Flexion Based on Pivot Point
Dear authors
The subject of the manuscript is relevant, but there are some issues in the manuscript that need to be addressed.
One of the main issues is equation 1, which is wrong.
From figure 3, it is AC^2 = BC^2 + AB^2 – 2(BC)(AB)cos(alpha)
Please check your computations accordingly.
Still in figure 3b) the horizontal dimensions of 80mm and 200mm, are the maximum values allowed by the device? Please add the unit (millimeters) to the caption of figure 3.
Please improve the “squares” in figure 3a), as they are not aligned between them.
The sentence in L10-1 is hard to follow, please rephrase.
What “kinematics” and why is the “storage” important?
The next sentence, about the leg scan, is out of context. Please rephrase.
L22, please remove “mathematical function”
L35, please briefly explain “standard way”
L62, please replace “everyone’s knee” by, for example, “patients’ knee”
L68, “consider the femoral joint as a flat movement.” How can a joint be a motion?
L71-4, the references are shown in a peculiar way, and it is difficult to follow their meaning. Are they all inside the “various authors”?
L74, “This prototype” is for all references above, or just for one?
L75 please confirm the name “Niethert” or “Nietert”?
At some points you use “defined” instead of “named”, e.g., L88, L182
At some points you use “defined” instead of “computed”, e.g., L251, L261
Please improve the paragraph in L71-81
The 60-40 ratio is very important for the development of the device, as you recognize in L306. Please reinforce its importance at the end of section 1, and, eventually, speculate about its change (say 55-45 or 65-35) on your results.
L105, KCP is undefined.
L123, please explain why you need the leg scan.
L128, replace by “and to store these data”
L142, replace by “(Eq. 1)”
First paragraph of section 3, please add a reference to figure 4 at the first mention of it.
L151, since your device uses a “Nietert mechanism”, please refer this fact early on.
L154, “These components” is too vague, please “number” them.
L153, what is the “proportion of the knee”?
The bottom part of figure 4 is missing
Please include the description found in table into the caption of figure 4, and delete table 1.
L171, L175, L197 and elsewhere, when you refer to a “letter” or “number” in any figure, include also the number of the figure.
L171-2, what is the “proportional to the medial part”?
L174, the “angle gamma” is the same as that of L172? From figure 3, it seems that the angle gamma is half of the angle between links 4 and 9.
How can you compute the angle gamma of figure 3 from the sensor at joint C of figure 3?
Are the angles in equation 2 the ones depicted in figure 3? Add this information in the text.
L184, what is the “module”?
Please include a sketch to explain equation 3.
L211, the common terminology is “Greater trochanter”.
L212, point 4 is not the bar, point 5 is the bar.
L212, “anatomical points” is not correct
L215, is “protruding anterior point of the knee region” correct?
Please improve figure 5, as the names on it are meaningless.
Furthermore, since you only fix the device by straps at mid-thigh and mid-shank, does the device (and its arms) remain stable as the subjects rotate the knee?
Looking at figure 7, the device protrudes a lot into the medial region. Is the gait still “normal” and comfortable? Please add a comment on the manuscript.
L257, is “on each knee in succession” correct?
From L258 until the end, please replace “group” by “set”, as “group” is usually referring to groups of subjects.
L260-1, replace by “trajectory and a cubic polynomial regression were computed. The similarity”
L265, replace by “the estimated intersection”
From L269 until the end, use the degree symbol, not the word.
L273-5, why do you use this equation? Please contextualize.
L276 and elsewhere, use always lowercase for “flexion” and “extension” in the middle of the sentences.
L277, what is the “error coefficient”?
Table 2, why do you need here the left and right legs?
It is not “weight”, it is “mass”.
Figures 8 and 9, please include the bottom of the plots 3D plots. Please include also 2D plots in the sagittal plane.
L296-303, please detail a little how do you may use these data/trajectories to do all these things.
L301, improve the citations [41-44]
L307, this mention to data storage is strange. The other devices, commercial or not, do not store the data?
L309-312, remove this paragraph, or improve it.
L317, you use “between”, therefore you must mention a second item, not just “development of medical devices based on standard measurements”
L323, remove “function”
References section. Confirm the references, as some of them have missing fields, e.g., journal name or volume.
Please include the DOI
Please use only the Spanish or the English title.
Author Response
Please find the details attached
Author Response File:
Author Response.pdf
Reviewer 4 Report
Comments and Suggestions for AuthorsThis manuscript presents the design and validation of a mechatronic device for non-invasive assessment of knee joint kinematics using the Pivot Point Knee Joint concept. The topic is relevant and timely, aligning well with Biomechanics’ focus on applied movement science, rehabilitation, and biomedical engineering. Overall, the paper is well organized and clearly written, with a strong rationale and promising potential for both clinical and engineering applications. The methodological design is appropriate for a proof-of-concept study, and the results demonstrate the feasibility of the proposed approach. However, several areas would benefit from clarification or refinement to enhance scientific rigor and transparency:
- The mechanical design is described in depth, but the testing and calibration procedures are less clearly outlined. It would strengthen the study to describe how repeatability, alignment accuracy, and calibration were verified.
- Please clarify how the potentiometer signals were acquired and processed, including details on sampling rate, filtering, and potential sources of measurement error.
- The authors should specify whether all measurements were performed by the same operator, as this affects inter-observer variability and reproducibility.
- The study includes only ten participants with relatively narrow anthropometric characteristics. While appropriate for a pilot validation, this limitation should be explicitly acknowledged in the discussion, emphasizing that broader testing is necessary before generalizing results.
- Several figures appear low in resolution, particularly Figures 8 and 9. Please improve image quality, especially for axis labels and titles, to enhance readability.
- The discussion could more explicitly reflect the limitations and assumptions of the device’s application.
- The statement on the potential for prosthetic and orthotic design is valid but should be expressed more cautiously, emphasizing the need for further validation studies.
- The sentence “Authors should discuss the results…” (line 309) appears to be an editing artifact and should be removed.
- Overall, the manuscript is clearly written, but minor English language editing would improve readability and technical precision. Simplifying phrasing in the methods and discussion sections would enhance flow.
Author Response
The comments and responses are attached.
Author Response File:
Author Response.pdf
Round 2
Reviewer 1 Report
Comments and Suggestions for AuthorsThere are still parts of the results section that shouldn’t be there; the results section still needs to be restructured.
The advantage of using the proportional mechanism is not clarified at all. Instead, it is simply mentioned that a potentiometer is used, which is not an advantage, especially when compared to affordable Hall effect encoders.
There is no comparison between the results obtained with the proposed mechanism and those obtained with preexisting systems.
It is suggested to add experiments.
Author Response
|
Comments 1: There are still parts of the results section that shouldn’t be there; the results section still needs to be restructured. Response 1: We thank the reviewer for this comment. The Results section has been revised and restructured to improve its clarity and organization. Redundant information has been removed and overlapping descriptions have been condensed.
|
|
Comments 2: The advantage of using the proportional mechanism is not clarified at all. Instead, it is simply mentioned that a potentiometer is used, which is not an advantage, especially when compared to affordable Hall effect encoders. There is no comparison between the results obtained with the proposed mechanism and those obtained with preexisting systems. |
|
Response 2: In the Introduction section, the following has been added: Several methods have been developed to determine the kinematics of the knee joint, ranging from early approaches to modern techniques employing advanced equipment. The most relevant include: Realeaux Method: Determines the instantaneous center of rotation, which describes a semicircular trajectory due to the progressive decrease in the curvature radius of the femoral condyles from extension to flexion. The procedure locates this center using two points defined on the medial tibial plateau (anterior and posterior sectors), identified in the sagittal plane obtained through radiography. The intersection of the perpendicular bisectors of the segments connecting these points establishes the tibiofemoral rotation center [36]. Blankevoort Method (1990): Defines the spatial trajectory of rotation axes during flexion–extension, considering the axial rotation of the tibia along its longitudinal axis. Its designation derives from the geometry described by the axes during displacement [23]. Four-Bar Linkage Model: Establishes that the instantaneous center of rotation is located at the intersection of the anterior and posterior cruciate ligaments. The model employs four bars representing the tibial plateau, femoral condyle, and both cruciate ligaments, with the intersection point defining the tibiofemoral rotation center [37]. Contact-Based Methods: Define the center of rotation as the midpoint of the shortest distance between the tibial plateau and femoral condyle surfaces, thereby establishing the rotation center at the contact point [38, 39]. Additionally, information was added to the Discussion section, as indicated below: The device developed to determine PPKJ demonstrates substantial advantages over traditional radiographic methods [36–39], as it enables a precise and non-invasive estimation of knee trajectory while avoiding exposure to ionizing radiation from X-rays, computed tomography, or fluoroscopy. Furthermore, the acquisition of knee kinematic data with the proposed device is performed while the patient is walking, in contrast to previous studies where measurements were obtained in a static position, disregarding the activation of muscles surrounding the knee joint. Conventional methods require sophisticated equipment for radiographic capture of the knee’s anatomical structure, which limits the number of samples that can be obtained and significantly increases costs. Additionally, the absence of an anatomical reference frame and the restriction to a reduced spatial environment may compromise the accuracy of tibiofemoral motion capture. |
|
Comments 3: It is suggested to add experiments. Response 3: For future work, we plan to include a larger and more diverse sample of participants, covering a wider range of anthropometric measurements. In addition, we intend to compare our approach with established methods to contrast and validate the obtained results. |
Author Response File:
Author Response.pdf
Reviewer 3 Report
Comments and Suggestions for AuthorsMechatronic Device for Accurate Characterization of Knee Flexion Based on Pivot Point
Dear authors
The manuscript improved from the previous version, but there are still too many issues in the manuscript that need to be addressed.
From figure 3 and paragraph L146-50, you have not account for the soft tissues surrounding the bones in these measurements. Will this introduce some errors in your claims and measurements? Will these be valid for all subjects, e.g. obese subjects?
L164-5, what is an “angular goniometer”? Is it manual or electronic? How do you use this goniometer? In static positions or in dynamic motions? Explain here or in L186.
L237, and figures 3, 4, 6 and 8: delta is not the angle to the horizontal. It is the “angle between the femur and tibia”, as you explain in L232-5
L266 and figure 7, the rotation point is not an anatomical landmark, therefore you have only 3 anatomical landmarks.
L267, replace “the joint rotation center” by “the patella”
L267-8, remove “the trajectory, and a cubic polynomial regression was computed” as these do not belong here.
L273-4, how do you know the exact location of the rotation center, in “the proportional mechanism (…) must be precisely located at the rotation center”?
L286, my concern was, and still is, about the medial protrusion of the device: in figure 8 b and c it is clear that the device, placed in the right knee, prevents the left knee from getting closer to the right knee. Is the gait still “normal”?
L294 and L304, why do you mention “walking” in one place and “squatting” in another?
L306 and L394ff versus L319ff, why do you need a cubic regression? And, if the cubic regression is paramount, as inferred from its’ mentioning in the conclusion, why do you use a peculiar fit in L319ff?
Figures 9 and 10, and L313, since the figures do not show 40 trajectories, where are the missing ones?
What are the y and z axis? Do they relate to equation 3? See also my comment about angle delta in L237.
Why do you have such a large spread of values for any given angle?
For full extension, i.e., 0degrees, why do you have such a large spread of y-z values?
Where is the cubic regression mentioned in L306-7?
In figure 10 there is at least one trajectory with a very small angular excursion, as it starts around 30degrees.
Figures 11 and 12, please use the same horizontal scale in all 4 plots; the same scale for the 2 Y-axis; the same scale for the 2 Z-axis.
Please add an arrow “upward and leftward” for the flexion-extension (figure 11) plots
Please add an arrow “downward and rightward” for the extension-flexion (figure 12) plots
L345-8 versus L361-4, is your device easily built for other populations, or is it only valid for the L361ff population?
L390-3, it is not clear from the main text how you can conclude this.
L394-7, where is this cubic regression shown? How can it be used?
From the review above, you still have a lot of work to do and to clearly present what you have done and why it is important and better than competing methods.
Comments on the Quality of English LanguageThe quality of English Language must be improved
A few examples:
L22, replace by “with a range of”
L29 and L31, replace by “design of customized prostheses”
L46, replace by “asymmetries”
L48, replace by “characterized by its axis of rotation in space, and, in simplified terms, it can be”
L50, replace by “movements”
L92, remove the parenthesis around PPKJ
L98, replace by “called the knee pivot point calibrator”
L128, replace by “a 3D scanner was used. By”
L134-5, replace by “with anatomical”
L137, replace by “store these data enabling its”
L140, replace by “specifically the kinematic joint C. The mechanism consists of”
L142-4, replace by “between pin groove 2 and pin groove 7. To comply with”
L149-50, replace by “Figure 3a”
L157, replace by “Where (see also Figure 3b):”
L169, replace by “the angular position”
L179, replace by “potentiometer readings (x) and the rotation angles (y) is described”
L188-9, replace by “to 50 Hz. This frequency”
L208, replace by “(1)”
L209, replace by “(2)”
L221, replace by “elements 1 and 5 of Figure 5 to determine”
L229, replace by “Knowing that the angles beta and gamma are equal, the sum of”
Equation 2, replace by “alpha + 2gamma = 180o”, where o is to be replaced by the degree symbol
L237, replace “AC distance” by “AC line”
L246-7, replace by “on a microSD card. Figure 6 shows”
L269, replace by “Bar 5 (Figure 4) must be”
L281, replace by “To ensure that there is no interference”
Throughout the manuscript, check the name “Nietert”. At least in L365 it is written “Neiter”…
Remove sentence in L170-1.
Remove sentence in L182-3.
Remove sentences in L225-8.
L237, remove “Displacement”
L251 to L258 repeats information previously shown. Remove these sentences.
Condense L195-L201, for example by saying “MSE (Mean Squared Error): 0.50595 and 0.9173, for sensors 1 and 2, respectively”
Author Response
|
Comments 1: From figure 3 and paragraph L146-50, you have not account for the soft tissues surrounding the bones in these measurements. Will this introduce some errors in your claims and measurements? Will these be valid for all subjects, e.g. obese subjects?
|
|
Response 1: We have improved the indicated paragraph.
The device was based on a 60-40 proportional mechanism (see Figure 3), coupled to both sides of the knee joint (see Figure 3-a), and aligned with anatomical reference points, specifically to locate the PPKJ and generate the points of the kinematic evolution of the knee joint during flexion and extension movements, and to store these data enabling its processing and analysis in subsequent stages. In the present study, the design and application of the device were considered only for participants without obesity, with a body weight between 48 and 59 kg and a height between 150 and 163 cm. Obesity could limit the correct positioning of the device by hindering the detection of anatomical reference points, which are located close to the bony structures, thus introducing potential errors in the measurements and in the claims derived from them. Therefore, the validity of the results presented here cannot be directly extrapolated to subjects with obesity, and this group remains the target of future specific studies.
|
|
Comments 2: L164-5, what is an “angular goniometer”? Is it manual or electronic? How do you use this goniometer? In static positions or in dynamic motions? Explain here or in L186. |
|
Response 2: Agree.
In this study, two sensors were used: 5 kΩ potentiometers for angular measurement (components 3 and 8, as detailed in Figure 4), together with a digital orthopedic angular goniometer (model GR311-L) to compare the results obtained. The goniometer is an electronic device with an accuracy of and a resolution of . It was aligned with the same anatomical reference points as the mechanical device and used to record knee flexion-extension angles in a series of static positions covering the functional range of motion, and these measurements were then compared with the potentiometer reading for calibration and validation.
Comments 3: L237, and figures 3, 4, 6 and 8: delta is not the angle to the horizontal. It is the “angle between the femur and tibia”, as you explain in L232-5 Response 3: We have, accordingly, It has been corrected.
Comments 4: L266 and figure 7, the rotation point is not an anatomical landmark, therefore you have only 3 anatomical landmarks. Response 4: You are right; we have modified it.
Comments 5: L267, replace “the joint rotation center” by “the patella” Response 5: We replaced
Comments 6: L267-8, remove “the trajectory, and a cubic polynomial regression was computed” as these do not belong here. Response 6: We removed
Comments 7: L273-4, how do you know the exact location of the rotation center, in “the proportional mechanism (…) must be precisely located at the rotation center”? Response 7: We have improved it. 2. The alignment of the device is based on three anatomical landmarks, as illustrated in Figure 7: the greater trochanter, patella, and the lateral malleolus. Installation is performed with the patient's leg in full extension. Bar 5 (Figure 4) must be positioned along an imaginary axis connecting the greater trochanter and the rotation center. Similarly, bar 1 (see Figure 4) must be aligned with the axis between the rotation center and the lateral malleolus. Additionally, the sensor of the proportional mechanism (component C in Figure 3-b and Detail 1 in Figure 4) is positioned at the rotation center by aligning the three identified anatomical landmarks. Finally, the patella positioner (see Figure 4) must be adjusted to coincide with the patella. Figures 8-a, 8-b, and 8-c depict the device in its operational configuration during the data acquisition process.
Comments 8: L286, my concern was, and still is, about the medial protrusion of the device: in figure 8 b and c it is clear that the device, placed in the right knee, prevents the left knee from getting closer to the right knee. Is the gait still “normal”? Response 8: We have improved it. Although the device presents a medial protrusion, no alterations in movement were observed during the trials, since during squatting the participants maintain a normal separation between both legs to preserve balance, so that no contact occurs between the contralateral leg and the device.
Once the device is placed, as indicated above, it is fixed with adhesive tape. To ensure that there is no interference between the clothing and the device, the device is placed directly on the skin surface of the lower extremity, and for a better fit, it is fixed with adhesive tapes between element 4 and the contour of the thigh, and in the same way between element 2 and the contour of the tibia. Elements 2 and 4 are flexible, allowing for a smooth and comfortable fit that avoids unnecessary displacement (see Figure 8). Although the device presents an anterior protrusion, no alterations in movement were detected during the trials, since during squatting the legs are kept at normal distance apart to maintain balance. Proper adjustment of the adhesive straps was sufficient to keep the device stable and aligned. To ensure reproducibility of the trials, supervision and guidance were provided by medical personnel specializing in prosthetic alignment.
Comments 9: L294 and L304, why do you mention “walking” in one place and “squatting” in another? Response 9: You are right; it has been corrected.
Comments 10: L306 and L394ff versus L319ff, why do you need a cubic regression? And, if the cubic regression is paramount, as inferred from its’ mentioning in the conclusion, why do you use a peculiar fit in L319ff? Response 10: It has been considered in the conclusion. By obtaining the trajectory from the knee joint pivot point, an alternative approach is proposed that may better match the joint’s kinematics. Based on these average trajectories, represented by the Rational Model function for the two movements studied (extension-to-flexion and flexion-to-extension), it is possible to improve the design of medical devices such as prostheses, exoskeletons, and lower-limb rehabilitation devices.
Comments 11: Figures 9 and 10, and L313, since the figures do not show 40 trajectories, where are the missing ones? Response 11: It has been corrected; it now shows 20 trajectories for extension and 20 trajectories for flexion.
Comments 12: What are the y and z axis? Do they relate to equation 3? See also my comment about angle delta in L237. Response 12: A specification has been added in the Results section, in the paragraph preceding Figure 8. 6. Once the data are obtained, they are processed to define the PPKJ during tibio-femoral displacement; these data are represented in the YZ sagittal plane as a function of the knee joint flexion angle.
Comments 13: Why do you have such a large spread of values for any given angle? Response 13: The following paragraph has been added. The spread of values for any given angle is associated with inter-subject variability, since each participant exhibits their own trajectory for each movement. In addition, there is anthropometric variability between individuals that could influence the results obtained.
Comments 14: For full extension, i.e., 0degrees, why do you have such a large spread of y-z values? Response 14: We thank the reviewer for this observation. The large dispersion observed for a given flexion angle is mainly due to inter-subject variability. The participants present different anthropometric characteristics (e.g., segment lengths, body mass, and proportions) and subject-specific knee joint geometries, which strongly influence the location of the PPKJ. In addition, small differences in marker placement and soft-tissue artefacts also contribute to the spread of the data. It is important to note that our objective was not to minimize this dispersion, but rather to capture the natural variability of tibiofemoral kinematics across individuals. Despite the dispersion, the overall trends of the trajectories can be considered consistent across participants.
Comments 15: Where is the cubic regression mentioned in L306-7? Response 15: We have added a description. Comments 16: In figure 10 there is at least one trajectory with a very small angular excursion, as it starts around 30degrees. Response 16: All trajectories lie within the same flexion range, and the viewing perspective of the plot has been improved.
Comments 17: Figures 11 and 12, please use the same horizontal scale in all 4 plots; the same scale for the 2 Y-axis; the same scale for the 2 Z-axis. Response 17: The figure was modified by adding the movement.
Comments 18: Please add an arrow “upward and leftward” for the flexion-extension (figure 11) plots Response 18: The figure was modified by adding the movement.
Comments 19: Please add an arrow “downward and rightward” for the extension-flexion (figure 12) plots Response 19: Agree. I/We have, accordingly, done/revised/changed/modified…..to emphasize this point.
Comments 20: L345-8 versus L361-4, is your device easily built for other populations, or is it only valid for the L361ff population? Response 20: The device is engineered for a target population with statures ranging from 1.50 to 1.63 meters and body masses between 48 and 59 kilograms; therefore, the proportional mechanism and its fastening system can be scaled from aproximate measurements, as long as the 60-40% ratio is maintained. That is, the device could be replicated in smaller versions, for example for pediatric populations, or in larger versions for taller participants, while always preserving the 60-40% proportional mechanism ratio defined inn this study. The knee joint’s 60–40% dimensional ratios, noted by Nietert and verified in the Otto Bock mechanism (series 743A8), are scalable and can therefore be used to develop smaller or larger designs according to the user’s anthropometric measurements. If the mechanism’s dimensions are altered such that this ratio is not respected, data acquisition will be incorrect, as the pivot point will be offset from its reference, invalidating the obtained data and hindering proper alignment of the mechanism with the anatomical reference points shown in Figure 8. It is important to store data for later analysis and comparison to facilitate the early detection of anomalies or as an assessment tool in a specific rehabilitation program. emphasize this point.
Comments 21: L390-3, it is not clear from the main text how you can conclude this. Response 21: Agree. I/We have, accordingly, done/revised/changed/modified…..to emphasize this point.
Comments 22: L394-7, where is this cubic regression shown? How can it be used? Response 22: The last paragraph of the conclusion has been improved.
|
Author Response File: