Evaluation of a New Tandem Mass Spectrometry Method for Sickle Cell Disease Newborn Screening

,
,
,
,
,
,
,
,
…
1
Laboratoire de Biologie Médicale Multi-Sites (LBMMS), Service de Biochimie et Biologie Moléculaire Grand Est, Hospices Civils de Lyon, 69500 Bron, France
2
Centre Régional de Dépistage Néonatal Auvergne Rhône-Alpes (CRDN AuRA), 69003 Lyon, France
3
Centre de Référence Constitutif des Maladies Constitutionnelles du Globule Rouge et de l’Erythropoïèse, 69003 Lyon, France
4
Inter-University Laboratory of Human Movement Sciences, LIBM, Universite Claude Bernard Lyon 1, UR 7424, 69622 Villeurbanne, France

Abstract

In France, sickle cell disease newborn screening (SCD NBS) has been targeted to at-risk regions since 1984, but generalization to the whole population will be implemented from November 2024. Although tandem mass spectrometry (MS/MS) is already used for the NBS of several inherited metabolic diseases, its application for SCD NBS has not been widely adopted worldwide. The aim of this study was to evaluate a dedicated MS/MS kit (Targeted MS/MS Hemo, ZenTech, LaCAR Company, Liege, Belgium) for SCD NBS and to compare the results obtained with those from an NBS reference center using matrix-assisted laser desorption/ionization time of flight (MALDI-TOF) and cation-exchange high-performance liquid chromatography (CE-HPLC, Variant NBS, Biorad Laboratories, Inc., Hercules, CA, USA) as confirmatory method. The MS/MS Hemo kit was used according to the manufacturer’s instructions and performed on a Waters Xevo TQ-D (Waters Corporation, USA). The software provided by the manufacturer was used for the calculation and analysis of peptide signal ratios. Among the 1333 samples, the results of 1324 samples were consistent with the HPLC and/or MALDI-TOF results (1263 FA, 50 FAS, 7 FAC, 1 FAO-Arab, and 3 FS). All the discordant results (one FAS on MS/MS vs. FA in CE-HPLC, one FA on MS/MS vs. FAS in CE-HPLC, seven FS on MS/MS vs. FAS in CE-HPLC) were corrected after modifying the peptide signal ratios thresholds, allowing the MS/MS Hemo kit to achieve near-100% sensitivity and specificity for SCD NBS. In conclusion, the MS/MS Hemo kit appears to be an effective method for SCD NBS, particularly for laboratories already equipped with MS/MS technology. However, these results should be confirmed in a larger cohort including a greater number of positive samples for SCD.

Article Metrics

Citations

Article Access Statistics

Multiple requests from the same IP address are counted as one view.