Article

Regional Ultrahigh-Resolution Rescan in a Clinical Whole-Body CT Scanner Using a Contact Detector Insert

by
Thomas C. Larsen
1,
Eric E. Bennett
1,
Dumitru Mazilu
1,
Marcus Y. Chen
2 and
Han Wen
1,*
1
Imaging Physics Laboratory, Biochemistry and Biophysics Center, Lung and Blood Institute, National Institutes of Health, Bethesda, MA, USA
2
Cardiovascular Branch, National Heart, Lung and Blood Institute, National Institutes of Health, Bethesda, MA, USA
*
Author to whom correspondence should be addressed.
Tomography 2019, 5(2), 233-238; https://doi.org/10.18383/j.tom.2019.00002
Submission received: 1 March 2019 / Revised: 2 April 2019 / Accepted: 11 May 2019 / Published: 1 June 2019

Abstract

Ultrahigh-resolution, low-dose rescans in a region of interest following a general screening computed tomography (CT) scan is motivated by the need to reduce invasive tissue biopsy procedures in cancer screening. We describe a new method to meet the conflicting demands of ultrahigh resolution, high-speed and ultralow-dose, and the first proof-of-concept experiment. With improving detector resolution, the limiting factor for the system resolution of whole-body CT scanners shifts to the penumbra of the source focal spot. The penumbra unsharpness is minimized by inserting flat-panel detector(s) that are in direct contact with the body. In the hybrid system, the detector insert and the CT detector acquire data simultaneously, whereby the standard CT images give the position and orientation of the detector insert(s) as needed for tomosynthesis reconstruction. Imaging tests were performed with a compact photon-counting detector insert on resolution targets of both high- and low-contrast as well as a mouse specimen, all inside a body phantom. Detector insert tomosynthesis provided twice the resolution of the CT scanner alone at the same dose concentration. The short 2-cm beam collimation of the tomosynthesis rescan gave an effective dose equivalent to 6% of an average CT scan in the chest or abdomen.
Keywords: CT re-scan; ultrahigh resolution; ultralow dose; contact detector insert; tomosynthesis; hybrid CT CT re-scan; ultrahigh resolution; ultralow dose; contact detector insert; tomosynthesis; hybrid CT

Share and Cite

MDPI and ACS Style

Larsen, T.C.; Bennett, E.E.; Mazilu, D.; Chen, M.Y.; Wen, H. Regional Ultrahigh-Resolution Rescan in a Clinical Whole-Body CT Scanner Using a Contact Detector Insert. Tomography 2019, 5, 233-238. https://doi.org/10.18383/j.tom.2019.00002

AMA Style

Larsen TC, Bennett EE, Mazilu D, Chen MY, Wen H. Regional Ultrahigh-Resolution Rescan in a Clinical Whole-Body CT Scanner Using a Contact Detector Insert. Tomography. 2019; 5(2):233-238. https://doi.org/10.18383/j.tom.2019.00002

Chicago/Turabian Style

Larsen, Thomas C., Eric E. Bennett, Dumitru Mazilu, Marcus Y. Chen, and Han Wen. 2019. "Regional Ultrahigh-Resolution Rescan in a Clinical Whole-Body CT Scanner Using a Contact Detector Insert" Tomography 5, no. 2: 233-238. https://doi.org/10.18383/j.tom.2019.00002

APA Style

Larsen, T. C., Bennett, E. E., Mazilu, D., Chen, M. Y., & Wen, H. (2019). Regional Ultrahigh-Resolution Rescan in a Clinical Whole-Body CT Scanner Using a Contact Detector Insert. Tomography, 5(2), 233-238. https://doi.org/10.18383/j.tom.2019.00002