Validation of Compressed Sensing Accelerated Magnetization Prepared Rapid Acquisition Gradient Echo for Fast and Reliable Brain Volume Measurement
Simple Summary
Abstract
1. Introduction
2. Materials and Methods
2.1. Study Population
2.2. Image Acquisition and Reconstruction
2.3. Quantitative Analysis
2.4. Qualitative Analysis
2.5. Statistical Analysis
3. Results
3.1. Study Cohort
3.2. Acquisition Time
3.3. Quantitative Analysis
3.4. Qualitative Analysis
4. Discussion
5. Conclusions
Supplementary Materials
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Conflicts of Interest
Abbreviations
| CS | Compressed Sensing |
| MPRAGE | Magnetization Prepared Rapid Acquisition Gradient Echo |
| MRI | Magnetic resonance imaging |
References
- Goodkin, O.; Pemberton, H.; Vos, S.B.; Prados, F.; Sudre, C.H.; Moggridge, J.; Cardoso, M.J.; Ourselin, S.; Bisdas, S.; White, M. The quantitative neuroradiology initiative framework: Application to dementia. Br. J. Radiol. 2019, 92, 20190365. [Google Scholar] [CrossRef] [Scilit]
- Giorgio, A.; De Stefano, N. Clinical use of brain volumetry. J. Magn. Reson. Imaging 2013, 37, 1–14. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Mugler, J.P., III; Brookeman, J.R. Rapid three-dimensional T1-weighted MR imaging with the MP-RAGE sequence. J. Magn. Reson. Imaging 1991, 1, 561–567. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Dieckmeyer, M.; Roy, A.G.; Senapati, J.; Wachinger, C.; Grundl, L.; Döpfert, J.; Bertran, P.F.; Lemke, A.; Zimmer, C.; Kirschke, J.S. Effect of MRI acquisition acceleration via compressed sensing and parallel imaging on brain volumetry. Magn. Reson. Mater. Phys. Biol. Med. 2021, 34, 487–497. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Griswold, M.A.; Jakob, P.M.; Heidemann, R.M.; Nittka, M.; Jellus, V.; Wang, J.; Kiefer, B.; Haase, A. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn. Reson. Med. 2002, 47, 1202–1210. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Bilgic, B.; Gagoski, B.A.; Cauley, S.F.; Fan, A.P.; Polimeni, J.R.; Grant, P.E.; Wald, L.L.; Setsompop, K. Wave-CAIPI for highly accelerated 3D imaging. Magn. Reson. Med. 2015, 73, 2152–2162. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lustig, M.; Donoho, D.; Pauly, J.M. Sparse MRI: The application of compressed sensing for rapid MR imaging. Magn. Reson. Med. 2007, 58, 1182–1195. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Mönch, S.; Sollmann, N.; Hock, A.; Zimmer, C.; Kirschke, J.S.; Hedderich, D.M. Magnetic resonance imaging of the brain using compressed sensing–quality assessment in daily clinical routine. Clin. Neuroradiol. 2020, 30, 279–286. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Longo, M.; Conklin, J.; Cauley, S.; Setsompop, K.; Tian, Q.; Polak, D.; Polackal, M.; Splitthoff, D.; Liu, W.; González, R. Evaluation of ultrafast wave-CAIPI MPRAGE for visual grading and automated measurement of brain tissue volume. Am. J. Neuroradiol. 2020, 41, 1388–1396. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Cohen, J.F.; Korevaar, D.A.; Altman, D.G.; Bruns, D.E.; Gatsonis, C.A.; Hooft, L.; Irwig, L.; Levine, D.; Reitsma, J.B.; de Vet, H.C.W.; et al. STARD 2015 guidelines for reporting diagnostic accuracy studies: Explanation and elaboration. BMJ Open 2016, 6, e012799. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Esiri, M.M. Ageing and the brain. J. Pathol. 2007, 211, 181–187. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lee, H.; Kim, H.W.; Lee, M.; Kang, J.; Kim, D.; Lim, H.K.; Lee, J.-Y.; Kim, E.; Kim, R.E. Evaluating brain volume segmentation accuracy and reliability of FreeSurfer and Neurophet AQUA at variations in MRI magnetic field strengths. Sci. Rep. 2024, 14, 24513. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Suh, C.H.; Jung, S.C.; Lee, H.B.; Cho, S.J. High-resolution magnetic resonance imaging using compressed sensing for intracranial and extracranial arteries: Comparison with conventional parallel imaging. Korean J. Radiol. 2019, 20, 487–497. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Sartoretti, T.; van Smoorenburg, L.; Sartoretti, E.; Schwenk, Á.; Binkert, C.A.; Kulcsár, Z.; Becker, A.S.; Graf, N.; Wyss, M.; Sartoretti-Schefer, S. Ultrafast intracranial vessel imaging with non-Cartesian spiral 3-dimensional time-of-flight magnetic resonance angiography at 1.5 T: An in vitro and clinical study in healthy volunteers. Investig. Radiol. 2020, 55, 293–303. [Google Scholar]
- Sartoretti, T.; Sartoretti, E.; van Smoorenburg, L.; Schwenk, Á.; Mannil, M.; Graf, N.; Binkert, C.A.; Wyss, M.; Sartoretti-Schefer, S. Spiral 3-dimensional T1-weighted turbo field echo: Increased speed for magnetization-prepared gradient echo brain magnetic resonance imaging. Investig. Radiol. 2020, 55, 775–784. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Koo, T.K.; Li, M.Y. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J. Chiropr. Med. 2016, 15, 155–163. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Biffen, S.C.; Warton, C.M.; Dodge, N.C.; Molteno, C.D.; Jacobson, J.L.; Jacobson, S.W.; Meintjes, E.M. Validity of automated FreeSurfer segmentation compared to manual tracing in detecting prenatal alcohol exposure-related subcortical and corpus callosal alterations in 9-to 11-year-old children. NeuroImage Clin. 2020, 28, 102368. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Akudjedu, T.N.; Nabulsi, L.; Makelyte, M.; Scanlon, C.; Hehir, S.; Casey, H.; Ambati, S.; Kenney, J.; O’Donoghue, S.; McDermott, E. A comparative study of segmentation techniques for the quantification of brain subcortical volume. Brain Imaging Behav. 2018, 12, 1678–1695. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Delattre, B.M.A.; Boudabbous, S.; Hansen, C.; Neroladaki, A.; Hachulla, A.-L.; Vargas, M.I. Compressed sensing MRI of different organs: Ready for clinical daily practice? Eur. Radiol. 2020, 30, 308–319. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Jaspan, O.N.; Fleysher, R.; Lipton, M.L. Compressed sensing MRI: A review of the clinical literature. Br. J. Radiol. 2015, 88, 20150487. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Toledano-Massiah, S.; Sayadi, A.; De Boer, R.; Gelderblom, J.; Mahdjoub, R.; Gerber, S.; Zuber, M.; Zins, M.; Hodel, J. Accuracy of the compressed sensing accelerated 3D-FLAIR sequence for the detection of MS plaques at 3T. Am. J. Neuroradiol. 2018, 39, 454–458. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Clarkson, M.J.; Cardoso, M.J.; Ridgway, G.R.; Modat, M.; Leung, K.K.; Rohrer, J.D.; Fox, N.C.; Ourselin, S. A comparison of voxel and surface based cortical thickness estimation methods. NeuroImage 2011, 57, 856–865. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lee, J.Y.; Oh, S.W.; Chung, M.S.; Park, J.E.; Moon, Y.; Jeon, H.J.; Moon, W.-J. Clinically available software for automatic brain volumetry: Comparisons of volume measurements and validation of intermethod reliability. Korean J. Radiol. 2020, 22, 405. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Knussmann, G.N.; Anderson, J.S.; Prigge, M.B.; Dean, D.C., III; Lange, N.; Bigler, E.D.; Alexander, A.L.; Lainhart, J.E.; Zielinski, B.A.; King, J.B. Test-retest reliability of FreeSurfer-derived volume, area and cortical thickness from MPRAGE and MP2RAGE brain MRI images. NeuroImage Rep. 2022, 2, 100086. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Meléndez, J.C.; McCrank, E. Anxiety-related reactions associated with magnetic resonance imaging examinations. JAMA 1993, 270, 745–747. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Zaitsev, M.; Maclaren, J.; Herbst, M. Motion artifacts in MRI: A complex problem with many partial solutions. J. Magn. Reson. Imaging 2015, 42, 887–901. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Andre, J.B.; Bresnahan, B.W.; Mossa-Basha, M.; Hoff, M.N.; Smith, C.P.; Anzai, Y.; Cohen, W.A. Toward quantifying the prevalence, severity, and cost associated with patient motion during clinical MR examinations. J. Am. Coll. Radiol. 2015, 12, 689–695. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Svennerholm, L.; Boström, K.; Jungbjer, B. Changes in weight and compositions of major membrane components of human brain during the span of adult human life of Swedes. Acta Neuropathol. 1997, 94, 345–352. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ferraro, P.M.; Gualco, L.; Costagli, M.; Schiavi, S.; Ponzano, M.; Signori, A.; Massa, F.; Pardini, M.; Castellan, L.; Levrero, F. Compressed sensing (CS) MP2RAGE versus standard MPRAGE: A comparison of derived brain volume measurements. Phys. Medica 2022, 103, 166–174. [Google Scholar] [CrossRef] [Scilit] [PubMed]




| Standard MPRAGE | CS MPRAGE | |
|---|---|---|
| TR/TE (ms) | 2300/2.98 | 2300/2.9 |
| Echo Spacing (ms) | 7.14 | 7.86 |
| Inversion time (ms) | 900 | 900 |
| Flip angle (°) | 9 | 9 |
| Acceleration Method | GRAPPA | Compressed sensing |
| Acceleration factor | 2 | 5 |
| Slice thickness (mm) | 1.0 | 1.0 |
| Number of slices | 176 | 176 |
| Acquisition time | 5′20″ | 1′46″ |
| Standard MPRAGE (Mean ± SD) | CS-MPRAGE (Mean ± SD) | p | ICC | ||
|---|---|---|---|---|---|
| Intracranial Volume | 1546.63 ± 132.97 | 1551.21 ± 132.88 | 0.32 | 0.988 | |
| Whole-brain | 1176.89 ± 105.36 | 1181.14 ±105.62 | 0.27 | 0.987 | |
| Gray matter | 557.12 ± 45.59 | 556.34 ±47.84 | 0.72 | 0.989 | |
| White matter | 469.15 ± 55.74 | 473.32 ±55.70 | 0.02 | 0.990 | |
| Frontal lobe | 177.30 ± 17.49 | 177.53 ± 17.94 | 0.72 | 0.987 | |
| Precentral gyrus | 28.11 ± 2.48 | 28.37 ± 2.63 | 0.70 | 0.939 | |
| Superior frontal gyrus | 46.50 ± 5.12 | 46.82 ± 5.43 | 0.72 | 0.975 | |
| Middle frontal gyrus | 43.89 ± 5.54 | 43.97 ± 5.68 | 0.88 | 0.990 | |
| Inferior frontal gyrus | 22.19 ± 2.48 | 22.13 ± 2.42 | 0.88 | 0.987 | |
| Frontal pole | 1.98 ± 0.26 | 1.92 ± 0.28 | 0.37 | 0.929 | |
| Parietal lobe | 124.73 ± 11.90 | 124.17 ± 12.09 | 0.31 | 0.979 | |
| Superior parietal lobule | 29.06 ± 2.83 | 28.87 ± 2.85 | 0.97 | 0.984 | |
| Inferior parietal lobule | 29.96 ± 3.65 | 29.67 ± 3.66 | 0.74 | 0.979 | |
| Postcentral gyrus | 19.24 ± 1.83 | 19.29 ± 3.67 | 0.91 | 0.952 | |
| Temporal lobe | 118.55 ± 10.19 | 118.25 ± 9.78 | 0.57 | 0.966 | |
| Superior temporal gyrus | 25.78 ± 2.77 | 25.76 ± 2.64 | 0.75 | 0.985 | |
| Middle temporal gyrus | 26.89 ± 3.04 | 26.84 ± 3.09 | 0.92 | 0.928 | |
| Inferior temporal gyrus | 25.52 ± 2.32 | 25.52 ± 2.25 | 0.84 | 0.977 | |
| Transverse temporal gyrus | 2.15 ± 0.37 | 2.11 ± 0.37 | 0.67 | 0.988 | |
| Occipital lobe | 48.83 ± 5.56 | 48.14 ± 5.51 | 0.02 * | 0.972 | |
| Lateral occipital cortex | 25.79 ± 3.17 | 25.28 ± 3.19 | 0.97 | 0.979 | |
| Medial occipital cortex | 23.04 ± 2.75 | 22.86 ± 2.74 | 0.97 | 0.961 | |
| Pericalcarine cortex | 3.12 ± 0.63 | 3.14 ± 0.66 | 0.72 | 0.952 | |
| Cerebellum | 141.99 ± 10.13 | 142.88 ± 10.15 | 0.21 | 0.951 | |
| Deep gray matter | |||||
| Caudate | 7.83 ± 0.97 | 7.89 ± 0.97 | 0.07 | 0.992 | |
| Pallidum | 4.61 ± 0.60 | 4.58 ± 0.58 | 0.42 | 0.954 | |
| Putamen | 11.34 ± 1.31 | 11.37 ± 1.32 | 0.26 | 0.995 | |
| Thalamus | 16.47 ± 1.37 | 16.61 ± 1.40 | 0.06 | 0.970 | |
| Medial temporal lobe | |||||
| Hippocampus | 7.87 ± 0.51 | 7.89 ± 0.50 | 0.65 | 0.958 | |
| Amygdala | 3.90 ± 0.36 | 3.98 ± 0.37 | 0.001 * | 0.958 | |
| Enthorhinal cortex | 4.03 ± 0.39 | 3.98 ± 0.45 | 0.006 * | 0.949 | |
| Parahippocampus | 4.13 ± 0.38 | 4.09 ± 0.38 | 0.25 | 0.941 | |
| Standard MPRAGE (Mean ± SD) | CS-MPRAGE (Mean ± SD) | p | ||
|---|---|---|---|---|
| CNR | ||||
| Left | 30.45 ± 11.21 | 49.53 ± 32.71 | 0.0001 | |
| Right | 30.73 ± 11.97 | 46.89 ± 30.40 | 0.0001 | |
| SNR | ||||
| Left Centrum Semiovale | 47.18 ± 12.24 | 50.95 ± 16.72 | 0.1834 | |
| Right Centrum Semiovale | 46.83 ± 12.39 | 50.32 ± 14.77 | 0.4772 | |
| Left Putamen | 32.22 ± 8.72 | 29.65 ± 10.16 | 0.1916 | |
| Right Putamen | 31.12 ± 9.62 | 28.44 ± 8.10 | 0.0707 | |
| Left Cerebellum | 36.41 ± 12.20 | 33.67 ± 10.03 | 0.1864 | |
| Right Cerebellum | 37.81 ± 11.24 | 34.83 ± 11.69 | 0.1946 | |
| Pons | 25.55 ± 7.03 | 25.8 ± 8.94 | 0.7699 | |
| Standard MPRAGE | CS-MPRAGE | p Value | |
|---|---|---|---|
| Overall image quality | 4.94 ± 0.24 | 4.54 ± 0.55 | <0.0001 |
| GW differentiation | 5.00 ± 0.00 | 4.91 ± 0.40 | 0.05 |
| DG differentiation | 5.00 ± 0.00 | 4.71 ± 0.51 | <0.0001 |
| Artifact | 5.00% (n = 4/80) | 41.25% (n = 33/80) | <0.0001 |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.
Share and Cite
Lee, J.B.; Ahn, H.S.; Joo, L.; Yim, Y. Validation of Compressed Sensing Accelerated Magnetization Prepared Rapid Acquisition Gradient Echo for Fast and Reliable Brain Volume Measurement. Tomography 2026, 12, 121. https://doi.org/10.3390/tomography12090121
Lee JB, Ahn HS, Joo L, Yim Y. Validation of Compressed Sensing Accelerated Magnetization Prepared Rapid Acquisition Gradient Echo for Fast and Reliable Brain Volume Measurement. Tomography. 2026; 12(9):121. https://doi.org/10.3390/tomography12090121
Chicago/Turabian StyleLee, Jung Bin, Hye Shin Ahn, Leehi Joo, and Younghee Yim. 2026. "Validation of Compressed Sensing Accelerated Magnetization Prepared Rapid Acquisition Gradient Echo for Fast and Reliable Brain Volume Measurement" Tomography 12, no. 9: 121. https://doi.org/10.3390/tomography12090121
APA StyleLee, J. B., Ahn, H. S., Joo, L., & Yim, Y. (2026). Validation of Compressed Sensing Accelerated Magnetization Prepared Rapid Acquisition Gradient Echo for Fast and Reliable Brain Volume Measurement. Tomography, 12(9), 121. https://doi.org/10.3390/tomography12090121