Real-Time Endoscopic Video Enhancement via Degradation Representation Estimation and Propagation

,
,
and
1
Department of Mechanical Engineering, Tsinghua University, Beijing 100084, China
2
Beijing Key Laboratory of Transformative High-End Manufacturing Equipment and Technology, Tsinghua University, Beijing 100084, China
3
State Key Laboratory of Tribology in Advanced Equipment, Tsinghua University, Beijing 100084, China
*
Authors to whom correspondence should be addressed.
This article belongs to the Section Medical Imaging

Abstract

Endoscopic images are often degraded by uneven illumination, motion blur, and tissue occlusion, which obscure critical anatomical details and complicate surgical manipulation. This issue is particularly pronounced in single-port endoscopic surgery, where the imaging capability of the camera is further constrained by limited working space. While deep learning-based enhancement methods have demonstrated impressive performance, most existing approaches remain too computationally demanding for real-time surgical use. To address this challenge, we propose an efficient stepwise endoscopic image enhancement framework that introduces an implicit degradation representation as an intermediate feature to guide the enhancement module toward high-quality results. The framework further exploits the temporal continuity of endoscopic videos, based on the assumption that image degradation evolves smoothly over short time intervals. Accordingly, high-quality degradation representations are estimated only on key frames at fixed intervals, while the representations for the remaining frames are obtained through fast inter-frame propagation, thereby significantly improving computational efficiency while maintaining enhancement quality. Experimental results demonstrate that our method achieves an excellent balance between enhancement quality and computational efficiency. Further evaluation on the downstream segmentation task suggests that our method substantially enhances the understanding of the surgical scene, validating that implicitly learning and degradation representation propagation offer a practical pathway for real-time clinical application.

1. Introduction

Minimally invasive surgery (MIS) has become a cornerstone of modern clinical practice, offering reduced surgical trauma, shorter recovery times, and improved postoperative outcomes compared with traditional open procedures [1]. Among various MIS techniques, endoscopic surgery plays a particularly critical role, as it enables surgeons to access deep or delicate anatomical regions through narrow working channels with minimal disruption of surrounding tissues [2]. For example, in spine surgery, endoscopic approaches have been increasingly adopted for the treatment of conditions such as lumbar disc herniation, spinal stenosis, and degenerative diseases [3]. Unlike open surgical procedures, MIS relies entirely on intraoperative endoscopic video as the sole source of visual feedback, making video quality a decisive factor for surgical safety and outcome.
However, endoscopic imaging is inherently constrained by complex in vivo conditions, which give rise to various forms of image degradation. The near-axis illumination of cold light sources often results in pronounced spatial non-uniformity in lighting, while strong surface reflectance and light scattering induced by intraoperative fluids further reduce both global and local image contrast. Further, blood contamination, surgical smoke, and instruments frequently enter the field of view, causing persistent dynamic occlusions, whereas continuous endoscope motion within confined spaces and a limited depth of field introduce motion blur and defocus. These degradation factors commonly coexist with temporal and spatial variability, leading to complex and non-stationary degradation patterns in endoscopic videos. Such compounded degradations substantially compromise the visibility of fine anatomical structures, hinder accurate surgical manipulation, and may increase the risk of complications.
To address these challenges, video enhancement techniques have been widely investigated. Conventional approaches [4,5,6,7] can provide modest improvements in brightness or contrast, but they often fail under severe degradations and may amplify noise or introduce artifacts [8]. More recently, deep learning-based methods have demonstrated remarkable success in image restoration tasks, including denoising, deblurring [9], and super-resolution [10,11], by leveraging large-scale data and learning complex degradation models. These techniques have been gradually extended to the domain of endoscopic imaging, with promising results in improving surgical visibility.
Nevertheless, several critical limitations remain. Most existing methods are designed and evaluated in offline settings, where computational efficiency is not a primary concern. As a result, they often involve complex models or iterative optimization procedures that are computationally expensive and unsuitable for deployment in the operating room. In real surgical environments, enhancement must be performed at video frame rates to provide surgeons with immediate visual feedback, typically requiring real-time processing at full HD (1080p) resolution and 25–30 fps with an end-to-end latency below 100 ms. The inability of current methods to achieve real-time restoration, therefore, represents a substantial barrier to clinical adoption.
The complex and dynamically evolving degradations in endoscopic images exhibit substantial temporal and spatial variability, making it challenging to achieve high-quality enhancement using a single end-to-end deep learning model without resorting to large and computationally expensive architectures. To overcome this limitation, we propose a stepwise enhancement framework that explicitly decouples degradation estimation from image enhancement. Specifically, an implicit degradation representation is introduced to compactly characterize the overall degradation state of each frame and to guide the enhancement process. Under the assumption that image degradation varies smoothly over short temporal intervals, high-quality degradation representations are estimated only on sparsely sampled key frames and stored in a degradation memory bank. For intermediate frames, the corresponding degradation representations are efficiently obtained through fast propagation from the memory bank, enabling substantial improvements in inference efficiency while maintaining enhancement quality. Qualitative and quantitative comparisons with other video or image enhancement methods demonstrate the superior performance and efficiency of the proposed method in enhancing endoscopic videos. Further evaluation on the downstream segmentation task suggests that our method substantially enhances the understanding of the surgical scene. A preliminary version of this work has been released as a preprint on arXiv [12].

3. Methodology

3.1. Video Enhancement Framework

The proposed endoscopic video enhancement framework is illustrated in Figure 1. In this framework, each frame of the input video sequence undergoes degradation representation modeling and enhancement generation. Specifically, for frames whose index is a multiple of T Δ , the system invokes the degradation-aware module to perform a full estimation, yielding high-precision degradation features. For frames that are not multiples of T Δ , the degradation representation is rapidly estimated through the DRPM along the temporal dimension and subsequently fed into the generator to enhance the current frame.
Figure 1. Framework of the real-time video enhancement. Each frame of the input video sequence undergoes degradation representation modeling and generative enhancement. Specifically, for frames whose index is a multiple of T Δ , the system invokes the degradation-aware module (DAM) to perform a full estimation, yielding high-precision degradation features. For frames that are not multiples of T Δ , the degradation representation is rapidly propagated through the degradation representation propagation module (DRPM) along the temporal dimension and subsequently fed into the single frame enhancement model to enhance the current frame. All the modules are detailed in the following sections.
This “key-propagation” design enables the framework to maintain accurate degradation representations without performing high-complexity modeling on every frame, thereby significantly reducing the overall computational load. More importantly, the propagation mechanism ensures strong temporal consistency and continuity across the video sequence, allowing the system to produce stable enhancement results while satisfying real-time requirements.

3.2. Single Frame Enhancement Model

We introduce a single-frame image enhancement model that is guided by degradation representation. Building upon the CycleGAN framework [31], our approach leverages cyclical consistency between degraded and restored images to construct the training paradigm. Within the generator, we design a dedicated degradation-aware module (DAM) to extract the implicit degradation characteristics of the input image, which are then used to guide the enhancement of the current frame. This design ensures that the enhancement process is explicitly aware of image degradation, leading to more reliable restoration in complex surgical scenarios.

3.2.1. Degradation-Aware Module

This module learns to characterize degradation representations through a contrastive learning strategy inspired by MoCo [32,33]. As illustrated in Figure 2a, given a batch of random degraded images X, the assumption is that only representations extracted from the same image should be identical, while those from different images should remain distinct. Given an image patch as the query patch, other patches from the same image can be considered as positive samples. In contrast, patches from other images are referred to as negative samples. To enforce this property, the degradation-aware module is trained using the InfoNCE loss [34] as (1), which measures the similarity among multiple degradation representations.
L con = − log exp ( d i q · d i k / τ ) ∑ j = 1 N exp ( d i q · d j k / τ )
Figure 2. The diagram of the DAM. (a) The DAM training workflow. (b) The structure of the DAM.
To strengthen the expressive capacity of the DAM, we replace conventional stacked convolutional layers with a series of residual blocks, as shown in Figure 2b. All other settings follow the configuration described in [33].
The extracted degradation representation d is high-dimensional and often redundant. To obtain a compact yet informative embedding, we introduce both channel attention and spatial attention mechanisms as (2), which selectively emphasize the most relevant degradation cues while suppressing irrelevant information (Figure 3b).
d ′ = σ ( W c 2 δ ( W c 1 d ) ) ∗ d d c = σ ( W s 1 d ′ ) ∗ d ′
where σ , δ represent the sigmoid and ReLU activation function separately.
Figure 3. Degradation guided enhancement module (DGEM). (a) The basic architecture of the DGEM, which includes four parts: Degradation representation compression, shallow feature extraction, feature modulation, and Reconstruction. (b) The degradation compression block, which implements both channel attention and spatial attention mechanisms. (c) The Degradation-aware Swin Transformer Block, which injects the degradation representation d c into the value component and modulates the input feature based on multi-head self-attention.
This architectural refinement allows the encoder to capture richer and more discriminative degradation features. Importantly, by producing a more accurate and compact representation of degradation, the DAM reduces the burden on the subsequent image enhancement module. As a result, the overall framework not only improves the quality of degradation modeling but also lowers the complexity of the enhancement stage, which is essential for achieving real-time endoscopic video enhancement.

3.2.2. Degradation Guided Enhancement Module

As shown in Figure 3a, building on the degradation representation generated by the DAM, we design a degradation guided enhancement module (DGEM) that generates the enhanced image in three sequential stages.
Shallow feature extraction. We extract low-level structural information from the input image using two stacked 3 × 3 convolutional layers. This step preserves local edges and textures, which are essential for fine-grained restoration.
Feature modulation. Inspired by the Swin Transformer [35], we adopt a shifted-window attention mechanism to capture both local and long-range dependencies. Building upon the standard multi-head attention mechanism, we introduce a degradation-aware modulation applied specifically to the value vectors in each attention head, as shown in Figure 3c. As shown in (3), by incorporating the degradation representation into the values, the attention module can adaptively emphasize or suppress features according to the estimated degradation in the input image.
V ^ = V ⊙ d MSA ( I ) = [ Concat h = 1 H ( Softmax ( Q h K h T d h ) V ^ ) ] W o
where d denotes the compressed degradation representation, Q h , K h , and V h correspond to the query, key, and value of the h_th attention head, and d h represents the dimensionality of each head.
Unlike modulating the query or key vectors, which would alter the attention distribution itself, modulating the Values directly adjusts the content carried through the attention mechanism, allowing the network to integrate degradation information without disrupting spatial relationships. This strategy enables more effective feature aggregation that is guided by the current degradation conditions, ultimately improving the fidelity and robustness of the enhanced image.
Reconstruction. Finally, the modulated features are passed into a sub-pixel convolution upsampling module, which ensures that the final output benefits from degradation-aware guidance while maintaining high spatial fidelity.

3.2.3. Cyclical Consistency

In the CycleGAN framework, style transfer between two image domains is achieved by employing two generators and two discriminators under the constraint of cyclical consistency. Inspired by this idea, we construct a cyclical consistent adversarial network between the low-quality and high-quality image domains, denoted as L , H , respectively.
As illustrated in Figure 4a, the enhancement process L → H is carried out by the combination of DAM and DGEM, while the degradation process H → L is directly implemented by predefined degradation models (PDMs) based on degradation parameters d p . For simplicity, we use G H to denote the combination of DAM, DGEM and use G L to denote PDMs, respectively. Meanwhile, the appearance discriminators D L , D H compete with the corresponding generators to correctly differentiate the synthetic images from the real images, denoted by (4) and (5). To encourage the generation of high-quality enhanced images, we extend the standard two-discriminator framework by adding a high-frequency discriminator D h f that focuses specifically on the high-frequency of the high-quality image domain. This design aims to guide the generator in preserving fine-grained details and texture information, and its corresponding adversarial loss is defined as (6).
L adv ( G H , D H ) = E x h ∼ p ( x h ) [ log ( D H ( x h ) ) ] + E x l ∼ p ( x l ) [ log ( 1 − D H ( G H ( x l ) ) ) ]
L adv ( G L , D L ) = E x l ∼ p ( x l ) [ log ( D L ( x l ) ) ] + E x h ∼ p ( x h ) [ log ( 1 − D L ( G L ( x h ) ) ) ]
L adv ( G H , D h f ) = E x h ∼ p ( x h ) [ log ( D h f ( F h p ( x h ) ) ) ] + E x l ∼ p ( x l ) [ log ( 1 − D h f ( F h p ( G H ( x l ) ) ) ) ]
where F h p denotes a highpass filter.
Figure 4. Diagram of the cyclical consistency. (a) The dataflow of the whole model. L → H : achieved by the DAM and the DGEM with an additional output d c . H → L : achieved based on PDMs and the regression of d p . (b) Calculation of the cyclical consistency loss. Solid lines indicate the dataflow when generating reconstructed synthetic images, and dashed lines indicate the cyclical consistency loss calculation.
It should be note that, shown as the purple part of Figure 4a, different multilayer perceptrons (MLPs) are utilized to regress degradation parameters involved in different image degradation models. These MLPs estimate explicit degradation parameters d p from the learned implicit degradation representation d c , which are then used to synthesize degraded images back. Taking the classical image blurring model I ^ = σ ⊗ I + n as an example, the regression module can estimate a spatially varying blur kernel σ ^ i , j together with a noise intensity field n ^ i , j . These estimated parameters are further used to degrade the enhanced images back to the low-quality domain L , thereby enabling training with paired high- and low-quality images. Importantly, by grounding the degradation process in physically meaningful models rather than in an unconstrained learned generator, our framework avoids introducing unrealistic artifacts and ensures that the cyclical consistency better reflects the actual imaging degradation.
As shown in Figure 4b, the cyclical consistency loss is defined as (7), which forces the reconstructed synthetic images G L ( G H ( l A ) ) and G H ( G L ( h B ) ) to be identical to their respective inputs l A and h B .
L cyc ( G H , D ) = E l A ∼ p ( A ) [ ‖ G L ( G H ( l A ) ) − l A ‖ 1 ] + E h B ∼ p ( B ) [ ‖ G H ( G L ( h B , d c A ) ) − h B ‖ 1 ] + E l A , h B ∼ p ( A , B ) [ ‖ D ( G L ( h B , d c A ) ) − D ( l A ) ‖ 1 ]
where the third term denotes the cyclical consistency of the degradation representation, and D in it represents the DAM. It should be mentioned that the degradation representation d c can be only produced by the DAM within G H . Consequently, in the term G L ( h B , d c B ) , the degradation representation d c B can only be replaced with d c A that derived from G H ( l A ) , which means that the degradation representation cycle contains only a single term.

3.3. Degradation Representation Propagation Module

Although the aforementioned single-frame image enhancement model can achieve high-quality endoscopic image restoration, its inference efficiency is insufficient to meet the real-time requirements of intraoperative observation. To address this, we balance the complexity of different model components against overall performance. Specifically, we manually adjust the number of residual blocks and Swin Transformer blocks to adjust the number of parameters and floating-point operations (FLOPs) for each module (detailed in Table 1), concentrating the primary computational cost in the degradation representation extraction stage.
Table 1. Number of parameters and FLOPs for each part of DGGAN.
Based on these adjustments, we further introduce a degradation representation propagation module (DRPM) to fully exploit temporal correlations across video frames. This module leverages the degradation representations of previous frames as implicit priors, propagating and updating them along the temporal dimension to efficiently estimate the degradation representations of future frames. Compared with modeling each frame independently, this propagation mechanism significantly reduces redundant computations by avoiding repeated processing of highly similar information across adjacent frames, thereby lowering overall computational complexity and ensuring real-time enhancement. Moreover, by maintaining inter-frame consistency of the degradation representations, the module enhances the stability and temporal coherence of degradation modeling across sequences.
In implementation, the DRPM is built upon a Transformer architecture as shown in Figure 5. Unlike conventional recurrent neural networks (RNNs) or 3D convolutional networks (3D-CNNs), the self-attention mechanism in Transformers enables flexible modeling of dependencies over a long temporal range, overcoming the locality constraints of convolution or the limited temporal span of RNNs. Through multi-head attention, the module can simultaneously capture multi-scale temporal correlations in the degradation representations, accurately modeling both local dynamics between adjacent frames and long-range dependencies across distant frames. As a result, employing a Transformer not only improves the accuracy and robustness of degradation representation estimation but also preserves temporal consistency and overall stability in the enhanced video sequence.
Figure 5. The architecture of the DRPM.
Importantly, to further improve the efficiency of degradation propagation while maintaining effectiveness, the module does not propagate the full degradation representation D directly. Instead, it uses a compressed representation D ^ as the temporal modeling unit, which is not directly illustrated in Figure 1. This strategy reduces feature dimensionality and computational redundancy while also enhancing the model’s generalization and real-time performance, providing an efficient degradation-aware support for continuous-frame endoscopic image enhancement.

3.4. Training Strategy

The training strategy is detailed in Algorithm 1. In the first N d epochs, the model first pre-trains the DAM using the contrastive loss L c o n on paired endoscopic images generated by artificial degradations, enabling it to capture effective degradation representations. After this stage, we switch to real-world data, training the model on unpaired low-quality and high-quality endoscopic images under the CycleGAN paradigm. During adversarial learning, to train the generators, the associated objective is minimized, while to train the discriminators, the associated objective is maximized. Please note that when the generators are optimized, the weights of the discriminators are fixed and vice versa. At each step, we conduct the following computations, including forward propagation, loss calculation, backward propagation, and updating weights of associated components. Finally, for video datasets, non-key frames are processed by replacing the DAM in the single-frame enhancement model with the DRPM. The training is then conducted using the same loss functions as described above, which guide the DRPM to propagate and generate reasonable degradation representations.
Algorithm 1: Training strategy
Input: A dataset of high quality endoscopic images X h , A dataset of low-quality endoscopic videos X l , A predefined degradation model PDM
  1
for epoch in 0 , N d  do
  2
      Sampling x h from X h
  3
      Generate x ^ l by degrading PDM ( x h )
  4
      Train DAM by minimizing Lcon
  5
end
  6
for epoch in N d , N d + N s  do
  7
      Sampling x h from X h
  8
      Sampling x l from X l as images
  9
      Train G L , G H by minimizing L adv + L cyc
10 
      Train D L , D H , D h f by maximizing L adv
11 
end
12 
for epoch in N d + N s , N  do
13 
      Sampling x h from X h
14 
      Sampling x l t , … , x l t + T from X l as videos
15 
      Replace DAM with DRPM
16 
      Train DRPM by minimizing L adv + L cyc
17 
      Train D L , D H , D h f by maximizing L adv
18 
end
The design of this two-stage dataset utilization scheme is motivated by two key considerations. On one hand, acquiring large-scale paired endoscopic data in clinical practice is extremely challenging, particularly for image pairs exhibiting diverse degradation types, which has a significant impact on the generalization capability of the DAM. On the other hand, models trained solely on artificially degraded pairs often generalize poorly to real surgical videos. By first initializing the degradation representation module with artificial pairs and then adapting the whole model with real unpaired data, our approach combines the advantages of both worlds and ensures better performance in realistic surgical environments.

4. Experiments and Results

In this section, we first evaluate the performance of the proposed method on the SCARED dataset [36] and compare it with several state-of-the-art deep learning-based approaches. We then assess the degradation-awareness capability of the proposed module, demonstrating the effectiveness of the learned degradation representations. Furthermore, we evaluate the performance of our method under different degradation patterns across various surgical scenarios. To gain deeper insights into the contribution of each component, we conduct ablation studies. Finally, we construct a simulated surgical environment to examine the real-time performance of the proposed approach.

4.1. Data Collection and Implementation Details

4.1.1. Datasets and Preprocessing

In this study, we construct two datasets to support model training and evaluation. For the high-resolution images, we adopt the SCARED dataset. Although originally designed for monocular depth estimation, we select 2295 clear and high-quality left-camera images to build a high-quality endoscopic image dataset. For the low-resolution videos, we utilize the spinal endoscopic surgery dataset (referred to as the SES dataset) collected at Peking University People’s Hospital, from which we extract 240 clips, each consisting of 120 frames. The choice of this dataset is motivated by the fact that Percutaneous Endoscopic Lumbar Discectomy (PELD) is performed in a narrow operative field filled with irrigation fluid, and the image is degraded severely, which may pose different levels of challenge to our method.
Specifically, the resolution of the SCARED dataset is 1280 × 1024 , while the spinal endoscopic dataset has resolutions of 720 × 1080 or 1920 × 1080 . In order to reduce the redundant areas of the images, such as pure black areas and the inner wall of the working channel, we crop the images into 320 × 320 patches to construct the low-quality endoscopic dataset. Finally, the SCARED dataset and the SES dataset are randomly split into training, validation, and testing sets following a 7:2:1 ratio, respectively.

4.1.2. Implementation Details

We train our model following the procedure outlined in Algorithm 1. Throughout the training process, we adopt the Adam optimizer with parameters set to β 1 = 0.9 and β 2 = 0.999 . During the training of the DAM, the learning rate is set to 5 × 10 − 5 . For the training of the single-frame enhancement model, the learning rates of G and D are set to 5 × 10 − 5 and 2 × 10 − 4 , respectively. For the DRPM, we configure the time interval for DAM-based degradation estimation to 15 frames, with the same learning rate as the previous stage. All experiments are conducted using two NVIDIA RTX 3090Ti GPUs for training, while a single NVIDIA RTX 3090Ti is employed for model testing.

4.2. Enhancement Performance and Efficiency of DGGAN

We compare our method with several open-source image enhancement algorithms, including degradation-specific approaches as well as two general-purpose methods (Restormer [37] and SwinIR [38]). For our framework, we evaluate both variants: the degradation-guided enhancement model based on DAM (DGGAN-DAM) and the degradation-guided enhancement model based on DRPM (DGGAN-DRPM).
We first conduct experiments on the SCARED dataset, where the low-quality test images are generated by applying degradations based on multiple degradation models, as defined below. Since the ground truth is available in this setting, we adopt PSNR [39] and SSIM [40] as full-reference metrics to evaluate image quality. We then conduct experiments on the spinal endoscopic dataset, where no ground truth is available. In this case, we employ no-reference metrics, including NIQE [41] and PIQE [42], to assess the performance of different models.
In addition, to evaluate the computational efficiency of each model, we reported two key complexity measures—the number of parameters and the floating point operations (FLOPs)—following common practice in super-resolution studies. To measure the computational efficiency of each model, we calculate two essential properties: the number of parameters and FLOPs (Floating-point operations per second). The number of parameters in a model describes its size, with fewer parameters indicating a more lightweight model. FLOPs can describe the computational complexity of the model and have become a widely used metric to measure the efficiency of models.

4.2.1. Degradation Models

On the SCARED dataset, we apply four degradation types frequently observed in endoscopic surgery: noise, motion blur, low illumination, and electrocautery smoke. For each type, the corresponding degradation model is implemented as summarized in Table 2, and during the training of our DGGAN model, the associated parameters are regressed to support the degradation process.
Table 2. The corresponding models for different degradation types.
It should be mentioned that although explicit degradation models and their parameters are calculated, we deliberately choose not to use these parameters as direct supervision signals. Instead, our framework learns to encode and exploit degradation representations that implicitly capture the degradation characteristics. This design enables the model to move beyond parameter-specific regression and achieve greater robustness and generalization across diverse degradations.
For the spinal endoscopy dataset, we further construct a degradation representation model (8) tailored to real surgical scenarios, taking into account four major aspects: low resolution, blurring, low illumination, and underwater imaging effects.
I ( x ) = ( β ( α J ( x ) ⊗ k ) γ T ( x ) + A ( 1 − T ( x ) ) ) ↓ s + n
The parameter ranges and degradation visual performance are shown in Appendix A.

4.2.2. Results on SCARED Dataset

The performance of multiple models on the SCARED dataset is shown in Table 3 and Figure 6. It is worth noting that since our DGGAN framework incorporates an inherent super-resolution step, for competing methods without an upsampling module, we adopt a strategy of first applying enhancement and then bicubic interpolation for resolution restoration. While this inevitably reduces their final performance, it keeps their inference efficiency comparable.
Table 3. Performance comparison in terms of PSNR (dB), SSIM, parameter counts, FLOPs, and running time (s) of the proposed DGGAN with the SOTAs on the SCARED dataset. Bold indicates the best result, underline indicates the second-best result. ↑ and ↓ denote whether a higher or lower value represents better performance, respectively.
Figure 6. Qualitative comparison of enhancement results on degraded endoscopic datasets. The first four rows correspond to the SCARED dataset under four degradation types: noise, motion blur, smoke, and low light. The last row shows results on our SES dataset.
For the denoising task, we compare against DnCNN [43], CBDNet [44], and Uformer [45]. As denoising represents the simplest enhancement scenario, the performance of all models is remarkable except DnCNN, since it was trained for Grayscale image denoising. Notably, our DGGAN-DRPM achieves results only slightly lower than Restormer and SwinIR, while requiring just 30 % and 7.3 % of their FLOPs, respectively.
For the deblurring task, we select DeblurGAN [46], MIMOUNet [47], and DBGAN [48]. In this case, DGGAN-DAM achieves the best overall performance, surpassing Restormer by 0.1083 in SSIM, while DGGAN-DRPM reaches 0.7785 SSIM, ranking just behind DGGAN-DAM and SwinIR.
For the low-light enhancement task, we compared against Zero-DCE [49], Retinex-Net [50], and EnlightenGAN [25]. Here, DGGAN-DAM nearly matches Restormer’s performance (a marginal SSIM drop of 0.0038 ), while DGGAN-DRPM ranks third overall, outperforming SwinIR by 0.0327 .
For the electrocautery smoke removal task, we selected DehazeNet [51], FFANet [52], and MSBDN [53]. Again, DGGAN-DAM achieves the best results, though DGGAN-DRPM underperforms compared to Restormer and SwinIR. We attribute this to the inherently stochastic nature of smoke diffusion, which limits the ability of the DRPM to effectively propagate degradation representations across frames.
Overall, these results highlight that even when compared against task-specific enhancement models under comparable efficiency settings, our framework consistently achieves an outstanding balance between computational efficiency and enhancement quality. In particular, for each degradation type, we deliberately choose relatively lightweight baseline models to ensure similar inference efficiency, thereby making the performance gains of our method more directly attributable to its superior balance between efficiency and effectiveness.
In addition, we present the PSNR curves over image frames under DRPM-based degradation representation propagation for four different degradation types, as illustrated in Figure 7. The curves exhibit a downward sawtooth pattern despite the degradation condition remaining unchanged. Specifically, after each DAM-based re-estimation, the restoration quality reaches a local peak. However, as the subsequent degradation representations are obtained through temporal propagation rather than re-estimation, error accumulation gradually occurs, leading to a progressive decline in PSNR. Once the DAM module is triggered again, the degradation representation is re-estimated, and the restoration quality correspondingly improves, forming the observed sawtooth trend.
Figure 7. The PSNR curves over image frames under DRPM-based degradation representation propagation for four different degradation types.

4.2.3. Results on SES Dataset

We conduct evaluations on our SES dataset subsequently. Although the performance can only be assessed using no-reference quality metrics, specifically NIQE and PIQE, the results in Table 4 and Figure 6 clearly demonstrate that our method significantly outperforms both Restormer and SwinIR. This finding highlights the strong capability of DGGAN in handling complex and diverse degradation scenarios. More importantly, in real surgical environments, the challenges differ markedly from those in general image enhancement tasks. While patient-specific anatomical and physiological conditions may vary, the types of visual degradations that occur during a given minimally invasive surgery tend to remain relatively consistent. For example, in spinal endoscopy surgery, we can systematically characterize the degradation process with a degradation model such as the one formulated in (8). Leveraging this insight, DGGAN is able to achieve reliable and real-time enhancement performance. This implies that, for a specific minimally invasive surgical procedure, it is sufficient to identify and model the major degradation types that are likely to occur. Once this is done, a dedicated enhancement model can be efficiently trained, enabling high-quality and real-time endoscopic video enhancement that directly benefits intraoperative visualization and surgical safety.
Table 4. The Evaluation results of NIQE/PIQE on our SES dataset. Bold indicates the best result, underline indicates the second-best result. ↓ denote a lower value.

4.2.4. Robustness to Abrupt Degradation Change

To evaluate the robustness of the proposed framework under sudden degradation transitions, we design a controlled experiment that simulates abrupt changes in imaging conditions within a continuous video sequence. Specifically, the degradation type and its parameters are intentionally switched at predefined frames, where random motion blur or noise is introduced or removed while low-illumination degradation is consistently present, thereby generating temporally adjacent segments with significantly different degradation characteristics. To enable this evaluation, the model is retrained using data containing combined noise, motion blur and low illumination degradations, ensuring that it can properly handle such composite degradation patterns. All other experimental settings remain identical to those described in the main experiments.
Quantitative and visual results shown in Figure 8 demonstrate that the proposed method maintains stable enhancement performance across most frames despite abrupt degradation shifts. However, a noticeable performance drop is observed in the first few frames immediately after each transition. The output quality in these frames is relatively lower than that in subsequent frames within the same segment, after which the performance rapidly recovers and stabilizes. This behavior indicates that the model possesses a certain degree of robustness to degradation discontinuities but still requires a short adaptation period to fully adjust to newly introduced degradation patterns.
Figure 8. Performance when abrupt degradation change. (a) Temporal variation of PSNR and SSIM over time. The purple arrows indicate frames in which the degradation representation is estimated by the DAM module, while the green line denotes the timing and pattern of degradation transitions. (b) Visual enhancement performance of frames before and after the time points marked by the yellow triangles in (a).
Further analysis reveals that this phenomenon mainly stems from the temporal propagation mechanism employed in the framework. At the moment of degradation change, the degradation representation of the current frame is inherited from preceding frames rather than being instantly re-estimated, which limits the responsiveness of the model to sudden variations. Once sufficient new-frame information is accumulated, the internal representation gradually aligns with the new degradation condition, leading to restored enhancement performance. These observations suggest that while the current design effectively balances efficiency and temporal consistency, incorporating a rapid re-estimation or change-detection mechanism could further improve adaptability to abrupt degradation transitions.

4.3. Degradation Representation Performance

Degradation representation learning is a key component of our DGGAN framework, as it enables the model to obtain an implicit estimation of endoscopic image degradations, which in turn guides the overall training process. To validate the effectiveness of this learned representation, we performed a visualization analysis.
Specifically, based on the degradation types defined in Table 2, we apply four levels of artificial degradations to high-quality endoscopic images, denoted as L1–L4. These degraded images are then fed into the DAM to extract their degradation representations. To facilitate interpretation, we further project the high-dimensional representations into a two-dimensional space using PCA [54] for visualization. As shown in Figure 9, the DAM successfully learns meaningful degradation features: different degradation types, such as noise and motion blur, form clearly separable clusters. Even in the case of more complex degradations typical of spinal endoscopic surgery, as modeled in (8), the DAM is still able to capture and distinguish degradation severity to a certain extent, as indicated by the arrow. These results confirm that the DAM provides effective and discriminative degradation representation learning, which forms the basis for robust enhancement performance in diverse surgical scenarios.
Figure 9. The degradation representation performance of the DAM under different degradation types. Arrows indicate the distribution of representations under different degradation levels.

4.4. Ablation Study

4.4.1. Effectiveness of the Degradation Representation

In this study, we investigate the role of degradation representation in our framework. To this end, we remove the pretraining stage of the DAM module in Algorithm 1 (referred to as DGGAN w/o p), which to some extent simulates the vanishing of degradation representation learning, and compare it with our full model. The visualization results of the learned degradation representations and the corresponding enhancement performance on the SES dataset are shown in Figure 10 and Table 5, respectively. Without pretraining, the model almost completely loses its ability to capture meaningful degradation representations, which directly leads to a significant drop in enhancement performance. We attribute this to the fact that, without pretraining, the DAM is treated merely as a generic feature extractor rather than a degradation-oriented encoder, thus losing its capacity to represent degradation effectively.
Figure 10. The degradation representation performance on SES dataset with or without pretraining. Arrows indicate the distribution of representations under different degradation levels.
Table 5. The enhancement performance with or without the DAM pretraining. Bold indicates the best result. ↑ and ↓ denote whether a higher or lower value represents better performance, respectively.

4.4.2. Effectiveness of the Degradation Propagation

In this part, we study the impact of different propagation intervals on model performance and inference efficiency. In our default setting, the propagation interval Δ T is set to 15. To further examine its influence, we test the model with Δ T values of 3, 5, 10, 20, and 30, and report the average performance and efficiency in Table 6. The results show that smaller Δ T values yield better enhancement quality but lower inference efficiency, whereas larger Δ T values achieve the opposite trade-off. This observation is consistent with our intuition: shorter intervals allow more accurate degradation estimation, while longer intervals reduce computational redundancy. Considering that endoscopic videos are typically recorded at 30 fps, we choose Δ T = 15 as a balance, which ensures satisfactory enhancement quality while maintaining real-time performance.
Table 6. The enhancement performance under different propagation interval Δ T . Bold indicates the best result. ↑ and ↓ denote whether a higher or lower value represents better performance, respectively.

4.4.3. Degradation Representation Injection

To validate the design choice of incorporating the degradation representation into the Value branch of the degradation-aware Swin Transformer block, we conduct an ablation study comparing different injection strategies within DGEM. Specifically, we evaluate four alternative conditioning schemes in addition to the used Value modulation mechanism: (1) Query modulation, (2) Key modulation, (3) cross-attention conditioning, and (4) direct feature concatenation. All variants share the same backbone architecture, training data, and optimization settings to ensure fair comparison, differing only in how the degradation representation is utilized.
As listed in Table 7, experimental results on DGGAN-DRPM and its variants show that Value modulation achieves the best trade-off between restoration quality and computational efficiency. Modulating the Query or Key branches leads to slightly inferior PSNR/SSIM performance and introduces higher sensitivity to degradation estimation errors, likely because these branches directly affect attention weight computation, amplifying mismatch effects. Cross-attention conditioning yields competitive enhancement quality; however, it significantly increases computational cost and memory consumption due to the additional attention operations. Direct concatenation exhibits the lowest performance, suggesting that naive feature fusion is insufficient for effectively guiding the attention mechanism.
Table 7. The enhancement performance using different degradation representation injection. Bold indicates the best result. ↑ and ↓ denote whether a higher or lower value represents better performance, respectively.
In contrast, Value modulation preserves the original attention structure while adaptively adjusting feature aggregation according to degradation characteristics. Since only the Value vectors are modulated, the attention weights remain stable, improving robustness to degradation representation noise while maintaining low computational overhead. These results demonstrate that the proposed conditioning strategy offers a balanced solution in terms of enhancement accuracy, stability, and inference speed.

4.4.4. Effectiveness of the Cyclical Consistency

In this study, we analyze the contribution of each component to the cyclical consistency loss. As defined in (7), the cyclical loss consists of three terms, L c l , L c h , and L c d . We conduct ablation experiments by removing each of these terms individually and report the results in Table 8. The results show that L c l and L c h have a major impact on the model’s performance, and removing either of them leads to a substantial drop in enhancement quality. In contrast, the role of L c d is relatively minor, as it mainly penalizes inconsistency in degradation representation. Specifically, L c d improves the stability of degradation estimation and prevents representation drift during propagation. Therefore, L c d serves as a structural regularization term that enhances robustness and theoretical coherence of the degradation modeling framework.
Table 8. The effectiveness of the cyclical consistency loss. Bold indicates the best result. ↑ and ↓ denote whether a higher or lower value represents better performance, respectively.

4.5. Enhanced Segmentation Performance

To quantitatively evaluate the impact of the proposed video enhancement method on downstream tasks, we conduct vessel segmentation comparative experiments on both the original degraded DRIVE [55] images and the enhanced images. The segmentation network FR-UNet [56] and pretrained model are applied in both settings to ensure a fair comparison, which means the segmentation model is not retrained on our enhanced dataset. As summarized in Table 9, after applying the proposed enhancement algorithm, the segmentation accuracy consistently improves across all evaluation metrics. Specifically, the Area under Curve (AUC) increases from 0.9517% to 0.9582%, while the Intersection-over-Union (IoU) improves by 2 percentage points.
Table 9. The vessel segmentation performance on both the origin and the enhanced DRIVE dataset using FR-UNet. Bold indicates the best result.
Qualitative results in Figure 11 further demonstrate that the enhanced images exhibit clearer tissue boundaries, improved local contrast, and reduced artifacts, which effectively facilitate the segmentation network’s discrimination between adjacent anatomical regions. These results validate that the proposed image enhancement algorithm not only improves visual quality but also provides a more reliable and discriminative input for tissue segmentation. This confirms the practical value of the enhancement module as a preprocessing step for downstream clinical perception tasks, particularly in challenging endoscopic surgical scenarios.
Figure 11. The enhanced performance on the DRIVE dataset and the downstream segmentation performance comparison.

5. Conclusions

We developed and validated an efficient stepwise endoscopic video enhancement framework. Our approach estimates degradation representations of images and propagates them across video frames, thereby enabling high-quality real-time enhancement of endoscopic videos. By achieving a superior balance between performance and efficiency compared with several state-of-the-art methods, our work demonstrates the value of degradation representations in image enhancement. Moreover, this degradation-guided enhancement paradigm implies that, for endoscopic images in a specific minimally invasive surgery, it is sufficient to evaluate and model the potential degradations to train an enhancement model, which can then deliver high-quality real-time enhancement during surgery.
Nevertheless, our method has certain limitations. On the one hand, to ensure real-time performance, the parameter size of the single-frame enhancement module was reduced, leading to a performance gap compared with the best-performing state-of-the-art methods. On the other hand, our current design employs a fixed interval for degradation feature estimation, which may limit optimal performance. For instance, sudden events such as tissue bleeding may not be captured promptly, resulting in suboptimal enhancement outcomes due to delayed degradation estimation.
Future research will concentrate on two aspects. The first is comprehensive end-to-end evaluation in real surgical settings to verify the practical feasibility of real-time application. The second involves leveraging the emerging capability of large language models (LLMs) for medical image analysis, as evidenced by prior studies [57,58], to incorporate LLM-guided or vision-language reasoning for high-level scene interpretation, adaptive key-frame scheduling, and controllable enhancement through surgeons’ natural-language constraints. These directions aim to enhance clinical reliability and accelerate real-world deployment.

Author Contributions

Conceptualization, H.X. and Z.N.; methodology, H.X.; software, H.X.; validation, H.X. and T.P.; investigation, T.P.; data collection, H.X. and T.P.; writing—original draft preparation, H.X.; writing—review and editing, H.X. and Z.N.; visualization, H.X. and T.P.; supervision, Z.N. and X.-J.L.; funding acquisition, Z.N. All authors have read and agreed to the published version of the manuscript.

Funding

This work is supported by Beijing Natural Science Foundation (L242163) and the Joint Fund for Regional Innovation and Development co-established by NSFC and Provincial Governments under Grant U23A20471.

Institutional Review Board Statement

Ethical review and approval were waived for this study due to the data used were obtained from the public databases.

Data Availability Statement

The data presented in this study are openly available in SCARED at https://endovissub2019-scared.grand-challenge.org/ (accessed on 3 May 2024).

Acknowledgments

The authors gratefully acknowledge the helpful comments and suggestions of the reviewers, which improved the presentation.

Conflicts of Interest

The authors declare no conflicts of interest.

Appendix A

This appendix summarizes the parameter ranges for each degradation type discussed in the manuscript and illustrate the endoscopic images under varying degradation levels in Figure A1.
Random noise: Random Gaussian noise with intensity in the range of [0, 30].
Motion blur: Random Gaussian blur where the kernel size k ∈ [ 3 , 19 ] , the standard deviation σ ∈ [ 0.2 , 4.0 ] , and the motion angle θ ∼ U ( − 45 ∘ , 45 ∘ ) .
Low light: The image is converted to HSV, where the normalized V channel is additively modulated within [−0.25, 0], then converted back to RGB with random noise in [0, 10].
Smoke: Fog is synthesized by adding a plasma-fractal haze field [59] with intensity α ∈ [ 1.0 , 3.0 ] and roughness β ∈ [ 1.4 , 2.0 ] , followed by contrast compression and clipping.
SES: The size of the isotropic Gaussian kernel k in (8) is fixed to 7 × 7 , and the range of the width of the kernel σ is set to [0.2, 4.0]. The noise level range n is set to [0, 30]. Then, the parameters α , β , γ are set in range of [0.9, 1.0], [0.8, 1.0] and [1.0, 2.5], respectively. The attenuation coefficients are set to β R ∈ [ 1.5 , 3.0 ] , β G ∈ [ 0.8 , 1.5 ] , β B ∈ [ 0.3 , 0.8 ] with β R > β G > β B . The background light B ∈ [ 0.05 , 0.8 ] , and the depth is normalized to [0, 1].
As illustrated in Figure 4, these parameters can cover the range of image degradation in PELD endoscope scenarios.
Figure A1. The endoscopic images under varying degradation levels.

References

  1. Fuchs, K. Minimally invasive surgery. Endoscopy 2002, 34, 154–159. [Google Scholar] [CrossRef] [Scilit]
  2. Mahfoz, T.B. Single-port versus multiport robotic surgery in head and neck procedures: A systematic review and meta-analysis of surgical parameters. Ann. Surg. Treat. Res. 2025, 109, 293–301. [Google Scholar] [CrossRef] [Scilit]
  3. Pan, M.; Li, Q.; Li, S.; Mao, H.; Meng, B.; Zhou, F.; Yang, H. Percutaneous endoscopic lumbar discectomy: Indications and complications. Pain Physician 2020, 23, 49. [Google Scholar] [CrossRef] [Scilit]
  4. Swee, S.K.; Chen, L.C.; Ching, T.S. Contrast Enhancement in Endoscopic Images Using Fusion Exposure Histogram Equalization. Eng. Lett. 2020, 28, 73–81. [Google Scholar]
  5. Tan, W.; Xu, C.; Lei, F.; Fang, Q.; An, Z.; Wang, D.; Han, J.; Qian, K.; Feng, B. An endoscope image enhancement algorithm based on image decomposition. Electronics 2022, 11, 1909. [Google Scholar] [CrossRef] [Scilit]
  6. Halidou, A.; Mohamadou, Y.; Ari, A.A.A.; Zacko, E.J.G. Review of wavelet denoising algorithms. Multimed. Tools Appl. 2023, 82, 41539–41569. [Google Scholar] [CrossRef] [Scilit]
  7. Yuan, Q.; Dai, S. Adaptive histogram equalization with visual perception consistency. Inf. Sci. 2024, 668, 120525. [Google Scholar] [CrossRef] [Scilit]
  8. Roy, S.; Bhalla, K.; Patel, R. Mathematical analysis of histogram equalization techniques for medical image enhancement: A tutorial from the perspective of data loss. Multimed. Tools Appl. 2024, 83, 14363–14392. [Google Scholar] [CrossRef] [Scilit]
  9. Yang, S.; Guo, T.; Zhan, R. Capsule Endoscopy Image Deblurring Method Based on Multi-scale Recurrent Network. J. Phys. Conf. Ser. 2023, 2640, 012003. [Google Scholar] [CrossRef] [Scilit]
  10. Song, X.; Tang, H.; Yang, C.; Zhou, G.; Wang, Y.; Huang, X.; Hua, J.; Coatrieux, G.; He, X.; Chen, Y. Deformable transformer for endoscopic video super-resolution. Biomed. Signal Process. Control 2022, 77, 103827. [Google Scholar] [CrossRef] [Scilit]
  11. Huang, W.; Liao, X.; Chen, H.; Hu, Y.; Jia, W.; Wang, Q. Deep local-to-global feature learning for medical image super-resolution. Comput. Med Imaging Graph. 2024, 115, 102374. [Google Scholar] [CrossRef] [Scilit]
  12. Xu, H.; Nie, Z.; Peng, T.; Pan, H.; Liu, X.J. DGGAN: Degradation Guided Generative Adversarial Network for Real-time Endoscopic Video Enhancement. arXiv 2025, arXiv:2512.07253. [Google Scholar] [CrossRef] [Scilit]
  13. Peng, L.; Liu, S.; Xie, D.; Zhu, S.; Zeng, B. Endoscopic video deblurring via synthesis. In Proceedings of the 2017 IEEE Visual Communications and Image Processing (VCIP); IEEE: New York, NY, USA, 2017; pp. 1–4. [Google Scholar]
  14. Tchaka, K.; Pawar, V.M.; Stoyanov, D. Chromaticity based smoke removal in endoscopic images. In Proceedings of the Medical Imaging 2017: Image Processing; SPIE: Bellingham, WA, USA, 2017; Volume 10133, pp. 463–470. [Google Scholar]
  15. Lin, J.; Jiang, M.; Pang, Y.; Wang, H.; Chen, Z.; Yan, C.; Liu, Q.; Wang, Y. A desmoking algorithm for endoscopic images based on improved U-Net model. Concurr. Comput. Pract. Exp. 2021, 33, e6320. [Google Scholar] [CrossRef] [Scilit]
  16. Zou, S.; Long, M.; Wang, X.; Xie, X.; Li, G.; Wang, Z. A CNN-based blind denoising method for endoscopic images. In Proceedings of the 2019 IEEE Biomedical Circuits and Systems Conference (BioCAS); IEEE: New York, NY, USA, 2019; pp. 1–4. [Google Scholar]
  17. Wu, J.; Yuan, Y.; Liu, L.; Cui, H.; Xu, T.; Zhou, M.; Chen, Z.; Xu, B. Research progress of frontier image processing in medical endoscopes. Prog. Med. Devices 2023, 1, 62–77. [Google Scholar] [CrossRef] [Scilit]
  18. Wang, Q.; Pan, N.; Xiong, W.; Lu, H.; Li, N.; Zou, X. Reduction of bubble-like frames using a RSS filter in wireless capsule endoscopy video. Opt. Laser Technol. 2019, 110, 152–157. [Google Scholar] [CrossRef] [Scilit]
  19. Almalioglu, Y.; Ozyoruk, K.B.; Gokce, A.; Incetan, K.; Gokceler, G.I.; Simsek, M.A.; Ararat, K.; Chen, R.J.; Durr, N.J.; Mahmood, F.; et al. EndoL2H: Deep super-resolution for capsule endoscopy. IEEE Trans. Med Imaging 2020, 39, 4297–4309. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  20. Zhang, T.; Yang, J. Transformer with hybrid attention mechanism for stereo endoscopic video super resolution. Symmetry 2023, 15, 1947. [Google Scholar] [CrossRef] [Scilit]
  21. Hayat, M.; Aramvith, S. E-SEVSR—Edge guided stereo endoscopic video super-resolution. IEEE Access 2024, 12, 30893–30906. [Google Scholar] [CrossRef] [Scilit]
  22. Hayat, M.; Aramvith, S.; Achakulvisut, T. Combined channel and spatial attention-based stereo endoscopic image super-resolution. In Proceedings of the TENCON 2023—2023 IEEE Region 10 Conference (TENCON); IEEE: New York, NY, USA, 2023; pp. 920–925. [Google Scholar]
  23. Hayat, M.; Aramvith, S. Saliency-aware deep learning approach for enhanced endoscopic image super-resolution. IEEE Access 2024, 12, 83452–83465. [Google Scholar] [CrossRef] [Scilit]
  24. An, Z.; Xu, C.; Qian, K.; Han, J.; Tan, W.; Wang, D.; Fang, Q. EIEN: Endoscopic image enhancement network based on retinex theory. Sensors 2022, 22, 5464. [Google Scholar] [CrossRef] [Scilit]
  25. Jiang, Y.; Gong, X.; Liu, D.; Cheng, Y.; Fang, C.; Shen, X.; Yang, J.; Zhou, P.; Wang, Z. Enlightengan: Deep light enhancement without paired supervision. IEEE Trans. Image Process. 2021, 30, 2340–2349. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  26. Guo, L.; Xiao, X.; Wu, C.; Zeng, X.; Zhang, Y.; Du, J.; Bai, S.; Xie, J.; Zhang, Z.; Li, Y.; et al. Real-time automated diagnosis of precancerous lesions and early esophageal squamous cell carcinoma using a deep learning model (with videos). Gastrointest. Endosc. 2020, 91, 41–51. [Google Scholar] [CrossRef] [Scilit]
  27. Yamada, M.; Saito, Y.; Imaoka, H.; Saiko, M.; Yamada, S.; Kondo, H.; Takamaru, H.; Sakamoto, T.; Sese, J.; Kuchiba, A.; et al. Development of a real-time endoscopic image diagnosis support system using deep learning technology in colonoscopy. Sci. Rep. 2019, 9, 14465. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  28. Zhu, M.; Chen, Z.; Yuan, Y. DSI-Net: Deep synergistic interaction network for joint classification and segmentation with endoscope images. IEEE Trans. Med Imaging 2021, 40, 3315–3325. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  29. Gunasekaran, H.; Ramalakshmi, K.; Swaminathan, D.K.; Mazzara, M. GIT-Net: An ensemble deep learning-based GI tract classification of endoscopic images. Bioengineering 2023, 10, 809. [Google Scholar] [CrossRef] [Scilit]
  30. Nie, Z.; Xu, M.; Wang, Z.; Lu, X.; Song, W. A Review of Application of Deep Learning in Endoscopic Image Processing. J. Imaging 2024, 10, 275. [Google Scholar] [CrossRef] [Scilit]
  31. Zhu, J.Y.; Park, T.; Isola, P.; Efros, A.A. Unpaired image-to-image translation using cycle-consistent adversarial networks. In Proceedings of the IEEE International Conference on Computer Vision; IEEE: New York, NY, USA, 2017; pp. 2223–2232. [Google Scholar]
  32. He, K.; Fan, H.; Wu, Y.; Xie, S.; Girshick, R. Momentum contrast for unsupervised visual representation learning. In Proceedings of the IEEE/CVF Conference on Computer Vision and Pattern Recognition; IEEE: New York, NY, USA, 2020; pp. 9729–9738. [Google Scholar]
  33. Wang, L.; Wang, Y.; Dong, X.; Xu, Q.; Yang, J.; An, W.; Guo, Y. Unsupervised degradation representation learning for blind super-resolution. In Proceedings of the IEEE/CVF Conference on Computer Vision and Pattern Recognition; IEEE: New York, NY, USA, 2021; pp. 10581–10590. [Google Scholar]
  34. van den Oord, A.; Li, Y.; Vinyals, O. Representation learning with contrastive predictive coding. arXiv 2018, arXiv:1807.03748. [Google Scholar]
  35. Liu, Z.; Lin, Y.; Cao, Y.; Hu, H.; Wei, Y.; Zhang, Z.; Lin, S.; Guo, B. Swin transformer: Hierarchical vision transformer using shifted windows. In Proceedings of the IEEE/CVF International Conference on Computer Vision; IEEE: New York, NY, USA, 2021; pp. 10012–10022. [Google Scholar]
  36. Allan, M.; Mcleod, J.; Wang, C.; Rosenthal, J.C.; Hu, Z.; Gard, N.; Eisert, P.; Fu, K.X.; Zeffiro, T.; Xia, W.; et al. Stereo correspondence and reconstruction of endoscopic data challenge. arXiv 2021, arXiv:2101.01133. [Google Scholar] [CrossRef] [Scilit]
  37. Zamir, S.W.; Arora, A.; Khan, S.; Hayat, M.; Khan, F.S.; Yang, M.H. Restormer: Efficient transformer for high-resolution image restoration. In Proceedings of the IEEE/CVF Conference on Computer Vision and Pattern Recognition; IEEE: New York, NY, USA, 2022; pp. 5728–5739. [Google Scholar]
  38. Liang, J.; Cao, J.; Sun, G.; Zhang, K.; Van Gool, L.; Timofte, R. Swinir: Image restoration using swin transformer. In Proceedings of the IEEE/CVF International Conference on Computer Vision; IEEE: New York, NY, USA, 2021; pp. 1833–1844. [Google Scholar]
  39. Kim, J.; Lee, J.K.; Lee, K.M. Accurate image super-resolution using very deep convolutional networks. In Proceedings of the IEEE Conference on Computer Vision and Pattern Recognition; IEEE: New York, NY, USA, 2016; pp. 1646–1654. [Google Scholar]
  40. Lai, W.S.; Huang, J.B.; Ahuja, N.; Yang, M.H. Deep laplacian pyramid networks for fast and accurate super-resolution. In Proceedings of the IEEE Conference on Computer Vision and Pattern Recognition; IEEE: New York, NY, USA, 2017; pp. 624–632. [Google Scholar]
  41. Mittal, A.; Soundararajan, R.; Bovik, A.C. Making a “completely blind” image quality analyzer. IEEE Signal Process. Lett. 2012, 20, 209–212. [Google Scholar] [CrossRef] [Scilit]
  42. Venkatanath, N.; Praneeth, D.; Maruthi, C.B.; Sumohana, S.C.; Swarup, S.M. Blind image quality evaluation using perception based features. In Proceedings of the 2015 Twenty First National Conference on Communications (NCC); IEEE: New York, NY, USA, 2015; pp. 1–6. [Google Scholar]
  43. Zhang, K.; Zuo, W.; Chen, Y.; Meng, D.; Zhang, L. Beyond a gaussian denoiser: Residual learning of deep cnn for image denoising. IEEE Trans. Image Process. 2017, 26, 3142–3155. [Google Scholar] [CrossRef] [Scilit]
  44. Guo, S.; Yan, Z.; Zhang, K.; Zuo, W.; Zhang, L. Toward convolutional blind denoising of real photographs. In Proceedings of the IEEE/CVF Conference on Computer Vision and Pattern Recognition; IEEE: New York, NY, USA, 2019; pp. 1712–1722. [Google Scholar]
  45. Wang, Z.; Cun, X.; Bao, J.; Zhou, W.; Liu, J.; Li, H. Uformer: A general u-shaped transformer for image restoration. In Proceedings of the IEEE/CVF Conference on Computer Vision and Pattern Recognition; IEEE: New York, NY, USA, 2022; pp. 17683–17693. [Google Scholar]
  46. Kupyn, O.; Budzan, V.; Mykhailych, M.; Mishkin, D.; Matas, J. Deblurgan: Blind motion deblurring using conditional adversarial networks. In Proceedings of the IEEE Conference on Computer Vision and Pattern Recognition; IEEE: New York, NY, USA, 2018; pp. 8183–8192. [Google Scholar]
  47. Cho, S.J.; Ji, S.W.; Hong, J.P.; Jung, S.W.; Ko, S.J. Rethinking coarse-to-fine approach in single image deblurring. In Proceedings of the IEEE/CVF International Conference on Computer Vision; IEEE: New York, NY, USA, 2021; pp. 4641–4650. [Google Scholar]
  48. Zhang, K.; Luo, W.; Zhong, Y.; Ma, L.; Stenger, B.; Liu, W.; Li, H. Deblurring by realistic blurring. In Proceedings of the IEEE/CVF Conference on Computer Vision and Pattern Recognition; IEEE: New York, NY, USA, 2020; pp. 2737–2746. [Google Scholar]
  49. Guo, C.; Li, C.; Guo, J.; Loy, C.C.; Hou, J.; Kwong, S.; Cong, R. Zero-reference deep curve estimation for low-light image enhancement. In Proceedings of the IEEE/CVF Conference on Computer Vision and Pattern Recognition; IEEE: New York, NY, USA, 2020; pp. 1780–1789. [Google Scholar]
  50. Wei, C.; Wang, W.; Yang, W.; Liu, J. Deep retinex decomposition for low-light enhancement. arXiv 2018, arXiv:1808.04560. [Google Scholar] [CrossRef] [Scilit]
  51. Cai, B.; Xu, X.; Jia, K.; Qing, C.; Tao, D. Dehazenet: An end-to-end system for single image haze removal. IEEE Trans. Image Process. 2016, 25, 5187–5198. [Google Scholar] [CrossRef] [Scilit]
  52. Qin, X.; Wang, Z.; Bai, Y.; Xie, X.; Jia, H. FFA-Net: Feature fusion attention network for single image dehazing. In Proceedings of the AAAI Conference on Artificial Intelligence; AAAI: Palo Alto, CA, USA, 2020; pp. 11908–11915. [Google Scholar]
  53. Dong, H.; Pan, J.; Xiang, L.; Hu, Z.; Zhang, X.; Wang, F.; Yang, M.H. Multi-scale boosted dehazing network with dense feature fusion. In Proceedings of the IEEE/CVF Conference on Computer Vision and Pattern Recognition; IEEE: New York, NY, USA, 2020; pp. 2157–2167. [Google Scholar]
  54. Abdi, H.; Williams, L.J. Principal component analysis. Wiley Interdiscip. Rev. Comput. Stat. 2010, 2, 433–459. [Google Scholar] [CrossRef] [Scilit]
  55. Staal, J.; Abràmoff, M.D.; Niemeijer, M.; Viergever, M.A.; Van Ginneken, B. Ridge-based vessel segmentation in color images of the retina. IEEE Trans. Med Imaging 2004, 23, 501–509. [Google Scholar] [CrossRef] [Scilit]
  56. Liu, W.; Yang, H.; Tian, T.; Cao, Z.; Pan, X.; Xu, W.; Jin, Y.; Gao, F. Full-resolution network and dual-threshold iteration for retinal vessel and coronary angiograph segmentation. IEEE J. Biomed. Health Inform. 2022, 26, 4623–4634. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  57. Boostani, M.; Bánvölgyi, A.; Zouboulis, C.C.; Goldfarb, N.; Suppa, M.; Goldust, M.; Lőrincz, K.; Kiss, T.; Nádudvari, N.; Holló, P.; et al. Large language models in evaluating hidradenitis suppurativa from clinical images. J. Eur. Acad. Dermatol. Venereol. JEADV 2025, 39, e1052–e1055. [Google Scholar] [CrossRef] [Scilit]
  58. Boostani, M.; Bánvölgyi, A.; Goldust, M.; Cantisani, C.; Pietkiewicz, P.; Lőrincz, K.; Holló, P.; Wikonkál, N.M.; Paragh, G.; Kiss, N.; et al. Diagnostic performance of GPT-4o and Gemini Flash 2.0 in acne and rosacea. Int. J. Dermatol. 2025, 64, 1881–1882. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  59. FLHerne. mapgen: Python Diamond-Square Map Generator. 2015. Available online: https://github.com/FLHerne/mapgen (accessed on 27 February 2026).
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.