Repeatability, Reproducibility and Inter-Device Agreement of Anterior and Posterior Corneal Parameters Measured with Visionix 130+ and Pentacam HR in Healthy Eyes

,
,
,
,
and
1
GI-2092 Optometry, Applied Physics Department (Optometry Area), Universidade de Santiago de Compostela, 15705 Santiago de Compostela, Spain
2
AC24 Optometry Group, Optometry Group, Instituto de Investigacion Sanitaria Santiago de Compostela (IDIS), 15706 Santiago de Compostela, Spain
3
Models of Optimization, Decision, Statistics and Applications Research Group (MODESTYA), Department of Statistics, Mathematical Analysis and Optimization, Universidade de Santiago de Compostela, 15705 Santiago de Compostela, Spain
*
Authors to whom correspondence should be addressed.
Photonics2026, 13(12), 1120;https://doi.org/10.3390/photonics13121120 
(registering DOI)
This article belongs to the Special Issue Recent Advances in Biomedical Optics and Biophotonics

Abstract

Assessing the validity and measurement performance of optical instrumentation is essential to ensure its clinical utility. Therefore, the study’s aims were to analyze the repeatability and reproducibility of two multi-diagnostic platforms (Visionix 130+ and Pentacam HR) in the measurement of corneal characteristics, and to assess inter-device agreement. Anterior and posterior corneal curvature, eccentricity and central corneal thickness were measured using both devices in 31 healthy subjects. Each participant was subjected to six measurements per device, divided equally between two observers (three per observer). Both devices showed moderate-to-high reliability for anterior parameters and central corneal thickness (ICC > 79.89). For posterior parameters, Pentacam HR maintained moderate-to-high reliability (ICC: 78.28–99.34), whereas Visionix 130+ revealed decreased reliability (ICC: 5.16–77.77). Regarding eccentricity, high reliability was observed for Pentacam HR (ICC: 96.55), while Visionix 130+ exhibited low reliability (ICC: 68.81). No agreement was found between the results provided by the two devices for most of the parameters assessed. It is concluded that both devices demonstrate good repeatability and reproducibility when assessing anterior corneal surface and central corneal thickness, with the exception of eccentricity measured by Visionix 130+. For posterior surface measurements, Pentacam HR demonstrates moderate-to-high reliability, while Visionix 130+ exhibits low reliability for most parameters. Finally, the results of both devices are not interchangeable.