Rethinking Processed EEG Monitoring in Children: Developmental Neurophysiology, Age-Dependent Index Validity, and Pediatric-Specific Validation

,
,
and
1
Department of Anesthesiology, University of Mississippi Medical Center, 2500 N State St, Jackson, MS 39216, USA
2
Department of Pediatrics, University of Mississippi Medical Center, Jackson, MS 39216, USA
*
Author to whom correspondence should be addressed.

Highlights

What are the main findings?
  • The anesthetized EEG reorganizes rapidly and partly non-monotonically during infancy and childhood, so the spectral, coherence, suppression, and entropy-related features entering a processed index are not developmentally stationary.
  • Pediatric studies report lower, higher, flattened, and paradoxical processed-index behavior across developmental strata and anesthetic conditions rather than one uniform direction of bias.
What are the implications of the main findings?
  • Clinical benefit from EEG-informed management supports the monitoring strategy that was evaluated rather than the numerical target embedded within it.
  • Pediatric validation should specify the developmental population, anesthetic agent, reference state, physiological signal, technical conditions, external population, and clinical context in which an output retains interpretable meaning.

Abstract

Processed electroencephalographic (pEEG) monitoring supports anesthetic titration in children, but the same numerical target may not represent an equivalent brain state across development. This focused narrative review examines developmental EEG physiology, age-dependent index performance, and the evidence needed for pediatric validation. Spectral power, coherence, discontinuity, suppression, and entropy-related features change with maturation and anesthetic context. Pediatric studies consequently report lower, higher, flattened, or paradoxical index behavior across age groups. We distinguish exposure validity, brain-state validity, and clinical validity: tracking anesthetic concentration, identifying an operationally defined state, and improving management are different claims. Clinical benefit from an EEG-informed strategy does not independently validate its numerical target. Newer computational approaches may improve prediction, but their interpretation remains tied to the reference used for training or validation. Regulatory clearance, independent replication, and clinical-outcome evidence also answer different questions. We propose an eight-domain framework to organize these requirements. Uncertainty about universal targets is greatest in infants and young children and should not be interpreted as evidence that pEEG lacks clinical value. Interpretation should integrate age, anesthetic drugs, raw and spectral EEG, signal quality, and clinical context rather than rely on a numerical value alone.

1. Introduction

Processed electroencephalographic (pEEG) monitoring summarizes features of cortical electrical activity as numerical outputs intended to support assessment of anesthetic state. The interpretation of fixed index ranges depends on whether the relationship between the EEG and the displayed value is transferable across patients. Age-related changes occur across the lifespan [1]; childhood presents a particular challenge because oscillatory features emerge, reorganize, and sometimes disappear as neural circuits mature [2].
Widely used indices such as the Bispectral Index (BIS) and Patient State Index (PSi) were developed and initially evaluated largely in adults, with subsequent pediatric studies spanning heterogeneous ages and anesthetic conditions [3,4,5,6]. Their ability to track exposure does not establish that a fixed numerical range has the same physiological meaning throughout development.
Under sevoflurane, theta and alpha activity emerge around 4 months, frontal alpha predominance around 6 months, and coherent frontal alpha around 10 months; early frontal slow-oscillation coherence is subsequently lost [2]. Total EEG power follows different trajectories under sevoflurane and propofol [7,8]. Discontinuity and suppression are influenced by developmental age, anesthetic exposure, and practice-related factors [9,10,11].
Studies report low, high, flattened, and paradoxical processed-index responses in different pediatric populations [12,13,14,15]. Differences in age, anesthetic exposure, co-administered drugs, measurement phase, and reference standard complicate comparison and leave equivalent calibration across ages unproven.
Recent EEG-guided trials have reduced sevoflurane exposure, although recovery findings vary [16,17]. Infant-specific machine-learning studies and expansion of a regulatory indication also warrant reassessment of what pediatric validation establishes [18,19].
Developmental limitations of processed EEG and the uncertain equivalence of index values across ages have been discussed previously [1,20,21,22,23]. This review builds on that work by separating exposure, brain-state, and clinical validity; distinguishing benefit from a monitoring strategy from validation of its numerical target; and proposing an eight-domain framework for future pediatric validation.

Literature Search and Review Approach

This focused narrative review examined developmental EEG physiology, age-dependent monitor performance, and pediatric validation. Targeted PubMed searches covered the literature available through 20 August 2026, with no lower publication-date limit. Pediatric terms (child, pediatric, paediatric, infant, neonate, adolescent) were combined with electroencephalography, anesthesia or sedation, and terms including Bispectral Index, Patient State Index, entropy, spectral edge frequency, spectrogram, density spectral array, coherence, spectral power, suppression, discontinuity, and depth of anesthesia. An illustrative combination of these concepts is: (child OR infant OR adolescent) AND (EEG OR electroencephalography) AND (anesthesia OR sedation) AND (“bispectral index” OR “patient state index”). Reference lists of relevant studies and reviews were also examined. One Japanese-language report (cited in Section 4) was assessed using its English-language PubMed abstract.
Priority was given to human pediatric studies addressing EEG development, monitor performance against an identifiable reference, or EEG-guided management. Original studies, reviews, letters, case reports, and regulatory documents were considered; foundational adult studies provided context. Section 4 presents selected studies relevant to age-related performance and validity; Table S1 compares study designs, reference standards, and age comparisons underlying methodological heterogeneity. Studies were classified by the reference assessed: anesthetic dose or concentration (exposure validity), an independently assessed clinical or physiological state (brain-state validity), another processed index (inter-index correlation or agreement), or effects of a monitoring strategy on management or outcomes (clinical validity). Categories identify the question assessed, not study quality or successful validation; a study may address more than one category.
Findings were synthesized narratively by developmental stage and validity question, without quantitative pooling. Study selection was purposive, and the tables are not exhaustive inventories. All four authors participated in study selection and assignment of studies to table categories. Relevant reports may have been missed, introducing selection bias. Heterogeneous reference standards and overlapping cohorts further limit comparisons across studies.

2. What Does a Processed EEG Index Measure?

A displayed pEEG value is the endpoint of a processing chain, and each stage of that chain can constrain what the number can mean. Understanding the chain clarifies why an index may behave sensibly in one population and unpredictably in another without any component having malfunctioned.
Acquisition comes first. A small number of frontally placed electrodes record scalp potential differences that reflect spatially weighted, volume-conducted neural activity together with non-neural contributions. Electrode contact and montage require explicit attention: the developmental study by Cornelissen et al. excluded poorly contacting ear electrodes and specified its re-referencing procedure [2]. Preprocessing follows: filtering defines the frequency content available for subsequent analysis, while artifact-handling and signal-quality procedures may identify, reject, downweight, or flag segments affected by electromyographic activity, movement, poor electrode contact, or electrical interference. The signal presented to subsequent feature extraction is therefore a conditioned scalp EEG rather than the originally acquired waveform, and the proportion of recording time judged usable is itself a variable that differs across patients and settings.
Feature extraction then reduces the conditioned signal to a set of quantitative descriptors. These typically include measures of spectral distribution, such as power within defined frequency bands or the spectral edge frequency below which a specified proportion of power lies. They also include measures of signal regularity or complexity, such as entropy-based descriptors, and measures of discontinuity, such as the proportion of time meeting criteria for suppression. Some descriptors have explicit mathematical definitions and can in principle be reproduced from the same recording if preprocessing, parameter definitions, and implementation are held constant. Others form part of proprietary processing pipelines [24].
The transformation stage combines selected descriptors into the displayed index. For proprietary commercial indices, elements of this architecture may be described publicly, but the complete transformation is generally not available for independent inspection. For BIS, for example, constituent subparameters including spectral, bispectral, and suppression-related features have been described, whereas the exact weighting and rules by which they contribute to the final index remain proprietary [24]. Because these transformations were developed and calibrated empirically, interpretation of their outputs reflects relationships between EEG features, anesthetic exposure, and clinical reference states observed in their development and validation populations. Foundational development and early validation work for widely used indices was conducted largely in adults [3,4]. Whether those relationships remain quantitatively transportable to a developing EEG is therefore an empirical question rather than an intrinsic property of the scale.
Reporting an output on a bounded numerical scale can create the impression of a directly measured physiological quantity, but the scale has no physical unit. Its interpretation depends on empirical relationships between the processed EEG and the reference conditions against which the index has been developed or evaluated.
Several properties of a processed measurement should be distinguished.
Transportability asks whether an index–state relationship established in one population holds in another. When the statistical properties of the input features differ between the development population and the application population, an index may continue to change systematically while its numerical calibration or state discrimination differs in the new population. This is a property of the relationship between populations rather than a defect in the algorithm, and it is the concept most directly engaged by pediatric application. Two pediatric studies make the point concretely. Fuentes and colleagues compared two commercial indices during the same dynamically changing sevoflurane exposures in the same children and found materially different effect-site equilibration and prediction error [25]. Khosravi and colleagues showed that monitor dynamics themselves bias the pharmacodynamic parameters inferred about a patient unless those dynamics are explicitly separated from the patient’s response [26]. The transformation, in other words, is not neutral: what is inferred about the child depends in part on which device is doing the inferring.
Correlation describes whether two indices vary together; it does not establish numerical agreement. Agreement concerns the differences between paired measurements, including systematic bias and limits of agreement. Interchangeability requires those differences to be acceptable for a defined clinical purpose. Devices can be highly correlated while assigning materially different values. Neither correlation nor agreement alone establishes whether either device identifies the brain state of interest.
Validity asks whether the output corresponds to the construct it is intended to identify—and therefore requires specifying which construct is meant. An index may track anesthetic exposure, distinguish operationally defined clinical states, or contribute to improved management, and evidence for one of these does not substitute for evidence of another. This distinction organizes the evidence assessed in the sections that follow.
Non-proprietary metrics occupy an intermediate position. Descriptors such as spectral edge frequency or suppression ratio are computationally defined and might be assumed to be platform-independent. In practice, the value reported depends on preprocessing choices, filter settings, and implementation details that vary between devices, and a named metric should therefore not be assumed to represent an identical measurement across systems. This point is developed in Section 6 and Section 9.
Processed indices can provide useful information, but interpretation depends on the populations, anesthetic conditions, and reference states in which they have been evaluated.

3. Development of the Anesthetized EEG Across Childhood

The spectral and spatial organization of the anesthetized EEG changes substantially during early development. These changes are summarized in Table 1 and Figure 1. The timing estimates come predominantly from sevoflurane studies and should not be assumed to apply unchanged to other agents. In neonates and former preterm infants, postnatal age alone is insufficient to characterize maturation; gestational age at birth, postmenstrual age at recording, and prematurity status should also be reported. For example, the principal sevoflurane developmental cohort excluded children born before 32 weeks [2].
Chronological age provides an incomplete account of the developmental background of the EEG. Genetic influences interact with preconception, prenatal, and postnatal conditions, including maternal–placental–fetal interactions, to shape brain development and its electrophysiological expression. Neonatal EEG therefore reflects a developmental history that begins before birth [27]. For processed EEG monitoring during anesthesia, this history provides the background against which drug-induced changes in brain state must be interpreted.
Table 1. Developmental evolution of the anesthetized EEG and implications for processed EEG interpretation.
Developmental periods are organizational bands, not discrete biological cutoffs. Cornelissen 2018 [2] explicitly identifies earlier reports on 36 infants [33] and discontinuity in 68 children [34] as analyses of selected subjects from its cohort; Agrawal 2019 [9] is a secondary discontinuity analysis. Markus 2021 [29] and Markus 2026 [31] also share a source cohort. These publications should not be counted as independent replications of the same findings (Table S2).
Neonatal period and earliest infancy. The most granular evidence in this period comes from a study of 62 infants examined under sevoflurane in age bands of under 1 month, 1–2 months, 3–5 months, and 6 months to 2 years [28]. In neonates, four standard quantitative descriptors—spectral edge frequency, burst-suppression ratio, relative beta ratio, and approximate entropy—showed minimal dose dependence. The oldest band showed more classically dose-related behavior, with the shift toward this pattern emerging at roughly 3–5 months. This describes one cohort under one agent across a limited concentration range rather than a universal biological threshold. Complementary evidence comes from 115 infants under one year studied under sevoflurane in bands of 0–3, 4–6, 7–9, and 10–12 months. The youngest group was dominated by relative delta activity, whereas relative alpha and beta rose markedly with age, especially by 4–6 months [29]. The 0–3-month values describe that band as a whole rather than neonates specifically. Amplitude-integrated approaches have separately characterized the depression of neonatal background activity during anesthesia [30]. The implication for scalar interpretation is direct: a transformation depending on frequency distribution, entropy, suppression, or the balance of beta and alpha activity is operating on substantially different inputs during the first months of life. Absence of a conventional dose–response relationship in a constituent feature does not establish absence of anesthetic effect; it indicates that the feature carrying dose information in older children may not be carrying it here.
Figure 1. Approximate developmental milestones under sevoflurane and selected lower age boundaries of trial enrollment and device indications. Theta and alpha activity emerge around 4 months, frontal alpha predominance around 6 months, and frontal alpha coherence around 10 months; early frontal slow-oscillation coherence is subsequently lost [2]. These are approximate observations, not discrete thresholds of monitoring validity. The displayed trial entry ages are 6 months for the density spectral array trial [35], 1 year for Long and Miyasaka [16,17], 3 years for Frelich [36], and 4 years for the multicenter BIS trial [37]. SedLine brain-function monitoring is indicated from 1 year, whereas its pediatric PSi indication begins at 4 years for sevoflurane or propofol [19]. The age axis is expanded during infancy and is not linear. No evidence-volume or accuracy gradient is shown. Figure 1 was rendered with aesthetic formatting assistance from Claude Opus 5 (Anthropic, PBC); content and design were determined by the authors.
Approximately 4 to 12 months: emergence and reorganization of oscillatory structure. A larger analysis from a related research program at Boston Children’s Hospital provides the most detailed developmental picture. Multichannel EEG was recorded during sevoflurane maintenance in 91 children up to approximately 3 years of age, with power and coherence mapped across cortical regions [2]. Six findings are reported. Slow-delta activity was present across all ages. Theta and alpha activity emerged around 4 months. Alpha power increased from approximately 4 to 10 months. Frontal alpha predominance emerged around 6 months. Frontal slow-oscillation coherence was present during approximately the first 6 months and was subsequently lost. Frontal alpha coherence emerged around approximately 10 months and persisted thereafter. The report identifies the earlier 36-infant study [33] and the 68-child discontinuity study [34] as analyses of selected subjects from this cohort; these are overlapping analyses, not independent confirmations.
Two features move in opposite directions across the same interval: an adult-associated frontal alpha anesthetic EEG signature is acquired while an earlier form of slow-oscillation coherence is lost. This creates a plausible basis for non-monotonic age effects in a multi-feature processed index, since a transformation weighting several coherence- or power-derived descriptors could shift. This would then reflect the composition of the maturing signal rather than the patient’s anesthetic state. This study did not test a commercial index, and the mechanism is therefore proposed rather than demonstrated. It still identifies a specific structural reason why age-dependent index behavior in this window might not follow a simple monotonic pattern.
Agent-specific trajectories during and beyond the first year. Two studies bear directly on whether developmental trajectories generalize across anesthetics. Under sevoflurane in a cohort spanning infancy to adulthood, EEG power increased through childhood to approximately age 6 before declining, and the alpha coherence characteristic of older subjects was not observed in the under-one-year subgroup [7]. Under propofol, children older than one year showed slow activity together with coherent frontal alpha organization; frontal alpha was not coherent in the under-one-year subgroup, and total EEG power peaked later, at approximately age 8 [8]. Both studies carry an important caveat: only four participants in each were younger than one year, so conclusions about infant coherence should be treated cautiously. Larger sevoflurane infant datasets from the Boston program provide the more substantial context for that age range. Cross-agent generalization in infancy remains considerably weaker than the citation count might suggest. Broad maturational patterns appear qualitatively similar across the two agents, but their quantitative trajectories differ, with power peaking at different ages. These agent-specific trajectories suggest that developmental calibration may require more than a single drug-independent age correction.
Discontinuity and suppression. Discontinuous activity has been characterized within the same Boston developmental program [34], and subsequently examined in a post-hoc analysis of previously recorded EEG from 65 children in that program [9]. That analysis reported that discontinuity was associated with younger age and with greater anesthetic depth. Age-matched children with discontinuity had received higher induction doses of propofol, and high-frequency power decreased before discontinuity in children four months and older. As a re-analysis of recordings from the same program, it extends rather than independently replicates those observations. A separate prospective cohort from Montefiore/Albert Einstein similarly found that lower sevoflurane-induced posterior alpha power during induction was associated with subsequent EEG discontinuity in infants [38]. Publication relationships among the developmental cohorts are set out in Supplementary Table S2.
Prevalence estimates for isoelectric events are also distinct. A pilot study of children aged 0–37 months reported such events in 32 of 51 patients, approximately 63% [10]. A subsequent multicenter study across 15 centers reported events in 206 of 648 patients (approximately 32%), with site prevalence varying from approximately 9% to 88% [11]. The pilot did not find a significant difference in overall incidence between age groups (p = 0.41) [10]. These estimates therefore should not be read as a simple age-dependent gradient. The marked between-site variation indicates that suppression and isoelectric activity are associated not only with developmental age but also with differences in anesthetic practice and exposure. Suppression therefore cannot be interpreted solely as a developmental marker or as a universal proxy for excessive anesthetic exposure. For processed indices, this matters because the spectral and amplitude characteristics of discontinuity vary with age, so suppression-derived contributions to a scalar index may also require age-sensitive interpretation.
Toddlerhood to later childhood. Development does not conclude at one year. A related analysis from the Charité program examined 147 recordings spanning approximately 1 month to 8 years. It reported that alpha and beta activity became more evident from around 6 months and increased in older age groups [31]. Its youngest group was small, and the analysis depended on proprietary Narcotrend-derived processing with imbalanced age groups. The 2026 report incorporated infant data from the RetroBaby cohort that also underlies Markus et al. 2021 [29]; exact participant-level reuse is not enumerated, so the infant findings should not be treated as an independent replication. Beyond infancy, slow-alpha organization becomes increasingly recognizable, while total power continues to change through childhood with agent-specific peaks at different ages [7,8], and age-sensitive quantitative features remain detectable well into later childhood [32]. The developmental challenge therefore changes with age rather than disappearing abruptly: older-child EEG is increasingly adult-like, but residual age, drug, and transition-state effects remain relevant to numerical calibration.
Development changes the features available to a processed algorithm, and different features follow different trajectories. This provides a physiological reason to test calibration across age groups; it does not, by itself, establish that an index is invalid or that raw EEG interpretation is clinically superior.

4. Age-Dependent Performance of Conventional Processed Indices

The evidence is organized by developmental period and by the question tested: exposure responsiveness, discrimination of an operational state, individual variability, or clinical benefit. Table 2 summarizes selected studies using the classification criteria in Section 1 (Literature Search and Review Approach). Age-stratified analyses are distinguished from broader pediatric studies used to illustrate measurement limitations.
Table 2. Key pediatric studies informing developmental validity of processed EEG monitoring.
The categories describe the reference tested, not proof of successful validation. Exposure validity uses dose or concentration; brain-state validity uses an independently assessed operational physiological or clinical state; clinical validity evaluates a monitoring intervention. Inter-index studies are labeled as correlation or agreement and do not independently validate brain state. Behavioral endpoints, including movement or nonresponse to stimulation, are not direct measures of consciousness. Table 2 is a purposive selection; further age-comparison studies appear in Table S1, and trials of raw-EEG, spectrogram, or density-spectral-array guidance [16,17,35] are compared in Section 7.
Neonates and early infancy. Exposure responsiveness is the dimension most directly examined here. Across several quantitative measures, conventional dose-dependent behavior is weak or atypical in the earliest months. In the age-banded sevoflurane study described earlier, four quantitative descriptors showed little dose dependence below approximately 3 months [28]. Rather than merely performing weakly, non-proprietary metrics behave in atypical ways. In a cohort ranging from 24 days to 14 years of age, spectral edge frequency rose rather than fell during anesthesia in infants under 6 months, with amplitude-integrated EEG offering only marginal additional information [44]. Calibration evidence is sparse, but a large sedation dataset found lower index cutoffs in infants aged 6 months and under when compared against an observational sedation scale [42]. State discrimination has been studied less extensively than exposure responsiveness. A case report of a 3-month-old ex-preterm infant described a Patient State Index of 3–10 with a high suppression ratio during clinically evident movement, illustrating that movement and a low processed index can coexist; movement alone does not identify consciousness or cortical state [51]. A single case, though, establishes only that such divergence is possible. An exploratory secondary analysis of the Yuan multicenter cohort [11] in 42 infants aged 3 months and under achieved approximately 68% internal accuracy against expired sevoflurane categories, which speaks to exposure rather than state [18]. What has been shown in this stratum is therefore narrow. Several conventional EEG features show limited or atypical relationships with anesthetic exposure in the earliest months, and little is established about absolute calibration or state discrimination.
Later infancy. Exposure responsiveness and behavioral discrimination remain less reliable than in older children, although several studies detect age-related differences within infancy. Within the first year, index values differ significantly across infant bands: a comparison of infants aged 3–6 versus 7–12 months found significant differences under steady-state sevoflurane, and the investigators advised caution below 6 months [52]. Exposure responsiveness remains less robust in infancy than in older children. Across multiple steady-state sevoflurane concentrations in 54 children, the 0–1-year group showed the smallest index change of any age band, together with markedly lower pre-awakening values [12]. Comparing 20 infants aged 1 month to 1 year with 40 older children, concentration relationships for spectral entropy and BIS were less clear in infants, while behavioral discrimination varied with induction versus emergence [41]. A five-band study spanning under 1 year to 13 years found that BIS served as a measure of arousal overall but performed better between 1 and 13 years than below 1 year, and attributed the age dependence to the underlying EEG features rather than to the index alone [43].
State responsiveness and inter-index correlation should also be considered separately. Entropy increased after awakening across age groups, demonstrating sensitivity to the state transition, whereas the correlation between Entropy and BIS was weaker in infants [40]. The latter is evidence of weaker correlation between two processed transformations, not independent evidence of poorer state discrimination.
These observations are not limited to a single algorithm. In a retrospective study of 62 children aged 1–36 months, both the Patient State Index and spectral edge frequency were lower, and the burst-suppression ratio was higher, in those under 1 year. The authors concluded that the index did not accurately assess anesthetic state in that group [14]. Recent propofol work in 49 infants aged 3–12 months is informative in a different way. Rather than establishing a calibration, it found that the appropriate spectral edge frequency value depended on the clinical endpoint being predicted [48]. That finding is further developed in Section 6.
Later infancy and toddlerhood: the direction of calibration difference. During later infancy and toddlerhood, the literature does not converge on a single direction of calibration difference. More recent observational studies have emphasized unexpectedly elevated processed-index values in children younger than two years. A prospective study of 50 children aged 4–24 months investigated this phenomenon precisely, finding index values frequently higher than intended during maintenance [15]. In a separate cohort of 111 children, Patient State Index values were significantly higher in children under age 2 than in older children given comparable age-adjusted anesthetic concentrations [47].
A prospective pharmacometric study spanning 1 month to 12 years provides a useful bridge across these age strata. Among 48 children given sevoflurane with fentanyl, BIS, State Entropy, and Response Entropy each predicted better as age increased. Both performance and inter-index correlation were weaker in infants, who trended toward lower intraoperative and awakening values, while toddlers trended higher [45]. The finding is important not because it establishes a universal direction of bias, but because opposite age-associated patterns occurred within the same study design—removing the need to attribute the discrepancy entirely to cohort differences.
State-related discrimination becomes more evident during this period, although performance remains metric- and age-dependent: the expected decrease in spectral edge frequency from awake to anesthetized states was evident only after approximately 2 years of age [44]. Inter-index correlation also varies with age. A study of children aged 6 months to 12 years found fair BIS–Patient State Index correlation in the 6-month-to-2-year and 8-to-12-year groups but no significant correlation in the 2-to-7-year group, with correlation weakening as anesthetic concentration rose [46].
Older children and adolescents. Group-level state discrimination is better characterized in older children than in infants. Simultaneous recording in 50 children aged 3–12 years found that BIS and the Patient State Index distinguished maintenance from recovery similarly, with prediction probabilities of 0.85 and 0.87, respectively [53]. A secondary analysis of 43 children aged 2–15 years from the 51-child spectral edge frequency 95 (SEF95) cohort [54] found the two indices to be strongly but nonlinearly related, with the greatest differences at loss and recovery of consciousness—that is, at the transitions of greatest clinical interest [55]. Index values also correlate with observational sedation scales in school-aged children [56]. However, individual-level precision remains limited even when group performance is good. Rodriguez et al. enrolled 87 children, but the paired BIS–clinical-sign analysis included 59 aged 4 months–14 years and did not demonstrate an age effect [39]. BIS was associated with clinical-sign categories during induction and emergence. Yet individual values overlapped substantially across those categories, and a BIS above 50 had only a 25% positive predictive value for movement at incision [39].
Anesthetic and exposure context. Across the pediatric literature, monitor performance also depends on anesthetic and sedative context. Index correlation with an observational sedation scale was poor when children received ketamine, or chloral hydrate combined with hydroxyzine and meperidine [57], and with ketamine or opioids in Malviya et al.’s pooled secondary analysis of four studies, which included data reported by McDermott and Sadhasivam [42,56,57]. Modeled behavior is also monitor-dependent: two commercial indices recorded during the same dynamically changing sevoflurane exposures in children aged 3–15 years showed materially different effect-site equilibration and prediction error [25]. Propofol equilibration has been shown to vary with age, though that analysis restricted its conclusions to children older than 1 year [58]. Separately, Long et al. observed that children younger than two years required higher end-tidal sevoflurane concentrations regardless of trial allocation [16]. This finding provides pharmacologic context for age-dependent monitoring but does not by itself establish an age-specific bias in the processed index.
Two features of this evidence base should be stated plainly. First, it is dominated by BIS, with substantial spectral entropy coverage and a growing but smaller Patient State Index literature. Evidence for the Cerebral State Index and Narcotrend in young children is too limited to support age-stratified conclusions. Generality claims should be constrained accordingly, although the age-dependent limitations reported for PSi below one year indicate that concerns about developmental calibration are not confined to BIS [14]. Second, age-specific validation evidence remains thinnest in the developmental strata in which processed-index interpretation is most uncertain.
In neonates and young infants, several conventional EEG features show weak or atypical relationships with anesthetic exposure [28,44]. Sedation-scale comparisons indicate lower index cutoffs at 6 months and under [42], but age-specific calibration and state-discrimination evidence remains limited. In later infancy and toddlerhood, indices respond to exposure, but the direction of age-related differences varies across cohorts [12,15,45,47], as does inter-index correlation [40,46]. In older children and adolescents, group-level discrimination between maintenance and recovery is better characterized [53], although individual values overlap [39]. Across ages, performance remains drug-dependent [42,57], and a consistent direction of absolute calibration difference has not been established.

5. Why Universal Numerical Targets Are Problematic

Here, developmental calibration means that the relationship between a displayed index value and a specified physiological or clinical reference remains comparable across developmental groups. It does not mean merely applying an age correction. Statistical calibration, by contrast, concerns agreement between predicted probabilities and observed event frequencies in a prediction model.
An index may track changing anesthetic exposure within a child without establishing that the same value denotes an equivalent state in an infant, toddler, and adolescent. Exposure responsiveness supports trend interpretation; equivalent absolute targets require age-stratified testing against the same defined reference.
The studies summarized in Table S1 differ in age bands, agents, co-administered drugs, exposure range, measurement phase, monitor architecture, and reference standard. Their directional differences do not themselves prove developmental instability, but this heterogeneity prevents a simple inference that universal targets have equivalent physiological meaning.
Reference standards also change the meaning of a result. Within one infant cohort, Yuan et al. reported different SEF95 values for three clinical endpoints [48]. These endpoints and their implications are examined in Section 6.
Clinical benefit addresses a separate question. Trials, including the positive BIS-guided Frelich trial [36], can support a management strategy without isolating the contribution or age equivalence of its numerical target. Section 7 distinguishes the interventions, comparators, and outcomes.
Processed indices can remain useful adjunctive trend monitors when interpreted with age, anesthetic exposure, signal quality, and raw or spectral EEG. The evidence supports this contextual use without establishing universal absolute targets or superiority of raw EEG guidance in infants.

6. Beyond Legacy Indices: Emerging Metrics and Computational Approaches

The problems outlined in Section 3, Section 4 and Section 5 have driven the search for alternatives to legacy composite indices. These alternatives include more mathematically defined or physiologically interpretable features, richer feature sets, and statistical or machine-learning models that can represent nonlinear relationships. This section asks whether such approaches resolve the developmental calibration problem. A recent infant study suggests the answer depends less on method sophistication than on what the method is asked to predict.
In infants aged 3 to 12 months receiving propofol, with 49 enrolled and 44 evaluable, estimated SEF95 values associated with a 50% probability of nonresponse (ED50) differed substantially by endpoint: 21.4 Hz for pacifier placement, 12.6 Hz for standardized electrical stimulation, and 8.5 Hz for laryngoscopy [48]. The same metric also correlated inversely with plasma propofol concentration. Within the same infant population, anesthetic, and EEG metric, the estimated ED50 differed approximately 2.5-fold between the least and most demanding clinical endpoints. The same EEG metric does not acquire one universal interpretation simply because it is mathematically defined.
This observation formalizes what may be termed the ground-truth problem. Candidate reference standards for a pediatric brain-state monitor range widely. These span drug exposure measures such as end-tidal and plasma concentration; behavioral endpoints such as responsiveness, sedation scales, movement, and response to standardized stimulation; and expert EEG phenotype classification, suppression, concordance with another processed index, and downstream clinical outcome. They are not simply noisy variants of a single, interchangeable latent variable. They measure different biological or clinical phenomena, so a metric calibrated against one of them has been calibrated against that phenomenon rather than against anesthetic state in general.
Two questions therefore need to be kept apart. Prediction accuracy asks how well a model predicts its chosen target. Construct validity asks whether that target is the construct the monitor is intended to estimate. A method can perform excellently on the first while leaving the second unaddressed. A computational model inherits the construct represented by its training or validation target; greater algorithmic sophistication does not by itself create a more valid reference standard.
Investigators have also evaluated mathematically defined or more physiologically interpretable quantitative EEG measures, including spectral edge frequency, entropy-based descriptors, suppression-related variables, amplitude-integrated EEG, and wavelet-derived measures. When preprocessing and implementation are specified, such features may offer a more inspectable relationship to the underlying signal than a proprietary composite index. This does not confer developmental validity. Four conventional descriptors showed little dose dependence in neonates [28]. Spectral edge frequency behaved paradoxically in infants under approximately six months, increasing rather than decreasing during anesthesia [44]. Permutation entropy fell with deeper anesthesia and produced a modelled dose–response comparable to a legacy index, but was itself validated against end-tidal concentration rather than against a state [59]. Moreover, the endpoint dependence demonstrated by Yuan et al. [48] involved a mathematically defined spectral metric rather than a proprietary composite index. Transparency of the metric and validity of the construct are separate properties.
A further complication is that a named feature is not necessarily a single measurement. In 51 children with 402 simultaneously paired observations, spectral edge frequency derived from two commercial platforms was substantially correlated but showed nontrivial between-device disagreement, with 95% limits of agreement of approximately −4.08 to +5.32 Hz [54]. Separately, technical work conducted outside pediatric settings has shown that user-selectable display filtering can change how the displayed value is calculated [60]. This amounts to technical background rather than pediatric validation evidence, but it points to a mechanism through which nominally identical metrics can diverge. A mathematically named feature is not necessarily a device-independent measurement unless preprocessing and implementation are comparable—a point that bears directly on any proposal to standardize pediatric monitoring around a common non-proprietary descriptor.
Kumar et al. reanalyzed data from the 15-center Yuan cohort [11], using 4574 one-minute EEG epochs from 42 infants aged 3 months and under to classify four expired-sevoflurane ranges [18]. Best internal accuracy was approximately 68%. Random epoch-level splits can place recordings from the same infant in both training and testing sets, allowing patient-specific signal characteristics to improve apparent performance without demonstrating generalization to a new patient. The four classes contained 262, 1868, 1470, and 974 epochs; the authors reported class-specific balanced accuracy and F1 scores as well as overall accuracy. Performance distributions across repeated splits are not patient-level confidence intervals, and patient-independent external validation was not performed. Thus, the study provides exploratory exposure-prediction evidence rather than brain-state validation.
A recent study of entropic features moves the target closer to a clinical state. In 77 participants aged 1 to 18 years receiving propofol, permutation entropy distinguished propofol-induced unresponsiveness from clinical recovery with approximately 96.6% overall accuracy, with reported toddler performance of approximately 94.7% compared with 88.9% for BIS. The entropic measures themselves changed with age during maintenance [32]. Because the target is an operational behavioral state transition rather than drug exposure, this represents a different kind of claim. Several limitations bound it: no children under one year were included, the study was conducted at a single center, thresholds were evaluated within the same population that generated them, and the transition between unresponsiveness and recovery is not equivalent to connected consciousness. The approach may therefore improve discrimination of the state it was designed to identify while leaving unanswered whether the same feature thresholds transport across development or to other constructs.
A Wavelet Index (WLi) has also been prospectively evaluated in 40 children aged 6 months to 6 years during stepwise sevoflurane administration. WLi tracked some concentration transitions comparably to BIS but showed a plateau between 0.6 and 1.3 age-adjusted minimum alveolar concentration (MAC), despite distinguishing other transitions [49]. Although age was not identified as a significant covariate within this cohort, the study evaluated concentration responsiveness rather than demonstrating developmentally invariant state calibration.
This literature spans three overlapping approaches. The first is compression of the EEG into a broadly applicable composite scalar, as in the legacy indices. The second is adoption of mathematically defined or physiologically interpretable features intended to retain a clearer connection to signal characteristics. The third is combination of multiple features within statistical or machine-learning models capable of representing nonlinear or age-sensitive relationships. These approaches coexist rather than succeed one another, and the progression concerns representation and prediction rather than reference standards. Replacing one universal scalar with another does not address the developmental construct-validity problem unless the calibration itself is developmental.
New features and models may improve prediction, but interpretation remains tied to the training target and population. Patient-level separation, class-specific performance, and uncertainty estimates are needed before internal accuracy is generalized to clinical use.
How accurately a metric predicts its chosen target is nonetheless a different question from whether a monitoring strategy built around it improves care. The clinical literature addresses the second question, and it does so through trials whose interventions and comparators differ considerably.

7. Clinical Outcomes: Does EEG-Informed Anesthetic Management Matter?

Overlapping meta-analyses generally support reduced volatile-anesthetic exposure, while delirium results depend on whether incidence or severity scores are assessed and which interventions are pooled [61,62,63,64]. BIS-specific syntheses did not find clear reductions in delirium scores [61,62]. Haidar et al. included nine trials and 1052 participants overall, but the delirium-incidence estimate came from four trials involving 513 participants: 27% versus 48%, risk ratio 0.56 (95% CI 0.37–0.84; I2 = 55%) [63]. The overall review population is not the denominator for every pooled outcome. This estimate also combines different forms of EEG guidance; it should not be attributed to a single index or target.
The meta-analyses share trials and should not be counted as independent replications. The following trial comparisons distinguish exposure endpoints, delirium incidence, and delirium severity; reported confidence intervals are retained where available.
BIS-guided titration. Frelich et al. [36] randomized 163 children aged 3–8 years to BIS 40–60 or clinician dosing based on MAC and end-tidal concentration. The primary outcome, delirium during postanesthesia care unit (PACU) stay, defined as a Pediatric Anesthesia Emergence Delirium (PAED) score >10, occurred in 11/86 (12.8%) versus 27/77 (35.1%) children (p = 0.001); severity scores at 10 and 30 min were also lower [36]. This is direct positive evidence for the tested BIS-guided strategy. The PONV publication is a companion report of the same trial [65]. In the multicenter Templeton trial, 180 children aged 4–18 years were randomized and 170 analyzed. BIS 45–60 was compared with clinician-guided care with BIS concealed. The primary maintenance-sevoflurane endpoint was lower across all three age strata, with reported mean differences of −0.3 to −0.4 percentage points and 95% confidence intervals excluding zero. Recovery endpoints were prespecified, but the published intention-to-treat analyses were post hoc. Overall recovery profiles were not clearly different, and the lower delirium-incidence finding was exploratory [37].
Raw and spectral EEG guidance. Long et al. [16] randomized 200 children aged 1–6 years to a raw-EEG/spectrogram strategy, with PSi 25–50 where feasible, or standard care with EEG concealed. Primary end-tidal sevoflurane outcomes were lower during induction (mean difference −0.88 percentage points, 95% CI −1.45 to −0.31) and maintenance (−0.15, 95% CI −0.25 to −0.05); PAED scores were similar [16]. Miyasaka et al. [17] compared EEG-pattern titration with a fixed 1.0-MAC strategy in children aged 1 to younger than 6 years. The primary delirium-incidence outcome occurred in 19/91 versus 30/86 children. The reported control-minus-intervention difference was 14 percentage points (interim-adjusted 96.65% CI −0.0019 to 28 percentage points; nominal p = 0.04) [17]. Thus, the observed reduction did not exclude no difference under the adjusted interval.
Density spectral array guidance. In de Heer et al. [35], 96 children aged 6 months–12 years were analyzed after DSA-guided titration or a predefined 2.3% end-tidal sevoflurane regimen with clinical adjustment permitted. The primary outcome, time to operating-room discharge readiness, had medians of 6 and 12 min, respectively (reported difference 95% CI −7 to 0 min; p = 0.041) [35]. This was a recovery-time endpoint, not a delirium-incidence trial.
Comparator design affects interpretation. The predefined concentration regimens in Miyasaka et al. [17] and de Heer et al. [35] may compare access to EEG information and individualized titration simultaneously. Comparisons with actively titrated clinical care address a different practical question.
Combined index and spectral-edge guidance. Kocaturk and Keles [50] randomized 100 children aged 4–10 years to SedLine guidance (PSi 25–50; secondary SEF target 10–15 Hz) or MAC and clinical-sign guidance. Delirium incidence, the primary outcome, was lower on PACU arrival (17% versus 43%), and severity scores were also lower. Maintenance sevoflurane differed by 0.10 percentage points (standard minus EEG group; 95% CI 0.006–0.190). Lower pain scores may complicate delirium interpretation, and no monitored EEG parameter independently predicted delirium [50].
Emergence delirium carries clinical importance and can be measured with reasonable reliability. Yet pain, anesthetic technique and exposure, and recovery conditions all shape it as well. It is also measured after the fact rather than during anesthesia, placing it downstream of the intraoperative brain state a monitor is meant to estimate. The pain difference in the trial above is a concrete illustration. A downstream outcome can demonstrate clinical usefulness of a strategy, but it is not a direct reference standard for intraoperative anesthetic brain state.
One further limitation applies across this literature. Modern positive outcome trials largely enroll from age 1 year, from age 4 years, or across broad mixed pediatric cohorts. The density spectral array trial enrolled from 6 months but did not report powered infant-specific outcome evidence. An older trial remains unusual in including a genuine 0–6-month subgroup, in which lower sevoflurane use did not translate into recovery benefit [6]. Infants are therefore not unstudied, but the clinical-outcome literature is weakest precisely where developmental calibration is most uncertain.
These trials support clinical usefulness of several EEG-informed strategies, including reductions in exposure and, in some settings, delirium. Their interventions, comparators, and endpoints differ, and positive strategy-level results do not establish equivalent numerical targets across development.

8. Current Directions in Pediatric Processed EEG Monitoring

The expansion of pediatric indications represents an important direction for commercial processed EEG monitoring. The July 2026 FDA clearance for SedLine provides a current example: the Patient State Index indication now includes children aged 4 years and older receiving sevoflurane or propofol, while EEG brain-function monitoring is indicated from age 1 year [19]. This development brings pediatric clinical evidence into the application of processed EEG during anesthesia.
The regulatory submission included data from 169 pediatric patients, with PSi evaluated against burst suppression, University of Michigan Sedation Scale scores, and anesthesia stages. The reported findings supported pediatric PSi responsiveness and the expanded indication [19]. For this review, the significance is the move toward defining the populations and anesthetic conditions in which a processed index can support assessment of brain function.
These developments complement the physiological and clinical studies discussed earlier. Regulatory indications, developmental characterization, and clinical-outcome studies address complementary questions. Continued evaluation across ages and anesthetic contexts can further clarify how numerical indices relate to the underlying EEG and contribute to anesthetic management, particularly in infants and early toddlers. The framework in Section 9 is intended to organize this continuing work.

9. What Should the Next Pediatric Brain-State Monitor Be Validated Against?

No single developmentally invariant reference standard can establish all aspects of pediatric pEEG validity. We propose eight complementary domains to organize evaluation (Table 3). The core items make a claimed use interpretable; additional tests strengthen generalizability or clinical usefulness. These are author-proposed priorities, not mandatory regulatory or consensus requirements. External validity concerns performance in new patients and settings; algorithmic validity concerns the design and integrity of model development and evaluation.
Table 3. Proposed multidomain framework for validation of pediatric anesthetic brain-state monitoring.
Core denotes minimum reporting proposed for interpreting the stated claim; additional denotes evidence that strengthens that claim or supports a broader one. These priorities are author-derived, not mandatory regulatory, guideline, or consensus standards.
Developmental validity. The greatest uncertainty in conventional index performance has repeatedly been observed early in development, yet many subsequent pharmacodynamic and outcome studies exclude infants or report only pooled pediatric estimates. Pharmacodynamic work has restricted its conclusions to children older than one year [58]. Recent work on entropic features begins at age one [32], a large multicenter index-guided trial began at age four [37], and a trial enrolling from six months did not report a powered infant-specific treatment effect [35]. Aggregate pediatric performance can therefore obscure clinically important age-specific variation in precisely the developmental subgroup for which a pediatric-specific algorithm would be most valuable. We propose that future algorithms be evaluated within developmental strata defined in advance. Prespecified strata could include neonatal or very early infancy, later infancy, toddlerhood, and older childhood or adolescence, with performance reported separately rather than only as a pooled pediatric estimate. The specific boundaries matter less than the principle, and biologically meaningful strata may not coincide with regulatory age cutoffs.
Drug validity. Most developmental EEG characterization remains sevoflurane-based. Propofol evidence has improved substantially [8,32,48], but dexmedetomidine, ketamine, nitrous oxide, and common multimodal combinations remain thinly represented, and historical work indicates that index performance can differ markedly with ketamine and with opioid co-administration [42,57]. Validation should therefore be agent-specific rather than assuming drug invariance. No monitor could reasonably be expected to be validated across every anesthetic technique. A more practical standard is to require explicit disclosure of the specific agents against which a model has actually been validated. Separate evidence for volatile anesthetics and for propofol should be regarded as a starting point rather than a complete account.
State validity. The endpoint-specific spectral edge frequency findings from Section 6 show why an explicitly defined construct is necessary here. Even within a single infant population, a single anesthetic, and a single metric, the value tied to a 50% probability of nonresponse varied substantially depending on which clinical endpoint was being predicted [48]. Responsiveness, movement, emergence, tolerance of stimulation, profound cortical suppression, and connected consciousness are not interchangeable. A pediatric brain-state algorithm cannot be more valid than the construct against which it is calibrated. Going forward, validation studies should specify, using prospectively defined operational endpoints, exactly which state or transition an algorithm aims to detect. They should also report performance separately for the transitions of clinical interest—loss and recovery of responsiveness, emergence, and profound suppression—rather than presenting a single value as an index of universal anesthetic depth. Connected consciousness, though conceptually important, is not presently a readily implementable reference standard in young children, and framing validation as though it were would obscure rather than resolve the problem.
Physiological validity. A scalar index should also be interpretable in relation to the EEG from which it is derived, particularly when developmental changes alter the spectral and temporal features available to the algorithm. Concordance between a processed value and the underlying EEG is distinct from concordance with a behavioral endpoint. For example, a case report of very low PSi with a high suppression ratio despite clinically evident movement shows that movement and a low processed index can coexist [51]. Movement alone does not identify consciousness or cortical state, and this report does not, by itself, establish disagreement between the index and the EEG. We propose that concurrent raw EEG and spectral information be retained during validation wherever feasible, so that algorithm output can be examined alongside the physiological signal from which it is derived. Expert raw-EEG interpretation does not constitute a definitive ground truth and itself requires age-appropriate reference information. It can nevertheless provide an additional validation layer, particularly when scalar values appear unexpected in their clinical context.
Technical validity. A displayed value reflects the entire acquisition and processing pipeline, not neurophysiology alone. Montage effects in small children have been noted without being resolved [44]. The between-device disagreement in spectral edge frequency described in Section 6 is one reason for caution. Together with evidence that display settings can alter how a derived value is computed, it indicates that a named metric should not be assumed to represent an identical measurement across platforms [54,60]. Future studies would strengthen their interpretability by reporting signal quality, artifact handling, EMG behavior, electrode and montage constraints, usable recording time, and the computation of derived metrics. When a metric is meant to be platform-independent, reproducibility across devices should also be demonstrated.
External validity. Recent computational work remains predominantly single-center and internally evaluated. Infant-specific machine-learning models have been developed using data partitions at the EEG-epoch rather than the patient level [18], and entropic feature thresholds have been evaluated within the same population that generated them [32]. Epoch-level performance may overestimate patient-level generalizability when correlated epochs from the same individual contribute to both model development and evaluation. A stronger development pathway would separate model development, locked validation, and independent patient-level external testing. Multicenter and developer-independent evaluation should be understood as important evidence of generalizability rather than as a regulatory threshold.
Clinical validity. Benefit has been observed with several different monitoring strategies, sometimes against actively titrated care and sometimes against a predefined concentration; in one trial, benefit occurred without any monitored parameter independently predicting the outcome. Clinical validity asks whether use of the monitoring strategy improves clinically relevant management or outcomes. Where feasible, a titrated comparator may help isolate the incremental contribution of the monitoring strategy, since comparison against a fixed anesthetic concentration may partly evaluate titration itself. Relevant outcomes may include anesthetic exposure, suppression, recovery, emergence quality, and safety; no single endpoint, including emergence delirium, is uniquely appropriate across settings.
Algorithmic validity. The evaluation should prevent information leakage and match the analysis to the intended claim. Core reporting includes the model version, reference labels, feature selection and tuning procedures, patient-level data separation, class distribution, and performance with uncertainty. A locked model and prespecified analysis strengthen prospective testing. Statistical calibration is relevant when probabilities are predicted; discrimination alone is insufficient. Comparisons with another proprietary index [46,53,55] can assess inter-index association but cannot serve as the sole reference for brain-state validity. Whether performance transfers to an independent population is addressed separately under external validity.
The domains are complementary rather than independent tests. A study need not evaluate every possible drug, population, or clinical outcome, but its claim should remain within the conditions actually tested. Infants and early toddlers remain the most important unresolved population.
Pediatric-specific calibration would represent an important advance. Pediatric validity, however, should ultimately describe more than the age range of the cohort in which an algorithm was developed or evaluated. It should describe the developmental stage, anesthetic agent, state being predicted, physiological signal, technical conditions, external population, and clinical context in which an output has been shown to retain meaning. Framed this way, the question is not whether processed EEG works in children, but how precisely the conditions of its interpretability can be specified.

10. Conclusions

Uncertainty about the interpretation of absolute pEEG values is greatest in infants and young children. Developmental changes in spectral power and coherence [2], discontinuity and suppression [9,10,11], and entropy-related measures [28,32] mean that the same numerical target cannot be assumed to represent an equivalent state across ages. This does not establish that pEEG lacks clinical value.
Processed indices should be interpreted in relation to developmental age, anesthetic agent, co-administered medication, and the state or endpoint being assessed. Evidence that an index tracks concentration or that a monitoring strategy improves outcomes supports those specific uses; it does not independently validate a universal numerical target.
In practice, numerical trends should be considered alongside raw EEG, spectral information, signal quality, and the clinical context. Future validation should define its population and reference state, test performance in independent patients, and report the conditions under which the output is useful. The proposed framework provides a structure for that evaluation without implying that one monitor or target is appropriate throughout childhood.

Supplementary Materials

The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/children13121665/s1: Table S1, Methodological heterogeneity across selected pediatric processed-EEG studies; Table S2, Cohort lineage and publication relationships. References [65,66] are also cited in the Supplementary Materials.

Author Contributions

Conceptualization, V.C.; methodology and literature search, V.C., F.K., H.A.B. and S.B.; evidence verification, V.C., F.K., H.A.B. and S.B.; writing—original draft preparation, V.C.; writing—review and editing, V.C., F.K. and H.A.B. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Data Availability Statement

No new data were created or analyzed in this study. Data sharing is not applicable to this article.

Acknowledgments

During the preparation of this manuscript, the authors used Claude Opus 5 (Anthropic, PBC, San Francisco, CA, USA) to assist with language polishing and aesthetic formatting of Figure 1. The authors independently designed the figures, verified all scientific claims, and take full responsibility for the final manuscript.

Conflicts of Interest

The authors declare no conflicts of interest. Specifically, the authors have no financial, consulting, advisory, or speaking relationship with any manufacturer of depth-of-anesthesia or electroencephalographic monitoring devices, and no such entity had any role in the conception, writing, or submission of this review.

References

  1. Purdon, P.L.; Pavone, K.J.; Akeju, O.; Smith, A.C.; Sampson, A.L.; Lee, J.; Zhou, D.W.; Solt, K.; Brown, E.N. The Ageing Brain: Age-dependent changes in the electroencephalogram during propofol and sevoflurane general anaesthesia. Br. J. Anaesth. 2015, 115, i46–i57. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  2. Cornelissen, L.; Kim, S.E.; Lee, J.M.; Brown, E.N.; Purdon, P.L.; Berde, C.B. Electroencephalographic markers of brain development during sevoflurane anaesthesia in children up to 3 years old. Br. J. Anaesth. 2018, 120, 1274–1286. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  3. Kearse, L.A., Jr.; Rosow, C.; Zaslavsky, A.; Connors, P.; Dershwitz, M.; Denman, W. Bispectral analysis of the electroencephalogram predicts conscious processing of information during propofol sedation and hypnosis. Anesthesiology 1998, 88, 25–34. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  4. Chen, X.; Tang, J.; White, P.F.; Wender, R.H.; Ma, H.; Sloninsky, A.; Kariger, R. A comparison of patient state index and bispectral index values during the perioperative period. Anesth. Analg. 2002, 95, 1669–1674. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  5. Denman, W.T.; Swanson, E.L.; Rosow, D.; Ezbicki, K.; Connors, P.D.; Rosow, C.E. Pediatric evaluation of the bispectral index (BIS) monitor and correlation of BIS with end-tidal sevoflurane concentration in infants and children. Anesth. Analg. 2000, 90, 872–877. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  6. Bannister, C.F.; Brosius, K.K.; Sigl, J.C.; Meyer, B.J.; Sebel, P.S. The effect of bispectral index monitoring on anesthetic use and recovery in children anesthetized with sevoflurane in nitrous oxide. Anesth. Analg. 2001, 92, 877–881. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  7. Akeju, O.; Pavone, K.J.; Thum, J.A.; Firth, P.G.; Westover, M.B.; Puglia, M.; Shank, E.S.; Brown, E.N.; Purdon, P.L. Age-dependency of sevoflurane-induced electroencephalogram dynamics in children. Br. J. Anaesth. 2015, 115, i66–i76. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  8. Lee, J.M.; Akeju, O.; Terzakis, K.; Pavone, K.J.; Deng, H.; Houle, T.T.; Firth, P.G.; Shank, E.S.; Brown, E.N.; Purdon, P.L. A prospective study of age-dependent changes in propofol-induced electroencephalogram oscillations in children. Anesthesiology 2017, 127, 293–306. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  9. Agrawal, U.; Berde, C.B.; Cornelissen, L. Electroencephalographic features of discontinuous activity in anesthetized infants and children. PLoS ONE 2019, 14, e0223324. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  10. Yuan, I.; Landis, W.P.; Topjian, A.A.; Abend, N.S.; Lang, S.S.; Huh, J.W.; Kirschen, M.P.; Mensinger, J.L.; Zhang, B.; Kurth, C.D. Prevalence of isoelectric electroencephalography events in infants and young children undergoing general anesthesia. Anesth. Analg. 2020, 130, 462–471. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  11. Yuan, I.; Xu, T.; Skowno, J.; Zhang, B.; Davidson, A.; von Ungern-Sternberg, B.S.; Sommerfield, D.; Zhang, J.; Song, X.; Zhang, M.; et al. Isoelectric electroencephalography in infants and toddlers during anesthesia for surgery: An international observational study. Anesthesiology 2022, 137, 187–200. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  12. Davidson, A.J.; Huang, G.H.; Rebmann, C.S.; Ellery, C. Performance of entropy and Bispectral Index as measures of anaesthesia effect in children of different ages. Br. J. Anaesth. 2005, 95, 674–679. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  13. Kim, H.S.; Oh, A.Y.; Kim, C.S.; Kim, S.D.; Seo, K.S.; Kim, J.H. Correlation of bispectral index with end-tidal sevoflurane concentration and age in infants and children. Br. J. Anaesth. 2005, 95, 362–366. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  14. Gao, Z.; Zhang, J.; Zhang, X.; Wang, L.; Huang, Y.; Yu, J. A retrospective study of the patient state index during general anesthesia in infants and young children. Clin. Pediatr. 2024, 63, 249–256. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  15. Yoon, S.B.; Park, J.B.; Kang, P.; Jang, Y.E.; Kim, E.H.; Lee, J.H.; Lee, H.C.; Kim, J.T.; Kim, H.S.; Ji, S.H. Electroencephalography and anesthetic depth in children under 2 years of age: A prospective observational study. Pediatr. Anesth. 2025, 35, 294–301. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  16. Long, M.H.Y.; Lim, E.H.L.; Balanza, G.A.; Allen, J.C., Jr.; Purdon, P.L.; Bong, C.L. Sevoflurane requirements during electroencephalogram (EEG)-guided vs standard anesthesia care in children: A randomized controlled trial. J. Clin. Anesth. 2022, 81, 110913. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  17. Miyasaka, K.W.; Suzuki, Y.; Brown, E.N.; Nagasaka, Y. EEG-guided titration of sevoflurane and pediatric anesthesia emergence delirium: A randomized clinical trial. JAMA Pediatr. 2025, 179, 704–712. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  18. Kumar, R.; Skowno, J.; von Ungern-Sternberg, B.S.; Davidson, A.; Xu, T.; Zhang, J.; Song, X.; Zhang, M.; Zhao, P.; Liu, H.; et al. Quantitative electroencephalogram and machine learning to predict expired sevoflurane concentration in infants. J. Clin. Monit. Comput. 2025, 39, 999–1014. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  19. U.S. Food and Drug Administration. 510(k) Premarket Notification K260880: SedLine Sedation Monitor; Decision date 10 July 2026; determination: Substantially equivalent; U.S. Food and Drug Administration: Silver Spring, MD, USA, 2026. Available online: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfPMN/pmn.cfm?ID=K260880 (accessed on 6 September 2026).
  20. Bong, C.L.; Balanza, G.A.; Khoo, C.E.H.; Tan, J.S.K.; Desel, T.; Purdon, P.L. A narrative review illustrating the clinical utility of electroencephalogram-guided anesthesia care in children. Anesth. Analg. 2023, 137, 108–123. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  21. de Heer, I.J.; Weber, F. Electroencephalographic Density Spectral Array monitoring in pediatric anesthesia: Clinical background and practical applications. Minerva Anestesiol. 2023, 89, 351–358. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  22. Yuan, I.; Bong, C.L.; Chao, J.Y. Intraoperative pediatric electroencephalography monitoring: An updated review. Korean J. Anesthesiol. 2024, 77, 289–305. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  23. Bong, C.L.; Yuan, I. The utility of electroencephalography in guiding general anesthesia in children. Anesth. Analg. 2026, 142, 1155–1168. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  24. Lee, H.C.; Ryu, H.G.; Park, Y.; Yoon, S.B.; Yang, S.M.; Oh, H.W.; Jung, C.W. Data driven investigation of bispectral index algorithm. Sci. Rep. 2019, 9, 13769. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  25. Fuentes, R.; Cortínez, L.I.; Struys, M.M.; Delfino, A.; Muñoz, H. The dynamic relationship between end-tidal sevoflurane concentrations, bispectral index, and cerebral state index in children. Anesth. Analg. 2008, 107, 1573–1578. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  26. Khosravi, S.; Hahn, J.O.; Dumont, G.A.; Ansermino, J.M. A monitor-decoupled pharmacodynamic model of propofol in children using state entropy as clinical endpoint. IEEE Trans. Biomed. Eng. 2012, 59, 736–743. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  27. Scher, M.S. The neural exposome influences the preterm fetal-to-neonatal connectome. Pediatr. Res. 2024, 95, 9–11. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  28. Hayashi, K.; Shigemi, K.; Sawa, T. Neonatal electroencephalography shows low sensitivity to anesthesia. Neurosci. Lett. 2012, 517, 87–91. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  29. Markus, M.; Nagelsmann, H.; Schneider, M.; Rupp, L.; Spies, C.; Koch, S. Peri- and intraoperative EEG signatures in newborns and infants. Clin. Neurophysiol. 2021, 132, 2959–2964. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  30. Stolwijk, L.J.; Weeke, L.C.; de Vries, L.S.; van Herwaarden, M.Y.A.; van der Zee, D.C.; van der Werff, D.B.M.; Benders, M.J.N.L.; Toet, M.; Lemmers, P.M.A. Effect of general anesthesia on neonatal aEEG—A cohort study of patients with non-cardiac congenital anomalies. PLoS ONE 2017, 12, e0183581. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  31. Markus, M.; Panagiotou, F.; Spies, C.; Koch, S. EEG dynamics in children before, during and after general anesthesia. Pediatr. Anesth. 2026, 36, 641–649. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  32. Zhang, Y.; Liang, Z.H.; Wang, X.; Zhang, N.; Zhu, H.T.; Wang, D.X.; Guo, X.Y.; Li, X.L.; Song, L.L. Age-dependent entropic features during propofol anesthesia in developing brain. Anesth. Analg. 2026, 142, 769–781. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  33. Cornelissen, L.; Kim, S.E.; Purdon, P.L.; Brown, E.N.; Berde, C.B. Age-dependent electroencephalogram (EEG) patterns during sevoflurane general anesthesia in infants. eLife 2015, 4, e06513. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  34. Cornelissen, L.; Bergin, A.M.; Lobo, K.; Donado, C.; Soul, J.S.; Berde, C.B. Electroencephalographic discontinuity during sevoflurane anesthesia in infants and children. Pediatr. Anesth. 2017, 27, 251–262. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  35. de Heer, I.J.; Raab, H.A.C.; de Vries, J.; Karaöz-Bulut, G.; Weber, F. The influence of electroencephalographic density spectral array guidance of sevoflurane administration on recovery from general anesthesia in children: A randomized controlled trial. Pediatr. Anesth. 2025, 35, 287–293. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  36. Frelich, M.; Lečbychová, K.; Vodička, V.; Ekrtová, T.; Sklienka, P.; Jor, O.; Straková, H.; Bílená, M.; Formánek, M.; Burša, F. Effect of BIS-guided anesthesia on emergence delirium following general anesthesia in children: A prospective randomized controlled trial. Anaesth. Crit. Care Pain Med. 2024, 43, 101318. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  37. Templeton, T.W.; Alex, G.; Eloy, J.D.; Stollings, L.; Ing, R.J.; Cheon, E.C.; Belani, K.; Breskin, I.; Sebel, P.S.; Taicher, B.M.; et al. BIS guided titration of sevoflurane in pediatric patients undergoing elective surgery: A randomized controlled trial. Pediatr. Anesth. 2025, 35, 277–286. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  38. Chao, J.Y.; Gutiérrez, R.; Legatt, A.D.; Yozawitz, E.G.; Lo, Y.; Adams, D.C.; Delphin, E.S.; Shinnar, S.; Purdon, P.L. Decreased electroencephalographic alpha power during anesthesia induction is associated with EEG discontinuity in human infants. Anesth. Analg. 2022, 135, 1207–1216. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  39. Rodriguez, R.A.; Hall, L.E.; Duggan, S.; Splinter, W.M. The bispectral index does not correlate with clinical signs of inhalational anesthesia during sevoflurane induction and arousal in children. Can. J. Anaesth. 2004, 51, 472–480. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  40. Davidson, A.J.; Kim, M.J.; Sangolt, G.K. Entropy and bispectral index during anaesthesia in children. Anaesth. Intensive Care 2004, 32, 485–493. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  41. Klockars, J.G.M.; Hiller, A.; Ranta, S.; Talja, P.; van Gils, M.J.; Taivainen, T. Spectral entropy as a measure of hypnosis in children. Anesthesiology 2006, 104, 708–717. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  42. Malviya, S.; Voepel-Lewis, T.; Tait, A.R.; Watcha, M.F.; Sadhasivam, S.; Friesen, R.H. Effect of age and sedative agent on the accuracy of bispectral index in detecting depth of sedation in children. Pediatrics 2007, 120, e461–e470. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  43. Jeleazcov, C.; Schmidt, J.; Schmitz, B.; Becke, K.; Albrecht, S. EEG variables as measures of arousal during propofol anaesthesia for general surgery in children: Rational selection and age dependence. Br. J. Anaesth. 2007, 99, 845–854. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  44. McKeever, S.; Johnston, L.; Davidson, A.J. An observational study exploring amplitude-integrated electroencephalogram and spectral edge frequency during paediatric anaesthesia. Anaesth. Intensive Care 2012, 40, 275–284. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  45. Sciusco, A.; Standing, J.F.; Sheng, Y.; Raimondo, P.; Cinnella, G.; Dambrosio, M. Effect of age on the performance of bispectral and entropy indices during sevoflurane pediatric anesthesia: A pharmacometric study. Pediatr. Anesth. 2017, 27, 399–408. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  46. Kim, D.; Kim, J.; Kim, I.; Gil, N.S.; Shin, Y.H.; Jeong, J.S. Correlation between bispectral index and patient state index in children under sevoflurane anesthesia. Pediatr. Anesth. 2022, 32, 740–746. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  47. Ricci, Z.; Robino, C.; Rufini, P.; Cumbo, S.; Cavallini, S.; Gobbi, L.; Brocchi, A.; Serio, P.; Romagnoli, S. Monitoring anesthesia depth with patient state index during pediatric surgery. Pediatr. Anesth. 2023, 33, 855–861. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  48. Yuan, I.; Garcia-Marcinkiewicz, A.G.; Zhang, B.; Ulrich, A.M.; Georgostathi, G.; Missett, R.M.; Lang, S.S.; Bruton, J.L.; Kurth, C.D. Electroencephalographic indices for clinical endpoints during propofol anesthesia in infants: An early-phase propofol biomarker-finding study. Anesthesiology 2024, 141, 353–364. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  49. Jiang, J.; Dan, Y.; Huang, Y.; Zheng, J.; Zhang, K. Prediction of EEG-derived wavelet index for end-tidal sevoflurane in pediatric anesthesia: A prospective, single-blind study. Pediatr. Anesth. 2025, 35, 657–663. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  50. Kocaturk, O.; Keles, S. EEG-guided sevoflurane anesthesia and emergence delirium in children: A randomized controlled trial. Pediatr. Anesth. 2026, 36, 1429–1436. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  51. Nagy, M.; Mahmoud, A.H.M.; AlDhuwaihy, A.; Alessa, M.E.; AlOsaimi, A.S.; Alshammeri, M.; Mahmoud, A. Intraoperative movement despite low SedLine Patient State Index in a preterm neonate: Evidence of algorithm limitation in neonatal EEG monitoring—A case report. Int. J. Med. Dev. Ctries. 2025, 9, 3428–3431. [Google Scholar] [CrossRef] [Scilit]
  52. Kawaraguchi, Y.; Fukumitsu, K.; Kinouchi, K.; Hirao, O.; Miyamoto, Y.; Taniguchi, A. Bispectral index (BIS) in infants anesthetized with sevoflurane in nitrous oxide and oxygen. Masui 2003, 52, 389–393. [Google Scholar] [PubMed]
  53. Jang, Y.E.; Kim, E.H.; Lee, J.H.; Kim, J.T.; Kim, H.S. Usefulness of bispectral index and patient state index during sevoflurane anesthesia in children: A prospective observational study. Medicine 2022, 101, e29925. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  54. Ricci, Z.; Gobbi, L.; La Rosa, E.; Filippini, E.; Lui, M.; Fischer, M.; Colosimo, D.; Romagnoli, S. Analysis of SEF95 derived from two processed EEG devices during pediatric anesthesia for non-cardiac surgery. J. Clin. Monit. Comput. 2026, 40, 317–322. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  55. Ricci, Z.; Colosimo, D. Analysis of BIS and patient state index in children undergoing general anesthesia. Pediatr. Anesth. 2026, 36, 104–106. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  56. Sadhasivam, S.; Ganesh, A.; Robison, A.; Kaye, R.; Watcha, M.F. Validation of the bispectral index monitor for measuring the depth of sedation in children. Anesth. Analg. 2006, 102, 383–388. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  57. McDermott, N.B.; VanSickle, T.; Motas, D.; Friesen, R.H. Validation of the bispectral index monitor during conscious and deep sedation in children. Anesth. Analg. 2003, 97, 39–43. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  58. Jeleazcov, C.; Ihmsen, H.; Schmidt, J.; Ammon, C.; Schwilden, H.; Schüttler, J.; Fechner, J. Pharmacodynamic modelling of the bispectral index response to propofol-based anaesthesia during general surgery in children. Br. J. Anaesth. 2008, 100, 509–516. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  59. Kim, P.J.; Kim, H.G.; Noh, G.J.; Koo, Y.S.; Shin, T.J. Usefulness of permutation entropy as an anesthetic depth indicator in children. J. Pharmacokinet. Pharmacodyn. 2015, 42, 123–134. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  60. Hight, D.; Kaiser, H.A.; Schuller, P.J. A display filter alters SEF95 calculation in the BIS monitor: A persistent design flaw with clinical and research implications. Anesthesiology 2026, 144, 234–236. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  61. de Oliveira, D.A.C.; Dos Santos Monteiro, J.; de Macêdo, N.R.; Tavares Costa Santos, A.H.; Rosa Neves, S.M. Bispectral index-guided anesthesia in children: A systematic review and meta-analysis. Anaesth. Crit. Care Pain Med. 2025, 44, 101607. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  62. Derylo, K.; Sannasi, N.; Rosamystica, M.; Pilia, E. Efficacy of bispectral index-guided sevoflurane administration in pediatric patients undergoing general anesthesia: An up-to-date systematic review and meta-analysis. Pediatr. Anesth. 2026, 36, 164–172. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  63. Haidar, L.; Abadi, F.; Sarieddine, T.; Vitali, A.; Zouein, F.A.; Massoud, G. EEG-guided anesthesia for the prevention of emergence delirium in children: A systematic review and meta-analysis. JAMA Pediatr. 2026, 180, 394–401. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  64. Javed, J.; Bacha, Z.; Brohi, F.; Sikandar, M.; Shah, A.; Khan, S.; Kakakhel, M.Z.J.; Shahzad, M.; Al-Badri, S.G.; Ullah, S.H.; et al. EEG-guided sevoflurane anesthesia vs. standard care in pediatric emergence delirium: A GRADE-assessed systematic review and meta-analysis with trial sequential analysis. J. Anesth. 2026, 40, 422–437, Correction in J. Anesth. 2026, 40, 869. https://doi.org/10.1007/s00540-026-03828-3. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  65. Frelich, M.; Sklienka, P.; Romanová, T.; Němcová, S.; Bílená, M.; Straková, H.; Lečbychová, K.; Jor, O.; Formánek, M.; Burša, F. The effect of BIS-guided anaesthesia on the incidence of postoperative nausea and vomiting in children: A prospective randomized double-blind study. BMC Anesthesiol. 2024, 24, 228. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  66. Bong, C.L.; Long, M.H.Y. Sevoflurane requirements during electroencephalogram (EEG)-guided vs standard anesthesia care in children: A randomized controlled trial. J. Clin. Anesth. 2023, 86, 111071. [Google Scholar] [CrossRef] [Scilit] [PubMed]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Article Metrics

Citations

Article Access Statistics

Article metric data becomes available approximately 24 hours after publication online.