Towards Reconstructable CF-LIBS Quantification: A Perspective on Reporting, Validation, and Reproducibility

,
,
,
,
,
and
1
Heyrovský Institute of the Czech Academy of Sciences, Dolejškova 3, 182 00 Prague, Czech Republic
2
Central European Institute of Technology, Brno University of Technology, Purkyňova 656/123, 612 00 Brno, Czech Republic
3
Institute of Physical Engineering, Faculty of Mechanical Engineering, Brno University of Technology, Technická 2, 616 69 Brno, Czech Republic
*
Author to whom correspondence should be addressed.
Chemosensors2026, 14(12), 262;https://doi.org/10.3390/chemosensors14120262 
(registering DOI)
This article belongs to the Section Analytical Methods, Instrumentation and Miniaturization

Abstract

Calibration free laser induced breakdown spectroscopy (CF-LIBS) estimates elemental composition using a plasma model, offering a practical approach where suitable reference standards are unavailable. This Perspective proposes reconstructability as a reporting objective, distinct from physical validity and fitness for purpose. Five checkpoints organize the calculation, illustrated through constructed examples and selected calculations from three published records. The constructed generator and reconstruction share a model; their agreement tests the reporting record, not physical validity. Under the specified ternary-system and uncertainty assumptions, six of nine convention slips move the light minor constituent outside its propagated 95% interval, and three indicators flag seven of nine slips. These counts are case specific. The two undetected linewidth slips are examined in 43 gated xenon–hydrogen spectra using hydrogen lines as a separate density diagnostic. For two bins with median instrumental-to-observed width ratios of 0.50 and 0.58, retaining the instrumental width gives densities 36–61% above the hydrogen-derived estimates and quadrature subtraction by 16–33%, consistent with the calculated effect. Four pulse-energy cases for one regolith simulant demonstrate reconstruction from measured spectra and comparison with X-ray fluorescence (XRF) on a common elemental mass-ratio basis. We distinguish core from recommended reporting items and supply a copyable template. Three analysts recover specified outputs using the supplied code; broader usability and interlaboratory performance remain to be evaluated.

Article Metrics

Citations

Article Access Statistics

Article metric data becomes available approximately 24 hours after publication online.