Rapid and Point-of-Care Detection of Alpha-Fetoprotein for Hepatocellular Carcinoma Surveillance and Management

,
,
and
1
Department of Hematology and Oncology, Karamay Central Hospital, Xinjiang Second Medical College, Karamay 834000, China
2
School of Biological Science and Medical Engineering, Southeast University, Nanjing 211189, China
*
Authors to whom correspondence should be addressed.

Abstract

Alpha-fetoprotein (AFP) measurement complements ultrasonography in hepatocellular carcinoma (HCC) surveillance and is also used for post-treatment follow-up. Current methods provide quantitative results but generally require serum or plasma separation and automated analyzers. Rapid and point-of-care testing (POCT) may reduce dependence on central laboratories and provide results during the same clinical visit. However, most reported rapid assays measure total AFP alone and focus on lowering the limit of detection, while broad-range quantification, complete sample-to-result procedures, AFP-L3% calculation, and separate PIVKA-II measurement are neglected. This review summarizes the clinical and technical requirements for rapid AFP testing and discusses recent lateral-flow, microfluidic, electrochemical, and optical methods. These methods are compared in terms of specimen type, reported assay time, quantitative range, readout, and biomarker output. Future development should focus on traceable, broad-range AFP quantification directly from whole blood and on longitudinal measurement of total AFP, AFP-L3%, and PIVKA-II to support HCC surveillance.

1. Introduction

Hepatocellular carcinoma (HCC) is the most common type of primary liver cancer. Early detection directly affects whether patients are eligible for potentially curative treatments, including surgical resection, liver transplantation, and local ablation [1,2]. Regular surveillance of high-risk populations, such as patients with cirrhosis or chronic hepatitis B virus infection, is therefore essential for improving early-stage HCC detection [3,4]. Ultrasonography is the recommended first-line method for HCC surveillance, but its sensitivity is influenced by lesion size, the nodular background of cirrhotic liver, patient obesity, and operator experience [5]. Serum alpha-fetoprotein (AFP) measurement is relatively simple and inexpensive, while serial measurements allow changes in AFP levels to be monitored over time. Therefore, AFP is widely used as an adjunct to ultrasonography [6]. A meta-analysis involving 13,367 patients with cirrhosis showed that the combination of AFP and ultrasonography increased the sensitivity for early-stage HCC detection from 45% to 63% [5]. Current American Association for the Study of Liver Diseases (AASLD) guidelines recommend semiannual surveillance with ultrasonography and serum AFP in eligible high-risk populations [2]. For patients undergoing HCC surveillance, AASLD recommends diagnostic evaluation with multiphasic contrast-enhanced computed tomography (CT) or magnetic resonance imaging (MRI) when AFP is ≥20 ng/mL or is rising [2]. However, AFP may also increase in nonmalignant liver diseases, including chronic hepatitis, cirrhosis, and hepatic regeneration, while some patients with HCC have persistently low AFP levels [5,6]. Thus, AFP is not used independently to confirm HCC. Instead, its clinical value lies in improving surveillance sensitivity when combined with ultrasonography, identifying patients who require diagnostic imaging, and monitoring treatment response and recurrence [6,7,8]. These clinical uses establish the need for repeated and reliable quantitative AFP measurement.
Clinical laboratories mainly measure total AFP using antibody-based immunoassays, including enzyme-linked immunosorbent assays (ELISA) [9], chemiluminescence immunoassays (CLIA) [10], electrochemiluminescence immunoassays (ECLIA) [11], and fluorescence immunoassays (FIA) [12,13]. Automated CLIA and ECLIA are widely used because they provide reliable quantification, broad detection ranges, high throughput, and established quality-control procedures [14,15]. However, routine laboratory testing generally requires serum or plasma separation and large automated analyzers in central laboratories. Rapid AFP assays and point-of-care testing (POCT) may provide quantitative results during the same clinical visit and reduce dependence on central laboratories [16]. To provide a practical alternative to central-laboratory testing, rapid assays must simplify sample processing and shorten the complete sample-to-result workflow without compromising quantitative performance.
Workflow improvement alone is insufficient if the assay does not provide clinically interpretable biomarker results. Typically, conventional immunoassays report total AFP and do not distinguish its glycoforms, which are variants of the same protein that differ in their attached glycans. AFP-L3 is the core-fucosylated AFP glycoform that reacts with Lens culinaris agglutinin and is more closely associated with HCC and aggressive tumor features [17,18,19,20,21]. Even at similar total AFP concentrations, patients can differ in the proportion present as AFP-L3. Expressing AFP-L3 as a percentage of total AFP (AFP-L3%) captures this difference and may therefore aid HCC risk stratification [22,23,24]. In addition, multiplex assays that separately report total AFP, AFP-L3, and protein induced by vitamin K absence or antagonist-II (PIVKA-II), also known as des-γ-carboxy prothrombin, may provide additional information beyond total AFP alone [20,21,25]. Therefore, the value of AFP POCT depends not only on rapid total AFP measurement, but also on whether clinically interpretable biomarker results can be obtained with limited sample processing and manual operation.
Until now, several methods have been developed for rapid and point-of-care AFP testing, including lateral-flow and vertical-flow immunoassays [26,27,28], microfluidic immunoassays [29,30], electrochemical sensors [31], surface-enhanced Raman scattering (SERS) assays [32], aptamer-based sensors [33], and optical-fiber sensors [34,35]. Existing reviews have mainly organized these methods according to sensing principles, signal-transduction mechanisms, recognition elements, or functional materials, with analytical performance often compared primarily by the limit of detection (LOD) [36,37,38]. However, many reported assays remain distant from practical clinical use because they depend on serum or plasma separation, involve multiple manual steps, or have limited quantitative ranges. Some clinical applications also require glycoform-specific or multiplex results. From the perspective of HCC surveillance and follow-up, these limitations indicate four important directions for AFP POCT: calculation of AFP-L3% [39], separate measurement of complementary biomarkers such as PIVKA-II [40], direct analysis of unprocessed whole blood [41,42], and greater automation of the complete sample-to-result workflow [43,44,45]. Figure 1 illustrates how four representative assays capture AFP and generate a signal. To the best of our knowledge, a focused review that evaluates recent AFP detection methods against these clinical and technical requirements is still lacking.
Figure 1. Operating principles of four representative assays for AFP detection. (a) Lateral-flow immunoassay. AFP binds to a detection antibody labeled with a gold nanotriangle. Capture of the complex at the test line (T) produces a visible band. C denotes the control line. (b) Microfluidic particle assay. AFP bridges magnetic and polystyrene microparticles. Magnetic separation retains the complexes, while free polystyrene particles accumulate at a downstream trap. The accumulation length decreases as the AFP concentration increases. (c) Electrochemical aptasensor. AFP binding to an electrode-bound aptamer decreases the peak current of the toluidine blue (TB) reporter. (d) Optical-fiber immunoassay. Sequential binding of AFP and a detection antibody at the fiber surface shifts the resonance wavelength. Orange spheres represent AFP.
In this review, we examine rapid and point-of-care detection of AFP for HCC surveillance and management. We searched Web of Science for publications on rapid and point-of-care AFP testing. We selected original studies to illustrate different approaches to sample handling, AFP measurement, or signal readout across the four assay categories. The AFP assay studies discussed in Section 2 span 2015–2026, with most published after 2023. The reported methods are organized as lateral-flow, microfluidic, electrochemical, and optical assays and are evaluated according to the four practical directions proposed above. For each method, we consider specimen type and preparation, total sample-to-result time, manual operation, reader portability, quantitative range, biomarker output, and agreement with central-laboratory methods [46,47]. Finally, we examine the key challenges in sample processing, assay integration, quantitative standardization, and clinical validation and discuss priorities for future development.

2. Rapid and Point-of-Care Detection Methods

Rapid AFP methods include lateral-flow immunoassays, microfluidic assays, electrochemical biosensors, and optical assays [48]. Most reported assays measure total AFP, whereas fewer distinguish AFP-L3 or combine AFP with other HCC biomarkers. AFP-L3 measurement requires glycoform-selective separation or capture, usually based on Lens culinaris agglutinin (LCA) binding [49], but the separated fraction can subsequently be quantified using the same optical or electrochemical readouts used for total AFP. The following sections are therefore organized by assay format or principal sensing architecture, while the biomarker output and complete testing workflow are evaluated within each section.

2.1. Lateral-Flow Immunoassays (LFIAs)

Lateral-flow immunoassay (LFIA) is the most established method for rapid AFP testing because capillary transport, immunoreaction, and signal localization occur on a disposable strip. Most AFP LFIAs use a double-antibody sandwich format in which labeled detection-antibody complexes are captured at the test line. Recent LFIA designs extend beyond visual detection of total AFP to include reader-based quantification, direct whole-blood testing, and multiplex biomarker measurement. However, gains in sensitivity or quantitative readout often require additional labels, signal-generation reactions, or dedicated readers, and only a small number of studies have integrated specimen processing with result calculation.

2.1.1. Colorimetric and Fluorescence LFIAs

Yang et al. [26] used antibody-labeled gold nanotriangles as visible reporters in a sandwich LFIA for serum AFP (Figure 2a). The test line was visible at 5 ng/mL AFP, and the ratio of test-line to control-line intensity provided a quantitative readout (Figure 2b–e). Aptamers provide an alternative to antibody pairs but do not by themselves resolve the workflow and measuring-range limitations of colorimetric LFIA. Ma et al. [50] developed an antibody-free strip using one AFP aptamer conjugated to AuNPs and another immobilized at the test line. The assay provided visual or reader-based results within 10 min, with a linear range of 10–100 ng/mL and an LOD of 10 ng/mL. However, it remained limited to total AFP in serum and had a relatively narrow quantitative range.
Fluorescence LFIAs replace visual color intensity with reader-based fluorescence measurement and are therefore better suited to quantitative testing [51,52]. An earlier europium(III)-chelate microparticle LFIA measured AFP in 5 μL of serum within 15 min over a range of 1–1000 IU/mL; in 284 serum samples, the results correlated with a commercial chemiluminescence immunoassay (r = 0.986) [53]. This study provided stronger clinical evidence than many later material-focused assays, although it still used separated serum. In addition to improving sensitivity, selection of the fluorescence wavelength can reduce optical interference from the sample matrix. Chen et al. [54] developed a second near-infrared-window (NIR-II) LFIA using BBTD-loaded polystyrene nanoparticles (BBTD@PS) and a laboratory-built portable reader (Figure 2f–k). At a fixed AFP concentration, the blood content of the prepared hemolytic sample was increased from 0% to 20%. The NIR-II BBTD@PS readout maintained a more stable T/C signal ratio than the readout obtained using visible-fluorescent rhodamine B-loaded polystyrene nanoparticles (RB@PS). The assay showed a range of 0.80–5700 ng/mL and an LOD of 0.24 ng/mL. This study demonstrated resistance to hemolysis-related optical interference and broad-range AFP quantification, but it evaluated prepared hemolytic samples rather than direct testing of unprocessed whole blood.
Figure 2. Representative colorimetric and fluorescence LFIAs for rapid AFP testing. (a) Sandwich LFIA using gold nanotriangles as visible labels. (b) Strip images obtained with serum samples containing 0–500 ng/mL AFP; the asterisk indicates the naked-eye LOD of 5 ng/mL. (c) Corresponding test- and control-line intensity profiles. (d) T/C responses across the tested AFP concentrations and (e) the linear calibration curve. Adapted with permission from Yang et al. [26]. (f) Conjugation of the capture antibody to NIR-II-emissive BBTD@PS. (g) Configuration of the laboratory-built portable NIR-II reader. (h) Strip architecture and sandwich detection of AFP in hemolytic samples. (i) Photographs and NIR-II fluorescence images obtained using prepared hemolytic samples with blood contents of 0–20% at a fixed AFP concentration. (j,k) Test-to-control signal ratios (IT/IC) obtained using visible-fluorescent RB@PS (j) and NIR-II-emissive BBTD@PS (k), respectively. The lower variation in the BBTD@PS signal indicates reduced optical interference from the hemolytic matrix. Adapted with permission from Chen et al. [54].
Commercial fluorescence LFIAs provide the clearest indication of current AFP POCT readiness. The iCHROMATM AFP assay quantitatively measures total AFP in whole blood, serum, or plasma. According to the manufacturer’s instructions, 30 μL of whole blood or 15 μL of serum or plasma is manually mixed with detection buffer and transferred to the test cartridge. After 15 min, a portable reader reports AFP concentration over a stated range of 5–350 ng/mL [55]. Whole-blood compatibility removes centrifugation, and automated signal processing reduces subjective interpretation. In 25 replicate measurements of a control material, the iCHROMA assay yielded a mean AFP concentration of 24.5 ng/mL and a within-run coefficient of variation of 9.8% [55]. This result describes repeatability near the 20 ng/mL referral threshold. Classification of patient samples on either side of that threshold was not evaluated. A separate comparison using external quality assessment materials yielded coefficients of determination of 0.9457–0.9738 against peer-group means for 12 laboratory immunoassays. However, mixing and cartridge loading remain manual, and the upper measuring limit may require dilution of markedly elevated samples. Moreover, the comparison was not based on paired clinical samples; correlation did not establish interchangeability. This distinction is important for serial AFP measurement, in which a method change could be mistaken for a biological change. The commercial assay also measures total AFP alone and does not provide AFP-L3% or separate PIVKA-II results.

2.1.2. Signal-Amplified and Instrument-Assisted LFIAs

Electrochemical [28], SERS [56], and pressure [41] readouts have been coupled with LFIA to improve analytical sensitivity or enable instrument-based quantification. These methods extend the analytical capability of the strip, but their POCT value depends on whether signal generation and readout are integrated without adding substantial manual work.
Shi et al. [28] developed an electrochemical LFIA using silver-containing labels and a screen-printed electrode (Figure 3a). Silver anodic stripping provided an LOD of 0.85 ng/mL within a total assay time of 20 min. The disposable electrode supported quantitative measurement, although an electrochemical reader remained necessary. SERS-LFIAs have used gold nanorods or gold nanobipyramids to obtain highly sensitive spectral signals [56]. Li et al. [57] reported an LOD of 0.72 pg/mL, but the complete testing time and performance in clinical HCC samples were not reported, and measurement required a Raman spectrometer and spectral analysis software. Xiao et al. [58] developed a portable SERS reader with an integrated multichannel LFIA reaction column for AFP, carcinoembryonic antigen (CEA), and prostate-specific antigen (PSA). The reader automatically scanned the test zones, and the method was examined in 45 clinical serum samples. Although serum preparation was still external and the biomarker combination was not specific to HCC, the study demonstrated automated multichannel signal acquisition.
Figure 3. Electrochemical- and pressure-readout LFIAs for quantitative AFP detection. (a) Structure and detection principle of the DMSNs/AuNPs@Ag-based electrochemical LFIA (e-LFIA), in which silver-containing labels captured at the test line are quantified by anodic stripping using a screen-printed electrode. Adapted with permission from Shi et al. [28]. (b) Pressure-based LFIA for AFP measurement using 1 μL of whole blood without external serum or plasma separation. AFP is captured by Au@Pt nanoparticle–antibody conjugates at the test line. The test line is subsequently excised and immersed in H2O2, and the pressure generated through Au@Pt-catalyzed O2 production is measured using a portable barometer. Adapted with permission from Tang et al. [41].
Direct whole-blood testing was demonstrated by Tang et al. [41], who coupled LFIA with nanozyme-catalyzed gas generation and a portable pressure meter (Figure 3b). The assay quantified AFP in 1 μL of unprocessed whole blood within 30 min, and measurements in 30 whole-blood samples were compared with hospital laboratory results. Its practical value arose from whole-blood compatibility and reduced off-device processing rather than from an exceptionally low LOD. Larger studies are needed to establish the effects of hematocrit, common interferents, reagent lots, and markedly elevated AFP concentrations.
Artificial intelligence (AI) has also been explored in LFIA design and readout. Wang et al. reviewed how models of capillary flow, analyte transport, and binding can be combined with AI to guide strip design [59]. For quantitative readout, Lu et al. used machine vision to measure cardiac biomarkers from strip images [60], while Du et al. analyzed signal development during the first 3 min to estimate analyte concentrations in COVID-19 and hepatitis B assays [61].

2.1.3. AFP-L3 and Multiplex LFIAs

LCA can be incorporated into LFIA to selectively capture fucosylated AFP. In 2023, Puangpila et al. [62] immobilized LCA at the test line to capture fucosylated AFP and completed an AFP-L3 assay within 10 min, with a linear range of 1.5–160 ng/mL in spiked serum. Because total AFP was not measured in parallel, the assay reported only the LCA-reactive fraction and could not calculate AFP-L3%. In comparison, multiplex LFIAs have measured AFP together with glycocholic acid, or PIVKA-II [63,64]. Among these combinations, AFP with PIVKA-II is more directly relevant to HCC. In 2026, Zuo et al. [64] developed HepatoAILFA, which combines an enzyme-accelerated ultrafast polydopamine deposition (EAUPD)-amplified LFIA for AFP and PIVKA-II measurement with smartphone-based image analysis (Figure 4a,b). Figure 4c,d compares the strip responses before and after amplification, and Figure 4e–h shows the corresponding calibration curves for AFP and PIVKA-II. The measured AFP and PIVKA-II concentrations were combined with age and sex in the F4-ASAD machine-learning model. The model was developed and validated using separate derivation and external-validation cohorts, and the complete HepatoAILFA workflow was subsequently evaluated in 90 clinical specimens, achieving an accuracy of 91.1%, a sensitivity of 93.3%, and a specificity of 86.7%. These values describe the diagnostic performance of the combined approach rather than the analytical performance of either biomarker assay.
Figure 4. HepatoAILFA for amplified AFP and PIVKA-II detection and model-based HCC risk assessment. (a) Overview of HepatoAILFA, which combines an EAUPD-amplified LFIA with the F4-ASAD model for personalized precise HCC diagnosis. (b) Strip configuration, assay procedure, and visual or smartphone-assisted quantitative readout. Following sample application, dopamine and H2O2 are added for horseradish peroxidase-catalyzed polydopamine deposition at the test lines. (c,d) Strip responses to different AFP and PIVKA-II concentrations without (c) and with (d) EAUPD amplification. (e,f) AFP calibration curves and linear fits with and without EAUPD amplification. (g,h) Corresponding calibration curves and linear fits for PIVKA-II. DCP in the figure denotes PIVKA-II. Adapted with permission from Zuo et al. [64].
Colorimetric strips offer the simplest visual readout, whereas fluorescence readers enable quantitative AFP measurement over a wider concentration range in the examples reviewed [26,50,53,54]. Other readouts offer different advantages. SERS can lower the detection limit but requires a Raman reader [56,57]. The pressure-based strip accepts 1 μL of unprocessed whole blood, but the operator must excise the test line and immerse it in a reagent before measuring the pressure [41]. The lectin-based strip measured AFP-L3 [62], but calculating AFP-L3% also requires a quantitative result for total AFP. The multiplex strip provided separate measurements of AFP and PIVKA-II [64].

2.2. Microfluidic Assays

Microfluidic devices use microscale channels or droplets to reduce sample and reagent consumption and to combine fluid handling with analytical reactions [65]. Flow can be driven passively by capillary force or actively by pumps, magnetic fields, centrifugal force, or electrical actuation, and quantitative detection may be optical or electrochemical [66]. In AFP testing, recent work ranges from simple channel-based visual assays to programmable digital microfluidics that perform glycoform separation and multistep immunoassays. The central issue is therefore not miniaturization alone, but how much of the sample-to-result workflow is actually performed on the device.

2.2.1. Channel-Based Microfluidic Assays

Channel-based devices can use passive capillary transport for simple operation or incorporate electrodes and compact readers for quantitative measurement [67]. These designs reduce fluid volume and device size, but incubation, magnetic manipulation, washing, or sample transfer may still occur outside the channel. In 2023, Wu et al. [68] developed a microfluidic particle-accumulation assay for visual AFP measurement in undiluted serum (Figure 5a). AFP formed sandwich complexes between capture-antibody-modified magnetic microparticles and detection-antibody-modified polystyrene microparticles. These complexes were retained by a magnetic separator, whereas unbound polystyrene microparticles continued into a narrow trapping channel. Because increasing AFP concentrations generated more magnetically retained complexes and left fewer unbound polystyrene microparticles, the accumulation length decreased with AFP concentration and provided a visual quantitative readout. Figure 5b,c shows the particle accumulation and calibration in buffer, while Figure 5d,e shows the selectivity tests. A rapid mode covered 0–1000 ng/mL after 15 min of incubation but had an LOD of 34.3 ng/mL, whereas a 70-min incubation lowered the LOD to 15.8 ng/mL and covered 0–500 ng/mL. This study therefore illustrates a practical trade-off between testing time and sensitivity around 20 ng/mL, a concentration commonly used to prompt further evaluation. Shen et al. [69] integrated electrochemical detection into a 3D-printed microfluidic device connected to a smartphone reader. A gold nanoparticle-doped three-dimensional reduced-graphene-oxide electrode provided a range of 0.1 pg/mL-200 ng/mL and an LOD of 0.045 pg/mL using approximately 9 μL of reagent. The compact reader and low reagent consumption supported portable measurement, but the complete testing time and off-chip incubation, washing, and transfer steps were not clearly reported. Moreover, the upper measuring limit of 200 ng/mL would require additional dilution steps.
Figure 5. Visual AFP quantification using a microfluidic particle-accumulation assay. (a) Working principle based on the formation of magnetic microparticle–AFP–polystyrene microparticle complexes and the accumulation of unbound polystyrene microparticles in a microfluidic channel. (b,c) Representative particle-accumulation images and calibration for AFP in buffer. (d,e) Selectivity evaluation against potentially interfering substances. Error bars represent the mean ± standard deviation (n = 3). Adapted with permission from Wu et al. [68].

2.2.2. Digital Microfluidic Assays

Digital microfluidics manipulates individual droplets on an electrode array. In electrowetting-on-dielectric (EWOD) devices, electrical potentials change surface wettability so that droplets can be dispensed, moved, merged, mixed, and separated according to a programmed sequence [66]. This capability is particularly relevant to AFP-L3 because lectin-based fractionation can be combined with subsequent immunoassays on the same device.
In 2025, an EWOD digital microfluidic assay was developed for total AFP, AFP-L3, and PIVKA-II measurement (Figure 6a–c) [70]. AFP-L3 separation from serum was completed within 10 min, followed by magnetic-particle immunoassays within 15 min; the complete analysis required approximately 30 min. About 2.4 μL of serum was used for each biomarker, and one chip analyzed all three biomarkers in five samples. The reported LODs were 0.24 ng/mL for total AFP and AFP-L3 and 1.89 ng/mL for PIVKA-II. Because total AFP and AFP-L3 were measured separately, AFP-L3% could be calculated, while PIVKA-II was reported as an independent quantitative result. Among the microfluidic methods reviewed here, this platform most directly addressed glycoform measurement, complementary biomarkers, and analytical automation. The remaining limitations were primarily related to specimen handling and clinical validation rather than analytical sensitivity. Serum preparation remained external, and the assay required magnetic particles, an EWOD controller, signal-generation reagents, and software-based calculation. The platform is therefore closer to a compact automated analyzer than to an instrument-free rapid test. Evaluation in 27 clinical serum samples showed no significant group-level differences between the two methods (Figure 6d,e), but this was insufficient to establish agreement with central-laboratory methods for longitudinal monitoring.
Figure 6. Electrowetting-on-dielectric digital microfluidic platform for rapid measurement of AFP, AFP-L3%, and PIVKA-II. (a,b) Photographs of the digital microfluidic platform and its temperature-control module. (c) Layout of the reagent, sample, waste, and incubation/detection areas on the EWOD chip. (d,e) Comparison of AFP, AFP-L3%, and PIVKA-II results obtained from 27 clinical serum samples using the digital microfluidic platform and conventional central-laboratory instruments. DCP in the figure denotes PIVKA-II; NS, not significant. Adapted with permission from Zhu et al. [70].
The particle-accumulation chip allowed visual quantification of AFP in undiluted serum, with a trade-off between speed and sensitivity. Extending the incubation time from 15 to 70 min lowered the detection limit from 34.3 to 15.8 ng/mL [68]. An electrode-equipped chip achieved a lower detection limit with approximately 9 μL of reagent, although the complete testing time was unclear [69]. Digital microfluidics offered a different advantage. Programmable droplet handling integrated AFP-L3 separation with immunoassays for AFP, AFP-L3, and PIVKA-II, but required a dedicated control instrument [70]. The representative microfluidic AFP assays described above all used serum rather than unprocessed whole blood [68,69,70]. Direct whole-blood testing would require reliable on-chip separation of blood cells or plasma, and separation efficiency may be affected by hematocrit, cell accumulation, channel geometry, and flow conditions [71].

2.3. Electrochemical Biosensors

Electrochemical AFP biosensors convert antibody- or aptamer-based recognition into changes in current, potential, conductance, or impedance [72]. Their signals are independent of sample color and ambient illumination, and disposable electrodes can be connected to compact potentiostats or smartphones [73,74]. Recent AFP studies have used conductive nanomaterial interfaces, ratiometric self-calibration, homogeneous reactions, paper-based devices, and dual signals for AFP-L3% calculation. These strategies address different limitations: interfacial engineering improves sensitivity, ratiometric measurement reduces signal fluctuation, homogeneous or paper-based designs simplify parts of the workflow, and glycoform-selective recognition expands biomarker output. Nevertheless, surface fouling, matrix effects, electrode-to-electrode variation, and the additional incubation and washing required by amplified assays remain important barriers to quantitative POCT [75].

2.3.1. Redox-Probe-Assisted Interfacial Aptasensors

AFP aptamers have been used in both electrochemical and optical assays [76]. Several interfacial AFP aptasensors have been described as label-free because neither AFP nor a secondary recognition reagent is chemically labeled. Their signals nevertheless depend on a soluble or electrode-associated redox probe whose electron transfer is altered by AFP binding. They are therefore considered here as redox-probe-assisted interfacial aptasensors. They reduce secondary-labeling steps, but quantitative performance depends strongly on electrode conductivity, aptamer immobilization, surface blocking, and control of nonspecific adsorption. The assay configuration can also affect apparent affinity: the same AFP aptamer yielded different affinity estimates in magnetic-bead binding and aptamer–antibody sandwich assays [77].
In 2021, Upan et al. [78] immobilized an AFP aptamer on a platinum nanoparticle/carboxylated graphene oxide-modified screen-printed electrode and monitored the binding-induced decrease in hydroquinone response. Conductive nanomaterials and bovine serum albumin (BSA) blocking improved the interfacial response, but the assay covered only 3–30 ng/mL and required a 45-min incubation and 100-fold serum dilution. Yaiwong et al. [79] subsequently developed a disposable aptasensor containing polyethyleneimine-coated AuNPs and electrode-associated toluidine blue indicator. AFP binding reduced the electrochemical response over a range of 0.01–50 ng/mL, with an LOD of 9.5 pg/mL in diluted serum. A small potentiostat connected to an Android smartphone enabled portable signal acquisition. However, serum was diluted 50-fold, and the 30-min incubation, washing, and transfer to measurement buffer remained manual.

2.3.2. Sandwich and Ratiometric Electrochemical Assays

Sandwich architectures introduce a second recognition event to improve specificity, whereas ratiometric designs use an internal reference signal to reduce variation in electrode response. Both approaches can improve quantitative reliability, but they generally require more reagents and reaction steps than interfacial aptasensors.
Chanarsa et al. [80] combined antibody capture with an AFP aptamer–methylene blue complex on a MoSe2/WSe2-modified screen-printed electrode. Although dual recognition and smartphone-based measurement improved specificity and portability, the assay required 50-fold serum dilution, multiple incubations, and repeated washing.
Sun et al. [81] addressed signal fluctuation using a ratiometric sandwich immunosensor with ferrocene (Fc) and toluidine blue (TB) as two electrochemical reporters (Figure 7a–c). AFP was quantified from the ratio of their peak currents, providing internal signal self-calibration over a range of 0.001–200 ng/mL. The results in clinical serum were consistent with a commercial ECLIA and were more accurate than those derived from the corresponding non-ratiometric signal. This study demonstrated that improving quantitative stability may be more clinically relevant than lowering the LOD alone. However, the nanomaterial-modified sandwich format still required a labeled secondary antibody, and a portable or automated workflow was not demonstrated.

2.3.3. Homogeneous and Paper-Based Electrochemical Assays

Homogeneous recognition can remove solid-phase separation and washing, whereas paper-based devices can incorporate fluid transport and electrodes into a disposable substrate. These approaches simplify different parts of the workflow but do not necessarily eliminate hazardous reagents, serum preparation, or sample dilution.
Li et al. [82] developed a homogeneous AFP assay based on nano-impact electrochemistry. AFP binding released aptamers from platinum nanoparticles and restored the current generated by nanoparticle-catalyzed hydrazine oxidation. Because recognition occurred in solution, this method avoided target labeling, solid-phase separation, and washing. It achieved an LOD of 50 pg/mL and was applied to diluted sera from patients with HCC. A sliding electrochemical paper-based device reported by Li et al. [83] integrated paper fluid handling with a flexible three-electrode configuration in a sliding electrochemical paper-based analytical device (ePAD). The assay covered 0.005–50 ng/mL, and the paper-based assay and chemiluminescence immunoassay (CLIA) were compared for positive or negative classification of 30 clinical serum samples (Figure 7d,e). Receiver operating characteristic (ROC) analysis yielded an area under the curve (AUC) of 0.992 for distinguishing liver cancer from normal controls (Figure 7f). This study advanced device integration and clinical method comparison, although it still required serum and had an upper measuring limit that would necessitate dilution of many AFP-positive samples. Homogeneous detection and paper-based integration have therefore reduced different manual steps, but neither has yet been combined with direct whole-blood processing and automated dilution.
Figure 7. Ratiometric and paper-based electrochemical biosensors for AFP detection. (a) Fabrication and detection principle of a ratiometric electrochemical immunosensor using Fc and TB signals. (b) Selectivity of the ratiometric immunosensor toward AFP. (c) Comparison of AFP concentrations in clinical serum samples measured by ECLIA, the ratiometric sensor, and the corresponding non-ratiometric sensor. Adapted with permission from Sun et al. [81]. (d,e) Classification of 30 clinical serum samples using ePAD and CLIA; “+” and “−” indicate positive and negative results, respectively. (f) ROC curve for distinguishing liver cancer from normal controls. Adapted with permission from Li et al. [83] *** p < 0.001. ECL in the figure denotes ECLIA.

2.3.4. Glycoform-Selective Electrochemical Assays

Electrochemical AFP-L3 assays require both selective recognition of the LCA-reactive fraction and independent measurement of total AFP if AFP-L3% is to be calculated. Li et al. [84] used LCA-functionalized silver nanoparticles to detect AFP-L3 directly over a range of 0.025–15 ng/mL, but total AFP was not measured. Figure 8 compares three subsequent strategies that generated separate electrochemical signals for total AFP and AFP-L3, thereby enabling AFP-L3% calculation.
Figure 8. Electrochemical strategies for AFP-L3% measurement. (a) Fe2Ni-MOF-based signal-off-to-on assay for the separate measurement of total AFP and AFP-L3. Adapted with permission from Tu et al. [85]. (b,c) Dilution linearity of two serum samples containing 102.8 ng/mL AFP with an AFP-L3% of 41.2% and 52.3 ng/mL AFP with an AFP-L3% of 6.9%, respectively; the dashed line indicates the 10% cutoff for AFP-L3%. Adapted with permission from Wei et al. [22]. (d) Ordered-labeling electrochemical aptasensor using CdS QDs to label total AFP and LCA-functionalized AgNPs to label AFP-L3 after Exo III-mediated probe removal, enabling AFP-L3% calculation. Adapted with permission from Zhao et al. [39].
Tu et al. [85] used a signal-off-to-on strategy based on an Fe2Ni metal–organic framework (Fe2Ni-MOF) (Figure 8a). The Fe2Ni-MOF amplified a decreasing hydroquinone signal for total AFP, whereas silver nanoparticles generated an increasing signal for AFP-L3. The strategy avoided the exonuclease-dependent probe-removal step and produced an AFP-L3 LOD of 69 pg/mL. Nevertheless, sequential recognition and separate calibration of total AFP and AFP-L3 were still required.
Wei et al. [22] introduced two nanoparticle-derived electrochemical signals for simultaneous measurement of total AFP and AFP-L3 in serum, enabling AFP-L3% calculation. Serial dilution of two serum samples, one containing 102.8 ng/mL AFP with an AFP-L3% of 41.2% and the other containing 52.3 ng/mL AFP with an AFP-L3% of 6.9%, showed that the calculated AFP-L3% remained relatively stable as the AFP concentration decreased (Figure 8b,c). The two samples also remained above and below the 10% threshold, respectively, across the tested dilutions.
Zhao et al. [39] further developed an ordered-labeling assay on a single aptamer-modified electrode (Figure 8d). Total AFP was first labeled using DNA probes functionalized with CdS quantum dots (CdS QDs). After exonuclease III (Exo III)-mediated removal of the DNA probes, AFP-L3 was recognized by LCA and labeled with silver nanoparticles (AgNPs). The respective ranges were 0.0008–40 ng/mL for total AFP and 0.004–40 ng/mL for AFP-L3, allowing AFP-L3% to be calculated from two electrochemical signals. This study directly addressed AFP-L3% rather than AFP-L3 alone, but sequential labeling, exonuclease treatment, LCA recognition, and two measurements produced a complex workflow. The upper limit of 40 ng/mL would also require dilution of many AFP-positive samples, potentially adding error to the calculated ratio.
Interfacial aptasensors can measure total AFP without a labeled secondary reagent, although the reported protocols still required serum dilution and manual incubation [78,79]. Sandwich assays add a second recognition step, while a ratiometric readout uses a reference signal to compensate for variation in electrode response [80,81]. Homogeneous recognition avoids solid-phase washing, and paper devices integrate fluid transport and electrodes into a disposable substrate [82,83]. For glycoform analysis, measuring total AFP and AFP-L3 separately allows the assay to report AFP-L3%. This requires calibration of both signals and, in some designs, sequential labeling [22,39,85].

2.4. Optical Assays

Optical readouts have already been discussed in the lateral-flow and microfluidic sections. This section focuses on AFP assays implemented using optical fibers, plasmonic substrates, microplates, or solution-phase reactions. Although these methods can provide rapid or sensitive signal acquisition, their complete workflows may still require incubation, washing, magnetic separation, or addition of signal-generating reagents [86].

2.4.1. Plasmonic Resonance Assays

Plasmonic sensors are sensitive to refractive-index changes near a metal surface, allowing biomolecular binding to be measured through changes in resonance wavelength or intensity [87]. In 2023, Jing et al. [34] developed an optical-fiber near-field-enhanced plasmonic resonance (NFE-PR) immunoprobe for AFP measurement. A capture antibody was immobilized on the fiber surface, and an unlabeled detection antibody was subsequently added to increase the resonance wavelength shift (Figure 9a,b). The reported LODs were 0.01 ng/mL using logistic calibration and 0.05 ng/mL using Langmuir calibration. Five serum samples containing approximately 2–40 ng/mL AFP were prepared by diluting or spiking commercial healthy-donor serum, and the results correlated with those obtained using a commercial magnetic nanoparticle chemiluminescence kit. However, this was not a method-comparison study using patient samples. The assay also required 4 mL of sample, a 15-min AFP incubation, rinsing, a further 15-min detection-antibody incubation, and a spectrometer-based optical setup. Its compact fiber probe therefore did not translate into a simple sample-to-result workflow. In the same year, Li et al. [88] developed a nanozyme-linked immunosorbent surface plasmon resonance (nano-ELISPR) assay using a gold–silver nanocup metasurface plasmon resonance (MetaSPR) chip and an Au@Pt nanoflower (Au@Pt NF)-labeled detection antibody. The nanozyme catalyzed 3,3′,5,5′-tetramethylbenzidine (TMB) oxidation, and the oxidized product etched the silver layer of the MetaSPR chip, producing an absorption change that could be measured with a portable reader (Figure 9c). The etching responses produced by Au@Pt NFs, Au@Pt alloys, and horseradish peroxidase (HRP) were compared (Figure 9d). The spectra, calibration curve, and time-dependent optical density (OD) responses over 0–100 pg/mL are shown in Figure 9e–g. The assay used approximately 20 μL of serum, required about 40 min, and achieved an AFP LOD of 1.5 pg/mL, corresponding to approximately 21.74 fM. In 60 clinical samples, AFP classification at a cutoff of 20 ng/mL showed a sensitivity of 86.7% and a specificity of 97.8%. The platform was also adapted to CEA and cancer antigen 125 (CA125), but the three biomarkers were measured in separate assays rather than simultaneously. It therefore demonstrated platform versatility rather than multiplex measurement of an HCC biomarker panel. Serum preparation also remained external to the assay.
Figure 9. Optical-fiber NFE-PR and nano-ELISPR assays for AFP detection. (a) Sensing structure of the optical-fiber NFE-PR immunoprobe. (b) Double-antibody sandwich immunoassay for AFP using the NFE-PR immunoprobe. Adapted with permission from Jing et al. [34]. (c) Nano-ELISPR assay based on Au@Pt NF-catalyzed TMB oxidation and etching of the silver layer on the MetaSPR chip. (d) Etching responses produced by Au@Pt NFs, Au@Pt alloys, and HRP. (e) Absorption spectra at AFP concentrations of 0–100 pg/mL. (f) Calibration curve for AFP detection (R2 = 0.996; LOD = 1.5 pg/mL). Data are presented as mean ± SD (n = 3). (g) Time-dependent OD responses at different AFP concentrations over 600 s. Adapted with permission from Li et al. [88].

2.4.2. Nanozyme-Amplified Colorimetric Assays

Nanozymes are nanomaterials with enzyme-like catalytic activity and can offer greater operational stability and lower production costs than natural enzymes [89]. In 2024, Tang et al. [90] developed a metal enzyme-linked immunosorbent assay using Cu-MOFs@AuPtNPs composites with peroxidase-like activity as the detection label. The double-antibody sandwich assay showed a linear range of 0.001–1000 ng/mL and an LOD of 0.86 pg/mL. The color change was visible to the naked eye, and the method was examined using clinical serum samples. Its broad range was more relevant to quantitative AFP measurement than that of many ultrasensitive assays. However, the plate-based method still required sequential incubation, washing, and color development. Naked-eye detection simplified endpoint readout but did not provide an integrated POCT workflow. Further transfer to a disposable cartridge or paper-based device would be required to reduce manual operation.

2.4.3. Chemiluminescence Assays

Chemiluminescence does not require an external excitation source and generally provides a high signal-to-background ratio, allowing light emission to be measured using relatively simple optical detectors [91]. In 2024, Liu et al. [92] developed a target-triggered chemiluminescence aptasensor using aptamer–luminol-modified magnetic graphene oxide and a ZnO nanoparticle–Au@CuMOF nanozyme. AFP binding released aptamer-linked luminol from the magnetic material, after which the nanozyme catalyzed the luminol–H2O2 reaction. The assay showed a linear range of 1.0 × 10−4–50 ng/mL and an LOD of 4.2 × 10−5 ng/mL and was applied to serum samples. Despite its high analytical sensitivity, the upper limit of 50 ng/mL would require dilution of many AFP-positive samples. The method also required magnetic separation, addition of chemiluminescent reagents, and an optical detector, none of which were integrated into a closed testing device. It is therefore better considered a sensitive laboratory aptasensor with potential for miniaturization than an established POCT assay.
Plasmonic assays detect AFP binding through a change in optical resonance and require an optical reader. Sample requirements differed substantially even within this group. The fiber-based assay used 4 mL of sample, whereas nano-ELISPR used approximately 20 μL and achieved a detection limit of 1.5 pg/mL [34,88]. Nanozyme-amplified colorimetry offered a visible endpoint, although the plate assay still required sequential incubation, washing, and color development [90]. The chemiluminescence assay reported by Liu et al. provided a sensitive readout without an excitation source, but required magnetic separation, reagent addition, and a light detector [92].

3. Discussion

3.1. Scenario-Based Comparison of Current Methods

The suitability of a rapid AFP assay depends on its intended setting. One proposed use of patient-operated testing is to flag a possible AFP elevation between scheduled surveillance visits. Tang et al. demonstrated AFP measurement in 1 μL of unprocessed whole blood [41], although patients’ ability to perform the test and interpret the result remains to be established. In this setting, a semiquantitative warning would prompt the patient to contact the clinical team for confirmatory serum AFP measurement and review of the result. A result below the warning threshold would leave the surveillance schedule unchanged.
In clinics that use ultrasound and AFP for routine surveillance, a quantitative total AFP assay provides the blood measurement needed [2]. Multiplex testing may add value when a clinic uses a multibiomarker score to assess HCC risk in patients with cirrhosis. In a prospective multicenter study, GALAD, which combines age, sex, AFP, AFP-L3%, and PIVKA-II, had a sensitivity of 62% for HCC diagnosed within the following 12 months, compared with 41% for AFP alone at the same specificity of 82% [93]. The study did not assess whether adding AFP-L3% and PIVKA-II to ultrasound plus total AFP improves HCC detection in routine surveillance. Whether such an addition would justify the extra complexity and cost of a multiplex POCT device therefore remains uncertain. For either single-marker or multiplex testing in a clinic, serum preparation, a compact reader, and a moderately complex cartridge may be acceptable if trained staff can complete the assay during the same visit. The commercial iCHROMA fluorescence LFIA provides near-patient total AFP measurement from whole blood using a portable reader [55]. For multiplex testing, a digital microfluidic platform has demonstrated combined measurement of total AFP, AFP-L3, and PIVKA-II, although it requires prepared serum and a dedicated controller [70]. A nanozyme-based colorimetric assay achieved a low LOD and a broad calibration range, but sequential incubation, washing, and color development still limit its use at the point of care [90]. Table 1 compares specimen requirements, testing speed, readout, and operating requirements across platform categories. Table 2 gives the reported detection limits, calibration ranges, assay or incubation times, and sample volumes for selected individual assays.
Table 1. Specimen types, readout methods, and operating requirements of rapid AFP testing platforms.
Table 2. AFP detection limits, calibration ranges, assay times, and sample volumes reported for selected assays.

3.2. Challenges to Clinical Translation

The first challenge is incomplete workflow integration. Portable readers do not establish POCT suitability when centrifugation, dilution, incubation, or washing remains external, whereas direct whole-blood testing requires control of hematocrit and matrix effects.
The second challenge is quantitative comparability. Differences among AFP immunoassays and the method- and matrix-dependent commutability of reference materials indicate that calibration alone may not ensure comparable results. For example, the iCHROMA AFP assay showed positive bias relative to laboratory peer-group means despite high correlations in external quality assessment materials [55]. To establish interchangeability, paired measurements in patient specimens should span the intended measuring range, and the observed bias and limits of agreement should be evaluated against clinically acceptable differences.
The third challenge is clinically appropriate validation. For assays that report AFP-L3%, both AFP-L3 and total AFP require independent calibration, and uncertainty in either measurement affects the calculated percentage. At low total AFP concentrations, AFP-L3 may approach its limit of quantification (LOQ) even when total AFP can still be measured precisely. When either measurement approaches its LOQ, small variations in the measured concentrations may shift the calculated AFP-L3% across a clinical decision threshold. Each platform should therefore establish the conditions under which it can reliably report AFP-L3%, with particular attention to precision near the intended threshold. The choice of validation samples also depends on the intended use. For example, recovery of AFP added to serum was used to assess matrix effects in a fluorescent aptamer assay [94]. Evaluating an assay for longitudinal surveillance requires serial patient samples to determine how reliably it tracks changes in endogenous AFP over time.

3.3. Opportunities

Future priorities should reflect the intended clinical application. LFIAs should focus on direct finger-prick whole-blood testing and simple semiquantitative interpretation. Microfluidic assays should integrate blood-cell separation and automatic dilution. Devices intended for multibiomarker risk assessment could also measure total AFP, AFP-L3, and PIVKA-II in parallel. Electrochemical biosensors should improve antifouling performance, electrode reproducibility, and internal calibration. Among optical methods, plasmonic assays should reduce sample volume and reader complexity, whereas nanozyme colorimetric and chemiluminescence assays should move toward disposable cartridges with automated reagent handling. These priorities support two complementary routes: a low-cost whole-blood alert test for patient-operated use and a compact, traceable multibiomarker analyzer for clinic- or hospital-based quantitative testing.

4. Conclusions

In summary, we reviewed recent advances in lateral-flow, microfluidic, electrochemical, and optical methods for rapid and point-of-care AFP testing. These methods were evaluated according to specimen type, sample-to-result time, workflow integration, quantitative range, readout portability, and biomarker output, with particular attention to direct whole-blood testing, AFP-L3% calculation, separate PIVKA-II measurement, and automation. This comparison indicates that future development should be guided by the application scenario. Patient-operated or community tests should prioritize low-cost, error-resistant whole-blood measurement and easily interpreted warning results, whereas clinic- or hospital-based platforms should provide traceable total AFP results over a broad measuring range. Measurement of AFP-L3% and PIVKA-II is relevant when the intended clinical assessment uses these markers. We expect that continued advances in these areas will promote the clinical translation of rapid AFP testing and improve HCC surveillance and management.

Author Contributions

Conceptualization, R.W. and J.H.; investigation, R.W., H.Z. and Q.Z.; writing—original draft preparation, R.W.; writing—review and editing, R.W., H.Z., Q.Z. and J.H.; supervision, R.W. and J.H. All authors have read and agreed to the published version of the manuscript.

Funding

This work was supported by the Natural Science Foundation of Xinjiang Province (2026D01C91), JiangSu ShuangChuang Team (JSSCTD202409), Jiangsu Young Elite Scientists Sponsorship Program (JSTJ-2025-252), Open Research Fund of State Key Laboratory of Digital Medical Engineering (2025-K01), the Fundamental Research Funds for Central Universities (2242026K30019), and the Young Scientists Fund of State Key Laboratory of Digital Medical Engineering (2242026K30019).

Institutional Review Board Statement

Not applicable.

Data Availability Statement

No new data were created or analyzed in this study.

Acknowledgments

The authors used ChatGPT (OpenAI, GPT-6) to help draft and revise portions of the text and to generate the schematic in Figure 1 (version: gpt-image). The authors take responsibility for the final text and figure.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Llovet, J.M.; Kelley, R.K.; Villanueva, A.; Singal, A.G.; Pikarsky, E.; Roayaie, S.; Lencioni, R.; Koike, K.; Zucman-Rossi, J.; Finn, R.S. Hepatocellular carcinoma. Nat. Rev. Dis. Primers 2021, 7, 6. [Google Scholar]
  2. Singal, A.G.; Llovet, J.M.; Yarchoan, M.; Mehta, N.; Heimbach, J.K.; Dawson, L.A.; Jou, J.H.; Kulik, L.M.; Agopian, V.G.; Marrero, J.A.; et al. AASLD Practice Guidance on prevention, diagnosis, and treatment of hepatocellular carcinoma. Hepatology 2023, 78, 1922–1965. [Google Scholar] [CrossRef] [Scilit]
  3. European Association for the Study of the Liver. EASL Clinical Practice Guidelines on the management of hepatocellular carcinoma. J. Hepatol. 2025, 82, 315–374. [Google Scholar] [CrossRef] [Scilit]
  4. Reig, M.; Forner, A.; Rimola, J.; Ferrer-Fàbrega, J.; Burrel, M.; Garcia-Criado, Á.; Kelley, R.K.; Galle, P.R.; Mazzaferro, V.; Salem, R.; et al. BCLC strategy for prognosis prediction and treatment recommendation: The 2022 update. J. Hepatol. 2022, 76, 681–693. [Google Scholar] [CrossRef] [Scilit]
  5. Tzartzeva, K.; Obi, J.; Rich, N.E.; Parikh, N.D.; Marrero, J.A.; Yopp, A.; Waljee, A.K.; Singal, A.G. Surveillance Imaging and Alpha Fetoprotein for Early Detection of Hepatocellular Carcinoma in Patients With Cirrhosis: A Meta-analysis. Gastroenterology 2018, 154, 1706–1718.e1. [Google Scholar] [CrossRef] [Scilit]
  6. Yeo, Y.H.; Lee, Y.-T.; Tseng, H.-R.; Zhu, Y.; You, S.; Agopian, V.G.; Yang, J.D. Alpha-fetoprotein: Past, present, and future. Hepatol. Commun. 2024, 8, e0422. [Google Scholar] [CrossRef] [Scilit]
  7. Mak, L.-Y.; Yuen, M.-F. Surveillance for Hepatocellular Carcinoma in Cirrhosis: End of Monopoly for Serum Alpha Fetoprotein. Gastroenterology 2025, 168, 217–219. [Google Scholar] [CrossRef] [Scilit]
  8. Li, M.; Hannan, L.M.; Goyal, L.; Bocobo, A.G.; Parks, A.L.; Bauer, K.; Baiev, I.; Dinicola, C.; Gordan, J.D.; Venook, A.P.; et al. Changes in alpha-fetoprotein across the systemic therapy continuum in advanced hepatocellular carcinoma—A real-world, multicenter study. Ther. Adv. Med. Oncol. 2024, 16, 17588359241297085. [Google Scholar] [CrossRef] [Scilit]
  9. Liu, Z.; Liu, X.; Pei, H.; Bao, K.; Su, B.; Cao, H.; Wu, L.; Chen, Q. Rapid and sensitive immunoassay for alpha-fetoprotein in serum by fabricating primary antibody–enzyme complexes using protein self-assembly. Anal. Methods 2023, 15, 1576–1583. [Google Scholar] [CrossRef] [Scilit]
  10. Zhang, L.; Jin, Y.; Li, B. Asymmetric Sulfur Vacancy-Engineered Cu–Co9S8 Nanozyme: A High-Performance Signal Label for Chemiluminescence Immunoassay. Anal. Chem. 2026, 98, 13516–13525. [Google Scholar] [CrossRef] [Scilit]
  11. Chan, H.L.Y.; Vogel, A.; Berg, T.; De Toni, E.N.; Kudo, M.; Trojan, J.; Eiblmaier, A.; Klein, H.-G.; Hegel, J.K.; Sharma, A.; et al. Performance evaluation of the Elecsys PIVKA-II and Elecsys AFP assays for hepatocellular carcinoma diagnosis. JGH Open 2022, 6, 292–300. [Google Scholar] [CrossRef] [Scilit]
  12. Liu, W.; Kang, Q.; Wang, P.; Zhou, F. Ratiometric fluorescence immunoassay based on MnO2–o-phenylenediamine–fluorescent carbon nanodots for the detection of α-fetoprotein via fluorescence resonance energy transfer. New J. Chem. 2022, 46, 1120–1126. [Google Scholar] [CrossRef] [Scilit]
  13. Chen, C.; Zhao, J.; Lu, Y.; Sun, J.; Yang, X. Fluorescence Immunoassay Based on the Phosphate-Triggered Fluorescence Turn-on Detection of Alkaline Phosphatase. Anal. Chem. 2018, 90, 3505–3511. [Google Scholar] [CrossRef] [Scilit]
  14. van Rossum, H.H.; Holdenrieder, S.; Ballieux, B.E.P.B.; Badrick, T.C.; Yun, Y.-M.; Zhang, C.; Patel, D.; Thelen, M.; Song, J.; Wojtalewicz, N.; et al. Investigating the Current Harmonization Status of Tumor Markers Using Global External Quality Assessment Programs: A Feasibility Study. Clin. Chem. 2024, 70, 669–679. [Google Scholar] [CrossRef] [Scilit]
  15. Zhang, J.; Zhao, J.; Wang, Q.; Zhang, R.; Yue, Y. Evaluating the Commutability of Reference Materials for α-Fetoprotein: Accurate Value Assignment With Multiple Systems and Trueness Verification. Ann. Lab. Med. 2024, 44, 507–517. [Google Scholar] [CrossRef] [Scilit]
  16. Liu, M.; Wen, Y. Point-of-care testing for early-stage liver cancer diagnosis and personalized medicine: Biomarkers, current technologies and perspectives. Heliyon 2024, 10, e38444. [Google Scholar] [CrossRef] [Scilit]
  17. Liu, K.; Wu, C.; Zhu, M.; Xu, J.; Lin, B.; Lin, H.; Liu, Z.; Li, M. Structural characteristics of alpha-fetoprotein, including N-glycosylation, metal ion and fatty acid binding sites. Commun. Biol. 2024, 7, 505. [Google Scholar] [CrossRef] [Scilit]
  18. Wang, Y.; Chen, H. Protein glycosylation alterations in hepatocellular carcinoma: Function and clinical implications. Oncogene 2023, 42, 1970–1979. [Google Scholar] [CrossRef] [Scilit]
  19. Lee, J.; Yeo, I.; Kim, Y.; Shin, D.; Kim, J.; Kim, Y.; Lim, Y.-S.; Kim, Y. Comparison of Fucose-Specific Lectins to Improve Quantitative AFP-L3 Assay for Diagnosing Hepatocellular Carcinoma Using Mass Spectrometry. J. Proteome Res. 2022, 21, 1548–1557. [Google Scholar] [CrossRef] [Scilit]
  20. Mehta, N.; Kotwani, P.; Norman, J.; Shui, A.; Li, P.J.; Saxena, V.; Chan, W.; Yao, F.Y. AFP-L3 and DCP are superior to AFP in predicting waitlist dropout in HCC patients: Results of a prospective study. Liver Transplant. 2023, 29, 1041–1049. [Google Scholar] [CrossRef] [Scilit]
  21. Kudo, M. Urgent Global Need for PIVKA-II and AFP-L3 Measurements for Surveillance and Management of Hepatocellular Carcinoma. Liver Cancer 2024, 13, 113–118. [Google Scholar] [CrossRef] [Scilit]
  22. Wei, T.; Zhang, W.; Tan, Q.; Cui, X.; Dai, Z. Electrochemical Assay of the Alpha Fetoprotein-L3 Isoform Ratio To Improve the Diagnostic Accuracy of Hepatocellular Carcinoma. Anal. Chem. 2018, 90, 13051–13058. [Google Scholar] [CrossRef] [Scilit]
  23. Li, Z.; Li, H.; Deng, D.; Liu, R.; Lv, Y. Mass Spectrometric Assay of Alpha-Fetoprotein Isoforms for Accurate Serological Evaluation. Anal. Chem. 2020, 92, 4807–4813. [Google Scholar] [CrossRef] [Scilit]
  24. Ma, H.; Sun, X.; Chen, L.; Cheng, W.; Han, X.X.; Zhao, B.; He, C. Multiplex Immunochips for High-Accuracy Detection of AFP-L3% Based on Surface-Enhanced Raman Scattering: Implications for Early Liver Cancer Diagnosis. Anal. Chem. 2017, 89, 8877–8883. [Google Scholar] [CrossRef] [Scilit]
  25. Hou, J.; Berg, T.; Vogel, A.; Piratvisuth, T.; Trojan, J.; De Toni, E.N.; Kudo, M.; Malinowsky, K.; Findeisen, P.; Hegel, J.K.; et al. Comparative evaluation of multimarker algorithms for early-stage HCC detection in multicenter prospective studies. JHEP Rep. 2025, 7, 101263. [Google Scholar] [CrossRef] [Scilit]
  26. Yang, C.; Zhou, Y.; Fan, C.; Wang, H.; Xu, J.; Wu, Z.; Song, Y.; Hu, Y.; Tian, M.; Liu, G. Gold nanotriangle-enhanced high-sensitive determination of alpha-fetoprotein via lateral flow immunoassay in serum. Microchim. Acta 2025, 192, 431. [Google Scholar] [CrossRef] [Scilit]
  27. Yakoh, A.; Mehmeti, E.; Kalcher, K.; Chaiyo, S. Hand-Operated, Paper-Based Rotational Vertical-Flow Immunosensor for the Impedimetric Detection of α-Fetoprotein. Anal. Chem. 2022, 94, 5893–5900. [Google Scholar] [CrossRef] [Scilit]
  28. Shi, D.; Yin, Y.; Li, X.; Yuan, J. Signal-Boosted Electrochemical Lateral Flow Immunoassay for Early Point-of-Care Detection of Liver Cancer Biomarker. ACS Sens. 2024, 9, 5293–5301. [Google Scholar] [CrossRef] [Scilit]
  29. Dai, B.; Yin, C.; Wu, J.; Li, W.; Zheng, L.; Lin, F.; Han, X.; Fu, Y.; Zhang, D.; Zhuang, S. A flux-adaptable pump-free microfluidics-based self-contained platform for multiplex cancer biomarker detection. Lab Chip 2021, 21, 143–153. [Google Scholar] [CrossRef] [Scilit]
  30. Li, X.; Pan, X.; Lu, J.; Zhou, Y.; Gong, J. Dual-modal visual/photoelectrochemical all-in-one bioassay for rapid detection of AFP using 3D printed microreactor device. Biosens. Bioelectron. 2020, 158, 112158. [Google Scholar] [CrossRef] [Scilit]
  31. Shang, Z.; Su, T.; Jin, D.; Xu, Q.; Hu, X.; Shu, Y. An integrated and flexible PDMS/Au film-based electrochemical immunosensor via Fe–Co MOF as a signal amplifier for alpha fetoprotein detection. Biosens. Bioelectron. 2023, 230, 115245. [Google Scholar] [CrossRef] [Scilit]
  32. Er, E.; Sánchez-Iglesias, A.; Silvestri, A.; Arnaiz, B.; Liz-Marzán, L.M.; Prato, M.; Criado, A. Metal Nanoparticles/MoS2 Surface-Enhanced Raman Scattering-Based Sandwich Immunoassay for α-Fetoprotein Detection. ACS Appl. Mater. Interfaces 2021, 13, 8823–8831. [Google Scholar] [CrossRef] [Scilit]
  33. Hu, Q.; Hu, S.; Li, S.; Liu, S.; Liang, Y.; Cao, X.; Luo, Y.; Xu, W.; Wang, H.; Wan, J.; et al. Boronate Affinity-Based Electrochemical Aptasensor for Point-of-Care Glycoprotein Detection. Anal. Chem. 2022, 94, 10206–10212. [Google Scholar] [CrossRef] [Scilit]
  34. Jing, J.; Liu, K.; Jiang, J.; Xu, T.; Xiao, L.; Zhan, X.; Liu, T. Optimally Configured Optical Fiber Near-Field Enhanced Plasmonic Resonance Immunoprobe for the Detection of Alpha-Fetoprotein. Adv. Sci. 2023, 10, 2207437. [Google Scholar] [CrossRef] [Scilit]
  35. Zhang, J.; Tian, Z.; Yao, Y.; Li, Z.; Li, H.; Qiu, S.; Rao, L.; Yan, B.; Wang, K.; Sang, X.; et al. An ultrasensitive microfiber biosensor for fast detection of alpha-fetoprotein in human serum samples. Talanta 2026, 305, 129608. [Google Scholar] [CrossRef] [Scilit]
  36. Niaki, N.M.; Hatefnia, F.; Heidari, M.M.; Tabean, M.; Mobed, A. Alpha-Fetoprotein (AFP) biosensors. Clin. Chim. Acta 2025, 573, 120293. [Google Scholar] [CrossRef] [Scilit]
  37. Shan, C.-W.; Chen, Z.; Han, G.-C.; Feng, X.-Z.; Kraatz, H.-B. Electrochemical immuno-biosensors for the detection of the tumor marker alpha-fetoprotein: A review. Talanta 2024, 271, 125638. [Google Scholar] [CrossRef] [Scilit]
  38. Ramachandran, L.; Abul Rub, F.; Hajja, A.; Alodhaibi, I.; Arai, M.; Alfuwais, M.; Makhzoum, T.; Yaqinuddin, A.; Al-Kattan, K.; Assiri, A.M.; et al. Biosensing of Alpha-Fetoprotein: A Key Direction toward the Early Detection and Management of Hepatocellular Carcinoma. Biosensors 2024, 14, 235. [Google Scholar] [CrossRef] [Scilit]
  39. Zhao, Y.; Liu, Q.; Qin, Y.; Cao, Y.; Zhao, J.; Zhang, K.; Cao, Y. Ordered Labeling-Facilitated Electrochemical Assay of Alpha-Fetoprotein-L3 Ratio for Diagnosing Hepatocellular Carcinoma. ACS Appl. Mater. Interfaces 2023, 15, 6411–6419. [Google Scholar] [CrossRef] [Scilit]
  40. Xing, K.; Wang, C.; Zhang, Q.; Tang, Y.; Yang, X. A signal-on molecularly imprinted electrochemiluminescent biosensor for simultaneous determination of HCC-concerned AFP and PIVKA-II. Sens. Actuators B Chem. 2026, 451, 139278. [Google Scholar] [CrossRef] [Scilit]
  41. Tang, G.; Liu, K.; Xia, L.; Chen, T.; Liman, A.; Feng, Q.; Zhou, R. An integrated pressure-based lateral flow immunoassay for microliter whole-blood AFP detection at the point of care. Biosens. Bioelectron. 2026, 299, 118398. [Google Scholar] [CrossRef] [Scilit]
  42. Lee, M.; Ahmed, S.; Tong, L.; Ulku, M.K.; Chaturvedi, N.; Tripathy, S.; Palanisamy, B.; Negahdary, M.; Hutcheson, J.; Cote, G.; et al. Advances in Whole Blood Separation Techniques for Point-of-Care Testing. ACS Sens. 2025, 10, 9203–9224. [Google Scholar] [CrossRef] [Scilit]
  43. Jesadabundit, W.; Khamcharoen, W.; Jampasa, S.; Chuaypen, N.; Tangkijvanich, P.; Chailapakul, O. DNase I-based signal amplification on sequential flow microfluidic platform for electrochemical detection of alpha-fetoprotein using ruthenium(III)-methylene blue redox cycling. Sens. Actuators B Chem. 2025, 444, 138461. [Google Scholar] [CrossRef] [Scilit]
  44. Land, K.J.; Boeras, D.I.; Chen, X.S.; Ramsay, A.R.; Peeling, R.W. REASSURED diagnostics to inform disease control strategies, strengthen health systems and improve patient outcomes. Nat. Microbiol. 2019, 4, 46–54. [Google Scholar] [CrossRef] [Scilit]
  45. Burrow, D.T.; Heggestad, J.T.; Kinnamon, D.S.; Chilkoti, A. Engineering innovative interfaces for point-of-care diagnostics. Curr. Opin. Colloid Interface Sci. 2023, 66, 101718. [Google Scholar] [CrossRef] [Scilit]
  46. Pum, J. Chapter Six—A Practical Guide to Validation and Verification of Analytical Methods in the Clinical Laboratory. In Advances in Clinical Chemistry; Makowski, G.S., Ed.; Elsevier: Amsterdam, The Netherlands, 2019; pp. 215–281. [Google Scholar]
  47. Xiao, Y.; Chambliss, A.B. A Review of Regulations, Terminology, and Practical Approaches to Reportable Range Verification. J. Appl. Lab. Med. 2023, 8, 773–788. [Google Scholar] [CrossRef] [Scilit]
  48. Mohammadinejad, A.; Kazemi Oskuee, R.; Eivazzadeh-Keihan, R.; Rezayi, M.; Baradaran, B.; Maleki, A.; Hashemzaei, M.; Mokhtarzadeh, A.; de la Guardia, M. Development of biosensors for detection of alpha-fetoprotein: As a major biomarker for hepatocellular carcinoma. TrAC Trends Anal. Chem. 2020, 130, 115961. [Google Scholar] [CrossRef] [Scilit]
  49. Kim, H.; Sohn, A.; Yeo, I.; Yu, S.J.; Yoon, J.-H.; Kim, Y. Clinical Assay for AFP-L3 by Using Multiple Reaction Monitoring–Mass Spectrometry for Diagnosing Hepatocellular Carcinoma. Clin. Chem. 2018, 64, 1230–1238. [Google Scholar] [CrossRef] [Scilit]
  50. Ma, M.; Zhang, M.; Wang, J.; Zhou, Y.; Zhang, X.; Liu, G. Rapid Detection of Alpha-Fetoprotein (AFP) with Lateral Flow Aptasensor. Molecules 2025, 30, 484. [Google Scholar] [CrossRef] [Scilit]
  51. Wang, C.; Hou, F.; Ma, Y. Simultaneous quantitative detection of multiple tumor markers with a rapid and sensitive multicolor quantum dots based immunochromatographic test strip. Biosens. Bioelectron. 2015, 68, 156–162. [Google Scholar] [CrossRef] [Scilit]
  52. Fang, C.C.; Chou, C.C.; Yang, Y.Q.; Wei-Kai, T.; Wang, Y.T.; Chan, Y.H. Multiplexed Detection of Tumor Markers with Multicolor Polymer Dot-Based Immunochromatography Test Strip. Anal. Chem. 2018, 90, 2134–2140. [Google Scholar] [CrossRef] [Scilit]
  53. Liang, R.L.; Xu, X.P.; Liu, T.C.; Zhou, J.W.; Wang, X.G.; Ren, Z.Q.; Hao, F.; Wu, Y.S. Rapid and sensitive lateral flow immunoassay method for determining alpha fetoprotein in serum using europium (III) chelate microparticles-based lateral flow test strips. Anal. Chim. Acta 2015, 891, 277–283. [Google Scholar] [CrossRef] [Scilit]
  54. Chen, R.; Zhou, X.; Wu, Y.; Liu, Q.; Liu, Q.; Huang, J.; Li, F. NIR-II emissive lateral flow immunoassay for accurate determination of tumor marker in hemolysis. Sens. Actuators B Chem. 2021, 328, 129050. [Google Scholar] [CrossRef] [Scilit]
  55. Bolodeoku, J.; Coker, O.; Bains, S.; Anyaeche, C.; Kim, T.K. An evaluation of the performance of the Point of Care Test iCHROMA™ AFP method: Precision and accuracy. J. Glob. Med. 2021, 1, e7. [Google Scholar] [CrossRef] [Scilit]
  56. Lu, L.; Yu, J.; Liu, X.; Yang, X.; Zhou, Z.; Jin, Q.; Xiao, R.; Wang, C. Rapid, quantitative and ultra-sensitive detection of cancer biomarker by a SERRS-based lateral flow immunoassay using bovine serum albumin coated Au nanorods. RSC Adv. 2020, 10, 271–281. [Google Scholar] [CrossRef] [Scilit]
  57. Li, J.; Fang, C.; Yao, Y.; Chen, L.; Lin, B.; Wang, Y.; Guo, L. Rapid antibody conjugation strategy via instant charge inversion of AuNBPs toward ultrasensitive SERS-LFIA detection of AFP. Microchem. J. 2024, 202, 110832. [Google Scholar] [CrossRef] [Scilit]
  58. Xiao, R.; Lu, L.; Rong, Z.; Wang, C.; Peng, Y.; Wang, F.; Wang, J.; Sun, M.; Dong, J.; Wang, D.; et al. Portable and multiplexed lateral flow immunoassay reader based on SERS for highly sensitive point-of-care testing. Biosens. Bioelectron. 2020, 168, 112524. [Google Scholar] [CrossRef] [Scilit]
  59. Wang, W.; Cao, C.; Li, Y.; Zhang, Y.; Huang, H.; Feng, S.; Wen, T.; You, M. Mechanistic modeling of flow-transport-reaction coupling in lateral flow assays: From capillary wicking to physics-AI hybrid design. TrAC Trends Anal. Chem. 2026, 205, 119159. [Google Scholar] [CrossRef] [Scilit]
  60. Lu, Y.; Peng, X.; Yin, Z.; Fan, X.; Fan, J.; Li, Y.M.G. AI-driven lateral flow immunoassay for point-of-care detection of cardiac biomarkers in acute myocardial infarction. Microchim. Acta 2026, 193, 323. [Google Scholar] [CrossRef] [Scilit]
  61. Du, J.; Cao, C.; Xue, Z.; Wang, W.; Lu, X.; Wei, Y.; Huang, J.; Zhao, L.; Wang, L.; Xu, F.; et al. AI-Enhanced Lateral Flow Assay Enables 3-Minute Quantitative Detection with Laboratory-Grade Accuracy. Anal. Chem. 2025, 97, 24196–24208. [Google Scholar] [CrossRef] [Scilit]
  62. Puangpila, C.; Anukulkich, N.; Chiapleam, S.; Intajan, B.; Jakmunee, J.; Pencharee, S. Development of lectin-based lateral flow assay for fucosylated alpha-fetoprotein. J. Cell. Biochem. 2023, 124, 1546–1556. [Google Scholar] [CrossRef] [Scilit]
  63. Jiang, S.; Chen, Y.; Liang, J.; Xiao, H.; Lin, M.; Cui, X.; Zhao, S. An AgPd NP-based lateral flow immunoassay for simultaneous detection of glycocholic acid and alpha-fetoprotein. Anal. Methods 2024, 16, 1508–1514. [Google Scholar] [CrossRef] [Scilit]
  64. Zuo, W.; Meng, X.; Zeng, S.; Wei, W.; Yang, Q.; Song, J.; Hu, Q.; Deng, S.; Zhang, D.; Dai, J.; et al. HepatoAILFA: Machine-Learning-Assisted Nano-enhanced Point-of-Care System for Personalized Precise Diagnosis of Hepatocellular Carcinoma. Nano Lett. 2026, 26, 384–394. [Google Scholar] [CrossRef] [Scilit]
  65. Adedokun, G.; Alipanah, M.; Fan, Z.H. Sample preparation and detection methods in point-of-care devices towards future at-home testing. Lab Chip 2024, 24, 3626–3650. [Google Scholar] [CrossRef] [Scilit]
  66. Xu, X.; Cai, L.; Liang, S.; Zhang, Q.; Lin, S.; Li, M.; Yang, Q.; Li, C.; Han, Z.; Yang, C. Digital microfluidics for biological analysis and applications. Lab Chip 2023, 23, 1169–1191. [Google Scholar] [CrossRef] [Scilit]
  67. Mazzaracchio, V.; Arduini, F. Smart microfluidic devices integrated in electrochemical point-of-care platforms for biomarker detection in biological fluids. Anal. Bioanal. Chem. 2026, 418, 459–481. [Google Scholar] [CrossRef] [Scilit]
  68. Wu, M.; Hartanto, H.; Wu, S.; Jiang, T.; Wang, G.; Chen, T.-H. Visualizing alpha-fetoprotein level in undiluted serum based on microfluidic particle accumulation. Sens. Actuators B Chem. 2023, 390, 133963. [Google Scholar] [CrossRef] [Scilit]
  69. Shen, Y.; Lu, W.; Fu, T.; Zhang, K.; Zhuo, X.; Wang, H.; Liu, C.; Xie, X.; Zhang, Q.; Ma, S. Portable smartphone-based microfluidic electrochemical immunosensor for ultrasensitive detection of alpha-fetoprotein in human serum. Electroanalysis 2023, 35, e202300056. [Google Scholar] [CrossRef] [Scilit]
  70. Zhu, W.; Qian, H.; Cao, S.; Xia, W.; Wang, X.; Jin, J.; Wang, X.; Zhang, H.; Liu, D.; Chen, Y. A new platform of electrowetting-on-dielectric digital microfluidics for rapid detection of early-stage Hepatocellular Carcinoma(HCC) specific biomarker. Anal. Chim. Acta 2025, 1336, 343533. [Google Scholar] [CrossRef] [Scilit]
  71. Sheybanikashani, S.; Zhou, J.; Papautsky, I. Blood microfluidics: Progress and challenges. Lab Chip 2026, 26, 1123–1147. [Google Scholar] [CrossRef] [Scilit]
  72. Thévenot, D.R.; Toth, K.; Durst, R.A.; Wilson, G.S. Electrochemical biosensors: Recommended definitions and classification. Biosens. Bioelectron. 2001, 16, 121–131. [Google Scholar] [CrossRef] [Scilit]
  73. Russo, M.J.; Han, M.; Desroches, P.E.; Manasa, C.S.; Dennaoui, J.; Quigley, A.F.; Kapsa, R.M.I.; Moulton, S.E.; Guijt, R.M.; Greene, G.W.; et al. Antifouling Strategies for Electrochemical Biosensing: Mechanisms and Performance toward Point of Care Based Diagnostic Applications. ACS Sens. 2021, 6, 1482–1507. [Google Scholar] [CrossRef] [Scilit]
  74. Wehmeyer, K.R.; White, R.J.; Kissinger, P.T.; Heineman, W.R. Electrochemical Affinity Assays/Sensors: Brief History and Current Status. Annu. Rev. Anal. Chem. 2021, 14, 109–131. [Google Scholar] [CrossRef] [Scilit]
  75. Wachholz, J.D.; Deroco, P.B.; Hryniewicz, B.M.; Kubota, L.T. Strategies for Electrochemical Point-of-Care Biosensors. Annu. Rev. Anal. Chem. 2025, 18, 307–333. [Google Scholar] [CrossRef] [Scilit]
  76. Liu, L.; Wang, H.; Xie, B.; Zhang, B.; Lin, Y.; Gao, L. Detection of Alpha-Fetoprotein Using Aptamer-Based Sensors. Biosensors 2022, 12, 780. [Google Scholar] [CrossRef] [Scilit]
  77. Lorenzo-Gómez, R.; González-Robles, D.; Miranda-Castro, R.; De-los-Santos-Álvarez, N.; Lobo-Castañón, M.J. On the Electrochemical Detection of Alpha-Fetoprotein Using Aptamers: DNA Isothermal Amplification Strategies to Improve the Performance of Weak Aptamers. Biosensors 2020, 10, 46. [Google Scholar] [CrossRef] [Scilit]
  78. Upan, J.; Youngvises, N.; Tuantranont, A.; Karuwan, C.; Banet, P.; Aubert, P.-H.; Jakmunee, J. A simple label-free electrochemical sensor for sensitive detection of alpha-fetoprotein based on specific aptamer immobilized platinum nanoparticles/carboxylated-graphene oxide. Sci. Rep. 2021, 11, 13969. [Google Scholar] [CrossRef] [Scilit]
  79. Yaiwong, P.; Anuthum, S.; Sangthong, P.; Jakmunee, J.; Bamrungsap, S.; Ounnunkad, K. A new portable toluidine blue/aptamer complex-on-polyethyleneimine-coated gold nanoparticles-based sensor for label-free electrochemical detection of alpha-fetoprotein. Front. Bioeng. Biotechnol. 2023, 11, 1182880. [Google Scholar] [CrossRef] [Scilit]
  80. Chanarsa, S.; Jakmunee, J.; Ounnunkad, K. A sandwich-like configuration with a signal amplification strategy using a methylene blue/aptamer complex on a heterojunction 2D MoSe2/2D WSe2 electrode: Toward a portable and sensitive electrochemical alpha-fetoprotein immunoassay. Front. Cell. Infect. Microbiol. 2022, 12, 916357. [Google Scholar] [CrossRef] [Scilit]
  81. Sun, H.-N.; Wang, M.; Tan, H.-S.; Liu, H.-P.; Liu, M.; Li, S.-S. Ratiometric electrochemical immunoassay based on 2D Co/Fe MOF decorated with toluidine blue and Fc-labeled Schiff base for accurate assay of alpha-fetoprotein in clinical serum. Talanta 2024, 273, 125876. [Google Scholar] [CrossRef] [Scilit]
  82. Li, J.; Liang, X.; Zhong, R.; Liu, M.; Liu, X.; Yan, H.-L.; Zhou, Y.-G. Clinically Applicable Homogeneous Assay for Serological Diagnosis of Alpha-Fetoprotein by Impact Electrochemistry. ACS Sens. 2022, 7, 3216–3222. [Google Scholar] [CrossRef] [Scilit]
  83. Li, Y.; Huang, R.; Duan, Y.; Deng, D.; Chen, H.; Xia, T.; Duan, Y.; Lei, H.; Luo, L. Ultrasensitive lab-on-paper electrochemical device via heterostructure copper/cuprous sulfide@N-doped C@Au hollow nanoboxes as signal amplifier for alpha-fetoprotein detection. Biosens. Bioelectron. 2025, 267, 116827. [Google Scholar] [CrossRef] [Scilit]
  84. Li, J.; Gao, T.; Gu, S.; Zhi, J.; Yang, J.; Li, G. An electrochemical biosensor for the assay of alpha-fetoprotein-L3 with practical applications. Biosens. Bioelectron. 2017, 87, 352–357. [Google Scholar] [CrossRef] [Scilit]
  85. Tu, W.; Liu, H.; Wang, J.; Wang, Y.; Wang, Z.; Dai, Z. Using signal off-to-on strategy for designing precise and ultrasensitive biosensor towards hepatocellular carcinoma through protein variant detection based on biocompatible bimetallic MOF. Biosens. Bioelectron. 2025, 280, 117429. [Google Scholar] [CrossRef] [Scilit]
  86. García-Ramírez, J.I.; Luna-Cervantes, M.; Izaguirre-Hernández, I.Y.; Hernández-Torres, J.; Juárez-Aguilar, E.; Thomas-Dupont, P.; Remes-Troche, J.M.; Zamora-Peredo, L. SERS-Based Immunoassay for α-Fetoprotein Biomarker Detection Using an Au-Ag Nanostars Platform. Biosensors 2025, 15, 632. [Google Scholar] [CrossRef] [Scilit]
  87. Hojjat Jodaylami, M.; Masson, J.-F.; Badia, A. Surface plasmon resonance sensing. Nat. Rev. Methods Primers 2025, 5, 47. [Google Scholar] [CrossRef] [Scilit]
  88. Li, R.; Fan, H.; Zhou, H.; Chen, Y.; Yu, Q.; Hu, W.; Liu, G.L.; Huang, L. Nanozyme-Catalyzed Metasurface Plasmon Sensor-Based Portable Ultrasensitive Optical Quantification Platform for Cancer Biomarker Screening. Adv. Sci. 2023, 10, 2301658. [Google Scholar] [CrossRef] [Scilit]
  89. Huang, Y.; Ren, J.; Qu, X. Nanozymes: Classification, Catalytic Mechanisms, Activity Regulation, and Applications. Chem. Rev. 2019, 119, 4357–4412. [Google Scholar] [CrossRef] [Scilit]
  90. Tang, S.; Cai, J.; Zhou, K.; Mei, Z.; Huang, D.; Liu, L.; Yang, L.; Yin, D.; Hu, L. Cu-MOFs@AuPtNPs nanozyme-based immunosorbent assay for colorimetric detection of alpha-fetoprotein. Anal. Methods 2024, 16, 6443–6450. [Google Scholar] [CrossRef] [Scilit]
  91. Roda, A.; Mirasoli, M.; Michelini, E.; Di Fusco, M.; Zangheri, M.; Cevenini, L.; Roda, B.; Simoni, P. Progress in chemical luminescence-based biosensors: A critical review. Biosens. Bioelectron. 2016, 76, 164–179. [Google Scholar] [CrossRef] [Scilit]
  92. Liu, J.; Wang, X.; Sun, Y.; Luo, C. A novel chemiluminescence sensor for alpha-fetoprotein detection based on an aptamer–luminol modified magnetic graphene oxide and copper-based MOF composite. Anal. Methods 2024, 16, 5723–5732. [Google Scholar] [CrossRef] [Scilit]
  93. Marsh, T.L.; Parikh, N.D.; Roberts, L.R.; Schwartz, M.E.; Nguyen, M.H.; Befeler, A.; Page-Lester, S.; Tayob, N.; Srivastava, S.; Rinaudo, J.A.; et al. A Phase 3 Biomarker Validation of GALAD for the Detection of Hepatocellular Carcinoma in Cirrhosis. Gastroenterology 2025, 168, 316–326.e6. [Google Scholar] [CrossRef] [Scilit]
  94. Liu, L.; Wang, H.; Sulemana, H.; Xie, B.; Gao, L. Detection of Alpha Fetoprotein Based on AIEgen Nanosphere Labeled Aptamer Combined with Sandwich Structure of Magnetic Gold Nanocomposites. Biosensors 2023, 13, 351. [Google Scholar] [CrossRef] [Scilit]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Article Metrics

Citations

Article Access Statistics

Multiple requests from the same IP address are counted as one view.