Abstract
Accurate transfer of implant position and maxillomandibular relationship determines complete-arch prosthesis fit. This in vitro study compared three intraoral scanners and two photogrammetry-based workflows (intraoral and extraoral), each including intraoral surface scanning and data merging. Edentulous maxillary and mandibular models (four implants each) were mounted on a semi-adjustable articulator, with the maxillomandibular relationship fixed; two operators (2 and 12 years of experience) performed 10 repetitions per system (100 records); an industrial scanner served as reference. Trueness was assessed as RMS deviation, inter-implant and inter-arch distances, and scan body angular deviation (Bonferroni-corrected two-way ANOVA on Yeo–Johnson-transformed data; α = 0.05). The recording system affected RMS deviation (F(4,90) = 49.67, p < 0.0001), with one scanner highest (369 ± 17 µm, p < 0.001); all but one within-arch distance, all four between-arch distances and all eight angular deviations differed among systems (p ≤ 0.010). Operator experience affected five distances (p ≤ 0.035) but not overall deviation (p = 0.665) or any angular variable (p ≥ 0.089), with interactions for overall deviation, five distances and two angular variables (p ≤ 0.036). Both photogrammetry workflows required a surface record and merge; their values describe the complete workflow, not photogrammetry alone. Precision and clinical outcomes were not assessed.