Abstract
Background/Objectives: Continuous non-invasive core temperature monitoring may benefit intensive care unit (ICU) patients, but the performance of Moni-Patch™, a wireless sensor based on a compensated heat-flux method, has not been evaluated in critically ill patients. We assessed its agreement, trending ability, and fever detection against pulmonary artery blood temperature. Methods: In this single-center retrospective study, we analyzed one-minute data from adult ICU patients in whom pulmonary artery blood, bladder, zero-heat-flux (Bair Hugger™), and Moni-Patch™ temperatures were recorded simultaneously. Agreement was assessed using mixed-effects Bland–Altman analysis for repeated measures, with patient-level bootstrap confidence intervals. Trending ability (four-quadrant and polar analyses of 60 min changes) and fever detection (≥38.0 °C) were also evaluated. Results: Twenty-five patients contributed 14,705 paired measurements. Moni-Patch™ showed a bias of −0.30 °C (95% limits of agreement, −1.10 to +0.51 °C), with 72.2% of readings within ±0.5 °C. Its mean absolute error did not differ significantly from that of Bair Hugger™ (difference, −0.08 °C; 95% CI, −0.22 to 0.08), whereas bladder temperature agreed more closely (bias, +0.18 °C; limits, −0.22 to +0.58 °C). Directional concordance was 90.6% (77.3% on intervals defined by blood temperature changes). A reading ≥38.0 °C detected fever with a sensitivity of 36.6% and a specificity of 98.6%. Conclusions: Moni-Patch™ showed limited trending ability, and its error did not differ significantly from that of a zero-heat-flux thermometer. However, given its systematic underestimation and wide limits of agreement, caution is warranted when using it to replace invasive core temperature monitoring or as the sole means of detecting fever.
1. Introduction
Body temperature is a basic vital sign in the intensive care unit (ICU). Both fever and hypothermia are associated with poor outcomes in critically ill patients, and temperature measurements guide the evaluation of infection and decisions on temperature management [1,2,3]. After major surgery, such as cardiovascular surgery, rewarming from intraoperative hypothermia and the onset of postoperative fever also need to be followed continuously [4].
Pulmonary artery blood temperature (hereafter, blood temperature) is regarded as the reference standard for core temperature, but it requires a pulmonary artery catheter and is therefore available only in a limited number of patients [5,6]. Alternative sites such as the bladder and esophagus are also invasive to varying degrees, and their readings are affected by urine output and probe position [6,7,8]. A non-invasive method that can estimate core temperature continuously without burdening the patient is therefore needed.
Zero-heat-flux (ZHF) thermometers have shown generally good agreement with reference temperatures in the perioperative setting [9,10,11,12]. In the ICU, however, their accuracy has been reported to decrease because of reduced peripheral perfusion, rapid changes in body temperature, and patient movement [13,14,15,16]. The Moni-Patch™ Deep Body Temperature Sensor System (Murata Manufacturing Co., Ltd., Kyoto, Japan) is a new device developed from heat-flux thermometry [17]. Unlike conventional ZHF devices, which heat the skin, it uses a compensated heat-flux method: several thermistors embedded in a small, flexible sensor attached to the neck are used to calculate core temperature from both the heat flux and the heat retained in the sensor. Its performance in critically ill patients, however, has not yet been evaluated. In addition, the clinical value of a continuous monitor depends on more than agreement in absolute values. The ability to follow the direction and magnitude of temperature changes (trending ability), including during rewarming, active cooling, and the onset of fever, when skin-based sensors may lag behind core temperature, and to detect fever is also important, but these properties have been reported less consistently than bias and limits of agreement. Existing systematic reviews have likewise focused mainly on bias and the 95% limits of agreement (LoA) [6,14].
We evaluated Moni-Patch™ as a non-invasive monitor that could supplement core temperature monitoring, particularly when invasive monitoring is unavailable, rather than as a replacement for it. The primary aim of this study was to assess the agreement between Moni-Patch™ and blood temperature in ICU patients. The secondary aims were to evaluate trending ability, based on temperature changes, and fever detection, and to compare Moni-Patch™ with bladder temperature and Bair Hugger™. To our knowledge, this is the first evaluation of Moni-Patch™ in critically ill patients, and it examines agreement in absolute values, trending ability, and fever detection in the same patients.
2. Materials and Methods
2.1. Study Design and Ethics
This single-center retrospective observational study was conducted in the eight-bed ICU of Yokohama City University Hospital, Japan. Data recorded during routine clinical care were anonymized before analysis. The study was approved by the hospital’s ethics committee (approval number F250700030; 12 August 2025) and was retrospectively registered, after data collection and before data analysis, with the University Hospital Medical Information Network Clinical Trials Registry (UMIN000059740; 11 November 2025). The requirement for individual informed consent was waived by the ethics committee because of the retrospective design of the study, in accordance with the Ethical Guidelines for Life Science and Medical Research Involving Human Subjects (Japan). Instead, an opt-out procedure was used: information about the study was publicly disclosed, and patients were given the opportunity to refuse the use of their data. No patient declined participation. The study is reported in accordance with the Guidelines for Reporting Reliability and Agreement Studies (GRRAS) [18], and the fever-detection analysis in accordance with STARD 2015 [19].
2.2. Patients
All patients aged 18 years or older who were admitted to the ICU between July 2024 and April 2025 and in whom blood, bladder, Bair Hugger™, and Moni-Patch™ temperatures were measured simultaneously and continuously were included. Pulmonary artery catheters were inserted solely on clinical indications for hemodynamic management, independently of this study. No exclusion criteria were applied.
2.3. Temperature Measurements
Blood temperature was measured with a thermodilution pulmonary artery catheter (Swan–Ganz CCO/CEDV, 7.5 Fr; Edwards Lifesciences, Tokyo, Japan) inserted via the internal jugular vein and was recorded continuously with the HemoSphere™ monitoring platform (Edwards Lifesciences, Irvine, CA, USA). Bladder temperature was measured continuously with a urinary catheter fitted with a temperature sensor and incorporated into a standard urine collection system (AlphaMed™; Covidien Japan, Tokyo, Japan).
The Bair Hugger™ temperature monitoring system (Solventum, Eagan, MN, USA) is based on the ZHF principle. A heater in the sensor warms the skin surface until heat flow from the skin to the environment is zero, so that the temperature of the tissue beneath the skin approaches core temperature [10]. The sensors were applied to the forehead, in accordance with the manufacturer’s instructions, by intensivists experienced in the use of the device.
The Moni-Patch™ Deep Body Temperature Sensor System is a heat-flux device that does not heat the skin [17]. In heat-flux thermometry, the heat flux through the skin surface is derived from the temperature difference across a material of known thermal resistance, and core temperature is estimated from skin temperature and this heat flux. Moni-Patch™ uses an extension of this approach, the compensated heat-flux method: multiple thermistors embedded in a flexible sensor are used to calculate both the heat flux and the heat retained in the sensor itself, and the two are combined to estimate core temperature. This design is intended to reduce the influence of external thermal disturbance and to allow stable continuous measurement without a separate heating unit.
The sensor measures 65 × 45 mm, is 10 mm thick, weighs 12 g, and is attached to the skin of the neck (Figure 1). Readings are transmitted via Bluetooth to a receiver and recorded on a monitor. One receiver and two sensors were used during the study period; which sensor was used in each patient could not be determined retrospectively. To avoid interference, the sensor was placed on the side of the neck opposite the insertion site of the central venous and pulmonary artery catheters. The sensors were applied and measurements started by intensivists trained in the use of the system. Moni-Patch™ and Bair Hugger™ readings were displayed to the clinical team and may have influenced clinical decisions, including the initiation of surface cooling.
Figure 1.
The Moni-Patch™ system. (A) Sensor; (B) receiver; (C) sensor attached to the neck. The sensor measures 65 × 45 mm, is 10 mm thick, weighs 12 g, and is attached to the skin of the neck. Readings are transmitted via Bluetooth to the receiver and recorded on a monitor.
ICU nurses who had received standardized training in the use of all temperature monitors routinely checked sensor attachment, signal stability, and detachment throughout the measurement period. Surface cooling was performed with ice packs placed in the axillary region on the attending physician’s instructions, away from the Moni-Patch™ sensor on the neck; the timing of application was not recorded in a time-resolved manner.
2.4. Data Acquisition and Processing
All four temperatures were recorded at one-minute intervals in the ICU clinical information system. All temperatures were transmitted to the ICU patient monitor, and the monitor’s clock was used as the time of measurement; the values recorded on the monitor every minute were collected. To ensure technically valid readings, data from periods of temporary device removal and from periods when the patient was outside the ICU were excluded. Because non-invasive heat-flux devices need time to stabilize after application, data from the first 10 min after sensor application were also excluded, following previous studies [20,21].
For each patient, up to 1000 time points at which all four readings were available, counted from the start of measurement, were analyzed. This upper limit was set so that long recordings from a few patients would not dominate the results. Data were counted from the start of measurement because the early phase of ICU care is when continuous core temperature monitoring is most needed and when pulmonary artery catheters are generally in place. Temperatures were recorded with a resolution of 0.1 °C. Age, sex, body size, primary diagnosis on ICU admission, and main treatments were extracted retrospectively from the electronic medical records.
2.5. Statistical Analysis
All analyses were performed in R (version 4.3.3; R Foundation for Statistical Computing, Vienna, Austria), with blood temperature as the reference. Mixed models were fitted with lme4 (version 1.1-35.1), receiver operating characteristic (ROC) analyses were performed with pROC (version 1.18.5), and figures were drawn with ggplot2 (version 3.4.4).
Agreement. To account for repeated measurements within patients, bias was estimated with a linear mixed model with the difference (each method—blood temperature) as the dependent variable and a random intercept for patient [22,23]. The LoA were calculated as bias ± 1.96 × √(between-patient variance + within-patient variance). Bias was estimated as the fixed intercept of the mixed model rather than as the simple mean of all measurements. In this estimate, the contribution of each patient is weighted by the between- and within-patient variances as well as by the number of measurements, so it does not equal the simple mean weighted by the number of measurements. All 95% confidence intervals (CIs) in this study were obtained by non-parametric bootstrapping to account for correlation between measurements within the same patient; patients were resampled with replacement (1000 replicates), and the percentile method was used. Proportional bias was assessed with a mixed model in which the difference was regressed on the mean of the two methods. Because this approach can produce a spurious slope when the measurement errors of the two methods differ, we also estimated the within-patient slope of each method against blood temperature, after confirming with Grubbs estimators that the error variance of blood temperature was negligible. As supplementary indices, the proportion of readings within ±0.5 °C (|difference| ≤0.5 °C, calculated from values rounded to 0.1 °C) [9,10,24,25,26], the mean absolute error (MAE), and Lin’s concordance correlation coefficient (CCC) [27] were calculated. Differences in MAE between methods were assessed with the bootstrap CI and with the Wilcoxon signed-rank test applied to per-patient MAE values.
Acceptance criteria. The clinical acceptance criterion (LoA within ±0.5 °C) and the descriptive guide for bias (±0.3 °C) were set before data analysis began, although they were not specified in the study protocol. This threshold follows the value widely used in previous agreement studies of thermometers and was chosen to allow comparison with those studies [9,20,21,28]. Clinically, an error of 0.5 °C exceeds the difference between commonly used fever thresholds (38.0 and 38.3 °C) [1] and could change the classification of fever or hypothermia; the criterion should therefore be regarded as pragmatic rather than stringent. For bias, |bias| ≤ 0.3 °C was used as a descriptive guide rather than an acceptance criterion, for two reasons: 0.3 °C is the difference between the fever thresholds of 38.0 °C used in the systemic inflammatory response syndrome (SIRS) criteria [29] and 38.3 °C used in ICU guidelines [1], and a device with |bias| > 0.3 °C could meet the ±0.5 °C LoA criterion only if the SD of the differences were <0.1 °C, comparable to the measurement resolution. The comparison of the Moni-Patch™ LoA with this criterion was the only confirmatory analysis; all other analyses were exploratory.
Analysis by temperature range. Measurements were divided into three strata according to the simultaneous blood temperature: ≤36.0 °C, 36.1–37.9 °C, and ≥38.0 °C [4,15].
Per-patient analysis. Bias, LoA, and the proportion of readings within ±0.5 °C were calculated for each patient.
Trending ability. For each patient, changes in each method and in blood temperature over 60 min (30 min in a secondary analysis) were calculated. Intervals were defined by clock time rather than as 60 consecutive measurements. Two time points exactly 60 min apart with readings at both ends formed one interval, and the end point of each interval was used as the start point of the next, so that intervals did not overlap. If no reading was available at the end point, the start point was moved to the next measurement; missing readings within an interval were allowed. The number of intervals therefore does not equal the number of measurements per patient divided by 60. Intervals in which the mean change of the two methods was less than 0.2 °C (two steps of the measurement resolution) were excluded. The concordance rate for the direction of change was calculated with four-quadrant analysis, and the angular bias, radial LoA, and proportion of changes within ±30° were calculated with polar plot analysis [30,31]. To compare all methods on a common set of intervals, intervals were also defined by a change in blood temperature of ≥0.2 °C, and exclusion thresholds of 0.1 and 0.3 °C were examined. Changes were calculated from values rounded to 0.1 °C, and exact (Clopper–Pearson) CIs were calculated for concordance rates. Because the polar-plot thresholds were proposed for cardiac output monitors and their clinical relevance for temperature has not been established, they were used descriptively.
Fever detection. Fever was defined as a blood temperature ≥38.0 °C. For each method, the area under the ROC curve (AUC) was calculated, together with the sensitivity, specificity, positive predictive value, and negative predictive value of a reading ≥38.0 °C. At the patient level, the proportion of febrile patients in whom the device reading reached ≥38.0 °C at least once during fever was calculated. The proportion of febrile time during which each device read below 38.0 °C (missed febrile time) was also calculated. Device-specific thresholds were derived with the Youden index and validated by leave-one-patient-out cross-validation. A threshold of 38.3 °C, used to define new fever in ICU guidelines [1], was examined as a sensitivity analysis; 38.0 °C was used as the primary threshold because a temperature above 38 °C is the fever criterion of the SIRS criteria [29] and of the Centers for Disease Control and Prevention/National Healthcare Safety Network (CDC/NHSN) surveillance definitions of postoperative surgical site infection [32], both relevant to our predominantly postoperative cohort, and because the same threshold was used in a previous validation of non-invasive thermometers in cardiac surgery ICU patients [4].
Sensitivity analyses. The following sensitivity analyses were performed: (1) down-sampling at 10 min intervals, in which each patient’s paired data were divided into 10 min time blocks and the first recorded time point in each block was used; (2) equal weighting of patients; (3) exclusion of the first 60 min of each record; (4) exclusion of patients on veno-arterial extracorporeal membrane oxygenation (VA-ECMO); (5) exclusion of patients who received surface cooling; (6) exclusion of both (4) and (5); and (7) leave-one-patient-out reanalysis. We also examined (8) agreement by time since the start of the record (0–4, 4–8, 8–12, 12–16, and >16 h); (9) models allowing patient-specific and temperature-dependent residual variance and serial correlation (exponential correlation structure fitted to 10 min data), with bias-corrected and accelerated (BCa) bootstrap intervals (5000 replicates); (10) time lag, by cross-correlation of 10 min changes and by shifting device readings by −10 to +40 min; (11) stratification by the mean of the two methods and by bladder temperature; and (12) an exploratory mixed model of the Moni-Patch™ error with the rate of temperature change, time since the start of the record, blood temperature, and surface cooling as covariates.
Confidence intervals. The 95% CIs for the proportion within ±0.5 °C, MAE, CCC, differences in MAE between methods, concordance rate, angular bias, proportion within ±30°, AUC, sensitivity, and specificity were obtained with the same patient-level bootstrap as those for bias and LoA.
Sample size. Because this was a retrospective study, the number of patients was determined by the number of eligible patients during the study period. As the precision of the LoA depends mainly on the number of patients, the uncertainty of the estimates is presented as 95% CIs.
Use of generative AI. Generative AI (Claude Opus 5.5, Anthropic) was used to assist with English editing, preparation of figures and tables, and writing R code for the statistical analysis. The authors specified all analyses, verified the code by re-running it and reproducing the key estimates, and take full responsibility for the content.
3. Results
3.1. Patient Characteristics
During the study period, 643 patients were admitted to the ICU, of whom 554 were aged 18 years or older. Of these, 25 patients in whom pulmonary artery blood, bladder, Bair Hugger™, and Moni-Patch™ temperatures were all measured were included; the remaining 529 patients did not have all four measurements. The median age was 68 years (interquartile range [IQR], 61–76), 19 patients (76.0%) were male, and the median Acute Physiology and Chronic Health Evaluation (APACHE) II score was 21 (IQR, 16–25) (Table 1). Twenty-three patients (92.0%) were admitted after cardiovascular surgery; 23 received mechanical ventilation and 24 received catecholamines. One patient was treated with VA-ECMO, and six (24.0%) received surface cooling. A total of 14,705 paired temperature measurements were analyzed, with a median of 578 (IQR, 335–946) per patient. The upper limit of 1000 time points was reached in two patients; all eligible data were analyzed in the others. Blood temperature ranged from 35.5 to 39.3 °C, and 16 patients (64.0%) developed fever (≥38.0 °C) during the analysis period. The analyzed records spanned a median of 14.3 h (IQR, 7.0–18.0), with readings available for 87.9% (IQR, 82.1–98.4) of this period; gaps >60 min occurred in 10 patients (maximum, 12.3 h).
Table 1.
Patient characteristics (n = 25).
3.2. Agreement with Blood Temperature
For Moni-Patch™, the bias was −0.30 °C (95% CI, −0.43 to −0.16), and the LoA were −1.10 °C (95% CI, −1.29 to −0.85) to +0.51 °C (95% CI, 0.33 to 0.66), which did not meet the predefined acceptance criterion (Table 2, Figure 2). The proportion of readings within ±0.5 °C was 72.2% (95% CI, 56.3–87.6), the MAE was 0.40 °C, and the CCC was 0.70. Variability in the error was larger between patients (standard deviation [SD], 0.33 °C) than within patients (SD, 0.24 °C). Within-patient residual SD increased with temperature (0.14 °C at 36.5 °C to 0.34 °C at 38.5 °C), and residuals were strongly autocorrelated; nevertheless, models allowing patient-specific or temperature-dependent variance and serial correlation gave LoA of −1.14 to −1.09 °C and +0.48 to +0.51 °C, and 96.2% of readings fell within the LoA (Table A4, Figure A2). The difference showed a positive slope against the mean of the two methods (0.19 °C/°C; 95% CI, 0.08–0.35); however, the within-patient slope of Moni-Patch™ against blood temperature was 1.03 (95% CI, 0.91–1.14), indicating that this slope largely reflects the greater measurement error of the device rather than true proportional bias (Table A5).
Table 2.
Agreement with pulmonary artery blood temperature (14,705 pairs, 25 patients).
Figure 2.
Bland–Altman plots for repeated measurements. (A) Moni-Patch™; (B) Bair Hugger™; (C) bladder temperature. Shading indicates the density of paired measurements in 0.1 °C × 0.1 °C bins. Black solid lines indicate the bias; dashed lines, the 95% LoA from the mixed model; orange lines, the regression line of the difference on the mean (see text for interpretation); and gray bands, ±0.5 °C. Per-patient plots are shown in Figure A4. LoA, limits of agreement.
For Bair Hugger™, the bias was −0.39 °C, the LoA were −1.28 to +0.50 °C, and 65.5% of readings were within ±0.5 °C. Bladder temperature showed the closest agreement, with a bias of +0.18 °C, LoA of −0.22 to +0.58 °C, and 97.8% of readings within ±0.5 °C.
The difference in MAE (Moni-Patch™ − Bair Hugger™) was −0.08 °C (95% CI, −0.22 to 0.08), and the patient-level comparison also showed no significant difference (p = 0.40). The MAE of Moni-Patch™ was 0.17 °C (95% CI, 0.03–0.31) larger than that of bladder temperature (p = 0.013) (Table A3).
3.3. Agreement by Temperature Range
In the ≥38.0 °C stratum (3883 pairs, 16 patients), the bias of Moni-Patch™ was −0.44 °C, and the proportion of readings within ±0.5 °C decreased to 60.0% (Table 3). Bair Hugger™ showed a similar decline (−0.46 °C, 50.9%), whereas the agreement of bladder temperature was maintained (+0.20 °C, 99.8%). In the 36.1–37.9 °C stratum, the bias of Moni-Patch™ was −0.25 °C, and 76.2% of readings were within ±0.5 °C. The ≤36.0 °C stratum included only 3 patients and 169 pairs, and these results are presented descriptively.
Table 3.
Agreement with pulmonary artery blood temperature by temperature range.
3.4. Per-Patient Agreement
The per-patient bias of Moni-Patch™ ranged from −0.93 to +0.34 °C, and |bias| was ≤0.5 °C in 18 of the 25 patients (Figure 3, Table A1). The 95% prediction interval for the bias in a new patient was −1.00 to +0.40 °C, and per-patient LoA were within ±0.5 °C in 6 of 25 patients (Bair Hugger™, 8; bladder temperature, 16). The proportion of readings within ±0.5 °C varied widely between patients, from 13.5% to 100% (median, 90.7%). For Bair Hugger™, the per-patient bias ranged from −1.22 to +0.18 °C, and |bias| was ≤0.5 °C in 14 patients. Bladder temperature agreed well in all patients except the patient on VA-ECMO (bias, −0.50 °C).
Figure 3.
Bias and 95% LoA in individual patients. Dots indicate the bias for each patient, horizontal lines the 95% LoA, and gray bands ±0.5 °C. The patient on VA-ECMO and the patients who received surface cooling are indicated. LoA, limits of agreement; VA-ECMO, veno-arterial extracorporeal membrane oxygenation.
3.5. Trending Ability
In total, 238 intervals of 60 min were obtained from the 25 patients; 96 (40.3%; 23 patients) were analyzed for Moni-Patch™, 93 for Bair Hugger™, and 82 for bladder temperature (Table 4, Figure 4). The concordance rate was 90.6% (95% CI, 82.2–97.8) for Moni-Patch™, 91.4% for Bair Hugger™, and 96.3% for bladder temperature. The angular bias was +13.6° (95% CI, 8.7–19.3), +12.4°, and +3.5°, and the proportions within ±30° were 76.0%, 83.9%, and 91.5%, respectively. On the common set of 97 intervals defined by blood temperature, concordance was 77.3% (95% CI, 68.1–85.9) for Moni-Patch™, 85.6% for Bair Hugger™, and 86.6% for bladder temperature, and the angular bias was close to zero for all methods (+0.8°, −0.2°, and −5.6°) (Table A7). For 30 min intervals, the concordance rate of Moni-Patch™ was 75.0%.
Table 4.
Trending ability.
Figure 4.
Trending ability for 60 min changes. (A–C) Four-quadrant plots; gray dots indicate changes within the exclusion zone (mean change <0.2 °C); the number of analyzed and excluded intervals is shown for each method, and the analyzed subsets differ between methods. (D–F) Polar plots; dashed lines indicate ±30°.
3.6. Fever Detection
Fever was present in 3883 pairs from 16 patients (Table 5, Figure 5). The AUC of Moni-Patch™ was 0.89 (95% CI, 0.81–0.94), but with a reading ≥38.0 °C as the threshold, its sensitivity was 36.6% (95% CI, 21.3–48.0) and its specificity was 98.6%. The sensitivity of Bair Hugger™ was also low (33.0%). Of the 16 febrile patients, a reading ≥38.0 °C was recorded at least once during fever in 9 patients with Moni-Patch™, 8 with Bair Hugger™, and 16 with bladder temperature. Moni-Patch™ read below 38.0 °C during 63.4% of febrile time (Bair Hugger™, 67.0%; bladder temperature, 0.8%). With the cross-validated threshold (37.3 °C), sensitivity was 89.2% (95% CI, 76.7–96.5) and specificity 69.5% (56.0–81.4); all 16 febrile patients were detected, and missed febrile time decreased to 10.8%. With fever defined as ≥38.3 °C, the sensitivity of Moni-Patch™ was 28.9% and specificity 99.2% (Table A8). In an exploratory episode-level analysis, the cross-validated threshold detected all 29 febrile episodes, compared with 17 (58.6%) at 38.0 °C, but false alarms increased from 0.6 to 1.9 per 24 h (Table A10).
Table 5.
Detection of fever (pulmonary artery blood temperature ≥38.0 °C).
Figure 5.
Receiver operating characteristic curves for fever detection. Fever was defined as a blood temperature ≥38.0 °C. Circles indicate the operating point at a device reading ≥38.0 °C and triangles at the cross-validated device-specific threshold, annotated with sensitivity and specificity.
3.7. Sensitivity Analyses
Across all sensitivity analyses, the bias of Moni-Patch™ ranged from −0.32 to −0.27 °C, and the proportion of readings within ±0.5 °C ranged from 69.9% to 76.4% (Table A2, Figure A1). After exclusion of the patients on VA-ECMO and those who received surface cooling, the difference in MAE between Moni-Patch™ and Bair Hugger™ remained nonsignificant (−0.08 °C; 95% CI, −0.26 to 0.09; patient-level bootstrap). After exclusion of the patient on VA-ECMO, the LoA for bladder temperature narrowed to −0.08 to +0.50 °C. By time since the start of the record, the bias of Moni-Patch™ was −0.27 °C at 0–4 h (25 patients; 80.0% within ±0.5 °C), −0.43 to −0.44 °C at 4–12 h, and −0.30 °C after 12 h, with no linear trend over time (Table A6). Shifting the device readings by −10 to +40 min did not reduce the within-patient SD of Moni-Patch™ (0.24–0.27 °C), and the cross-correlation of 10 min changes peaked at 0–2 min, whereas bladder temperature lagged blood temperature by approximately 25 min (Figure A3). The bias of Moni-Patch™ was similar during rising (>0.3 °C/h), stable, and falling (<−0.3 °C/h) phases (−0.29, −0.30, and −0.30 °C) (Table A9). In the exploratory model, none of the rate of temperature change, time since the start of the record, blood temperature, or surface cooling was associated with the Moni-Patch™ error (Table A9).
4. Discussion
To our knowledge, this is the first study to compare the performance of Moni-Patch™ in critically ill patients simultaneously with blood temperature, bladder temperature, and a ZHF thermometer. Moni-Patch™ read on average approximately 0.3 °C lower than blood temperature, and its LoA spanned approximately 1.6 °C, failing to meet the predefined acceptance criterion. Its error did not differ significantly from that of the existing ZHF thermometer and was larger than that of bladder temperature. The direction of 60 min changes agreed in 90.6% of intervals with a mean change ≥0.2 °C, but in 77.3%, intervals were defined by blood temperature changes.
Clinical implications. An average negative bias of approximately 0.3 °C, present in 20 of 25 patients, is large enough to alter clinical decisions near the threshold used to define fever. Of the measurements with a blood temperature ≥ 38.0 °C, Moni-Patch™ read ≥38.0 °C in only 37%, and in about 40% of febrile patients it never reached 38.0 °C during fever. Because fever prompts evaluation for infection [1], these missed episodes are an important caveat for clinical use. Missed fever may delay the recognition of postoperative infection, prevent identification of clinically relevant thresholds, and add uncertainty during active temperature management. The discrepancy between the relatively high AUC (0.89) and the low sensitivity mainly reflects miscalibration due to the negative bias: a cross-validated threshold of 37.3 °C reduced the missed febrile time from 63% to 11%, at the cost of lower specificity (69.5%). This exploratory threshold, which also increased false alarms (Table A10), was derived from the same cohort and requires external validation before clinical use. In addition, the bias varied considerably between patients (−0.93 to +0.34 °C), so a uniform correction is unlikely to eliminate the error in individual patients. A model allowing a patient-specific slope gave wider LoA (−1.27 to +0.76 °C at 37.5 °C; Table A4), because the slope of a new patient is also unknown; as these slopes were estimated over narrow temperature ranges and partly reflect measurement noise, this finding further indicates that the error in an individual patient cannot be predicted from pooled estimates. In a previous study of cardiac surgery ICU patients, a ZHF sensor also underestimated core temperature and detected hyperthermia with a sensitivity of 0.35 [4], similar to our findings.
Trending ability. Moni-Patch™ did not meet the polar-plot criteria. Its concordance of 90.6% was obtained in intervals selected by the device’s own changes; on intervals defined by blood temperature, concordance fell to 77.3% and the positive angular bias disappeared, suggesting that the apparent overstatement of changes reflected interval selection and measurement noise. In addition, about 60% of hourly intervals showed changes <0.2 °C, so trending results apply only to the minority of periods with meaningful change. Because no established criteria exist for temperature trends, these thresholds [30,31] were used descriptively.
Reduced accuracy at high temperatures. At high temperatures, the negative bias increased for both Moni-Patch™ and Bair Hugger™. Similar findings have been reported for other non-invasive thermometers [4,13,15]. Part of this pattern may reflect regression to the mean, because the strata were defined by the reference: stratified by the mean of the two methods, the bias of Moni-Patch™ at ≥38.0 °C was −0.24 °C, although it remained −0.43 °C when stratified by bladder temperature (Table A5). Peripheral vasoconstriction during the rising phase of fever could increase the skin–core gradient [13,14,15,16], but this explanation is speculative. However, we did not directly assess peripheral perfusion, sweating, or the phase of fever, and the mechanism remains unclear.
Comparison with Bair Hugger™. Moni-Patch™ showed numerically slightly better bias, LoA, and proportion of readings within ±0.5 °C, but the patient-level analysis showed no significant difference. Because treating individual measurements as independent can make even small differences statistically significant, we based our inference on the patient as the unit of analysis. However, with only 25 patients, the study may have lacked the power to detect a difference, and these results do not demonstrate equivalence between the two devices. Bair Hugger™ performed worse than in previous ICU studies (bias −0.12 °C, LoA −0.76 to +0.51 °C [13]; pooled population LoA −0.93 to +0.98 °C [14]), possibly because it was used in ventilated, sedated patients receiving vasopressors, many of whom were febrile. Similarity to a comparator that itself performed poorly is therefore not reassuring.
Bladder temperature and the reference standard. Bladder temperature showed the closest agreement and is currently the most reliable alternative in patients with an indwelling urinary catheter [6,7]. In the patient on VA-ECMO, bladder temperature read approximately 0.5 °C lower than blood temperature; because most venous return bypasses the pulmonary circulation and the circuit contains a heat exchanger, blood temperature may not have been a valid reference, but excluding this patient did not change the results for Moni-Patch™ (Table A2). Bladder temperature also narrowly failed the ±0.5 °C criterion, indicating that the criterion is demanding; nevertheless, the LoA width of Moni-Patch™ was twice that of bladder temperature (1.6 vs. 0.8 °C). IABP does not alter blood temperature, so blood temperature remained a valid reference in the four patients with IABP; however, their low cardiac output and peripheral vasoconstriction may have increased the error of skin-based estimates [13,14,15,16], which could not be examined separately.
Limitations. First, this was a single-center retrospective study with only 25 patients. Although the number of measurements was large, the precision of agreement estimates depends on the number of patients, and the 95% CIs of the LoA were wide. Second, 92% of the patients had undergone cardiovascular surgery, so the results cannot be directly applied to patients with sepsis, trauma, or neurocritical illness. All patients were Japanese. Third, only three patients contributed data in the low temperature range, and performance during targeted temperature management (TTM) could not be assessed. Fourth, ambient temperature, sweating, and indices of peripheral perfusion were not evaluated. Fifth, one Moni-Patch™ receiver and two sensors were provided free of charge by the manufacturer for clinical use; because the sensor used in each patient was not recorded, between-device variability could not be evaluated, and measurements were not blinded. Sixth, the stabilization period after sensor application was set at 10 min, but the optimal duration was not examined; nevertheless, excluding the first 60 min did not change the results. Seventh, data were limited to periods when all four signals were available, and measurements were serially dependent; the number of rapid temperature changes and of hypothermic readings was small, so performance under these conditions was not adequately assessed. Eighth, blood temperature itself may be affected by cold infusions or rewarming through central lines. Ninth, catecholamines and sedatives were used in 24 and 23 patients, respectively, but their agents and doses changed frequently; the exact times of axillary ice-pack application, replacement, and removal and of sensor replacement could not be obtained retrospectively; and sensor detachment and signal loss were not recorded systematically. The effects of these factors, and of IABP in four patients, on the error could therefore not be analyzed. Therefore, further validation is needed before the present findings can be generalized to other ICU populations.
5. Conclusions
In critically ill Japanese patients, predominantly after cardiovascular surgery, Moni-Patch™ read on average approximately 0.3 °C lower than blood temperature, with marked between-patient variability, and its LoA exceeded the clinical acceptance criterion. Its error did not differ significantly from that of the existing ZHF thermometer but was larger than that of bladder temperature, its trending ability did not meet polar-plot criteria, and its sensitivity for fever at 38.0 °C was low. It should therefore not replace invasive core temperature monitoring or be used as the sole means of detecting fever. Device-specific thresholds and validation in other ICU populations are needed.
Supplementary Materials
The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/jcm15239285/s1, File S1: R code used for the statistical analyses; File S2: De-identified dataset analyzed in this study; File S3: STARD 2015 Checklist.
Author Contributions
Conceptualization, M.I. and N.Y.; methodology, M.I. and N.Y.; formal analysis, M.I. and N.Y.; investigation, M.I., Y.S., Y.M., S.T., K.K., S.S., T.M., S.O. and N.Y.; data curation, M.I. and N.Y.; writing—original draft preparation, M.I. and N.Y.; writing—review and editing, all authors; visualization, M.I., Y.M. and N.Y.; project administration, M.I. and N.Y. All authors have read and agreed to the published version of the manuscript.
Funding
This research received no external funding.
Institutional Review Board Statement
The study was conducted in accordance with the Declaration of Helsinki and was approved by the Ethics Committee of Yokohama City University Hospital (approval number F250700030; date of approval: 12 August 2025).
Informed Consent Statement
Patient consent was waived by the Ethics Committee of Yokohama City University Hospital because of the retrospective design of the study; an opt-out procedure was used instead.
Data Availability Statement
The R code used for the analysis (File S1) and the de-identified dataset analyzed in this study (File S2) are provided as Supplementary Materials.
Acknowledgments
We thank the nurses and clinical engineers of the ICU at Yokohama City University Hospital for their cooperation with data collection and device management during routine clinical care. We are also grateful to Masashi Yokose for his clinical support and cooperation and to the attending physicians who cared for the patients.
Conflicts of Interest
Murata Manufacturing Co., Ltd. provided one Moni-Patch™ receiver and two sensors free of charge to our ICU in June 2024 for clinical use; this study was planned retrospectively after its use, and there was no contract or agreement with the manufacturer concerning this study. The manufacturer had no role in the design of the study, data analysis, interpretation of the results, or preparation of the manuscript. The authors declare no other financial or non-financial conflicts of interest. S.T. is a Guest Editor of this Special Issue and will not be involved in the editorial handling, peer review, or decision-making process for this manuscript.
Abbreviations
The following abbreviations are used in this manuscript:
| ICU | intensive care unit |
| ZHF | zero-heat-flux |
| LoA | limits of agreement |
| ROC | receiver operating characteristic |
| CI | confidence interval |
| MAE | mean absolute error |
| CCC | concordance correlation coefficient |
| AUC | area under the curve |
| VA-ECMO | veno-arterial extracorporeal membrane oxygenation |
| IQR | interquartile range |
| APACHE | Acute Physiology and Chronic Health Evaluation |
| SD | standard deviation |
| TTM | targeted temperature management |
| BMI | body mass index |
| CTEPH | chronic thromboembolic pulmonary hypertension |
| IABP | intra-aortic balloon pump |
| PPV | positive predictive value |
| NPV | negative predictive value |
| BCa | bias-corrected and accelerated |
| CDC | Centers for Disease Control and Prevention |
| NHSN | National Healthcare Safety Network |
| SIRS | systemic inflammatory response syndrome |
| STARD | Standards for Reporting of Diagnostic Accuracy Studies |
Appendix A
Figure A1.
Leave-one-patient-out analysis. Proportion of readings within ±0.5 °C for each method after each patient was omitted in turn.
Table A1.
Agreement with pulmonary artery blood temperature in individual patients.
Table A2.
Sensitivity analyses.
Table A3.
Comparison of absolute measurement error between methods.
Table A4.
Limits of agreement of Moni-Patch™ under alternative variance and correlation structures, with BCa bootstrap intervals.
Figure A2.
Diagnostics of the random-intercept model for Moni-Patch™. (A) Within-patient residuals against blood temperature (hexagons, density; line, smoother); (B) normal Q–Q plot of residuals; (C) mean within-patient autocorrelation of residuals by lag; (D) normal Q–Q plot of patient random intercepts.
Figure A3.
Time-lag analysis. (A) Within-patient SD of the difference from blood temperature when device readings were shifted by −10 to +40 min (positive values, device later); (B) cross-correlation between 10 min changes in each method and in blood temperature by lag.
Figure A4.
Differences between Moni-Patch™ and blood temperature plotted against their mean in individual patients. Each panel shows one patient (pt, patient number); points are paired measurements (darker shading indicates overlapping points), horizontal lines indicate zero difference, and gray bands ±0.5 °C.
Table A5.
Proportional bias (within-patient slope against blood temperature) and agreement stratified by the mean of the two methods and by bladder temperature.
Table A6.
Agreement by time since the start of the record.
Table A7.
Trending ability on common intervals defined by blood temperature and with alternative exclusion thresholds (0.1 and 0.3 °C).
Table A8.
Fever detection with fever defined as ≥38.0 °C and ≥38.3 °C, cross-validated device-specific thresholds, and missed febrile time.
Table A9.
Bias of Moni-Patch™ by phase of temperature change and exploratory mixed model of the Moni-Patch™ error.
Table A10.
Exploratory episode-level fever detection, delay to detection, and false alarms.
References
- O’Grady, N.P.; Alexander, E.; Alhazzani, W.; Alshamsi, F.; Cuellar-Rodriguez, J.; Jefferson, B.K.; Kalil, A.C.; Pastores, S.M.; Patel, R.; van Duin, D.; et al. Society of Critical Care Medicine and the Infectious Diseases Society of America Guidelines for Evaluating New Fever in Adult Patients in the ICU. Crit. Care Med. 2023, 51, 1570–1586. [Google Scholar] [CrossRef] [Scilit]
- Laupland, K.B.; Shahpori, R.; Kirkpatrick, A.W.; Ross, T.; Gregson, D.B.; Stelfox, H.T. Occurrence and Outcome of Fever in Critically Ill Adults. Crit. Care Med. 2008, 36, 1531–1535. [Google Scholar] [CrossRef] [Scilit]
- Evans, L.; Rhodes, A.; Alhazzani, W.; Antonelli, M.; Coopersmith, C.M.; French, C.; Machado, F.R.; Mcintyre, L.; Ostermann, M.; Prescott, H.C.; et al. Surviving Sepsis Campaign: International Guidelines for Management of Sepsis and Septic Shock 2021. Intensive Care Med. 2021, 47, 1181–1247. [Google Scholar] [CrossRef] [Scilit]
- Engelbart, G.; Brandt, S.; Scheeren, T.; Tzabazis, A.; Kimberger, O.; Kellner, P. Accuracy of Non-Invasive Sensors Measuring Core Body Temperature in Cardiac Surgery ICU Patients: Results from a Monocentric Prospective Observational Study. J. Clin. Monit. Comput. 2023, 37, 1619–1626. [Google Scholar] [CrossRef] [Scilit]
- Robinson, J.; Charlton, J.; Seal, R.; Spady, D.; Joffres, M.R. Oesophageal, Rectal, Axillary, Tympanic and Pulmonary Artery Temperatures during Cardiac Surgery. Can. J. Anaesth. 1998, 45, 317–323. [Google Scholar] [CrossRef] [Scilit]
- Cutuli, S.L.; See, E.J.; Osawa, E.A.; Ancona, P.; Marshall, D.; Eastwood, G.M.; Glassford, N.J.; Bellomo, R. Accuracy of Non-Invasive Body Temperature Measurement Methods in Adult Patients Admitted to the Intensive Care Unit: A Systematic Review and Meta-Analysis. Crit. Care Resusc. 2021, 23, 6–13. [Google Scholar] [CrossRef] [Scilit]
- Fallis, W.M. Monitoring Urinary Bladder Temperature in the Intensive Care Unit: State of the Science. Am. J. Crit. Care 2002, 11, 38–45. [Google Scholar] [CrossRef] [Scilit]
- Wartzek, T.; Mühlsteff, J.; Imhoff, M. Temperature Measurement. Biomed. Tech. 2011, 56, 241–257. [Google Scholar] [CrossRef] [Scilit]
- Eshraghi, Y.; Nasr, V.; Parra-Sanchez, I.; Van Duren, A.; Botham, M.; Santoscoy, T.; Sessler, D.I. An Evaluation of a Zero-Heat-Flux Cutaneous Thermometer in Cardiac Surgical Patients. Anesth. Analg. 2014, 119, 543–549. [Google Scholar] [CrossRef] [Scilit]
- West, N.; Cooke, E.; Morse, D.; Merchant, R.N.; Görges, M. Zero-Heat-Flux Core Temperature Monitoring System: An Observational Secondary Analysis to Evaluate Agreement with Naso-/Oropharyngeal Probe during Anesthesia. J. Clin. Monit. Comput. 2020, 34, 1121–1129. [Google Scholar] [CrossRef] [Scilit]
- Soehle, M.; Dehne, H.; Hoeft, A.; Zenker, S. Accuracy of the Non-Invasive Tcore™ Temperature Monitoring System to Measure Body Core Temperature in Abdominal Surgery. J. Clin. Monit. Comput. 2020, 34, 1361–1367. [Google Scholar] [CrossRef] [Scilit]
- Morettini, E.; Turchini, F.; Tofani, L.; Villa, G.; Ricci, Z.; Romagnoli, S. Intraoperative Core Temperature Monitoring: Accuracy and Precision of Zero-Heat Flux Heated Controlled Servo Sensor Compared with Esophageal Temperature during Major Surgery; the ESOSPOT Study. J. Clin. Monit. Comput. 2020, 34, 1111–1119. [Google Scholar] [CrossRef] [Scilit]
- Bräuer, A.; Fazliu, A.; Perl, T.; Heise, D.; Meissner, K.; Brandes, I.F. Accuracy of Zero-Heat-Flux Thermometry and Bladder Temperature Measurement in Critically Ill Patients. Sci. Rep. 2020, 10, 21746. [Google Scholar] [CrossRef] [Scilit]
- Conway, A.; Bittner, M.; Phan, D.; Chang, K.; Kamboj, N.; Tipton, E.; Parotto, M. Accuracy and Precision of Zero-Heat-Flux Temperature Measurements with the 3M™ Bair Hugger™ Temperature Monitoring System: A Systematic Review and Meta-Analysis. J. Clin. Monit. Comput. 2021, 35, 39–49. [Google Scholar] [CrossRef] [Scilit]
- Thomas, S.S.; Flickinger, K.L.; Elmer, J.; Callaway, C.W. Evaluation of Non-Invasive Sensors for Monitoring Core Temperature. J. Clin. Monit. Comput. 2025, 39, 1137–1142. [Google Scholar] [CrossRef] [Scilit]
- Idei, M.; Nomura, T. Accuracy of a New Zero-Heat-Flux Cutaneous Thermometer (SpotOn™) in Pediatric Intensive Care Patients. Tokyo Women’s Med. Univ. J. 2019, 3, 51–57. [Google Scholar] [CrossRef]
- Murata Manufacturing Co., Ltd. Moni-Patch™ Product Information. Available online: https://medical.murata.com/ja-jp/medical/products/moni-patch (accessed on 5 November 2026).
- Kottner, J.; Audigé, L.; Brorson, S.; Donner, A.; Gajewski, B.J.; Hróbjartsson, A.; Roberts, C.; Shoukri, M.; Streiner, D.L. Guidelines for Reporting Reliability and Agreement Studies (GRRAS) Were Proposed. J. Clin. Epidemiol. 2011, 64, 96–106. [Google Scholar] [CrossRef] [Scilit]
- Bossuyt, P.M.; Reitsma, J.B.; Bruns, D.E.; Gatsonis, C.A.; Glasziou, P.P.; Irwig, L.; Lijmer, J.G.; Moher, D.; Rennie, D.; de Vet, H.C.W.; et al. STARD 2015: An Updated List of Essential Items for Reporting Diagnostic Accuracy Studies. BMJ 2015, 351, h5527. [Google Scholar] [CrossRef] [Scilit]
- Pedersen, C.; Munch, P.; Kjaergaard, J.; Grønlykke, L.; Bräuer, A. Accuracy of a Zero-Heat-Flux Thermometer in Cardiac Surgery, a Prospective, Multicentre, Method Comparison Study. Sci. Rep. 2024, 14, 3169. [Google Scholar] [CrossRef] [Scilit]
- Ehlers, U.E.; Ulmer, J.; Keller, M.; Klein, C.; Pietsch, U. Comparison of Continuous Temperature Measurement Methods in the Intensive Care Unit: Standard Bladder Catheter Measurements versus Non-Invasive Transcutaneous Sensors. J. Clin. Monit. Comput. 2025, 39, 193–203. [Google Scholar] [CrossRef] [Scilit]
- Bland, J.M.; Altman, D.G. Agreement between Methods of Measurement with Multiple Observations per Individual. J. Biopharm. Stat. 2007, 17, 571–582. [Google Scholar] [CrossRef] [Scilit]
- Parker, R.A.; Weir, C.J.; Rubio, N.; Rabinovich, R.; Pinnock, H.; Hanley, J.; McCloughan, L.; Drost, E.M.; Mantoani, L.C.; MacNee, W.; et al. Application of Mixed Effects Limits of Agreement in the Presence of Multiple Sources of Variability: Exemplar from the Comparison of Several Devices to Measure Respiratory Rate in COPD Patients. PLoS ONE 2016, 11, e0168321. [Google Scholar] [CrossRef] [Scilit]
- Pesonen, E.; Silvasti-Lundell, M.; Niemi, T.T.; Kivisaari, R.; Hernesniemi, J.; Mäkinen, M.T. The Focus of Temperature Monitoring with Zero-Heat-Flux Technology (3M Bair-Hugger): A Clinical Study with Patients Undergoing Craniotomy. J. Clin. Monit. Comput. 2019, 33, 917–923. [Google Scholar] [CrossRef] [Scilit]
- Brandes, I.F.; Tirilomis, T.; Nemeth, M.; Wieditz, J.; Bräuer, A. Intraoperative Zero-Heat-Flux Thermometry Overestimates Nasopharyngeal Temperature by 0.39 °C: An Observational Study in Patients Undergoing Congenital Heart Surgery. J. Clin. Monit. Comput. 2025, 39, 205–215. [Google Scholar] [CrossRef] [Scilit]
- Nemeth, M.; Lovric, M.; Asendorf, T.; Bräuer, A.; Miller, C. Intraoperative Zero-Heat-Flux Thermometry Overestimates Esophageal Temperature by 0.26 °C: An Observational Study in 100 Infants and Young Children. J. Clin. Monit. Comput. 2021, 35, 1445–1451. [Google Scholar] [CrossRef] [Scilit]
- Lin, L.I. A Concordance Correlation Coefficient to Evaluate Reproducibility. Biometrics 1989, 45, 255–268. [Google Scholar] [CrossRef] [Scilit]
- Niven, D.J.; Gaudet, J.E.; Laupland, K.B.; Mrklas, K.J.; Roberts, D.J.; Stelfox, H.T. Accuracy of Peripheral Thermometers for Estimating Temperature: A Systematic Review and Meta-Analysis. Ann. Intern. Med. 2015, 163, 768–777. [Google Scholar] [CrossRef] [Scilit]
- Bone, R.C.; Balk, R.A.; Cerra, F.B.; Dellinger, R.P.; Fein, A.M.; Knaus, W.A.; Schein, R.M.; Sibbald, W.J. Definitions for Sepsis and Organ Failure and Guidelines for the Use of Innovative Therapies in Sepsis. Chest 1992, 101, 1644–1655. [Google Scholar] [CrossRef] [Scilit]
- Critchley, L.A.; Lee, A.; Ho, A.M. A Critical Review of the Ability of Continuous Cardiac Output Monitors to Measure Trends in Cardiac Output. Anesth. Analg. 2010, 111, 1180–1192. [Google Scholar] [CrossRef] [Scilit]
- Critchley, L.A.; Yang, X.X.; Lee, A. Assessment of Trending Ability of Cardiac Output Monitors by Polar Plot Methodology. J. Cardiothorac. Vasc. Anesth. 2011, 25, 536–546. [Google Scholar] [CrossRef] [Scilit]
- Centers for Disease Control and Prevention. National Healthcare Safety Network (NHSN) Patient Safety Component Manual, Chapter 9: Surgical Site Infection (SSI) Event. 2026. Available online: https://www.cdc.gov/nhsn/pdfs/pscmanual/9pscssicurrent.pdf (accessed on 21 November 2026).
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.