Abstract
Leadless pacemakers were developed to reduce the risk of infection associated with conventional pacing systems by eliminating the subcutaneous generator pocket and transvenous leads. Over the past decade, clinical studies, including five-year follow-up data, have shown low absolute infection rates and favourable complication and revision profiles relative to conventional pacemakers. Their use has also expanded to atrial and dual-chamber pacing, and they are increasingly used following the removal of infected transvenous systems. However, important challenges remain. Although uncommon, leadless pacemaker infections have been reported, including cases of bacteraemia and infective endocarditis involving newer atrial and helix-fixation devices. Diagnosis can be difficult because there are no visible pocket-related signs of infection. Management is also unsettled. In the Micra investigational study all systemic infectious events were adjudicated as unrelated to the device or procedure, so they do not demonstrate that device-related bacteraemia can be treated without extraction, and no comparative evidence defines the balance between device removal and retention. Device retrieval presents another challenge, particularly after prolonged implantation, when tissue encapsulation can make extraction more difficult and increase procedural risks. Reports of atrial device embolisation have raised further concerns about retrieval techniques. This narrative review examines the mechanisms underlying the lower infection risk of leadless pacemakers, reported infections and their microbiology, diagnostic and management strategies, and current retrieval techniques. It also discusses the role of leadless pacing following extraction of infected conventional systems and highlights areas where further research is needed, particularly with the increasing use of atrial and dual-chamber devices.
1. Introduction
Cardiac implantable electronic device (CIED) infection is a serious complication associated with significant morbidity, prolonged hospitalisation, increased healthcare costs, and higher mortality, particularly when infection becomes systemic [1,2]. In conventional pacemakers, the subcutaneous generator pocket and transvenous leads are the main sites associated with infection. These components provide surfaces for bacterial attachment and potential routes for infection to spread [3]. Leadless pacemakers were developed to reduce these risks by eliminating both components. They consist of a small, self-contained device implanted directly into the right ventricle or, more recently, the right atrium [4].
The first large multicentre leadless experience was the LEADLESS II study of the Nanostim device [5], followed by the pivotal Micra investigational trial, which established the feasibility and short-term safety of transcatheter ventricular pacing [4]. The technology has since advanced considerably. Five-year registry data have provided evidence of long-term performance, with low rates of major complications and system revision compared with conventional transvenous systems [6]. Further developments include retrievable helix-fixation devices [7], atrioventricular-synchronous single-chamber pacing, and dual-chamber leadless systems [8,9]. Absolute infection rates have been low across clinical trials and registries (Section 3, Table 1), supporting the use of leadless pacemakers in patients at increased risk of infection and those requiring pacing after removal of an infected transvenous system. In a United States readmissions analysis involving 41,058 leadless pacemaker implantations, 7.9% were performed following transvenous lead removal for CIED infection [10].
Table 1.
Absolute infection outcomes from principal leadless pacemaker trials and registries.
Despite these advantages, two important challenges remain. First, leadless pacemaker infection, although rare, is increasingly recognised, with published cases of bacteraemia and infective endocarditis. Diagnosis can be difficult because these devices lack the visible pocket-related signs typically associated with conventional pacemaker infections. A third, related question, is how an established leadless infection should be managed, since removal is not always straightforward and retention has sometimes succeeded. Second, device retrieval remains challenging, particularly after prolonged implantation, when tissue encapsulation may complicate extraction and increase procedural risk. Retrieval may be required because of infection, device malfunction, dislodgement, embolisation, or battery depletion.
This narrative review examines the current evidence on leadless pacemaker infection and retrieval, including clinical presentation, diagnosis, management, and procedural outcomes. Particular attention is given to newer atrial and dual-chamber systems, for which long-term evidence remains limited. Two recent reviews have summarised the clinical performance of leadless pacemakers broadly [15] and the reported cases of leadless infective endocarditis up to mid-2024 [16]. The present review is distinct in scope: it concentrates on device infection and its management together with retrieval and reimplantation, applies an explicit endocarditis-classification framework to the reported cases (Table 2), separates the question of relative from absolute infection risk (Table 1), and incorporates the most recent atrial, dual-chamber, and helix-fixation retrieval data.
Table 2.
Published cases of leadless pacemaker device infection, classified by the 2023 Duke-ISCVID criteria.
2. Methods
This article is an invited narrative review and was not registered or conducted as a systematic review. We followed the principles of the Scale for the Assessment of Narrative Review Articles (SANRA) for transparency of reporting. The electronic search was run in PubMed from database inception to October 2026. Search terms combined device terms (“leadless pacemaker”, “leadless cardiac pacemaker”, “Micra”, “Aveir”, “Nanostim”) with infection terms (“infection”, “infective endocarditis”, “endocarditis”, “bacteraemia”, “sepsis”, “pocket infection”, “lead infection”, “vegetation”) and retrieval terms (“retrieval”, “extraction”, “explantation”, “removal”). Searches using “Micra” were field-restricted to the device to exclude records referring to the bacterium Parvimonas micra. The reference lists of retrieved articles and of relevant consensus documents were hand-searched. Records were screened by title and abstract, and the full texts of potentially relevant articles were reviewed; discrepancies were resolved by discussion among the authors. Only English-language reports were included. We did not formally search Embase, Scopus, Web of Science, or society (HRS, EHRA, APHRS) conference abstracts; this restriction to PubMed and reference-list screening is acknowledged as a limitation, and the case tabulation in Table 2 is therefore presented as illustrative rather than as a complete census. Priority was given to randomised trials, registries, and case reports describing uncommon or newly recognised complications. Cases included in Table 2 were limited to reports in which the leadless pacemaker itself was considered the site of infection, and each was classified using the 2023 Duke-International Society for Cardiovascular Infectious Diseases (Duke-ISCVID) criteria [28] and the EHRA 2020 consensus definitions [1]. No formal restrictions were applied regarding publication date.
3. The Infection Burden of Transvenous Systems and the Rationale for Leadless Pacing
Conventional transvenous cardiac implantable electronic devices (CIEDs) carry a recognised risk of infection, including generator pocket infection and lead-related endocarditis. Pocket haematoma is an additional complication that increases the risk of subsequent infection. Repeated procedures, such as generator replacements and lead revisions, further increase this risk by requiring repeated access to the device pocket [3]. Once a transvenous system becomes infected, complete device and lead removal is recommended. However, extracting chronically implanted leads carries risks, including vascular injury and myocardial damage [1,2].
Leadless pacemakers were developed to reduce these complications by eliminating the generator pocket and transvenous leads. The absolute infection risk reported for leadless devices is low. No device-related infection was observed in the pivotal leadless trials enrolling more than 3000 patients, against an expected transvenous infection rate of approximately 0.8% to 2.1% [11]. These absolute figures, summarised in Table 1, should be read with the caveat that trial and registry follow-up is finite and that infection may be under-ascertained. Available evidence generally supports a lower risk of infection compared with conventional systems, although leadless devices have their own procedural risks, particularly vascular complications and cardiac perforation. A systematic review of 18 studies involving 2496 leadless implants reported implantation success of 95.5% to 100%, device- or procedure-related death in 0.3%, any complication in 3.1%, and pericardial tamponade in 1.4%, with infection in fewer than 1% of patients; its meta-analysis against transvenous pacing was limited to haematoma, pericardial effusion, device dislocation, any complication, and death, none of which differed significantly, and it did not compare infection rates [29]. A larger meta-analysis of 55 studies involving 286,349 patients, including contemporary Micra and Aveir devices, reported significantly lower risks of pneumothorax (relative risk 0.29) and lead dislodgement (relative risk 0.21) with leadless devices and described infection rates as negligible without reporting a comparative infection estimate, while cardiac tamponade (relative risk 2.36) and access-site complications (relative risk 2.82) were more frequent [14]. Additional meta-analyses have reported comparable implantation success rates while identifying differences in early procedural complications [30,31]. These findings support the consideration of leadless pacing in patients at increased risk of infection, including those requiring pacing after extraction of infected transvenous systems [32] and patients with previous valve interventions [33]. The two United States readmissions analyses cited here [10,32] are separate reports drawn from the same National Readmissions Database cohort and should not be read as independent datasets.
Although several meta-analyses have compared infection rates between leadless and conventional pacemakers, less attention has been given to the clinical presentation and management of infection involving leadless devices. This review therefore focuses on the characteristics of reported infections, diagnostic challenges, treatment approaches, and device retrieval rather than repeating existing pooled analyses.
When interpreting the available evidence, it is important to distinguish between the relative risk of infection, its absolute incidence, and clinical outcomes once infection occurs. Comparative studies and registries provide information on infection rates, although differences in patient selection and follow-up duration limit direct comparisons. A further caveat applies to the five-year Micra post-approval registry, which compared major complications and system revisions with a historical transvenous cohort only to 36 months and did not perform a head-to-head comparison of infection rates [6]; its contribution is the demonstration of a very low absolute infection rate rather than a comparative infection estimate. In contrast, published case reports and small case series offer useful insights into clinical presentation, microbiology, and management but cannot reliably establish infection incidence or comparative treatment effectiveness. Furthermore, most long-term data come from ventricular leadless pacemakers, while evidence for newer atrial and dual-chamber systems remains limited. Findings from established ventricular platforms should therefore be applied cautiously to these newer devices. Absolute infection outcomes from principal leadless pacemaker trials and registries are shown in Table 1.
4. Mechanisms of Infection Resistance and the Encapsulation Paradox
The lower infection risk associated with leadless pacemakers may be explained by several complementary mechanisms [11]. First, eliminating the subcutaneous generator pocket removes a common site of CIED infection. Second, the absence of transvenous leads reduces the foreign-body surface available for bacterial adhesion and biofilm formation. It also eliminates the connection between the device pocket and the venous system. Third, the leadless capsule presents a much smaller total surface area for microbial adhesion than a transvenous system, and it sits in a high-velocity intracardiac flow environment that is less favourable to bacterial colonisation. Fourth, the parylene-coated titanium surface has been shown in bench studies to reduce bacterial adhesion compared with bare titanium [11]. Fifth, the transfemoral delivery of a fully enclosed device reduces skin and glove contact and handling relative to a pocket procedure [11]. Finally, following implantation, the leadless pacemaker gradually becomes surrounded by fibrous and endothelial tissue, a process known as encapsulation. This tissue covering may reduce direct exposure of the device to circulating microorganisms. Direct comparative or preclinical challenge data establishing the relative contribution of each mechanism are lacking, and these remain plausible but unproven explanations.
Evidence of leadless pacemaker encapsulation initially came from animal studies and postmortem examinations. In 2026, a retrospective study using contrast-enhanced computed tomography provided an early in vivo assessment of the extent and distribution of encapsulation in patients with implanted devices [34]. However, the study did not investigate whether encapsulation protects against infection. Its potential role in preventing bacterial attachment and haematogenous infection therefore remains uncertain.
The encapsulation paradox is that the same tissue response may act in opposite directions. On one hand, a mature capsule could shield the device surface from circulating organisms, which is consistent with the observation that most reported infections presented early: six of the twelve tabulated cases (Table 2) occurred within approximately four months of implantation, before encapsulation is expected to be complete, and several arose in devices implanted shortly after extraction of an infected transvenous system. On the other hand, the same capsule binds the device to the endocardium and is the principal obstacle to later retrieval (Section 8).
As tissue develops around the pacemaker, it can become firmly attached to the endocardium, making extraction more difficult and increasing the risk of myocardial injury. Interestingly, a reported case of fatal leadless pacemaker endocarditis involving a device implanted three years earlier found no encapsulation at extraction [16]. Although this observation raises questions about the relationship between encapsulation and infection, a single case cannot establish a protective effect.
Other factors remain poorly understood, including biofilm formation on the titanium device surface and the potential influence of helix-fixation mechanisms on infection risk. Further research is needed to determine whether microbiological markers or host-response assessments could help identify patients at increased risk of device infection. Emerging approaches to more individualised diagnosis and management of CIED infections may contribute to this area [35]. Artificial intelligence and electronic health record-based algorithms have also been explored for CIED infection risk stratification and early detection, although their clinical utility in leadless pacemaker recipients remains unestablished [36].
5. The Clinical Spectrum of Leadless Device Infection
Although leadless pacemaker infections are rare, an increasing number of cases have been reported, showing a wider range of clinical presentations than initially recognised. Most available evidence comes from individual case reports and small case series, making it difficult to determine the true incidence of infection. Publication bias is also possible, as unusual cases or those with successful treatment outcomes may be more likely to be reported. Table 2 summarises the published cases that we identified of leadless pacemaker infection, each classified by the 2023 Duke-ISCVID criteria as definite or possible device-related infection [28]. Because the search was limited to PubMed and reference lists, the table is illustrative rather than a complete census; it incorporates the first reported infected Micra device, described by Koay and colleagues in 2016 [17], together with the cases reviewed by Mourad and colleagues [16,17,18,19,20,21,22,23,24,25,26,27].
Two definitional points follow from the table. First, the inclusion criterion is that the leadless device itself was the reported site of infection; where the device was implicated on imaging but device cultures were negative, as in the Candida case of Ellison and colleagues [19], the case is classified as possible rather than definite device infection. Second, two reports [20,27] describe concomitant infection of transvenous hardware and a leadless device; these are retained but flagged as concomitant, because the leadless device was not the sole focus.
Several findings emerge from the reported cases. Staphylococcus aureus was the causative organism in ten of the twelve reported cases, and no case was caused by a coagulase-negative staphylococcus. This microbiological pattern, characterised by the predominance of Staphylococcus aureus and the absence of coagulase-negative staphylococci, suggests that haematogenous seeding may play a greater role in leadless pacemaker infection than in conventional pocket-related CIED infection. Clinical presentations vary from persistent or recurrent bacteraemia without an identifiable source [18,23] to device-associated vegetations and infective endocarditis [21,24]. Severe complications and death have also been reported. In one case, a patient developed MRSA endocarditis three years after leadless pacemaker implantation, complicated by cerebral and septic pulmonary emboli. Despite device removal and clearance of bacteraemia, the patient died from multiorgan failure [16].
Fungal infections have also been reported, with Candida albicans identified in two cases. One involved a helix-fixation device with pathologically confirmed fungal endocarditis, managed with vacuum-assisted removal of the vegetation, device extraction, and replacement [24]. The other involved a Micra pacemaker removed because of persistent Candida albicans fungaemia and a device-associated mass suspected to be a vegetation, although cultures from the extracted device were negative [19]. Unusual organisms, including rapidly growing non-tuberculous mycobacteria, may also need consideration when routine microbiological investigations fail to identify a pathogen [37].
Recent reports have expanded our understanding of leadless pacemaker infection. In 2026, the first case involving an atrial leadless pacemaker was published. A 70-year-old man developed MRSA bacteraemia three months after Aveir AR implantation. Imaging identified a cystic mass attached to the device, requiring percutaneous extraction followed by aspiration of residual material [25]. Infection has also been reported with helix-fixation ventricular devices, including a case in which an unexpected rise in pacing threshold was an important early diagnostic finding [26]. These reports, together with an earlier case of fungal infection involving a helix-fixation device [24], confirm that infection is not limited to the original tine-fixation platforms. Furthermore, simultaneous infection of leadless and transvenous pacing systems has been described, highlighting the importance of assessing all implanted hardware when investigating suspected device infection [27].
Other reported complications include tine fracture and device dislodgement associated with bioprosthetic tricuspid valve endocarditis [38]. Recurrent device-related infections have also led to the use of sequential leadless pacing systems in individual patients [39]. Importantly, two reports describe a leadless device becoming infected after implantation into a field that was not yet sterile: in one, a Micra was placed alongside abandoned, infected transvenous leads and both became culture-positive for MRSA [20], and in another, recurrent MRSA bacteraemia followed implantation shortly after extraction, prompting device retrieval [18]. These cautionary cases, rather than examples of elective early reimplantation, underline the importance of infection control before a new device is placed. As newer pacing systems become more widely used, larger studies will be needed to establish their long-term infection and complication rates. Early multicentre data on the electrical performance and short-term outcomes of Aveir VR devices are becoming available, although longer follow-up is required [40].
6. Diagnostic Challenges
Diagnosing leadless pacemaker infection can be difficult because these devices lack the generator pocket associated with conventional pacemakers. Consequently, visible signs such as pocket inflammation, swelling, or erosion are absent. Infection should therefore be considered in patients with persistent or recurrent bacteraemia without an identifiable source, particularly when Staphylococcus aureus is isolated [18,23]. In this context, and by analogy with transvenous systems, Staphylococcus aureus bacteraemia in a leadless pacemaker recipient without an alternative source should prompt active evaluation of the device as a possible focus rather than being presumed unrelated, even though the absolute probability of device involvement is lower than for transvenous hardware. Unexplained changes in device function may provide additional clues. For example, an increase in pacing threshold was an important finding in a reported case of helix-fixation leadless pacemaker infection [26].
Imaging plays an important role in diagnosis. Diagnosis should be framed within a formal endocarditis framework: the 2023 Duke-ISCVID criteria [28] and the 2023 European Society of Cardiology (ESC) endocarditis guidelines [41] both now recognise CIED leads and intracardiac device material as diagnostic elements, and we have applied the Duke-ISCVID criteria to classify the cases in Table 2. Transthoracic and transoesophageal echocardiography are used to assess for vegetations and other features of infective endocarditis. Intracardiac echocardiography may provide additional information by allowing direct visualisation of the device and its surrounding structures, particularly when assessing small intracardiac devices [42,43]. A specific difficulty is that an echo-dense mass on a leadless device may represent either a vegetation or a bland thrombus, and the two cannot always be distinguished on imaging alone; the Candida case in Table 2, in which a device-associated mass was accompanied by negative device cultures, illustrates this uncertainty [19]. When conventional echocardiography is inconclusive, advanced imaging techniques, including nuclear imaging used in conventional CIED infections, may be considered. However, the small size and metallic construction of a leadless capsule limit the spatial resolution and specificity of 18F-FDG positron-emission tomography and radiolabelled-leucocyte imaging for a device of this size, and evidence supporting their diagnostic accuracy specifically in leadless pacemaker infections remains limited. Emerging approaches to more individualised infection diagnosis may help improve identification of device-related infection in the future [35]. In some cases, the diagnosis remains uncertain until infected material is retrieved and examined, as demonstrated in a reported case of fungal leadless pacemaker endocarditis [24].
7. Management Principles
Management of leadless pacemaker infection requires an individualised approach, as the available evidence is largely observational. The main principles are similar to those used for conventional CIED infections and include infection source control, appropriate antimicrobial therapy, and multidisciplinary decision-making involving electrophysiology, infectious diseases, cardiac imaging, and cardiac surgery [1,2,44]. Neither the 2026 multisociety consensus statement [2] nor the 2026 Delphi consensus [44] provides a leadless-specific algorithm; the former addresses leadless devices within its general lead-management and extraction recommendations and supports percutaneous retrieval of infected devices where feasible, while the latter offers expert-consensus guidance on prevention and treatment of CIED infection that is applicable, but not specific, to leadless systems. The recommendations below are therefore graded by their basis: guideline, observational, or expert opinion.
A central and unresolved question is whether every infected leadless device must be removed, or whether, in selected patients, device-related bacteraemia can be treated without extraction. The observational evidence cuts both ways. In the Micra investigational study, 16 of 720 patients developed 21 serious infectious events (bacteraemia or endocarditis) at a mean of 4.8 months after implant; all were adjudicated as unrelated to the device or procedure, were managed with antimicrobial therapy without device removal, and no persistent bacteraemia was seen after antibiotic cessation [12]. Because these events were not device infections, they show only that systemic infection in a patient with a leadless pacemaker need not mandate device removal; they do not establish that an infection involving the device itself can be treated without extraction. Across the five-year post-approval registry of 1809 patients there were no device removals for infection, and the single possible device-related endocarditis resolved with antibiotics alone [6]. These observational data therefore speak to systemic infection in patients with a leadless device rather than to established device infection, for which no evidence supports retention. By contrast, the tabulated case reports (Table 2), which are weighted towards extracted and more severe presentations, show that once a device vegetation or persistent, relapsing bacteraemia is established, antibiotics alone frequently fail and extraction is required. A pragmatic synthesis is that Staphylococcus aureus or persistent bacteraemia with an imaged device vegetation should be managed as device infection and treated with extraction where feasible, whereas transient, cleared bacteraemia without a device vegetation may reasonably be managed medically with close surveillance.
A practical five-step approach is outlined in Figure 1. First, leadless pacemaker infection should be suspected in patients with persistent or recurrent bacteraemia without an identifiable source, suspected infective endocarditis, a device-associated mass, or unexplained pacing dysfunction. Blood cultures should be drawn and empirical antimicrobial therapy started at this point, before and independent of the multidisciplinary discussion. Second, investigations should combine microbiological findings with transthoracic and transoesophageal echocardiography, supported by additional imaging when needed. It is important to distinguish infection involving the leadless pacemaker from infection of surrounding cardiac structures or coexisting transvenous devices. The pathway then branches: a device-associated vegetation directs management towards extraction, whereas bacteraemia without a demonstrable vegetation may be managed with culture-directed antibiotics and surveillance only when it is transient and clears promptly. Persistent or relapsing bacteraemia, particularly with Staphylococcus aureus, should be treated as device infection and directed towards extraction even when no vegetation is seen.
Figure 1.
Stepwise approach to the patient with a leadless pacemaker and suspected infection. Blood cultures are drawn and empirical antibiotics started before multidisciplinary discussion; management branches on the presence of a device-associated vegetation; and reimplantation follows bacteraemia clearance. TTE, transthoracic echocardiography; TOE, transoesophageal echocardiography; EP, electrophysiology.
Third, a multidisciplinary team should assess the findings and discuss treatment options. Fourth, the most appropriate source-control strategy should be determined, including whether device extraction is necessary and feasible. The decision should consider the possibility of percutaneous retrieval, the need for vegetation debulking, procedural risks, the causative organism, the extent of infection, and the patient’s overall clinical condition. Where the extraction risk is prohibitive, vegetation debulking or chronic suppressive therapy may be used as a bridge or palliative measure, but this leaves the infected device in situ and is not a route to reimplantation of that device.
Finally, culture-directed antimicrobial therapy should be planned, with treatment duration guided by the organism and extent of infection. For device-related endocarditis, a course of the order of four to six weeks from the first negative blood culture is consistent with transvenous CIED practice, and reimplantation, when required, should be deferred until bacteraemia has cleared, conventionally by at least two weeks [1]. If permanent pacing remains necessary, the timing and type of replacement system should also be considered.
The need for device removal remains an important question in managing leadless pacemaker infection. In conventional transvenous systems, complete hardware removal is recommended when device infection is confirmed (guideline) [1,2]. However, extracting an infected leadless pacemaker can be more challenging because of its intracardiac position and potential tissue encapsulation. The decision must therefore balance the need for infection source control against procedural risks and the patient’s clinical condition.
Several approaches have been reported, including vacuum-assisted removal of infected vegetations followed by device extraction [24], combined transcatheter vegetation removal and leadless pacemaker implantation during the same procedure [45], and individualised decisions regarding device retrieval [21,27]. Although these reports demonstrate the feasibility of different techniques, evidence supporting device retention or vegetation removal alone as definitive treatment remains limited and is largely confined to transient bacteraemia without a device vegetation, as discussed above. Complete device removal should therefore remain the preferred approach for confirmed device infection whenever it can be performed safely.
Following extraction, prolonged culture-directed antimicrobial therapy is often required. The timing of reimplantation also needs careful consideration, as early placement of a new device before adequate infection control may increase the risk of recurrent infection [23]. A Delphi consensus published in 2026 provides further recommendations on the prevention and management of CIED infections, including areas not fully addressed by existing guidelines (expert opinion) [44]. Patients requiring anaesthesia or surgical intervention may also need additional perioperative planning, particularly when significant comorbidities are present [46].
8. Retrieval and Extraction of Leadless Devices
Leadless pacemaker retrieval may be required for several reasons, including infection, device malfunction, dislodgement, embolisation, battery depletion, or the need for a pacing system upgrade. Retrieval can be classified as acute, performed during or shortly after implantation, or delayed, involving devices that have been implanted for longer periods. Delayed retrieval is generally more challenging because of tissue encapsulation around the device [47,48]. It is important to note that retrieval is not the only option at the end of device service: for the Micra, the established strategy at battery depletion or for most revisions is to abandon the device in situ and implant a second leadless pacemaker, reserving extraction for infection, embolisation, or a limited number of other indications [6,49].
Encapsulation develops as fibrous and endothelial tissue gradually surrounds the pacemaker and attaches it to the endocardium. This can make extraction difficult and increase the risk of myocardial injury, perforation, or cardiac tamponade. A recent computed tomography study provided in vivo evidence of the extent and distribution of encapsulation, offering useful information for understanding the challenges of delayed retrieval [34].
Several retrieval techniques have been described. These include transcatheter removal of tine-fixation ventricular devices [48], different approaches to Micra retrieval [50], snare-assisted extraction through alternative venous access [51], and removal of chronically implanted devices before upgrading to conduction-system pacing [52]. Helix-fixation devices present additional technical challenges, including the possibility of helix fracture during extraction [53]. The largest dedicated chronic-retrieval experience to date comes from the worldwide helix-fixation programme reported by Neuzil, Exner, and colleagues, in which 234 chronic retrieval attempts (at least six weeks after implantation) in 233 subjects achieved an overall retrieval success of 87.6% (205 of 234), with failed attempts most often due to inability to engage the device docking feature [54]. This provides the most robust estimate available that mid-term and long-term helix-fixation devices can usually, but not always, be removed percutaneously. Large registries have also provided data on device revisions and their outcomes [49], while comparative studies have examined extraction complications in leadless and conventional transvenous systems [55]. In cases of infection, percutaneous retrieval using steerable sheaths and snares has been frequently reported, sometimes combined with vacuum-assisted aspiration of associated vegetations [18,19,21,22].
Although the available studies differ in design and patient populations, several provide useful information on retrieval outcomes, as summarised in Table 3. In a single-centre study of nine patients, transcatheter retrieval of tine-fixation leadless pacemakers was successful in eight patients (88.9%), with a mean implantation duration of 3.1 years [48]. A French multicentre registry involving 8994 Micra implantations reported 100 device revisions (1.1%). Of these, 30 involved attempted device removal, with a success rate of 97% (29/30) and no major complications reported. Most removals occurred relatively early, with a median implantation duration of 81.5 days and 80% performed within the first year [49].
Long-term data from the Micra post-approval registry, which followed 1809 patients for a median of 51.1 months, showed a 4.9% all-cause system-revision rate and a 4.5% major-complication rate at five years, both lower than those observed in the transvenous comparator cohort [6]. System revisions in this registry were predominantly device upgrades or reprogramming for elevated thresholds rather than extraction; there were no removals for infection, and the only major complication associated with an explant followed entanglement of the device with an inferior vena cava filter [6]. More recent reports have addressed retrieval of newer atrial devices. A six-centre series described percutaneous retrieval of embolised atrial leadless pacemakers [56], while a separate four-case series reported successful Micra retrieval for complications including increased pacing thresholds, pulmonary artery embolisation, and tine fracture, using alternative retrieval tools where a dedicated system was unavailable [50].
These findings suggest that percutaneous retrieval is feasible in selected patients, including some with chronically implanted devices. However, differences in device type, indication for removal, implantation duration, and reported outcomes limit direct comparisons between studies. The available success rates should therefore be interpreted cautiously, particularly when considering retrieval of older or heavily encapsulated devices.
Table 3.
Procedural cohorts and registries informing leadless pacemaker retrieval and revision outcomes.
A distinct and increasingly reported problem is embolisation of atrial leadless devices, which can migrate to the pulmonary circulation and require specific percutaneous retrieval strategies [56,57,58]. The 2026 multisociety expert consensus statement update, which explicitly extends its scope to leadless pacing and to devices implanted outside the vascular system, provides the most current framework for lead and device management and extraction and should be read alongside the device-specific technical literature [2].
9. Leadless Pacing as a Reimplantation Strategy After Infected-System Removal
Leadless pacemakers are increasingly used to restore pacing after the removal of infected transvenous systems. This approach avoids the need for a new generator pocket and transvenous leads, which may reduce the risk of recurrent infection. In a United States readmissions analysis, patients receiving leadless pacemakers after extraction of infected transvenous systems were generally younger and had more comorbidities than those undergoing primary leadless pacemaker implantation [10]. Direct registry evidence supports this strategy: in the Micra post-approval registry, 105 patients received a leadless device within 30 days of extraction of an infected CIED, with a 99% implantation success rate and, importantly, no device explanted for infection and no recurrent infection requiring removal on follow-up [13].
An early systematic review supported the feasibility of leadless pacemaker implantation following removal of infected transvenous systems [59]. Subsequent observational studies have reported outcomes in this patient group [10], with comparative data suggesting favourable results for leadless pacing compared with conventional transvenous reimplantation [32].
An important consideration is the timing of reimplantation. Leadless pacemakers may be implanted during the same procedure as extraction or after infection has been adequately controlled. Studies comparing immediate and delayed implantation have examined their safety and associated complications, although the available evidence remains limited [60,61]. Same-session extraction and reimplantation has been reported and may be particularly useful in pacemaker-dependent patients, as it reduces the need for prolonged temporary pacing [62,63].
However, immediate reimplantation carries a potential risk of introducing a new device before the infection has resolved. Cases of recurrent bacteraemia following early leadless pacemaker implantation into a non-sterile field highlight this concern [18,20]. The decision should therefore consider the severity of infection, microbiological findings, pacing requirements, and the patient’s overall condition. Further prospective studies are needed to establish the safest timing of leadless pacemaker implantation following extraction of infected systems.
10. Prevention, Patient Selection, and the Dual-Chamber and Atrial Era
Preventing leadless pacemaker infection requires appropriate patient selection, careful implantation techniques, and the judicious use of antimicrobial prophylaxis. Antibiotic strategies should account for individual infection risk while avoiding unnecessary treatment. Recent research has reassessed the role of prophylactic antibiotics in electrophysiology procedures considered to have a low infection risk, highlighting the need for more targeted preventive strategies [64]. Leadless pacing is frequently chosen specifically for patients at high infection risk, including those with prior CIED infection, end-stage renal disease, immunosuppression, or limited vascular access, and the preventive evidence base is largely extrapolated from transvenous practice rather than generated in these leadless-specific high-risk groups.
Advances in leadless pacing have expanded treatment options beyond single-chamber ventricular devices. These include single-chamber systems capable of atrioventricular-synchronous pacing [8] and dual-chamber leadless pacemakers, whose feasibility was established in the pivotal investigational study of the Aveir dual-chamber system and which have since shown encouraging one-year safety outcomes and maintenance of atrioventricular synchrony in clinical practice [9,65]. Real-world studies have compared dual-chamber leadless systems with conventional transvenous pacemakers [66], while early procedural studies have assessed their performance and safety [67]. Atrial leadless pacing has also emerged as an option for patients with sinus node dysfunction [68]. Beyond pacing alone, modular leadless systems that pair a leadless pacemaker with a subcutaneous defibrillator to deliver antitachycardia pacing are in early clinical use [69]; their infection and retrieval profiles, involving communication between two implanted components, are not yet defined and warrant specific surveillance.
Although these developments have broadened the use of leadless pacing, they introduce additional considerations regarding infection and device retrieval. Dual-chamber systems involve two intracardiac devices, while some newer platforms use active helix-fixation mechanisms. These features may create additional technical challenges during extraction. Furthermore, long-term infection and retrieval outcomes remain poorly understood for newer devices. Recent reports of atrial leadless pacemaker infection and infections involving helix-fixation systems demonstrate the need for continued surveillance and longer-term clinical studies [25,26].
11. Current Challenges and Future Directions
Important gaps remain in the understanding of leadless pacemaker infection and retrieval. Table 4 summarises the principal research priorities in leadless pacemaker infection and retrieval and the uncertainties they aim to address.
Table 4.
Research priorities in leadless pacemaker infection and retrieval.
Two priorities deserve emphasis. First, the field lacks prospective registries that capture infection and retrieval as prespecified endpoints across atrial, dual-chamber, and helix-fixation platforms, without which absolute incidence and comparative management outcomes cannot be established. Second, the threshold at which an infected but encapsulated device should be retained and treated medically rather than extracted remains undefined, and is the single management question most likely to change practice.
12. Conclusions
Leadless pacemakers have reduced several infection risks associated with conventional transvenous pacing systems and provide an important option for patients requiring pacing after removal of infected devices. However, infection remains possible, including with newer atrial and helix-fixation systems, and its management is not yet standardised: retention with antibiotics has succeeded for transient bacteraemia, whereas an established device vegetation generally requires extraction. Diagnosis can be challenging because typical signs of generator pocket infection are absent, making clinical suspicion, microbiological investigation, formal endocarditis criteria, and cardiac imaging particularly important. Device retrieval presents additional difficulties, especially after prolonged implantation when tissue encapsulation may increase procedural risk, although contemporary series show that chronic devices can usually be removed percutaneously. Larger prospective studies are needed to improve risk assessment, guide treatment decisions, and establish safer approaches to device removal and reimplantation, particularly for atrial and dual-chamber systems.
Author Contributions
Conceptualization, I.A. and M.Z.; methodology, I.A. and A.A.; investigation and literature review, I.A., A.A., F.M., O.F., E.J. and R.C.; writing, original draft preparation, I.A. and A.A.; writing, review and editing, I.A., A.A., F.M., O.F., E.J., R.C., G.M. and M.Z.; supervision, G.M. and M.Z. All authors have read and agreed to the published version of the manuscript.
Funding
The British Heart Foundation (CH/12/1/29419) supports the University of Leicester, with funding matched to this award to Mustafa Zakkar. Additional support was provided by the Leicester NIHR Biomedical Research Centre (NIHR203327).
Institutional Review Board Statement
Not applicable. This review did not involve new studies on human participants or animals.
Informed Consent Statement
Not applicable.
Data Availability Statement
No new data were created or analyzed in this study. Data sharing is not applicable to this article.
Acknowledgments
ChatGPT-6 (OpenAI, San Francisco, CA, USA) was used solely to assist with language editing, grammar, and improving the clarity and readability of the manuscript. All AI-assisted content was critically reviewed and verified by the authors, who take full responsibility for the accuracy, integrity, and scientific content of the manuscript. Generative AI was not used for data collection, statistical analysis, or interpretation of the results.
Conflicts of Interest
The authors declare no conflicts of interest.
References
- Blomström-Lundqvist, C.; Traykov, V.; Erba, P.A.; Burri, H.; Nielsen, J.C.; Bongiorni, M.G.; Poole, J.; Boriani, G.; Costa, R.; Deharo, J.C.; et al. European Heart Rhythm Association (EHRA) international consensus document on how to prevent, diagnose, and treat cardiac implantable electronic device infections-endorsed by the Heart Rhythm Society (HRS), the Asia Pacific Heart Rhythm Society (APHRS), the Latin American Heart Rhythm Society (LAHRS), International Society for Cardiovascular Infectious Diseases (ISCVID) and the European Society of Clinical Microbiology and Infectious Diseases (ESCMID) in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS). Europace 2020, 22, 515–549. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Cha, Y.M.; El-Chami, M.F.; Liu, C.F.; Andreychuk, L.J.; Beaver, T.M.; Bergen, K.M.; Berul, C.I.; Birgersdotter-Green, U.M.; Breitenstein, A.; Epstein, L.M.; et al. 2026 HRS/AHA/APHRS/EHRA/IDSA/LAHRS/PACES/STS expert consensus statement update on cardiovascular implantable electronic device lead management and extraction. Heart Rhythm. 2026, 23, e1716–e1792. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Zimetbaum, P.J.; Ferro, E.G.; Secemsky, E.A.; Karchmer, A.W.; Kramer, D.B. Complications Associated with Transvenous Cardiac Implantable Electronic Devices: Recognition and Management: A Narrative Review. Ann. Intern. Med. 2025, 178, 1605–1615. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Reynolds, D.; Duray, G.Z.; Omar, R.; Soejima, K.; Neuzil, P.; Zhang, S.; Narasimhan, C.; Steinwender, C.; Brugada, J.; Lloyd, M.; et al. A Leadless Intracardiac Transcatheter Pacing System. N. Engl. J. Med. 2016, 374, 533–541. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Reddy, V.Y.; Exner, D.V.; Cantillon, D.J.; Doshi, R.; Bunch, T.J.; Tomassoni, G.F.; Friedman, P.A.; Estes, N.A.M.; Ip, J.; Anderson, J.; et al. Percutaneous Implantation of an Entirely Intracardiac Leadless Pacemaker. N. Engl. J. Med. 2015, 373, 1125–1135. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- El-Chami, M.F.; Garweg, C.; Clementy, N.; Al-Samadi, F.; Iacopino, S.; Martinez-Sande, J.L.; Roberts, P.R.; Tondo, C.; Johansen, J.B.; Vinolas-Prat, X.; et al. Leadless pacemakers at 5-year follow-up: The Micra transcatheter pacing system post-approval registry. Eur. Heart J. 2024, 45, 1241–1251. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Reddy, V.Y.; Exner, D.V.; Doshi, R.; Tomassoni, G.; Bunch, T.J.; Friedman, P.; Estes, N.A.M.; Neužil, P.; de la Concha, J.F.; Cantillon, D.J. 1-Year Outcomes of a Leadless Ventricular Pacemaker: The LEADLESS II (Phase 2) Trial. JACC Clin. Electrophysiol. 2023, 9, 1187–1189. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Garweg, C.; Chinitz, J.S.; Marijon, E.; Haeberlin, A.; Winter, S.; Iacopino, S.; Curnis, A.; Breitenstein, A.; Hussin, A.; Mela, T.; et al. A leadless ventricular pacemaker providing atrioventricular synchronous pacing in the real-world setting: 12-Month results from the Micra AV post-approval registry. Heart Rhythm 2024, 21, 1939–1947. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Knops, R.E.; Ip, J.E.; Doshi, R.; Exner, D.V.; Defaye, P.; Canby, R.; Bongiorni, M.G.; Shoda, M.; Hindricks, G.; Neužil, P.; et al. One-Year Safety and Performance of a Dual-Chamber Leadless Pacemaker. Circ. Arrhythm. Electrophysiol. 2025, 18, e013619. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Agarwal, S.; Patel, H.P.; Asad, Z.U.A.; Munir, M.B.; Sugrue, A.; Tan, N.Y.; Munoz, F.D.; Padmanabhan, D.; DeSimone, D.; Mulpuru, S.K.; et al. Real-World Outcomes of Leadless Pacemaker Implantation After Transvenous Lead Removal for Infected Cardiac Implantable Electronic Devices in the United States. Am. J. Cardiol. 2025, 253, 10–13. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- El-Chami, M.F.; Bonner, M.; Holbrook, R.; Stromberg, K.; Mayotte, J.; Molan, A.; Sohail, M.R.; Epstein, L.M. Leadless pacemakers reduce risk of device-related infection: Review of the potential mechanisms. Heart Rhythm 2020, 17, 1393–1397. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- El-Chami, M.F.; Soejima, K.; Piccini, J.P.; Reynolds, D.; Ritter, P.; Okabe, T.; Friedman, P.A.; Cha, Y.M.; Stromberg, K.; Holbrook, R.; et al. Incidence and outcomes of systemic infections in patients with leadless pacemakers: Data from the Micra IDE study. Pacing Clin. Electrophysiol. 2019, 42, 1105–1110. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- El-Chami, M.F.; Johansen, J.B.; Zaidi, A.; Faerestrand, S.; Reynolds, D.; Garcia-Seara, J.; Mansourati, J.; Pasquie, J.L.; McElderry, H.T.; Roberts, P.R.; et al. Leadless pacemaker implant in patients with pre-existing infections: Results from the Micra postapproval registry. J. Cardiovasc. Electrophysiol. 2019, 30, 569–574. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Soleimani, H.; Dastjerdi, P.; Hosseini Mohammadi, N.S.; Najdaghi, S.; Shayesteh, H.; Narimani Davani, D.; Yarmohammadi, H.; Eslami, M.; Mollazadeh, R. Safety and efficacy of leadless pacemakers: A systematic review and meta-analysis. Heart Rhythm O2 2026, 7, 598–618. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Antoun, I.; Alkhayer, A.; Somani, R. Leadless Pacemakers: Clinical Advances, Outcomes, and Comparison with Conventional Systems. Br. J. Hosp. Med. 2026, 87, 50278. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Mourad, A.; Niehaus, E.D.; Fowler, V.G., Jr.; Holland, T.L. Leadless pacemaker infective endocarditis: A case report and review of the literature. Clin. Microbiol. Infect. 2024, 30, 1199–1201. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Koay, A.; Khelae, S.; Wei, K.K.; Muhammad, Z.; Mohd Ali, R.; Omar, R. Treating an infected transcatheter pacemaker system via percutaneous extraction. HeartRhythm Case Rep. 2016, 2, 360–362. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Nozoe, M.; Yoshida, D.; Nagatomo, D.; Suematsu, N.; Kubota, T.; Okabe, M.; Yamamoto, Y. Successful percutaneous retrieval of a micra transcatheter pacing system at 8 weeks after implantation. J. Arrhythm. 2018, 34, 653–655. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ellison, K.; Hesselson, A.; Ayoub, K.; Leung, S.; Gurley, J. Retrieval of an infected leadless pacemaker. HeartRhythm Case Rep. 2020, 6, 863–866. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Okada, A.; Shoda, M.; Tabata, H.; Kobayashi, H.; Shoin, W.; Okano, T.; Yoshie, K.; Kato, K.; Motoki, H.; Kuwahara, K. Simultaneous infection of abandoned leads and newly implanted leadless cardiac pacemaker: Why did this occur? J. Cardiol. Cases 2021, 23, 35–37. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Adamek, K.E.; Haque, N.; Martinez-Parachini, J.R.; Ayoub, K.; Richardson, T.D. Percutaneous aspiration and removal of infected leadless pacemaker vegetation. J. Cardiovasc. Electrophysiol. 2022, 33, 2658–2662. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Mor-Welch, T.; Butcher, C.; Sporton, S.; Dhinoja, M. The use of a needle’s eye snare to extract a leadless pacemaker in a patient with a persistent bacteremia. HeartRhythm Case Rep. 2022, 8, 699–701. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Patel, H.; Harrell, S.; Hreibe, H.; Sharkawi, M.; AlJaroudi, W. A Rare Case of Persistent Bacteremia: Leadless Micra Pacemaker Endocarditis. Case Rep. Cardiol. 2023, 2023, 8326020. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Jia, K.Q.; Beerkens, F.J.; Zhang, C.; Itagaki, S.; Dukkipati, S.; Reddy, V.Y.; Musikantow, D.R. Fungal Leadless Pacemaker Endocarditis Managed by Percutaneous Vacuum-Assisted Evacuation of Vegetation, Pacemaker Extraction, and Replacement. JACC Case Rep. 2024, 29, 102586. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Karki, S.; Rosenberg, J.; Abraham, N.; Fatima, A.; Kawasaki, R.; Metzl, M. First Reported Case of Infection Involving a Leadless Atrial Pacemaker. JACC Case Rep. 2026, 31, 106071. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Nakata, K.; Inamura, Y.; Ohya, H.; Takagi, T.; Matsumura, Y.; Inaba, O. Late-onset infection of helix-fixation leadless pacemaker diagnosed by pacing threshold elevation and successfully managed with retrieval and reimplantation. HeartRhythm Case Rep. 2026, 12, 331–333. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Phan, J.; Al Solaia, N.; Aydin, A.; Alkadri, J.; Golian, M. Double trouble in device infection: Concomitant transvenous and leadless pacemaker infection. HeartRhythm Case Rep. 2026, 12, 1054–1057. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Fowler, V.G.; Durack, D.T.; Selton-Suty, C.; Athan, E.; Bayer, A.S.; Chamis, A.L.; Dahl, A.; DiBernardo, L.; Durante-Mangoni, E.; Duval, X.; et al. The 2023 Duke-International Society for Cardiovascular Infectious Diseases Criteria for Infective Endocarditis: Updating the Modified Duke Criteria. Clin. Infect. Dis. 2023, 77, 518–526. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Darlington, D.; Brown, P.; Carvalho, V.; Bourne, H.; Mayer, J.; Jones, N.; Walker, V.; Siddiqui, S.; Patwala, A.; Kwok, C.S. Efficacy and safety of leadless pacemaker: A systematic review, pooled analysis and meta-analysis. Indian Pacing Electrophysiol. J. 2022, 22, 77–86. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Habboush, S.; Elmoursi, A.; Gadelmawla, A.F.; Masoud, A.T.; Khalil, M.; Sheashaa, H.; Merza, N.; Massoud, A.T. Transvenous Compared with Leadless Pacemakers: A meta-analysis comparing TP versus LP. Cardiol. Rev. 2026, 34, 123–129. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Dai, H.; Liu, H.; Gao, C.; Han, J.; Meng, J.; Liu, P.; Zhang, M.; Li, D.; Guo, W. Comparison of Postoperative Outcomes between Leadless and Conventional Transvenous Pacemakers Implantation: An Up-to-Date Meta-analysis. Rev. Cardiovasc. Med. 2024, 25, 359. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Agarwal, S.; Patel, H.P.; Asad, Z.U.A.; Munir, M.B.; Sugrue, A.; Tan, N.Y.; Munoz, F.D.; Padmanabhan, D.; DeSimone, D.; Mulpuru, S.K.; et al. Leadless pacemaker vs transvenous pacemaker in patients undergoing transvenous lead removal for infected cardiac implantable electronic devices in the United States. Heart Rhythm 2025, 22, e813–e815. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Gad, A.S.; Eissa, O.A.; Awad, L.; Elkhouly, A.; Altaamreh, S.M.; Saeed, T.Z. Leadless vs. Conventional Transvenous Pacemakers in Valve Intervention Patients: A Systematic Review & Meta-Analysis. Pacing Clin. Electrophysiol. 2026, 49, 435–448. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Andreae, A.E.; Harowicz, M.R.; Koweek, L.M.; Henry, T.S.; Al-Khatib, S.M.; Frazier-Mills, C.; Friedman, D.J.; Pokorney, S.D.; Loring, Z.; Solomon, J.B.; et al. Frequency and Extent of Leadless Pacemaker Encapsulation. JACC Clin. Electrophysiol. 2026, ahead of print. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Da Costa, A.; Groussin, P.; Barengo, A.; Yvorel, C.; Mohammed, R.; Romeyer, C.; Boukhris, M.; Benali, K. Cardiac implantable electronic device infections: Emerging paradigms in precision prevention and personalized management. Trends Cardiovasc. Med. 2026, ahead of print. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Antoun, I.; Alkhayer, A.; Abdelrazik, A.; Eldesouky, M.; Thu, K.M.; Dhutia, H.; Somani, R.; Ng, G.A. Artificial Intelligence and the Future of Cardiac Implantable Electronic Devices: Diagnostics, Monitoring, and Therapy. J. Clin. Med. 2025, 14, 8824. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Nuaimi, M.; Ubosy, S.; Crespo, A.; Tello, D.; Carlan, S. Mycobacterium smegmatis Infection in a Cardiovascular Implantable Electronic Device in an Immunocompetent Patient. Cureus 2026, 18, e115768. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Rommens, O.; Willems, R.; Voros, G.; Frederiks, P.; Garweg, C. Leadless pacemaker tine fracture and dislodgement in infective tricuspid valve endocarditis: A case report. Eur. Heart J. Case Rep. 2025, 9, ytaf480. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Chinybayeva, A.; Marat, G.; Suleymen, Z.; Abdrakhmanov, A. Case Report: Sequential leadless pacing after battery depletion in a patient with recurrent device-related infection. Front. Cardiovasc. Med. 2026, 13, 1856082. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Yang, J.D.; Li, R.G.; Liu, X.B.; Xue, X.L.; Zhang, J.H.; Hu, Y.M.; Zhang, B.J.; Tong, L.; Luo, H.; Shen, M.; et al. Multicenter Study of the Electrical Characteristics and Short-Term Outcomes of the Aveir VR Leadless Pacemaker. J. Am. Heart Assoc. 2026, 15, e050579. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Delgado, V.; Ajmone Marsan, N.; de Waha, S.; Bonaros, N.; Brida, M.; Burri, H.; Caselli, S.; Doenst, T.; Ederhy, S.; Erba, P.A.; et al. 2023 ESC Guidelines for the management of endocarditis. Eur. Heart J. 2023, 44, 3948–4042. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Matsumoto, K.; Tanaka, N.; Mori, H.; Ikeda, Y.; Kato, R. Intracardiac Echocardiography-Guided Implantation of Aveir VR Leadless Pacemaker. J. Arrhythm. 2026, 42, e70303. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Kikuchi, S.; Takada, Y.; Satomi, K.; Yazaki, Y. Contrast-Free Dual-Chamber Leadless Pacemaker Implantation Using Intracardiac Echocardiography and 3-Dimensional Electroanatomical Mapping. JACC Case Rep. 2026, 31, 110246. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Baldauf, B.; Bode, K.; Biffi, M.; Rapacciuolo, A.; Zucchelli, G.; Bongiorni, M.G.; Casorelli, E.; Mitacchione, G.; Hohendanner, F.; Mangoni, E.D.; et al. Delphi consensus recommendations for preventing and treating cardiac implantable electronic device infections beyond current guidelines. Sci. Rep. 2026, 16, 13135. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- O’Neill, P.; Osler, B.; Junarta, J.; Zivan, T.; Hoeltzel, G.; Ford, R.; Flomenberg, P.; Greenspon, A.; Mehrotra, P. Combined transcatheter vegetectomy and leadless pacemaker implantation for endocarditis and complete heart block. Future Cardiol. 2024, 20, 359–363. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Uhor, F.; Haeberlin, A.; Knapp, J. Update on the perioperative management of cardiac implantable electronic devices. Anaesthesiologie 2026, 75, 287–300. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Tajstra, M.; Dyrbuś, M.; Kowalski, O.; Sterliński, M.; Breitenstein, A.; Hofer, D. Transvenous leadless pacemaker extraction: When, who, where, and how? Kardiol. Pol. 2026, 84, 174–187. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Funasako, M.; Hála, P.; Janotka, M.; Šorf, J.; Machová, L.; Petrů, J.; Chovanec, M.; Škoda, J.; Šedivá, L.; Šimon, J.; et al. Transcatheter non-acute retrieval of the tine-based leadless ventricular pacemaker. Europace 2024, 26, euae256. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Jacon, P.; Benkilani, M.; Bisson, A.; Marijon, E.; Da Costa, A.; Haddad, C.; Behar, N.; Cesari, O.; Lellouche, N.; Mansourati, V.; et al. Micra leadless pacemaker revisions: Incidence, characteristics, and outcomes from a multicentre French cohort. Europace 2025, 27, euaf291. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Guo, Q.; Gao, H.; Deng, Z.; Li, S.; Zhou, S.; Wang, J.; Chen, Y. Lessons learned from Micra leadless pacemaker retrieval: A case series. Eur. Heart J. Case Rep. 2026, 10, ytag539. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Horbal, P.J.; Savona, S.; Afzal, M.R.; Augostini, R.; Gunda, S. First case report of leadless pacemaker snaring and replacement via internal jugular approach. HeartRhythm Case Rep. 2026, 12, 766–768. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Zhang, Y.; Li, K.; Chen, A.L.; Yang, W.; Chen, X.; Liu, Z. Extraction of Micra Leadless Pacemaker and Upgrade to Left Bundle Branch Pacing: A Case Report. Pacing Clin. Electrophysiol. 2026, ahead of print. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Parkash, R.; Abdelwahab, A. Helix fracture during atrial leadless pacemaker explantation. HeartRhythm Case Rep. 2026, 12, 1039–1042. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Neuzil, P.; Exner, D.V.; Knops, R.E.; Cantillon, D.J.; Defaye, P.; Banker, R.; Friedman, P.; Hubbard, C.; Delgado, S.M.; Bulusu, A.; et al. Worldwide Chronic Retrieval Experience of Helix-Fixation Leadless Cardiac Pacemakers. J. Am. Coll. Cardiol. 2025, 85, 1111–1120. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Sahyouni, A.; Jimenez, A.; Alrahal, J.; Knight, A.; Chen, R.; Maddineni, T.; Lugo, L.A.; Koneru, A.; Ramey, K.; Subbiondo, R.J. Comparative Analysis of Extraction Complications in Transvenous Versus Leadless Pacemakers. Cureus 2026, 18, e108392. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Zaidi, S.M.; Supple, G.; Marchlinski, F.E.; D’Souza, B.; Berkowitz, E.; Rashba, E.; Katz, M.; Feigenblum, D.; Schaller, R.D. Percutaneous retrieval of embolized atrial leadless pacemakers: Early multicenter experience. Heart Rhythm 2026, 23, 2470–2475. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Fukui, T.; Saeki, H.; Miyoshi, M.; Ogasawara, N. How Should an Embolized Atrial Leadless Pacemaker Be Retrieved from the Pulmonary Artery? Pacing Clin. Electrophysiol. 2026, 49, 1744–1747. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Yasuoka, R.; Tanaka, M.; Noda, T.; Nakazawa, G.; Kurita, T. Intraprocedural tail-first pulmonary artery migration of a leadless pacemaker: A protective sleeve-assisted helix-side retrieval strategy-a case report. Eur. Heart J. Case Rep. 2026, 10, ytag657. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Tan, M.C.; Tan, J.L.; Tay, S.T.; Sorajja, D.; Scott, L.; Cha, Y.M.; Russo, A.M.; Hussein, A.; Lee, J.Z. A Systematic Review of Short-Term Outcomes of Leadless Pacemaker Implantation After Transvenous Lead Removal of Infected Cardiac Implantable Electronic Device. Am. J. Cardiol. 2023, 203, 444–450. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Nadeem, B.; Sedrakyan, S.; Fatima, A.; Baig, M.M.A.; Ahmed, A.; Sherwani, M.R.K.; Wylie, J. Outcomes of concurrent and delayed leadless pacemaker implantation following extraction of infected cardiovascular implantable electronic device. J. Interv. Card. Electrophysiol. 2025, 68, 1523–1529. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Inoue, N.; Ito, Y.; Imaizumi, T.; Morikawa, S.; Murohara, T. Assessment of adverse events stratified by timing of leadless pacemaker implantation with cardiac implantable electronic devices extraction due to infection: A systematic review and meta-analysis. J. Arrhythm. 2025, 41, e13208. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Maille, B.; Behar, N.; Jacon, P.; Hourdain, J.; Franceschi, F.; Koutbi, L.; Tovmassian, L.; Bierme, C.; Seder, E.; Klein, V.; et al. Two-in-one procedure for transvenous lead extraction and leadless pacemaker reimplantation in pacemaker-dependent patients with device infection: Streamlined patient flow. Europace 2024, 26, euae162. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Hamayel, H.; Awwad, F.; Abutaqa, M.; Baker, M.A.; Ramadan, A.; Daralammouri, Y.; Marie, I. Combined Lead Extraction and Leadless VDD Pacemaker Implantation in d-TGA Patient After Senning Procedure. JACC Case Rep. 2026, 31, 106866. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Sileshi, R.M.; Mathes, K.; Kosuri, S.; Gupta, A.; Phillips, N.; Pimentel, R.; Emert, M.; Soma, S.; Ramirez, R.; Dendi, R.; et al. A Reassessment of Prophylactic Antibiotics in Low-Infection-Risk Electrophysiology Procedures. Pacing Clin. Electrophysiol. 2026, ahead of print. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Defaye, P.; Reddy, V.Y.; Ip, J.E.; Doshi, R.N.; Exner, D.V.; Canby, R.C.; Bongiorni, M.G.; Shoda, M.; Hindricks, G.; Rashtian, M.Y.; et al. Atrioventricular synchrony maintained by a dual-chamber leadless pacemaker in real-world conditions. Europace 2026, 28, euag012. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lo, M.; Amin, A.K.; Kowalski, M.; Kroman, A.; Mathew, D.; Parikh, V.; Weiner, S.; Joseph, J.M.; Ganz, L.; Mortada, M. Leadless Versus Transvenous Dual-Chamber Pacemakers: Real-World Evidence from AVEIR DR Coverage with Evidence Development Study. J. Cardiovasc. Electrophysiol. 2026, 37, 544–553. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Velaga, S.T.; Jia, K.; Rimsky, E.; Saouma, S.; Shah, R.; Akhrass, P.; Kowalski, M.; Parikh, V. Procedural and short-term performance outcomes of dual-chamber leadless pacemakers from a single-center experience. J. Interv. Card. Electrophysiol. 2026, 69, 627–635. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Nilsson, K.R., Jr.; Birgersdotter-Green, U.; Saleem-Talib, S.; Bettampadi, D.; Yang, Z.; Diaz, J.O.; Gabriels, J.K.; Schaller, R.D.; Shukla, H.H.; Kühne, M.; et al. Leadless atrial versus transvenous pacing for sinus node dysfunction: 1-year outcomes from the leadless ARRIVE real-world study. Europace 2026, 28, euag072. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Knops, R.E.; Lloyd, M.S.; Roberts, P.R.; Wright, D.J.; Boersma, L.V.A.; Doshi, R.; Friedman, P.A.; Neuzil, P.; Blomstrom-Lundqvist, C.; Bongiorni, M.G.; et al. A modular communicative leadless pacing-defibrillator system. N. Engl. J. Med. 2024, 391, 1402–1412. [Google Scholar] [CrossRef] [Scilit] [PubMed]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.