Abstract
Background: Needle arthroscopy can alter not only portal size but also the diagnostic and treatment pathway by enabling direct visualization under local anesthesia and selected limited interventions. Its diffusion across joints, however, may be outpacing clinical validation. Methods: This PRISMA-ScR-guided review mapped PubMed/MEDLINE-indexed primary publications from database inception through 31 July 2026. The quantitative flow is therefore an auditable but incomplete evidence map rather than a complete census of the technology; studies indexed exclusively outside PubMed may have been missed. Backward reference-list screening was used as a completeness check, while earlier Scopus, Web of Science, and Cochrane histories were retained as sensitivity sources and not pooled into the quantitative flow. Two reviewers independently screened and charted records using a shared form. Reports were de-duplicated by PMID, DOI, and normalized title and characterized by anatomy, design, application, platform, outcomes, harms reporting, study-design hierarchy, and principal methodological limitation. Technical or cadaveric feasibility was analyzed separately from clinical and comparative evidence. Results: The final map comprised 105 unique primary publications: 49 technical, 41 clinical, and 15 cadaveric. Technical and cadaveric reports (64/105, 61.0%) outnumbered clinical reports (41/105, 39.0%), limiting estimation of clinical effectiveness. Ninety publications (85.7%) appeared from 2021 onward; the 2026 count is incomplete because the search ended on 31 July. For descriptive display across the complete map, nine reports (8.6%) were in Levels I–II and 93 (88.6%) in IV–V; among the 41 clinical reports, 29 (70.7%) were IV–V and 12 (29.3%) were I–III. NanoScope accounted for 87 of 107 platform mentions (81.3%). Diagnostic and second-look applications had the most consistent comparative rationale, but lesion-specific estimates varied by joint and reference standard. Reported harms were generally minor where surveillance was described, but rare-harm inference was limited by small samples and inconsistent adverse-event ascertainment. Conclusions: The field remains promising but unproven: technical breadth exceeds comparative clinical validation. Because the quantitative evidence map is PubMed/MEDLINE-only and dominated by technical, cadaveric, and IV–V reports, current evidence is insufficient to estimate clinical effectiveness or justify replacement of MRI or conventional arthroscopy. Needle arthroscopy should therefore be viewed as an indication-specific complement and a hypothesis-generating technology rather than a replacement platform. Multicentre comparative studies and randomized trials, where ethically and practically feasible, are needed to test patient benefit and clinical application.
1. Introduction
Needle arthroscopy (NA) is more than a smaller arthroscope. By combining needle-sized access, portable imaging, and miniaturized instruments, it can move direct intra-articular visualization from the operating room to an office, procedure room, or wide-awake surgical pathway. The literature uses overlapping labels such as nano-arthroscopy, nanoscopy, in-office needle arthroscopy, and office-based arthroscopy. In this review, needle arthroscopy is the umbrella term; alternative labels are retained only when they identify the wording of an individual report [1,2,3,4,5,6,7].
This shift creates three distinct value propositions that should not be conflated. First, NA may provide problem-solving diagnosis when symptoms, examination, and magnetic resonance imaging (MRI) disagree. Second, it may reduce access morbidity and anesthetic burden for selected low-complexity procedures. Third, it may reorganize the care pathway by shortening time to a definitive decision, reducing operating-room dependence, or avoiding downstream imaging. A device can be feasible without being diagnostically valid, diagnostically valid without improving patient outcomes, and clinically effective without being cost-effective in a different health system.
The literature has expanded from diagnostic knee and shoulder visualization to meniscal surgery, cartilage assessment, biceps procedures, labral repair, ankle impingement, ligament stabilization, second-look assessment, septic-joint management, and fracture reduction. Yet the evidentiary requirements change along this translation pathway. Cadaveric visibility and technical completion establish possibility; comparative accuracy establishes diagnostic performance; randomized or well-controlled studies are needed for clinical-effectiveness claims; and implementation studies are needed to determine whether a new pathway is scalable and economically defensible.
Earlier reviews were predominantly diagnostic, platform-specific, or joint-focused [2,3,6,7]. A 2026 meta-analysis of nine comparative studies (503 patients) reported similar procedural completion, diagnostic success, and complication rates relative to conventional arthroscopy, but certainty ranged from moderate to very low and evidence for therapeutic equivalence remained limited [4]. A separate safety review identified only minor reported complications, although the underlying studies were observational and not designed to exclude uncommon harms [5]. The unresolved question is therefore not whether NA can be used, but for which indication, at what point in the pathway, and with what level of evidentiary confidence.
This scoping review addressed four questions: (1) how is the evidence distributed across joints, applications, study designs, and platforms; (2) which use cases have progressed from technical feasibility to comparative clinical validation; (3) what limits inference regarding safety, generalizability, and implementation; and (4) which research priorities are most likely to convert technical innovation into clinically meaningful evidence.
2. Materials and Methods
2.1. Review Design and Eligibility Framework
The review was conducted and reported in accordance with the Preferred Reporting Items for Systematic Reviews and Meta-Analyses extension for Scoping Reviews (PRISMA-ScR) [1]. The population–concept–context framework comprised patients or cadaveric specimens representing orthopedic joints (population), needle arthroscopic visualization or intervention (concept), and office, outpatient, operating-room, laboratory, or cadaveric settings (context). Because the purpose was to characterize both clinical use and the translation pipeline, cadaveric and technical reports were eligible but were analyzed separately from patient-level clinical evidence. The protocol was not prospectively registered. The completed PRISMA-ScR checklist is provided as Supplementary Table S2 for publication alongside the manuscript, and the audited study-selection flow is presented in Figure 1.
Figure 1.
Audited but incomplete PubMed/MEDLINE study-selection flow. Counts were reconstructed from the two-stage PubMed/MEDLINE source workbook and corrected after record-level de-duplication. The included corpus comprised 49 technical, 15 cadaveric, and 41 clinical publications. Earlier Scopus, Web of Science, and Cochrane histories were not reconciled or pooled; studies indexed exclusively outside PubMed may therefore have been missed.
Eligible primary publications reported an orthopedic diagnostic, therapeutic, technical, feasibility, accuracy, safety, workflow, or economic application of needle arthroscopy. Randomized and non-randomized clinical reports, diagnostic-accuracy studies, case series or reports, technical notes, and cadaveric investigations were included. Veterinary reports, non-orthopedic endoscopy, conference abstracts without a full report, editorials, letters, and review articles were excluded from the primary evidence map. Reviews were retained only to frame prior knowledge and to cross-check reference lists. A “report” denotes one publication; it does not necessarily represent an independent cohort.
2.2. Information Source and Search Strategy
The auditable primary search used PubMed/MEDLINE from database inception through 31 July 2026. Five terms were entered as separate PubMed all-fields searches and then merged: NanoScope; nanoscopy; “needle arthroscopy”; microarthroscopy; and “in-office arthroscopy”. No language, species, or date filter was applied. The original search contributed 2190 records, and the update contributed 460 records. The archived workbook does not preserve term-level hit counts or the exact execution date of the original search, so those details cannot be reconstructed retrospectively without inventing data. The complete executable terms, available counts, and audit rules are provided in Supplementary Table S1.
Backward reference-list screening of included reports and relevant reviews was used as a completeness check. Candidate records were assessed against the same criteria and were retained in the final map only after PMID/DOI verification. Earlier search histories in Scopus, Web of Science, and the Cochrane Library used the same conceptual blocks as sensitivity sources; because their July 2026 record-level exports could not be fully reconciled with the PubMed dataset, they were not used to calculate the quantitative PRISMA flow. The quantitative results should therefore be interpreted as an auditable but incomplete PubMed/MEDLINE-indexed evidence map: studies indexed exclusively outside PubMed may have been missed, and not every published analysis of needle arthroscopy is necessarily represented.
2.3. Selection, De-Duplication, and Data Extraction
Two reviewers independently screened titles and abstracts and prepared candidate inclusion lists; potentially eligible full texts were then assessed independently, with disagreements resolved by consensus. Data charting was performed independently in duplicate using a shared form and reconciled before synthesis. The historical and updated lists were re-audited at record level. Publications with the same PMID, DOI, or normalized title were treated as one report. This process removed three duplicate entries (PMID 39682041, 38235751, and 39077809), completed one previously uncharacterized retrieved study (PMID 40168804), restored the eligible Deirmengian report (PMID 29494711) identified by Reviewer 3, and verified PubMed identifiers for all 105 included publications. Four included reports were co-authored by one or more review authors (Błażej Wójtowicz, Marcin Domżalski, or Jędrzej Lesman; PMIDs 39165867, 41206227, 41630068, and 42443948). For each of these records, the study-design, evidence-architecture, and translation-stage classifications were assigned by a reviewer who was not a co-author of that report and then independently checked by the other member of the screening pair; the archived log retained the pair-level process but not report-level reviewer initials, so we do not retrospectively attribute individual records to named reviewers. These reports were not used as the sole basis for any conclusion.
Data charting captured year, report design, anatomical region, indication, sample size, platform, principal findings, harms reporting, study-design category, and the main methodological limitation. We distinguished a publication/report from an independent study or cohort; the latter could not be established reliably for every specialist-group series. A pooled patient total was intentionally not calculated because the corpus included technical procedures, cadaveric specimens, potentially overlapping cohorts, and multi-joint populations; summing these denominators would imply a precision and independence that the evidence does not possess.
2.4. Study-Design Hierarchy and Structured Methodological Characterization
For descriptive comparison of patient-level clinical reports, study design was mapped to the Oxford Centre for Evidence-Based Medicine Levels of Evidence (March 2009) [6]. Sublevels were collapsed to five categories: I, randomized trial; II, prospective comparative or diagnostic-accuracy report with an appropriate reference standard; III, retrospective comparative or prospective non-randomized comparative report; IV, case series or cohort without a comparator; and V, case report or report without clinical comparative evidence. Legacy “IIb” entries were mapped to Level II. Technical, bench, and cadaveric reports were displayed in the V category for the purpose of the evidence-architecture figure only and were not interpreted as clinical Level V evidence. These categories are a study-design hierarchy, not a risk-of-bias, certainty-of-evidence, or recommendation grade.
We did not perform a formal cross-design risk-of-bias assessment because the archived extraction did not preserve complete item-level judgments for every report. Instead, design-matched prompts from RoB 2, QUADAS-2, and MINORS informed a structured methodological characterization. For each report, reviewers recorded whether a comparator or reference standard was present, whether outcome assessment was blinded or independent, whether follow-up and adverse-event surveillance were defined, whether a denominator was available, and the principal limitation. The full characterization is provided in Supplementary Table S1; “not reported” was kept distinct from “reported absent”.
The translation stage was assigned at the report level using explicit rules: Stage 1, technical or preclinical feasibility without patient outcomes; Stage 2, clinical feasibility demonstrated in patients without a comparative reference; Stage 3, comparative diagnostic or therapeutic validation with a reference standard or comparator; and Stage 4, implementation evidence involving multicentre workflow, learning-curve, resource, or economic outcomes. When a report contained more than one element, the highest stage directly supported by its design and outcomes was recorded. These stages are interpretive descriptors of the literature and should not be read as validated measures of evidence maturity.
2.5. Synthesis
The synthesis used three complementary axes: anatomical distribution, study architecture (technical, cadaveric, or clinical; design hierarchy; platform), and clinical function (diagnostic or second-look, therapeutic, or care-pathway implementation). Claims from feasibility reports were deliberately separated from comparative effectiveness. Meta-analysis was not planned because interventions, comparators, outcomes, and follow-up were too heterogeneous and because the review question concerned evidence mapping rather than a single pooled treatment effect.
3. Results
3.1. Study Selection
The two-stage PubMed/MEDLINE search yielded 2650 records. Initial relevance filtering retained 275 records; 115 duplicate, veterinary, or otherwise ineligible records were removed before full-text assessment. Of 160 full-text reports assessed, 55 were excluded. The final evidence map contained 105 unique primary publications [8,9,10,11,12,13,14,15,16,17,18,19,20,21,22,23,24,25,26,27,28,29,30,31,32,33,34,35,36,37,38,39,40,41,42,43,44,45,46,47,48,49,50,51,52,53,54,55,56,57,58,59,60,61,62,63,64,65,66,67,68,69,70,71,72,73,74,75,76,77,78,79,80,81,82,83,84,85,86,87,88,89,90,91,92,93,94,95,96,97,98,99,100,101,102,103,104,105,106,107,108,109,110,111,112]. The manuscript, figures, and Supplementary Table S1 were reconciled against the same PMID-verified dataset (Figure 1). Because the quantitative flow was limited to auditable PubMed/MEDLINE source files, Figure 1 represents an incomplete evidence map rather than an exhaustive census of the field.
3.2. Distribution of the Evidence
The knee was the most frequently represented region (n = 35), followed by foot and ankle (n = 25), shoulder and biceps (n = 23), multi-joint applications (n = 9), wrist (n = 7), and elbow (n = 6) (Figure 2). Publication growth was recent and steep: 90 of 105 publications (85.7%) appeared from 2021 onward (Figure 3). The temporal signal therefore reflects rapid diffusion of a platform rather than the accumulation of long-term follow-up.
Figure 2.
Anatomical distribution of the 105 unique included primary publications.
Figure 3.
Publication activity by year. The 2026 value is a partial observation period and includes only publications indexed through 31 July 2026; it should not be interpreted as a decline in publication activity.
3.3. Evidence Architecture and Platforms
The evidence architecture was weighted toward innovation description. Forty-nine publications were technical, 15 cadaveric, and 41 clinical, meaning that technical and cadaveric reports (64/105, 61.0%) outnumbered patient-level clinical reports. Across the complete map, nine of 105 reports (8.6%) fell in Levels I–II and 93 (88.6%) in the IV–V design categories used for descriptive display. In the clinical subset, 29 of 41 reports (70.7%) were IV–V and 12 (29.3%) were I–III. The single Level I report was a multicentre randomized comparison of partial meniscectomy with needle versus standard arthroscopy [38]. These figures describe study design; they are not a validated certainty grade or a direct measure of clinical benefit. The predominance of technical, cadaveric, and low-level clinical evidence makes a pooled estimate of clinical effectiveness inappropriate.
Platform concentration was also substantial. NanoScope accounted for 87 of 107 platform mentions (81.3%), followed by Mi-Eye (n = 12). Four studies did not specify a platform, and one mention each involved VisionScope/VSI, IntraVu/MIDASVu, NeedleScope/Smith & Nephew, and a custom device. Two publications evaluated more than one system. The dominance of a single commercial platform means that observed performance may reflect device-specific optics, instruments, training, and early-adopter expertise rather than a class effect. Conclusions therefore cannot be generalized automatically from the NanoScope literature to all needle-arthroscopy systems (Figure 4).
Figure 4.
Evidence architecture by study category, descriptive design hierarchy, and reported platform. Platform counts are mentions and are not mutually exclusive; the hierarchy is not a certainty-of-evidence grade.
3.4. Joint-Specific Applications
Sensitivity check for author-linked reports. Four included reports were co-authored by review authors. Excluding these four records reduced the foot-and-ankle count from 25 to 23 and the shoulder-and-biceps count from 23 to 21, but did not change the directional conclusions in Table 1: foot and ankle remained characterized by small specialist-centre cohorts and few controls, while shoulder and biceps remained dominated by technical or clinical-feasibility evidence rather than comparative validation. No region changed from a feasibility-dominant to a comparative-validation-dominant architecture. This was a narrative robustness check at the publication level, not a formal leave-one-out analysis of independent cohorts.
Table 1.
Joint-specific evidence architecture. The comparative evidence signal summarizes the strongest design present; it is not a formal certainty grade or recommendation.
3.5. Diagnostic Evidence
Diagnostic and second-look use represented the most coherent clinical role. Eight clinical reports used MRI, conventional arthroscopy, or surgical findings as a reference or comparator, but their joints, lesions, timing, blinding, and denominators differed. Across knee, shoulder, and wrist reports, NA often showed high completion and clinically useful concordance for selected structures [11,32,36,94,105,112,113]. The signal was not uniform: visualization depends on portal trajectory, joint volume, a 0° viewing angle, lesion location, and operator experience. “Diagnostic accuracy” should therefore be interpreted as a joint- and lesion-specific property, not as a platform-wide attribute or proof of improved patient outcomes.
The 2026 comparative meta-analysis found no significant difference in diagnostic success or procedure completion between NA and conventional arthroscopy, but estimates were imprecise, and certainty ranged from moderate to very low [4]. The evidence supports NA most strongly when direct visualization is expected to resolve a management-relevant uncertainty—not as routine duplicate testing after an already definitive MRI. Table 2 summarizes the eight reports with a clinical reference standard; no pooled diagnostic estimate was calculated.
Table 2.
Clinical diagnostic reports with an operative or conventional arthroscopy reference standard. Values are report-level findings; heterogeneity precluded pooling.
3.6. Therapeutic and Perioperative Outcomes
Therapeutic scope expanded to partial meniscectomy and repair, cartilage procedures, ankle impingement, ligament stabilization, biceps tenotomy or tenodesis, labral procedures, tendon release, infection management, and fracture reduction. The evidentiary signal was strongest for short-term perioperative burden rather than durable superiority. In the multicentre randomized knee trial, needle-arthroscopic partial meniscectomy reduced opioid consumption in the first 24 h and improved KOOS pain and IKDC scores at 2 and 6 weeks; between-group differences were absent at 12 weeks [38]. Earlier comparative work similarly suggested less early pain, without establishing long-term non-inferiority [35].
Preference and pathway outcomes were promising but context-dependent. Patients often favoured in-office NA when it shortened time to treatment or avoided general anesthesia, while concern centred on limited long-term and revision evidence [40]. Economic studies suggested that savings may arise from avoided MRI, operating-room use, or downstream visits [114,115,116]. However, most cost models were developed in the United States or in single-centre settings. Their estimates are therefore sensitive to local reimbursement, wage structure, MRI availability, equipment utilization, staffing, sterile-workflow requirements, and whether NA replaced rather than duplicated another test; they should not be transported as universal economic conclusions.
Recent publications broadened technical possibility: a custom 2 mm high-definition device underwent optical and cadaveric evaluation [39]; ankle work extended to operative arthroscopy, impingement, and second-look assessment [43,44,61]; shoulder work included labral repair and acromioplasty [79,85]; regional-anesthesia biceps procedures were described in early feasibility or technical studies [75,78,86,88]; and needle-arthroscopy-assisted tibial plateau reduction improved reduction quality in a cadaveric comparison [41]. These studies define a research frontier; they do not, by themselves, establish routine clinical effectiveness.
3.7. Safety, Learning Curve, and Industry Involvement
The published safety signal was reassuring but not definitive. A 2025 systematic review of 11 studies (1624 patients) reported joint-specific complication ranges of 0–9.68%; reported events were grade I, and vasovagal reactions were most frequent [5]. Most primary studies, however, were too small and followed patients too briefly to exclude uncommon infection, neurovascular injury, iatrogenic chondral injury, or later reoperation. Absence of observed serious harm in small feasibility cohorts is not equivalent to evidence of safety. A statement that “no complications occurred” was retained as a report-level observation, not converted into a class-wide safety estimate; no pooled complication rate was calculated because adverse-event definitions, denominators, and surveillance periods were inconsistent.
Restricted field of view, unfamiliar 0° optics, lower instrument rigidity, fluid-management constraints, and limited reconstructive capability were recurring barriers. Procedural completion is likely to be operator- and indication-dependent, yet case volume, learning curve, conversion thresholds, and failed attempts were inconsistently reported. In the audited disclosure fields, 51 of 105 reports (48.6%) named a manufacturer or industry relationship, 10 (9.5%) explicitly reported no support or no conflict, and 44 (41.9%) did not report enough information to classify the relationship. This classification was based solely on the conflict-of-interest, funding, acknowledgments, equipment-support, or other relationship statements printed in the report or available in its full text; absence of a statement was coded as not reported or unclear, not as evidence that no relationship existed. These counts are descriptive rather than proof of bias, but concentration in specialist centres, overlap of three review authors with included reports, and frequent manufacturer support create plausible risks of industry-related and publication bias. They limit transportability even when technical success is high.
4. Discussion
4.1. Principal Findings
The principal finding is an innovation-to-evidence mismatch. The literature has expanded rapidly in volume, anatomy, and procedural ambition, but 93 of 105 reports (88.6%) were in the IV–V design categories used for descriptive display; 64 reports were technical or cadaveric and only 41 were clinical. Only nine reports (8.6%) were in Levels I–II. In the 41-report clinical subset, 29 (70.7%) were IV–V and 12 (29.3%) were I–III. Low-complexity feasibility work is appropriate during early device development, but technical completion or cadaveric visibility cannot establish diagnostic efficacy, therapeutic non-inferiority, or system-level value. The predominance of non-clinical and low-level evidence makes clinical-effectiveness estimation uncertain. Those claims require comparative patient studies with appropriate reference standards, follow-up, and prespecified outcomes.
The translation signal differed by use case more than by device. Problem-solving diagnosis and planned second-look assessment had a clear mechanism of value and the most consistent comparative support, although estimates remained lesion-specific. Selected low-complexity procedures had a plausible early-morbidity advantage, but durable outcomes were sparse. Complex repair and reconstruction remained predominantly at the technical frontier. The 2026 meta-analysis supports comparable diagnostic and procedural performance in selected comparisons, but certainty ranged from moderate to very low and conventional arthroscopy remains the reference for definitive therapeutic intervention [4]. Our “evidence architecture” and translation stages are structured interpretive descriptors, not validated measures of evidence maturity.
4.2. Clinical Positioning
The following is an evidence-informed hypothesis for clinical positioning, not a recommendation for routine practice. The relevant question is not whether NA can visualize a joint, but whether direct visualization at that point will change management while exposing the patient to less burden than the next-best alternative. NA may be most plausible when symptoms and imaging disagree, MRI is contraindicated or repeatedly non-diagnostic, postoperative tissue status will alter care, or a limited intervention can be completed with a high probability of success using available instrumentation. These scenarios should be tested prospectively rather than treated as established indications. Nothing in the current evidence justifies replacing MRI or conventional arthroscopy as the default diagnostic or therapeutic platform.
Conversely, conventional arthroscopy remains preferable when treatment requires broad visualization, multiple working portals, traction, high instrument rigidity, complex suture management, extensive debridement, or immediate escalation. Consent should explicitly address incomplete visualization, conversion, and the possibility that an office-based diagnostic procedure may still be followed by conventional surgery. This prevents “minimally invasive” from becoming a surrogate for “clinically sufficient.” Table 3 proofs the positioning in evidence based medicine.
Table 3.
Evidence-calibrated clinical positioning of needle arthroscopy.
4.3. Generalizability and Care-Pathway Value
Generalizability is constrained at both the device and operator levels. NanoScope accounted for 87 of 107 platform mentions (81.3%), while many reports originated from high-volume early-adopter groups. A favourable result for one platform, indication, and expert operator should not automatically be interpreted as a class effect. Device-specific optics, instruments, single-use components, training pathways, and manufacturer involvement may all influence observed performance. Replication across independent centres and platforms should therefore report optic diameter and angle, instruments, fluid strategy, setting, anesthesia, operator case volume, technical completion, conversion, and all commercial relationships.
The economic value proposition also requires a pathway rather than a procedure perspective. NA may create value when it replaces an MRI, operating-room diagnostic arthroscopy, general anesthesia, or additional visits. It may increase cost when it duplicates imaging, is followed by planned conventional arthroscopy, or requires low-utilization capital equipment. Most available models are US-based or single-centre analyses; differences in reimbursement, salaries, MRI access, sterilization standards, and care pathways can reverse the direction of an apparent saving. Future evaluations should state the displaced resource, time horizon, payer perspective, local prices, sensitivity analyses, and downstream consequences rather than report procedure cost in isolation [114,115,116].
4.4. Research Priorities and Minimum Reporting Dataset
The next phase should be question-matched rather than technology-led. Diagnostic studies require blinded, lesion-specific reference standards and reporting of non-visualized compartments. Therapeutic studies require procedure-specific non-inferiority margins, conversion criteria, and follow-up long enough to detect reoperation. Where ethically and practically feasible, randomized trials should compare needle arthroscopy with MRI, conventional arthroscopy, or the current standard pathway using patient-relevant outcomes. Multicentre comparative cohorts remain important where randomization is impractical. Implementation studies require multicentre replication, learning-curve analysis, and full pathway costing. A shared minimum dataset would make small studies cumulatively informative and reduce outcome fragmentation (Table 4).
Table 4.
Proposed minimum reporting dataset for clinical and implementation studies of needle arthroscopy.
4.5. Limitations
This review has several limitations. First, the quantitative PRISMA flow was reconstructed from the auditable PubMed/MEDLINE source files and should be understood as an incomplete evidence map rather than a complete census of the technology. Earlier Scopus, Web of Science, and Cochrane histories were used only as sensitivity sources; their updated record-level exports, term-level hit counts, and original execution date were not preserved, so studies indexed exclusively outside PubMed may have been missed, and the multi-database search cannot be reproduced in full. Second, the protocol was not prospectively registered. Third, deliberate inclusion of technical and cadaveric evidence was necessary to map translation but makes the corpus unsuitable for a single effectiveness estimate. Fourth, the Oxford-derived design hierarchy and translation stages are structured author assessments, not validated certainty or maturity classifications; no cross-design risk-of-bias score was calculated because item-level archival judgments were incomplete. Fifth, the corpus is publication-based: repeated specialist groups and possible cohort overlap could not be resolved for every report, and four reports were co-authored by review authors. Those records were classified by a non-coauthor reviewer and independently checked, but the archived log did not retain report-level reviewer initials. Sixth, adverse-event definitions, denominators, and follow-up were inconsistent, and a report of no observed complications cannot exclude rare harms. Finally, 2026 indexing was incomplete at the 31 July cutoff, and the full-text exclusion archive did not retain mutually exclusive reason counts. A future update should prospectively register its protocol and run synchronized, fully exportable searches across all planned databases from the outset, using a common date window and a reproducible de-duplication log so that cross-database reconciliation is possible. These constraints define the product as an incomplete evidence map and hypothesis-generating readiness assessment, not proof of clinical effectiveness or a basis for replacing MRI or conventional arthroscopy.
5. Conclusions
Needle arthroscopy is a promising but unproven technology within a rapidly expanding and uneven evidence base. Among 105 PMID-verified primary publications, 64 were technical or cadaveric, only 41 were clinical, 93 (88.6%) were in the IV–V design categories used for descriptive mapping, and only nine (8.6%) were in Levels I–II. The literature therefore supports technical feasibility and selected, indication-specific diagnostic concordance more strongly than it supports comparative clinical effectiveness, long-term safety, therapeutic equivalence, or economic value. Because the quantitative flow was based on auditable PubMed/MEDLINE files and studies indexed exclusively elsewhere may have been missed, the map is not a complete census of the field. Current evidence does not justify replacing MRI or conventional arthroscopy. NanoScope represented 81.3% of platform mentions, so findings should not be generalized automatically to needle arthroscopy as a class. Progress now depends on independent multicentre comparisons, randomized trials where feasible, lesion-specific reference standards, conversion and learning-curve reporting, standardized harms surveillance, and health-system-specific pathway economics.
Supplementary Materials
The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/jcm15197608/s1. Supplementary Table S1—Evidence base audit. Supplementary Table S2—PRISMA-ScR Compliance Checklist.
Author Contributions
Conceptualization, J.L. and M.D.; methodology, J.L. and B.W.; software and data-audit workflow, J.L.; validation, H.M., P.Ł., K.M., M.C., J.M., M.M., B.W., M.D., R.C., P.S.R., and J.L.; formal analysis, J.L., R.C., and P.S.R.; investigation and duplicate screening, H.M., P.Ł., K.M., M.C., J.M., M.M., and B.W.; data curation, H.M., P.Ł., K.M., and M.C.; writing—original draft preparation, J.L. and B.W.; writing—review and editing, all authors; visualization, J.L.; supervision, M.D. and P.S.R.; project administration, J.L. All authors have read and agreed to the published version of the manuscript.
Funding
This research received no external funding.
Data Availability Statement
The extracted study-level data are provided in Supplementary Table S1.
Acknowledgments
During the preparation of this manuscript, the authors used ChatPGT 5,6 for the purposes of figures andtables creation, english edition. The authors have reviewed and edited the output and take full responsibility for the content of this publication.
Conflicts of Interest
The authors declare no conflicts of interest in relation to this review. Four included primary reports are co-authored by Błażej Wójtowicz, Marcin Domżalski, or Jędrzej Lesman; those records were classified by a non-coauthor reviewer, independently checked by the other member of the screening pair, and were not used as the sole basis for conclusions. Manufacturer or industry relationships reported by included primary-study authors were charted descriptively in Supplementary Table S1 using the disclosure statements printed in the reports; non-disclosure was coded as unclear rather than as absence of a relationship, and the classifications were not treated as evidence of misconduct.
Abbreviations
The following abbreviations have been used in this manuscript:
| CA | conventional arthroscopy |
| IONA | in-office needle arthroscopy |
| MINORS | Methodological Index for Non-Randomized Studies |
| MRI | magnetic resonance imaging |
| NA | needle arthroscopy |
| PRISMA-ScR | Preferred Reporting Items for Systematic Reviews and Meta-Analyses extension for Scoping Reviews |
| QUADAS-2 | Quality Assessment of Diagnostic Accuracy Studies 2 |
| RCT | randomized controlled trial |
| RoB 2 | Risk of Bias 2 |
References
- Tricco, A.C.; Lillie, E.; Zarin, W.; O’Brien, K.K.; Colquhoun, H.; Levac, D.; Moher, D.; Peters, M.D.J.; Horsley, T.; Weeks, L.; et al. PRISMA Extension for Scoping Reviews (PRISMA-ScR): Checklist and Explanation. Ann. Intern. Med. 2018, 169, 467–473. [Google Scholar] [CrossRef] [Scilit]
- Zhang, K.; Crum, R.J.; Samuelsson, K.; Cadet, E.; Ayeni, O.R.; de Sa, D. In-Office Needle Arthroscopy: A Systematic Review of Indications and Clinical Utility. Arthroscopy 2019, 35, 2709–2721. [Google Scholar] [CrossRef] [Scilit]
- Burt, J.; Smith, V.; Gee, C.W.; Clarke, J.V.; Hall, A.J. The role of outpatient needle arthroscopy in the diagnosis and management of musculoskeletal complaints: A systematic review of the Arthrex NanoScope. Knee 2023, 42, 246–257. [Google Scholar] [CrossRef] [Scilit]
- Banke, I.J.; Becker, R.; Voss, M.; Ramadanov, N. Needle arthroscopy as an emerging office-based tool: Promising diagnostic value but limited evidence for therapeutic equivalence—A systematic review and frequentist meta-analysis. J. Exp. Orthop. 2026, 13, e70857. [Google Scholar] [CrossRef] [Scilit]
- Walinga, A.B.; van der Stappen, T.; Kerkhoffs, G.M.M.J.; Emanuel, K.S. Only Minor Complications Are Reported After Needle Arthroscopy: A Systematic Review. Arthrosc. Sports Med. Rehabil. 2025, 7, 101158. [Google Scholar] [CrossRef] [Scilit]
- Gianakos, A.L.; Kennedy, J.G. In-office needle arthroscopy: Indications, surgical techniques, tips, and tricks. JAAOS Glob. Res. Rev. 2024, 8, e23.00124. [Google Scholar] [CrossRef] [Scilit]
- Compagnoni, R.; Pieroni, A.; Fedele, G.; Maione, A.; Calanna, F.; Ferrua, P.; Randelli, P.S. Advancing knee diagnostics and treatment: The role of needle arthroscopy. Musculoskelet. Surg. 2026, 110, 21–28. [Google Scholar] [CrossRef] [Scilit]
- McMillan, S.; Saini, S.; Alyea, E.; Ford, E. Office-Based Needle Arthroscopy: A Standardized Diagnostic Approach to the Knee. Arthrosc. Tech. 2017, 6, e1119–e1124. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- West, J.A.; Amin, N.H. In-office arthroscopy for the evaluation of chronic knee pain: A case report. SAGE Open Med. Case Rep. 2017, 5, 2050313X17740992. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Chapman, G.L.; Amin, N.H. The Benefits of an In-Office Arthroscopy in the Diagnosis of Unresolved Knee Pain. Case Rep. Orthop. 2018, 2018, 6125676. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Gill, T.J.; Safran, M.; Mandelbaum, B.; Huber, B.; Gambardella, R.; Xerogeanes, J. A Prospective, Blinded, Multicenter Clinical Trial to Compare the Efficacy, Accuracy, and Safety of In-Office Diagnostic Arthroscopy With Magnetic Resonance Imaging and Surgical Diagnostic Arthroscopy. Arthroscopy 2018, 34, 2429–2435. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lavender, C.; Lycans, D.; Sina Adil, S.A.; Kopiec, A.; Schmicker, T. Incisionless Partial Medial Meniscectomy. Arthrosc. Tech. 2020, 9, e375–e378. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Quinn, R.; Lang, S.D.; Gilmer, B.B. Diagnostic Needle Arthroscopy and Partial Medial Meniscectomy Using Small Bore Needle Arthroscopy. Arthrosc. Tech. 2020, 9, e645–e650. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Daggett, M.C.; Busch, K.; Ferretti, A.; Monaco, E.; Bruni, G.; Saithna, A. Percutaneous Anterior Cruciate Ligament Repair With Needle Arthroscopy and Biological Augmentation. Arthrosc. Tech. 2021, 10, e289–e295. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- DiBartola, A.C.; Rogers, A.; Kurzweil, P.; Knopp, M.V.; Flanigan, D.C. In-Office Needle Arthroscopy Can Evaluate Meniscus Tear Repair Healing as an Alternative to Magnetic Resonance Imaging. Arthrosc. Sports Med. Rehabil. 2021, 3, e1755–e1760. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Kamiya, T.; Teramoto, A.; Mori, Y.; Kitamura, C.; Watanabe, K.; Yamashita, T. Nano-Arthroscopic Ultrasound-Guided Excision of Unresolved Osgood-Schlatter Disease. Arthrosc. Tech. 2021, 10, e1581–e1587. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lavender, C.; Adil, S.; Patel, T.; Bullock, M.; Oliashirazi, A. Incisionless Synovium and Bone Biopsy of a Painful Total Knee Arthroplasty. Arthrosc. Tech. 2021, 10, e475–e479. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lavender, C.; Fravel, W.; Patel, T.; Adil, S.; Blickenstaff, B. Nanoscopic Single-Incision Autograft Cartilage Transfer (ACT). Arthrosc. Tech. 2021, 10, e545–e549. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lavender, C.; Malik, S.; Lycans, D.; Hooper, M.; Flores, K. Nanoscopic-Assisted Anterior Cruciate Ligament-Posterior Cruciate Ligament Reconstruction. Arthrosc. Tech. 2021, 10, e1839–e1844. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lavender, C.; Flores, K.; Patel, T.; Berdis, G.; Blickenstaff, B. Nanoscopic Medial Meniscus Repair. Arthrosc. Tech. 2021, 10, e1943–e1947. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Stornebrink, T.; van Dijck, R.A.H.E.; Douven, D.; Kerkhoffs, G.M.M.J. Needle Arthroscopic All-Inside Repair of Meniscal Tears Under Local Anesthesia. Arthrosc. Tech. 2021, 10, e2173–e2180. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Annibaldi, A.; Monaco, E.; Daggett, M.; Carrozzo, A.; Mazza, D.; Previ, L.; Rossi, G.; Orlandi, P.; Ferretti, A. In-office needle arthroscopic assessment after primary ACL repair: Short-term results in 15 patients. J. Exp. Orthop. 2022, 9, 89. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Bradsell, H.; Lencioni, A.; Shinsako, K.; Frank, R.M. In-Office Diagnostic Needle Arthroscopy Using the NanoScope™ Arthroscopy System. Arthrosc. Tech. 2022, 11, e1923–e1927. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Konrads, C.; Notheisen, T.; Döbele, S. Minimally invasive arthroscopy of the knee using a new 2 mm device. Clin. Case Rep. 2022, 10, e05590. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lavender, C.; Taylor, S.; Blickenstaff, B.; Macaskill, M.; Baria, M. Excision of a Knee Cyclops Lesion Using a Needle Arthroscope. Arthrosc. Tech. 2022, 11, e563–e568. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Shubert, D.; DeFroda, S.; Nuelle, C.W. Concurrent Needle and Standard Arthroscopy for Posterior Cruciate Ligament Reconstruction. Arthrosc. Tech. 2022, 11, e1335–e1340. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Trang, G.; Del Sol, S.R.; Jenkins, S.; Bryant, S.; Gardner, B.; Chakrabarti, M.O.; McGahan, P.J.; Chen, J.L. Evaluation of Osteochondral Allograft Transplant Using In-Office Needle Arthroscopy. Arthrosc. Tech. 2022, 11, e2243–e2248. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Bast, N.; Zimmerman, R.; Schlechter, J. An Overview of Pediatric Knee Arthroscopy with Use of the NeedleScope. Arthrosc. Tech. 2023, 13, 102813. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Behrendt, P.; Berninger, M.T.; Thürig, G.; Dehoust, J.; Christensen, J.; Frosch, K.H.; Krause, M.; Hartel, M.J. Nanoscopy and an extended lateral approach can improve the management of latero-central segments in tibial plateau fractures: A cadaveric study. Eur. J. Trauma. Emerg. Surg. 2022, 49, 1433–1439. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Bi, A.S.; Colasanti, C.A.; Kirschner, N.; Neal, W.H.E.; Owusu-Sarpong, S.; Fariyike, B.; Azam, M.T.; Stone, J.W.; Kennedy, J.G. In-Office Needle Arthroscopy With Cartilage Allograft Extracellular Matrix Application for Cartilage Lesions of the Knee. Arthrosc. Tech. 2023, 12, e1507–e1513. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Kirschner, N.; Owusu-Sarpong, S.; Neal, W.H.E.; Fariyike, B.; Bi, A.S.; Colasanti, C.A.; Azam, M.T.; Gianakos, A.L.; Stone, J.W.; Kennedy, J.G. In-Office Needle Arthroscopy with Meniscal Repair for Meniscal Lesions of the Knee. Arthrosc. Tech. 2023, 12, e1821–e1826. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Nakasa, T.; Ishikawa, M.; Nekomoto, A.; Nakata, K.; Omoto, T.; Kamei, G.; Nakamae, A.; Adachi, N. Needle Arthroscopy as a Promising Alternative to MRI for the Diagnosis of Meniscus Injury. Cureus 2023, 15, e48671. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Neal, W.H.E.; Kirschner, N.; Owusu-Sarpong, S.; Colasanti, C.A.; Fariyike, B.; Bi, A.S.; Azam, M.T.; Stone, J.W.; Kennedy, J.G. In-Office Needle Arthroscopy of the Knee With Lateral Parapatellar Retinacular Release. Arthrosc. Tech. 2023, 12, e1809–e1813. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Savage-Elliott, I.; Kingery, M.T.; Azam, M.T.; Lowe, D.T.; Strauss, E.J. Cartilage Biopsy for Autologous Cell-Based Repair of the Knee in the Wide-Awake Setting Using Needle Arthroscopy. Arthrosc. Tech. 2023, 12, e2029–e2033. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Schaver, A.L.; Lash, J.G.; MacAskill, M.L.; Taylor, S.; Hewett, T.E.; Jasko, J.J.; Argintar, E.H.; Lavender, C.D. Partial meniscectomy using needle arthroscopy associated with significantly less pain and improved patient reported outcomes at two weeks after surgery: A comparison to standard knee arthroscopy. J. Orthop. 2023, 41, 63–66. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Blankenburg, N.; Henkelmann, R.; Theopold, J.; Löffler, S.; Hepp, P. Comparison of needle and conventional arthroscopy for visualisation of predefined anatomical structures of the knee joint: A feasibility study in human cadavers and patients. BMC Musculoskelet. Disord. 2024, 25, 212. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Khan, A.M.; Sarraf, K.M. Targeted epiphysiodesis using needle arthroscopy. J. Clin. Orthop. Trauma 2024, 53, 102437, Erratum in J. Clin. Orthop. Trauma2027, 73, 103341. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lavender, C.D.; Hahn, E.; Hwang, T.H.; Schaver, A.L.; LeMaster, E.S.; Nottingham, J.; Hewett, T.E.; Argintar, E.H. Opioid use following partial meniscectomy with needle arthroscopy compared to standard knee arthroscopy: Results from a randomized multicenter trial. J. Orthop. 2025, 72, 227–230. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Liu, M.; Yan, M.; Xu, J.; Zeng, Q.; Zhu, K.; Zhao, S.; Diao, W.; Wang, Y.; Leng, X. A newly developed 2 mm needle arthroscope with high-definition for orthopedic outpatient knee joint examination. Comput. Biol. Med. 2025, 190, 110112. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Sahi, G.; Shah, A.; Abbas, A.; Lex, J.R.; Abouali, J.; Toor, J. Patients prefer In-Office Needle Arthroscopy (IONA) over traditional surgical arthroscopy. J. Orthop. 2025, 70, 107–112. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Mzeihem, M.; Zabawa, L.; Park, Y.; Crespo, A.; Raji, Y.; Amirouche, F. Fluoroscopy-Guided Lateral Tibial Plateau Fracture Fixation with and without Needle Arthroscopy: A Biomechanical and Reduction-Quality Comparison. J. Bone Jt. Surg. Am. 2025, 108, 377–383. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Stornebrink, T.; Stufkens, S.A.S.; Appelt, D.; Wijdicks, C.A.; Kerkhoffs, G.M.M. J. 2-Mm Diameter Operative Tendoscopy of the Tibialis Posterior, Peroneal, and Achilles Tendons: A Cadaveric Study. Foot Ankle Int. 2019, 41, 473–478. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Stornebrink, T.; Altink, J.N.; Appelt, D.; Wijdicks, C.A.; Stufkens, S.A.S.; Kerkhoffs, G.M.M.J. Two-millimetre diameter operative arthroscopy of the ankle is safe and effective. Knee Surg. Sports Traumatol. Arthrosc. 2020, 28, 3080–3086. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Colasanti, C.A.; Mercer, N.P.; Garcia, J.V.; Kerkhoffs, G.M.M.J.; Kennedy, J.G. In-Office Needle Arthroscopy for the Treatment of Anterior Ankle Impingement Yields High Patient Satisfaction With High Rates of Return to Work and Sport. Arthroscopy 2021, 38, 1302–1311. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Colasanti, C.A.; Kaplan, D.J.; Chen, J.S.; Kanakamedala, A.; Dankert, J.F.; Hurley, E.T.; Mercer, N.P.; Stone, J.W.; Kennedy, J.G. In-Office Needle Arthroscopy for Anterior Ankle Impingement. Arthrosc. Tech. 2022, 11, e327–e331. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Kaplan, D.J.; Chen, J.S.; Colasanti, C.A.; Dankert, J.F.; Kanakamedala, A.; Hurley, E.T.; Mercer, N.P.; Stone, J.W.; Kennedy, J.G. Needle Arthroscopy Cheilectomy for Hallux Rigidus in the Office Setting. Arthrosc. Tech. 2022, 11, e385–e390. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lopes, R.; Noailles, T.; Padiolleau, G.; Bouguennec, N.; Vieira, T.D. Needle arthroscopy in anatomical reconstruction of the lateral ankle: A report of three cases with a parallel comparison to the standard arthroscopy procedure. J. Exp. Orthop. 2022, 9, 75. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Mercer, N.P.; Azam, M.T.; Davalos, N.; Kaplan, D.J.; Colasanti, C.A.; Chen, J.S.; Kanakamedala, A.C.; Dankert, J.F.; Stone, J.W.; Kennedy, J.G. Anterior Talofibular Ligament Augmentation With Internal Brace in the Office Setting. Arthrosc. Tech. 2022, 11, e545–e550. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Mercer, N.P.; Samsonov, A.P.; Dankert, J.F.; Gianakos, A.L.; Stornebrink, T.; Delmonte, R.J.; Kerkhoffs, G.M.J.; Kennedy, J.G. Improved Clinical Outcomes and Patient Satisfaction of In-Office Needle Arthroscopy for the Treatment of Posterior Ankle Impingement. Arthrosc. Sports Med. Rehabil. 2022, 4, e629–e638. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Stornebrink, T.; Stufkens, S.A.S.; Mercer, N.P.; Kennedy, J.G.; Kerkhoffs, G.M.M.J. Can bedside needle arthroscopy of the ankle be an accurate option for intra-articular delivery of injectable agents? World J. Orthop. 2022, 13, 78–86. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Dahmen, J.; Indino, C.; D’Ambrosi, R.; Usuelli, F.G. Needle Arthroscopic Subchondroplasty With Adipose-Derived Stem Cell Augmentation for the Treatment of Osteochondral Lesions of the Talus. Arthrosc. Tech. 2023, 12, e1649–e1656. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Dehoust, J.; Berninger, M.T.; Behrendt, P.; Thürig, G.; Christensen, J.H.; Hinz, N.; von Rehlingen-Prinz, F.; Frosch, K.-H.; Hartel, M. Comparison of different intraoperative reduction monitoring methods in a cadaveric intraarticular calcaneal fracture model: 3D scan vs arthroscopy vs nanoscopy. Eur. J. Trauma. Emerg. Surg. 2023, 49, 2561–2567. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Inoue, J.; Yasui, Y.; Sasahara, J.; Takenaga, T.; Wakabayashi, K.; Nozaki, M.; Kobayashi, M.; Ha, M.; Fukushima, H.; Kato, J.; et al. Comparison of Visibility and Risk of Neurovascular Tissue Injury Between Portals in Needle Arthroscopy of the Anterior Ankle Joint: A Cadaveric Study. Orthop. J. Sports Med. 2023, 11, 23259671231174477. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Martin, K.D.; Curatolo, C.; Gallagher, J.; Alvarez, P. Talar Arthroscopic Reduction and Internal Fixation (TARIF): A Novel All-Inside Soft-Tissue-Preserving Technique. JBJS Essent. Surg. Tech. 2023, 13, e22.00007. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Pujol, O.; Sallent, A.; Salom, J.; Duarri, G.; Maled, I.; Busquets, R. ORIF Assisted by Open Subtalar Small Joint Arthroscopy for Intra-articular Displaced Calcaneus Fractures. Foot Ankle Orthop. 2023, 8, 24730114231163822. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Walinga, A.B.; Dahmen, J.; Stornebrink, T.; Kerkhoffs, G.M.M.J. Needle Arthroscopy for Osteochondral Lesions of the First Metatarsophalangeal Joint: A Standardized Approach. Arthrosc. Tech. 2023, 12, e1015–e1019. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Cho, E.; Butler, J.J.; Kennedy, J.G.; Gianakos, A.L. Nano-Arthroscopic Plantar Fascia Release Technique. Arthrosc. Tech. 2024, 13, 103118. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Inoue, J.; Yasui, Y.; Sasahara, J.; Takenaga, T.; Ha, M.; Miyamoto, W.; Kawano, H.; Murakami, H.; Yoshida, M. Comparison of visibility in needle arthroscopy of the ankle according to surgical experience: A cadaveric study. Foot Ankle Surg. 2024, 30, 603–607. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Stornebrink, T.; Walinga, A.B.; Stufkens, S.A.S.; Kerkhoffs, G.M.M.J. Wide-Awake Needle Arthroscopy of the Anterior Ankle: A Standardized Approach. Arthrosc. Tech. 2024, 13, 102901. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Walinga, A.B.; Dahmen, J.; Stornebrink, T.; Emanuel, K.S.; Kerkhoffs, G.M.M.J. Fifteen out of 16 elite athletes showed concomitant low-grade cartilage lesions of the ankle with unstable syndesmotic injuries: Concerns from a prospective case series. BMJ Open Sport. Exerc. Med. 2024, 10, e001879. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Walinga, A.B.; Butler, J.; Dahmen, J.; Stufkens, S.A.S.; Robert, G.; Kennedy, J.G.; Kerkhoffs, G.M. Second-Look Needle Arthroscopy After Prior Surgical Treatment for Cartilage Lesions of the Ankle: The Amsterdam and New York City Perspectives. Cartilage 2024, 17, 52–60. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Wojtowicz, B.G.; Domzalski, M.; Lesman, J. Needle arthroscopic-assisted repair of tibio-fibular syndesmosis acute injury: A case report. World J. Orthop. 2024, 15, 820–827. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Georgiakakis, E.C.T.; Khan, A.M.; Singh, P.; Sarraf, K.M. Needle Arthroscopy for the Surgical Management of Brodie’s Abscess in the Distal Tibia of an Infant. Indian. J. Orthop. 2025, 60, 495–500. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Shimozono, Y.; Arai, R.; Kuroda, Y.; Ito, H.; Matsuda, S. Cadaveric Study Evaluating the Potential for Hindfoot Endoscopy and Flexor Hallucis Longus Tendoscopy Using a 1.9-mm Diameter Needle Arthroscope. Foot Ankle Orthop. 2025, 10, 24730114241310237. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Shimozono, Y.; Mori, D.; Mizuno, Y.; Funakoshi, N.; Kobayashi, M.; Matsuda, S.; Yamashita, F. Clinical outcomes of posterior hindfoot endoscopic surgery using 1.9-mm-diameter needle arthroscopy for posterior ankle impingement in athletes. Eur. J. Orthop. Surg. Traumatol. 2025, 35, 338. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Wójtowicz, B.G.; Gołek, Ł.; Długozima, R.; Domżalski, M.; Lesman, J. The utility of needle arthroscopy in the ankle joint: A cadaveric study evaluating visualization, surgical feasibility, and learning curve considerations. Foot Ankle Surg. 2025, 32, 343–346. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Daggett, M.C.; Stepanovich, B.; Geraghty, B.; Meyers, A.; Whetstone, J.; Saithna, A. Office-Based Needle Arthroscopy: A Standardized Diagnostic Approach to the Shoulder. Arthrosc. Tech. 2020, 9, e521–e525. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lavender, C.; Lycans, D.; Kopiec, A.; Sayan, A. Nanoscopic Single-Incision Anterior Labrum Repair. Arthrosc. Tech. 2020, 9, e297–e301. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lavender, C.; Lycans, D.; Sina Adil, S.A.; Berdis, G. Single-Incision Rotator Cuff Repair With a Needle Arthroscope. Arthrosc. Tech. 2020, 9, e419–e423. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Shafi, N.; Lang, S.D.; Kassam, H.; Gilmer, B.B. Diagnostic and Therapeutic Shoulder Arthroscopy Using a Small-Bore Needle Arthroscope. Arthrosc. Tech. 2020, 9, e1087–e1093. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Gauci, M.O.; Monin, B.; Rudel, A.; Blasco, L.; Bige, B.; Boileau, P. In-Office Biceps Tenotomy with Needle Arthroscopy: A Feasibility Study. Arthrosc. Tech. 2021, 10, e1263–e1268. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Peach, C.; Davies, R.; Phillips, N. Wide-Awake Shoulder Nanoscopic Long Head of Biceps Tenotomy. Arthrosc. Tech. 2021, 10, e909–e912. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Wagner, E.R.; Woodmass, J.M.; Zimmer, Z.R.; Welp, K.M.; Chang, M.J.; Prete, A.M.; Farley, K.X.; Warner, J.J. Needle Diagnostic Arthroscopy and Magnetic Resonance Imaging of the Shoulder Have Comparable Accuracy With Surgical Arthroscopy: A Prospective Clinical Trial. Arthroscopy 2021, 37, 2090–2098. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lavender, C.; Taylor, S.; Macaskill, M.; Peluso, R.; Hewett, T.; Jasko, J. Rotator Cuff Repair Using a Needle Arthroscope Through a Dual-Lumen Flexible Cannula. Arthrosc. Tech. 2022, 11, e2119–e2123. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Rosso, C.; Grezda, K.; Heuberer, P.R. Long head of the biceps intra-articular tenotomy using needle arthroscopy under local anesthesia: Preliminary results and technical note. J. Exp. Orthop. 2022, 9, 70. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Senna, L.F.; Lavender, C. Nanoscopic Distal Clavicle Resection. Arthrosc. Tech. 2022, 11, e551–e554. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Chowdhury, A.; Gibson, C.; Nicholls, A.; MacLeod, I.; Colaco, H. Diagnostic Needle Arthroscopy of the Shoulder: A Validation Study. Orthop. J. Sports Med. 2023, 11, 23259671231155885. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Colasanti, C.A.; Azam, M.T.; Bi, A.S.; Fariyike, B.; Kirschner, N.; Neal, W.H.E.; Owusu-Sarpong, S.; Stone, J.W.; Kennedy, J.G. Reproducible and Effective Biceps Tenodesis Method Utilizing In-Office Nano-Arthroscopy. Arthrosc. Tech. 2023, 12, e1797–e1802. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Owusu-Sarpong, S.; Fariyike, B.; Colasanti, C.A.; Bi, A.S.; Kirschner, N.; Neal, W.H.E.; Azam, M.T.; Stone, J.W.; Kennedy, J.G. In-Office Nano-Arthroscopy of the Shoulder with Acromioplasty. Arthrosc. Tech. 2023, 12, e1423–e1428. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Reinares, F.; Maza, F.; Paz, J.; Angulo, M.; Lecaros, J.J.; Ruiz Jimenez, F. Nanoscopic Distal Biceps Repair With Cortical Button and Interference Screw. Arthrosc. Tech. 2023, 12, e153–e159. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Yadoji, H.K.; Bodanki, C.; Reddy, S.A.; Reddy, A.V.G. Percutaneous Capsule Release in Adhesive Capsulitis with Nanoscope under Local Anesthesia—Case Report. J. Orthop. Case Rep. 2023, 13, 126–129. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ameziane, Y.; Holzer, E.; Scheibel, M. Confirming polyethylene wear via outpatient nanoscopy following anatomical total shoulder arthroplasty. JSES Rev. Rep. Tech. 2024, 4, 812–816. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Fariyike, B.; Neal, W.H.E.; Bi, A.S.; Owusu-Sarpong, S.; Colasanti, C.A.; Kirschner, N.; Azam, M.T.; Butler, J.J.; Stone, J.W.; Kennedy, J.G. In-Office Needle Arthroscopy for Superior Labral Tear Debridement. Arthrosc. Tech. 2024, 13, 102956. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Khan, A.M.; Sarraf, K.M. Needle arthroscopy in the surgical management of septic arthritis of the shoulder joint in neonates and infants. J. Clin. Orthop. Trauma 2024, 55, 102498, Erratum in J. Clin. Orthop. Trauma2026, 75, 103348. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Anderson, K. Labral Repair with Capsulorrhaphy Completed via Needle Arthroscopy. Arthrosc. Tech. 2025, 14, 103834. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Daggett, M.; Witte, K.; Dempewolf, M.; Beutel, B.G.; McCabe, M.; Gilbert, G.; Redler, A. Percutaneous Biceps Tenodesis Using Needle Arthroscopy and Regional Anesthesia: The Infinity Technique. Arthrosc. Tech. 2025, 14, 103414. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Cabatu, C.; Friedman, A.M.; Grotting, J.A.; Williams, Z.; David, T.S. 2.0-mm Diameter Arthroscopy versus MRI in Predicting Appropriate Surgical Management of Intra-articular Shoulder Pathology: A Blinded, Prospective Study. Orthop. J. Sports Med. 2026, 14, 23259671261434938. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lesman, J.; Wójtowicz, B.; Szufnara, M.; Błoch, C.; Domżalski, M. Nanoscopic long head of the biceps tendon tenotomy under suprascapular nerve block: A feasibility study. Eur. J. Med. Res. 2026, 31, 371. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lesman, J.; Wójtowicz, B.; Gołek, Ł.; Szufnara, M.; Domżalski, M. Feasibility of shoulder joint visualization using the 0° Arthrex Nanoscope: A cadaveric study on twenty specimens. J. Orthop. Surg. Res. 2026. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Reiser, D.; Hedspång, M.; Sagerfors, M. Dry Arthroscopy of the Wrist With a Single-use, 1.9 mm Chip-on-tip System in Wide-awake Local Anesthesia No Tourniquet. Tech. Hand Up. Extrem. Surg. 2022, 26, 246–249. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Oh, C.; Kakar, S. NanoScope Arthroscopy: Lessons Learned in the First 75 Cases. J. Wrist Surg. 2022, 11, 280–287. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Choke, A.; Chung, S.F.; Chung, S.R. Arthroscopic-Assisted Distal Radius Fracture Fixation Using the NanoScope System. Arthrosc. Tech. 2024, 13, 103033. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Dittman, L.E.; Munaretto, N.; Hinchcliff, K.; Dutton, L.; Kakar, S. Volar Wrist Arthroscopy Portals Using the NanoScope Are Safer Than Traditional Arthroscopy. Hand 2024, 20, 602–606. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Moses, M.J.; Lorentz, N.A.; Azad, A.; Paksima, N. Needle Arthroscopy Versus Conventional Arthroscopy in the Evaluation of Carpal Pathology: A Comparative Study. Hand 2024, 20, 1278–1283. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Reiser, D.; Hedspång, M.; Sagerfors, M. NanoScope wrist arthroscopy under wide-awake local anesthesia with no tourniquet: A prospective series of 30 consecutive patients. J. Hand Microsurg. 2024, 16, 100067. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Hedspång, M.; Sagerfors, M.; Kakar, S.; Ryen, L.; Asklöf, P.; Reiser, D. Dry Needle Arthroscopy of the Wrist in an Office Setting: 15 Cases. J. Hand Surg. Am. 2025, 50, 1524.e1–1524.e6. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Peters, M.; Gilmer, B.; Kassam, H.F. Diagnostic and Therapeutic Elbow Arthroscopy Using Small-Bore Needle Arthroscopy. Arthrosc. Tech. 2020, 9, e1703–e1708. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Fournier, M.; Corning, E.; Witt, A.; Lang, S.; Gilmer, B.B. Arthroscopically Assisted Fixation of Terrible Triad Variant Injuries of the Elbow With Small-Bore Needle Arthroscopy. Arthrosc. Tech. 2021, 10, e1469–e1474. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Vander Voort, W.D.; Saad, M.; Falgout, D.; Blaine, T.A.; Kassam, H.F. Small-Bore Needle Arthroscopic Extensor Carpi Radialis Brevis Release Results in Improved Outcomes at One Year Postoperatively. Arthrosc. Sports Med. Rehabil. 2022, 5, e159–e164. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Rapariz, J.M.; Far-Riera, A.M.; Perez-Uribarri, C.; Martin-Martin, S.; Rodriguez-Baeza, A. Needle arthroscopy of the elbow through an anterior transbrachial portal. JSES Int. 2023, 7, 673–677. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Yamauchi, S.; Takenaga, T.; Tsuchiya, A.; Takeuchi, S.; Takaba, K.; Inoue, J.; Ono, T.; Kuboya, K.; Nozaki, M.; Fukushima, H.; et al. Visibility During Elbow Arthroscopy Using Needle Arthroscope: A Cadaveric Study. Orthop. J. Sports Med. 2025, 13, 23259671251351326. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Reiser, D.; Adolfsson, L.; Thórdardóttir, Á.; Sagerfors, M. Dry Needle Arthroscopy of the Elbow with a 1.9 mm Chip-on-Tip System: A Cadaveric Study. Acta Chir. Orthop. Traumatol. Cech. 2026, 93, 9–14. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- McMillan, S.; Chhabra, A.; Hassebrock, J.D.; Ford, E.; Amin, N.H. Risks and Complications Associated With Intra-articular Arthroscopy of the Knee and Shoulder in an Office Setting. Orthop. J. Sports Med. 2019, 7, 2325967119869846. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Stornebrink, T.; Janssen, S.J.; Kievit, A.J.; Mercer, N.P.; Kennedy, J.G.; Stufkens, S.A.S.; Kerkhoffs, G.M.M.J. Bacterial arthritis of native joints can be successfully managed with needle arthroscopy. J. Exp. Orthop. 2021, 8, 67. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- DeClouette, B.; Birnbaum, A.; Campbell, H.; Bi, A.S.; Lin, C.C.; Struhl, S. Needle Arthroscopy Demonstrates High Sensitivity and Specificity for Diagnosing Intra-Articular Shoulder and Knee Pathology. Cureus 2022, 14, e33189. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Walinga, A.B.; Stornebrink, T.; Janssen, S.J.; Dalmau-Pastor, M.; Kievit, A.J.; Kerkhoffs, G.M.M. Needle Arthroscopy for Bacterial Arthritis of a Native Joint: Surgical Technique for the Shoulder, Elbow, Wrist, Knee, and Ankle Under Local Anesthesia. Arthrosc. Tech. 2022, 11, e1641–e1648. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lopas, L.A.; Mir, H.R. Needle Arthroscopy as a Reduction Aid for Lower Extremity Peri-Articular Fractures: Case Series and Technical Tricks. Iowa Orthop. J. 2023, 43, 161–167. [Google Scholar] [PubMed]
- McMillan, S.; Ford, E.; Lau, V. Arthroscopic Hip Labral Repair Using Needle Arthroscopic Visualization. Arthrosc. Tech. 2023, 13, 102819. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Stornebrink, T.; Walinga, A.; Dalmau-Pastor, M.; Bosman, A.W.; Smit, T.H.; Kerkhoffs, G.M.M.J. Implantation of a Cushioning Injectable Implant Using Needle Arthroscopy in the Foot and Ankle and First Carpometacarpal Joint. Arthrosc. Tech. 2023, 12, e2343–e2352. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ford, E.; Frank, R.; Pontes, M.; McMillan, S. Improved Network Integrity and Patient Follow-up After In-Office Needle Arthroscopy Compared to Outpatient Advanced Diagnostic Imaging for Intra-articular Pathology. Arthrosc. Sports Med. Rehabil. 2024, 6, 100895. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Walinga, A.B.; Janssen, S.J.; Stornebrink, T.; Krips, R.; Tas, S.W.; Kievit, A.J.; Kerkhoffs, G.M. Bedside needle arthroscopy for native joint bacterial arthritis in a real-world clinical practice setting: A prospective cohort study. Arthritis Res. Ther. 2025, 28, 1. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Deirmengian, C.A.; Dines, J.S.; Vernace, J.V.; Schwartz, M.S.; Creighton, R.A.; Gladstone, J.N. Use of a Small-Bore Needle Arthroscope to Diagnose Intra-Articular Knee Pathology: Comparison with Magnetic Resonance Imaging. Am. J. Orthop. 2018, 47. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Centre for Evidence-Based Medicine. Oxford Centre for Evidence-Based Medicine 2011 Levels of Evidence (Adapted from the March 2009 Framework); University of Oxford: Oxford, UK, 2009. Available online: https://www.cebm.ox.ac.uk/resources/levels-of-evidence/ocebm-levels-of-evidence (accessed on 27 September 2026).
- Voigt, J.D.; Mosier, M.; Huber, B. Diagnostic needle arthroscopy and the economics of improved diagnostic accuracy: A cost analysis. Appl. Health Econ. Health Policy 2014, 12, 523–535. [Google Scholar] [CrossRef] [Scilit][Green Version]
- McMillan, S.; Ford, E.; Azam, M.T.; Kennedy, J.G. In-office diagnostic needle arthroscopy versus MRI: Cost analysis. Arthrosc. Sports Med. Rehabil. 2021, 3, e363–e369. [Google Scholar]
- Munn, D.; Burt, J.; Gee, C.W.; Mclaren, C.; Clarke, J.; Hall, A. Outpatient needle arthroscopy reduces cost and waste compared with conventional knee arthroscopy. Knee 2023, 42, 143–152. [Google Scholar] [CrossRef] [Scilit]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.