Reliability of Tracheal Temperature as a Measurement of Core Body Temperature During Cardiac Surgery Using Cardiopulmonary Bypass

,
,
,
and
1
Department of Anesthesiology and Pain Medicine, Asan Medical Center, University of Ulsan College of Medicine, Seoul 05505, Republic of Korea
2
Department of Anesthesiology and Pain Medicine, Kyungpook National University Hospital, Daegu 41944, Republic of Korea
3
Department of Anesthesiology and Pain Medicine, Anam Hospital, Korea University College of Medicine, Seoul 02841, Republic of Korea
*
Author to whom correspondence should be addressed.
This article belongs to the Section Anesthesiology

Abstract

Background: To compare tracheal temperature (TET) with nasopharyngeal temperature (TNP) in patients undergoing cardiac surgery using cardiopulmonary bypass (CPB). Methods: TET was measured using a thermistor in the cuff of an endotracheal tube and TNP was monitored using an esophageal stethoscope. Depending on the management of the CPB strategy, the operation was divided into four periods (pre-CPB, cooling, rewarming, and post-CPB). A Bland–Altman analysis was carried out to compare TET with TNP in each period. The concordance correlation coefficient for repeated measures analysis and various time lags was used to validate the time lag maximizing the concordance correlation coefficient between the two methods. Results: A total of 7191 pairs of temperature recordings acquired from 24 patients were included in the analysis. During steady normothermia, 81.7% (95% confidence interval [CI], 79.8–83.5%) of the pairs had a difference within ±0.5 °C, with a mean bias of −0.28 °C and limits of agreement (LOA) of −0.74 °C to 0.18 °C. The LOA during the cooling and rewarming phase of CPB were −1.13 °C to 0.51 °C and −0.91 °C to 1.29 °C, respectively. The mean bias and LOA throughout the entire operation were −0.10 °C and −0.98 °C to 0.77 °C, respectively. Throughout the entire operation, −2 min of time lag on TET maximized the concordance correlation coefficient (0.94 [95% CI, 0.92–0.96] to 0.95 [95% CI, 0.93–0.96]), indicating an earlier response of TET than TNP. Conclusions: TET could be an alternative to TNP during cardiac surgery using CPB.

Article Metrics

Citations

Article Access Statistics

Multiple requests from the same IP address are counted as one view.