Abstract
Early detection of malignancy is imperative, yet existing diagnostic approaches struggle to identify small peripheral lesions. This study evaluated a novel imaging modality, heat diffusion analysis, to assess its ability to differentiate between malignant and normal lung tissue. Considering that lung cancer is the leading cause of cancer-related mortality worldwide, lung tumors were induced in mice in a preclinical ex vivo model to evaluate the proposed technology. The HTOScan System was used to analyze the thermal characteristics of 60 sites from excised lungs, including normal and abnormal regions. The algorithm classified pixels as high- or low-risk for malignancy. The HTOScan System demonstrated a high accuracy of 97%, with 94% sensitivity and 98% specificity compared to the gold standard of histopathology. The technology successfully differentiated abnormal from normal tissue ex vivo based on differences in thermal diffusivity. This proof-of-concept study suggests that combining heat diffusion imaging techniques with machine learning algorithms could enable the HTOScan System to identify malignant lesions accurately with high confidence. The technique shows promise as a real-time decision support tool for cancer detection, pending further in vivo validation. This novel functional-imaging approach could improve the identification of peripheral lesions and the guidance of biopsies during bronchoscopy.
1. Introduction
Cancer remains a leading cause of death globally, with solid tumors being the most prevalent form [1,2]. While survival rates are improving in countries with robust healthcare systems, thanks to advances in early detection technologies, many solid cancers still go undetected in their early stages [2,3]. Given the high prevalence of lung nodules and the severe repercussions of incorrect or delayed diagnosis, research in this area is significant, emphasizing the importance of early diagnosis (e.g., [4]).
Current lung tissue sampling techniques include surgical resection, CT-guided transthoracic needle biopsy (CT-TTNB), and bronchoscopy biopsy [4]. Although bronchoscopy biopsy offers a better safety profile than CT-TTNB [5], its diagnostic yield (DY) has been limited, particularly for smaller tumors [6]. Various guided bronchoscopy technologies have emerged, allowing real-time localization, including radial probe endobronchial ultrasound (REBUS) [7], virtual bronchoscopy navigation (VBN) [8], electromagnetic navigational bronchoscopy (ENB) [9], digital tomosynthesis [10], and cone-beam computed tomography (CBCT) [11] to address this issue. These advanced technologies have shown improvements in DY, with some achieving rates comparable to those reported for the CT-TTNB series, where the pooled diagnostic yield is about 88.9%, and recent reviews estimate that the range from 85% to 95% [4,11,12,13]. However, they still face limitations, including radiation exposure, CT-to-body divergence, catheter deflection, and breathing variability [13,14,15]. The variability in reported DY across studies and the lack of standardized guidelines for reporting and estimating DY further complicate comparisons and analyses of these techniques [16]. Overcoming these challenges while maintaining minimally invasive procedures requires collaborative decision-making and local expertise, as the existing literature lacks a firm prescription for patients [4,5,8].
This study introduces the HTOScan System (a trade name for a prototype developed by HTO Bioimaging Ltd.). This novel functional thermal imaging modality provides a real-time decision-support tool for the surface diagnosis of superficial mucosal tissue characteristics in hollow organs during microscopic exploration procedures, such as bronchoscopy, where obtaining conclusive biopsy samples remains an important challenge [16]. Accordingly, this system can also serve as a guidance tool to improve diagnostic abilities by localizing the excision with the highest detection probability. The HTOScan System aims to analyze the heat-diffusion properties of biological tissues to differentiate cancerous from normal tissues by identifying differences in their heat-transfer rates [17,18].
Thermal diffusivity, a crucial thermophysical property, indicates how quickly heat transfers through various substances. The thermal diffusivity of viable biological tissues is affected by their composition, morphology, and vascular networks [19,20,21,22]. The calculated rate at which heat transfers throughout a material is termed thermal diffusivity [23]. Well-established tumor tissue exhibits an increase of approximately one degree Celsius compared to healthy neighboring tissue [22] due to increased metabolic activity, faster growth processes, and increased blood vessel usage [21,24].
This temperature difference forms the basis of the HTOScan System technology, which leverages active thermal imaging to detect temperature fluctuations of designated lesions, potentially serving as a practical decision-support tool for early cancer diagnosis. The widely used Pennes bioheat model [25], which accounts for factors such as metabolism and perfusion, describes the effect of blood flow on tissue temperature [23,24].
When comparing malignant and healthy tissues, differences in composition, tissue morphology, and vascular networks result in variations in heat transfer rates, thereby affecting their thermophysical properties [19,20,22,26]. These differences in heat transfer rates are the key principle that the HTOScan System utilizes for cancer detection [17,21].
A recent pilot study using the HT Vista system (a trade name for a prototype developed by HTO Bioimaging LTD.), a pet-compatible version of the HTOScan System developed for veterinary use, demonstrated highly efficient real-time diagnostics for detecting malignancy in subcutaneous and cutaneous masses in canines [27]. The system achieved 90% accuracy compared with cytology or biopsy results, correctly classifying 62 out of 69 masses, with only 2 false negatives (lymph nodes diagnosed with lymphoma). The overall performance parameters, including accuracy, sensitivity, specificity, positive predictive value, and negative predictive value, were promising, with values of 90%, 93%, 88%, 83%, and 95%, respectively, for all mass types.
These promising results from viable animal tissues support the potential for implementing the method in viable human tissues. Consequently, this study presents a preliminary proof-of-concept (POC), an ex vivo trial to assess the accuracy of thermal diffusion imaging performed by the HTOScan System for cancer detection in lung tissues immediately after extraction. By utilizing thermal diffusion imaging parameters, the study aims to evaluate the presence of cancerous lesions, potentially paving the way for a novel approach to early lung cancer diagnosis.
2. Materials
2.1. Cell Lines and Animals
Given lung cancer’s prevalence, its cell characteristics (such as squamous cell carcinoma, adenocarcinoma, or large-cell carcinoma), its surface localization, and the ease of extraction in mouse models, lung cancer was selected for this ex vivo case study. Continued progress in modeling lung cancer in mice has provided new insights into the molecular pathways governing human lung cancer and has also enabled the development of alternative tools for testing treatments against this malignancy [28]. Although there are considerable differences between mice and humans, particularly in genetics, physiology, and immunology, a more thorough understanding of these differences and their effects on whole-organism function will provide insights into relevant disease mechanisms and potential drug targets for further clinical investigation [29]. The development of various mouse models of both small cell lung cancer (SCLC) and non-small cell lung cancer (NSCLC) to mimic human lung cancer supports the use of mice to demonstrate the feasibility of utilizing the HTOScan System for early lung cancer detection [30].
Pharmaseed, a CRO specializing in preclinical studies and operating in Ness Ziona, Israel, was responsible for conducting the HT bioimaging study. The study model included 30 female mice aged 9 to 11 weeks, of whom 28 received injections, and 2 served as controls. The cancer cells used in the study were American Type Culture Collection (ATCC) Lewis Lung Carcinoma (LLC) cells, administered intravenously in 100 μL of cell solution, with growth media prepared according to ATCC’s protocol. Marinating LLC cell lines followed Pharmaseed’s Standard Operating Procedures (SOPs), Nos. 400.31 [31], 402.6 [32], and 403.5 [33], adhering to acceptable practice.
Data collection and statistical analysis were spread across four consecutive cycles to ensure tumor progression and to gather additional information. Cycle duration ranged from 24 days after the first day of each cycle. The ideal lung tumor size range was 1.5 to 5 mm, facilitating easy detection and consistent with the system’s operational capability. Mice were sacrificed at different time points to determine the optimal termination day, ensuring that the tumor did not spread or prove fatal. There were twice-daily and once-weekend morbidity and mortality inspections, with body weight measured before as per Pharmaseed’s SOP No. 010.21 [34].
The animals were observed for toxic/adverse symptoms after LLC cell injection and three times a week for clinical symptoms until study termination. Any irregularities, such as changes in the injection site, skin, fur, eyes, mucous membranes, respiratory secretions and excretions, autonomic activity, movement, posture, strange behavior, tremors, convulsions, sleep, and coma, were documented.
2.2. Heat Diffusion Imaging System
The HTOScan System uses heat-diffusion imaging to detect differences between normal and malignant tissues based on their unique thermal properties. The system shown in Figure 1 comprises a unique setup of an infrared (IR) camera that records a thermal video, an optical camera to register the examined region, and a high-power ultraviolet light-emitting diode (LED) with a wavelength of 360 nm as a heat source. The thermal imaging system used in this study was the Optris PI 450 (Optris GmbH, Berlin, Germany; thermal sensitivity 40 mK), operated at 27 FPS. The camera was factory calibrated and stabilized prior to each recording via a warm-up period. Radiometric consistency during acquisition was maintained using the camera’s automatic nonuniformity correction mechanism, following the manufacturer’s guidelines. The control software regulates the cameras, heat source, and data storage while ensuring tissue safety [35].
Figure 1.
HTOScan System hand-held prototype.
The HTOScan System has undergone successful safety testing, meeting Israeli and international standards for medical electrical systems, electromagnetic compatibility, and non-laser light source safety. Additionally, the software life cycle validation has been assured [36].
3. Methods
3.1. Algorithmic Methodology
The HTOScan System’s algorithmic methodology involves four steps: (1) Data acquisition, (2) Feature extraction and pre-processing, (3) Feature selection, and (4) Lesion classification. The classifier’s performance was evaluated by comparison with the gold standard, histological diagnosis. Also, a preliminary sample preparation step involving lung excision via standard anterior neck dissection and thoracotomy for ex vivo scanning is part of the ex vivo procedure. The excised lungs (15 mm × 20 mm, with 1.5 mm–5 mm lesions) with the trachea on top were then placed on an examination board (Figure 2) and prepared for scanning.
Figure 2.
Excised lungs placed on the examination board.
3.1.1. Data Acquisition
Following sample preparation and system setup, each mouse specimen was logged into the HTOScan System software using a unique Case ID, with details recorded on lesion characteristics. The excised lung samples underwent an ex vivo thermal diffusion using the following protocol:
- The HTOScan System was positioned 70 mm from the lung sample to ensure proper thermal camera focus, as shown in Figure 3. A manual Z-axis lab jack and a high-emissivity blackbody reference board were used to ensure stable positioning and minimize thermal-reflection artifacts.Figure 3. HTOScan System setup.
- After a 3-s baseline temperature recording, a heat source was activated for 20 s or until the sample reached 41 °C [37].
- Throughout the heating process, both infrared video (382 pixels × 288 pixels) and visible-light cameras (20 mm × 20 mm field of view) continuously captured the examined area and the temperature decay.
- The system’s mainframe recorded all relevant data regarding the operation and procedure.
- A standard protocol-based disinfection procedure was performed after each scan to prepare the system for subsequent testing.
3.1.2. Feature Extraction and Pre-Processing
The HTOScan System captures and records thermal videos at 382 pixels × 288 pixels over time. The algorithm receives the characteristic thermal signals as input for a detailed analysis of every pixel. These signals track temperature evolution from the start of the case’s trial, covering the steady-state, heating, and decay phases. Figure 4 depicts the temporal measurements of thermal signals from two sample pixels in a single specimen.
Figure 4.
Raw data of two pixels, 27 frames per second (normal tissue (orange), abnormal tissue (blue)).
Three sites, each 15 pixels by 15 pixels, were selected for analysis in each scan: two presumed normal and one abnormal from each of the 18 injected mice, and three normal sites from each of the two control mice, yielding a total of 60 sites for analysis. For each pixel, whether from a presumed healthy region or a suspected lesion area, a set of 264 distinct features was extracted and standardized, spanning three broad functional categories. The first category comprises thermal and biophysical features derived from Pennes’ bioheat transfer model, capturing pixel-level heat diffusion dynamics, including perfusion-related heat dissipation and metabolic heat generation patterns. The second category includes frequency-domain features extracted from the Fourier transform of the temporal thermal signal at each pixel, encoding periodic and spectral characteristics of the tissue’s thermal response that are not discernible in the raw time-domain signal. The third category encompasses statistical and signal-processing features, including descriptors of signal amplitude, temporal variability, asymmetry, and dynamic behavior across the acquisition period. Together, these three categories provide a comprehensive multi-domain physiological profile for each pixel, serving as the discriminative basis on which the SVM classifier operates. Consequently, the resulting dataset for each site included 264 feature columns and 225 rows of pixels. It is worth noting that although the study encompasses 60 sites, the SVM classifier operates at the pixel level, such that the effective training set comprises 13,500 pixel-level observations across the full dataset, which presents a sample-to-feature ratio that is substantially more favorable than the site count alone implies, and well-suited to the margin-maximization framework of the SVM classifier.
The features were then normalized against the specimen’s baseline, representing the relative deviation of the initial feature value from its average across all pixels across all three sites, to create new normalized features. This normalization allowed for a direct comparison of the 264 feature values between an assumed healthy baseline and pathological sites, quantifying various aspects of the pixel’s physiological profile. Presumptively normal and abnormal sites were selected visually based on macroscopic tissue morphology. Histopathology subsequently proved whether they were abnormal or normal by confirming the presence or absence of malignant cells.
3.1.3. Feature Selection
Statistical analysis techniques like ANOVA (analysis of variance) and Shannon’s entropy (a measure of data uncertainty) determined ten significant features that distinguish abnormal from normal tissues, whereas additional features contributed very little beyond these ten significant features. The analysis yielded a table with dimensions of 225 pixels by ten features, summarizing the key thermal characteristics for each site. It is important to note that applying ANOVA and Shannon entropy for feature selection served as a domain characterization step rather than a classifier optimization procedure. These filter methods assess the univariate statistical properties of each feature independently of the SVM classifier and its training procedure, with no feedback from the classification pipeline to the feature selection outcome. The resulting ten features were retained based on their biophysical and statistical significance rather than their ability to maximize classifier accuracy, and the aggressive reduction from 264 to 10 features yields a sample-to-feature ratio of 1350 pixel-level observations per feature, substantially reducing the risk of overfitting and limiting the practical magnitude of any optimism that could be attributed to the pre-cross-validation feature selection step.
3.1.4. Lesion Classification
Analysis of temporal thermal signals from each pixel at the sites revealed key thermal characteristics of the tissues. The HTOScan System’s algorithm uses a support vector machine (SVM) classifier to analyze these feature datasets. The SVM classifier was selected for its well-documented suitability for small, high-dimensional biomedical datasets [38], its strong generalization capabilities under limited sample conditions [39], and its continued application in recent thermal imaging classification literature [40]. This classifier, implemented using Python’s (v3.8) ‘scikit-learn’ library with a Gaussian Radial Basis Function kernel, categorized each pixel as 1—High Risk (abnormal) or 0—Low Risk (normal), thus creating a spatial analysis. Default hyperparameters were retained throughout (C = 1.0, gamma = ‘scale’), and no additional hyperparameter optimization was performed. The training dataset comprised 60 sites, of which 42 were normal and 18 abnormal, reflecting a moderate class imbalance (~70/30). Given this ratio, no class reweighting was applied, and model performance did not indicate systematic bias toward the majority class. Following pixel-level classification, site-level diagnosis was determined by aggregating individual pixel classifications into a single diagnostic decision using a predefined abnormal-pixel-proportion threshold. This two-stage design is intentional: pixel-level classification exploits the spatial granularity of the thermal feature space, while site-level evaluation serves as the diagnostic inference unit, explicitly accounting for the spatial non-independence of pixels within a site.
Lesion classification was validated using the leave-one-out cross-validation (LOOCV) approach at the animal level, in which all imaging sites from a single mouse were held out in each iteration, while the remaining mice were used for training. This design is suitable for small datasets, prevents intra-animal data leakage, and yields biologically meaningful results to estimate the model’s predictive performance on new data. Combining the resulting site classifications and true labels produced a confusion matrix that served as the basis for computing the algorithm’s overall performance metrics.
The HTOScan System generates a malignancy probability utilizing spatial relationships. This heat diffusion map uses a functional color scheme: red pixels indicate a higher likelihood of abnormality, and blue pixels indicate normality. The underlying algorithm performs a pixel-wise classification, assigning each pixel a malignancy probability score between 0.0 and 1.0. The resulting HDI map visually represents the tissue’s cancerous or healthy state.
3.2. Histopathology Assessment
Patho-Logica performed histopathological analysis using standard protocols for inverted block sectioning and H&E staining. Key details include:
- Passive formalin injection to avoid artifacts.
- Systematic sectioning along pulmonary arteries:
- Six longitudinal sections per lung (three dorsal, three ventral).
- 150 μm intervals between sections.
- Approximately 5 μm section thickness.
- Pathological examination and lesion mapping (Figure 5a).Figure 5. Histopathology assessment: (a) schematic representation of the lung compartmentalization. The lung specimen was divided into four quadrants by two perpendicular axes centered on the trachea and left bronchus, which serve as the anatomical reference point for compartmentalization (b) a set of 3 pathological lesion slides.
- Additional sectioning to capture the lesion surface effectively (Figure 5b):
- Longitudinal horizontal sections on each lung surface.
- Approximately 100 μm intervals.
- Three sections per side, six total per mouse.
- Imaging with Olympus BX60 microscope at 1.25× magnification.
4. Results
The study analyzed 30 mice: 18 were designated experimental animals, and 2 served as healthy controls. During the tumor culturing period, an additional set of 10 mice was used to determine the optimal tumor growth conditions before analysis, including the minimum detectable lesion size and the optimal sacrifice interval to obtain tumors above the threshold by sacrificing these mice at different time points throughout the tumor culture period. Two normal and one abnormal site from each of the 18 injected mice, and three normal sites from each of the two control mice, yielded a total of 60 sites for analysis.
The HTOScan System’s algorithm classified site thermal features using mathematical, physical, and signal-processing methods. Figure 6 presents the discriminative power of the extracted feature set by illustrating the progressive separation between normal (blue) and abnormal (orange/red) pixels across one, two, and three feature dimensions, respectively. In the left panel, a single representative feature already produces a partial but meaningful distributional separation between the two-pixel populations, demonstrating that individual features carry a genuine physiological signal. The center panel reveals that combining two features substantially enhances this separation, with the abnormal pixel cluster forming a visually distinct trajectory from the normal cluster in two-dimensional feature space. The right panel demonstrates that in three-dimensional feature space, the two populations resolve into geometrically distinct volumetric clusters with markedly reduced overlap. This progressive improvement in class separability across panels provides direct empirical support for the construction of a rich, high-dimensional feature space, as the cumulative discriminative power of the combined feature set substantially exceeds that of any individual feature in isolation, thereby justifying the 264-feature approach adopted in this study.
Figure 6.
Features extracted by the HTOScan algorithm across all pixels of all sites. Orange/Red—Abnormal pixels, Blue—Normal Pixels; (Left) one-dimensional histogram of a single representative feature, (Center) two-dimensional scatter plot of two features, and (Right) three-dimensional feature space visualization. The HTOScan System collects total pixel identification to construct heat diffusion maps with a resolution of 0.1 mm per pixel.
A comparison with the pathology results for the same tissue sample enabled the HTOScan System’s abnormality predictions for each pixel to be validated. The analysis used clinically identified locations, each 15 × 15 pixels, captured from each specimen, to demonstrate the system’s ability to correctly classify lesions. Figure 7 demonstrates an example of a specimen with marked abnormal and normal sites, illustrating the anatomical overlay between the optical image and the static infrared (IR) image acquired prior to thermal excitation.
Figure 7.
Specimen marked with normal (green outline) and abnormal (red outline) sites: (a) optical image, (b) IR image.
The HTOScan System diffusion maps were correlated with the pathology findings, as demonstrated in representative cases (Figure 8a–e). The comparison demonstrated for each case includes the optical image (left side) of the macroscopic specimen, the HTOScan System diffusion map (in the middle) (ranging from blue (normal) to red (abnormal)) superimposed on the optical image, and the pathology slide (right side). To preserve pixel orientation, the trachea and the outer outlines of both lungs were used as reference points, as shown in Figure 8.
Figure 8.
Representative cases of the HTOScan System diffusion maps correlating with the pathology findings: (a) Example #1, (b) Example #2, (c) Example #3, (d) Example #4, and (e) Example #5. Each case presents the optical image (left), the corresponding HTOScan heat diffusion map (center), and the H&E-stained histopathological slide (right). The diffusion map employs a thermal color scale ranging from red, indicating higher malignancy likelihood, to blue, indicating normal tissue.
The HTOScan System diffusion maps encompassed all abnormal areas described in the histopathology analysis in this trial.
Appendix A summarizes the HTOScan System’s algorithmic prediction based on data from all sites investigated, compared with the histopathology results. The validation procedure comprised 20 iterations, with one sample (i.e., a mouse and not a lesion) excluded from the dataset. The performance analysis for all sites was demonstrated based on the calculated sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), accuracy, and receiver operating characteristic (ROC) curve. The trained classifier’s performance was measured using the omitted patient’s data and compared with the gold standard (histological diagnosis). The analysis results were summarized in a confusion matrix, as shown in Table 1. Accordingly, the system’s performance was evaluated, yielding the parameters presented in Table 2.
Table 1.
Confusion matrix: Abnormal (positive) vs. normal (negative) per site.
Table 2.
System’s performance parameters per site.
The results demonstrated high accuracy, with only 1 false positive and 1 false negative out of 60 total sites analyzed across 20 mice. The ROC graph showed an area under the ROC curve (AUC) of 0.92 (Figure 9).
Figure 9.
ROC curve.
5. Discussion
This study aimed to demonstrate the HTOScan System’s functional imaging platform’s ability to distinguish between malignant (abnormal) and benign (normal) tissue using an ex vivo preclinical study. The ex vivo nature of this study enhances its robustness, as lung tissue extraction demonstrates the technology’s ability to distinguish structural changes in tissues independent of vascularity and metabolic activity. The results suggest that each tissue’s unique thermal-physical properties, heat capacity, conductivity, and density maintain their characteristics in normal and cancerous tissue ex vivo, enabling differentiation between cancerous and non-cancerous tissues. These findings are supported by high diagnostic performance, as indicated by 94% sensitivity and 98% specificity, 94% PPV and 98% NPV, and an overall 97% accuracy. It should be noted that the ex vivo design, rather than representing a limitation, strengthens the proof-of-concept by demonstrating that the HTOScan System achieves tissue discrimination based solely on intrinsic thermal-physical properties in the absence of perfusion and metabolic activity. These two physiological processes are known to amplify further thermal contrast between malignant and normal tissue in vivo. The ability to distinguish tissue types without relying on these additional signals therefore establishes a conservative performance baseline, suggesting that in vivo conditions are likely to provide additional discriminative information that may further enhance classification performance.
The high efficacy of the system indicates superior diagnostic performance relative to reported limitations of several light-based diagnostic approaches, which have low diagnostic accuracy and high false-positive and moderate false-negative rates. Thus, as previously mentioned, it may offer an advantage over standalone guided bronchoscopy localization methods, such as REBUS, VBN, and related approaches [41]. While these performance metrics suggest a potential advantage over existing diagnostic approaches, it is important to note that a formal qualitative and quantitative comparison with related technologies is currently limited. To the best of our knowledge, there are no directly comparable imaging-based diagnostic platforms that rely on similar thermal physical characterization principles. Therefore, direct benchmarking against equivalent methods is not yet feasible. Instead, the current results should be interpreted as a proof-of-concept demonstration, establishing the foundation for future studies that will enable systematic comparative evaluation as additional technologies and datasets become available.
The HTOScan System data analysis utilizes a proprietary closed-source AI platform. Since the primary goal of this POC study was to demonstrate feasibility, the selected features were based on tissue characteristics and properties, without attempting to represent biological processes. While feature selection was applied globally prior to cross-validation, its filter-based nature and the aggressive reduction to 10 features from an initial set of 264 are considered unlikely to meaningfully affect the validity of the reported diagnostic metrics, as leakage-induced bias in SVM-based classifiers has been shown to converge toward zero as the number of retained features approaches 10 [42]. Nonetheless, as previously mentioned, because mouse lungs closely mimic human lung cancer, the data gathered are clinically relevant and of medical importance. However, the limited data present challenges for algorithm training, which requires additional data before implementation in human tissue. The sample size of 60 sites derived from approximately 20 animals reflects a pragmatic choice to enable an initial feasibility assessment, balancing resource constraints with the need to obtain a preliminary statistical signal. This study was not designed for definitive inference but rather to explore early trends that may justify a larger-scale investigation. Notably, a sample size in this range is often considered sufficient for approximate application of the central limit theorem, supporting basic distributional assumptions, although this heuristic does not replace a formal power-based design. The reported performance metrics should therefore be interpreted within this context, and validation in larger and more diverse datasets remains a necessary step prior to clinical translation. By adopting a site-based analysis rather than a pixel-based one, the impact of data scarcity is mitigated. A site-based methodology is inherently more resilient to data limitations. It therefore does not require an extensive dataset to derive robust conclusions, in contrast to pixel-level analyses, which are more susceptible to sparse data and may require a larger dataset for robust results. Moreover, it offers a benefit over histopathology, as a few slices may not reveal surface lesions. The controlled preclinical design mitigates potential sources of bias. Though regions of interest (ROIs) were determined visually, they were subsequently confirmed histopathologically. Also, normal tissue characterization was established through dual-source validation: systematic sampling from tumor-bearing animals was supplemented by samples from dedicated healthy control mice. This approach provides a robust, representative normal-tissue baseline for the algorithm’s classification metrics.
6. Conclusions
This proof-of-concept study evaluated the HTOScan System’s functional imaging modality and its AI-based algorithm in an ex vivo preclinical lung cancer mouse model, demonstrating high diagnostic performance in differentiating cancerous from non-cancerous lung tissue based on intrinsic thermal-physical tissue properties.
While these findings are promising, several limitations warrant acknowledgment. The study was conducted ex vivo on a small dataset, and although the site-based analytical approach mitigates some constraints imposed by limited sample size, further validation with larger, more diverse datasets is necessary prior to clinical translation. Additionally, establishing well-characterized normal-tissue baselines by comparing with control samples will strengthen the algorithm’s discriminative reliability. Future implementations should consider incorporating feature selection within each cross-validation fold to satisfy the independence requirements of rigorous cross-validation frameworks fully.
Looking ahead, the HTOScan System holds potential across multiple clinical applications. Near-term priorities include in vitro validation studies and expansion of the training dataset to enhance algorithmic robustness. Subsequently, in vivo investigation during oncological surgery represents a compelling avenue, particularly for real-time surgical margin assessment and intraoperative tissue identification. Ultimately, successful translation of this technology could provide surgeons with an objective, real-time tool for estimating malignancy probability, complementing existing diagnostic modalities and improving patient outcomes.
Author Contributions
A.G., A.O. and G.P. contributed to the study’s conception and design. H.T. developed the system design. S.G. and M.T. performed the thermodynamic calculations. S.T., G.A. and A.Y. performed the algorithmic analysis. S.G. and G.P. wrote the first draft of the manuscript, and all authors commented on previous versions. All authors have read and agreed to the published version of the manuscript.
Funding
This research received no external funding.
Institutional Review Board Statement
This study complied with “The Israel Animal Welfare Act” and obtained Ethics Committee approval #IL-19-11-449 from the “The Israel Board for Animal Experiments” on 19 November 2019. All animal program personnel assisted the AV in overseeing animal health and husbandry. Pharmaseed is a preclinical CRO, working according to the OECD principles of Good Laboratory Practice ENV/MC/CHEM (98)17. Pharmaseed was responsible for performing the HT bioimaging study according to protocol HTB002EF and the ethics approval. Pharmaseed conducts its preclinical studies at Ness–Ziona, Israel. Its facility houses a vivarium with a surgery suite and several procedure rooms, and it is inspected routinely by the Israeli National Council for Experiments on Animal Subjects and the Israel G.L.P. compliance monitoring unit. The Israel Laboratory Accreditation Authority (ISRAC) authorized conducting toxicity studies in compliance with the OECD principles of G.L.P. An Attending Veterinarian (AV) examined the animals to determine their fitness for the study before the initiation of the experimental phase and then inspected them daily for any signs of morbidity or mortality. The Study Director ensured that all the research and technical personnel were adequately trained and experienced in implementing the required procedures.
Informed Consent Statement
Not applicable.
Data Availability Statement
The datasets presented in this article are not readily available because the HTOScan System data analysis utilizes a proprietary, closed-source AI platform. Requests to access the datasets should be directed to the corresponding author.
Conflicts of Interest
Authors Gal Aviram, Assaf Gur, Shani Toledano and Gabriel Polliack are employed by H.T. BioImaging Ltd. The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Appendix A
Table A1.
Analysis per site: pathology vs. the HTOScan System’s algorithm prediction. Green indicates concordance between histopathological findings and the HTOScan System prediction; red indicates discordance.
References
- Sung, H.; Ferlay, J.; Laversanne, M.; Soerjomataram, I.; Jemal, A.; Bray, F. Global Statistics 2020: GLOBOCAN Estimates of Incidence and Mortality Worldwide for 36 Cancers in 185 Countries. CA A Cancer J. Clin. 2021, 71, 209–249. [Google Scholar] [CrossRef] [Scilit]
- World Health Organization (WHO). Cancer Facts. 3 February 2022. Available online: https://www.who.int/news-room/fact-sheets/detail/cancer (accessed on 14 February 2023).
- World Health Organization (WHO). Cancer. Available online: https://www.who.int/health-topics/cancer#tab=tab_1 (accessed on 14 February 2023).
- Balasubramanian, P.; Abia-Trujillo, D.; Barrios-Ruiz, A.; Garza-Salas, A.; Koratala, A.; Chandra, N.C.; Lee-Mateus, A.Y.; Labarca, G.; Fernandez-Bussy, S. Diagnostic yield and safety of diagnostic techniques for pulmonary lesions: Systematic review, meta-analysis and network meta-analysis. Eur. Respir. Rev. 2024, 33, 240046. [Google Scholar] [CrossRef] [Scilit]
- Magnini, A.; Fissi, A.; Cinci, L.; Landini, N.; Nardi, C. Diagnostic accuracy of imaging-guided biopsy of peripheral pulmonary lesions: A systematic review. Acta Radiol. 2024, 65, 1222–1237. [Google Scholar] [CrossRef] [Scilit]
- Nadig, T.R.; Thomas, N.; Nietert, P.J.; Lozier, J.; Memoli, J.; Pastis, N.; Silvestri, G. Guided bronchoscopy for the evaluation of pulmonary lesions: An updated meta-analysis. Chest 2023, 163, 1589–1598. [Google Scholar] [CrossRef] [Scilit]
- Tian, S.; Li, X.; Liu, J.; Wang, X.; Chen, H.; Dai, Z.; Chen, Q.; Shi, H.; Li, Y.; Huang, H.; et al. Radial endobronchial ultrasound-guided bronchoscopy for the diagnosis of peripheral pulmonary lesions: A systematic review and meta-analysis of prospective trials. Heliyon 2024, 10, e29446. [Google Scholar] [CrossRef] [Scilit]
- Kops, S.E.; Heus, P.; Korevaar, D.A.; Damen, J.A.; Idema, D.L.; Verhoeven, R.L.; Annema, J.T.; Hooft, L.; van der Heijden, E.H. Diagnostic yield and safety of navigation bronchoscopy: A systematic review and meta-analysis. Lung Cancer 2023, 180, 107196. [Google Scholar] [CrossRef] [Scilit]
- Folch, E.E.; Bowling, M.R.; Pritchett, M.A.; Murgu, S.D.; Nead, M.A.; Flandes, J.; Krimsky, W.S.; Mahajan, A.K.; LeMense, G.P.; Murillo, B.A.; et al. NAVIGATE 24-month results: Electromagnetic navigation bronchoscopy for pulmonary lesions at 37 centers in Europe and the United States. J. Thorac. Oncol. 2022, 17, 519–531. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Katsis, J.; Roller, L.; Aboudara, M.; Pannu, J.; Chen, H.; Johnson, J.; Lentz, R.J.; Rickman, O.; Maldonado, F. Diagnostic yield of digital tomosynthesis-assisted navigational bronchoscopy for indeterminate lung nodules. J. Bronchol. Interv. Pulmonol. 2021, 28, 255–261. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Li, Z.; Xu, S.; Zhang, Y.; Shi, J.; Li, Z.; Xu, S.; Zhang, Y.; Shi, J. Efficacy and safety of cone-beam computed tomography-guided bronchoscopy for peripheral pulmonary lesions: A systematic review and meta-analysis. J. Thorac. Dis. 2025, 17, 551–563. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Constantinescu, A.; Stoicescu, E.R.; Iacob, R.; Chira, C.A.; Cocolea, D.M.; Nicola, A.C.; Mladin, R.; Oancea, C.; Manolescu, D. CT-guided transthoracic core-needle biopsy of pulmonary nodules: Current practices, efficacy, and safety considerations. J. Clin. Med. 2024, 13, 7330. [Google Scholar] [CrossRef] [Scilit]
- Yang, H.; Huang, J.; Zhang, Y.; Guo, J.; Xie, S.; Zheng, Z.; Ma, Y.; Deng, Q.; Zhong, C.; Li, S. The diagnostic performance and optimal strategy of cone beam CT-assisted bronchoscopy for peripheral pulmonary lesions: A systematic review and meta-analysis. Pulmonology 2024, 31, 2420562. [Google Scholar] [CrossRef] [Scilit]
- Pritchett, M.A.; Bhadra, K.; Calcutt, M.; Folch, E. Virtual or reality: Divergence between preprocedural computed tomography scans and lung anatomy during guided bronchoscopy. J. Thorac. Dis. 2020, 12, 1595–1611. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Wijma, A.N.; Casal, R.F.; Cheng, G.Z.; Einsiedel, P.F.; Fantin, A.; Hall, D.J.; Herth, F.J.; Ng, C.S.; Pritchett, M.A.; Shah, P.L.; et al. Radiation principles, protection, and reporting for interventional pulmonology: A world association of bronchology and interventional pulmonology white paper. Respiration 2024, 103, 707–722. [Google Scholar] [CrossRef] [Scilit]
- Gonzalez, A.V.; Silvestri, G.A.; Korevaar, D.A.; Gesthalter, Y.B.; Almeida, N.D.; Chen, A.; Gilbert, C.R.; Illei, P.B.; Navani, N.; Pasquinelli, M.M.; et al. Assessment of advanced diagnostic bronchoscopy outcomes for peripheral lung lesions: A Delphi consensus definition of diagnostic yield and recommendations for patient-centered study designs. An official American Thoracic Society/American College of Chest. Am. J. Respir. Crit. Care Med. 2024, 209, 634–646. [Google Scholar]
- D’Alessandro, G.; Tavakolian, P.; Sfarra, S. A review of techniques and bio-heat transfer models supporting infrared thermal imaging for diagnosis of malignancy. Appl. Sci. 2024, 14, 1603. [Google Scholar] [CrossRef] [Scilit]
- Pal, U.M.; GK, A.V.; Varma, M.; Vaidya, J.S.; Pandya, H.J. Thermo-optic measurements and their inter-dependencies for delineating cancerous breast biopsy tissue from adjacent normal. J. Biophotonics 2021, 14, e202100041. [Google Scholar] [CrossRef] [Scilit]
- Camilleri, J.S.; Farrugia, L.; Curto, S.; Rodrigues, D.B.; Farina, L.; Dingli, G.C.; Bonello, J.; Farhat, I.; Sammut, C.V. Review of thermal and physiological properties of human breast tissue. Sensors 2022, 22, 3894. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Bianchi, L.; Cavarzan, F.; Ciampitti, L.; Cremonesi, M.; Grilli, F.; Saccomandi, P. Thermophysical and mechanical properties of biological tissues as a function of temperature: A systematic literature review. Int. J. Hyperth. 2022, 39, 297–340. [Google Scholar] [CrossRef] [Scilit]
- Gutierrez, C.; Owens, A.; Medeiros, L.; Dabydeen, D.; Sritharan, N.; Phatak, P.; Kandlikar, S.G. Breast cancer detection using enhanced IRI-numerical engine and inverse heat transfer modeling: Model description and clinical validation. Sci. Rep. 2024, 14, 3316. [Google Scholar] [CrossRef] [Scilit]
- Lozano, A., III; Hayes, J.C.; Compton, L.M.; Azarnoosh, J.; Hassanipour, F. Determining the thermal characteristics of breast cancer based on high-resolution infrared imaging, 3D breast scans, and magnetic resonance imaging. Sci. Rep. 2020, 10, 10105. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Jiji, L.M. Heat Conduction; Springer Berlin: Heidelberg, Germany, 2009. [Google Scholar]
- Sritharan, N.; Gutierrez, C.; Perez-Raya, I.; Gonzalez-Hernandez, J.L.; Owens, A.; Dabydeen, D.; Medeiros, L.; Kandlikar, S.; Phatak, P. Breast cancer screening using inverse modeling of surface temperatures and steady-state thermal imaging. Cancers 2024, 16, 2264. [Google Scholar] [CrossRef] [Scilit]
- Pennes, H.H. Analysis of Tissue and Arterial Blood Temperatures in the Resting Human Forearm. J. Appl. Physiol. 1948, 1, 93–122. [Google Scholar] [CrossRef] [Scilit]
- Huang, D.; Fang, Y.; Zheng, W.; Peng, Y.; Liu, J.; She, J.; Chen, C.; Yue, Y. Leveraging Thermal Properties Data toward Efficient Photothermal Therapy of Mammary Glands. Adv. Ther. 2023, 6, 2200348. [Google Scholar] [CrossRef] [Scilit]
- Dank, G.; Buber, T.; Polliack, G.; Aviram, G.; Rice, A.; Yehudayoff, A.; Kent, M.S. A pilot study for a non-invasive system for detection of malignancy in canine subcutaneous and cutaneous masses using machine learning. Front. Vet. Sci. 2023, 10, 1109188. [Google Scholar] [CrossRef] [Scilit]
- De Seranno, S.; Meuwissen, R. Progress and applications of mouse models for human lung cancer. Eur. Respir. J. 2010, 35, 426–443. [Google Scholar] [CrossRef] [Scilit]
- Rydell-Törmänen, K.; Johnson, J.R. The Applicability of Mouse Models to the Study of Human Disease. Mouse Cell Cult. Methods Protoc. 2019, 1940, 3–22. [Google Scholar]
- Safari, R.; Meuwissen, R. Practical Use of Advanced Mouse Models for Lung Cancer. In Mouse Models of Cancer: Methods and Protocols; Springer: New York, NY, USA, 2015; pp. 93–124. [Google Scholar]
- Division of Comparative Medicine. Rodent Husbandry, SOP#400.31; University of South Florida: Tampa, FL, USA, 2022. [Google Scholar]
- Division of Comparative Medicine. Sentinel Rodent Exposure, SOP#402.6; University of South Florida: Tampa, FL, USA, 2017. [Google Scholar]
- Division of Comparative Medicine. Environmental Enrichment and Social Housing, SOP#:403.5; University of South Florida: Tampa, FL, USA, 2022. [Google Scholar]
- Division of Comparative Medicine. Handling, Storage, and Retrieval of Records and Data, SOP#010.21; University Of South Florida: Tampa, FL, USA, 2022. [Google Scholar]
- Zamanian, A.; Hardiman, C.J. Electromagnetic Radiation and Human Health: A Review of Sources and Effects. High Freq. Electron. 2005, 4, 16–26. [Google Scholar]
- The Standards Institution of Israel, Electrical Safety, Medical Devices. Available online: https://www.sii.org.il/en/medical-equipment-safety (accessed on 15 September 2022).
- Hall, E.J.; Giaccia, A.J. Radiobiology for the Radiologist, 8th ed.; Lippincott Williams & Wilkins: Philadelphia, PA, USA, 2018. [Google Scholar]
- Selvam, J.; Madhavan, M. Automatic Detection and Classification of Thyroid Nodules by Thermal Imaging. Biomed. Pharmacol. J. 2025, 18, 1422–1431. [Google Scholar] [CrossRef] [Scilit]
- Guido, R.; Ferrisi, S.; Lofaro, D.; Conforti, D. An overview on the advancements of support vector machine models in healthcare applications: A review. Information 2024, 15, 235. [Google Scholar] [CrossRef] [Scilit]
- Amaya-Tejera, N.; Gamarra, M.; Zurek, E. A distance-based kernel for classification via Support Vector Machines. Front. Artif. Intell. 2024, 7, 1287875. [Google Scholar] [CrossRef] [Scilit]
- Matsumoto, Y.; Kho, S.S.; Furuse, H. Improving diagnostic strategies in bronchoscopy for peripheral pulmonary lesions. Expert Rev. Respir. Med. 2024, 18, 581–595. [Google Scholar] [CrossRef] [Scilit]
- Shim, M.; Lee, S.H.; Hwang, H.J. Inflated prediction accuracy of neuropsychiatric biomarkers caused by data leakage in feature selection. Sci. Rep. 2021, 11, 7980. [Google Scholar] [CrossRef] [Scilit]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.