Development of a Mathematical Model to Determine the Stability of Osteosynthesis in Pertrochanteric Fractures

,
,
,
and
1
University Clinic of Traumatology, Orthopedic Diseases, Anesthesia, Resuscitation, Intensive Care and Emergency Center (TOARILUC), Clinical Center “Mother Teresa”, 1000 Skopje, North Macedonia
2
Clinic for Orthopaedics and Traumatology “Academician Prof. Dr. Milorad Mitković”, University Clinical Center Nis, 18000 Nis, Serbia
3
Faculty of Medicine, University of Nis, 18104 Nis, Serbia
4
Faculty of Mechanical Engineering, University “Ss. Cyril and Methodius” in Skopje, 1000 Skopje, North Macedonia
This article belongs to the Section Biomedical Engineering

Abstract

Background and Objectives: Determining the mechanical stability of osteosynthesis in pertrochanteric fractures remains a critical challenge in orthopedic biomechanics. The aim of this study was to develop a mathematical model for quantifying the stability of osteosynthesis and to establish criteria for its evaluation under physiological loading conditions. Materials and Methods: A mathematical model describing the biomechanical behavior of a proximal femur with a pertrochanteric fracture stabilized using a cephalomedullary nail (CMN) was developed. The model integrates force equilibrium, stress–strain relationships, and loading conditions representative of early functional rehabilitation. The theoretical framework was implemented in MATLAB/Simulink R2025b and complemented by finite element analysis to determine stress distribution, deformation patterns, and stability-related parameters of the bone–implant system. Results: The developed mathematical model enabled a quantitative assessment of osteosynthesis stability through the evaluation of key mechanical indicators, including displacement, stress distribution, and safety factor within the fixation system. Critical stress zones in the implant and surrounding bone were identified, allowing analysis of load transfer mechanisms. Finite element simulations showed that improved fixation mechanics reduced peak implant stresses, limited displacement at the fracture site, and increased the safety factor of the fixation construct, resulting in a more uniform load distribution in the surrounding bone and enhanced overall stability of the osteosynthesis system. Conclusions: The proposed mathematical model provides a systematic approach for determining the stability of osteosynthesis in pertrochanteric fractures. It offers a theoretical basis for optimizing implant design and fixation strategies, with potential applications in preclinical evaluation and surgical planning.

1. Introduction

Around 1.5 million hip fractures occur each year, and this number is expected to increase to 6.3 million by 2050 (Abrahamsen et al. [1]). In elderly populations, the decline in bone mineral density and structural strength often results in progressive degeneration of bones and joints (Imwinkelried [2]). Among the different types of hip fractures, pertrochanteric fractures constitute nearly half of all cases, making them a major clinical concern. Despite the widespread use of internal fixation as the standard treatment, these fractures are associated with a high incidence of complications. Femoral fractures represent serious injuries that predominantly affect elderly and senile patients, most of whom present with multiple comorbidities (Lee et al. [3]). Early mobilization is essential in this population to reduce the risk of somatic complications associated with prolonged immobility. It is generally accepted that patients with femoral fractures can be mobilized only following surgical intervention, preferably using the least invasive approach possible (Wu et al. [4]).
Replacement of damaged or lost bone and joint structures following injury is commonly achieved using artificial implants or prostheses. It is well established that bone tissue could regenerate, and when implants are used to stabilize and align bone fractures, the tissue can gradually heal and restore itself over time.
Pertrochanteric fractures occur in the proximal femur between the greater and lesser trochanters and are commonly observed in elderly patients, most often as a result of low-energy falls associated with osteoporosis. Typical clinical features include hip pain, limb shortening, and external rotation of the affected leg (Lu et al. [5]).
Li et al. [6] present a finite element analysis comparing three distinct internal fixation implants used in the treatment of femoral fractures. In the paper, the authors designed a plate that combines the advantages of both cannulated screws (CCS) and the sliding hip screw. The design incorporates three dynamic screws positioned at a fixed angle to a side plate, referred to as the Slide Compression Anatomic Plate–Femoral Neck (SCAP-FN). The SCAP-FN provides stable fixation and demonstrates improved performance in terms of stress distribution, reduction in stress peaks, and control of rotational angles. In the study by Cheol-Jeong Kim et al. [7], finite element analysis was used to compare the biomechanical stability of various implants for femoral fracture treatment. The results of this study provide valuable insights that could significantly inform decision-making regarding surgical strategies for intertrochanteric fractures. Jiantao Li et al. [8] performed a comparative analysis of the biomechanical behavior of the Medial Sustainable Nail (MSN) and the Proximal Femoral Nail Antirotation (PFNA) in the treatment of AO/OTA 31-A2.3 fractures. Finite element analysis (FEA) was conducted using Abaqus 6.14, applying an axial load of 2100 N to evaluate von Mises stress distribution and overall displacement in both implant models. Similarly, Radek Bartoska et al. [9] employed a numerical model combined with finite element analysis to examine different positioning configurations of intramedullary hip nails (IMHN/PFH) within the proximal femur. The objective was to identify implant positions that exhibit a higher likelihood of mechanical failure, providing valuable guidance for optimizing implant placement and enhancing fixation stability. Furthermore, finite element analysis was utilized to compare the biomechanical performance of a conventional titanium alloy Gamma nail and a biodegradable magnesium alloy bionic Gamma nail for the treatment of intertrochanteric fractures (Li et al. [10]). A 3D model of the proximal femur with a simulated 31A1 intertrochanteric fracture was developed from CT data. Two implant types were analyzed to evaluate the effect of bone ingrowth on stress distribution after fixation.
Bai et al. [11] had used FEA in the von Mises stress analysis in order to determined maximum and minimum stresses, were to assess bone ingrowth influence on stress distribution in the proximal femur post-fixation in 31A1 trochanteric fractures.
Mahaisavariya et al. [12] reported about comparison between intramedullary and extramedullary fixation of basocervical femoral fractures using FEA. Basocervical femoral fractures are similar to pertrochanteric fractures. They reported higher implant loads in DHS compared with intramedullary nailing but observed stress concentration at the interface between the lag screw and the nail body, mainly within internal intersection planes rather than on the model surface. This may explain the differences relative to our study.
Recent advances in additive manufacturing have enabled the development of personalized orthopedic implants with optimized mechanical performance. Selective laser melting (SLM) techniques have been successfully applied to fabricate individualized femur implants, allowing precise control over geometry and internal structure to improve load distribution and biomechanical compatibility, as reported in the study of Zhang et al. [13]. In addition, optimization of the 3D printing process components, such as high-performance powder-spreading arms, has been shown to enhance the reliability and efficiency of metal additive manufacturing systems, further supporting the production of complex biomedical devices (Zhang et al. [14], Hachimi et al. [15]). These advances highlight the potential of integrating advanced manufacturing techniques with implant design to achieve improved clinical outcomes.
The literature review further demonstrate that the instability of current osteosynthesis methods stems primarily from ineffective approaches to fixation mechanics. When fixation mechanics are appropriately addressed, osteosynthesis can provide stable fracture fixation even under early functional loading of the injured limb—fully in line with the fundamental principles of mechanics. We therefore suggest that adopting the principle of active fixation with tension may serve as a promising direction for advancing osteosynthesis techniques in the management of this complex and socially significant pathology.
This paper focuses on the theoretical analysis of a dynamic model system based on empirical calculations, mathematical modeling, simulations performed in MATLAB/Simulink R2025b, virtual testing using finite element analysis and experimental validation. This study investigates the biomechanical stability of pertrochanteric fractures treated with a trapezoidal nail featuring integrated dual-screw interlocking (InterTAN), with a focus on assessing fracture fixation stability.

2. Materials and Methods

A dynamic model of the knee system (DMKS) was developed to analyze the movements of the lower and upper leg, which determine the loads on the joints and other parts of the limbs (Figure 1). The key feature of the DMKS (hip–knee–foot) is that it treats the knee not as an isolated element, but as an integral part of the leg. It accounts for the influence of forces from the body, thigh, and lower leg. The model links the system’s coordinates and center of mass with the projections of the support reaction acting on the foot, making it possible to calculate the forces acting on the body, thigh, and lower leg. Compared with other knee system models, DMKS has the advantage of clearly representing the time dependence of the calculated values. Another strength of DMKS is that the graphical solutions of the differential equations illustrate how the angles between the hip and the vertical, and between the body and the vertical, change over time. This visualization highlights both the dynamics of the movements and the interdependence between these angles.
Figure 1. A dynamic model of the knee system (DMKS).
In prosthetics, a key challenge is the accurate assessment of loads and forces acting on different parts of the body during movement, particularly on the knee and hip joints. To address this, various experimental setups—such as motion simulators—are used.
However, such experiments are often time-consuming and expensive, and therefore not always feasible. Moreover, this research method cannot account for all the factors that occur in real human motion. For this reason, a more effective and economical alternative is the development of a mathematical model capable of accurately describing the movements of the knee and hip.
A dynamic model of the knee system (DMKS) is schematically represented in Figure 1. The dynamic biomechanical model consists of the pelvis, hip joint, femur bone, knee joint, shinbone and contact of the foot with the ground.
  • Q —ground reaction force acting on the foot;
  • Q x —projection of the ground reaction force on the x—axis;
  • Q y —projection of the ground reaction force on the y—axis;
  • m b —body weight;
  • m f —femur weight;
  • m s —shinbone weight;
  • l b —body height;
  • l f —femur height;
  • l s —shinbone height;
  • α —the angle between the body and the vertical Y-axis, which varies during gait and changes over time α = α t ;
  • β —the angle between the femur and the vertical Y-axis, which varies over time β = β t ;
  • γ —the angle between the shinbone and the vertical Y-axis, which varies over time γ = γ t ;
J b = 1 2 m b + m f + m s l b 2 + l f + l s 2
—the moment of inertia of the patient’s body relative to the ground reference frame;
J f = 1 12 m f + m s l f 2 + l s 2
—the moment of inertia of the patient’s femur relative to the ground reference frame;
J s = 1 12 m s l s 2 2
—the moment of inertia of the patient’s shinbone relative to the ground reference frame.
Variation in the foot’s x and y coordinates in accordance with the dynamic model presented in Figure 1.
x = x 0 − l f s i n β + l s s i n γ
y = l f c o s β + l s c o s γ
The x and y coordinates of the foot change over time.
The ground reaction force acting on the foot can be calculated by equation:
Q = Q x 2 + Q y 2
The values of x c ¨ and y c ¨ are calculated from the differential equations of motion of the center of mass of the system shown in Figure 1.
x c ¨ = Q x m b + m f + m s
y c ¨ = Q y − m b + m f + m s g m b + m f + m s
The coordinates x c and y c   of the center of mass of the considered system shown in Figure 1 depends on time and are calculated using the following equations:
x c = x ( m b + m f + m s ) − m b   r b   s i n α + m f   r f   s i n β − m s   r s   s i n γ                                     m b + m f + m s
y c = y ( m b + m f + m s ) − m b   r b   c o s α + m f   r f   c o s β − m s   r s   c o s γ                                     m b + m f + m s
r b ,   r f and r s —represents distance between hip joint to center of body mass, femur and shinbone, and g—represents acceleration due to gravity.
Of particular interest in analyzing the knee structure shown in Figure 1 are the forces acting on the upper and lower leg, which vary in both magnitude and direction during walking and are functions of time. To determine these forces, we apply Lagrange’s equations of motion.
The forces on the upper and lower leg cannot be examined without also considering the forces acting on the body while walking. Of primary importance is the hip contact force Ff acting on the upper leg, which is transmitted to the hip joint according to the principle of action and reaction. This force varies in both magnitude and direction over a half-period during normal single-leg walking. The half-period is assumed to be 8 s, making this force a key factor in the analysis. Accordingly, Figure 2 model normal single-leg walking using an 8 s half-period.
Figure 2. Diagram showing the variation in hip contact force F f for patient 90 [kg]/180 [cm].
The functional (active) force acting on the hip, Ff, is determined as a function of time t, as well as the patient’s weight and height, using the MATLAB/Simulink R2025b software package. Unlike in previous calculations where it was assumed constant, the force Ff (measured in %BW, percentage of body weight) is variable and depends on the patient’s weight.
F f = J f β ¨ − r f m f + m s g + x ¨ c o s β + y ¨ s i n β + r s m s γ ¨ c o s β − γ − γ ˙ 2 s i n β − γ
A patient who weighs 90 kg and is 180 cm tall (shortly written as (90/180)) has the following values:
m f = 12.971   kg ,     m s = 3.831 kg ,   l f = 0.495   m ,   l s = 0.4059   m ,     r f = 0.2178   m ,     r s = 0.168   m ,     J f = 0.5977   Nm 2
By substituting the obtained values of β and γ into Equation (8), we obtain the tibial force F f in (%BW), which is the quantity of interest.
For a person weighing 90 kg and 180 cm tall, a Simulink model was developed to determine the hip contact force during normal single-leg walking over an 8 s half-period.
The MATLAB Simulink block diagram is used to solve Equation (8). The Simulink block diagram presented in Figure 3 is used to compute the hip force acting on the femur F f , for a patient weighing 90 kg and measuring 180 cm in height.
Figure 3. Simulink block diagram for computing of the hip contact force acting on the femur F f for patient 90 [kg]/180 [cm].
Determination of the angles β and γ of the upper and lower leg as functions of time (t), as well as the patient’s weight and height, using the MATLAB/Simulink R2025b software package.
The resulting equations for a patient weighing 90 kg and 180 cm tall are presented below:
− 0.36 β ¨ c o s β − β ˙ sin β − 0.42 γ ¨ c o s γ − γ ˙ sin γ = 1.27
0.36 − β ¨ s i n β − β ˙ cos β − 0.42 − γ ¨ s i n γ − γ ˙ cos γ = 0.27
By solving Equations (9) and (10) using the MATLAB/Simulink R2025b software package, we obtain the following results (see Figure 4) for the angles:
Figure 4. Diagram which shows the variation in the angle between the femur and the vertical Y-axis β and the angle between the shinbone and the vertical Y-axis γ .
β = 40°, and γ ranges from 0° to 5° and from 5° to (–60°).
The diagrams (see Figure 5) below show the results calculated for six (6) patient cases using MATLAB/Simulink R2025b:
Figure 5. Diagrams of hip contact forces generated for six patient cases during normal walking over an 8 s half-period.
  • 65/168 = (112.5 [%BW] ÷ 125   % BW ) = 12.5 [%BW] difference minimum/maximum
  • 75/175 = (132 [%BW] ÷ 148   % BW ) = 16 [%BW] difference minimum/maximum
  • 82/177 = (146,5 [%BW] ÷ 164   % BW ) = 17.5 [%BW] difference minimum/maximum
  • 90/180 = (165 [%BW] ÷ 185   % BW ) = 20 [%BW] difference minimum/maximum
  • 100/185 = (197.5 [%BW] ÷ 222.5   % BW ) = 25 [%BW] difference minimum/maximum
  • 120/195 = (245 [%BW] ÷ 285   % BW ) = 40 [%BW] difference minimum/maximum
In 10 ÷ 30% of cases during osteosynthesis, non-union of femoral pertrochanteric fractures occurs due to an inadequate calculation of the implant (screw) tightening force F f i x . Therefore, in this paper, the tightening force F f i x was calculated based on the rotation of the femoral head and the displacement (slip) of the fracture fragment surfaces, using actual values obtained from the calculations described above.
Below, we present the formulas for calculating the force balance that governs the potential rotation and sliding of the bone fragments. The goal is to determine the parameters for osteosynthesis in which the active force of the implant counteracts and balances the effects of the functional load on the limb. Rotation of the bone fragments is prevented if the rotational moment ( M r o t   is less than the stabilizing moment M s t z b .
M r o t   ≤ M s t z b
The moment is the product of the force and the distance from its point of application, and to prevent rotation, the following condition must be met:
F f · L f ≤ F f i x · L f i x
This equation provides the stability criterion for osteosynthesis of femoral pertrochanteric fractures according to the CMN principle.
The rotation of the bone fragments under load occurs about their lower contact point, and accordingly, the torques are defined relative to this point. The lower contact point of the bone fragments, with respect to which the moments of the functional load and the active force of the implant were calculated, is shown in Figure 6.
Figure 6. Lower contact point (L) of the bone fragments, relative to which the moments of the functional load and the active force of the implant are determined.
As a result, considering the fracture characteristics and force parameters, the formula for calculating rotation is as follows (Figure 7b):
F f · O A ¯ · cos β + 8 ° ≤ F f i x · C O ¯ · sin γ
Figure 7. (a) Balance of forces F f and F f i x , (b) Radiometric indicators.
The relative sliding of the bone fragments is determined by the relationship between the applied tangential loads and the resisting forces. We assume that the resistance force is proportional to the normal loads. Since the shear components are the tangential components of the load from body weight and the implant force, the system remains in stable equilibrium when:
F f t a n g < F f i x t + F f i x n
The formula, derived using vector analysis, is as follows:
F f · sin α + 8 ° < N · k r − F f i x · cos γ
N—the normal reaction force, which is the sum of the normal components of the functional load and the implant in the fracture plane.
By substituting N into the general formula, we obtained the final expression for calculating displacement across the width.
F f · sin α − 8 ° ≤ k r · F f · cos α − 8 ° + F f i x · sin γ − F f i x · cos γ
where F f is the magnitude of the load on the femoral head, F f i x is the magnitude of the force in the implant, O A ¯ , C O ¯ , α ,   β and γ are radiometric indicators (Figure 7b), and k r is the coefficient of resistance to bone fragment displacement (introduced by analogy with the coefficient of friction). The coefficient k r = 0.85 was determined experimentally, as described in detail in (Shugol et al. [16]).
Figure 7a illustrates that the resultant force, formed by the sum of the implant (screw) force and the functional force, should pass through the fracture plane. This is a critical condition for successful osteosynthesis.
Changes in the geometric parameters also determine the implant’s position relative to the fracture plane on the femur, which is a crucial factor for the success of osteosynthesis (Figure 8).
Figure 8. Variation in geometric parameters and the position of the implant tightening force F f i x depending on the location of the femoral fracture plane: (a) fracture location 1 and (b) fracture location 2.
The following geometric parameters (Figure 9) were determined:
Figure 9. The geometrical parameters.
  • α —the inclination angle of the fracture plane relative to the horizontal plane, is between 40° and 90°;
  • γ —angle between the direction of the fixator and the fracture plane, is between 150° and 30°;
  • O A ¯ —the distance from the base of the femoral head to the fracture plane was determined, is between 0 and 34 mm;
  • O C ¯ —the distance from the point where the implant intersects the fracture plane to the lowest point of contact between the bone fragments reaches 50 mm;
  • β —the angle of inclination of the Adams arch relative to the horizontal plane is 38°.
If α = 60 ° ; β = 38 ° ; γ = 120 ° ; O A ¯ = 10   mm ; O C ¯ = 37   mm ; and k r = 0.85 , then the implant tightening force will be F f i x = 350   N and the functional (active) force acting on the hip will be F f = 1665   N (185 [%BW] × 900 [N] = 1665 [N]) for a patient weighing 90 kg and 180 cm tall, expressed in terms of displacement along the width of the fracture plane. For rotation about point O, the resulting fixation force F f i x = 362   N   serves as a key parameter in assessing the stability of osteosynthesis.
The described compression fixation method corresponds to the concept of Stabilized Compression Fixation (SCF). SCF prevents the rotation of bone fragments, and its fixation mechanism relies on an optimized force that represents a combination of the minimum and maximum effective forces acting on the implant, which vary over an 8 s period. The application of SCF overcomes the fundamental limitations of existing fixation techniques. Future research will focus on further exploring this optimization approach to determine the optimal force applied by the implant (screw).
Dynamic axial compression between bone fragments under functional loading will be ensured, as illustrated in Figure 10. The figure clearly demonstrates that the initial optimization condition is defined by maintaining a compression force between the fragments that balances the minimum required stability under the specific functional load.
Figure 10. Osteosynthesis scheme based on the principle of fixation through tension.
The objective of mathematical modeling is to identify the conditions under which the active force generated by the implant counteracts the potential slippage of the fracture components resulting from the extensibility of the functional load.
For the purpose of virtual testing using finite element analysis (FEA), the upper portion of the femur, including a pertrochanteric fracture, was modeled. The femur model comprises an inner layer of cancellous bone enclosed by a cortical bone shell, with a thickness varying between 4 and 7 mm. In the finite element model, both cancellous and cortical bone were assumed to behave as linear, elastic, and homogeneous materials.
In healthy young adults (aged 20–40 years), the elastic (Young’s) modulus of cortical bone typically ranges from 18 to 30 GPa, as reported in the study of Beisekenov et al. [17]. With aging (approximately beyond 65–70 years), these values often decrease to about 10–25 GPa due to reductions in mineral density and microstructural changes. Factors such as age, biological condition (e.g., osteoporosis), sex, and specimen location significantly influence variations in both the elastic modulus and Poisson’s ratio. The elastic (Young’s) modulus of cancellous bone in healthy young adults (aged 20–40 years) typically ranges from 1.5 to 3 GPa, whereas in older adults (aged 60–70 years) these values often decrease to 0.8–1.2 GPa (Sansalone et al. [18]). The elastic modulus of cancellous bone has a negligible effect on stress distribution in the finite element analysis, whereas the elastic modulus of cortical bone has a significant impact.
The elastic (Young’s) modulus of cancellous bone typically ranges from 0.8 to 3 GPa; in this study, a value of 3 GPa was used with a Poisson’s ratio of 0.3. For cortical bone, an average elastic modulus of 20 GPa was adopted, also with a Poisson’s ratio of 0.3. The cortical layer forms a dense outer shell and represents the strongest part of the bone, consisting of approximately 95% mineral content.
Titanium alloy (Ti-6Al-4V) was selected for the implant, as it is a biocompatible material widely used in orthopedic applications due to its excellent corrosion resistance, non-toxicity, and favorable mechanical properties. In addition, its non-magnetic nature makes it particularly suitable for medical use. In the finite element model, the titanium alloy was also considered linear, elastic, and homogeneous. The corresponding material properties are summarized in Table 1.
Table 1. Material properties of osteosynthesis implant.
A reverse engineering approach was employed to generate a 3D model of the femur bone from computed tomography (CT) data. CT scanning was performed using a Siemens spiral CT scanner (Siemens, Berlin, Germany) with a slice thickness of 2 mm. A three-dimensional solid model of the femur bone was created using specialized medical image processing software. The SolidWorks 2024 software package was used to create solid models, perform finite element analysis, and conduct virtual testing. In SolidWorks Simulation, each part and assembly was discretized using solid tetrahedral elements. Mesh convergence was performed automatically using the adaptive mesh refinement function. The complete assembly consisted of 3,751,815 nodes and 2,575,066 elements, while the implant model contained 257,592 nodes and 133,501 elements.
The internal osteosynthesis implant (Figure 11), which features an intramedullary trapezoidal nail with integrated interlocking through two screws, consists of the following primary components: the trapezoidal nail, a two-screw integrated interlocking system, and fixation screws. The implant geometry was created as a modified conceptual model inspired by publicly available design features of intramedullary nail systems, including those produced by Smith & Nephew. The model does not represent any specific commercial product and was used exclusively for numerical simulation.
Figure 11. Internal osteosynthesis implant, intramedullary trapezoidal nail with integrated interlocking through two screws: (a) 3D model, (b) 3D mesh model.
The implant system shown in Figure 11 is specifically designed as a trochanteric portal intramedullary nail, tailored for fractures of the proximal femur. It offers anatomically shaped trapezoidal implants and includes an integrated interlocking screw option to enhance stability and resistance to femoral head rotation during both the intraoperative and postoperative phases. Additionally, the integrated interlocking screw configuration provides rotational stability to the femoral head and neck segment, increasing resistance to cutout.
The complete configuration of the internal fixator 3D model, along with its corresponding finite element analysis (FEA) mesh, is shown in Figure 11. This implant is constructed from titanium alloy (Ti-6Al-4V). In the finite element analysis, the fixation screw is represented in a simplified form.
The model contains multiple components in contact with one another. Table 2 presents the coefficient of friction values corresponding to various contact pairs.
Table 2. Coefficient of friction values for various contact pair types (Mercan et al. [19], Mircheski et al. [20], Ma et al. [21]).
In the SOLIDWORKS Simulation model, contact interactions between cancellous and cortical bone, cancellous–cancellous bone, and cortical–cortical bone at the fracture plane were defined as No Penetration contacts with a friction coefficient of 0.5. The fracture plane itself was modeled as a No Penetration contact with sliding disabled. Contact between bone and implant was also defined as No Penetration, with a friction coefficient of 0.3, while interactions between implant components were modeled as No Penetration contacts with a friction coefficient of 0.15.
Based on the dynamic model system (DMKS), schematically illustrated in Figure 1, and the concept of Stabilized Compression Fixation (SCF), shown in Figure 6, Figure 7, Figure 8, Figure 9 and Figure 10, developed in the previous section and described in detail through mathematical modeling, the corresponding equations, and numerical simulations implemented in MATLAB/Simulink, the following results were obtained: (a) the maximum active force acting on the hip ( F f = 1665   N ) and (b) the axial tightening force of the compression screw ( F f i x = 362   N ).
The initial and boundary conditions (loading scenarios) used in the finite element analysis were defined based on the results obtained from the dynamic model system (DMKS) and the concept of stabilized compression fixation (SCF). In the first stage, a preliminary axial tightening force ( F f i x = 362   N ) was applied to the compression screw (Figure 12). In the second stage, this preload was maintained, and an additional vertical load was applied to the system. The maximum value of force acting on the femur is 1665 N (Figure 12). During load application, the cylindrical end of the femur was rigidly fixed, and a uniformly distributed load was applied to the spherical surface of the femoral head (Figure 12) (Morgan et al. [22]).
Figure 12. Initial and boundary conditions for finite element analysis.
The mathematical model implemented as a simulation diagram in MATLAB/Simulink R2025b calculates the maximum force acting on the femur for patients with different anthropometric characteristics (65/168, 75/175, 82/177, 90/180, 100/185, and 120/195). For each anthropometric case, the femur has different geometric dimensions, and implant sizes are selected accordingly to match the patient’s anatomy. Due to the geometric and structural complexity of both the femur and the implant, only one representative case is analyzed using the finite element method. Specifically, the case of a patient with a body weight of 90 kg and a height of 180 cm is selected. The resulting stress and deformation distributions obtained from the finite element analysis are presented below.

3. Results

Three-dimensional finite element analysis (FEA) was employed to evaluate the stress and deformation fields. Figure 13 presents the von Mises stress distribution and deformation field in the cancellous bone, cortical bone of the femur, and the osteosynthesis implants.
Figure 13. 3D Finite element analysis of the assembled implant: (a) stress distribution, (b) cross-section and (c) deformation.
The maximum stress obtained in the assembly implant is 425.7 MPa, while the maximum deformation reaches 9.7 mm. The femoral assembly with the internal osteosynthesis implant (Figure 13), incorporating an intramedullary trapezoidal nail, demonstrates favorable results and satisfactory static performance.
The stress and deformation fields of the osteosynthesis implant with an intramedullary trapezoidal nail are presented in Figure 14. In this configuration, the maximum stress reaches 425.7 MPa, while the maximum deformation is 9.265 mm. The safety factor for the internal implant with the trapezoidal nail is 1.94, indicating a high level of structural reliability and enhanced patient safety. The implant safety factor of 1.94 was calculated as the ratio of the tensile strength of the Ti-6Al-4V titanium alloy (827 MPa) to the maximum implant stress obtained from finite element analysis (425.7 MPa). The safety factor reported here refers to loads on the dynamic system during normal walking, without additional loading. A safety factor under these conditions indicates that the implant is likely to withstand forces associated with running, jumping, or other high-impact activities. However, further studies are needed to evaluate the safety factor under additional loads and to more accurately assess the implant’s overall durability.
Figure 14. Three-dimensional FEA shows stress distribution in the internal osteosynthesis implant.
Table 3 presents a comparison of the finite element results obtained in the present study with those reported in previous studies (Kim et al. [7], Ma et al. [22]). For similar geometric configurations, friction coefficients, implant material, implant type, and loading conditions, the numerical results are generally consistent. The maximum implant stresses reported in the study of Kim et al. [7] and study of Ma et al. [22] were 455.2 MPa and 373.2 MPa, respectively, whereas the maximum implant stress obtained in the present analysis was 425.7 MPa. In addition, the safety factor in the present study was 1.94, compared with 2.21 and 1.81 in the study of Kim et al. [7] and study of Ma et al., respectively. The comparative analysis of safety factors indicates that the proposed theoretical methodology, combining mathematical modeling, MATLAB Simulink, and finite element analysis, offers a useful framework for assessing fracture fixation stability.
Table 3. Comparison of the results of this study with those reported in studies (Kim et al. [7], Ma et al. [22]).
To describe the healing process, a simulation-based model was developed in MATLAB Simulink R2025b. Figure 15 illustrates the overall structure of the model, in which the input parameters include the patient’s weight, height, and age. The model outputs comprise the time-dependent evolution of gait force, stress, and micromotion associated with the pertrochanteric fracture. Specifically, the model comprises the following modules: patient parameters (including the osteoporosis factor), muscle strength model, fracture parameters, femoral dynamic parameters, implant parameters (CMN), micromovement, healing progress, healing model, implant performance, fatigue limit, healing index, risk assessment, and limit conditions. The model demonstrates strong potential for supporting clinical decision-making, optimizing implant design, and enabling advanced applications in orthopedic biomechanics.
Figure 15. The healing process of pertrochanteric fracture.

4. Experimental Validation of the Implant

The implant used for the treatment of pertrochanteric fractures was tested using a custom-designed experimental device developed specifically for this study (see Figure 16). The test was conducted on a sample representing a patient with a body weight of 90 kg. The device consisted of a wooden component in the form of an angled block and several steel components. The implant stem was fixed to the angled block in a position that ensured a vertical orientation of the sliding screws, with their tips directed upward. The steel assembly included two vertical bars secured to the wooden component and additionally connected by horizontal bars passing through a supporting clamp. An adjustable clamp was mounted on the vertical bars, through which a threaded rod was inserted. An anti-rotation lever was attached to one end of the threaded rod, while a tension lever was screwed onto the threaded section. The opposite end of the threaded rod was connected to a portable electronic dynamometer (Portable Electronic Scale, Ansenym, Guangzhou, China). The dynamometer was linked to the sliding screws by a cable.
Figure 16. Model of the testing device used in the study. (1) Anti-rotation lever; (2) threaded rod; (3) tension lever; (4) adjustable clamp; (5) vertical bar; (6) supporting clamp; (7) horizontal bar; (8a) portable electronic dynamometer; (8b) direction of the force measured by portable electronic dynamometer; (9a) integrated electronic dynamometer; (9b) direction of the force measured by integrated electronic dynamometer; (10) cap; (11) implant stem; (12) sliding screw; and (13) angled block.
Figure 17 shows the relationship between the applied load and the experimentally obtained elongation using a custom-designed experimental device.
Figure 17. Model loading curve.
The conclusions of the testing are as follows:
  • At loads up to 1600 N, implant migration does not exceed 2 mm.
  • Implant migration without an increase in load occurs at loads above 1665 N.
  • No model failure or implant deformation was detected.
Mathematical modeling demonstrates that osteosynthesis based on the CMN method requires an implant force of up to 362 N to achieve stable fixation of the proximal femoral head. However, the strength of the tissue in the tested specimen was nearly twice the minimum strength reported in the elderly population (390 N/cm2 vs. 181 N/cm2, calculated as 362 N/2 cm2, Sansalone et al. [18]). The fixation surface area of the central end of the implant within the femoral head reaches 2 cm2.
The absence of implant failure under a load of 1665 N indicates a safety margin of more than 4.6 times (1665 N/362 N = 4.6).
The experiment was conducted to eliminate potential bias in the data arising from differences in the strength of the bone specimens. The load was applied only until rotation was detected, rather than until displacement of the model fragments across their width occurred, which would have partially fractured them. The experimental results were in full agreement with the calculations.
Simulation studies were also conducted in MATLAB R2025b to analyze the in-hospital healing process of osteosynthesis by monitoring fatigue behavior through the stress state over a 30-day period (see Figure 18). A progressive reduction in stress is observed throughout the 30 days.
Figure 18. Stress evolution during healing.

5. Discussion

This study investigates an intramedullary osteosynthesis approach based on a trapezoidal wedge with integrated dual interlocking screws (InterTAN), which demonstrated satisfactory fixation stability. The results provide biomechanical justification for applying the cephalomedullary nail (CMN) principle in femoral pertrochanteric fractures, supported by biomechanical calculations, mathematical modeling, virtual testing, and experimental validation for the evaluation of controlled fixation and implant performance.
The proposed mechatronic methodology incorporates mathematical and simulation modeling of an active fixation system, in which the interlocking screws are subjected to controlled tension to enhance overall construct stability. Active tensioned fixation may offer a promising strategy for improving osteosynthesis techniques in this complex clinical condition. The mathematical and biomechanical model provides a robust theoretical foundation for implant design and surgical planning. Its clinical applicability is further enhanced by the potential for integration into surgical software platforms, as illustrated in Figure 15.
A virtual biomechanical analysis was performed to investigate the behavior of the active fixation system using an intramedullary trapezoidal wedge with integrated dual interlocking screws. Multiple simulations were conducted to evaluate implant performance under tension-controlled fixation by analyzing stress distribution within the construct. Based on mathematical modeling, MATLAB Simulink R2025b calculations, and finite element analysis, the observed stress distributions and deformation patterns support the hypothesis that the proposed mechatronics methodology can effectively assess mechanical stability under loading conditions. These findings were further confirmed by experimental testing.
One of the main goals of this study is to determine the tightening force required to achieve stable fixation, which directly relates to primary post-operative stability. While the implant–bone interface is modeled using “No Penetration” contacts with friction, the applied interlocking screw tension effectively simulates the stabilizing effect on micromotion and press-fit conditions. This approach allows assessment of initial construct stability, which is critical for successful osseointegration. Incorporating these considerations highlights the relevance of the methodology and provides guidance for optimizing fixation strategies in pertrochanteric fracture.
To evaluate the computational efficiency of the proposed methodology, the actual computation times were recorded for both the DMKS and FEA simulations. The DMKS simulations, covering six different anthropometric cases, required approximately 8–10 min on a standard desktop workstation, while the FEA simulation for the representative patient model (90 kg, 180 cm) required approximately 380 min. These results demonstrate that the integrated computational framework enables quantitative assessment of osteosynthesis stability while remaining feasible for practical research applications.
The analysis was based on mathematical modeling, finite element simulations, and experimental verification with measurement equipment and included biological factors and fracture healing. It remains to be confirmed under conditions of load associated with daily activities.

6. Conclusions

This study presents a mechatronic approach to investigating osteosynthesis in pertrochanteric fractures and evaluates its biomechanical behavior through mathematical modeling, MATLAB/Simulink R2025b simulations, finite element analysis, and experimental validation. The primary objective of this study was to investigate the biomechanical and functional characteristics of pertrochanteric fractures treated using the CMN method with a mechatronic approach, and to validate the proposed methodology through experimental testing. The proposed methodology enables a quantitative assessment of fixation stability based on key mechanical indicators, including functional hip force, fracture fragment displacement, stress distribution in the implant and surrounding bone, and the safety factor of the fixation construct. To achieve this objective, a dynamic model of the knee system (DMKS) was developed to analyze the kinematics of the pelvis, femur and shinbone, which determine the loading conditions of the lower-limb joints, including the hip joint. Using mathematical modeling and MATLAB/Simulink R2025b simulations, the functional hip force (Ff) was determined as a time-dependent parameter influenced by patient body weight and height. The kinematic behavior of the femur and shinbone was analyzed, with the angles β and γ expressed as functions of time and patient anthropometric characteristics. The results indicate that the functional force acting on the hip varies throughout the gait cycle rather than remaining constant, as assumed in simplified analytical models. The developed model further enabled evaluation of the relationship between functional loading, implant tightening force ( F f i x ), femoral head rotation, and fracture fragment displacement, providing a biomechanical basis for assessing controlled fixation conditions. The analysis was conducted for a representative patient model (90 kg, 180 cm), providing a biomechanical basis for the assessment of controlled fixation.
Three-dimensional finite element analysis enabled a detailed quantitative evaluation of stress–strain distributions, deformation patterns, and safety factors for implants with complex geometries and configurations.
Experimental validation was performed to confirm the controlled fixation methodology and to verify the reliability of the computational predictions and implant performance under controlled loading conditions.
The integrated computational framework provides a robust and quantitative methodology for the biomechanical assessment and preliminary design evaluation of novel orthopedic implant concepts.

Author Contributions

Conceptualization, I.M. (Igor Merdzanoski) and M.M.; methodology, I.M. (Igor Merdzanoski) and I.M. (Ivan Mickoski); software, I.M. (Ile Mircheski).; validation, I.M. (Igor Merdzanoski), M.M. and M.S.; formal analysis, I.M. (Igor Merdzanoski), I.M. (Ivan Mickoski), and I.M. (Ile Mircheski); investigation, I.M. (Igor Merdzanoski); resources, I.M. (Igor Merdzanoski) and M.S.; writing—original draft preparation, I.M. (Igor Merdzanoski), I.M. (Ivan Mickoski) and I.M. (Ile Mircheski).; writing—review and editing, M.M.; visualization, I.M. (Ile Mircheski); supervision, I.M. (Ivan Mickoski); All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Data Availability Statement

The datasets generated and/or analyzed during the current study are available from the corresponding author upon request.

Conflicts of Interest

The authors declare no conflicts of interest.

Abbreviations

The following abbreviations are used in this manuscript:
FEAFinite element analysis
%BWPercentage of body weight
CMNCephalomedullary Nail
DMKSDynamic model of the knee system
SCFStabilized Compression Fixation

References

  1. Abrahamsen, B.; van Staa, T.; Ariely, R.; Olson, M.; Cooper, C. Excess mortality following hip fracture: A systematic epidemiological review. Osteoporos. Int. 2009, 20, 1633–1650. [Google Scholar] [CrossRef] [Scilit]
  2. Imwinkelried, T. Mechanical properties of open-pore titanium foam. J. Biomed. Mater. Res. Part A 2007, 81, 964–970. [Google Scholar] [CrossRef] [Scilit]
  3. Lee, C.; Kelley, B.; Gurbani, A.; Stavrakis, A.-I. Strategies for pertrochanteric fracture reduction and intramedullary nail placement: Technical tips and tricks. J. Am. Acad. Orthop. Surg. 2022, 30, 867–878. [Google Scholar] [CrossRef] [Scilit]
  4. Wu, H.-F.; Chang, C.-H.; Wang, G.-J.; Lai, K.-A.; Chen, C.-H. Biomechanical investigation of dynamic hip screw and wire fixation on an unstable intertrochanteric fracture. Biomed. Eng. Online 2019, 18, 49. [Google Scholar] [CrossRef] [Scilit]
  5. Lu, Y.; Uppal, H.S. Hip Fractures: Relevant Anatomy, Classification, and Biomechanics of Fracture and Fixation. Geriatr. Orthop. Surg. Rehabil. 2019, 10, 1–10. [Google Scholar] [CrossRef] [Scilit]
  6. Li, J.; Zhao, Z.; Yin, P.; Zhang, L.; Tang, P. Comparison of three different internal fixation implants in treatment of femoral neck fracture-a finite element analysis. J. Orthop. Surg. Res. 2019, 14, 76. [Google Scholar] [CrossRef] [Scilit]
  7. Kim, C.-J.; Lee, J.-S.; Goh, T.-S.; Shin, W.-C.; Lee, C. Finite element analysis of fixation stability according to reduction position for internal fixation of intertrochanteric fractures. Sci. Rep. 2024, 14, 19214. [Google Scholar] [CrossRef] [Scilit]
  8. Li, J.; Han, L.; Zhang, H.; Zhao, Z.; Su, X.; Zhou, J.; Li, C.; Yin, P.; Hao, M.; Wang, K.; et al. Medial sustainable nail versus proximal femoral nail antirotation in treating AO/OTA 31-A2.3 fractures: Finite element analysis and biomechanical evaluation. Injury 2019, 50, 648–656. [Google Scholar] [CrossRef] [Scilit]
  9. Bartoska, R.; Baca, V.; Horak, Z.; Hrubina, M.; Skala-Rosenbaum, J.; Marvan, J.; Kachlik, D.; Dzupa, V. The importance of intramedullary hip nail positioning during implantation for stable pertrochanteric fractures: Biomechanical analysis. Surg. Radiol. Anat. 2016, 38, 577–585. [Google Scholar] [CrossRef] [Scilit]
  10. Li, M.; Zhao, K.; Ding, K.; Cui, Y.-W.; Cheng, X.-D.; Yang, W.-J.; Hou, Z.-Y.; Zhang, Y.-Z.; Chen, W.; Hu, P.; et al. Titanium Alloy Gamma Nail versus Biodegradable Magnesium Alloy Bionic Gamma Nail for Treating Intertrochanteric Fractures: A Finite Element Analysis. Orthop. Surg. 2021, 13, 1513–1520. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  11. Bai, H.; Liu, L.; Duan, N.; Xue, H.; Sun, L.; Li, M.; Li, Z.; Zhang, K.; Wang, Q.; Huang, Q. Biomechanical evaluation of three implants for treating unstable femoral intertrochanteric fractures: Finite element analysis in axial, bending and torsion loads. Front. Bioeng. Biotechnol. 2023, 11, 1279067. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  12. Mahaisavariya, C.; Jitprapaikulsarn, S.; Mahaisavariya, B.; Chantarapanich, N. Torsional stability of fixation methods in basicervical femoral neck fractures: A biomechanical study. J. Orthop. Surg. Res. 2024, 19, 371. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  13. Zhang, G.; Li, J.; Li, J.; Zhou, X.; Xie, J.; Wang, A. Selective laser melting molding of individualized femur implant: Design, process, optimization. J. Bionic Eng. 2021, 18, 128–137. [Google Scholar] [CrossRef] [Scilit]
  14. Zhang, G.; Li, J.; Zhou, X.; Zhou, Y.; Xie, J.; Bai, Y. Optimization design of high-performance powder-spreading arm for metal 3D printers. Micromachines 2025, 16, 1194. [Google Scholar] [CrossRef] [Scilit]
  15. Hachimi, T.; Ait Hmazi, F.; Arhouni, F.E.; Zekriti, N.; Doghmi, H.; Majid, F. Damage evolution and fracture behavior in 3D-printed materials: Effects of crack propagation and progressive layer degradation. Prog. Addit. Manuf. 2026, 11, 2591–2610. [Google Scholar] [CrossRef] [Scilit]
  16. Shugol, G.-B.; Demakov, S.-L.; Shugol, I.-G. Osteosynthesis of Fractures of the Femoral Neck, Based on the Use of the Principle of Active Fixation by Contraction; UGMU: Ekaterinburg, Russia, 2014. [Google Scholar]
  17. Beisekenov, N.; Sadenova, M.; Azamatov, B.; Syrnev, B. Analysis of Biomechanical Characteristics of Bone Tissues Using a Bayesian Neural Network: A Narrative Review. J. Funct. Biomater. 2025, 16, 168. [Google Scholar] [CrossRef] [Scilit]
  18. Sansalone, V.; Naili, S.; Bousson, V.; Bergot, C.; Peyrin, F.; Zarka, J.; Laredo, J.D.; Haïat, G. Determination of the heterogeneous anisotropic elastic properties of human femoral bone: From nanoscopic to organ scale. J. Biomech. 2010, 43, 1857–1863. [Google Scholar] [CrossRef] [Scilit]
  19. Mercan, N.; Eravsar, E.; Safali, S.; Ugur, L.; Ozdemir, A. Biomechanical insights through finite element analysis of Bennett fracture fixation: A comparative study of four surgical techniques. J. Orthop. Surg. Res. 2025, 20, 90. [Google Scholar] [CrossRef] [Scilit]
  20. Mircheski, I.; Gradisar, M. 3D finite element analysis of porous Ti-based alloy prostheses. Comput. Methods Biomech. Biomed. Eng. 2016, 19, 1531–1540. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  21. Ma, X.; Chen, J.; Hu, Y.; Fan, Z.-R. Biomechanical analysis of anatomical intramedullary nails and proximal femoral anti-rotation-II intramedullary nails for the treatment of unstable intertrochanteric fractures in older patients. Res. Sq. 2025, preprint. [Google Scholar]
  22. Morgan, E.-F.; Unnikrisnan, G.-U.; Hussein, A.-I. Bone Mechanical Properties in Healthy and Diseased States. Annu. Rev. Biomed. Eng. 2018, 20, 119–143. [Google Scholar] [CrossRef] [Scilit] [PubMed]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.