Abstract
Background: Directional hemispherical reflectance (DHR) is a precise method for evaluating skin reflectance and is widely used in dermatological, photobiological, and cosmetic or pharmaceutical research. Reflectance measurements may support emissivity-related interpretation, particularly in the infrared range, being influenced by chromophore content, epidermal structure, and physiological factors such as hydration, pigmentation, and surface heterogeneity. However, most in vivo studies have focused on limited spectral ranges or selected anatomical sites. This study aimed to assess skin directional hemispherical reflectance across a broad spectral range and to provide an integrated dataset supporting emissivity-related interpretation in the infrared region. Methods: The study included 20 women aged 22–50 years (27 ± 9 years) with Fitzpatrick skin phototypes II–III. Reflectance measurements were performed at 14 anatomical sites using an ET 100 emissometer (1.9–21 µm) and an SOC 410 Solar DHR reflectometer (335–2500 nm). Infrared measurements were conducted at incidence angles of 20° and 60° to assess angular effects. Data were statistically analyzed. Results: The lowest reflectance values were observed within 335–380 nm, 1700–2500 nm, and 1.5–21 µm, whereas the highest reflectance was recorded in the 590–720 nm and 700–1100 nm bands. Reflectance symmetry between body sides was observed. In the infrared range, reflectance decreased with increasing wavelength, while mid- and far-infrared values were more uniform across locations. The highest reflectance values were noted for the thigh, calf (crural region), forearm, and palmar surface of the hand, whereas the lowest values were observed in the neck, abdominal region, and dorsal surface of the hand. Measurements at 60° incidence yielded higher reflectance values than those at 20°, while preserving spatial patterns. Conclusions: Directional hemispherical reflectance provides a robust approach for assessing skin reflectance across a broad spectral range. Reflectance depends on wavelength, anatomical location, and physiological factors, including epidermal thickness, pigmentation, and sebum presence. The integrated analysis of spectral, anatomical, and angular variability may support improved interpretation of skin optical properties and contribute to reference data for biomedical and infrared imaging applications.
1. Introduction
The growing interest in optical methods for the assessment of skin properties has contributed to the increasing importance of directional hemispherical reflectance (DHR) in biomedical sciences, cosmetology, and materials research. This method enables a precise characterization of how skin reflects radiation across a broad spectral range, including visible light, ultraviolet (UV) radiation, and infrared (IR) radiation. Consequently, DHR has become a valuable tool for evaluating skin properties, the effects of active substances, the efficacy of photoprotective products, and for supporting diagnostic approaches based on thermal radiation analysis [1,2].
Emissivity is a key physical parameter describing the radiative properties of skin and is defined as the ratio of radiation emitted by a surface to that emitted by an ideal blackbody at the same temperature. Human skin exhibits exceptionally high emissivity, typically ranging from 0.92 to 0.99, with a reference value of 0.95 commonly adopted in experimental studies [1,2]. Such high emissivity causes the skin to behave as a quasi-blackbody in the IR range, which underlies the application of thermography in medicine and cosmetology. Thermal imaging is widely employed in the assessment of inflammatory processes, vascular disorders, and injuries, as well as in aesthetic medicine, where it supports the evaluation of injection procedures and the monitoring of treatment outcomes, including those involving botulinum toxin [3,4,5]. Skin emissivity is influenced by multiple surface and biological factors, including the presence of sweat, sebum, or topical products. Moreover, clinical conditions such as wounds, scars, burns, and skin dryness may locally alter emissivity values [3,6,7,8,9,10,11]. This parameter plays a critical role in non-invasive temperature measurements, physiological investigations of metabolism and thermoregulation, and thermal imaging, where accuracy depends on appropriate emissivity assumptions [5,7,12,13,14,15]. Emissivity is intrinsically linked to the absorption and reflection of radiation [9].
According to Kirchhoff’s law of thermal radiation, for objects in thermal equilibrium, emissivity equals absorptivity. For opaque materials, absorptivity may be approximated as 1 − reflectance, which provides the theoretical basis for the indirect determination of emissivity from directional hemispherical reflectance (DHR) measurements. In the present study, this relationship was considered primarily for the infrared range, where human skin can be treated as approximately opaque. In contrast, in the ultraviolet, visible, and near-infrared ranges, skin reflectance is influenced not only by absorption but also by transmission and subsurface scattering. Therefore, measurements in these spectral bands were interpreted primarily as reflectance characteristics rather than direct estimates of emissivity [16,17,18,19].
The interaction of skin with electromagnetic radiation, including reflection, absorption, and transmission, is governed by the presence of key chromophores such as melanin, hemoglobin, and water [16,17]. Melanin, primarily located in the epidermis, exhibits a broad absorption spectrum spanning the UV, visible, and near-infrared (NIR) regions [20,21,22]. Hemoglobin, depending on its oxygenation state, demonstrates characteristic absorption peaks within the visible range. Water, in turn, strongly absorbs radiation in the IR region [20,23,24,25,26]. These chromophores determine both the spectral absorption profile of skin and its reflectance characteristics. Therefore, their optical behavior must be considered in skin optics analyses and in the interpretation of measurements obtained using DHR [27,28].
DHR describes the fraction of radiation incident from a specific direction that is reflected into the entire hemisphere above a surface. Unlike simple directional reflectance measurements, DHR accounts for both specular and diffuse reflection components, which is particularly relevant for biological tissues such as skin, a structurally heterogeneous, rough, and highly scattering surface. Human skin exhibits low DHR values in the infrared range, primarily due to high absorption and minimal reflection of thermal radiation. Consequently, the skin behaves predominantly as a heat emitter rather than a reflector, a property fundamental to thermal imaging techniques. DHR measurements enable the evaluation of skin hydration, structural characteristics, and physiological condition through the analysis of reflectance patterns, facilitating the identification of surface irregularities, dryness, and pathological alterations. This method is also applied in the assessment of cosmetic product performance, including formulations designed to modulate skin reflectance. DHR is of particular importance in photoprotection research, as mineral filters such as zinc oxide and titanium dioxide increase hemispherical reflectance in the ultraviolet range, thereby supporting the evaluation of photoprotective efficacy [29,30,31,32,33].
The reflective properties of skin are strongly influenced by its type and physiological condition. Healthy, well-hydrated skin typically exhibits more uniform light reflection, whereas dry, keratinized, or damaged surfaces promote increased scattering. Oily skin, in contrast, is associated with higher directional reflectance, contributing to the appearance of excessive surface shine. Understanding the relationship between skin structure and reflectance characteristics is essential for the precise selection of cosmetic products and skincare procedures, while also supporting the development of non-invasive dermatological diagnostic techniques. Collectively, these considerations highlight directional hemispherical reflectance as a key tool in the investigation of skin optical properties and emissivity. Its expanding application across cosmetology, dermatology, medicine, and imaging technologies further underscores its relevance in contemporary skin assessment methodologies [34,35,36,37,38]. However, previous in vivo studies have generally examined selected spectral ranges, specific interventions, or limited anatomical areas. Comprehensive datasets combining UV–Vis–NIR reflectance with mid- and far-infrared measurements across multiple anatomical sites remain scarce. In addition, most existing studies have focused on isolated determinants of skin reflectance, such as specific physiological factors or selected clinical interventions. For example, studies such as Banyś et al. 2023 [6] have demonstrated the role of sebum in modulating reflectance, particularly in the near-infrared range, while Odrzywołek et al. [10] have investigated treatment-induced changes in reflectance and skin structure in pathological conditions. However, these approaches do not capture the combined influence of spectral, anatomical, and geometrical factors within a unified experimental framework.
The novelty of the present study lies in the integrated in vivo assessment of directional hemispherical reflectance across a broad spectral range, multiple anatomical locations, and, in the infrared region, different angles of incidence. This comprehensive approach enables a more complete characterization of skin optical behavior and provides a reference framework for the interpretation of reflectance-based measurements in biomedical and cosmetic applications.
Accordingly, the aim of this study was to assess directional hemispherical reflectance of human skin across 14 anatomical regions using complementary reflectometric and emissometric measurements covering wavelengths from 335 nm to 21 µm.
2. Materials and Methods
The study included 20 women aged 22–50 years (mean age: 27 ± 9 years). All participants presented Fitzpatrick skin phototypes II–III. Inclusion criteria comprised age ≥ 18 years, female sex, and provision of informed consent. Exclusion criteria included the presence of dermatological lesions, tattoos at the measurement sites, and age < 18 years. Participation was voluntary and anonymous, and all participants provided written informed consent. Skin reflectance measurements were performed in each participant using an emissometer and a reflectometer. Fourteen measurement sites representing different anatomical regions were selected: right anterior neck region, left anterior neck region, right posterior neck region, right forearm, left thigh, left calf, left lower leg (crural region), abdominal region superior to the umbilicus, right hand (first web space, dorsal side), right hand (dorsal side beneath the middle finger), right hand (palmar side beneath the middle finger), right hand (fourth web space, dorsal side), forehead (frontal region), and left cheek (buccal region).
Skin reflectance in the UV, visible, and NIR ranges was assessed using a portable SOC 410 Solar DHR reflectometer (Surface Optics Corporation, San Diego, CA, USA). The device measures the total hemispherical reflectance across seven discrete spectral bands within the 335–2500 nm range: 335–380 nm, 400–540 nm, 480–600 nm, 590–720 nm, 700–1100 nm, 1000–1700 nm, and 1700–2500 nm. The device measured a circular area with a diameter of 12.7 mm. The reflectometer was calibrated prior to measurements using Diffuse Calibration Coupon and Specular Calibration Coupon standards, traceable to the National Institute of Standards and Technology (NIST).
Reflectance measurements in the mid- and far-infrared ranges were performed using an ET 100 emissometer (Surface Optics Corporation, San Diego, CA, USA). The instrument enables reflectance measurements at two angles of incidence: 20° and 60°. Measurements were conducted across six spectral bands spanning wavelengths from 1.5 µm to 21.0 µm: 1.5–2.0 µm, 2.0–3.5 µm, 3.0–4.0 µm, 4.0–5.0 µm, 5.0–10.5 µm, and 10.5–21.0 µm. The measurement aperture of the emissometer was also 12.7 mm in diameter. Prior to measurements, the emissometer was calibrated using a Gold Calibration Coupon standard, compliant with NIST requirements. Both devices are equipped with integrating spheres designed to collect reflected radiation from the skin surface and integrate the signal over the hemispherical space above the sample. Detectors located within the spheres register the reflected radiation within specific spectral bands, which is subsequently converted into electrical signals.
The SOC 410 Solar DHR reflectometer provides measurements at a fixed incidence angle of 20°, which was therefore used for the UV–Vis–NIR range. The ET 100 emissometer enables measurements at two predefined incidence angles (20° and 60°); both were included to examine whether the angular configuration affects infrared reflectance values and spatial reflectance patterns across anatomical sites. Measurements at normal incidence were not performed because this geometry was not available in the predefined optical configurations of the instruments used. Therefore, the selected angles reflected the manufacturer-defined measurement geometries rather than arbitrary experimental choices. Emissivity-related interpretation was restricted to the infrared range, in which skin may be considered approximately opaque for practical measurement purposes. For the UV–Vis–NIR range, the study focused on directional hemispherical reflectance itself, without applying a direct ε = 1 − R conversion, due to the contribution of subsurface scattering and partial transmission at shorter wavelengths. Therefore, reflectance data in these spectral regions were not directly interpreted as emissivity values. During all measurements, the probe was placed in direct contact with the skin and positioned as consistently as possible relative to the local skin surface. Measurements were obtained from anatomically accessible areas that allowed stable probe placement. For curved surfaces, such as the forearm or calf, care was taken to place the probe on the flattest available local area. Given the fixed measurement aperture (12.7 mm), the recorded signal corresponded to a defined local skin area; however, slight variations in the effective local surface normal and illumination geometry may occur and should be considered a limitation of in vivo measurements.
All measurements were performed under controlled environmental conditions (ambient temperature: 21 °C; relative humidity: 45%), following a minimum 15 min acclimatization period. Measurements were initiated by placing the measurement head in direct contact with the skin surface and activating the acquisition trigger. Prior to signal acquisition, the cleanliness and condition of the skin at each measurement site were visually assessed. For facial regions (forehead and cheek), makeup removal was performed, followed by gentle drying of the skin. No solvent-based degreasing protocol was applied. Three repeated measurements were obtained at each anatomical site to improve measurement reliability and data consistency.
Statistical Analysis
The obtained data were entered into an Excel 2016 spreadsheet (Microsoft Corporation, Redmond, WA, USA) and verified for completeness. Statistical analyses were performed using Statistica 13 software (TIBCO Software Inc., Palo Alto, CA, USA). The normality of data distribution was assessed using the Shapiro–Wilk test. As normal distribution was not observed for all variables, non-parametric tests were applied. Quantitative data are presented as medians, minimum and maximum values, and quartiles. Comparisons of quantitative variables across multiple anatomical locations were conducted using the Friedman’s ANOVA test. Statistical significance was set at p < 0.05. In addition to p values, effect size estimates for Friedman’s ANOVA were calculated using Kendall’s coefficient of concordance (W) to quantify the strength of differences across anatomical sites. Because the primary aim of the study was to assess overall regional variability across the spectrum, the main analysis focused on global comparisons rather than pairwise post hoc testing.
3. Results
3.1. Directional Hemispherical Reflectance of Skin Across Different Anatomical Regions Within the Spectral Bands of 335–380 nm, 400–540 nm, 480–600 nm, 590–720 nm, 700–1100 nm, 1000–1700 nm, and 1700–2500 nm at an Incidence Angle of 20°, Measured Using the SOC 410 Solar DHR Reflectometer
Skin reflectance exhibited the lowest values within the 335–380 nm and 1700–2500 nm spectral bands (Table 1). The highest reflectance values were observed in the 590–720 nm and 700–1100 nm ranges. Statistical analysis demonstrated significant differences in reflectance between anatomical regions across all investigated wavelength bands (p < 0.001) (Table 1 and Table 2). Within the ultraviolet range (335–380 nm), the highest reflectance values were recorded for the right forearm, forehead (frontal region), left cheek (buccal region), and the palmar surface of the hand beneath the middle finger. The lowest reflectance values were observed for the dorsal surface of the hand beneath the middle finger and the fourth web space of the hand. In the 400–540 nm band, the highest reflectance values were found for the right forearm, the abdominal region superior to the umbilicus, and the first web space of the hand. The lowest reflectance values were observed for the palmar surface of the hand beneath the middle finger and the posterior neck region. Within the 480–600 nm range, the highest reflectance values were recorded for the right forearm and the abdominal region superior to the umbilicus, whereas the lowest values were observed for the dorsal surface of the hand beneath the middle finger and the posterior neck region. In the 590–720 nm band, the highest reflectance values were observed for the palmar surface of the hand beneath the middle finger and the abdominal region superior to the umbilicus. The lowest reflectance values occurred in the posterior neck region and the fourth web space of the hand. Within the 700–1100 nm spectral range, the highest reflectance values were found for the abdominal region superior to the umbilicus and the left thigh. The lowest reflectance values were recorded for the anterior neck region, the posterior neck region, and the fourth web space of the hand. In the 1000–1700 nm band, the highest reflectance values were observed for the left thigh, right forearm, and left calf. The lowest reflectance values were noted for the dorsal and palmar surfaces of the hand beneath the middle finger. Within the 1700–2500 nm spectral band, the highest reflectance values were recorded for the left lower leg (crural region), left calf, and left thigh, whereas the lowest values were consistently observed for the dorsal and palmar surfaces of the hand beneath the middle finger. Across the entire investigated spectral range, the highest reflectance values were generally observed for the forearm and abdominal regions, while the lowest reflectance values were associated with the dorsal hand surface beneath the middle finger and the posterior neck region (Table 1 and Table 2).
Table 1.
Median skin directional hemispherical reflectance values measured at an incidence angle of 20° across the following spectral bands: 335–380 nm, 400–540 nm, 480–600 nm, 590–720 nm, 700–1100 nm, 1000–1700 nm, and 1700–2500 nm. Measurements were obtained at the following anatomical sites: right anterior neck region, left anterior neck region, posterior neck region, right forearm, left thigh, left calf, left lower leg (crural region), abdominal region superior to the umbilicus, right hand (first web space, dorsal side), right hand (dorsal side beneath the middle finger), right hand (palmar side beneath the middle finger), right hand (fourth web space, dorsal side), forehead (frontal region), and left cheek (buccal region). Statistical significance: *** p < 0.001. Color intensity reflects reflectance magnitude, with red indicating the highest values.
Table 2.
Percentage comparison of median directional hemispherical reflectance values across anatomical sites, expressed relative to the lowest median reflectance value observed at each spectral band (335–380 nm, 400–540 nm, 480–600 nm, 590–720 nm, 700–1100 nm, 1000–1700 nm, and 1700–2500 nm). Measurements were obtained at the following anatomical sites: right anterior neck region, left anterior neck region, posterior neck region, right forearm, left thigh, left calf, left lower leg (crural region), abdominal region superior to the umbilicus, right hand (first web space, dorsal side), right hand (dorsal side beneath the middle finger), right hand (palmar side beneath the middle finger), right hand (fourth web space, dorsal side), forehead (frontal region), and left cheek (buccal region). Color intensity reflects reflectance magnitude, with red indicating the highest values.
3.2. Directional Hemispherical Reflectance Measured at an Incidence Angle of 20° Across the Infrared Spectral Bands of 1.5–2.0 µm, 2.0–3.5 µm, 3.0–4.0 µm, 4.0–5.0 µm, 5.0–10.5 µm, and 10.5–21.0 µm, Obtained Using the ET 100 Emissometer
Statistical analysis demonstrated significant differences in directional hemispherical reflectance between anatomical regions across all investigated infrared spectral bands (p < 0.001) (Table 3 and Table 4). At an incidence angle of 20°, emissometric measurements revealed that reflectance values within the 1.5–2.0 µm band were consistently the highest across all anatomical sites, with median values ranging from 0.062 to 0.093 (Table 3). The highest reflectance values were observed for the left thigh, left calf, left lower leg (crural region), and right forearm, whereas the lowest values were recorded for the dorsal and palmar surfaces of the hand beneath the middle finger. Within the 2.0–3.5 µm spectral band, reflectance values ranged from 0.041 to 0.059. Higher reflectance values were again observed for the left thigh, left calf, and left lower leg (crural region), while lower values were associated with the dorsal and palmar surfaces of the hand beneath the middle finger, the fourth web space of the hand, and the posterior neck region. Across the spectral bands from 3.0–4.0 µm to 10.5–21.0 µm, median reflectance values were comparable, ranging from 0.030 to 0.045. Despite the similarity in absolute reflectance values, statistically significant differences between anatomical regions were observed (p < 0.001). In these wavelength ranges, the highest reflectance values were recorded for the left thigh and left lower leg (crural region), whereas the lowest values were observed for the posterior neck region and the anterior neck regions.
Table 3.
Median directional hemispherical reflectance values measured at an incidence angle of 20° across infrared spectral bands: 1.5–2.0 µm, 2.0–3.5 µm, 3.0–4.0 µm, 4.0–5.0 µm, 5.0–10.5 µm, and 10.5–21.0 µm. Measurements were obtained at the following anatomical sites: right anterior neck region, left anterior neck region, posterior neck region, right forearm, left thigh, left calf, left lower leg (crural region), abdominal region superior to the umbilicus, right hand (first web space, dorsal side), right hand (dorsal side beneath the middle finger), right hand (palmar side beneath the middle finger), right hand (fourth web space, dorsal side), forehead (frontal region), and left cheek (buccal region). Statistical significance: *** p < 0.001. Color intensity reflects reflectance magnitude, with red indicating the highest values.
Table 4.
Percentage comparison of median directional hemispherical reflectance values measured at an incidence angle of 20° across anatomical sites, expressed relative to the lowest median reflectance value observed at each infrared spectral band (1.5–2.0 µm, 2.0–3.5 µm, 3.0–4.0 µm, 4.0–5.0 µm, 5.0–10.5 µm, and 10.5–21.0 µm). Measurements were obtained at the following anatomical sites: right anterior neck region, left anterior neck region, posterior neck region, right forearm, left thigh, left calf, left lower leg (crural region), abdominal region superior to the umbilicus, right hand (first web space, dorsal side), right hand (dorsal side beneath the middle finger), right hand (palmar side beneath the middle finger), right hand (fourth web space, dorsal side), forehead (frontal region), and left cheek (buccal region). Color intensity reflects reflectance magnitude, with red indicating the highest values.
3.3. Directional Hemispherical Reflectance Measured at an Incidence Angle of 60° Across the Infrared Spectral Bands of 1.5–2.0 µm, 2.0–3.5 µm, 3.0–4.0 µm, 4.0–5.0 µm, 5.0–10.5 µm, and 10.5–21.0 µm, Obtained Using the ET 100 Emissometer
Median reflectance values obtained at an incidence angle of 60° were consistently higher than those measured at 20° across all investigated spectral bands (Table 3 and Table 5). Statistical analysis demonstrated significant differences in directional hemispherical reflectance between anatomical regions at an incidence angle of 60° across all wavelength bands (p < 0.001) (Table 5 and Table 6). Within the 1.5–2.0 µm spectral band, reflectance values were the highest among all investigated infrared ranges, with median values ranging from 0.091 to 0.130 (Table 5). The highest reflectance values were observed for the left thigh, left calf, and left lower leg (crural region), whereas the lowest values were recorded for the dorsal and palmar surfaces of the hand beneath the middle finger. Comparison of reflectance patterns obtained at incidence angles of 20° and 60° revealed consistent spatial trends. Anatomical regions exhibiting the highest and lowest reflectance values remained unchanged, although absolute median reflectance values differed (Table 4 and Table 6). Within the 2.0–3.5 µm band, reflectance values ranged from 0.064 to 0.092. Higher reflectance values were again observed for the left thigh, left calf, and left lower leg (crural region), while lower values were associated with the dorsal and palmar surfaces of the hand beneath the middle finger and the anterior neck region. In the 3.0–4.0 µm spectral band, reflectance values were lower compared with other infrared ranges measured at 60°, ranging from 0.049 to 0.079. The highest reflectance values were recorded for the left calf and left lower leg (crural region), whereas the lowest values were observed for the abdominal region superior to the umbilicus and the anterior neck regions. Across the spectral bands from 4.0–5.0 µm to 10.5–21.0 µm, median reflectance values were comparable, ranging from 0.055 to 0.079. Despite the similarity in absolute reflectance values, statistically significant differences between anatomical regions were observed (p < 0.001). In these wavelength ranges, the highest reflectance values were recorded for the left calf and left lower leg (crural region), whereas the lowest values were observed for the abdominal region superior to the umbilicus, dorsal hand surface beneath the middle finger, posterior neck region, and anterior neck regions.
Table 5.
Median directional hemispherical reflectance values measured at an incidence angle of 60° across infrared spectral bands: 1.5–2.0 µm, 2.0–3.5 µm, 3.0–4.0 µm, 4.0–5.0 µm, 5.0–10.5 µm, and 10.5–21.0 µm. Measurements were obtained at the following anatomical sites: right anterior neck region, left anterior neck region, posterior neck region, right forearm, left thigh, left calf, left lower leg (crural region), abdominal region superior to the umbilicus, right hand (first web space, dorsal side), right hand (dorsal side beneath the middle finger), right hand (palmar side beneath the middle finger), right hand (fourth web space, dorsal side), forehead (frontal region), and left cheek (buccal region). Statistical significance: *** p < 0.001. Color intensity reflects reflectance magnitude, with orange indicating the highest values.
Table 6.
Percentage comparison of median directional hemispherical reflectance values measured at an incidence angle of 60° across anatomical sites, expressed relative to the lowest median reflectance value observed at each infrared spectral band (1.5–2.0 µm, 2.0–3.5 µm, 3.0–4.0 µm, 4.0–5.0 µm, 5.0–10.5 µm, and 10.5–21.0 µm)Measurements were obtained at the following anatomical sites: right anterior neck region, left anterior neck region, posterior neck region, right forearm, left thigh, left calf, left lower leg (crural region), abdominal region superior to the umbilicus, right hand (first web space, dorsal side), right hand (dorsal side beneath the middle finger), right hand (palmar side beneath the middle finger), right hand (fourth web space, dorsal side), forehead (frontal region), and left cheek (buccal region). Color intensity reflects reflectance magnitude, with orange indicating the highest values.
Across the full investigated spectrum, three general patterns were observed. First, reflectance was lowest in the shortest UV band and in the longest near-infrared/infrared bands, whereas the highest values were found in the visible-to-near-infrared range. Second, regional differences were consistent, with the forearm, thigh, calf, and crural region generally showing higher reflectance than the neck and dorsal hand regions. Third, in the infrared range, reflectance values decreased and became more uniform with increasing wavelength, while measurements obtained at 60° were consistently higher than those recorded at 20°.
4. Discussion
The evaluation of human skin reflectance represents an important aspect of research in thermoregulation, photobiology, and physiology, and plays a significant role in the development of thermal imaging technologies. DHR is considered one of the most precise methods for assessing skin reflective properties, as it enables accurate quantification of radiation reflected from the skin surface and, consequently, supports the indirect estimation of skin emissivity [7,13,39]. The present study provides a unique integration of full-spectrum in vivo reflectance measurements across multiple anatomical regions, which, to the best of our knowledge, has not been systematically reported in previous studies.
The main novel contribution of the present study lies in the full-spectrum in vivo mapping of directional hemispherical reflectance across 14 anatomical regions using two complementary instruments covering wavelengths from 335 nm to 21 µm. Unlike previous reports typically focused on selected spectral windows, specific cosmetic interventions, or single anatomical regions, the present work provides a broad comparative dataset showing how reflectance patterns vary simultaneously with wavelength, anatomical location, and measurement angle in the infrared range. This comprehensive approach may support future reference datasets and improve the interpretation of skin optical measurements in biomedical, cosmetic, and thermal imaging applications. In contrast to previous studies, the present work did not focus on a single intervention, selected skin condition, or limited spectral range. Its added value lies in combining UV–Vis–NIR reflectometry with mid- and far-infrared emissometric assessment within the same participants and anatomical protocol. Therefore, the novelty of this study is not the confirmation of individual optical principles, but the integrated in vivo comparison of their anatomical and spectral expression across the skin surface.
In contrast to previous studies that examined either individual physiological determinants of reflectance or treatment-related changes, the present study provides a multi-dimensional in vivo dataset integrating spectral, anatomical, and angular variability within a single experimental framework. While earlier works, including Banyś et al. [7] and Odrzywołek et al. [11], contributed important insights into specific aspects of skin optics, they did not address the combined effects of these variables. The added value of the present study lies in establishing a comprehensive reference dataset, which enables a more holistic interpretation of skin optical properties and may serve as a baseline for future interventional, diagnostic, and modeling studies.
The findings are consistent with the expected wavelength-dependent behavior of skin reflectance; however, the present study extends this understanding by demonstrating its simultaneous variation across anatomical regions, spectral ranges, and measurement geometries within a single in vivo experimental framework. Reflectance values were higher within the near-infrared range (approximately 1.5–3 µm), whereas a reduction in reflectance was observed in the mid- and far-infrared regions (above 3 µm), corresponding to increased skin emissivity. Reflectance values also exhibited regional variability. Anatomical sites such as the left thigh, left calf, and left lower leg (crural region) demonstrated higher reflectance compared with areas characterized by thinner skin, distinct structural properties, or increased pigmentation, including the posterior neck region and the palmar and dorsal surfaces of the hands. These observations support the notion that skin emissivity is not a constant parameter but rather reflects the combined influence of multiple physiological and structural factors, including hydration status, stratum corneum thickness, and chromophore composition.
The application of directional hemispherical reflectance enables a precise evaluation of the variability of skin optical properties, which may be of particular relevance in aesthetic medicine, dermatology, exercise physiology, and the development of infrared-based systems for physiological monitoring. Importantly, DHR measurements provide information not only on the reflective behavior of skin but also indirectly on its physiological state, surface structure, and potential pathological alterations [4,7,12,39,40]. In the present study, reflectance values exhibited a decreasing trend with increasing wavelength within the near- and mid-infrared ranges. Simultaneously, within the mid- and far-infrared bands (3.0–21.0 µm), reflectance values demonstrated relatively low variability in absolute magnitude across anatomical regions. This observation may be attributed to the increasing absorption of infrared radiation by water and collagen structures within the dermis [24,26,40,41]. These findings are consistent with established knowledge indicating that water and collagen represent the dominant absorbers in the infrared spectrum. As a result, skin exhibits high absorption and correspondingly low reflectance within these wavelength ranges [26].
Skin reflectance within the UV, visible, and NIR ranges is influenced by multiple factors, including sebum presence, measurement location, skin structure, and layer thickness. Previous studies have demonstrated that structural modifications of the epidermis may significantly alter reflectance characteristics. Stolecka-Warzecha et al. [27] reported that exfoliation procedures, such as microdermabrasion, lead to substantial changes in reflectance values, particularly within the UV and IR ranges, which were attributed to a reduction in stratum corneum thickness. The observed decrease in reflectance within the 480–600 nm band following microdermabrasion was explained by increased radiation penetration into deeper skin layers, including the basal layer and dermis, where chromophore concentration (melanin, oxyhemoglobin, and deoxyhemoglobin) is highest, resulting in enhanced absorption [26]. In the present study, reflectance values obtained across the UV–NIR spectral range (335–2500 nm) demonstrated a distribution comparable to pre-treatment values reported in the aforementioned study, supporting the reproducibility of reflectance patterns in healthy skin. Comparable findings were reported by Odrzywołek et al. [40], who investigated reflectance characteristics of both psoriatic and healthy skin within the 335–2500 nm range. In healthy skin, the lowest reflectance values were observed within the 335–380 nm and 1700–2500 nm bands, whereas the highest values occurred within the 590–720 nm and 700–1100 nm ranges, consistent with the results obtained in the present study. This agreement suggests the presence of characteristic reflectance patterns that may be common to healthy skin across populations, potentially supporting the development of reference databases for medical imaging applications.
The relationship between sebum content and skin reflectance has also been well documented. Banyś et al. [6] demonstrated an increase in reflectance with increasing wavelength up to the NIR range, followed by a reduction in the mid-infrared region. These observations align with the findings of the present study. Notably, the strongest association between sebum presence and reflectance was reported within the 700–1100 nm band. In agreement with these findings, sebaceous gland-rich regions such as the forehead and cheek exhibited elevated reflectance values within this spectral range. This observation further highlights the importance of appropriate skin preparation, including cleansing and degreasing, prior to light-based procedures such as laser therapy or phototherapy.
Because no solvent-based degreasing procedure was performed, residual surface sebum may have contributed to inter-site reflectance differences, especially in sebaceous-rich regions such as the forehead and cheek. Thus, the observed regional differences should be interpreted as reflecting both intrinsic anatomical variation and the natural condition of the skin surface at the time of measurement. Although ambient temperature, relative humidity, and acclimatization time were standardized, other physiological factors that may influence skin reflectance were not directly quantified. These include skin hydration status, circadian variation, recent physical activity, and short-term changes in cutaneous blood flow. Therefore, part of the observed inter-site variability may also reflect transient physiological conditions at the time of measurement.
Additional support for the stability of skin optical properties was provided by Mickoś et al. [42], who investigated the photoprotective properties of plant oils. Reflectance values reported for the forearm within the 1000–1700 nm and 1700–2500 nm bands were comparable to those obtained in the present study. Minor discrepancies may be attributed to interindividual variability and biological heterogeneity. Collectively, the present findings confirm that directional hemispherical reflectance analysis provides a precise and reliable approach for evaluating skin optical properties across a broad spectral range.
The primary objective of this study was a comparative analysis of skin reflectance across anatomical locations rather than the estimation of population parameters, which justifies the applied sample size. The study has several limitations that should be considered when interpreting the findings. First, the sample size was relatively small and included only women with Fitzpatrick skin phototypes II–III. Consequently, the results should not be directly generalized to men, individuals representing broader ethnic diversity, or subjects with darker or lighter skin phototypes. However, the homogeneity of the study group reduced melanin-related variability and allowed more controlled comparison of anatomical-site differences in reflectance. Although the homogeneity of the study group with respect to sex and skin phototype may limit generalizability, it allowed for controlled assessment by reducing melanin-related variability in reflectance measurements. Reflectance measurements may represent a valuable tool in dermatological monitoring, cosmetic assessment, and emissivity-related investigations. Moreover, the limited availability of studies employing directional hemispherical reflectance in the context of skin emissivity highlights the novelty of the present work and underscores the need for further research in this field.
The present findings support the proposed hypotheses indicating that both skin reflectance and emissivity are dependent on anatomical location and radiation wavelength. The results additionally suggest a tendency toward reduced and more uniform reflectance values within the mid- and far-infrared spectral ranges. Given the limited availability of directly comparable studies, further investigations involving larger populations and a broader range of physiological variables are warranted. The continued development of reflectance-based methodologies, particularly those employing directional hemispherical reflectance, may contribute to a more comprehensive understanding of the mechanisms governing skin optical properties and facilitate the advancement of novel diagnostic and therapeutic approaches.
5. Conclusions
The present study provides an integrated in vivo dataset of directional hemispherical reflectance of female skin across 14 anatomical regions and a broad spectral range from 335 nm to 21 µm. The results confirm that anatomical differences in skin reflectance are not uniform across the electromagnetic spectrum, but depend on the spectral band and, in the infrared range, on the measurement angle. The most pronounced regional differentiation was observed in selected UV–NIR and short-wave infrared bands, whereas mid- and far-infrared reflectance values were generally lower and more spatially uniform. These findings extend previous skin reflectance studies by combining broad spectral coverage with multi-site anatomical mapping and angular infrared assessment within the same experimental protocol. The dataset may support future reference values for biomedical optics, cosmetic assessment, and thermal imaging applications. Further studies including larger and more diverse populations are required to determine the generalizability of these patterns.
Author Contributions
Conceptualization, M.H.-P. and S.W.; methodology, S.W., P.Z. and M.H.-P.; formal analysis, S.W.; investigation, P.Z.; resources, M.H.-P.; data curation, M.H.-P. and P.Z.; writing—original draft preparation, J.W. and P.Z.; writing—review and editing, J.W. and M.H.-P.; supervision, S.W.; project administration, S.W. All authors have read and agreed to the published version of the manuscript.
Funding
Medical University of Silesia BNW-1-108/K/4/I.
Institutional Review Board Statement
The study was conducted at the Department of Basic Biomedical Sciences, Medical University of Silesia, Katowice, Poland. The study protocol was designed in accordance with the Declaration of Helsinki and approved by the Bioethics Committee of the Medical University of Silesia (PCN/CBN/0022/KB1/27/III/16/17/21; approval date: 29 June 2021).
Informed Consent Statement
Informed consent was obtained from all subjects involved in the study.
Data Availability Statement
The datasets used and/or analysed during the current study are available from the corresponding author upon reasonable request.
Conflicts of Interest
The authors declare no conflicts of interest.
Abbreviations
The following abbreviations are used in this manuscript:
| DHR | Directional hemispherical reflectance |
| UV | Ultraviolet |
| IR | Infrared |
| NIR | Near-infrared |
References
- Wilczyński, S.; Deda, A.; Koprowski, R.; Banyś, A.; Błońska-Fajfrowska, B. The use of directional reflectance measurement for in vivo assessment of protective properties of cosmetics in the infrared radiation range. Photochem. Photobiol. 2017, 93, 1303–1311. [Google Scholar] [CrossRef] [Scilit]
- Fernández-Cuevas, I.; Bouzas Marins, J.C.; Arnáiz Lastras, J.; Gómez Carmona, P.M.; Piñonosa Cano, S.; García-Concepción, M.Á.; Sillero-Quintana, M. Classification of factors influencing the use of infrared thermography in humans: A review. Infrared Phys. Technol. 2015, 71, 28–55. [Google Scholar] [CrossRef] [Scilit]
- Charlton, M.; Stanley, S.A.; Whitman, Z.; Wenn, V.; Coats, T.J.; Sims, M.; Thompson, J.P. The effect of constitutive pigmentation on the measured emissivity of human skin. PLoS ONE 2020, 15, e0241843. [Google Scholar] [CrossRef] [Scilit]
- Bałamut, K.; Kaszuba, N.; Kaszuba, M.; Stanek, A.; Cholewka, A.J. Zastosowanie termowizji w medycynie estetycznej—Praca pilotażowa. Inż. Fiz. Med. 2021, 10, 329–331. [Google Scholar]
- Modrzejewska, A.; Parafiniuk, M. Zastosowanie termografii w medycynie—Przegląd literatury. Pomeranian J. Life Sci. 2018, 64, 29–32. [Google Scholar] [CrossRef] [Scilit]
- Banyś, A.; Hartman-Petrycka, M.; Kras, K.; Kamińska, M.; Krusiec-Świdergoł, B.; Popielski, P.; Lebiedowska, A.; Wilczyński, S. The influence of sebum on directional reflectance of the skin. Appl. Sci. 2023, 13, 2838. [Google Scholar] [CrossRef] [Scilit]
- Dell’Isola, G.B.; Cosentini, E.; Canale, L.; Ficco, G.; Dell’Isola, M. Noncontact body temperature measurement: Uncertainty evaluation and screening decision rule to prevent the spread of COVID-19. Sensors 2021, 21, 346. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Yoshida, K.; Okiyama, N. Estimation of reflectance, transmittance, and absorbance of cosmetic foundation layer on skin using translucency of skin. Opt. Express 2021, 29, 40038–40050. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Knapek, D. Bezpieczeństwo—Termowizja a pomiar temperatury ciała. Ochr. Bezp. 2020, 3, 29–31. [Google Scholar]
- Odrzywołek, W.; Deda, A.; Zdrada-Nowak, J.; Błońska-Fajfrowska, B.; Wcisło-Dziadecka, D.; Wilczyński, S. Effect of narrow-band ultraviolet B therapy of psoriasis vulgaris on skin directional reflectance, skin density, and epidermal thickness. Appl. Sci. 2023, 13, 9311. [Google Scholar] [CrossRef] [Scilit]
- Keenan, E.; Gethin, G.; Flynn, L.; Watterson, D.; O’Connor, G. Investigating point-to-point emissivity variation of wounded tissue. Physiol. Meas. 2017, 38, 37. [Google Scholar]
- Verstockt, J.; Verspeek, S.; Thiessen, F.; Tjalma, W.A.; Brochez, L.; Steenackers, G. Skin cancer detection using infrared thermography: Measurement setup, procedure, and equipment. Sensors 2022, 22, 3327. [Google Scholar] [CrossRef] [Scilit]
- Bauer, J.; Dereń, E. Standaryzacja badań termograficznych w medycynie i fizykoterapii. Inż. Biomed. 2014, 20, 11–20. [Google Scholar]
- Żyłka, W.; Gala-Błądzińska, A. Zastosowanie termowizji w diagnostyce chorób naczyń u chorych hemodializowanych. Acta Bio-Opt. Inform. Med. 2016, 22, 121–129. [Google Scholar]
- Saiko, G. Skin temperature: The impact of perfusion, epidermis thickness, and skin wetness. Appl. Sci. 2022, 12, 7106. [Google Scholar] [CrossRef] [Scilit]
- Cooksey, C.C.; Tsai, B.K.; Allen, D.W. Spectral reflectance variability of skin and attributing factors. Proc. SPIE 2015, 9461, 510–517. [Google Scholar]
- Verdel, N.; Marin, A.; Milanič, M.; Majaron, B. Physiological and structural characterization of human skin in vivo using combined photothermal radiometry and diffuse reflectance spectroscopy. Biomed. Opt. Express 2019, 10, 944–960. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Dziewiecki, A. Zasady bezstykowego pomiaru temperatury z użyciem nowoczesnych pirometrów. Inż. Bezpieczenstwa Obiektów Antropog. 2017, 1, 36–40. [Google Scholar]
- Hagen, N. Review of thermal infrared polarimetry, I: Theory. Opt. Eng. 2022, 61, 070902. [Google Scholar] [CrossRef] [Scilit]
- Li, S. Human Skin Characterization and Analysis Based on Hyperspectral Reflectance Using Machine Learning. Ph.D. Thesis, Université de Lyon, Lyon, France, 2021. Available online: https://theses.hal.science/tel-03412032/document (accessed on 8 April 2025).
- Solano, F. Photoprotection and skin pigmentation: Melanin-related molecules and some other new agents obtained from natural sources. Molecules 2020, 25, 1537. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Zonios, G.; Dimou, A.; Bassukas, I.; Galaris, D.; Tsolakidis, A.; Kaxiras, E. Melanin absorption spectroscopy: New method for noninvasive skin investigation and melanoma detection. J. Biomed. Opt. 2008, 13, 014017. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Patel, A.D.; Chopra, R.; Avram, M.; Sakamoto, F.H.; Kilmer, S.; Anderson, R.R.; Ibrahimi, O.A. Updates on lasers in dermatology. Dermatol. Clin. 2024, 42, 33–44. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Mustafa, M.B.; Hamoudi, W.K.; Khashan, K.S. Non-ablative laser skin rejuvenation by targeting indigenous chromophores. J. King Saud. Univ. Sci. 2024, 36, 103516. [Google Scholar] [CrossRef] [Scilit]
- Yakimov, B.P.; Davydov, D.A.; Fadeev, V.V.; Budylin, G.S.; Shirshin, E.A. Comparative analysis of the methods for quantitative determination of water content in skin from diffuse reflectance spectroscopy data. Quantum Electron. 2020, 50, 41–46. [Google Scholar] [CrossRef] [Scilit]
- Tanaka, Y.; Tsunemi, Y.; Kawashima, M.; Tatewaki, N.; Nishida, H. Objective assessment of skin tightening in Asians using a water-filtered near-infrared (1000–1800 nm) device with contact-cooling and freezer-stored gel. Clin. Cosmet. Investig. Dermatol. 2013, 6, 167–176. [Google Scholar] [CrossRef] [Scilit]
- Stolecka-Warzecha, A.; Brandys, A.; Bożek, M.; Lebiedowska, A.; Błońska-Fajfrowska, B.; Wilczyński, S. Use of directional-hemispherical reflectance to identify female skin features in response to microdermabrasion treatment. Appl. Sci. 2022, 12, 12087. [Google Scholar] [CrossRef] [Scilit]
- Spigulis, J.; Oshina, I.; Berzina, A.; Bykov, A. Smartphone snapshot mapping of skin chromophores under triple-wavelength laser illumination. J. Biomed. Opt. 2017, 22, 091508. [Google Scholar] [CrossRef] [Scilit]
- Stolecka-Warzecha, A.; Wilczyński, S.; Pawlus, A.; Lebiedowska, A.; Chmielewski, Ł.; Niezgoda, Z. The use of hemispheric directional reflectance method to verify the usefulness of filters protecting the skin against infrared radiation. Clin. Cosmet. Investig. Dermatol. 2023, 16, 2663–2675. [Google Scholar] [CrossRef] [Scilit]
- Bernard, V.; Staffa, E.; Mornstein, V.; Bourek, A. Infrared camera assessment of skin surface temperature: Effect of emissivity. Phys. Med. 2013, 29, 583–591. [Google Scholar] [CrossRef] [Scilit]
- Keenan, E.; Gethin, G.; Flynn, L.; Watterson, D.; O’Connor, G.M. Enhanced thermal imaging of wound tissue for better clinical decision making. Physiol. Meas. 2017, 38, 1104–1115. [Google Scholar] [CrossRef] [Scilit]
- Qassem, M.; Kyriacou, P.A. Investigating skin barrier function utilizing reflectance NIR spectroscopy. In Proceedsing of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society; IEEE: New York, NY, USA, 2014; pp. 3735–3738. [Google Scholar]
- Cole, C.; Shyr, T.; Ou-Yang, H. Metal oxide sunscreens protect skin by absorption, not by reflection or scattering. Photodermatol. Photoimmunol. Photomed. 2016, 32, 5–10. [Google Scholar] [CrossRef] [Scilit]
- Ezerskaia, A.; Ras, A.; Bloemen, P.; Pereira, S.F.; Urbach, H.P.; Varghese, B. High sensitivity optical measurement of skin gloss. Biomed. Opt. Express 2017, 8, 3981–3992. [Google Scholar] [CrossRef] [Scilit]
- Lentner, A.; Wienert, V. A new method for assessing the gloss of human skin. Ski. Pharmacol. 1996, 9, 184–189. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Yudovsky, D.; Pilon, L. Retrieving skin properties from in vivo spectral reflectance measurements. J. Biophotonics 2011, 4, 305–314. [Google Scholar] [CrossRef] [Scilit]
- Sakuma, T.H.; Maibach, H.I. Oily skin: An overview. Ski. Pharmacol. Physiol. 2012, 25, 227–235. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Winkler, D.; Eyssele, T.; Glanzer, A.; Wirsching, H.; Klein, E.; Welzel, J.; Schuh, S. Use of non-invasive optical imaging techniques in dermatology. J. Dtsch. Dermatol. Ges. 2025, 23, 1430–1443. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Odrzywołek, W.; Deda, A.; Zdrada, J.; Wcisło-Dziadecka, D.; Lipka-Trawińska, A.; Błońska-Fajfrowska, B.; Wilczyński, S. Assessment of psoriatic skin features using non-invasive imaging technique. Processes 2022, 10, 985. [Google Scholar] [CrossRef] [Scilit]
- Kasprzyk-Kucewicz, T.; Stanek, A.; Kazior, M.; Szurko, A.; Sieroń, K.; Cholewka, A. Zastosowanie obrazowania termicznego w ocenie wydolności organizmu sportowców przy użyciu ergometru wioślarskiego. Inż. Fiz. Med. 2020, 9, 66–68. [Google Scholar]
- Park, E.; Lee, Y.J.; Lee, C.; Eom, T.J. Effective photoacoustic absorption spectrum for collagen-based tissue imaging. J. Biomed. Opt. 2020, 25, 056002. [Google Scholar] [CrossRef] [Scilit]
- Mickoś, E.; Michalak, M.; Hartman-Petrycka, M.; Banyś, A.; Babczyńska, P.; Koprowski, R.; Wilczyński, S. Assessing skin photoprotection in the infrared range: The reflectance profiles of cold-pressed plant oils. Cosmetics 2025, 12, 80. [Google Scholar] [CrossRef] [Scilit]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.