Abstract
In this manuscript, an acute preclinical feasibility evaluation was used to examine an intravascular ultrasound (IVUS) diagnostic device incorporating mechanically rotating catheters with frequencies of 30 MHz (PU3030/PU303018) and 15 MHz (PU8015) developed for peripheral intravascular imaging. The configurations accommodate 0.014, 0.018, and 0.035 inch guidewires. Transducer simulations and bench measurements characterized the two acoustic designs of the piezoelectric ultrasonic transducers for the IVUS catheters; the wire-phantom experiments and animal experiments directly evaluated PU303018 and PU8015. In wire-phantom experiments, the reported phantom bounds were an imaging radius ≥10 mm and axial resolution ≤100 μm for PU303018, and an imaging radius ≥20 mm and axial resolution ≤300 μm for PU8015. In animal experiments, six healthy pigs underwent acute evaluation: three contributed 12 stented vessel segments to a commercial-device comparison, and three underwent dedicated histopathological assessment. Pullbacks at 1, 3, 6, and 10 mm/s were explored. Using the stent cross-sectional area obtained using two commercial IVUS catheters (Eagle Eye Platinum, Visions PV.035) as the reference, the relative errors corresponding to 1 mm/s, 3 mm/s, 6 mm/s, and 10 mm/s pullback speeds were (4.28 ± 1.67)%, (3.33 ± 2.58)%, (3.80 ± 2.55)%, and (2.73 ± 2.05)%, respectively. The lumen/intimal interface, vascular branch structures, stent struts, and the positional relationship between the stent and the lumen were the main evaluation contents, and the image quality of all indicators was good and basically consistent with that of the commercial IVUS catheters. Safety results show that all six experimental animals successfully completed the experiment, with no device-related adverse events; postoperative DSA re-examination found no vascular dissection, perforation, thrombosis, air embolism, or significant blood flow slowing. In histopathological examination, only mild vessel wall injury and slight inflammatory response were observed locally in individual vessels; considering the experimental procedure, these are more likely related to mechanical stimulation caused by vascular access establishment and guidewire/catheter manipulation; no thrombosis, necrosis, or severe inflammatory response was observed in any target vessel. The findings support acute preclinical feasibility in healthy porcine vessels with implanted stents.
1. Introduction
Peripheral arterial disease (PAD) is a life-threatening and limb-threatening manifestation of systemic atherosclerosis associated with high risks of myocardial infarction, stroke, amputation, disability, and premature mortality [1,2]. In 2023, an estimated 316.5 million individuals globally were living with PAD, corresponding to a prevalence of 6.6% among adults aged 25 years or older [1]. Atherosclerotic plaque deposition in the intima is the primary mechanism underlying peripheral arterial stenosis. Vessel occlusion can be further promoted by intraluminal thrombi, medial calcification, and intimal thickening [3,4]. When medium-sized peripheral arteries become blocked, the downstream tissues become ischemic. This condition may manifest as claudication, atypical leg discomfort, or impaired mobility, and revascularization is frequently needed to enhance quality of life and prevent amputation [4].
Digital Subtraction Angiography (DSA) is a commonly used imaging modality in PAD. It is a vascular imaging method based on X-ray and digital image processing technology that can display blood vessels in real time. However, it mainly presents the two-dimensional contours of blood vessels and cannot comprehensively show vessel inner wall structure, lesion distribution, or stent apposition after deployment [5,6,7,8]. In addition to DSA, Intravascular Ultrasound (IVUS) is often performed with another medical imaging modality [9]. In order to perform IVUS, a catheter with a miniature ultrasonic transducer mounted at its tip is introduced via arterial or venous puncture and advanced into the target vascular region. Then, the cross-sectional vascular images can be acquired by scanning through the miniature ultrasonic transducer, which can display the vessel wall and lumen structures from an intraluminal perspective, thereby providing complementary information for lesion assessment and optimization of interventional treatment [5,10,11,12]. IVUS can clearly visualize vascular morphology, lesion characteristics, and their distribution within the vessel wall in real time. It can also be used to measure vessel inner diameter and cross-sectional area, thereby guiding device selection and postprocedural optimization [6,7,8,13,14]. In recent years, IVUS has also been increasingly used in the diagnosis of peripheral vascular diseases [6,7,12,13,15,16,17,18,19,20].
Current IVUS technology is predominantly side-looking, using either mechanically rotated single-element transducers or solid-state arrays, to provide cross-sectional morphological information for lesion assessment and stent guidance. Many efforts have focused on improving the performance of IVUS. Janjic et al. reported an 80-element PZT matrix transducer with a front-end ASIC requiring only four cables, enabling 3D FL-IVUS imaging [21]. Cabrera-Munoz et al. fabricated a 32-element 30 MHz phased-array transducer within an 8F catheter, demonstrating forward-looking imaging of structures ahead of the catheter [22]. More recently, Collins et al. developed a dual-resonance 16/32 MHz single-element transducer with a single electrical connection, integrated with a robotically steerable guidewire, to achieve simultaneous high-resolution and high-penetration imaging [4]. In parallel, micromachined ultrasonic transducers (MUTs), particularly CMUTs, have attracted increasing attention for IVUS owing to their wide bandwidth, suitability for high-density arrays, and compatibility with CMOS integration, although high-frequency operation and flexible array fabrication remain challenging [23]. Beyond morphological imaging, IVUS has also been combined with functional and elasticity imaging. Shih et al. integrated IVUS with acoustic radiation force elasticity imaging using a dual-element transducer to assess the stiffness of plaques and vessel walls [24]. Hong et al. designed a dual-mode catheter that combines a forward-looking transducer for PW Doppler flow measurement with a side-looking transducer for B-mode imaging, enabling one-time acquisition of both morphological and functional information [25]. Furthermore, many IVUS catheter products are available on the market, such as Visions PV.035 [26], Visions PV.018 [27], Eagle Eye Platinum [28], etc.
Peripheral arteries and veins vary substantially in anatomic scale and access requirements. The internal diameter of the adult carotid artery is generally 6–8 mm, whereas central veins such as the superior vena cava may reach 15–25 mm [29,30]. These differences motivate trade-offs among imaging depth, spatial resolution, catheter profile, and guidewire compatibility. For carotid arteries, IVUS catheters with a relatively high center frequency are typically required to achieve higher imaging quality, whereas for the superior vena cava, IVUS catheters with a relatively low center frequency are needed to provide greater imaging depth. Therefore, it is of great clinical significance and commercial value to develop multi-specification IVUS catheters for a single IVUS diagnostic device that are compatible with different vessel diameters and guidewire platforms. In large-diameter veins and central venous lesions, these multi-specification IVUS catheters can improve the accuracy of stenosis identification and stent positioning [31,32,33]. Several commercial IVUS platforms already support multiple catheter configurations and frequencies [26,27,28]. The independent engineering development, transducer design and bench measurements, and acute preclinical feasibility evaluation of a multi-specification system tailored for complex peripheral interventions remain useful for developing and evaluating the IVUS catheters for peripheral intravascular imaging.
To accommodate the complex peripheral vascular interventional environment, an IVUS diagnostic device with multi-specification IVUS catheters (PU3030, PU303018 and PU8015) for peripheral intravascular imaging is proposed. The PU3030 and PU303018 catheter designs have a diameter of 3 Fr and a center frequency of 30 MHz, with compatibility for 0.014- and 0.018 inch guidewires, respectively. PU8015 catheters have a diameter of 8 Fr, a center frequency of 15 MHz, and compatibility with a 0.035 inch guidewire. The comparison between the proposed multi-specification IVUS catheters and the prior art is summarized in Table 1. The development follows established multi-catheter IVUS technology, with incremental refinements to catheter profile, frequency, and piezoelectric composite implementation. The reported pulse-echo measurements for the shared 30 MHz acoustic design are a sensitivity level of −54.41 dB, a center frequency of 29.71 MHz, a −6 dB bandwidth of 53.94%, and a pulse length of 108.77 ns. Corresponding values for the 15 MHz design are −48.66 dB, 16.73 MHz, 49.21%, and 201.21 ns. PU303018 and PU8015 are evaluated directly in wire-phantom experiments and animal experiments, with Eagle Eye Platinum and Visions PV.035 as commercial comparators. Finally, the proposed IVUS catheters are obtained through gross anatomy and histopathological examinations of the experimental animals.
Table 1.
The design specifications of PU3030, PU303018 and PU8015 and the comparison with prior art.
2. Materials and Methods
2.1. The IVUS Imaging System
The proposed IVUS imaging system includes the V20i IVUS diagnostic device designed by Acoustic Life Science Co., Ltd. (Shanghai, China) and single-use IVUS catheters (PU3030, PU303018, and PU8015). As shown in Figure 1, the V20i IVUS diagnostic device mainly consists of an imaging console, a monitor, a bedside connector, and a catheter controller. The imaging console, monitor, and bedside connector together can provide functions such as digital transmission control, image processing, image display, and a user input interface. The catheter controller performs a series of functions, including transducer rotation control, ultrasonic wave excitation and reception, ultrasonic echo amplification, ultrasonic echo acquisition, image preprocessing, and digital transmission control.
Figure 1.
The structure of the V20i IVUS diagnostic device.
The single-use IVUS catheter consists of an imaging catheter, accessories, and a sterile bag, while the imaging catheter consists of a transducer, a transducer base, a drive shaft, and a connector assembly. Based on the structure of ultrasonic transducers, there are two types of IVUS catheters, which are a mechanical/rotational catheter and a solid-state catheter [10,34]. The mechanical/rotational IVUS catheter uses a single-element ultrasonic transducer mounted at the catheter tip, which is rotated by a flexible drive shaft to transmit and receive ultrasonic waves and thereby generate cross-sectional vessel images. The solid-state IVUS catheter, by contrast, uses a multi-element ultrasonic transducer array mounted circumferentially around the catheter tip and operated as a phased array, enabling cross-sectional vessel imaging through electronic beam sweeping around the vessel circumference. Given the simple structure and technological maturity of the mechanical/rotational IVUS catheter, the proposed IVUS catheters adopt this design. To accommodate the complex peripheral vascular interventional environment in PAD, three catheters are designed: PU3030, PU303018, and PU8015, as shown in Figure 2. Among them, PU3030 and PU303018 are high-frequency catheters intended for smaller-diameter carotid arteries, whereas PU8015 is a low-frequency catheter intended for larger-diameter central veins. Their design specifications are summarized in Table 1. During vascular image acquisition, the IVUS catheter is first introduced into the target vessel by puncture. A drive shaft rotates the ultrasonic transducer at the catheter tip for rotational scanning. Under excitation by the excitation circuit, the transducer emits ultrasonic waves, which are reflected by tissue structures in different layers of the vessel to form ultrasonic echoes. The ultrasonic imaging system then processes the echo signals received by the transducer to generate images. The V20i IVUS diagnostic device operates at a constant frame rate of 30 frames per second (fps) across different pullback speeds. Consequently, the longitudinal sampling interval is approximately 0.033 mm/frame at 1 mm/s, 0.1 mm/frame at 3 mm/s, 0.2 mm/frame at 6 mm/s, and 0.333 mm/frame at 10 mm/s. The V20i IVUS diagnostic device and the PU3030, PU303018, and PU8015 catheters are reported to have obtained medical device registration certificates from the National Medical Products Administration (NMPA), China (Registration Nos. 20253060067, 20263061029, and 20263060900).
Figure 2.
The overview of the single-use IVUS catheters: (a) PU3030; (b) PU303018; (c) PU8015.
2.2. The Design of the Piezoelectric Ultrasonic Transducer
The ultrasonic transducer is the core component of an IVUS catheter and is responsible for converting electrical energy into acoustic energy and vice versa. It is a multilayer structure consisting of a piezoelectric layer, a backing layer, and a matching layer, which are bonded through epoxy. To achieve higher electroacoustic conversion efficiency and better imaging depth and resolution, the piezoelectric layer is fabricated from a piezoelectric composite. The matrix material of the piezoelectric composite is PZT5H from CTS Corporation, whose properties are listed in Table 2 [35], and epoxy is used as the filler material. In the dice-and-fill process, ceramic pillars are formed by dicing, the kerfs are filled with epoxy and cured, and the composite is lapped to the required thickness. The dimensions of the piezoelectric composite for PU8015 are 2 mm × 0.7 mm × 0.1 mm with a ceramic volume fraction of 70%, whereas those for PU3030 and PU303018 are 0.8 mm × 0.45 mm × 0.05 mm with a ceramic volume fraction of 75%. The backing is a tungsten–epoxy mixture, and the matching layer is epoxy. The piezoelectric composite and backing layer are made by Acoustic Life Science Co., Ltd.
Table 2.
The material properties of PZT5H.
The resonant frequency of the piezoelectric material can generally be determined using the frequency constant Nt given in Table 2. Considering the effects of the matching layer and backing layer on the ultrasonic transducer, simulations are performed to design the dimensions of the ultrasonic transducer. PiezoCAD (v3.06, Sonic Concepts, Woodinville, WA, USA) is used for pulse-echo design calculations, with the backing and matching-layer inputs listed in Table 3. In PiezoCAD, the acoustic stack of the transducers is modeled using the Krimholtz–Leedom–Matthaei (KLM) equivalent circuit model. The initial thickness of the piezoelectric layer is set to a half-wavelength for the target center frequency, while the matching layer is set to a quarter-wavelength. Through iterative simulation, the structural parameters are optimized to achieve the desired center frequency, −6 dB bandwidth, and pulse length. The simulation outputs are summarized in Figure 3 and Table 4. In Table 4, LS denotes the pulse-echo loop gain sensitivity level in decibels; FC denotes center frequency, BW the fractional −6 dB bandwidth, and PL the pulse duration at −20 dB. To evaluate the acoustic field distribution, a spatial impulse response (SIR) model is established in Field II. The transducer aperture is modeled as a piston/rectangular element approximated by subdivided rectangular facets to satisfy the far-field accuracy criterion. The simulation medium is set as water with a sound speed of 1540 m/s and a sampling frequency of 100 MHz. The axial acoustic pressure attenuation and focal depth are extracted from the simulated field to characterize the axial acoustic field. The axial acoustic pressure attenuation results are shown in Figure 4. The simulated axial maxima are 2.98 mm for the 30 MHz design and 9.34 mm for the 15 MHz design.
Table 3.
The material parameters of the matching layer and backing layer.
Figure 3.
The simulated pulse-echo signals and their frequency spectrum: (a) PU3030/PU303018 acoustic design; (b) PU8015 acoustic design.
Table 4.
The simulation performance of PU3030, PU303018 and PU8015.
Figure 4.
The simulated axial acoustic pressure attenuation results: (a) PU3030/PU303018 acoustic design; (b) PU8015 acoustic design.
3. Results and Analysis
3.1. Characterization of the Piezoelectric Ultrasonic Transducer
The impedance of the piezoelectric composite and the piezoelectric ultrasonic transducer used in the IVUS catheters are characterized by an impedance analyzer (Keysight E4990A, Keysight Technologies, Santa Rosa, CA, USA), and the measurement results are shown in Figure 5. As shown in Figure 5a,b, for the piezoelectric composite of PU3030/PU303018, the resonant frequency (fr) of the piezoelectric ultrasonic transducer is 32.11 MHz, the anti-resonant frequency (fantir) is 39.74 MHz, and the effective electromechanical coupling coefficient (Keff) is 58.9%; for the piezoelectric composite of PU8015, the resonant frequency (fr) of the piezoelectric ultrasonic transducer is 15.03 MHz, the anti-resonant frequency (fantir) is 19.58 MHz, and the effective electromechanical coupling coefficient (Keff) is 61.09%. The Keff values of the fabricated piezoelectric composites are all greater than the kt of the matrix material (PZT5H). Due to the addition of the matching layer, two resonance peaks can be observed in the impedance measurement results. For the piezoelectric ultrasonic transducer of PU3030/PU303018, the 1st resonant frequency (fr) of the piezoelectric ultrasonic transducer is 25.66 MHz, the anti-resonant frequency (fantir) is 28.69 MHz; the 2nd resonant frequency of the piezoelectric ultrasonic transducer is 36.77 MHz, and the anti-resonant frequency is 40.30 MHz. For the piezoelectric ultrasonic transducer of PU8015, the 1st resonant frequency (fr) of the piezoelectric ultrasonic transducer is 12.88 MHz, the anti-resonant frequency (fantir) is 21.36 MHz; the 2nd resonant frequency of the piezoelectric ultrasonic transducer is 18.64 MHz, the anti-resonant frequency is 21.36 MHz.
Figure 5.
The impedance measurement results of the IVUS catheters: (a) impedance of PU3030/PU303018 piezoelectric composite, (b) impedance of PU8015 piezoelectric composite, (c) impedance of PU3030/PU303018 piezoelectric ultrasonic transducer, and (d) impedance of PU8015 piezoelectric ultrasonic transducer.
The pulse-echo characteristics of the piezoelectric ultrasonic transducer are also measured. The physical measurement setup is shown in Figure 6a; for clarity, the schematic of the pulse-echo configuration is illustrated in Figure 6b. The piezoelectric ultrasonic transducer and a reflection target are placed in a transparent water tank. The reflection target is fixed at a distance of 3.5 mm from the piezoelectric ultrasonic transducer. A dual pulser/receiver (DPR500, JSR Ultrasonics, Pittsford, NY, USA) is used to generate pulse signals that excite the piezoelectric ultrasonic transducer to transmit an ultrasonic wave and is also used to receive echo signals. An oscilloscope (Tektronix, Inc., Beaverton, OR, USA) connected to the dual pulser/receiver is used to display the received echo signals. The pulse-echo measurement is calibrated in accordance with the standard YY/T 1142-2013 [36] methods of measuring the frequency of medical ultrasonic equipment and probe. The representative measured waveforms and numerical results are shown in Figure 7 and Table 5. The difference in acoustic impedance between the PU3030/PU303018 and PU8015 acoustic designs leads to a certain difference in the reference voltage used when calculating their respective LS values. For the PU3030/PU303018 acoustic design, the measured center frequency is 29.71 MHz versus 31.783 MHz in simulation, a difference of −2.073 MHz (−6.52% relative to simulation); bandwidth is 53.94% versus 55.27%, a reduction of 1.33 percentage points. For the PU8015 acoustic design, the measured center frequency is 16.73 MHz versus 15.41 MHz in simulation, a difference of +1.32 MHz (+8.57%); bandwidth is 49.21% versus 69.13%, a reduction of 19.92 percentage points (28.82% relative to the simulated bandwidth). Several factors may contribute to this discrepancy. First, the PiezoCAD utilizes a KLM equivalent circuit model, which assumes homogeneous and isotropic materials. For piezoelectric composites with complex pillar-and-epoxy structures, this model cannot fully account for the lateral modes, internal scattering, and effective electromechanical coupling, potentially contributing to differences between simulated and measured bandwidth. In addition, fabrication tolerances, such as slight variations in the kerf width, epoxy filling uniformity, and layer thicknesses, may cause the physical device to deviate from the ideal simulated geometry and alter the measured bandwidth.
Figure 6.
The pulse-echo measurement setup of IVUS catheters: (a) physical measurement setup; (b) schematic of the pulse-echo configuration. The water tank contains the IVUS transducer and reflection target, separated by 3.5 mm. The DPR500 pulser/receiver supplies excitation and receives echoes; the oscilloscope records the received electrical waveform.
Figure 7.
The representative pulse-echo measurement waveforms of the IVUS catheters: (a) PU3030/PU303018 acoustic design, (b) PU8015 acoustic design.
Table 5.
Representative pulse-echo measurement results corresponding to Figure 7.
The acoustic characteristics of the piezoelectric ultrasonic transducer are measured by a hydrophone. The physical measurement setup is shown in Figure 8a; for clarity, the schematic of the acoustic-field measurement configuration is illustrated in Figure 8b. The piezoelectric ultrasonic transducer is placed at the center of a transparent water tank. The dual pulser/receiver is also used to generate pulse signals that excite the piezoelectric ultrasonic transducer. Then, the ultrasonic wave generated by the piezoelectric ultrasonic transducer is measured through a hydrophone (NH0500, Precision Acoustics Ltd., Dorchester, Dorset, UK). The hydrophone is calibrated in accordance with the standard IEC 62127-2 [37] Ultrasonics Hydrophones Part 2: Calibration for ultrasonic fields up to 40 MHz. The hydrophone is driven by a mechanical structure that can scan a certain region. The oscilloscope connected to the hydrophone is also used to display the received ultrasonic wave. The measurement results are shown in Figure 9. As shown in Figure 9, the ultrasonic transducer of PU3030/PU303018 has a focal depth of about 3 mm, while the ultrasonic transducer of PU8015 has a focal depth of about 10 mm, which is close to that of the simulation.
Figure 8.
The acoustic measurement setup of the IVUS catheters: (a) physical measurement setup, (b) schematic of the acoustic-field measurement configuration. The water tank contains the IVUS catheter and the scanned NH0500 hydrophone. The DPR500 excites the transducer, the positioning mechanism moves the hydrophone, and the oscilloscope records the hydrophone output. No reflection target is used in this field-scan arrangement.
Figure 9.
The acoustic measurement results of the IVUS catheters: (a) PU3030/PU303018 acoustic design, (b) PU8015 acoustic design.
3.2. Experiment of Wire Phantom
To evaluate the imaging radius, axial resolution, lateral resolution, longitudinal and lateral geometric position accuracy, and image geometric distortion of the proposed IVUS catheters, wire-phantom experiments are performed in water at pullback speeds of 1 mm/s, 3 mm/s, 6 mm/s, and 10 mm/s according to the requirements of Appendix A of the standard YY/T 1659-2019 [38]. As shown in Table 1, the PU3030 and PU303018 catheters are consistent in main technical parameters such as outer diameter and center frequency and differ only in guidewire compatibility; therefore, PU303018 is used to represent the shared 30 MHz acoustic design in these experiments. Since PU303018 and Eagle Eye Platinum are both small-outer-diameter IVUS catheters, while PU8015 and Visions PV.035 are mainly intended for large-diameter vessels and 0.035 inch guidewire platforms, Eagle Eye Platinum is selected as the corresponding reference for PU303018, and Visions PV.035 is selected as the corresponding reference for PU8015. The transducer center frequency is 20 MHz for the Eagle Eye Platinum catheter and 10 MHz for the Visions PV.035 catheter.
The measurement setup for the wire phantoms is similar to the pulse-echo measurement setup, except that the DPR500 Dual Pulser/Receiver and Tektronix oscilloscope are replaced by the V20i IVUS diagnostic device, and the reflection target is replaced by different wire phantoms. The wire phantom used to evaluate the imaging radius contains 20 target wires with an interval of 1 mm between adjacent wires. It is placed at a distance of 1 mm from the transducer during the measurement. The imaging radius is determined by the number of clearly resolved target wires at four different pullback speeds. The wire phantoms used to evaluate the axial resolution contain only two target wires, with intervals of 100 μm, 200 μm, and 300 μm, respectively. During the measurement, PU303018 is evaluated with the phantom placed at distances of 2 mm and 5 mm from the transducer, while PU8015 is evaluated at distances of 5 mm and 10 mm. The axial resolution is determined by the minimum distance between the two target wires that can be clearly resolved at four different pullback speeds. The evaluation for lateral resolution is similar to that for axial resolution, except that the intervals between the two target wires are 400 μm, 600 μm, 800 μm, 1000 μm, 1250 μm and 1500 μm, respectively. The wire phantoms used to evaluate the longitudinal and lateral geometric position accuracy also contain only two target wires, with an interval of 7 mm. PU303018 is evaluated with the phantom at a distance of 5 mm from the transducer, and PU8015 is evaluated at a distance of 10 mm. Geometric position accuracy is determined by the relative error between the wire interval measured by the V20i IVUS diagnostic device at four pullback speeds and the actual 7 mm distance. The target wires in the wire phantoms used to evaluate image geometric distortion form a rectangle. The rectangle dimensions are 3 mm × 4.5 mm for PU303018 and 16 mm × 23 mm for PU8015. During the measurement, the transducer passes through the center of the rectangle. The image geometric distortion is determined by the degree of distortion between the rectangular shape measured by the V20i IVUS diagnostic device at four pullback speeds and the actual shape. In addition, wire phantom imaging results from two commercial IVUS catheters (Eagle Eye Platinum and Visions PV.035) serve as references. The measurement procedures for Eagle Eye Platinum are identical to those for PU303018, and the measurement procedures for Visions PV.035 are identical to those for PU8015.
Figure 10, Figure 11, Figure 12 and Figure 13 illustrate the wire-phantom images, and Table 6 reports the available bounds. At different pullback speeds, PU303018 has an imaging radius ≥10 mm; an axial resolution ≤ 100 μm (at radii of 2 mm and 5 mm); and a lateral resolution ≤0.4 mm (at a radius of 2 mm) and ≤0.6 mm (at a radius of 5 mm). PU8015 has an imaging radius ≥20 mm; an axial resolution ≤ 300 μm (at radii of 5 mm and 10 mm); and a lateral resolution ≤0.8 mm (at a radius of 5 mm) and ≤1.5 mm (at a radius of 10 mm). For PU303018 and PU8015, the longitudinal geometric position accuracy is ≤10%, the lateral geometric position accuracy is ≤10%, and the image geometric distortion is ≤10%. As can be seen from the results in Figure 10, Figure 11, Figure 12 and Figure 13 and Table 6, the performance of PU303018 and PU8015 is similar to that of Eagle Eye Platinum and Visions PV.035.
Figure 10.
Representative wire-phantom images from PU303018: (a) imaging radius; (b) axial resolution; (c) lateral resolution.
Figure 11.
Representative wire-phantom images from Eagle Eye Platinum: (a) imaging radius; (b) axial resolution; (c) lateral resolution.
Figure 12.
Representative wire-phantom images from PU8015: (a) imaging radius; (b) axial resolution; (c) lateral resolution.
Figure 13.
Representative wire-phantom images from Visions PV 0.035: (a) imaging radius; (b) axial resolution; (c) lateral resolution.
Table 6.
The results of the wire-phantom experiment.
3.3. Experimental Setup of Animal Experiment
The animal experiments in this manuscript are designed with reference to the ARRIVE 2.0 guidelines [39]. Porcine vascular models are often used in the preclinical evaluation of vascular interventional devices because their vascular anatomy and hemodynamic characteristics are similar to those of humans [40,41]. In this manuscript, six healthy pigs are selected for animal experiments (license number: SYXK (Shanghai) 2021-0019). No sex restriction is applied to the pigs used in the experiment, and their weights range from 63.6 to 72.8 kg. The experimental pigs are housed in stainless steel cages placed above the floor, and the temperature and humidity of the animal room are monitored and controlled. During the experimental period, the animals are provided with an appropriate amount of qualified feed daily and have free access to drinking water. This experiment is reviewed and approved by the Experimental Animal Welfare and Ethics Committee (approval number: XNM-YX-20251121-01). According to the “Guidance for Animal Study of Medical Devices Part II: Study Design and Quality Assurance” [42], an adequate number of animals ensures the reliability of the experiment results and complies with the 3R principles (Replacement, Reduction, Refinement). If multiple observation time points are involved, the number of animals at each time point ensures the reliability of the results. Furthermore, the estimation of the animal number accounts for potential premature animal death due to various reasons.
Animals with poor health conditions or obvious physical abnormalities are not used in experiments. Animals deemed unsuitable for surgery by veterinary or surgical personnel are also excluded. The exclusion criteria for experimental animals are as follows:
- (1)
- Animals with various diseases (e.g., active consumptive diseases, hematological and metabolic diseases, infectious diseases, and zoonoses);
- (2)
- Animals with poor health conditions or obvious abnormal clinical manifestations during the quarantine period;
- (3)
- Animals found to have iodine allergy during surgery, or whose angiography shows anatomical structures inconsistent with the requirements of the animal model;
- (4)
- Intraoperative death of experimental animals caused by improper procedural manipulation unrelated to the test device;
- (5)
- Other complications or death unrelated to the test device.
Ultimately, six animals are used in the experiment.
In animal experiments, the carotid artery, renal artery, popliteal vein, subclavian vein, brachiocephalic vein, superior vena cava, and jugular vein are selected from six experimental pigs as target vessels, and catheters of corresponding specifications are used for endovascular experiments. As described in Section 3.2, only the PU303018 and PU8015 catheters are used in animal experiments. The proposed V20i IVUS diagnostic device is used as an experimental device, and an intravascular ultrasound diagnostic system manufactured by Volcano Corporation is used as a control device; it is used with two IVUS catheters: Eagle Eye Platinum and Visions PV.035.
The animal experiment is divided into two groups:
(1) Efficacy group: Three experimental animals are used to simulate clinical use scenarios, and safety evaluation does not include histopathological examination. Under identical experimental conditions, the experimental device and the control device are used to perform IVUS examinations and evaluations of the target vessels before and after stent implantation. The mean stent cross-sectional area and image quality are measured, with image quality expressed as the rate of excellent and good image quality. On this basis, the efficacy of the experimental device is evaluated under different pullback speeds and maximum numbers of pullbacks, and the safety of the experimental device is preliminarily assessed.
(2) Safety group: Three experimental animals are selected to simulate clinical use scenarios, and safety evaluation includes histopathological examination. The experimental device is used to perform IVUS examinations and evaluations of the corresponding target vessels in the experimental animals. The animals are then euthanized for observation, and histopathological examination of the treated vessels is performed to further verify the safety of the experimental device.
All procedures are performed according to the laboratory standard operating procedures. The animals are injected with Zoletil 50 to induce anesthesia and then connected to an anesthesia machine. Anesthesia is maintained by inhalation of 1–5% isoflurane. The femoral artery and iliac vein are punctured successively, and sheaths are inserted. Anticoagulants are administered, and a guidewire is then advanced to the target vessel segment. In a randomized order, the IVUS catheters of the experimental device or the control device are advanced over the guidewire to the distal segment of the target vessel. The experimental details of the efficacy group and safety group for pullback imaging are as follows:
(1) Efficacy group: The experimental device undergoes pullback imaging at a pullback speed of 10 mm/s, whereas the control device undergoes pullback imaging at a pullback speed of 1 mm/s. One stent is implanted in the middle segment of selected vessels, such as the carotid artery, popliteal vein, subclavian vein, and jugular vein. The experimental device performs two pullback imaging runs at pullback speeds of 1 mm/s, 3 mm/s, 6 mm/s, and 10 mm/s, respectively, while the control device performs pullback imaging at 1 mm/s.
(2) Safety group: The experimental device performs two pullback imaging runs at pullback speeds of 1 mm/s, 3 mm/s, 6 mm/s, and 10 mm/s, respectively, and then performs one additional pullback imaging run at 10 mm/s.
In both groups, each PU303018 catheter completes pullback imaging in the carotid artery and renal artery, with 9 pullback images at each site, totaling 18 pullback images. Each PU8015 catheter completes pullback imaging in the popliteal vein, jugular vein, subclavian vein, and the brachiocephalic vein combined with the superior vena cava, with 9 pullback images at each site, totaling 36 pullback images. After the procedure, the animals are euthanized under deep anesthesia by intravenous injection of a high-concentration potassium chloride solution. Animals that reach the study endpoint undergo gross necropsy to examine the body surface, glands, and organs for abnormalities.
Animals are randomly allocated to the efficacy group or the safety group by investigators not involved in the surgical procedures. Within each animal, the order of catheter testing (experimental device vs. control device) and the sequence of target vessel imaging are randomized using a balanced design to minimize potential confounding from temporal trends or operator-related factors. Due to the differences between the experimental device and the control device, blinding of the operators and evaluators is not feasible. Therefore, an open-label design is adopted in this study. To mitigate potential assessment bias, an independent evaluation panel is established, and the operating physicians are not involved in the image quality evaluation.
The efficacy of the experimental device under different pullback speeds and maximum pullback cycles is assessed by the mean stent cross-sectional area and image quality (expressed as the excellent and good image quality rate). The stent cross-sectional area (mm2) is defined as the area enclosed by the stent boundaries. After stent implantation in the target vessels of the experimental animals, the experimental and control devices are randomly applied to perform automatic pullback imaging of the stented target vessels. The experimental device undergoes one automatic pullback at each of the following speeds: 1 mm/s, 3 mm/s, 6 mm/s, and 10 mm/s, while the control device undergoes one pullback at 1 mm/s. The pullback starts at a point 10 mm distal to the implanted stent and ends at a point 10 mm proximal to the stent after the stent struts completely disappear. The imaging video files of both the experimental and control devices are saved and analyzed by an independent evaluation panel consisting of two experts to obtain the stent cross-sectional area data. The stent cross-sectional area is measured by replaying the IVUS recordings. The measurement range begins when at least 50% of the stent struts are first visualized and ends when the stent struts completely disappear for the first time. Within this measurement range for the carotid artery, popliteal vein, jugular vein, and subclavian vein, every 5 mm of automatic pullback imaging constitutes one measurement segment for the stent cross-sectional area. The mean stent cross-sectional area at each pullback speed is calculated using the following equation:
where S denotes the stent cross-sectional area measured for a single segment, and n denotes the number of selected measurement segments. IBM SPSS Statistics 28 is used for statistical analysis. Continuous data are usually expressed as mean ± standard deviation, while categorical data are expressed as frequencies and percentages. The IVUS images of the same vessel after stent implantation are used as the evaluation objects, thereby minimizing the influence of individual animal differences and vascular anatomical differences. The relative error of stent cross-sectional area is used to evaluate measurement agreement, with the control device results as the reference [8,14], according to the following formula:
where δr denotes the relative error of the mean stent cross-sectional area, Sed denotes the mean area measured by the experimental device, and Scd denotes the mean area measured by the control device. A predefined acceptable error limit is set at 10%. One-sample t-tests are performed on the relative error of the mean stent cross-sectional area at different pullback speeds to examine whether the mean error of each group is below 10%. To evaluate inter-device agreement, 95% confidence intervals (CIs) for the relative errors are calculated. Additionally, a Bland–Altman analysis is performed on the paired mean cross-sectional areas of the 12 vessel segments to calculate the mean difference and 95% limits of agreement (LoA). To illustrate the hierarchical data structure (multiple segments per animal), individual animal data points are color-coded in the Bland–Altman plots, allowing visual assessment of intra-animal variability. Two independent clinicians with extensive experience in IVUS examination evaluate the IVUS image quality of the experimental and control devices before and after stent implantation. The evaluation items include the lumen/intimal interface, external elastic membrane interface, anatomical structure of vascular branches, stent struts, and the positional relationship between stent struts and the vessel lumen (stent strut-related items are evaluated only after stent implantation). The final overall evaluation result for each image is determined as the lower of the two reviewers’ scores for the same image. The image-quality assessment criteria are provided in Table 7. The overall IVUS image quality is rated as “Excellent” when all observable sub-items are rated as excellent, “Good” when one or more sub-items are rated as good, and “Poor” when one or more sub-items are rated as poor. The excellent and good image quality rate is calculated using Equation (3):
where NE&G denotes the percentage of images rated as “Excellent” and “Good”, NE denotes the number of images rated as “Excellent”, NG denotes the number of images rated as “Good”, and NT denotes the total number of evaluated images.
Table 7.
Image-quality assessment criteria.
The safety evaluation includes hemodynamic and body temperature monitoring, digital subtraction angiography (DSA) examination, the incidence of catheter-related adverse events, the device defect rate, gross anatomical observation, and histopathological examination.
- (1)
- Hemodynamic and body temperature monitoring: Intraoperative vital signs (body temperature), hemodynamic parameters (diastolic/systolic blood pressure, heart rate, and oxygen saturation), and device-related adverse events are observed and recorded for each IVUS catheter before the initial pullback, during pullbacks at different speeds (recorded when the catheter reaches the midpoint of the vessel), and after the final pullback.
- (2)
- DSA examination: Immediately after the procedure, angiography images are obtained to evaluate the treated vessels for dissection, thrombosis/air embolism, perforation, aneurysm formation, and other abnormalities. Follow-up angiography of the target arteries, distal lower extremity vessels, target veins, and pulmonary arteries is performed to assess blood flow patency, filling defects, and slowed blood flow. Concurrently, gross anatomical observation of the distal lower extremity arteries and pulmonary arteries is conducted to check for the presence of thrombosis.
- (3)
- Incidence of catheter-related adverse events: This is defined as the ratio of the number of adverse events related to the single-use peripheral intravascular imaging catheters (including test and control devices) to the total number of catheter deliveries. Catheter-related adverse events include, but are not limited to, animal death, vascular occlusion, perforation, and dissection requiring intervention.
- (4)
- Device defect rate: This is defined as the ratio of the number of unreasonable risk events endangering health and safety (including mislabeling, catheter quality issues, and system failures) to the total number of catheter deliveries.
- (5)
- Gross anatomical observation: Routine gross visual observation and photographic recording are performed on the treated vessels and major organs (heart, liver, spleen, lungs, bilateral kidneys, and brain). The appearance and cross-sections of each organ are visually inspected and photographed for archiving.
- (6)
- Histopathological examination: Histopathological analysis is performed on the treated vessels.
3.4. Experiment Result of Animal Experiment
To evaluate the measurement agreement of the experimental device for implanted stents at different pullback speeds, the mean stent cross-sectional area measured by the control device at a pullback speed of 1 mm/s serves as the reference value. The three animals in the efficacy group are selected, and the carotid artery, popliteal vein, jugular vein, and subclavian vein are chosen, resulting in a total of 12 stent-implanted vessel segments. The mean stent cross-sectional area is measured by replaying the post-stent implantation images, and the relative errors between the experimental device (at 1, 3, 6, and 10 mm/s) and the control device (at 1 mm/s) are calculated. The measurement results and measurement errors are shown in Table 8 and Table 9. The results show that the mean stent cross-sectional areas measured by the experimental device are close to those obtained by the control device at all pullback speeds. The relative errors are (4.28 ± 1.67)%, (3.33 ± 2.58)%, (3.80 ± 2.55)%, and (2.73 ± 2.05)%, respectively. A one-sample t-test is used to verify whether these relative errors are significantly less than the predefined acceptable error limit of 10%. It should be noted that the primary purpose of the one-sample t-test here is to perform a single-group non-inferiority assessment, verifying whether the measurement errors significantly deviate from the fixed 10% clinical acceptable limit. However, because multiple vessel segments are evaluated within the same animals, the observations exhibit a hierarchical structure. The individual measurement errors for all vessel segments across the three animals are detailed in Table 9 to transparently present the data distribution and intra-animal variability. The 95% confidence intervals (CIs) for all pullback speeds are well below the 10% threshold (e.g., [3.22%, 5.34%] at 1.0 mm/s). To further evaluate inter-device absolute agreement, a Bland–Altman analysis is performed using the paired mean areas of the 12 vessel segments (Figure 14). The results show small mean biases of 0.83, 1.03, 0.44, and 0.56 mm2 at 1.0, 3.0, 6.0, and 10.0 mm/s, respectively, with 95% limits of agreement (LoA) within clinically acceptable ranges. The data points from the three animals are color-coded in the plots, visually confirming that the high agreement is consistent across different animals. The above results demonstrate that the experimental device has good and comparable measurement agreement under different automatic pullback speeds [14].
Table 8.
Measurement results of mean stent cross-sectional area (unit: mm2).
Table 9.
Measurement errors of mean stent cross-sectional area.
Figure 14.
Bland–Altman plot for the mean stent cross-sectional area at a pullback speed of 1.0 mm/s. The x-axis represents the mean area measured by the experimental and control devices (PU303018 vs. Eagle Eye Platinum for the carotid artery, and PU8015 vs. Visions PV.035 for the veins), while the y-axis represents the difference between the two devices (Experimental−Control). The solid black line indicates the mean bias (0.83 mm2), and the red dashed lines represent the 95% limits of agreement (LoA: −4.65 to 6.32 mm2). Data points are color-coded by animal ID (orange for P-25-705, blue for P-25-703, and green for P-25-694) to visually illustrate the hierarchical data structure and inter-animal variability.
Both the experimental device and the control device complete IVUS imaging of the predetermined target vessels. IVUS images suitable for evaluation are obtained for the carotid artery, renal artery, popliteal vein, subclavian vein, brachiocephalic vein, superior vena cava, and jugular vein. The representative IVUS images of the target vessels before and after stent implantation are shown in Figure 15 and Figure 16. The images show that both the experimental device and the control device display the lumen morphology and stent structure before and after stent implantation at the corresponding vessel sites, and both meet the image quality evaluation requirements. The stent-implanted vessel segments involved in the efficacy evaluation use stents with different diameter specifications, including 6 mm, 7 mm, 10 mm, 12 mm, and 14 mm. Stent size is selected according to the vessel dimensions measured by angiography and IVUS imaging, thereby covering evaluation scenarios for IVUS image quality and quantitative measurement after implantation of stents of different sizes.
Figure 15.
Representative IVUS images acquired with the experimental device before and after stent implantation: (a) carotid artery before stent implantation, (b) carotid artery after stent implantation, (c) subclavian vein before stent implantation, (d) subclavian vein after stent implantation, (e) popliteal vein before stent implantation, (f) popliteal vein after stent implantation.
Figure 16.
Representative IVUS images acquired with the control device before and after stent implantation: (a) carotid artery before stent implantation, (b) carotid artery after stent implantation, (c) subclavian vein before stent implantation, (d) subclavian vein after stent implantation, (e) popliteal vein before stent implantation, (f) popliteal vein after stent implantation.
The evaluation results for the excellent and good rates of IVUS image quality are shown in Table 10. As shown in Table 10, for the carotid artery, popliteal vein, jugular vein, and subclavian vein, the lumen/intimal interface, external elastic membrane (EEM) interface, and anatomical structures of vascular branches are clearly imaged and easily identifiable before and after stent implantation. After stent implantation, the stent struts and their positional relationship with the vessel lumen are clearly displayed and easily identifiable. The excellent and good image quality rates for both the PU303018 and PU8015 test catheters are 100%, with no obvious difference observed between the experimental and control devices. Since both groups achieve an excellent and good image quality rate of 100%, the data show complete consistency. Consequently, Fisher’s exact test detects no variability, indicating no statistically significant difference in the distribution of “excellent” and “good” ratings between the two groups. The evaluation results show that the experimental device can display the lumen boundary, intimal structure, external elastic membrane (EEM) interface (in arteries), vascular branch anatomy, and stent strut morphology, and can reflect the positional relationship between the stent and the vessel lumen. Under all pullback speed conditions, no difference in image quality grade is observed between the experimental device and the control device.
Table 10.
Evaluation results of excellent and good IVUS image quality rates.
3.5. Results of the Safety Evaluation
The recorded hemodynamic and body-temperature summaries are shown in Table 11, Table 12 and Table 13, with the reported exploratory within-procedure comparisons in Table 14 and Table 15. These results show that all vital signs of the experimental animals fluctuate within the normal range during the surgical procedure. In the efficacy group, comparisons of vital signs at different time points (during pullback at different speeds, after the final imaging, and before the initial imaging) show no significant differences in body temperature, heart rate, systolic blood pressure, diastolic blood pressure, or oxygen saturation (p > 0.05). In the safety group, comparisons of vital signs at the same time points reveal statistically significant differences in body temperature and invasive arterial pressure at specific time points (during or after pullback compared with before pullback). However, the invasive arterial pressure shows no abnormal fluctuations and remains within the clinically acceptable range. These slight fluctuations are mainly affected by deep anesthesia in the animals and are unrelated to the test device. At other speeds, no significant differences are observed in body temperature, heart rate, systolic blood pressure, diastolic blood pressure, or oxygen saturation (p > 0.05). These results indicate that the body temperature, heart rate, systolic blood pressure, diastolic blood pressure, and oxygen saturation of the animals are relatively stable during the surgical procedure, and the test catheters do not adversely affect the vital signs of the animals.
Table 11.
Intraoperative animal vital signs (Efficacy group—Before stent implantation/Preoperative).
Table 12.
Intraoperative animal vital signs (Efficacy group—After stent implantation/Postoperative).
Table 13.
Intraoperative animal vital signs (Safety group).
Table 14.
Statistical analysis of significant p values for intraoperative animal vital signs (Efficacy group).
Table 15.
Intraoperative animal vital signs (Safety group—After stent implantation/Postoperative).
Preclinical evaluation of vascular interventional devices usually requires attention to vascular patency, dissection, perforation, thrombosis, and device-related adverse events [40,43]. Representative preprocedural and immediate postprocedural DSA images are shown in Figure 17. Immediate postoperative follow-up angiography of the experimental animals, compared with preoperative images, reveals no abnormalities such as vascular dissection, thrombosis, perforation, or aneurysm. During the surgical procedures, no adverse events such as vascular occlusion, perforation, dissection, or death are observed in any animal, and the incidence of catheter-related adverse events is 0%. Furthermore, no unreasonable risk events that endanger health and safety, such as device mislabeling, catheter quality issues, or system failures, are observed during the procedures, and the device defect rate is 0%.
Figure 17.
Representative DSA images: (a) right carotid artery preoperatively; (b) right carotid artery immediately postoperatively; (c) right jugular vein preoperatively; (d) right jugular vein immediately postoperatively.
Histopathological evaluation is an important component of preclinical safety evaluation of vascular interventional devices [40,41,43,44]. The six experimental animals included in this experiment underwent immediate postoperative re-examination and were then euthanized for gross anatomy. The observation results show that, in terms of efficacy and safety, no obvious abnormalities such as infection or injury related to the experimental device were observed in the major organs (heart, liver, spleen, lungs, bilateral kidneys, and brain) or the treated vessels of the experimental animals in either group.
The histopathological analysis results for the different blood vessels are as follows:
Renal artery: Mild inflammatory cell infiltration is observed in only one vascular section. No thrombosis is found in the lumen of any vessel. Mild vessel wall injury is observed between the intima and the internal elastic lamina in some (6/9) vessel sections.
Carotid artery: No inflammatory cell infiltration or vessel wall injury is observed in any vessel, and no thrombosis is found in the lumen of any vessel.
Popliteal vein: Mild inflammatory cell infiltration and mild vessel wall injury (located between the intima and the internal elastic lamina) are observed in only one vascular section. No thrombosis is found in the lumen of any vessel.
Subclavian vein: No inflammatory cell infiltration or vessel wall injury is observed in any vessel, and no thrombosis is found in the lumen of any vessel.
Superior vena cava: No inflammatory cell infiltration or vessel wall injury is observed in any vessel, and no thrombosis is found in the lumen of any vessel.
Brachiocephalic vein: Mild inflammatory cell infiltration is observed in individual vessels, and no thrombosis is found in the lumen of any vessel. Mild vessel wall injury is observed in only one vascular section, with the injury located between the intima and the internal elastic lamina.
Jugular vein: No inflammatory cell infiltration or vessel wall injury is observed in any vessel, and no thrombosis is found in the lumen of any vessel.
The mild injury and inflammatory response mainly result from slight damage to endothelial cells of the vessel wall caused by device manipulation, such as guidewire delivery, during access establishment. This response is mild, self-healing, and relatively common, and is unrelated to the experimental device.
The three animals in the safety group are euthanized after immediate postoperative re-examination, and the target vessels are harvested for histopathological examination. The target vessels are shown in Figure 18, and the histopathological sections of the target vessels are shown in Figure 19. Figure 19a shows the proximal segment of the left renal artery, with black arrows indicating vessel wall injury; Figure 19b shows the proximal segment of the right carotid artery, with no obvious abnormalities; Figure 19c shows the distal segment of the left popliteal vein, with no obvious abnormalities; and Figure 19d shows the proximal segment of the right subclavian vein, with no obvious abnormalities.
Figure 18.
The representative anatomical images of the experimental blood vessels: (a) renal artery, (b) carotid artery, (c) popliteal vein, (d) subclavian vein.
Figure 19.
The representative histopathological sections of the experimental blood vessels (H&E staining, original magnification ×40): (a) renal artery, (b) carotid artery, (c) popliteal vein, (d) subclavian vein.
4. Conclusions and Limitations
In this manuscript, a mechanically rotating IVUS catheter family with 30 MHz and 15 MHz acoustic designs and three guidewire configurations is implemented on the V20i platform. Simulations and measurements of the piezoelectric ultrasonic transducers in the IVUS catheters show that the high-frequency catheter has a focal depth of approximately 3 mm, while the low-frequency catheter has a focal depth of approximately 10 mm, as characterized in water. In wire-phantom experiments, PU303018 achieves an imaging radius ≥10 mm and an axial resolution ≤100 μm, and PU8015 achieves an imaging radius ≥20 mm and an axial resolution ≤300 μm. Animal experiments evaluate the acute imaging performance of the proposed IVUS catheters in different vascular sites: PU303018 is used for the carotid and renal arteries, and PU8015 is used for the popliteal vein, subclavian vein, jugular vein, brachiocephalic vein, and superior vena cava; stents with diameters of 6–14 mm are used in the stent-implanted vessel segments. The mean stent cross-sectional area obtained by two commercial IVUS catheters (Eagle Eye Platinum, Visions PV.035) is used as the reference for evaluating the measurement results of the proposed IVUS catheters at pullback speeds of 1 mm/s, 3 mm/s, 6 mm/s, and 10 mm/s. The relative errors corresponding to the four speeds are (4.28 ± 1.67)%, (3.33 ± 2.58)%, (3.80 ± 2.55)%, and (2.73 ± 2.05)%, respectively. The measurement error does not show an obvious upward trend with increasing pullback speed. This indicates that the proposed IVUS catheters maintain relatively stable stent cross-sectional area measurement performance at different automatic pullback speeds under the experimental conditions. Image quality evaluation further verifies the in vivo imaging capability of the proposed IVUS catheters. This research uses the lumen/intimal interface, vascular branch structures, stent struts, and the positional relationship between the stent and the lumen as the main evaluation contents; the image quality of all indicators is good and basically consistent with that of the commercial IVUS catheters. Safety results show that all six experimental animals successfully complete the experiment, with no device-related adverse events; postoperative DSA re-examination finds no vascular dissection, perforation, thrombosis, air embolism, or significant blood flow slowing. In histopathological examination, only mild vessel wall injury and slight inflammatory response are observed locally in individual vessels; considering the experimental procedure, these are more likely related to mechanical stimulation caused by vascular access establishment and guidewire/catheter manipulation; no thrombosis, necrosis, or severe inflammatory response is observed in any target vessel.
This manuscript has several limitations. The most notable limitation is the small animal sample size (ns = 6). Because the animals are divided into an efficacy group (ns = 3) and a safety group (ns = 3), the statistical power of the comparative analyses is limited. Consequently, the findings of this manuscript should be interpreted as preliminary feasibility data rather than definitive evidence of clinical superiority. Future studies with larger cohorts are required to validate these results. The direct comparison between the experimental and control devices is limited by their inherent physical differences, particularly the mismatched center frequencies (30/15 MHz vs. 20/10 MHz) and different system architectures. While this comparison reflects a pragmatic clinical evaluation of the system’s applicability, it cannot isolate the specific contribution of frequency versus transducer design or image processing to the final image quality. Future studies with strictly frequency-matched controls would be beneficial to delineate these factors. Another significant limitation of this manuscript is the use of healthy porcine vessels. Peripheral artery disease is typically characterized by complex atherosclerotic plaques, calcification, and severe stenosis, which can significantly alter ultrasound penetration and image interpretation. Therefore, the current study does not establish the lesion identification or measurement performance of the proposed IVUS system in diseased vessels. Future studies employing validated animal models with induced atherosclerosis or calcification are necessary to fully assess the clinical applicability of this system.
Author Contributions
Conceptualization, T.Z.; methodology, T.Z.; software, Q.L.; validation, T.Z., Q.L. and S.J.; formal analysis, Q.L.; investigation, S.J.; resources, T.Z.; data curation, T.Z.; writing—original draft preparation, T.Z.; writing—review and editing, Q.L. and Z.W.; visualization, T.Z.; supervision, X.L. and Z.W.; project administration, S.J.; funding acquisition, X.L. All authors have read and agreed to the published version of the manuscript.
Funding
This research was supported by Hubei Medical Devices Quality Supervision and Test Institute, grant number SMDTKL-2025-2-01.
Institutional Review Board Statement
The animal study protocol was approved by the Experimental Animal Welfare and Ethics Committee of Shanghai Xinova Medical Research Co., Ltd. (protocol code XNM-YX-20251121-01 and date of approval 21 November 2025). The animal experiments were commissioned to and performed by Shanghai Xinova Medical Research Co., Ltd. in accordance with the approved protocol.
Informed Consent Statement
Not applicable.
Data Availability Statement
The original contributions presented in this study are included in the article. Further inquiries can be directed to the corresponding authors.
Acknowledgments
The authors wish to thank Shanghai Acoustic Life Science Co., Ltd., for the fabrication of the IVUS diagnostic device and IVUS catheters. The authors also wish to thank the Hubei Medical Devices Quality Supervision and Test Institute for supporting the measurement environment of the IVUS catheters.
Conflicts of Interest
All authors (Tao Zhou, Qian Li, Shiqi Jia, Xiang Li, and Zhipeng Wu) were employed by Acoustic Life Science Co., Ltd. The authors declare that the research was conducted in the absence of any other commercial or financial relationships that could be construed as potential conflicts of interest.
References
- Søgaard, M.; Golledge, J. Peripheral arterial disease: Beyond prevalence estimates to equitable care. Lancet Glob. Health 2026, 14, 104004. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Criqui, M.H.; Matsushita, K.; Aboyans, V.; Hess, C.N.; Hicks, C.W.; Kwan, T.W.; McDermott, M.M.; Misra, S.; Ujueta, F. Lower extremity peripheral artery disease: Contemporary epidemiology, management gaps, and future directions: A scientific statement from the American Heart Association. Circulation 2021, 144, e171–e191. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Narula, N.; Dannenberg, A.J.; Olin, J.W.; Bhatt, D.L.; Johnson, K.W.; Nadkarni, G.; Min, J.; Torii, S.; Poojary, P.; Anand, S.S.; et al. Pathology of peripheral artery disease in patients with critical limb ischemia. J. Am. Coll. Cardiol. 2018, 72, 2152–2163. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Collins, G.C.; Brumfiel, T.; Bercu, Z.L.; Desai, J.P.; Lindsey, B.D. Dual-resonance (16/32 MHz) piezoelectric transducer with a single electrical connection for forward-viewing robotic guidewire. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2022, 69, 1428–1441. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Loffroy, R.; Falvo, N.; Galland, C.; Fréchier, L.; Ledan, F.; Midulla, M.; Chevallier, O. Intravascular ultrasound in the endovascular treatment of patients with peripheral arterial disease: Current role and future perspectives. Front. Cardiovasc. Med. 2020, 7, 551861. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Secemsky, E.A.; Parikh, S.A.; Kohi, M.; Lichtenberg, M.; Meissner, M.; Varcoe, R.; Holden, A.; Jaff, M.J.; Chalyan, D.; Clair, D.; et al. Intravascular ultrasound guidance for lower extremity arterial and venous interventions. EuroIntervention 2022, 18, 598–608. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Secemsky, E.A.; Aronow, H.D.; Kwolek, C.J.; Meissner, M.; Muck, P.E.; Parikh, S.A.; Winokur, R.S.; George, J.C.; Salazar, G.; Murphy, E.H.; et al. Intravascular ultrasound use in peripheral arterial and deep venous interventions: Multidisciplinary expert opinion from SCAI/AVF/AVLS/SIR/SVM/SVS. J. Vasc. Interv. Radiol. 2024, 35, 335–348. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Shammas, N.W.; Radaideh, Q.; Shammas, W.J.; Daher, G.E.; Rachwan, R.J.; Radaideh, Y. The role of precise imaging with intravascular ultrasound in coronary and peripheral interventions. Vasc. Health Risk Manag. 2019, 15, 283–290. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Arthurs, Z.M.; Bishop, P.D.; Feiten, L.E.; Eagleton, M.J.; Clair, D.G.; Kashyap, V.S. Evaluation of peripheral atherosclerosis: A comparative analysis of angiography and intravascular ultrasound imaging. J. Vasc. Surg. 2010, 51, 933–939. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Peng, C.; Wu, H.; Kim, S.; Dai, X.; Jiang, X. Recent advances in transducers for intravascular ultrasound (IVUS) imaging. Sensors 2021, 21, 3540. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Marteslo, J.P.; Makary, M.S.; Khabiri, H.; Flanders, V.; Dowell, J.D. Intravascular ultrasound for the peripheral vasculature—Current applications and new horizons. Ultrasound Med. Biol. 2020, 46, 216–224. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Khedr, M.; Megaly, M.; Elgendy, I.Y. Intravascular ultrasound in endovascular interventions for peripheral artery disease. Curr. Treat. Options Cardiovasc. Med. 2023, 25, 347–358. [Google Scholar] [CrossRef] [Scilit]
- Secemsky, E.A.; Mosarla, R.C.; Rosenfield, K.; Kohi, M.; Lichtenberg, M.; Meissner, M.; Varcoe, R.; Holden, A.; Jaff, M.R.; Chalyan, D.; et al. Appropriate use of intravascular ultrasound during arterial and venous lower extremity interventions. JACC Cardiovasc. Interv. 2022, 15, 1558–1568. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Soney, H.; Kakkilaya, A.; Vazquez, D.F.; Banerjee, R.; Rosol, Z.; Tsai, S.; Banerjee, S. Reproducibility of femoropopliteal artery intravascular ultrasound imaging in patients with peripheral artery disease. Am. J. Cardiol. 2023, 199, 1–6. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Divakaran, S.; Parikh, S.A.; Hawkins, B.M.; Chen, S.; Song, Y.; Banerjee, S.; Rosenfield, K.; Secemsky, E.A. Temporal trends, practice variation, and associated outcomes with IVUS use during peripheral arterial intervention. JACC Cardiovasc. Interv. 2022, 15, 2080–2090. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Tsujimura, T.; Iida, O.; Takahara, M.; Soga, Y.; Yamaoka, T.; Fujihara, M.; Kawasaki, D.; Ichihashi, S.; Kozuki, A.; Nanto, S.; et al. Clinical impact of intravascular ultrasound-guided fluoropolymer-based drug-eluting stent implantation for femoropopliteal lesions. JACC Cardiovasc. Interv. 2022, 15, 1569–1578. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Kumar, A.; Shariff, M.; Majmundar, M.; Stulak, J.M.; Anavekar, N.; Deshmukh, A.; Bashir, R. Intravascular ultrasound during endovascular intervention for peripheral artery disease, by severity, location, device, and procedure. Am. J. Cardiol. 2024, 225, 41–51. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Marulanda, K.; Genovese, E.A. Adjunctive utilization of intravascular ultrasound in peripheral arterial disease treatment. Ann. Vasc. Surg. 2024, 107, 195–207. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Guan, X.; Han, H.; Xu, H. Cutting-edge progress of intravascular ultrasound in lower-extremity vascular interventions. Ultrasonography 2025, 44, 5–18. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lee, S.-J.; Kim, T.-H.; Lee, J.-H.; Ahn, C.-M.; Lee, S.-H.; Lee, Y.-J.; Kim, B.-K.; Hong, M.-K.; Jang, Y.; Park, H.-W.; et al. Intravascular ultrasound-guided vs angiography-guided drug-coated balloon angioplasty in patients with complex femoropopliteal artery disease. JACC Cardiovasc. Interv. 2025, 18, 558–569. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Janjic, J.; Tan, M.; Daeichin, V.; Noothout, E.; Chen, C.; Chen, Z.; Chang, Z.-Y.; Beurskens, R.H.S.H.; van Soest, G.; van der Steen, A.F.W.; et al. 2D ultrasound transducer with front-end ASIC and low cable count for 3D forward-looking intravascular imaging: Performance and characterization. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2018, 65, 1832–1844. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Cabrera-Munoz, N.E.; Eliahoo, P.; Wodnicki, R.; Jung, H.; Chiu, C.T.; Williams, J.A.; Kim, H.H.; Zhou, Q.; Shung, K.K. Forward-looking 30-MHz phased-array transducer for peripheral intravascular imaging. Sens. Actuators A Phys. 2018, 280, 145–163. [Google Scholar] [CrossRef] [Scilit]
- Wang, J.; Zheng, Z.; Chan, J.; Yeow, J.T.W. Capacitive micromachined ultrasound transducers for intravascular ultrasound imaging. Microsyst. Nanoeng. 2020, 6, 73. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Shih, C.-C.; Chen, P.-Y.; Ma, T.; Zhou, Q.; Shung, K.K.; Huang, C.-C. Development of an intravascular ultrasound elastography based on a dual-element transducer. R. Soc. Open Sci. 2018, 5, 180138. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Hong, J.; Su, M.; Yu, Y.; Zhang, Z.; Liu, R.; Huang, Y.; Mu, P.; Zheng, H.; Qiu, W. A dual-mode imaging catheter for intravascular ultrasound application. IEEE Trans. Med. Imaging 2019, 38, 657–663. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Philips. Visions PV.035 Digital IVUS Catheter. Available online: https://www.usa.philips.com/healthcare/product/HCIGTD88901/visions-pv-035-digital-ivus-catheter (accessed on 21 September 2026).
- Philips. Visions PV.018 Digital IVUS Catheter. Available online: https://www.usa.philips.com/healthcare/product/HCIGTD86700/visions-pv-018-digital-ivus-catheter (accessed on 8 November 2026).
- Philips. Eagle Eye Platinum Digital IVUS Catheter. Available online: https://www.usa.philips.com/healthcare/product/HC85900P/eagle-eye-platinum-digital-ivus-catheter (accessed on 21 September 2026).
- Krejza, J.; Arkuszewski, M.; Kasner, S.E.; Weigele, J.; Ustymowicz, A.; Hurst, R.W.; Cucchiara, B.L.; Messe, S.R. Carotid artery diameter in men and women and the relation to body and neck size. Stroke 2006, 37, 1103–1105. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Mahlon, M.A.; Yoon, H.-C. CT angiography of the superior vena cava: Normative values and implications for central venous catheter position. J. Vasc. Interv. Radiol. 2007, 18, 1106–1110. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Saleem, T.; Raju, S. Comparison of intravascular ultrasound and multidimensional contrast imaging modalities for characterization of chronic occlusive iliofemoral venous disease: A systematic review. J. Vasc. Surg. Venous Lymphat. Disord. 2021, 9, 1545–1556.e2. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Montminy, M.L.; Thomasson, J.D.; Tanaka, G.J.; Lamanilao, L.M.; Crim, W.; Raju, S. A comparison between intravascular ultrasound and venography in identifying key parameters essential for iliac vein stenting. J. Vasc. Surg. Venous Lymphat. Disord. 2019, 7, 801–807. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ulloa, J.G.; Gelabert, H.A.; O’Connell, J.B.; Patel, R.; Rigberg, D.A. Intravascular ultrasonography provides more sensitive detection of subclavian vein stenosis than venography in patients presenting with Paget-Schroetter syndrome. J. Vasc. Surg. Venous Lymphat. Disord. 2021, 9, 1145–1150.e1. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- McDaniel, M.C.; Eshtehardi, P.; Sawaya, F.J.; Douglas, J.S., Jr.; Samady, H. Contemporary clinical applications of coronary intravascular ultrasound. JACC Cardiovasc. Interv. 2011, 4, 1155–1167. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- CTS Corporation. Piezoelectric Products. Available online: https://www.ctscorp.com/Products/Piezoelectric (accessed on 21 September 2026).
- YY/T 1142-2013; Methods for Measuring the Frequency of Medical Ultrasonic Equipment and Probes. Standards Press of China: Beijing, China, 2013.
- IEC 62127-2; Ultrasonics—Hydrophones—Part 2: Calibration for Ultrasonic Fields up to 40 MHz. IEC: Geneva, Switzerland, 2007.
- YY/T 1659-2019; General Technical Requirements for Intravascular Ultrasound Diagnostic Equipment. Standards Press of China: Beijing, China, 2019.
- Percie du Sert, N.; Hurst, V.; Ahluwalia, A.; Alam, S.; Avey, M.T.; Baker, M.; Browne, W.J.; Clark, A.; Cuthill, I.C.; Dirnagl, U.; et al. The ARRIVE guidelines 2.0: Updated guidelines for reporting animal research. PLoS Biol. 2020, 18, e3000410. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Perkins, L.E.L.; Rippy, M.K. Balloons and stents and scaffolds: Preclinical evaluation of interventional devices for occlusive arterial disease. Toxicol. Pathol. 2019, 47, 297–310. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- An, S.; Kim, J.; Lee, D.; Kim, M.; Byun, K.; Yun, J.; Jeong, W.; Son, D. Biocompatibility and patency of a novel titanium vascular anastomotic device in a pig jugular vein. Sci. Rep. 2021, 11, 17512. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- National Medical Products Administration. Medical Device Animal Study Registration Review Guiding Principles Part II: Study Design and Quality Assurance; Announcement No. 75 of 2021; NMPA: Beijing, China, 2021; Available online: https://www.cmde.org.cn/flfg/zdyz/fbg/fbgwy/20210929091828268.html (accessed on 8 November 2026).
- Perkins, L.E.L.; Tu, M. Preclinical evaluation of vascular closure devices. Front. Cardiovasc. Med. 2024, 11, 1502909. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- ISO 10993-1:2018; Biological Evaluation of Medical Devices—Part 1: Evaluation and Testing Within a Risk Management Process. ISO: Geneva, Switzerland, 2018.
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |