Voltammetric pH Determination and Water Quality Screening—A Novel Method Based on Iridium Dioxide Electrodes

and
Department of Mechanical Engineering and Environmental Technology, University of Applied Sciences Amberg-Weiden (OTH), Kaiser-Wilhelm-Ring 23, 92224 Amberg, Germany
*
Author to whom correspondence should be addressed.

Abstract

Water quality is one of the United Nations’ sustainable development goals. In the face of climate change, early-warning systems for ecosystem tipping points are becoming increasingly important. For online monitoring of environmental pollution, the pH value serves as a reliable parameter that provides valuable insights into water treatment and processing. This article reviews low-cost, miniaturised alternatives to the widely used glass electrode, and outlines novel methods of pH determination using transient signals rather than established potentiometry. The tutorial demonstrates how to measure pH and conductivity using voltammetric redox peaks of IrO2 electrodes. It shows how solution resistance can be derived from cyclic voltammograms without the need for separate conductivity measurements. Finally, the impacts of electrode ageing and calibration are illustrated, based on experimental results. The outlook expands the current understanding of practical pH—a value based on reference buffer solutions—and highlights the challenges associated with future coulometric pH measurement based on proton counting.

1. Introduction

In August 2022, large quantities of dead fish were found along the Oder River in Germany and Poland. The cause was poisoning resulting from an algal bloom, the rapid growth of which was promoted by increases in pH, salinity and water temperature. The first indication of the event was a rise in conductivity at a real-time monitoring station in Frankfurt (Oder). Three days later, an increase in chlorophyll-a concentration from 20 to 140 µg L–1 was detected, pointing to a massive phytoplankton bloom (Prymnesium parvum). One recommendation for preventing future incidents of this kind was the expansion of continuous, publicly accessible, and precise monitoring systems, coupled with mandatory communication of pollution events across the borders of international river basin districts [1,2].
The 2030 Agenda for Sustainable Development, published by the United Nations in 2015, is an action plan for people, the planet, and prosperity. Goal 6 focuses on ensuring the safe and sustainable use of water and reducing water pollution [3]. Adequate wastewater and water treatment are becoming increasingly important in highly industrialised and urbanised areas [4]. To measure the success of these methods, the UN introduced indicator 6.3.2; this encompasses the parameters of dissolved oxygen, salinity, total oxidised nitrogen and nitrate, orthophosphate and pH [5]. The European Union mandates regular monitoring of drinking water quality under EU Directive 2020/2184 [6]. Routine monitoring covers colour, turbidity, taste, odour, microbiological parameters, pH and conductivity.

1.1. Developments Beyond the Glass Electrode

The glass electrode is widely considered the gold standard for pH determination [7]. Its operation relies on measuring the potential difference across a thin glass membrane that is in contact with the electrolyte under investigation on its outer surface and with a reference electrolyte on its inner surface. When hydrated, the membrane forms a swollen silicate gel layer capable of exchanging sodium ions for protons from the external electrolyte. The resulting interfacial potential follows the Nernst equation (E ~ pH). In practice, however, an asymmetry potential arises in the zero-current state, alongside an unknown diffusion potential at the reference electrode’s glass frit. Therefore, a linear two-point calibration with pH standard buffer solutions is performed. This means that the glass electrode does not allow for an accurate measurement of the absolute pH value, but, rather, relies on a practical pH scale based on buffer solutions.
Despite its widespread use, the glass electrode exhibits fundamental limitations that complicate its application in automated, continuous measurement systems [8]: (i) At pH values above 9, an alkaline error occurs; rising Na+ concentrations affect the interfacial potential. In general, glass electrodes do not yield accurate readings in highly concentrated and highly dilute solutions. (ii) The high impedance of glass electrodes requires high-impedance amplifier circuits for measurement. (iii) Due to brittle glass membranes, there is a risk of mechanical failure, particularly in outdoor applications. (iv) Hydrofluoric acid attacks silicate glasses, rendering the glass electrode unsuitable for such solutions. (v) A lack of miniaturisation hinders the integration of the electrode into compact sensor platforms. (vi) Electrode slope and standard potential are subject to drift and ageing processes over time, requiring regular recalibration.
The limitations of the glass electrode have driven the development of solid-state pH sensors since the 1970s. To date, no single Chem-FET sensor has been fully convincing in terms of robustness, suitability for miniaturisation, cost-efficiency, and accuracy. Simple disposable sensors are being considered for medical analysis and water monitoring.
The pH-dependent redox processes of metal oxides [9] are particularly suitable for this purpose, as they can be miniaturised [10,11] and applied in hydrofluoric acid solutions [12]. Near-Nernstian behaviour with respect to pH and chemical stability across the entire pH range has been reported for TiO2, RuO2, RhO2, SnO2, Ta2O5, PtO2-IrO2 [13,14], Bi2O3 [15], and other metal oxides. A typical sensor consists of a working electrode and a reference electrode; for instance, a screen-printed Ag|AgCl electrode. pH calibration is performed via linear regression of the open-circuit potential (OCP) in standard buffer solutions.

1.2. Voltammetric pH Measurement and Proton Counting

Most of the reported sensors measure the open-circuit potential (OCP) [16]. The shift in the Ir(III/IV) redox potential peak in the cyclic voltammogram was identified early as a pH-sensitive signal [17]. Nevertheless, the methodology has not been sufficiently investigated to date [18,19].
The voltammetric method offers several advantages over OCP measurement: (i) The possibility of internal calibration [20] by utilizing the difference between the Ir(III/IV) peak and the hydrogen adsorption peaks on iridium, e.g., in seawater, at pH 7.5 to 8.5. (ii) The pH sensitivity and standard potential of RuO2 and IrO2 electrodes are influenced by the conductivity and interference from salt ions [21]. Consequently, such a sensor would require calibration tailored to the matrix, constant ionic strength conditions, or simultaneous conductivity measurement [22].
This paper provides a tutorial to voltammetric pH measurement, focusing on IrO2. It discusses possible sources of error, and addresses the effects of electrolyte conductance and electrode aging. We demonstrate that the shift in the Ir(III/IV) peak can be used to simultaneously determine both pH and electrolyte conductivity, thereby directly capturing two parameters required for water quality monitoring.
Cyclic voltammetry (CV) is likely suitable as a method for “proton counting.” However, despite numerous attempts over a period of three decades [23,24], we have not yet succeeded in quantitatively deriving the exact pH value based on peak heights or the electrical charge.

2. Iridium Dioxide as a pH Probe

2.1. From Redox Chemistry to pH Determination

The thermodynamically stable phases at a metallic iridium electrode in aqueous solutions, as shown in the Pourbaix diagram (Figure 1), indicate oxidation states of 0, II, III, IV, and VI. The theoretical open-circuit potential is determined by the Ir(III/IV) redox couple at E0 = 926 mV (Equation (1)).
I r 2 O 3 + H 2 O ⇌ 2 I r O 2 + 2 H + + 2 e −
Figure 1. Potential-pH equilibrium diagram for the iridium–water system, at 25 °C. Modified after Pourbaix [25]. Line ③: Ir + 2 H2O ⇌ IrO2 + 4 H+ + 4 e– (E/V = 0.926 − 0.0591⋅pH). Line (a): hydrogen evolution; line (b): oxygen evolution; line ①: limit of the domain of relative predominance of Ir(III).
Depending on the water content, however, an iridium(III)-oxide surface may be conceived as a type of hydroxide, 2 Ir(OH)3 ≡ Ir2O3 ⋅ 3 H2O ≡ 2 IrO(OH) + 2 H2O, allowing the redox equation to be simplified (Equation (2)).
I r O ( O H ) ⇌ I r O 2 + H + + e −
Several decades ago, Gottesfeld (1978) [26] formulated Equation (3), which was used by Kötz [27] to model the oxidative deprotonation of the hydrated iridium oxide layer involving proton-coupled electron transfer (Equations (4) and (5))—a process that precedes the oxygen evolution reaction (OER).
I r ( O H ) n ⇌ I r O x ( O H ) n − x + x H + + x e −
I r I I I ( O H ) 3 ⇌ I r I V O ( O H ) 2 + H + + e −
I r I V O ( O H ) 2 ⇌ I r V O 2 ( O H ) + H + + e −
Later, operando measurements [28] revealed a maximum oxidation state of Ir(V), independent of whether the IrO2 was generated electrochemically or thermally.
The change in oxidation state is coupled to the uptake or release of protons; consequently, the equilibrium potential depends on the pH value [29]. In the Nernst Equation (6), E0 represents the standard potential (in V versus the standard hydrogen electrode at 25 °C, 101,325 Pa), R = 8.3144 J mol–1K–1 is the molar gas constant, F = 96,485 C mol–1 is the Faraday constant, T the absolute temperature (in K), and a is the activity of the oxidizing or reducing species (dimensionless value; effective concentration divided by 1 mol L–1).
E = E 0 − R T F ln a ( I r I I I ) a ( I r I V ) ⋅ a ( H + )
With regard to Equation (2), where the charge number of the electrode reaction z = 1, a(IrIII) = a(IrIV), and T = 298 K, the factor dE/dpH = −59 mV is often referred to as the pH sensitivity, as it corresponds to the slope of the linear relationship between electrode potential and pH value.
E = E 0 − RT ln 10 F p H = E 0 − 59.16   m V ⋅ p H
Since the exact standard potential E0 in the unknown solution is not known, in practice, absolute pH measurements are not possible; standard buffer solutions are required for calibration.

2.2. Voltammetric Peak Potential

In 1955, galvanostatic charging curves provided information on the amount of adsorbed hydrogen and oxygen on iridium wires in sulfuric acid [30]. The (probably) first cyclic voltammogram of an iridium wire, recorded in 1961, revealed the pH-dependent shift of the Ir(III/IV) redox peak in both acidic and alkaline solutions [31]. In the 1970s, metal dissolution and oxide-layer formation during potential cycling became obvious, and it was observed that the expected Nernstian slope of dE/dpH = −59 mV was exceeded.
The cyclic voltammograms in Figure 2a,b show fundamentally the same regions for an iridium rod and an IrO2 film.
Figure 2. Cyclic voltammograms of (a) metallic iridium wire after 50 voltammetric pre-cycles, at different scan rates (100 mV/s to 1000 mV/s), and (b) IrO2 film on titanium support, without pretreatment. Reference electrode: Ag|AgCl|3M KCl (0.2 V SHE), counter electrode: platinum sheet, electrolyte: 0.5 mol L−1 H2SO4. Green points: Ir(III/IV) oxidation peak, blue points: Ir(III/IV) reduction peak. (c) IrO2 film on a quartz crystal microbalance (20 mV/s). Mass changes during the Ir(III/IV) redox reaction around 0.9 V SHE (dotted lines and arrow). (d) Linear correlation of Ir(III/IV) peak current and scan rate. Illustrations based on the authors’ measurements [this work].
  • The reversible redox peaks of Ir(III/IV) and Ir(IV/V) [32] appear rather sharp on metallic iridium, whereas the broad signals for IrO2 reflect the heterogeneity of the oxide surface.
  • In the double-layer region, no redox reactions take place. Electric current is determined by the double-layer charging, and is significantly influenced by the electrode surface, particularly in the case of porous oxides.
  • Hydrogen underpotential deposition at low potentials (above 0 V SHE) occurs predominantly on metallic electrodes. The electric charge of the reduction peak reflects the electrochemically active surface area [33]—a relationship that, unfortunately, holds only approximately for oxides [34,35].
  • In the hydrogen (HER) or oxygen (OER) evolution region, respectively, the electrolyte undergoes irreversible decomposition. In the HER region, the IrO2 coating frequently detaches from the substrate.

2.3. Electrochemical Valence

An IrO2 electrode on conductive titanium tracks was characterized using cyclic voltammetry combined with simultaneous quartz-crystal microbalance measurements in Figure 2c. The initial IrO2 film mass was 0.1823 mg cm−2 for an electrode area of 1.287 cm2. The shift in the resonance frequency, described by the Sauerbrey equation [36], Δm = −Δf/C (where C = 56.6 Hz cm2 µg−1), indicates the mass change during the redox reaction.
The Ir(III/IV) current peak in Figure 2c was fitted using a Gaussian distribution, to determine the voltammetric charge Q. The mass change Δm was calculated by subtracting the values at two points, one before and one after the current peak. The most rapid mass change occurs in the region of 0.72 V vs. Ag/AgCl (0.92 V SHE). The measured average mass-to-charge ratio, Δm/ΔQ = 757.86 µg/C, corresponds almost exactly to the species [H(H2O)4]+, theoretical: M(H9O4+)/F = 757.30 µg/C. This confirms the number of exchanged electrons (z = 1) in the redox Equation (2). With this, the oxidation of an iridium centre could be explained by the dissociative adsorption of water and proton displacement.
[ I r I I I ] ⋅ H 2 O ⇌ [ I r I V ] O H + H + + e −

2.4. Voltammetric Peak Currents

The Ir(III/IV) transition is a surface-bound, proton-coupled, Faradaic reaction. The redox-active centres are immobilized within an IrO2 film, so that, unlike with a dissolved redox mediator, there is no mass transfer limitation of a dissolved species to the electrode. Thin films of surface-bound species display reversible, diffusion-free behaviour at low sweep rates. In contrast, thick hydrated IrO2 films exhibit pronounced diffusion-controlled kinetics in the cathodic process, determining the rate at high scan rates.
A linear plot of peak current versus the square root of scan rate indicates a freely diffusing species in solution. A surface-adsorbed species scales linearly with v, instead.
  • Surface-bound species: peak current increases with scan rate: I p ∝ v ; ideal peak separation: Δ E P = E P , o x − E p , r e d = 0 ; width at half-maximum: 90.6 z   m V [37].
  • Diffusion-controlled case: I P ∝ v 1 2 (Randles–Ševčík); ideal peak separation: Δ E P = 2.218 R T z F   ≈ 57 z m V ( 25 ° C ) (reversible transition) [38].
For the electrodes investigated in this study, the regime of surface-bound species is relevant (see Figure 2d). Thin films ensure short transport paths within the film, preventing the formation of a limiting concentration gradient. Consequently, the ideal peak separation for the system is 0 V, rather than 57/z mV. This value is crucial for the subsequent determination of the uncompensated resistance (see below).
If the concentration of iridium in the active layer were precisely known, absolute pH measurements could be performed using Faraday’s law: Q = F⋅ n(H+). Unfortunately, in practice, there is at best only a vague relationship between peak current and pH value (see Section 4.5). Furthermore, there must be an excess of active iridium surface sites that interact with protons from the solution.

2.5. The Problem of Electrolyte Conductivity

All electrolyte solutions exhibit a certain conductivity that depends on their ionic strength. While diluted acids and bases, salt solutions and seawater possess excellent-to-good conductivity, the conductivity of tap water is low. In a three-electrode cell configuration, the ohmic voltage drop across the uncompensated electrolyte resistance (Re) creates a discrepancy between the applied potential and the actual electrode potential, relative to the reference electrode. The same applies to the shift in anodic and cathodic peak potentials (EP).
E p , o x = E p , o x ∗ + I p , o x R e
E p , r e d = E p , r e d ∗ − | I p , r e d | ⋅ R e
Note the negative sign of the cathodic current: I p , o x > 0 and I p , r e d < 0 . The anodic peak potential shifts towards more positive values, while the cathodic potential shifts towards more negative values. The peak separation increases with rising ohmic voltage drop, whereas the mean value E1/2 of the peak potentials remains constant.
E 1 / 2 = E p , o x + E p , r e d 2 = E p , o x ∗ + E p , r e d ∗ 2 + R e 2 ( I p , o x − | I p , r e d | )
In the surface-bound, reversible case (I ~ v), the ideal half-wave potential E 1 / 2 ∗ = ( E p , o x ∗ + E p , r e d ∗ ) /2 coincides with the formal potential E0’ which depends on the pH value, owing to the proton-coupled nature of the reaction. The contribution of the ohmic voltage drop becomes zero precisely when the peak currents are equal in magnitude; this is the case for an ideal reversible surface-bound redox process. To assess the error in pH measurement caused by the uncompensated electrolyte resistance, it is useful to express the total current in terms of the peak current ratio. The closer the peak current ratio is to unity, the more the electrolyte resistance can be neglected.
E 1 / 2 = E 1 / 2 ∗ + I p , o x R e 2 ⋅ ( 1 − | I p , r e d | I p , o x )   ≈ E 1 / 2 ∗

2.6. Resistance Estimation Based on Peak Separation—A Novel Approach

The peak potentials of the oxidation reaction and the reduction reaction are separated by the distance ΔEP (Equation (13)).
Δ E p = E p , o x − E p , r e d = ( E p , o x ∗ − E p , r e d ∗ ) + R e ⋅ ( I p , o x + | I p , r e d | )
The ideal separation for the reversible, surface-bound redox process is Δ E P ∗   ( E p , o x ∗ − E p , r e d ∗ )   = 0. The kinetics of quasi-reversible electron transfer contribute a residual separation (ΔEP,kin); however, this is small—provided the heterogeneous reaction rate is high or the scan rate is low—and remains independent of solution resistance. If only the specific resistance of the solution is varied, only the ohmic potential drop changes, while the kinetic contribution remains as a constant intercept. Therefore, in the reversible limiting case (ΔEP,kin ≈ 0), the uncompensated electrolyte resistance simplifies to Equation (14):
R e = Δ E p − Δ E p , k i n I P , o x − I p , r e d ≈ E p , o x − E p , r e d I p , o x − I p , r e d
Practical applications are outlined in Section 5.

3. Electrode Fabrication and Oxide Formation

3.1. Coating Processes for IrO2 pH-Electrodes

Until the late 1970s, research on iridium oxide focused on coating technologies for electrolysis, redox chemistry, and electrochromic displays. Since 1980, the number of publications on IrO2 electrodes for pH sensors has steadily risen, to over 160 per year.
Over the years, various methods for producing active iridium oxide layers have been developed, as shown in Figure 3c.
Figure 3. (a) Extrapolated standard potential as a function of the reported Nernst slope; the latter were clustered hierarchically. Average value (in green): –56.9 mV pH–1, 848.7 mV SHE (anhydrous IrO2); –74.5 mV pH–1, 946.2 mV SHE (hydrous IrO2 ∙ xH2O). Median value (red): −58.5 mV pH–1, 875.5 mV SHE (anhydrous); –72.6 mV pH–1, 941.6 mV SHE (hydrous). See Appendix A. (b) Number of publications on IrO2 electrodes for pH sensing from 1980 to mid-2026. Right axis: this survey (arrow). (c) Absolute frequency of coating methods for anhydrous and hydrous IrO2: AIROF: anodic iridium oxide films; EIROF: electrodeposited iridium oxide films; SIROF: sputtered iridium oxide films; oxidized metallic iridium wires; pre-synthesized IrO2 particles; thermal decomposition of precursors (IrCl3, etc.).
The data in Appendix A can be classified into six methodical categories: (i) anodic iridium oxide films (26.3% of all publications), (ii) electrodeposited iridium oxide films (40.7% of all publications), (iii) sputtered iridium oxide films, (iv) metallic iridium treated thermally or by melt oxidation (15.4%), (v) pre-synthesized iridium oxide particles, and (vi) thermally prepared iridium oxide films (17.6%).
  • Originally, most electrodes were fabricated by the anodic iridium oxide film (AIROF) process [39,40]. Metallic iridium is oxidized via continuous voltammetric cycling. In the potential range between 0 V and 1.5 V SCE in 0.5-molar sulfuric acid at a scan rate of 150 mV/s, colour changes correspond to the redox state: Ir(OH)n is transparent, while IrOx(OH)n−x is dark (Gottesfeld [26]). Voltage pulses ranging from −0.25 V to 1.25 V at a frequency of 0.5 Hz correlate with the oxidation state and the number of pulses [41].
  • Reactive sputtering was introduced in 1979. In this process, high-energy ions from an oxygen-containing discharge eject atoms from an iridium target; these atoms subsequently deposit as a thin IrO2 layer, onto a substrate. Sputtered iridium oxide films (SIROFs) exhibit superior stability during oxygen evolution compared to AIROFs. Furthermore, the Ir(III/IV) redox peak in 0.5 mol L−1 H2SO4 appears at 0.80 V RHE, instead of 0.95 V RHE [42].
  • In the case of thermally prepared iridium oxide films, the particles form directly on the electrode from a precursor. In 1981, Trasatti and co-workers [43] prepared dimensionally stable anodes (DSAs) using this method. IrCl3 is dissolved in water or isopropanol, applied to a titanium substrate, dried at 60–70 °C, and subsequently sintered at temperatures between 300 °C and 550 °C. The best agreement with Pourbaix theory (59 mV pH−1 and E0 = 0.95 V SHE) was found at 400 °C.
    The Pechini method employs a sol–gel synthesis based on polymer precursors. IrCl3 is added to a mixture of citric acid and ethylene glycol (ratio 1:4.65) at 60 °C. The resulting solution is applied to the substrate and subjected to heat treatment (110 °C for 30 min, 250 °C for 20 min, 400 °C for 10 min) [44,45]. IrCl4 [46], H2IrCl6 or isopropoxide can also be used as precursors.
  • Thermal oxidation and melt oxidation of iridium. The thermal oxidation of iridium wires—achieved by heat-treating at 800 °C in an oxygen-rich atmosphere following immersion in 2 mol L–1 NaOH (Hitchman [12])—yields a material resistant to hydrofluoric acid. In the melt oxidation process, the iridium wire is oxidized for 5 h at 870 °C in an aluminium crucible lined with gold foil and covered with Li2CO3 powder; subsequently, the solid carbonate is removed using diluted HCl, and the oxidized wire is dried overnight at 120 °C [47]. The pH-sensitivity of such oxidized iridium is different from AIROFs.
  • Electrodeposited iridium oxide films (EIROFs) were developed by Yamanaka in 1989 [48]: 0.15 g IrCl4⋅H2O (dissolved in 100 mL water), 1 mL of aqueous H2O2 (30 wt.%), and 0.5 g of oxalic acid (C2H2O2·H2O) are added sequentially to 100 mL of water; subsequently, the solution is adjusted to pH 10.5, using anhydrous K2CO3. After two days, anodic deposition (1.6–2 A/m2, at 15 °C) yields a thin film (approximately 0.1 µm) with high cycling stability (>8∙106, 0.6 V, 1 Hz) for electrochromic displays.
    Alternatively, 0.13 g Na2IrCl6∙6H2O in 10 mL 0.1 M HCl is boiled, with moderate stirring and the addition of water, to obtain the diaquatetrachloridoiridate(III) ion [Ir(H2O)2Cl4]–. Prior to coating, dissolved oxygen is purged from the solution using nitrogen, and the solution is made slightly alkaline (0.05 to 0.1 M OH−) by adding 6 M NaOH. This results in the formation of air-sensitive iridium(III)-oxide. Electrodeposition is then carried out over 10 to 20 cycles, in the potential range of −0.3 to 1.0 V vs. Ag|AgCl|3 M KCl, at a scan rate of 100 mV/s [49].
  • Pre-synthesized iridium oxide particles are applied to a substrate using polymer binders [50]: a mixture of 0.03 g K2IrCl6 and 0.05 g sodium hydrogen citrate sesquihydrate in 50 mL of water is repeatedly adjusted to a pH of 7.5, using 0.25 M NaOH, and heated to 95 °C under constant stirring (30 min) until the pH stabilizes after cooling to room temperature. Finally, the solution is maintained at 95 °C for 2 h while oxygen is bubbled through it, until it turns a deep-blue colour. Excess citrate is removed via anion-exchange (DOWEX 2X8-50, for 30 min) and the solution is diluted to 100 mL, yielding colloidal IrO2 particles with an estimated size of 10 to 20 nm.
    An IrO2 nanocomposite electrode is fabricated using a compression moulding process, combining the particles with carbon black, and either PVC/DMF or PMMA [51,52]. The electrode surface can be abraded using SiC sandpaper. The nanoparticle suspension in isopropanol (1:1) is suitable for inkjet printing, for example, onto a flexible ITO/PET substrate [53].
  • Electrospun IrO2 nanofibers [54] are produced by forcing a solution (0.19 g IrCl4 + 4 mL DMF + 0.87 g PVP) through a 23-gauge needle (20 kV, 0.3 mL h−1) onto an aluminium foil collector. After drying and calcination (in air, at 900 °C for 3 h), nanofibers with a diameter of approximately 133 nm are obtained. Finally, the material is applied to a previously cleaned glassy carbon electrode using the drop-casting method.

3.2. Influence of Coating Methods on pH Calibration

The aim of this literature review is to determine mean values for the typical Nernst slope and the rest potential of IrO2 electrodes (with different water content).
Interestingly, hydrated IrO2⋅xH2O layers exhibit significantly higher pH sensitivity than anhydrous IrO2 [55]. Katsube (1982, [56]) observed that the open-circuit potential decreases after prolonged storage of the electrodes in water (750 h), a phenomenon attributable to hydrogenation. As the oxidation state increases due to cycling loading, AIROF electrodes show a rising OCP and increased pH sensitivity (low oxidation state: −62.0 mV/pH, E0 = 492 mV; high oxidation state: −73.6 mV/pH, E0 = 824 mV) (Olthuis [41]).
Figure 3a illustrates the influence of the water content when Nernstian slopes from the literature are divided into two groups using one-dimensional hierarchical cluster analysis (via the Python package scipy.cluster.hierarchy; scipy version 1.18.1 [57]). This computational method gradually arranges similar data points into a tree-like structure of nested clusters (a dendrogram), without the number of clusters needing to be predetermined. The method is well suited for exploratory data analysis and visualization, but is computationally intensive, with large datasets, and sensitive to outliers. Each value begins as its own group. In each step, the two groups whose values are closest to one another (defined by Ward’s criterion) are merged. Specifically, the pair is chosen for merging whose combination results in the smallest increase in the variance within the resulting group. This process is repeated until all values are consolidated into a single group; the sequence of these merges is represented as a tree, known as a dendrogram. The final number of groups is then determined by making a cut through this tree.
The Nernst slopes and open-circuit potentials described in the literature (Appendix A) are, indeed, classified into two groups, in Figure 3a. In the first group of 68 values, the average sensitivity of (−56.9 ± 1.5) mV/pH, close to the ideal Nernstian slope (59.16 mV at 25 °C), confirms the electrochemical valence of z = 1 (see Equation (2) for anhydrous IrO2). The OCP lies close to the thermodynamic potential of 926 mV SHE described by Pourbaix for the Ir2O3/IrO2 reaction. The second cluster of 49 values (IrO2 ⋅ xH2O) tends toward higher electrode potentials and sensitivities: (−74.5 ± 2.23) mV/pH.
In summary, thermal methods primarily yield crystalline, anhydrous IrO2. Precipitation methods in alkaline solution produce IrO2 ⋅ xH2O with a residual Ir(III) content. Hydrous IrO2 ⋅ xH2O is produced almost exclusively by electrochemical coating processes, such as AIROF and EIROF. Thermally prepared IrO2 films exhibit better mechanical and chemical stability in aqueous solutions.

4. Numerical Evaluation of Cyclic Voltammograms

This section deals with the computer-aided analysis of cyclic voltammograms, which enables automated voltammetric pH measurement. At the same time, the electrolyte resistance is determined from the CV, eliminating the need for concurrent impedance measurements. Statistical methods are used to show which quantity in the cyclic voltammogram best represents the pH value. As a demonstration dataset, voltammograms were recorded at various pH levels and conductivity values.

4.1. Experimental Setup for pH Titration and Statistical Analysis

The IrO2 working electrode (area: 1 cm2, active layer: 0.144 mg/cm2) was prepared by drop-casting a solution (125 mg IrCl3 + 20 mL acetone + 5 mL 1 M HCl) onto a cleaned high-purity titanium foil, followed by heat treatment (500 °C, 2 h, in air). Preparation details are given in the Supplementary Materials. The electrochemical cell further comprised an Ag|AgCl|3 M KCl reference electrode and a platinum-sheet counter electrode.
To vary the pH value, a Britton–Robinson type universal buffer solution—characterized by a linear relationship between the pH value and the volume of added NaOH [58,59] and consisting of 16.02 g/L citric acid, 10.37 g/L KH2PO4, 9.236 g/L TRIS, 4.714 g/L boric acid, 5.722 g/L glycine, and 9.8 mL/L HCl (32%)—was titrated with sodium hydroxide solution. Following the addition and mixing of a small amount of 1 M NaOH under stirring, and an equilibration phase (30 s, so that the liquid stops swirling), the upper and lower switching potentials for cyclic voltammetry were determined, and CV measurements were performed within the specified potential window, at various scan rates.
To achieve different conductivities across various experiments, the initial buffer solution was diluted with deionized water (1:1, 1:10, 1:5, 1:25, 1:2, 1:1). This experiment results in a parameter space spanning pH values from 2 to 12 and conductivities from 1.05 to 25.79 mS/cm, subdivided into 61 steps and comprising a total of 1830 cyclic voltammetry cycles. The experimental details are outlined in the Supplementary Materials.

4.2. What Quantity Best Reflects pH Changes?

The redox peaks in the cyclic voltammogram [60] can be fitted to a Gaussian normal distribution [61,62,63] using the PYTHON packages numpy, scipy and pygmo [64,65]. The statsmodels package was used for regressions and statistical analyses [66]. The graphs were created with matplotlib [67], and some colour gradients are from Crameri [68,69]. Figure 4 illustrates the workflow:
Figure 4. Cyclic voltammogram of an IrO2 electrode (titanium support) versus an Ag|AgCl|3 M KCl reference and a platinum-sheet counter electrode. Scan rate: 100 mV/s, electrolyte: universal buffer solution (pH 2.3, 14.58 mS/cm). The Ir(III/IV) redox peak was fitted using a Gaussian distribution function: oxidation (green area), reduction (blue). Double-layer correction (dotted line).
  • Load the data and determine the potential reversal point to separate the oxidation and reduction cycles. Smooth the rising and falling edges with a Savitzky-Golay filter (polynomial order: 3, window: 0.1 V).
  • Locate the peak potential, find the Ir(III/IV) redox peak, and select the low points at each peak position as the fitting window for the rising slope (oxidation). Curve fitting is performed using a Gaussian distribution and weighted least squares.
  • Determine: peak potential, peak current, faradaic and capacitive current components, and voltammetric charge. The procedure is analogous for the falling slope (reduction).
To quantify the strongest dependencies between the voltammetric parameters and the pH value (in Figure 5), the Pearson correlation coefficient (Equation (15)) was calculated using the corrcoef function from the PYTHON package numpy.
r = s X Y ~ s X ~ s Y ~
Figure 5. Pearson correlation coefficients for positive and negative trends in the cyclic voltammogram as function of pH at (a) low scan rate (100 mV/s) and (b) fast scan rate (1000 mV/s): ∎ strong correlation, □ no correlation. Solution conductivity κ, resistivity ρ, cycle age. I: input parameters for detecting collinearity, II: raw and averaged potential variables, III: peak standard deviations and differential potential variables, IV: all variables related to the electric current (oxidation and reduction). E electrode potential, I current: P = peak, Ox = oxidation, Red = reduction, ½ half-step, 0 = zero crossing (I = 0). Ru uncompensated electrolyte resistance, Q voltammetric charge, κ electrolyte conductivity, ϱ resistivity.
Here, s X Y ~ represents the empirical covariance between the quantities X and Y, while s X ~ and s Y ~ are the respective standard deviations of X and Y. The strength of the correlation is classified as follows: weak (|r| < 0.5), moderate (0.5 ≤ |r| < 0.8), strong (|r| ≥ 0.8), or perfectly linear (|r| = 1).
Figure 5 compiles the Pearson correlation coefficients for slow and fast scan rates.
I: The first three rows show the internal collinearity of the input parameters within each titration block. The dilution steps were performed in a randomised order. The calculated resistivity and hydronium ion concentration serve as reference values.
II: At low scan rates, the peak potentials (Ep,ox, Ep,red) and zero-crossing potentials (E01, E02) correlate strongly with the pH value. This also applies to the mean values (E1/2 and E0,Avg) and reflects the expected potential shift behaviour described by the Nernst equation (E ~ pH). This correlation weakens at high scan rates, indicating that lower scan rates are better suited for pH measurement.
III: The standard deviations of the fitted normal distributions (σOx, σRed), the potential window (ΔE), and the calculated uncompensated resistance (Ru) show only weak correlation with the pH value at low scan rates.
At high scan rates, potential and resistance correlate with the resistivity in accordance with Ohm’s law (ΔE ~ R ~ ϱ). σOx and σRed also appear to correlate moderately with resistivity ϱ. The zero-crossings and the shape of the voltammogram show a slight correlation with hydronium ion concentration (c(H+) = 10–pH) and the cycle age of the electrode.
IV: No useful correlation is observed between peak currents and pH value (see also Section 4.5). During repeated oxidation and reduction, peak current and voltammetric charge correlate only with electrode age; their absolute values decrease with the number of cycles. This is intuitively understandable for a surface-bound, proton-coupled reaction in buffered solutions without mass transport limitations, given the excess of protons present. Consequently, the reduction in active sites due to aging or material wear results in smaller current peaks. Reliable pH measurements, therefore, require calibration that accounts for the gradual change in the electrode surface.

4.3. Influence of Electrolyte Conductivity and Electrode Aging

Figure 6a,b compares the cyclic voltammograms in highly conductive and poorly conducting electrolytes at similar pH values and at different scan rates. In electrolytes with low conductivity, oxidation peaks shift toward more positive potentials and reduction peaks toward more negative potentials. This effect is more pronounced at high scan rates, meaning the redox peaks no longer occur at the same potential.
Figure 6. Cyclic voltammograms of an IrO2 electrode (on titanium support) at 25 °C in (a) high-conductivity universal buffer solution (pH 2.35, 14.58 mS/cm), second cycle (the first cycle was discarded) at various scan rates (100 to 1000 mV/s). (b) Low-conductivity universal buffer (1:25 dilution, pH 2.87, 1.125 mS/cm, starting cycle: 932). Peak potentials of IrIII/IrIV oxidation (green dots) and reduction (blue dots). (c) Comparison of the new electrode (cycles two to seven) and the aged electrode (cycle 1532 etc., in orange) in universal buffer (undiluted, pH 2.35, 14.58 mS/cm) at a scan rate of 100 mV/s. (d) Shift of the cyclic voltammogram (at 100 mV/s) as function of solution pH. The current intensity reflects the interfering differences in conductivity.
Electrode aging (shown in Figure 6c) leads to a decrease in peak current and voltammetric charge, a result of the loss of active surface area during long-term operation. The measurements mentioned above were performed more or less in immediate succession; the peak separation is therefore attributable to the electrolyte, rather than to any aging of the electrodes. The low Pearson coefficient (0.12 in Figure 5) confirms the negligible influence of aging across the entire series of measurements. Aging has a greater impact on the peak current than on the peak potential.
Increasing the pH value shifts the entire cyclic voltammogram toward more negative potentials (Figure 6d). This is consistent with the theory of the Nernst equation. The linear relationship between the numerically determined peak current and the scan rate (IP ~ v) confirms the kinetic regime of a surface-bound reaction (see Section 2.4).
In fact, for a given pH value, conductivity and comparable cycle age, the peak currents follow Equation (16), employing a global least-squares minimization (Differential Evolution, PYTHON package scipy).
I p = a ⋅ v n + b
The exponent is n = 0.874 ± 0.001 for the oxidation peak and n = 0.857 ± 0.001 for the reduction peak. This trend indicates mixed kinetics with a strong tendency toward a surface reaction, as is to be expected for Ir(III/IV). For low scan rates, Figure 2d shows near-perfect linearity (n → 1).

4.4. Differential Conductance and Capacitance

The electrolyte resistance can be interpreted as the reciprocal of the slope between data points in the cyclic voltammogram: R = (dI/dE)–1. However, this diagram exhibits infinitely high signals as potential changes approach zero (Figure 7).
Figure 7. IrO2 electrode in universal buffer solution at 25 °C at a scan rate of v = 100 mV/s. (a) Differential conductance, (b) differential resistance, (c) differential capacitance: ① IrIII/IrIV oxidation, ② double-layer region. (d) Peak current as function of pH at a new and at a used electrode: +IrIII/IrIV oxidation, + reduction.
Differentiation of the cyclic voltammogram with respect to the electrode potential yields the differential conductance, G = dI/dE = 1/R. A zero-crossing is observed in place of the Ir(III/IV) redox peak. Practical application: see Section 6.
The ohmic voltage drop appears as the difference between the peak potentials, and thus constitutes the dominant component of the separation.
Differential capacitance, C = dQ/dU = I/v, is qualitatively similar to the original cyclic voltammogram. Serving as a measure of the active surface area, it is composed of the double-layer capacitance and the Faradaic capacitance of the redox process. Unfortunately, the advantages of these diagrams for practical pH measurement remain limited.

4.5. Is There a Correlation Between Peak Current and pH?

Based on the Randles–Ševčík equation of voltammetry, a linear relationship between peak current and hydronium ion concentration (or: log IP ~ pH) would be expected—if the proton were a freely diffusing species. However, Figure 7d shows a linear dependence of the Ir(III/IV) peak current on pH, although the data exhibit significant scatter.
In unbuffered solutions, the voltammetric peak height (current maximum) and peak width (half-width at half-height) decrease with increasing pH (Yuen [17]); in buffered solutions, only the Ir(III/IV) peak potential changes, while the shape (peak height and width) remains constant across all pH levels (cf. Figure 6d).
Unfortunately, the gradual aging of the electrode—manifesting as a loss of active surface area—complicates the situation. Consequently, absolute amperometric or coulometric pH measurements are not readily feasible, as the concentration of the active iridium species is unknown. There is neither an IrO2 reference standard electrode nor a practical method for determining the electrode surface area as part of a calibration procedure.
A clear correlation can also be observed between the voltammetric charge and the pH value; however, the scatter in the measured values—due to various influencing factors and electrode aging—is too great for practical utility. The anodic peak charge appears to be a measure of the electrochemically active surface area [70,71]; thin IrO2 coatings show a linear correlation of peak current and scan rate (IP ~ v), whereas thick layers are controlled by diffusion at high scan rates (IP ~ v ). However, there is currently no reference standard for an IrO2 electrode with reproducible voltammetric charge for a calibration procedure. Moreover, the hydrogen reduction region reflects the electrode surface area, yet a reliable reference value for such absolute measurements is lacking.
It follows from the foregoing that absolute pH measurement using potentiometric, amperometric, or coulometric methods is not technically feasible. For voltammetric pH measurement based on redox potentials, reference buffer solutions are still required for calibration.

5. Practical Voltammetric pH Determination

5.1. Mean Peak Potential and Calibration

Recent literature correlates the mean value of the Ir(III/IV) oxidation and reduction peak potentials with pH (e.g., [72,73]); for reversible reactions and negligible uncompensated resistance, this mean value lies close to the half-wave potential—and, thus, the equilibrium potential—of the redox reaction (Bard [38]). Accuracy could be improved by using square-wave voltammetry (SWV), albeit at the cost of greater deviations attributable to the ionic strength of the sample matrix [74]. For internal calibration, the mean potential of the Ir(III/IV) reaction can be subtracted from the potential of hydrogen underpotential deposition; this method is suitable for use in seawater within the pH range of 7.5 to 8.5 (Chen [20]). For unknown pH values between pH 0 and pH 14, however, there is currently no alternative to calibration using standard buffer solutions.

5.2. Proof of Linearity

As shown in Section 4.2, the peak potentials, the zero-crossing potentials, and the potential difference between the Ir(III/IV) oxidation and the zero-crossing, (Ep,ox − E0), reflect the pH value. The mean value, EP,1/2 = (EP,ox + EP,red)/2, is calculated from the oxidation and reduction peak potentials. The pH value of the unknown solution is determined from the calibration curve. Figure 8 shows the excellent linearity between electrode potential and pH in various regions of the cyclic voltammogram.
Figure 8. pH-calibration based on the mean peak potential, determined in various regions of the cyclic voltammogram at an IrO2 electrode in universal buffer solutions at a scan rate of 100 mV/s. 1 (Grey) upper zero-crossing potential (E02), correlating with the onset of oxygen evolution. 2 (Red) average value of the zero-crossing potentials: (E01 + E02)/2. 3 (Blue) Ir(III/IV) oxidation and (light blue) reduction (E1/2). 4 (Orange) lower zero-crossing potential (E01), reflecting the onset of the hydrogen adsorption region (cf. Figure 4).
The measured sensitivity is slightly higher than the expected Nernstian slope of 59.16 mV/pH (25 °C). This is because the hydration state of iridium oxide strongly influences pH sensitivity. Hydrous IrO2 ∙ x H2O layers exhibit a sensitivity of nearly −90 mV/pH, whereas this value changes to almost −60 mV/pH for anhydrous layers (see Section 3.2). The thermal decomposition of iridium chloride, as employed in this study, yields weakly hydrated iridium dioxide; however, these layers are more stable over the long term than hydrous powders.
The cyclic voltammograms were recorded at a scan rate of 100 mV/s using a voltage window that reveals both the Ir(III/IV) oxidation and reduction. Peak potentials were determined using a peak-detection algorithm such as find_peaks from the PYTHON package scipy. To avoid erroneous peak recognition-caused by noise, it is advisable to use a filter configured for a peak width of approximately 0.1–0.2 V: for example, find_peaks (I_smooth, width = Points_per_V*0.1).

5.3. pH-Determination in Real Samples

The calibration results with two standard solutions of known specified pH and conductivity are shown in Table 1. For the unknown samples, five CV cycles were recorded at a scan rate of 100 mV/s, using the same electrochemical cell setup. Buffer solutions of known pH values were used as some ‘unknown solutions’ to assess the quality and error of the voltammetric method (see Table 2). The unknown pH values of beverages were determined, as well. The maximum deviation from the known pH value of the IUPAC standards is 0.46, with the value measured by the sensor consistently lying above the pH reference value. In comparison, the glass electrode exhibits a maximum measurement error of 0.31. In both cases, the error increases in an alkaline environment.
Table 1. Simple 2-point calibration measurement using two standard buffer solutions. Calibration line: EP1/2/V= –0.06333 pH + 0.6627. Measurement uncertainty of EP1/2 is ±0.00012 V (in italics).
Table 2. Voltammetric pH determination. IUPAC primary standard solutions (with known pH values), and beverage samples (unknown pH). Glass electrode for comparison only.
The high resistance of tap water makes peak evaluation difficult. The five CV cycles consistently yield the same value: pH 9.15 ± 0.14. It should be noted that the glass electrode also tends to drift in water toward an alkaline response.

6. Conductivity Measurement by Cyclic Voltammetry

Alternating current methods are typically used for conductivity measurements. The following section demonstrates that approximate resistance values for the test solution can also be derived from the cyclic voltammogram.

6.1. Application Example of a Simple, Practice-Oriented Method

As shown in Section 2.4, the ideal peak separation for the reversible, surface-bound redox process is EP,ox − EP,red = 0. Assuming that the kinetics of the Ir(III/IV) redox reaction generates only a negligible charge transfer resistance, then the peak separation is mainly caused by the uncompensated electrolyte resistance between working and reference electrode (see Equation (14)). For calibration, the cyclic voltammogram was measured in solutions of known resistivity (Figure 9).
Figure 9. (a) Calibration curve of the solution resistance R = 1/G (determined from the peak separation of the Ir(III/IV) redox peak at an IrO2 electrode; scan rate 1000 mV/s), as function of the known resistivity ϱ (reciprocal of conductivity κ) for various standard buffer solutions. (b) Conductance G as the slope of the straight line connecting the oxidation (red) and reduction (blue) peaks.
To test the IrO2 sensor, various potassium chloride solutions were analysed, using five voltammetric cycles at a scan rate of 1000 mV/s. Conductance was calculated based on the peak voltages and currents (Table 3). A simple two-point calibration of the conductivity was performed using two KCl standard solutions. Then, the conductivity of the test samples (with known reference conductivity) was determined, using the calibration curve. The deviation between the measured values and the literature values ranges from 0.009 to 2.384 mS cm−1, with the latter value falling outside the calibration range.
Table 3. Voltammetric conductivity (κ) determination in electrolyte solutions. Two-point calibration line: κ = 0.6435 cm–1 ∙ G − 0.4956 mS cm–1, where G = (IP,ox − Ip,red)/(EP,ox − Ep,red) is the conductance from the cyclic voltammogram. Values of a commercial conductivity meter and impedance measurement (EIS) in the same solution are for comparison only.
This result demonstrates—quite strikingly, and, perhaps, surprisingly for some CV users—that the uncompensated solution resistance can be determined from the peak spacing at IrO2 electrodes, thereby eliminating the need for impedance measurements.

6.2. The Difference Between Differential and Integral Conductance

The mean conductance G in Figure 9b accounts for both the oxidation and the reduction peak, whereas the differential conductance dI/dE represents the derivative along the cyclic voltammogram. The two values differ significantly in numerical terms, and possess different physicochemical significance.
Figure 10a yields an integral conductance of G = ΔIP/ΔEP = 0.0796 S, whereas differential conductance is dI/dU = 0.0022 S at 0.707 V. In simplified terms, G primarily reflects the changes in the electrolyte and the double layer, whereas the differential value, dI/dE, tends to capture the changes in kinetic limitations at the electrode. At the chemical equilibrium of a redox reaction, the reaction rate—and, consequently, the change in current—approaches zero (dI = 0). At high scan rates, G and dI/dE increase, due to higher double-layer charge and discharge currents (I = C v).
Figure 10. Differential and integral conductance of an IrO2 electrode on titanium support in 0.5 M H2SO4, derived from the cyclic voltammogram: (a) at a scan rate of 100 mV/s, (b) at 1000 mV/s: G = 0.2574 S, and dI/dU = 0.0177 S. ◌ maximum oxidation and reduction currents, ● half-wave potential, determined by equation (12). Green: oxidation curve, blue: reduction curve.

7. A Preliminary Attempt at Calibration-Free pH Measurement

For many technical applications, it is often sufficient to distinguish between acidic and basic conditions on a rough pH scale, without requiring a precise numerical pH value. The following hypothetical approach assumes that IrO2·xH2O exists with an unknown composition, and that its composition ages over time. The water content varies, depending on the manufacturing process and the history of the electrode’s use; this is reflected in the differing Nernstian slopes reported in the literature (Appendix A). The variety of IrO2 electrodes is reduced here to a representative model electrode, which exhibits a Nernst slope that remains to be defined.
Any given anhydrous or hydrated IrO2 electrode will deviate, to a greater or lesser extent, from the literature mean value of 64 ± 6 mV/pH. The medians for anhydrous and hydrated IrO2 (Figure 3a) largely confirm this value: (58.5 + 72.6)/2 ≈ 65.6 mV.
Based on this proposed Nernstian slope (dE/dpH = 64 mV), as well as the rounded standard potential of an IrO2 electrode (E0 = 0.9 V vs. SHE), an attempt is made to determine the pH value of unknown solutions using cyclic voltammetry, without the need for prior calibration. In this approach, only the Ir(III/IV) redox potential, E1/2 (calculated as the mean of the oxidation and reduction peak potentials) was measured relative to a silver/silver chloride electrode (Eref = 0.2 V SHE).
p H = − ( E 1 / 2 + E r e f ) − 0.9   V 0.064   V
As shown in Table 4, the pH value calculated according to Equation (17) deviates from the target values by approximately half a pH unit to less than one pH unit.
Table 4. Attempt of calibration-free pH measurement in IUPAC standard solutions of known pH, using average values for Nernst slope and standard potential from literature (see Appendix A). SSE = silver/silver chloride reference electrode.
The next consideration is based on the hypothesis that the zero-crossings of the cyclic voltammogram (I = 0) correspond approximately to the thermodynamic equilibrium potentials for hydrogen and oxygen evolution. These potentials are theoretically separated by 1.23 V. If the observed separation is greater, this is attributable to the resistance of the electrode–electrolyte interface. Consequently, the measured cyclic voltammogram is scaled to a 1.23 V range. The deviation from the rest potential (0.9 V) provides a direct indication of the pH value, provided that the Nernst slope is defined in advance.
p H = − ( E 1 / 2 + E r e f − 0.9   V ) ⋅ ( 1.23   V E 02 − E 01 ) 0.064   V
Here, E01 and E02 denote the points at which the cyclic voltammogram intersects the axis at zero current. Using Equation (18) improves pH measurement with an IrO2 electrode across the entire pH range, although at the cost of accuracy in the region around pH 7.
Note that this is a hypothetical model experiment, intended to demonstrate the impact of an arbitrary setting of the Nernst slope on the accuracy of the pH value.

8. Visionary Concepts of Counting Protons

The following experiment provides visual evidence of the proton that is released during the Ir(III/IV) redox reaction (see Equations (1) and (2)). This method is not intended for pH measurement, but to explain the processes at the IrO2 electrode.

8.1. Optical Detection of Proton Release

In a very small volume of electrolyte, the change in pH at an IrO2 electrode during cyclic voltammetry can be monitored using the indicator methyl red. A planar sensor—consisting of an IrO2 working electrode, an IrO2 pseudo-reference electrode, and a gold counter electrode—was wetted with 50 µL of sulfuric acid (1 µmol L−1) containing methyl red. During the oxidation of Ir(III) to Ir(IV), the solution becomes more acidic, and the indicator turns red (Figure 11a, ① → ②). Upon the reduction of Ir(IV) to Ir(III), methyl red loses its colour (② → ③), as the IrO2 film takes up protons or hydroxide ions are formed at the counter electrode. RGB values of the electrode surface, captured via a Raspberry Pi camera, reveal the protonated form of methyl red in acidic solution (detected in the green channel at 523–526 nm) and the deprotonated form in alkaline solution (at 427–437 nm). The transition range lies between pH 4.4 and 6.0, in theory at pH 5.1.
Figure 11. (a) Cyclic voltammetry at an IrO2 working electrode against an IrO2 reference electrode and a gold counter electrode in 50 µL sulfuric acid (1 µmol/L) at a scan rate of 100 mV/s. Absorbance of the green channel of the RGB values: A(t) = –log G(t)/G0 (in green). (b) Potential step experiment from 0 V (for 20 s) to 0.8 V (20–80 s, above Ir(III/IV) redox potential) and down to –0.15 V (80 s to 140 s, below Ir(III/IV) peak). ① → ⑥ Methyl red changes from yellow–green colour (pH > 6.2) to red (pH < 4.4). Note the small red coloration on the black electrode in sample ②.
The absorbance of the green channel, A’ = –log(G/G0), relative to the initial state G0 prior to potential change, increases when the indicator turns red in acidic solution and decreases in alkaline solution. Calibration of the RGB colours against pH was omitted because this investigation is about detecting voltammetrically released protons, i.e., a relative pH change.

8.2. Limitations of Coulometric pH Determination

Typically, an acidic analyte contains a different number of hydronium ions than the IrO2 electrode itself. Until the proton storage capacity is exhausted, the voltammetric charge at the IrO2 electrode corresponds to the amount of substance reacted: Q = F n(H+). For instance, 50 µL of H2SO4 (containing 0.1 nmol H+) would require a charge of 9.65 µC. At pH 2, the voltammetric charge associated with the Ir(IV/III) reduction peak was measured as 102.2 µC (see Section 4). Consequently, it should be possible to determine the pH value (or potential H+ concentration; that is, dissociated and undissociated H+) according to Faraday’s law, rather than measuring it indirectly via the electrode potential. In this approach, the choice of reference electrode would become unimportant, and a two-point calibration could be replaced by the proton storage capacity of the IrO2 layer.
Jansod et al. [75] describe a polymer membrane (containing a lipophilic hydrogen-ion receptor and a cation exchanger). A potential pulse drives protons from an analyte layer (approximately 100 µm thick) into an HCl compartment (10 mmol/L) that is terminated by an Ag|AgCl contact. Integrating the current over the duration of the pulse yields the number of protons removed and, according to Faraday’s law, the change in pH within the layer.
Coulometric pH determination would require a material that remains stable in its deprotonated state when in contact with the sample under open-circuit conditions; it would also need to possess a high storage capacity—large, relative to the proton content of the sample—and be capable of selectively taking up protons without simultaneously reducing water. IrO2 exhibits weaknesses in these respects. Furthermore, there are additional practical obstacles: the counter electrode (where hydrogen evolution generates a comparable amount of OH− ions), atmospheric CO2 (approximately 0.7 nmol of dissolved inorganic carbon in 50 µL), and an H+ offset originating from water contained in the film itself. In the case of buffer solutions, protons would be stored within the electrode film, meaning the resulting measurement would reflect the acid capacity, rather than the pH value. A conductive polymer based on quinones, which functions as a ‘proton battery’ in aqueous acid [76], would be an obvious candidate for this purpose.
Consequently, an absolute pH measurement traceable to Faraday’s law is currently not feasible.

9. Conclusions

A practical voltammetric method for pH measurement was outlined, which utilizes the pH-dependent potential shift at an IrO2 electrode. It could be used analogously to a pH glass electrode, and requires calibration with standard buffer solutions.

9.1. Feasibility of Voltammetric pH Measurement

The voltammetric pH measurement described is not yet sufficiently reliable for monitoring water quality in any medium. However, the method opens up fundamental possibilities for applications when exact pH values and conductance values are not required. In the future, miniaturised IrO2 electrodes—potentially in combination with other metal oxides—could play a role in pH sensor arrays and ‘electronic tongues.’
  • The half-wave potential of the Ir(III)/Ir(IV) redox peak exhibits a strong linear relationship with the pH value. An analysis of literature values from the past 45 years yields a mean Nernstian slope of (−64 ± 6) mV/pH as a rough reference value for IrO2-pH electrodes.
  • The voltammetric signals allow a cyclic voltammogram to be corrected for the interfering ohmic resistance, which shifts the redox potentials away from their theoretical values.
  • Absolute pH measurements or the use of peak currents are currently not feasible, due to the gradual aging of the electrode surface (loss of active area).
  • ‘Proton counting’ with the help of voltammetric charge remains a vision, as long as there is no reliable reference standard for the active proton-exchanging electrode surface.

9.2. Sustainability of IrO2 Electrodes

Compared to the well-established glass electrode, the IrO2 electrode exhibits the following specific characteristics:
  • Reusability, operational lifespan and replacement frequency: in this study, the IrO2 electrode was operated for more than 1500 cycles without any loss of functionality; however, a 49% decrease in peak current and a –18 mV shift in the half-wave potential were observed. For medical applications and use in contaminated environments, single-use IrO2 sensors could be more cost-effective than glass electrodes.
  • Material availability: IrO2 and its precursor compounds are expensive and scarce materials (approx. 320 EUR/gram); however, they are required only in sub-milligram quantities (in this study: <0.15 mg/cm2). At the end of its service life, IrO2 can be chemically recycled and, like used electrodes from water electrolysis, reused as part of a circular economy.
  • Fabrication costs and energy consumption: compared to a glass electrode, an IrO2 sensor requires fewer components and materials. There is no need for an internal liquid electrolyte, a glass membrane, or a KCl filling, which eliminates the risk of breakage associated with glass electrodes. Furthermore, the production and processing of IrO2 electrodes are less energy-intensive than glass processing. Reducing the electrode surface area allows for additional material savings; consequently, IrO2 is particularly attractive for microelectrodes.
  • Maintenance and calibration requirements: for both the glass electrode and the IrO2 sensor, a two-point calibration using standard buffer solutions is sufficient. The glass electrode can only be used after the membrane has been soaked, and it must be stored in a protective solution. The IrO2 sensor can be stored dry.
  • Reproducibility and long-term stability: the glass electrode currently exhibits superior stability in pH measurements, although a certain degree of potential drift does occur. For both the glass electrode and the IrO2 electrode, aging effects are compensated by a proper calibration process.
  • In terms of performance in realistic water matrices and monitoring reliability, the glass electrode holds the advantage. However, it is expected that future sensor systems based on various metal-oxide electrodes will be able to overcome the current drawbacks associated with IrO2. The glass electrode cannot be used in solutions containing hydrofluoric acid, and exhibits greater error in alkaline media and pure water. pH measurement in concentrated strong acids and bases has not yet been successfully achieved using potentiometric methods, which highlights the need for new approaches.
In summary, the extent to which an IrO2 electrode can replace a conventional glass electrode depends on the specific application and the associated costs. For future applications, e.g., in an ‘electrochemical tongue’, which can provide far more than just a single numerical pH value, metal-oxide electrodes are likely to prove superior to glass electrodes. Although IrO2 is considered the most stable oxide among the platinum-group metal oxides, it can also be replaced by other metal oxides. The voltammetric method described here functions independently of the specific electrode material; the essential requirement is a reversible redox process coupled with the uptake or release of a hydronium ion, whether occurring within the material itself or involving the surrounding solution.

Supplementary Materials

The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/su182312119/s1, Figure S1: XRD spectrum of the IrO2 electrode; Figure S2: photographs of the test setup; Figure S3: Range of measured parameters; Figure S4: Overlapping cyclic voltammograms from five consecutive measurements; Table S1: Measurement uncertainty; Figure S5: Additional example, proof of linearity; Table S2: Calibration measurement; Table S3: voltammetric pH measurement of beverage samples.

Author Contributions

Writing—original draft preparation, review and editing, J.S. and P.K. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable; studies not involving humans or animals.

Data Availability Statement

The original contributions presented in this study are included in the article/Supplementary Materials. Further inquiries can be directed to the corresponding author(s).

Conflicts of Interest

The authors declare no conflicts of interest.

Symbols and Abbreviations

CVCyclic voltammetry
Eelectrode potential, usually measured versus that of an Ag|AgCl reference electrode
FFaraday constant: 96485 C/mol
Ielectric current, P = peak (in A)
OCPOpen-circuit potential
oxelectrochemical oxidation, release of electrons
Qvoltammetric charge (in C = As)
Rmolar gas constant: 8.3144 J mol–1K–1; ohmic resistance (in Ω)
redelectrochemical reduction, uptake of electrons
Tabsolute temperature (in K)
RT⋅ln(10)/FNernst slope dE/dpH, often called ‘pH sensitivity’: 0.059 V (at 25 °C)
vscan rate (in V/s)
zelectrochemical valency, number of electrons in the redox equation

Appendix A

Table A1. Narrative literature overview of iridium oxide pH sensors: preparation methods, substrates, precursors, solvents, Nernstian slope (sensitivity), extrapolated electrode potential E at pH 0 versus reference electrode at 25 °C, pH range, and measurement method. CV = Ir(III/IV) half-wave potential. Calculated standard potentials (E0 vs. SHE) are shown in Figure 3, where the reference potentials of 210 mV for Ag|AgCl, and 241 mV for SCE are added. Potentially inaccurate temperature data in the literature (e.g., by 5 K) would lead to an error in sensitivity of 1 − 293 K/298 K ≈ 1.7%.

References

  1. Sługocki, Ł.; Czerniawski, R. Water Quality of the Odra (Oder) River before and during the Ecological Disaster in 2022: A Warning to Water Management. Sustainability 2023, 15, 8594. [Google Scholar] [CrossRef] [Scilit]
  2. Schulte, C.; Abbas, B.; Engelke, C.; Fischer, H.; Henneberg, S.; Hentschel, H.; Jekel, H.; Jeske, R.; Pietsch, K.; Schöll, F.; et al. Fish Die-off in the Oder River; German Environment Agency (UBA): Dessau-Roßlau, Germany, 2022. [Google Scholar]
  3. United Nations General Assembly. Transforming Our World: The 2030 Agenda for Sustainable Development; United Nations General Assembly: New York, NY, USA, 2015. [Google Scholar]
  4. Ahmed, U.; Mumtaz, R.; Anwar, H.; Mumtaz, S.; Qamar, A.M. Water Quality Monitoring: From Conventional to Emerging Technologies. Water Supply 2020, 20, 28–45. [Google Scholar] [CrossRef] [Scilit]
  5. UN-Water An Introduction to SDG Indicator 6.3.2: Proportion of Bodies of Water with Good Ambient Water Quality. 2026. Available online: https://www.unwater.org/our-work/sdg-6-integrated-monitoring-initiative/indicator-632-proportion-bodies-water-good-ambient (accessed on 1 November 2026).
  6. Available online: https://eur-lex.europa.eu/legal-content/EN/TXT/PDF/?uri=CELEX:32020L2184 (accessed on 1 November 2026).
  7. Graham, D.J.; Jaselskis, B.; Moore, C.E. Development of the Glass Electrode and the pH Response. J. Chem. Educ. 2013, 90, 345–351. [Google Scholar] [CrossRef] [Scilit]
  8. Wiora, A.; Wiora, J. Over One-Year Long-Term Laboratory Tests of pH Electrodes in Terms of Industrial Applications Checking Stabilities of Their Parameters and Their Influence on Uncertainties of Measurements. Sensors 2018, 18, 4102. [Google Scholar] [CrossRef] [Scilit]
  9. Kurzweil, P. Metal Oxides and Ion-Exchanging Surfaces as pH Sensors in Liquids: State-of-the-Art and Outlook. Sensors 2009, 9, 4955–4985. [Google Scholar] [CrossRef] [Scilit]
  10. Zhuiykov, S. Solid-State Sensors Monitoring Parameters of Water Quality for the next Generation of Wireless Sensor Networks. Sens. Actuators B Chem. 2012, 161, 1–20. [Google Scholar] [CrossRef] [Scilit]
  11. Martínez-Máñez, R.; Soto, J.; García-Breijo, E.; Gil, L.; Ibáñez, J.; Gadea, E. A Multisensor in Thick-Film Technology for Water Quality Control. Sens. Actuators A Phys. 2005, 120, 589–595. [Google Scholar] [CrossRef] [Scilit]
  12. Hitchman, M.L.; Ramanathan, S. Potentiometric Determination of Proton Activities in Solutions Containing Hydrofluoric Acid Using Thermally Oxidized Iridium Electrodes. Analyst 1991, 116, 1131. [Google Scholar] [CrossRef] [Scilit]
  13. Fog, A.; Buck, R.P. Electronic Semiconducting Oxides as pH Sensors. Sens. Actuators 1984, 5, 137–146. [Google Scholar] [CrossRef] [Scilit]
  14. Dong, Q.; Sun, X.; He, S. Iridium Oxide Enabled Sensors Applications. Catalysts 2021, 11, 1164. [Google Scholar] [CrossRef] [Scilit]
  15. Hassan, S.S.M.; Fathy, M.A.; Moussa, I.; Obaida, M.; Kamel, A.H. A novel miniaturized pH potentiometric electrode based on a nano bismuth oxide film deposited on a fluorine doped nano tin oxide glass substrate. Sens. Actuators B Chem. 2023, 380, 133397. [Google Scholar] [CrossRef] [Scilit]
  16. Manjakkal, L.; Szwagierczak, D.; Dahiya, R. Metal Oxides Based Electrochemical pH Sensors: Current Progress and Future Perspectives. Prog. Mater. Sci. 2020, 109, 100635. [Google Scholar] [CrossRef] [Scilit]
  17. Yuen, M.F.; Lauks, I.; Dautremont-Smith, W.C. PH Dependent Voltammetry of Iridium Oxide Films. Solid State Ion. 1983, 11, 19–29. [Google Scholar] [CrossRef] [Scilit]
  18. Chalupczok, S.; Kurzweil, P.; Hartmann, H.; Schell, C.H. The Redox Chemistry of Ruthenium Dioxide: A Cyclic Voltammetry Study, Review and Revision. Int. J. Electrochem. 2018, 2018, 1273768. [Google Scholar] [CrossRef] [Scilit]
  19. Wachta, I.; Balasubramanian, K. Electroanalytical Strategies for Local pH Sensing at Solid–Liquid Interfaces and Biointerfaces. ACS Sens. 2024, 9, 4450–4468. [Google Scholar] [CrossRef] [Scilit]
  20. Chen, Y.; Compton, R. A Calibration-Free pH Sensor Using an In-Situ Modified Ir Electrode for Bespoke Application in Seawater. Sensors 2022, 22, 3286. [Google Scholar] [CrossRef] [Scilit]
  21. Chalupczok, S.; Kurzweil, P.; Hartmann, H. Impact of Various Acids and Bases on the Voltammetric Response of Platinum Group Metal Oxides. Int. J. Electrochem. 2018, 2018, 1697956. [Google Scholar] [CrossRef] [Scilit]
  22. Chawang, K.; Bing, S.; Chiao, J.-C. Effects of Viscosity and Salt Interference for Planar Iridium Oxide and Silver Chloride pH Sensing Electrodes on Flexible Substrate. Chemosensors 2022, 10, 371. [Google Scholar] [CrossRef] [Scilit]
  23. Kurzweil, P. Low temperature proton conducting mixed metal oxide supercapacitor. In Meeting Abstracts: Vol. 96-2, Fall Meeting San Antonio, Texas, October 6–11; The Electrochemical Society: Pennington, NJ, USA, 1996; p. 825. [Google Scholar]
  24. Kurzweil, P. Precious Metal Oxides for Electrochemical Energy Converters: Pseudocapacitance and pH Dependence of Redox Processes. J. Power Sources 2009, 190, 189–200. [Google Scholar] [CrossRef] [Scilit]
  25. Pourbaix, M. Atlas of Electrochemical Equilibria in Aqueous Solutions. National Accociation of Corrosion Engineers; Cebelcor: Brussels, Belgium, 1974. [Google Scholar]
  26. Gottesfeld, S.; McIntyre, J.D.E.; Beni, G.; Shay, J.L. Electrochromism in Anodic Iridium Oxide Films. Appl. Phys. Lett. 1978, 33, 208–210. [Google Scholar] [CrossRef] [Scilit]
  27. Kötz, R.; Neff, H.; Stucki, S. Anodic Iridium Oxide Films. J. Electrochem. Soc. 1984, 131, 72–77. [Google Scholar] [CrossRef] [Scilit]
  28. Saveleva, V.A.; Wang, L.; Teschner, D.; Jones, T.; Gago, A.S.; Friedrich, K.A.; Zafeiratos, S.; Schlögl, R.; Savinova, E.R. Operando Evidence for a Universal Oxygen Evolution Mechanism on Thermal and Electrochemical Iridium Oxides. J. Phys. Chem. Lett. 2018, 9, 3154–3160. [Google Scholar] [CrossRef] [Scilit]
  29. Burke, L.D.; Mulcahy, J.K.; Whelan, D.P. Preparation of an Oxidized Iridium Electrode and the Variation of Its Potential with pH. J. Electroanal. Chem. Interfacial Electrochem. 1984, 163, 117–128. [Google Scholar] [CrossRef]
  30. Breiter, M.; Knorr, C.A.; Völkl, W. Über die bei verschiedenen stationären Überspannungswerten auf Edelmetall-elektroden adsorbierten, mittels Ladekurven bestimmbaren Wasserstoffmengen. Z. für Elektrochem. Berichte der Bunsenges. für Phys. Chem. 1955, 59, 681–688. [Google Scholar] [CrossRef] [Scilit]
  31. Böld, W.; Breiter, M. Untersuchung des anodischen Aufbaus und der kathodischen Reduktion der Sauerstoffbelegung am Rhodium und Iridium. Electrochim. Acta 1961, 5, 169–179. [Google Scholar] [CrossRef] [Scilit]
  32. Lam, A.; Sun, Q.; Polus, Z.; Buek, C.; McCrea, K.; Mahoney, I.; Wegge, L.; Zeng, Z.; McGowan, M.; Yu, H.; et al. Structure of Iridium Oxides and Their Oxygen Evolution Electrocatalysis in Acidic Media. ACS Catal. 2025, 15, 8414–8425. [Google Scholar] [CrossRef] [Scilit]
  33. Trasatti, S.; Petrii, O.A. Real Surface Area Measurements in Electrochemistry. J. Electroanal. Chem. 1992, 327, 353–376. [Google Scholar] [CrossRef] [Scilit]
  34. Łukaszewski, M.; Soszko, M.; Czerwiński, A. Electrochemical Methods of Real Surface Area Determination of Noble Metal Electrodes—An Overview. Int. J. Electrochem. Sci. 2016, 11, 4442–4469. [Google Scholar] [CrossRef] [Scilit]
  35. Wei, C.; Sun, S.; Mandler, D.; Wang, X.; Qiao, S.Z.; Xu, Z.J. Approaches for Measuring the Surface Areas of Metal Oxide Electrocatalysts for Determining Their Intrinsic Electrocatalytic Activity. Chem. Soc. Rev. 2019, 48, 2518–2534. [Google Scholar] [CrossRef] [Scilit]
  36. Sauerbrey, G. Verwendung von Schwingquarzen zur Wägung dünner Schichten und zur Mikrowägung. Z. Phys. 1959, 155, 206–222. [Google Scholar] [CrossRef] [Scilit]
  37. Laviron, E. General Expression of the Linear Potential Sweep Voltammogram in the Case of Diffusionless Electro-chemical Systems. J. Electroanal. Chem. 1979, 101, 19–28. [Google Scholar] [CrossRef] [Scilit]
  38. Bard, A.J.; Faulkner, L.R.; White, H.S. Electrochemical Methods: Fundamentals and Applications, 3rd ed.; John Wiley & Sons: Hoboken, NJ, USA, 2022. [Google Scholar]
  39. Rand, D.A.J.; Woods, R. Cyclic Voltammetric Studies on Iridium Electrodes in Sulphuric Acid Solutions. J. Electroanal. Chem. Interfacial Electrochem. 1974, 55, 375–381. [Google Scholar] [CrossRef]
  40. Schiavone, L.M.; Dautremont-Smith, W.C.; Beni, G.; Shay, J.L. Electrochromic Iridium Oxide Films Prepared by Reactive Sputtering. Appl. Phys. Lett. 1979, 35, 823–825. [Google Scholar] [CrossRef] [Scilit]
  41. Olthuis, W.; Robben, M.A.M.; Bergveld, P.; Bos, M.; van der Linden, W.E. pH Sensor Properties of Electrochemically Grown Iridium Oxide. Sens. Actuators B Chem. 1990, 2, 247–256. [Google Scholar] [CrossRef] [Scilit]
  42. Beni, G.; Schiavone, L.M.; Shay, J.L.; Dautremont-Smith, W.C.; Schneider, B.S. Electrocatalytic Oxygen Evolution on Reactively Sputtered Electrochromic Iridium Oxide Films. Nature 1979, 282, 281–283. [Google Scholar] [CrossRef] [Scilit]
  43. Ardizzone, S.; Carugati, A.; Trasatti, S. Properties of thermally prepared iridium dioxide. J. Electroanal. Chem. 1981, 126, 287–292. [Google Scholar] [CrossRef]
  44. Da Silva, G.M.; Lemos, S.G.; Pocrifka, L.A.; Marreto, P.D.; Rosario, A.V.; Pereira, E.C. Development of Low-Cost Metal Oxide pH Electrodes Based on the Polymeric Precursor Method. Anal. Chim. Acta 2008, 616, 36–41. [Google Scholar] [CrossRef] [Scilit]
  45. Pechini, M.P.; North, A. Method of Preparing Lead and Alkaline Earth Titanates and Niobates and Coating Method Using the Same to form a Capacitor. US Patent 3,330,697, 11 July 1967. [Google Scholar]
  46. Nishio, K.; Tsuchiya, T. Electrochromic Thin Films Prepared by Sol-Gel Process. Sol. Energy Mater. Sol. Cells 2001, 68, 279–293. [Google Scholar] [CrossRef] [Scilit]
  47. Yao, S.; Wang, M.; Madou, M. A pH Electrode Based on Melt-Oxidized Iridium Oxide. J. Electrochem. Soc. 2001, 148, H29. [Google Scholar] [CrossRef] [Scilit]
  48. Yamanaka, K. Anodically Electrodeposited Iridium Oxide Films (AEIROF) from Alkaline Solutions for Electrochromic Display Devices. Jpn. J. Appl. Phys. 1989, 28, 632. [Google Scholar] [CrossRef] [Scilit]
  49. Baur, J.E.; Spaine, T.W. Electrochemical Deposition of Iridium (IV) Oxide from Alkaline Solutions of Iridium(III) Oxide. J. Electroanal. Chem. 1998, 443, 208–216. [Google Scholar] [CrossRef] [Scilit]
  50. Hara, M.; Lean, J.T.; Mallouk, T.E. Photocatalytic Oxidation of Water by Silica-Supported Tris(4,4′-Dialkyl-2,2′-Bipyridyl)Ruthenium Polymeric Sensitizers and Colloidal Iridium Oxide. Chem. Mater. 2001, 13, 4668–4675. [Google Scholar] [CrossRef] [Scilit]
  51. Kim, W.; Chung, K.C.; Park, J. Surface Renewable Hydrogen Ion-Selective Polymeric Composite Electrode Containing Iridium Oxide. Bull. Korean Chem. Soc. 2005, 26, 1565–1568. [Google Scholar] [CrossRef] [Scilit]
  52. Park, J.; Kim, M.; Kim, S. Surface Renewable Nano-Iridium Oxide Polymeric Composite pH Electrodes. Sens. Actuators B Chem. 2014, 204, 197–202. [Google Scholar] [CrossRef] [Scilit]
  53. Jović, M.; Hidalgo-Acosta, J.C.; Lesch, A.; Bassetto, V.C.; Smirnov, E.; Cortés-Salazar, F.; Girault, H.H. Large-Scale Layer-by-Layer Inkjet Printing of Flexible Iridium-Oxide Based pH Sensors. J. Electroanal. Chem. 2018, 819, 384–390. [Google Scholar] [CrossRef] [Scilit]
  54. Dong, Q.; Song, D.; Huang, Y.; Xu, Z.; Chapman, J.H.; Willis, W.S.; Li, B.; Lei, Y. High-Temperature Annealing Enabled Iridium Oxide Nanofibers for Both Non-Enzymatic Glucose and Solid-State pH Sensing. Electrochim. Acta 2018, 281, 117–126. [Google Scholar] [CrossRef] [Scilit]
  55. Burke, L.D.; Lyons, M.E.; O’Sullivan, E.J.M.; Whelan, D.P. Influence of hydrolysis on the redox behavior of hydrous oxide films. J. Electroanal. Chem. 1981, 122, 403–407. [Google Scholar] [CrossRef]
  56. Katsube, T.; Lauks, I.; Zemel, J.N. pH-Sensitive Sputtered Iridium Oxide Films. Sens. Actuators 1981, 2, 399–410. [Google Scholar] [CrossRef] [Scilit]
  57. Virtanen, P.; Gommers, R.; Oliphant, T.E.; Haberland, M.; Reddy, T.; Cournapeau, D.; Burovski, E.; Peterson, P.; Weckesser, W.; Bright, J.; et al. SciPy 1.0: Fundamental Algorithms for Scientific Computing in Python. Nat. Methods 2020, 17, 261–272. [Google Scholar] [CrossRef] [Scilit]
  58. Britton, H.T.S.; Robinson, R.A. CXCVIII.—Universal Buffer Solutions and the Dissociation Constant of Veronal. J. Chem. Soc. 1931, 134, 1456–1462. [Google Scholar] [CrossRef] [Scilit]
  59. Ebihara, A.; Kawamoto, S.; Shibata, N.; Yamaguchi, T.; Suzuki, F.; Nakagawa, T. Development of a Modified Britton-Robinson Buffer with Improved Linearity in the Alkaline pH Region. Biojournal Sci. Technol. 2016, 3, 1–9. [Google Scholar]
  60. Steegstra, P.; Busch, M.; Panas, I.; Ahlberg, E. Revisiting the Redox Properties of Hydrous Iridium Oxide Films in the Context of Oxygen Evolution. J. Phys. Chem. C 2013, 117, 20975–20981. [Google Scholar] [CrossRef] [Scilit]
  61. Honeychurch, M.J.; Rechnitz, G.A. Voltammetry of Adsorbed Molecules. Part 1: Reversible Redox Systems. Electroanalysis 1998, 10, 285–293. [Google Scholar] [CrossRef]
  62. Clark, R.A.; Bowden, E.F. Voltammetric Peak Broadening for Cytochrome c /Alkanethiolate Monolayer Structures: Dispersion of Formal Potentials. Langmuir 1997, 13, 559–565. [Google Scholar] [CrossRef] [Scilit]
  63. Rowe, G.K.; Carter, M.T.; Richardson, J.N.; Murray, R.W. Consequences of Kinetic Dispersion on the Electrochemistry of an Adsorbed Redox-Active Monolayer. Langmuir 1995, 11, 1797–1806. [Google Scholar] [CrossRef] [Scilit]
  64. Harris, C.R.; Millman, K.J.; Van Der Walt, S.J.; Gommers, R.; Virtanen, P.; Cournapeau, D.; Wieser, E.; Taylor, J.; Berg, S.; Smith, N.J.; et al. Array Programming with NumPy. Nature 2020, 585, 357–362. [Google Scholar] [CrossRef] [Scilit]
  65. Biscani, F.; Izzo, D. A Parallel Global Multiobjective Framework for Optimization: Pagmo. J. Open Source Softw. 2020, 5, 2338. [Google Scholar] [CrossRef] [Scilit]
  66. Seabold, S.; Perktold, J. Statsmodels: Econometric and Statistical Modeling with Python; SciPy: Austin, TX, USA, 2010; pp. 92–96. [Google Scholar]
  67. Hunter, J.D. Matplotlib: A 2D Graphics Environment. Comput. Sci. Eng. 2007, 9, 90–95. [Google Scholar] [CrossRef] [Scilit]
  68. Crameri, F. Geodynamic Diagnostics, Scientific Visualisation and StagLab 3.0. Geosci. Model Dev. 2018, 11, 2541–2562. [Google Scholar] [CrossRef] [Scilit]
  69. Crameri, F.; Shephard, G.E.; Heron, P.J. The Misuse of Colour in Science Communication. Nat. Commun. 2020, 11, 5444. [Google Scholar] [CrossRef] [Scilit]
  70. Conway, B.E.; Mozota, J. Surface and Bulk Processes at Oxidized Iridium Electrodes—II. Conductivity-Switched Behaviour of Thick Oxide Films. Electrochim. Acta 1983, 28, 9–16. [Google Scholar] [CrossRef] [Scilit]
  71. Burke, L.D.; Whelan, D.P. A voltammetric investigation of the charge storage reactions of hydrous iridium oxide layers. J. Electroanal. Chem. 1984, 162, 121–141. [Google Scholar] [CrossRef] [Scilit]
  72. Steegstra, P.; Ahlberg, E. Influence of Oxidation State on the pH Dependence of Hydrous Iridium Oxide Films. Electrochim. Acta 2012, 76, 26–33. [Google Scholar] [CrossRef] [Scilit]
  73. Chaisiwamongkhol, K.; Batchelor-McAuley, C.; Compton, R.G. Optimising Amperometric pH Sensing in Blood Samples: An Iridium Oxide Electrode for Blood pH Sensing. Analyst 2019, 144, 1386–1393. [Google Scholar] [CrossRef] [Scilit]
  74. Chen, Y.; Li, X.; Li, D.; Batchelor-McAuley, C.; Compton, R.G. A Simplified Methodology: pH Sensing Using an in Situ Fabricated Ir Electrode under Neutral Conditions. J. Solid State Electrochem. 2021, 25, 2821–2833. [Google Scholar] [CrossRef] [Scilit]
  75. Jansod, S.; Afshar, M.; Crespo, G.; Bakker, E. Alkalinization of Thin Layer Samples with a Selective Proton Sink Membrane Electrode for Detecting Carbonate by Carbonate Selective Electrodes. Anal. Chem. 2016, 88, 3444–3448. [Google Scholar] [CrossRef] [Scilit]
  76. Wiorek, A.; Cuartero, M.; De Marco, R.; Crespo, G. Polyaniline Films as Electrochemical-Proton Pump for Acidification of Thin Layer Samples. Anal. Chem. 2019, 91, 14951–14959. [Google Scholar] [CrossRef] [Scilit]
  77. Kinoshita, K.; Madou, M.J. Electrochemical Measurements on Pt, Ir, and Ti Oxides as pH Probes. J. Electrochem. Soc. 1984, 131, 1089–1094. [Google Scholar] [CrossRef] [Scilit]
  78. Hitchman, M.L.; Ramanathan, S. Evaluation of Iridium Oxide Electrodes Formed by Potential Cycling as pH Probes. Analyst 1988, 113, 35. [Google Scholar] [CrossRef] [Scilit]
  79. Tarlov, M.J.; Semancik, S.; Kreider, K.G. Mechanistic and Response Studies of Iridium Oxide pH Sensors. Sens. Actuators B Chem. 1990, 1, 293–297. [Google Scholar] [CrossRef] [Scilit]
  80. Kreider, K. Iridium Oxide Thin-Film Stability in High-Temperature Corrosive Solutions. Sens. Actuators B Chem. 1991, 5, 165–169. [Google Scholar] [CrossRef] [Scilit]
  81. Hitchman, M.L.; Ramanathan, S. A Field-Induced Poising Technique for Promoting Convergence of Standard Electrode Potential Values of Thermally Oxidized Iridium pH Sensors. Talanta 1992, 39, 137–144. [Google Scholar] [CrossRef] [Scilit]
  82. VanHoudt, P.; Lewandowski, Z.; Little, B. Iridium Oxide pH Microelectrode. Biotech Bioeng. 1992, 40, 601–608. [Google Scholar] [CrossRef] [Scilit]
  83. Pásztor, K.; Sekiguchi, A.; Shimo, N.; Kitamura, N.; Masuhara, H. Iridium Oxide-Based Microelectrochemical Transistors for pH Sensing. Sens. Actuators B Chem. 1993, 12, 225–230. [Google Scholar] [CrossRef] [Scilit]
  84. Osaka, A.; Takatsuna, T.; Miura, Y. Iridium Oxide Films via Sol-Gel Processing. J. Non-Cryst. Solids 1994, 178, 313–319. [Google Scholar] [CrossRef] [Scilit]
  85. Kinlen, P.J.; Heider, J.E.; Hubbard, D.E. A Solid-State pH Sensor Based on a Nafion-Coated Iridium Oxide Indicator Electrode and a Polymer-Based Silver Chloride Reference Electrode. Sens. Actuators B Chem. 1994, 22, 13–25. [Google Scholar] [CrossRef] [Scilit]
  86. Kreider, K.G.; Tarlov, M.J.; Cline, J.P. Sputtered Thin-Film pH Electrodes of Platinum, Palladium, Ruthenium, and Iridium Oxides. Sens. Actuators B Chem. 1995, 28, 167–172. [Google Scholar] [CrossRef] [Scilit]
  87. Hendrikse, J.; Olthuis, W.; Bergveld, P. A Drift Free Nernstian Iridium Oxide pH Sensor. In Proceedings of the Proceedings of International Solid State Sensors and Actuators Conference (Transducers’97); IEEE: Chicago, IL, USA, 1997; Volume 2, pp. 1367–1370. [Google Scholar]
  88. Marzouk, S.A.M.; Ufer, S.; Buck, R.P.; Johnson, T.A.; Dunlap, L.A.; Cascio, W.E. Electrodeposited Iridium Oxide pH Electrode for Measurement of Extracellular Myocardial Acidosis during Acute Ischemia. Anal. Chem. 1998, 70, 5054–5061. [Google Scholar] [CrossRef] [Scilit]
  89. Wipf, D.O.; Ge, F.; Spaine, T.W.; Baur, J.E. Microscopic Measurement of pH with Iridium Oxide Microelectrodes. Anal. Chem. 2000, 72, 4921–4927. [Google Scholar] [CrossRef] [Scilit]
  90. Hara, M.; Waraksa, C.C.; Lean, J.T.; Lewis, B.A.; Mallouk, T.E. Photocatalytic Water Oxidation in a Buffered Tris(2,2′-Bipyridyl)Ruthenium Complex-Colloidal IrO2 System. J. Phys. Chem. A 2000, 104, 5275–5280. [Google Scholar] [CrossRef] [Scilit]
  91. Wang, M.; Yao, S. Carbonate-Melt Oxidized Iridium Wire for pH Sensing. Electroanalysis 2003, 15, 1606–1615. [Google Scholar] [CrossRef] [Scilit]
  92. Wang, M.; Yao, S.; Madou, M. A Long-Term Stable Iridium Oxide pH Electrode. Sens. Actuators B Chem. 2002, 81, 313–315. [Google Scholar] [CrossRef] [Scilit]
  93. Marzouk, S.A.M. Improved Electrodeposited Iridium Oxide pH Sensor Fabricated on Etched Titanium Substrates. Anal. Chem. 2003, 75, 1258–1266. [Google Scholar] [CrossRef] [Scilit]
  94. Ges, I.; Ivanov, B.; Schaffer, D.; Lima, E.; Werdich, A.; Baudenbacher, F. Thin-Film IrO pH Microelectrode for Microfluidic-Based Microsystems. Biosens. Bioelectron. 2005, 21, 248–256. [Google Scholar] [CrossRef] [Scilit]
  95. Juodkazytė, J.; Šebeka, B.; Valsiunas, I.; Juodkazis, K. Iridium Anodic Oxidation to Ir(III) and Ir(IV) Hydrous Oxides. Electroanalysis 2005, 17, 947–952. [Google Scholar] [CrossRef] [Scilit]
  96. Marshall, A.; Børresen, B.; Hagen, G.; Sunde, S.; Tsypkin, M.; Tunold, R. Iridium Oxide-Based Nanocrystalline Particles as Oxygen Evolution Electrocatalysts. Russ. J. Electrochem 2006, 42, 1134–1140. [Google Scholar] [CrossRef] [Scilit]
  97. O’Hare, D.; Parker, K.H.; Winlove, C.P. Metal–Metal Oxide pH Sensors for Physiological Application. Med. Eng. Phys. 2006, 28, 982–988. [Google Scholar] [CrossRef] [Scilit]
  98. El-Giar, E.E.-D.M.; Wipf, D.O. Microparticle-Based Iridium Oxide Ultramicroelectrodes for pH Sensing and Imaging. J. Electroanal. Chem. 2007, 609, 147–154. [Google Scholar] [CrossRef] [Scilit]
  99. Mayorga Martinez, C.C.; Madrid, R.E.; Felice, C.J. A pH Sensor Based on a Stainless Steel Electrode Electrodeposited With Iridium Oxide. IEEE Trans. Educ. 2009, 52, 133–136. [Google Scholar] [CrossRef] [Scilit]
  100. Elsen, H.A.; Monson, C.F.; Majda, M. Effects of Electrodeposition Conditions and Protocol on the Properties of Iridium Oxide pH Sensor Electrodes. J. Electrochem. Soc. 2009, 156, F1. [Google Scholar] [CrossRef] [Scilit]
  101. Zimer, A.M.; Lemos, S.G.; Pocrifka, L.A.; Mascaro, L.H.; Pereira, E.C. Needle-like IrO/Ag Combined pH Microelectrode. Electrochem. Commun. 2010, 12, 1703–1705. [Google Scholar] [CrossRef] [Scilit]
  102. Park, J.; Kim, J.; Quan, H. A Surface Renewable Iridium Oxide–Glass Composite Hydrogen Ion Electrode. Microchem. J. 2010, 95, 102–106. [Google Scholar] [CrossRef] [Scilit]
  103. Carroll, S.; Baldwin, R.P. Self-Calibrating Microfabricated Iridium Oxide pH Electrode Array for Remote Monitoring. Anal. Chem. 2010, 82, 878–885. [Google Scholar] [CrossRef] [Scilit]
  104. Huang, W.-D.; Cao, H.; Deb, S.; Chiao, M.; Chiao, J.C. A Flexible pH Sensor Based on the Iridium Oxide Sensing Film. Sens. Actuators A Phys. 2011, 169, 1–11. [Google Scholar] [CrossRef] [Scilit]
  105. Steegstra, P.; Ahlberg, E. In Situ pH Measurements with Hydrous Iridium Oxide in a Rotating Ring Disc Configuration. J. Electroanal. Chem. 2012, 685, 1–7. [Google Scholar] [CrossRef] [Scilit]
  106. Nguyen, C.M.; Huang, W.-D.; Rao, S.; Cao, H.; Tata, U.; Chiao, M.; Chiao, J.-C. Sol-Gel Iridium Oxide-Based pH Sensor Array on Flexible Polyimide Substrate. IEEE Sens. J. 2013, 13, 3857–3864. [Google Scholar] [CrossRef] [Scilit]
  107. Prats-Alfonso, E.; Abad, L.; Casañ-Pastor, N.; Gonzalo-Ruiz, J.; Baldrich, E. Iridium Oxide pH Sensor for Biomedical Applications. Case Urea–Urease in Real Urine Samples. Biosens. Bioelectron. 2013, 39, 163–169. [Google Scholar] [CrossRef] [Scilit]
  108. Huang, F.; Jin, Y.; Wen, L.; Mu, D.; Cui, M. Effects of Thermal Oxidation Cycle Numbers and Hydration on IrOx pH Sensor. J. Electrochem. Soc. 2013, 160, B184–B191. [Google Scholar] [CrossRef] [Scilit]
  109. Contu, F.; Vega-Arroyo, M.; Taylor, R. The Fabrication and Calibration of an Iridium pH Micro-Sensor for Biological Applications. Int. J. Mater. Sci. 2014, 4, 8. [Google Scholar] [CrossRef] [Scilit]
  110. Kuo, L.-M.; Chou, Y.-C.; Chen, K.-N.; Lu, C.-C.; Chao, S. A Precise pH Microsensor Using RF-Sputtering IrO2 and Ta2O5 Films on Pt-Electrode. Sens. Actuators B Chem. 2014, 193, 687–691. [Google Scholar] [CrossRef] [Scilit]
  111. Kim, T.Y.; Yang, S. Fabrication Method and Characterization of Electrodeposited and Heat-Treated Iridium Oxide Films for pH Sensing. Sens. Actuators B Chem. 2014, 196, 31–38. [Google Scholar] [CrossRef] [Scilit]
  112. Huang, X.; Ren, Q.; Yuan, X.; Wen, W.; Chen, W.; Zhan, D. Iridium Oxide Based Coaxial pH Ultramicroelectrode. Electrochem. Commun. 2014, 40, 35–37. [Google Scholar] [CrossRef] [Scilit]
  113. Zimer, A.M.; Medina da Silva, M.; Machado, E.G.; Varela, H.; Mascaro, L.H.; Pereira, E.C. Development of a Versatile Rotating Ring-Disc Electrode for in Situ pH Measurements. Anal. Chim. Acta 2015, 897, 17–23. [Google Scholar] [CrossRef] [Scilit]
  114. Yu, J.; Khalil, M.; Liu, N.; Lee, R. Iridium Oxide-Based Chemical Sensor for in Situ pH Measurement of Oilfield-Produced Water under Subsurface Conditions. Ionics 2015, 21, 855–861. [Google Scholar] [CrossRef] [Scilit]
  115. Khalil, M.; Wang, S.; Yu, J.; Lee, R.L.; Liu, N. Electrodeposition of Iridium Oxide Nanoparticles for pH Sensing Electrodes. J. Electrochem. Soc. 2016, 163, B485–B490. [Google Scholar] [CrossRef] [Scilit]
  116. Uria, N.; Abramova, N.; Bratov, A.; Muñoz-Pascual, F.-X.; Baldrich, E. Miniaturized Metal Oxide pH Sensors for Bacteria Detection. Talanta 2016, 147, 364–369. [Google Scholar] [CrossRef] [Scilit]
  117. Rouhi, J.; Kakooei, S.; Ismail, M.C.; Karimzadeh, R.; Mahmood, M.R. Development of Iridium Oxide Sensor for Surface pH Measurement of a Corroding Metal under Deposit. Int. J. Electrochem. Sci. 2017, 12, 9933–9943. [Google Scholar] [CrossRef] [Scilit]
  118. Yang, J.; Kwak, T.J.; Zhang, X.; McClain, R.; Chang, W.-J.; Gunasekaran, S. Digital pH Test Strips for In-Field pH Monitoring Using Iridium Oxide-Reduced Graphene Oxide Hybrid Thin Films. ACS Sens. 2016, 1, 1235–1243. [Google Scholar] [CrossRef] [Scilit]
  119. Yang, X.; De Angeli, M.; Fu, J.; Dubey, S.; Chiao, J.-C. Miniature pH Sensors on Ultra-Flexible Substrates. In Proceedings of the 2017 IEEE SENSORS; IEEE: Glasgow, UK, 2017; pp. 1–3. [Google Scholar]
  120. Ratanaporncharoen, C.; Tabata, M.; Kitasako, Y.; Ikeda, M.; Goda, T.; Matsumoto, A.; Tagami, J.; Miyahara, Y. pH Mapping on Tooth Surfaces for Quantitative Caries Diagnosis Using Micro Ir/IrOx pH Sensor. Anal. Chem. 2018, 90, 4925–4931. [Google Scholar] [CrossRef] [Scilit]
  121. Pan, Y.; Sun, Z.; He, H.; Li, Y.; You, L.; Zheng, H. An Improved Method of Preparing Iridium Oxide Electrode Based on Carbonate-Melt Oxidation Mechanism. Sens. Actuators B Chem. 2018, 261, 316–324. [Google Scholar] [CrossRef] [Scilit]
  122. Shahrestani, S.; Ismail, M.C.; Kakooei, S.; Beheshti, M.; Zabihiazadboni, M.; Zavareh, M.A. Iridium Oxide pH Sensor Based on Stainless Steel Wire for pH Mapping on Metal Surface. IOP Conf. Ser. Mater. Sci. Eng. 2018, 328, 012014. [Google Scholar] [CrossRef] [Scilit]
  123. Park, J.; Jeong, Y.; Kim, J.; Gu, J.; Wang, J.; Park, I. Biopsy Needle Integrated with Multi-Modal Physical/Chemical Sensor Array. Biosens. Bioelectron. 2019, 148, 111822. [Google Scholar] [CrossRef] [Scilit]
  124. Zea, M.; Moya, A.; Fritsch, M.; Ramon, E.; Villa, R.; Gabriel, G. Enhanced Performance Stability of Iridium Oxide-Based pH Sensors Fabricated on Rough Inkjet-Printed Platinum. ACS Appl. Mater. Interfaces 2019, 11, 15160–15169. [Google Scholar] [CrossRef] [Scilit]
  125. Mingels, R.H.G.; Kalsi, S.; Cheong, Y.; Morgan, H. Iridium and Ruthenium Oxide Miniature pH Sensors: Long-Term Performance. Sens. Actuators B Chem. 2019, 297, 126779. [Google Scholar] [CrossRef] [Scilit]
  126. Yang, X.; Chawang, K.; Chiao, J.-C. Wearable Iridium Oxide pH Sensors for Sweat pH Measurements. In Proceedings of the 2019 IEEE SENSORS; IEEE: Montreal, QC, Canada, 2019; pp. 1–4. [Google Scholar]
  127. Liu, B.; Zhang, J. A Ruthenium Oxide and Iridium Oxide Coated Titanium Electrode for pH Measurement. RSC Adv. 2020, 10, 25952–25957. [Google Scholar] [CrossRef] [Scilit]
  128. Huang, F.; Wang, Q.; Bi, P.; Yu, Z.; Wang, W.; Wang, S.; Wen, L.; Jin, Y. Application of Cyclic Thermal-Oxidized IrOx Electrode in pH Detection of Zn/Steel Galvanic Process in 3.5 Wt.% NaCl Solution. J. Electrochem. Soc. 2020, 167, 167501. [Google Scholar] [CrossRef] [Scilit]
  129. Sun, Z.; Ma, Q.; Wang, Y.; Pan, Y. Effects of Structures on the Sensing Properties of Long-Term Stable IrOx pH Electrodes. J. Electrochem. Soc. 2020, 167, 047501. [Google Scholar] [CrossRef] [Scilit]
  130. Zhou, L.; Cheng, C.; Li, X.; Ding, J.; Liu, Q.; Su, B. Nanochannel Templated Iridium Oxide Nanostructures for Wide-Range pH Sensing from Solutions to Human Skin Surface. Anal. Chem. 2020, 92, 3844–3851. [Google Scholar] [CrossRef] [Scilit]
  131. Marsh, P.; Manjakkal, L.; Yang, X.; Huerta, M.; Le, T.; Thiel, L.; Chiao, J.-C.; Cao, H.; Dahiya, R. Flexible Iridium Oxide Based pH Sensor Integrated With Inductively Coupled Wireless Transmission System for Wearable Applications. IEEE Sens. J. 2020, 20, 5130–5138. [Google Scholar] [CrossRef] [Scilit]
  132. Cheng, C.; Wu, Y.; Li, X.; An, Z.; Lu, Y.; Zhang, F.; Su, B.; Liu, Q. A Wireless, Ingestible pH Sensing Capsule System Based on Iridium Oxide for Monitoring Gastrointestinal Health. Sens. Actuators B Chem. 2021, 349, 130781. [Google Scholar] [CrossRef] [Scilit]
  133. Xi, Y.; Guo, Z.; Wang, L.; Xu, Q.; Ruan, T.; Liu, J. Fabrication and Characterization of Iridium Oxide pH Microelectrodes Based on Sputter Deposition Method. Sensors 2021, 21, 4996. [Google Scholar] [CrossRef] [Scilit]
  134. Chawang, K.; Bing, S.; Chiao, J.-C. Printable and Flexible Iridium Oxide-Based pH Sensor by a Roll-to-Roll Process. Chemosensors 2023, 11, 267. [Google Scholar] [CrossRef] [Scilit]
  135. Ruggiero, B.N.; Weidner, A.R.; Notestein, J.M.; Seitz, L.C. Observing Local pH Changes Using a Rotating Ring-Disk Electrode Functionalized with a Potentiometric pH-Sensing Probe. J. Phys. Chem. C 2023, 127, 20640–20651. [Google Scholar] [CrossRef] [Scilit]
  136. Yoon, E.S.; Park, H.J.; Kil, M.S.; Kim, J.; Lee, K.G.; Choi, B.G. Preparation of Nanopillar Array Electrode of Iridium Oxide for High Performance of pH Sensor and Its Real-time Sweat Monitoring. Bull. Korean Chem. Soc. 2023, 44, 528–535. [Google Scholar] [CrossRef] [Scilit]
  137. Lam, H.-H.; Phan, H.-P.; Trung, D.-B. Tailor-Made Iridium Oxide (IrOx) Coated Titanium Electrode: pH Sensor towards Potentiometric Titration. Chem. Eng. Trans. 2023, 106, 163–168. [Google Scholar] [CrossRef]
  138. Ashton, R.; Silver, C.D.; Bird, T.W.; Coulson, B.; Pratt, A.; Johnson, S. Enhancing the Repeatability and Sensitivity of Low-Cost PCB, pH-Sensitive Field-Effect Transistors. Biosens. Bioelectron. 2023, 227, 115150. [Google Scholar] [CrossRef] [Scilit]
  139. Xiao, W.; Dong, Q. Iridium Oxide and Cobalt Hydroxide Microfluidic-Based Potentiometric pH Sensor. Microchim Acta 2023, 190, 457. [Google Scholar] [CrossRef] [Scilit]
  140. Shen, Y.; Pan, Y.; Chen, Q.; Sun, Z.; Cai, W.-J.; Jiang, Z.-P.; Qu, F.; Wang, X.; Chen, Y. Development of SPEEK-Coated IrOx Sensor for High-Biomass Aquatic pH Monitoring. ACS Sens. 2024, 9, 3367–3376. [Google Scholar] [CrossRef] [Scilit]
  141. Marsh, P.; Huang, M.-H.; Xia, X.; Tran, I.; Atanassov, P.; Cao, H. Polarization Conforms Performance Variability in Amorphous Electrodeposited Iridium Oxide pH Sensors: A Thorough Surface Chemistry Investigation. Sensors 2024, 24, 962. [Google Scholar] [CrossRef] [Scilit]
  142. Jírů, J.; Hybášek, V.; Michalcová, A.; Korbelová, K.; Koláčný, L.; Fojt, J. Preparation, Characterization and Electrochemical Response of Nanostructured TiAlV with Potentiostatically Deposited IrOx as a pH Sensor for Rapid Detection of Inflammation. Chemosensors 2024, 12, 109. [Google Scholar] [CrossRef] [Scilit]
  143. Lee, J.; Soltis, I.; Tillery, S.A.; Lee, S.H.; Kim, H.; Yeo, W.-H. Long-Term Stable pH Sensor Array with Synergistic Bilayer Structure for 2D Real-Time Mapping in Cell Culture Monitoring. Biosens. Bioelectron. 2024, 254, 116223. [Google Scholar] [CrossRef] [Scilit]
  144. Chawang, K.; Bing, S.; Kwon, K.Y.; Chiao, J.-C. Miniaturized Iridium Oxide Microwire pH Sensor for Biofluid Sensing. Chemosensors 2024, 12, 168. [Google Scholar] [CrossRef] [Scilit]
  145. Tabata, M.; Yoshida, H.; Kobayashi, R.; Sugimura, Y.; Yamashita, N.; Negishi, Y.; Yamada, M.; Saito, T.; Miyahara, Y. Influence of Surface Roughness on the Response Time of Iridium Oxide pH Sensors. Sens. Actuators B Chem. 2025, 422, 136647. [Google Scholar] [CrossRef] [Scilit]
  146. Du, C.; Zhou, H.; Jee, S.; Mauzeroll, J. Good Readings Come in Threes: Understanding Electrodeposited Iridium Oxide for a Reproducible pH Microsensor Performance. Anal. Chem. 2025, 97, 16877–16884. [Google Scholar] [CrossRef] [Scilit]
  147. Sun, Y.; He, P.; Xu, Z.; Qin, G.; Liu, R.; Chen, B.; Sun, X.; Dong, Y. Iridium Oxide Thin Film Flexible Ocean pH Sensor Based on Electroplating Deposition Process. Microchem. J. 2025, 215, 114454. [Google Scholar] [CrossRef] [Scilit]
  148. Madhu, S.; Ramasamy, S.; Choi, J. Wireless IrO2 Electrochemical Sensor System for Real-Time pH Monitoring in Microliter Volumes. ACS Meas. Sci. Au 2026, 6, 238–248. [Google Scholar] [CrossRef] [Scilit]
  149. Sun, Y.; Chen, B.; He, P.; Xu, Z.; Dong, Y.; Liu, S. Mechanically Flexible pH Sensor Demonstrating High Sensitivity in Seawater and Long-Term Operational Stability for Ocean Environmental Applications. Colloids Surf. A Physicochem. Eng. Asp. 2026, 749, 141238. [Google Scholar] [CrossRef] [Scilit]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Article Metrics

Citations

Article Access Statistics

Multiple requests from the same IP address are counted as one view.