To the Editor:
Biodegradable fixation is becoming a viable alternative to traditional fixation techniques. In recent medical literature, a degree of success has been reported in several clinical trials involving long-term follow-up care of patients with absorbable fixation [1,2,3,4,5,6,7,8,9,10]. The most common absorbable materials in use are polyglycolide (Biofix (Acufex Microsurgical, Inc, Norwood, MA)) and poly-p-dioxanon (Orthosorb (Johnson & Johnson Orthopaedic, Inc, New Brunswick, NJ)). There are, however, complications involved in the use of these types of absorbable fixation devices, as evidenced by the following case report.
Biofix absorbable fixation products consisting of polyglycolide have a well documented foreignbody reaction profile. Sterile sinus formation at approximately 3 months postoperatively occurs at a rate of 3% to 8% [1,2]. Osteolytic changes in the pin tract of the implant occur in up to 50% of cases [3]. These complications generally have little effect on the healing process after surgery [4].
Orthosorb products consisting of poly-p-dioxanon, however, have only one previously documented foreign-body reaction in the literature [5].
The patient in this case was diagnosed with a delayed union caused by foreign-body reaction to Orthosorb pins. This was based on the histologic, radiographic, and clinical findings. Osteomyelitis was ruled out because of a negative histologic examination for infection and negative intraoperative cultures. Displacement caused by partial weightbearing could contribute to the clinical picture; however, a more significant bony callus and greater deviation of the fourth metatarsal head would be expected if this were the only etiology involved.
Case Report
A 40-year-old female with no pertinent medical history sought treatment for a longstanding tender callus under the fourth metatarsal head of the right foot. The patient had repeated debridements with accommodative padding with little or no relief. After several months of unsuccessful conservative therapy, the patient underwent an elevating V-osteotomy of the fourth metatarsal head. Fixation was accomplished with two 1.3-mm Orthosorb pins. These pins were placed in a crosswise fashion through the osteotomy site. Postoperative radiographs revealed adequate approximation of the osteotomy. No gapping or deviation of the capital fragment was seen. The patient was placed in a below the knee cast with a cane with partial weightbearing.
At the first postoperative visit 6 days later, slight erythema of the operative site was seen. No drainage or dehiscence was noted. The patient began empiric oral antibiotic therapy for a possible postoperative infection. Plain film radiographs revealed adequate apposition of the osteotomy with no lucency present (Figure 1). A below the knee cast was reapplied and the patient continued partial weightbearing with a cane. Two weeks postoperatively, the incision site had persisting local erythema with maceration present. No significant drainage was seen. The sutures were removed and a surgical shoe was dispensed. The patient continued the oral antibiotics.
Figure 1.
Anteroposterior radiograph of the right foot 6 days after surgery. Note the adequate apposition of the fourth metatarsal osteotomy fixated with Orthosorb pins.
Five weeks postoperatively, the incision area was ulcerated with increased erythema. Serous drainage was noted from a superficial sinus that did not track to deep tissues. The patient related increased pain on ambulation in the area of the fourth metatarsal head of the right foot for the past several days. Cultures of the drainage revealed no growth.
Seven weeks postoperatively, the incision was dry without sinus, local edema, or erythema. The patient related continued pain in the location of the osteotomy during cane-assisted ambulation. Radiographs at this time revealed a transverse area of lucency at the osteotomy site of the fourth metatarsal (Figure 2). No significant displacement or rotation of the fourth metatarsal head was present. No apparent bony callus formation was noted. The patient was continued on the current treatment plan with suspicion of foreign-body reaction as the cause of symptoms and radiographic changes the patient was experiencing.
Figure 2.
Anteroposterior radiograph of the right foot 7 weeks after surgery. Note the lucent area in the fourth metatarsal osteotomy site.
One month later, in the eleventh week after surgery, the patient related gradually increasing levels of pain at the operative site aggravated by ambulation. Significant pain could be elicited on palpation of the plantar aspect of the metatarsal head, which was not previously present. The incision area appeared to be healed without drainage, erythema, or any sign of inflammation. Radiographs revealed increased area of lucency at the osteotomy site in comparison to the previous films (Figure 3). The osteotomy edges were separated by an approximately 5-mm gap, constituting a delayed union of the fourth metatarsal. No gross angulational or rotational deformities of the fourth metatarsal head were noted, however. An open reduction with internal fixation and a biopsy of the fourth metatarsal of the right foot were performed.
Figure 3.
Anteroposterior radiograph of the right foot 11 weeks after surgery. Note the increased size of the lucent area in the fourth metatarsal caused by the foreign-body reaction.
Intraoperatively, the affected metatarsal neck area was seen with hypertrophied fibrous tissue surrounding the delayed union site. The capital fragment was noted to have significant mobility against the shaft of the fourth metatarsal. Using a combination of blunt and sharp instrumentation, this tissue was removed and a serous exudate with multiple small, apparently osseus fragments was noted. Also, several small, transparent fragments of foreign-body material were observed in the delayed union site. Bony gapping consistent with previous radiographic measurements was present. The osteotomy edges appeared without necrosis which would be expected if infection were present.
An osteotome and mallet were used to resect the edges of the osteotomy site. This bone and fibrous tissue were sent for histopathology and culture and sensitivity. The gap between the osteotomy edges was evacuated of all debris and fibrous tissue. The osteotomy was then approximated, impacted, and fixated with 0.045 Kirschner wire. The patient was placed in a nonweightbearing below the knee cast.
The results of the surgical pathology report indicated that a focus of foreign-body reaction to a basophilic, suture-like material was present with bone, cartilage, and fibrous tissue. Multinucleated giant cells were noted with the presence of clear, refractile matter consistent with partially absorbed internal fixation material (Figure 4). No polymorphonuclear leukocytes were present. Cultures intraoperatively were negative, which in light of the histopathologic findings, ruled out infection as an etiology for the delayed union of the fourth metatarsal.
Figure 4.
Light photomicrograph of focal tissue reaction. Note the presence of a multinucleated giant cell (white arrow) and clear, basophilic, suture-like Orthosorb material (black arrow) (H&E × 200).
Summary
While it is evident that poly-p-dioxanon is less reactive than polyglycolide, recent studies, including the case above, have shown that Orthosorb pins are not exempt from the same type of foreign-body reaction as the Biofix products.
References
- HIRVENSALO, E; BOSTMAN, O; TORMALA, P. ET AL: Chevron osteotomy fixed with absorbable polyglycolide pins. Foot Ankle 1991, 11, 212. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- BOSTMAN, O; HIRVENSALO, E; VAINIONPAA, S. ET AL: Degradable polyglycolic rods for the internal fixation of displaced bimalleolar fractures. Int Orthop 1990, 14, 1. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- BOSTMAN, O. Osteolytic changes accompanying degrada-tion of absorbable fracture fixation implants. J Bone Joint Surg 1991, 73B, 679. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- GERBERT J: Effectiveness of absorbable fixation devices in Austin bunionectomies. JAPMA 1992, 82, 189. [CrossRef] [Scilit] [PubMed]
- KALLA, TP; JANZEN, DL. Orthosorb: a case of foreign-body reaction. J Foot Ankle Surg 1995, 34, 366. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- STRYCKER ML: Biodegradable internal fixation. J Foot Ankle Surg 1995, 34, 82. [CrossRef] [Scilit] [PubMed]
- LAVERY, LA; PETERSON, JD; POLLACK, R; et al. Risk of complications of first metatarsal head osteotomies with biodegradable pin fixation: Biofix versus Orthosorb. J Foot Ankle Surg 1994, 33, 334. [Google Scholar] [PubMed]
- BRUNETTI, VA; TREPAL, MJ. JULES KT: Fixation of the Austin osteotomy with bioresorbable pins. J Foot Surg 1991, 30, 56. [Google Scholar] [PubMed]
- FRIEND, GJ; GRACE, KS; STONE, HA. Cortical screws ver-sus absorbable pins for fixation of short z-bunionec-tomy. J Foot Ankle Surg 1994, 33, 411. [Google Scholar] [PubMed]
- HETHERINGTON, VJ; SHIELDS, SL; WILHELM, KR. ET AL: Absorbable fixation of first ray osteotomies. J Foot Ankle Surg 1994, 33, 290. [Google Scholar] [PubMed]