Abstract
The authors present a brief history of the development of the Ilizarov external fixator, and the classic uses associated with it. A preliminary study of 53 patients using 56 frames is presented. The Ilizarov external fixator is successfully used in these cases in place of traditional fixation.
Although the Ilizarov technique has been used by surgeons since the early 1950s, it has been underutilized in podiatric surgery. The basic technique, which uses the tension-stress effect, is useful in the correction of complex foot and ankle deformities. The authors have successfully used the Ilizarov technique because it provides excellent stability, yet is adaptable and allows the patient to bear weight postoperatively. This is an alternative to traditional methods of internal fixation, which require periods of non-weightbearing.
History of Technique
The Ilizarov technique of lengthening osseous and soft tissue structures is well known throughout the world. The concept of the tension-stress effect was born following the application of the external fixation apparatus as a side effect of stable fixation [1,2]. The need for reconstructive extremity surgery during and following World War II led to the development of this technique.
In 1951, Ilizarov began using Kirschner wires in conjunction with homemade ring devices for the stabilization of complex fractures. The ring devices of the time, although of crude construction, were similar to those of modern design with fewer holes. The technique involved the insertion of Kirschner wires through the fracture fragments and through the aspects proximal and distal to the fracture site. These Kirschner wires were then fastened to the external fixator using grooved buckles [3,4]. Following the application of the wires and rings, the rings were then connected with telescoping rods that created immobilization of the limb and stabilization of the fracture. Ilizarov continued his use of the frame with no components supplied by the medical industry.
In the early 1960s, Ilizarov began to use his device for limb lengthening following a Z-type osteotomy. He later eliminated the Z-type osteotomy in favor of a simple corticotomy. The corticotomy provided for smaller incisions and less dissection than the Zosteotomy, yet yielded a similar result. The idea of limb lengthening through distraction was developed when treating a patient with a painful short amputation stump and knee contractures. Ilizarov’s original plan was to perform a corticotomy and slowly distract the distal remaining bone in preparation for a later bone graft. The patient was to distract the stump through the telescoping rods attaching the rings together. Ilizarov then went on vacation, leaving the patient to continue the distraction. Following his return, radiographs were obtained of the patient’s stump. To Ilizarov’s surprise, the bone in the distraction site had filled in completely, eliminating the need for the planned bone graft. Lacking research facilities, Ilizarov began to experiment with this technique on several additional patients with successful results [3,4].
The Tension-stress Effect
The tension-stress effect, also known as the law of tension-stress, is the stimulation and active growth of osseous and soft tissue structures by gradual traction. This effect was first shown accidentally while distracting the stump from a below the knee amputation for the purpose of later bone grafting. Through canine experimentation, Ilizarov has proven that osseous and soft tissue structures can proliferate, resulting in elongation, if distracted at the recommended rate. When distracted correctly, these structures will show no signs of stretching, indicating that growth has occurred [1,2,5]. Tissues distracted slowly act as those seen around the growth centers in children. The ability to “turn on” a growth hormone-mediated process through the use of distraction has been proven in both human and canine models [1,2].
The first extensive model of the tension-stress effect was completed on 480 adult dogs in five stability groups. Corticotomies were performed with Ilizarov fixators of varying stability [1,2,6]. Distraction began 5 to 7 days after application and continued for 28 days. The results showed that osteogenic activity was directly correlated to the stability of the fixation. The canine limbs with the greatest mobility of the osseous fragments had the least osteogenic activity [1,2,6].
It was further noted that with preservation of periosteal and endosteal blood supply, the osteogenesis became more intense. The group of canine limbs with the greatest mobility of the osseous fragments had a high predilection for pseudoarthrosis. This pseudoarthrosis was also noted among those canines whose limbs were distracted too quickly in a longitudinal or transverse fashion [1,2,6].
Histologic examination of the canine osseous tissues reveals an elongation of the fibroblast-like cells found in the central growth zone or center of the distraction. These fibroblast cells are always oriented parallel to the vector of distraction. It was further observed that osteoblastic activity was occurring both proximally and distally to the central growth zone during distraction. Further examination of the soft tissue structures surrounding the osseous distraction site revealed a tissue genesis effect rather than tissue stretching [1,2,6].
Examination of the arterial system revealed the formation of new blood vessel networks in the direction of the distraction vector. Skeletal and smooth muscle showed signs of growth. New myofibrils and sarcomeres were identified in skeletal muscle and cytoplasmic organelles had developed in the smooth muscle. Nerve tissue exhibited growth through the proliferation of axons with the elongation of Schwann’s cells now forming a new myelin sheath. Finally, the epidermis showed mitotic features in the basal layers. With the understanding of the effects of the law of tension-stress, Ilizarov studied the proper distraction rate next [1,2,6].
The canine model for studying distraction rate consisted of 120 animals whose limbs were distracted with stable Ilizarov external fixators at different rates. One half of the canine osteotomies were completed with transection of the medullary canal, while the remaining animals were subjected to an osteoclasis technique leaving most of the medullary canal intact. Limbs distracted at 1 mm per day in four 0.25 mm intervals revealed a small regeneration zone as compared with animals where the distraction was performed at a faster rate per day, or at 1 mm per day in one interval. In animals that had undergone an osteotomy with destruction of the medullary canal, there was a higher tendency for fibrous tissue formation in the distraction site than with the osteoclasis technique [2].
Finally, with the use of an autodistractor set to distract 1 mm per day in 60 equal intervals, a lengthened extremity with soft tissues histologically identical to the normal canine tissues was produced. This proves that distraction-mediated osteogenesis and soft tissue genesis appear the same histologically as hormone-mediated growth with the proper frequency of distraction. These studies underline the importance of the rate of distraction, the stability of the extremity, and the effects of preserving the medullary canal and periosteal blood supply during the lengthening of extremities. Stimulation of tissue growth occurs according to the law of tension-stress and it is halted when the distraction period ends [1,2,8,9,10].
Uses of the Ilizarov External Fixation System
The goal in the application of the Ilizarov fixation system is to provide an effective method for stabilization and correction of deformities while using minimally traumatic techniques. More than 600 effective treatment methods have been developed with the invention of Ilizarov’s external fixator and his experimentation [7]. The main uses for the Ilizarov external fixator are the treatment of fractures, osteomyelitis, soft tissue contractures, limb lengthening, and complex foot deformities [2,4,6,9,10,11,12,13,14,15,16].
Fractures
The surgical treatments of complex injuries, including pilon, tibial plateau, and calcaneal fractures, are now more common occurrences in the western hemisphere because of the options for reconstruction of the extremity that exist with external fixation. Application of an external fixator allows for osseous stabilization while leaving room for debridement of necrotic tissues. Furthermore, soft tissue coverage can be performed around or through the fixator. Using an external fixator in cases of open fractures may reduce the number of follow-up surgical procedures. In the case of contaminated open fractures, where extensive osseous debridement is required, a bone transport technique can be used. This technique, often used in the case of osteomyelitis, preserves the length of the limb while creating a new segment of bone through corticotomy. The tension-stress effect is used here as in the case of limb lengthening. Additionally, with the use of bone transport, soft tissue can be transported to cover the defect resulting from the traumatic event [5,17].
Payley and Fischgrund [18] have reported the use of the Ilizarov external fixator for the reduction and fixation of calcaneal fractures. Their choice of external fixation, rather than internal fixation, is based on its ability to provide stabilization for weightbearing. The technique of Payley and Fischgrund involves the application of the external fixator followed by a small incision laterally for the visualization and manipulation of the fracture fragments.
The use of ligamentotaxis (the realignment of fractures through the use of soft tissue traction) with the Ilizarov technique allows for smaller incisions with less dissection for the reduction of fractures. Stable fixation of these fractures using Ilizarov external fixation allows for early range of motion, rehabilitation, and weightbearing. The use of component external fixation allows for modification without loss of stability during this process [19,20]. The use of ring fixators in combination with hybrid monorail systems for the stabilization of extremity fractures has been described recently by Weiner et al [21]. This form of fixation provides stability with a lower profile of the external fixator while still adhering to Ilizarov’s principles. The Ilizarov external fixator and its hybrids provide fracture fragment stability with immobilization of the limb even with early weightbearing.
Osteomyelitis
The treatment of osteomyelitis and septic nonunions often involves surgical debridement in combination with antibiotic therapy. In many cases, the use of osseous debridement in long bones causes a loss of stability in the extremity. Frequent debridements coupled with bone grafting techniques and internal fixation often require a nonweightbearing status. Stable external fixation allows for bone grafting or bone and soft tissue transport while allowing a weightbearing status [4,9]. Green [17] and Ilizarov have used bone transport and septic fusion techniques in an effort to obtain both an osseous union and clinical cure for the osteomyelitis. The use of bone grafting or bone transport in the case of an osteomyelitic defect will not sterilize the wound but will isolate the cultured bacteria, sending it into a dormant state. The use of bone transport increases the vascularity of the area so that arthrodesis can occur in the presence of bacteria [17,22]. If the patient refractures the transported area, the dormant bacteria would most likely begin to proliferate once again. In most cases, the patient lives free of symptoms throughout life [17]. Bone transport can be used in sterile nonunions, septic nonunions, and osteomyelitis [2,9].
Soft Tissue Contractures
A new strategy for soft tissue lengthening was developed by Ilizarov for the reduction of deformities and prevention of soft tissue contractures. The weightbearing characteristics of the fixator allow for immobilization of the affected limb while allowing motion at the surrounding joints. Through the use of the tension-stress effect, all soft tissues can be slowly lengthened with the proper distraction technique [2,23]. Grill and Franke [23] later used the fixator in the treatment of talipes equinovarus. The talipes equinovarus deformity was reduced with application of the fixator and distraction alone. The management of contractures, secondary to burns, has been described using the Ilizarov external fixator. The device is applied and distraction begins, creating soft tissue growth to a satisfactory result [13]. Hsu et al [16] have reported the use of the Ilizarov technique in the soft tissue realignment of a patient with hereditary progressive dystonia, also known as Segawa disease. The use of an external fixator may eliminate the need for surgical incisions in the treatment of soft tissue contractures.
Limb Lengthening
Limb lengthening was first accomplished in the early 1900s by Codivilla. Since that time, many techniques of extremity lengthening have been suggested, as reviewed by Paterson [24]. Currently, the techniques of Wagner [25], Ilizarov, and De Bastiani are used throughout the world for long bone elongation. The Wagner technique involves a full diaphyseal osteotomy with distraction of soft tissues followed by bone grafting with internal fixation. A third procedure is needed to remove the internal fixation [25,26,27]. De Bastiani et al [28,29] favor distraction of the epiphyseal plate (chondrodiatasis), and the use of an external monorail fixator.
Ilizarov used his external ring fixator for stabilization and distraction of the limb following corticotomy [1,2,30]. Although first credited with physeal distraction, Ilizarov used the corticotomy technique in children and adults so as not to disturb the natural growth centers [31]. With the Ilizarov external fixator, three-dimensional (3-D) distraction is accomplished. This allows for correction of angular deformity and limb length discrepancy with full weightbearing. The distraction rate is 1 mm per day in four 0.25-mm increments spaced throughout the day [2,30]. Following full correction, the fixator is dynamized prior to its removal. This entails the removal of tension from the wires and accessory rings from the limb, preventing refracture. The fixator may be removed following radiographic evidence of osseous incorporation [1,2,32].
Each of the three techniques used for limb lengthening involves the use of external fixation. Wagner [25] and De Bastiani [26] differ from Ilizarov by using unilateral monorail fixators, and allowing only partial weightbearing. In each case of osseous lengthening through distraction, the tension-stress effect, as described by Ilizarov, is used.
Complex Foot Deformities
Ilizarov external fixators have been used for the correction of a variety of complex foot deformities. Examples range from cavus-type deformities to reconstructive cases following trauma, osteomyelitis, bone tumors, and soft tissue contractures. In each case, the surgeons have selected the external fixator because of its adaptability, immobilization, rehabilitation characteristics, and the smaller incisions that can be used. In some cases, standard procedures for correcting rearfoot and midfoot deformities have been by obviated with procedures more suited to the use of the tension-stress effect and the rotational abilities of the external fixator. Most authors agree that the frame construction in the foot and ankle area is more complex than the midshaft tibia and femur, because of the multiplanar angulations often seen in the foot and ankle. The surgeon must account for multiple joints in a small region. Rosman and Brown [32] suggest constructing a 3-D rubber model of the leg preoperatively so that the fixator can be completely fabricated prior to surgery. This process provides time for preoperative surgical planning and reduces operating time.
Ilizarov [33], Grant et al [34], Laughlin and Calhoun [35], and Payley [14], and have each described multiple fixator constructions for foot and ankle deformities used at their facilities. These authors describe the U-shaped osteotomy for the correction of the rearfoot complex and the V-shaped osteotomy to correct the midfoot- to-rearfoot relationship [10,14,33,34,35]. These approaches to reconstructive foot and ankle surgery are tailored to external fixation rather than the use of internal fixation. However, many of the osseous procedures previously used in the reconstruction of foot and ankle deformities may still be used with the Ilizarov external fixator. The fixator is more adaptable than internal fixation. The fixator may be adjusted postoperatively by adding or repositioning Kirschner wires, and by compressing or distracting the fixator through the telescoping rods.
Complications
External fixation is noted throughout the literature as having a higher complication rate than internal fixation, because of the nature of external fixation. In external fixation, wires or half pins are used transdermally to create stabilization of osseous tissues and immobilization of the limb. However, while creating this stability, an open wound is present, and will be colonized by the normal skin flora, providing these organisms an opportunity to become pathogenic.
Complications have been categorized in the literature as minor and major. Minor complications include: pin tract infection, pin tract irritation, and pain associated with pin positioning [18,23,31,36,37]. Major complications have included: nonunion, delayed union, malunion, deep infection, osteomyelitis, nerve or vascular injury, joint contractures or subluxation, wound dehiscence, and reflex sympathetic dystrophy [18,37,38].
Methods and Materials
Ilizarov external fixators have been used by the authors since 1992 to correct various forms of foot, ankle, and leg pathology in 53 patients. A total of 56 external fixators has been applied to 24 male patients and 29 female patients. Three children, two females, age 8 and 14 years, and one male, age 14 years, accounted for three fixator applications. Presently, eight of the 53 patients remained in nine frames. Of the 56 Ilizarov external fixators used, one of the authors performed 20 ankle fusions, 12 tibial lengthening, one fibular lengthening, eight triple arthrodeses, five subtalar fusions, five tibiocalcaneal fusions, three tibial nonunion repairs, one soft tissue lengthening, one Cole osteotomy, one pilon fracture, one calcaneal distraction, one Charcot foot and ankle reconstruction, and repair of one malunion (Table 1).
Table 1.
Diagnosis and Treatment.
Surgical Technique
In each case, the patient was given the choice between internal fixation and external fixation. Radiographs were obtained preoperatively for accurate construction of the external fixators. In the case of limb length discrepancy, computed tomography scout films were obtained comparing angulation and length of the affected limb against the opposite limb. A preoperative surgical planning session occurred prior to each case and all fixators were individually constructed before surgery. The fixator used in the triple arthrodesis consisted of one foot plate and one ring attached by four threaded rods (Fig. 1). The fixator for ankle fusion consisted of two rings and one foot plate connected by four threaded rods (Fig. 2). The fixator construction for limb lengthening procedures consisted of at least three rings and as many as five rings connected by four threaded rods. Hinges were used as necessary for angular correction and two mechanical motors for distraction in most cases (Fig. 3). All external fixators for foot and ankle procedures consisted of at least one foot plate and one ring. Wires were inserted according to the anatomical guidelines stated by Green [36], Barral et al [39], and Catagni [40]. Intraoperative fluoroscopy and radiographs were used to determine osseous positioning and positioning of the wires.
Figure 1.
Ilizarov fixator for triple arthrodesis.
Figure 2.
Ilizarov fixator used in ankle arthrodesis.
Figure 3.
Ilizarov fixator used in angular correction and limb lengthening.
Radiographs were obtained postoperatively to ensure proper positioning of the fixator and the limb. Each patient was permitted to walk on the fixator with weightbearing immediately following surgery. All patients remained in the hospital for 2 to 3 days with the first dressing change on the day of discharge. The patients then changed their own dressings on a daily basis for the length of time the fixator was in place. Additionally, patients undergoing lengthening-type procedures were instructed in the use of the mechanical motors attached to their fixators and were responsible for the distraction.
Lengthening began 5 to 7 days postoperatively at a rate of 0.25 mm four times per day, totaling 1 mm per day. Radiographs were then obtained every 2 weeks with the exception of patients with lengthening procedures, in which cases radiographs were obtained weekly. Fixators were removed following radiographic and clinical evidence of union. All fixators were dynamized prior to their removal. External fixators were used on the patient for an average period of 3.8 months with the minimum being 6 weeks and the maximum 7 months. Follow-up care continued on all patients for 2.5 years radiographically (Table 2).
Table 2.
Data on Ilizarov Fixated Procedures
Results
Final results are available for 45 patients using 47 external fixators. Osseous union was achieved in 41 of 45 patients yielding a 91% rate of arthrodesis with 43 fixators. Each of these patients continues to walk at this time, with some requiring the use of orthotic devices.
The authors have recognized complications in three categories: major, intermediate, and minor. Major complications are composed of failure to achieve osseous fusion, osseous failure, or loss of limb. Two patients have undergone a second application of the external fixator. One patient is able to walk symptom-free on a tibial pseudoarthrosis. One patient is currently using a double upright brace while considering further treatment for her symptomatic tibiocalcaneal pseudoarthrosis. Following failure to obtain osseous union in an ankle fusion, one patient successfully underwent refusion with internal fixation. She is currently undergoing a limblengthening procedure with the external fixator. One patient elected to have a below the knee amputation following a 4-month immobilization of her limb in the fixator. This patient had undergone three previous surgeries resulting in 3 years of immobilization.
Intermediate complications included three wound dehiscences, one posterior ulceration secondary to edema and the external fixator, and two nerve injuries secondary to wire placement. The wound dehiscences on three patients were allowed to granulate until healed with only superficial debridement and wound care. The ulcer secondary to edema was treated by removal of the posterior portion of the ring. This wound was also allowed to granulate until healed. Two patients required a second procedure for the removal of neuromas caused by pin placement. One patient has required a fixator revision following the breakage of multiple wires, and completed osseous fusion.
Minor complications include pin-tract irritation, pain from pin placement, and pin-tract infection. All patients initially experienced pin-tract irritation symptoms. These symptoms subsided as the patient recovered postoperatively. Pin-tract infections were treated symptomatically with oral antibiotics as necessary. No pins required removal secondary to infection. Several patients have broken one wire requiring only removal of that wire. All patients report satisfaction with their ambulatory status using the external fixator.
Discussion
External fixation is now regaining popularity. The answer to all reconstructive and trauma surgery cannot be found within the Ilizarov external fixator. A well prepared surgical plan and a full understanding of the patient’s deformity is necessary for choosing the proper fixation. The authors have used the Ilizarov external fixation system because of its stability, adaptability, and weightbearing characteristics. Patients who need reconstructive lower extremity surgery, and who would experience difficulty in complying with a nonweightbearing recovery period, may benefit from the use of the external fixator. Excellent fixation was obtained intraoperatively and maintained postoperatively in most of the authors’ cases.
The majority of patients experience some type of minor complication with the external fixation system. However, the benefits of the system outweigh the negatives in most cases. The authors have had success using external fixation in the arthrodesis of joints, angular correction, lengthening of limbs, correction of soft tissue contractures, and other procedures. A learning curve exists and increased experience with the Ilizarov fixator has diminished the authors’ complication rate. Additionally, the operative time has decreased as has the size of the incisions in each of the procedures. In cases of rearfoot and ankle reconstruction, a two-incision approach has been modified to a one-incision technique. A oneincisional approach allows for a reduced amount of soft tissue dissection. This is secondary to the improved use of the fluoroscope by the authors and use of the adaptability found with the Ilizarov external fixator.
Complications can be influenced by factors other than the external fixator itself. The patient must comply with dressing changes, observation of pin tracts, and distraction or compression techniques as prescribed. Procedures such as Charcot foot and ankle reconstruction provide complexity by including multiple fusion sites and increased dissection. Reconstructive surgery, especially secondary or tertiary procedures, carry a higher complication rate than virgin procedures. These complications include wound dehiscence, painful scar formation, formation of hematomas and seromas, infection, and nonunion or avascular necrosis of bone.
The Ilizarov external fixator is not a substitute for internal fixation universally. Once a surgeon has mastered all the types of fixation, the external fixation becomes another tool for the best management of the limb deformity. Not all patients will psychologically tolerate an external fixator of any kind, and not all deformities are corrected best with external fixation. Ilizarov corrected many deformities with external fixation in the 1950s because that was the only method available, not necessarily because it was the best method available. With the new forms of absorbable fixation, stainless steel internal fixation, and the Ilizarov external fixator, the modern physician has a large inventory of fixation methods at his or her disposal. The Ilizarov external fixator is an excellent method of fixation that provides stability, adaptability, and full weightbearing and should be considered as a fixation option.
Acknowledgments
Kimberly LaBianco, MSSA, and Nancy Vickers, BM.
References
- ILIZAROV GA: The tension-stress effect on the genesis and growth of tissues: part I. the influence of stability of fixation and soft tissue preservation. Clin Orthop 238: 249, 1989.
- ILIZAROV GA: The tension-stress effect on the genesis and growth of tissues. part II. the influence of stability of the rate and frequency of distraction. Clin Orthop 239: 263, 1989.
- GOLYAKHOVSKY V, ILIZAROV G: The magician from Kurgan. Bull Hosp Jt Dis 48: 12, 1988.
- GREEN S: Editorial comment. Clin Orthop 280: 2, 1992.
- SCHWARTSMAN V, MCMURRAY M, MARTIN S: The Ilizarov method: the basics. Contemp Orthop 19: December 1989.
- LEHMAN W, GRANT A, ATAR D: “The Use of Distraction Osteogenesis (Ilizarov) in Complex Foot Deformities,” in Disorders of the Foot and Ankle, ed by MH Jahss, WB Saunders, Philadelphia, 1991.
- ILIZAROV G: The principals of the Ilizarov method. Bull Hosp Jt Dis 48: 1, 1988.
- ILIZAROV GA: The Tension-Stress Effect on the Genesis and Growth of Tissues, ed by GA Ilizarov, Springer-Verlag, Berlin, 1992.
- DENDRINOS GK, KONTOS S, LYRITSIS E: Use of the Ilizarov technique for treatment of non-union of the tibia associated with infection. J Bone Joint Surg 77A: 835, 1995.
- GRANT AD, ATAR D, LEHMAN WB: The Ilizarov technique in correction of complex foot deformities. Clin Orthop 280: 94, 1992.
- JOHNSON EE, WELTMER J: The Ilizarov ankle arthrodesis. Clin Orthop 280: 160, 1992.
- HAWKINS BJ, LANGERMAN RJ, ANGER DM, ET AL: The Ilizarov technique in ankle fusion. Clin Orthop 303: 217, 1994.
- CALHOUN JH, EVANS EB, HERNDON DN: Techniques for the management of burn contractures with the Ilizarov fixator. Clin Orthop 280: 217, 1992.
- PAYLEY D: The correction of complex foot deformities using Ilizarov’s distraction osteotomies. Clin Orthop 293: 97, 1993.
- KARGER C, GUILLE JT, BOWEN R: Lengthening of congenital lower limb deficiencies. Clin Orthop 291: 236, 1993.
- HSU KY, KUO KN, HSU WW, ET AL: Correction of foot deformity by the Ilizarov method in a patient with Segawa disease. Clin Orthop 314: 199, 1995.
- GREEN SA: Osteomyelitis: the Ilizarov perspective. Orthop Clin North Am 22: 515, 1991.
- PAYLEY D, FISCHGRUND J: Open reduction and circular external fixation of intraarticular calcaneal fractures. Clin Orthop 290: 125, 1993.
- MORANDI M: “The Treatment of Complex Plateau Fractures with the Ilizarov External Fixator,” in Current Concepts of Lower Extremity Trauma External Fixation, Nashville, May 10-11, 1996.
- KULDJANOV D: “Reduction Technique of Ilizarov,” in Current Concepts of Lower Extremity Trauma External Fixation, Nashville, May 10-11, 1996.
- WEINER LS, KELLEY M, YANG E, ET AL: The use of combination internal fixation and hybrid external fixation in severe proximal tibia fractures. J Orthop Trauma 9: 244, 1995.
- CRENSHAW AH: “Delayed Union and Nonunion of Fractures,” in Campbell’s Operative Orthopedics, 8th Ed, Vol 2, ed by AH Crenshaw, Mosby-Yearbook, St Louis, 1992.
- GRILL F, FRANKE J: The Ilizarov distractor for the correction of relapsed or neglected clubfoot. J Bone Joint Surg 69B: 593, 1987.
- PATERSON D: Leg-lengthening procedures: a historical review. Clin Orthop 250: 27, 1990.
- WAGNER H: Operative lengthening of the femur. Clin Orthop 136: 125, 1978.
- HOOD RW, RISEBOROUGH EJ: Lengthening of the lower extremity by the Wagner method: a review of the Boston’s Children Hospital experience. J Bone Joint Surg 63A: 1122, 1981.
- PAYLEY D: Current techniques of limb lengthening. J Pediatr Orthop 8: 73, 1988.
- DE BASTIANI G, ALDEGHERI R, RENZI-BRIVIO L, ET AL: Limb lengthening by distraction of the epiphyseal plate. J Bone Joint Surg 68B: 545, 1986.
- DE BASTIANI G, ALDEGHERI R, RENZI-BRIVIO L, ET AL: Chondrodiastasis-controlled symetrical distraction of the epiphyseal plate limb lengthening in children. J Bone Joint Surg 68B: 550, 1986.
- ILIZAROV GA: Clinical application of the tension-stress effect for limb lengthening. Clin Orthop 250: 8, 1990.
- PAYLEY D: Editorial comment. Clin Orthop 250: 2, 1990.
- ROSMAN M, BROWN K: Preoperative Ilizarov frame construction for correction of ankle and foot deformities. J Pediatr Orthop 11: 238, 1991.
- ILIZAROV GA: “Treatment of Disorders of the Foot,” in Transosseous Osteosynthesis, ed GA Ilizarov, Springer-Verlag, Berlin, 1992.
- GRANT AD, ATAR D, LEHMAN WB: Ilizarov technique in correction of foot deformities: a preliminary report. Foot Ankle 11: 1, 1990.
- LAUGHLIN RT, CALHOUN JH: Ring fixators for reconstruction of traumatic disorders of the foot and ankle. Orthop Clin North Am 26: 287, 1995.
- GREEN S: Ilizarov external fixation technical and anatomic considerations. Bull Hosp Jt Dis 48: 28, 1988.
- VELAZQUEZ RJ, BELL DF, ARMSTRONG PF, ET AL: Complications of use of the Ilizarov technique in the correction of limb deformities in children. J Bone Joint Surg 75A: 1148, 1993.
- PAYLEY D: Problems, obstacles and complications of limb lengthening by the Ilizarov technique. Clin Orthop 250: 81, 1990.
- BARRAL JP, GIL DR, VERGARA SS: “Atlas for the Insertion of Transosseous Wires,” in Operative Principles of Ilizarov, ed by AB Maiocchi, J Aronson, Medi Surgical Video, Milan, Italy, 1991.
- CATAGNI MA: “The Anatomic Location for Insertion of Wires and Half-Pins,” in Advances in Ilizarov Apparatus Assembly, ed by AB Maiocchi, Medicalplastic srl, Milan, Italy, 1994.
© 1996 American Podiatric Medical Association