Abstract

Background: Musculoskeletal ultrasound has been demonstrated to be an accurate measurement tool in the diagnostic evaluation of plantar fasciitis. Previous authors have shown a high degree of reliability between different machines and well-trained, experienced operators in evaluating the thickness and echogenicity of the plantar fascia. We aimed to demonstrate through this study that accurate ultrasonic measurements of the plantar fascia can be reliably reproduced by inexperienced operators as well by demonstrating reliability of measurements between medical student and physician operators. Methods: Fifty-one healthy participants volunteered for this study. Each participant had both feet examined. The Lumify musculoskeletal ultrasound unit with a 12- to 4-MHz linear transducer was used for this evaluation. Three independent images of each participant’s foot were taken by each operator and de-identified. For blinding, each de-identified image was then evaluated later by each evaluator using the ImageJ processing program. Plantar fascia thickness was measured at both the insertion on the calcaneus and a point 1 cm distal from the insertion. Each evaluator’s three measurements at each point of the participant’s fascia were then averaged for use in statistical analysis. An intraclass correlation coefficient (ICC) and Bland-Altman plots were used for statistical and visual interpretation of the results. Results: Using the ICC method, the results indicated excellent reliability between operators of different experience levels. The ICC for each rater was 0.967 when evaluating the same image of the plantar fascia at 1 cm from the insertion and 0.804 when evaluating the same fascia using six different images. Conclusions: The excellent reliability indicates the ease of use and practicality of the musculoskeletal ultrasound for clinical practice. For the most reliable results, we recommend that the point of measurement should be 1 cm from the insertion.
Plantar fasciitis is one of the most common causes of heel pain, affecting an estimated 10% of the general population [1] Typical clinical features include a gradual onset of inferior heel pain that is often worse during the patient’s first steps in the morning or after a period of inactivity. Although diagnostic imaging can visualize confirmatory features, diagnosis of this condition remains primarily clinical in most practices.
As a noninvasive diagnostic technique, interest in ultrasound examination of the plantar fascia is growing. Thickening of the plantar fascia corresponds to symptoms in most individuals, who also show changes in the echogenicity of the fascia. Such changes could be linked to histologic studies demonstrating an increase in mucoid deposition and plantar fascia collagen degeneration [2]. Ultrasound assessment of the plantar fascia allows for real-time imaging without the risk of ionizing radiation, time delay, or high cost of other modalities [3,4] Increased thickness and decreased echogenicity are correlated with symptomatic plantar fasciitis, especially when compared bilaterally with an asymptomatic foot [5]. Past studies have shown high reliability of various ultrasound platforms with trained operators in determining structural measurements [6,7]. However, many such studies examined a relatively small data set and did not include differences between medical student and physician operators [4,6].
The purpose of this study was to determine interrater reliability of ultrasound measurements of the plantar fascia using a large data set of 51 students (102 feet) from the Des Moines University College of Podiatric Medicine and Surgery. Measurements were compared between an MS2 student (S.G.) with limited training and an attending DPM (M.J.). In addition, a small side study was conducted simultaneously in which 53 measurements of a participant in the control group were recorded by podiatric medical students to document the accuracy of first-time users after brief training. We plan to show that clinical ultrasound can be used as a reliable tool for obtaining accurate measurements of the plantar fascia even in the hands of relatively untrained operators.

Materials and Methods

Sample 

Fifty-one healthy participants from the College of Podiatric Medicine and Surgery program volunteered for this study. Inclusion criteria were 1) being over the age of 18, 2) having no injuries in the lower limb in the previous 2 years, and 3) lack of heel pain or discomfort at the time of evaluation. The participants read and signed the informed consent form after the data collection procedures were thoroughly explained. Two operators were involved in all of the participants’ imaging and evaluations. Operator 1 was a DPM MS2 student (S.G.) with no previous experience in sonogram analysis of the plantar fascia who was trained specifically for this study. Operator 2 was a board-certified DPM (M.J.) with 7 years of experience in musculoskeletal ultrasound. This study was approved by the institutional review board of Des Moines University.

Procedures 

The Lumify musculoskeletal ultrasound unit with a 12- to 4-MHz linear transducer (Koninklijke Philips Electronics N.V., Amsterdam, Netherlands) was used. Participants were placed in a supine position with their feet hanging off the edge of the exam chair. Imaging of the plantar fascia consisted of three independent images per operator per foot per participant. The images were obtained using a preset musculoskeletal setting for the device; namely, 12-MHz frequency and depth of 3.5 cm, with the focus on the level of the plantar fascia at the insertion on the plantar calcaneus (1 cm), and gain set at 50. These values were not adjusted by any operator during any of the imaging. The images were oriented and taken using the long axis exclusively. This was selected to accommodate the inexperienced users as well as the utility in measuring the thickness of the fascia. The focus of our analysis was evaluation of the thickness of the plantar fascia; as such, the longitudinal view was thought to be sufficient. Although an exact landmark area for visualization was not set for each foot, the medial band was visualized by design and the inexperienced users were trained to visualize the plantar medial heel in the same manner. Each image was then coded, saved, and later evaluated by each blinded operator using the ImageJ processing program (National Institutes of Health, Bethesda, Maryland). The distal (deep) edge of the plantar fascia insertion on the calcaneus was identified. From this level, a 1-cm line was drawn toward the distal insertion (Fig. 1A). The thickness of the plantar fascia was measured at this point (Fig. 1B). The thickness of the plantar fascia was also measured at the proximal edge of this line (deemed the insertion) (Fig. 1C). Thickness was measured as the distance between hyperechoic edge points of a line oriented perpendicularly to the plantar fascia longitudinal axis. Each operator self- and cross-evaluated the images separately and independently.
Figure 1. Plantar fascia thickness measurement in a healthy 25-year-old woman. A 1-cm line was drawn from the point of insertion (A). A line was drawn between the hyperechoic lines of the plantar fascia 1 cm away, indicating the thickness (B). A line was drawn between the hyperechoic lines at the insertion, indicating the thickness (C).
A total of 612 images from 102 feet (51 participants) were obtained. For blinding purposes, each rater viewed the de-identified images stored under a code name on a secure server on a different date than the scans themselves. Each data set was blinded from the other rater until final analysis. The average of three measurements from each foot was used to establish a reference measurement for each rater, which in turn was used for the purpose of statistical analysis.
Regarding the side study, after a brief demonstration, an examination of a control plantar fascia was performed by MS1, MS2, and MS3 students. An image was obtained by the students using the same musculoskeletal ultrasound as the evaluators. Next, a 1-cm line was drawn from the insertion using the Lumify tablet and the thickness was recorded from the hyperechoic to hyperechoic line.

Statistical Analysis 

The analysis of variance between our two raters was determined by the intraclass correlation coefficient (ICC), which has been used in prior studies of similar design [3,4,8]. The ICC values were interpreted as poor when below 0.5, as moderate when between 0.5 and 0.75, as good when between 0.75 and 0.90, and as excellent when above 0.90 [8]. Using MedCalc software (MedCalc, Ostend, Belgium), the data set was analyzed to produce the ICC for two raters measuring the same structure and again for two raters when evaluating the same image. The precision between raters in the visual inspection of the images was also analyzed through the error distribution across the predicted value using the Bland-Altman plot (Figs. 2, 3) [9]. The plantar fascia thickness was determined using a mean of three measurements, for which a higher reliability would be expected compared with a single measurement [4,6,7] The values obtained by the student participants in the side study were compared using the SD from the mean; a simpler method was chosen for this endpoint, as each student participant obtained only one image and measurement.
Figure 2. Bland-Altman plot of the pooled data from the average measurements of the plantar fascia at 1 cm from the insertion by rater 1 and rater 2.
Figure 3. Bland-Altman plot showing the pooled data from the average measurements of the plantar fascia at the insertion on the calcaneus by rater 1 and rater 2.

Results

A total of 51 students from the Des Moines University College of Podiatric Medicine and Surgery participated in the study. There were 26 male participants and 25 female participants, ranging in age from 22 to 39 (mean ± SD 25.9 ± 2.9). The average measurement of the plantar fascia was 0.311 ± 0.02 cm at 1 cm distal to the insertion and 0.387 ± 0.025 cm at the insertion. The average difference between the left and right feet of the same patient when evaluated by the same operator was 0.021 ± 0.019 cm at a point 1 cm distal to the insertion and 0.106 ± 0.120 cm at the insertion.
Each plantar fascia was imaged a total of six times (three per investigator) and the saved images were used to measure the thickness of the plantar fascia at the insertion and again 1 cm distal to the insertion. Reliability of the measurement system was checked using the ICC for each investigator when evaluating the same image. The ICC value for the investigators when viewing the same image of the plantar fascia at 1 cm from the insertion was 0.967 (95% confidence interval [CI], 0.957–0.975) and improved to 0.983 (95% CI, 0.978–0.987) with the average of two measurements. At the insertion of the plantar fascia, the ICC was 0.9425 (95% CI, 0.925–0.956). Table 1 shows the ICC values of the investigators when measuring the same patient.
Table 1. Intraclass Correlation Coefficient and 95% Confidence Interval for Each Rater’s Measured Data Set and Pooled Sets
The ICC between raters when evaluating the same foot (different images) was then calculated for each data set, as each rater recorded their own measurements blinded to the other’s data pool. As the reliability of evaluating the same images was quite high, we deemed it appropriate to include a pooled data set as well.
A Bland-Altman plot of the pooled data from the average measurements of the plantar fascia at 1 cm from the insertion by rater 1 and rater 2 is shown in Figure 2. A second Bland-Altman plot showing the pooled data of the insertion from both raters is shown in Figure 3.
A secondary endpoint for the study was to track the SDs of first-time measurements by participants in the study when evaluating a control foot (right foot of rater 2). The average measurement of rater 2’s plantar fascia when imaged by students who were participants in this study was 0.324 ± 0.034 cm.

Discussion

Regarding the mean measurements of the participants, we believe that our data set represents a consistency of measurements for these two locations with previous authors [6,7,10]. Analysis of the ICC for both raters’ data sets and the pooled set suggests that the measurement taken at 1 cm from the insertion of the plantar fascia is reliable between raters of vastly different experience. Using an average of two measurements would raise the reliability to “good” in all cases (including the insertional measurements, which our data show to be slightly less reliable between raters).
When analyzing the differences between the left and right foot measurements, we can see a minimal (submillimeter) difference between the left and right feet at the reference point 1 cm from the insertion and a slightly higher difference between the left and right measurements of the insertion itself. We believe that this is consistent with our ICC values of the measurement of the fascia itself and indicative of the difficulty in obtaining consistent measurements at the point of insertion.
Additionally, through a simplified analysis of the measurements of the control foot by each study participant, we believe that the low SD of this set can be correlated with the intuitiveness of the musculoskeletal ultrasound in measuring the plantar fascia. Similar measurements were consistently produced by even first-time users with reliability.
Some limitations of our study are the use of only two raters for the ICC component as well as the nonstandardized area of measurement (there was no exact point of reference on the plantar heel—landmark approach—established before the study, although users were trained to use the medial insertion). Regarding the comparison of the left and right feet, our measurements were taken of asymptomatic individuals and were expected to be similar between individuals’ feet. Prior studies have alluded to the use of the ICC as a potential source of issues because of the natural variability of the structures being measured. This potentially underrepresents the true reliability of the rating system or raters [11]. The supine position of measurement, although much more practical in the clinical setting, may also be responsible for more unreliable measurements and difficulty visualizing the structures [10].

Conclusions

We believe that our results have established several important points. First, the interrater reliability for the two blinded raters when evaluating the same images was excellent. We believe this indicates the ease and reliability in identifying the anatomical structure of the plantar fascia on any given ultrasound image. Second, the interrater reliability of measurements of the plantar fascia at 1 cm distal to the insertion was rated “high” via the ICC; somewhat less reliable were the measurements taken at the insertion. We recommend using the point 1 cm from the insertion for clinical utility. Third, the relative inexperience of rater 2 had little to no bearing on the interrater reliability of these measurements, which are consistent with prior studies between two experienced raters [4,6,7]. We hope that this finding is encouraging to current practitioners who may be inexperienced using clinical ultrasound and hesitant to use this modality because of concerns about accuracy of measurements.
Financial Disclosure: None reported.
Conflict of Interest: None reported.

References

  1. Mohseni-Bandpei MA, Nakhaee M, Mousavi ME, et al: Application of ultrasound in the assessment of plantar fascia in patients with plantar fasciitis: a systematic review. Ultrasound Med Biol 40: 1737, 2014; doi: 10.1016/j.ultrasmedbio.2014.03.001.
  2. Lemont H, Ammirati KM, Usen N: Plantar fasciitis: a degenerative process (fasciosis) without inflammation. JAPMA 93: 234, 2003; doi: 10.7547/87507315-93-3-234.
  3. Radwan A, Wyland M, Applequist L, et al: Ultrasonography, an effective tool in diagnosing plantar fasciitis: a systematic review of diagnostic trials. Int J Sports Phys Ther 11: 663, 2016. 
  4. Crofts G, Angin S, Mickle KJ, et al: Reliability of ultrasound for measurement of selected foot structures. Gait Posture 39: 35, 2014; doi: 10.1016/j.gaitpost.2013.05.022.
  5. Boussouar A, Meziane F, Crofts G: Plantar fascia segmentation and thickness estimation in ultrasound images. Comput Med Imaging Graph 56: 60, 2017; doi: 10.1016/j.compmedimag.2017.02.001.
  6. Heinen EH, Lima KMME, Correia R, et al: Reliability in ultrasound measurements of plantar aponeurosis thickness. Foot (Edinb) 46: 101749, 2021; doi: 10.1016/j.foot.2020.101749.
  7. Cheng JW, Tsai WC, Yu TY, et al: Reproducibility of sonographic measurement of thickness and echogenicity of the plantar fascia. J Clin Ultrasound 40: 14, 2012; doi: 10.1002/jcu.20903.
  8. Koo TK, Li MY: A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med 15: 155, 2016; doi: 10.1016/j.jcm.2016.02.012.
  9. Bland JM, Altman D: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1: 307, 1986; doi: 10.1016/S0140-6736(86)90837-8.
  10. Granado MJ, Lohman EB, Daher NS, et al: Effect of gender, toe extension position, and plantar fasciitis on plantar fascia thickness. Foot Ankle Int 40: 439, 2019; doi: 10.1177/1071100718811631.
  11. Rathleff MS, Mølgaard CM, Olesen JL: Intra- and interobserver reliability of quantitative ultrasound measurement of the plantar fascia. J Clin Ultrasound 39: 128, 2011; doi: 10.1002/jcu.20787.