Abstract
Accurate differentiation between malignant and benign pulmonary nodules remains a significant clinical challenge. MicroRNAs (miRNAs) are stably present in serum and exhibit early tumorigenic changes, making them ideal biomarkers for liquid biopsy-based cancer detection. However, as the classic method for miRNA detection, conventional reverse transcription-quantitative PCR (RT-qPCR) is limited by multi-step procedures, enzyme dependence, and reproducibility that can be affected by batch-to-batch enzyme variability. To address these limitations, we developed a platform integrating locked nucleic acid (LNA)-enhanced magnetic bead (MB) capture, on-bead hybridization chain reaction (HCR) amplification, and flow cytometric dual-parameter readout for the detection of serum miR-182, a marker for non-small cell lung cancer (NSCLC). By assembling the LNA-modified hairpin onto MBs, a simplified one-step detection of serum microRNA can be achieved, which triggers HCR amplification upon miR-182 recognition. The optimized assay achieved a detection limit of 0.48 pM with excellent specificity against 9 potentially interfering miRNAs. Compared to RT-qPCR (56 manual operation steps, 500 μL serum, area under the receiver operating characteristic curve (AUC) = 0.7887, 95% confidence interval (CI): 0.6680~0.9094), our method reduces the number of manual operation steps to 19 and requires just 1 μL of serum input, while achieving a favorable diagnostic AUC of 0.8140 in 61 clinical samples (31 lung cancer and 30 non-cancer controls, 95% CI: 0.7097~0.9183); these comparisons refer to manual steps and serum input volume, not to total assay time. This LNA-HCR-flow cytometry platform provides an operationally simplified approach for serum miRNA detection, with intra-day and inter-day CV values below 10% under the tested conditions. Collectively, this platform shows potential as a non-invasive auxiliary diagnostic approach for serum miRNA detection. However, the present study is an exploratory proof-of-concept, and the current cohort cannot yet establish clinical utility for the specific differential diagnosis of malignant versus benign pulmonary nodules.
1. Introduction
Lung cancer remains the leading cause of cancer-related mortality globally, making early and accurate diagnosis essential for enhancing patient outcomes [1,2,3]. Currently, low-dose computed tomography (LDCT) is the primary screening modality [4,5,6]; however, it has limited specificity in distinguishing malignant pulmonary nodules from their benign counterparts, leading to a high rate of false-positive findings and unnecessary invasive procedures [7,8,9]. This limitation leads to a significant number of patients with pulmonary nodules undergoing potentially unnecessary invasive procedures, contributing to follow-up anxiety [7]. Liquid biopsy represents a promising non-invasive molecular diagnostic approach that can aid in distinguishing between malignant and benign pulmonary nodules [10,11,12,13]. Among these, miRNAs have attracted considerable interest as a novel class of liquid biopsy biomarkers. miRNAs are stably present in serum, and their expression profiles undergo characteristic alterations during early tumorigenesis, positioning them as ideal ‘molecular fingerprints’ for cancer [14,15]. Among the various miRNAs, miRNA-182 (miR-182) has been consistently found to be overexpressed in the serum of patients with non-small cell lung cancer (NSCLC) [16,17]. Its upregulation is significantly correlated with tumor invasion, metastasis, and poor prognosis. Notably, one study reported that serum miR-182 levels were significantly higher in patients with malignant pulmonary nodules than in those with benign nodules with an AUC = 0.781, demonstrating promising diagnostic sensitivity and specificity [18]. Consequently, miR-182 presented as a potentially valuable biomarker for differentiating between benign and malignant pulmonary nodules. However, conventional RT-qPCR requires RNA extraction and reverse transcription prior to amplification. This multi-step process is prone to batch-to-batch enzyme variability, which can lead to an elevated and poorly controlled coefficient of variation (CV) [19]. Moreover, the low abundance and high sequence homology among circulating miRNAs pose additional challenges for reliable clinical detection.
To address these limitations, enzyme-free isothermal amplification strategies based on nucleic acid assembly have been developed. These strategies have been increasingly adapted to miRNA detection in complex biological matrices; for example, Nakai et al. recently reported an HCR-based approach for miRNA detection in serum [20]. Synthetic nucleic acid probes provide a basis for enzyme-free amplification, thereby avoiding potential sources of enzyme-associated variability [21,22]. Among these, hybridization chain reaction (HCR) is a classic enzyme-free amplification method that is particularly well-suited for microRNA detection. Upon recognition of a specific target miRNA, HCR selectively triggers the autonomous and sequential assembly of two stable hairpin DNA probes into long DNA nanostructures under isothermal conditions [23]. Conventional HCR-based detection often incorporates a fluorophore–quencher pair within the hairpin structure. In the closed state, fluorescence is quenched via fluorescence resonance energy transfer (FRET); upon target-triggered HCR assembly, the hairpin opens, restoring a strong fluorescent signal. Despite its utility, FRET-based HCR struggles to simultaneously achieve high sensitivity and specificity in complex serum matrices due to non-specific interactions and background interference. To facilitate detection in clinical samples, Pingarrón et al. developed a magnetic bead-based HCR assay [24]. In this design, magnetic beads bind to an initiator strand upon target miRNA recognition, subsequently triggering HCR. By conjugating horseradish peroxidase (HRP) to the hairpin probes, detection was achieved via tetramethylbenzidine (TMB) substrate reaction. However, although this strategy replaces the amplification step with an enzyme-free mechanism, it still relies on enzymatic activity (HRP) for signal readout. Importantly, this method demonstrates that magnetic bead-based nucleic acid probes can be effectively utilized for the isolation and detection of microRNA in complex clinical samples, such as blood. However, immobilizing nucleic acid probes onto magnetic bead surfaces inevitably introduces steric hindrance, which compromises probe-target interactions and reduces detection sensitivity. Therefore, enhancing the hybridization efficiency of surface-immobilized probes with target miRNAs is critical for clinical applications.
Locked nucleic acid (LNA) is an artificially modified nucleotide analog in which a methylene bridge connects the 2′-O and 4′-C positions of the ribose, locking the sugar ring into a rigid “C3′-endo” conformation [25]. This modification confers ultra-high binding affinity, single-base mismatch discrimination, and enhanced resistance to enzymatic degradation in serum [26,27,28]. Recently, Gao et al. reported an LNA-based droplet digital PCR (ddPCR) assay for detecting miR-192 in hepatocellular carcinoma liquid biopsies, demonstrating that LNA probes increased positive droplet counts by 32% [29]. Estrela and colleagues further showed that LNA-modified magnetic beads could efficiently and specifically capture miRNAs from serum, followed by electrochemical detection after elution [30]. This indicates that LNA-functionalized magnetic beads are highly promising for miRNA isolation and enrichment in liquid biopsy applications. Thus, the LNA probe-modified magnetic beads exhibit excellent performance in isolating and enriching microRNA from serum, representing a highly promising platform for liquid biopsy of microRNA.
In this study, we developed an LNA-enhanced magnetic bead platform for the simultaneous isolation, enrichment, and on-bead HCR-based detection of serum miR-182. Specifically, we designed a biotinylated DNA hairpin probe (LH1-biotin) targeting miR-182, with three LNA residues incorporated into the sticky end to enhance target recognition. This probe was immobilized onto streptavidin-coated magnetic beads via biotin-streptavidin interaction, yielding MB-LH1. Upon exposure to miR-182, the LH1 hairpin recognizes and binds the target, leading to hairpin opening. The opened structure then triggers HCR upon transfer to a solution containing two fluorescently labeled hairpin probes, resulting in the assembly of long DNA nanostructures on the magnetic bead surface. After magnetic separation, the signal-bearing beads were redispersed in buffer and analyzed by flow cytometry. We evaluated the sensitivity, specificity, and recovery rate of MB-LH1 using spike-in experiments in serum samples, demonstrating that our LNA-enhanced HCR-flow cytometry platform shows potential for diagnostic accuracy for NSCLC. Notably, unlike conventional RT-qPCR, which typically requires approximately 56 manual operation steps and 500 μL of serum, our MB-LH1 platform reduces the workflow to 19 manual operation steps and consumes just 1 μL of serum input, substantially reducing operational complexity and sample volume requirements; these comparisons refer to manual steps and serum input volume, not to total assay time.
2. Materials and Methods
2.1. Materials
Tris-(hydroxymethyl)-aminomethane (Tris), ethylenediaminetetraacetic acid (EDTA), and magnesium chloride (MgCl2) were procured from Sinopharm Group (Shanghai, China). Streptavidin magnetic beads (MBs) were obtained from Beyotime Biotechnology (Shanghai, China). A 40% acrylamide/Bis solution, DNA marker A (ranging from 25 to 500 bp), 4SGelred (10,000× in water), and all customized nucleic acid sequences were acquired from Sangon Biotech (Shanghai, China). The specific probe sequences, modifications, and their positions are listed in Supplementary Tables S1 and S3. All water (18.2 MΩ·cm at 25 °C) used in the experiments was treated with diethyl pyrocarbonate. The serum/plasma small RNA extraction kit was sourced from Neohalo Biotechnology (Hangzhou, China). The miRNA first strand cDNA synthesis kit (utilizing the stem-loop method) and the miRNA Unimodal SYBR qPCR master mix kit were obtained from Vazyme (Nanjing, China).
2.2. Clinical Study Subjects and Sample Collection
The collection of clinical samples was conducted in accordance with a protocol approved by the Institutional Review Board of Tongji Hospital, which is affiliated with Tongji Medical College of Huazhong University of Science and Technology (Project ethics number: TJ-IRB202402010). The inclusion and exclusion criteria are detailed in the Supplementary Information. Serum samples were collected and subjected to centrifugation at 4 °C for 10 min at 3000 rpm. The supernatant was then frozen and stored at −80 °C.
2.3. Flow Cytometer Model and Parameters
Flow cytometric analysis was performed on a CytoFLEX 3 flow cytometer (Beckman Coulter, Brea, CA, USA) equipped with a 488 nm laser. The detector gains were set as follows: forward scatter (FSC) = 150, side scatter (SSC) = 80, and FITC = 80. FSC and SSC were used to identify the magnetic bead population and to exclude aggregated beads and debris, respectively. The FITC channel (fluorescein isothiocyanate-area, FITC-A) was used to quantify the fluorescence signal on individual beads. For each sample, a total of 10,000 events were acquired. The bead population was first gated on the FSC vs. SSC plot, and single beads were further selected to exclude aggregates. The fluorescence intensity of individual beads was then recorded in the FITC-A channel. Positive beads were defined as those with FITC-A intensity exceeding the 95th percentile of the blank control, which corresponds to a background (false-positive) rate of ≤5%. The percentage of positive beads (positive%) and the mean FITC-A intensity (MFI) were used for quantitative analysis.
2.4. Preparation of Buffers
50 mM Tris buffer: Tris base (6.05 g) was dissolved in water (800 mL), and the pH was adjusted to 7.5 with HCl (5 M). Then the solution was filled up to 1 L with water.
Suspension buffer: 20 mM Tris buffer with 3 mM MgCl2.
Hybridization reaction buffer: 1× phosphate-buffered saline (PBS) with 5 mM MgCl2.
2.5. Preparation of MB-LH1
LH1-biotin was initially annealed in a suspension buffer. The magnetic beads (MBs) were washed three times using Tris buffer. Subsequently, the MBs precipitate was mixed with the previously mentioned LH1-biotin (Ratio: 0.5 nmol hairpin per 1 mg MBs). After 2 h, the MBs underwent magnetic separation and were gently washed three times with PBS. The resulting MB-LH1 was then redispersed in 50 μL of suspension buffer.
2.6. HCR Testing of miR-182 Solution or Serum Sample
A total of 20 μL of MB-LH1 was combined with either 1 μL of miR-182 solution or 1 μL of serum sample. The mixture was then supplemented with hybridization reaction buffer to achieve a final volume of 100 μL and incubated with vortex shaking for 30 min. Following this, magnetic attraction was applied for 1 min to separate the components, and the supernatant was completely discarded. Subsequently, 2 μL HCR solution (5 μmol L−1 H1 and 5 μmol L−1 H2-FAM) was added, and the hybridization reaction buffer was adjusted to 100 μL before incubating with vortex shaking for an additional 4 h, as previously described [23]. After this incubation, magnetic attraction was again applied for 1 min, and the supernatant was discarded. The resulting mixture was washed twice with 100 μL of buffer, subjected to magnetic attraction for 1 min, and the supernatant was discarded once more. Finally, the pellet was resuspended in 300 μL of Tris buffer and analyzed using a flow cytometer.
2.7. Nucleic Acid Electrophoresis
Nucleic acid samples were separated by electrophoresis on a 12% native polyacrylamide gel prepared in 1× Tris-borate-EDTA (TBE) buffer (89 mM Tris-borate, 2 mM EDTA, pH 8.3). After polymerization at room temperature, the comb was removed, and the wells were rinsed with 1× TBE buffer. Each nucleic acid sample (50–200 ng) was mixed with 6× DNA loading buffer (5:1 ratio) and loaded onto the gel. Electrophoresis was conducted in 1× TBE running buffer at a constant voltage of 120 V (12 V/cm) for 45–50 min in an ice-water bath, with the current remaining above 20 mA for most of the run. The ice-water bath was used to dissipate Joule heating during electrophoresis, thereby preventing thermal accumulation that could lead to band smearing and uneven migration. Following electrophoresis, the gel was stained with 4SGelred and visualized under UV transillumination. The melting temperatures (Tm) of the intermolecular miR-182/H1 and H1/H2 duplexes were estimated under the electrophoresis and hybridization reaction conditions (Table S2). The predicted Tm values exceeded the temperatures used during electrophoresis and HCR, suggesting that the intermolecular duplexes were thermodynamically stable under the corresponding experimental conditions.
2.8. RT-qPCR
Serum small RNA was extracted according to the procedures outlined in the instruction manual for Serum/Plasma small RNA extraction kits. The extracted small RNA samples were immediately subjected to the subsequent reverse transcription step, which was performed using the miRNA 1st Strand cDNA Synthesis Kit via the stem-loop method. The relative levels of miRNA were quantified using the miRNA Unimodal SYBR qPCR Master Mix.
2.9. Statistical Analysis
Statistical analyses were performed using GraphPad Prism (Version 10.4.1). Quantitative data from replicate analytical measurements are presented as mean ± standard deviation (SD), unless otherwise indicated. The number and type of replicates are specified in the corresponding figure legends and table notes. Independent experiments refer to separately performed assay runs, whereas technical replicates refer to repeated measurements of the same sample under identical experimental conditions. For clinical analyses, n represents the number of individual participants. Statistical comparisons were performed using an unpaired two-tailed t-test or one-way analysis of variance (ANOVA), as appropriate. Differences were considered statistically significant at p < 0.05. Receiver operating characteristic (ROC) curves were compared using DeLong’s test. The limit of detection (LOD) was determined using the blank signal threshold calculated as the mean plus three standard deviations (mean + 3 SD).
3. Results
3.1. HCR-Based Detection Principle
To achieve rapid separation, signal amplification, and accurate detection of lung cancer-associated miRNAs in serum, a DNA-based sensor composed of HCR components that mediate fluorescence signal amplification detected by flow cytometry was developed. As previously described [23], two hairpin strands (H1 and H2) for HCR amplification specifically targeting miR-182 were designed, with H1 modified with a 3′-terminal Biotin-TEG (Tables S1 and S2). To enhance capture efficiency while maintaining specificity, three deoxyribonucleotides in the sticky end of H1-biotin—which was designed to be complementary to miR-182—were replaced with locked nucleic acids (LNAs), generating LH1-biotin. LH1-biotin was then immobilized onto streptavidin-coated magnetic beads via biotin-streptavidin interaction, yielding MB-LH1 (Figure 1A). Upon miR-182 capture, the hairpin structure of MB-LH1 opens, triggering HCR with H1 and H2-FAM on the bead surface (Figure 1B). This process generates fluorescent double-stranded DNA nanostructures, enabling signal amplification and fluorescence readout. In practical sample detection, the target microRNA can be captured by adding serum to the MB-LH1 buffer system, followed by magnetic separation to obtain beads capable of initiating HCR. The magnetically separated beads are then redispersed in a reaction solution containing H1 and H2-FAM to initiate HCR amplification. After a final magnetic separation, the fluorescent signal on the beads was analyzed by flow cytometry (Figure 1C).
Figure 1.
(A) Assembly route of LNA-modified magnetic beads. (B) The capture and HCR amplification of microRNA using MB-LH1. (C) Clinical sample testing process.
Nucleic acid binding and HCR assembly were further validated by gel electrophoresis (Figure 2). No HCR amplification was triggered by the annealed hairpins (H1, H2, and LH1, Lane 5) alone. Faint bands (Lane 2, Lane 3, and Lane 4) observed between 50~100 bp likely corresponded to residual hairpin dimers, which did not interfere with target capture or subsequent HCR. Weak binding to miR-182 was exhibited by H1 alone (Lane 6), while no binding was observed for H2 (Lane 7). In contrast, enhanced binding to miR-182 was observed for LH1 (Lane 8) in the gel assay. Efficient triggering of HCR by miR-182 was confirmed by the high-molecular-weight assembly products observed in Lane 9.
Figure 2.
Electrophoresis image of HCR nucleic acids.
3.2. Optimization of the Detection Conditions
LH1-biotin and streptavidin-coated magnetic beads (MB-SA) were assembled in PBS at a ratio of 0.5 nmol LH1 per 1 mg MB-SA. An assembly efficiency of 73% was determined, based on quantification of residual LH1 in the supernatant using a UV spectrophotometer. The resulting MB-LH1 was stored at 4 °C at a concentration of 50 mg/mL.
Optimization was first performed for the target capture and HCR amplification steps. To evaluate the capture capacity of MB-LH1, varying volumes of MB-LH1 were incubated with 1 μM miR-182 for 1 h. After magnetic separation, the captured beads were incubated with 1 μM H2-FAM for 4 h. The residual H2-FAM fluorescence in the supernatant was then measured, which inversely correlates with capture efficiency. As shown in Figure 3A,B, the fluorescence reduction plateaued when the MB-LH1 volume reached 0.8~1.2 μL, indicating that 1 μL of MB-LH1 was sufficient for capturing 1 μM miR-182.
Figure 3.
Optimization of incubation and amplification conditions. (A,B) Volume of MB-LH1 used for miR-182 capture. (C,D) Capture time of miR-182 with MB-LH1. (E,F) Volume of the HCR amplification solution. (G,H) HCR amplification time. Data in panels (B,D,F,H) are presented as mean ± SD from three independent experiments (n = 3). Representative fluorescence spectra are shown in panels (A,C,E,G).
Capture time was next optimized (Figure 3C,D). A marked increase in capture efficiency was observed from 10 to 30 min, after which it plateaued between 1 and 4 h. Only marginal improvement was yielded by extending capture time to 20 h. Therefore, 30 min was selected as the optimal capture time. The effect of HCR incubation temperature was also evaluated. As shown in Figure S1 (Supplementary Materials), incubation at 4 °C, room temperature, and 37 °C yielded comparable fluorescence signals, with no significant differences observed. Room temperature was therefore selected to simplify the operational workflow and avoid the need for strict temperature control.
The effect of the HCR volume (buffer containing 5 μM H1 and 5 μM H2-FAM) on signal amplification was then examined (Figure 3E,F). The fluorescence signal plateaued when the reaction volume reached 2 μL. Finally, the HCR amplification time was optimized. The fluorescence signal peaked at 4 h and decreased slightly after 20 h (Figure 3G,H). Based on these results, an HCR volume of 2 μL and an amplification time of 4 h were selected for subsequent experiments.
3.3. Analytical Performance of the MB-LH1 Sensor
Under the optimized conditions, the analytical performance of MB-LH1 was evaluated using flow cytometry. Serial dilutions of miR-182 were mixed with MB-LH1, processed according to the established protocol, and analyzed by flow cytometry. Representative results from three independent experiments are shown in Figure 4A. As the miR-182 concentration increased, the FITC-A (fluorescein isothiocyanate-area) peak was progressively shifted to the right.
Figure 4.
(A) Flow cytometry results of different concentrations of miR-182 using MB-LH1-HCR from three independent experiments. (B) Curve fitted to the mean fluorescence intensity of magnetic beads using a one-site binding model (inset: linearized fitting curve). (C) Curve fitted to the percentage of positive magnetic beads using a logarithmic model based on the Poisson-like binding behavior at low concentrations (inset: linearized fitting curve). Positive% was calculated using the 95th percentile of the blank control as the positive-bead threshold. LOD values were calculated as mean blank signal + 3 SD. Each data point in panels (B,C) represents the mean ± SD of three independent experiments (n = 3).
The relationship between mean fluorescence intensity (MFI) and miRNA concentration was fitted to a one-site binding model:
where Y is the MFI, X is the target concentration, Bmax is the maximum specific binding signal, Kd is the dissociation constant, and NS is the slope of non-specific binding. As shown in Figure 4B, fitting the data (1.56 pM to 200 pM) to this model yielded Bmax = 8448, Kd = 5.674, NS = 16.31, Background = 13,053, with R2 = 0.989. The limit of detection (LOD), calculated as the mean blank signal plus three times the standard deviation (mean + 3 SD), was determined to be 0.82 pM.
The data were also analyzed using the percentage of positive beads (positive%), a common metric in flow cytometry. The magnetic bead population was first gated on an FSC vs. SSC plot based on bead size and granularity. Single beads were then selected by gating on FSC-A vs. FSC-H to exclude doublets and aggregates. The FITC-A fluorescence of single beads was recorded. Positive beads were defined as those with FITC-A intensity exceeding the 95th percentile of the blank control, which corresponds to a background (false-positive) rate of ≤5% and a blank positive% of approximately 5%. The percentage of positive beads (positive%) was used as the quantitative readout. A logarithmic relationship between positive% and target concentration was observed:
Fitting the data (1.56 pM to 200 pM) to this model yielded k = 2.726, b = 0.95, c = 5.167, with R2 = 0.992 (Figure 4C). The corresponding LOD, calculated using the same mean + 3 SD criterion, was determined to be 0.48 pM.
The specificity of MB-LH1 was next evaluated. Given the high sequence homology among circulating miRNAs, 9 potentially interfering miRNA biomarkers (including lung cancer-associated miRNAs, Figure 5A,B, Table S3) were tested. At a uniform concentration of 50 pM, both MFI and the positive% confirmed that MB-LH1 exhibited excellent specificity for miR-182, with negligible cross-reactivity to the other miRNAs tested.
Figure 5.
Specificity of MB-LH1-HCR for miR-182 against potentially interfering miRNAs. (A) Mean fluorescence intensity (MFI). (B) Percentage of positive magnetic beads. Each dot represents one independent experiment (n = 5 per condition), and the horizontal lines indicate the mean values.
Having validated the assay in buffer, its performance in a more complex matrix was assessed through spike-in experiments using healthy human serum samples that tested negative for miR-182 (Table 1). Each of the three serum samples was spiked with miR-182 at 10, 40, and 160 pM. After MB-LH1 capture and flow cytometric analysis, recoveries ranged from 96.7% to 112.3%, with relative standard deviations (RSDs) ranging from 0.45% to 4.51%.
Table 1.
Recovery and precision of miR-182 measurements in spiked serum samples.
In summary, the analytical characteristics of the MB-LH1-HCR-flow cytometry platform were as follows: an LOD of 0.82 pM (MFI) and 0.48 pM (positive%), a tested dynamic range of 1.56–200 pM (R2 = 0.989 for MFI and 0.992 for positive%), spike-in recovery of 96.7–112.3% with RSD of 0.45–4.51% in serum, intra-day CV of 5.26–8.38% and inter-day CV of 6.57–9.88%, and negligible cross-reactivity against nine interfering miRNAs at 50 pM. These values are summarized in the Results text and in Table 1, Tables S8 and S9.
3.4. Clinical Sample Diagnostic Performance
The clinical diagnostic performance of our method was validated using 61 samples from a recently collected case–control cohort. As a comparator, miR-182 levels were simultaneously measured by RT-qPCR. Baseline characteristics are summarized in Table S4. The cohort included 31 lung cancer (LC) samples (27 early-stage, 4 advanced-stage) and 30 non-cancer (NC) samples (10 benign lung disease, BLD and 20 healthy controls). No significant differences in age, BMI, smoking history, or alcohol consumption were observed between the two groups.
As shown in Figure 6A, RT-qPCR revealed significantly higher miR-182 levels in the LC group than in the NC group (p < 0.0001). However, the lowest miR-182 level in the LC group was lower than the lowest level in the NC group, indicating substantial overlap between the two groups at the individual sample level. ROC analysis yielded an AUC of 0.7887 (95% CI: 0.6680–0.9094, Figure 6B). At the optimal cutoff of 1.724 determined by the Youden index, RT-qPCR achieved a sensitivity of 64.52% (20/31), a specificity of 90.00% (27/30), an accuracy of 77.05% (47/61), a positive predictive value (PPV) of 86.96% (20/23), a negative predictive value (NPV) of 71.05% (27/38), and a Youden index of 0.5452, with a confusion matrix of true positive (TP) = 20, false negative (FN) = 11, true negative (TN) = 27, and false positive (FP) = 3 (Table S6).
Figure 6.
Performance study of miR-182 in lung cancer diagnosis. (A) Relative microRNA levels in the LC group and NC group measured by RT-qPCR. (B) ROC analysis of the RT-qPCR results. (C) miRNA concentrations derived from the MFI in the LC group and NC group measured by MB-LH1-HCR. (D) ROC analysis of miRNA concentrations derived from the MFI measured by MB-LH1-HCR. (E) miRNA concentrations derived from the positive% in the LC group and NC group measured by MB-LH1-HCR. (F) ROC analysis of miRNA concentrations derived from the positive% measured by MB-LH1-HCR. Positive% values were calculated using the 95th percentile of the blank control as the positive-bead threshold. Each dot in panels (A,C,E) represents an individual participant (LC, n = 31; NC, n = 30). Horizontal lines indicate the median. ****: p < 0.0001.
Using our MB-LH1-HCR-flow cytometry method, statistically significant differences between the LC and NC groups were revealed by both the MFI (Figure 6C) and the positive% (Figure 6E). ROC analysis was performed by converting the corresponding MFI or positive% into miRNA concentrations (Figure 6D,F, Table S5). An AUC of 0.7995 (95% CI: 0.6917–0.9072) was yielded by ROC analysis of MFI, while an AUC of 0.8140 (95% CI: 0.7097–0.9183) was obtained from ROC analysis of positive%. At the optimal cutoff of 3.073 for MFI, the assay achieved a sensitivity of 48.39% (15/31), a specificity of 96.67% (29/30), an accuracy of 72.13% (44/61), a PPV of 93.75% (15/16), an NPV of 64.44% (29/45), and a Youden index of 0.4505, with a confusion matrix of TP = 15, FN = 16, TN = 29, and FP = 1. At the optimal cutoff of 3.154 for positive%, the assay achieved a sensitivity of 54.84% (17/31), a specificity of 96.67% (29/30), an accuracy of 75.41% (46/61), a PPV of 94.44% (17/18), an NPV of 67.44% (29/43), and a Youden index of 0.5151, with a confusion matrix of TP = 17, FN = 14, TN = 29, and FP = 1 (Table S6).
No statistically significant differences were shown by pairwise comparisons using DeLong’s test among the three ROC analyses (p > 0.05). Taken together, these results indicate that all three readouts show a statistically significant group-level difference between LC and NC, but that substantial individual-level overlap remains, which limits the diagnostic performance of the assay in its current form.
Only 1 μL of serum input per test was required by our method, in contrast to the 500 μL required for RT-qPCR (Table S7); this comparison refers to serum input volume, not to total assay time. The entire procedure required only 19 manual operation steps compared to approximately 56 steps for RT-qPCR (Table S7); this comparison refers to manual operation steps, not to total assay time. The analytical precision of our method was further evaluated by assessing intra-day and inter-day variability (Tables S8 and S9). For MFI-based detection, the intra-day CV ranged from 5.35% to 8.38% (mean: 6.72%), and the inter-day CV ranged from 7.65% to 9.44% (mean: 8.83%). For positive%-based detection, the intra-day CV ranged from 5.26% to 7.68% (mean: 6.22%), and the inter-day CV ranged from 6.57% to 9.88% (mean: 8.75%). In a previous study, RT-qPCR was reported to exhibit CV values of 39% and 50% for whole-blood and PBMC samples, respectively [31]; these values were not measured in the present study.
4. Discussion
In this study, we developed a streamlined LNA-enhanced magnetic bead platform (MB-LH1) that integrates target isolation, enrichment, and on-bead HCR amplification for the flow cytometric detection of serum miR-182. The design rationale centered on three key elements: LNA-modified capture probes for enhanced target recognition in complex matrices, on-bead HCR for enzyme-free signal amplification, and flow cytometric readout for quantitative analysis.
The incorporation of three LNAs into the sticky end of the capture probe proved critical for achieving specific capture of miR-182. In the gel electrophoresis assay, the LNA-modified probe (LH1) showed enhanced binding to miR-182 compared with its unmodified DNA counterpart (H1). This gel-based evidence is semi-quantitative. Although the Tm values of the unmodified intermolecular duplexes were estimated computationally (Table S2), the experimental Tm and dissociation constant (Kd) of the LNA-modified probe–target duplex were not determined in this study. The observed enhancement is consistent with the known thermodynamic properties of LNA-DNA/RNA duplexes: the pre-organized C3′-endo conformation of LNA reduces the entropic penalty of duplex formation, resulting in significantly higher melting temperatures and binding affinities [25,27]. This enhanced affinity is particularly advantageous for direct capture from crude serum, where competing nucleic acids and proteins are abundant and target miRNA concentrations are extremely low. The ability to bypass RNA extraction and reverse transcription not only simplified the workflow but also preserved the integrity of the target miRNA and minimized sample loss [30].
Consistent with this enhanced affinity, the capture and amplification conditions were further optimized to clarify why a short capture time and a small reaction volume were sufficient. Capture occurs at the magnetic bead interface, where the LNA-modified LH1 probes are locally concentrated and the diffusion distance for miR-182 is short. As the MB-LH1 volume increased, the residual H2-FAM fluorescence in the supernatant plateaued at 0.8–1.2 μL, indicating that 1 μL of MB-LH1 already provided sufficient capture sites for 1 μM miR-182. The capture time-course showed a marked increase from 10 to 30 min, followed by a plateau between 1 and 4 h and only marginal improvement up to 20 h, consistent with rapid surface hybridization of the short 18-bp complementary region under adequate mixing. For HCR, the signal plateaued at a 2 μL reaction volume and peaked at 4 h, reflecting the balance between local hairpin availability, on-bead HCR propagation, and steric crowding on the bead surface. These results indicate that the selected 30 min capture and 4 h HCR conditions are kinetically sufficient under the present assay format, rather than arbitrary choices.
The enzyme-free nature of HCR amplification eliminates the requirement for amplification enzymes and thereby avoids potential sources of enzyme-associated variability [21,22]. Furthermore, performing HCR on the magnetic bead surface enabled extensive washing steps to remove non-specifically bound components, while the magnetic separation streamlined the entire process without requiring centrifugation or column purification. This combination of features likely contributed to the low background signal and high specificity observed in our assay.
For quantitative readout, flow cytometry enabled dual-parameter analysis. Although MFI is a conventional metric for quantifying bead-bound fluorescence, the positive% method yielded a lower LOD (0.48 pM vs. 0.82 pM) and a better correlation (R2 = 0.992 vs. 0.989). This advantage can be explained by the statistical nature of target binding at low concentrations. When the target concentration is much lower than the available binding sites, the number of bound targets per bead follows a Poisson distribution. By applying a threshold (95th percentile of blank control) to define positive beads, the positive% method effectively excluded weakly fluorescent background events that would otherwise contribute to the MFI; the LOD values, in contrast, were calculated using the mean + 3 SD of the blank control. This single-bead level analysis provided a more discrete and statistically robust metric for quantifying rare binding events, which is particularly relevant for low-abundance miRNA detection in clinical specimens [15].
The AUC of our method (0.8140) was numerically similar to that of RT-qPCR (0.7887). However, because the two workflows differ fundamentally in serum input volume, sample processing, and readout format, this comparison is exploratory and should not be interpreted as evidence of superiority. The difference did not reach statistical significance (DeLong’s test, p > 0.05), and substantial individual-level overlap remained between the LC and NC groups for both methods. The observed AUC is consistent with previous reports for single miRNAs as lung cancer biomarkers (typically 0.7–0.9) [32,33] and is comparable to the performance of serum miR-182 in earlier studies (AUC = 0.781) [18]. This consistency validated the analytical reliability of our assay and confirmed that miR-182 remains a relevant diagnostic marker for NSCLC.
The analytical characteristics of the present platform are within the ranges commonly reported for magnetic bead-based miRNA sensors. The LOD of sub-picomolar level and the dynamic range of 1.56–200 pM are consistent with the performance of bead-based miRNA assays reported in the literature, and the spike-in recovery (96.7–112.3%) and CV values (<10%) are within the ranges generally regarded as acceptable for serum-based assays. Because these comparisons involve different targets, detection formats, and cohort designs, they are intended as a qualitative context rather than as a quantitative benchmark, and no claim of superiority over specific literature values is made.
The reduction in manual operation steps (19 vs. 56) and serum input volume (1 μL vs. 500 μL) represented a practical advantage for clinical translation; these comparisons refer to manual steps and serum input volume, respectively, and do not imply a shorter total assay time. The elimination of RNA extraction and reverse transcription—major sources of systematic error and variability in RT-qPCR workflows [19]—may have contributed to the observed group-level difference. Fewer handling steps reduced the cumulative risk of pipetting errors, sample contamination, and reagent variability, enhancing the overall robustness of the assay. These features are particularly valuable in clinical settings where sample availability is limited, such as pediatric patients or serial monitoring applications.
The intra-day and inter-day precision of our platform was demonstrated by CV values below 10% for both MFI- and positive%-based detection. Positive%-based detection showed slightly lower mean intra-day and inter-day CVs (6.22% and 8.75%, respectively) than MFI-based detection (6.72% and 8.83%, respectively). Previous studies have reported substantial variability in RT-qPCR-based miRNA detection [31]. However, differences in sample matrices, analytical procedures, and experimental designs preclude a direct comparison with the CV values obtained in the present study. Although the elimination of reverse transcription and enzymatic amplification, together with the reduction in manual handling steps, may reduce potential sources of analytical variability, the relative precision of the two methods remains to be established through a direct comparative study under matched experimental conditions.
In this exploratory cohort, DeLong’s test showed no statistically significant difference among the three ROC curves. Because the two workflows differ fundamentally in serum input volume (500 μL vs. 1 μL), sample processing, and readout format, this comparison is not head-to-head and should not be interpreted as evidence of superiority or formal non-inferiority. Nevertheless, the observation suggests that our simplified platform has the potential to provide diagnostic performance comparable to RT-qPCR while offering clear operational advantages.
Several limitations should be acknowledged. First, the sample size (n = 61) was relatively modest, and the cohort was derived from a single institution. Second, the current assay targeted only miR-182; while single-miRNA biomarkers have demonstrated clinical utility, multiplexed panels may offer enhanced diagnostic accuracy [17]. Third, the case–control design provided proof-of-concept validation but did not fully reflect the real-world diagnostic challenge of distinguishing malignant from benign nodules in an unselected screening population. Prospective validation in larger, independent cohorts will be essential to confirm clinical utility and assess the assay’s performance in the intended use setting. Additionally, the multi-channel detection capability of flow cytometry renders our platform readily adaptable for multiplexed detection, which could further enhance sensitivity and specificity by incorporating additional lung cancer-associated miRNAs.
5. Conclusions
In this study, we developed a streamlined LNA-enhanced magnetic bead platform (MB-LH1) that integrates target isolation, enrichment, and on-bead hybridization chain reaction (HCR) amplification for the flow cytometric detection of serum miR-182. The incorporation of three LNAs into the sticky end of the hairpin capture probe enabled high-affinity, sequence-specific capture directly from crude serum, eliminating the need for RNA extraction and reverse transcription. The on-bead HCR amplification provided a robust enzyme-free alternative to traditional amplification methods, achieving a detection limit of 0.48 pM with excellent specificity against 9 potentially interfering miRNAs. Flow cytometric readout enabled dual-parameter quantification, with the positive% method demonstrating a slightly higher R2 (0.992) and lower mean CVs (intra-day CV: 6.22%; inter-day CV: 8.75%) compared to MFI-based analysis. Clinical validation using 61 case–control samples showed that our method distinguished lung cancer patients from non-cancer controls with an AUC of 0.8140 (95% CI: 0.7097–0.9183), comparable to RT-qPCR (AUC = 0.7887, 95% CI: 0.6680–0.9094), while substantially reducing operational complexity—requiring only 19 manual operation steps and 1 μL of serum input per test versus 56 steps and 500 μL for RT-qPCR; these comparisons refer to manual steps and serum input volume, not to total assay time. The platform demonstrated good intra-day and inter-day precision, with CV values below 10% under the tested conditions. Its relative precision compared with RT-qPCR remains to be established in future studies. The RT-free and enzyme-free design may help minimize systematic variability. The proposed LNA-HCR-flow cytometry platform offers an operationally simplified, highly specific approach with good analytical precision for serum miRNA detection. It shows potential as a non-invasive auxiliary diagnostic tool for differentiating malignant from benign pulmonary nodules; however, the present study is an exploratory proof-of-concept, and the current cohort cannot yet establish clinical utility for this specific differential diagnosis. The platform also has potential for further multiplexing to enhance diagnostic accuracy in lung cancer and other diseases. Future work will focus on validating the method in larger prospective cohorts and expanding the panel of target miRNAs to improve sensitivity and specificity in clinical practice.
Supplementary Materials
The following supporting information can be downloaded at https://www.mdpi.com/article/10.3390/s26237412/s1, Figure S1: Effect of HCR incubation temperature on the fluorescence signal. HCR was performed at room temperature (RT), 4 °C, and 37 °C. No significant differences were observed among the three conditions; Table S1: Nucleic acid sequences of HCR and primer sequences used in RT-PCR; Table S2: Thermodynamic parameters and Tm values of the intermolecular duplexes; Table S3: The microRNA sequences used for specificity testing; Table S4: Basic information of study subjects; Table S5: Concentrations of miR-182 derived from the MFI or the positive%; Table S6: Optimal cutoff, sensitivity, specificity and confusion matrices of three methods for microRNA 182 in lung cancer diagnosis; Table S7: Comparison table of operation steps between MB-LH1-HCR and RT-PCR; Table S8: Intra-day and inter-day precision of the MB-LH1-HCR-flow cytometry assay for serum miR-182 detection based on mean fluorescence intensity (MFI); Table S9: Intra-day and inter-day precision of the MB-LH1-HCR-flow cytometry assay for serum miR-182 detection based on the percentage of positive beads.
Author Contributions
Conceptualization, W.Z. and L.C.; methodology, W.Z. and X.X.; validation, X.X. and H.R.; formal analysis, T.Y. and Q.B.; investigation, Y.W. (Yi Wang), S.X. and Y.F.; resources, C.H., Y.W. (Yuhan Wang) and F.C.; data curation, H.H.; writing—original draft preparation, W.Z. and X.X.; writing—review and editing, W.Z. and L.C.; visualization, W.Z. and X.X.; supervision, J.L. and Q.Y.; project administration, W.Z.; funding acquisition, W.Z. and L.C. All authors have read and agreed to the published version of the manuscript.
Funding
This work was supported by the National Natural Science Foundation of China (Grant No. 82272418, 22404062).
Institutional Review Board Statement
The collection of clinical samples was conducted in accordance with a protocol approved by the Institutional Review Board of Tongji Hospital, which is affiliated with Tongji Medical College of Huazhong University of Science and Technology (Ethics number: TJ-IRB202402010, approved on 2 February 2024).
Informed Consent Statement
Informed consent was obtained from all subjects involved in the study.
Data Availability Statement
The raw and processed data generated in this study are available from the corresponding author upon reasonable request. Restrictions apply due to the protection of patient privacy.
Conflicts of Interest
The authors declare no conflicts of interest.
Abbreviations
The following abbreviations are used in this manuscript:
| AUC | area under the receiver operating characteristic curve |
| BLD | benign lung disease |
| CI | confidence interval |
| CV | coefficient of variation |
| ddPCR | droplet digital PCR |
| EDTA | ethylenediaminetetraacetic acid |
| FITC | fluorescein isothiocyanate |
| FITC-A | fluorescein isothiocyanate-area |
| FN | false negative |
| FP | false positive |
| FRET | fluorescence resonance energy transfer |
| FSC | forward scatter |
| HCR | hybridization chain reaction |
| HRP | horseradish peroxidase |
| LC | lung cancer |
| LDCT | low-dose computed tomography |
| LNA | locked nucleic acid |
| LOD | limit of detection |
| MB | magnetic bead |
| MB-LH1 | LNA-modified hairpin probe immobilized on magnetic beads |
| MB-SA | streptavidin-coated magnetic beads |
| MFI | mean fluorescence intensity |
| miRNA | microRNA |
| NC | non-cance |
| NPV | negative predictive value |
| NSCLC | non-small cell lung cancer |
| PBS | phosphate-buffered saline |
| PPV | positive predictive value |
| ROC | receiver operating characteristic |
| RSD | relative standard deviation |
| RT-qPCR | reverse transcription-quantitative PCR |
| SSC | side scatter |
| TBE | Tris-borate-EDTA |
| TMB | tetramethylbenzidine |
| TN | true negative |
| TP | true positive |
| Tris | Tris-(hydroxymethyl)-aminomethane |
References
- Henschke, C.I.; Yip, R.; Shaham, D.; Markowitz, S.; Deval, J.C.; Zulueta, J.J.; Seijo, L.M.; Aylesworth, C.; Klingler, K.; Andaz, S.; et al. A 20-year follow-up of the International Early Lung Cancer Action Program (I-ELCAP). Radiology 2023, 309, e231988. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Leiter, A.; Veluswamy, R.R.; Wisnivesky, J.P. The global burden of lung cancer: Current status and future trends. Nat. Rev. Clin. Oncol. 2023, 20, 624–639. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Bray, F.; Laversanne, M.; Sung, H.; Ferlay, J.; Siegel, R.L.; Soerjomataram, I.; Jemal, A. Global cancer statistics 2022: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA Cancer J. Clin. 2024, 74, 229–263. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Humphrey, L.L.; Deffebach, M.; Pappas, M.; Baumann, C.; Artis, K.; Mitchell, J.P.; Zakher, B.; Fu, R.; Slatore, C.G. Screening for lung cancer with low-dose computed tomography: A systematic review to update the US Preventive Services Task Force recommendation. Ann. Intern. Med. 2013, 159, 411–420, Correction in Ann. Intern. Med. 2013, 159. https://doi.org/10.7326/0003-4819-159-9-201311050-00019. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Tammemagi, M.C.; Lam, S. Screening for lung cancer using low dose computed tomography. BMJ 2014, 348, g2253. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Jonas, D.E.; Reuland, D.S.; Reddy, S.M.; Nagle, M.; Clark, S.D.; Weber, R.P.; Enyioha, C.; Malo, T.L.; Brenner, A.T.; Armstrong, C.; et al. Screening for lung cancer with low-dose computed tomography: Updated evidence report and systematic review for the US Preventive Services Task Force. JAMA 2021, 325, 971–987. [Google Scholar] [PubMed]
- Ost, D.E.; Gould, M.K. Decision making in patients with pulmonary nodules. Am. J. Respir. Crit. Care Med. 2012, 185, 363–372. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- MacMahon, H.; Naidich, D.P.; Goo, J.M.; Lee, K.S.; Leung, A.N.; Mayo, J.R.; Mehta, A.C.; Ohno, Y.; Powell, C.A.; Prokop, M.; et al. Guidelines for management of incidental pulmonary nodules detected on CT images: From the Fleischner Society 2017. Radiology 2017, 284, 228–243. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Bartlett, E.C.; Silva, M.; Callister, M.E.; Devaraj, A. False-negative results in lung cancer screening—Evidence and controversies. J. Thorac. Oncol. 2021, 16, 912–921. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Pantel, K.; Alix-Panabières, C. Circulating tumour cells in cancer patients: Challenges and perspectives. Trends Mol. Med. 2010, 16, 398–406. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Diaz LAJr Bardelli, A. Liquid biopsies: Genotyping circulating tumor DNA. J. Clin. Oncol. 2014, 32, 579–586. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Li, W.; Liu, J.B.; Hou, L.K.; Yu, F.; Zhang, J.; Wu, W.; Tang, X.-M.; Sun, F.; Lu, H.-M.; Deng, J.; et al. Liquid biopsy in lung cancer: Significance in diagnostics, prediction, and treatment monitoring. Mol. Cancer 2022, 21, 25. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ren, F.; Fei, Q.; Qiu, K.; Zhang, Y.; Zhang, H.; Sun, L. Liquid biopsy techniques and lung cancer: Diagnosis, monitoring and evaluation. J. Exp. Clin. Cancer Res. 2024, 43, 96. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Schwarzenbach, H.; Hoon, D.S.; Pantel, K. Cell-free nucleic acids as biomarkers in cancer patients. Nat. Rev. Cancer 2011, 11, 426–437. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Mitchell, P.S.; Parkin, R.K.; Kroh, E.M.; Fritz, B.R.; Wyman, S.K.; Pogosova-Agadjanyan, E.L.; Peterson, A.; Noteboom, J.; O’Briant, K.C.; Allen, A.; et al. Circulating microRNAs as stable blood-based markers for cancer detection. Proc. Natl. Acad. Sci. USA 2008, 105, 10513–10518. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Stenvold, H.; Donnem, T.; Andersen, S.; Al-Saad, S.; Busund, L.T.; Bremnes, R.M. Stage and tissue-specific prognostic impact of miR-182 in NSCLC. BMC Cancer 2014, 14, 138. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Abdipourbozorgbaghi, M.; Vancura, A.; Radpour, R.; Haefliger, S. Circulating miRNA panels as a novel non-invasive diagnostic, prognostic, and potential predictive biomarkers in non-small cell lung cancer (NSCLC). Br. J. Cancer 2024, 131, 1350–1362. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Zhu, W.; Zhou, K.; Zha, Y.; Chen, D.; He, J.; Ma, H.; Liu, X.; Le, H.; Zhang, Y. Diagnostic value of serum miR-182, miR-183, miR-210, and miR-126 levels in patients with early-stage non-small cell lung cancer. PLoS ONE 2016, 11, e0153046. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Zedan, Y.; Konoshenko, M.Y.; Bryzgunova, O.E.; Ilyushchenko, A.A.; Danilova, Y.M.; Gorbunkov, S.D.; Zykov, K.A.; Laktionov, P.P. Blood-based miRNA panels for timely detection of non-small-cell lung cancer: From biomarker discovery to clinical translation. Int. J. Mol. Sci. 2025, 26, 12035. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Nakai, M.; Watanabe, Y.; Koda, M.; Sakamoto, C.; Hasegawa, T.; Chuang, H.S.; Sakamoto, H. Highly sensitive hybridization chain reaction-based miRNA detection technology using diffusivity analysis of fluorescent probe-modified miRNA particles. Sensors 2026, 26, 713. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Zhao, Y.; Chen, F.; Li, Q.; Wang, L.; Fan, C. Isothermal amplification of nucleic acids. Chem. Rev. 2015, 115, 12491–12545. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Bi, S.; Yue, S.; Zhang, S. Hybridization chain reaction: A versatile molecular tool for biosensing, bioimaging, and biomedicine. Chem. Soc. Rev. 2017, 46, 4281–4298. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Dirks, R.M.; Pierce, N.A. Triggered amplification by hybridization chain reaction. Proc. Natl. Acad. Sci. USA 2004, 101, 15275–15278. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Torrente-Rodríguez, R.M.; Campuzano, S.; Montiel, V.R.-V.; Montoya, J.; Pingarrón, J. Sensitive electrochemical determination of miRNAs based on a sandwich assay onto magnetic microcarriers and hybridization chain reaction amplification. Biosens. Bioelectron. 2016, 86, 516–521. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Obika, S.; Nanbu, D.; Hari, Y.; Morio K-i In, Y.; Ishida, T.; Imanishi, T. Synthesis of 2′-O,4′-C-methyleneuridine and -cytidine. Novel bicyclic nucleosides having a fixed C3,-endo sugar puckering. Tetrahedron Lett. 1997, 38, 8735–8738. [Google Scholar] [CrossRef] [Scilit]
- Várallyay, E.; Burgyán, J.; Havelda, Z. MicroRNA detection by northern blotting using locked nucleic acid probes. Nat. Protoc. 2008, 3, 190–196. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Válóczi, A.; Hornyik, C.; Varga, N.; Burgyan, J.; Kauppinen, S.; Havelda, Z. Sensitive and specific detection of microRNAs by northern blot analysis using LNA-modified oligonucleotide probes. Nucleic Acids Res. 2004, 32, e175. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Neely, L.A.; Patel, S.; Garver, J.; Gallo, M.; Hackett, M.; McLaughlin, S.; Nadel, M.; Harris, J.; Gullans, S.; Rooke, J. A single-molecule method for the quantitation of microRNA gene expression. Nat. Methods 2006, 3, 41–46. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Liu, X.; Zhang, Y.; Narisu, N.; Wu, F.; Zhang, K.; Sun, Y.; Gao, H. A robust ddPCR assay for the absolute quantification of miR-192-5p in hepatocellular carcinoma liquid biopsies. Clin. Chem. Lab. Med. 2026, 64, 717–731. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Wang, K.H.; Chen, Y.-Y.; Wang, C.-H.; Hsu, K.-F.; Chau, L.-K.; Wang, S.-C.; Chen, Y.-L. Ultrasensitive amplification-free detection of circulating miRNA via droplet-based processing of SERS tag–miRNA–magnetic nanoparticle sandwich nanocomplexes on a paper-based electrowetting-on-dielectric platform. Analyst 2024, 149, 1981–1987. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Krepelkova, I.; Mlocka, M.; Nytra, M.; Kucera, R.; Topolcan, O.; Fuchsova, R.; Hrncirova, P.; Kolar, V.; Pesta, M.; Vlachova, V.; et al. Evaluation of miRNA detection methods for the analytical characteristic necessary for clinical utilization. Biotechniques 2019, 66, 277–284. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Yi, M.; Liao, Z.; Deng, L.; Xu, L.; Tan, Y.; Liu, K.; Chen, Z.; Zhang, Y. High diagnostic value of miRNAs for NSCLC: Quantitative analysis for both single and combined miRNAs in lung cancer. Ann. Med. 2021, 53, 2178–2193. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Moretti, F.; D’Antona, P.; Finardi, E.; Barbetta, M.; Dominioni, L.; Poli, A.; Gini, E.; Noonan, D.M.; Imperatori, A.; Rotolo, N.; et al. Systematic review and critique of circulating miRNAs as biomarkers of stage I-II non-small cell lung cancer. Oncotarget 2017, 8, 94980. [Google Scholar] [CrossRef] [Scilit] [PubMed]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.